UNIT AND DIRECT SUPPORT MAINTENANCE MANUAL (INCLUDING REPAIR PARTS AND SPECIAL TOOLS LIST) FOR
IMPROVED CHEMICAL AGENT MONITOR (ICAM) (NSN 6665-01-357-8502) (EIC:5AB) WARNING - This document contains export-controlled technical data whose export is restricted by the Army Export Control Act (Title 22, U.S.C., Sec 2751 et seq) or executive order 12470. Violations of these export laws are subject to sever criminal penalties. DISTRIBUTION STATEMENT C: Distribution authorized to U.S. Government agencies and their contractors to protect technical or operational information. This determination was made on 12 November, 1991. Other requests for this document will be referred to: Tech Director, U.S. Army Edgewood Research, Development and Engineering Center, ATTN: SCBRD-ENL-V, Aberdeen Proving Ground, MD 21010-5423 DESTRUCTION NOTICE - Destroy by any method that will prevent disclosure of contents or reconstruction of the document.
HEADQUARTERS, DEPARTMENT OF THE ARMY 9 JUNE 1998
TM 3-6665-343-23&P
WARNING
RADIATION HAZARD
NICKEL-63 (Ni-63) Wipe test must be performed before and after any maintenance or testing of CAM. Contamination can become airborne by heated air and smoke from fire. Stand upwind of fire to avoid inhalation of possible contamination. The Chemical Agent Monitor (CAM) contains a beta radiation source in the quantity which requires a Nuclear Regulatory Commission (NRC) license to possess items which contain the source. The CAM contains 10 millicuries of Nickel-63. The Nickel-63 plated cylinder is the source which ionizes chemical agents allowing for their detection. If the CAM is lost or stolen, notify the NBC Officer/NCO and the local Radiation Protection Officer (RPO) immediately. Lost or stolen CAMs are reportable to the Nuclear Regulatory Commission (NRC) by the license holder. Report must be made through the local RPO to the Armament and Chemical Acquisition and Logistics Activity (ACALA) Safety Office, DSN 793-2965/2995/6228. The source is totally enclosed and protected by the CAM case. However the CAM is potentially dangerous if broken. SAFETY PRECAUTIONS 1.
Handle the CAM carefully.
2.
If CAM case is broken or cracked, inform the NBC Officer/NCO, and if available, the local Radiation Protection Officer (RPO).
3.
A CAM damaged beyond the repair capability of direct support should be wrapped in a plastic bag and shipped in original packing container, if available, to depot maintenance for repair. If skin contact is made with any area thought to be contaminated with Nickel-63, wash immediately with nonabrasive soap and water.
4.
Follow safety procedures for storage, shipment, and disposal in accordance with this manual, local regulations, AR 710-3 and AR 385-11.
a
TM 3-6665-343-23&P WARNING LITHIUM SULFUR DIOXIDE BATTERIES: Do not attempt to decontaminate CAM batteries. Dispose of batteries according to TB 43-0130 and local Standing Operating Procedures (SOP). FLAMMABLE AND CORROSIVE HAZARD Lithium sulfur dioxide batteries contain lithium, sulfur dioxide and an electrolyte. Sulfur dioxide is an irritant gas. The electrolyte is flammable and highly corrosive. Do not immerse in water or decon solution, crush, or burn batteries. Do not attempt to recharge batteries or store at temperatures above 158°F (70°C). If a battery is mishandled or misused, the lithium may rapidly vent out, carrying with it the sulfur dioxide gas and the electrolyte, and it may heat up. If this happens, stay away until the smell of sulfur is gone. If you have to move the battery, move it outside by using a shovel or long tongs. Wear suitable protection when handling suspect batteries. A rapidly venting battery may create an explosive and corrosive hazard which may cause damage to skin and eyes. If skin or eyes come in contact with the electrolyte, wash thoroughly with generous amounts of water and seek medical attention. Dispose of batteries according to TB 43-0130 and local SOP. WARNING
DANGER-HIGH VOLTAGE-1000 VOLTS DC 1000 volts DC is present on the Printed Circuit Board (PCB) when monitor case is removed. Danger of electrical shock exists if PCB is touched while CAM is operating. CAUTION Do not decontaminate CAM or its accessories with M258A1 or M280 Decontamination Kits. These kits may cause false positives and temporarily render the CAM inoperative. FIRST AID For first aid information, refer to FM 1-11.
b
TM 3-6665-343-23&P TECHNICAL MANUAL No. 3-6665-343-23&P
HEADQUARTERS Department of the Army Washington, DC, 9 June 1998 UNIT AND DIRECT SUPPORT MAINTENANCE MANUAL (INCLUDING REPAIR PARTS AND SPECIAL TOOLS LIST) FOR IMPROVED CHEMICAL AGENT MONITOR (ICAM) NSN 6665-01-357-8502 (EIC:5AB) REPORTING ERRORS AND RECOMMENDING IMPROVEMENTS
You can help improve this manual. If you find any mistakes or know of a way to improve the procedures, please let us know. Mail your letter, DA Form 2028 (Recommended Changes to Publications and Blank Forms), or DA Form 2028-2 located in the back of this manual directly to Technical Director, U.S. Army Edgewood Research, Development and Engineering Center, ATTN: SCBRD-ENL-V, Aberdeen Proving Ground, MD 21010-5423. A reply will be furnished to you. WARNING - This document contains export-controlled technical data whose export is restricted by the Army Export Control Act (Title 22, U.S.C., Sec 2751 et seq) or Executive Order 12470. Violations of these export laws are subject to severe criminal penalties. DISTRIBUTION STATEMENT C: Distribution authorized to U.S. Government agencies and their contractors to protect technical or operational information. This determination was made on 12 November 1991. Other requests for this document will be referred to Technical Director, U.S. Army Edgewood Research, Development and Engineering Center, ATTN: SCBRD-ENL-V, Aberdeen Proving Ground, MD 21010-5423 DESTRUCTION NOTICE - Destroy by any method that will prevent disclosure of contents or reconstruction of the document. TABLE OF CONTENTS Page
HOW TO USE THIS MANUAL---------------------------------------------------------
iii
CHAPTER 1
INTRODUCTION--------------------------------------------------------------------------
1-1
Section I
General Information ---------------------------------------------------------------------
1-1
Section II
Equipment Description and Data -----------------------------------------------------
1-4
Section III
Repair Parts, Special Tools; Test, Measurement --------------------------------and Diagnostic Equipment (TMDE); and Support Equipment
1-7
Section IV
Principles of Operation-------------------------------------------------------------------
1-8
i
TM 3-6665-343-23&P TABLE OF CONTENTS (CONT) Page CHAPTER 2
UNIT MAINTENANCE INSTRUCTIONS ------------------------------------------------------------2-1
Section I
Service Upon Receipt ------------------------------------------------------------------------------------2-1
Section II
Preventive Maintenance Checks and Services (PMCS)-----------------------------------------2-3
Section III
Unit Troubleshooting--------------------------------------------------------------------------------------2-4
Section IV
Unit Maintenance Procedures ------------------------------------------------------------------------ 2-10
Section V
Preparation for Storage or Shipment --------------------------------------------------------------- 2-10
CHAPTER 3
DIRECT SUPPORT MAINTENANCE INSTRUCTIONS-----------------------------------------3-1
Section I
Radiological Safety Instructions------------------------------------------------------------------------3-1
Section II
Service Upon Receipt ------------------------------------------------------------------------------------3-4
Section III
Direct Support Troubleshooting---------------------------------------------------------------------- 3-20
Section IV
Direct Support Maintenance Procedures ---------------------------------------------------------- 3-78
Section V
Preparation for Storage or Shipment ------------------------------------------------------------- 3-142
APPENDIX A
REFERENCES-------------------------------------------------------------------------------------------- A-1
APPENDIX B
MAINTENANCE ALLOCATION CHART ---------------------------------------------------------- B-1
APPENDIX C
REPAIR PARTS AND SPECIAL TOOLS LIST -------------------------------------------------- C-1 Illus/ Figure Introduction ------------------------------------------------------------------------------------------------ C-1
Section I
Repair Parts List --------------------------------------------------------------------------------------- C-1-1
C-1
Section II
Monitor, Chemical Agent (ICAM) ------------------------------------------------------------------ C-1-1
C-2
Group 00
Monitor Kit, CAM -------------------------------------------------------------------------------------- C-2-1
C-3
Group 01
0101 Monitor, Chemical Agent --------------------------------------------------------------------- C-3-1
C-4
010101 Module Assembly, Monitor --------------------------------------------------------------- C-4-1
C-5
01010101 Monitor Assembly ----------------------------------------------------------------------- C-5-1
C-6
0101010101 Module, Drift Tube ------------------------------------------------------------------- C-6-1
C-7
010102 Case Assembly, Monitor ------------------------------------------------------------------ C-7-1
C-8
0102 Carrying Case Assembly--------------------------------------------------------------------- C-8-1 Section III
Special Tools List (Not Applicable)
Section IV
National Stock Number, Part Number, and Reference Designator Indexes ------------- C-I-1
APPENDIX D
EXPENDABLE AND DURABLE ITEMS LIST---------------------------------------------------- D-1
Section I
Introduction ------------------------------------------------------------------------------------------------ D-1
Section II
Expendable and Durable Items List------------------------------------------------------------------ D-2
APPENDIX E
NUCLEAR REGULATORY COMMISSION REQUIREMENTS ------------------------------ E-1 INDEX ------------------------------------------------------------------------------------------------- Index-1
ii
TM 3-6665-343-23&P HOW TO USE THIS MANUAL This manual contains a number of features that enable the user to locate information easily. The table of contents contains a general list of information sections. A comprehensive cross-referenced index is located at the end of the manual. It will also help the user locate specific information. Chapter 1 contains introductory information. Topics of special importance include equipment description and data, location and description of major components, safety, and principles of operation. Chapter 2 contains information at the unit maintenance level. Topics of special importance include service upon receipt, preventive maintenance checks and services, troubleshooting, maintenance procedures and preparation for storage or shipment. Chapter 3 contains information at the direct support maintenance level. Topics of special importance include radiological safety instructions, service upon receipt, troubleshooting, maintenance, and preparation for storage or shipment. When a CAM arrives at Direct Support for repair, use the information in Chapter 3 to test, troubleshoot, repair and retest the CAM. Use the following steps as a guide: 1. Get the equipment required for testing and troubleshooting a CAM. This equipment is listed in para 3-12. Then prepare the Diagnostic Test Set (for use in testing and troubleshooting the CAM) as instructed in para 3-14. 2. Perform the operational check of the CAM presented in para 3-15. If you come to a part of the operational check that cannot be accomplished, refer to the symptom for the problem. (Para 3-11 lists possible symptoms that could occur during the operational check.) Turn to the page in the manual for troubleshooting the symptom. 3. Follow the troubleshooting logic tree for the symptom. The tree will refer you to a maintenance procedure or to further testing. Further testing consists of more detailed tests: cell test, leak test, pump test, and waveform test. 4. Turn to the maintenance procedure and perform the indicated maintenance action. The procedure begins with an initial setup: a listing of the tools and materials needed to complete the task, and the equipment condition prior to performance of the maintenance task. The equipment conditions are a list of the maintenance tasks that must be accomplished before the present task is started. Do the equipment condition tasks first. 5. The specific maintenance task is then performed. The task usually consists of removal and installation procedures. Once the specific task is completed, return to the beginning of the task at the equipment conditions listing. Reverse the order of equipment condition steps to place the CAM back into a fully assembled condition. 6. Return again to the troubleshooting tree for further direction in trouble analysis or additional testing. Tasks in the troubleshooting tree end by performing the operational check to see that the CAM has been completely returned to operational status. 7. Perform wipe test before returning to user (para 3-7). A Repair Parts and Special Tools List (RPSTL) that lists and authorizes spares and repair parts, tools, special tools, Test, Measurement and Diagnostic Equipment (TMDE), and support equipment required for performance of maintenance of the CAM is supplied in Appendix C. Refer to page C-1 for detailed information on using the RPSTL.
iii
TM 3-6665-343-23&P
1. Carrying Case Assembly 2. Carrying Harness Assembly 3. Battery Cap Assembly
4. 5. 6. 7.
Environmental Cap Buzzer Chemical Agent Monitor Nozzle Protective Cap
Improved Chemical Agent Monitor (CAM)
1-0
TM 3-6665-343-23&P CHAPTER 1 INTRODUCTION Section I. GENERAL INFORMATION 1-1.
SCOPE. NOTE The monitor, page 1-0 item 6, will show an NSN on the case; however, the monitor itself is not stocked, stored, or issued. To obtain a new monitor the next higher assembly must be requisitioned. (See system NSN on the front of this manual.) a. Type of Manual. Unit and Direct Support Maintenance including Repair Parts and Special Tools List. b. Model Number and Equipment Name. No model number; Improved Chemical Agent Monitor (ICAM).
c. Purpose of Equipment. Used by ground forces to search out clean areas, to search and locate contamination on personnel, equipment, ships' structures, aircraft and land vehicles, buildings and terrain, and to monitor the boundaries of collective protection. The CAM responds to nerve and mustard agent vapors down to the lowest concentrations that could affect personnel over a short period. d. Special Limitations on Equipment. The CAM is a monitor. It is not a detector; therefore, it can become contaminated and overloaded. The CAM only reports conditions at the front of the inlet nozzle. It is only a point monitor and cannot give a realistic assessment of the vapor hazard over an area from one position. Maintenance must be performed in a clean work area and rubber gloves worn to avoid contaminating CAM. 1-2.
MAINTENANCE FORMS, RECORDS, AND REPORTS.
a. Maintenance. Department of the Army forms and procedures used for equipment maintenance will be those prescribed by DA PAM 738-750, The Army Maintenance Management System (TAMMS) as contained in the Maintenance Management Update. b. Testing and Tracking. For Department of Army procedures for tracking and wipe testing the cell (a component of the CAM) through its life cycle, refer to AR 710-3, Chapter 3. c. Shipments/Receipts. When transporting a CAM, refer to AR 385-11. Also see para 3-62 and 3-63. d. Report of Packaging and Handling Deficiencies. Fill out and forward SF 364 (Report of Discrepancy) as prescribed in AR 735-11-2. e. Discrepancy In Shipment Report. Fill out and forward Discrepancy in Shipment Report (SF 361) as prescribed in AR 55-38. f.
Radiological Accident Reports, RCS DD-SD (AR) 1168. Fill out and forward as prescribed in AR 385-
40. 1-3.
DESTRUCTION OF ARMY MATERIEL TO PREVENT ENEMY USE. Destroy the CAM in accordance with TM 43-0002-31, procedures for destruction of alarm systems.
1-1
TM 3-6665-343-23&P 1-4.
PREPARATION FOR STORAGE OR SHIPMENT. Refer to para 3-62 and 3-63 for special instructions concerning storage or shipment.
1-5.
NOMENCLATURE CROSS-REFERENCE LIST. Common Name Battery Cap Assembly Connector Assembly Display Assembly Environmental Cap Long and Short Pillar Nozzle Seal
1-6.
LIST OF ABBREVIATIONS. AC A/D CAM DC DC-DC DMM DTS EHT G H IMS LCD LED NRC PCB RAM Vdc °C °F
1-7.
Official Nomenclature Retainer Battery Connector Receptacle Optoelectronic Display Protective Dust Cap Threaded Standoff Spacer Sleeve
Alternating Current Analog/Digital Chemical Agent Monitor Direct Current Direct Current to Direct Current Digital Multimeter Diagnostic Test Set Extra High Tension Nerve Agent Mustard Agent Ion Mobility Spectrometry Liquid Crystal Display Light Emitting Diode Nuclear Regulatory Commission Printed Circuit Board Random Access Memory Volts Direct Current Degrees Centigrade (Celsius) Degrees Fahrenheit
REPORTING EQUIPMENT IMPROVEMENT RECOMMENDATIONS (EIR).
If your CAM needs improvement, let us know. Send us a SF 368 (Product Quality Deficiency Report). You, the user, are the only one who can tell us what you do not like about your equipment. Let us know why you do not like the design or performance. Mail it to the address specified in DA PAM 738-750. A reply will be furnished to you. 1-8.
SAFETY, CARE, AND HANDLING.
a. Rules and Regulations. The CAM contains a radioactive source which is controlled by the U.S. Nuclear Regulatory Commission (NRC) under Title 10 Code of Federal Regulations. AR 385-11 and 700-64 implement NRC regulations. Army-wide possession and use of the CAM is authorized by an NRC Byproduct Materials License issued to the Armament and Chemical Acquisition and Logistics Activity, Attn: AMSTA-AC-SF, Rock Island, IL 61299-7630. The license is issued on the basis of statements concerning procedures established for the life-cycle control of the CAM. The CAM units are issued to Army activities through the Armament and Chemical Acquisition and Logistics Activity National Inventory Control Point, (AMSTA-AC-CTC-D). The serial numbers of the CAM and drift tube module are tracked by AR 710-3, Chapter 4, Section 2. Please notify your Cell Serialization Surety Officer (CSSO) to process the serial number transactions with the DOD Central Registry.
1-2
TM 3-6665-343-23&P b. Wipe test. The wipe test (para 3-7) must be performed upon receipt and before and after all maintenance and testing is complete. c. Posting Requirements. Federal law requires certain notices, instructions, and standards be made available to all users of items which contain licensed radioactive material. In lieu of posting this information, pages are attached at the back of this technical manual for reference. These pages contain instructions about Title 10 Code of Federal Regulations, NRC Form 3 (Notice to Employees) and Public Law 93-438. d. Repair/Disposition. Unserviceable CAM will be returned to depot. e. Emergency Procedures. NOTE In any emergency event, radioactive contamination must be considered to be present until determined otherwise. (1)
Fire. In a fire emergency, the basic concern is possible airborne contamination carried out of the flames by the heated air and in the smoke. Firefighters should fight the fire from upwind, and wear portable air systems, if available. After the fire has been extinguished, debris must be surveyed for the presence of equipment containing Nickel-63 sources. The RADIAC meter AN/VDR-2 (with the beta window open) is suitable for detecting the location of the Nickel-63 sources. Wipes must be taken and evaluated to determine contamination location.
(2)
Equipment Destruction. In an equipment destruction emergency, parts of the equipment must be retrieved, and surveys for possible contamination must be done.
(3)
Accident Response in General. (a) (b) (c) (d)
1-9.
Remove injured and spectators. Establish exclusion area. Determine extent of radiation with AN/VDR-2 (or equivalent). Notify authorities (NBC NCO or officer, and, if available, local RPO).
CORROSION PREVENTION AND CONTROL (CPC).
a. Corrosion Prevention and Control (CPC) of Army Materiel is a continuing concern. It is important that any corrosion problems with this item be reported so that the problem can be corrected and improvements can be made to prevent the problem in future items.
1-3
TM 3-6665-343-23&P 1-9.
CORROSION PREVENTION AND CONTROL (CPC) (CONT).
b. While corrosion is typically associated with rusting of metals, it can also include deterioration of materials such as rubber and plastic. Unusual cracking, softening, swelling, or breaking of these materials may be a corrosion problem. c. Such problems should be reported using SF 368 (Quality Deficiency Report). Mail it to the address specified in DA PAM 738-750. Use of key words such as corrosion, rust, deterioration, or cracking will assure that the information is identified as a CPC problem. 1-10.
CALIBRATION. The CAM does not require calibration. Section II. EQUIPMENT DESCRIPTION AND DATA
1-11.
EQUIPMENT CHARACTERISTICS, CAPABILITIES, AND FEATURES.
a. Portable, hand-held instrument designed to determine and indicate the hazard from nerve or mustard agent vapor present in the air. b. Held in either hand while dressed in any level of nuclear, biological, and chemical (NBC) and environmental protective clothing. c. Used to search out clean areas, to search for and locate contamination on personnel, equipment, ship's structures, aircraft and land vehicles, buildings and terrain, and to monitor the boundaries of collective protection. d. Responds to nerve and mustard agent vapors down to the lowest concentrations that could affect personnel over a short period. e. Operates on one battery. Battery life varies with frequency of use and with temperature. At 68°F (20°C), a fresh battery will last at least 12 hours. f.
Can be operated on an external power supply.
g. Easy to operate: only two controls. h. Relative vapor hazard level indicated on a Liquid Crystal Display (LCD).
1-4
TM 3-6665-343-23&P 1-12.
LOCATION AND DESCRIPTION OF MAJOR COMPONENTS.
a. Nozzle Protective Cap (1). Protects nozzle assembly and ensures that clean air is sampled when CAM is started. b. Nozzle Assembly (2). Chemical agent vapors are drawn in through this inlet for analysis by the CAM. A heater wire is wound around the tubular section of the molding. Nozzle assembly is secured to CAM with a special bayonet type locking ring. A special tool from the diagnostic test set is used to remove it from the CAM. c. Case-Front End (3). Provides a common attachment point for the nozzle protective cap, nozzle assembly, nozzle holder assembly, and drift tube module. d. Display Assembly (4). Display is a liquid crystal device back-lit by light emitting diodes (LEDs). Display shows operating status and hazard level, and aids in indicating a malfunction. The ON/OFF and G/H mode switches are part of the assembly. The mode switch controls the polarity of the cell to enable the cell to handle either positive nerve agent ions or negative mustard agent ions. Two sealed rubber pads cover the switch push-buttons. e. Battery Contact Assembly (5). Provides spring loaded contact to mate with the battery terminals and interconnect battery power to the CAM. f. Case Assembly (6). Houses display assembly and battery contact assembly. Case assembly fits over the monitor module assembly. g. Battery Cap Assembly (7). Bayonet fitting cap which retains the battery.
1-5
TM 3-6665-343-23&P 1-12.
LOCATION AND DESCRIPTION OF MAJOR COMPONENTS (CONT).
h. Locknut (8). Locknut is removed to gain access to the internal parts of the CAM. It also acts as a retainer for the disc that contains the serial number of the drift tube module and CAM lot number. i. Environmental Cap Assembly (9). Consists of a tether-bayonet fitting cap protecting the connector assembly. Provides storage for nozzle protective cap during operation. j.
Connector Assembly (10). Provides diagnostic electronic connections for use during maintenance.
k. Sieve Pack Assembly (11). A molecular sieve which filters the CAM recirculatory air. I. mechanism.
Pump Assembly (12). Consists of two silicone rubber diaphragm pumps driven by a DC motor and yoke
m. Drift Tube Module (13). Consists of a cell assembly, manifold assembly, and membrane assembly. The cell assembly, which contains the radioactive source, is in the form of a stack of components held together with stainless steel rods and a retaining clip. Two of the rods are drilled with holes to provide airflow paths for the drift and source regions of the cell; the rods also link the return airflow through the molecular sieve assembly and pump assembly. A membrane assembly is located at the forward end of the cell assembly. The membrane assembly is a thin silicone rubber layer that forms a division between the outside air and the controlled environment inside the analysis section of the CAM. The membrane permits chemical agent vapor molecules to permeate the cell assembly. Membrane is heated to about 250°F (120°C). The serial number of the drift tube module is located on a disc near the locknut (8) on the end of the CAM. The serial number is also etched on the manifold block assembly of the drift tube module.
1-6
TM 3-6665-343-23&P n. Printed Circuit Board (PCB) (14). A printed circuit board that contains the microprocessor. o. Monitor Module Assembly (15). Consists of a locknut, sieve pack assembly, pump assembly, drift tube module and printed circuit board. p. Confidence Sample (16). Used to test the CAM for its ability to detect G and H. The end with a round cross-section is marked G and contains nerve agent simulant. The ribbed end is marked H and contains mustard agent simulant. Housed in carrying harness pocket. q. Buzzer (17). Attaches to electrical connector and provides an audible alarm when CAM displays two or more bars. Section III. REPAIR PARTS; TOOLS; SPECIAL TOOLS; TEST MEASUREMENT AND DIAGNOSTIC EQUIPMENT (TMDE); AND SUPPORT EQUIPMENT 1-13.
COMMON TOOLS AND EQUIPMENT.
For authorized common tools and equipment, refer to the Modified Table of Organization and Equipment (MTOE) applicable to your unit. 1-14.
SPECIAL TOOLS, TMDE, AND SUPPORT EQUIPMENT.
Special tools and TMDE are listed in the Maintenance Allocation Chart in Appendix B of this manual. Support equipment is not needed. Additionally authorized items are listed in TM 3-6665-343-10. 1-15.
REPAIR PARTS. Repair parts are listed and illustrated in the Repair Parts and Special Tools List in Appendix C of this manual.
1-7
TM 3-6665-343-23&P Section IV. PRINCIPLES OF OPERATION 1-16.
INTRODUCTION.
The CAM samples air in the immediate vicinity of the nozzle for the presence of G or H chemical agents. Air sample conditions a short distance away from the CAM may be quite different, and a change in wind direction could quickly bring a hazardous level of agent vapor to a previously safe area. An internal motor/pump assembly (sample pump) continually draws air in through the nozzle assembly. The air is then passed over a heated membrane assembly before being exhausted back into the atmosphere via an exhaust vent at the side of the nozzle assembly. Any agent vapor present in the air permeates the membrane assembly and the cell assembly. In the cell assembly, the vapor molecules are electrically charged (ionized) by a weak radioactive source in order that the vapor molecules can be detected. These charged particles are used to produce an electronic signal by means of a process called Ion Mobility Spectrometry (IMS). Agent ions in the cell assembly are swept towards a collector electrode which produces an electrical impulse for each ion received. The current produced by this electrode is analyzed by a microprocessor which drives the display assembly. From the characteristics of this signal, CAM determines the concentration of the nerve or mustard agent present and indicates the vapor hazard by displaying a number of bars on the display assembly. The closed cell assembly system air is circulated by a recirculating pump through molecular sieves which keep the system dry and chemically clean. The G/H mode push button switch reverses the voltages in the cell assembly to suit nerve or mustard agent vapor which have opposite charges when ionized. Acetone vapor is introduced into the closed circulatory system to stabilize the response of the CAM in the nerve agent mode.
1-8
TM 3-6665-343-23&P
Principles of Operation
1-9
TM 3-6665-343-23&P
The CAM has two operating modes, selectable by means of the G/H mode push-button switch: in the G mode, CAM monitors for nerve agents; in the H mode, CAM monitors for mustard agents. The selected mode is indicated on the display assembly by a G or H. An ON/OFF push-button switch applies 6 Vdc battery power to the CAM. Heater controllers and heaters keep the temperature of the nozzle, membrane, and cell constant for agent detection. A DC-DC converter changes 6 Vdc input to positive or negative 1 kv for cell operation. An analog-to-digital converter changes the varying cell analog output to digital for processing in the microprocessor. The microprocessor output is shown on the liquid crystal display. 1-10
TM 3-6665-343-23&P
THREE DOTS The three dots on the far right of the display are used normally only as a diagnostic indication. If at any time during normal operation the processor loses the Reactant Ion Peak (RIP) (i.e., the RIP, always monitored, does not meet the preset height and position criteria), the three dots will be activated. This would normally happen briefly after a mode change or during excessive dosage of agent or simulant (as they have a greater charge affinity than air). During this period, the unit will still operate normally as the processor is using the last known reference for the RIP (from RAM). If, however, the processor does not find a RIP again within about 30 minutes of its disappearance (normally this would be caused only by either gross contamination or equipment failure), a fault condition signal is given to the operator. This failure signal comprises all eight bars and the WAIT symbol flashing at a frequency of approximately 2Hz.
1-11/(1-12 blank)
TM 3-6665-343-23&P CHAPTER 2 UNIT MAINTENANCE INSTRUCTIONS Section I. SERVICE UPON RECEIPT 2-1.
INTRODUCTION.
This section provides the information the unit maintenance technician needs to inspect and service the Chemical Agent Monitor (CAM) before issuing it to the operator. Table 2-1. Service Upon Receipt – CAM
LOCATION/ITEM
CAM
ACTION
a.
Preparation of work area. (1) Clean the work surface. (2) Cover work surface with paper. (3) Secure paper to work surface with tape, ensuring all seams are sealed.
b.
Open packaging carton (1) from top.
c.
Lift the CAM Monitor Kit (2) out of the packaging carton and set it aside until further directed. DO NOT OPEN ANY PACKAGES OR CONTAINERS AT THIS TIME.
d.
Save the empty container for future use.
2-1
TM 3-6665-343-23&P Table 2-1. Service Upon Receipt - CAM (Cont) LOCATION/ITEM CAM (Cont)
ACTION e. Open the CAM Monitor Kit: (1) Squeeze buckle and pull up on strap (1) to open case lid (2); verify that there are two filtered nozzle package assemblies (3) in the case lid (2). DO NOT OPEN FILTERED NOZZLE PACKAGE ASSEMBLIES UNTIL REQUIRED FOR USE. (2) Lift carrying harness (4); verify that two filtered nozzle packages (5) are located in the carrying harness. DO NOT OPEN FILTERED NOZZLE PACKAGE ASSEMBLIES UNTIL REQUIRED FOR USE. (3) Verify the following items are located in the interior pockets of the carrying case (6): nozzle protective cap (7), packaged carrying harness assembly (8), confidence sample (9), buzzer (10), and strap (11). Remove strap (11) from pocket and attach to rings at ends of carrying case (6). (4) Remove small handle strap (12) from carrying case (6). (5) Remove large handle strap (13) from carrying case (6). (6) Lift CAM (14) from carrying case (6). (7) Open monitor polyethylene bag (15) and remove CAM (14). (8) Remove and discard two desiccant bags (16) from CAM (14). f.
Inspect the equipment for damage incurred during shipment. If the equipment has been damaged, report the damage on SF 364, Report of Discrepancy.
g. Check the equipment against the packing slip to see if the shipment is complete. Report all discrepancies in accordance with the instructions of DA PAM 738-750, The Army Maintenance Management System (TAMMS) Update. h. Turn in to Direct Support to perform wipe and waveform tests. i.
2-2
Install battery (para 3-8 steps d, f, and g only). Perform self-test (para 3-9) and confidence test (para 3-10).
TM 3-6665-343-23&P Section II. PREVENTIVE MAINTENANCE CHECKS AND SERVICES (PMCS) 2-2.
UNIT PROCEDURES.
Your Preventive Maintenance Checks and Services (PMCS) table lists the procedures you are required to perform. The interval column tells you when to do the check or service. The procedures column tells how to do the required PMCS. Carefully follow these instructions. PREVENTIVE MAINTENANCE CHECKS AND SERVICES FOR CAM Item No.
Interval
Item to Check/Service
Procedure
1
Weekly
CAM
Perform a self-test and a confidence test. Run the CAM until it passes the confidence test, but at least 30 minutes minimum. Run the CAM at least 5 additional minutes after the display bars clear. NOTE This procedure is also operator PMCS. See para 2-2 of TM 3-6665343-10.
2
Annually*
*
CAM
Process transactions IAW AR 710-3 and return CAM with transactions to Direct Support for wipe test.
Wipe test to be performed annually or at change in ownership if it has not been performed in the past six months. NOTE Before you send the CAM for DS maintenance, monitor it for radiation leaks using a RADIAC set, if available. If a leak is detected, place the CAM in a plastic bag before sending it to DS If a RADIAC is not available, or if CAM is damaged, place the CAM in a plastic bag before sending it to DS.
2-3
TM 3-6665-343-23&P Section III. UNIT TROUBLESHOOTING 2-3.
INTRODUCTION.
Use the following index to locate trouble symptoms. You should perform the test/inspections and corrective actions in the order listed, until the malfunction is corrected. If malfunction is not corrected by listed corrective actions, then forward the CAM to Direct Support (para 2-6). SYMPTOM INDEX
Symptom
Troubleshooting Procedure Page
1.
BACKLIGHT NOT OPERATING ..........................................................................................................
2-5
2.
BL SHOWN ON DISPLAY DURING OPERATION ...............................................................................
2-5
3.
DISPLAY DOES NOT COME ON OR DISPLAY DISAPPEARS ............................................................
2-5
4.
DISPLAY FLASHES ON AND OFF.......................................................................................................
2-5
5.
DOES NOT CHANGE MODE FROM G TO H OR H TO G ...................................................................
2-6
6.
DOES NOT CLEAR DOWN TO ZERO OR ONE BAR IN 2 MINUTES (OR 5 MINUTES FOR FIVE OR MORE BARS) ...................................................................................
2-6
7.
DOES NOT CLEAR DOWN TO ZERO OR ONE BAR BEFORE SHUTDOWN ......................................
2-7
8.
DOES NOT RESPOND TO CONFIDENCE SAMPLE (SIMULANT) ......................................................
2-7
9.
ONE OR MORE DISPLAY INDICATORS DO NOT SHOW DURING SELF-TEST.................................
2-8
10. REMAINS IN SELF-TEST FOR MORE THAN 30 SECONDS ...............................................................
2-8
11. REMAINS IN WAIT FOR LONGER THAN 2 MINUTES ........................................................................
2-9
12. BUZZER DOES NOT SOUND WHEN DISPLAY SHOWS TWO BARS OR MORE................................
2-9
2-4
TM 3-6665-343-23&P Table 2-3. Unit Troubleshooting MALFUNCTION TEST OR INSPECTION CORRECTIVE ACTION 1. BACKLIGHT NOT OPERATING. Step 1.
No test or inspection required; proceed to corrective action. Turn CAM off. Forward CAM to Direct Support (para 2-6).
2. BL SHOWN ON DISPLAY DURING OPERATION. Step 1.
No test or inspection required; proceed to corrective action. Turn CAM off. Replace battery. Forward CAM to Direct Support (para 2-6).
3. DISPLAY DOES NOT COME ON OR DISPLAY DISAPPEARS. Step 1.
No test or inspection required; proceed to corrective action. Turn CAM off. Remove battery cap assembly.
Step 2.
Check that battery is present. Install battery with terminal end toward display. Press ON/OFF push-button switch. Turn CAM off. Replace battery. Forward CAM to Direct Support (para 2-6).
4. DISPLAY FLASHES ON AND OFF. Step 1.
No test or inspection required; proceed to corrective action. Make sure filtered nozzle standoff is removed and protective cap assembly is attached on front of CAM case. Turn CAM off. Wait 2 seconds.
Step 2.
Perform self-test (para 3-9). Look for BL indication after the self-test is complete. Replace battery if necessary.
2-5
TM 3-6665-343-23&P Table 2-3. Unit Troubleshooting (Cont) MALFUNCTION TEST OR INSPECTION CORRECTIVE ACTION 4. DISPLAY FLASHES ON AND OFF (CONT). Let CAM run for 20 minutes. Turn CAM off. Wait 2 seconds. Step 3.
Perform self-test (para 3-9) and confidence test (para 3-10). Forward CAM to Direct Support (para 2-6).
5. DOES NOT CHANGE MODE FROM G TO H OR H TO G. Step 1.
No test or inspection required; proceed to corrective action. Press G/H mode push-button switch.
Step 2.
Check that display does not indicate change in mode. Forward CAM to Direct Support (para 2-6).
6. DOES NOT CLEAR DOWN TO ZERO OR ONE BAR IN 2 MINUTES (OR 5 MINUTES FOR FIVE OR MORE BARS). Step 1.
No test or inspection required; proceed to corrective action. Remove the filtered nozzle standoff. Install nozzle protective cap onto CAM front end (para 3-10, steps h through j). Let CAM run for 15 minutes. Clean vertical face of case front-end with a cloth dampened with water.
Step 2.
Repeat self-test (para 3-9) and confidence test (para 3-10). Forward CAM to Direct Support (para 2-6).
2-6
TM 3-6665-343-23&P Table 2-3. Unit Troubleshooting (Cont) MALFUNCTION TEST OR INSPECTION CORRECTIVE ACTION 7. DOES NOT CLEAR DOWN TO ZERO OR ONE BAR BEFORE SHUTDOWN. Step 1.
No test or inspection required; proceed to corrective action. Make sure CAM runs with protective cap on nozzle for at least 1 hour. Remove protective cap from CAM nozzle and twist (clockwise) onto CAM environmental cap. Turn CAM off. Decontaminate the CAM nozzle assembly with a cloth dampened with water. Install protective cap onto nozzle. Turn CAM on. The number of bars displayed should decrease with the cap on. If number of bars increases, the nozzle protective cap is contaminated. If CAM does not clear within 15 minutes, replace nozzle protective cap. Turn CAM on and let run for an additional 15 minutes. Forward CAM to Direct Support (para 2-6).
8. DOES NOT RESPOND TO CONFIDENCE SAMPLE (SIMULANT). Step 1.
No test or inspection required; proceed to corrective action. Make sure CAM runs for at least 15 minutes. CAM may require an extra warm-up period if it has been in storage. Wait 10 seconds and repeat the 1 second exposure. This can be done up to five times. Run CAM 5 minutes.
Step 2.
Repeat confidence test (para 3-10) using another confidence sample, if available. Remove filtered nozzle standoff. Install nozzle protective cap onto CAM front end (para 3-10, steps h through j). Turn CAM off. Wait 2 seconds.
Step 3.
Perform self-test (para 3-9). Let CAM operate for 1 hour. (Disregard display during the hour.) Turn CAM off. Wait 2 seconds.
2-7
TM 3-6665-343-23&P Table 2-3. Unit Troubleshooting (Cont) MALFUNCTION TEST OR INSPECTION CORRECTIVE ACTION 8. DOES NOT RESPOND TO CONFIDENCE SAMPLE (SIMULANT)(CONT). Step 4.
Repeat self-test (para 3-9) and confidence test (para 3-10). Remove filtered nozzle standoff; install nozzle protective cap onto CAM front end (para 3-10, steps h through j). Turn CAM on. Let CAM run 30 minutes. Turn CAM off. Wait 2 seconds.
Step 5.
Repeat self-test (para 3-9) and confidence test (para 3-10). Forward CAM to Direct Support (para 2-6).
9. ONE OR MORE DISPLAY INDICATORS DO NOT SHOW DURING SELF-TEST. NOTE Either G or H mode show on the display, never both at the same time. Step 1.
No test or inspection required; proceed to corrective action. Turn CAM off. Wait 2 seconds.
Step 2.
Repeat self-test (para 3-9). Forward CAM to Direct Support (para 2-6).
10. REMAINS IN SELF-TEST FOR MORE THAN 30 SECONDS. Step 1.
No test or inspection required; proceed to corrective action. Turn CAM off. Wait 2 seconds.
Step 2.
Repeat self-test (para 3-9). Forward CAM to Direct Support (para 2-6).
2-8
TM 3-6665-343-23&P Table 2-3. Unit Troubleshooting (Cont) MALFUNCTION TEST OR INSPECTION CORRECTIVE ACTION 11. REMAINS IN WAIT FOR LONGER THAN 2 MINUTES. Step 1.
No test or inspection required; proceed to corrective action. Turn CAM off. Wait 2 seconds.
Step 2.
Repeat self-test (para 3-9). Let CAM run up to 30 minutes. If CAM remains in WAIT again, let CAM run another 30 minutes. Turn CAM off. Wait 2 seconds.
Step 3.
Repeat self-test (para 3-9). If CAM does not clear within 2 minutes, replace nozzle protective cap assembly. Turn CAM on. Let CAM run for up to 30 minutes. Forward CAM to Direct Support (para 2-6).
12. BUZZER DOES NOT SOUND WHEN DISPLAY SHOWS TWO BARS OR MORE. Step 1.
Replace the buzzer battery.
Step 2.
Repeat confidence test (para 3-10). Forward CAM to Direct Support (para 2-6).
2-9
TM 3-6665-343-23&P Section IV. UNIT MAINTENANCE PROCEDURES 2-4.
INTRODUCTION. Unit maintenance is limited to inspection and fault verification. Section V. PREPARATION FOR STORAGE OR SHIPMENT
2-5.
STORAGE. a. Remove battery from CAM. Store battery separately from CAM. Install and turn (clockwise) battery cap on CAM.
b. Store CAM in rooms/areas/sections designated for storage of radioactive materials which are free from the danger of flooding, outside the danger radius of flammables or explosives, and secured against unauthorized removal. c.
Post the area with CAUTION - RADIOACTIVE MATERIAL signs as required.
d. The signs shall be yellow with magenta letters and radiation symbol at least 8 inches by 10 inches. Post signs on outside of lockable containers. e. Storage of a single CAM must be in a secure area; the area does not need to be marked. f. 2-6.
Process transactions as required by AR 710-3. SHIPMENT.
a. Remove battery from CAM. Install and turn (clockwise) battery cap on CAM. b. Place CAM in carrying case assembly and return to Direct Support.
2-10
TM 3-6665-343-23P CHAPTER 3 DIRECT SUPPORT MAINTENANCE INSTRUCTIONS Section I. RADIOLOGICAL SAFETY INSTRUCTIONS WARNING RADIATION HAZARD
NICKEL-63 (Ni-63) Wipe test must be performed before and after any maintenance or testing of Chemical Agent Monitor (CAM). CAUTION To avoid contamination of equipment DO NOT: Smoke, eat, or drink in work area. Use aftershave or perfume. Use cleaning agents, floor wax sprays, adhesives, paint, or similar items. Use insect or skin repellents. Handle internal parts without wearing rubber gloves. NOTE CAM performance may be affected by EMI when attached to the DTS. Avoid performing maintenance tasks near equipment emitting high electromagnetic radiation. 3-1.
WORK AREA.
INITIAL SETUP Tools/Equipment RADIAC Set AN/VDR-2 (or equivalent) Materials/Parts Paper (Item 5, App D) Tape (Item 8, App D) Gloves (Item 4, App D) Plastic Bag (Item 1, App D) Wipe Test Kit (Item 10, App D) Ball Point Pen (Item 6, App D) Soap (Item 7, App D) 3-1
TM 3-6665-343-23P 3-1.
WORK AREA (CONT). a. Preparation of Work Area. (1) Clean the work surface. (2) Cover work surface with paper. (3) Secure paper to work surface with tape, insuring all seams are sealed. b. Work Area Cleanup and Decontamination. NOTE Work area must be wiped and the wipe surveyed with AN/VDR-2, or equivalent, for presence of Nickel-63 contamination at conclusion of work EACH day that maintenance operations are performed. Surveys must be documented. (1) Record wipe test information IAW para 3-7. a. Identify the area (e.g., Bldg 51) on the form. Set aside vial and form. (2) Wear disposable gloves. (3) Fold paper inward to hold in contamination and place in plastic bag. (4) Take a wipe test of the area with a moistened wipe test filter from the wipe test kit (Item 10, App D). Evaluate the wipe test filter with AN/VDR-2, or equivalent. It should read no more than twice background. If the reading is higher, stop cleanup and contact the RPO. (5) Dispose of gloves in plastic bag. Seal bag. (6) Using a ball point pen, mark: "RADIOACTIVE MATERIAL Ni-63" on a piece of tape and apply the tape to the plastic bag. Set aside bag pending lab results of wipe test. (7) Place the wipe test filter in the prepared vial. Send this wipe test for evaluation to the Rock Island Laboratory identified in para 3-7, along with the CAM wipes. (8) Wash hands for 2 minutes with soap at conclusion of maintenance or testing of CAM and at conclusion of any work area cleanup operations. (9) Dispose of bag as regular or radioactive waste, depending upon lab test results.
3-2.
ACCOUNTABILITY.
For instructions on completing necessary transactions, contact your Cell Serialization Surety Officer (CSSO) in accordance with AR 710-3, Chapter 4, Section 2. A transaction is required for each of the following: a. Inventory loss. b. Suspected loss or theft.
3-2
TM 3-6665-343-23P c.
Receipt.
d. Shipment. e. Demilitarization. f.
Wipe test result.
g. Removal of a drift tube module from CAM. h. Insertion of a drift tube module into CAM. 3-3.
FIRE AND EXPLOSION EMERGENCIES. a. Fire Emergencies. WARNING RADIATION HAZARD
NICKEL-63 (Ni-63) Contamination can become airborne by heated air and smoke from fire. Stand upwind of fire to avoid inhalation of possible contamination. (1) Notify all personnel to evacuate area or building and to stand upwind of heat and smoke. (2) Notify fire department of fire and possible hazards, and that self-contained breathing apparatus should be worn. (3) Turn off all ventilation equipment. (4) Close all doors and windows. (5) Notify your NBC NCO or Officer, or RPO. b. Explosion Emergencies. (1) Care for injured (FM 21-11). (2) Notify your NBC NCO or Officer, or RPO. (3) Notify medical team. (4) Cordon off area. 3-3
TM 3-6665-343-23P Section II. SERVICE UPON RECEIPT
3-4.
INTRODUCTION.
This section describes unpacking, checking unpacked equipment, wipe test, assembly and preparation of CAM for use, self-test, and confidence test. 3-5.
UNPACKING. a. Preparation of Work Area. (1) Clean the work surface. (2) Cover work surface with paper. (3) Secure paper to work surface with tape, insuring all seams are sealed. b. Open packaging carton (1) from top.
c. Lift the CAM Monitor Kit (2) out of the packaging carton and set it aside until further directed: DO NOT OPEN ANY PACKAGES OR CONTAINERS AT THIS TIME. d. Save the empty container for future use.
3-4
TM 3-6665-343-23P e. Open the CAM Monitor Kit: (1) Pull up on strap (1) to open case lid (2); verify that there are two filtered nozzle package assemblies (3) in the case lid (2). DO NOT OPEN FILTERED NOZZLE PACKAGE ASSEMBLIES UNTIL REQUIRED FOR USE. (2) Lift carrying harness (4); verify that two filtered nozzle packages (5) are located in the carrying harness. DO NOT OPEN FILTERED NOZZLE PACKAGE ASSEMBLIES UNTIL REQUIRED FOR USE. (3) Verify the following items are located in the interior pockets of the carrying case (6): nozzle protective cap (7), packaged carrying harness assembly (8), confidence sample (9), buzzer (10), and strap (11). Remove strap (11) from pocket and attach to rings at ends of carrying case (6). (4) Remove small handle strap (12) from carrying case (6). (5) Remove large handle strap (13) from carrying case (6). (6) Lift CAM (14) from carrying case (6). (7) Open monitor polyethylene bag (15); remove CAM (14). (8) Remove and discard two desiccant bags (16) from CAM (14).
3-5
TM 3-6665-343-23P 3-6.
CHECKING UNPACKED EQUIPMENT.
a. Inspection. Inspect the equipment for damage incurred during shipment. If the equipment has been damaged, report the damage on SF 364, Report of Discrepancy. b. Check the equipment against the packing slip to see if the shipment is complete. Report all discrepancies in accordance with the instructions of DA PAM 738-750, The Army Maintenance Management System (TAMMS) Update. 3-7.
WIPE TEST. WARNING Do not lick envelopes to seal them or ingestion of Nickel-63 may result. NOTE Perform wipe test annually or at change in ownership if it has not been performed in the past six months. a. Recording Information. (1) If needed, make additional copies of the RIA Wipe Test Request Form that is supplied in the wipe test kit (Item 10, App D). (2) Record the following information on the Wipe Test Request Form: complete mailing address, POC, DSN or commercial phone and fax numbers, date wipes were taken, and reporting activity DODAAC. Use a ball-point pen (Item 6, App D). (3) Using the felt marker provided in the kit, assign a number to the top of the cap of the plastic vial. NOTE Do not mark on the side of the plastic vial or the wipe test may not be readable. (4) Mark the same number in the column marked "VIAL NO." on the Wipe Test Request Form along with the corresponding owning activity DODAAC. Indicate an A for annual wipe or M for maintenance wipe. b. Wipe Test Procedures. NOTE To avoid mixing up samples, perform each wipe test after recording each CAM's serial number on the Wipe Test Request Form. (1) Wear disposable gloves. (2) Twist (counterclockwise) and remove environmental cap (1) from CAM (2). Let environmental cap hang on its tether (3).
3-6
TM 3-6665-343-23P (3) Record the CAM serial number from the side of the CAM (4) and the drift tube module serial number from the disc on the end of the CAM (view A) on the same line of the Wipe Test Request Form that the vial number is recorded on. (4) Moisten a wipe test filter with 2 or 3 drops of water from the kit or clean tap water. (5) Wipe the exterior of the CAM (2) around the nozzle protective cap assembly (5), along the case seal and around the environmental cap(1). (6) Using a RADIAC set, AN/VDR-2 (or equivalent) check the wipe test filter for contamination as follows: NOTE Perform the following procedures in an area that is free from all radiation, except for normal background radiation.
* This is a manufacturer number (it will vary) VIEW A
(a) Adjust AN/VDR-2 (or equivalent) to measure 0 to 0.5 mR/hr. (b) Open the beta shield of the probe of the AN/VDR-2 (or equivalent). (c) Place wipe test filter approximately 1/4 in. in front of the probe and note the indication. DO NOT TOUCH THE PROBE WITH THE WIPE TEST FILTER. WARNING Any sustained reading of the AN/VDR-2 (or equivalent meter) that is twice background may indicate Nickel-63 contamination. If the reading is over twice background, discontinue use of the CAM, place it in a plastic bag and store it in a secure radioactive material storage area until the wipe test is analyzed by the laboratory, to verify that Nickel-63 contamination is present. Notify your RPO, and have supply personnel coordinate shipment with the ACALA item manager, DSN 793-5961/5916. If the sustained reading is less than twice background level, the CAM may be returned to use. The wipe test will be verified by the laboratory stated in step (9). (7) Place wipe test filter into the plastic vial and screw the cap on the vial.
3-7
TM 3-6665-343-23P 3-7.
WIPE TEST (CONT). WARNING Do not lick packaging envelopes to seal them or ingestion of Ni-63 may result.
(8) Place the numbered vial(s) along with the Wipe Test Request Form in the pre-addressed packaging envelope provided or other sturdy packaging. Seal the package. Mark the envelope: MAIL ROOM - DO NOT OPEN. (9) Mail samples to the following address: Commander, Rock Island Arsenal Radiation Test Lab/Radiation Wipe Test Samples ATTN: SIORI-SEL/RIA RPO Rodman Ave. Bldg 210, Room 407 Rock Island, IL 61299-5000 (10) Results of the wipe test evaluated by the laboratory will be sent to the address entered on the Wipe Test Request Form. (11) The CAM can be used immediately after wipe testing unless leakage of Nickel-63 is suspected or an accident has occurred. (12) Proceed with work area cleanup and Decontamination (para 3-1b) or Direct Support troubleshooting (para 3-11) as appropriate.
3-8.
ASSEMBLY AND PREPARATION OF CAM FOR USE. a. Assemble carrying harness to CAM: (1) If not already buckled, remove twists from carrying harness (1) and fasten buckle (2). (2) Place small handle strap (3) around forward part of handle (4); snap the end (5) of the carrying harness that contains filtered nozzle package assembly (6) to the small handle strap (3). (3) Place large handle strap (7) around rear section of handle (4) as shown. Snap end (8) of carrying harness to large handle strap (7).
3-8
TM 3-6665-343-23P
b. Remove vapor proof barrier (1) from confidence sample (2). Place confidence sample (2) in carrying harness pocket (3).
c.
Twist (counterclockwise) and remove battery cap assembly (1) from monitor case assembly (2).
d. Pull desiccant bag (3) from CAM (4) and discard.
3-9
TM 3-6665-343-23P 3-8.
ASSEMBLY AND PREPARATION OF CAM FOR USE (CONT). e. Install battery (1) into CAM. Ensure that terminal end of battery is inserted first. NOTE If the CAM makes a humming sound and/or if the display shows an indication when the battery cap assembly is installed in the next step, press the ON/OFF (2) push-button switch one time. f.
Install and turn (clockwise) battery cap assembly (3) onto CAM.
g. Perform self-test (para 3-9). If CAM fails self-test, refer to troubleshooting (para 3-11). h. Remove battery. Prepare DTS for use (para 3-14). Perform waveform test (para 3-61). Let CAM run until it reaches peak heights (step m) of 2 volts in each mode, up to 8 hours.
3-10
i.
Perform confidence test (para 3-10). If CAM fails confidence test, refer to troubleshooting (para 3-11).
j.
Let CAM run at least 2 hours. Disconnect CAM from DTS.
TM 3-6665-343-23P WARNING FLAMMABLE AND CORROSIVE HAZARD
Lithium-sulfur dioxide batteries contain lithium, sulfur dioxide and an electrolyte. Sulfur dioxide is an irritant gas. The electrolyte is flammable and highly corrosive. Do not immerse in water or decon solution, crush, or burn batteries. Do not attempt to recharge batteries or store at temperatures above 158°F (70°C). If a battery is mishandled or misused, the lithium may rapidly vent out, carrying with it the sulfur dioxide gas and the electrolyte, and it may heat up. If this happens, stay away until the smell of sulfur is gone. If you have to move the battery, move it outside by using a shovel or long tongs. Wear suitable protection when handling suspect batteries. A rapidly venting battery may create an explosive and corrosive hazard which may cause damage to skin and eyes. If the skin or eyes come in contact with the electrolyte, wash thoroughly with generous amounts of water, and seek medical attention. Dispose of batteries according to TB 43-0130 and local SOP.
3-11
TM 3-6665-343-23P 3-8.
ASSEMBLY AND PREPARATION OF CAM FOR USE (CONT). k.
Place CAM (1) into carrying case (2).
I.
Close lid (3) of carrying case (4). Slide hand along hook-and-loop fastener (5) to seal carrying case closed.
m. Clip cover buckle (6) into case buckle (7). n. Forward CAM to unit maintenance for distribution to operators.
3-12
TM 3-6665-343-23P 3-9.
SELF-TEST. a. Ensure that nozzle protective cap assembly (1) is in position on CAM (2).
b. Press ON/OFF push-button switch (3) and observe the display for the following indications (1) through (8); if indications are not shown, refer to troubleshooting section (para 3-11). (1) H mode (4) is shown; if G mode is shown, press G/H mode push-button switch (5). Verify H mode is shown. (2) Markers A and B (6) are shown. (3) All eight bars (7) are shown. (4) Three vertical dots (8) are shown. (5) BL (9) is shown. (6) WAIT (10) is shown.
3-13
TM 3-6665-343-23P 3-9.
SELF-TEST (CONT). (7) Display clears from self-test after 30 seconds (H mode, WAIT, and A and B markers remain).
NOTE Self-test is now complete. (8) Display clears from WAIT mode within 2 minutes. NOTE In very cold conditions, the battery may not immediately reach operating level (BL may not go out). If BL is still displayed after 5 minutes, repeat step b. An extended warm up period may be necessary if a CAM has been in storage 30 days or longer. Allow CAM to run until it will pass confidence test (para 3-10), but not more than 72 hours.
3-14
TM 3-6665-343-23P 3-10.
CONFIDENCE TEST. CAUTION Nozzle protective cap must be stored on environmental cap when not on the front of the CAM. Do not touch the nozzle assembly or the new filtered nozzle standoff; touching could contaminate them. Remove and install nozzle protective cap and nozzle standoffs as quickly as possible to avoid dust or other contamination from entering the CAM protective cap or nozzle.
a. Twist (counterclockwise) and remove nozzle protective cap (1) from front of CAM (2); place nozzle protective cap onto CAM environmental cap (3) and twist clockwise.
3-15
TM 3-6665-343-23P 3-10.
CONFIDENCE TEST (CONT). b. Place a filtered nozzle standoff onto CAM nozzle assembly, as follows: (1) Pull one filtered nozzle package assembly (1) from pocket (2) of carrying harness. (2) Peel back covering (3) from top of filtered nozzle package assembly (1) until one filtered nozzle standoff (4) is exposed. (3) Press CAM nozzle assembly (5) into exposed filtered nozzle standoff (4) and remove. (4) Lay covering (3) back in place across top of filtered nozzle package assembly (1). Slide package assembly (1) back into pocket (2) of carrying harness.
c. Remove environmental cap (6) and nozzle protective cap (7) combination by twisting counterclockwise. Install buzzer (9) onto electrical connector by twisting the ring (8) clockwise.
3-16
TM 3-6665-343-23P NOTE Make sure CAM display indicates H Mode. Your confidence sample may be different in appearance and operation (depending on when it was produced), but will provide the same result. d. Remove confidence sample from carrying harness and perform H confidence test as follows: (1) Grasp confidence sample (1) with H end valve exposed.
CAUTION Do not allow the CAM to sample the confidence sample for more than 1 second. Longer than 1 second will saturate the CAM with vapor. It is only necessary that at least three bars show for test verification (do not attempt to have all bars show). (2) Insert CAM nozzle assembly (2) in H end of confidence sample (1) for 1 second, depressing valve (confidence sample should touch filtered nozzle standoff). (3) Remove confidence sample (1). (4) Verify that at least three (3) bars are displayed and buzzer sounds after a few seconds. Three dots may appear momentarily; ignore them. If at least three bars do not appear, wait 10 seconds and repeat the 1 second exposure. You may do this up to five times, waiting at least 10 seconds between each try. If at least three bars still do not appear, refer to troubleshooting symptom 17 (para 3-32). If the buzzer does not sound, refer to troubleshooting symptom 21 (para 3-36).
3-17
TM 3-6665-343-23P 3-10.
CONFIDENCE TEST (CONT). (5) Wait for display to clear. Display should clear to either zero or one bar within 2 minutes. (If display shows five or more bars, CAM may need approximately 5 minutes to clear.)
e. Press G/H mode push-button switch (1); verify mode changes from H to G (2). WAIT may be displayed for several seconds. When WAIT has disappeared, proceed. The three dots may also appear following the mode change.
f.
Perform G confidence test the same as the H confidence test using the G end of the confidence sample.
g. Place confidence sample (3) into pocket (4) of carrying harness (5).
3-18
TM 3-6665-343-23P
h. Remove and discard filtered nozzle standoff (1) from nozzle assembly (2). i.
Twist (counterclockwise) and remove nozzle protective cap assembly (3) from environmental cap (4).
j.
Twist and install nozzle protective cap assembly (3) to CAM case front-end (5).
k.
Remove buzzer (7) from electrical connector and install environmental cap (4) on electrical connector.
CAUTION Do not switch CAM off when one or more bars are showing because this action stops the purging of any contaminant remaining; this would extend the warm up time on future use. NOTE CAM normally clears down to either zero or one bar within 2 minutes. If CAM is contaminated, it may take at least 1 hour to clear down to zero bars. I. Observe display (6). If display shows zero bars, press G/H push-button switch to change modes of operation. Observe display (6). m. Make sure CAM has cleared down to zero bars in both modes. Press ON/OFF push-button switch to turn CAM off. If CAM fails to clear down, in either mode, within 1 hour, refer to troubleshooting (para 3-11).
3-19
TM 3-6665-343-23P Section III. DIRECT SUPPORT TROUBLESHOOTING 3-11.
INTRODUCTION. WARNING
RADIATION HAZARD
NICKEL-63 (NI-63) Wipe test must be performed before and after any maintenance or testing of CAM. When a CAM is sent to DS for repair, the performance of the CAM is first checked by an operational check. When a fault occurs during the operational check, the maintainer goes to the troubleshooting logic tree for the symptom of the fault. The troubleshooting logic tree leads the maintainer through step-by-step procedures to determine that there is a malfunction and then to isolate the trouble to the malfunctioning assembly or part. The logic tree will reference tests and maintenance procedures (in section IV) where needed. At any point during the referenced tests that the indications are not as required, the maintainer returns to the logic tree. The logic tree will tell the maintainer what to do if the test has failed. Once the CAM has been repaired, it is again checked by the operational check to determine that the CAM is ready for a mission.
3-20
TM 3-6665-343-23P Use the following index to locate information in this troubleshooting section. Para
Subject
EQUIPMENT REQUIRED ................................................................................................................. CLEAN WORK SURFACE ................................................................................................................ PREPARATION OF DIAGNOSTIC TEST SET FOR USE................................................................... OPERATIONAL CHECK ................................................................................................................... TROUBLESHOOTING SYMPTOMS
1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. 12. 13. 14. 15. 16. 17. 18. 19. 20. 21.
MECHANICALLY BROKEN PUSH-BUTTON SWITCH OR CRACKED OR BROKEN DISPLAY .................................................................................................... PUMP NOT RUNNING ..................................................................................................... CAM AMPS NOT BETWEEN 3.6 AND 6.0, OR CAM DOES NOT COME ON ................... ALL INDICATIONS ARE NOT DISPLAYED DURING SELF-TEST .................................... DISPLAY DOES NOT COMPLETE SELF-TEST OR BL IS DISPLAYED WHEN CAM IS POWERED BY DTS ............................................................................... WAIT IS DISPLAYED LONGER THAN 2 MINUTES ......................................................... BL NOT SHOWN WHEN BATT LOW SWITCH IS PRESSED .......................................... NO DISPLAY OR BL DISPLAYED AFTER SELF-TEST WHEN CAM IS POWERED BY BATTERY ............................................................................................... NOZZLE FLOW IS LESS THAN 200 ml/min ..................................................................... GATE WIDTH, GATE HEIGHT, OR PULSE WIDTH NOT WITHIN SPECIFICATION ............................................................................................................. NOISE LEVEL IN WAVEFORM TEST IS EXCESSIVE ..................................................... SECONDARY PEAKS PRESENT IN WAVEFORM TEST OR CAM WILL NOT CLEAR DOWN ................................................................................................................ DC OFFSET IS GREATER THAN 1V................................................................................ (G OR H) PEAK HEIGHTS OR RATIO OF TIMES NOT WITHIN SPECIFICATION ............................................................................................................. CELL DOES NOT PASS CELL TEST ............................................................................... MONITOR MODULE FAILS LEAK TEST ......................................................................... DOES NOT RESPOND TO CONFIDENCE SAMPLE ....................................................... DISPLAY FLASHES ON AND OFF .................................................................................. DOES NOT CHANGE MODE FROM G TO H OR H TO G ............................................... NO DISPLAY OR BL DISPLAYED AFTER SELF-TEST WHEN CAM IS POWERED BY DTS ......................................................................................................... BUZZER DOES NOT SOUND WHEN DISPLAY SHOWS TWO BARS OR MORE ............
3-12 3-13 3-14 3-15
3-16 3-17 3-18 3-19 3-20 3-21 3-22 3-23 3-24 3-25 3-26 3-27 3-28 3-29 3-30 3-31 3-32 3-33 3-34 3-35 3-36
3-21
TM 3-6665-343-23P 3-12.
EQUIPMENT REQUIRED. The following equipment is required to troubleshoot the CAM: Tool Kit, Electronic Equipment TK-105/G Oscilloscope, Dual Trace, AN/USM-488 Multimeter, Digital, AN/PSM-45A (DMM) CAM Diagnostic Test Set Wrench, Torque (4-120 lb in. 3/8 in. drive) Screwdriver, Torque (10-120 oz in.) Adapter Socket (30A/F 1/2 in. drive) Bit, Screwdriver (1/4 in. A/F Hex Shank, cross-tipped) Blade, Screwdriver (1/4 in. A/F Hex Shank) Bit, Screwdriver (1/4 in. A/F Hex Shank, posidrive) Power Supply
3-13.
CLEAN WORK SURFACE.
Clean the work surface (Para 3-1). Place clean paper (Item 5, App D) across the top of the surface. 3-14.
PREPARATION OF DIAGNOSTIC TEST SET FOR USE. a. Release 12 knurled case lid knobs and remove case lid. b. Inspect inside lid knobs and remove case lid. (1) Container (for soap solution) (1) (2) Chassis assembly (2) (3) Nozzle extraction tool (3) (4) Battery contact assembly tool (4) (5) Nozzle flexible hose (5) (6) Cell flexible hose (6) (7) Partition (7) with the following items: (a) (b) (c) (d) (e) (f) (g) (h) (i)
Battery contact extraction tool (8) H.V. monitor adapter assembly (9) H.V. leads, test assembly (black and red) (9A) CAM power lead (W1) (10) Cell heater lead assembly (W5) (11) Display connector (W2) (12) Main cord set (W3) (13) Cell lead (W4) (14) Cell assembly to body assembly tool (15) (posidrive screwdriver) (j) Membrane Extraction Tool (16)
3-22
TM 3-6665-343-23P
3-23
TM 3-6665-343-23P 3-14.
PREPARATION OF DIAGNOSTIC TEST SET FOR USE (CONT).
c. Using the container of soap solution from the lid of the DTS, remove and fill the bulb (1) of the LEAK TEST flowmeter approximately three-quarters full of solution and replace bulb on flowmeter. Return bottle to lid of DTS for storage. d. Check that the 115 or 230 switch (2) is switched to proper voltage; if incorrect, remove two screws (3) and shield (4). Set switch to proper voltage. Reinstall shield and two screws. WARNING Make sure DTS SUPPLY switch (13) is down (off) to avoid electrical shock. e. Connect main cord set W3 (5) to DTS connector (6) and then to 115 V ac source power. f.
Connect W1P1 end of CAM power lead W1 to DTS CAM I/O connector (7).
g. Connect W2P2 end of display connector W2 to DTS 25 PIN connector (8). h. Connect W4P1 end of cell lead W4 to DTS 9 PIN connector (9). NOTE The meter indicates arbitrary units of measure - not actual volts or amperes. Therefore, the reading limits specified in the troubleshooting procedures should only be associated with use of the DTS. i.
Connect tubing end of nozzle flexible hose to DTS FLOW LEAK TEST NOZZLE fitting (10).
j.
Connect one end of cell flexible hose to DTS FLOW LEAK TEST CELL fitting (11). CAUTION Avoid using excessive force when removing or installing the bubble flowmeter. It is a fragile piece of glassware.
k.
Vigorously squeeze bulb (1) of LEAK TEST flowmeter until bubbles reach the top.
I.
Set R405 CURRENT switch (12) to NORM.
m. Press DTS SUPPLY switch (13) up to ON. n. Hold TEST DISPLAY switch (14) to TEST; all display indications should come on. o. Press DTS SUPPLY switch (13) down to OFF. 3-15. OPERATIONAL CHECK. a. Prepare DTS for use (para 3-14). b. Prepare work area (para 3-1). c.
3-24
Perform visual inspection for cracked or broken components of CAM.
TM 3-6665-343-23P
3-25
TM 3-6665-343-23P 3-15.
OPERATIONAL CHECK (CONT). d. Remove CAM environmental cap assembly; let it hang on the tether. e. Remove battery cap assembly. Ensure battery is removed. Reinstall battery cap assembly. f.
Connect W1P2 to CAM connector assembly. NOTE The CAM is protected from excessive current draw by the DTS automatic power shut off. To reset the DTS, press the supply switch off and then on again.
g. Press DTS SUPPLY switch (1) up to ON; SUPPLY indicator lamp comes on. If CAM does not come on, press CAM ON/OFF push-button switch. h. Listen for pump running. If pump is not running, perform troubleshooting symptom 2 (para 3-17). i.
On DTS, press CAM AMP switch (2).
j. DTS meter should indicate 3.6 to 6.0 units. If DTS meter does not indicate 3.6 to 6.0 units, perform troubleshooting symptom 3 (para 3-18). k. Press CAM ON/OFF push-button switch to turn CAM off. Wait 2 seconds. Press ON/OFF push-button switch again to start CAM in self-test. I.
Monitor CAM display for the following indications: (1) All eight bars are shown. (2) Markers A and B are shown. (3) Three vertical dots are shown. (4) BL is shown. (5) WAIT is shown. (6) H or G is shown. (7) If above 6 steps are not indicated, perform troubleshooting symptom 4 (para 3-19). (8) The display partially clears after 30 seconds; H or G, A and B markers, and WAIT remain. (9) After completion of self-test, press G/H push-button; opposite indication of step 6 above is shown. WAIT should clear within 2 minutes. If it does not clear within 2 minutes, perform troubleshooting symptom 6 (para 3-21). NOTE Disregard any bars that may reappear on CAM display for purposes of this test.
m. Press CAM VOLTS switch (3). Press and hold BATT LOW switch (4) until BL shows on CAM display. Wait 2 seconds and then release BATT LOW switch (4). IF BL is not shown on display while BATT LOW switch is pressed refer to troubleshooting symptom 7 (para 3-22).
3-26
TM 3-6665-343-23P NOTE If testing a monitor module assembly, proceed to step w. n. Press CAM ON/OFF push-button switch to turn CAM off. o. Disconnect CAM power lead W1 from CAM connector assembly. p. Install battery into CAM. q. Press CAM ON/OFF switch to turn CAM on, if necessary. CAM should complete self-test and BL should not remain lit. If CAM display does not light or BL remains lit after self test refer to troubleshooting symptom 8 (para 3-23). r.
Press CAM ON/OFF switch to turn CAM off.
s.
Remove battery from CAM.
t.
Connect DTS CAM power lead W1 to CAM connector assembly.
u. On DTS, set V2 (5) to NOZZLE INLET FLOW. Let CAM continue to run.
3-27
TM 3-6665-343-23P 3-15. OPERATIONAL CHECK (CONT). NOTE Remove and install nozzle protective cap as quickly as possible to avoid dust or other contamination from entering CAM protective cap. v. Twist (counterclockwise) and remove nozzle protective cap assembly from front of CAM; place nozzle protective cap assembly onto CAM environmental cap and twist clockwise. w. Push nozzle end of nozzle flexible hose over CAM nozzle assembly. x. Read FLOW 2 flowmeter. Reading should be greater than 200 (ml/min). Read flowmeter across top of float. If reading is less than 200, perform troubleshooting symptom 9 (para 3-24).
y. Remove nozzle flexible hose from CAM nozzle assembly. Twist (counterclockwise) and remove nozzle protective cap assembly from CAM environmental cap; place nozzle protective cap assembly onto front of CAM and twist clockwise. z. Perform waveform test (para 3-61). When a symptom occurs, refer to the appropriate troubleshooting procedure (para 3-11). aa. Perform confidence test (para 3-10). ab. Tun off DTS. Remove I/O cable W1 from CAM. Install environmental cap. ac. Perform wipe test (para 3-7) upon completion of all maintenance, before CAM is returned to user. ad. Remove DTS from use. (1) Remove all cables/connections attached to the front panel except cell flexible hose (1). (2) Pull bulb (2) from LEAK TEST flowmeter. Dump out any solution remaining in bulb. Attach bellows to cell flexible hose (1). Set valve V3 (3) to cell LEAK TEST. Squeeze bellows and look for liquid in bubble flowmeter. Continue squeezing bellows until no more liquid enters flowmeter. Remove cell flexible hose from front panel.
ae. Clean up and decontaminate work area (para 3-1b).
3-28
TM 3-6665-343-23P 3-16.
SYMPTOM 1. MECHANICALLY BROKEN PUSH-BUTTON SWITCH OR CRACKED OR BROKEN DISPLAY.
para 3-41
para 3-15
3-29
TM 3-6665-343-23P 3-17.
SYMPTOM 2. PUMP NOT RUNNING.
para 3-60
para 3-47
para 3-47
para 3-59
para 3-31
para 3-40
para 3-15
3-30
TM 3-6665-343-23P 3-18.
SYMPTOM 3. CAM AMPS NOT BETWEEN 3.6 AND 6.0 OR CAM DOES NOT COME ON. NOTE The DTS will shut off power to the CAM if the current is too high. If the meter rises, then drops to zero when the CAM is tugged on, the current is too high. Answer the question below "NO" and proceed to (B).
page 3-33
para 3-40
3-31
TM 3-6665-343-23P 3-18.
SYMPTOM 3. CAM AMPS NOT BETWEEN 3.6 AND 6.0 OR CAM DOES NOT COME ON (CONT).
para 3-41
para 3-43
para 3-15
3-32
TM 3-6665-343-23P
para 3-52
para 3-41
para 3-15
3-33
TM 3-6665-343-23P 3-18.
SYMPTOM 3. CAM AMPS NOT BETWEEN 3.6 AND 6.0 OR CAM DOES NOT COME ON (CONT).
para 3-53 para 3-50
para 3-15
para 3-53
para 3-53
para 3-15
para 3-52
para 3-40
3-34
TM 3-6665-343-23P
para 3-41
para 3-15
3-35
TM 3-6665-343-23P 3-18.
SYMPTOM 3. CAM AMPS NOT BETWEEN 3.6 AND 6.0 OR CAM DOES NOT COME ON (CONT).
para 3-60
para 3-61
para 3-58
TM 3-6665-332-13&P
3-36
TM 3-6665-343-23P
para 3-60
para 3-49
para 3-59
para 3-31
para 3-40
para 3-15
3-37
TM 3-6665-343-23P 3-18.
SYMPTOM 3. CAM AMPS NOT BETWEEN 3.6 AND 6.0 OR CAM DOES NOT COME ON (CONT).
para 3-47
para 3-59
para 3-31
para 3-40
para 3-15
3-38
TM 3-6665-343-23P
para 3-47
para 3-43
para 3-43
para 3-49 para 3-59
para 3-31
para 3-40
para 3-15
3-39
TM 3-6665-343-23P 3-19.
SYMPTOM 4. ALL INDICATIONS ARE NOT DISPLAYED DURING SELF-TEST.
para 3-40
3-40
TM 3-6665-343-23P
para 3-43
para 3-41
para 3-15
3-41
TM 3-665-343-23P 3-20.
SYMPTOM 5. DISPLAY DOES NOT COMPLETE SELF-TEST OR BL IS DISPLAYED WHEN CAM IS POWERED BY DTS.
para 3-43
para 3-15
3-42
TM 3-6665-343-23P 3-21.
SYMPTOM 6. WAIT IS DISPLAYED LONGER THAN 2 MINUTES.
para 3-15
para 3-15
para 3-31
3-43
TM 3-6665-343-23P 3-21.
SYMPTOM 6. WAIT IS DISPLAYED LONGER THAN 2 MINUTES (CONT).
para 3-61
para 3-15
3-44
TM 3-6665-343-23P 3-22.
SYMPTOM 7. BL NOT SHOWN WHEN BATT LOW SWITCH IS PRESSED.
para 3-43
para 3-15
3-45
TM 3-6665-343-23P 3-23.
SYMPTOM 8. NO DISPLAY OR BL DISPLAYED AFTER SELF-TEST WHEN CAM POWERED BY BATTERY.
para 3-42
3-46
para 3-15
TM 3-6665-343-23P
para 3-42
para 3-42
para 3-41
para 3-15
3-47
TM 3-6665-343-23P 3-24.
SYMPTOM 9. NOZZLE FLOW IS LESS THAN 200 ML/MIN.
para 3-52
para 3-53
para 3-53
para 3-53
para 3-52
para 3-52
para 3-15
para 3-40
para 3-15 para 3-60
page 3-50
3-48
TM 3-6665-343-23P
para 3-60
para 3-40
para 3-59
para 3-31
para 3-40
para 3-15
3-49
TM 3-6665-343-23&P 3-24.
SYMPTOM 9. NOZZLE FLOW IS LESS THAN 200 ML/MIN (CONT).
para 3-47
para 3-44
para 3-59
para 3-49
para 3-48
para 3-15
para 3-43
para 3-59
3-50
TM 3-6665-343-23&P
para 3-31
para 3-40
para 3-15
3-25.
SYMPTOM 10. GATE WIDTH, GATE HEIGHT, OR PULSE WIDTH NOT WITHIN SPECIFICATION.
para 3-43
para 3-15
3-51
TM 3-6665-343-23&P 3-26. SYMPTOM 11. NOISE LEVEL IN WAVEFORM TEST IS EXCESSIVE.
para 3-40
3-52
TM 3-6665-343-23&P
para 3-61
para 3-40
para 3-15
App D
para 3-61
para 3-40
para 3-15
3-53
TM 3-6665-343-23&P 3-26.
SYMPTOM 11. NOISE LEVEL IN WAVEFORM TEST IS EXCESSIVE (CONT).
para 3-47
para 3-47
para 3-59
para 3-31
para 3-61
3-54
TM 3-6665-343-23&P
para 3-40 para 3-43 para 3-15
para 3-40
para 3-15
para 3-31
para 3-49
para 3-59
para 3-40
para 3-15
3-55
TM 3-6665-343-23&P 3-27.
SYMPTOM 12. SECONDARY PEAKS PRESENT IN WAVEFORM TEST OR CAM WILL NOT CLEAR DOWN.
para 3-52
App D
para 3-53
para 3-59
para 3-31
3-56
TM 3-6665-343-23&P
para 3-52
para 3-61
para 3-15
para 3-60
3-57
TM 3-6665-343-23&P 3-27.
SYMPTOM 12. SECONDARY PEAKS PRESENT IN WAVEFORM TEST OR CAM WILL NOT CLEAR DOWN (CONT).
para 3-47
para 3-59
para 3-47
para 3-59
para 3-31
para 3-40
para 3-15
3-58
TM 3-6665-343-23&P
para 3-31
3-59
TM 3-6665-343-23&P 3-27.
SYMPTOM 12. SECONDARY PEAKS PRESENT IN WAVEFORM TEST OR CAM WILL NOT CLEAR DOWN (CONT).
para 3-61
para 3-44
para 3-40
para 3-61
3-60
para 3-15
TM 3-6665-343-23&P
para 3-49
para 3-59
para 3-31
para 3-40
para 3-15
3-61
TM 3-6665-343-23&P 3-28.
SYMPTOM 13. DC OFFSET IS GREATER THAN IV.
para 3-40
para 3-58
para 3-30
para 3-43
para 3-15
3-62
TM 3-6665-343-23&P 3-29.
SYMPTOM 14. (G OR H) PEAK HEIGHTS, OR RATIO OF TIME NOT WITHIN SPECIFICATION.
para 3-44
para 3-59
para 3-31
para 3-61
3-63
TM 3-6665-343-23&P 3-29.
SYMPTOM 14. (G OR H) PEAK HEIGHTS, OR RATIO OF TIME NOT WITHIN SPECIFICATION (CONT).
para 3-60
para 3-40
para 3-15
para 3-47
para 3-49
para 3-59
para 3-31
para 3-40
para 3-15
3-64
TM 3-6665-343-23&P
para 3-59
para 3-31
para 3-43
para 3-61
para 3-40
para 3-15
3-65
TM 3-6665-343-23&P 3-30. SYMPTOM 15. CELL DOES NOT PASS CELL TEST.
para 3-49
para 3-44
para 3-59
para 3-31
para 3-40
para 3-15
3-66
TM 3-6665-343-23&P 3-31. SYMPTOM 16. MONITOR MODULE FAILS LEAK TEST.
para 3-47
para 3-60
para 3-47
para 3-47
3-67
TM 3-6665-343-23&P 3-31.
SYMPTOM 16. MONITOR MODULE FAILS LEAK TEST (CONT).
para 3-59
para 3-31
para 3-40
para 3-15
3-68
TM 3-6665-343-23&P
para 3-53
para 3-47
para 3-44
para 3-59
para 3-49
para 3-44
para 3-59
para 3-31 para 3-40
para 3-15
3-69
TM 3-6665-343-23&P 3-32.
SYMPTOM 17. DOES NOT RESPOND TO CONFIDENCE SAMPLE. para 3-39
para 3-10
para 3-15
para 3-10
para 3-10
para 3-9
para 3-10
3-70
TM 3-6665-343-23&P
para 3-10
para 3-15
para 3-10
para 3-9
para 3-10
3-71
TM 3-6665-343-23&P 3-32.
SYMPTOM 17. DOES NOT RESPOND TO CONFIDENCE SAMPLE (CONT).
para 3-44
para 3-15
para 3-59
para 3-31
para 3-40
para 3-15
3-72
TM 3-6665-343-23&P 3-33.
SYMPTOM 18. DISPLAY FLASHES ON AND OFF.
para 3-10
para 3-9
para 3-9
para 3-15
para 3-10
3-73
TM 3-6665-343-23&P 3-34.
SYMPTOM 19. DOES NOT CHANGE MODE FROM G TO H OR H TO G.
para 3-40
para 3-58
para 3-41
para 3-40
para 3-15
3-74
TM 3-6665-343-23&P
para 3-30
para 3-43
para 3-15
3-75
TM 3-6665-343-23&P 3-35.
SYMPTOM 20. NO DISPLAY OR BL DISPLAYED AFTER SELF-TEST WHEN CAM IS POWERED BY DTS.
TM 3-6665-332-13&P
para 3-40
para 3-41
para 3-43
3-76
para 3-15
TM 3-6665-343-23&P 3-36.
SYMPTOM 21. BUZZER DOES NOT SOUND WHEN DISPLAY SHOWS TWO BARS OR MORE.
para 3-43
para 3-15
3-77
TM 3-6665-343-23&P Section IV. DIRECT SUPPORT MAINTENANCE PROCEDURES 3-37. INTRODUCTION. WARNING RADIATION HAZARD
NICKEL-63 (Ni-63) Wipe test must be performed before and after any maintenance or testing of CAM. This section consists of general maintenance instructions, detailed Direct Support maintenance procedures, and performance tests. Read and understand the general instructions before beginning the detailed procedures. Use the following index to locate Direct Support maintenance procedures and performance tests. Performance Tests
Para
CELL TEST......................................................................................................................................... LEAK TEST ........................................................................................................................................ PUMP TEST ....................................................................................................................................... WAVEFORM TEST .............................................................................................................................
3-58 3-59 3-60 3-61
Assembly/Part - Maintenance Procedures
Para
CONFIDENCE SAMPLE REPLACEMENT .......................................................................................... MONITOR CASE ASSEMBLY REPLACEMENT.................................................................................. DISPLAY ASSEMBLY REPLACEMENT.............................................................................................. BATTERY CONTACT ASSEMBLY REPLACEMENT........................................................................... PRINTED CIRCUIT BOARD REPLACEMENT AND ADJUSTMENT .................................................... SIEVE PACK ASSEMBLY REPLACEMENT........................................................................................ CONNECTOR ASSEMBLY REPLACEMENT ...................................................................................... PRINTED WIRING BOARD REPLACEMENT...................................................................................... PUMP ASSEMBLY REPLACEMENT .................................................................................................. FILTER ASSEMBLY REPLACEMENT................................................................................................. DRIFT TUBE MODULE REPLACEMENT AND ALIGNMENT .............................................................. NOZZLE HOLDER ASSEMBLY REPLACEMENT ............................................................................... CASE FRONT-END REPLACEMENT ................................................................................................. NOZZLE ASSEMBLY REPLACEMENT............................................................................................... MEMBRANE ASSEMBLY REPLACEMENT ........................................................................................ ENVIRONMENTAL CAP ASSEMBLY REPLACEMENT....................................................................... BATTERY CAP ASSEMBLY REPLACEMENT .................................................................................... CARRYING CASE ASSEMBLY REPLACEMENT................................................................................ CARRYING CASE STRAP REPLACEMENT .......................................................................................
3-39 3-40 3-41 3-42 3-43 3-44 3-45 3-46 3-47 3-48 3-49 3-50 3-51 3-52 3-53 3-54 3-55 3-56 3-57
3-78
TM 3-6665-343-23&P 3-38.
GENERAL MAINTENANCE INSTRUCTIONS. The following general maintenance procedures are to be followed whenever performing maintenance on a CAM.
a. Inspect CAM, including case and display window for any cracked or broken parts. Replace any cracked or broken parts. b. Inspect all removed assemblies/parts for tears, breaks, indications of wear, and corrosion. Replace any parts showing these faults. c.
Observe all parts for cleanliness. If parts appear to have dust on them, blow the dust off.
d. Check the threads of all removed screws for the presence of Tuflok. If Tuflok no longer remains in the threads, replace the screw(s). e. Prior to working on a CAM, prepare a clean work surface. Follow work preparation instructions per Chapter 3, Section I. f. DO NOT remove a new CAM assembly/part from its packaging until you are ready to install it. This will keep the assembly/part clean and dry. g. Open the pneumatic loop of a CAM (opening the sieve pack, pump assembly, drift tube module, membrane, nozzle assembly) only long enough to make required tests and/or replacements.
3-79/(3-80 blank)
TM 3-6665-343-23&P 3-39.
CONFIDENCE SAMPLE REPLACEMENT.
This task covers: Replacement of Confidence Sample. INITIAL SETUP Tools None Materials/Parts Confidence Sample Equipment Condition None REPLACEMENT 1. Open vapor proof barrier (1) around confidence sample (2). 2. Pull confidence sample (2) from vapor proof barrier (1). Inspect confidence sample for damage and verify that valves move. 3. Place new confidence sample (2) into pocket of carrying harness (3).
3-81
TM 3-6665-343-23&P 3-40.
MONITOR CASE ASSEMBLY REPLACEMENT.
This task covers: Removal and installation of Monitor Case Assembly. INITIAL SETUP Tools Electronic Equipment Tool Kit TK-105/G 1-3/16 Socket Adapter Torque Wrench Clean Work Surface (para 3-13) Materials/Parts Monitor Case Assembly Insulation Sheet, Electrical Gloves (Item 4, App D) Equipment Condition CAM turned off, DTS turned off, and all cables disconnected from CAM. REMOVAL 1. Remove battery cap assembly (1) by twisting (counterclockwise). 2. Remove battery (2). 3. Remove environmental cap assembly (3) with tether by twisting (counterclockwise). 4. Remove locknut (4), drift tube module lot number disc (5), and CAM lot number disc (6) from connector assembly (7). Save discs and locknut. 5. Push lightly on connector assembly (7) until monitor module assembly (8) separates from monitor case assembly (9). Pull monitor module assembly (8) (about 2 inches) out of monitor case assembly (9) until all of display connector (10) is fully visible. Check that case seal (11) is in place on case front-end (12); if not, press case seal into place (lip of seal faces groove of case front-end). 6. Grip tab (13) and disconnect flexi connector (14); gently pull on tab, as shown, to remove flexi connector. 7. Holding flexi connector (14) up and clear, as shown, completely separate monitor module assembly (8) from monitor case assembly (9). Locate O-ring seal (15) on connector assembly (7). If O-ring is not seen, look for it inside case assembly (9) and pull it out. 8. Inspect O-ring seal; if damaged, replace. 9. Place O-ring seal into groove of connector assembly (7). (It fits loosely.) Place CAM lot number disc (6), drift tube module lot number disc (5), and locknut (4) onto connector assembly (7). 10. Remove and discard insulation sheet (16) from around monitor module assembly (8).
3-82
TM 3-6665-343-23&P
3-83
TM 3-6665-343-23&P 3-40.
MONITOR CASE ASSEMBLY REPLACEMENT (CONT).
INSTALLATION 1. Install new insulation sheet (16) around monitor module assembly (8) and hold it in place: position as shown. NOTE In steps 2 through 5, keep monitor module assembly tilted with the rear connector facing upward so that the O-ring seal (15) remains in place. 2. Remove locknut (4), drift tube module lot number disc (5), and CAM lot number disc (6) from connector assembly (7). 3. Fold flexi connector (14) out of the way above the front access area of the monitor case assembly (9) and hold the flexi connector in place. 4. Partially (until the display connector (10) is still just accessible) slide monitor module assembly (8) with insulation sheet (16) into monitor case assembly (9). 5.
Using tab (13) to hold flexi connector (14), gently press flexi connector (14) to mate with display connector (10) until flush with PCB. Look under tab (13) at connector to be sure sockets have not been pushed out of connector; if they are out, press sockets gently into connector with small flat-tip screwdriver. Slide monitor module assembly (8) completely into monitor case assembly (9).
6. Hold CAM vertically and look through hole in end at the connector assembly (7) to see if O-ring seal (15) is in the groove around connector assembly (7). 7. Install CAM lot number disc (6), drift tube module lot number disc (5), and locknut (4) onto connector assembly (7). NOTE Insure discs are not engaged in connector threads. When properly assembled, one to two threads will extend through the locknut. 8. Check that the gap between the case front-end (12) and the monitor case assembly (9) has closed and that the case seal (11) is not pinched. Tighten locknut (4) until tight; torque to 87 in. lb. 9. Install environmental cap assembly with tether (3) by twisting (clockwise). 10. Install battery (2) (terminal end goes in first). 11. Install battery cap assembly (1) by twisting (clockwise).
3-84
TM 3-6665-343-23&P
3-85
TM 3-6665-343-23&P 3-41. DISPLAY ASSEMBLY REPLACEMENT.
This task covers: Removal and installation of Monitor Case Assembly. INITIAL SETUP Tools Electronic Equipment Tool Kit TK-105/G Hand Screw Starter Clean Work Surface (para 3-13) Materials/Parts Display Assembly Gloves (Item 4, App D) Equipment Conditions CAM turned off, DTS turned off, and all cables disconnected from CAM. Monitor case assembly removed (para 3-40). REMOVAL 1. Hold monitor case assembly (1) and flexi connector (2), as shown, and remove retaining clip (3). To remove clip, press against the "U" shape of the clip. 2. Press display board (4) and flex cable (5) down into case assembly (1) to clear the display board from the retaining channels (6). 3. Remove two screws (7); lift out display assembly (8). 4. Lift up tab and pull connector (9) from display assembly (8). INSTALLATION 1. Reach through front of monitor case assembly and pull out the battery terminal coiled cable (10) and connector (9). 2. Insert connector (9) into plug on new display assembly (8). 3. Use a 3/16 inch flat-tip screwdriver to press coil of wire (10) into handle of monitor case assembly (1); coil should be fully inside handle. NOTE Switches must be under rubber boots. 4. Press switches (11) into place. Ensure that the switches are located correctly beneath the rubber switch covers (12) on the monitor case assembly. Install two screws (7). Press ON/OFF and G/H switches several times to be sure both switches click when pressed. 5. Slide the display board (4) into position in the retaining channel (6) in the monitor case assembly (1). 6. Install retaining clip (3) into retaining channels.
3-86
TM 3-6665-343-23&P
CAUTION To avoid damage to the equipment, ensure that the ends of the clip are fully inserted in the retaining channels and the clip does not contact any pins on the display assembly.
3-87
TM 3-6665-343-23&P 3-42.
BATTERY CONTACT ASSEMBLY REPLACEMENT.
This task covers: Removal and installation of Monitor Case Assembly. INITIAL SETUP Tools Electronic Equipment Tool Kit TK-105/G Battery Contact Extraction Tool (from DTS) Battery Contact Assembly Tool (from DTS) Clean Work Surface (para 3-13) Materials/Parts Battery Contact Assembly Gloves (Item 4, App D) Twine (Item 9, App D) Equipment Conditions Monitor case assembly removed (para 3-40). Display assembly removed (para 3-41). REMOVAL 1. Hook battery contact extraction tool (1) under the plastic portion of terminal assembly (2). 2. Pull the battery contact assembly (3) out of the case assembly handle (4).
3-88
TM 3-6665-343-23&P
INSTALLATION 1.
Tie twine to 2-pin connector (5) and coil of battery contact assembly (3), and feed twine into case assembly handle (4) as shown in Step 1. 2. While pulling twine out the front of case assembly handle, use the battery contact extraction tool (1) to guide battery contact assembly (3) into case assembly handle (4) as shown in Step 2. 3. Fit the terminal assembly (2) onto the battery contact assembly tool (6) as shown in Step 3. NOTE Two tabs (7) should be in vertical position as shown in the exploded view on previous page to allow the terminal assembly to snap in place. 4. Vertically align two tabs (7) on terminal assembly (2) with mating "tabs" in case assembly handle (4). Observe that the coil is in position for battery contact extraction tool (1) to reach and pull coil through front of case assembly handle (4). Push and turn battery contact assembly until it snaps into place at the display end of the battery compartment.
5. Pull connector (5) and coil through case assembly handle (4) with battery contact extraction tool (1).
3-89/(3-90 blank)
TM 3-6665-343-23&P 3-43. PRINTED CIRCUIT BOARD REPLACEMENT AND ADJUSTMENT.
This task covers: Removal and installation of Monitor Case Assembly. INITIAL SETUP Tools Electronic Equipment Tool Kit TK-105/G Posidrive Torque Screwdriver Oscilloscope Diagnostic Test Set Multimeter Test Lead Set Clean Work Surface (para 3-13) Materials/Parts Printed Circuit Board Gloves (Item 4, App D) Equipment Condition CAM turned off, DTS turned off, and all cables disconnected from CAM. Monitor case assembly removed (para 3-40). REMOVAL 1. Remove two screws (1). The connector assembly (2) will hinge on two lower screws (3). 2. Gently remove PCB connector (4) from connector assembly (2); use a small flat-tip screwdriver, as shown.
3-91
TM 3-6665-343-23P 3-43. PRINTED CIRCUIT BOARD REPLACEMENT AND ADJUSTMENT (CONT). CAUTION While performing the next procedure, avoid excessive flexing of the cell connector film wire to prevent damage to the equipment. 3. Disconnect cell connector (5) by pulling straight back from PCB assembly. Disconnect three electrical connectors (6, 7 and 8) from PCB assembly. 4. Remove four screws (9).
NOTE In the following steps, avoid hitting the PCB connectors on the monitor assembly. 5. Hold PCB assembly (10) while gently pulling it back, out of the case front-end (11) (approximately 1/8 inch).
CAUTION While performing the next procedure, avoid excessive flexing of the board-side coaxial cable to prevent damage to the equipment. 6. Use a small flat-tip screwdriver to remove coaxial pin (12) and socket (13) from clip (14) on the PCB assembly (10). Separate coaxial pin and socket. Press coaxial socket (13) back into clip (14). 3-92
TM 3-6665-343-23&P INSTALLATION NOTE If installing the same PCB assembly and drift tube module, proceed with steps 2, 3, and 6 through 10 only. 1. If installing a new PCB assembly or new drift tube module, rotate the adjustment screw of VR3 fully counterclockwise (until it clicks). 2. Grip PCB assembly (10) between top and bottom edges, and insert coaxial pin (12) into socket (13). Place pin and socket into clip (14). 3. Place tongue of PCB assembly (10) into the slot in the case front-end. Ensure the pump lead is positioned as shown below. Secure PCB with four screws (9); do not tighten screws. Thread PCB connector (4) between sieve pack assembly (15) and the connector assembly (2); connect PCB connector (4). (Align and insert bottom pins of connector first.)
3-93
TM 3-6665-343-23P 3-43.
PRINTED CIRCUIT BOARD REPLACEMENT AND ADJUSTMENT (CONT).
INSTALLATION (CONT). NOTE VR3 is a potentiometer that adjusts the current draw by the membrane heater. 4. Adjustment of VR3. Make the following adjustment if a new PCB or drift tube module is being installed; if not a new PCB or drift tube module, proceed to step 5. WARNING DANGER - HIGH VOLTAGE - 1,000 VOLTS DC Serious injury can occur during this process if proper procedures are not followed. During this adjustment, power will be applied to the flexiboard. 1000 Volts DC is present. Do not use metal work surface. Do not use metal blade tools to make adjustment. Variable resistor VR3 should be adjusted with alignment tool. CAUTION If VR3 is not adjusted, cell damage may result. (a) Connect DMM to R405 test sockets. Switch leads at DMM, if necessary, to measure current. Set DMM to 2000 mA DC range. Set R405 CURRENT switch to TEST position. (b) Connect W1P1 end of CAM power lead W1 to DTS CAM I/O connector and W1P2 end of W1 to CAM connector assembly. (c) Connect W2P2 end of display connector W2 to DTS 25 PIN connector. Connect W2P1 end of W2 to PCB connector P2 on monitor module assembly. (d) Place monitor module assembly in chassis assembly of DTS.
3-94
TM 3-6665-343-23&P
3-95
TM 3-6665-343-23P 3-43. PRINTED CIRCUIT BOARD REPLACEMENT AND ADJUSTMENT (CONT). INSTALLATION (CONT). (e) Connect pump motor connector (8) to P6. (f) Press DTS SUPPLY switch up to ON. Press TEST DISPLAY ON switch to turn monitor module assembly on. (g) Wait 60 seconds and measure and record the "baseline current" in milliamperes shown on the DMM. (h) Press TEST DISPLAY ON switch to turn monitor module assembly off. CAUTION When adjusting VR3 in the next step, ensure that the adjustment is made slowly so that the calculated current is NOT exceeded. If the current is excessive, the membrane heater may be destroyed. (i) Connect cell filmwire connector (5) to P4. Press TEST DISPLAY ON switch to turn monitor module assembly on. Adjust VR3 until DMM indicates baseline current recorded in step 4(g) plus 76 ±5 mA. (j) Press TEST DISPLAY ON switch to turn monitor module assembly off. (k) Connect nozzle heater connector (6) to P1. To the baseline current recorded in step 4(g) add 186 mA. (I) Press TEST DISPLAY ON switch to turn monitor module assembly on. Wait 2 minutes. Verify that current indicates the value of step 4(k) ± 40 mA. If not, repeat adjustment starting at step (e). (m) Press TEST DISPLAY ON switch to turn monitor module assembly off. Disconnect cell filmwire (5). Connect heaters/PCB connector (7) to P3. Connect cell filmwire (5). (n) Disconnect display connector (16) from P2. Disconnect CAM power lead W1 from connector assembly (2). 5. Disconnect the DMM from the DTS and set R405 CURRENT switch to NORM.
3-96
TM 3-6665-343-23&P
3-97
TM 3-6665-343-23P 3-43.
PRINTED CIRCUIT BOARD REPLACEMENT AND ADJUSTMENT (CONT).
INSTALLATION (CONT). 6. Visually inspect that there is clearance between the components of the PCB assembly (10) and the monitor assembly. 7. Connect heater/PCB connector (7) (if not connected previously). 8. Connect electrical connectors (5, 6 and 8) to PCB assembly (if not connected previously). 9. Tighten four screws (9); torque screws to 90 oz. in. 10. Hinge connector assembly (2) into place; secure with two screws (1). Torque screws to 50 oz. in.
NOTE VR1 is a potentiometer used to adjust amplification of the output of the drift tube module. Whenever the PCB assembly or drift tube module is replaced, VR1 must be adjusted to tune the PCB assembly to the drift tube module. It is important that this adjustment be done when the CAM has a good sieve pack installed. Peak heights will drop as sieves age. If adjustment is attempted with nearly depleted sieves, the CAM will not perform correctly. It is critical that the CAM pass all criteria in the waveform test except peak height. If not, troubleshooting should be performed prior to attempting this adjustment. 3-98
TM 3-6665-343-23&P 11. Adjustment of VR1. Make the following adjustment if a new PCB assembly or new drift tube module is being installed; if not a new PCB assembly or drift tube module, complete the assembly of the CAM. (a) If not already in place, place monitor module assembly into chassis assembly. If not connected, connect display connector from DTS to P2. (b) Connect CAM power lead W1 from DTS to connector assembly. Connect ground wire of power lead to ground stud on chassis assembly. (c) Set DTS SUPPLY switch to ON; if monitor module assembly is not running, press TEST DISPLAY ON. Let monitor module assembly run for 1 hour. (d) Perform waveform test (para 3-61), ignoring any amplitude measurements. Perform troubleshooting if necessary. When completed with the waveform test, return here. (The oscilloscope will be used in the adjustment of VR1.) (e) Look at both the G and H reactant ion peaks (RIP). When switching modes for this adjustment, wait 2 minutes after each switchover. Set VR1 so that the average of the two peaks is as close as you can get to 3.1 V keeping both peaks between 2.4 and 3.7 V.
3-99
TM 3-6665-343-23&P 3-44.
SIEVE PACK ASSEMBLY REPLACEMENT.
This task covers: Removal and installation of Monitor Case Assembly. INITIAL SETUP Tools Electronic Equipment Tool Kit TK-105/G Torque Screwdriver Clean Work Surface (para 3-13) Materials/Parts Sieve Pack Assembly Gloves (Item 4, App D) Equipment Condition CAM turned off, DTS turned off, and all cables removed from CAM. Monitor case assembly removed from CAM (para 3-40).
WARNING
RADIATION HAZARD
NICKEL-63 (Ni-63) This CAM contains a beta radiation source in the quantity which requires a Nuclear Regulatory Commission license to possess items which contain the source. The CAM contains 10 millicuries of Nickel-63. REMOVAL NOTE Have the replacement sieve pack assembly ready, to enable reassembly to be completed as quickly as possible. This will minimize degradation of the sieve. 1. Disconnect acetone source heater connector (1). 2. Loosen sieve pack retaining screw (2). 3-100
TM 3-6665-343-23&P CAUTION Perform next procedure carefully to avoid damage to PCB. 3. Withdraw the sieve pack assembly (3) from the monitor assembly (4) by sliding the sieve pack to the rear and lifting upward. 4. Remove two screws (5). Remove acetone source (6) and dispose of as regular waste. 5. Dispose of sieve pack (7) as radioactive waste.
INSTALLATION CAUTION Handle the acetone source by metal cap only to avoid contaminating it. 1. Remove sieve pack (7) and acetone source (6) and o-ring (8) from packaging. Handle the acetone source by metal cap only. Do not touch the tube. 2. Slide o-ring (8) up acetone source (6) until it is seated in slot right under the cap. 3. Insert the acetone source (6) into the sieve pack (7) with wire facing forward, and secure with two screws (5). Torque screws to 65 oz. in. 4. Assemble o-ring (9) over wire of acetone source heater connector wire and the sieve front seal. 5. Insert the sieve pack assembly (3) down into the monitor assembly (4), then slide the sieve pack assembly (3) forward to engage with the manifold assembly (10). 6. Secure sieve pack assembly (3) by tightening retaining screw (2). 7. Connect acetone source heater connector (1). 3-101
TM 3-6665-343-23&P 3-45.
CONNECTOR ASSEMBLY REPLACEMENT.
This task covers: Removal and installation of Monitor Case Assembly. INITIAL SETUP Tools Electronic Equipment Tool Kit TK-105/G Torque Screwdriver Clean Work Surface (para 3-13) Materials/Parts Connector Assembly Gloves (Item 4, App D) Equipment Condition CAM turned off, DTS turned off, and all cables disconnected from CAM. Monitor case assembly removed (para 3-40). PCB removed from monitor module assembly (para 3-43). REMOVAL 1. Remove screw (1) and threaded standoff (2). 2. Remove connector assembly (3) from monitor assembly (4). INSTALLATION 1. Install connector assembly (3) to monitor assembly (4). 2. Secure connector assembly (3) with screw (1) and threaded standoff (2); torque to 50 oz. in.
3-102
TM 3-6665-343-23&P 3-46.
PRINTED WIRING BOARD REPLACEMENT.
This task covers: Removal and installation of Monitor Case Assembly. INITIAL SETUP Tools Electronic Equipment Tool Kit TK-105/G Clean Work Surface (para 3-13) Materials/Parts Gloves (Item 4, App D) Equipment Condition Monitor case assembly removed (para 3-40).
CAUTION While performing the next procedure, avoid excessive flexing of the cell connector filmwire to prevent damage to the equipment. REMOVAL 1. Disconnect cell connector (1) by pulling straight back from PCB assembly. Disconnect electrical connectors (2) and (3) from printed wiring board (5). Disconnect third electrical connector (4) from PCB. 2. Remove screw (6) and spacer (7). INSTALLATION 1. Position spacer (7) between chassis (8) and printed wiring board (5) and secure printed wiring board (5) to chassis (8) with screw (6). 2. Connect two electrical connectors (2) and (3) to the printed wiring board (5). Connect third electrical connector (4) to PCB. Connect cell connector (1) to PCB.
3-103
TM 3-6665-343-23&P 3-47.
PUMP ASSEMBLY REPLACEMENT.
This task covers: Removal and installation of Monitor Case Assembly. INITIAL SETUP Tools Electronic Equipment Tool Kit TK-105/G Clean Work Surface (para 3-13) Materials/Parts Pump Assembly Gloves (Item 4, App D) Equipment Condition CAM turned off, DTS turned off, and all cables disconnected from CAM. Monitor case assembly removed (para 3-40). PCB assembly removed from monitor module assembly (para 3-43). Sieve pack assembly removed (para 3-44). REMOVAL 1. Feed the pump motor cable (1) through the chassis plate until it is clear. 2. While holding the monitor assembly (2), push forward and then lift the pump retaining plate (3) upward. Once the plate is free enough to maneuver, withdraw it rearward. 3. Pull the pump (4) and pump supporting plate (5) away from the monitor assembly (2). 4. Separate the pump and pump support plate. 5. Ensure the two seals (6) are removed.
3-104
TM 3-6665-343-23&P INSTALLATION 1. Fit the pump support plate (5) onto the pump assembly (4). Ensure the rear seals are properly aligned against the pump support plate. 2. While holding the monitor assembly (2) vertically, fit two seals (6) into manifold assembly (7). 3. Position the pump support plate (5) (with the notched end toward the bottom of the CAM) and the pump assembly (4) into the monitor assembly (2) and press forward with slight pressure, ensuring seals are correctly located at the front. 4. While maintaining pressure, install the pump retaining plate (3) by locating the lugs into the side slots, pushing forward and then downward. NOTE Tuck wiring into L slot in chassis plate. Ensure sleeving aligns with slot.
3-105
TM 3-6665-343-23&P 3-48.
FILTER ASSEMBLY REPLACEMENT.
This task covers: Removal and installation of Monitor Case Assembly. INITIAL SETUP Tools Electronic Equipment Tool Kit TK-105/G Clean Work Surface (para 3-13) Materials/Parts Filter Assembly Tiedown Strap Gloves (Item 4, App D) Equipment Condition CAM turned off, DTS turned off, and all cables removed from CAM. Monitor case assembly removed (para 3-40).
NOTE Discard all materials as normal waste. REMOVAL 1. Cut tie down straps (1) and discard. 2. Pull filter tubes (2) from inlet connection seal (3), and sealing tube (4) and discard. 3. Inspect inlet connection seal (3), sealing tubes (4) and exhaust tubes (5) and (6) for cracks, tears, and holes. (a) If sealing tube (4) is damaged, remove sieve pack and pump assembly. Pull sealing tube from the rear of the manifold assembly (7) and discard. (b) If inlet connection seal (3) is damaged, proceed as follows: (1) Remove the case front-end. (2) Pull and stretch inlet connection seal (3) and cut off as close to the drift tube (8) as possible. (3) Remove sealing plug (9), push remainder of inlet connection seal (3) through the front of the drift tube (8) with screw driver and discard. (c) If either exhaust tube (5) or (6) is damaged, remove by pulling apart from sealing tube (4) and drift tube (8) and discard.
3-106
TM 3-6665-343-23&P INSTALLATION 1. If sealing tube is removed, gently slide new tube (4) into manifold assembly (7). Install pump and sieve pack. If inlet connection seal is removed, push new seal (3) through the front of the drift tube (8) until it snaps into place. Install sealing plug (9) and case front-end. If exhaust tubes are removed, install new tubes (5) and (6) by pushing into sealing tube (4) and drift tube (8). 2. Fit straight side of filter (10) into inlet connection seal (3). Fit bent side of filter into sealing tube (4). 3. Place new tie down straps (1) around drift tube module (8), filter assembly (10), and exhaust tube (5). Fasten down strap.
NOTE Position tie down strap fasteners at bottom of drift tube.
3-107
TM 3-6665-343-23&P 3-49.
DRIFT TUBE MODULE REPLACEMENT AND ALIGNMENT.
This task covers: Removal and installation of Monitor Case Assembly. INITIAL SETUP Tools Electronic Equipment Tool Kit TK-105/G Torque Screwdriver Clean Work Surface (para 3-13) Materials/Parts Drift Tube Module Sieve Pack Assembly Gloves (Item 4, App D) Contact Enhancer (Item 3, App D) Equipment Conditions CAM turned off, DTS turned off and all cables disconnected from CAM. Monitor case assembly removed (para 3-40). PCB removed from monitor module assembly (para 3-43). Sieve pack assembly removed (para 3-44). Pump assembly removed (para 3-47). Nozzle assembly removed (para 3-52).
WARNING RADIATION HAZARD
NICKEL-63 (Ni-63) This CAM contains a beta radiation source in the quantity which requires a Nuclear Regulatory Commission license to possess items which contain the source. The CAM contains 10 millicuries of Nickel-63.
3-108
TM 3-6665-343-23&P REMOVAL NOTE Sieve pack assembly is always replaced when drift tube module is replaced. 1. Disconnect drift tube heater connector (1). 2. Remove three screws (2). 3. Remove threaded standoff (3). 4. Slide coaxial cable (4) through grommet (5). 5. Remove chassis assemblies (6) from monitor assembly.
6. Loosen three screws (7) from case front-end (8). 7. Pull assembly (9) from case front-end (8).
3-109
TM 3-6665-343-23&P 3-49.
DRIFT TUBE MODULE REPLACEMENT AND ALIGNMENT (CONT).
REMOVAL (CONT). CAUTION Shims may be located on the screws to be removed in the next step. It is important that the same number of shims are returned to the same screw position (keeps CAM in alignment). Therefore, tag/mark the number of shims relative to each screw position as the shims are removed. 8. While holding nozzle holder assembly (10) against assembly (9), remove shim (retaining sleeve) (11) and shim(s) (nozzle holder/front cover) (12) from screw (7). Tag/mark shims (12) as they are removed. Do this for each of three screws. 9. Gently pull exhaust (rubber) tube (13) from port. 10. Hold assembly vertically, as shown (screw threads down), and carefully pull nozzle holder assembly (10) from assembly (9). 11. Remove three screws (7).
NOTE Proceed with next step ONLY if it has been determined that drift tube is faulty. 12. Pack drift tube module disc with drift tube module, using packaging from new drift tube module. 13. Seal drift tube module package.
3-110
TM 3-6665-343-23&P INSTALLATION CAUTION Do not remove drift tube module from packaging until necessary to avoid contamination. NOTE Wipe test (para 3-7) must be performed after a new drift tube module is installed. 1. Remove drift tube module disc from drift tube module packaging; save disc for later installation. Remove drift tube module from packaging. Insert three screws (7) through assembly (9). 2. Obtain new sealing plug from drift tube module kit. Press sealing plug (14) into place. 3. Feed exhaust (rubber tube) (13) through nozzle holder assembly (10) and attach to port. Position nozzle holder assembly (10) to assembly (9) as it passes over three screws (7). 4. Place shim(s) (nozzle holder/front cover) (12) onto screws (7) according to tags/marks established at disassembly. Place shim(s) (retaining sleeves) (11) on each screw (7) to hold shims (12) in place. (In case of loss, spare shims are provided in the drift tube module kit.)
3-111
TM 3-6665-343-23&P 3-49.
DRIFT TUBE MODULE REPLACEMENT AND ALIGNMENT (CONT).
INSTALLATION (CONT). 5. Remove old seal (holder/case) (15) and discard. 6. Obtain new seal (holder/case) (15) from drift tube module kit. Place seal (holder/case) (15) in groove in case front-end (8). 7. Holding drift tube module together, mate it to case front-end, as shown. Secure with three screws (7); torque to 80 oz. in.
8. Ensure by visual inspection that seal (holder/case) (15) is seated and sealing correctly. 9. Install nozzle assembly (para 3-50). Install nozzle protective cap assembly. 10. Install chassis right hand and left hand assemblies (6) with three screws (2) and torque to 30 oz. in. 11. Install threaded standoff (3). 12. Hinge connector assembly (16) into place; secure with two screws (17). Torque to 30 oz. in.
3-112
TM 3-6665-343-23&P ALIGNMENT 1. Remove nut (18) and o-ring (19) from connector assembly (16). 2. Check that case seal (20) is correctly oriented and in position on case front-end (8); if not, position case seal on case front-end.
NOTE In the following step, be careful not to damage the connectors. 3. Place old insulation sheet around chassis; be sure connectors are inside sheet. 4. Slide monitor assembly (21) into monitor case assembly (22). 5. With hand pressure, mate the monitor case assembly (22) to the case front-end (8). The connector assembly (16) will extend out through the hole in the rear of the monitor case assembly. 6. Visually inspect that the monitor case assembly meets the case seal (20) and case front-end evenly. No gaps should appear. If there is a gap, proceed as follows: (a) Remove monitor assembly from monitor case assembly. Move chassis (6) in an attempt to rectify the gap. (b) Repeat alignment steps (5) and (6).
3-113
TM 3-6665-343-23&P 3-49.
DRIFT TUBE MODULE REPLACEMENT AND ALIGNMENT (CONT).
ALIGNMENT (CONT). 7. If alignment is satisfactory, remove monitor assembly (21) from monitor case assembly and tighten three screws (2); torque to 90 oz. in. 8. Insert PCB retaining screw into threaded standoff (3) and torque to 90 oz. in. Hold standoff (3) with pliers and remove screw. Repeat alignment steps (2) thru (6) as a final visual check; if not in alignment, loosen screws (2) and threaded standoff (3) and repeat alignment steps (2) thru (9). 9. Remove monitor assembly (21) from monitor case assembly. Remove and discard old insulation sheet. 10. Feed coaxial cable (4) through grommet (5). 11. Connect drift tube heater connector (1) as shown.
FINAL INSTALLATION 1. Remove two screws (17). 2. Dip coaxial connector (4) in contact enhancer.
3-114
TM 3-6665-343-23&P 3-50.
NOZZLE HOLDER ASSEMBLY REPLACEMENT.
This task covers: Removal and installation of Monitor Case Assembly. INITIAL SETUP Tools Electronic Equipment Tool Kit TK-105/G Torque Screwdriver Clean Work Surface (para 3-13) Materials/Parts Nozzle Holder Assembly Gloves (Item 4, App D) Equipment Condition CAM turned off, DTS turned off and all cables disconnected from CAM. Monitor case assembly removed (para 3-40). PCB removed from monitor module assembly (para 3-43). Nozzle assembly removed (para 3-52). REMOVAL 1. Loosen three screws (1) from case front-end (2). CAUTION Shims are located on the screws to be removed in the next step. It is important that the same shims are returned onto the same screws (keeps CAM in alignment). Therefore, tag/mark the shims relative to each screw position as the shims are removed. 2. Pull assembly (3) from case front-end (2).
NOTE: CHASSIS ASSEMBLY REMOVED FOR CLARITY
3-115
TM 3-6665-343-23&P 3-50.
NOZZLE HOLDER ASSEMBLY REPLACEMENT (CONT).
REMOVAL (CONT). 3. Remove end of exhaust (rubber) tube (4) from nozzle holder assembly (5) at the port. 4. While holding nozzle holder assembly (5) against assembly (3), remove shim (retaining sleeve) (6) and shim(s) (nozzle holder/front cover) (7) from screw (1). Tag/mark shims (7) as they are removed. Do this for each of three screws. 5. Hold assembly vertically, as shown (screw threads down), and carefully pull nozzle holder assembly (5) from assembly (3).
NOTE: FILTER INLET PLUGGED
3-116
TM 3-6665-343-23&P INSTALLATION 1. Feed exhaust (rubber) tube (4) through nozzle holder assembly (5). 2. Insert three screws (1) through assembly (3) and nozzle holder assembly (5). Place correct shim(s) (nozzle holder/front cover) (7) and shim retaining sleeve (6) onto each screw (1). 3. Place exhaust (rubber) tube (4) onto port of nozzle holder assembly (5).
4. Check that seal holder/case (8) is located properly in the case front-end (2); if not, locate seal correctly.
3-117
TM 3-6665-343-23&P 3-50.
NOZZLE HOLDER ASSEMBLY REPLACEMENT (CONT).
INSTALLATION (CONT). 5. Mate assembly (3) to case front-end (2) as shown. 6. Tighten three screws (1) into case front-end (2); torque to 80 oz. in. 7. Ensure by visual inspection that seal (holder/case) (8) is seated and sealed correctly.
NOTE: CHASSIS ASSEMBLY REMOVED FOR CLARITY
3-118
TM 3-6665-343-23&P 3-51.
CASE FRONT-END REPLACEMENT.
This task covers: Removal and installation of Monitor Case Assembly. INITIAL SETUP Tools Electronic Equipment Tool Kit TK-105/G Torque Screwdriver Clean Work Surface (para 3-13) Materials/Parts Case Front-End Gloves (Item 4, App D) Equipment Conditions CAM turned off, DTS turned off, and all cables disconnected from CAM. Monitor case assembly removed (para 3-40). PCB removed from monitor module assembly (para 3-43). Nozzle assembly removed (para 3-52). REMOVAL 1. Loosen three screws (1) from case front-end (2) and pull assembly (3) from case front-end (2). 2. Remove nozzle holder/case seal (4) from case front-end (2). 3.
Remove case seal (5) from case front-end (2).
INSTALLATION 1. Place case seal (5) on case front-end (2). 2. Install nozzle holder/case seal (4) to case front-end (2) and mate assembly (3) to case front-end (2). 3. Tighten three screws (1) into case front-end (2); torque to 80 oz. in. 4. Ensure by visual inspection that seal (holder/case) (4) is seated and sealed correctly. 3-119
TM 3-6665-343-23&P 3-52.
NOZZLE ASSEMBLY REPLACEMENT.
This task covers: Removal and installation of Monitor Case Assembly. INITIAL SETUP Tools Electronic Equipment Tool Kit TK-105/G Nozzle Extraction Tool (from DTS) Clean Work Surface (para 3-13) Materials/Parts Nozzle Assembly Gloves (Item 4, App D) REMOVAL 1. Twist (counterclockwise) and remove nozzle protective cap assembly (1) from front of CAM (2).
2. Using nozzle extraction tool, loosen nozzle assembly. (a) Insert slots of nozzle extraction tool (3) into mating slots of locking ring (inside case front-end (4)). (b) Turn tool counterclockwise (about 1/4 turn) until it stops. (c) Turn tool clockwise very slightly and remove tool.
3-120
TM 3-6665-343-23&P 3. Remove nozzle assembly and associated parts: (a) Hold CAM by handle. (b) Grab nozzle (5) and gently work nozzle and associated parts out by pulling and moving nozzle slightly up and down. CAUTION Puncturing the membrane will damage the drift tube module. Do not get tweezers or any objects near the membrane. If the membrane appears damaged or ruptured, replace membrane assembly (para 3-53). (c) Ensure that seal (nozzle to drift tube) (6) comes out with nozzle. If seal does not come out, remove seal gently using a pair of tweezers. DO NOT GET TWEEZERS NEAR THE MEMBRANE.
4. Remove four parts from nozzle assembly (5): (a) Locking ring (7). (b) Nozzle thrust washer (8). (c) Seal (nozzle/holder) (9). (d) Seal (nozzle/drift tube) (6).
3-121
TM 3-6665-343-23&P 3-52.
NOZZLE ASSEMBLY REPLACEMENT (CONT).
REMOVAL (CONT). 5. Check the following on the CAM: (a) Inspect that two contacts (11) are in the nozzle holder assembly. Also inspect for presence of three locking tabs (12) on nozzle holder assembly. If contacts or tabs are not present, replace nozzle holder assembly (para 3-50). (b) Inspect membrane (13) for excessive dust. If parts of the grid cannot be seen under the membrane with the naked eye, the dust is excessive. If dust is excessive, replace membrane (para 3-53). (c) Inspect for presence of three locking tabs (14) on the case front-end (15); if not present, replace case front-end (para 3-51).
INSTALLATION 1.
Place parts onto nozzle (5). Ensure parts are positioned correctly as shown. Inspect each part for breaks or tears before assembly; replace part if required. (a) Seal (nozzle/drift tube) (6). (b) Seal (nozzle/holder) (9). (c) Nozzle thrust washer (8). (d) Locking ring (7).
3-122
TM 3-6665-343-23&P 2.
Place nozzle assembly with parts into nozzle removal tool. Locking tabs on nozzle removal tool fit into slots on locking ring.
3.
Install nozzle assembly with parts into CAM: (a) Ensure nozzle assembly is inserted straight into CAM. Gently rotate tool back and forth until nozzle assembly with parts slips into CAM; locking ring must be inside case front-end. (b) Rotate nozzle removal tool (4) clockwise about 1/4 turn until locking ring snaps into place.
4.
Twist (clockwise) and install nozzle protective cap assembly (1) onto front of CAM (2).
3-123
TM 3-6665-343-23&P 3-53.
MEMBRANE ASSEMBLY REPLACEMENT.
This task covers: Removal and installation of Monitor Case Assembly. INITIAL SETUP Tools Electronic Equipment Tool Kit TK-1 05/G Membrane Extraction Tool (DTS Lid) Clean Work Surface (para 3-13) Materials/Parts Membrane Assembly Gloves (Item 4, App D) Equipment Condition Nozzle assembly removed (para 3-52).
WARNING
RADIATION HAZARD
NICKEL-63 (Ni-63) This CAM contains a beta radiation source in the quantity which requires a Nuclear Regulatory Commission license to possess items which contain the source. The CAM contains 10 millicuries of Nickel-63. WARNING Removal of the membrane assembly "opens" the pneumatically "sealed" drift tube module, therefore indirectly exposing the Ni-63 ionizing source. To avoid exposure, minimize the time spent performing this task. For further information on the Nuclear Regulatory Commission License, see Appendix E. CAUTION Removal of the membrane assembly exposes CAM's highly sensitive sensing device to outside contamination. Ensure that gloves are worn and appropriate precautions are taken to avoid equipment contamination. Avoid touching or getting gloves near membrane. A special membrane extraction tool is provided in the DTS lid. DO NOT ATTEMPT TO USE ANY OTHER TOOL. The membrane extraction tool has a built-in "stop" to avoid damaging the membrane; do not attempt to adjust or modify this tool. 3-124
TM 3-6665-343-23&P CAUTION Do not rotate membrane during removal as it may damage the heater grid. REMOVAL 1.
Slowly lower "opened" tool into the nozzle holder assembly and carefully engage its pins with the membrane holder's slots. Be very careful not to rotate the membrane during this action, as it may damage the heater grid lying directly behind it.
2.
Very gently pull the membrane out of its seat using only a straight ahead motion-DO NOT ROTATE. It may be necessary to gently rock the membrane extraction tool from side to side while gently pulling forward.
3.
Dispose of membrane as radioactive waste.
INSTALLATION 1.
Grasp new membrane from packaging with membrane tool.
2.
Press membrane firmly into place DO NOT ROTATE. Remove the tool. NOTE If it is suspected that the heater grid has become damaged, perform the membrane heater (MH) resistance part of the cell test (para 3-58) to confirm correct resistance.
3-125
TM 3-6665-343-23&P 3-54.
ENVIRONMENTAL CAP ASSEMBLY REPLACEMENT.
This task covers: Removal and installation of Monitor Case Assembly. INITIAL SETUP Materials/Parts Environmental Cap Assembly REMOVAL 1. Twist (counterclockwise) and remove environmental cap assembly (1) from CAM (2). 2. Separate environmental cap assembly (1) and retaining strap from CAM. INSTALLATION 1. Attach environmental cap assembly to CAM using retaining strap (pass cap through loop in strap). 2. Twist (clockwise) and install environmental cap onto CAM (2).
3-126
TM 3-6665-343-23&P 3-55.
BATTERY CAP ASSEMBLY REPLACEMENT.
This task covers: Removal and installation of Monitor Case Assembly. INITIAL SETUP Materials/Parts Battery Cap Assembly REMOVAL Twist (counterclockwise) and remove battery cap assembly (1) from CAM. INSTALLATION Twist (clockwise) and install new battery cap assembly (1) onto CAM.
3-127
TM 3-6665-343-23&P 3-56.
CARRYING CASE ASSEMBLY REPLACEMENT.
This task covers: Replacement of Confidence Sample. INITIAL SETUP Materials/Parts Carrying Case Assembly REMOVAL Remove all items from old carrying case assembly (1). INSTALLATION Place all items into new carrying case assembly (1).
3-57.
CARRYING CASE STRAP REPLACEMENT.
This task covers: Removal and installation of Monitor Case Assembly. INITIAL SETUP Materials/Parts Carrying Case Strap REMOVAL Remove old carrying case strap (2) from carrying case assembly (1). INSTALLATION
3-128
TM 3-6665-343-23&P Install new carrying case strap (2) onto new carrying case assembly (1). 3-58. a.
CELL TEST. Set DTS SUPPLY switch to off. Remove CAM power lead W1 and ground wire from connector assembly and W2 from display connector P2. WARNING To avoid shock, DO NOT apply power to the CAM from the DTS for this test.
b.
Disconnect flexi connector from P4 of PCB assembly.
c.
Connect cell lead W4 from DTS to flexi connector of drift tube module, as shown, with the white dot on connector W4 on the underside (not visible from above).
d.
Use a small flat-tip screwdriver to remove the coaxial plug and socket from the clip on flexiboard (see illustration). Separate coaxial plug and socket. Press board side coaxial socket back into clip (see illustration). Attach coaxial socket from cell lead to coaxial plug of CAM, as shown.
3-129
TM 3-6665-343-23&P 3-58.
CELL TEST (CONT). e. Set DMM for resistance measurements and connect test lead plugs to DTS DMM sockets (1).
f.
3-130
Select and measure each of the cell resistances:
SET DMM OHMS
ON DTS PRESS
DMM ACCEPTABLE READOUT
200
MH RESISTANCE
7.3 TO 11.4
2000K
SG TO CASE RESISTANCE
900 TO 1100
20M
FG/MG INSULATION
GREATER THAN 10
20M
SG/CE INSULATION
GREATER THAN 10
20M
MH TO CASE
GREATER THAN 10
TM 3-6665-343-23&P g. Connect multimeter test leads to megger sockets on DTS as described below. SET DMM
CONNECT DMM LEADS TO
DMM ACCEPTANCE READOUT IN M OHMS
20M
SS and MH
GREATER THAN 20
20M
FG and CASE
GREATER THAN 20
20M
SS and CASE
GREATER THAN 20
h.
Remove cell lead coaxial socket from coaxial plug of CAM. Use a small flat-tip screwdriver to remove the board side coaxial socket from the clip on the CAM. Join CAM coaxial plug and socket. Press joined coaxial plug and socket back into clip.
i.
Remove cell lead W4 of DTS from flexi connector of drift tube module.
j.
Connect flexi connector to P4 of PCB.
k.
If cell does not pass resistance test, refer to para 3-30.
3-131
TM 3-6665-343-23&P 3-59.
LEAK TEST. a. Place monitor module assembly into chassis assembly (from DTS). b. On DTS, set SUPPLY switch to off. SUPPLY indicator green lamp goes off. c.
Connect CAM power lead W1 to connector assembly on monitor module assembly.
d. Connect ground wire clip of CAM power lead to ground stud (located below connector) on chassis assembly. e. Connect display connector W2 to connector P2 on monitor module assembly. f.
Set SUPPLY switch to on: SUPPLY indicator green lamp comes on. If CAM comes on, press DTS TEST DISPLAY ON switch to turn CAM off.
g. On DTS, set valve V3 to CELL LEAK TEST. Ensure VENT is closed.
NOTE Do not run this test until the monitor module assembly is at ambient temperature. Temperature changes of the monitor module assembly could cause invalid test results. h. Push and turn Luer fitting on CELL flexible hose into breather outlet on sieve pack assembly. i.
3-132
Squeeze bulb of LEAK TEST flowmeter vigorously until bubbles reach top of flow tube.
TM 3-6665-343-23&P
CAUTION If bubbles go down the flow tube in the next step (instead of up) at a rate of 50 mm (5 cm) per minute or more, stop the test immediately by pressing the DTS TEST DISPLAY ON switch to turn the CAM off. Bubbles should not flow backward into the DTS; they could contaminate the DTS. j.
Press TEST DISPLAY ON switch to turn module assembly on.
k.
Squeeze bulb of LEAK TEST flowmeter gently to form bubbles in flow tube.
I.
During first 30-40 seconds of CAM operation, ensure that bubbles rise rapidly in the LEAK TEST flowmeter; bubble movement then stabilizes.
m. Open VENT and allow bubbles to move to a convenient measurement line. Close VENT.
3-133
TM 3-6665-343-23&P 3-59.
LEAK TEST (CONT). (1) Bubble must not move up or down more than 10 mm in 1 minute. (2) If the leak rate is greater than 10 mm per minute, let CAM run for 5 minutes and repeat test.
CAUTION Open vent and remove Luer fitting from sieve pack breather outlet before removing electrical power from the monitor module assembly. Failure to remove the fitting first could cause a bubble to be sucked into the DTS. n. Open VENT. o. Twist and pull Luer fitting from drift tube module breather outlet. p. Press TEST DISPLAY ON switch to turn monitor module assembly to off. q. Set DTS SUPPLY switch to off. Remove display connector W2 from P2. Remove CAM power lead W1 and ground wire from CAM.
3-134
TM 3-6665-343-23&P 3-60.
PUMP TEST.
a. Remove pump assembly (para 3-47). Inspect pump seals for apparent distortion or damage; replace seals if distorted or damaged. b. On DTS, set PUMP TEST ON switch to OFF. If not already on, set DTS SUPPLY switch to ON. c.
Install and connect pump assembly to DTS mount: (1) Remove two seals from bottom of pump assembly. (2) Remove two wing nuts and top plate from DTS. (3) Observe underneath top plate. Set top plate over pump assembly so that top plate seal indentations mate with pump assembly seals. (4) Observe DTS fixture and fit pump assembly and top plate of DTS pump mounting assembly so that label "front" is on front of top plate. (5) Install two wing nuts and clamp pump into DTS pump mounting assembly. (6) Connect pump assembly connector to DTS connector.
d. Set DTS for pump flow test: (1) Flow valve V1 to PUMP RECIRC FLOW. (2) Flow valve V2 to PUMP INLET FLOW. (3) Flow valve V3 to CELL LEAK TEST. e. Turn restrictor knob (located at bottom of flow 1 meter) fully clockwise, then back off two turns. f.
On DTS, set PUMP TEST ON switch to ON.
g. Adjust restrictor knob on DTS FLOW 1 meter to obtain a setting of -58 mbars on the VACUUM gauge. The flowmeters should indicate ml/min as follows (read across top of flowmeter float): • •
FLOW 1 - Greater than 285 FLOW 2 - Greater than 395
3-135
TM 3-6665-343-23&P 3-60.
PUMP TEST (CONT). h. On the DTS, press PUMP AMPS switch. Meter should indicate less than 8.9 units. i.
On DTS, set PUMP TEST ON switch to OFF to turn pump off. NOTE Do not proceed until the pump assembly is at room temperature. Temperature changes of the pump assembly could cause invalid test results.
j.
Squeeze bulb of LEAK TEST flowmeter vigorously until bubbles reach top of flow tube.
k.
On DTS, set valve V3 to PUMP LEAK TEST.
I.
Set valve V1 to PUMP LEAK.
m. Run the pump leak test by performing the next steps sequentially and quickly: (1) Set PUMP TEST ON switch to ON. (2) Squeeze bulb of LEAK TEST flowmeter to get some bubbles into the flow tube. (Position of bubble can be changed by opening and closing vent.) n. Run test for 45 seconds. Bubble should stabilize within 30 seconds. o. Set VENT to open position, set PUMP TEST ON switch to OFF. p. Remove pump assembly from DTS pump mounting assembly. q. Install test block on DTS.
3-136
TM 3-6665-343-23&P 3-61.
WAVEFORM TEST. NOTE The CAM should be exposed to the test area ambient environment for 2 hours prior to the beginning of this test. Ensure DTS is prepared for use. a. If CAM is in case: (1) Make sure cable W1 is connected to connector assembly. (2) If DTS is not on, turn it on. (3) If CAM is not operating, press CAM ON/OFF push-button switch to turn CAM on. b. If CAM is in chassis assembly: (1) Make sure cable W1 is connected to connector assembly and ground wire is connected to ground stud on chassis assembly. (2) Connect W2 cable. (3) If DTS is not on, turn it on. (4) Press DTS TEST DISPLAY ON switch to turn CAM on. c.
Make sure CAM has been running for at least 10 minutes.
GATE WIDTH AND HEIGHT.
NOTE All oscilloscope configurations cannot be covered here. Instructions given are accurate for most oscilloscopes. Maintainer may have to make adjustments peculiar to some equipment. d. Connect oscilloscope signal cables to oscilloscope jacks on DTS. e. Set up oscilloscope as shown below. Do not measure waveform until 10 minute period is over. f. Measure gate width and height. Move beginning pulse height to top graticule line. Width should be 180 ± 20 µs rectangular pulse with a height of 5 ± 0.6 V as shown. If gate width or height is not within specification, refer to troubleshooting symptom 10, (para 3-25).
3-137
TM 3-6665-343-23&P 3-61.
WAVEFORM TEST (CONT). SET TO
CONTROL VERTICAL MODE CH1 BOTH CH2 BW LIMIT INVERT CH2 VOLTS/DIV CH2 AC GND DC HORIZONTAL MODE A ALT B A AND B SEC/DIV A TRIGGER NORM SLOPE A TRIG BW A SOURCE A EXT COUPLING
CH2 In In 1 DC A 20µ s In Out FULL Int DC
Connect IMS cable from DTS to CH1 input Connect GATE cable from DTS to CH2 input PULSE WIDTH. g. Set up oscilloscope as shown below. h. Measure pulse width. Width should be 30±10 ms. If pulse width is not within specification, refer to troubleshooting symptom 10 (para 3-25).
3-138
CONTROL
SET TO
HORIZONTAL MODE A AND B SEC/DIV
5 ms
TM 3-6665-343-23&P NOISE. i.
Set up oscilloscope as shown below.
j. Examine G and H mode outputs for noise. Noise level should not exceed 300 mV peak-to-peak. If noise level in waveform test is excessive, refer to troubleshooting symptom 11 (para 3-26).
CONTROL VERTICAL MODE CH1 BOTH CH2 ADD ALT CHOP CH1 VOLTS/DIV CH1 AC GND DC CH2 AC GND DC HORIZONTAL MODE A AND B SEC/DIV A TRIGGER NORM
SET TO CH1 CHOP 0.5 AC AC 2 ms (noise) 1 ms (peak position and width) In
3-139
TM 3-6665-343-23&P 3-61.
WAVEFORM TEST (CONT).
SECONDARY PEAKS.
NOTE Secondary peaks are any visible peak other than the normal RIP. They may look much like the peaks that show when a CAM is exposed to simulant, but do not diminish and disappear within a few minutes. A secondary peak due to contamination can occur anywhere on the waveform, depending on the source of the contamination. However, do not consider a peak that occurs within 1 ms from the center of the RIP to be a secondary peak. A corresponding bar reading on the CAM may or may not be present. To become familiar with the types of peaks contamination causes, observe the response of a good CAM to simulant. k. Set CH1 VOLTS/DIV to 1. Set A and B SEC/DIV to 2 ms. Observe for secondary peaks occurring after the main peak in both the G and H modes. There should not be any secondary peaks. Secondary peaks indicate the presence of contamination (see below). If secondary peaks are present in waveform test, remove nozzle protective cap. If secondary peaks significantly decrease in amplitude, clean nozzle assembly area with a dampened cheesecloth and install a new protective cap. If secondary peaks are still present in waveform test, refer to troubleshooting symptom 12 (para 3-27). I. Set CH1 to DC and measure dc offset from zero in both the G and H modes: offset should be less than 1 V dc. If DC offset is greater than 1 V dc, refer to troubleshooting symptom 13, para 3-28. NOTE Wait a few seconds after switching modes to make peak measurements. m. Set CH1 to AC. Set A and B SEC/DIV on oscilloscope to 1 ms. Make the following reactant ion peak (RIP) measurements: (1) H mode peak height: greater than 1.5 V peak. (2) G mode peak height: greater than 1.5 V peak. n. If (G or H) peak heights are not within specifications, refer to troubleshooting symptom 14 (para 3-29).
3-140
TM 3-6665-343-23&P RIP RATIO. Measure time of G mode RIP position. Measure time of H mode RIP position. Using the equation below calculate ratio of G mode RIP timing to H mode RIP timing. Ratio should be 1.17 ± 0.09. If RIP ratio is not within specifications, refer to troubleshooting symptom 14 (para 3-29).
RATION =
G RIP POSITION TIMING H RIP POSITION TIMING
3-141
TM 3-6665-343-23&P Section V. PREPARATION FOR STORAGE OR SHIPMENT 3-62.
STORAGE. a. Store CAM in a secure area when unattended. b. Post the area with CAUTION - RADIOACTIVE MATERIAL signs as required. c.
The signs shall be yellow with magenta letters and radiation symbol at least 8 inches by 10 inches. Post signs on outside of lockable containers.
d. Storage of a single CAM must be in a secure area; the area does not need to be marked. 3-63.
SHIPMENT. Refer to AR 385-11 for receiving, unloading, and movement of radioactive materials. a. Packing. (1) Pack the CAM in the original shipping pack and over pack with a cardboard box or wooden box. CAUTION Cushioning and overpacking should be chemically clean so that no outgassing occurs. If possible include desiccant and double wrap the CAM with vapor barrier and polyethylene bag. (2) If the original shipping pack is not available, use available cushioning and over pack with a cardboard or wooden box. (3) The Government Bill of Lading or other shipping document should be completed in accordance with 49 Code of Federal Regulations, and Army and local regulations. b. Instructions for shipment of any CAM marked XXX will be provided by the point of contact below: Armament and Chemical Acquisition and Logistics Activity ATTN: AMSTA-AC-CTC-D Rock Island, IL 61299-7630 DSN 793-5961 COM (309) 782-5961
3-64.
ACCOUNTABILITY.
Accountability will be maintained at the Property Book level. For instructions on completing transactions contact your Cell Serialization Surety Office (CSSO) in accordance with AR 710-3, Chapter 4, Section 2 for radiation tracking requirements. A transaction is required for each of the following: a. b. c. d. e. f. g.
3-142
Inventory loss Suspected loss or theft Receipt Shipment Demilitarization Wipe test result Removal of drift tube module from CAM.
TM 3-6665-343-23&P APPENDIX A REFERENCES A-1.
SCOPE. This appendix lists service publications referenced in this manual.
A-2.
A-3.
ARMY REGULATIONS. AR 55-38 ...................................................................
Reporting of Transportation Discrepancies in Shipments
AR 385-11 .................................................................
Ionizing Radiation Protection (Licensing, Control, Transportation, disposal, and Radiation Safety)
AR 385-40 .................................................................
Accident Reporting and Records
AR 700-64 .................................................................
Radioactive Commodities in the DoD Supply Systems
AR 710-3 ...................................................................
Asset and Transaction Reporting System
AR 735-11-2 ..............................................................
Reporting of Item and Packaging Discrepancies
COMMON TABLE OF ALLOWANCES. CTA 50-970 ............................................................... CTA 8-100 .................................................................
A-4.
DA PAMPHLETS. DA PAM 738-750.......................................................
A-5.
Functional Users Manual for The Army Maintenance Management System (TAMMS)
FIELD MANUALS. FM 3-5....................................................................... FM 21-11...................................................................
A-6.
Expendable and Durable Items (Except: Medical Class V, Repair Parts and Heraldic Items) Army Medical Department Expendable and Durable Items
NBC Decontamination First Aid for Soldiers
FORMS. DA Form 2028-2 ........................................................ RCS DD-SD(AR) 1168............................................... SF 361 ...................................................................... SF 364 ...................................................................... SF 368 ......................................................................
A-1
Recommended Changes to Equipment Technical Publications Radiological Accident Report Transportation Discrepancy Report Report of Discrepancy Product Quality Deficiency Report
TM 3-6665-343-23&P A-7.
SOLDIER TRAINING PUBLICATIONS. STP 21-1-5MCT ........................................................
A-8.
Soldier's Manual of Common Tasks, Skill Level 1
TECHNICAL MANUALS. TM 3-6665-343-10 .....................................................
Operator's Manual for Improved Chemical Agent Monitor (CAM)
TM 3-6665-332-13&P ............................................... . Operators and Direct Support Maintenance Manual (Including RPSTL) for Diagnostic Test Set (DTS). TM 43-0002-31 .......................................................... TM 750-244-2 ............................................................ A-9.
A-2
Destruction of Chemical Weapons and Defense Equipment to Prevent Enemy Use Procedure for Destruction of Electronics Materiel to Prevent Enemy Use (Electronics Command)
TECHNICAL BULLETINS. TB 43-0130................................................................
Instructions for the Safe Handling and Identification of US Army Communications-Electronics Command Managed Lithium-Sulfur Dioxide Batteries
TB 700-4 ...................................................................
Decontamination of Facilities and Equipment
TM 3-6665-343-23&P APPENDIX B MAINTENANCE ALLOCATION CHART (MAC)
Section I. INTRODUCTION B-1.
THE ARMY MAINTENANCE SYSTEM MAC. a. This introduction (Section I) provides a general explanation of all maintenance and repair functions authorized at various maintenance levels under the standard Army Maintenance System concept. b. The Maintenance Allocation Chart (MAC) in Section II designates overall authority and responsibility for the performance of maintenance functions on the identified end item or component. The application of the maintenance functions to the end item or component will be consistent with the capacities and capabilities of the designated maintenance levels which are shown on the MAC in column (4) as: Unit - includes two subcolumns, C (operator/crew) and O (unit) maintenance. Direct Support - includes an F subcolumn. General Support - includes an H subcolumn. Depot - includes a D subcolumn. c. Section III lists the tools and test equipment (both special tools and common tool sets) required for each maintenance function as referenced from Section II. d. Section IV contains supplemental instructions and explanatory notes for a particular maintenance function. B-2. MAINTENANCE FUNCTIONS. Maintenance functions will be limited to and defined as follows: a. Inspect. To determine the serviceability of an item by comparing its physical, mechanical, and/or electrical characteristics with established standards through examination (e.g., by sight, sound, or feel). b. Test. To verify serviceability by measuring the mechanical, pneumatic, hydraulic, or electrical characteristics of an item and comparing those characteristics with prescribed standards. c. Service. Operations required periodically to keep an item in proper operating condition; i.e., to clean (includes decontaminate, when required), to preserve, to drain, to paint, or to replenish fuel, lubricants, chemical fluids or gases. d. Adjust. To maintain or regulate, within prescribed limits, by bringing into proper or exact position, or by setting the operating characteristics to specified parameters. e. Align. To adjust specified variable elements of an item to bring about optimum or desired performance. f. Calibrate. To determine and cause corrections to be made or to be adjusted on instruments or test, measuring, and diagnostic equipments used in precision measurement. Consists of comparisons of two instruments, one of which is a certified standard of known accuracy, to detect and adjust any discrepancy in the accuracy of the instrument being compared.
B-1
TM 3-6665-343-23&P B-2. MAINTENANCE FUNCTIONS (CONT). g. Remove/Install. To remove and install the same item when required to perform service or other maintenance functions. Install may be the act of emplacing, seating, or fixing into position a spare, repair part, or module (component or assembly) in a manner to allow the proper functioning of an equipment or system. h. Replace. To remove an unserviceable item and install a serviceable counterpart in its place. "Replace" is authorized by the MAC and assigned maintenance level is shown as the 3rd position code of the SMR code. i. Repair. The application of maintenance services including fault location/troubleshooting, removal/installation, and disassembly/assembly procedures, and maintenance actions to identify troubles and restore serviceability to an item by correcting specific damage, fault, malfunction, or failure in a part, subassembly, module (component or assembly), end item, or system. j. Overhaul. That maintenance effort (service/action) prescribed to restore an item to a completely serviceable/operational condition as required by maintenance standards in appropriate technical publications (i.e. DMWR). Overhaul is normally the highest degree of maintenance performed by the Army. Overhaul does not normally return an item to like new condition. k. Rebuild. Consists of those services/actions necessary for the restoration of unserviceable equipment to a like new condition in accordance with original manufacturing standards. Rebuild is the highest degree of materiel maintenance applied to Army equipment. The rebuild operation includes the act of returning to zero those age measurements (hours, miles, etc.) considered in classifying Army equipment/components. B-3. EXPLANATION OF COLUMNS IN THE MAC, SECTION II. a. Column (1), Group Number. Column (1) lists functional group code numbers, the purpose of which is to identify maintenance significant components, assemblies, subassemblies, and modules with the next higher assembly. b. Column (2), Component/Assembly. Column (2) contains the item names of components, assemblies, subassemblies, and modules for which maintenance is authorized. c. Column (3), Maintenance Function. Column (3) lists the functions to be performed on the item listed in column (2). d. Column (4), Maintenance Level. Column (4) specifies, by the listing of a work time figure in the appropriate subcolumn(s), the level of maintenance authorized to perform the function listed in column (3). This figure represents the active time required to perform that maintenance function at the indicated level of maintenance. If the number or complexity of the tasks within the listed maintenance function vary at different levels, appropriate work time figures will be shown for each level. The work time figure represents the average time required to restore an item (assembly, subassembly, component, module, end item, or system) to a serviceable condition under typical field operating conditions. This time includes preparation time (including any necessary disassembly/assembly time), troubleshooting/fault location time, and quality assurance/quality control time in addition to the time required to perform the specific tasks identified for the maintenance functions authorized in the Maintenance Allocation Chart. The symbol designations for the various maintenance levels are as follows: C ---------------------O ---------------------F----------------------H ---------------------L----------------------D ----------------------
B-2
Operator or Crew Unit Maintenance Direct Support Maintenance General Support Maintenance Specialized Repair Activity (SRA) Depot Maintenance
TM 3-6665-343-23&P e. Column (5), Tools and Equipment. Column (5) specifies, by code, those common tool sets (not individual tools), and special tools, TMDE, and support equipment required to perform the designated function. f. Column (6), Remarks. This column shall, when applicable, contain a letter code, in alphabetic order, which shall be keyed to the remarks contained in Section IV. B-4.
EXPLANATION OF COLUMNS IN TOOL AND TEST EQUIPMENT REQUIREMENTS, SECTION III.
a. Column (1), Reference Code. The tool and test equipment reference code correlates with the code used in the MAC, Section II, Column 5. b. Column (2), Maintenance Level. The lowest level of maintenance authorized to use the tool or test equipment. c. Column (3), Nomenclature. Name or identification of the tool or test equipment. d. Column (4), National Stock Number. The National stock number of the tool or test equipment. e. Column (5), Tool Number. The manufacturer's part number.
B-5.
EXPLANATION OF COLUMNS IN REMARKS, SECTION IV. a. Column (1), Reference Code. The code recorded in column 6, Section II.
b. Column (2), Remarks. This column lists information pertinent to the maintenance function being performed as indicated in the MAC, Section II.
B-3
TM 3-6665-343-23&P Section II. MAINTENANCE ALLOCATION CHART FOR FOR IMPROVED CHEMICAL AGENT MONITOR (ICAM) (1)
(2)
(3)
GROUP NUMBER
COMPONENT/ ASSEMBLY
MAINTENANCE FUNCTION
00
01
0101
UNIT C O
Monitor, Chemical Agent (ICAM)
Inspect Repair
0.1 0.1
Monitor Kit, CAM
Inspect Repair
0.1 0.1
Monitor, Chemical Agent
Inspect Test Service Repair
0.1 0.1 0.5 0.1
(4) MAINTENANCE LEVEL DIRECT GENERAL SUPPORT SUPPORT F H
(5) TOOLS AND EQUIPMENT REF CODE
DEPOT D
(6)
REMARKS CODE
0.1 0.2
A
0.2
A
0.5
1,3,4,5,6
0.2
9,10,11
010101
Module Assembly, Monitor
Test Repair
0.5 1.2
1,3,4,5,6 1,7,8
01010101
Monitor Assembly
Repair
0.8
1,3,4,5,6,8
0101010101
Module, Drift Tube
Test Replace Repair
0.3 0.1 0.2
6 1,8 1,4,5,6
0101010102
Connector Receptacle
Replace Repair
0.2 0.1
8 1
01010102
Printed Circuit Board
Test Adjust Replace
1.2 1.4
1,3,4,5,6 1,8
010102
Case Assembly, Monitor
Replace Repair
0.2 0.5
1,9,10,11 1,12
0102
Carrying Case Assembly
Replace Repair
0.1 0.1
B
C
B-4
1.0
TM 3-6665-343-23&P Section III.
TOOL AND TEST EQUIPMENT REQUIREMENTS
(1) TOOL OR TEST EQUIPMENT REFERENCE CODE
(2)
(3)
(4)
(5)
MAINTENANCE LEVEL
NOMENCLATURE
NATIONAL/NATO STOCK NUMBER
TOOL NUMBER
1
F
Tool Kit, Electronic Equipment
5180-00-610-8177
TK-105/G
2
F
RADIAC Set
6665-01-222-1425
AN/VDR-2 or equivalent
3
F
Test Lead Set
6665-01-121-0510
861
4
F
Oscilloscope, dual trace
6625-01-187-7847
AN/USM-488 or equivalent
6625-01-265-6000 5
F
Multimeter, digital
AN/PSM-45A or equivalent
6
F
Diagnostic Test Set
6665-99-000-3023
464-003
7
F
Power Supply
6130-01-139-2514
PP-2309C/U
8
F
Torque Wrench
5120-00-943-0941
9
F
Torque Wrench
5120-01-396-6071
10
F
Adapter
5120-00-240-8703
11
F
Socket
5120-00-293-0093
12
F
Starter, Screw, Hand
5120-00-965-0626
Section IV. (1) REFERENCE CODE
REMARKS (2) REMARKS
A
Operator repair consists of replacing specified items using spares stored in the carrying case. Direct Support repair consists of replacing subassemblies.
B
Operator test consists of self-test and confidence test. Direct Support test consists of wipe test for radioactive contamination and diagnostic tests to identify faulty components.
C
Depot Maintenance consists of test only.
B-5/(B-6 blank)
TM 3-6665-343-23&P APPENDIX C UNIT AND DIRECT SUPPORT MAINTENANCE REPAIR PARTS AND SPECIAL TOOLS LIST Current as of 9 June 1998 SECTION I. INTRODUCTION 1. SCOPE. This RPSTL lists and authorizes spares and repair parts; special tools; special test, measurement, and diagnostic equipment (TMDE), and other special support equipment required for performance of Unit and Direct Support maintenance of the Improved Chemical Agent Monitor (ICAM). It authorizes the requisitioning, issue, and disposition of spares, repair parts and special tools as indicated by the source, maintenance and recoverability (SMR) codes. 2. GENERAL. This repair parts and special tools list is divided into the following sections: a. Section II. Repair Parts List. A list of spares and repair parts authorized by this RPSTL for use in the performance of maintenance. The list also includes parts which must be removed for replacement of the authorized parts. Parts lists are composed of functional groups in ascending alphanumeric sequence, with the parts in each group listed in ascending figure and item number sequence. Bulk materials are listed by item name in FIG. BULK at the end of the section. Repair parts kits are listed separately in their own functional group within Section II. Repair parts for reparable special tools are also listed in Section II. Items listed are shown on the associated illustration. b. Section III. Special Tools List. A list for special tools, special TMDE, and other special support equipment authorized by this RPSTL (as indicated by Basis of Issue (BOI) information in DESCRIPTION AND USABLE ON CODE (UOC) column) for the performance of maintenance. c. Section IV. Cross-reference Index. A list, in National Item Identification Number (NIIN) sequence, of all National stock numbers (NSN) appearing in the listings, followed by a separate list in alphanumeric sequence of all part numbers appearing in the listings. National stock numbers and part numbers are cross-referenced to each illustration figure and item number appearance. The figure and item number index lists figure and item numbers in alphanumeric sequence and cross-references NSN, CAGEC and part numbers. 3. EXPLANATION OF COLUMNS (SECTIONS II AND III). a. ITEM NO. (Column (1)). Indicates the number used to identify items called out in the illustration.
C-1
TM 3-6665-343-23&P 3. EXPLANATION OF COLUMNS (SECTIONS II AND III) (CONT). b. SMR CODE (Column (2)). The Source, Maintenance, and Recoverability (SMR) code is a 5-position code containing supply/requisitioning information, maintenance level authorization criteria, and disposition instruction, as shown in the following breakout: Source Code XX
Maintenance Code
Recoverability Code
XX
X
1st two positions
How to get an item
3rd position
4th position
Who can install replace, or use the item
Who can do complete repair* on the item
Who determined disposition action on an unserviceable item
*Complete Repair: Maintenance capacity, capability and authority to perform all corrective maintenance tasks of the "Repair" function in a use/user environment in order to restore serviceability to a failed item. (1) Source Code. The source code tells you how to get an item needed for maintenance, repair, or overhaul of an end item/equipment. Code PA PB PC** PD PE PF PG
Explanation
Stocked items; use the applicable NSN to request/requisition items with these source codes. They are authorized to the level indicated by the code entered in the 3rd position of the SMR code.
** NOTE: Items coded PC are subject to deterioration. KD KF KB
MO-(Made at Unit AVUM Level MF-(Made at DS/AVUM Level) MH-(Made at GS Level) ML-(Made at Specalized Repair Act (SRA)) MD-(Made at Depot)
C-2
Items with these codes are not to be requested/requisitioned individually. They are part of a kit which is authorized to the maintenance level indicated in the 3rd position of the SMR code. The complete kit must be requisitioned and applied.
Items with these codes are not to be requested/requisitioned individually. They must be made by bulk material which is identified by the part number in the DESCRIPTION AND USABLE ON CODE (UOC) column and listed in the Bulk Material group of the repair parts and special tools list in the RPSTL. If the item is authorized to you by the 3rd position code of the SMR code, but the source code indicates it is made at a higher level, order the item from the higher level of maintenance.
TM 3-6665-343-23&P AO-(Assembled at Unit AVUM Level AF-(Assembled at DS/AVUM Level) AH-( Assembled at GS Level) AL-(Assembled at Specialized Repair Act (SRA)) AD-(Assembled at Depot)
Items with these codes are not to be requested/requisitioned individually. The parts that make up the assembled item must be requisitioned or fabricated and assembled at the level of maintenance indicated by the source code. If the 3rd position code of the SMR code authorizes you to replace the item, but the source code indicates it is assembled at a higher level, order the item from the higher level of maintenance.
XA - Do not requisition an "XA" coded item. Order its next higher assembly. (Also, refer to the NOTE below.) XB - If an "XB" coded item is not available from salvage, order it using the CAGEC and part number given. XC - Installation drawing, diagram, instruction sheet, field service drawing, that is identified by manufacturer's part number. XD - Item is not stocked. Order an "XD" coded item through normal supply channels using the CAGEC and part number given, if no NSN is available. NOTE: Cannibalization or controlled exchange, when authorized, may be used as a source of supply for items with the above source codes, except for those sources coded "XA" or those aircraft support items restricted by requirements of AR 750-1. (2) Maintenance Code. Maintenance codes tell you the level(s) of maintenance authorized to USE and REPAIR support items. The maintenance codes are entered in the third and fourth positions of the SMR Code as follows: (a) The maintenance code entered in the third position tells you the lowest maintenance level authorized to remove, replace, and use an item. The maintenance code entered in the third position will indicate authorization to one of the following levels of maintenance. Code
Application/Explanation
C
-Crew or operator maintenance done within Unit or Aviation Unit maintenance.
O
-Unit or Aviation Unit level can remove, replace, and use the item.
F
-Direct support or Aviation Intermediate level can remove replace, and use the item.
H
-General support level can remove, replace, and use the item.
L
-Specialized repair activity can remove, replace, and use the item.
D
-Depot level can remove, replace, and use the item.
(b) The maintenance code entered in the fourth position tells whether or not the item is to be repaired and identifies the lowest maintenance level with the capability to do complete repair (i.e., perform all authorized repair functions). (NOTE: Some limited repair may be done on the item at a lower level of maintenance, if authorized by the Maintenance Allocation Chart (MAC) and SMR codes.) This position will contain one of the following maintenance codes.
C-3
TM 3-6665-343-23&P 3. EXPLANATION OF COLUMNS (SECTIONS II AND III) (CONT). Code
Application/Explanation
O
-Unit or Aviation Unit is the lowest level that can do complete repair of the item.
F
-Direct support or Aviation Intermediate is the lowest level that can do complete repair of the item.
H
-General support is the lowest level that can do complete repair of the item.
L
-Specialized repair activity (designate the specialized repair activity) is the lowest level that can do complete repair of the item.
D
-Depot is the lowest level that can do complete repair of the item.
Z
-Nonrepairable. No repair is authorized.
B
-No repair is authorized. (No parts or special tools are authorized for the maintenance of a "B" coded item.) However, the item may be reconditioned by adjusting, lubricating, etc., at the user level.
(3) Recoverability Code. Recoverability codes are assigned to items to indicate the disposition action on unserviceable items. The recoverability code is entered in the fifth position of the SMR Code as follows: Recoverability Codes
Application/Explanation
Z
-Nonrepairable item. When unserviceable, condemn and dispose of the item at the level of maintenance shown in 3rd position of SMR Code.
O
-Reparable item. When uneconomically reparable, condemn and dispose of the item at Unit or Aviation Unit level.
F
-Reparable item. When uneconomically reparable, condemn and dispose of the item at the Direct support or Aviation Intermediate level.
H
-Reparable item. When uneconomically reparable, condemn and dispose of the item at the General support level
D
-Reparable item. When beyond lower level repair capability, return to Depot. Condemnation and disposal of item not authorized below Depot level.
L
-Reparable Item. Condemnation and disposal not authorized below Specialized Repair Activity (SRA).
A
-Item requires special handling or condemnation procedures because of specific reasons (e.g., precious metal content, high dollar value, critical material, or hazardous material). Refer to appropriate manuals/directories for specific instructions.
c. CAGEC (column (3)). The Commercial and (Govemment Entity Code (CAGEC) is a 5-digit code which is used to identify the manufacturer, distributor, or Government agency, etc., that supplies the item. d. PART NUMBER (Column (4)). Indicates the primary number used by the manufacturer (individual, company, firm, corporation, or Government activity), which controls the design and characteristics of the item by means of its engineering drawings, specifications standards, and inspection requirements to identify an item or range of items. C-4
TM 3-6665-343-23&P NOTE: When using an NSN to requisition an item, the item may have a different part number from the part number listed. e. Description and Usable On Code (UOC) (Column (5)). This column includes the following information: (1) The Federal item name and, when required, a minimum description to identify the item. (2) Part numbers for bulk materials are referenced in this column in the line item entry for the item to be manufactured/fabricated. (3) The statement "END OF FIGURE" appears just below the last item description in Column 5 for a given figure in both Section II and Section III. f. QTY (Column (6)). The QTY (quantity per figure column) indicates the quantity of the item used in the breakout shown on the illustration figure, which is prepared for a functional group, subfunction group, or an assembly. A IV" appearing in this column in lieu of a quantity indicates that the quantity is variable and the quantity may vary from application to application. 4. EXPLANATION OF COLUMNS (SECTION IV). a. NATIONAL STOCK NUMBER (NSN) INDEX. (1) STOCK NUMBER column. This column lists the NSN by National Identification Number NSN (NIIN) sequence. The NIIN consists of the last nine digits of the NSN. When using (i.e., 5305-01-674-1467) NIIN this column to locate an item, ignore the first 4 digits of the NSN. However, the complete NSN should be used when ordering items by stock number. (2) FIG. column. This column lists the number of the figure where the item is identified/located. The figures are in numerical order in Section II and III. (3) ITEM column. The item number identifies the item associated with the figure listed in the adjacent FIG. column. This item is also identified by the NSN listed on the same line. b. PART NUMBER INDEX. Part numbers in this index are listed by part number in ascending alphanumeric sequence (i.e., vertical arrangement of letter and number combination which places the first letter or digit of each group in order A through Z, followed by the numbers 0 through 9 and each following letter or digit in like order). (1) CAGEC column. The Commercial and Government Entity Code (CAGEC) is a 5-digit code used to identify the manufacturer, distributor, or Government agency, etc., that supplies the item. (2) PART NUMBER column. Indicates the primary number used by the manufacturer (individual, firm, corporation, or Government activity), which controls the design and characteristics of the item by means of its engineering drawings, specifications standards, and inspection requirements to identify an item or range of items. (3) STOCK NUMBER column. This column lists the NSN for the associated part number and manufacturer identified in the PART NUMBER and CAGEC columns to the left.
C-5
TM 3-6665-343-23&P 4. EXPLANATION OF COLUMNS (SECTION IV) (CONT). (4) FIG. Column. This column lists the number of the figure where the item is identified/located in Section II and III. (5) ITEM column. The item number is that number assigned to the item as it appears in the figure referenced in the adjacent figure number column. c. FIGURE AND ITEM NUMBER INDEX. (1) FIG. column. This column lists the number of the figure where the item is identified/located in Section II and III. (2) ITEM column. The item number is that number assigned to the item as it appears in the figure referenced in the adjacent figure number column. (3) STOCK NUMBER column. This column lists the NSN for the item. (4) CAGEC column. The Commercial and Government Entity Code (CAGEC) is a 5-digit code used to identify the manufacturer, distributor, or Government agency, etc., that supplies the item. (5) PART NUMBER column. Indicates the primary number used by the manufacturer (individual, firm, corporation, or Government activity), which controls the design and characteristics of the item by means of its engineering drawings, specification standards, and inspection requirements to identify an item or range of items. 5. SPECIAL INFORMATION. Not Applicable. 6. HOW TO LOCATE REPAIR PARTS. a. When National Stock Number or Part Number Is Not Known. (1) First. Using the table of contents, determine the functional group or subfunctional group to which the item belongs. This is necessary since figures are prepared for functional groups and subfunctional groups, and listings are divided into the same groups. (2) Second. Find the figure covering the functional group or subfunctional group to which the item belongs. (3) Third. Identify the item on the figure and use the Figure and Item Number Index to find the NSN. b. When National Stock Number or Part Number is Known. (1) First. Using the National Stock Number or the part Number Index, find the pertinent National Stock Number or Part Number. The NSN index is in National Item Identification Number (NIIN) sequence (see 4.a). The part numbers in the Part Number Index are listed in ascending alphanumeric sequence (see 4.b). Both indexes cross-reference you to the illustration/figure and item number of the item you are looking for. (2) Second. Turn to the figure and item number, verify that the item is the one you are looking for, then locate the item number in the repair parts list for the figure.
C-6 /(C-7 blank)
TM 3-6665-343-23&P Section II. REPAIR PARTS LIST
Figure C-1. Monitor, Chemical Agent (ICAM)
TM 3-6665-343-23&P SECTION II (1) (2) ITEM SMR NO CODE
(3) CAGEC
(4) PART NUMBER
(5)
(6)
DESCRIPTION AND USABLE ON CODES (UOC)
QTY
GROUP 00: MONITOR, CHEMICAL AGENT (ICAM) FIG. C-1. MONITOR, CHEMICAL AGENT (ICAM) 5-15-17100 1 2 3
XAFFA PACZZ PCFZZ
81361 81361 81361
5-15-17012 442-674 442-642
MONITOR KIT, CAM ................................................. CAP, NOZZLE, PROTEC........................................... CONFIDENCE SAMPLE............................................ END OF FIGURE
C-1-1
1 1 1
TM 3-6665-343-23&P
Figure C-2. Monitor Kit, CAM
TM 3-6665-343-23&P SECTION II (1) (2) ITEM SMR NO CODE
(3) CAGEC
(4) PART NUMBER
(5)
(6)
DESCRIPTION AND USABLE ON CODES (UOC)
QTY
GROUP 01: MONITOR KIT, CAM FIG. C-2. MONITOR KIT, CAM 5-15-17012 1 2 3 4 5
PACZZ XAFFA PAFZZ PAFFF PACZZ
81361 81361 81361 81361 81361
442-570 5-15-17013 5-15-13900 5-15-18520 442-546
HARNESS ASSEMBLY, CA................................................. MONITOR, CHEMICAL AG ................................................. BUZZER ............................................................................. CASE, CARRYING, CHEMI ................................................ FILTERED NOZZLE PAC .................................................... END OF FIGURE
C-2-1
2 1 1 1 4
TM 3-6665-343-23&P
Figure C-3. Monitor, Chemical Agent
TM 3-6665-343-23&P SECTION II (1) (2) ITEM SMR NO CODE
(3) CAGEC
(4) PART NUMBER
(5)
(6)
DESCRIPTION AND USABLE ON CODES (UOC)
QTY
GROUP 0101: MONITOR, CHEMICAL AGENT FIG. C-3. MONITOR, CHEMICAL AGENT 5-15-17013 1 2 3 4 5 6 7
PACZZ XAFFA PAFZZ PAFZZ PAFZZ XAFZZ XAFZZ
81361 81361 81361 81361 81361 81361 81361
442-674 5-15-17014 442-582 442-675 442-492 442-772 5-15-17129
8 9
PCFFF PAFZZ
81361 81361
5-15-17015 442-391
CAP, NOZZLE, PROTECTIVE ............................................. MODULE ASSEMBLY, MON ............................................... INSULATION SHEET, EL .................................................... RETAINER, BATTERY ........................................................ CAP, PROTECTIVE DUST .................................................. CAM DISC........................................................................... DRIFT TUBE MDL DISC...................................................... (Supplied with P/N 5-15-17130, PAGE C-5-1) CASE ASSEMBLY, MONITOR ............................................ SEAL, NONMETALLIC ........................................................ END OF FIGURE
C-3-1
1 1 1 1 1 1 1 1 1
TM 3-6665-343-23&P
Figure C-4. Module Assembly, Monitor
TM 3-6665-343-23&P SECTION II (1) (2) ITEM SMR NO CODE
(3) CAGEC
(4) PART NUMBER
(5)
(6)
DESCRIPTION AND USABLE ON CODES (UOC)
QTY
GROUP 010101: MODULE ASSEMBLY, MONITOR FIG. C-4. MODULE ASSEMBLY, MONITOR 5-15-17014
1 2 3 4
XAFFA PCFDD PAFZZ PAFZZ
81361 81361 81361 81361
5-15-17030 5-15-17150 442-519 442-494
MONITOR ASSEMBLY........................................................ PRINTED CIRCUIT BOA ..................................................... SCREW, MACHINE ............................................................. SCREW, MACHINE ............................................................. END OF FIGURE
C-4-1
1 1 4 2
TM 3-6665-343-23&P
Figure C-5. Monitor Assembly
TM 3-6665-343-23&P SECTION II (1) (2) ITEM SMR NO CODE
(3) CAGEC
(4) PART NUMBER
(5)
(6)
DESCRIPTION AND USABLE ON CODES (UOC)
QTY
GROUP 01010101: MONITOR ASSEMBLY FIG. C-5. MONITOR ASSEMBLY 5-15-17030
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23
PAFZZ XAFZZ XAFZZ PAFZZ PAFZZ XDFZZ PAFZZ PAFFF XAFZZ PAFZZ PAFZZ PAFZZ XDFZZ PAFZZ PAFZZ PAFZZ PAFZZ PAFZZ PAFZZ PAFZZ PAFZZ PAFZZ PAFFA XAFZZ PAFZZ
81361 81361 81361 81361 81361 81361 81361 81361 81361 96906 96906 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361
5-15-17033 5-15-17108 5-15-17114 5-15-17054 5-15-17089 5-15-17049 5-15-17052 442-314 442-NPN MS9068-023 MS27512-10E 442-489 5-15-17047 442483 5-15-17125 5-15-17127 442-519 5-15-17040 5-15-17044 5-15-17034 442-377 5-15-17051 5-15-17130 5-15-17129 442-073
24 25 26
KFFZZ PAFZZ KFFZZ
81361 81361 81361
442-724 442-445 442-706
27
KFFZZ
81361
442-750
28
KFFZZ
81361
442-516
29 30 31 32 33 34
PAFZZ PAFZZ PAFZZ PAFZZ PAFZZ PAFZZ
81361 81361 81361 81361 81361 81361
442-633 442-514 442-310 442-512 442-522 442-521
SIEVE PACK ASSEMBLY ................................................... ..O-RING ............................................................................. ..ACETONE SOURCE ASSEMBLY...................................... STANDOFF, THREADED .................................................... GROMMET, NONMETALIC ................................................ CHASSIS L.H. ASSEMBLY.................................................. STANDOFF, THREADED .................................................... CONNECTOR RECEPTACLE ............................................. ..CONNECTOR, CHASSIS PLATE ...................................... ..PACKING, REFORMED .................................................... ..NUT, PLAIN EXAGON ....................................................... SCREW, MACHINE ............................................................. CHASSIS, R.H. ASSEMBLY ................................................ SCREW, MACHINE ............................................................. PRINTED WIRING BOARD ................................................. SPACER, SLEEVE .............................................................. SCREW, MACHINE ............................................................. BRACKET, MOUNTING....................................................... PLATE, MOUNTING ............................................................ PUMP, ROTARY ................................................................. SEAL, NONMETALLIC ........................................................ SCREW, MACHINE ............................................................. MODULE, DRIFT TUBE....................................................... ..CELL MODULE ................................................................. ..MODULE KIT, DRIFT ........................................................ .. ..PLUG, SEALING............................................................. .. ..SHIM, NOZZLE HOLDER (1) C-5-26.......................... .. ..SHIM, RETAINING SLEEVE (3) C-5-27 ......................... .. ..SEAL, NOZZLE HOLDER (1) C-5-28 ......................... PLUG, SEALING PART OF KIT P/N 442-073....................... NOZZLE HOLDER ASSEMBLY ........................................... SHIM, NOZZLE HOLDER .................................................... PART OF KIT P/N 442-073 SHIM, RETAINING SLEEVE................................................ PART OF KIT P/N 442-073 SEAL, NOZZLE HOLDER.................................................... PART OF KIT P/N 442-073 CASE FRONT-END ............................................................. SEAL, NONMETALLIC ........................................................ NOZZLE ASSEMBLY .......................................................... SPACER, SLEEVE .............................................................. WASHER, FLAT .................................................................. RING, LOCK, KEYED .......................................................... END OF FIGURE C-5-1
1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 3 1 1 1 2 3 1 1 1
1 1 1 3 1 1 1 1 1 1 1
TM 3-6665-343-23&P
Figure C-6. Module, Drift Tube
TM 3-6665-343-23&P SECTION II (1) (2) ITEM SMR NO CODE
(3) CAGEC
(4) PART NUMBER
(5)
(6)
DESCRIPTION AND USABLE ON CODES (UOC)
QTY
GROUP 0101010101: MODULE, DRIFT TUBE FIG. C-6. MODULE, DRIFT TUBE 5-15-17130 1 2 3 4 5 6 7
PAFZZ PAFZZ PAFZZ PAFZZ PAFZZ PAFZZ PAFZZ
81361 96906 81361 81361 81361 81361 81361
442-313 MS3367-7-9 5-15-17086 5-15-17032 5-15-17042 5-15-17041 5-15-17043
MEMBRANE ASSEMBLY ................................................... STRAP, TIEDOWN, ELECT................................................. TUBING, NONMETALLIC ................................................... FILTER ASSEMBLY, DRI .................................................... TUBING, NONMETALLIC .................................................... TUBING, NONMETALLIC .................................................... SEAL, NONMETALLIC SP.................................................. END OF FIGURE
C-6-1
1 2 2 1 1 1 1
TM 3-6665-343-23&P
Figure C-7. Case Assembly, Monitor
TM 3-6665-343-23&P SECTION II (1) (2) ITEM SMR NO CODE
(3) CAGEC
(4) PART NUMBER
(5)
(6)
DESCRIPTION AND USABLE ON CODES (UOC)
QTY
GROUP 010102: CASE ASSEMBLY, MONITOR FIG. C-7. CASE ASSEMBLY, MONITOR 5-15-17015
1 2 3 4 5
PAFZZ PCFZZ XAFZZ PAFZZ PAFZZ
81361 81361 81361 81361 81361
442-419 5-15-17021 5-15-17016 442-726 441-519
CLIP, RETAINING ............................................................... DISPLAY, OPTOELECTRON IC .......................................... CASE ASSEMBLY............................................................... CONTACT, ASSEMBLY BA................................................. SCREW, MACHINE ............................................................. END OF FIGURE
C-7-1
1 1 1 1 2
TM 3-6665-343-23&P
Figure C-8. Carrying Case Assembly
TM 3-6665-343-23&P SECTION II (1) (2) ITEM SMR NO CODE
(3) CAGEC
(4) PART NUMBER
(5)
(6)
DESCRIPTION AND USABLE ON CODES (UOC)
QTY
GROUP 0102: CARRYING CASE ASSEMBLY FIG. C-8. CARRYING CASE ASSEMBLY 5-15-18520 1 2
XAFZZ PAFZZ
81361 81361
5-15-NPN 5-15-18517
CASE, CARRYING .............................................................. STRAP, WEBBING .............................................................. END OF FIGURE
C-8-1
1 1
TM 3-6665-343-23&P Section III. SPECIAL TOOLS LIST. Not Applicable.
C-9-1
SECTION IV
TM 3-6665-343-23&P CROSS-REFERENCE INDEXES NATIONAL STOCK NUMBER INDEX
STOCK NUMBER
FIG.
5330-00-182-3115 5975-00-570-9598 5310-01-127-2137 4320-01-380-0297 9330-01-380-0281 9330-01-380-0282 9330-01-380-1331 6665-01-380-1680 6665-01-380-1681 6665-01-380-1688 5980-01-380-7627 5330-01-380-7678 6665-01-380-8449 6665-01-380-9903 6665-01-380-9974
C-5 C-6 C-5 C-5 C-6 C-6 C-6 C-5 C-3 C-6 C-7 C-6 C-5 C-3 C-1 C-3 C-5 C-3 C-3 C-1 C-2 C-5 C-5 C-7 C-5 C-5 C-5 C-5 C-7 C-5 C-4 C-6 C-5 C-5 C-5 C-5 C-2 C-4 C-5 C-4 C-5 C-7
6665-01-381-5085 5340-01-382-0201 6665-01-382-3203 6665-01-382-7081 6665-01-382-7084 6665-01-382-8029 6665-01-382-8038 5340-01-382-8040 6665-01-382-8526 6665-01-382-8546 5330-01-382-8555 6665-01-382-8576 5999-01-383-2792 5998-01-383-2795 5998-01-383-4219 6665-01-383-6195 6665-01-384-0287 5935-01-384-9748 5330-01-385-6397 5305-01-388-2289 6665-01-388-4269 5305-01-388-5080 5305-01-388-5169
ITEM 10 2 11 20 6 7 3 1 8 4 2 5 23 9 2 1 21 4 5 3 5 33 34 1 25 KIT 31 29 4 15 2 1 32 8 30 12 1 4 13 3 17 5
STOCK NUMBER
FIG.
ITEM
5970-01-389-7173 5365-01-393-6329 5325-01-393-7104 5305-01-393-7790 5340-01-394-5377 5305-01-394-5378 6350-01-394-9916 5340-01-395-0132 5340-01-395-2789 6665-01-417-9307 5340-01-433-6375 5305-01-451-2799
C-3 C-5 C-5 C-5 C-5 C-5 C-2 C-5 C-5 C-2 C-8 C-5
3 16 5 22 19 18 3 7 4 4 2 14
C-I-1
SECTION IV
TM 3-6665-343-23&P CROSS-REFERENCE INDEXES PART NUMBER INDEX
CAGEC
PART NUMBER
STOCK NUMBER
FIG.
ITEM
C-5 C-6 C-5 C-5 C-5 C-5 C-6 C-5 C-5 C-3 C-7 C-5 C-5 C-5 C-5 C-4 C-5 C-5 C-5 C-5 C-4 C-5 C-7 C-5 C-5 C-2 C-2 C-3 C-5 C-1 C-1 C-3 C-3 C-5 C-5 C-7 C-5 C-3 C-8 C-2 C-1 C-2 C-3 C-3 C-7 C-7 C-4 C-6
11 2 10 9 KIT 31 1 8 21 9 1 25 14 12 5 4 13 32 30 28 3 17 5 34 33 5 1 3 29 3 2 1 4 26 24 4 27 6 1 3 1 2 2 8 3 2 1 4
96906 96906 96906 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361
MS27512-10E MS3367-7-9 MS9068-023 442-NPN 442-073 442-310 442-313 442-314 442-377 442-391 442-419 442-445 442-483 442-489 442-492 442-494
5310-00-127-2137 5975-00-570-9598 5330-00-182-3115
81361 81361 81361 81361
442-512 442-514 442-516 442-519
6665-01-384-0287 5330-01-385-6397
81361 81361 81361 81361 81361 81361 81361 81361
442-521 442-522 442-546 442-570 442-582 442-633 442-642 442-674
6665-01-382-8038 6665-01-382-8029 6665-01-382-7084 6665-01-388-4269 5970-01-389-7173 6665-01-382-8576 6665-01-382-7081 6665-01-380-9974
81361 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361
442-675 442-706 442-724 442-726 442-750 442-772 5-15-NPN 5-15-13900 5-15-17012 5-15-17013 5-15-17014 5-15-17015 5-15-17016 5-15-17021 5-15-17030 5-15-17032
5340-01-382-0201
6665-01-382-8546 6665-01-382-8555 6665-01-383-6195 5936-01-384-9748 6665-01-381-5085 6665-01-380-9903 5340-01-382-8040 6665-01-382-5826 5305-01-451-2799 5305-01-388-2289 6665-01-382-3203 5305-01-388-5080
5305-01-388-5169
5999-01-383-2792
6350-01-394-9916
6665-01-380-1681 5980-01-380-7626 6665-01-380-1688 C-I-2
SECTION IV
TM 3-6665-343-23&P CROSS-REFERENCE INDEXES PART NUMBER INDEX
CAGEC 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361
PART NUMBER 5-15-17033 5-15-17034 5-15-17040 5-15-17041 5-15-17042 5-15-17043 5-15-17044 5-15-17047 5-15-17049 5-15-17051 5-15-17052 5-15-17054 5-15-17086 5-15-17054 5-15-17108 5-15-17114 5-15-17125 5-15-17127 5-15-17129 5-15-17130 5-15-17150 5-15-18517 5-15-18520
STOCK NUMBER
FIG.
ITEM
6665-01-380-1680 4320-01-380-0279 5340-01-394-5387 9330-01-380-0281 5330-01-380-7678 9330-01-380-0282 5340-01-394-5377
C-5 C-5 C-5 C-6 C-6 C-6 C-5 C-5 C-5 C-5 C-5 C-5 C-6 C-5 C-5 C-5 C-5 C-5 C-3 C-5 C-4 C-8 C-2
1 20 18 6 5 7 19 13 6 22 7 4 3 5 2 3 15 16 7 23 2 2 4
5305-01-393-7790 5340-01-395-0132 5340-01-395-2789 9330-01-380-1331 5325-01-393-7104
5998-01-383-2795 5365-01-393-6329 6665-01-380-8449 5998-01-383-4219 5340-01-433-6375 6665-01-417-9307
C-I-3
SECTION IV
TM 3-6665-343-23&P CROSS-REFERENCE INDEXES FIGURE AND ITEM NUMBER INDEX
FIG.
C-1 C-1 C-1 C-2 C-2 C-2 C-2 C-2 C-3 C-3 C-3 C-3 C-3 C-3 C-3 C-3 C-3 C-4 C-4 C-4 C-4 C-5 C-5 C-5 C-5 C-5 C-5 C-5 C-5 C-5 C-5 C-5 C-5 C-5 C-5 C-5 C-5 C-5 C-5 C-5 C-5 C-5 C-5 C-5 C-5 C-5 C-5 C-5
ITEM
1 2 3 1 2 3 4 5 1 2 3 4 5 6 7 8 9 1 2 3 4 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27
STOCK NUMBER
CAGEC
6665-01-380-9974 6665-01-382-7081 6665-01-388-4269
81361 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361 96906 96906 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361
6350-01-394-9916 6665-01-417-9307 6665-01-382-7084 6665-01-380-9974 5970-01-389-7173 5340-01-382-0201 6665-01-382-3203
6665-01-380-1681 6665-01-380-9903 5998-01-383-4219 5305-01-388-5169 5305-01-388-5080 6665-01-380-1680
5340-01-395-2789 5325-01-393-7104 5340-01-395-0132 5935-01-384-9748 5330-00-182-3115 5310-01-127-2137 5305-01-388-2289 5305-01-451-2799 5998-01-383-2795 5365-01-393-6329 5305-01-388-5169 5340-01-394-5378 5340-01-394-5377 4320-01-380-0279 6665-01-381-5085 5305-01-393-7790 6665-01-380-8449 6665-01-382-8526
C-I-4
PART NUMBER 5-15-17012 442-674 442-642 442-570 5-15-17013 5-15-13900 5-15-18520 442-546 442-674 5-15-17014 442-582 442-675 442-492 442-772 5-15-17129 5-15-17015 442-391 5-15-17030 5-15-17150 442-519 442-494 5-15-17033 5-15-17108 5-15-17114 5-15-17054 5-15-17089 5-15-17049 5-15-17052 442-314 442-NPN MS9068-023 MS27512-10E 442-489 5-15-17047 442-483 5-15-17125 5-15-17127 442-519 5-15-17040 5-15-17044 5-15-17034 442-377 5-15-17051 5-15-17130 442-724 442-445 442-706 442-750
SECTION IV
TM 3-6665-343-23&P CROSS-REFERENCE INDEXES FIGURE AND ITEM NUMBER INDEX
FIG.
C-5 C-5 C-5 C-5 C-5 C-5 C-5 C-5 C-6 C-6 C-6 C-6 C-6 C-6 C-6 C-7 C-7 C-7 C-7 C-7 C-8 C-8
ITEM
28 29 30 31 32 33 34 KIT 1 2 3 4 5 6 7 1 2 3 4 5 1 2
STOCK NUMBER
CAGEC
6665-01-382-8576 5330-01-385-6397 6665-01-382-8555 6665-01-384-0287 6665-01-382-8029 6665-01-382-8038 6665-01-382-8546 6665-01-383-6195 5975-00-570-9598 9330-01-380-1331 6665-01-380-1688 5330-01-380-7678 9330-01-380-0281 9330-01-380-0282 5340-01-382-8040 5980-01-380-7626
81361 81361 81361 81361 81361 81361 81361 81361 81361 96906 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361 81361
5999-01-383-2792 5305-01-388-5169 5340-01-433-6375
C-l-5/(C-I-6 blank)
PART NUMBER 442-516 442-633 442-514 442-310 442-512 442-522 442-521 442-073 442-313 MS3367-7-9 5-15-17086 5-15-17032 5-15-17042 5-15-17041 5-15-17043 442-419 5-15-17021 5-15-17016 442-726 442-519 5-15-NPN 5-15-18517
TM 3-6665-343-23&P APPENDIX D EXPENDABLE AND DURABLE ITEMS LIST
Section I. INTRODUCTION D-1.
SCOPE.
This appendix lists expendable and durable items that you will need to operate and maintain the Chemical Agent Monitor (CAM). This listing is for information only and is not authority to requisition the listed items. These items are authorized to you by CTA 50-970, Expendable/Durable Items (except medical, Class V repair parts and heraldic items) or CTA 8-100, Army Medical Department Expendable/Durable Items. D-2.
EXPLANATION OF COLUMNS.
a. Column (1) - Item Number. This number is assigned to the entry in the listing and is referenced in the narrative instructions to identify the item (e.g., "Use gloves, Item 4, App. D "). b. Column (2) - Level. This column identifies the lowest level of maintenance that requires the listed item (C-Operator/Crew, O - Unit Maintenance, F - Direct Support Maintenance). c. Column (3) - National Stock Number. This is the National stock number assigned to the item which you can use to requisition it. d. Column (4) - Description. Indicates the item name, description, (CAGEC) and part number. This provides the other information you need to identify the item. e. Column (5) - Unit of Measure (U/M). this code shows the physical measurement or count of an item, such as gallon (GL), each (EA), dozen (DZ), gross (GR), etc. If the unit of measure differs from the unit of issue, requisition the lowest unit of issue that will satisfy your requirements.
D-1
TM 34665-343-23&P
Section II. EXPENDABLE AND DURABLE ITEMS LIST (4) DESCRIPTION CAGEC AND PART NUMBER
(5) U/M
(1) ITEM NUMBER
(2) LEVEL
(3) NATIONAL STOCK NUMBER
1
F
8105-01-011-2564
Bag, Heat Sealable (81349) MIL-B-117
EA
2
F
8305-00-267-3015
Cloth, Cheesecloth (22527) 06-665-17
YD
3
F
5999-21-909-9984
Contact Enhancer (38948) 007-065-015
BT
4
F
8415-00-682-6786
Gloves, Disposable (96717) Pinkies
PR
5
F
8135-00-160-7770
Paper, Kraft, Untreated (58536) A-A-203
FT
6
F
7520-01-060-5820
Pen, Ball-Point (83421) FMFO-004
EA
7
F
8520-00-228-0598
Soap, Toilet (81348) P-S-624
OZ
8
F
7510-00-266-6712
Tape, Pressure Sensitive, Adhesive (19203) 8783476
YD
9
F
4020-00-243-3156
Twine, Fibrous (80244) TT881TYPE1
FT
10
F
6665-01-447-5639
Wipe Test Kit (81361) 5-15-18960
KT
D-2
TM 3-6665-343-23&P APPENDIX E NUCLEAR REGULATORY COMMISSION REQUIREMENTS INFORMATION ABOUT TITLE 10 AND THE LICENSE E-1.
RULES AND REGULATIONS/NRC LICENSE.
The U.S. Nuclear Regulatory Commission (NRC) sets standards and issues licenses for the safe use, storage, and possession of items containing radioactive material. The CAM contains the radioisotope Nickel-63. CAMs should be used and stored in accordance with the conditions of the NRC license issued to Armament and Chemical Acquisition and Logistics Activity, ATTN: AMSTA-AC-SF, Rock Island, IL 61299-7630. Current copies of the NRC regulations (Title 10 Code of Federal Regulation) and the licenses are available from the licensee upon request. E-2.
NOTICE TO EMPLOYEES.
Form NRC-3, Notice to Employees, contained on the following page, is provided for reference whenever the radioactive material described by this technical manual is used. E-3.
DISPOSAL. Unserviceable CAMs will be turned in to depot for disposal as radioactive waste.
E-4.
IDENTIFICATION OF RADIOACTIVE MATERIAL.
The CAM contains the radioactive source Nickel-63. The CAM is labeled on two sides to indicate radioactive material is present. E-5.
PART 21.
Title 10 CFR part 21 requires a responsible officer of the licensed organization to promptly evaluate and report defects and noncompliances which relate to substantial safety hazards. Failure to comply may result in personal fines to these officers. a. Substantial safety hazard examples: Exposure to, or release of, licensed radioactive material Major degradation of essential safety related equipment Major deficiencies involving design or use of licensed material b. Responsible Officer or Radiological Protection Officer. Any suspected defect or noncompliance should be reported in writing or telephonically promptly to the following: Armament and Chemical Acquisition and Logistics Activity ATTN: AMSTA-AC-SF Rock Island, IL 61299-7630 DSN: 793-2965/2995/2962 COMMERCIAL: (309) 782-2965/2995/2962 AFTER DUTY HOURS: DSN 793-3005/6587 COMMERCIAL: (309) 782-3005/6587
E-1
TM 3-6665-343-23&P
E-2
TM 3-6665-343-23&P
INDEX
PAGE
SUBJECT A Abbreviations -------------------------------------------------------------------------------------------------------------------------------Accountability (DS)------------------------------------------------------------------------------------------------------------------------Accountability (Unit)-----------------------------------------------------------------------------------------------------------------------Army Maintenance System--------------------------------------------------------------------------------------------------------------Assembly and Preparation of CAM for Use ----------------------------------------------------------------------------------------C
1-2 3-142 3-2 B-1 3-8
Calibration -----------------------------------------------------------------------------------------------------------------------------------Cell Test -------------------------------------------------------------------------------------------------------------------------------------Common Table of Allowances----------------------------------------------------------------------------------------------------------Common Tools and Equipment --------------------------------------------------------------------------------------------------------Confidence Test ---------------------------------------------------------------------------------------------------------------------------Corrosion Prevention and Control ------------------------------------------------------------------------------------------------------
1-4 3-129 A-1 1-7 3-15 1-3
D DA Pamphlets -----------------------------------------------------------------------------------------------------------------------------Destruction of Army Materiel to Prevent Enemy Use -----------------------------------------------------------------------------Diagnostic Test Set, Preparation for Use --------------------------------------------------------------------------------------------Direct Support Maintenance Procedures --------------------------------------------------------------------------------------------Battery Cap Assembly Replacement---------------------------------------------------------------------------------------Battery Contact Assembly Replacement----------------------------------------------------------------------------------Carrying Case Assembly Replacement ----------------------------------------------------------------------------------Carrying Case Strap Replacement ----------------------------------------------------------------------------------------Case-Front End Replacement -----------------------------------------------------------------------------------------------Confidence Sample Replacement -----------------------------------------------------------------------------------------Connector Assembly Replacement-----------------------------------------------------------------------------------------Display Assembly Replacement---------------------------------------------------------------------------------------------Drift Tube Module Replacement and Alignment-------------------------------------------------------------------------Environmental Cap Assembly Replacement -----------------------------------------------------------------------------Filter Assembly Replacement------------------------------------------------------------------------------------------------General Maintenance Instructions -----------------------------------------------------------------------------------------Membrane Assembly Replacement ----------------------------------------------------------------------------------------Monitor Case Assembly Replacement-------------------------------------------------------------------------------------Nozzle Assembly Replacement ---------------------------------------------------------------------------------------------Nozzle Holder Assembly Replacement------------------------------------------------------------------------------------Printed Circuit Board Replacement and Adjustment -------------------------------------------------------------------Printed Wiring Board and Replacement-----------------------------------------------------------------------------------Pump Assembly Replacement-----------------------------------------------------------------------------------------------Sieve Pack Assembly Replacement ---------------------------------------------------------------------------------------Direct Support Troubleshooting --------------------------------------------------------------------------------------------------------Equipment Required -----------------------------------------------------------------------------------------------------------Operational Check--------------------------------------------------------------------------------------------------------------Preparation of Diagnostic Test Set for Use-------------------------------------------------------------------------------Troubleshooting Symptoms 1. Mechanically broken push-button switch or cracked or broken display ----------------------------------2. Pump not running------------------------------------------------------------------------------------------------------
Index-1
A-1 1-1 3-22 3-78 3-127 3-88 3-128 3-128 3-119 3-81 3-102 3-86 3-108 3-126 3-106 3-79 3-124 3-82 3-120 3-115 3-91 3-103 3-104 3-100 3-20 3-22 3-24 3-22 3-29 3-3
TM 3-6665-343-23&P D (Cont) 3. 4. 5. 6. 7. 8. 9. 10. 11. 12. 13. 14. 15. 16. 17. 18. 19. 20. 21.
CAM AMPS not between 3.6 and 6.0, or CAM does not come on ------------------------------------All indications are not displayed during self-test -----------------------------------------------------------Display does not complete self-test or BL is displayed when CAM is powered by DTS --------------------------------------------------------------------------------------------------WAIT is displayed longer than 2 minutes--------------------------------------------------------------------BL not shown when batt low switch is pressed ------------------------------------------------------------No display or BL displayed after self-test when CAM is powered by battery ----------------------Nozzle flow is less than 200 ml/min --------------------------------------------------------------------------Gate width, gate height, or pulse width not within specification----------------------------------------Noise level in waveform test is excessive -------------------------------------------------------------------Secondary peaks present in waveform test or CAM will not clear down ----------------------------DC offset is greater than 1 V-----------------------------------------------------------------------------------(G or H) peak heights, or ratio of times not within specification ---------------------------------------Cell does not pass cell test -------------------------------------------------------------------------------------Monitor module fails leak test ----------------------------------------------------------------------------------Does not respond to confidence sample --------------------------------------------------------------------Display flashes ON and OFF ----------------------------------------------------------------------------------Does not change mode from G to H or H to G ----------------------------------------------------------No display or BL displayed after self-test when CAM powered by DTS -----------------------------Buzzer does not sound when display shows two bars or more -----------------------------------------
3-31 3-40 3-42 3-43 3-45 3-46 3-48 3-51 3-52 3-56 3-62 3-63 3-66 3-67 3-70 3-73 3-74 3-76 3-77
E Equipment Characteristics, Capabilities and Features --------------------------------------------------------------------------- 1-4 Equipment Description and Data ------------------------------------------------------------------------------------------------------ 1-4 Expendable and Durable Items List---------------------------------------------------------------------------------------------------- D-2 F Field Manuals ------------------------------------------------------------------------------------------------------------------------------Fire and Explosion Emergencies -----------------------------------------------------------------------------------------------------Forms ---------------------------------------------------------------------------------------------------------------------------------Forms, Records and Reports -----------------------------------------------------------------------------------------------------------
A-1 3-3 A-1 1-1
G General Maintenance Instructions (DS) ---------------------------------------------------------------------------------------------- 3-79 H How To Use This Manual ---------------------------------------------------------------------------------------------------------------- iii L Leak Test ---------------------------------------------------------------------------------------------------------------------------------- 3-132 Location and Description of Major Components ----------------------------------------------------------------------------------- 1-5 M Maintenance Allocation Chart (MAC)-------------------------------------------------------------------------------------------------- B-4
Index-2
TM 3-6665-343-23&P N Nomenclature Cross-Reference List ------------------------------------------------------------------------------------------------- 1-2 Nuclear Regulatory Commission Requirements ----------------------------------------------------------------------------------- E-1 O Operational Check ------------------------------------------------------------------------------------------------------------------------- 3-24 P Preparation for Storage or Shipment (DS) ------------------------------------------------------------------------------------------Preparation for Storage or Shipment (Unit) ----------------------------------------------------------------------------------------Preparation of Diagnostic Test Set for Use -----------------------------------------------------------------------------------------Principles of Operation ------------------------------------------------------------------------------------------------------------------Pump Test ---------------------------------------------------------------------------------------------------------------------------------Purpose of Equipment---------------------------------------------------------------------------------------------------------------------
3-142 2-10 3-22 1-8 3-135 1-1
R Radiological Safety Instructions -------------------------------------------------------------------------------------------------------References ---------------------------------------------------------------------------------------------------------------------------------Repair Parts and Special Tools List---------------------------------------------------------------------------------------------------Section I Introduction ---------------------------------------------------------------------------------------------------------------------Section II Repair Parts List -------------------------------------------------------------------------------------------------------------Section III Special Tools List -----------------------------------------------------------------------------------------------------------Section IV Cross-Reference Indexes ------------------------------------------------------------------------------------------------Reporting Equipment Improvement Recommendations (EIR)------------------------------------------------------------------Regulations ----------------------------------------------------------------------------------------------------------------------------------
3-1 A-1 C-1 C-1 C-1-1 C-9-1 C-1-1 1-2 A-1
S Safety, Care and Handling -------------------------------------------------------------------------------------------------------------Self-Test ---------------------------------------------------------------------------------------------------------------------------------Service Upon Receipt -------------------------------------------------------------------------------------------------------------------Shipment (DS) -----------------------------------------------------------------------------------------------------------------------------Shipment (Unit) ----------------------------------------------------------------------------------------------------------------------------Soldier Training Publications ----------------------------------------------------------------------------------------------------------Special Tools, TMDE and Support Equipment ------------------------------------------------------------------------------------Stock Number Cross Reference Indexes --------------------------------------------------------------------------------------------Storage (DS)--------------------------------------------------------------------------------------------------------------------------------Storage (Unit)-------------------------------------------------------------------------------------------------------------------------------T Technical Bulletins ------------------------------------------------------------------------------------------------------------------------Technical Manuals ------------------------------------------------------------------------------------------------------------------------Tests Cell Test---------------------------------------------------------------------------------------------------------------------------Leak Test -------------------------------------------------------------------------------------------------------------------------Pump Test -----------------------------------------------------------------------------------------------------------------------Waveform Test -----------------------------------------------------------------------------------------------------------------Three Dots ----------------------------------------------------------------------------------------------------------------------------------
1-2 3-13 2-1 3-142 2-10 A-2 1-7 C-1-1 3-142 2-10
A-2 A-2 3-129 3-132 3-135 3-137 1-11
Index-3
TM 3-6665-343-23&P T (Cont) Tool and Test Equipment Requirements---------------------------------------------------------------------------------------------- B-5 Training Publications, Soldier ----------------------------------------------------------------------------------------------------------- A-2 Troubleshooting Symptoms -------------------------------------------------------------------------------------------------------------- 3-21 U Unit Maintenance Instructions ---------------------------------------------------------------------------------------------------------Unit Preventive Maintenance Checks and Services (PMCS) ------------------------------------------------------------------Unit Troubleshooting----------------------------------------------------------------------------------------------------------------------Troubleshooting Symptoms 1. Backlight not operating --------------------------------------------------------------------------------------------------2. BL shown on display during operation -------------------------------------------------------------------------------3. Display does not come on or display disappears -----------------------------------------------------------------4. Display flashes on and off ----------------------------------------------------------------------------------------------5. Does not change mode from G to H or H to G --------------------------------------------------------------------6. Does not clear down to zero or one bar in 2 minutes -----------------------------------------------------------7. Does not clear down to zero or one bar before shutdown ------------------------------------------------------8. Does not respond to confidence sample (simulant) --------------------------------------------------------------9. One or more display indicators do not show during self-test ---------------------------------------------------10. Remains in self-test for more than 30 seconds --------------------------------------------------------------------11. Remains in wait for longer than 2 minutes --------------------------------------------------------------------------12. Buzzer does not sound when display shows two or more bars ------------------------------------------------Unpacked Equipment, Checking ------------------------------------------------------------------------------------------------------Unpacking ---------------------------------------------------------------------------------------------------------------------------------W Waveform Test ----------------------------------------------------------------------------------------------------------------------------Wipe Test ---------------------------------------------------------------------------------------------------------------------------------Work Area ---------------------------------------------------------------------------------------------------------------------------------Work Surface, Clean ----------------------------------------------------------------------------------------------------------------------
Index-4
2-1 2-3 2-4 2-5 2-5 2-5 2-5 2-6 2-6 2-7 2-7 2-8 2-8 2-9 2-9 3-6 3-4 3-137 3-6 3-1 3-22
DISTRIBUTION: To be distributed in accordance with the initial requirements for SPECIAL DISTRIBUTION for TM 3-6665-343-23&P
By Order of the Secretary of the Army:
DENNIS J. REIMER General, United States Army Chief of Staff Official:
JOEL B. HUDSON Administrative Assistant to the Secretary of the Army 04749
I U.S. GOVERNMENT PRINTING OFFICE : 1998-646-036/60505
PIN: 076560-000