Polydoros IT Wiring Diagrams
Short Description
polidoros it...
Description
SIEMENS
AKTIENGESELLSCHAFT Medical Solutions Inde x Type No. / Board-No.
Surve sheet to wiring diagram
Product:
A
Sheet No. Revision-No.
1A 1
Type No.:
Polydoros IT
000 00
X2287 Author Department
Änd.Mittlg.Nr.
Date
-
02.08.05
Schnapperelle CO PTS
Print Nr. COPL-136.844.03.01.02
X2287 - A - 57 60 074 Type No.
2
00
3
00
4
00
5
00
6
00
7
00
8
00
Print No.
9
00
10
00
Type No.
11
00
12
00
13
00
14
00
15
00
16
00
17
00
18
00
18
00
19
00
20
00
21
00
Inde x
-
01
Part number
Rev .
Part number
Rev .
Part number
Rev .
-
Print No. Type No.
Inde x
-
-
Inde x
RevisionNo.000 Dept.: CO PTS
Date: Checked: Als Betriebsgeheimnis anvertraut/Disclosed as trade secret, Alle Re chte vorbehalten/ All rights reserved, Uebersichtsblatt zum Schaltplan / Surve sheet to wiring diagram, Erzeugnis / Produkt, Sach-Nr./ Part No, Blatt-Nr./ Sheet no., Ausführungs-Nr./ Version No., Typen-Nr./ Type No., Änd.-Mittlg.-Nr./ Change nitice No., Datum / Date, Ersteller / Issuer, Abtlg./ Department, Index / Index, Geprüft / Checked, U-Art / Doc.type, U-Stand / Doc.status, Druck Nr./ Print No., Ersatz für / Replacement for
Part-No.
57 60 074 Replacement for:
Art
EFS -
Teil
- 01S -
Version
00
Type-No.
- X2287 Sheet 01A
s A
B
C
Table of contents
4
3
Only for internal use appoimted. All rights reserved.
Copyright © SIEMENS AG All rights reserved.
2
Author: Date: Schnapperelle CO PTS 02.08.05
4
Page 1A
Cover page
1
Overview of pages
2
Functional Overview Polydoros IT
3
Overview of the Generator
4
External connections
5
Schematic Polydoros IT 30 / 55
6
Schematic Polydoros IT 65 / 80
7
Power supply
8
D500 Circuit board 77 59 819
9
D500 Circuit board 57 60 165
10
LEDs / Jumpers D500
11
Test points D500
12
D501 Circuit board 77 59 835
13
D501 Circuit board 57 60 181
14
LEDs / switches D501
15
D470 Filament
16
D115 Rotation Anode Circuit RAC
17
D510 Inverter IT 30/55
18
D510 Inverter IT 65/80
19
Connection Single Tank
20
Connection Optitop Tube
21
Option Energy Storage Unit
3
2
1
1
Sheet No. 1 Description
RevisionNo.00
A
B
Overview of pages
Part-No
Art
57 60 074
EFS
Teil
Version
01S
00 C
Type-No.
X2287
s B
A
D500 Generator
C
D510 kV Inverter
Controller board kV Control 4
4
Main Inverter Control
≈
MCB *1 + Dongle *2 RAC Control
I_Roh Control I FilamentControl
FIL Inverter Control
UDC D511
Monitoring
Resonant circuit
3
D501 Interface
Only for internal use appointed. All rights reserved.
3
D470 Filament inverter
24V dc Supply
≈
AEC IONTOMAT
D115 Rotating Anode Circuit
Copyright © SIEMENS AG All rights reserved.
I/O Interface
≈ Tube
2
Control Console *3
Author: Date: Schnapperelle CO PTS 02.08.05
2
External connections HVT M16 Linefilter
Transformer optional 1
additional in STTA
L1 L2 L3
*1 Inclusive SW Generator and OGP Data for Sta ndard desk *2 SW Optionen *3 Touch panel with SD Card for SW and OGP's A
B
Sheet No. 2
RevisionNo.00 Description
1
Functional Overview Polydoros IT
Part-No.
Art
57 60 074
EFS
Teil
Version
01S
00 C
Type-No.
X2287
s C
B
A
Overview of the Generator Cabinet
4
4
3
3
Only for internal use appointed. All rights reserved.
Copyright © SIEMENS AG All rights reserved.
Overview of the Control Consoles
2
2
Author: Date: Schnapperelle CO PTS 02.08.05
1
Control console (standard) / organ console (option)
Touch panel 1
Sheet No. 3
RevisionNo.00 Description
A
B
Overview of the Generator
Part-No.
Art
57 60 074
EFS
Teil
Version
01S
00 C
Type-No.
X2287
s C
B
A
X4
V_P15V
F2 (5A)
6 5
F4 (1A)
-15V
6
ANA_GND
7
+5V
8
GND
9
ANA_GND
Workstation
24V
2 4
PREP
6
STANDBY
CONS_IN_2 CONS_IN_1
2
XRAY
3
24V, 5W
24V
Raster_In Xray_On
K8 24V
Select_G2 K12
HK_G2
24V
GND24
5 24V, 5W
6
STANDBY
Only for internal use appointed. All rights reserved.
7
24V, 5W
K14 24V PLANI_TIME_E
1 2
Copyright © SIEMENS AG All rights reserved.
PLANI_TIME_C 24V DOOR_STATE_IN 24V
TUBE: TEMP
3 4
T_MAX_TUBE_IN
TUBE: PRESSURE
Energie Storage Unit:READY
7 8
ESU_READY
24V
Schnapperelle CO PTS 02.08.05
Ready_In_G1
24V PLANI_END_S_IN ( Plani_Time_A ) PLANI_STARTPOS_IN ( Plani_Time_B ) PLANI_8GRAD_SYNC_IN ( Plani_Time_C ) PLANI_30GR_SYNC_IN ( Plani_Time_D ) RASTER_IN (Grid) PLANI_HK (Exposure_Request)
V_24V_1 GND24
Dose Area Product_Unit (KermaX) Iontomat_Chamber A
X14
Ready_In_G2
X33
V_24V_1
24V, 5W
24V GND
1 2 3 4
Xray_End_PF1 Xray_End_PF2 GND24
X22
Iontomat_Chamber B
24V
G2 OEM
Xray_End
X34
V_24V_1 Gen. Purpuse In
X23
2
1 2 3 4 5 6 7 8
HK_G2 Grid_G2 HK_G2_PF1 HK_G2_PF2
24V
G1 OEM
X31
GND24 Select_G2
24V
KK2.G1/G2
1 2 3 4 5 6 7 8
HK_G1 Grid_G1 HK_G1_PF1 HK_G1_PF2 V_24V_1
24V
3
X30
Select_G1
24V
KK 1B.G1/G2
18 20
GND24
HK: Exposure_Request VK: Prep PF: Potential Free
GND24
P_MAX_TUBE_IN
1 2 3 4 5 6 7 8
Author: Date:
PLANI_TIME_D
Signal description:
24V
5 6
X10
2
6
14 15 16 17
PLANI_TIME_B
X8
6
Control Console
X25
PLANI_TIME_A
8
DOOR
13
9 12 14 16 20
VK
4
PREP
13
2 5 7 8
HK_G1
1 24V, 5W
7 14
X24
GND24 Select_G1
X7
7
14
S2 Generator ON
8
Workstation
3
X19 W515
CONS_OUT_2
K114
7
230V, 100W
4
CONS_OUT 1
K112
5
230V, 100W
D501 Interface Board
K108
3
230V, 100W
S1 Generator OFF/ON
K113
1 XRAY
W503
F3 (0.5A)
X6
230V, 100W
Sensor T3
2 1
VCC
F1 (5A)
4
X3 W503
5
+15V
X1
Power Supply
Gen. Purpuse Out GND24
1 2 3 4
not used
1
1
Sheet No. 4
RevisionNo.00 Description
A
B
External connections
Part-No.
Art
57 60 074
EFS
Teil
Version
01S
00 C
Type-No.
X2287
s A
C
B
4
4
3
3
2
2
1
1
Only for internal use appointed. All rights reserved.
Copyright © SIEMENS AG All rights reserved.
Author: Date: Schnapperelle CO PTS 02.08.05
Sheet No. 5 Description
RevisionNo.00 A
B
Schematic Polydoros IT 30 / 55
Part-No.
Art
57 60 074
EFS
Teil
Version
01S
00 C
Type-No.
X2287
s A
C
B
4
4
3
3
2
2
1
1
Only for internal use appointed. All rights reserved.
Copyright © SIEMENS AG All rights reserved.
Author: Date: Schnapperelle CO PTS 02.08.05
Sheet No. 6 Description
RevisionNo.00 A
B
Schematic Polydoros IT 65 / 80
Part-No.
Art
57 60 074
EFS
Teil
Version
01S
00 C
Type-No.
X2287
RevisionNo.00
A
1
B
Description
T1
Clamp
4-5-6
1-2-3
T 3,15A
F2
C
400VA
T3
F33 T 6,3A F31 T 0,8A
Schnapperelle CO PTS 02.08.05
T 3,15A
F1
Author: Date:
L3
L1
480V
440V
Voltage
P 15V A
Part-No.
Art
Teil
Version
57 60 074
EFS
01S
00
C Poti 5V
(1) (3)
T 1,25A
F3
+24V
X21
(6)
X4
T 1A
T 5A (5)
F4
F2
-15V
X13
+15V
+15V (7) GND
X12
T 5A
F1 (8)
+5V
GND (7) (9)
+2,5V +3,5V
D501
X17
ribbon cable 16 pole
ribbon cable 50 pole
X3
X2
D115
D470
+15V (47) -15V (13) GND
15V / 50kHz
15V / 50kHz
X20
X21
X1
B
+15 V +5V -15V GND
X4
(4)
(3)
(2)
(1)
Power Supply
X3 230V AC
23VAC
X3
X2
X16
+2,5V +3,5V
Only for internal use appointed. All rights reserved.
X1
N 15V A VCC
+5V (1) GND (2) +15V (3) -15V (4)
3
X1
X2
D500
4
Connection
X1
+15V (1) GND (2)
X10
2
X10
Copyright © SIEMENS AG All rights reserved.
+5V (25) GND +15V (30) -15V (26)
A
D-SUB plug 15 pole
D510
s C
4
3
2
1
Sheet No. 7
Power supply
X2287
Type-No.
s B
C
1
S3 Hardware configuration X11
V4
LEDs Workstation Selection
OFF
ON 4
LEDs Dominat
X16
V6
X4
X5
V45
V3
A
V49 V54
4
V7
Test Points IONTOMAT
8
LEDs Flex
D300 MCB3
ON
V21
Dongle D502 (Option)
3
1
Only for internal use appointed. All rights reserved.
S1 Service
(further info see page "Dongle" )
OFF
D300/X11 PIN 3 A-B for Download SW from SPC
3
8
Test points Tube unit current
Service PC or Canon PC connection
VCC
Ground P_15_A
D500 part no.: 77 59 819
Copyright © SIEMENS AG All rights reserved.
X1
X15 V230
2 Test point Starter, kV controller
X302 3 2 1 Jumper Hardware configuration 1
2
OFF ON
X12
1
S4 Ignition
Test point mA_Meas
X110
X 311
X186
Test point Voltage not on partno.: 5760165
X210
X13
X221
Schnapperelle CO PTS 02.08.05
2 X14
X301 Jumper Hardware configuration
X100
X150
Author: Date:
N_15_A
123
X 319
P_15_D
X300 1 2 3 Jumper Hardware configuration
Test points Heating
V67
S2 mA_Measureswitch 1
X10
Sheet No. 8 Description
RevisionNo.00 A
B
D500 Circuit board 77 59 819
Part-No.
Art
57 60 074
EFS
Teil
Version
01S
00 C
Type-No.
X2287
s B
C
LEDs Dominat
1
S3 Hardware configuration X11
V45
LEDs Workstation Selection
OFF
ON
4
V4
X16
V6
X4
X5
V3
A
V49
4
V54
V7
Test Points IONTOMAT
8
LEDs Flex
D300 MCB3
Dongel D502 (Option)
1
Only for internal use appointed. All rights reserved.
S1 Service
(further info see page Dongle)
3
OFF
3
D300/X11 PIN 3 A-B for Download SW from SPC
ON
V21
8
Test points Tube unit current
Service PC or Canon PC connection
Ground Copyright © SIEMENS AG All rights reserved.
D500 part no.: 57 60 165 X1
X15
Test points Heating
2
2 Test point Starter, kV controller
2
X110
Test point mA_Meas
OFF ON
1
X186
S4 Ignition
X210
X13
X221
X150
Schnapperelle CO PTS 02.08.05
X14
X100
Author: Date:
V67
S2 mA_Measureswitch
1 X12
1
X10
Sheet No. 9 Description
RevisionNo.00 A
B
D500 Circuit board 57 60 165
Part-No.
Art
57 60 074
EFS
Teil
Version
01S
00 C
Type-No.
X2287
s B
A
C
LEDs
4
FLEX Logistical component not programmed; no function of board is possible Flashes when D500 is ready Heating inverter start signal RAC start signal Inverter start signal Radiation display (UTact>20kV) Radiation request RAC -Error Error in the primary inverter Error in heating inverter Error with power supply voltages Over-voltage error (+/- kV > 82kV) Service switch switched on; generator goes into service mode Pre-heating (both focuses are heated) active Small focus active
V7 V8 V9 V10 V11 V12 V13 V14 V15 V16 V17 V18 V19 V20 V21
red green green green green yellow yellow red red red red red yellow yellow yellow
Flex Fault Cycle LED Start FIL Inv Start RAC_Inv Start Inv XRAY_ON SE RAC_Inv_Fault Inv_Fault FIL_Inv_Fault Supply_Volt UT_MAX Service Mode Pre Head Small Focus
V4 V6 V3
yellow yellow yellow
Dominat If inverting is active: left dominant selected, otherwise the right one (Change after prep.) Middle dominant selected (Change after prep.) If inverting is active: right dominant is selected, otherwise the left on (Change after prep.)
V45 V49 V54
yellow yellow yellow
AP 0 AP 1 AP 2
Workstation Selection Workstation Selection 0: off for Channel A (table), on for Channel B (Bucky W.S.) Workstation Selection 1: always off Workstation Selection 2: always on
V67
yellow
mA_Meas
mA_Meas mA_Meas switch in position 2
V230 V231 V232 V233 V234
yellow yellow yellow yellow yellow
3,3V P15V_A P15V_D N15V_A VCC
3,3V P15V_A P15V_D N15V_A VCC
3
4
3
Voltage Only for internal use appointed. All rights reserved.
Switches and Jumper Copyright © SIEMENS AG All rights reserved.
S1
S2
2
S4
1 Author: Date:
2, 3, 4
Schnapperelle CO PTS 02.08.05
5 6 7 8
X300 X301 1
X302
S1 Service Normal condition Generator goes into the service mode after a restart
OFF ON
mA_Meas Position 1 LED V67 off -> Normal mode Position 2 LED V67 on -> tube current can be measured directly between 1 and 2 at test point X221 OFF ON
Ignition Ignition pulses not transmitted to inverter; no high voltage Normal condition
OFF ON ON OFF ON ON OFF ON ON
S3 Hardware configuration With Single Tank With high voltage transformer Must be ON With power storage unit No power storage unit Must be ON With MDM Detector No MDM Detector Must be ON
1 and 2 2 and 3 1 and 2 2 and 3 1 and 2 2 and 3
POLYDOROS POLYDOROS POLYDOROS POLYDOROS POLYDOROS POLYDOROS
2
Jumper (not on board partno.: 57 60 165) IT 65/80 IT (S) 30/55 IT 65/80 IT (S) 30/55 IT 65/80 IT (S) 30/55
1
On Board D300 D300/X11 PIN 3 A-B Just for Download SW from SPC/ see Update Instructions
Sheet No. 10 Description
RevisionNo.00 A
B
Leds / Jumpers D500
Part-No.
Art
57 60 074
EFS
Teil
Version
01S
00 C
Type-No.
X2287
s B
A
new old designation
X170 4
X120
X319 X317 X315 X313 X305 X307 X309 X311
A_GND Vin Vout Fout MP4 MP3 MP2 Start_Inv
VION DL_IN
SWR
C
Test Points Iontomat Analog ground Input of Iontomat U/F converter Iontomat U/F converter voltage output Iontomat drift measurement capability Iontomat U/F converter possible frequency
4
not used not used not used Inverter start signal; for triggering during troubleshooting; low active Voltage Test point (not on board partno.: 57 60 165) Ground Analog Ground Digital Ground + 3,3V (+/- 5%) not adjustable VCC + 5V range (5,0 V - 5,2 V) adjustable + 15V Analog (+/- 5%) not adjustable - 15V Analog (+/- 5%) not adjustable + 15V Digital (+/- 5%) not adjustable
GND A_GND GND10 VCC3.3 VCC P_15V_A N_15V_A P_15V_D
Tube current (IT) 3
Only for internal use appointed. All rights reserved.
Copyright © SIEMENS AG All rights reserved.
X243
ITact
X242 X241 X240
ITnom Threshold A_GND
Actual value of tube current: < 8mA: 1V= 1mA < 80mA: 1V=10mA > 80mA: 1V=100mA Ref. value of tube current (as a reference for falling load) Threshold DAC output for debug Analog ground
X150
A_GND F_Rot I_Max I_Load
Starter, kV controller (RAC) Analog ground Tube unit remanence signal Starter max. current Stator current / 1V = 10A
F_ROT I_ANL
X100
Analog ground U/F converter controller output Inverter control frequency (OutA or OutB) / range: 38-100kHz Control frequency OutA / range: 19-50 kHz
X210
I_Load I_Last I_Load nom I_Load act A_GND
RMS-value of Inverter oscillation current / 1V = 50A Oscillating current reference value of controller: / 1V = 50A Oscillation current actual value / 1V = 50A Analog ground
UZ_IST
X110
UDCact Ucon Upre A_GND
Intermediate circuit voltage / 1V = 100V (565V +/-10%) Controller output with pre-control Controller output without pre-control Analog ground
X186
P_UT N_UT UTact A_GND
KV+_ACT KV-_ACT KV_IST
Positive tube unit voltage / 1V = 10kV Negative tube unit voltage / 1V = 10 KV Composite tube unit voltage / 1V = 20kV Analog ground
I_Fact Y_LIM I_Load
F _RMS
Author: Date:
X140
F_INV 1
3
A_GND U_VFO F_INV OutA
2
Schnapperelle CO PTS 02.08.05
I_T_ACT
2 MAS 1 MAS
X221
RevisionNo.00
Heating (FIL) RMS value of heating current / 1V = 1.5A (Stand By ~2A) depends on Tube type Limited controller output of heating controller Oscillation current actual value / 1V = 1.5A Control frequency of heating inverter (OutA or OutB) / range: 30-100kHz; Signal is low active mA_Meas tube current between X220.1 and X220.2 / mA_Meas Switch has to be in Position 2 tube current between X220.1 and X220.2 / mA_Meas Switch has to be in Position 2
All possible test points that are labeled on the board are listed in the tables. Sometimes the test points are labeled, but there is no signal present on the test points. Only the test points defined under "Designation old" are relevant for service. Otherwise the test points are explicitly listed in the documents. A
2
B
1
Sheet No. 11 Description
Test points D500
Part-No.
Art
57 60 074
EFS
Teil
Version
01S
00 C
Type-No.
X2287
s B
4
X24
X150
D470
X111
X22
X21
X14
X25
1
X31
Main Contactors X19 Console
Test point Voltage
V24 Power Fail 2 V24 Power Fail 1
ON
F1 F2
X3
V33
X1
X2
V32
X4
OFF
F3
D501 part no.: 77 59 835
S1
S2
X33 X112
V24
1
F4
V34
2
Schnapperelle CO PTS 02.08.05
Ground
X102
Author: Date:
3
1
V 31
X20
1
X10
Copyright © SIEMENS AG All rights reserved.
1 Only for internal use appointed. All rights reserved.
X8
3
X34
X40
X5
X17
1
X30
X23
1 2
3
X110
Iontomat C
1
1
Iontomat A
X7
8
X6
4
C
8
Iontomat B
A
2
1
1
Sheet No. 12 Description
RevisionNo.00
A
B
D501 Circuit board 77 59 835
Part-No.
Art
57 60 074
EFS
Teil
Version
01S
00 C
Type-No.
X2287
s B
A
C
4 X7
8
1
X14
D470
X21
X25
X22
X23
X24
1
Iontomat A
X6
8
Iontomat B
4
Main Contactors
X17
X30
3
3
X40
X5
X31 X8 X19 Console
S2
X3
F2 V34
Author: Date:
Ground
X102
F4
Schnapperelle CO PTS 02.08.05
D501 part no.: 57 60 181
F1
V32 V33
X1
X2
ON
F3
OFF
2
X4
S1
V31
V24_Power fail
X20
X33 X112
Copyright © SIEMENS AG All rights reserved.
X10
Only for internal use appointed. All rights reserved.
2
1
1
Sheet No. 13 Description
RevisionNo.00
A
B
D501 Circuit board 57 60 181
Part-No.
Art
57 60 074
EFS
Teil
Version
01S
00 C
Type-No.
X2287
s B
A
C
LEDs (not on board partno.: 5760181) 4
V69 V70 V71 V72 V73 V127 V128
red green green green green green red
V_24V Power Fail_1 V_24V_D VCC V_N15V V_P15V V_24V_2 V_24V Power Fail_2
24V permanent Power +5V VCC Power -15V Power +15V Power 24V for 80kW conductor Power
4
Switches S1
OFF ON
The generator cannot be switched on. Required only in a service situation. The generator can be switched on. Normal switch position.
S2
When working in the generator cabinet, it is sometimes helpful to switch on the generator with the S2 switch. In addition, the generator cabinet can be completely initialized without The control console connected for troubleshooting if the switch is pressed and held.
3
3
Test Points (not on board partno.: 5760181)
Only for internal use appointed. All rights reserved.
Voltage
Copyright © SIEMENS AG All rights reserved.
X123 X121 X119 X117 X115 X113
GND24 V_24V_2 V_24V_Dauer V_P15V V_N15V VCC
24V Supply for 80 kW conductors 24 Volt permanent (+/- 5%) not adjustable + 15V Analog (+/- 5%) not adjustable - 15V Analog (+/- 5%) not adjustable VCC + 5V range (5,0 V - 5,2 V) adjustable
X206 X204 X202 X200
GND MP3 MP2 MP1
not used not used not used
2
2
Author: Date:
External Connection
Schnapperelle CO PTS 02.08.05
1
X6 X7 X8 X14 X19 X22_A X23_B X24 X25 X30 X31 X32
Lamps 230V AC Lamps 24V DC Door contact / Tube temp Dose Area Product Unit Control Console IONTOMAT A IONTOMAT B KK1B. G1/G2 KK2.G1/G2 G1 OEM G2 OEM Xray_End
KermaX Bucky table Bucky wall Stand Tomo unit
1
Sheet No. 14 Description
RevisionNo.00
A
B
LEDs / Switches D501
Part-No.
Art
57 60 074
EFS
Teil
Version
01S
00 C
Type-No.
X2287
s C
B
A
I_max_FIL
X3.5 Driver A
4
4
X3.1
X6.6/5 RTN_FIL
D500 X13 Driver B
X6.3 FIL_Small
X3.3 X6.1 FIL_Large L= 49µH
I_max_FIL
L= 2*3,3µH
High voltage acquisition D532 / D530 .X1 FIL Small .X3 FIL Large
X3.5 X3.13
Test point: X140: FIL / I_load
X3.15
Swing current acquisition
X4.1 UDC_CEN X4.3 UDC_NEG
D510 X3
X4.6 UDC_POS
3
3
Snubber circuit VCC X1.1 / X2.1 UHF
D115 X21 D501 X17
Only for internal use appointed. All rights reserved.
X1.3 / X2.3
GND
Copyright © SIEMENS AG All rights reserved.
Standby: Test point D500.X140: FIL / I_load (1V 1,5A) 2
2
1
1
Author: Date: Schnapperelle CO PTS 02.08.05
Sheet No. 15 Description
RevisionNo.00
A
B
D470 Filament
Part-No.
Art
57 60 074
EFS
Version
Teil
01S
00 C
Type-No.
X2287
s B
A
C
D510.X3 4 X5 U_DC + Driver 4
X20.41 Driver 5 X20.39 X20.43
ANL 5
Gate5
Driver 3
D500.X12 X20.37 X20.45
Test point: X150: RAC / I_load
X21
D470.X1 3
ANL 3
X21
ANL 1
ANL 4
*
Gate4
U
X7
V
X8
W
X9
terminal blocks AN_ I
Driver 2 ANL 2
Gate2
AN_0
Gate3
Driver 0
Driver 1 X20.35
4
X6 U_DC -
ANL 0
Gate0
AN_II
Gate1
+15 V GND
Current acquisition
3
Only for internal use appointed. All rights reserved.
Copyright © SIEMENS AG All rights reserved.
2
2
Author: Date: Schnapperelle CO PTS 02.08.05
*Option 2 phase Stators
1
Sheet No. 16 Description
RevisionNo.00
A
B
D115 Rotation Anode Circuit RAC
Part-No.
Art
Teil
Version
57 60 074
EFS
01S
00 C
Type-No.
X2287
1
s C
B
A
D511
D500.X10 Monitoring U_Bmax, I_Load, U CE, P15V, N15V, VCC, auxiliary supply (50kHz)
L_par
ϑ
Heatsink
ϑ
X10 X20
X21
4
X10
X23
X22
X34
X35
Snubber circuit
4
U_DC + I_LOAD U_Bmax
Driver B
Driver A C_s
L1 L2 L3
Out_B
Out_A
Driver A
L1
Driver B
U_DC -
Charging
X44
L3
X46
X32
X33
D500 X14 3
Only for internal use appointed. All rights reserved.
3
Test Points X210: I_Load X186: P_UT X186: N_UT X186: UTact
X13
X7
X240: IT_act
X15
D530
Copyright © SIEMENS AG All rights reserved.
kV + kV -
HVTransformer 2
2 Operating point 81kV 100mA Test Point D500.X210: I_Load
Operating point 100kV 500mA Test Point D500.X210: I_Load
Author: Date: Schnapperelle CO PTS 02.08.05
1
1
Sheet No. 17 Description
RevisionNo.00
A
B
D510 Inverter IT30/55
Part-No.
Art
57 60 074
EFS
Version
Teil
01S
00 C
Type-No.
X2287
s B
A
C
D511
D500.X10 Monitoring U_Bmax, I_Load, U CE, P15V, N15V, VCC, auxiliary supply (50kHz)
TF 1
TF 2 L_par
ϑ
Heatsink
ϑ
X10 X20
X21
4
X10
X23
X22
X34
X35
Snubber circuit
4
U_DC + X40
I_LOAD U_Bmax
Driver B
Driver A C_s
L1 L2 L3
Out_B
Out_A
Driver A
Driver B
X42
L1
U_DC -
Charging
X44
L3
X46
X32
X33
D500 X14
3
K10
Test Points X210: I_Load X186: P_UT X186: N_UT X186: UTact
3
X240: IT_act Only for internal use appointed. All rights reserved.
X13
X7
X40
X4
X42
X5
X34
X1
X44
X2
CBuffer
X15
D580 80kW Opt ion
D530
D501
Copyright © SIEMENS AG All rights reserved.
CS
D501. X34
kV + kV -
HVTransformer 2
Operating point 81kV 100mA Test Point D500.X210: I_Load
2
Operating point 99kV 800mA Test Point D500.X210: I_Load
Author: Date: Schnapperelle CO PTS 02.08.05
1
1
Sheet No. 18 Description
RevisionNo.00 A
B
D510 Inverter IT 65/80
Part-No.
Art
57 60 074
EFS
Teil
Version
01S
00 C
Type-No.
X2287
s B
A
C
Single Tank Tube Assembly
GENERATOR 4
4
Shield
0
D115.X8
I
D115.X7
II
D115.X9
M16 PE
D470.X6
D500.X14
3
M
PE
heating
X1 / 3
Temp.
temparature / pressure switch
X7
pressure 3
X11
X32
D510
X33
X13
Only for internal use appointed. All rights reserved.
Copyright © SIEMENS AG All rights reserved.
2
2
1
1
Author: Date: Schnapperelle CO PTS 02.08.05
Sheet No. 19 Description
RevisionNo.00
A
B
Connection Single Tank
Part-No.
Art
57 60 074
EFS
Teil
Version
01S
00 C
Type-No.
X2287
s C
B
A
OPTITOP
GENERATOR 4
4
Shield
0
K3.R3
I
K5.1
II
K3.R5
M16 PE
M
PE
D501.X8.5 D501.X8.6
Oil pressure switch
pressure
D501.X8.3 D501.X8.4
temparature switch
Temp.
3
3
kV + HV- Tank
kV -
Only for internal use appointed. All rights reserved.
Copyright © SIEMENS AG All rights reserved.
2
2
1
1
Author: Date: Schnapperelle CO PTS 02.08.05
Sheet No. 20 Description
RevisionNo.00
A
B
Connection Optitop-Tube
Part-No.
Art
57 60 074
EFS
Version
Teil
01S
00 C
Type-No.
X2287
s A
C
B
4
4
3
3
2
2
1
1
Only for internal use appointed. All rights reserved.
Copyright © SIEMENS AG All rights reserved.
Author: Date: Schnapperelle CO PTS 02.08.05
Sheet No. 21
RevisionNo.00 Description
A
B
OptionEnergy Storage Unit
Part-No.
Art
57 60 074
EFS
Teil
Version
01S
00 C
Type-No.
X2287
View more...
Comments