TP-N01013 Determination of Gardner Color Number

November 12, 2018 | Author: m.cj1 | Category: Glasses, Color, Patent, Information, Nature
Share Embed Donate


Short Description

Download TP-N01013 Determination of Gardner Color Number...

Description

LUBRIZOL TEST PROCEDURE TP-N01013 Edition : November 1, 2011 2011 Previous Edition : October 7, 1996 1996

Determin Determin ati on of Gardne rdn er Color Numbe umb er Scope

Procedure

This method covers the measurement of color of varnishes, oils, resins, nitrocellulose lacquers, fatty acids, polymerized fatty acids, and similar transparent colored liquids by comparison with arbitrarily numbered glass standards. standards. A sample's sample's color can can indicate batch contamination contamination and and reproducibility. This method should conform to all of the requirements outlined in the ASTM method D1544.

1. Fill sample tube with sample to be tested and place in the test compartment of comparator. For solid samples, first fill a sample tube with pieces of the sample and melt in a hot water bath. Do not not heat to a temperature more than 5.6°C (10°F) above its cloud point.

Theory

2. Rotate the color wheel disks in the comparator.

 A specified sample tube is filled with the solution to be measured and placed in a comparator or  suitable photometer. Color comparisons using a comparator are made by simply revolving a disc with mounted glass color standards until a color match is obtained between the sample being tested and one of the glass standards. The glass color standards are based on a color scale equivalent to the 1933 Gardner color standard. The color scale conforms to ASTM specification D-1544. The measurement can also be made using a specially calibrated photometer which determines Gardner color by comparing transmitted light to international color standards.

Equipment 1. Gardner color comparator: Gardner Model 212 or equivalent. or

LICO 50 Photometer, and sample cuvettes suitable for the LICO 50, available from Dr. Bruno Lange GmbH Berlin, Willstatterstr. 11, Postfach 19 02 29, 4000 Dusseldorf 11, Telephone (0211)52 88-0, Telefax (0211) 5288-106.

For the Color Comparator Method:

3. Determine which glass standard most closely matches the sample in terms of brightness and saturation. Ignore hue differences. For the LICO 50 Photometer Method   (see the instrument manual for detailed descriptions of the principles of operation and instructions): 1. Fill a round cuvette with approximately 3 mL of the sample to be analyzed. 2. Press the "Gardner" program key to enter into Gardner color determination mode. 3. Insert the cuvette into the LICO 50 sample compartment. 4. The Gardner color value will be directly measured and displayed by the instrument. Read and record this value.

Data Reporting Report the color number of the matching standard, or the directly read Gardner value, as N1013. Enter Pass or Fail as N1013B.

2. Glass sample tube. 3. Heating bath. Lub rizol Advanc ed Materials , Inc. / 9911 9911 Brecks vil le Road, Cleveland , Ohio 44141-3247 44141-3247 / TEL: TEL: 800.379.5389 800.379.5389 or 216.447.50 216.447.5000 00 The information contained herein is being furnished for informational purposes only, upon the express condition that the User makes its own assessment of the appropriate use of such information. While the information contained herein is believed to be reliable, no representations, guarantees or warranties of any kind are made as to its accuracy, suitability for a particular application or the results to be obtained herefrom. Lubrizol Advanced Materials, Inc. ("Lubrizol") cannot guarantee how any products associated with this information will perf orm in

combination with other substances or in the User’s process. Due to variations in methods, conditions and equipment used commercially in processing these materials, no warranties or guarantees are made as to the suitability of the information or products for the applications disclosed. Lubrizol shall not be liable and the User assumes all risk and responsibility for any use or handling of any material beyond Lubrizol's direct control. LUBRIZOL MAKES NO WARRANTIES, EXPRESS OR IMPLIED, INCLUDING, BUT NOT LIMITED TO,

THE IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. It is the User’s sole responsibility to determine if there are any issues relating to patent infringement of any component or combination of components relating to the supplied information. Nothing contained herein is to be considered as permission, recommendation, nor as an inducement to practice any patented invention without permissio n of the patent owner.

For further information, please visit: www.lubrizol.com/personalcare Lubrizol Advanced Materials, Inc. is a wholly owned subsidiary of The Lubrizol Corporation  All trademarks owned by The Lubrizol Lubrizol Corporation © Copyright 2013 / The Lubrizol Corporation

Lubr izol Test Procedure TP-NO1013 Determination of Gardner Color Number Page 2 of 2

Perfo rmance Data 1.

Range: For the Color Comparator; 0 - 18 Gardner color numbers. For the LICO 50 photometer; 0.1 - 8 Gardner color numbers.

2.

Precision: The standard deviation reported in the  ASTM method D1544-80 is 0.5 color number.

3.

Operator time: 5 minutes.

4.

Elapsed time: 5 minutes.

5.

Date of last revision: 01/29/96.

6.

Plants currently using this procedure: 1, 6, 9, 10, 12, Germany.

References 1.

Annual Book of ASTM Standards, Volume 29, Method D-1544-80, pp. 201-203, 1982.

2.

Operating Manual, LICO 50 Photometer.

NOTE: The reference listed below was NOT reviewed for this revision but was included in the 12/14/78 revision of this method. 3.

Gardner, H. A., Stewart, J. R., American Paint and Manufacturers Associated Circular, No. 444, pp. 274-400, 1933.

View more...

Comments

Copyright ©2017 KUPDF Inc.
SUPPORT KUPDF