SBA-733Plus Manual
SBA-733 Plus Semi-Automatic Biochemistry Analyzer
Operation Manual Statement Sunostik Company has the right to modify the Operation Manual without prior notice. Sunostik Company has the right to change the technologies without prior notice. Sunostik Company has the right to revise the specification of the product without prior notice. Sunostik Company don’t guarantee anything about this document; including the liability for guarantee that suggested the marketability and compatibility for same special items Sunostik Company will be responsible for the instrument’s safety and reliability ,but only when you make sure the following items you has done correctly: ▲Assembly operations, expansion, re-tune, improve and repair must be done by Sunostik company’s engineer. ▲The related Electrical equipment must be satisfy with the national standards. ▲You must run the instrument under the correct operational guidance. In the following situation, Sunostik Company will not be responsible for the instrument’s safety and reliability. ▲The instrument were Disassembled, stretched, re-commissioned by not Sunostik Company’s engineer. ▲Repair, modify and dismantle the instrument with non-authorized by Sunostik company. ▲The process of Running the instrument is against to the 《Operation Manual》,and caused the instrument damage.
After-Service
Free service object: When your instruments need doing free technical service, your instruments must be satisfy with the Sunostik company warranty service ordinance. Charge service object: 1
SBA-733Plus Manual
1 2
Your instrument is not satisfied with the Sunostik company warranty service ordinance. In the following situation, even your instrument is in warranty, we will not doing free service to you: Damage by the human, un-correctly using, adopt volt was un-normal, Irresistible natural disasters, replace the parts and consumables are not allowed by Sunostik or repair by the person that Non-authorized officer of Sunostik company.
Contact Us Company name: CHANGCHUN SUNOSTIK MEDICAL TECHNOLOGY CO.,LTD Address: 733 Kunshan Road, Economy and Technology Development Zone Changchun City, Jilin Province, China. Free Tel: 0086 0431 85823168 0086-4008-168-169 0086-800-846-8622 Fax: +86-431-84651234 Website: www.sunostik.com
2
SBA-733Plus Manual
Preface Thank you for your purchase and use our SBA-733 Plus semi-automatic biochemical analyzer. This operation manual will guide you how to properly run this instrument . Before using, please carefully read and understand the contents of this operation manual for operating the equipment properly, After reading, please safekeeping the operation manual and placed to the right position for easy getting. Product name:Semi-Automatic Biochemistry Analyzer Model:SBA-733 Plus Product registration number: Kyrgyz armed Food Drug Administration (quasi-) character No. 2007, No. 2400064 Product Standard Number:YZB/JI0185-2006 Production license number:Ji Food Drug Administration Xu 20,051,008 Number of firearms production Registered Address:733 Kunshan Road, Economy and Technology Development Zone of Changchun City, Jilin Province, China. Production address:733 Kunshan Road, Economy and Technology Development Zone of Changchun City, Jilin Province, China. The operation manual Release Date: August 2009. Version:1.0. File Number:SUNO/SOP-733P-01
Intended Use: SBA-733 Plus semi-automatic biochemical analyzer is used in clinical diagnosis instrument,the product can be used for routine biochemical test items:Carbohydrates, enzymes, proteins, lipids, drug abuse, therapeutic drugs testing, as well as emergency and special biochemical tests, specific proteins, the immune class category and electrolyte testing.
User
The target audience of this manual is for the professional stuff that working in clinical laboratory. The following people are included: The staff who operating the SBA-733 Plus in the daily working. The staff who doing maintenance of this equipment. The staff who is learning how to operating this equipment.
Content This operation manual is for the SBA-733 Plus semi-automatic biochemical analyzer. The purpose of this operation manual is mainly for helping users understand the SBA-733 Plus semi-automatic biochemical analyzer’s working principle, structure, operational method, routine maintenance, problem fixing and so on. Please follow this instructions to operate this instrument.
3
SBA-733Plus Manual
Sign Symbol signs and terminology Based on the different content, we using some symbol signs for marking, please notice each symbol sign’s meaning correctly when you read thie user manual. These signs are as follow:
Sign
Description
1.Remind the users following the operation manual, otherwise, may cause personal injury. 2.Remind the users following the operation manual, otherwise, may damage the system or affect the test results. 3.Explain the important information of operation by step. or other contents needs the users attention. Remind the users following the operation manual, otherwise, there is the dangerous of biological contamination.
Label and screen printing instructions The meaning of label and screen printing on the SBA-733 Plus Semi-Automatic Biochemistry Analyzer as below:
Sign
Description CE Logo In Vitro Diagnostic Equipment Risk of biological contamination Risk of electric shock
Manufacturing Date Product Serial Number Warning: high-temperature
4
SBA-733Plus Manual
Picture The pictures in this operation manual are only as the description or give a demonstration, not for other purposes.
Safety Precautions In order to use this system safely, please read the following safety precautions. Any breach of the right operate process may destroy the system damage or personal injury.
Warning If the user does not run the instrument as the operation manual told you, this may destroy the equipment and affect the test results.
Prevent electric shock In order to prevent electric shock, please strictly to following precautions as below:
Warning When the instrument’s power was on, if you found liquid went inside the instrument or the instrument was leaking by itself, if the liquid into the system or the systems internal leakage, Please cut off the power supply immediately, and contact Sunstik's customer service department or distributors. Liquid may cause electric shock and destroy the system.
Biochemical risk protection For protecting the biochemistry risk, please following the below precautions:
Risk of biological contamination Incorrect use of the sample may lead to infection. Do not touch the sample, mixtures and liquid waste directly use hand, please wear gloves, work clothes to avoid being infected when operation, if necessary, please wear protective glasses. If the sample contact with the skin accidentally ,please processing following the user standard immediately. Some agents have a strong acid or alkali. Please be careful using , to prevent the hand and close thing to contact the reagents directly .if the reagents contact with the hands and clothing accidentally, please washing with soap and water immediately. If the reagent gets in the eyes accidentally ,please washing with water immediately and consult an ophthalmologist immediately. 5
SBA-733Plus Manual
Waste Fluid treatment In order to prevent the waste fluid cause the environmental pollution and physical harm, please following below items when dealing with waste fluid.
Risk of biological contamination Some substances from Reagents, Liquid quality control, Calibration solution , Strengthen the cleaning fluid and waste fluid are control by the pollution regulations and emission standards. Please following the local emission standards and consultation with the relevant reagent manufacturers or distributors.
Attention Items In order to operate the equipment correctly and effectively, please pay more attention to the following items. System Usage
Warning This equipment applies to serum, plasma, urine and cerebrospinal fluid. Out of this scope, please consult SUNOSTIK. Please reference other clinical symptom or other test results, when estimate the clinical result.
Environment
Attention Please set the equipment in the appointed environment followed by instruction. Out of the appointed environment, the test result will be incorrect and broke the system. If want to change the system, please connect SUNOSTIK service hotline or local distributor.
6
SBA-733Plus Manual
Electromagnetic wave and Interference Prevention Attention Please do not put a noise equipment near to the system. Please turn off any electromagnetic equipments in the room, where the system in. Such as mobile phone, radio equipment and so on. And do not use any CRT monitor near the system. Noise, electromagnetic wave can be cause the malfunction and affect the result.
System Use Attention (1)Please performance the equipment followed by instruction. Incorrect use will cause the wrong results; even the system and person will be harmed. (2)First time, please check the control to confirm the equipment is run. (3)After turn off the main power 10 seconds, open again. Otherwise, the system will be protected. If it is, please turn off the main power then turn on again.
System Maintenance Please maintain the system regularly followed by below notes. Otherwise, lifetime and characteristics will be affected.
Attention (1)Please maintain the system followed “System Maintenance” in the instruction. Incorrect maintenance will cause the wrong results, affect its characteristics and lifetime, even system and person can be harmed. (2)The surface will accumulate dust in long time. Use soft cloth to put in the water and twist it, then rub the surface lightly. Immerge the cloth in to soap liquid if necessary. Do not immerge the cloth in ethanol or related solvents. At last, use the dry cloth to clean.Before cleaning, please close the power and pull out the plug. During cleaning, please prevent water goes into the system, otherwise, system and person can be harmed. (3)Do performance the control analysis, after change the main parts, such as source lamp.
Parameters Set Attention Before use this equipment, please program the volumes of sample, reagent, wavelength and so on, please follow by the instruction sheet and reagent sheet. 7
SBA-733Plus Manual
Content AFTER-SERVICE .........................................................................................................1 PREFACE ......................................................................................................................3 USER ...........................................................................................................................3 CONTENT .....................................................................................................................3 SIGN ............................................................................................................................4 LABEL AND SCREEN PRINTING INSTRUCTIONS ..............................................................4 SAFETY PRECAUTIONS .................................................................................................5 PREVENT ELECTRIC SHOCK ..........................................................................................5 BIOCHEMICAL RISK PROTECTION .................................................................................5 WASTE FLUID TREATMENT ...........................................................................................6 ATTENTION ITEMS .......................................................................................................6 ENVIRONMENT ............................................................................................................6 ELECTROMAGNETIC WAVE AND INTERFERENCE PREVENTION ......................................7 SYSTEM USE ................................................................................................................7 SYSTEM MAINTENANCE ..............................................................................................7 PARAMETERS SET ........................................................................................................7 CHAPTER 1 INSTRUMENT INSTALLATIONS ......................................................10 1.1 UNPACKING .........................................................................................................10 1.2 INSTALLATION ENVIRONMENT..............................................................................11 1.3 TEMPERATURE AND HUMIDITY REQUIREMENT .....................................................12 1.4 POWER SUPPLY REQUIREMENT .............................................................................12 1.5 WASTE LIQUID BOX ..............................................................................................12 CHAPTER 2 EQUIPMENT OVERVIEW ..................................................................13 2.1 PHYSICAL FORM ..................................................................................................13 2.2 BACK FORM .........................................................................................................13 2.3 MEMBRANE KEYBOARD .......................................................................................14 2.4 DISPLAY SCREEN .................................................................................................15 2.5 THERMAL PRINTER ..............................................................................................15 2.6 RS232 INTERFACE ...............................................................................................15 2.7 PIPETTE AND IMBIBITION SWITCH ........................................................................16 2.8 COLORIMETRY SYSTEM .......................................................................................16 CHAPTER 3 SOFTWARE INTRODUCTION ...........................................................18 3.1 SELF-TEST ...........................................................................................................18 3.2 WINDOW DESCRIPTION ........................................................................................19 3.3 PARAMETERS SETTING .........................................................................................21 3.4 APPLICATION MENU.............................................................................................22 CHAPTER 4 PROJECT PROGRAMMING ...............................................................27 4.1 PROGRAMING INTRODUCTION ...........................................................................27 8
SBA-733Plus Manual 4.2 BASIC PARAMETER ............................................................................................27 4.3 CALIBRATION PARAMETER ................................................................................29 4.4 QC DATA ...........................................................................................................30 CHAPTER 5 SAMPLE TESTING ..............................................................................31 5.1 5.2 5.3 5.4
MAIN MENU ......................................................................................................31 END-POINT ........................................................................................................33 KINETICS METHOD ............................................................................................37 OTHER METHODOLOGY .....................................................................................38
CHAPTER 6 STATISTIC SYSTEM ...........................................................................39 6.1 INPUT QC PARAMETERS .......................................................................................39 6.2 QC TEST ..............................................................................................................39 6.3 QC VIEW .............................................................................................................39 6.4 PRINT QC CHART ............................................................................................40 CHAPTER 7 MAINTENANCE ..................................................................................41 7.1 REPLACE PRINTING PAPER ....................................................................................41 7.2 REPLACE FUSE .....................................................................................................41 7.3 REPLACE LAMP ....................................................................................................42 7.4 REPLACE PUMP PIPE .............................................................................................43 7.5 DAILY MAINTENANCE .........................................................................................45 7.6 CLEAN FLOW CELL ...............................................................................................46 APPENDIX A COMMON FAULT ANALYSIS AND CLEARANCE......................47 APPENDIX B TECHNICAL PARAMETERS............................................................49 APPENDIX C ACCESSORIES / CONSUMABLES..................................................52
9
SBA-733Plus Manual
Chapter 1 instrument installations Warning SBA-733 Plus are required to install by SUNOSTIK’s engineer or by the people who was authorized by Sunostik company.
1.1 Unpacking On receiving this instrument, check whether the packing is in perfect condition and that the seals are intact. Open the box and take out its content with care (Figure- 1.1). Read this chapter carefully before installing the instrument. Do not throw the packing material away. In case of the re-shipment, use it again to guarantee good transport.
10
SBA-733Plus Manual
Attention To avoid the un-necessary argument, please check each thing in side the box, make sure everything that wrote on the packing list is prepared.
Packing List Description
Quantity
Main unit
1
Fuse(2A)
4
Power cable(3 cores)
1
User’ Manual
1
Thermal paper(58mm)
1
Wash solution(100ml)
1
Waste bottle
1
Dust cover
1
Ground connection(1.5m)
1
Silicon tube
1
serial port line(1.5m)
1
1.2 Installation environment Indoor installation only Table should be flat(gradient less than 1/200) and no vibration The table at least able to withstand the weight of 80 KG Well ventilated place It will be better without less dust Avoid the direct sunlight Avoid to put near the source of wind and heat Non-corrosive and flammable gases; No noise source and power supply interference Far from the brush-type motors and the instrument that often switch on and off Do not close to the issue of electromagnetic devices, such as mobile phones, radio transceivers, etc.; ◆ Place for working is less than 2000 m altitude ◆ ◆ ◆ ◆ ◆ ◆ ◆ ◆ ◆ ◆ ◆
Attention The working environment should be well ventilated, if necessary, ventilation equipment 11
SBA-733Plus Manual can be used. However, should avoid blowing air directly to the system.
1.3 Temperature and Humidity requirement Temperature should be between 0℃to 40℃(Fluctuations less than ±2℃/H) The humidity for system store should be 30% RH ~ 80% RH and also non-condensing。 If the environment is out of this range, please using air condition to cooling.
Warning If the system is idle for a long time do not use, this may speed up the system’s mechanical and circuit aging process, so run the instrument on a regular time is required.
1.4 Power supply requirement ◆ ◆ ◆ ◆ ◆
230V~, 50Hz, 3-core power cable, good grounding; Well-grounded power outlet to provide the necessary power supply。 Should be has at least 3 meters distance between the instrument and the power supply plug。 The current of power supply plug should be more than 1 A Choose the right fuse
1.5 Waste liquid box Choose the right length of the waste pipe that you need. Put one end into the waste liquid bottle. Put anther end into the waste liquid exit of the equipment. You can see below picture for details.
Risk of biological contamination Make sure wear gloves when install the pipe of waste liquid bottle. Put on work clothes to prevent be contaminated. Put on the protection glassed if necessary
0
Waste liquid exit
废液瓶
12
Waste liquid bottle
SBA-733Plus Manual
Chapter 2 Equipment Overview 2.1 Physical Form ①
⑤
② ⑥
③
④
① printer cover ② incubation bit cover display ⑥ keyboard
③ pipette
④ imbibitions key
⑤ liquid crystal
2.2 Back Form
①
②
③
④
① power switch ② power supply connector outlet ⑥ peristaltic pump
⑤
③ RS232 interface
⑥ ④ fan
⑤ waste
Attention The environment should be ventilation to make the emitting system work normally, if necessary use the ventilation device. But please prevent the air be blew to the system directly, otherwise the data will be affected.
13
SBA-733Plus Manual
2.3 Membrane keyboard
Three parts of the keyboard: numeric, letter and function keys. Numeric keys: The keys 【0】~【9】and【.】constitute the numeric keyboard. They serve for inputting and revising the numerical data. Letter keys: The keys 【A】~【Z】and【-】constitute the letter keyboard. They serve for inputting and revising the letter parameter information. Function keys: Perform routine work and activate some functions. ◆ 【Clear】to delete the mistype data. ◆ 【Wash】to washing the flow cell, in order to reduce the cross-contamination. The peristaltic pump will be high speed rolling when this key is pressed, at the same time plenty of distilled water will be sipped in to rinse the flow cell, again press the key to stop the operation. This function is suggested to be proceeded before and after the test to reduce the cross-contamination, and improve the test accuracy. ◆ 【】、【∧】、【∨】to move the bar to change the focus. ◆ 【Esc】to exit the program or back to the main menu. ◆ 【Enter】to write or confirm the information. ◆ 【Absorb】the key is in the left front of the equipment, to run the peristaltic pump to sip the sample. ◆ 【F1】、 【F2】、 【F3】、 【F4】match with the words which are displayed at the bottom of LCD separately, different functions in different displays. Blank means no function.
14
SBA-733Plus Manual
2.4 Display Screen
Characteristics: 240×128 dot matrix LED backlight design Toshiba T6963C controller On-chip 32K Byte SRAM
2.5 Thermal Printer
Characteristics: 384dots/line Resolution 8dots/mm Max print speed 75mm/s Built-in thermistor to check the needle of the printer. Built-in sylvatron to check the paper.
2.6 RS232 Interface Go though the RS-232 interface to connect the equipment with computer. You can program a better report by using the enclosed CD software.
For details, please refer
to the instruction sheet in the enclosed CD. Meanwhile, you can upgrade the software 15
SBA-733Plus Manual through this interface. (no need to open the cover) Conveniently.
2.7 Pipette and Imbibition Switch When absorb the sample, put the pipette into the bottom of cuvette to prevent absorbing the bubble during do this action. After pressing the switch, the imbibition pump will self-start to absorb the sample into the flow cell. Take away the pipette from the cuvette quickly, when the imbibition device stopped. At this time in order to improve the test accurate, the air will be absorbed in this device, to reduce the cross-contamination between two contiguous samples.
Biological Pollutant Risk When doing the test, please mark sure the gloves and working clothes are wore, if necessary take on the prevention glasses.
Attention When checking the sample, please mark sure the pipette is straight to prevent the result accuracy is affected.
2.8 Colorimetry System Colorimetric system includes peristaltic pump, quartz flow cell (32ul), optional system and mechanism gearing, wavelength adjustment system and Peltier . The system is controlled by CPU, temperature control and wavelength lookup both be proceeded under the CPU. To be Noted that you’d better take out the silicon tube from the peristaltic pump when the instrument will not be used in long time to avoid the tube block.
16
SBA-733Plus Manual System components are as below:
Warning The colorimetric system is the main part of this equipment, unprofessional untrained person don’t open it randomly. Otherwise the properties will be affected.
High Temperature Warning When change the lamp, mark sure the power is off over than 20 minutes, otherwise your hands must be burned by the high temperature of the lamp.
17
SBA-733Plus Manual
Chapter 3 Software Introduction 3.1 Self-test After open the power switch, (see ① in the picture of
the equipment back form
picture), the display is as below:
The self-test will be started 5 seconds later, to test the memory, printer, filter and sampling circuit separately and display the status real time. The below picture will be displayed after self-test completely. Flash Memory:16.5Mbit RAM 32Kbit Printer test OK Filter test OK Sample Test OK Lamp Test OK Test Self OK
The temperature adjustment will be started after the self-test, refer to below picture. Main menu will be indicated after the temperature adjustment, or you can enter the main menu by pressing any key directly. At this time, the temperature adjustment is kept going without any display.
18
SBA-733Plus Manual
Temperature adjusting…
20%
Press any key to enter the main menu
Attention You’d better to start the test after complete the temperature adjustment to make the test accuracy.
Attentions for Err or Alarm: "err: #02 Cannot find flash":Flash is broken or connection is malfunction. "Warning: #01 Printer has no paper":No paper in the printer or connection is malfunction. Works can not be affected by this alarm. "Warning: #02 Sample test failed":A/D read failure. Malfunction of pick-board or the connection is malfunction between pick-board and main-board. "Warning: #03 Open lamp failure":Lamp is off or low A/D.
3.2 Window Description Window Format: All windows include in this system with the same format, such as main menu(below picture): There are three parts in the window: title bar up the window to display the subjects and time, middle is the customer area to indicate the parameter list and submenu information, status bar down the window to indicate shortcut key and cue information.
19
SBA-733Plus Manual
Main Menu
07/08/01 14:36:20
Sample Test Main Menu Description: There are 6 submenus under the main menu; cursor is rest on the “sample test” position default. You can move the cursor with keys【】 【∧】 【∨】, the related function also will be indicated, at the same time the related functions will be displayed on status bar. Press【Enter】 to enter the submenu.
Attention Press the shortcut keys【1】~【6】to enter the submenu directly.(From left to right; from up to down). Shortcut key “R” means to delete the results; “P” means to print; “T” means set time; “M” means to A/D; “U” means to update the window; “H” means the basic information.
The following is the brief introduction for each submenu: ⑴Sample Test: It is the beginning of testing one sample. After programming one item, press the key to test. ⑵ Absorbance: It is specially ready for testing the absorbance. You can test the absorbance after enter the menu. The wavelength, temperature, absorption volume and delay time can be programmed. There will be a hint when you exit the submenu. ⑶ Programming: You can write all the programs of 100 tests under this menu. You can also revise the program under this menu. ⑷ Parameter setting: There are four parameters can be set under this menu, such as measure device, sipping volume adjustment, air sipping and language. ⑸ Application menu: There are 6 functions for using, such as patient record, printer setting, time, system maintenance, software update and help. 20
SBA-733Plus Manual ⑹ QC query: Under this menu, you can check the saved QC information for 100 test items. There are 3 QC type under each item, each QC type can save the QC result 100 days. There are 2 modes for QC indication (QC data list and QC chart), both can be printed out. You can evaluate the accuracy of the test items through the QC chart, the instrument also can automatically calculate out the QC data’s average, differentiation and variance of the certain month.
3.3 Parameters setting Under the main menu to select parameters setting or directly press the shortcut key 【4】to enter this window, see as below:
Parameter System
07/08/01 14:36:20
Test Device: Flow cell Absorption Volume Calibration: 100% Air Volume: 80uL Language: Chinese
Sample Test Measure device setting The instrument can available for flow cell and colorimetric cup, the default device is flow cell. The samples sipping and draining is controlled by the peristaltic pump under the flow cell selection, convenience and efficiency; the peristaltic pump will stop operation under the colorimetric cup selection, the pump push button should be pressed for confirmation under the manually sample adding. The advantages of colorimetric cup are highly-transmittance and few cross-contaminations, etc.
Absorption volume calibration If there is incorrect, you can adjust the parameter to keep the accuracy. This parameter is direct ratio with the absorption volume, default 100%.
Air sipping setting This parameter is used for setting the length of air, the default is 80uL. It can be adjusted base on the pipette length, air can not be sipped into the flow cell is the condition. 21
SBA-733Plus Manual
Attention The absorption volume calibration and air sipping setting should be correct, incorrect setting will cause the wrong results.
Language The operation language is available for many languages. After selection, it works immediately.
3.4 Application Menu To select the application menu or directly press the shortcut key【5】to enter this window under the main menu, see below picture:
Application Menu
07/08/01 14:36:20
Check the record Application menu instruction There are 6 submenus under the main menu. You can move the cursor with the directional keys by pressing the buttons to select different submenus. At the same time the related functions will be displayed on status bar. Press【Enter】 to enter the submenu. The following is the brief introduction for each submenu: ⑴ Patient record: Enter this window to find the details of the patient test result and test time, to print or delete. See below picture:
22
SBA-733Plus Manual
Record
07/08/01 14:36:20
View Record Print Record Delete Record
Input Code | Record query: To input the code No. (Range of code No. is 1~200). Press key 【Enter】to start query. Comprehensive print: Input two code numbers and press key 【Enter】to print all the records between this two codes. Comprehensive delete: Input two code numbers and press key 【Enter】to delete all the records between this two codes. ⑵ Print setting: Enter this submenu to set the print parameter, operate the printer and check the status of the printer, see below picture:
System parameter
07/08/01 14:36:20
Lab Name: Auto Print: Print Style: Print State:
Test
SUNOSTIK On Standard Standby
Paper
Lab Name: The default is SUNOSTIK, it can be adjusted according to the customer request, and the name will be printed in the test report. (It is workable with the standard print type). Auto-print: The test result can be auto-printed out after complete the test if this function is in application. Print Style: There are two print type is optional, default is standard type, there is 23
SBA-733Plus Manual blank between the 2 test results to make sure all the content can be indicated out of the paper outlet. You can select economy type to save the paper, but the result content will be compactness.。 Print Status: The item is for read only to indicate the printer status. Self-test: The shortcut key【F1】is used for testing the paper and checking the printer. Paper Moving: The shortcut key【F2】is used for moving paper, this function can be acted when the printer is installed. Repress it to stop the moving. ⑶ Time System: Using the keys【】to move the cursor to revise the date and time after enter this submenu. See below picture:
Time System
07/08/01 14:36:20
07/08/01 14:36:20
Attention The time system is very important for working normally, incorrect setting will mistake the test result and QC data, please take it in serious.
⑷ System Maintenance: Enter in this submenu to check the A/D value in certain wavelength. See below picture: The default filter is 340nm; A/D value is the last number. Press key【Enter】to bounce out the wavelength list, you must select a wavelength to start the sampling,
and keep the A/D value continuous and in real
time.
24
SBA-733Plus Manual
System Maintenance
07/08/01 14:36:20
Filter: 340 A/D value: 167.1452
⑸ Software Update: You can follow the below picture to apply the latest software to upgrade the functions. See below picture:
Update
07/08/01 14:36:20
Connecting pc through RS232
Please restart the power for running the new procedures after update.
Download Guide: Use the matched connect line to connect the computer through RS-232 interface(Rear view③postion),Open the updated software that we provide to download and update .Software interface as below:
⑧
①
⑦ ⑥ ② ⑤ ③
⑨
25
④
SBA-733Plus Manual Help: ③⑤⑥⑦ position : Set up parameters according to the map,⑧ position :According to the port NO that contacting by equipment and computer for confirmation. After set up these parameters,Click ① to select the program file ,the extension of the file is .hex, Such as SBA733.hex。Open the equipment and click ④。After a few seconds,,There is presentation on ⑨ that show the connecting well or not ,if display "Reading of PART ID Success" ,it means connect successful,please click ② to download the file at this time .At the same time there is a blue progress bar showing on ⑨.when the progress bar run to the right side .download completely .If display "Reading of PART ID Failed",that means connection failure .please check the line connection carefully, Whether the line contact with a solid ,to connect again after the check.
Attention Please shut off the power of the equipment before download the Software ,use the matched connect line to connect the computer and equipment. It may burn out the equipment when connection with the electricity.
⑹ Help:Under this interface you can find the basic information of the equipment and the contact information of our company, see as below.
Help
07/08/01 14:36:20
Model:SBA-733 NO:SUNO070801001 Vender:Sunostik Medical Technology Co .,Ltd Version:1.0 TEL:4008-168-169
www.sunostik.com Please contact with us freely if you have any questions.
26
SBA-733Plus Manual
Chapter 4 Project Programming 4.1
Programming Introduction
Please select the application menu under the main menu or directly press the shortcut key【3】 to enter into project selection interface, input the item No and press 【Enter】item enter into edit interface, see as below:
ALT
07/08/01 14:36:20 Item name: ALT Test Mode: Kinetics Main wavelength: 340 Vice-wavelength: 700 Test temperature: 37
TEST
End page
PRINT
Next Page
Press【】key to move the cursor to switch the focus parameter;Press the shortcut key【F1】、 【F2】、 【F3】、 【F4】 to achieve the function that opposite on the status bar ;Press【Enter】key to edit the focus parameter ,at this time ,the cursor is moved to the editing window or the sub-window ;you can use【Clear】key to delete the mistake during the editing;;After the editing ,press【Esc】key to exit the submenu and give up the content which is edited . press【Enter】key to confirm and exit the editing window or the sub-window, back to the
4.2
above interface.
Basic Parameter
Item Name: Input the English name of the item with the form of edit-box,All the test items of this equipment is open that you can organize the project scheduling according to your personal habits .For example, Installed the frequently used items in front ,installed the related items next to the adjacent number. In order to improve your work efficiency. Test Mode: Edit with the form of sub-window, the method include end-point, kinetics, fixed time, differential and multi-standard, etc., different test method with different data list for comparing, Does not display the extra parameters , so edit any of these methods are 27
SBA-733Plus Manual very simple and clear. Decimal digits: Edit with the form of edit-window, range: 0 – 4,default is 2, This parameter is used to trim the number of the decimal digits of the test result, only for controlling the only measurement results display and print, and will not really change the results of precision ,so there is no rounding calculation and maintain the original data accuracy. For example: if you input “0” in the decimal digits,the test result showed that “32” (assuming the true results of 32.6789),when you modify this parameter is 4, the results can still be a true showing of 32.6789. Main Wavelength: Edit with the form of sub-window ,There are 7 wavelength for your selection :340、 405、505、546、578、620、670.please select the wavelength according to the user’s manual of the reagent. Vice Wavelength: Edit with the form of sub-window, There are 7 wavelength for your selection:340、 405、505、546、578、620、670.please follow the user’s manual of the reagent to select the wavelength correctly when the reagent requires to use of dual-wavelength for testing. If the parameter is selected as "×" or the same with main wavelength, then the vice-wavelength not be used. Test temperature: Edit with the form of sub-window,there are three items for your selection:25、30、 37. please select this parameter according to the user’s manual of the reagent correctly. Absorbance Volume: Edit with the form of edit-window,Range:200 – 1500 uL,programmed by 0.1 step.Absorbance volume is too few will affect the test precision ,absorbance volume is too much will waste the reagent .please setup consideration Delay Time: Edit with the form of edit-window,Range:3 – 999 S,please set up this parameter according to the user’s manual of the reagent correctly. Incubate Time: Edit with the form of edit-window,Range:1 – 999 S, please set up this parameter according to the user’s manual of the reagent correctly Measure Time: Edit with the form of edit-window,Range:30 – 999 S, please set up this parameter according to the user’s manual of the reagent correctly. Thickness Unit: 28
SBA-733Plus Manual Edit with the form of sub-window .there are 13 ordinary Units for selection. please set up this parameter according to the user’s manual of the reagent correctly. Reference Upper Limit: Edit with the form of edit-window, This is the upper limit of the reference range .please set up this parameter according to the user’s manual of the reagent correctly. Reference lower Limit: Edit with the form of edit-window, This is the lower limit of the reference range .please set up this parameter according to the user’s manual of the reagent correctly Correction coefficient: Edit with the form of edit-window. When the test results is generally low or too high, you can use this parameter to amend the results.
Attention Default number is 1,To setting this coefficient should be according to the huge statistic data, it should be applied in serious.
4.3
Calibration parameter
Calibration Mode: Edit with the form of sub-window,there are two modes for selection:STANDARD and FACTOR. If you select standard ,the standard liquid should be used during the test. The equipment will calculate the factor and saved it for the future automatically. If you select factor, you can apply the factor for the test directly to save the standard liquid, there is no need to set the calibration. If the test mode is Kinetics, the calibration will be inefficacy and hided automatically. Factor Value: If the calibration mode is Factor, The factor value must be input for calculation. If the calibration mode is standard, the factor value will be inefficacy and hided automatically. Standard quantity: If the test method is multi-standard, that need to set the quantity of the standard liquid, other test method , this parameter will be inefficacy and hided automatically.
29
SBA-733Plus Manual 1.
Standard Thickness:
If the calibration mode is standard, the standard thickness should be input for calculation. If the calibration is Factor, this parameter will inefficacy and hided automatically. If the test method is Kinetics, this parameter will be inefficacy and hided automatically. The standard thickness will be show X when the test method is multi-standard. 2.
Standard value X:
The standard value X should be input under the multi-standard test method(each thickness of the standard liquid),The maximum range is 2-6 , it is according to the set of standard quantity. Under the non multi-standard method, this parameter will be inefficacy and hided automatically.
Attention The setting of the standard value has to be followed from small value to big value , or it will cause the mistake result.
4.4
QC data
Target H、Target M、Target L: Edit with the form of edit-window, separate as QC high value, QC middle value, QC low value. please set up this parameter according to the user’s manual of the reagent correctly Standard difference H、Standard difference M、Standard difference L: Edit with the form of edit-window, separate as the standard difference for QC high value, middle value and low value, to be applied for plot of the QC chart. please set up this parameter according to the user’s manual of the reagent correctly
30
SBA-733Plus Manual
Chapter 5 Sample Testing 5.1
Main Menu MENU
07/08/01 14:36:20
Sample Test Press 【Enter】or shortcut key【1】to enter the following menu on above picture, namely the “sample test”, see below picture:
ITEM 01 AST 02 ALT 03 a-HBDH 04 LDH 05 CK
07/08/01 06 07 08 09 10
Input:|
CL-,MB ALP
α-AMYS γ-GT GLU END/P
NEXT/P
Please input the test item No under this menu. if the item is not exist on the interface , press shortcut key【F3】 【F4】,according to the next page indication on the screen to search the item which you need, total are 10 pages cycling, After input the No, press 【Enter】to enter the baseline test interface.
Attention For example: If you want to test “AST”,please press the key【1】and 【ENTER】enter to the baseline test interface ,see as below. 31
SBA-733Plus Manual
AST
07/08/01 14:36:20 Baseline
Absorb the water The above picture is in case of the measure device is flow cell, and the information indicated in the status bar is “Absorb the distilled water”,please prepare the distilled water and press the Absorbance button, the peristaltic pump start to working. The indicated content will be “absorbing…”. After absorb the distilled water, the status bar will display “withdraw the cuvette ” with one beep. Please take out the pipette from the cuvette at this time, and the analyzer will absorb in the certain air (the air absorbance item is included in the system parameter) to separate the different liquid, In order to avoid the cross-contamination. If the measure device select the “Colorimetric cup”, The above status bar will show “add the distilled water ”.You should put the colorimetric cup full filled with the distilled water and press the button. There is no absorb procedure when use colorimetric cup the mentioned as above. It is the same for adding the sample, we will not mention again. The delay time will be come out after finish the absorbance , in order to level off the distilled water,, and the status bar will show “in delay 00X”,
the unit for X is the second and count
down .The status bar will be show “in A/D value reading” after the delay time .If the baseline read failed ,it will show below picture.
AST
07/08/01 14:36:20 Read baseline failure!! Mean = …
CV% = …
Absorb the distilled water
32
SBA-733Plus Manual The reason for baseline reading fail is as below: 1.
If CV%(coefficient of variation)exceed 5%,it means
the set-up time of the
analyzer is too short and the analyzer is not in steady, please wait a moment and again try. Or it is caused by the aging of the peristaltic pump,the liquid is in fluid status, please replace the peristaltic pump and try again. 2.
If Mean(Average value)under 150,maybe caused by the bad connection of the flow cell or the colorimetric cup,or there are some air bladder in the distilled water, please deal with it and again try.
If read the baseline successful, it will enter into the different interface base on the selected test item; the following will introduce of the procedure for each test method.
5.2
End-point
The interface of end-point testing as below(take GLU for example,only for your reference).
GLU
07/08/01 14:36:20
Parameter Blank = 0 Factor = 1746
Black
Standard
Sample
QC
The default parameter will be the last test data of blank and factor. you can directly to start the sample or QC test if the indicated data is in your requirement. If the parameter does not match your inquire or you want to reset the parameter for increasing the test result .If not, you can press the shortcut key according to the content of the status bar, reread, blank and standard. below is the instruction for the four interfaces: Press【F1】to enter the test interface for blank test, see below picture:
33
SBA-733Plus Manual
GLU
07/08/01 14:36:20
Blank
Blank = Factor = Absorb the blank liquid The work procedure is the same with absorbing the distilled water , we won’t mention again, it will return the up menu after reading blank, the data in the system is updated, if the result is not fit for you, you can retest it follow the same procedure .retest the blank. Press【F2】to enter into the interface of standard, see as below:
GLU
07/08/01 14:36:20
Standard
Standard = Factor = Absorb the standard liquid The work procedure is same with reading the baseline procedure, we won’t mention again. It will return the up menu after the reading standard, the standard and factor parameter is updated at this time, now it finish setting the standard. If the result is not fit for you, you can retest it follow the same procedure. Reset the standard. Press【F3】 to enter into the interface of sample test, see as below:
34
SBA-733Plus Manual
GLU
07/08/01 14:36:20
Sample
ID:001|
Absorbance = Result= Absorb the sample The default patient No. is 001,the No. can be edited when the cursor in flash. Press Key 【Clear】to delete, and use numeric key to edit. The range of the No. is 1--999 (the first 200 patient the test result can be saved automatically). Please press the button to absorb the sample. The procedure is same as absorbing the distilled water . we won’t mention again. It will come into the below window after absorb the sample , the result will be sent to the computer through RS-232. If the auto-print function is on turn(this function is edited in the print editing), the test report will be automatically printed out .Otherwise, please press【F4】to print the test report.
GLU
07/08/01 14:36:20
Sample
ID:001|
Absorbance = 0.5547 Test Result = 30.22 Next
Retest
Reread
Print
You can select the retest or reread function if you are not satisfied with this result. Retest: Does not to change the patient No. and re-process the above procedure .need to absorb the sample again Re-read: Does not to change the patient No, and no need to absorb the sample again, reread the current sample If you want to proceed the other sample test under the same item, please press shortcut key【F1】to enter the test interface and the patient No. will be add 1
35
SBA-733Plus Manual automatically . it will test the next sample with reprocess the above procedure. The next test also can be followed with the same procedure. You also can directly press the button to start the new sample test without press【F1】,the patient No. will be added 1 automatically and enter to the procedure of absorbance and testing, in order to simplify the operation.
Attention Please press 【Wash】 when the test is finished ,to avoid the cross-contamination .Press【Wash】again to stop washing.
Press 【F4】 to enter the interface of QC control ,see as below.
GLU
07/08/01 14:36:20
QC
H
M L
Absorbance = Result = Absorb the QC liquid The QC type is default as the normal QC value, see above picture, the cursor is on the M point. If you’d like to do the abnormal value, use【】to move the cursor. The procedure of absorbing and reading is the same as the reading baseline , we won’t mention again, , it will come into the below window after reading the QC:
GLU
07/08/01 14:36:20
QC
Absorbance = 0.5547 Result = 30.22 Blank
Stand
Sample
QC
If you have some doubt for the above QC result or you’d like to do the other type of 36
SBA-733Plus Manual QC, press【F4】to reprocess the above procedure. Under the same item , type and date, only the latest QC result will be saved.
5.3
Kinetics Method
The interface of kinetics testing as below(take AST for example,only for your reference).
AST
07/08/01 14:36:20
Parameter
Blank = 0 Factor = 1746 Sample
Black
QC
The blank data that display in the data list is default as the latest test result, you can directly start the sample test or QC if the blank data is in your requirement. If the parameter is no fit for this test or want to retest the blank in order to increase the test accuracy. please re-read the blank data base on the indication on the status bar. Compare with End-point, it is no need to make the standard .so it is no standard optional .we are only to introduce the interface of sample test, the other procedure please refer to the End-point test method. Press 【F3】to inter the interface of sample test .see as below
AST
07/08/01 14:36:20 ID:001| Absorb: 0 Factor:1746
Absorb the sample On the left side of workspace is an coordinate system which put time on X-axis and absorbance on Y-axis in order to draw a reactive curve .The control no. on the left side of workspace defaults to 1,and could be edited freely ,the same as end-point 37
SBA-733Plus Manual method .Absorbance could real-time display its present value in the whole testing process of sample .Factor is an assigned numerical value under control of program. Please press aspirate liquid key to inhale the sample ,then it will enter the phases of hatching and testing samples. The following window will come out after test.
AST
07/08/01 14:36:20 ID:001| d A/m: 0.0207 Result:36.21
Next
Retest
Reread
print
The curve in coordinate system is a calculated linear regression which reflects sample’s changing trend and the starting absorbance value .If you would like to print、 re-read、re-test or test other samples ,please see end point method.
5.4
Other methodology
About the operating interface and operation method of two points ,please see end point method. Differential belongs to the expansion of end point, about the operating interface and operation method ,please see end point method. Multi-standard also belongs to the expansion of end point, about the operating interface and operation method ,please see end point method.
38
SBA-733Plus Manual
Chapter 6 Statistic System 6.1 Input QC parameters In order to ensure authenticity and validity of QC data ,QC test is programmed and tested with sample test at the same time .About editing of QC parameters, please see item programming.
6.2 QC Test In order to ensure authenticity and validity of QC data ,QC test is programmed and tested with sample test at the same time .About editing of QC parameters, please see item programming.
6.3 QC View Press QC View under main menu or directly press shortcut key【6】to enter into application control menu ,please see following picture:
ALT
07/08/01
14:36:20
Input month: |
High
Mediam
Low
Input your demanded month ,press【F1】、【F2】、【F3】to choose QC type. The following is list of QC data:
39
SBA-733Plus Manual
ALT
07/08/01
32.18 28.98 36.97 36.01 26.11
+1SD +1SD +2SD +2SD - 1SD
Chart
Print
14:36:20 07/07/01 07/07/02 07/07/03 07/07/04 07/07/05
Next
Back
Shortcut keys 【F1】、 【F2】、 【F3】、 【F4】separately stand for different functions
on
above status bar. Press shortcut key【F1】will draw out following QC chart.
ALT
07/08/01
14:36:20 X 32.2218
3sd 2sd 1sd X -1sd -2sd -3sd
SD 1.0548
1
Data
10
20
Print
30
Back
CV% 1.0256
Next
6.4 PRINT QC CHART In above two QC interface ,press shortcut key【F2】to print QC report and QC chart separately.
40
SBA-733Plus Manual
Chapter 7 Maintenance 7.1 Replace printing paper When installing printing paper, please cut the end of paper to a correct shape according to following picture. Extend the printing paper to feeder of the printer.
correct
Incorre
When the instrument is on working(when power is turned on ,it the printer processing power ),at this time, send printing paper to the feeder under the paper delivery roll ,press【paper feed】,the printing paper will come out automatically.
7.2 Replace fuse Fuse ① is on the rear of instrument ,it was installed together with power button, Rating current is 6A.
1 When replacing fuse, firstly twist off the cap ,remove old fuse and install a new one equipped together with instrument.
41
SBA-733Plus Manual
Warning If you replaced a new fuse and turned on power ,the fuse was burned at Once .Under this situation ,please don’t change a higher current fuse any More ,turn off the instrument and get in touch with our maintenance engineer at once.
7.3 Replace lamp The instrument uses a 6V/10W halogen lamp ,life span is about 2000 hours ,because the lamp is in high usage frequency , after a period of usage the lamp will be not bright enough or can’t work any more ,at such time ,you need to replace a new lamp. First, twist off the cap, loosen two inner hexangular set screws(2)be used to fix component parts of lamp base (4) with an inner hexagon spanner,remove the component parts and remove halogen lamp (1)from lamp base(3).Install a new halogen lamp on the lamp base ,and install the component parts on the lamp holder ,fasten the screws with inner hexagon spanner .Please see the following pictures to check detailed installing position.
3
4
1
2
High temperature warning When replacing lamp ,you must wait 20 minutes after turning off the power ,or your hands may be scalded by high temperature comes from the surface of lamp.
42
SBA-733Plus Manual Ajust light source: When you need to replace lamp, loosen fixed screws under the instrument to open upper case .Please pay more attentions when removing upper case to prevent electric shock .To place upper case down from the right side(facing the instrument direction). Loosen hexangular set screws with inner hexagon spanner .Install the new lamp as original way .Before fixing lamp base ,assure the strongest light of lamp is just in the centre position .You must assure filament is perpendicular to bottom board .After proper adjustments ,please fix lamp base with inner hexagon spanner.
Adjustment and detection: After fixing lamp ,please monitor A/D Value under maintenance window when standard filter is 340nm, the A/D value should be larger than 120nm at least ,if this value is too small ,please readjust position of lamp .please see below operating interface:
System maintenance
07/08/01 14:36:20
Filter: 340 A/D value: 189.1452
7.4 Replace pump pipe Peristaltic pump located in the rear of instrument ,If you find the instrument don’t aspirate liquid or the volume disagree with pre-set value(software setting didn’t work),please remove peristaltic pump and check the reason .One possible reason is the pump pipe blocks or a long time using and wearing happens to peristaltic pump self ,both of these make an increasing interspace between shaft and three pulleys ,at such situation ,you need to replace peristaltic pump or pump pipe. The way to remove peristaltic pump: first ,disconnect the pump from pipeline ,then press the pipe joint outside pump with your left hand ,at the same time ,pull the pump pipe out with hard strength. 43
SBA-733Plus Manual After disconnecting the pump ,press the clips on both sides of peristaltic pump ,at the same time, pull the pump pipe out with hard strength. Please take the following steps for your reference on how to exchange pump pipe.
44
SBA-733Plus Manual
7.5 Daily Maintenance Never use cleaning agent or abrasive cleansers, use clean water or fabric dipped with neutral soap . In case of any reagent or corrosive substances were splashed on instrument, clean it at once with damp cloth and clean water. It is necessary to clean liquid circuit on correct way after finishing one group of tests and daily job .Clean with plenty of distilled water after finishing one group of tests. Clean liquid circuit with equipped cleaning agent everyday after work,clean tube with plenty of distilled water circular to empty liquid circuit. If the end of suction tube is broken ,you could cut down several millimeters vertically and clean the cut part. If the tube is seriously run of date ,please change a new one and use equipped instrument components If long time don’t use the instrument ,please empty liquid circuit, then pull out waster liquid tube and store it suitable. Clean surface of instrument with non-abrasive cleansers every month. Wipe off dust fell outside the fan every month. Clean the tube and flow cell after finished tests everyday to avoid dirty precipitate. Don’t pull flow cell out of instrument unless cleaning the flow cell .Do pull and plug the bolt slightly .also pay attentions to placing position of flow cell ,never install it on oppositely direction and lead to damage. A canonical operation way could extend service life of instrument .Never press soft keyboards with sharp nail on which way the keyboard could be easily broken .You should press it with finger belly . When installing printing paper ,if a bit of paper was left in printer ,please don’t pull it out forcibly,because such way could easily damage the printer .The right way is press【Paper】key to send out paper.
45
SBA-733Plus Manual
Warning Most maintenance were carried on under power off state ,so please maintain it after turning off power for safety.Never spray any liquid like reagent to instrument or electronic accessories.
Biology contamination danger Do wear gloves and cloth to prevent from being infected ,wear protective glasses if necessary.
7.6 Clean flow cell Both misusing of flow cell and drawing it out frequently all may cause a cut or a permanent damage on surface of the flow cell .Dust on the surface may decline accuracy of measurement .So please don’t draw flow cell out in normal circumstances. Wipe flow cell with lens paper instead of hard material to avoid the cut of surface.Pay special attention to installing direction after cleaning to avoid an unnecessary damage due to an opposite installation .Please see following pictures for your reference:
The side marked with arrow faces front operating interface.
Attention When wiping flow cell ,do choose the soft material like lens paper to prevent cutting light-hole which will affect the Visible Light Transmission .At the same time ,pay attentions to installing direction.
46
SBA-733Plus Manual
Appendix A Common Fault Analysis and Clearance Warning: When settling troubles ,please handle it with great care ,check whether deal it after power is turned off
Examples of trouble
Reasons analysis
Eliminate methods
Power light don’t work
1 power cable is broken or on a poor contact 2 Fuse blows out 3 Power light fails
1 Replace or inspect the power cable 2 Replace the fuse 3 Replace the switch
LCD don’t work
1 Have not pluses negative voltage of LCD 2 Flat cable is off 3 LCD is broken
1 Inspect the negative voltage of LCD 2 Check if flat cable and LCD are in good contacts 3 Replace the LCD
Peristaltic pump don’t work or don’t aspirate liquid
1 Control circuit broke down 2 Aspirating liquid switch is broken or on a poor contact 3 Peristaltic pump is broken 4 Inside pump pipe blocks.
1 Check the control circuit 2 Adjust or replace injector port 3 Replace the peristaltic pump 4 Replace the pump pipe
Test abnormity
1 There are bubbles or dirt in flow cell 2 Reagent turns bad 3 Selected filter don’t confirm to this test. 4 Incorrect operation or large error 5 Halogen lamp is broken or with an insufficient light source 6 Couveuse’s temperature is inaccurate
1 Clean the flow cell repeatedly with distilled water and check autosampler system 2 Check whether the reagent is out of date. 3 Check whether your selected filter confirms to test requests. 4 inspect the whole operating process and test again. 5 Check the halogen lamp and test the voltage of it(6V more or less) 6 Check whether couveuse’s temperature confirms to the requested test temperature ,if the error is big ,adjust temperature to requested test temperature.
Keyboards don’t work
1 Control panel is broken 2 Fast booting and Shutting down lead system halted
1 Check the control panel and replace it. 2 Reboot the instrument (turn power off and then reboot it)
47
SBA-733Plus Manual
Printer don’t work or can’t print any word
1 The instrument is not online. 2 Check printer’s power. 3 Inadequate operation 4 Wrong installation way of print paper 5 Printing head broke down 6 Internal wiring or Semiconductor Component was broken 7 Check locking bar’s position
1 Check Signal Connection(Printer cable) 2 Check whether voltage is 5V 3 Turn off the power and re-operate 4 Check a correct installation to make sure thermal face upwards 5 Replace the printing head 6 Overhaul the internal wiring and semiconductor component 7 It should be in locked state 1 Check whether the voltage fluctuation of power is so large
Bad Repeatability and unstable reading
1 Check whether light source is stable 2 Whether flow cell is contaminated 3 Liquid absorption leads to some bubbles. 4 Long staged air leads to suction some bubbles
2 Clean the flow cell with cleaning fluid equipped together with instrument. 3 Calibrate the liquid absorption refers to chapter “Temperature and liquid absorption” 4 Use sample instead of air when aspirate staged air.
During the test,when press keys of “delay time”and “test time”,time stand still
Incubation slot don’t heat or overheat
1 Malfunction of chip in clock or battery is exhausted
2 temperature sensor is broken or circuits used for transmit signal are broken. 3 Something happens to temperature controlling circuit(Open the cover and see whether instruction light still work)
48
1 Check whether there is a clock item under main menu ,if not, enter into “service menu” and reset it.
1 Replace the 2 Replace the sensor or circuit used for transmit signal 3 Replace the broken component or replace the whole temperature controlling circuit.
SBA-733Plus Manual
Appendix B Technical Parameters Light source:6V/10W imported quartz halogen lamp. Standard filter :including 7 filter, 40nm、405nm、505nm、546nm、578nm、 620nm、670nm. Linearity range: 0~3.000O.D Cross contamination rate: ≤1% Blank :the instrument could automatic store the blank value when testing. Factor :the instrument could auto recalculate the factor when
without
modifying any program parameter. Interface :there are 48 keys in main keyboard used for inputting figures、 letters or execute other quick function;240X128 dot-matrix LCD New functions on ungrading and updating system(IAP function) The instrument could update program at any time through RS-232
serial
port with equipped serial port cable. This function provide a good support to users on using the latest software version. memory unit(memory system) Main program was saved in inside of CPU. Programming contents and results were saved in Super Flash Memory Memory unit could save Parameters under 100 test.including :item name、test method、wavelength choice、temperature、 unit、reference range、delay time、test time、aspiration volume、decimal digits、 standard deviation、QC target value .You could edit and modify above items freely. Memory unit could save QC information under 100 items ,there are three kinds of QC type under every item ,every QC type could save 100 results. Parameters like
blank and factor under every item could be auto
saved ,the saved factor could be displayed and modified under program interface .You could use these parameters directly for next time. Measuring Mode: End point :you can calculate with factors directly ,this method could save reagent and standard liquid. Kinetics :this method could real-time display absorbance of 49
SBA-733Plus Manual sample ,auto display response curve After the immunoreaction . Tow points :used to calculate changing parameter of absorbance between two intervals. Differential :you can use reagent blank(6 tubes)or not use reagent blank(4 tubes) Multi-standard :used for nonlinear sample tests. QC You could set parameters(QC target value) of three QC types when editing programs. The 100 test items all have function of QC. You could view and print QC chart and QC parameters(average value, standard deviation, coefficient of variation ) Delay time: Programmable Object Select (0~999s). Test time: Programmable Object Select (0~999s),test time of rate assay and two points are 30s~999s. Quartz windowed flow cell Capacity is 32ul,diameter is 1cm. Special design on shape could decline crossed contamination rate and easy maintenance. Min suction volume is 300uL. Temperature control Optional temperatures are : 25℃、30℃、 37℃ Precision: ±0.2℃ Accuracy: ±0.1 Aspiration system Imported peristaltic pump was installed in instrument. This aspiration system was installed in the front-end ,you could replace peristaltic pump without opening the cover ,this design is easy for maintenance .Aspiration volume was controlled by program,users could modify and adjust this volume by themselves. Min aspiration volume is 300uL,Max aspiration volume is 1400uL, the volume could be continuously Adjusted. Power source : Voltage:230V~ Frequency:50Hz Power:MAX 70VA environmental condition: 50
SBA-733Plus Manual Working temperature:10~35℃ Environment humidity:≤80% Volume: 430(length)×320(width)×220(height) Weight: 10 Kg
51
SBA-733Plus Manual
Appendix C Accessories / Consumables Please use accessories made or recommended by SANOSTIK to ensure personal safety and a stable system performance. If necessary,please get in touch with aftermarket department of SUNOSTIK or rural distributor.
Name of parts lamp peristaltic pipe flow cell pump fuse thermal paper
Specifications
Position
remarks
6V/10W
light path
≥2000 hours
Φ3*Φ2*160m on back panel of instrument optical path 1cm DC12V
6A/220V φ5×20
incubate position
damaged damaged
on back panel of instrument
damaged
on back panel of instrument
damaged
printer
cleaning fluid
Consumables Consumables
52