KXO30-50F Perfrmance Check

April 11, 2020 | Author: Anonymous | Category: Tabung Vakum, Beralih, Sinar-X, Perlindungan Radiasi, Barang Manufaktur
Share Embed Donate


Short Description

Download KXO30-50F Perfrmance Check...

Description

No. 2J313-002E"P

TOSHIBA

PERFORMANCE Ct-tECKMANUAL FOR DIAGNOSTIC X-RAY GENERATOR MODELS KXO-3OR, KXO-30F AND KXO-50F (23313-OOZE*P>

TQSHIRA

CORPORATION

%? TOSHIBA CORPORATION 1986 ALLRIGHTS RESERVED

No.

2J313-002E*P

REVISION RECORD REV.

DATE!

REASON

PAGE /AUTHOR

CHANGED

No.

DOC PRODUCI

--__-_-

INI.

'86 Apr. 18

NC

Mr. Wakai

:A

:B

:C

:D

FE

:F

'G

*H

.I

'J

:K

FL

FM

:N

:O

'87 Sep. 11

Correction

'88 Feb. 20

Correction

'88 May 11

Correction

'88 Dec. 23

Correction

'89 May 26

Correction

'89 Nov. 27

Correction

'90 Mar. 22

Correction

'90 May 21

Correction

'90 Jul. 20

Correction

'90 Sep. 20

Correction

'90 Nov. 30

Correction

'91 Oct. 15

Correction

'92 Jan.

Correction

03/'92

Correction

Mr. Wakai

Mr. Ishiyama

Mr. Kimura

Mr. Ishiyama

Mr. Kimura

Mr. Kimura

Mr. Kimura

Mr. Mochizuki

Mr. Ishiyama

Mr. Ishiyama

Mr. Kimura

Mr. Kimura

Mr. Kimura NC Mr. Ishiyama

.P

111'93

P.25, 52

Correction Mr. Ishiyama

*

R-l

No,

25313-002E*J

CONTENTS

Page 1.

INTRODUCTION

2.

PREPARATION

3.

______________-_-__---~-------~~~-~--~--~~-------~

1

______________________________I_________-~~-~---~--

1

2.1

Test Tools

2.2

Precautions

2.3

Radiation

BEFORE

TURNING

3.1

Visual

3.2

Power

4.

SYSTEM

5.

PERFORMANCE 5.1

5.2

Instruments

---_-_-----__----_----

Required

____-________-___-__~-~----~~----~--~_~___~___

2

_____________-______-------------___-----

3

POWER

ON -------_--_------------------a--_--a--_

5

-----_--_-_-----_----------------~~-~~-~

5

Source

5

_--_---_--_------__----~-~~~------~-_-~--_~-_

5

_---__--___--_--_-_----~~---~~~--~-~--~-~-~-----~

CHECK Test

-_-_-_---_____-_--__-_~-~-~_--~-~--~____-__-_

6

___-__-_____-_--____-~-~~~-~-~~~-----~~~--~~~~~~

8

5.1.1

Power

5.1.2

Checking

voltage

5.1.3

Checking

operations

5.1.4

Checking

the switching

5.1.5

Checking

the starter

circuit

5.1.6

Checking

the battery

_---_-_-_____--____-----------

Radiography

1

Safety

Inspection

SETTING

Unit

and

ON -----_---__-_-----_------------------

Adjustment

-----

-------__------------~~----~-----and display

circuits

----------

8 10

--------------------

11

----------------------

11

circuit

-_------_-_---___-----__-_--_------

Radiographic

tube current

adjustment

--------------

15

5.2.3

Radiographic

tube voltage

adjustment

--------------

24B

5.2.4

Checking

Automatic

5.4

Fluoroscopy

Exposure

Adjustment

5.4.1

Measuring

5.4.2

Fluoroscopic

-------------------

29

----------------------

30

______-__------------

30

-------------------

30

timer

Control

(KXO-30F/SOF)

instrument tube

-a-

connection

current

-_-_-

13

5.2.2

the radiographic

_---_______---

13

Measuring

Checking

connection

12

5.2.1

5.3

instrument

8

adjustment

--------------

30



No.

2J313-002E”B

CONTENTS

Page

5.4.3 5.5

ic

Fluoroscop

HHS Tests

tube

voltage

adjl .is tment

31

--_--_-_---___

32

____________________-______-_-_--_~--_--_~-~_~~~

5.5.1

Fluoroscopic

5.5.2

Reproducibility

5.5.3

Automatic

exposure

control

limit

5.5.4

Automatic

exposure

minimum

exposure

entrance and

exposure linearity

rate

32

---------------

32

--___________________

32

__________________ time

32

----------

6.

SYSTEM ACCURACY ___--_---__--__-____--_-~-_______-~~--~------~-

33

7.

ATTACHED DATA __-__-------_-_-___-___-~_--~_-----~-~-~~-~~~----

33

7.1

PWB Setting

7.2

Error

Code

and Table

Adjustment

Tables

_________-__---_-__-----

--_______________________________-__--_

8.

MAINTENANCE SCHEDULE __-___-________-_-_----_----_-e-----------w

9.

APPENDICES

-__-____-___-___-_____________-___~__--__----~-~

-b-

33 ---

70

71 ---_

71

No. 2J313-002E*A

1.

INTRODUCTION This manual applies to the diagnostic X-Ray Generators, models KXO-30R, KXO-30F, and KXO-50F, and describes the system setting, the electrical and test to ensure the proper X-ray performance check, and the adjustment exposure precision. Before adjusting the generator as described in this manual, check that the equipment has been installed according to the installation manual (2C313-020E) and understand operating method given in the operation manual. Refer to the wiring diagram the adjustment or test.

supplied

with

These tests are to be performed after requirements of the Radiation Control and all regulations required in Title Regulation.

the equipment

when

performing

installation to meet the for Health and Safety Act of 1968 21 of the Code of Federal

PREPARATION

2. 2.1

Test Tools and Instruments Required Prepare the measuring instruments given

Measuring No.

below.

instrument Remarks

Hodel

Item 1

; AC voltmeter

2

I ’ Digital

3

, Oscilloscope

Full range:

300/6OOV

voltmeter Tektronix

434

Or equivalent Or equivalent

4

X-ray tube voltmeter (bleeder resistance included)

ALCO kV201(D)

5

X-ray secondary tube ammeter

ALCO MA-12OlDC



Or equivalent

i I / Acrylic phantom

180mm thickness

Extension PWB (28 AB type) -l-

No.

2.2 2.2.1

2J313-002E*M

Precautions Introduction in X-ray work is protection the most important considerations X-ray equipment properly used does the hazards of X-radiation. careless or improper usage can be not pose any danger; however, it is essential that all service or operating dangerous. Therefore, personnel of this equipment must fully understand the problem themselves. They should also be thoroughly familiar and acquainted with the instructions in this manual pertaining to proper use of the equipment.

One of against

2.2.2

Maximum permissible

dose

Various studies on the effects of X-ray radiation have provided the foundation for establishing maximum permissible dose of X-radiation. The results of these studies used by the National Council on Radiation Protection and the International Comittee on Radiation Protection to develop recommendat ions for the maximum permissible does (MPD). The I.C.R.P. recommendations can be interpreted to say that the maximum permissible dose (MPD) to the most critical organs, accumulated at any by the number of age, shall not exceed 5 x 10 -2Sv (5 rems} multiplied years beyond age 18 and the dose in any 13 consecutive weeks shall not for those involved in radiological Thus, exceed 3 x 10s2Sv (3 rems}. works, accumulation MPD equals 5(N-18) x 10W2Sv {5(N-18) rems), where N For the hands, forearms, and is greater than 18. is the age in years the MPD is set at 75 x 10T2Sv and ankles of radiologists, feet, (75 2.2.3

rems}

a year.

Manufacturer’s Although

this

responsibility equipment

incorporates

protection

against

X-radiation

design cannot provide complete other than the useful beam, practical does not compel the operator protection. The design of the equipment or his assistants to take adequate precautions, or prevent the possibility of authorized or unauthorized persons carelessly, unwisely to direct or secondary or unknowingly exposing themselves or others Therefore, only authorized and properly trained persons radiation. should operate this equipment. All persons authorized to use the equipment must be cognizant of the The equipment is sold danger of excessive exposure to X-radiation. Medical Equipment with the understanding of the TOSHIBA CORPORATION. neither assume nor accept any Division, its agents and representatives, which may result from exposure responsibility for any injury or danger TOSHIBA CORPORATION does not accept any to X-radiation. Furthermore, responsibility for over-exposure of patients or personnel to the X-radiation generated by this equipment due to poor operating no responsibility will be assumed for techniques or procedures. Also, any machine that has not been properly serviced and maintained in accordance with the manufacturer’s technical manual or which has been modified or tampered with in any way.

-

2 -

No.

2.3

Radiation (1)

2J313-002E*M

Safety

Radiation

protection

Since effects harmful to the health may be experienced by anyone amount of X-radiation, the greatest exposed to an excessive consideration should be given to protection against the direct Some of these effects are primary beam from the X-ray tube. The best a period of months or years. cumulative and may extend over is to stay out of the primary beam for X-ray protection safety rules at all times. beam produces a secondary or Any object in the path of the primary The intensity of this scattered radiation scattered radiation. depends upon the type and amount of the primary beam and the atomic This number of the objects or tissues struck by the primary beam. radiation may actually be of greater intensity than the image-forming radiation reaching the film. Since radia

this tion,

radiation protective

dangerous i s as potenti ally to meas ures mus t be taken

beam a s the primary it. sa fegua rd against

out with lead shielding These protective measures may be carried Items such as movable lead screens and lead-impregnated devices. can be used to minimize possible dangerous rubber gloves and aprons devices have a lead thickness It is imperative that these exposures . they should be used by those of a least 0.25 mm or the equivalent; persons exposed to radiation fields of 12.9 x 10-7C/(kg=h) {5mR/h) or During servicing or while operating this equipment a person’s more. or other parts of the body should unprotected hands, wrists, arms, never be exposed to the primary X-ray beam. (2)

Monitoring Monitoring adequacy improper radiation

of

personnel

a valuable crosscheck of personnel provides It can aid in revealing of radiation safety. radiation protection practices, and potentially exposure situations.

of the inadequate

or

serious

This should be confirmed by measurement in rads. Adequate protection measurement should be taken to determine that the X-ray radiation at A less any location does not exceed the accepted tolerance dose. means of confirming this protection is but frequently used, accurate, strategic locations for a given to place unexposed film at several The use of fluorescent screens, period prior to film development. to the darkenss, may also aid after the eyes have become accommodated in locating inadequate protect ion areas. Another common method of radiation monitoring is the use of film badges. These devices consist of an X-ray sensitive film inserted into a badge which incorporates metal filters of varying degrees of Even though this device can only measure the X-ray transparency . radiation reaching the particular area of the body it is worn on, it does furnish a clue to the amount of radiation that is being received. All these regarding

measures are X-ray protect

unless meaningless ion are known and

-

3 -

the rules are rigidly

of safety followed.

No.

Refer to the following requirements.

recommendations

for specific

2J313-002E;kB

safety

(3) Recommendations It is vitally important that everyone associated with X-ray work become fully acquainted with the recommendations of the National Bureau of Standards, the National Council on Radiation Protection, The control and the International Committee on Radiation Protection. of diagnostic medical X-ray equipment varies in detail from state to of the National state; however, in general, the recommendations Council on Radiation Protection based on the other authorities, are recommendations in this manual. However, prior to operation, all personnel qualified and authorized to operate this apparatus should become familiar with and, where pertinent, conform to the established recommendations contained in the documents provided by the The current sources of information are aforementioned authorities. listed for reference. a. National Council on Radiation Protection No. 33 (Medical Gamma-Ray protection for Energies up to 10 MEV-Equipment and Use).

X-Ray and Design

b. National Bureau of Standards Handbook No. 76 (Medical X-Ray Refer to N.C.R.P. Report protection up to Three Million Volts). No. 33. C.

Current Recommendations Radiation Protection.

of the International

-4-

Committee

on

No.

3.

BEFORE

TURNING

POWER

ON

Minimum check points before Do not turn the main switch breaker ON before the check 3.1

Visual

visual

inspection

(1) All setscrews loose. Various

relays

(3) All connectors (4) All PWBs 3.2

Power

Note:

SYSTEM Refer

unit

items

of the units

and distribution

should

be securely

of cables

are securely

terminal

should

inserted

assembly

inserted

not

are as follows: screws

into the

should

not be

sockets.

be loose.

in place.

Source

. Since prior power

4.

turning power on are given below. of the distribution panel and the power is completed.

Inspection

Internal

(2)

2J313-002E*B

the cables were connected for the three-phase 415 V power source to shipment, reconnect the cables according to the installed supply (in accordance with section 6.4 of manual No. 2C313-020E). In the case of a single-phase power source, change the charging (Change the TDR operating time within the time of the capacitor. power unit from 30 to 60 seconds.)

SETTING to Table

7.1 "Setting

and Adjustment

-5-

Table".

No. 2J313-002EfcD

5.

PERFORMANCE CHECK The performance check flowchart is given below: Note: Check to see that the shielding terminal of the cable (for serial transmission) is connected to the cabinet for noise prevention. If it is not connected, error 700 (a communication error> may occur.

PWB unit with control box

1

SPIO or SIO-1

PWB

Chassis

BSXlZ-0276(5 pairs, 7.2$, 10m)

[I.

\

Y

-6-

Do not connect the panel or stand-side shielding terminal.

No.

Start

i

1

2J313-002EkB

1 Cable power

Note:

connections supply.

must

comply

with

the

1 Connect

cables

in

accordance

with

the

system

connection

diagram. I

I

connected with Only the power supplies system cables can be used externally.

Note:

I

I

Disconnect

the

connector

from

the

trigger

PWB within

the

INV unit.

t System setting “when installed”

(for all columns

items indicated of tables in

Note : The charging power supply.

v

Energizing

i

I

as “necessary” section 7.1. > time

in

must

the

comply

with

the

)

I t

) Unit

test

I sequence Filament setting

(sections

5.1.1

5.1.6)

check

I

heating

check

rConnect

to

the

connector

data

and heating

(cable

length

and

tube)

t Power

to

off

the

I

trigger

PWB within

the

INV unit.

I Connect the connector citor voltage become

Radiographic

& fluoi!oscopic

adjustment

(sections

5.2

after the capaless than 5V.

and

5.4)

ABC/AEC adjustment

ATR adjustment

‘I (

End

Note : Enter )

the

adjustment

data

in

the

data

sheet.

No. 2J313-092E

5.1 5.1.1

Unit

Test

Power

ON

Turn ON the main switch of the distribution panel and the power unit Check for abnormality by turning the breaker within the control box. panel or stand power switch ON/OFF. Within approximately 60 seconds after the capacitor unit inside the control approx. 30 seconds). Note:

5.1.2

After the power is turned off, approximately four minutes.

Checking

the power switch is turned on, box is charged (three-phase:

discharging

is completed

in

voltage

(1) KXO-30R (a)

Voltage

within

the panel

Panel Cont 1 PWB Between CH2 (+5 V> and CHl

(GND):

If the above allowab!e voltage the VR of the stabilized power Control

box:

Pl witnin

Make sure that +24 v+3 - v. (b) Voltages Mother

the voltage

within

value source

Unit

V

is not indicated, given below.

adjust

1

between

the control

PWB of PWB Unit

. Between . Between . Between

Power

+5 V,O.l

CH3

(+24 V> and CHl

(GND)

box

1

Jl (IAB) and Jl (28AB) : +5 V+O.l V J16 (2AB) and Jl (28AB): +15 ~~0.1 V J16 (3AB) and Jl (28AB): -15 VzO.1 V

If the above allowable voltage values are not indicated, the VRs of the stabilized power source given below. p2 (+5 v), ~3 (2) KXO-30F (a>

is

(+I5 V> and P4 (-15 V> within

Power

Unit

adjust

1

or 50 F

Voltages . Desk

within display

Between

CH2

the stand PWB (+5 V> and CHl

(GND):

+5 V+O.l V

If the above allowable voltage value is not indicated, adjust Make sure that the VR of the stabilized power source (PSZ). V. the voltage between CH3 (+24 V> and CHl (GND) is +24 V+3 -

-8-

No.

. Desk

control

Between

PWB

P2 (+5

V> and

pl

If the above allowable the VR of the stabilized (b)

Voltages Mother

within

(3)

ths

PWB 2 of

. Between . Between . Between

PJl PJl PJl

2J313-002E*B

(GND):

voltage power

control

PWB Unit

(4AB) (2AB) (3AB)

+5 V+O.l value source

v

is not (PSl).

indicated,

box 2

and PJl and PJl and PJl

(28AB): (28AB): (28AB):

+5 V+O.l V +15 V~0.1 V -15 V+_O.l V

If the above allowable voltage values the VRs of the stabilized power source

are not indicated, given below.

Pl

within

(+5

LV detection

V),

P2 (+15 circuit

V),

adjust

and P3 (-15

V)

Power

adjust

Unit

2

adjustment

in following Adjust VR2, VR3 and VR4 on INV3 PWB as instructed Tl on the power unit table only when the input tap o f transformer This adjustment , however, is in the control box indicates 200 V. been set. not required when other taps have already

VR to be 1 adjusted

Measurement point

1 VR2

TP2

VR3 VR4

Tl

j

220 V,

200 v

415

V,

v

+4.75

v

v

+5.25

V

TP3

+4.87

TP4,

+5.35

/

Setting

-9-

380 v,

v

v

LED3 through After adjustment, the LV levels with respect to value) (see the table below).

tap

+4.50

+4.75

/

input

2

prior

LED7 of the INV-3 PW board the nominal value (Tl input

to

440

V /

shipment indicates tap

No. 2J313-OOZEW

Nominal

Other values

200V

nominal = Vs

LV>zzoVQc

LV>V$$-10%

220v~Lv>214v

INV-3

Radiography flu scopy disabled area

LV LEVEL

s m % i LED7

Operating area

vstlo%~Lv>vst5%

u R64

+i

9 LED6

vs t5% 2 LVL vs -5%

214’LV2195

Optimum

operat

u

04

R62

,Q u LED5

Operating

Vs-5%>LVLVs-10%

195>LV 2 190

area

:

R61

c z

5.1.3

Checking (1)

operations

and

Radiographic

(2)

the

Checking Repeating indications . . . . . .

circuits

settings

condition individual

reading ID No.

settings and indications operation manuals.

when

the radiographic according to the

turning

turning equal

power those

power

conditions operation

of

the

in and from manual, check

units

an for

the

ON/OFF

that ON, OFF, and ON, check before turning power OFF.

the

following

R kV value

R mAs value R mA value

R TIME value TEC. SW SF/LF

Make sure that the contents turning power ON and OFF. (4)

u R60

LED3

ID function Writing and appropriate ID function.

(3)

display

condition

Check radiographic as described in

zu% ii:

Radiography fluoroscopydisabledarea

vs-lO%>LV

19O>LV

LED4

Overload Check X-ray X-ray

written

in

the

ID are

not

destroyed

by

protection

function under conditions of the the overload protection tube maximum rating, filament characteristics, kV limit, generator rating (40 kW, 60 kW).

-

10 -

set and

No. 2J313-002E"cA

5.1.4

Checking

the

(1)

potential

High

switching

circuit

changer

(KXO-30F

or

50F)

(HRU-A) of the Operating the changer selecting TECHNIQUE via the panel or changer changes over normally. (2)

Focus

change

Checking There

the

are

(a)

Normal

(b)

High

(c)

Anode

panel

starter

three

of

stand

types

of for

make

for

for

HSS is controlled by the starter frequency wiring, X-ray

indication

focus

anode

radiography

rotation

rotation

tube

fluoroscopy

the is

rotation

as

(Named

NSS)

radiography

NSS and F-ROT are controlled The starter X-ray control. (50 or 60 Hz).

Check the installed

stand,

generator sure

by

that

the

varies

with

the

set

circuit

rotation speed

voltage

by R-MA

Check that the tube current.

5.1.5

high

(Named

(Named

shown

HSS)

F-ROT)

by the normal speed frequency equals the

high speed 180 Hz.

starter

phase advance capacitor tubes as described below.

below.

starter within the power source frequency

unit

(RS-118A:

value,

and

set

option),

time

of

and

the

[NSS and F-ROT modes] ( 1)

Check tubes (a) (b)

the type depending

XS-U (standard stator) XS-AA (1 ow impedance

For detailed tube technical (2)

of the X-ray tube combined. on the X-ray tube stator.

Phase

advance

The relation X-ray tube

X-ray

stator data.

tube

refer

to

are

two

types

phase below.

Capacity

advance

the

capacitor

Capacitor the power

TOSHIBA U coil

24uF

Cl

TOSHIBA AA coil

44uF

Cl

individual

value

No. within unit

+ c2

cl: 24uF, C2: 20uF

-

11 -

of

tube X-ray

value

between the type is given

type

TOSHIBA X-ray tube stator) TOSHIBA X-ray

characteristics,

capacitor

There

and

the

2J313-002E”cA

No.

(3)

Time

setting

The relation given below.

between

the

I

X-ray

tube

I / /

type

The oscilloscope measured.

(a)

Check

(b)

Set

that the

is

power

on

to

X-ray

tube

type

is

1

time

be

isolated

observe

from

the

betweenj

power unit within storage mode)

and

the

circuit

to

be

off.

oscilloscope

Block TB2 of the div, 0.2 sec/div,

and

+o. 1 set -0

should

power

time

Boost

I 1 0.8 I

TOSHIBA U coil

Note:

boost

select

Tube

the

and / X2 control

box.

of

Terminal

(100

VI

1.

(c)

Turn

(d)

Set a low kV value, a low mA value, Normal Speed Rotation (NSS) mode.

(e)

Put into the oscilloscope

and a short-time

Ready state. Observe the and check t’ne boost time.

waveform

on

to

keep

in

the

[HSS mode 1 (a)

5.1.6

Set the stator coil, start voltage, and control time control voltage, installation manual. speed starter

(b)

Select

(c)

Set a kV value, Rotation (HSS)

(d)

Put into rotation

Checking

the

the

TECHNIQUE of

the

and

a mA value, mode.

the Ready according

state, to the

tube

as

to

R start time, specified in

F start time, the high

be checked.

a time

to

keep

in

and check the high speed said installation manual.

High

Speed

anode

battery

Each unit PWB is provided with a chargeable which stores the memory contents when power both terminals and check the voltage between given be low.

-

12 -

battery or turns off. according

dry battery, Turn power to the table

off,

No. 2J313-OOZE*1

Equipment

KXO-30F/50F

name

Cont

1 PWB

Voltage 3.6

V to 2.4

Heat 2 PWB of PWB Unit 1 within the control box

2.5

V

Stand

3.6

V to 2.4

Panel

KXO-30R

Unit

desk

control

PWB

Heat 2 PWB of PWB Unit 2 within the control box

5.2

Radiography

V (Chargeable)

V (Chargeable)

2.5 V

Adjustment

The adjustment method for tube voltage and current is given below. The R.mA and R.kV adjustment procedures are given separately below. and R.kV should be adjusted simultaneously at each mA step. 5.2.1

Measuring

instrument

The measuring

R.mA

connection

instruments

should

be connected

as shown

below.

(1) KXO-30R Set up the required units as illustrated the tube voltage and current. Control

source

generator

\ -9

and adjust

--

Secondary tube current meter (to be mounted to the X-ray tube) \

Anode Bleeder resistance

DG-303

Power ground terminal

both

box

G--J Power

below

/

Storage oscilloscope

-

13 -

\ f-A-X-ray tube

Cathode1

No.



Set up the required units tube current and readjust It is not necessary voltmeter-ammeter

Control

is

as it

to perform used for

illustrated if necessary.

below,

this procedure measurement.

box

25313-002E*l

check

if

the

the

X-ray

tube

Secondary tube current meter (to be mounted to the X-ray tube)

Anode X-ray tube

High

Power source

voltage -_:l 1

--

5

--Jr

Cathode

I

1 Power ground terminal

-I

. Do not generate X-rays Prior to checking the (within the TNV unit) l

with Tl and T2 opened. disconnect X-ray sequence, connector or the capacitor.

Note : Before disconnecting the capacitor voltage

the is

- 13A -

connector or 5 V or less.

the

capacitor,

TRG PW board

check

that

No.

(2)

KXO-30F

and

25313-002E*J

50F Secondary tube current meter (to be mounted to the X-ray tube) DG - 50F

Power

source

Control box

n \ &-7

Tl

Tl

_ T2

T2

---

High voltage gener-

lnode

ator

Cathode

Bleeder resistance

X-ray tube Cathod

--A

Tube voltmeter

-f-

Power ground terminal

Storage oscilloscope

. Do not generate X-rays to checking the . Prior (within the INV unit) Note:

with Tl and T2 opened. disconnect X-ray sequence, connector or the capacitor.

Before disconnecting the capacitor voltage

-

the is

14 -

connector or 5 V or less.

the

capacitor,

TRG PW board

check

that

No.

5.2.2

Radiographic tube current adjustment (Applicable to products manufactured from April, 1989 to Refer to “Radiographic tube current adjustment” described for products manufactured in and after July, 1990.) (1)

Operation

switch

and

adjustment

The HEAT 2 PW board, which with the operation switches below:

cl>

Mode indic

x2>

Mode

tor

has an adjustment function, and indicators give in the

I

SW

MODE



Adjustment



S/R

switches

switch

-

S/R

-

June, 1990. on page 24B

mode

TEST

selec

15 -

25313-002E*J

is provided drawing

25313-002E*J

No.



Mode indicator from 0 to This LED changes mode selector switch. Each mode is to be used as

1 to

2 and

follows

finally

to

according

to

7 with the

the

adjustment

type* Mode

I

0

Adjustment

Indicator

I

1

2

Type

moo

Standby

a.0

tube current Radiographic (Point adjustment >

adjustmnet

000

Radiographic tube (Line adjustment)

current

adjustment

current

adjustment

state

3

moo

4

00.

Fluoroscopic

5

0.0

PRH and

6

000

Cable

7

000

Tube setting error>

tube

PRF level

length

adjustment

setting (recovery

from

a data

0 ON (Lighted) Pressing

the

S/R

switch

causes

any mode

to

be

reset

to

Mode

0.

Mode selector switch The mode selector switch is used to select the mode and to check the main heating level. Each time the switch lever is set in the downward position, the The mode mode is switched from 0 to 1 to 2 and finally to 7. is then reset from 7 to 0. When the switch lever is set in the upward position, only the filament heating circuit becomes ready for main heating so that the heating adjustment level can be checked operations (in Modes 0, 1 and 2). Adjustment switches These switches are level and to input In Modes 1, code.

If

that

the

A, sl

caused

caused

or decrease cable length

the heating and the tube

switch

is

pressed,

voltage

at

Eo increases.

switch

is

pressed,

voltage

at

Eo decreases.

switch

is

pressed,

the

voltage

increases

ten

times

the

voltage

decreases

ten

times

by pressing switch

that

used to increase the high-voltage 2 and 4,

is

the pressed,

by pressing

(Note)

the

- 16 -

NO. 25313-002E*J

Note:

Eo is a check pin check pin AGND is Eo corresponds to transformer.

on the HEAT 1 PWB. used as the standard the primary voltage

Voltage for of

at

the

that at Eo. the filament

Whenever each of these switches is pressed, the buzzer will will increase or sound and, in Mode 1 or 2, the voltage decrease to approximately 0.05 V (to approximately 0.5 V when If the switch continues to be the tenfold switch is pressed). the buzzer will automatically start sounding so that pressed, the result can be achieved as if the switch were pressed by the number of times the buzzer sounded. for using these switches to input the The procedures in the high-voltage cable length or the tube code are described section

(2)

concerning

adjustment

methods

in

each

mode.

S/R switch This switch is used to reset Mode 0 from any mode or to input the high-voltage cable length and the tube code simultaneously using the adjustment switch. The same result can be achieved by setting the S/R switch lever in either the upward or downward position.

Adjustment Mode

Mode

procedure

6

7

Tube

code

setting /

Mode Mode

2 1

Radiographic adjustment

Mode

tube

current

tube

current

__,__ I I I I I I I I I

Mode 4

Fluoroscopic adjustment

(Not

required

First perform setting for (Only the cable three tubes. length setting will not cause the heating data to change. > Note: When the X-ray generator is used with an X-ray tube with a very small focus, such as the X-ray tube for magnification radiodo not connect graphy > tube H.T. cables until the tube code is set. Failure to do so may damage the tube focus.

for

KXO-30R) -

17 -

5

PRH/PRF

level

adjustment

No.

(3) Setting

the

high

voltage

cable

length:

Mode

6

The standard cable length is 15 m (for three tubes). change the cable length depending on the system.

--+



Set

Mode

6 with

Set the tube X-ray control If

the

No. to unit.

swith 1) cl 0

equipment

the

goes

is

into

mode

selector

the

data

If

necessary,

switch.

be changed

via

the

operation

the

indicator

input

state.

pressed,

25313-002E*J

panel

blinks,

of

and

the

the

Enter the new cable length using the adjustment switches and the S/R switch. If the two-digit (00 through 99) input is completed, the mode resulting in the condition in . indicator stops blinking, [Example:

6 ml

(Do not the

/ cl

First time

press switches)

aI

I

[-ZJ

4

[4

S/R

0

Press 4

Press L

[z.][-zj m 1-q

Second time

0 If

Perform voltage

0

[ZJ 0

0

0

Note:

Press

the

data

is

the operations cable length

When the S/R

switcll

the

tube.

0

1

1

correctly (2) for

set,

BCD code

0 1

0000 0001

2 3

0010 0011

4 5

0100 0101

6 7

0110 0111

8 9

1000 1001 -

0

aI

S/R

a buzzer

goes

through when changing the 2nd or 3rd tube.

adjustment is completed, or set Mode 7 with the

Value

Press ---f

18 -

reset mode

on. the

Mode 0 by pressing selector switch to

high-

the set

No.

(4)

Tube

code

setting:

Mode

7 Select set as initial setting. from Table 5.2.2 and perform

The DRX-2903HD (for 3 tubes) is each tube number to be combined setting by the following steps. --+

for

of

setting,

tube setting Note : If data adjustment into memory. The small

above

procedure

focuss.

is

the

is is

used

When the

setting, the buzzer again.

2nd

and

press

3rd

the

the indicator sounds and

goes the

tubes.

S/R

switch.

(Restores

completed after other adjustments, deleted and preset data are stored

to

set

the

switch

is

data

with

pressed

both in

step

large ,

the

and data

with either a large or small focus can be set. This setting can be used when the tube to be combined is not listed in Table 5.2.2 and large and small-focus data are to be substituted or when either large or small-focus data with different tube data, is to be reset. for example) is input for a blank tube name If any code (00 0000, the buzzer will sound for about two seconds, column in Table 5.2.2, thus indicating that there is no tube for the input code.

-

19 -

No.

23313-002E*J

is noted during the inputting operation or it is If an input error desired to input data once again, reset the Mode 7 initial state (with the indicator lit) by pressing the S/R switch several times. (5)

Radiographic

tube

current

adjustment

For the radiographic tube current adjustment, Mode 1 is used to ad just each adjustment point, and Mode 2 is used to adjust the tube current (MA) for each line. The radiographic tube current by first adjusting 60 kV in Mode 2 adjustment can be facilitated and then adjusting 40 kV, 90 kV and 120 kV in Mode 1. Adjustment

for

each

Adjustment for 120 kV in Mode When the adjustment The are

point

40, 1.

in Mode

60,

1

Adjustment

90,

for

each

Adjustment

for

MA line

in Mode

60 kV in Mode

2

2.

tube code is set for mammography in Mode 7, four points at 20, 25, 32 and 45 kV are required.

adjustment points for 20, 25, 32 and 45 kV.

mammography

The line adjustment should be a; 25 kV.

for

mammography

25kV 20kV 1 I]

800mA 1 Adjustment for 20, 45 kV in Mode 1.

25,

Adjustment

32 and

I for

25 kV in Mode 2.

tubes) is set before shipment. The standard data (for three Correct the data depending on combinations of tubes. Since the preheating and preflash are not checked, adjust the (Generate X-rays current after the filament heating is stabilized. by pressing the READY switch and the X-ray switch at intervals of them simultaneously. Take note (Do not press more than 2 seconds. of the heating data as shown in table 10. (1) in APPENDLX 01 after adjustment.)

- 20 -

No.

Set Mode

1 with

the mode

selector

25333-OOZE*5

switch.

Set the tube number (1, 2, or 3) to be adjusted operation panel of the X-ray control unit.

via

the

Set R.kV and R.mA to be adjusted in the same way. Some combinations of R.kV-R.mA are limited to Maximum Power (30 kW/SO kW). R.kV: 40, 60, 90, 120 kV or 20, 25, 32, 45 kV R.mA: 10, 13, 16, 20, 25, 32, 40, 50, 64, 80, 100, 125, 160, 200, 250, 320, 400, 500, 640 mA :

Correct

the data with

the

/c i/( cl ;

increase)

or

A, '1 (decrease) n

switches (Note 1) by generating X-rays under the individually set conditions in order to adjust the radiographic tube Adjust all conditions current. to be used in this way. Press the Ready SW and the X-ray SW simultaneously the set conditions again to see if it is necessary preheating or calibration of preflash level.

and confirm to perform

After checking the above, if the tube voltage is stabilized within a specific time, Item (6) "Preflash level adjustment" level adjustment" are not required. and (7) "Preheating When completing Mode 0. Note

1: When

adjustments,

press

switch

the

increased the B]

the S/R switch

or decreased or [q

switch

is used,

ten times

to return

the value

as compared

to

can be

to when

is used.

in this mode causes 40, 60, 90 and 120 kV to The adjustment simultaneously increase or decrease. is the same as Mode 1, except that this Mode 2 adjustment adjustment should be made with the R.kV tube voltage set to 60 kV. When the X-ray tube for mammography is used, the tube current adjustment in Mode 2 should be done at 25kV. (6) Preflash

level

adjustment

(PRF): Mode

5

The preflash level should be adjusted only when the tube current does not rise within the specified time in (5) . Preflash is intended to increase the X-ray tube filament rise rate. After the READY switch is pressed, the preflash is done prior to main heating. ~Preflash Heating

I

level

Main heating

Preheating

I READY w ON )) - 21 -

Time

No.

25313-002E*J

time is fixed and the amount of For this unit, the preflash preflash can be obtained by multiplying the amount of main heating by the preflash factor. the preflash level is adjusted by changing the preflash Therefore, factor. The preflash factor curve is generated by preflash factors at three The three points are the maximum, minimum and middle points. points of the filament current on the X-ray tube current characteristics curve. The three preflash factors, PRFl, PRF2 and PRF3 are to be selected with

the adjustment Tube

Tube

and

[,,[a, Current

[Tg

switches*

Characteristics

factor

current (l.mQ

Middle

Minimum Filament PRFl, PRF2, and PRF3 selected as follows:

Maximum

Current

(A)

for the three

adjustment

points

can be

Plot the X-ray conditions under which the preflash level is to be If the curve. adjusted, on the tube current characteristics If the filament current is near the maximum value, adjust PRFl. filament current is near the middle or minimum value, adjust PRF2 or PRF3, respectively. If the filament current is between the minimum and middle values, If the filament current is between the adjust PRF3 and PRF2. middle and maximum values, adjust PRF2 and PRFl. --+

Set Mode

the mdoe

5 with

selector

switch.

Set conditions under which the preflash level is short in (5) , via the operation panel of the X-ray control unit.

Select

the preflash

m,

m,

PRF3

PRFZ

sequence

factor

or [z]

to be

switch.

adjusted

The mode

PRFl

proceeds

to the next

- 22 -

step.

by

pressing

indicator

the

blinks,

and the

No.

While checking the tube current rise the READY and X-ray switches, change using

the

[z]

or

Increase L

[a

switch

1).

Decrease

When the adjustment S/R switch. 1:

by simultaneously pressing the preflach factor by

(Note

When the S/R switch is pressed, and the state in is reset. another focus or another factor, adjusted.

Note

25313-OOZE*5

When the

is

the indicator In this state, tube preflash

completed,

[z]

or

[A

reset

switch

Mode

is

stops blinking another preflash factor can be

0 by

pressed,

pressing

the

the

buzzer

Increase Decrease will sound and the preflash factor will increase or If either switch continues to be decrease by 1%. the buzzer will automatically start sounding pressed, so that the result can be achieved as if the switch were pressed by the number of times the buzzer sounded.

( 7) Preheating

level

adjustment

(PRH) : Mode

5

If the adjustment This adjustment is not necessary in principle. in (6) cannot be carried out satisfactorily depending on the system, this adjustment should be carried out as an auxiliary adjustment to (6).

r

Set Mode

5 with

Set the conditions

the

switch

Press

sequence

is

proceeds

11,

selector

switch.

the

next

switch

preflash panel

the

pressed,

to

the Increase

(Note

mode

under which the via the operation

not satisfactory, unit. When the

I

the

to

mode

of

blinks

is

and

stop. adjust

the

preheating

level

Decrease

and check

When the adjustment S/R switch. 1:

adjustment X-ray control

indicator

th e

tube

current

rise

by generating

When the S/R switch is pressed, the indicator In this state, and the state returns to . another tube preflash factor can be adjusted.

Note

level the

When the m

is

completed,

or [x]switch

reset

is

Mode

pressed,

X-rays.

stops blinking another focus

0 by pressing

the

buzzer

or

the

will

If either sound and the preheating level will change. switch continues to be pressed, the buzzer will sounding so that the result can be automatically start achieved as if the switch were pressed by the number of times the buzzer sounded. -

23 -

No.

(8)

Changing

the

The following adjusted to

Focal

Size

for

Fluorography:

Mode

25313-002E*J

6

procedures permit to change the focal size the small focus (SF) at the time of shipment.



Set

Mode

6 with

the



Use the operation the number of the

mode

selector

already

switch.

panel of the X-ray control tube whose fluorographic

unit to specify focal size needs

to

be changed. switch enabling

mode,

the

To adjust the switch only. For

the

to

place

following

focal

size

focus

(SF),

small

the

mode

to

the

large

tilt

the

focus

S/R

switch.

A buzzer procedure

sounds and the indicator stops has been performed correctly

Reset to mode “0” by means of the size adjustment has been completed.

the

focal

size

in

blinking

steps.

the

To change

indicator

to

S/R

(LF),

switch,

tilt

while

blinking so far. switch

the

if

when

pressing

the

the

J

L

the

focal

size

to

focal

LF Do not

To change

S/R

press

any

switch

Tilt

SF

I+ (9)

Recovery If

from

a data

any

error

data

has

occurred,

recover

When the S/R switch is pressed alarm stops and Mode 7 is set.

(2)

When the

A, sl

recovery

routine [z]

down

with

the

this

switch

Tilt

error



When the

Hold

switch

is

pressed,

while

the

the

following

alarm

indicator

procedure:

sounds,

blinks

the

and

starts. and

switches

[q

are

pressed

simultaneously,

the indicator goes out and the recovery routine (for Upon regeneration of an error check code) is executed. the indicator lights up and the completion of execution, sounds.

the

buzzer

Reset Mode 0 by pressing the S/R switch and check the data. the power supply is turned off in the adjustment mode during is judged to have be destroyed tube current adjustment , data -

24 -

If

No.

2J313-O02E”CJ

when the power supply is turned on again, and the alarm sounds. (This is because it is not possible to distinguish data change recover by using the from data breakdown. > In such cases, After recovery, data adjusted prior to above procedure. turning off the power supply are all effective. Note:

is simply to regenerate the error The recovery routine but not to repair the destroyed data. check code, it is necessary to check the data after Therefore, recovery. If the check shows that the data are changed, correct If data have been completely in Mode 1, 2 or 4. In this it is necessary to reset the tube. destroyed, the setting must start with the setting of the case, high-voltage cable length.

it

> Alarm

Stored data check The stored data is checked by heat data when the power supply is turned on. Data is checked in each tube and each focus (for data blocksTube 1 LF, Tube 1 SF . . . Tube 3 SF). If an in which an error has attempt is made to use the data block an alarm (intermittent beeping sound) and an been detected, To recover from this state, error message will be generated. execute recovery procedure (9) from a data error.

Set condition check than 150 kV or the tube If the set tube voltage becomes greater an alarm (intermittent beeping current is not set correctly, If the tube voltage or the tube sound) will be generated. the alarm will be removed. current is set correctly, Maximum filament current check smaller than the The heating output data is set to a value maximum filament current to prevent the X-ray tube filament If the heating data exceeds the maximum from breaking down. filament current value when the READY switch is pressed, the to the preheating level and an alarm heating level will be set (intermittent beeping sound) and an error message will be generated.

-

24A -

No.

5.2.3

Radiographic (Applicable (1) Operation The HEAT

tube current adjustment to products manufactured switch

and adjustment

in and after

July,

25313-002E"J

1990.)

mode

function, is provided 2 PW board, which has an adjustment switches and indicators given in the drawing

with the operation below:

xl>

Mode

indicator

C TEST



Mode

selector

switch MODE



x4>

Adjustment

switches

-

S/R

S/R switch

- 24B -

No.



Mode indicator This LED changes from 0 to mode selector switch. Each mode is to be used as

1 to

2 and

follows

finally

to

according

to

25313-002E*J

7 with the

the

adjustment

type* Mode

Adjustment

Indicator

I

0

**

l

1

2

Standby

Radiographic tube (Point adjustment)

current

adjustmnet

00.

Radiographic tube (Line adjustment)

current

adjustment

Fluoroscopic

current

adjustment

moo

4

00.

T-- 0.0

6 7

I

tube

PRH and PRF level

adjustment

000

Compensation

for

000

Tube error

(recovery

0 Pressing

state

0.0

3

5

Type

the

S/R

switch

setting >

tube

current from

drop a data

ON (Lighted)

causes

any

mode

to

be

reset

to

Mode

0.

Mode selector switch The mode selector switch is used to select the mode and to check the main heating level. Each time the switch lever is set in the downward position, the The mode mode is switched from 0 to 1 to 2 and finally to 7. is then reset from 7 to 0. When the switch lever is set in the upward position, only the filament heating circuit becomes ready for main heating operations so that the heating adjustment level can be checked (in Modes 0, 1 and 2). Adjustment switches These switches are level and to input code. In Modes 1,

If that If that

the

7 0

caused the

/) LI

caused

used to increase the high-voltage 2 and 4,

or decrease cable length

switch

is

pressed,

voltage

at Vset

switch

is

pressed,

voltage

at

switch

is

pressed,

the

by

pressing

switch

is

by pressing -

the pressed, the 24C -

y4 cl

voltage

the heating and the tube

increases.

Vset

decreases. (Note > increases ten times

switch. the

voltage

decreases

ten

times

No.

Note:

Vset is a check pin on check pin AGND is used

25313-OOZE”5

the HEAT 1 PWB. Voltage as the standard for that

at at

the Vset.

Whenever each of these switches is pressed, the buzzer will sound and, in Mode 1 or 2, the voltage will increase or If the switch is held down, the buzzer will autodecrease. matically sound at fixed intervals; the number of times the buzzer sounds is the same as pressing the switch the same number of times. The procedure for using these switches to input the tube code is described in the section concerning adjustment methods in that mode. S/R switch This switch is used to reset Mode 0 from any mode or to input the high-voltage cable length and the tube code simultaneously using the adjustment switch. The same result can be achieved by setting the S/R switch lever in either the upward or downward position. (2)

Adjustment Mode

7

Tube

Mode Mode

procedure

code

setting

2 1

Mode

Radiographic adjustment

tube

5

current

Mode

4

Note

1:

When the X-ray generator is used with an X-ray tube with a such as the X-ray tube for magnification very small focus, radiography, do not connect tube H.T. cables until the tube code is set. Failure to do this may damage the tube focus.

Note

2:

Diagnostic X-ray generators manufactured after April 1990 is turned ON without connecting cause an error if the power X-ray tubes. Be sure to remove the HEATlA PWB and to ground 12B on the HEAT2 PWB before turning the power ON.

-

24D -

No.

(3)

Tube

code

setting:

Mode

7 Select set as initial setting. from Table 5.2.2 and perform

The DRX-2903HD (for 3 tubes) is each tube number to be combined setting by the following steps.

Set

Mode



Set the

the tube number to X-ray control unit.

When the the

the

code

is

code

selector

be

into

goes

the binary 5.2.2.

[Example]

mode

switch

/) CJ

equipment

Enter Table

Binary

7 with

25313-002E*J

set

switch.

through

the

operation

panel

pressed,

the

mode

indicator

blinks,

the

input

code

and

state.

to

corresponding

of

the

tube

to

be

used

from

DRX-1203B Do not

press

a

[d

00 0010

0

Tilt &

0

First

time

S/R

? Eold

down

0

Note:

this

setting

After completion Mode 0. >

The small

0

Upon completion of correct Shortly after that, out. indicator lights up once

Perform

for

of

If tube setting adjustment data into memory.

above

procedure

focus.

is

the

is is

used

Tilt

S/R

0

setting, the buzzer

the indicator sounds and

goes the

again. and

tilt

3rd

the

tubes.

S/R

switch.

(Restores

completed after other adjustments, deleted and preset data are stored

to

set

switch

When the

1

A

2nd

setting,

Note:

this 1 switch

the is

data pressed

with in

both

large

step

,

the

and data

with either a large or small focus can be set. is not listed This setting can be used when the tube to be combined in Table 5.2.2 and large and small-focus data are to be substituted or when either large or small-focus data with different tube data, is to be reset. for example) is input for a blank tube name If any code (00 0000, the buzzer will sound for about two seconds, column in Table 5.2.2, thus indicating that there is no tube for the input code.

-

24E -

No.

25313-002E*J

If an input error is noted during the inputting operation or it is desired to input data once again, reset the Mode 7 initial state (with the indicator lit) by pressing the S/R switch several times. (4)

Radiographic

tube

current

adjustment

For the radiographic tube current adjustment, Mode and Mode 2 is used to adjust each adjustment point, current (MA) for each line. The radiographic tube adjustment can be facilitated by first adjusting 60 and then adjusting 50 kV, 90 kV and 120 kV in Mode Adjustment

for

each

Adjustment 120 kV in

Mode

for

point

50, 1.

60,

in Mode

1

for

each

Adjustment

for

Adjustment

90,

1 is used to adjust the tube current kV in Mode 2 1.

MA line

in

Mode

60 kV in Mode

2

2.

When the tube code is set for mammography in Mode 7, four adjustment points at 20, 25, 32 and 45 kV are required. The are

adjustment points for 20, 25, 32 and 45 kV.

mammography

The line adjustment should be at 25 kV.

for

mammography

1 20kV 1 25kV ( 32kV 1 45kV 20mA 1 I

50mA lOOmA

lOOmA

l

I

I

I

. . . 800mA 1

800mA

Adjustment for 20, 45 kV in Mode 1.

25,

32 and

Adjustment

for

25 kV in Mode

2.

tubes) is set before shipment. The standard data (for three Correct the data depending on combinations of tubes. Since the preheating and preflash are not checked, adjust the (Generate X-rays current after the filament heating is stabilized. by pressing the READY switch and the X-ray switch at intervals of Take note (Do not press them simultaneously. more than 2 seconds. of the heating data as shown in table 10. (1) in APPENDIX 01 after adjustment.)

-

24F -

No.



Set Mode 1 with



Set the operation

:I

Set

the

mode

selector

switch.

tube number (1, 2, or 3) to panel of the X-ray control

R.kV

and

R.mA to

be

23313-OOZE*5

adjusted

in

be adjusted unit. the

via

the

same way.

Some

combinations of R.kV-R.mA are limited to Maximum Power (30 kW/50 kW). R.kV: 50, 60, 90, 120 kV or 20, 25, 32, 45 kV R.mA: 10, 13, 16, 20, 25, 32, 40, 50, 64, 80, 100, 125, 160, 200, 250, 320, 400, 500, 640 mA

L-

Correct

the

data

with

the

y/ cl ;

(increase)

switches (Note 1) by generating X-rays set conditions in order to adjust the current. Adjust all conditions to be

or

/(/I n

(decrease)

under the individually radiographic tube used in this way.

Press the Ready SW and the X-ray SW simultaneously the set conditions again to see if it is necessary preheating or calibration of preflash level.

and confirm to perform

if the tube voltage is stabilized After checking the above, Item (5) “Preflash level adjustment” within a specific time, and (6) “Preheating level adjustment” are not required. When completing Mode 0. Note

1:

press

adjustments,

switch

When the increased the

[q

the

or

decreased

or [q

switch

is

ten

times

is

used.

S/R

switch

to

used,

the

as

compared

return

value to

can

to

be

when

The adjustment in this mode causes 50, 60, 90 and 120 kV to simultaneously increase or decrease. Mode 2 adjustment is the same as Mode 1, except that this adjustment should be made with the R.kV tube voltage set to 60 kV. When the X-ray tube for mammography is used, the tube current adjustment in Mode 2 should be done at 25kV. (5)

Preflash

level

adjustment

(PRF):

Mode

5

The preflash level should be adjusted only when the tube current does not rise within the specified time in (4) . Preflash is intended to increase the X-ray tube filament rise rate. the preflash is done prior to After the READY switch is pressed, main heating. / Preflash . Main heating I

Heating level

Preheating

1

READY mON ))

- 24G -

Time

No. 25313-002E*J

and the amount of the preflash time is fixed For this unit, preflash can be obtained by multiplying the amount of main heating by the preflash factor. the preflash level is adjusted by changing the preflash Therefore, factor. is generated by preflash factors at three The preflash factor curve The three points are the maximum, minimum and middle points. points of the filament current on the X-ray tube current characteristics curve. PRFl, PRF2 and PRF3 are to be selected The three preflash factors, with

the

[d,

adjustment Tube

Tube

and [z]

14,

Current

Characterist

switches. its

current

Pre flash factor

WI

Filament PRFl, PRF2, selected as

Maximum

Middle

Mi nimum

Current

and PRF3 for follows:

the

(A)

three

adjustment

points

can

be

Plot the X-ray conditions under which the preflash level is to be If the on the tube current characteristics curve. adjusted, If the filament current is near the maximum value, adjust PRFl. filament current is near the middle or minimum value, adjust PRF2 or PRF3, respectively. If the filament current is between the minimum and middle values, If the filament current is between the adjust PRF3 and PRF2. middle and maximum values, adjust PRF2 and PRFl.

Set

Set

Mode

5 with

conditions the

, via Select

the

m,

[a,

PRF3

PRFZ

sequence

the

mdoe

under operation

preflash or

proceeds

selector

which the preflash level is panel of the X-ray control factor

[a

switch.

to

switch.

be

adjusted

The mode

PRFl to

the

next

<

-

24H -

step.

by

indicator

short unit.

pressing blinks,

in

(5)

the and

the

No.

Change the (Note 1). L

preflach

factor

by using

S/R Note

or

@

switch

Increase Decrease the indicator In this state,

When the S/R switch is pressed, and the state in is reset. factor, another focus or another adjusted.

When the

[z]

the

25313-002E*J

adjustment

is

tube

preflash

reset

completed,

stops blinking another pref lash

Mode

factor

can

be

0 by pressing

the

switch. 1:

When the

[a

or

[q

Increase

switch

is

pressed,

the

buzzer

Decrease

will sound and the preflash factor will increase or If either switch continues to be decrease by 1%. pressed, the buzzer will automatically start sounding so that the result can be achieved as if the switch were pressed by the number of times the buzzer sounded. (6)

Preheating

level

adjustment

(PRH):

Mode

5

If the adjustment This adjustment is not necessary in principle. in (5) cannot be carried out satisfactorily depending on the this adjustment should be carried out as an auxiliary system, adjustment to (5). ---+

Set

Mode

5 with

the

Set the conditions not satisfactory, unit.

When the the Press

mode

switch.

under which the via the operation

switch

sequence

selector

is

proceeds

pressed,

to

the

switch

the

preflash panel

the

next to

mode

of

level the

adjustment is X-ray control

indicator

blinks

and

stop. adjust

the

preheating

level

Increase Decrease (Note

l),

th e

and check

When the S/R and the state another tube

1:

current

rise

by generating

the indicator switch is pressed, In this state, returns to . preflash factor can be adjusted.

When the adjustment S/R switch. Note

tube

is

When the[z]or

completed,

14

switch

reset

is

Mode

pressed,

X-rays.

stops blinking another focus

0 by pressing

the

buzzer

or

the

will

If either sound and the preheating level will change. switch continues to be pressed, the buzzer will can be automatically start sounding so that the result achieved as if the switch were pressed by the number of times the buzzer sounded.

-

241

-

25313-002E*J

No.

(7)

Changing

the

Set

Size

Mode 6 with

the

Use the operation the number of the be changed.

Press

the

mode,

enabling

To adjust switch

for

Fluorography:

Mode 6

to change the focal size procedures permit the small focus (SF) at the time of shipment.

The following adjusted to

Focal

the

the

For

the

the

Ed .

to

place

following

focal

size

pressing

small

selector

switch.

panel of the X-ray control tube whose fluorographic

switch

yh cl

while

mode

the

mode

unit to specify focal size needs

indicator

in

to

blinking

steps.

to

the

large

focus

(LF),

tilt

the

S/R

switch.

the

focus

already

(SF),

tilt

the

S/R

switch,

while

pressing

switch.

A buzzer procedure

sounds and ‘the indicator stops has been performed correctly

Reset to mode “0” by means of the size adjustment has been completed.

To change

the

focal

size

to

S/R

LF [Z]

switch

14 Hold

To change

the

focal

size

to

blinking so far.

for

tube

current

The dropout tube current can However, into the generator. level is too low or too high, procedures below.

drop:

Mode

Select

Enter

the

Press

the

tube

current

I/

switch.

I/a

-

mode which

selector is

The mode

245

focal

fl

G

3 switch

Ti;Tt

down

T

this

switch

Tilt

be compensated by entering a tube code if the preset compensation current it can be adjusted by following the



the

this

the

6

On the operating panel of the generator, for which the preset compensation current adjusted. 6 with

[a

the

SF

cl>

Mode

when

down

Hold Compensation

if

-

to

specify level

the tube No. needs to be

switch. be corrected.

indicator

starts

(See

Note

blinking.

1. >

No.

Change

the

(increase

preset

compensation

current

ecrease)

> and

level

switches.

When the current level is correct, the indicator stops blinking.

using

(See

set

the

S/R

25313-002E*J

the

Notes switch

yd cl

2 to

4.)

down;

that the adjusted current level for Generate X-rays to confirm (See Note 5.) the dropout tube current is correct. Repeat Procedure if the compensation current level requires or repeat from Procedure if it is further adjustment, the compensation current level for other necessary to correct tube currents.

(9)

Set

the

Note

1:

Note

2:

Note

3:

Note

4:

Note

5:

Recovery If

S/R

down

to

set

When either

the

y [Y]

Mode

0.

switch

Or [Zl

is

pressed

once,

the buzzer sounds once and the compensation current If the level is increased, or decreased, by one step. will automatically switch is held down, the buzzer sound at fixed intervals; the number of times the buzzer sounds is the same as pressing the switch the same number of times. When the compensation current level value reaches the the buzzer emits alternating upper or lower limit, short and long tones. time should be longer than 200ms. The X-ray exposure any

error

data

has

error recover

occurred,

When the S/R switch is tilted alarm stops and Mode 7 is set. switch recovery

to

Tube current values of 20, 50, 100, 200, 500 and 64OmA can be entered in Procedure . The tube voltage should be set to 90kV. Increase the compensation current level to correct for a fall in tube current during X-ray exposure and decrease it to correct for a rise in tube current.

from

a data

switch

routine

is

pressed,

with

while

the

the

the

following

alarm

indicator

procedure:

sounds,

the

blinks

and

the

starts. switches

are

pressed

simultaneously,

the indicator goes out and the recovery routine (for Upon regeneration of an error check code) is executed. the indicator lights up and the completion of execution, sounds.

-

24K -

buzzer

No.

25313-002E*J

Reset Mode 0 by tilting the S/R switch and check the data. If the power supply is turned off in the adjustment mode during data is judged to have be destroyed tube current adjustment, when the power supply is turned on again, and the alarm sounds. (This is because it is not possible to distinguish data change from data breakdown.) In such cases, recover by using the data adjusted prior to After recovery, above procedure. turning Note:

(10)

off

the

power

supply

are

all

effective.

The recovery routine is simply to regenerate the error but not to repair the destroyed data. check code, it is necessary to check the data after Therefore, recovery. If the check shows that the data are changed, correct it If data have been completely in Mode 1, 2 or 4. In this it is necessary to reset the tube. destroyed, the setting must start with the setting of the tube case, code.

Alarm

Stored data check The stored data is checked by heat data when the power supply Data is checked in each tube and each focus (for is turned on. If an Tube 1 LF, Tube 1 SF . . . Tube 3 SF). data blocksattempt is made to use the data block in which an error has an alarm (intermittent beeping sound) and an been detected, To recover from this state, error message will be generated. execute recovery procedure (9) from a data error.



Set condition check If the set tube voltage becomes greater than 150kV or the tube an alarm (intermittent beeping current is not set correctly, If the tube voltage or the tube sound) will be generated. the alarm will be removed. current is set correctly,

current check Maximum filament The heating output data is set to a value smaller than the maximum filament current to prevent the X-ray tube filament If the heating data exceeds the maximum from breaking down. filament current value when the READY switch is pressed, the heating level will be set to the preheating level and an alarm (intermittent beeping sound) and an error message will be generated.

-

24L -

No. 2J313-002E*P

If the alarm sounds, turn off the READY switch decrease the heating level. Table 5.2.2

Tube setting

. No.

Tube name

Binary code

No.

Binary code

Tube name

0

00 0000

32

10 0000

DRX-3535HC

1

00 0001

33

10 0001

DRX-3545HC

2

00 0010

DRX-1203B

34

10 0010

3

00 0011

DRX-1403B

35

10 0011

DRX-3603HD

4

00 0100

DRX-1415B

36

10 0100

DRX-3624HC

5

00 0101

37

10 0101

DRX-3634HC

6

00 0110

DRX-1603B

38

10 0110

1

00 0111

1 DRX-1615B

39

10 0111

’ DlU+3724HD/HC

8

00 1000

DRX-1725B

40

10 1000

DRX-3745HD/H

9

00 1001

DEU-1924HD

41

10 1001

10

00 1010

DRX-1724E

42

10 1010

11

00 1011

j

43

10 1011

/

12

00 1100

I DRX-2603HD

44

10 1100

( DRX-4234HC-S

1 DRX-2625C

45

10 1101

DRX-4634HC

7

1

13

1

00

1101

14

1

00

1110

15 16 17

/ 1

18

46

00 1111

1 DRX-2703HD

47

01 0000

1 DRX-2724HD

48



I

10 1110 10 1111

ISAPPHIRE 150TH

11 0000

IDXB-0335CS(A197)

11 0001

/DXB-0324CS(A192)

01 0001

/

49

01 0010

I

50

11 0010

DXB-0434ES(A256)

’ DFU-2825C

51

11 0011

DRX-B1124B-MO

52

11 0100

DRX-B1146B-MO

11 0101

DRX-B1356EC-MO

19

I

01 0011

20



01 0100

/

21

01 0101

DRX-2903HD

53

22

01 0110

DEU-2924HD

54

23

01 0111

24

01 1000

56

j

11 1000

25

I 01 1001

57

'

11 1001

DRX-8316HD

26



01 1010

58

11 1010

DRX-8336HD

/

55

(

11 0110 11 0111

27

01 1011

DRX-3103D

59

11 1011

28

01 1100

DRX-3224D

60

11 1100

29

01 1101

DRX-3224HD-H

61

11 1101

30

01 1110

62

11 1110

31

01 1111

63

11 1111

DRX-3335HC

1

- 25 -

No.

5.2.3

Radiographic (1) Prior Prior table

tube voltage

2J313-OOZE*1

adjustemnt

adjustment to X-ray 5.2.3.

exposure,

Table

Power

set SW1 on the INV-2

PWB

according

to

5.2.3

supply

Setting of SW1

position

Single-phase 200 V (190 to 215 V>

0

Single-phase

1

220 V

(216 to 240 V) Three-phase

380 V

4

(380 to 400 V) Three-phase

415 V

5

(401 to 425 V> Three-phase (426 to

440 V

6

1

Set SW5 on the INV2 PWB according to the table is required for PWBs without SW5.j

Length of the longest among those connected the three X-ray tubes

cable to

Setting of SW5

to 15m

r-

position

3, 2, 1 or 0

16 to 20m

2, 1 or 0

21 to 25m

1 or 0

26 to

0

- 26 -

below.

(No setting

No.

(2)

Tube

volt

age

25313-002E*I

adjustment

Perform tube voltage adjustments for each tube current setting. Although tube voltatges, 30 kV, 40 kV, 60 kV, 90 kV, 120 kV and 150 kV, are to be adjusted in principle, any voltage between tube voltage sections (see Note 2) can be adjusted. If the tube current value differs greatly from the value preset by means of the operation panel, the tube voltage cannot be adjusted. In such cases, therefore, the tube current adjustment should also be made along with the tube voltage adjustment. The delay time of the tube voltmeter should be 20 msec. [Adjustment within the INV-1 PWB using a storage oscilloscope for waveform comparison.

T

Tube voltage waveform

TP2

T Fine-adjust it CCW

VR3 by turning (towards "1").

i Gradual

increase

T Fine-adjust

(2) -we--_

I-

Gradual

decrease

it CW

-I

_L

(3)

F= Gradual

decrease

VR3 by turning (towards "3").

Gradual

-

decrease

28 -

If a similar waveform is observed, do not turn VR3. I In a single-phase 2OOV i system, the waveform may show a gradual decrease I under a large load. I

1

No. 25313-002Ek4L

5.2.4

Checking

the

Check that radiographic

radiographic

time correctly

the radiographic timer operates times as described below.

(1)

terminal Connect the monitor output input terminal of the oscilloscope

(2)

Turn

power

(3) Select

the

tube

to

which

kV, mA, and X-ray.

measuring

time

(6) The maximum radiation

Definition

power

and

the

defined time

as

as

specified

shown

tolerance

is

if

is

in

the

Fig. given

Item

connected.

7.2.2 (2), and

oscilloscope,

and

5.2.4. in

6.3.

off.

of

The radiation increases to decreases to

is

values

instrument

Read each radiation time from the waveform on check that it is within the allowable range. radiation

time

the

(5)

Turn

set

on.

Set the generate

(7)

several

of the tube voltmeter with a coaxial cable.

(4)

The

for

radiation time

is

75% of its 75% of its

time the interval peak value, peak value.

from Point A, where the tube voltage to Point B, where the tube volt age

/ 100%

75%

75% \

I

I

High potential waveform

Tolerance

Tolerance Fig.

5.2.4

-

29 -

No.

5.3

Checking

the

Automatic

The followings is installed.

5.4.1

check

should

(1)

Press

the

(2)

Check Refer

that the settings to the AEC-lOA/lOB

(3)

Perform manua 1,

@

Set

@

Set the If this timer.

@

Start

as

select

is

adjustment

control

is

(AEC)

performed.

the

AEC-1OA/lOB

installation

below.

kV and mA values. than an appropriate AEC radiation time. functions as an AEC backup or desk timer

generation.

(KXO-30F/50F)

instrument

connect

the

exposure

AEC.

in

given

Adjustment

Measur ing

X-ray

specified

time value longer the panel occurs,

X-ray

Fluoroscopic

to

automatic

of the AEC unit are correct. installation manual, 2C376-519E.

procedure

appropriate

when

after

TECHNIQUE button

the AEC test 26376-519E.

Fluoroscopy

Control

only

be performed

radiation

Connect 5.4.2

necessary

The

The

5.4

are

Exposure

2J313-002E;kB

measuring tube

ion

instrument

current

as

adjustment:

specified Mode

in

5.2.1.

4

can be adjusted individually point In Mode 4, each adjustment setting the f luoroscopic tube (F.kV) or all data concerning fluoroscopic tube current can be increased or decreased at

When F. kV is set to 60kV in Mode 4, all data can be increased or decreased at one time. Any F.mA setting can be made.

When F.kV is set to 50, 90, or 120kV in Mode 4, each adjustment point can be adjusted separately.

j

5OkV /

90kV

2mA

I

I I

I !

3rnA

I

I

I

50kV

/ 120kV

lmll

-

by the one time.

30 -

/

90kV

/ 120k

253 13-002E+B

No.

the

Therefore, adjusting

points

with

and

120kV

For

the

The

heating data after in APPENDIX 01.

11. (1)

tube

fluoroscopic

all

adjustment

F.kV

to

set

be

can

60kV

and

facilitated

then

by

adjusting

first

50,

90,

points. operation

Select

switches

(T u b e

SPOT

and

adjustment should

adjustment

1)

via

the

modes, refer to 5.2.2 (1). be recorded as shown in table

operation

panel

of

the

X-ray

control

unit. (2) (3)

Set

Mode

Set

the

4 with F.kV

and

part of the Combinations:

(4)

F.kV:

50,

90,

05,

1, the

switches

Note

3,

values

yA cl

by

order

generating

If

the

X-ray

correct

tube

Operation is (1)

the

is

completed,

or

or

tube

voltage

tube

current

below

VR6

[z]

tube

(2)

is

(VR6 VR6

within

7% of

If

tube

the

(1) above, radiographic possible the

off, In addition shown

,& Cl

under

operation of

on

ten

over,

this

Mode

is

F-unit.

pressing

the

as

compared

not

tube

adjustment

PWB)

the

INV-1

PWB so

all

value

be

can

to

possible

current

must

procedure

mA and

that

under in the

S/R

be

when

the

and

to

obtain

be

adjusted.

voltage

operation

the

measured

the conditions fluoroscopic

2,

(2)

kV values,

check

that

value of mode.

of

90kV

they

the

and

are

within 7% after adjustment PWB as in the case with the This adjustment, however, is

adjustment. When this mode fluoroscopic mode. tube voltage adjustment is possible. is performed for each voltage adjustment 1,

the

value.

voltage accuracy is not yet use SW2 and SW3 on the INV-2

0.5,

the

procedure.

INV-1

kV accurate

at

by

set current.

respectively.

may the

on

the

this

0

used,

times

used,

it

case

main

setting

tube

individual tube

only in the radiographic

below

spot part

(decrease)

fluoroscopic

switch

voltmeter becomes 90kV+lkV LED4 should indicate sT8”

1mA. By manually

and

reset

is

adjustment

adjustment

Fine-adjust

the

operation

adjustment

the

is

supplementary

the

decreased

In

voltage.

(1)

with

the

X-rays

ad just

switch

Fluoroscopic

adjusted or

(increase)

to

[z]

the

increased

5.4.3

be

table

4mA

adjustment

When

switch.

to

R/F

with 1)

in

1:

selector

120kV 2,

(Note

When the switch.

F.mA

data

conditions (5)

mode

conventional

F.mA: Correct

the

3 and

4mA respectively.

is

turned section

No.

Tube

voltage

As

for

section

50kV

operating

procedure

not to Note : Take care at the boundaries 5.5

to

63kV

of

64kV

SW2 and

to

SW3,

obtain the very at each section.

95kV refer

2J313-002E*B

96kV to

discontinuous

to

127kV

5.2.3. tube

voltage

HHS Tests After all of the preliminary and final tests to be necessary for the installation system have been successfully completed, the HHS test phase may be performed. The

5.5.1

HHS test

is

Fluoroscopic (1)

Refer

necessary entrance

to

Refer

to

Automatic Refer

to

and

Automatic Refer

to

the

ORH compliance.

required

and

rejection

limit

are

shown

required

and

rejection

limit

are

shown

linearity

APPENDIX 04

exposure

equipment

control

limit

APPENDIX 06

The test method, APPENDIX 06. 5.5.4

checking

rate

equipment

The test method, in APPENDIX 04. 5.5.3

finally

APPENDIX 03

Reproducibility (1)

when

exposure

The test method, in APPENDIX 03. 5.5.2

only

exposure

equipment

minimum

required

and

exposure

time

required

and

rejection

limit

are

shown

in

rejection

limit

are

shown

in

APPENDIX 07

The test method, APPENDIX 07.

equipment

-

32 -



No.

6.

SYSTEM

ACCURACY

I

I

Technique

No.

1

Applicable

factor

/ Maximum allowable deviation I

range

I

I

I

1

c

(Radiographic

TRadiographic 3

mA

/

all

I I

-+lO%

kV

1

all

~

27%

i

-+3ms

i

120%

24 to 99ms

I I

-+20%

0.1

to 9.999sec

I

-+lO%

0.2

to 2.9 mA

-+0.2mA

3.0

to 4.0 mA

-+lO%

/ Radiographic time Setting and indication in A.E.C.

1

1 Setting and indication in Ii I !ABC

6

7.

7.1

3 to llms for lOOmA

, I

for

'Fluoroscopic kV Setting and indication in ABC ) Fluoroscopic kV , time setting / and total time / indication

ATTACHED

lOOmA

or more or more

I

I

5

1

I

12 to 23ms

4 -/ Fluoroscopic

c

2J313-002E*B

all

I

I

I

-+7% I / I I

I I

+o _% -3

1 /

I

I

DATA

PWB Setting

and Adjustment

Tables

PWBs which are not listed in 7.1 require no setting and adjustment For the automatic exposure control PWB, refer they are installed. AEC-lOA/lOB installation manual.

- 33 -

when to the

No. 2J313-002E":D

(Panel) [PANEL CONT l] or TW CONT] l/8

PWB name:

A:

KXO-30R

B:

KXO-3OF/5OF

Equipment

When

Setting and adjustment

installed

B 0

VRl:

Adjusting the backup battery operating point There is not a VRl for KXO-30R=B.

Unnecessary

0

SWl:

Setting the serial transmission baud rate

Unnecessary

Switch settings

I 2

1

ON

Baud rate

OFF

OFF

ON

1200

ON

/ OFF 1 ON

2400

OFF

/ OFF 1 OFF

4800

OFF

9600

)

OFF

ON

ON

[W CONT SUB] l/7 or [W CONT 2/8

PWB name

i

4

OFF ON ,

3

Equipment B

/ I i-

Setting the X-ray generation buzzer timing

Jumper Jl: l

Open:

l

Short:

The buzzer goes on during radiation. goes on for approximately after the radiation is completed.

The buzzer

Necessary

50msec

t

L ON OFF

Necessary

Setting the system

sw2:

EP 1 2

'&El 3 4

5

Q

PU 6 7

8

A

Turn this switch ON only for USA. NOTE (Set this switch to OFF) MAMMO TOM0 GENE BKY 2 BKY 1

L

1

Turn switches used for the individual TECHNIQUES ON. (Set the other switches to OFF.)

Unused (Set this switch to OFF.)

NOTE _ This switch can be used for 100% of OLP testing only in the USA. ON: 100% Rating OFF: 90% Rating 100% Rating can be used for only testing; this switch shouid be

1

i

turned

OFF after

completing

the

- 34 -

100% Rating

test.

No. 2J313-002E9cD

PWB name:

[W CONT SUB] TW CONTl 4/8

3/7 or

A:

KXO-30R

B:

Equipment

When installed

Setting A 0

B j sw4:

KXO-3OF/5OF

For BKY 2 TECHNICQUiZ

Necessary

I 0

sw5:

For GENE

TECHNIQUE

Necessary

0

SW6:

For TOM0

TECHNIQUE

Necessary

For MAMMO Contents settings

Necessary

TECHNIQUE

of these of SW3.

switches

are

- 36 -

the same as the

No. 2J313-002E>kD

PWB name:

[W CONT SUB] 4/7 or [W CONT] S/8

KXO-30R

A:

B:

KXO-30F/SOF

Equipment

When installed

Setting

Sw8:

Necessary

Setting Tube 1

ONP1

OFF

U

P

Cl 51 Q

Q

P

-2

3

4

7

8

5

6

LL

X-ray tube type (12 types max.> Refer to next page. ISetting

the upper limit of the X-ray tube voltage. These switches can set tube voltages 15kV (maximum) lower than the maximum X-ray tube voltage.

I

l-i-4 SW No.

Upper limit

4

3

c

1 OFF

1 OFF

1 Max. -0kV

1OFF

/ ON

1 Max. -5kV

ON

OFF

Max. -1OkV

ON

ON

Max. -15kV

Setting the starter mode

SW No.

--r---l 1

1 ON

I

Mode

2

1 ON 1 Automatic change-mode

1

1 OFF 1 ON 1 High speed only mode

I

] OFF 1 OFF1 Normal speed only mode

1

I - 37 -

No. 2J313-002E"N

Sw8 (for USA and CANADA) SW No. 5

SW No. 6

SW No. 7

SW No. 8

CODE

TUBE TYPE

ON

OFF

ON

ON

B

DXB-0434ES(A256>(USA>

ON

ON

OFF

ON

D

DRX-3724HD/HC(USA)

ON

ON

ON

OFF

E

DRX-3624HC(USA)

ON

ON

ON

ON

F

DXB-0324CS(A192)(USA)

SW8 (for other count ies except USA and CANADA) SW No. 7

SW No. 8

CODE

TUBE TYPE

OFF

OFF

0

DXB-0324CS(Al92)

ON

OFF

2

DRX-3224HD-H

OFF

OFF

ON

ON

3

DRX-3224HD

OFF

ON

OFF

OFF

4

DRX-2924HD

OFF

ON

OFF

ON

5

DRX-3724HD

OFF

ON

ON

OFF

6

DRX-2703HD/HC

ON

OFF

OFF

OFF

8

DRX-1924HD

ON

OFF

OFF

ON

9

DRX-1603B

ON

OFF

ON

ON

B

DXB-0434EScA256)

ON

ON

OFF

OFF

C

DRX-2724HD

F

DRX-1724B

1

ON

ON

PWB name:

ON

ON

A:

[W CONT SUB] 5/7 or [W CONT] 6/8

Equipment Setting A 0

B sw9:

Setting Tube 2

- 38 -

KXO-30R

B:

KXO-30F/50F

When installed

No. 2J313-002E"D

[W CONT SUB] 6/7 or [W CONT] 7/8

PWB name:

A:

KXO-30R

B:

KXO-30F/50F

Equipment Setting A

B SWlO:

0

Necessary

Setting the display time of actual R-time when AEC is ON

-This value is set before shipment.

/ sw11:

0 1

Setting the UP/DOiJN key speed

1

I I

i

-~ -

39 -

Necessary

No. 2J313-002E*K

PWB name:

[W CONT SUB] 7/7 or [W CONT] 8/8

A:

KXO-30R

B:

KXO-3OF/5OF

Equipment

When installed

Setting A 0

B sw12:

Limiting the output tions of the user's

SW setting

depending on the condipower supply

Necessary

Contents

F . . . . . . . No limit to power up to 40kW

supply

0 . . . . . . . No limit to power up to 30kW

supply

1 , . . . . . . OLP at

for 51 msec or more 25 to 30kW

2 .,...*. OLP for 51 msec or more at 20 to 30kW

0

sw13: Setting

the exposure-time offset value The offset value added to the preset exposure time is the actual exposure time. Added time (msec> = O.Lt(Msec) x (setting

SW value)

The offset value can be set in 0.4 msec units in the range of 0 to 6.0 msec.

- 40 -

Necessary

No.

2J313-002E'kB

(Stand) [DESK CONTROL]

PWB name:

KXO-30R

A:

l/7

B:

KXO-3OF/5OF

Equipment

When installed

Setting A j

B

SW1

thru SW3:

SWl: sw2: sw3:

Baud

rate

for serial

For communication with the control For output to the printer For option RS-232C

Switch

box

settings Baud

1

rate

4

OFF

OFF

ON

ON

1200

Unnecessary

Jumper

0

Connect A-B, DSRA

0

Unnecessary

transmission

)

sw4:

-

jumers

as follows:

RTSA -RTSB, DSRB

Printer

CTSA -CTSB,

DTRA-

DTRB,

Necessary

connection

I

(OFF before

- 41 -

shipment)

No.

2J313-002E*B

(Stand) PWB name:

[DESK CONTROL]

2/7

A:

JLXO-30R

B:

KXO-30F/50F

Equipment

When installed

Setting A

B 0

sw5:

Setting

Setting

SW6:

user

Unnecessary

Destination

country

0

Japan

1

USA

2

Countries other Japan and USA

3 z F

0

for

Setting

than

Undefined

the exposure-time

offset

value

The offset value added to the preset time is the actual exposure time. Added time (msec> = O.Lt(msec) x (setting

exposure

SW value)

The offset vcllue can be set in 0.4 msec units in the range of 0 to 6.0 msec.

- 42 -

Necessary

No. 2J313-002E

PWB name:

[DESK CONTROL]

3/7

A:

KXO-30R

B:

KXO-30F/50F

Equipment

When installed'

Setting A

1 B 1 0 I

sW6:

Necessary

Setting the R-time offset

1 1 The switch sets the offset value of the set time.

(vs actual time)

I I ,

I I I I

o

SW7 thru SW12:

Setting the minimum time of mAs for each TECHNIQUE

I ’ Necessary

I /These switches set minimum time of mAs for each TECHNIQUE. , SW7 : i SW8 :

/

SPOT TECHNIQUE BKYl TECHNIQUE BKY2 TECHNIQUE GENE TECHNIQUE TOM0 TECHNIQUE MAMMO TECHNIQUE

SW9 : 1 SWlO: SWll: sw12:

Switch settings 112

3

OFF / OFF

4

OFF / OFF

5 OFF

6

7

Minimum time (msec>

81

OFF ’ OFF

OFF 1

0

OFF i OFF / OFF 1 OFF , OFF, OFF

OFF

ON

(

1

OFF / OFF 1 OFF

ON

OFF (

2

OFF

OFF

OFF

i

I 1

:

1 /ON 1 / ON

ON

1 ON

ON

1 ON

1 ON I / ON I 1 ON

ON

ON

/ ON I / ON

ON

ON

ON

/ I

OFF

ON

/

253

/ ON

ON

OFF j

254

ON

ON

ON

255

- 43 -

No. 2J313-002E*K

PWB name:

A:

[DESK CONTROL] 4/7

KXO-30R

B:

KXO-30F/SOF

Equipment

When installed

Setting A

B 0

sw13:

0

sw14:

01

02

0

Unnecessary

Output kW limit Set the limits if necessary customer's power conditions KXO-30F SW No.

Necessary due

to the

Limitation

5

6

7

8

ON

ON

ON

ON

No limit

up to 40kW

OFF

OFF

OFF

OFF

No limit

up to 30kW

OFF

OFF

OFF

ON

OLP

for Slmsec

or more

at 25kW

to 30kW

OFF

OFF

ON

OFF

OLP

for 5lmsec

or more

at 20kW

to .30kW

KXO-50F SW NO. Limitation

5

6

7

8

ON

ON

ON

ON

No

limit

up to 60kW

OFF

OFF

OFF

OFF

No

limit

up to 50kW

OFF

OFF

OFF

ON

OLP

for Slmsec

or more

at 45kW

to 50kW

OFF

OFF

ON

OFF

OLP

for Slmsec

or more

at 40kW

to 50kW

OFF

OFF

ON

ON

OLP

for Slmsec

or more

at 35kW

to 50kW

OFF

ON

OFF

OFF

OLP

for 5lmsec

or more

at 30kW to 50kW

sw15:

ON OFF

Selectable

APR program

Necessary

setting

00000000 12345678 I

Not us!d

A





‘l

1

Not used MAMMO TOMOGRAPHY

BAS1C

- 44 -

Table Bucky, Wall Bucky {General, Spot-shot I

NO.

Equipment

2J313-002E*C

When installed

Setting B 0

sw16

to 21:

max. mA code for mAs analysis when 1.1. is magnified

sw16: sw17:

upper 4 bits lower 4 bits

max. mA code for Normal size (X.1.)

SW18: sw19:

upper lower

4 bits 4 bits

max. mA code for Magnification-l (1.1.)

sw20: sw21:

upper 4 bits lower 4 bits

max. mA code for Magnification-2 (1.1.)

mA code

I

(Hex.)

mA (mA) 10

OOH OlH 02H 03H

13 16 20

OFH

! 320

For single size I.T., SW16 and SW17.

set the max.

- 44A -

mA code with

Necessary

No.

PWB name:

A:

[DESK CONTROL] 5/7

KXO-30R

B:

23313-002EAG

KXO-30F/50F

Equipment

When installed

Setting A

B 0

SW28:

Selecting TECHNIQUE to be operated when the APR is installed. (NOTE) Only one of 1 to 7 on this SW can be selected when SW30 or SW31 is used. ON OFF

1

PO 1 2

0 3

0 4

P 5

RRP 6 7

a



1



A

,

AL

1 I

/

//

1 I__

I

I

Unused

Necessary



All TECHNIQUES are operated. MAMMO TECHNIQUE TOMOGRAPHY TECHNIQUE SPOT TECHNIQUE

j

I

BKY2 TECHNIQUE BKYl TECHNIQUE GEN/CASSETTE TECHNIQUE

0

SW29: ON : OFF:

0

SW30 through SW31 are valid. SW30 through SW31 do not function. (The switch is set to OFF before shipment.)

Jumper Jl: Short:

0

sw30:

Necessary

Approving setting

Selecting KATAKANA

indication

Unnecessary

APR data are indicated in KATAKANA.

Copying APR data from ROM to RAM.

If this switch is turned ON, data equivalent to TECHNIQUE selected by SW28 is copied into the RAM area. (NOTE) Only one of 1 to 7 on SW28 can be used.

- 45 -

Necessary

No.

PWB name

[DESK CONTROL]

r

6/7

A:

KXO-30R

B:

KXO-30F/50F

Equipment

When installed

Setting A

B 0

Setting required only when combinations of DT-GCU/KDU .

is used with

Setting of positions/organs corresponding AU-APR operation selector switches Switches

to be used:

6, 7 and 8 combinations of SW13 1 . .. . . . . . .. 2 .......... 3 .......... 4 .......... 5 .......... 6 .......... 7 .......... 8 of SW28 0 .......... 1 .......... SW32 ........... SW17 combinations 1 .......... 2 .......... 3 .......... 4 .......... 5 .......... 6 .......... 7 .......... SW7 combinations 1 to 10, 15 or 20 . 0 to 12 . . . . . . . . . . .

.

AU-APR

2J313-002E*B

6, 7 and 8 of SW13, 8 of SW28, SW32, SW17

When

to the

and SW7

Contents for SW1 of AU-APR unit for SW2 of AU-APR unit for SW3 of AU-APR unit for SW4 of AU-APR unit for SW5 of AU-APR unit for SW6 of AU-APR unit the maximum in 1.1. fluoroscopy mode Contents AU-APR spot radiography mode AU-APR fluoroscopy mode Data writing Contents Selecting the head Selecting the breast Selecting the abdomen Selecting the spinal cord Selecting the lumbar vertebrae Selecting the arms Selecting the legs Contents Selecting an organ from those set with SW17 Determining the maximum mA code for 1.1. fluoroscopy Setting Setting Setting Setting Setting Setting Setting

Setting procedure a First set the numeric on SW13 corresponding to the switch number on the AU-APR unit to be set. SW17 and SW7 to the (@ Set SW28 (spot radiography/fluoroscopy), desired position/organ numbers. 3 Press the SW32 pushbutton switch (for write operation). If the buzzer the setting is completed. 4 If the buzzer sounds, 8 a SW setting error has occurred, and go back does not sound, to @ . @ Terminate the setting operation by setting 6, 7 and 8 of SW13 to "0". @ After the setting operation, reset SW13, SW28, SW17 and SW7 tc their original position.

- 46 -

No. 25313-OOZE"0

PWB name:

[DESK coNTRoLl

A:

7/7

KXO-30R

B:

KXO-30F/50F

Equipment

When installed

Setting A

B 0

SW32:

0

SW33:

Unnecessary

Setting the combined conventional R/F table

This switch sets the types of conventional R/F table.

Switch settings 12

3

4

5

6

7

8

Conventional R/F table

OFF

OFF

OFF

OFF

OFF

OFF

OFF

OFF

No combination

OFF

OFF

OFF

OFF

OFF

OFF

OFF

ON

DT-GCU/KDU/DUA

OFF

OFF

OFF

OFF

OFF

OFF

ON

OFF

DT-GCS

1 OFF

OFF

OFF

OFF

OFF

OFF ION 8

ON

DT-KEL

OFF

OFF

OFF

OFF

OFF

ON

OFF

OFF

DT-KEL-2

OFF 1 OFF

OFF

OFF

OFF

ON

ON

DT-MCS-2

OFF j OFF

OFF

OFF

OFF

ON

OFF (ON

OFF

DT-MCI-3 AUX.

i ON

ON

ON

ON

\

ON

ON

-

47 -

ON

ON

AUX.

Necessary

No.

PWB name:

[DESK DISPLAY] l/6

A:

KXO-30R

B:

2J313-002EfcB

KXO-30F/50F

Equipment

When installed

Setting A

B 0

Jumper Jl: Short: Open :

0

SWl:

Q

P

2

3

P 4

Necessary

Necessary

System setting

ONEl1

OFF

Setting the X-ray generation buzzer timing The buzzer goes on for approximately 50msec after radiography is completed. The buzzer goes on during radiography.

u u 5

6

El P 7

8

ATR MAMMO

TOM0 GENE BKY2 BKYl SPOT

1 Turn l

switches used for the individual TECHNIQUES ON, and set the other switches to OFF.

NOTE This switch can be used for 100% of OLP testing only in the USA. ON: 100% Rating OFF: 90% Rating 100% Rating can be used for only testing; this switch should be turned OFF after completing the 100% Rating test.

-

48 -

No.

PWB name:

[DESK DISPLAY]

A:

2/6

KXO-30R

B:

KXO-30F/50F

Equipment

When installed

Setting

SWZ: Setting

SPOT RADIOGRAPHY

Necessary

TECHNIQUE

Setting the maximum mA value small focus) of mAs 1

Setting the maximum mA value large focus) of mAS

t Steps

(for the 1 (for the) / I

0 to 3 Contents

Set vlaue 0

2J313-002EfcB

. . . . . mAs

decomposition by the maximum mA which is determined by the tube/unit

1 . . . . . mAs decomposition by mA at a step below the maximum mA which is determined by the tube/unit 2 . . . . . mAs decomposition below the maximum by the tube/unit

by m.A at two steps mA which is determined

3 . . . . . mAs decomposition below the maximum by the tube/unit

by mA at three steps mA which is determined

Setting ON: OFF:

t

the AEC detector SOLIDET detector Image intensifier or thin detector

Turn this switch ON when AEC is usable ON: OFF: AEC is unusable

-

49 -

photo

using

pickup

AEC.

No. 2J313-002E

PWB name:

[DESK DISPLAY]

3/6

Equipment , B A /

/

A:

KXO-30R

B:

KXO-30F/50F

When installed

Setting

0

sw3:

For BKY

1 TECHNIQUE

Necessary

0

sw4:

For BKY

2 TECHNIQUE

Necessary

sw5:

For GENE

TECHNIQUE

Necessary

0

SW6:

For TOM0

TECHNIQUE

Necessary

0

sw7:

For MAMMO

Contents settings

Necessary

TECHNIQUE

of these of SWZ.

switches

are

the same as the

iI

- 50 -

No. 23313-002E

[DESK DISPLAY] 4/6

PWB name:

A:

KXO-30R

B:

KXO-3OF/5OF

Equipment A

When installed

Setting

B

L

0

;

sw8:

Necessary

Setting Tube 1

ON a

E

u u a u

1

2

3

J

I

OFF

4

5

0 E

6

7

8

+

X-ray tube type (16 types max.> Refer to next page. 1

Setting the upper limit of the X-ray tube voltage. These switches can set tube voltages 15kV (maximum) lower than the maximum X-ray tube voltage.

Upper limit /

/SW.

Setting the starter mode

I SW 1

No.

I

I

I

Mode ,2

I

I~_ON ; ON OFF 1 ON

Automatic change-mode

I

! High speed only mode I

OFF j OFF / Normal speed only mode

-

51 -

.

No. 2J313-002E*P

TUBE 1 TUBE 2 SWlO: TUBE 3 (for USA and CANADA) Sw8: sw9:

SW No. 5

SW No. 6

SW No. 7

SW No. 8

CODE

TUBE TYPE

OFF

OFF

OFF

ON

1

DXB-0434ES(A256>(USA>

ON

OFF

OFF

OFF

8

DRX-3535HC/HD

ON

ON

OFF

OFF

C

DRX-4634HC(USA)

ON

ON

OFF

ON

D

DRX-3724HD/HC(USA)

ON

ON

ON

OFF

E

DRX-3624HC(USA)

ON

ON

. ON

ON

F

DXB-0324CS(A192>(USA>

SW8: TUBE 1 sw9: TUBE 2 SWlO: TUBE 3 (for other countries except USA and CANADA) TUBE TYPE

SW No. 5

SW No. 6

SW No. 7

SW No. 8

CODE

OFF

OFF

OFF

OFF

0

DRX-1924HD

OFF

OFF

OFF

ON

1

DXB-0434EScA256)

OFF

OFF

ON

OFF

2

DRX-2724HD

OFF

OFF

ON

ON

3

DRX-1615B

OFF

ON

OFF

OFF

4

DRX-1603B

OFF

ON

OFF

ON

5

DRX-B1356EC-MO

OFF

ON

ON

OFF

6

DXB-0324CS(A192)

OFF

ON

ON

ON

7

DRX-3745HD-H

ON

OFF

OFF

OFF

8

DRX-3535HD

ON

OFF

OFF

ON

9

DRX-3224HD-H

ON

OFF

ON

OFF

A

DRX-3224HD

ON

OFF

ON

ON

B

DRX-2924HD

ON

ON

OFF

OFF

C

DRX-3724HD

ON

ON

OFF

ON

D

DRX-2703HD/HC

ON

ON

ON

OFF

E

DRX-ll24B-MO

ON

ON

ON

ON

F

DRX-1724B i

1

- 52 -

No. 2J313-002E*F

PWB name:

[DESK DISPLAY I 516

A:

KXO-30R

B:

Equipment

when

Setting A

KXO-30F/%)F

installed

B 0

sw9:

Setting

Tube

2

Necessary

0

SW10: Setting

Tube

3

Necessary

Contents settings

of of

these SW~.

switches

are

the

same

as

the

* -

52-1

-

No. 25313-002E

PWB name:

A:

[DESK DISPLAY] 6/6

KXO-30R

B:

KXO-30F/50F When

installed Necessary

Setting the display time of a ctual R-time when AEC is ON.

SW11 .

/ Setting [ Display time (

I

0

!

1 2 3 4

!

0 1 2 3 4

1 I

-Th

is va lue is set fore shipment.

I

F -

I

15sec.

1

0

I

-I-

I

Setting the UP/DOWN key speed

- 53 -

Necessary

No.

2J313-002E:cA

(Control box)

[CPU1

PWB name:

A:

KXO-30R

B:

Equipment

When installed

Setting A

KXO_30F/50F

B, 0

1 Xl

:

1

Short

I Unnecessary

I

I

X26:

X26-

Al3

X27:

X27-

+5

/ so -

RAM-S6

s2

1 I

Selecting Ml ROM 27128

: Short

Ml address:

0000 'L 3FFF

.- Short

RAM address: 6000 ?J 7FFF

[FH/INV]

PWB name: Equipment'

Setting q---J

0

/

0

/ JP-C I

I



i

~

Open : Short:

Clock division by three Clock division by two

(Set before shipment "short")



- 54 -

When installed Unnecessary

No. 2J313-002EkD

PWB name:

A: KXO-30R

1

[ SIO-2

B:

KXO-3OF/5OF

There is no SIO-2 PWB in the KXO-3OR=B, KXO-30F=B or KXO-50F=B. Equipment

When installed

Setting

A 0

0

Jumper:

Specifying the RX and TX buffer addresses

All short

All short

LRZ-D LRl-C LTlw---+B LT2-A

HRZ-F HRl-E HTZ-B HTl-A

Unnecessary

(Set before shipment "short") PWB name:

[SIO-11 or [SPIO]

Equipment

When

Setting

installed

Aj 0

0

There is no Jumper in the KXO-30R=B, KXO-30F=B or KXO-SOF=B.

Jumper:

Unnecessary

J: Short T-G: Short SWl:

11

PWB name:

Unnecessary

Baud rate generator

2

3

4

Baud rate

Equipment

ON j OFF

OFF

OFF

4800

30R

ON / OFF

ON

OFF

9600

30F/50F

IPIN

Setting

L

- 55 -

When installed

No. 25313-OOZE

PWB name:

A:

[IFI

KXO-30R

B:

KXO-30F/50F

Equipment

When installed

Setting and adjustment A 0

B

/ Necessary

’ Jumpers 52 through 54

IO

, Connect one jumper of 52, 53, 54 corresponding ; the minimum HU depending on combined tubes.

52: 53: c 54:

: 0

'

0

;Jl

PWB name:

Unnecessar}

. .

Short (only for USA.) Unnecessary

Adjusting VRl through VR4

i

0

140 KHU 200 KHU 300 KHU

Jumper Jl

!

0

to

[APR IF]

Equipment

When installed

Setting A I

B



0

I Jumpers Jl, 52:

These switches set BKYl or BKY2 when using Bucky for 1 TOM0 TECHNIQUE (Jl and 52 should be set 1 simultaneously.) l

I I

Necessary

Setting TOMO, BUCKY selection

When using TOM0 Bucky at BKYl TECHNIQUE, connect C and Bl.

’ . When using TOM0 Bucky at BKY2 TECHNIQUE, connect C and B2. !

- 56 -

,

No. 2J313-002E"cL

PWB name:

A:

MQI

KXO-30R

B:

KXO-30F/5OF

Equipment

When installed

Setting and adjustment A

B 0

Jumper Jl:

0

SWl:

Necessary

Short for F. Rotation

Necessary

Setting continuous radiography TECHNIQUE

Unused

SPOT BKYl BKY2 GENE TOM0 MAMMO 105mm spot camera

When selecting the continuous radiography mode for TECHNIQUES, turn the switches corresponding to the individual TECHNIQUES ON. , (The switches of three TECHNIQUES of SPOT, TOMO, and 105mm Spot Camera are set to ON before shipment.) 0

0

VRl:

Adjusting the set-up time of the normal speed starter

Unnecessary

This switch adjusts the set-up time of the normal speed starter. VRl is set to 0.8sec before shipment.

PWB name:

[RY

SEQl

Equipment Setting A

B

0

0

Jumper:

Setting the type of stator coil of X-ray tube TUBE 1 TUBE 2

o------ ---Q TUBE 3

If a combined tube is the XS-AA coil type or the Eimac tube, short the pins of the tube.

- 57 -

When installed Necessary

No. 2J313-OOZE*M

PWB name:

A:

[F DIG] l/3

KXO-30R

B:

KXO-30F/50F

Equipment Setting and adjustment A

B 0

SW1 thru SW5:

Setting the dose limit in F kV Auto mode

These switches should be set for USA and other countries where the dose limit is imposed. An example for USA (limited to 2.58 x lo-IC/(kg*min) {lOR/min) is given below. These switches set F.kV values which exceed the limit of 2.58 x 10_3C/(kg*min) {lOR/min) at F.mA values of I 0.5mA, l.OmA, 2.0mA, 3.OmA, and 4.0mA.

F*kV SW1 =A SW2 =B SW3 = c SW4 = D SW5 = E

F mA When setting the switches. measure F.kV values exceeding the limti of 2.58 x 10-3C/(kg=min) (lOR/min} at individual mA values (binary), and set them in the individual switches.

- 58 -

When installed Necessary

No. 2J313-002E*M

PWB name:

A:

[F DIG] 2/3

KXO-30R

B:

KXO-30F/50F

Equipment Setting and adjustment A

B

When installed

125 110 100 0 85 80

The above figure gives an example in which SW1 through SW5 are set to 125kV, IlOkV, lOOkV, 85kV, and 80kV respectively. 0

/

,946, sw7:

Setting the dose limit in Fully Auto mode

/ ~ These switches should be set for USA and other 1 countries where the dose limit is imposed.

/( SW7 requires no setting.) ’

~ In Fully Auto mode , the F.kV and F.mA values are ~ linked, and the F.kV value should be set for dose ( limit control. 1 When setting SW6, measure the F.kV value in Fully 1 Auto mode which exceeds the limit of 2.58 x 10-X /C/(kg*min) (lOR/min), and set the value (binary) I in SW6.

-

59 -

Necessary

No.

PWB name:

[F DIG]

KXO-30R

A:

3/3

B:

KXO-30F/50F

Equipment

When installed

Setting B

A

0

I

’/

SW8,

; i j

When SW8 is set to "l", a mode is set which enables two F.ABC parameters to be set with the DIP switches (SW11 and SW12 on the ATR.DIG PWB). .

i

When SW8 is set to any other mined parameter is set.

SW9:

Setting the F.ABC full-auto mode

parameter

value,

1 When SW9 is set to "O", F.kV/FmA 50kV/O.ZmA to 125kV/4mA. I 1 When SW9 is set to "l", F.kV/F.mA ; 50kV/0.2mA to 125kV/2mA.

I

ZJ313-002E$cB

mode

and the

Necessary

the predeter-

changes

changes

from

from

1 I / I

SWlO:

O ) I (

F-ON

test

I 1 Unnecessary

switch

The fluoroscopy can be done by this Test switch instead of the F-ON signal from the conventional R/F table.

- 60 -

No.

PWB name: Equipment

A:

[F ANA]

T

Setting

KXO-30R

B:

JXXO-30F/50F

When installed

and adjustment

B 0

VRl:

Setting

manual

F.kV

2J313-002E*L

Unnecessary

control

Connect 5kS2 variable resistors (same as the F. ABC control of R/F table) to FASTON terminals Tl, T2, and T3. Adjust VRl to set the T2 voltage to +lO.OV when turning the variable resistor fully clockwise (equivalent to F.kV=125kV selected).

SWl, VR3,

VR4,

Adjustment

VR5,

VR6,

Unnecessary

VR7

procedure

1

Set SW1

to "1".

2

The set mode is used to adjust the D/A converters (1C and 2C). In this mode, adjust VR6 and VR7 so that the voltage at Pll and P12 is set to a value in the range of 9.961 to lO.OOOV.

3

Next,

4

Adjust

VR4

to set the P8 voltage

to - lO.OOV.

5

Adjust

VR5

to set the P9 voltage

to +9.961V.

6

Adjust VR3 to set the Pll voltage in the range of 9.961 to lO.OOOV.

7

After "0" .

set SW1 to "2".

to a value

the adjustment is completed, reset SW1 to (If not, the software will not operate.)

Jl: Selecting the jumper for ABC mode (Video signal type or photomultiplier . Video signal type : Jump "V" side . Photomultiplier type: Jump "P" side 0

VR8:

Necessary type) connector connector

ABC gain adjustment VR for video signal type Adjust the ABC gain according to the system requirements.

I

I - 61 -

ABC

Necessary for video signal type ABC

No. 25313-002E

PWB name:

[ATR DIG] 1/3

A:

KXO-30R

B:

KXO-30F/SOF

Equipment

When installed

Setting A

B

1

0

5, 9:

SWl,

Setting AEC backup mAs values

’ Necessary

For SPOT-SHOT TECHNIQUE For photofluorography TECHNIQUE For special TECHNIQUE

SWl: sw5: sw9: +EC

is automatically selected for ATR, and backup mAs 'values can be set by these switches. The switches are 'for the TECHNIQUES given above.

1

I

sw1,5,9

ZFE3 12 A

, , I

(

PEI

00 3

5

‘: I

4 4 i I

25

24

23

h

27

26

LIP 6 ,I



P 7 ih

;

I

~

!

22

21

1

8 A i

20

(

I

[Example]

;

: ds

,;;-;-1

*2 I

2510

0

30/o

0,o

35

1

0

0

I ’

1

1

1

111

!011/0



40 j O'O/ 1 0 ~45/O'Oll;O j 50 10 0~1~1~0~0

j

i

Ii 1

i

0

o/0(1

I

12i

I l/O!0

2.

l:‘“f~al

0)1'

19H

1'0

1EH

1: ON

l/

23H

0: OFF,

o/

28H 2DH

I I I !

/ I

11110 1

1 1'01

1

Max* )

1'1'1 /

1

I lil

1

- 62 -

1

/ ;

I

32H

i iI255

~

/

1

FFH

I

No. 25313-002E

PWB name:

A:

[ATR DIG] z/3

JLXO-30R

B:

KXO-30F/5OF

1

E;i.,,;t

When installed

Setting

0

I

/

SWZ, 6, 10:



SW2 : SW6 : SWlO:

Setting the upper limit of R.kV

Necessary

For SPOT-SHOT TECHNIQUE For photofluorography TECHNIQUE For special TECHNIQUE 1

These switches set the upper limit of the R.kV value when converting F.kV to R.kV. , I I

1 I

SW2,6,10

(Jp$ ““EIPPUUiiP

:

a,

12 /

h

h

m

A

A

Ii

4 ‘1

27

26

25

24

23

22

21

20

34567

[Example] Ic

Hexadecimal

Binary kV .I

I ~

2’

26

125

zL

! 2”

~

70;

0~1~0~0~0~1~1~0~

i

801

0

j

9oi

0

l/O 1

/ 100 , 0 / l/

12’

1

12” ~

0i

11i~oii~o~

(H)

46H

1

1101 0

I-9

0101

50H

I

, 1 / / 0 1 0 / 11 0 / 01 1

~AH 64H 6EH

~110~0~1/1,0'1/1~1~0' I 5

I

:ii

IMax.

1

o’.

1

5

oil~l’o~ I

150

96H ,

I O !

jsw3, 7, 11:

I I /

;“w:

..:

SWll:

Necessary

Setting the lower limit of R.kV

For SPOT-SHOT TECHNIQUE For photofluorography TECHNIQUE For special TECHNIQUE

I

'These switches set the lower limit of the R.kV value when converting F.kV to R.kV. (The setting procedure /is the same as the upper limit setting described iabove.) I I

I

,

- 63 -

I

No. 2J313-002EfcB

PWB name:

[ATR DIG] 3/3

-r---j

A:

JLXO-30R

B:

KXO-30F/50F

Equipment A

B



0

I

1

i I

When installed

Setting

Necessary

SW4, 8: Setting the maximum mA value of mAs SW4 : SW8:

For spot film TECHNIQUE For photofluorography TECHNIQUE

j In ATR mode, the equipment separates the backup mAs I value which is set in SWl, 5, 9 using the mA value

l

/which is set by these switches. When performing small I focus radiography in ATR mode, set this value to the 1 maximum mA value of small focus.

J,- 640mA is not applicable

0

to KXO-30F.

SW11, SW12: Setting the F.ABC parameters (effective only when SW8 and SW9 on the F.DIG PWB are set to "1" and "0" respectively) Parameter P (for control sensitivity) / SWll: Parameter I (for reference of output j sw12: amount) I

1

- 64 -

( When / necessary

No. 25313-OOZE

PWB name:

A:

[ATR ANA] 1/2

KXO-30R

KXO-30F/50F

B:

Equipment

When installed

Setting and adjustment A

B

Necessary P5 input voltage ranges from +3.33V (50kV) to +8.33V (125kV) and P6 input voltage ranges from +0.5V (0.2mA) to +lOV (4.OmA). The final output voltage (PIO) is +lOV (150kV) max. VRl, VR2, and VR3 should not be set because they are unused. VR31 is used for radiography at a constant R.kV value with the ATR function stopped temporarily in SPOT (Example: Esophagus radiography) 1 Radiography mode. Adjust the P9 voltage with 26B grounded.

0

IVR4 through VR7 adjustment (for spot film radiography) / I

I

P7 voltage

PlO voltage

IL

Correction VR7 ‘\ \ i$

A

/

with F.mA

k

When F.d

N’

> 1mA

When F.,ti < 1Lti

VR4 1

-

F.kV ,~/@ lVR4 F.kV

I

Adjust VR5 to set the P8 voltage to OV when F.mA=lmA. ! VR8 through VR13 adjustment (for spot film radiography)i These VRs correct R.kV values for film division and image intensifier (1.1.) magnification. Division and 1.1. magnification !

Adjustment VR

!

Division by 2

VR8

Division by 3

VR9

Division by 4

VRlO

,

Spare

VRll

1 j

Ml

VR12

M2

VR13

iVR14 through VR17 adjustment (for 1.1. photofluorography) These VRs have the same functions as VR4 through VR7 for spot-film radiography.

- 65 -

No. 25313-002EfcB

PWB name:

[ATR ANA]

KXO-30R

A:

2/z

B:

KXO-30F/50F

Equipment Setting

When installed

and adjustment

A VR18 through VRZO adjustment (for 1.1. photofluorography) These VRs have the same functions and 13 for spot film radiography.

0



1

j

PWB name:

as VRll,

12,

VR21 through VR24 adjustment (for AUX TECHNIQUE) These VRs have the same functions as VR4 through VR7 for spot film radiography. VR25 through VR30 adjustment (for AUX TECHNIQUE) These VRs have the same functions as VR8 through VR13 for spot film radiography.

[HEAT

A:

11

KXO-30R

B:

KXO-30F/50F

When installed

Setting

I

r

0

!

0

j VR2:

Frequency

adjustment

o

/

o

1 VR3:

Frequency

adjustment

/ VR4:

Offset

o 1-0

/Unnecessary I I IUnnecessary

adjustment

A:

[HEAT 21

PWB name:

/ iunnecessary

KXO-30R

B:

KXO-30F/50F

When installed

Setting

r I

0



0

0

I

o

I

’l/

VR12:

D/A converter

span

j

Unnecessary

I

j

128/256:

128 crossover

span

- 66 -

1 Unnecessary

No.

PWB name:

A

B

Adjustment point

0

0

CNNA

0

0

0

0

Equipment .

KXO-30R

A:

[INV-11

Contents

B:

2J313-002EkI

KXO-30F/50F Adjustment when installed

of adjustment

Unnecessary

;'(OVER CURRENT detection level setting +8V (TPl) VR2 KXO-50F=B or KXO-30F=B (VRl,VR2) * For the KXO-30R=B, VR3

Tube

voltage

waveform

m I

Note *: In the case of single-phase power supply: If 160mA or more is set, the tube waveform shifts slightly downwards, this cannot be adjusted with If VR3 is used VR3. for adjustment, the tube voltage waveform may become abnormal under a small load.

Tube

0

0

VR5

Tube overvoltage at 165kV

0

0

VR6

Tube voltage output adjustment in fluoroscopic mode Set the measured tube voltage to 90kV with 90kV and 1mA set and LED4 on the INV-2 PWB indicating "8".

Necessary

0

0

VR7

Error signal level -0.6 to 1OV (TP~)

Unnecessary

0

0

VR8

Frequency adjustment When Pin 7 at location IC: 2.5kHz at TP12 Variable-width timing

20kV

detection

to 1V (TP2)

Unnecessary

VR4

VR9,VRll

gain

In the case of a singlephase power supply, the '""=I tube voltage waveform may gradually descend under a heavy load.

0

0

signal

r

0

0

voltage

When necessary (fine adjustment

adjustment

If the tube voltage waveform under a small load * (about 5OmA) is overshoot or undershoot, Set the adjust VR3. waveform by a tube voltmeter to be the same as that at TP2.

Unnecessary

level

Detection

adjustment

adjustment

- 67 -

Unnecessary

Unnecessary 2D is 5V,

of phase

control

Unnecessary

No. 25313-002E*I

A:

[INV-21 l/2

PWB name:

KXO-30R

B:

KXO-30F/50F Adjustment when installed

Equipment Setting

and adjustment

A

B

0

0

D/A converter adjustment VRl, VR3, VR5: Zero-position adjustment Gain adjustment VR2, VR4:

Unnecessary

0

0

SW2,

Necessary at tube voltage adjustment

0

0

I

SW3:

Switches to write the tube voltage adjustment data (kV shift data) in radiographic or fluoroscopic mode into the .RAM on the PWB. kV to be set: For each section 20kV to 31kV, 32kV to 47kV, 48kV to 63kV, 64kV to 95kV, 96kv to 127kV, and 128kV to 150kV, kV shift data can be written into the RAM for each mA and each tube.

Necessary

SW1 POWER source

I

/

o

0

1

SW4:

Setting

Position

19 2oov

0

l$ 220v

1

3cP 380V

4

39 415v

5

39 440v

6

Tnverter

operation

disable

of SW1

switch

When this switch is set to "DIS" (DISABLE), LED3 lights up and the INV-UNIT operation is set this switch to "EN". disabled. Normally, Note:

I

I

If SW4 is set t0 "EN" (ENABLE), LED3 will light up in the following cases: @

If LV

@

If SW1

is not

@

If the

INV unit

@

During

charging

is too high

or too low

set as specified

:

- 68 -

fuse

is blown

Necessary

No. 25313-OOZE*1

PWB name:

[INV-21

2/2

A:

KXO-30R

B:

KXO-30F/50F

Equipment Setting A

B

0

0

SW5:

(YWA1977

Adjustment when installed

and adjustment

only)

Necessary

Length of the longest among those connected the three X-ray tubes

cable to

to 15m

Setting of SW5

position

3, 2, 1 or 0

16 to 20m

2, 1 or 0

21 to 25m

1 or 0

26 to

0

-

68A -

No. 2J313-002E*D

PWB name:

A:

[INV-31

KXO-30R

B:

Adjustment when installed

Equipment Setting

and adjustment

A

B

0

0

VRl

When the power voltage Line voltage signal: has the nominal value, signal at TPl is 5Vp.

0

0

VR2

TP2:

VR3

VR4

VR5

0

0

0

(

0

I

I

1

I

I I 1

cases>

TP3: +4.87V +4.75V

(at 14 2OOV), (in all other

cases>

TP4:

(at l$ 2OOV), (in all other

cases>

+5.5OV +5.5OV

VR6

TP6:

CNNA

Set to "NEGA"

SW1

Turn

(at l$ 2OOV), (in all other

= 1V for tube current

Unnecessary cases> Unnecessary

lOOmA

1 Ucnecessary

j

0

0

(at lb 2OOV), (in all other

TP5:

Unnecessary

Necessary

+4.75V +4.5OV

+5.35V +5.25V

HO-3OF/5OF

on 1 of SW1 for a single-phase power Turn it off for all other power suPPlY* supplies.

it is unnecessary to set SWl, for KXO-30R=B, KXG30F=B or KXO-5OF.

Necessary

the

I

-

-

69 -

No.

7.2

1

Error

Code

,

1



.k

_k

Code

Table Meaning

Transmission and the box

system error is not carried

(Data communication out correctly)

between

the

198

RAM error

(Warning.

RAM on

the

panel

or

desk

is

defective)

199

RAM error

(CPU stop.

RAM on

the

panel

or

desk

is

defective)

201

TECHNIQUE setting

202

Tube No. setting not set correctly)

301

TECHNIQUE setting

700

Box-side

CPU or

913

1 Normal starter / was not ready) I

914

; Over MAS I 1 Filament heater

88 89 t

2J313-002Ey:B

i Filament 1 greater

switch

(on

tghe

(Tube

error

switch

panel

setting

(on

the

(Current

RAM down

desk)

switch

panel

serial-transmission error

or

or

on

the

desk)

is

CPU down was

applied

(Heating

heater current abnormal than the specified value)

data (The

is

the

set

panel

not

(Either

when

are

not

panel/desk

or

desk

is

set

CPU malfunctions normal

starter

abnormal)

heating

current

has

become

1

1

90

Filament hardware

heater output is abnormal)

95

Over current (The discharge, etc. )

96

Over

kV (Output

kV value

97

Fuse

blown

fuse

98

Line voltage smaller than

(The

section

inverter

of

abnormal

thyristor

has the

become INV-UNIT

(Heating

70 -

system

has

malfunctioned

due

greater

than

specified

is

the

to

tube

value

blown)

(The power source voltage error the specified value and X-rays

-

output

has become greater cannot be generated.

or

No.

8.

MAINTENANCE

SCHEDULE described

The schedule of periodic maintenance procedures conducted for both the X-ray tube and generator listed below. First Servicing Second Servicing Note:

2J313-002E*B

-------------________

system

here

with

shall

be

the frequency

6 month after installation 1 yaer after installation

All subsequent servicing shall be performed every 12 months When a new X-ray tube is installed, repeat the first thereafter. servicing.

APPENDICES LIST APPENDIX

01

Recording

Form

APPENDIX

02

Relations Between Focal Tube Current Range

APPENDIX

03

Test Method

of Fluoroscopic

APPENDIX

04

Test

of Reproducibility

APPENDIX

05

The Setting Technique Factor for Reproducibility and Linearity Test without

Method

for Tests Spot

Size and

Entrance

Exposure

Rate

and Linearity

A.E.C.

APPENDIX

06

Test

Method

of Automatic

Exposure

Control

Limit

APPENDIX

07

Test Method Time

of Automatic

Exposure

Minimum

Exposure

- 71 -

No.

APPENDIX

01

Recording

Form

2J313-OOZE*B

for Tests

This

data sheets are to be filled out and copied at the time of the Record test results in copied sheets and return installation and maintenance. to: -------------

U.S.A.

TOSHIBA

MEDICAL

SYSTEM

DEVISION

AMERICA INC. 2441 MICHELLE CALIFORNIA 92680 The

---------

others

TOSHIBA'S

Service

Date:

2.

Installed Branch

Maintenance

by:

Office:

Address: 3.

Hospital: Address:

4.

Serial 0

No. of X-ray

Generator

DC-30R:

q DC-30F:

5.

0

DC-50F:

17

DG-30F:

q

DG-5OF:

X-ray

Tube

to be Combined

Manufacturer

6.

Tube

1

Tube

2

Tube

3

X-ray

Table

Model

name

to be Combined

Manufacturer: Model Serial

name: No.:

- 72 -

DRIVE,

TUSTIN,

representative

: Installation,

1.

OF TOSHIBA

Serial

No.

Inspection

No.

X-ray

beam

limiting

device

combined

2J313-002E~kB

with

Manufacturer: Model

name:

Serial: 8

Power

source 0 0

(1) Power

phase

(2) Power

frequency

q 5OHz 0

(3) Power G 0 9.

Single phase Three phase

60Hz

voltage 0 22ov i? 415V

2OOV 38OV

Preliminary

test

0

440v

3

48OV

results Result

Item Section Section Section Section Section Section Section

Section

Tube

3.2 5.1.2 5.1.3 5.1.4 5.1.5 5.1.6 7.1

5.1.5

No.

Power Source Checking voltage and display circuits Checking operations Checking the switching circuit Checking the starter circuit Checking the battery PWB Setting and Adjustment Tables

Check

on starter

Tube model name

circuit

Normal speed boost time

1

- 73 -

starter (set>

OK, OK, OK, OK, OK, OK, OK,

NG NG NG NG NG NG NG

No.

10.

Radiographic

KV, mA

and time

2J313-002E*K

test

TUBE NO. Tube model High

(1) R.kV

potential

cable

R.Time:

and R.mA

I

lx”

71A L

kV

mA

m

length

kV

mA

0.05

kV

set

mA

kV

mA

13 16

L

I

J

- 74 -

No.

TUBE

2J313-OOZE*K

NO.

Tube model High

R.kV

potential

cable

length

R.Time:

and R.mA

m

0.05

set

140

120

kV

-

75 -

mA

kV

mA

No.

2J313-002E*K

TUBE NO.

R.kV

10

and

Tube

model

High

potential

cable

m

R.Time:

R.mA

kV

length

mA

kV

mA

kV

13 16 20 25 32 40 50 64 80 100 125 160 200 250 320 400 500 640

-

76

-

mA

0.05

set

kV

mA

kV

mA

2J313-002E"B

No.

(2)

R.time

R.time

R.kV

0.005sec

60kV

Result

R.mA

1

(set>

320mA

111 +

lOOmA

90kV

1

0.2sec

11.

Fluoroscopic

test

(KXO-30F

or 50F) Tube model Table

model

F. KV and F. mA

/ I

1

I I I j

2

12.

1

F. ABC

mL4

mA

mA

mA

mA

mA

mA

mA

InA I

mA 3

In U.S.A, limit.



/ mA

4 Note

m&4 / I

test

should

be performed

within

the entrance

test

. Thickness

of the acrylic

cm

phantom:

V

. Peack value of the video signal waveforms: (including the signal pedestals) . Voltage

at P6 of the PTW unit:

requirements: . Fluoroscopic (during stable operation of the ABC circuit)

V kV,

- 77 -

mA

exposure ,

No.

13.

Automatic exposure control test (in combination with the automatic AEC-lOA/B) Record manual

14.

exposure

2J313-002E*B

. unit, model

control

test results, using data sheets attached to the installation (No. 2C376-519E). for A.E.C. UNIT, model AEC-lOA/B,

ATR adjustment (1) X-ray X-ray

(KXO-30F

or 50F) (FFD

table:

>

tube:

Grid: Intensifying

screen:

Film: (2) Set value R.kV BACK

kV kV

Clamp

DOWN: UP: UP mAs:

(3) Correction

ITAS

value

Correction

Correction mA correction

Division correction

0.5mA 1mA 2mA 3mA 4mA

0

value kV kV kV kV kV kV kV kV kV kV kV

l/2 l/3 l/4 l/6 Ml M2

- 78 -

No.

(4) ATR

data

in FKV-AUTO

--

:

--.-

---

-

i L :; -. . . :-

;_

I!

;_

__

.:

.I

-

F_

f

_. __~

._

.

____L_~-.-

gheet:~‘;~___i_I~”

ATR.d.ata

.-... . . _ _. _-.

-_ .. _.._-.

. -_--

-----

:I

:

::f

_~~~:-

s-___._.-----_ -1. -f -_

--!

._. -. _. :-

.. .-

.--.

I I

_ .

__

_

__. ..__._. . _._ . .

I

I

I

.

.

_.

. _

--‘i

t-

.

. .

_

I

I

I

. , I-

:

__

-.-

&_...- cm22--. -_--_ . . -_. _ ____ . _.________I___-_-_ . _ --.--_ _- ---.__. _ ._____-----Acrylic phantom thickness :L.LL--L- :.__________ ._._.---. __ ] ~ _-

li --__l&___---_-16

-_-

(mA: Manual)

._. . _. - ._ . _._-..__ _ _.

.

._______.&_______

mode

2J313-002E*B

_~____~&-_-__2O__

-L_

T Acrylic phantom 12 cm

Fluoroscopic

conditions

mA

kV

14 16 18 20 22 24

-

79 -

Radiographic

kV

conditions

mA

No.

15.

23313-002E*M

HHS Test The following tests are applied to the X-ray generator. The HSS test is necessary only for ORH regulations.

15.1

Entrance

exposure

(1) Installed (2) Measured a> Manual

rate

system: entrance

limit 1. Under exposure

tube

2. Over

tube

3. C-arm

rate

mode

b) ABC mode Brightness Entrance

control: exposure

limit

F.mA

limit

F.kV

maximum rate:

.

position (R/min) (1 R/min = 2.58 x lo-*C/(kg*min)) mA kV

- 80 -

No.

15.2

Reproducibility

kV

and linearity

60 kV

without

automatic

exposure

TUBE

NO.

TUBE

TYPE

2J313-002E*M

control

60 kV

60 kV

60 kV

60 kV

mA

InA

mA

mA

mA

mA

sec.

S

S

S

S

S

Focus Line

regulation

Exposure No.

I

I

-+%

I

-+%

I

I I

-+%

-+%

1

mR

mR

mR

mR

mR

2

mR

mR

mR

mR

mR

3

mR

mR

mR

mR

mR

4

mR

mR

mR

mR

mR

5

mR

mR

mR

mR

mR

6

mR

mR

mR

mR

mR

7

mR

mR

mR

mR

mR

8

mR

mR

mR

mR

mR

9

mR

mR

mR

mR

mR

10

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

x S

-+%

I

I

I

I

I

I

I

C

x

I

mR/mAs

mR/mAs

1

mR/mAs

1

mR/mAs

I

1

mR/mAs

1

I

Xl +X2 (1 mR = 2.58 x 10-7C/kg) Note:

Make a sufficient results.

number

of copies

- 81 -

of this

sheet

to record

test

No.

TUBE

NO.

TUBE

TYPE

2J313-002E*M

kV

90 kV

90 kV

90 kV

90 kV

90 kV

mA

mA

mA

mA

mA

mA

sec.

S

S

S

S

S

Focus Line

regulation

Exposure No.

‘j;-

+ -

%

+ -

%

-+

%

+ -

%

-+

%

1

mR

mR

mR

mR

mR

2

mR

mR

mR

mR

mR

3

mR

mR

mR

mR

mR

4

mR

mR

mR

mR

mR

5

mR

mR

mR

mR

mR

6

mR

mR

mR

mR

mR

7

mR

mR

mR

mR

mR

8

mR

mR

mR

mR

mR

9

mR

mR

mR

mR

mR

10

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR/mAs

mR/mAs

mR/mAs

mR/mAs

mR/mAs

S

C ?i I)il - x21 Xl + x2 (1 mR = 2.58 x IO-7C/kg)

- 82 -

No. 2J313-002EkM

TUBE NO. TUBE

TYPE

kV

140 kV

140 kV

140 kV

140 kV

140 kV

mA

mA

mA

mA

mA

mA

sec.

S

S

S

S

S

Focus Line

regulation

Exposure No.

x

+ -

%

+ -

%

-+

%

+ -

%

+ -

%

1

mR

mR

mR

mR

mR

2

mR

mR

mR

mR

mR

3

mR

mR

mR

mR

mR

4

mR

mR

mR

mR

mR

5

mR

mR

mR

mR

mR

6

mR

mR

mR

mR

mR

7

mR

mR

mR

mR

mR

8

mR

mR

mR

mR

mR

9

mR

mR

mR

mR

mR

10

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR

mR/mAs

mR/mAs

mR/mAs

mR/mAs

mR/mAs

S

C 51

Xl +X2

(1 mR = 2.58 x lo-7C/kg)

- 83 -

No.

15.3

Automatic

exposure

Setting technique Measured mAs 15.4

Automatic

exposure

(1) AEC Type: Measured (2) AEC Type:

Field

control

limit.

minumum

mA

50 kV

factors:

exposure

time

phototimer time Solidtimer

Measured

time

1 2 3 1+2 1+3 2+3 /1+2+3/

- 84 -

ms

2J313-002E*B

No.

APPENDIX

02

RELATIONSHIP

BETWEEN

FOCAL

SPOT

SIZE AND TUBE

2J313-002E9cB

CURRENT

RANGE

I

Small X-ray

Large

-r

focus

tube Focal

type

Al92

focus

(EIMK)

spot

The current

size

range

Focal size

spot

0.6 x 0.6mm

20 to 320mA

1.2 x 1.2mm

Tube

current

range

400

to 1OOOmA

Tube voltage range

40 to 150 kV 1

-

85

-

No,

APPENDIX 03

2J313-002E*M

TEST METHOD OF FLUOROSCOPIC ENTRANCE EXPOSURE RATE

A.

Objective Fluoroscopic equipment which is provided with automatic exposure rate of tube potential and current control is not operable at any combination rate in excess of 2.58 x lo-3C/ (kgemin) which results in an exposure of the useful beam enters the {lOR/min} at the point where the center (See the section 1020.32 (d).) patient.

B.

Test method (1) The position as follows. a.

Testing

of

test

when

the

equipment

source

is

and

below

flip

t

I

\

diagnostic

i

table

system

Spot

film

Lead

sheet(300

device x 300 x Zmm)

Aluminum attenuation (200 x 200 x 19mm)

3oomm

assembly

table

Imaging

I

X-ray

block

4Dosimeter

1

P Beam-limiting X-ray

the

Set

2.

Place the attenuation block (200 x 200 x 19mm) at the approximate center of the table and 150mm (6in) above tabletop and conform the center of the X-ray field to observing the image. imaging

to

3.

Set

4.

Place field image.

5.

Place the lead sheet X-ray field between system.

the and

the

system

horizontal

tube

1.

the

tabletop

device

to

probe of the 1Omm (0.4in)

position.

300mm (12in)

dosimeter above the

at the tabletop

above

86 -

tabletop.

center of the by observing

(300 x 300 x 2mm) at the attenuation block

-

the

the and

the it by

X-ray the

center of the the imaging

is

2J313-002E*B

No.

b.

Testing

when

the

source

is

above

the

table

Beam-limiting

device

Aluminum attenuation (200 x 200 x 19mm)

Spot

Lead

sheet

(300

x 300

film

device

Imaging

to

the

horizontal

sys tern

position.

Set

7 _.

If of be

3.

Place the attenuation block (200 x 200 x 19mm) at the center of the table and 150mm (6in) above approximate the center of the X-ray field to tabletop and conform observing the image.

5.

tabletop

x 2mm)

1.

4.

the

the source-to-image the beam-limiting placed provided

block

distance (SID) is variable, set the as close to the tabletop as it device shall not be closer than 30Omm. it

probe of the 30Omm (12in)

dosimeter above the

at

the

center

Place field image.

the and

Place X-ray

(300 x 300 x 2mm) at the lead sheet field on the tabletop.

-

57 -

tabletop

by

the

of

the

the it by

X-ray

observing

center

the

of

the

end can

2J313-002E*M

No.

Testing table)

c.

in

a C-arm

type

fluoroscope

(in

a Gyroscopic

-

4

diagnostic

Imaging Spot

A

system

film

device

Lead sheet (300 x 300 x Zmm) Aluminum attenuation (200 x 200 x 19mm)

300mm 850.5mm (MIN. SID)

Beam-limiting X-ray

t

1.

If necessary, primary beam.

2.

Place the probe of the surface of the imaging

3.

Place the aluminum attenuation block (200 lead sheet (200 x 200 x 2mm) at the center shown in the above figure. open

shutter

the

tabletop

dosimeter system.

(2)

Fully

(3)

Set the Rotary DIP SW, SW5 on the to the position 0. This setting means that the X-ray entrance exposure rate limit.

(4)

Push

(5)

Select the not select

(6)

Set the F.mA control film device.

(7)

Perform

(8)

Monitorig

(9)

When the dosimeter (kgamin) {9R/min}, record it.

(10)

Repeat

the

the

position

blades

of

out

of

the

path

the

the

input

x 200 x 19mm) and the of the X-ray field as

device.

DESK CONTROL PWB in is

of

from

beam-limiting

generator

device

tube

300mm (12in)

the

block

the

excluded

control

from

desk

the

POWER ON switch. spot-film the “ABC”

radiography mode on the for

technique spot film

4mA and F.kV

on the device.

control

for

control

panel,

5OkV on

the

spot

fluoroscopy. increase

a dosimeter,

Steps

(6)

the

reading reaches read the setting

to

(9)

at

the

-

tube

to

125kV.

the limit level 23.22 x lo-*C/ F.kV on the control panel and

F.mA setting

88 -

voltage

of

3,

2,

1 and

0.5mA.

and

No.

(11)

Set the limit kV which is 1 for each F.mA setting. numbers are shown below.

F.mA

i

in Step (9) according to the Table settings and the settings switch

SW No. on the F.CONTROL control box

setting

0.5

recorded The F.mA

2J313-002E*M

DIG.

PWB

in the

SW1

mA 1mA

SW2

2mA

SW3

3mA

SW4

4m.A

SW5

(12)

Select the F.ABC and MA-AUTO mode switches. Set the brightness control to the minimum position.

(13)

Perform

(14)

Monitoring

(15)

When the dosimeter reading reaches the limit level 23.22 x lo-4C/ (kgemin) {9R/min), read the indicated F.kV and F.mA on the control panel and record them.

(16)

Set the limit kV recorded above according to Table 1. Set the limit mA recorded above according to Table 2. The setting switches are shown below.

fluoroscopy. a dosimeter,

(17)

Note:

limit

mA

limit

the brightness

SW No. on the F.CONTROL control box

Limit

kV

increase

DIG. PWB

setting

in the

SW6 SW7

!

Set the Rotary DIP SW, SW5 on the DESK CONTROL to the position 1. When

control.

the DIP

switch,

-

the power

89 -

should

PWB

in the control

be off.

desk

No.

Table

1

Limit

kV setting

2J313-002E*B

code

(1; ON, 0; OFF) F.kV

-

1

2

SW. b

3

No. 5

6

7

8

F.kV

1

2

3

“;’

‘1.

6

7

8

I 50

0

0

1

t

I

I 51 lolo 52

10

1

’ l/1lOll:o

54

0,o

1

55

0

0

010

58

011/o /

1

0

1 1

)

1

(

1

IO

0

I

i

61

0

0

1

1 ; 1

‘l/O

62

0

I 0

1

1 / 1 ( 1 / 1

Note:

When

l~o;O,o

1

1 0

0

0 0

1

74

0

1

75

0

76

0

1

77

0

l/O

1

/

/

/ O/l

93

Ojl

0

11

0

94

0

1

1

95

0

10

01

96

0

1

0

10

IO

L

jl

010 1

1

Oil

1

/

1

1 1

/ 0

0

0

0

0

III

l’lil/l/l 1010

I

011

l/O

0

1 0 /

1

/ 1 1 0

I

1

1

1o7

/

0

/

1

1 1

1

0

/

1

/

!

: 0

l/

1

j

0

1

108

1

j

109

O/l

110

0[1

0

/

0

i0j

011

99 100

1

0

1/00/0/1~0

1

j

l

011

0

1

011/011

(

1

ojo

11

111

0

112

O/l

1

j 0

1 1 1 0

1 0

/

1

/

/

/

1

/

1

1 1

0

10

I

10 1



1

11

ljOlO(

-?

l/l

0

0

I

0

1

1

1 1

o/oJo

0

1

80

0

1 / O/

1

81

0

10

1;O

82

011

83

0’1/0/L~0~0/1

84

0

l/O

85

0

1

011/o

86

0

1

0

1

87

0’1

0

l,r)iiil

1

1 l/O

1

I

ilk

Oil

11

115

O/l

1

1

O’Oll/l

116

0

1

1

0

1

101110/l 1

O/

1

o/1/+

1

O/l /

0

/

0

10

0

0111

0

Ijo;1/0 1

0

-

90 -

on

/

11

11

119

0

1

1

01

120

01111

I/

121

Ojl

1

lql;olo~l

0

122

O/l

1

I

IjO

1

123

0

1

1

1

1

121,

0

1

1

I

1

125

oj

1

lllll

all

1

f

I 1 /

0

0

1

11

0

1

1

i

turn

O/l/O

0

/

j 0 / 1/ 1j 0 / I

0

118

0

010/i

125kV,

1

1/1/o

I 1 I 1

0

10

0

011

I

/o

)0~1,1/0/0,0/0/1

198

1)

j

79

than

1 1’1

1 1

0

I

/O/l

1

0

01

0

,l 1

/

0

I

o/1 0

0

o/

73

kV is more

,oo

10

0

j 0

~]

im

!

1

011 !

97

1(1,1 ’

72

1011

0

,

1

j

90

O/l

O/l

1/110!1,1/

0

I.0

(

91

l\l)Oil

0

71

89

, IO I

1

92 111’

60

10

o/101 I

/

I

[ 69

I/

(

10’0

1 011

j

011

l/l

IO,

88

0

,

/

10’0

I59

0 1

ijOll/l/O

1

/

57

I

I

o/o

/

1

I

II

53

0

0

, l!O/Ojl/lj

1

O/l

56

0

I

/

1I 0

switches.

/

0 j 0 1

I

1

0

0

I

1

1

0

10

(

0’ .’

No.

Table

F.mA

SW. No. 4 5

1

2

3

0.2

0

0

0

0

0.3

0

0

0

0.4

0

0

0.5

0

0.6

2

Limit

mA setting

code

SW 7

SW. No. 4 5

6

7

8

F*mA

1

2

3

0

0

1

0

2.6

0

0

0

1

0

0

0

1

1

2.7

0

0

0

0

0

0

1

0

0

2.8

0

0

0

0

0

0

1

0

1

2.9

0

0

0

0

0

0

1

1

0

0.7

0

0

0

0

0

1

1

1

3.1

0.8

0

0

0

0

1

0

0

0

0.9

0

0

0

0

1

_O

0

1.0

0

0

0

0

I

0

1.1

0

0

0

0

1

0

6

7

8

1

0

1

0

1

1

0

1

1

0

1

1

1

0

0

0

0

1

1

1

0

1

0

0

0

1

3.2

0

0

1

0

1

3.3

0

0

1

0

0

0

0

1

1

0

3.4

0

0

0

I

0

1

1

0

1 0

( 0

/ 11

1

0

0

1

0

0

1

3.5

I

1.2

0

LL

0

0

1

1

0

0

1

1

0

1

1

1

1

o(

1.3

0

0

0

0

1.4

0

0

0

0

1.5

0

0

0

0

1

1

1

1

1.6

0

0

0

1

0

0

0

0

1.7

0

0

0

1

0

0

0

1

1.8

0

0

0

1

0

0

1

0

1.9

0

0

0

1

0

0

1

1

2.0

0

0

0

1

0

1

0

0

2.1

0

0

0

1

0

1

0

1

2.2

0

0

0

1

0

1

1

0

2.3

0

0

0

1

0

1

1

i

2.4

0

0

0

1

1

0

0

0

2.5

0

0

0

1

1

0

0

1

1: Note:

ON, 0: OFF When mA is

more

than

I

4.OmA,

0

I

0

1

1 0 1 0 1 1 1 0 1 0

4.0

91 -

I

I

3.8

turn -

0 0 I 1 / 0 / 0 1 1 j

3.6

0

I

2J313-002E*K

on all

0

0

switches.

1

0

1

1 1 1 1

0

0

0

No.

2J313-002E*M

(18) After adjustment of the entrance exposure rate limit, repeat the Steps (6) to (8) for the F.mA setting of 4, 3, 2, 1 and 0.5mA and record the tube voltage and the dosime ter readings at the limit points in manual mode (19) Perform current

fluoroscopy in the FULL-AUTO mode and record and dosimeter readings at the limit point.

the tube voltage,

(20) If the dosimeter reading exceeds the rejection limit, follows and then repeat the Steps (18) to (19). (a) Manual mode If the test results are not within the rejection the limit kV which has been set in Step (11).

as

limit,

decrease

(c) FULL-AUTO mode If the test results are not within the rejection limit, the limit kV and mA which have been set in Step (16).

decrease

(b) kV-AUTO (F.ABC) Same as above.

C.

proceed

Test

instruments

required

(1) Dosimeter This measurement a. Keithley 006A).

mode

requires

model

35020

one of the following

with

96020

b. Victreen model 550 with 550-3, to TOSHIBA APPENDIX 026A.) C.

Victreen APPENDIX

(2) Aluminum

model 555 with 008A.j

attenuation

lmA,

chamber

550-4,

test

(Refer

550-5

instruments:

to TOSHIBA

or 550-6

1OmA or 1DAS chamber

APPENDIX

probe

(Refer

(Refer

to TOSHIBA

block

200 x 200 x 19 (mm> (3) Lead

sheet

300 x 300 x 2 (mm> D.

Rejection

limit

Maximum entrance dose limit: 23.22 x 10_4C/(kg=min) {9R/min). The maximum limit means a reading value on the dosimeter specified C.

-

92 -

in item

No. 2J313-002E*B

APPENDIX A.

04

TEST

METHOD

OF REPRODUCIBILITY

AND LINEARITY

Objective are performed A series of 10 radiation exposure measurements the coefficient of variation and the average exposure ratio. Section 1020.31 (b) and cc>.>

B. B-l

Test

to calculate (See the

method

Testing

without

automatic

exposure

control

(1) Set the X-ray apparatus and test following block diagram.

instrument

according

to the

s Dosimeter

Focus detector distance: 1OOcm Collimator

Primary voltmeter

:

High-voltage generator 1

(2) Turn (3) Set and two and

on the equipment

X-ray control

r~:7.

Power source

&

line switch

and calibrate

test

instruments.

the technique factors (radiographic kV, mA and time selectors) at intervals of one or during the 10 consecutive measurements, factors are adjusted to alternate settings minutes the technique reset the test setting after each measurement.

For a detailed description of this step refer to APPENDIX 05 "Combination of Specified Technique Factors" in this manual. (4) While confirming that the respective percent line voltage regulations are within +l% of the mean value for all the 10 measurements, record the value of a dosimeter. (5) The coefficient of variation "C", 1020.30(b)(8), of the 10 exposure measurements is calculated using the following equation;

where

Xi = the exposure measurement x = mean value of the exposure measurements S = estimated standard deviation

-

93 -

No.

(6)

Confirm

that

the

above

product

(7)

The average ratios of exposure product (mR/mAs), (1020.31(c)), calculated using the following

“C”

is

B-2

= average mR/mAs values tube current settings.

Confirm

above

Testing

with

the

automatic

product

exposure

X- ray apparatus block diagram.

( 1) Set

the following

than

0.045.

in

x lo-TC/kg)

where : xl and x2 consecutive that

greater

to the indicated milliampere-second of the 10 exposure measurements equation;

(1 mR = 2.58

(8)

not

2J313-002E*M

is

not

obtained

greater

at

than

each

0.09

of

two

times.

control

and

instruments

test

according

to

the

Dosimeter

Beam limiting

device __--------_____

___------

-----_ Aluminum attenuation block (200 x 200 x 38mm) SID

(i.e.

Source Image 1ooomm

Distance)

Primary voltmeter Power

H.P. Igenerator

(2)

Set the cassette type) posterior

(3)

Set the tube film density

(4)

Adjust

(5)

Adjust the the inside

(6)

Set

the

X-ray control

(

]

the

with film (Sakura to the detector as

voltage regulator

AEC so beam size

timer

of

for for

that

the

limiting of the the

90kV, “0”

A type) and screen shown in the above

the tube posit ion

film

density

X-ray

control

-

94 -

to

current for lOOmA and and make an exposure. of

device that the attenuation block.

(Toshiba figure.

0.85

to

exposure

0.3sec.

or

1.15 field

longer.

is

EM

the

got.

conforms

to

No.

(7)

Take

10

consecutive

percent

measurements

line-voltage All

measurement. within

-+l%

of

Measurement

/ X-ray 1

I(

conditions The The au

t

I

3

(Tube

1 the

tube

I 1 The

maximum

voltage tube

proper

time

current

to

1 The

proper

; p roper

tube

time

This func

(1)

time t ion.

can

to

be

Exposure

that

greater

than

by

equipment

to

b. C.

1 “0”

Keithley

b.

position

I

/

“0”

pos it ion

time.

that

by

of

1 “0”

position

I

an

ON/OFF

for

equation

!

the

oscilloscope SIGNAL”

with

terminal

a of

storage each

of AEC.) plural detecting fields, test For example, test should field.

detecting fields

get

2.

using

diagram with

“C”

can

1,

the

detector

with

3

fields.

the in

10 exposure step 6 of

measurements, Paragraph B-l,

the

following

dosimeters.

is

not

0.045.

model

35020

one with

of 96020

chamber

(Refer

to

TOSHIBA

APPENDIX

> 550-4,

550 with 550-3, Victreen model to TOSHIBA APPENDIX 026A.j Victreen

model

APPENDIX

008A.

555

with

lmA,

1OmA or

550-5

or

550-6

1DAS chamber

(Refer

probe

(Refer

to

TOSHIBA

>

measurement

ALCO primary 012A.

,

required

Electrical a.

I positions

change

“F:q:POSURE

Radiation measurement This measurement requires

006A.

got,

measured

the product using the

Density

time)

time to 0.3

proper

condition

each

table.

change

current

3 detecting

following

level.

connection the detector

taken

to

be

the

the

system the

not

the

is

the of

the

get

and

be

level.

proper

be

point

done

of

for

got,

The

measurements.

current,

step.) not be

can

should

Confirm calculated

a.

(2)

all

according

proper

the

except

should be

the

voltage

AEC. (See In the case

Test

done

Tube

this can

that

Measuring

C.

be

current

the tube

Note:

to

voltage

/ the

(8)

for

value

voltage,

tube

When

i

mean

should

(See NOTE in proper time

the

The

I

regulation values for

within a time period of 1 hour. shall be determined for each line-voltage regulation shall percent

minimum tube voltage of the system tube current that can get the proper omatic exposure control time of 0.1

1 sec.>. ’ When

12

the

ZJ313-002E$cB

voltmeter

model

PV-401

(Refer

to

TOSHIBA

APPENDIX

>

Tektronix 059B. >

storage

oscilloscope

-

model

95 -

464

(Refer

to

TOSHIBA

APPENDIX

1

No.

D.

Rejection

limit

(1)

Coefficient

(2)

Average

of exposure

variation

c = < 0.045

ratio

1x1 - x2/ Xl

-

+ x2

96 -

2 0.09

2J313402E*B

No.

APPENDIX 05 THE SETTING TECHNIQUE FACTOR FOR REPRODUCIBILITY TEST WITHOUT A. E. C. 1.

2J313-002E+B

AND LINEARITY

General which determines how the combination of technique fundamental idea, d, is to select a peak tube potential factors for the test are selecte middle and high kV) and the product of tube current and exposure (low, time (approximately 5 mAs and 30 mAs).

The

During a series of 10 radiation controls for technique factors reset to the test setting after According

60kV 1OmA 3.0sec

to

the

following

’ I1

-

1

4th exposure

all

variable settings

exposures

are

1

13.0sec

]

:Alternate setting(l))

(Test setting)

1

Resetting 60kV 1OmA 3.0sec

/I

(Alternate setting(2))

-/

10 radiation

procedure,

13.0sec

(Test setting)

Setting 70kV 13mA 2.31sec

exposure measurements, are adjusted to alternate each measurement.

I-

_I

/ (Alternate setting(l))

(Test setting)

_ ____---

----

1

I

Test

Procedure

-

97

-

1OmA / 3.0sec / (Test setting)

-*

and

measured.

No.

2.

Combination

No. Test Settings

of Specified

1 60kV 160mA 0.188sec

Technique

2

2J313-OOZE*K

Factors

4

3

6

7

60kV 20OmA 0.15sec

60kV

60kV

60kV

60kV

60kV

250mA 0.12sec

32OmA 0.094sec

400mA 0.075sec

500mA 0.06sec

640mA 0.047sec

Alternate Settings (1)

50kV 125mA ,0.24sec

50kV 160mA 0.188sec

5OkV 200mA 0.15sec

50kV 250mA 0.12sec

50kV 320mA 0.094sec

50kV 40OmA 0.075sec

50kV 500mA 0.06sec

Alternate Settings (2)

70kV ,200mA 0.15sec

70kV 250mA 0.12sec

70kV 320mA 0.094sec

70kV 400mA 0.075sec

70kV 500mA 0.06sec

70kV 640mA 0.047sec

70kV 640mA 0.047sec

No.

8

T

9

11

10

12

13

14

Test Settings

90kV 1OmA 0.5sec

90kV 13mA 0.385sec

90kV 16mA 0.313sec

90kV 20mA 0.25sec

90kV 25mA 0.2sec

90kV 32mA 0.157sec

90kV 40mA 0.125sec

Alternate Settings (1)

80kV 1OmA 0.5sec

80kV 1OmA 0.5sec

80kV 0.385sec

80kV 16mA 0.313sec

80kV 20mA 0.25sec

80kV 25mA 0.2sec

80kV 32mA 0.157sec

Alternate Settings (2)

1OOkV 13mA 0.385sec

1OOkV 16mA 0.313sec

1OOkV 20mA 0.25sec

1OOkV 25mA 0.2sec

1OOkV 32mA 0.157sec

1OOkV 40mA 0.125sec

1OOkV 50mA O.lsec

No.

15

13mA

18

17

16

19

Test Settings

140kV 50mA O.lsec

140kV 64mA 0.079sec

140kV 80mA 0.063ec

140kV lOOmA 0.05sec

140kV 125mA 0.04sec

Alternate

130kV

130kV

130kV 64mA 0.079sec

130kV 80mA 0.063sec

130kV lOOmA 0.05sec

145kV lOOmA 0.05sec

145kV 125mA 0.04sec

145kV 160mA 0.032sec

(Note)

No. 6 and No. 7 should

be tested

-

98 -

only

for KXO-50F.

No.

APPENDIX

TEST

06

METHOD

OF AUTOMATIC

EXPOSURE

CONTROL

2J313-002E*B

LIMIT

A. Objective To determine that the product of X-ray tube current and exposure time is limited to not more than 6OOmAs per exposure when an automatic exposure control is provided.

B . Test

method

(1) Connect

High resistance voltage MA meter divider (KV-201) (MA-1201DC) /

as follows.

Power SUPPl

X-ray tube --&

\)

High-voltage cables

I

4

KV indicating unit (KV-201)

Exposure time counter (TD-3/-3A) (2) Turn

on the main

(3) Press the control. (4) Turn (5) Set

circuit

technique

on the automatic

switch

which

exposure

can use

control

the POWER

ON pushbutton.

the automatic

exposure

switch.

the exposure condition as follows. : 50kV R.kV : more than GOOmAs R.mA and time

(6) Remove the detector X-ray field. (7) Make back

and press

breaker

for automatic

an exposure and record up operates.

exposure

the mA reading

-

99 -

control

outside

and exposure

the

time when

No.

The operation of over-mAs lamp and the buzzer. (8) Verify

C. Test

that

equipment

the product

: &CO time

(3) High voltage meter

D. Rejection

circuit

of indicated

is indicated

value

is not

over

by the BACK UP

540mAs.

required

(1) MA meter (2) Exposure counter

back-up

2J313-OOZE$cB

Model

MA-120lDC

: ALCO Model

TD-3/-3A

: ALCO Model

KV-201

(Refer (Refer

(Refer

to TOSHIBA to TOSHIBA

to TOSHIBA

APPENDIX APPENDIX

APPENDIX

004B.) 018A.)

003A.)

limit

The product of tube current and exposure time measured on the mA meter the exposure time counter must not exceeded the following limits: Maximum

product

of tube current

and exposure

-

100

-

time

limit

:

540mAs.

and

No.

APPENDIX 07 A.

2J313-002E*B

TEST METHOD OF AUTOMATIC EXPOSURE MINIMUM EXPOSURE TIME

Objective When the X-ray tube potential is equal to or greater than 50kV, the minimum time for field emission equipment rated for pulsed operation shall exposure less than a time interval equivalent to two pulses, and the be equal to or time for all other equipment shall be equal to or less minimum exposure than l/60 second or a time interval required to deliver SmAs, whichever is greater (See 102.31(a)(3)(ii)).

B.

Test (1)

method Set

the

X- ray

apparatus

and

test

instruments

according

X-ray

X-ray

Beam limiting

I

I

device

Acrylic

tube

phantom

to

Fig.

2.

detector

(5cm)

Power SUPPlY I

I

L

I

High resistante voltage divider (KV-201)

H.P. transformer

.

2

AEC selector

diagram

of

test

(2)

Turn

(3)

Set the density

the tube voltage for 90kV, regulator for “0” position.

tube

(4)

Set

acrylic

: 5cm)

(5)

Perform

the

the

Block

phantom and

an exposure

Confirm

that

the

exposure

Measured time on the than l/120 second.

method

switch.

(thickness observe

the

time

is

within

time

exposure

-

101

counter

-

current

actual

In the case of the detector with multiple be taken to each combination of detecting should be done under 7 conditions (23 fields. (6)

-

,

L

L Fig.

on

Automatic exposure control

X-ray control

in

for

lOOmA and

front

of

exposure

detector.

fields, test should For example, test the

rejection

should

film

time.

detecting fields. 1 = 7) for

the

the

the

detector

with

limits.

be equal

to

or

less

3

C.

Test

(2)

reqE:lrea

eqci?men:

Es?osure

tine

-+a couLl,,r

: AK0

?lodel

I’D-I/

-3A

(Eiefer

time

counter

to

TOSF-:I3A

_AI??EIiDIX

0!33.?

D. Rejection

limit

The exposure time following limits. Measured

measured

zirne

on the

5 l/120

exposure

second

-

102

-

must

no:

exceed

the

AP?ENDIX A list

TEST

OS of

test

TOQLS

AND

equipment

required Test

Test

INSTRUMENTS

REQUIRED

is shown for

equipment

below. installation

equipment Remarks

Item

Type I Full

AC Voltmeter Oscilloscope High

Voltage

Analyzer

Tektronix

464

Machlett Dynalizer

II

Digital

V 0 M

Simpson

Primary

Voltmeter

ALCO PV-401

Acrylic phantom Water phantom

or equivalent

or equivalent

thickness

are necessary

300/6OOV

or equivalent

260

or

test equipment The following the above equipment.

range:

cnly

for

18Omm

HHS test

in addition

'Test equipment Remarks Item

Type . Keithley model 35020 with 96020 chamber . Victreen model 550 with 550-3, 550-4, 550-S or 550-6 probe. . Victreen model 555 with ImA, lOmA or 1DAS chamber

Dosemeter

Lead

300 x 300 x 2 (mm>

sh.eet

Al attenuation

200 x 200 x 19 (mm)

block -

Al attenuat:on Acrylic

Note:

These

phantom

are used

for

block

(Note)

thickness

(Note)

testing

200 x 200 x 38 (mm>

of Automatic

- 103 -

Exposure

5Omm

Control.

to

TOSHIBA

TOSHiBA

CORPORATlON

1385,SHIMOISHIGAMI,OTAWARA-SHI,TOCHiGl-KEN

324-8550,JAPAN

View more...

Comments

Copyright ©2017 KUPDF Inc.
SUPPORT KUPDF