MMP Choice MMED6000DP - MS PDF
January 26, 2023 | Author: Anonymous | Category: N/A
Short Description
Download MMP Choice MMED6000DP - MS PDF...
Description
MMED6000DP/S6-CSN801 Service Manual
Copyright @ 2005 Choice Electronic Technology Co. Ltd. Edition: MMED6000DP-CSN801 Issued date: Aug.2005
MMED60000DP (S6) Service Manual
COPYRIGHT Beijing Choice Electronic Technology Technology Co., Ltd. (hereinafter Ltd. (hereinafter called Beijing Choice) owns all right to this unpublished work and intends to maintain m aintain this work as confidential. Beijing Choice may also seek to maintain this work as an unpublished copyright. This publication is to be used solely for the purpose of reference, operation, maintenance, or repair of Choice equipment. No part of this can be disseminated for other purposes. In the event of inadvertent inadvertent or deliberate publication, Beijing Choice intends to enforce its right to this work under copyright laws as a published work. Those having access to this work may m ay not copy, use, or disclose the information in this work unless expressly authorized by Beijing Choice to do so. All information information contained contained in this publication is believed to be correct. Choice shall not not be liable for errors contained herein nor for incidental or consequential damages in connection with the furnishing, performance, performance, or use of this material. This publication may refer to information and protected by copyrights or patents and does not convey any license under the patent rights of Beijing Choice, nor the rights of others. Beijing Choice does not assume any liability arising out of any infringements of patents p atents or other rights of third parties. Contentt of the manual is subject to changes without prior notice. Conten PROPERTY OF BEIJING CHOICE ELECTRONIC TECHNOLOGY CO., LTD. ALL RIGHTS RESERVED
MMED60000DP (S6) Service Manual
CONTENT Chapter 1 Introduction
5
General
5
Description of abbreviation
5
Explanation Explanatio n of symbols
5
Special feature
6
Appearance of monitor
7
Specification
13
Chapter 2 Special operation in menu
17
Chapter 3 Monitor function and principle
19
Power supply unit
19
CSN801 board
19
Chapter 4 Component and structure Component
21 21
Structure Chapter 5 Troubleshooting Troubleshooting Diagram of solutions to various troubles Troubleshooting Error Message
21 30 30 40 43
Chapter 6 Calibration of monitor
44
Required testing tool
44
Power on the monitor
44
Front panel test
44
Audio system system test
45
Test of automatic storage
45
Alarm test
45
ECG test
46
Respiration test
47
Temperature test
47
NIBP test
47
SPO2 test
49 50
Chapter 7 Adjustment procedure
MMED60000DP (S6) Service Manual
ECG adjustment NIBP adjustment SpO2 adjustment IBP Calibration ET-CO2 calibration Chapter 8 Disassembly procedure Monitor Disassembly Battery Removal Chapter 9 Cleaning and Disinfection Maintenance check General cleaning Sterilization Precondition and cleaning TEMP sensor cleaning and disinfection SpO2 sensor cleaning and disinfection Chapter 10 Warranty Warranty and repair Exemption and restriction Customer guarantee guarantee Non-warranty and non-replacement policy Customer special warranty period Repackaging
50 50 50 52 54 55 55 55 57 57 57 58 59 59 60 61 61 62 62 62 62 62
MMED60000DP (S6) Service Manual
CHAPTER 1 INTRODUCTION 1.1 General MMED6000DP is a Portable Patient Monitor that has abundant monitoring functions and is used for the clinical monitoring of adult, pediatric and neonate. Besides, the user may select the different parameter configuration according to different requirements. MMED6000DP MMED6000 DP can be connected to the central working station to form a network monitoring system. MMED6000DP can monitor vital signals as ECG, RR, SpO2, PR, NIBP, TEMP, Dual-channel IBP (optional), CO2 (optional). It integrates parameter measuring modules, display and recorder in one device, featuring in compactness, lightweight and portability. Large true color display provides clear view of 6 waveforms w aveforms and full monitoring paramete parameters. rs.
1.2 Description of Abbreviation HR: SPO2: NIBP: IBP: S: M: D: PR: RR: TEMP1: TEMP2: CO2:
heart rate 2-channel ECG waveform arterial oxygen saturation non-invasive blood pressure invasive blood pressure systolic mean blood pressure diastolic pulse rate respiration rate tempe mperature channel 1 tempe mperature channel 2 end-tidal CO2 concentration
1.3 Explanation of Symbols Attention Refer to the relevant the prompt. Resistant defibrillator BF type equipment
Class II Equipment Heart Beat Detected
Power on/off
Rotated knob Counter-clockwise and clockwise
Alarm on
Alarm off
Sound on
Silence
MMED60000DP (S6) Service Manual
System setup m enu
Trend menu
Equipotential grounding terminal
AC 50/60 HZ
Fuse
This symbol indicates that the instrument is IEC 60601-1 Type CF equipment. The unit displaying this symbol contains an F-type isolated (floating) p patient atient applied part providing a high degree of degree of protection against shock, and is suitable for use during defibrillation. This item is compliant with Medical Device Directive 93/42/EEC of 14 June 1993,, a directive of the European Economic Community. 1993
1.4 Special Features l
Portable, compact, AC power and internal rechargeable rechargeabl e battery;
l
Automatically Automatic ally and rapid charge, 80-90% power will be recharged within 3 hours;
l
Resistant high-frequency high-frequency electrotome electrotome design, design, reliable and special special module module is used used in operation room;
l
Optional inner printer with 3 channels channels wave or external printer;
l
Adjust volume more accurately by digital system;
l
Menu design adopts Huffman decode, decode, o operating perating more effectively;
l
Support Ethernet, Ethernet, wireless LAN and could connect with Central Central monitoring system.
MMED60000DP (S6) Service Manual
1.5
Appearance of Monitor
1.5.1. Screen display u 4 channels
(1)
(2)
(4)
(3) (6)
(5)
Fig.1-1 The display of MMED6000DP patient monitor is TFT LCD, which displays the patient
parameters, waveforms, historical data and monitor status. The screen is divided into five areas: (1) monitor status, (2) waveforms, (3) the table of historical data, (4) parameters (5) menu as well as (6) frame of menu status.
Fig. 1-2
MMED60000DP (S6) Service Manual
l
l
Monitoring status frame Bed No.: indicates the bed number of the patient being monitored when the patient monitor is connected with the workstation. DEMO: indicates the device is in DEMO monitoring mode now. In clinical application, this function is not recommended because the DEMO will mislead the hospital workers to treat the waveform and parameter as actual data of the patient, which may result in delay of treatment or mistreatment. Start time: indicates the time powered on the monitor, it shows in “16:25”. Current time: indicates the current date and time, it shows in “05/16/2004, 10:10” Waveforms displaying area The waveform area can maximally display 6 waveforms. The colour of waveforms matches with the parameter’s on the right of the screen. For the maximum configuration, the waveforms provided by the system for selection are (from up to bottom): 2-lead ECG waveforms SpO2 waveform RESP waveform There are two means means to gain RESP RESP waveform, trans-thoracic impedance impedance and Et-CO2 (optional). If operator selects Et-CO2 (side stream) to monitor the patient’s respiration, respiration, the CO2 waveform will instead of the RESP waveform acquired by Trans-thoracic impedance. 2 IBP waveforms(option) waveforms(option) If operator selects IBP operation, the table of historical data will be instead of 2 IBP waveforms.
l
Numeric displaying displaying area The parameter area lies to the right side of the graphic g raphic area, which includes: includes: ECG: heart rate (unit: beats/minute) NIBP: from left to right, there are Systolic pressure, Mean pressure and Diastolic
pressure (unit: mmHg or kPa) IBP: the BP of channel 1 and 2. From left to right, there are Systolic pressure, Mean pressure and Diastolic pressure (unit: mmHg or kPa) SpO2: SpO2(unit: %) pulse rate (unit: beats/minute) RESP: respiration rate (unit: breaths/minute) TEMP: temperature of channel 1 and 2 (unit: Et CO2: unit: mmHg or kPa
or
)
l
Menu The menu always occupies the fixed position on the bottom of the screen, user can set the system and perform operation.
l
Frame of menu status
MMED60000DP (S6) Service Manual
Silence on/off: showing the status of silence on or off
Alarm on/off: shows the status of alarm on
or off
,
Trend: the function is for operator observing the patient’s latest 72 hours change, including historical data table, HR, NIBP, SpO2, ST segment.
System setup: you can configure various aspects of the monitor, including system time, simulation, print setup, color, display wave, language, color, etc.. Menu bar: it shows the different menus of every operation.
Other modes: IBP EXT:
ECG EXT:
MMED60000DP (S6) Service Manual
1.5.2
Function buttons
(1) (2) (3) (4) (5) (6)
(7)
(8)
Fig.1-3
1 Alarm ALARM button, pressing the button will prohibit all the technically audio alarm and physically audio alarm for 3 minutes. The audio alarm function will be restored automatically after 3 minutes or be activated when new alarm occurs. As pressing the
button, the icon in menu status shows
which indicates all the audio alarms
have been shut off.
NOTE When “
” mark appears, the system can not give out the audio alarm
prompt. Therefore, the operator should use this function c carefully arefully..
2 Silence SILENCE button, whe when pr pre essin sing the bu buttton, thi this
mar mark app appears in th the men menu
status, indicating that all kinds of sound including including the audio alarm and heart beep have been manually muted until the button has been pressed again, the system will imme mmediately re restore th the nor normal mal sta status,
and th the e ico icon sho shows as
NOTE When pressing “SILENCE”, the system cannot give out the audio alarm and heart beep; therefore, the operator should be considerate in using the function.
3 Freeze FREEZE button, pressing the button can freeze the waveform of ECG. The screen displays two traces of ECG waveform. When one trace is active, pressing the FREEZE
MMED60000DP (S6) Service Manual
button will freeze another trace. Pressing “FREEZE” button again will restore the normal monitor status.
4 Start/Cancel START/CANCEL ST ART/CANCEL BP measuring measuring button, pressing the button will inflate the cuff to start a new NIBP measurement. When measuring, press it to cancel the measurement and deflate the cuff, including the “CYCLE” and “STAT (NOT STOP)” measuring mode.
5 Print PRINT button, pressing this button will w ill motivate the recorder or d desktop esktop printer to output the result if the monitor is equipped with them. (6) Menu Main MENU button, pressing the button will exit the submenu and refresh the screen.
7 Rotary Knob
The operator operator uses the rotary knob to select the menu item and modify the setup. It can be rotated clockwise or counter-clockwise and pressed like other buttons. The operator uses the knob to realize the operations.
The method of using the knob to execute the operation:
Fig. 1-4
Rotary knob is just like the cursor of computer. When operator rotates the knob on the icon where the operation is wanted, the icon will be automatic highlight. Then pressing the knob, operator will open the setup menu of the corresponding parameter so as to set up the menu. (8) Power Supply Indicator
Indicating the power supplying state of the monitor
Orange: 220Vac main power supplying
Orange and flash: 220Vac main power supplying and battery charging
Green and flash: internal battery power supplying and 50% power has been consumed.
1.5.3. Side Panel
MMED60000DP (S6) Service Manual
(2) (1)
(3)
(4) (7)
(5)
(6)
Fig.1-5 (1) ECG socket connect ECG 5-lead wire with 6 PIN connector
(2) NIBP socket conne connect ct NIBP cuff with extension extension tube (3)(4) IBP IBP (IBP1, (IBP1, IBP2) socket:
Single, Dual channel channel (optional) (optional)
(5)(6) Temperature Temperature socket: connecting connecting the the TEMP probe probe (7)SPO2 socket conne connect ct the SpO2 sensor with wi th extension cable (8) Et CO2 socket connect the nasal cannula with connector.
(8)
Fig.1-6
Fig.1-7
1.5.4 Recorder
Fig. 1-8
MMED60000DP (S6) Service Manual
A thermal dot matrices recorder r ecorder with standard 50mm (+1/-1) wide printout p rintout paper is used for MMED6000DP portable patient monitor. NOTE The thermal array recorder should be installed by the adequate technician. technician.
1.5.5 Rear panel
(1)
(2) (3) (5)
(6) (4) Fig.1-9
(1) Power switch
(2) NET: Internet access point: connecting with the central monitoring system through the standard RJ-45. (3) CRT: exteriorly connecting screen and standard VGA. (4) Printer: exteriorly connecting printer, which type is designated by the manufacturer. (5) AC Power supply: supply: 100-230 100-230 (VAC), (VAC), 50/60(Hz); Fuse: Fuse: standard –T 1.6A (6) Equivalent electric ground access for connecting with the hospital’s grounding system.
1.6 Specification Environment The operation environment should comply with the following f ollowing conditions: Operating Temperature: 5 to 45 Relative Humidity: 0 to 80%, non-condensing Height of sea level: -500m ~ 4600m The transport and storage environment should comply with the following conditions: Storage Temperature: -20 to 70 Relative Humidity: 0 to 93%, non-condensing Height of sea level: -500 m~ 13100m
MMED60000DP (S6) Service Manual
Display Type: Screen Size:
640X480 pixel color TFT 10.4 inch diagonal
Di s pla played yed P ara arame mete ters rs Time: Alarms:
Battery-backed quartz crystal clock
ECG: Heart Rate: NIBP: Pulse Oximet meter: Respiration Rate: Trends:
ECG Waveform Scale, displayed lead Derived from ECG OR Spo2 Pressure (systolic, mean and diastolic) Pulse Rate, Pul Pulse wave wavefo forrm, and percent saturat ration. Respiration rate derived from ECG. Heart rate, respiration rate, NIBP (systolic, mean, diastolic), Temperatures and SpO2 Tw o channels Traces A or B
Temperatures: Trace Freeze
High and low limits selectable on patient patient parameters
ECG Protected against defibrillator and electro surgery potentials Standard Lead I II III aVL aVR aVF V1 V2 V3 V4 V5 V6 Display Gain Scales 5mm/mV 7mm/mV 10mm/mV 15mm/mV 20mm/mV 25mm/mV Sweep Speed 12mm/s 25mm/s 50mm/s Inpu putt Resistance > 5M Ohm (at 10 Hz, Hz, not includ udin ing g patient cabl ble e) Frequency Re Respo pon nse 0.05Hz-100Hz (3 (3dB) Common Mode >60dB Rejection Ratio Electrode Offset Maximum 0.3V Potential Baseline Recove verry
View more...
Comments