MANUAL DO OPERADOR COULTER T-890 VERSÃO RESUMIDA E TRADUZIDA
APRESENTAÇÃO DO APARELHO COUTER T-890 (VISTA FRONTAL – ABERTO) 11
12
13
14
15
10
9
8
7
Legendas 1 – PAINEL DE MANOMETROS 2 – BOTÕES DE AJUSTE DE PRESSÃO 3 – WHOLE BLOOD – PIPETAR 4 – ESCALA DE LEDS – PRESSAO VÁCUO 5 – BOTAO DE REGULAGEM DO VÁCUO 6 – ORIFICIO BANHO HGB 7 – ORIFICIO BANHO WBC 8 – POWER – BOTAO DE LIGAR
9 – TECLADO FRONTAL 10 - DISPLAY 11 – BANHO WBC 12 – BANHO HBG 13 – MANOMETRO VÁCUO 14 – MANOMETRO 30 PSI 15 – MANOMETRO 05 PSI 16 – MANOMETRO 60 PSI
16
INSTRUÇÕES SOBRE O EQUIPAMENTO – COUTER SERIE T (T-890) (T- 890)
TECLAS DO PAINEL FRONTAL E SUAS FUNÇÕES DRAIN – ESVAZIA (DRENA) AS CAMARAS DE CONTAGEM RINSE – ENCHE (RINSA) AS CAMARAS DE CONTAGEM COM COM CLEANS LYSE – DISP DISPEN ENSA SA 1,0 1,000 mL DE HE HEMO MOLI LISA SANT NTE E (LYS (LYSE) E)NA NA CAMA CAMARA RA DE CONT CONTAG AGEM EM DE LEUCOCITOS
DIL DISP – DISPENSA 1,00 mL DE SOLUÇÃO ISOTONICA (ISOTON) NA PIPEPA (AGULHA) DE ASPIRAÇÃO
CLEAR – APLICA PRESSÃO DE 05 PSI NAS ABERTURAS E ESVAZIA A CAMARA DE VÁCUO E A CAMARA DE ESGOTO PRINT – IMPRIME O RESULTADO DA ULTIMA AMOSTRA ANALISADA
RES 1 E RES 2 - PRESSIONADAS SIMULTANEAMENTE PROVOCAM UM RESET NO APARELHO – ZERAM QUALQUER MENSAGEM DE ERRO OU FLAG
START UP – PREPARA O APARELHO PARA O INICIO DA ROTINA SHUT DN – PR PREP EPAR ARA A O AP APAR AREL ELHO HO PA PARA RA SE SER R DE DESL SLIG IGAD ADO. O. RE REAL ALIZ IZAR AR ESTA ESTA FU FUNÇ NÇÃO ÃO SOMENTE NO FINAL DA ROTINA
XMT – SOMENTE É USADO QUNDO O APARELHO ESTÁ CONECTADO A UM COMPUTADOR ESTERNO OU INTERFACIADO A OUTRO PROGRAMA EXTERNO
F – ACIONA AS FUNÇÕES INDIVIDUAIS TEST – USADO PARA TESTE DO BACKGROUD DOS REAGENTES. DIGITA-SE TEST 01, ENTER DIR – – ENTRA NO DIRETORIO ADV – AVANÇA PELO DIRETORIO SEL – SELECIONA OPÇÕES NO DIRETORIO.
I – INICIALIZANDO O APARELHO 1 – LIGUE O APARELHO PRESSIONANDO A TECLA POWER 2 – LIGUE A IMPRESSORA 3 – VERIFIQUE OS REAGENTES 4 – AGUARDE ALGUNS MINUTOS (aproximadamente 30 minutos – tempo aquecimento) 5 – VERIFIQUE OS MANOMETROS
POSIÇÃO NOMAL DOS MANOMETROS DA ESQUERDA PARA A DIREITA -
VÁCUO: ACIMA DE –20 POLEGADAS (NÃO AJUSTÁVEL) 30 PSI: + OU – 0,2 PONTOS (AJUSTAVEL NO BOTÃO ABAIXO) 05 PSI: + OU – 0,2 PONTOS (AJUSTAVEL NO BOTÃO ABAIXO) 60 PSI: + OU – 5,0 PONTOS (NÃO AJUSTAVEL) BARRA DE LEDS: 4 OU 5 ACESOS, AJUSTAR NO BOTAO ABAIXO DA BANDEJA
6 – DIGITE A DATA E PRESSIONE ENTER 02 02 VEZES 7 – PRESSIONE START UP 8 – AGUARDE ATÉ AIMPRESSÃO DO “SYSTEM TEST” 9 – VERIFIQUE VERIFIQUE O “BACKGROUND” (IMPRESSÃO DO START UP) -
WBC: MAXIMO 0,3 RBC: MAXIMO 0,03 PLT: MAXIMO 3,0
OBS.: CASO OS VALORE SEJAM SUPERIORES A ESTES REPITA O TESTE DE BACKGROUND: PRESSIONE TEST 01 ENTER.
II – INSTRUÇÕES PARA INICIO DOS TRABALHOS – CONTAGEM CELULAR 1 – SE O NUMERO DE AMOSTRA ANTES DO “START UP”, DIGITE SEGUINDO A SEQUENCIA: DIR, SEL, ENTER, NÚMERO DA PRIMEIRA AMOSTRA, ENTER. OBS.: O APARELHO SELECIONA E IMPRIME AUTOMATICAMENTE O NUMERO DA AMOSTRA EM SEQUENCIA QUANDO DIGITADO DESTA FORMA. 2 – COLOQUE A AMOSTRA HOMOGENEIZADA NA PIPETA (AGULHA) E PRESSIONE A TECLA WHHOLE BLOOD, ATRAS DA PIPETA 3 – QUANDO VOCE OUVIR O BIP DO APARELHO (NO DISPLAY APARECER: WIPE TIP), RETIRE A AMOSTRA E LIMPE COM CUIDADO A PIPETA. 4 – APÓS A IMPRESSAO DOS RESULTADOS, COLOQUE A OUTRA AMOSTRA E REPITA OS PASSOS ANTERIORES.
III – INSTRUÇÕES PARA DESLIGAR O EQUIPAMENTO: 1 – LIMPAR O ESGOTO DA PIPETA DIGITANDO F10 – ENTER INJETE INJETE 10 mL DE ÁGUA SANITARIA PELO ORIFICIO DA LIMPEZA DA PIPETA. 2 – PARA ENCERRAR E FECHAR O BRAÇO DA PIPETA, PRECIONE JUNTO RES 1 E RES 2 E AGUARDE ATÉ APARECER TEST Nº 001 NO DISPLAY E PRESSIONE ENTER. LOGO APÓS APERTE CLEAR. 3 – PRESSIONE A TECLA SHUT DN. 4 – AGUARDE ATÉ O COMPRESSOR DESLIGAR. 5 – LIMPE COM CUIDADO AS PARTES EXTERNAS (PIPETA, BANDEIJA, TAMPA) COM GAZE UMEDECIDA COM ÁGUA DESTILADA. 6 – DESLIGUE O APARELHO E A IMPRESSORA.
IV – INSTRUÇÕES PARA TROCA DE REAGENTES 1 – SEMPRE QUE ALGUM DOS REAGENTES ACABAR, O APARELHO MOSTRARA NO DISPLAY
CHANGE UNI-T-PACK 2 – PRESSIONE SIMULTANEAMENTE AS TECLAS RES 1 E RES 2 3 – AGUARDE APARECER NO DISPLAY TEST Nº 001 E PRESSIONE ENTER 4 – SUBISTITUA O REAGENTE 5 – DIGITE F 14 E ENTER , AGUARDE A IMPRESSAO DO “BACKGROUD”. 6 – CASO O REAGENTE FOR TROCADO ANTES DO APARELHO ACUSAR, FAZER SOMENTE O ITEM 5.
V – INSTRUÇÕES PARA CALIBRAÇÃO DO APARELHO O APARELHO ACEITA A CALIBRACAO DE 5 PARAMETROS: WBC, RBC, HGB, MCV E PLT O HCT É CALCULADO A PARTIR DOS PARAMETROS RBC E MCV A ACLIBRACAO É FEITA COM FATORES QUE FICAM ARMAZENADOS NA MEMORIA.
APÓS O “START UP”, COM UM RESULTADO DE BACKGROUND ACEITÁVEL, PODE SER INDICADA A CALIBRAÇÃO.
1 – DIGITE F 13 (PARA IMPRIMIR OS FATORES DE CALIBRAÇÃO ATUAIS) 2 – PASSAR UM SANGUE CONTROLE PARA AMBIENTAR O EQUIPAMENTO OBS.: SEGUIR AS INSTRUÇÕES DE USO DO SANGUE CONTROLE O SORO CONTROLE PODE SER SUBSTITUIDO POR UM SANGUE HUMANO NORMAL 3 – PASSAR 5 VEZES O MESMO SANGUE CONTROLE (DE PREFERENCIA “NORMAL”) 4 – CONFERIR OS RESULTADOS COM A TABELA DE REFERENCIA, NO CASO DE CONTROLE COMERCIAL. 5 – CALCULAR O NOVO FATOR
4,40 (RESULTADO TABELA) 1,000 (FATOR ATUAL)
FÓRMULA FATOR ATUAL X RESULTADO TABELA (MEDIDO) = NOVO FATOR RESULTADO ATUAL
1,000 X 4,40 = 1,047 (NOVO FATOR)
4,20 6 – DIGITE DIR, ADV, ADV (CALIBRATE), SEL, 890, ENTER, DIGITE O NOVO FATOR NO PARAMETRO QUE NECESSITA SER ALTERADO E PRESSIONE ENTER.
VI – MANUTENCAO PREVENTIVA MENSAL VI – I – PROCEDIMENTO DE LIMPEZA DAS ABERTURAS E “SAMPLING VALVE”
→ PREPARAR UMA SOLUCAO DE LIMPEZA: 50% ÁGUA SANITÁRIA, 50% ÁGUA DESTILADA. 1 – PRESSIONE A TECLA DRAIN 2 – UTILIZANDO AS ABERTURAS SUPERIORES DOS BANHOS E COM O AUXILIO DE UMA SERINGA, INJETE SOLUCAO DE LIMPEZA NOS BANHOS DE WBC E RBC, ATÉ O NÍVEL DE ENTRADA, 3 – DIGITE F 09 + ENTER 4 – APÓS O TERMINO DA FUNCAO, DEIXAR OS BANHOS CHEIOS E DE MOLHO POR 15 MINUTOS.
DURANTE ESTE PERIODO, PROCEDA A LIMPEZA DA “SAMPLING VALVE” 1 – DIGITE F 07 + ENTER 2 – APÓS O TERMINO DA FUNCAO, RETIRE O KNOB PRETO (PARAFUSO PRETO) QUE PRENDE A VALVULA. 3 – REMOVA AS TRÊS PARTES DA VALVULA.
OBS.: NÃO REMOVA AS MANGUEIRAS. 4 – COLOQUE UM RECIPIENTE ABAIXO DA VÁLVULA PARA RECOLHER OS RESÍDUOS DA LIMPEZA. 5 – INJETE CLEANS (SOLUCAO DE LIMPEZA) NOS ORIFICIOS E NA SUPERFICIE DAS TRÊS PARTES DA VALVULA, 6 – UMEDEÇA UMA GAZE COM CLEANS OU SOLUÇÃO DE LIMPEZA E PASSE POR TODAS AS SUPERFICIES. 7 – ENXAGUE AS SUPERFICIES CO AGUA DESTILADA E INJETA NOS ORIFICIOS TAMBÉM AGUA DESTILADA. 8 – SEQUE AS PARTES COM UM LENÇO DE PAPEL OU PAPEL HIGIENICO MACIO 9 – PINQUE UMA COTA DE OLEO MINERAL ESM TODAS AS SUPERFICIES QUE TENHAM CONTATO ENTRE SI E MONTE NOVAMENTE A VALVULA, OBSERVANDO A ORDEM E POSICIONANDO NA GUIA DE ENCAIXE, 10 – RECOLOQUE O KNOB PRETO (PARAFUSO PRETO) APERTENDO POR TRES VOLTAS A PARTIR DO MOMENTO EM QUE ELE INICIAR O APERTO, 11 – DIGITE F 05 + ENTER , A VALVULA VAI GIRAR PARA A DIREITA 12 – DIGITE F 06 + ENTER, A VALVULA VAI GIRAR PARA A ESQUERDA 13 – LIMPEZA DA VALVULA CONCLUIDA
RETORNE PARA OS PRECEDIMENTOS DE LIMPEZA DAS ABERTURAS E DOS BANHOS 5 – PRESSIONE F 10 + ENTER 6 – COM O AUXILIO DE UMA SERINGA, INJETE HIPOCLORITO DE SÓDIO A 2% NO ESGOTO DA PIPETA (BRAÇO DA PIPETA) ATÉ A CAMARA DE RESTO CHEGAR A METADE (CAMARA SITUDA A ESQUERDA DA PIPETA, ABAIXO DOS BANHOS) 7 – PRESSIONE RES 1 E RES 2 SIMULTANEAMENTE E AGUARDE O BRAÇO SER RECOLHIDO E PRESSIONE ENTER 8 – PRESSIONE CLEAR, E AGUARDE 9 – PRESSIONE DRAIN, E AGUARDE 10 – UTILIZANDO AS ABERTURAS SUPERIORES DOS BANHOS E COM O AUXILIO DE UMA SERINGA, INJETE ÁGUA DESTILADA NOS BANHOS DE WBC E RBC, ATÉ O NÍVEL DE ENTRADA, 11 – DIGITE F 09 + ENTER 12 – PRESSIONE DRAIN
13 – REPITA OS PASSOS 10, 11, 12 POR TRES VEZES 14 – DIGITE F 08 + ENTER 15 – APÓS O TERMONO DESTA FUNÇÃO, PRESSIONE DRAIN 16 – PRESSIONE RINSE 17 – PRESSIONE SHUT DOWN 18 – APÓS O TERMINO DESTA FUNÇÃO, AGUARDE 15 MINUTOS, 19 – APÓS ESTE PERIODO, ATIVE OS COMPRESSORES PRESSIONANDO A TECLA CE 20 – PRESSIONE START UP 21 – AO TERMINO DESTA FUNÇÃO, VERIFIQUE SE OS VALORES DE BACKGROUD ESTAO DENTRO DOS LIMITES. 22 – PARA RESINICIAR A ROTINA, CILCE UMA AMOSTRA DE SANGUE QUALQUER A FIM DE AMBIENTAR (CARREAR) O APARELHO.
VII – MENSAGENS DE ERRO – MAIS COMUNS
PNEUMATIC LOW: BAIXA PRESSAO PNEUMATICA – VERIFICAR O MANOMETRO DE 30 PSI CHANGE UNI-T-PACK: FIM DE REAGENTE – VERIFICAR O FIM DE ALGUM REAGENTE OU SE HÁ BOLHAS DE AR NA TUBULAÇÃO – PROCEDER COMO INSTRUI A TROCA DE REAGENTES
UNI-T-PACK LOW ADV TO CONTINUE: CONTADOR INTERNO DE REAGENTES INDICA QUE A BOMBONA DE SOLUÇÃO ISOTONICA ESTÁ CHEGANDO AO FIM, PROCEDER COMO INSTRUI A TROCA DE REAGENTES
PRINTER NOT READY: IMPRESSORA NÃO RESPONDE, VERIFICAR SE A IMPRESSORA ESTA LIGADA, SE NÃO HÁ FALTA DE PAPEL OU SE ESTE NÃO TRANCOU.
+15 OU –15 OU +5 OU +24 HIGH OU LOW: PROBLEMAS ELETRONICOS COM A FONTE DE ALIMENTÇÃO. DESLIGAR O APARELHO E AGUARDAR ALGUNS MINUTOS E TENTAR RELIGAR NOVAMENTE. PERSITINDO O PROBLEMA, CONTATE A ASSISTENCIA TECNICA.
VIII – CODIGOS DE ALARMES – FLAGS
+++ - OCORRE QUANDO HÁ UM VALOR MUITO ALTO DE ALGUM DOS PARAMETROS DE CONTAGEM. VALORES MUITO DIFERENTES ENTRE AS TRES - - - - REJEIÇÃO CONTAGENS.
L OU H – USADO QUANDO A CONTAGEM ESTÁ MUITO BAIXA OU ALTA (LIMITES DEFINIDOS PELO OPERADOR – PROGRAMADOR) + - RESULTADO MUITO ALTO * - ALARME PARA VERIFICAR VERIFICAR RESULTADOS (ALGUM PARAMETRO ESTA FOR FORA A DA NORMALIDADE)
- CÁLCULOS NÃO REALIZADOS, PROBLEMAS INTERNOS DO APARELHO.
IX – CÓDIGOS DE ALARMES DE PLAQUETAS
R1 – EXCESSIVO NÚMERO DE FRAGMENTOS E INTERFERENCIA ELETRONICA R2 – MÁ DISTRIBUIÇÃO DE PLAQUETAS R3 – HEMÁCIAS MUITO PEQUENAS PODENDO SER CONFUNDIDAS COM PLAQUETAS RM – ALARME EM DUAS OU MAIS REGIÕES.
X – DIRETÓRIO PRESSIONE DIR PARA PARA ENTRAR NO DIRETÓRIO, ADV PARA AVANÇAR OU ALTERAR, SEL PARA SELECIONAR A OPÇÃO.
“SELECT FUNCTION” DIR “SET DATA & TEST” – ENTRAR COM A DATA E NÚMERO DE TESTE ADV “SHOW RESULTS” – MOSTRA NO DISPLAY OS RESULTADOS ADV “CALIBRATE” – ACESSO AOS FATORES DE CALIBRAÇÃO ADV “AUTO-CAL” – CALIBRAÇÃO COM CALIBRADOR COMERCIAL ADV “DVM” MULTIMETRO DIGITAL INTERNO ADV “CONFIGURATION” – CONFIGURAÇÃO PARA PRINT (IMPRESSORA), DATA (DADOS) E SERVICE (SERVIÇOS) ADV “HGB BLANK ADJUST” – AJUSTE DE BRANCO DA HGB (**) ADV “ACTIONS LIMITS” – AJUSTE DE AÇÃO PARA IDENTIFICAÇÃO “L” OU “H” DOS RESULTADOS ADV “TEST NUMBER MODE” – OPERAÇÃO PARA NUMERAÇÃO DE TESTES MANULA OU AUTOMATICO
(**) – AJUSTE DE BRANCO DA HGB ( 7,00 ) DIGITE: DIR, ADV (6X), SEL, 890, ENTER, (7), (0), (0), ENTER. AGUARDE ATÉ APARECER – SELECT FUNCTION.