KXO30-50F Perfrmance Check
Short Description
Download KXO30-50F Perfrmance Check...
Description
No. 2J313-002E"P
TOSHIBA
PERFORMANCE Ct-tECKMANUAL FOR DIAGNOSTIC X-RAY GENERATOR MODELS KXO-3OR, KXO-30F AND KXO-50F (23313-OOZE*P>
TQSHIRA
CORPORATION
%? TOSHIBA CORPORATION 1986 ALLRIGHTS RESERVED
No.
2J313-002E*P
REVISION RECORD REV.
DATE!
REASON
PAGE /AUTHOR
CHANGED
No.
DOC PRODUCI
--__-_-
INI.
'86 Apr. 18
NC
Mr. Wakai
:A
:B
:C
:D
FE
:F
'G
*H
.I
'J
:K
FL
FM
:N
:O
'87 Sep. 11
Correction
'88 Feb. 20
Correction
'88 May 11
Correction
'88 Dec. 23
Correction
'89 May 26
Correction
'89 Nov. 27
Correction
'90 Mar. 22
Correction
'90 May 21
Correction
'90 Jul. 20
Correction
'90 Sep. 20
Correction
'90 Nov. 30
Correction
'91 Oct. 15
Correction
'92 Jan.
Correction
03/'92
Correction
Mr. Wakai
Mr. Ishiyama
Mr. Kimura
Mr. Ishiyama
Mr. Kimura
Mr. Kimura
Mr. Kimura
Mr. Mochizuki
Mr. Ishiyama
Mr. Ishiyama
Mr. Kimura
Mr. Kimura
Mr. Kimura NC Mr. Ishiyama
.P
111'93
P.25, 52
Correction Mr. Ishiyama
*
R-l
No,
25313-002E*J
CONTENTS
Page 1.
INTRODUCTION
2.
PREPARATION
3.
______________-_-__---~-------~~~-~--~--~~-------~
1
______________________________I_________-~~-~---~--
1
2.1
Test Tools
2.2
Precautions
2.3
Radiation
BEFORE
TURNING
3.1
Visual
3.2
Power
4.
SYSTEM
5.
PERFORMANCE 5.1
5.2
Instruments
---_-_-----__----_----
Required
____-________-___-__~-~----~~----~--~_~___~___
2
_____________-______-------------___-----
3
POWER
ON -------_--_------------------a--_--a--_
5
-----_--_-_-----_----------------~~-~~-~
5
Source
5
_--_---_--_------__----~-~~~------~-_-~--_~-_
5
_---__--___--_--_-_----~~---~~~--~-~--~-~-~-----~
CHECK Test
-_-_-_---_____-_--__-_~-~-~_--~-~--~____-__-_
6
___-__-_____-_--____-~-~~~-~-~~~-----~~~--~~~~~~
8
5.1.1
Power
5.1.2
Checking
voltage
5.1.3
Checking
operations
5.1.4
Checking
the switching
5.1.5
Checking
the starter
circuit
5.1.6
Checking
the battery
_---_-_-_____--____-----------
Radiography
1
Safety
Inspection
SETTING
Unit
and
ON -----_---__-_-----_------------------
Adjustment
-----
-------__------------~~----~-----and display
circuits
----------
8 10
--------------------
11
----------------------
11
circuit
-_------_-_---___-----__-_--_------
Radiographic
tube current
adjustment
--------------
15
5.2.3
Radiographic
tube voltage
adjustment
--------------
24B
5.2.4
Checking
Automatic
5.4
Fluoroscopy
Exposure
Adjustment
5.4.1
Measuring
5.4.2
Fluoroscopic
-------------------
29
----------------------
30
______-__------------
30
-------------------
30
timer
Control
(KXO-30F/SOF)
instrument tube
-a-
connection
current
-_-_-
13
5.2.2
the radiographic
_---_______---
13
Measuring
Checking
connection
12
5.2.1
5.3
instrument
8
adjustment
--------------
30
’
No.
2J313-002E”B
CONTENTS
Page
5.4.3 5.5
ic
Fluoroscop
HHS Tests
tube
voltage
adjl .is tment
31
--_--_-_---___
32
____________________-______-_-_--_~--_--_~-~_~~~
5.5.1
Fluoroscopic
5.5.2
Reproducibility
5.5.3
Automatic
exposure
control
limit
5.5.4
Automatic
exposure
minimum
exposure
entrance and
exposure linearity
rate
32
---------------
32
--___________________
32
__________________ time
32
----------
6.
SYSTEM ACCURACY ___--_---__--__-____--_-~-_______-~~--~------~-
33
7.
ATTACHED DATA __-__-------_-_-___-___-~_--~_-----~-~-~~-~~~----
33
7.1
PWB Setting
7.2
Error
Code
and Table
Adjustment
Tables
_________-__---_-__-----
--_______________________________-__--_
8.
MAINTENANCE SCHEDULE __-___-________-_-_----_----_-e-----------w
9.
APPENDICES
-__-____-___-___-_____________-___~__--__----~-~
-b-
33 ---
70
71 ---_
71
No. 2J313-002E*A
1.
INTRODUCTION This manual applies to the diagnostic X-Ray Generators, models KXO-30R, KXO-30F, and KXO-50F, and describes the system setting, the electrical and test to ensure the proper X-ray performance check, and the adjustment exposure precision. Before adjusting the generator as described in this manual, check that the equipment has been installed according to the installation manual (2C313-020E) and understand operating method given in the operation manual. Refer to the wiring diagram the adjustment or test.
supplied
with
These tests are to be performed after requirements of the Radiation Control and all regulations required in Title Regulation.
the equipment
when
performing
installation to meet the for Health and Safety Act of 1968 21 of the Code of Federal
PREPARATION
2. 2.1
Test Tools and Instruments Required Prepare the measuring instruments given
Measuring No.
below.
instrument Remarks
Hodel
Item 1
; AC voltmeter
2
I ’ Digital
3
, Oscilloscope
Full range:
300/6OOV
voltmeter Tektronix
434
Or equivalent Or equivalent
4
X-ray tube voltmeter (bleeder resistance included)
ALCO kV201(D)
5
X-ray secondary tube ammeter
ALCO MA-12OlDC
’
Or equivalent
i I / Acrylic phantom
180mm thickness
Extension PWB (28 AB type) -l-
No.
2.2 2.2.1
2J313-002E*M
Precautions Introduction in X-ray work is protection the most important considerations X-ray equipment properly used does the hazards of X-radiation. careless or improper usage can be not pose any danger; however, it is essential that all service or operating dangerous. Therefore, personnel of this equipment must fully understand the problem themselves. They should also be thoroughly familiar and acquainted with the instructions in this manual pertaining to proper use of the equipment.
One of against
2.2.2
Maximum permissible
dose
Various studies on the effects of X-ray radiation have provided the foundation for establishing maximum permissible dose of X-radiation. The results of these studies used by the National Council on Radiation Protection and the International Comittee on Radiation Protection to develop recommendat ions for the maximum permissible does (MPD). The I.C.R.P. recommendations can be interpreted to say that the maximum permissible dose (MPD) to the most critical organs, accumulated at any by the number of age, shall not exceed 5 x 10 -2Sv (5 rems} multiplied years beyond age 18 and the dose in any 13 consecutive weeks shall not for those involved in radiological Thus, exceed 3 x 10s2Sv (3 rems}. works, accumulation MPD equals 5(N-18) x 10W2Sv {5(N-18) rems), where N For the hands, forearms, and is greater than 18. is the age in years the MPD is set at 75 x 10T2Sv and ankles of radiologists, feet, (75 2.2.3
rems}
a year.
Manufacturer’s Although
this
responsibility equipment
incorporates
protection
against
X-radiation
design cannot provide complete other than the useful beam, practical does not compel the operator protection. The design of the equipment or his assistants to take adequate precautions, or prevent the possibility of authorized or unauthorized persons carelessly, unwisely to direct or secondary or unknowingly exposing themselves or others Therefore, only authorized and properly trained persons radiation. should operate this equipment. All persons authorized to use the equipment must be cognizant of the The equipment is sold danger of excessive exposure to X-radiation. Medical Equipment with the understanding of the TOSHIBA CORPORATION. neither assume nor accept any Division, its agents and representatives, which may result from exposure responsibility for any injury or danger TOSHIBA CORPORATION does not accept any to X-radiation. Furthermore, responsibility for over-exposure of patients or personnel to the X-radiation generated by this equipment due to poor operating no responsibility will be assumed for techniques or procedures. Also, any machine that has not been properly serviced and maintained in accordance with the manufacturer’s technical manual or which has been modified or tampered with in any way.
-
2 -
No.
2.3
Radiation (1)
2J313-002E*M
Safety
Radiation
protection
Since effects harmful to the health may be experienced by anyone amount of X-radiation, the greatest exposed to an excessive consideration should be given to protection against the direct Some of these effects are primary beam from the X-ray tube. The best a period of months or years. cumulative and may extend over is to stay out of the primary beam for X-ray protection safety rules at all times. beam produces a secondary or Any object in the path of the primary The intensity of this scattered radiation scattered radiation. depends upon the type and amount of the primary beam and the atomic This number of the objects or tissues struck by the primary beam. radiation may actually be of greater intensity than the image-forming radiation reaching the film. Since radia
this tion,
radiation protective
dangerous i s as potenti ally to meas ures mus t be taken
beam a s the primary it. sa fegua rd against
out with lead shielding These protective measures may be carried Items such as movable lead screens and lead-impregnated devices. can be used to minimize possible dangerous rubber gloves and aprons devices have a lead thickness It is imperative that these exposures . they should be used by those of a least 0.25 mm or the equivalent; persons exposed to radiation fields of 12.9 x 10-7C/(kg=h) {5mR/h) or During servicing or while operating this equipment a person’s more. or other parts of the body should unprotected hands, wrists, arms, never be exposed to the primary X-ray beam. (2)
Monitoring Monitoring adequacy improper radiation
of
personnel
a valuable crosscheck of personnel provides It can aid in revealing of radiation safety. radiation protection practices, and potentially exposure situations.
of the inadequate
or
serious
This should be confirmed by measurement in rads. Adequate protection measurement should be taken to determine that the X-ray radiation at A less any location does not exceed the accepted tolerance dose. means of confirming this protection is but frequently used, accurate, strategic locations for a given to place unexposed film at several The use of fluorescent screens, period prior to film development. to the darkenss, may also aid after the eyes have become accommodated in locating inadequate protect ion areas. Another common method of radiation monitoring is the use of film badges. These devices consist of an X-ray sensitive film inserted into a badge which incorporates metal filters of varying degrees of Even though this device can only measure the X-ray transparency . radiation reaching the particular area of the body it is worn on, it does furnish a clue to the amount of radiation that is being received. All these regarding
measures are X-ray protect
unless meaningless ion are known and
-
3 -
the rules are rigidly
of safety followed.
No.
Refer to the following requirements.
recommendations
for specific
2J313-002E;kB
safety
(3) Recommendations It is vitally important that everyone associated with X-ray work become fully acquainted with the recommendations of the National Bureau of Standards, the National Council on Radiation Protection, The control and the International Committee on Radiation Protection. of diagnostic medical X-ray equipment varies in detail from state to of the National state; however, in general, the recommendations Council on Radiation Protection based on the other authorities, are recommendations in this manual. However, prior to operation, all personnel qualified and authorized to operate this apparatus should become familiar with and, where pertinent, conform to the established recommendations contained in the documents provided by the The current sources of information are aforementioned authorities. listed for reference. a. National Council on Radiation Protection No. 33 (Medical Gamma-Ray protection for Energies up to 10 MEV-Equipment and Use).
X-Ray and Design
b. National Bureau of Standards Handbook No. 76 (Medical X-Ray Refer to N.C.R.P. Report protection up to Three Million Volts). No. 33. C.
Current Recommendations Radiation Protection.
of the International
-4-
Committee
on
No.
3.
BEFORE
TURNING
POWER
ON
Minimum check points before Do not turn the main switch breaker ON before the check 3.1
Visual
visual
inspection
(1) All setscrews loose. Various
relays
(3) All connectors (4) All PWBs 3.2
Power
Note:
SYSTEM Refer
unit
items
of the units
and distribution
should
be securely
of cables
are securely
terminal
should
inserted
assembly
inserted
not
are as follows: screws
into the
should
not be
sockets.
be loose.
in place.
Source
. Since prior power
4.
turning power on are given below. of the distribution panel and the power is completed.
Inspection
Internal
(2)
2J313-002E*B
the cables were connected for the three-phase 415 V power source to shipment, reconnect the cables according to the installed supply (in accordance with section 6.4 of manual No. 2C313-020E). In the case of a single-phase power source, change the charging (Change the TDR operating time within the time of the capacitor. power unit from 30 to 60 seconds.)
SETTING to Table
7.1 "Setting
and Adjustment
-5-
Table".
No. 2J313-002EfcD
5.
PERFORMANCE CHECK The performance check flowchart is given below: Note: Check to see that the shielding terminal of the cable (for serial transmission) is connected to the cabinet for noise prevention. If it is not connected, error 700 (a communication error> may occur.
PWB unit with control box
1
SPIO or SIO-1
PWB
Chassis
BSXlZ-0276(5 pairs, 7.2$, 10m)
[I.
\
Y
-6-
Do not connect the panel or stand-side shielding terminal.
No.
Start
i
1
2J313-002EkB
1 Cable power
Note:
connections supply.
must
comply
with
the
1 Connect
cables
in
accordance
with
the
system
connection
diagram. I
I
connected with Only the power supplies system cables can be used externally.
Note:
I
I
Disconnect
the
connector
from
the
trigger
PWB within
the
INV unit.
t System setting “when installed”
(for all columns
items indicated of tables in
Note : The charging power supply.
v
Energizing
i
I
as “necessary” section 7.1. > time
in
must
the
comply
with
the
)
I t
) Unit
test
I sequence Filament setting
(sections
5.1.1
5.1.6)
check
I
heating
check
rConnect
to
the
connector
data
and heating
(cable
length
and
tube)
t Power
to
off
the
I
trigger
PWB within
the
INV unit.
I Connect the connector citor voltage become
Radiographic
& fluoi!oscopic
adjustment
(sections
5.2
after the capaless than 5V.
and
5.4)
ABC/AEC adjustment
ATR adjustment
‘I (
End
Note : Enter )
the
adjustment
data
in
the
data
sheet.
No. 2J313-092E
5.1 5.1.1
Unit
Test
Power
ON
Turn ON the main switch of the distribution panel and the power unit Check for abnormality by turning the breaker within the control box. panel or stand power switch ON/OFF. Within approximately 60 seconds after the capacitor unit inside the control approx. 30 seconds). Note:
5.1.2
After the power is turned off, approximately four minutes.
Checking
the power switch is turned on, box is charged (three-phase:
discharging
is completed
in
voltage
(1) KXO-30R (a)
Voltage
within
the panel
Panel Cont 1 PWB Between CH2 (+5 V> and CHl
(GND):
If the above allowab!e voltage the VR of the stabilized power Control
box:
Pl witnin
Make sure that +24 v+3 - v. (b) Voltages Mother
the voltage
within
value source
Unit
V
is not indicated, given below.
adjust
1
between
the control
PWB of PWB Unit
. Between . Between . Between
Power
+5 V,O.l
CH3
(+24 V> and CHl
(GND)
box
1
Jl (IAB) and Jl (28AB) : +5 V+O.l V J16 (2AB) and Jl (28AB): +15 ~~0.1 V J16 (3AB) and Jl (28AB): -15 VzO.1 V
If the above allowable voltage values are not indicated, the VRs of the stabilized power source given below. p2 (+5 v), ~3 (2) KXO-30F (a>
is
(+I5 V> and P4 (-15 V> within
Power
Unit
adjust
1
or 50 F
Voltages . Desk
within display
Between
CH2
the stand PWB (+5 V> and CHl
(GND):
+5 V+O.l V
If the above allowable voltage value is not indicated, adjust Make sure that the VR of the stabilized power source (PSZ). V. the voltage between CH3 (+24 V> and CHl (GND) is +24 V+3 -
-8-
No.
. Desk
control
Between
PWB
P2 (+5
V> and
pl
If the above allowable the VR of the stabilized (b)
Voltages Mother
within
(3)
ths
PWB 2 of
. Between . Between . Between
PJl PJl PJl
2J313-002E*B
(GND):
voltage power
control
PWB Unit
(4AB) (2AB) (3AB)
+5 V+O.l value source
v
is not (PSl).
indicated,
box 2
and PJl and PJl and PJl
(28AB): (28AB): (28AB):
+5 V+O.l V +15 V~0.1 V -15 V+_O.l V
If the above allowable voltage values the VRs of the stabilized power source
are not indicated, given below.
Pl
within
(+5
LV detection
V),
P2 (+15 circuit
V),
adjust
and P3 (-15
V)
Power
adjust
Unit
2
adjustment
in following Adjust VR2, VR3 and VR4 on INV3 PWB as instructed Tl on the power unit table only when the input tap o f transformer This adjustment , however, is in the control box indicates 200 V. been set. not required when other taps have already
VR to be 1 adjusted
Measurement point
1 VR2
TP2
VR3 VR4
Tl
j
220 V,
200 v
415
V,
v
+4.75
v
v
+5.25
V
TP3
+4.87
TP4,
+5.35
/
Setting
-9-
380 v,
v
v
LED3 through After adjustment, the LV levels with respect to value) (see the table below).
tap
+4.50
+4.75
/
input
2
prior
LED7 of the INV-3 PW board the nominal value (Tl input
to
440
V /
shipment indicates tap
No. 2J313-OOZEW
Nominal
Other values
200V
nominal = Vs
LV>zzoVQc
LV>V$$-10%
220v~Lv>214v
INV-3
Radiography flu scopy disabled area
LV LEVEL
s m % i LED7
Operating area
vstlo%~Lv>vst5%
u R64
+i
9 LED6
vs t5% 2 LVL vs -5%
214’LV2195
Optimum
operat
u
04
R62
,Q u LED5
Operating
Vs-5%>LVLVs-10%
195>LV 2 190
area
:
R61
c z
5.1.3
Checking (1)
operations
and
Radiographic
(2)
the
Checking Repeating indications . . . . . .
circuits
settings
condition individual
reading ID No.
settings and indications operation manuals.
when
the radiographic according to the
turning
turning equal
power those
power
conditions operation
of
the
in and from manual, check
units
an for
the
ON/OFF
that ON, OFF, and ON, check before turning power OFF.
the
following
R kV value
R mAs value R mA value
R TIME value TEC. SW SF/LF
Make sure that the contents turning power ON and OFF. (4)
u R60
LED3
ID function Writing and appropriate ID function.
(3)
display
condition
Check radiographic as described in
zu% ii:
Radiography fluoroscopydisabledarea
vs-lO%>LV
19O>LV
LED4
Overload Check X-ray X-ray
written
in
the
ID are
not
destroyed
by
protection
function under conditions of the the overload protection tube maximum rating, filament characteristics, kV limit, generator rating (40 kW, 60 kW).
-
10 -
set and
No. 2J313-002E"cA
5.1.4
Checking
the
(1)
potential
High
switching
circuit
changer
(KXO-30F
or
50F)
(HRU-A) of the Operating the changer selecting TECHNIQUE via the panel or changer changes over normally. (2)
Focus
change
Checking There
the
are
(a)
Normal
(b)
High
(c)
Anode
panel
starter
three
of
stand
types
of for
make
for
for
HSS is controlled by the starter frequency wiring, X-ray
indication
focus
anode
radiography
rotation
rotation
tube
fluoroscopy
the is
rotation
as
(Named
NSS)
radiography
NSS and F-ROT are controlled The starter X-ray control. (50 or 60 Hz).
Check the installed
stand,
generator sure
by
that
the
varies
with
the
set
circuit
rotation speed
voltage
by R-MA
Check that the tube current.
5.1.5
high
(Named
(Named
shown
HSS)
F-ROT)
by the normal speed frequency equals the
high speed 180 Hz.
starter
phase advance capacitor tubes as described below.
below.
starter within the power source frequency
unit
(RS-118A:
value,
and
set
option),
time
of
and
the
[NSS and F-ROT modes] ( 1)
Check tubes (a) (b)
the type depending
XS-U (standard stator) XS-AA (1 ow impedance
For detailed tube technical (2)
of the X-ray tube combined. on the X-ray tube stator.
Phase
advance
The relation X-ray tube
X-ray
stator data.
tube
refer
to
are
two
types
phase below.
Capacity
advance
the
capacitor
Capacitor the power
TOSHIBA U coil
24uF
Cl
TOSHIBA AA coil
44uF
Cl
individual
value
No. within unit
+ c2
cl: 24uF, C2: 20uF
-
11 -
of
tube X-ray
value
between the type is given
type
TOSHIBA X-ray tube stator) TOSHIBA X-ray
characteristics,
capacitor
There
and
the
2J313-002E”cA
No.
(3)
Time
setting
The relation given below.
between
the
I
X-ray
tube
I / /
type
The oscilloscope measured.
(a)
Check
(b)
Set
that the
is
power
on
to
X-ray
tube
type
is
1
time
be
isolated
observe
from
the
betweenj
power unit within storage mode)
and
the
circuit
to
be
off.
oscilloscope
Block TB2 of the div, 0.2 sec/div,
and
+o. 1 set -0
should
power
time
Boost
I 1 0.8 I
TOSHIBA U coil
Note:
boost
select
Tube
the
and / X2 control
box.
of
Terminal
(100
VI
1.
(c)
Turn
(d)
Set a low kV value, a low mA value, Normal Speed Rotation (NSS) mode.
(e)
Put into the oscilloscope
and a short-time
Ready state. Observe the and check t’ne boost time.
waveform
on
to
keep
in
the
[HSS mode 1 (a)
5.1.6
Set the stator coil, start voltage, and control time control voltage, installation manual. speed starter
(b)
Select
(c)
Set a kV value, Rotation (HSS)
(d)
Put into rotation
Checking
the
the
TECHNIQUE of
the
and
a mA value, mode.
the Ready according
state, to the
tube
as
to
R start time, specified in
F start time, the high
be checked.
a time
to
keep
in
and check the high speed said installation manual.
High
Speed
anode
battery
Each unit PWB is provided with a chargeable which stores the memory contents when power both terminals and check the voltage between given be low.
-
12 -
battery or turns off. according
dry battery, Turn power to the table
off,
No. 2J313-OOZE*1
Equipment
KXO-30F/50F
name
Cont
1 PWB
Voltage 3.6
V to 2.4
Heat 2 PWB of PWB Unit 1 within the control box
2.5
V
Stand
3.6
V to 2.4
Panel
KXO-30R
Unit
desk
control
PWB
Heat 2 PWB of PWB Unit 2 within the control box
5.2
Radiography
V (Chargeable)
V (Chargeable)
2.5 V
Adjustment
The adjustment method for tube voltage and current is given below. The R.mA and R.kV adjustment procedures are given separately below. and R.kV should be adjusted simultaneously at each mA step. 5.2.1
Measuring
instrument
The measuring
R.mA
connection
instruments
should
be connected
as shown
below.
(1) KXO-30R Set up the required units as illustrated the tube voltage and current. Control
source
generator
\ -9
and adjust
--
Secondary tube current meter (to be mounted to the X-ray tube) \
Anode Bleeder resistance
DG-303
Power ground terminal
both
box
G--J Power
below
/
Storage oscilloscope
-
13 -
\ f-A-X-ray tube
Cathode1
No.
Set up the required units tube current and readjust It is not necessary voltmeter-ammeter
Control
is
as it
to perform used for
illustrated if necessary.
below,
this procedure measurement.
box
25313-002E*l
check
if
the
the
X-ray
tube
Secondary tube current meter (to be mounted to the X-ray tube)
Anode X-ray tube
High
Power source
voltage -_:l 1
--
5
--Jr
Cathode
I
1 Power ground terminal
-I
. Do not generate X-rays Prior to checking the (within the TNV unit) l
with Tl and T2 opened. disconnect X-ray sequence, connector or the capacitor.
Note : Before disconnecting the capacitor voltage
the is
- 13A -
connector or 5 V or less.
the
capacitor,
TRG PW board
check
that
No.
(2)
KXO-30F
and
25313-002E*J
50F Secondary tube current meter (to be mounted to the X-ray tube) DG - 50F
Power
source
Control box
n \ &-7
Tl
Tl
_ T2
T2
---
High voltage gener-
lnode
ator
Cathode
Bleeder resistance
X-ray tube Cathod
--A
Tube voltmeter
-f-
Power ground terminal
Storage oscilloscope
. Do not generate X-rays to checking the . Prior (within the INV unit) Note:
with Tl and T2 opened. disconnect X-ray sequence, connector or the capacitor.
Before disconnecting the capacitor voltage
-
the is
14 -
connector or 5 V or less.
the
capacitor,
TRG PW board
check
that
No.
5.2.2
Radiographic tube current adjustment (Applicable to products manufactured from April, 1989 to Refer to “Radiographic tube current adjustment” described for products manufactured in and after July, 1990.) (1)
Operation
switch
and
adjustment
The HEAT 2 PW board, which with the operation switches below:
cl>
Mode indic
x2>
Mode
tor
has an adjustment function, and indicators give in the
I
SW
MODE
Adjustment
S/R
switches
switch
-
S/R
-
June, 1990. on page 24B
mode
TEST
selec
15 -
25313-002E*J
is provided drawing
25313-002E*J
No.
Mode indicator from 0 to This LED changes mode selector switch. Each mode is to be used as
1 to
2 and
follows
finally
to
according
to
7 with the
the
adjustment
type* Mode
I
0
Adjustment
Indicator
I
1
2
Type
moo
Standby
a.0
tube current Radiographic (Point adjustment >
adjustmnet
000
Radiographic tube (Line adjustment)
current
adjustment
current
adjustment
state
3
moo
4
00.
Fluoroscopic
5
0.0
PRH and
6
000
Cable
7
000
Tube setting error>
tube
PRF level
length
adjustment
setting (recovery
from
a data
0 ON (Lighted) Pressing
the
S/R
switch
causes
any mode
to
be
reset
to
Mode
0.
Mode selector switch The mode selector switch is used to select the mode and to check the main heating level. Each time the switch lever is set in the downward position, the The mode mode is switched from 0 to 1 to 2 and finally to 7. is then reset from 7 to 0. When the switch lever is set in the upward position, only the filament heating circuit becomes ready for main heating so that the heating adjustment level can be checked operations (in Modes 0, 1 and 2). Adjustment switches These switches are level and to input In Modes 1, code.
If
that
the
A, sl
caused
caused
or decrease cable length
the heating and the tube
switch
is
pressed,
voltage
at
Eo increases.
switch
is
pressed,
voltage
at
Eo decreases.
switch
is
pressed,
the
voltage
increases
ten
times
the
voltage
decreases
ten
times
by pressing switch
that
used to increase the high-voltage 2 and 4,
is
the pressed,
by pressing
(Note)
the
- 16 -
NO. 25313-002E*J
Note:
Eo is a check pin check pin AGND is Eo corresponds to transformer.
on the HEAT 1 PWB. used as the standard the primary voltage
Voltage for of
at
the
that at Eo. the filament
Whenever each of these switches is pressed, the buzzer will will increase or sound and, in Mode 1 or 2, the voltage decrease to approximately 0.05 V (to approximately 0.5 V when If the switch continues to be the tenfold switch is pressed). the buzzer will automatically start sounding so that pressed, the result can be achieved as if the switch were pressed by the number of times the buzzer sounded. for using these switches to input the The procedures in the high-voltage cable length or the tube code are described section
(2)
concerning
adjustment
methods
in
each
mode.
S/R switch This switch is used to reset Mode 0 from any mode or to input the high-voltage cable length and the tube code simultaneously using the adjustment switch. The same result can be achieved by setting the S/R switch lever in either the upward or downward position.
Adjustment Mode
Mode
procedure
6
7
Tube
code
setting /
Mode Mode
2 1
Radiographic adjustment
Mode
tube
current
tube
current
__,__ I I I I I I I I I
Mode 4
Fluoroscopic adjustment
(Not
required
First perform setting for (Only the cable three tubes. length setting will not cause the heating data to change. > Note: When the X-ray generator is used with an X-ray tube with a very small focus, such as the X-ray tube for magnification radiodo not connect graphy > tube H.T. cables until the tube code is set. Failure to do so may damage the tube focus.
for
KXO-30R) -
17 -
5
PRH/PRF
level
adjustment
No.
(3) Setting
the
high
voltage
cable
length:
Mode
6
The standard cable length is 15 m (for three tubes). change the cable length depending on the system.
--+
Set
Mode
6 with
Set the tube X-ray control If
the
No. to unit.
swith 1) cl 0
equipment
the
goes
is
into
mode
selector
the
data
If
necessary,
switch.
be changed
via
the
operation
the
indicator
input
state.
pressed,
25313-002E*J
panel
blinks,
of
and
the
the
Enter the new cable length using the adjustment switches and the S/R switch. If the two-digit (00 through 99) input is completed, the mode resulting in the condition in . indicator stops blinking, [Example:
6 ml
(Do not the
/ cl
First time
press switches)
aI
I
[-ZJ
4
[4
S/R
0
Press 4
Press L
[z.][-zj m 1-q
Second time
0 If
Perform voltage
0
[ZJ 0
0
0
Note:
Press
the
data
is
the operations cable length
When the S/R
switcll
the
tube.
0
1
1
correctly (2) for
set,
BCD code
0 1
0000 0001
2 3
0010 0011
4 5
0100 0101
6 7
0110 0111
8 9
1000 1001 -
0
aI
S/R
a buzzer
goes
through when changing the 2nd or 3rd tube.
adjustment is completed, or set Mode 7 with the
Value
Press ---f
18 -
reset mode
on. the
Mode 0 by pressing selector switch to
high-
the set
No.
(4)
Tube
code
setting:
Mode
7 Select set as initial setting. from Table 5.2.2 and perform
The DRX-2903HD (for 3 tubes) is each tube number to be combined setting by the following steps. --+
for
of
setting,
tube setting Note : If data adjustment into memory. The small
above
procedure
focuss.
is
the
is is
used
When the
setting, the buzzer again.
2nd
and
press
3rd
the
the indicator sounds and
goes the
tubes.
S/R
switch.
(Restores
completed after other adjustments, deleted and preset data are stored
to
set
the
switch
is
data
with
pressed
both in
step
large ,
the
and data
with either a large or small focus can be set. This setting can be used when the tube to be combined is not listed in Table 5.2.2 and large and small-focus data are to be substituted or when either large or small-focus data with different tube data, is to be reset. for example) is input for a blank tube name If any code (00 0000, the buzzer will sound for about two seconds, column in Table 5.2.2, thus indicating that there is no tube for the input code.
-
19 -
No.
23313-002E*J
is noted during the inputting operation or it is If an input error desired to input data once again, reset the Mode 7 initial state (with the indicator lit) by pressing the S/R switch several times. (5)
Radiographic
tube
current
adjustment
For the radiographic tube current adjustment, Mode 1 is used to ad just each adjustment point, and Mode 2 is used to adjust the tube current (MA) for each line. The radiographic tube current by first adjusting 60 kV in Mode 2 adjustment can be facilitated and then adjusting 40 kV, 90 kV and 120 kV in Mode 1. Adjustment
for
each
Adjustment for 120 kV in Mode When the adjustment The are
point
40, 1.
in Mode
60,
1
Adjustment
90,
for
each
Adjustment
for
MA line
in Mode
60 kV in Mode
2
2.
tube code is set for mammography in Mode 7, four points at 20, 25, 32 and 45 kV are required.
adjustment points for 20, 25, 32 and 45 kV.
mammography
The line adjustment should be a; 25 kV.
for
mammography
25kV 20kV 1 I]
800mA 1 Adjustment for 20, 45 kV in Mode 1.
25,
Adjustment
32 and
I for
25 kV in Mode 2.
tubes) is set before shipment. The standard data (for three Correct the data depending on combinations of tubes. Since the preheating and preflash are not checked, adjust the (Generate X-rays current after the filament heating is stabilized. by pressing the READY switch and the X-ray switch at intervals of them simultaneously. Take note (Do not press more than 2 seconds. of the heating data as shown in table 10. (1) in APPENDLX 01 after adjustment.)
- 20 -
No.
Set Mode
1 with
the mode
selector
25333-OOZE*5
switch.
Set the tube number (1, 2, or 3) to be adjusted operation panel of the X-ray control unit.
via
the
Set R.kV and R.mA to be adjusted in the same way. Some combinations of R.kV-R.mA are limited to Maximum Power (30 kW/SO kW). R.kV: 40, 60, 90, 120 kV or 20, 25, 32, 45 kV R.mA: 10, 13, 16, 20, 25, 32, 40, 50, 64, 80, 100, 125, 160, 200, 250, 320, 400, 500, 640 mA :
Correct
the data with
the
/c i/( cl ;
increase)
or
A, '1 (decrease) n
switches (Note 1) by generating X-rays under the individually set conditions in order to adjust the radiographic tube Adjust all conditions current. to be used in this way. Press the Ready SW and the X-ray SW simultaneously the set conditions again to see if it is necessary preheating or calibration of preflash level.
and confirm to perform
After checking the above, if the tube voltage is stabilized within a specific time, Item (6) "Preflash level adjustment" level adjustment" are not required. and (7) "Preheating When completing Mode 0. Note
1: When
adjustments,
press
switch
the
increased the B]
the S/R switch
or decreased or [q
switch
is used,
ten times
to return
the value
as compared
to
can be
to when
is used.
in this mode causes 40, 60, 90 and 120 kV to The adjustment simultaneously increase or decrease. is the same as Mode 1, except that this Mode 2 adjustment adjustment should be made with the R.kV tube voltage set to 60 kV. When the X-ray tube for mammography is used, the tube current adjustment in Mode 2 should be done at 25kV. (6) Preflash
level
adjustment
(PRF): Mode
5
The preflash level should be adjusted only when the tube current does not rise within the specified time in (5) . Preflash is intended to increase the X-ray tube filament rise rate. After the READY switch is pressed, the preflash is done prior to main heating. ~Preflash Heating
I
level
Main heating
Preheating
I READY w ON )) - 21 -
Time
No.
25313-002E*J
time is fixed and the amount of For this unit, the preflash preflash can be obtained by multiplying the amount of main heating by the preflash factor. the preflash level is adjusted by changing the preflash Therefore, factor. The preflash factor curve is generated by preflash factors at three The three points are the maximum, minimum and middle points. points of the filament current on the X-ray tube current characteristics curve. The three preflash factors, PRFl, PRF2 and PRF3 are to be selected with
the adjustment Tube
Tube
and
[,,[a, Current
[Tg
switches*
Characteristics
factor
current (l.mQ
Middle
Minimum Filament PRFl, PRF2, and PRF3 selected as follows:
Maximum
Current
(A)
for the three
adjustment
points
can be
Plot the X-ray conditions under which the preflash level is to be If the curve. adjusted, on the tube current characteristics If the filament current is near the maximum value, adjust PRFl. filament current is near the middle or minimum value, adjust PRF2 or PRF3, respectively. If the filament current is between the minimum and middle values, If the filament current is between the adjust PRF3 and PRF2. middle and maximum values, adjust PRF2 and PRFl. --+
Set Mode
the mdoe
5 with
selector
switch.
Set conditions under which the preflash level is short in (5) , via the operation panel of the X-ray control unit.
Select
the preflash
m,
m,
PRF3
PRFZ
sequence
factor
or [z]
to be
switch.
adjusted
The mode
PRFl
proceeds
to the next
- 22 -
step.
by
pressing
indicator
the
blinks,
and the
No.
While checking the tube current rise the READY and X-ray switches, change using
the
[z]
or
Increase L
[a
switch
1).
Decrease
When the adjustment S/R switch. 1:
by simultaneously pressing the preflach factor by
(Note
When the S/R switch is pressed, and the state in is reset. another focus or another factor, adjusted.
Note
25313-OOZE*5
When the
is
the indicator In this state, tube preflash
completed,
[z]
or
[A
reset
switch
Mode
is
stops blinking another preflash factor can be
0 by
pressed,
pressing
the
the
buzzer
Increase Decrease will sound and the preflash factor will increase or If either switch continues to be decrease by 1%. the buzzer will automatically start sounding pressed, so that the result can be achieved as if the switch were pressed by the number of times the buzzer sounded.
( 7) Preheating
level
adjustment
(PRH) : Mode
5
If the adjustment This adjustment is not necessary in principle. in (6) cannot be carried out satisfactorily depending on the system, this adjustment should be carried out as an auxiliary adjustment to (6).
r
Set Mode
5 with
Set the conditions
the
switch
Press
sequence
is
proceeds
11,
selector
switch.
the
next
switch
preflash panel
the
pressed,
to
the Increase
(Note
mode
under which the via the operation
not satisfactory, unit. When the
I
the
to
mode
of
blinks
is
and
stop. adjust
the
preheating
level
Decrease
and check
When the adjustment S/R switch. 1:
adjustment X-ray control
indicator
th e
tube
current
rise
by generating
When the S/R switch is pressed, the indicator In this state, and the state returns to . another tube preflash factor can be adjusted.
Note
level the
When the m
is
completed,
or [x]switch
reset
is
Mode
pressed,
X-rays.
stops blinking another focus
0 by pressing
the
buzzer
or
the
will
If either sound and the preheating level will change. switch continues to be pressed, the buzzer will sounding so that the result can be automatically start achieved as if the switch were pressed by the number of times the buzzer sounded. -
23 -
No.
(8)
Changing
the
The following adjusted to
Focal
Size
for
Fluorography:
Mode
25313-002E*J
6
procedures permit to change the focal size the small focus (SF) at the time of shipment.
Set
Mode
6 with
the
Use the operation the number of the
mode
selector
already
switch.
panel of the X-ray control tube whose fluorographic
unit to specify focal size needs
to
be changed. switch enabling
mode,
the
To adjust the switch only. For
the
to
place
following
focal
size
focus
(SF),
small
the
mode
to
the
large
tilt
the
focus
S/R
switch.
A buzzer procedure
sounds and the indicator stops has been performed correctly
Reset to mode “0” by means of the size adjustment has been completed.
the
focal
size
in
blinking
steps.
the
To change
indicator
to
S/R
(LF),
switch,
tilt
while
blinking so far. switch
the
if
when
pressing
the
the
J
L
the
focal
size
to
focal
LF Do not
To change
S/R
press
any
switch
Tilt
SF
I+ (9)
Recovery If
from
a data
any
error
data
has
occurred,
recover
When the S/R switch is pressed alarm stops and Mode 7 is set.
(2)
When the
A, sl
recovery
routine [z]
down
with
the
this
switch
Tilt
error
When the
Hold
switch
is
pressed,
while
the
the
following
alarm
indicator
procedure:
sounds,
blinks
the
and
starts. and
switches
[q
are
pressed
simultaneously,
the indicator goes out and the recovery routine (for Upon regeneration of an error check code) is executed. the indicator lights up and the completion of execution, sounds.
the
buzzer
Reset Mode 0 by pressing the S/R switch and check the data. the power supply is turned off in the adjustment mode during is judged to have be destroyed tube current adjustment , data -
24 -
If
No.
2J313-O02E”CJ
when the power supply is turned on again, and the alarm sounds. (This is because it is not possible to distinguish data change recover by using the from data breakdown. > In such cases, After recovery, data adjusted prior to above procedure. turning off the power supply are all effective. Note:
is simply to regenerate the error The recovery routine but not to repair the destroyed data. check code, it is necessary to check the data after Therefore, recovery. If the check shows that the data are changed, correct If data have been completely in Mode 1, 2 or 4. In this it is necessary to reset the tube. destroyed, the setting must start with the setting of the case, high-voltage cable length.
it
> Alarm
Stored data check The stored data is checked by heat data when the power supply is turned on. Data is checked in each tube and each focus (for data blocksTube 1 LF, Tube 1 SF . . . Tube 3 SF). If an in which an error has attempt is made to use the data block an alarm (intermittent beeping sound) and an been detected, To recover from this state, error message will be generated. execute recovery procedure (9) from a data error.
Set condition check than 150 kV or the tube If the set tube voltage becomes greater an alarm (intermittent beeping current is not set correctly, If the tube voltage or the tube sound) will be generated. the alarm will be removed. current is set correctly, Maximum filament current check smaller than the The heating output data is set to a value maximum filament current to prevent the X-ray tube filament If the heating data exceeds the maximum from breaking down. filament current value when the READY switch is pressed, the to the preheating level and an alarm heating level will be set (intermittent beeping sound) and an error message will be generated.
-
24A -
No.
5.2.3
Radiographic (Applicable (1) Operation The HEAT
tube current adjustment to products manufactured switch
and adjustment
in and after
July,
25313-002E"J
1990.)
mode
function, is provided 2 PW board, which has an adjustment switches and indicators given in the drawing
with the operation below:
xl>
Mode
indicator
C TEST
Mode
selector
switch MODE
x4>
Adjustment
switches
-
S/R
S/R switch
- 24B -
No.
Mode indicator This LED changes from 0 to mode selector switch. Each mode is to be used as
1 to
2 and
follows
finally
to
according
to
25313-002E*J
7 with the
the
adjustment
type* Mode
Adjustment
Indicator
I
0
**
l
1
2
Standby
Radiographic tube (Point adjustment)
current
adjustmnet
00.
Radiographic tube (Line adjustment)
current
adjustment
Fluoroscopic
current
adjustment
moo
4
00.
T-- 0.0
6 7
I
tube
PRH and PRF level
adjustment
000
Compensation
for
000
Tube error
(recovery
0 Pressing
state
0.0
3
5
Type
the
S/R
switch
setting >
tube
current from
drop a data
ON (Lighted)
causes
any
mode
to
be
reset
to
Mode
0.
Mode selector switch The mode selector switch is used to select the mode and to check the main heating level. Each time the switch lever is set in the downward position, the The mode mode is switched from 0 to 1 to 2 and finally to 7. is then reset from 7 to 0. When the switch lever is set in the upward position, only the filament heating circuit becomes ready for main heating operations so that the heating adjustment level can be checked (in Modes 0, 1 and 2). Adjustment switches These switches are level and to input code. In Modes 1,
If that If that
the
7 0
caused the
/) LI
caused
used to increase the high-voltage 2 and 4,
or decrease cable length
switch
is
pressed,
voltage
at Vset
switch
is
pressed,
voltage
at
switch
is
pressed,
the
by
pressing
switch
is
by pressing -
the pressed, the 24C -
y4 cl
voltage
the heating and the tube
increases.
Vset
decreases. (Note > increases ten times
switch. the
voltage
decreases
ten
times
No.
Note:
Vset is a check pin on check pin AGND is used
25313-OOZE”5
the HEAT 1 PWB. Voltage as the standard for that
at at
the Vset.
Whenever each of these switches is pressed, the buzzer will sound and, in Mode 1 or 2, the voltage will increase or If the switch is held down, the buzzer will autodecrease. matically sound at fixed intervals; the number of times the buzzer sounds is the same as pressing the switch the same number of times. The procedure for using these switches to input the tube code is described in the section concerning adjustment methods in that mode. S/R switch This switch is used to reset Mode 0 from any mode or to input the high-voltage cable length and the tube code simultaneously using the adjustment switch. The same result can be achieved by setting the S/R switch lever in either the upward or downward position. (2)
Adjustment Mode
7
Tube
Mode Mode
procedure
code
setting
2 1
Mode
Radiographic adjustment
tube
5
current
Mode
4
Note
1:
When the X-ray generator is used with an X-ray tube with a such as the X-ray tube for magnification very small focus, radiography, do not connect tube H.T. cables until the tube code is set. Failure to do this may damage the tube focus.
Note
2:
Diagnostic X-ray generators manufactured after April 1990 is turned ON without connecting cause an error if the power X-ray tubes. Be sure to remove the HEATlA PWB and to ground 12B on the HEAT2 PWB before turning the power ON.
-
24D -
No.
(3)
Tube
code
setting:
Mode
7 Select set as initial setting. from Table 5.2.2 and perform
The DRX-2903HD (for 3 tubes) is each tube number to be combined setting by the following steps.
Set
Mode
Set the
the tube number to X-ray control unit.
When the the
the
code
is
code
selector
be
into
goes
the binary 5.2.2.
[Example]
mode
switch
/) CJ
equipment
Enter Table
Binary
7 with
25313-002E*J
set
switch.
through
the
operation
panel
pressed,
the
mode
indicator
blinks,
the
input
code
and
state.
to
corresponding
of
the
tube
to
be
used
from
DRX-1203B Do not
press
a
[d
00 0010
0
Tilt &
0
First
time
S/R
? Eold
down
0
Note:
this
setting
After completion Mode 0. >
The small
0
Upon completion of correct Shortly after that, out. indicator lights up once
Perform
for
of
If tube setting adjustment data into memory.
above
procedure
focus.
is
the
is is
used
Tilt
S/R
0
setting, the buzzer
the indicator sounds and
goes the
again. and
tilt
3rd
the
tubes.
S/R
switch.
(Restores
completed after other adjustments, deleted and preset data are stored
to
set
switch
When the
1
A
2nd
setting,
Note:
this 1 switch
the is
data pressed
with in
both
large
step
,
the
and data
with either a large or small focus can be set. is not listed This setting can be used when the tube to be combined in Table 5.2.2 and large and small-focus data are to be substituted or when either large or small-focus data with different tube data, is to be reset. for example) is input for a blank tube name If any code (00 0000, the buzzer will sound for about two seconds, column in Table 5.2.2, thus indicating that there is no tube for the input code.
-
24E -
No.
25313-002E*J
If an input error is noted during the inputting operation or it is desired to input data once again, reset the Mode 7 initial state (with the indicator lit) by pressing the S/R switch several times. (4)
Radiographic
tube
current
adjustment
For the radiographic tube current adjustment, Mode and Mode 2 is used to adjust each adjustment point, current (MA) for each line. The radiographic tube adjustment can be facilitated by first adjusting 60 and then adjusting 50 kV, 90 kV and 120 kV in Mode Adjustment
for
each
Adjustment 120 kV in
Mode
for
point
50, 1.
60,
in Mode
1
for
each
Adjustment
for
Adjustment
90,
1 is used to adjust the tube current kV in Mode 2 1.
MA line
in
Mode
60 kV in Mode
2
2.
When the tube code is set for mammography in Mode 7, four adjustment points at 20, 25, 32 and 45 kV are required. The are
adjustment points for 20, 25, 32 and 45 kV.
mammography
The line adjustment should be at 25 kV.
for
mammography
1 20kV 1 25kV ( 32kV 1 45kV 20mA 1 I
50mA lOOmA
lOOmA
l
I
I
I
. . . 800mA 1
800mA
Adjustment for 20, 45 kV in Mode 1.
25,
32 and
Adjustment
for
25 kV in Mode
2.
tubes) is set before shipment. The standard data (for three Correct the data depending on combinations of tubes. Since the preheating and preflash are not checked, adjust the (Generate X-rays current after the filament heating is stabilized. by pressing the READY switch and the X-ray switch at intervals of Take note (Do not press them simultaneously. more than 2 seconds. of the heating data as shown in table 10. (1) in APPENDIX 01 after adjustment.)
-
24F -
No.
Set Mode 1 with
Set the operation
:I
Set
the
mode
selector
switch.
tube number (1, 2, or 3) to panel of the X-ray control
R.kV
and
R.mA to
be
23313-OOZE*5
adjusted
in
be adjusted unit. the
via
the
same way.
Some
combinations of R.kV-R.mA are limited to Maximum Power (30 kW/50 kW). R.kV: 50, 60, 90, 120 kV or 20, 25, 32, 45 kV R.mA: 10, 13, 16, 20, 25, 32, 40, 50, 64, 80, 100, 125, 160, 200, 250, 320, 400, 500, 640 mA
L-
Correct
the
data
with
the
y/ cl ;
(increase)
switches (Note 1) by generating X-rays set conditions in order to adjust the current. Adjust all conditions to be
or
/(/I n
(decrease)
under the individually radiographic tube used in this way.
Press the Ready SW and the X-ray SW simultaneously the set conditions again to see if it is necessary preheating or calibration of preflash level.
and confirm to perform
if the tube voltage is stabilized After checking the above, Item (5) “Preflash level adjustment” within a specific time, and (6) “Preheating level adjustment” are not required. When completing Mode 0. Note
1:
press
adjustments,
switch
When the increased the
[q
the
or
decreased
or [q
switch
is
ten
times
is
used.
S/R
switch
to
used,
the
as
compared
return
value to
can
to
be
when
The adjustment in this mode causes 50, 60, 90 and 120 kV to simultaneously increase or decrease. Mode 2 adjustment is the same as Mode 1, except that this adjustment should be made with the R.kV tube voltage set to 60 kV. When the X-ray tube for mammography is used, the tube current adjustment in Mode 2 should be done at 25kV. (5)
Preflash
level
adjustment
(PRF):
Mode
5
The preflash level should be adjusted only when the tube current does not rise within the specified time in (4) . Preflash is intended to increase the X-ray tube filament rise rate. the preflash is done prior to After the READY switch is pressed, main heating. / Preflash . Main heating I
Heating level
Preheating
1
READY mON ))
- 24G -
Time
No. 25313-002E*J
and the amount of the preflash time is fixed For this unit, preflash can be obtained by multiplying the amount of main heating by the preflash factor. the preflash level is adjusted by changing the preflash Therefore, factor. is generated by preflash factors at three The preflash factor curve The three points are the maximum, minimum and middle points. points of the filament current on the X-ray tube current characteristics curve. PRFl, PRF2 and PRF3 are to be selected The three preflash factors, with
the
[d,
adjustment Tube
Tube
and [z]
14,
Current
Characterist
switches. its
current
Pre flash factor
WI
Filament PRFl, PRF2, selected as
Maximum
Middle
Mi nimum
Current
and PRF3 for follows:
the
(A)
three
adjustment
points
can
be
Plot the X-ray conditions under which the preflash level is to be If the on the tube current characteristics curve. adjusted, If the filament current is near the maximum value, adjust PRFl. filament current is near the middle or minimum value, adjust PRF2 or PRF3, respectively. If the filament current is between the minimum and middle values, If the filament current is between the adjust PRF3 and PRF2. middle and maximum values, adjust PRF2 and PRFl.
Set
Set
Mode
5 with
conditions the
, via Select
the
m,
[a,
PRF3
PRFZ
sequence
the
mdoe
under operation
preflash or
proceeds
selector
which the preflash level is panel of the X-ray control factor
[a
switch.
to
switch.
be
adjusted
The mode
PRFl to
the
next
<
-
24H -
step.
by
indicator
short unit.
pressing blinks,
in
(5)
the and
the
No.
Change the (Note 1). L
preflach
factor
by using
S/R Note
or
@
switch
Increase Decrease the indicator In this state,
When the S/R switch is pressed, and the state in is reset. factor, another focus or another adjusted.
When the
[z]
the
25313-002E*J
adjustment
is
tube
preflash
reset
completed,
stops blinking another pref lash
Mode
factor
can
be
0 by pressing
the
switch. 1:
When the
[a
or
[q
Increase
switch
is
pressed,
the
buzzer
Decrease
will sound and the preflash factor will increase or If either switch continues to be decrease by 1%. pressed, the buzzer will automatically start sounding so that the result can be achieved as if the switch were pressed by the number of times the buzzer sounded. (6)
Preheating
level
adjustment
(PRH):
Mode
5
If the adjustment This adjustment is not necessary in principle. in (5) cannot be carried out satisfactorily depending on the this adjustment should be carried out as an auxiliary system, adjustment to (5). ---+
Set
Mode
5 with
the
Set the conditions not satisfactory, unit.
When the the Press
mode
switch.
under which the via the operation
switch
sequence
selector
is
proceeds
pressed,
to
the
switch
the
preflash panel
the
next to
mode
of
level the
adjustment is X-ray control
indicator
blinks
and
stop. adjust
the
preheating
level
Increase Decrease (Note
l),
th e
and check
When the S/R and the state another tube
1:
current
rise
by generating
the indicator switch is pressed, In this state, returns to . preflash factor can be adjusted.
When the adjustment S/R switch. Note
tube
is
When the[z]or
completed,
14
switch
reset
is
Mode
pressed,
X-rays.
stops blinking another focus
0 by pressing
the
buzzer
or
the
will
If either sound and the preheating level will change. switch continues to be pressed, the buzzer will can be automatically start sounding so that the result achieved as if the switch were pressed by the number of times the buzzer sounded.
-
241
-
25313-002E*J
No.
(7)
Changing
the
Set
Size
Mode 6 with
the
Use the operation the number of the be changed.
Press
the
mode,
enabling
To adjust switch
for
Fluorography:
Mode 6
to change the focal size procedures permit the small focus (SF) at the time of shipment.
The following adjusted to
Focal
the
the
For
the
the
Ed .
to
place
following
focal
size
pressing
small
selector
switch.
panel of the X-ray control tube whose fluorographic
switch
yh cl
while
mode
the
mode
unit to specify focal size needs
indicator
in
to
blinking
steps.
to
the
large
focus
(LF),
tilt
the
S/R
switch.
the
focus
already
(SF),
tilt
the
S/R
switch,
while
pressing
switch.
A buzzer procedure
sounds and ‘the indicator stops has been performed correctly
Reset to mode “0” by means of the size adjustment has been completed.
To change
the
focal
size
to
S/R
LF [Z]
switch
14 Hold
To change
the
focal
size
to
blinking so far.
for
tube
current
The dropout tube current can However, into the generator. level is too low or too high, procedures below.
drop:
Mode
Select
Enter
the
Press
the
tube
current
I/
switch.
I/a
-
mode which
selector is
The mode
245
focal
fl
G
3 switch
Ti;Tt
down
T
this
switch
Tilt
be compensated by entering a tube code if the preset compensation current it can be adjusted by following the
the
this
the
6
On the operating panel of the generator, for which the preset compensation current adjusted. 6 with
[a
the
SF
cl>
Mode
when
down
Hold Compensation
if
-
to
specify level
the tube No. needs to be
switch. be corrected.
indicator
starts
(See
Note
blinking.
1. >
No.
Change
the
(increase
preset
compensation
current
ecrease)
> and
level
switches.
When the current level is correct, the indicator stops blinking.
using
(See
set
the
S/R
25313-002E*J
the
Notes switch
yd cl
2 to
4.)
down;
that the adjusted current level for Generate X-rays to confirm (See Note 5.) the dropout tube current is correct. Repeat Procedure if the compensation current level requires or repeat from Procedure if it is further adjustment, the compensation current level for other necessary to correct tube currents.
(9)
Set
the
Note
1:
Note
2:
Note
3:
Note
4:
Note
5:
Recovery If
S/R
down
to
set
When either
the
y [Y]
Mode
0.
switch
Or [Zl
is
pressed
once,
the buzzer sounds once and the compensation current If the level is increased, or decreased, by one step. will automatically switch is held down, the buzzer sound at fixed intervals; the number of times the buzzer sounds is the same as pressing the switch the same number of times. When the compensation current level value reaches the the buzzer emits alternating upper or lower limit, short and long tones. time should be longer than 200ms. The X-ray exposure any
error
data
has
error recover
occurred,
When the S/R switch is tilted alarm stops and Mode 7 is set. switch recovery
to
Tube current values of 20, 50, 100, 200, 500 and 64OmA can be entered in Procedure . The tube voltage should be set to 90kV. Increase the compensation current level to correct for a fall in tube current during X-ray exposure and decrease it to correct for a rise in tube current.
from
a data
switch
routine
is
pressed,
with
while
the
the
the
following
alarm
indicator
procedure:
sounds,
the
blinks
and
the
starts. switches
are
pressed
simultaneously,
the indicator goes out and the recovery routine (for Upon regeneration of an error check code) is executed. the indicator lights up and the completion of execution, sounds.
-
24K -
buzzer
No.
25313-002E*J
Reset Mode 0 by tilting the S/R switch and check the data. If the power supply is turned off in the adjustment mode during data is judged to have be destroyed tube current adjustment, when the power supply is turned on again, and the alarm sounds. (This is because it is not possible to distinguish data change from data breakdown.) In such cases, recover by using the data adjusted prior to After recovery, above procedure. turning Note:
(10)
off
the
power
supply
are
all
effective.
The recovery routine is simply to regenerate the error but not to repair the destroyed data. check code, it is necessary to check the data after Therefore, recovery. If the check shows that the data are changed, correct it If data have been completely in Mode 1, 2 or 4. In this it is necessary to reset the tube. destroyed, the setting must start with the setting of the tube case, code.
Alarm
Stored data check The stored data is checked by heat data when the power supply Data is checked in each tube and each focus (for is turned on. If an Tube 1 LF, Tube 1 SF . . . Tube 3 SF). data blocksattempt is made to use the data block in which an error has an alarm (intermittent beeping sound) and an been detected, To recover from this state, error message will be generated. execute recovery procedure (9) from a data error.
Set condition check If the set tube voltage becomes greater than 150kV or the tube an alarm (intermittent beeping current is not set correctly, If the tube voltage or the tube sound) will be generated. the alarm will be removed. current is set correctly,
current check Maximum filament The heating output data is set to a value smaller than the maximum filament current to prevent the X-ray tube filament If the heating data exceeds the maximum from breaking down. filament current value when the READY switch is pressed, the heating level will be set to the preheating level and an alarm (intermittent beeping sound) and an error message will be generated.
-
24L -
No. 2J313-002E*P
If the alarm sounds, turn off the READY switch decrease the heating level. Table 5.2.2
Tube setting
. No.
Tube name
Binary code
No.
Binary code
Tube name
0
00 0000
32
10 0000
DRX-3535HC
1
00 0001
33
10 0001
DRX-3545HC
2
00 0010
DRX-1203B
34
10 0010
3
00 0011
DRX-1403B
35
10 0011
DRX-3603HD
4
00 0100
DRX-1415B
36
10 0100
DRX-3624HC
5
00 0101
37
10 0101
DRX-3634HC
6
00 0110
DRX-1603B
38
10 0110
1
00 0111
1 DRX-1615B
39
10 0111
’ DlU+3724HD/HC
8
00 1000
DRX-1725B
40
10 1000
DRX-3745HD/H
9
00 1001
DEU-1924HD
41
10 1001
10
00 1010
DRX-1724E
42
10 1010
11
00 1011
j
43
10 1011
/
12
00 1100
I DRX-2603HD
44
10 1100
( DRX-4234HC-S
1 DRX-2625C
45
10 1101
DRX-4634HC
7
1
13
1
00
1101
14
1
00
1110
15 16 17
/ 1
18
46
00 1111
1 DRX-2703HD
47
01 0000
1 DRX-2724HD
48
’
I
10 1110 10 1111
ISAPPHIRE 150TH
11 0000
IDXB-0335CS(A197)
11 0001
/DXB-0324CS(A192)
01 0001
/
49
01 0010
I
50
11 0010
DXB-0434ES(A256)
’ DFU-2825C
51
11 0011
DRX-B1124B-MO
52
11 0100
DRX-B1146B-MO
11 0101
DRX-B1356EC-MO
19
I
01 0011
20
’
01 0100
/
21
01 0101
DRX-2903HD
53
22
01 0110
DEU-2924HD
54
23
01 0111
24
01 1000
56
j
11 1000
25
I 01 1001
57
'
11 1001
DRX-8316HD
26
’
01 1010
58
11 1010
DRX-8336HD
/
55
(
11 0110 11 0111
27
01 1011
DRX-3103D
59
11 1011
28
01 1100
DRX-3224D
60
11 1100
29
01 1101
DRX-3224HD-H
61
11 1101
30
01 1110
62
11 1110
31
01 1111
63
11 1111
DRX-3335HC
1
- 25 -
No.
5.2.3
Radiographic (1) Prior Prior table
tube voltage
2J313-OOZE*1
adjustemnt
adjustment to X-ray 5.2.3.
exposure,
Table
Power
set SW1 on the INV-2
PWB
according
to
5.2.3
supply
Setting of SW1
position
Single-phase 200 V (190 to 215 V>
0
Single-phase
1
220 V
(216 to 240 V) Three-phase
380 V
4
(380 to 400 V) Three-phase
415 V
5
(401 to 425 V> Three-phase (426 to
440 V
6
1
Set SW5 on the INV2 PWB according to the table is required for PWBs without SW5.j
Length of the longest among those connected the three X-ray tubes
cable to
Setting of SW5
to 15m
r-
position
3, 2, 1 or 0
16 to 20m
2, 1 or 0
21 to 25m
1 or 0
26 to
0
- 26 -
below.
(No setting
No.
(2)
Tube
volt
age
25313-002E*I
adjustment
Perform tube voltage adjustments for each tube current setting. Although tube voltatges, 30 kV, 40 kV, 60 kV, 90 kV, 120 kV and 150 kV, are to be adjusted in principle, any voltage between tube voltage sections (see Note 2) can be adjusted. If the tube current value differs greatly from the value preset by means of the operation panel, the tube voltage cannot be adjusted. In such cases, therefore, the tube current adjustment should also be made along with the tube voltage adjustment. The delay time of the tube voltmeter should be 20 msec. [Adjustment within the INV-1 PWB using a storage oscilloscope for waveform comparison.
T
Tube voltage waveform
TP2
T Fine-adjust it CCW
VR3 by turning (towards "1").
i Gradual
increase
T Fine-adjust
(2) -we--_
I-
Gradual
decrease
it CW
-I
_L
(3)
F= Gradual
decrease
VR3 by turning (towards "3").
Gradual
-
decrease
28 -
If a similar waveform is observed, do not turn VR3. I In a single-phase 2OOV i system, the waveform may show a gradual decrease I under a large load. I
1
No. 25313-002Ek4L
5.2.4
Checking
the
Check that radiographic
radiographic
time correctly
the radiographic timer operates times as described below.
(1)
terminal Connect the monitor output input terminal of the oscilloscope
(2)
Turn
power
(3) Select
the
tube
to
which
kV, mA, and X-ray.
measuring
time
(6) The maximum radiation
Definition
power
and
the
defined time
as
as
specified
shown
tolerance
is
if
is
in
the
Fig. given
Item
connected.
7.2.2 (2), and
oscilloscope,
and
5.2.4. in
6.3.
off.
of
The radiation increases to decreases to
is
values
instrument
Read each radiation time from the waveform on check that it is within the allowable range. radiation
time
the
(5)
Turn
set
on.
Set the generate
(7)
several
of the tube voltmeter with a coaxial cable.
(4)
The
for
radiation time
is
75% of its 75% of its
time the interval peak value, peak value.
from Point A, where the tube voltage to Point B, where the tube volt age
/ 100%
75%
75% \
I
I
High potential waveform
Tolerance
Tolerance Fig.
5.2.4
-
29 -
No.
5.3
Checking
the
Automatic
The followings is installed.
5.4.1
check
should
(1)
Press
the
(2)
Check Refer
that the settings to the AEC-lOA/lOB
(3)
Perform manua 1,
@
Set
@
Set the If this timer.
@
Start
as
select
is
adjustment
control
is
(AEC)
performed.
the
AEC-1OA/lOB
installation
below.
kV and mA values. than an appropriate AEC radiation time. functions as an AEC backup or desk timer
generation.
(KXO-30F/50F)
instrument
connect
the
exposure
AEC.
in
given
Adjustment
Measur ing
X-ray
specified
time value longer the panel occurs,
X-ray
Fluoroscopic
to
automatic
of the AEC unit are correct. installation manual, 2C376-519E.
procedure
appropriate
when
after
TECHNIQUE button
the AEC test 26376-519E.
Fluoroscopy
Control
only
be performed
radiation
Connect 5.4.2
necessary
The
The
5.4
are
Exposure
2J313-002E;kB
measuring tube
ion
instrument
current
as
adjustment:
specified Mode
in
5.2.1.
4
can be adjusted individually point In Mode 4, each adjustment setting the f luoroscopic tube (F.kV) or all data concerning fluoroscopic tube current can be increased or decreased at
When F. kV is set to 60kV in Mode 4, all data can be increased or decreased at one time. Any F.mA setting can be made.
When F.kV is set to 50, 90, or 120kV in Mode 4, each adjustment point can be adjusted separately.
j
5OkV /
90kV
2mA
I
I I
I !
3rnA
I
I
I
50kV
/ 120kV
lmll
-
by the one time.
30 -
/
90kV
/ 120k
253 13-002E+B
No.
the
Therefore, adjusting
points
with
and
120kV
For
the
The
heating data after in APPENDIX 01.
11. (1)
tube
fluoroscopic
all
adjustment
F.kV
to
set
be
can
60kV
and
facilitated
then
by
adjusting
first
50,
90,
points. operation
Select
switches
(T u b e
SPOT
and
adjustment should
adjustment
1)
via
the
modes, refer to 5.2.2 (1). be recorded as shown in table
operation
panel
of
the
X-ray
control
unit. (2) (3)
Set
Mode
Set
the
4 with F.kV
and
part of the Combinations:
(4)
F.kV:
50,
90,
05,
1, the
switches
Note
3,
values
yA cl
by
order
generating
If
the
X-ray
correct
tube
Operation is (1)
the
is
completed,
or
or
tube
voltage
tube
current
below
VR6
[z]
tube
(2)
is
(VR6 VR6
within
7% of
If
tube
the
(1) above, radiographic possible the
off, In addition shown
,& Cl
under
operation of
on
ten
over,
this
Mode
is
F-unit.
pressing
the
as
compared
not
tube
adjustment
PWB)
the
INV-1
PWB so
all
value
be
can
to
possible
current
must
procedure
mA and
that
under in the
S/R
be
when
the
and
to
obtain
be
adjusted.
voltage
operation
the
measured
the conditions fluoroscopic
2,
(2)
kV values,
check
that
value of mode.
of
90kV
they
the
and
are
within 7% after adjustment PWB as in the case with the This adjustment, however, is
adjustment. When this mode fluoroscopic mode. tube voltage adjustment is possible. is performed for each voltage adjustment 1,
the
value.
voltage accuracy is not yet use SW2 and SW3 on the INV-2
0.5,
the
procedure.
INV-1
kV accurate
at
by
set current.
respectively.
may the
on
the
this
0
used,
times
used,
it
case
main
setting
tube
individual tube
only in the radiographic
below
spot part
(decrease)
fluoroscopic
switch
voltmeter becomes 90kV+lkV LED4 should indicate sT8”
1mA. By manually
and
reset
is
adjustment
adjustment
Fine-adjust
the
operation
adjustment
the
is
supplementary
the
decreased
In
voltage.
(1)
with
the
X-rays
ad just
switch
Fluoroscopic
adjusted or
(increase)
to
[z]
the
increased
5.4.3
be
table
4mA
adjustment
When
switch.
to
R/F
with 1)
in
1:
selector
120kV 2,
(Note
When the switch.
F.mA
data
conditions (5)
mode
conventional
F.mA: Correct
the
3 and
4mA respectively.
is
turned section
No.
Tube
voltage
As
for
section
50kV
operating
procedure
not to Note : Take care at the boundaries 5.5
to
63kV
of
64kV
SW2 and
to
SW3,
obtain the very at each section.
95kV refer
2J313-002E*B
96kV to
discontinuous
to
127kV
5.2.3. tube
voltage
HHS Tests After all of the preliminary and final tests to be necessary for the installation system have been successfully completed, the HHS test phase may be performed. The
5.5.1
HHS test
is
Fluoroscopic (1)
Refer
necessary entrance
to
Refer
to
Automatic Refer
to
and
Automatic Refer
to
the
ORH compliance.
required
and
rejection
limit
are
shown
required
and
rejection
limit
are
shown
linearity
APPENDIX 04
exposure
equipment
control
limit
APPENDIX 06
The test method, APPENDIX 06. 5.5.4
checking
rate
equipment
The test method, in APPENDIX 04. 5.5.3
finally
APPENDIX 03
Reproducibility (1)
when
exposure
The test method, in APPENDIX 03. 5.5.2
only
exposure
equipment
minimum
required
and
exposure
time
required
and
rejection
limit
are
shown
in
rejection
limit
are
shown
in
APPENDIX 07
The test method, APPENDIX 07.
equipment
-
32 -
’
No.
6.
SYSTEM
ACCURACY
I
I
Technique
No.
1
Applicable
factor
/ Maximum allowable deviation I
range
I
I
I
1
c
(Radiographic
TRadiographic 3
mA
/
all
I I
-+lO%
kV
1
all
~
27%
i
-+3ms
i
120%
24 to 99ms
I I
-+20%
0.1
to 9.999sec
I
-+lO%
0.2
to 2.9 mA
-+0.2mA
3.0
to 4.0 mA
-+lO%
/ Radiographic time Setting and indication in A.E.C.
1
1 Setting and indication in Ii I !ABC
6
7.
7.1
3 to llms for lOOmA
, I
for
'Fluoroscopic kV Setting and indication in ABC ) Fluoroscopic kV , time setting / and total time / indication
ATTACHED
lOOmA
or more or more
I
I
5
1
I
12 to 23ms
4 -/ Fluoroscopic
c
2J313-002E*B
all
I
I
I
-+7% I / I I
I I
+o _% -3
1 /
I
I
DATA
PWB Setting
and Adjustment
Tables
PWBs which are not listed in 7.1 require no setting and adjustment For the automatic exposure control PWB, refer they are installed. AEC-lOA/lOB installation manual.
- 33 -
when to the
No. 2J313-002E":D
(Panel) [PANEL CONT l] or TW CONT] l/8
PWB name:
A:
KXO-30R
B:
KXO-3OF/5OF
Equipment
When
Setting and adjustment
installed
B 0
VRl:
Adjusting the backup battery operating point There is not a VRl for KXO-30R=B.
Unnecessary
0
SWl:
Setting the serial transmission baud rate
Unnecessary
Switch settings
I 2
1
ON
Baud rate
OFF
OFF
ON
1200
ON
/ OFF 1 ON
2400
OFF
/ OFF 1 OFF
4800
OFF
9600
)
OFF
ON
ON
[W CONT SUB] l/7 or [W CONT 2/8
PWB name
i
4
OFF ON ,
3
Equipment B
/ I i-
Setting the X-ray generation buzzer timing
Jumper Jl: l
Open:
l
Short:
The buzzer goes on during radiation. goes on for approximately after the radiation is completed.
The buzzer
Necessary
50msec
t
L ON OFF
Necessary
Setting the system
sw2:
EP 1 2
'&El 3 4
5
Q
PU 6 7
8
A
Turn this switch ON only for USA. NOTE (Set this switch to OFF) MAMMO TOM0 GENE BKY 2 BKY 1
L
1
Turn switches used for the individual TECHNIQUES ON. (Set the other switches to OFF.)
Unused (Set this switch to OFF.)
NOTE _ This switch can be used for 100% of OLP testing only in the USA. ON: 100% Rating OFF: 90% Rating 100% Rating can be used for only testing; this switch shouid be
1
i
turned
OFF after
completing
the
- 34 -
100% Rating
test.
No. 2J313-002E9cD
PWB name:
[W CONT SUB] TW CONTl 4/8
3/7 or
A:
KXO-30R
B:
Equipment
When installed
Setting A 0
B j sw4:
KXO-3OF/5OF
For BKY 2 TECHNICQUiZ
Necessary
I 0
sw5:
For GENE
TECHNIQUE
Necessary
0
SW6:
For TOM0
TECHNIQUE
Necessary
For MAMMO Contents settings
Necessary
TECHNIQUE
of these of SW3.
switches
are
- 36 -
the same as the
No. 2J313-002E>kD
PWB name:
[W CONT SUB] 4/7 or [W CONT] S/8
KXO-30R
A:
B:
KXO-30F/SOF
Equipment
When installed
Setting
Sw8:
Necessary
Setting Tube 1
ONP1
OFF
U
P
Cl 51 Q
Q
P
-2
3
4
7
8
5
6
LL
X-ray tube type (12 types max.> Refer to next page. ISetting
the upper limit of the X-ray tube voltage. These switches can set tube voltages 15kV (maximum) lower than the maximum X-ray tube voltage.
I
l-i-4 SW No.
Upper limit
4
3
c
1 OFF
1 OFF
1 Max. -0kV
1OFF
/ ON
1 Max. -5kV
ON
OFF
Max. -1OkV
ON
ON
Max. -15kV
Setting the starter mode
SW No.
--r---l 1
1 ON
I
Mode
2
1 ON 1 Automatic change-mode
1
1 OFF 1 ON 1 High speed only mode
I
] OFF 1 OFF1 Normal speed only mode
1
I - 37 -
No. 2J313-002E"N
Sw8 (for USA and CANADA) SW No. 5
SW No. 6
SW No. 7
SW No. 8
CODE
TUBE TYPE
ON
OFF
ON
ON
B
DXB-0434ES(A256>(USA>
ON
ON
OFF
ON
D
DRX-3724HD/HC(USA)
ON
ON
ON
OFF
E
DRX-3624HC(USA)
ON
ON
ON
ON
F
DXB-0324CS(A192)(USA)
SW8 (for other count ies except USA and CANADA) SW No. 7
SW No. 8
CODE
TUBE TYPE
OFF
OFF
0
DXB-0324CS(Al92)
ON
OFF
2
DRX-3224HD-H
OFF
OFF
ON
ON
3
DRX-3224HD
OFF
ON
OFF
OFF
4
DRX-2924HD
OFF
ON
OFF
ON
5
DRX-3724HD
OFF
ON
ON
OFF
6
DRX-2703HD/HC
ON
OFF
OFF
OFF
8
DRX-1924HD
ON
OFF
OFF
ON
9
DRX-1603B
ON
OFF
ON
ON
B
DXB-0434EScA256)
ON
ON
OFF
OFF
C
DRX-2724HD
F
DRX-1724B
1
ON
ON
PWB name:
ON
ON
A:
[W CONT SUB] 5/7 or [W CONT] 6/8
Equipment Setting A 0
B sw9:
Setting Tube 2
- 38 -
KXO-30R
B:
KXO-30F/50F
When installed
No. 2J313-002E"D
[W CONT SUB] 6/7 or [W CONT] 7/8
PWB name:
A:
KXO-30R
B:
KXO-30F/50F
Equipment Setting A
B SWlO:
0
Necessary
Setting the display time of actual R-time when AEC is ON
-This value is set before shipment.
/ sw11:
0 1
Setting the UP/DOiJN key speed
1
I I
i
-~ -
39 -
Necessary
No. 2J313-002E*K
PWB name:
[W CONT SUB] 7/7 or [W CONT] 8/8
A:
KXO-30R
B:
KXO-3OF/5OF
Equipment
When installed
Setting A 0
B sw12:
Limiting the output tions of the user's
SW setting
depending on the condipower supply
Necessary
Contents
F . . . . . . . No limit to power up to 40kW
supply
0 . . . . . . . No limit to power up to 30kW
supply
1 , . . . . . . OLP at
for 51 msec or more 25 to 30kW
2 .,...*. OLP for 51 msec or more at 20 to 30kW
0
sw13: Setting
the exposure-time offset value The offset value added to the preset exposure time is the actual exposure time. Added time (msec> = O.Lt(Msec) x (setting
SW value)
The offset value can be set in 0.4 msec units in the range of 0 to 6.0 msec.
- 40 -
Necessary
No.
2J313-002E'kB
(Stand) [DESK CONTROL]
PWB name:
KXO-30R
A:
l/7
B:
KXO-3OF/5OF
Equipment
When installed
Setting A j
B
SW1
thru SW3:
SWl: sw2: sw3:
Baud
rate
for serial
For communication with the control For output to the printer For option RS-232C
Switch
box
settings Baud
1
rate
4
OFF
OFF
ON
ON
1200
Unnecessary
Jumper
0
Connect A-B, DSRA
0
Unnecessary
transmission
)
sw4:
-
jumers
as follows:
RTSA -RTSB, DSRB
Printer
CTSA -CTSB,
DTRA-
DTRB,
Necessary
connection
I
(OFF before
- 41 -
shipment)
No.
2J313-002E*B
(Stand) PWB name:
[DESK CONTROL]
2/7
A:
JLXO-30R
B:
KXO-30F/50F
Equipment
When installed
Setting A
B 0
sw5:
Setting
Setting
SW6:
user
Unnecessary
Destination
country
0
Japan
1
USA
2
Countries other Japan and USA
3 z F
0
for
Setting
than
Undefined
the exposure-time
offset
value
The offset value added to the preset time is the actual exposure time. Added time (msec> = O.Lt(msec) x (setting
exposure
SW value)
The offset vcllue can be set in 0.4 msec units in the range of 0 to 6.0 msec.
- 42 -
Necessary
No. 2J313-002E
PWB name:
[DESK CONTROL]
3/7
A:
KXO-30R
B:
KXO-30F/50F
Equipment
When installed'
Setting A
1 B 1 0 I
sW6:
Necessary
Setting the R-time offset
1 1 The switch sets the offset value of the set time.
(vs actual time)
I I ,
I I I I
o
SW7 thru SW12:
Setting the minimum time of mAs for each TECHNIQUE
I ’ Necessary
I /These switches set minimum time of mAs for each TECHNIQUE. , SW7 : i SW8 :
/
SPOT TECHNIQUE BKYl TECHNIQUE BKY2 TECHNIQUE GENE TECHNIQUE TOM0 TECHNIQUE MAMMO TECHNIQUE
SW9 : 1 SWlO: SWll: sw12:
Switch settings 112
3
OFF / OFF
4
OFF / OFF
5 OFF
6
7
Minimum time (msec>
81
OFF ’ OFF
OFF 1
0
OFF i OFF / OFF 1 OFF , OFF, OFF
OFF
ON
(
1
OFF / OFF 1 OFF
ON
OFF (
2
OFF
OFF
OFF
i
I 1
:
1 /ON 1 / ON
ON
1 ON
ON
1 ON
1 ON I / ON I 1 ON
ON
ON
/ ON I / ON
ON
ON
ON
/ I
OFF
ON
/
253
/ ON
ON
OFF j
254
ON
ON
ON
255
- 43 -
No. 2J313-002E*K
PWB name:
A:
[DESK CONTROL] 4/7
KXO-30R
B:
KXO-30F/SOF
Equipment
When installed
Setting A
B 0
sw13:
0
sw14:
01
02
0
Unnecessary
Output kW limit Set the limits if necessary customer's power conditions KXO-30F SW No.
Necessary due
to the
Limitation
5
6
7
8
ON
ON
ON
ON
No limit
up to 40kW
OFF
OFF
OFF
OFF
No limit
up to 30kW
OFF
OFF
OFF
ON
OLP
for Slmsec
or more
at 25kW
to 30kW
OFF
OFF
ON
OFF
OLP
for 5lmsec
or more
at 20kW
to .30kW
KXO-50F SW NO. Limitation
5
6
7
8
ON
ON
ON
ON
No
limit
up to 60kW
OFF
OFF
OFF
OFF
No
limit
up to 50kW
OFF
OFF
OFF
ON
OLP
for Slmsec
or more
at 45kW
to 50kW
OFF
OFF
ON
OFF
OLP
for Slmsec
or more
at 40kW
to 50kW
OFF
OFF
ON
ON
OLP
for Slmsec
or more
at 35kW
to 50kW
OFF
ON
OFF
OFF
OLP
for 5lmsec
or more
at 30kW to 50kW
sw15:
ON OFF
Selectable
APR program
Necessary
setting
00000000 12345678 I
Not us!d
A
’
”
‘l
1
Not used MAMMO TOMOGRAPHY
BAS1C
- 44 -
Table Bucky, Wall Bucky {General, Spot-shot I
NO.
Equipment
2J313-002E*C
When installed
Setting B 0
sw16
to 21:
max. mA code for mAs analysis when 1.1. is magnified
sw16: sw17:
upper 4 bits lower 4 bits
max. mA code for Normal size (X.1.)
SW18: sw19:
upper lower
4 bits 4 bits
max. mA code for Magnification-l (1.1.)
sw20: sw21:
upper 4 bits lower 4 bits
max. mA code for Magnification-2 (1.1.)
mA code
I
(Hex.)
mA (mA) 10
OOH OlH 02H 03H
13 16 20
OFH
! 320
For single size I.T., SW16 and SW17.
set the max.
- 44A -
mA code with
Necessary
No.
PWB name:
A:
[DESK CONTROL] 5/7
KXO-30R
B:
23313-002EAG
KXO-30F/50F
Equipment
When installed
Setting A
B 0
SW28:
Selecting TECHNIQUE to be operated when the APR is installed. (NOTE) Only one of 1 to 7 on this SW can be selected when SW30 or SW31 is used. ON OFF
1
PO 1 2
0 3
0 4
P 5
RRP 6 7
a
’
1
’
A
,
AL
1 I
/
//
1 I__
I
I
Unused
Necessary
’
All TECHNIQUES are operated. MAMMO TECHNIQUE TOMOGRAPHY TECHNIQUE SPOT TECHNIQUE
j
I
BKY2 TECHNIQUE BKYl TECHNIQUE GEN/CASSETTE TECHNIQUE
0
SW29: ON : OFF:
0
SW30 through SW31 are valid. SW30 through SW31 do not function. (The switch is set to OFF before shipment.)
Jumper Jl: Short:
0
sw30:
Necessary
Approving setting
Selecting KATAKANA
indication
Unnecessary
APR data are indicated in KATAKANA.
Copying APR data from ROM to RAM.
If this switch is turned ON, data equivalent to TECHNIQUE selected by SW28 is copied into the RAM area. (NOTE) Only one of 1 to 7 on SW28 can be used.
- 45 -
Necessary
No.
PWB name
[DESK CONTROL]
r
6/7
A:
KXO-30R
B:
KXO-30F/50F
Equipment
When installed
Setting A
B 0
Setting required only when combinations of DT-GCU/KDU .
is used with
Setting of positions/organs corresponding AU-APR operation selector switches Switches
to be used:
6, 7 and 8 combinations of SW13 1 . .. . . . . . .. 2 .......... 3 .......... 4 .......... 5 .......... 6 .......... 7 .......... 8 of SW28 0 .......... 1 .......... SW32 ........... SW17 combinations 1 .......... 2 .......... 3 .......... 4 .......... 5 .......... 6 .......... 7 .......... SW7 combinations 1 to 10, 15 or 20 . 0 to 12 . . . . . . . . . . .
.
AU-APR
2J313-002E*B
6, 7 and 8 of SW13, 8 of SW28, SW32, SW17
When
to the
and SW7
Contents for SW1 of AU-APR unit for SW2 of AU-APR unit for SW3 of AU-APR unit for SW4 of AU-APR unit for SW5 of AU-APR unit for SW6 of AU-APR unit the maximum in 1.1. fluoroscopy mode Contents AU-APR spot radiography mode AU-APR fluoroscopy mode Data writing Contents Selecting the head Selecting the breast Selecting the abdomen Selecting the spinal cord Selecting the lumbar vertebrae Selecting the arms Selecting the legs Contents Selecting an organ from those set with SW17 Determining the maximum mA code for 1.1. fluoroscopy Setting Setting Setting Setting Setting Setting Setting
Setting procedure a First set the numeric on SW13 corresponding to the switch number on the AU-APR unit to be set. SW17 and SW7 to the (@ Set SW28 (spot radiography/fluoroscopy), desired position/organ numbers. 3 Press the SW32 pushbutton switch (for write operation). If the buzzer the setting is completed. 4 If the buzzer sounds, 8 a SW setting error has occurred, and go back does not sound, to @ . @ Terminate the setting operation by setting 6, 7 and 8 of SW13 to "0". @ After the setting operation, reset SW13, SW28, SW17 and SW7 tc their original position.
- 46 -
No. 25313-OOZE"0
PWB name:
[DESK coNTRoLl
A:
7/7
KXO-30R
B:
KXO-30F/50F
Equipment
When installed
Setting A
B 0
SW32:
0
SW33:
Unnecessary
Setting the combined conventional R/F table
This switch sets the types of conventional R/F table.
Switch settings 12
3
4
5
6
7
8
Conventional R/F table
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
No combination
OFF
OFF
OFF
OFF
OFF
OFF
OFF
ON
DT-GCU/KDU/DUA
OFF
OFF
OFF
OFF
OFF
OFF
ON
OFF
DT-GCS
1 OFF
OFF
OFF
OFF
OFF
OFF ION 8
ON
DT-KEL
OFF
OFF
OFF
OFF
OFF
ON
OFF
OFF
DT-KEL-2
OFF 1 OFF
OFF
OFF
OFF
ON
ON
DT-MCS-2
OFF j OFF
OFF
OFF
OFF
ON
OFF (ON
OFF
DT-MCI-3 AUX.
i ON
ON
ON
ON
\
ON
ON
-
47 -
ON
ON
AUX.
Necessary
No.
PWB name:
[DESK DISPLAY] l/6
A:
KXO-30R
B:
2J313-002EfcB
KXO-30F/50F
Equipment
When installed
Setting A
B 0
Jumper Jl: Short: Open :
0
SWl:
Q
P
2
3
P 4
Necessary
Necessary
System setting
ONEl1
OFF
Setting the X-ray generation buzzer timing The buzzer goes on for approximately 50msec after radiography is completed. The buzzer goes on during radiography.
u u 5
6
El P 7
8
ATR MAMMO
TOM0 GENE BKY2 BKYl SPOT
1 Turn l
switches used for the individual TECHNIQUES ON, and set the other switches to OFF.
NOTE This switch can be used for 100% of OLP testing only in the USA. ON: 100% Rating OFF: 90% Rating 100% Rating can be used for only testing; this switch should be turned OFF after completing the 100% Rating test.
-
48 -
No.
PWB name:
[DESK DISPLAY]
A:
2/6
KXO-30R
B:
KXO-30F/50F
Equipment
When installed
Setting
SWZ: Setting
SPOT RADIOGRAPHY
Necessary
TECHNIQUE
Setting the maximum mA value small focus) of mAs 1
Setting the maximum mA value large focus) of mAS
t Steps
(for the 1 (for the) / I
0 to 3 Contents
Set vlaue 0
2J313-002EfcB
. . . . . mAs
decomposition by the maximum mA which is determined by the tube/unit
1 . . . . . mAs decomposition by mA at a step below the maximum mA which is determined by the tube/unit 2 . . . . . mAs decomposition below the maximum by the tube/unit
by m.A at two steps mA which is determined
3 . . . . . mAs decomposition below the maximum by the tube/unit
by mA at three steps mA which is determined
Setting ON: OFF:
t
the AEC detector SOLIDET detector Image intensifier or thin detector
Turn this switch ON when AEC is usable ON: OFF: AEC is unusable
-
49 -
photo
using
pickup
AEC.
No. 2J313-002E
PWB name:
[DESK DISPLAY]
3/6
Equipment , B A /
/
A:
KXO-30R
B:
KXO-30F/50F
When installed
Setting
0
sw3:
For BKY
1 TECHNIQUE
Necessary
0
sw4:
For BKY
2 TECHNIQUE
Necessary
sw5:
For GENE
TECHNIQUE
Necessary
0
SW6:
For TOM0
TECHNIQUE
Necessary
0
sw7:
For MAMMO
Contents settings
Necessary
TECHNIQUE
of these of SWZ.
switches
are
the same as the
iI
- 50 -
No. 23313-002E
[DESK DISPLAY] 4/6
PWB name:
A:
KXO-30R
B:
KXO-3OF/5OF
Equipment A
When installed
Setting
B
L
0
;
sw8:
Necessary
Setting Tube 1
ON a
E
u u a u
1
2
3
J
I
OFF
4
5
0 E
6
7
8
+
X-ray tube type (16 types max.> Refer to next page. 1
Setting the upper limit of the X-ray tube voltage. These switches can set tube voltages 15kV (maximum) lower than the maximum X-ray tube voltage.
Upper limit /
/SW.
Setting the starter mode
I SW 1
No.
I
I
I
Mode ,2
I
I~_ON ; ON OFF 1 ON
Automatic change-mode
I
! High speed only mode I
OFF j OFF / Normal speed only mode
-
51 -
.
No. 2J313-002E*P
TUBE 1 TUBE 2 SWlO: TUBE 3 (for USA and CANADA) Sw8: sw9:
SW No. 5
SW No. 6
SW No. 7
SW No. 8
CODE
TUBE TYPE
OFF
OFF
OFF
ON
1
DXB-0434ES(A256>(USA>
ON
OFF
OFF
OFF
8
DRX-3535HC/HD
ON
ON
OFF
OFF
C
DRX-4634HC(USA)
ON
ON
OFF
ON
D
DRX-3724HD/HC(USA)
ON
ON
ON
OFF
E
DRX-3624HC(USA)
ON
ON
. ON
ON
F
DXB-0324CS(A192>(USA>
SW8: TUBE 1 sw9: TUBE 2 SWlO: TUBE 3 (for other countries except USA and CANADA) TUBE TYPE
SW No. 5
SW No. 6
SW No. 7
SW No. 8
CODE
OFF
OFF
OFF
OFF
0
DRX-1924HD
OFF
OFF
OFF
ON
1
DXB-0434EScA256)
OFF
OFF
ON
OFF
2
DRX-2724HD
OFF
OFF
ON
ON
3
DRX-1615B
OFF
ON
OFF
OFF
4
DRX-1603B
OFF
ON
OFF
ON
5
DRX-B1356EC-MO
OFF
ON
ON
OFF
6
DXB-0324CS(A192)
OFF
ON
ON
ON
7
DRX-3745HD-H
ON
OFF
OFF
OFF
8
DRX-3535HD
ON
OFF
OFF
ON
9
DRX-3224HD-H
ON
OFF
ON
OFF
A
DRX-3224HD
ON
OFF
ON
ON
B
DRX-2924HD
ON
ON
OFF
OFF
C
DRX-3724HD
ON
ON
OFF
ON
D
DRX-2703HD/HC
ON
ON
ON
OFF
E
DRX-ll24B-MO
ON
ON
ON
ON
F
DRX-1724B i
1
- 52 -
No. 2J313-002E*F
PWB name:
[DESK DISPLAY I 516
A:
KXO-30R
B:
Equipment
when
Setting A
KXO-30F/%)F
installed
B 0
sw9:
Setting
Tube
2
Necessary
0
SW10: Setting
Tube
3
Necessary
Contents settings
of of
these SW~.
switches
are
the
same
as
the
* -
52-1
-
No. 25313-002E
PWB name:
A:
[DESK DISPLAY] 6/6
KXO-30R
B:
KXO-30F/50F When
installed Necessary
Setting the display time of a ctual R-time when AEC is ON.
SW11 .
/ Setting [ Display time (
I
0
!
1 2 3 4
!
0 1 2 3 4
1 I
-Th
is va lue is set fore shipment.
I
F -
I
15sec.
1
0
I
-I-
I
Setting the UP/DOWN key speed
- 53 -
Necessary
No.
2J313-002E:cA
(Control box)
[CPU1
PWB name:
A:
KXO-30R
B:
Equipment
When installed
Setting A
KXO_30F/50F
B, 0
1 Xl
:
1
Short
I Unnecessary
I
I
X26:
X26-
Al3
X27:
X27-
+5
/ so -
RAM-S6
s2
1 I
Selecting Ml ROM 27128
: Short
Ml address:
0000 'L 3FFF
.- Short
RAM address: 6000 ?J 7FFF
[FH/INV]
PWB name: Equipment'
Setting q---J
0
/
0
/ JP-C I
I
’
i
~
Open : Short:
Clock division by three Clock division by two
(Set before shipment "short")
’
- 54 -
When installed Unnecessary
No. 2J313-002EkD
PWB name:
A: KXO-30R
1
[ SIO-2
B:
KXO-3OF/5OF
There is no SIO-2 PWB in the KXO-3OR=B, KXO-30F=B or KXO-50F=B. Equipment
When installed
Setting
A 0
0
Jumper:
Specifying the RX and TX buffer addresses
All short
All short
LRZ-D LRl-C LTlw---+B LT2-A
HRZ-F HRl-E HTZ-B HTl-A
Unnecessary
(Set before shipment "short") PWB name:
[SIO-11 or [SPIO]
Equipment
When
Setting
installed
Aj 0
0
There is no Jumper in the KXO-30R=B, KXO-30F=B or KXO-SOF=B.
Jumper:
Unnecessary
J: Short T-G: Short SWl:
11
PWB name:
Unnecessary
Baud rate generator
2
3
4
Baud rate
Equipment
ON j OFF
OFF
OFF
4800
30R
ON / OFF
ON
OFF
9600
30F/50F
IPIN
Setting
L
- 55 -
When installed
No. 25313-OOZE
PWB name:
A:
[IFI
KXO-30R
B:
KXO-30F/50F
Equipment
When installed
Setting and adjustment A 0
B
/ Necessary
’ Jumpers 52 through 54
IO
, Connect one jumper of 52, 53, 54 corresponding ; the minimum HU depending on combined tubes.
52: 53: c 54:
: 0
'
0
;Jl
PWB name:
Unnecessar}
. .
Short (only for USA.) Unnecessary
Adjusting VRl through VR4
i
0
140 KHU 200 KHU 300 KHU
Jumper Jl
!
0
to
[APR IF]
Equipment
When installed
Setting A I
B
’
0
I Jumpers Jl, 52:
These switches set BKYl or BKY2 when using Bucky for 1 TOM0 TECHNIQUE (Jl and 52 should be set 1 simultaneously.) l
I I
Necessary
Setting TOMO, BUCKY selection
When using TOM0 Bucky at BKYl TECHNIQUE, connect C and Bl.
’ . When using TOM0 Bucky at BKY2 TECHNIQUE, connect C and B2. !
- 56 -
,
No. 2J313-002E"cL
PWB name:
A:
MQI
KXO-30R
B:
KXO-30F/5OF
Equipment
When installed
Setting and adjustment A
B 0
Jumper Jl:
0
SWl:
Necessary
Short for F. Rotation
Necessary
Setting continuous radiography TECHNIQUE
Unused
SPOT BKYl BKY2 GENE TOM0 MAMMO 105mm spot camera
When selecting the continuous radiography mode for TECHNIQUES, turn the switches corresponding to the individual TECHNIQUES ON. , (The switches of three TECHNIQUES of SPOT, TOMO, and 105mm Spot Camera are set to ON before shipment.) 0
0
VRl:
Adjusting the set-up time of the normal speed starter
Unnecessary
This switch adjusts the set-up time of the normal speed starter. VRl is set to 0.8sec before shipment.
PWB name:
[RY
SEQl
Equipment Setting A
B
0
0
Jumper:
Setting the type of stator coil of X-ray tube TUBE 1 TUBE 2
o------ ---Q TUBE 3
If a combined tube is the XS-AA coil type or the Eimac tube, short the pins of the tube.
- 57 -
When installed Necessary
No. 2J313-OOZE*M
PWB name:
A:
[F DIG] l/3
KXO-30R
B:
KXO-30F/50F
Equipment Setting and adjustment A
B 0
SW1 thru SW5:
Setting the dose limit in F kV Auto mode
These switches should be set for USA and other countries where the dose limit is imposed. An example for USA (limited to 2.58 x lo-IC/(kg*min) {lOR/min) is given below. These switches set F.kV values which exceed the limit of 2.58 x 10_3C/(kg*min) {lOR/min) at F.mA values of I 0.5mA, l.OmA, 2.0mA, 3.OmA, and 4.0mA.
F*kV SW1 =A SW2 =B SW3 = c SW4 = D SW5 = E
F mA When setting the switches. measure F.kV values exceeding the limti of 2.58 x 10-3C/(kg=min) (lOR/min} at individual mA values (binary), and set them in the individual switches.
- 58 -
When installed Necessary
No. 2J313-002E*M
PWB name:
A:
[F DIG] 2/3
KXO-30R
B:
KXO-30F/50F
Equipment Setting and adjustment A
B
When installed
125 110 100 0 85 80
The above figure gives an example in which SW1 through SW5 are set to 125kV, IlOkV, lOOkV, 85kV, and 80kV respectively. 0
/
,946, sw7:
Setting the dose limit in Fully Auto mode
/ ~ These switches should be set for USA and other 1 countries where the dose limit is imposed.
/( SW7 requires no setting.) ’
~ In Fully Auto mode , the F.kV and F.mA values are ~ linked, and the F.kV value should be set for dose ( limit control. 1 When setting SW6, measure the F.kV value in Fully 1 Auto mode which exceeds the limit of 2.58 x 10-X /C/(kg*min) (lOR/min), and set the value (binary) I in SW6.
-
59 -
Necessary
No.
PWB name:
[F DIG]
KXO-30R
A:
3/3
B:
KXO-30F/50F
Equipment
When installed
Setting B
A
0
I
’/
SW8,
; i j
When SW8 is set to "l", a mode is set which enables two F.ABC parameters to be set with the DIP switches (SW11 and SW12 on the ATR.DIG PWB). .
i
When SW8 is set to any other mined parameter is set.
SW9:
Setting the F.ABC full-auto mode
parameter
value,
1 When SW9 is set to "O", F.kV/FmA 50kV/O.ZmA to 125kV/4mA. I 1 When SW9 is set to "l", F.kV/F.mA ; 50kV/0.2mA to 125kV/2mA.
I
ZJ313-002E$cB
mode
and the
Necessary
the predeter-
changes
changes
from
from
1 I / I
SWlO:
O ) I (
F-ON
test
I 1 Unnecessary
switch
The fluoroscopy can be done by this Test switch instead of the F-ON signal from the conventional R/F table.
- 60 -
No.
PWB name: Equipment
A:
[F ANA]
T
Setting
KXO-30R
B:
JXXO-30F/50F
When installed
and adjustment
B 0
VRl:
Setting
manual
F.kV
2J313-002E*L
Unnecessary
control
Connect 5kS2 variable resistors (same as the F. ABC control of R/F table) to FASTON terminals Tl, T2, and T3. Adjust VRl to set the T2 voltage to +lO.OV when turning the variable resistor fully clockwise (equivalent to F.kV=125kV selected).
SWl, VR3,
VR4,
Adjustment
VR5,
VR6,
Unnecessary
VR7
procedure
1
Set SW1
to "1".
2
The set mode is used to adjust the D/A converters (1C and 2C). In this mode, adjust VR6 and VR7 so that the voltage at Pll and P12 is set to a value in the range of 9.961 to lO.OOOV.
3
Next,
4
Adjust
VR4
to set the P8 voltage
to - lO.OOV.
5
Adjust
VR5
to set the P9 voltage
to +9.961V.
6
Adjust VR3 to set the Pll voltage in the range of 9.961 to lO.OOOV.
7
After "0" .
set SW1 to "2".
to a value
the adjustment is completed, reset SW1 to (If not, the software will not operate.)
Jl: Selecting the jumper for ABC mode (Video signal type or photomultiplier . Video signal type : Jump "V" side . Photomultiplier type: Jump "P" side 0
VR8:
Necessary type) connector connector
ABC gain adjustment VR for video signal type Adjust the ABC gain according to the system requirements.
I
I - 61 -
ABC
Necessary for video signal type ABC
No. 25313-002E
PWB name:
[ATR DIG] 1/3
A:
KXO-30R
B:
KXO-30F/SOF
Equipment
When installed
Setting A
B
1
0
5, 9:
SWl,
Setting AEC backup mAs values
’ Necessary
For SPOT-SHOT TECHNIQUE For photofluorography TECHNIQUE For special TECHNIQUE
SWl: sw5: sw9: +EC
is automatically selected for ATR, and backup mAs 'values can be set by these switches. The switches are 'for the TECHNIQUES given above.
1
I
sw1,5,9
ZFE3 12 A
, , I
(
PEI
00 3
5
‘: I
4 4 i I
25
24
23
h
27
26
LIP 6 ,I
”
P 7 ih
;
I
~
!
22
21
1
8 A i
20
(
I
[Example]
;
: ds
,;;-;-1
*2 I
2510
0
30/o
0,o
35
1
0
0
I ’
1
1
1
111
!011/0
’
40 j O'O/ 1 0 ~45/O'Oll;O j 50 10 0~1~1~0~0
j
i
Ii 1
i
0
o/0(1
I
12i
I l/O!0
2.
l:‘“f~al
0)1'
19H
1'0
1EH
1: ON
l/
23H
0: OFF,
o/
28H 2DH
I I I !
/ I
11110 1
1 1'01
1
Max* )
1'1'1 /
1
I lil
1
- 62 -
1
/ ;
I
32H
i iI255
~
/
1
FFH
I
No. 25313-002E
PWB name:
A:
[ATR DIG] z/3
JLXO-30R
B:
KXO-30F/5OF
1
E;i.,,;t
When installed
Setting
0
I
/
SWZ, 6, 10:
’
SW2 : SW6 : SWlO:
Setting the upper limit of R.kV
Necessary
For SPOT-SHOT TECHNIQUE For photofluorography TECHNIQUE For special TECHNIQUE 1
These switches set the upper limit of the R.kV value when converting F.kV to R.kV. , I I
1 I
SW2,6,10
(Jp$ ““EIPPUUiiP
:
a,
12 /
h
h
m
A
A
Ii
4 ‘1
27
26
25
24
23
22
21
20
34567
[Example] Ic
Hexadecimal
Binary kV .I
I ~
2’
26
125
zL
! 2”
~
70;
0~1~0~0~0~1~1~0~
i
801
0
j
9oi
0
l/O 1
/ 100 , 0 / l/
12’
1
12” ~
0i
11i~oii~o~
(H)
46H
1
1101 0
I-9
0101
50H
I
, 1 / / 0 1 0 / 11 0 / 01 1
~AH 64H 6EH
~110~0~1/1,0'1/1~1~0' I 5
I
:ii
IMax.
1
o’.
1
5
oil~l’o~ I
150
96H ,
I O !
jsw3, 7, 11:
I I /
;“w:
..:
SWll:
Necessary
Setting the lower limit of R.kV
For SPOT-SHOT TECHNIQUE For photofluorography TECHNIQUE For special TECHNIQUE
I
'These switches set the lower limit of the R.kV value when converting F.kV to R.kV. (The setting procedure /is the same as the upper limit setting described iabove.) I I
I
,
- 63 -
I
No. 2J313-002EfcB
PWB name:
[ATR DIG] 3/3
-r---j
A:
JLXO-30R
B:
KXO-30F/50F
Equipment A
B
’
0
I
1
i I
When installed
Setting
Necessary
SW4, 8: Setting the maximum mA value of mAs SW4 : SW8:
For spot film TECHNIQUE For photofluorography TECHNIQUE
j In ATR mode, the equipment separates the backup mAs I value which is set in SWl, 5, 9 using the mA value
l
/which is set by these switches. When performing small I focus radiography in ATR mode, set this value to the 1 maximum mA value of small focus.
J,- 640mA is not applicable
0
to KXO-30F.
SW11, SW12: Setting the F.ABC parameters (effective only when SW8 and SW9 on the F.DIG PWB are set to "1" and "0" respectively) Parameter P (for control sensitivity) / SWll: Parameter I (for reference of output j sw12: amount) I
1
- 64 -
( When / necessary
No. 25313-OOZE
PWB name:
A:
[ATR ANA] 1/2
KXO-30R
KXO-30F/50F
B:
Equipment
When installed
Setting and adjustment A
B
Necessary P5 input voltage ranges from +3.33V (50kV) to +8.33V (125kV) and P6 input voltage ranges from +0.5V (0.2mA) to +lOV (4.OmA). The final output voltage (PIO) is +lOV (150kV) max. VRl, VR2, and VR3 should not be set because they are unused. VR31 is used for radiography at a constant R.kV value with the ATR function stopped temporarily in SPOT (Example: Esophagus radiography) 1 Radiography mode. Adjust the P9 voltage with 26B grounded.
0
IVR4 through VR7 adjustment (for spot film radiography) / I
I
P7 voltage
PlO voltage
IL
Correction VR7 ‘\ \ i$
A
/
with F.mA
k
When F.d
N’
> 1mA
When F.,ti < 1Lti
VR4 1
-
F.kV ,~/@ lVR4 F.kV
I
Adjust VR5 to set the P8 voltage to OV when F.mA=lmA. ! VR8 through VR13 adjustment (for spot film radiography)i These VRs correct R.kV values for film division and image intensifier (1.1.) magnification. Division and 1.1. magnification !
Adjustment VR
!
Division by 2
VR8
Division by 3
VR9
Division by 4
VRlO
,
Spare
VRll
1 j
Ml
VR12
M2
VR13
iVR14 through VR17 adjustment (for 1.1. photofluorography) These VRs have the same functions as VR4 through VR7 for spot-film radiography.
- 65 -
No. 25313-002EfcB
PWB name:
[ATR ANA]
KXO-30R
A:
2/z
B:
KXO-30F/50F
Equipment Setting
When installed
and adjustment
A VR18 through VRZO adjustment (for 1.1. photofluorography) These VRs have the same functions and 13 for spot film radiography.
0
’
1
j
PWB name:
as VRll,
12,
VR21 through VR24 adjustment (for AUX TECHNIQUE) These VRs have the same functions as VR4 through VR7 for spot film radiography. VR25 through VR30 adjustment (for AUX TECHNIQUE) These VRs have the same functions as VR8 through VR13 for spot film radiography.
[HEAT
A:
11
KXO-30R
B:
KXO-30F/50F
When installed
Setting
I
r
0
!
0
j VR2:
Frequency
adjustment
o
/
o
1 VR3:
Frequency
adjustment
/ VR4:
Offset
o 1-0
/Unnecessary I I IUnnecessary
adjustment
A:
[HEAT 21
PWB name:
/ iunnecessary
KXO-30R
B:
KXO-30F/50F
When installed
Setting
r I
0
’
0
0
I
o
I
’l/
VR12:
D/A converter
span
j
Unnecessary
I
j
128/256:
128 crossover
span
- 66 -
1 Unnecessary
No.
PWB name:
A
B
Adjustment point
0
0
CNNA
0
0
0
0
Equipment .
KXO-30R
A:
[INV-11
Contents
B:
2J313-002EkI
KXO-30F/50F Adjustment when installed
of adjustment
Unnecessary
;'(OVER CURRENT detection level setting +8V (TPl) VR2 KXO-50F=B or KXO-30F=B (VRl,VR2) * For the KXO-30R=B, VR3
Tube
voltage
waveform
m I
Note *: In the case of single-phase power supply: If 160mA or more is set, the tube waveform shifts slightly downwards, this cannot be adjusted with If VR3 is used VR3. for adjustment, the tube voltage waveform may become abnormal under a small load.
Tube
0
0
VR5
Tube overvoltage at 165kV
0
0
VR6
Tube voltage output adjustment in fluoroscopic mode Set the measured tube voltage to 90kV with 90kV and 1mA set and LED4 on the INV-2 PWB indicating "8".
Necessary
0
0
VR7
Error signal level -0.6 to 1OV (TP~)
Unnecessary
0
0
VR8
Frequency adjustment When Pin 7 at location IC: 2.5kHz at TP12 Variable-width timing
20kV
detection
to 1V (TP2)
Unnecessary
VR4
VR9,VRll
gain
In the case of a singlephase power supply, the '""=I tube voltage waveform may gradually descend under a heavy load.
0
0
signal
r
0
0
voltage
When necessary (fine adjustment
adjustment
If the tube voltage waveform under a small load * (about 5OmA) is overshoot or undershoot, Set the adjust VR3. waveform by a tube voltmeter to be the same as that at TP2.
Unnecessary
level
Detection
adjustment
adjustment
- 67 -
Unnecessary
Unnecessary 2D is 5V,
of phase
control
Unnecessary
No. 25313-002E*I
A:
[INV-21 l/2
PWB name:
KXO-30R
B:
KXO-30F/50F Adjustment when installed
Equipment Setting
and adjustment
A
B
0
0
D/A converter adjustment VRl, VR3, VR5: Zero-position adjustment Gain adjustment VR2, VR4:
Unnecessary
0
0
SW2,
Necessary at tube voltage adjustment
0
0
I
SW3:
Switches to write the tube voltage adjustment data (kV shift data) in radiographic or fluoroscopic mode into the .RAM on the PWB. kV to be set: For each section 20kV to 31kV, 32kV to 47kV, 48kV to 63kV, 64kV to 95kV, 96kv to 127kV, and 128kV to 150kV, kV shift data can be written into the RAM for each mA and each tube.
Necessary
SW1 POWER source
I
/
o
0
1
SW4:
Setting
Position
19 2oov
0
l$ 220v
1
3cP 380V
4
39 415v
5
39 440v
6
Tnverter
operation
disable
of SW1
switch
When this switch is set to "DIS" (DISABLE), LED3 lights up and the INV-UNIT operation is set this switch to "EN". disabled. Normally, Note:
I
I
If SW4 is set t0 "EN" (ENABLE), LED3 will light up in the following cases: @
If LV
@
If SW1
is not
@
If the
INV unit
@
During
charging
is too high
or too low
set as specified
:
- 68 -
fuse
is blown
Necessary
No. 25313-OOZE*1
PWB name:
[INV-21
2/2
A:
KXO-30R
B:
KXO-30F/50F
Equipment Setting A
B
0
0
SW5:
(YWA1977
Adjustment when installed
and adjustment
only)
Necessary
Length of the longest among those connected the three X-ray tubes
cable to
to 15m
Setting of SW5
position
3, 2, 1 or 0
16 to 20m
2, 1 or 0
21 to 25m
1 or 0
26 to
0
-
68A -
No. 2J313-002E*D
PWB name:
A:
[INV-31
KXO-30R
B:
Adjustment when installed
Equipment Setting
and adjustment
A
B
0
0
VRl
When the power voltage Line voltage signal: has the nominal value, signal at TPl is 5Vp.
0
0
VR2
TP2:
VR3
VR4
VR5
0
0
0
(
0
I
I
1
I
I I 1
cases>
TP3: +4.87V +4.75V
(at 14 2OOV), (in all other
cases>
TP4:
(at l$ 2OOV), (in all other
cases>
+5.5OV +5.5OV
VR6
TP6:
CNNA
Set to "NEGA"
SW1
Turn
(at l$ 2OOV), (in all other
= 1V for tube current
Unnecessary cases> Unnecessary
lOOmA
1 Ucnecessary
j
0
0
(at lb 2OOV), (in all other
TP5:
Unnecessary
Necessary
+4.75V +4.5OV
+5.35V +5.25V
HO-3OF/5OF
on 1 of SW1 for a single-phase power Turn it off for all other power suPPlY* supplies.
it is unnecessary to set SWl, for KXO-30R=B, KXG30F=B or KXO-5OF.
Necessary
the
I
-
-
69 -
No.
7.2
1
Error
Code
,
1
’
.k
_k
Code
Table Meaning
Transmission and the box
system error is not carried
(Data communication out correctly)
between
the
198
RAM error
(Warning.
RAM on
the
panel
or
desk
is
defective)
199
RAM error
(CPU stop.
RAM on
the
panel
or
desk
is
defective)
201
TECHNIQUE setting
202
Tube No. setting not set correctly)
301
TECHNIQUE setting
700
Box-side
CPU or
913
1 Normal starter / was not ready) I
914
; Over MAS I 1 Filament heater
88 89 t
2J313-002Ey:B
i Filament 1 greater
switch
(on
tghe
(Tube
error
switch
panel
setting
(on
the
(Current
RAM down
desk)
switch
panel
serial-transmission error
or
or
on
the
desk)
is
CPU down was
applied
(Heating
heater current abnormal than the specified value)
data (The
is
the
set
panel
not
(Either
when
are
not
panel/desk
or
desk
is
set
CPU malfunctions normal
starter
abnormal)
heating
current
has
become
1
1
90
Filament hardware
heater output is abnormal)
95
Over current (The discharge, etc. )
96
Over
kV (Output
kV value
97
Fuse
blown
fuse
98
Line voltage smaller than
(The
section
inverter
of
abnormal
thyristor
has the
become INV-UNIT
(Heating
70 -
system
has
malfunctioned
due
greater
than
specified
is
the
to
tube
value
blown)
(The power source voltage error the specified value and X-rays
-
output
has become greater cannot be generated.
or
No.
8.
MAINTENANCE
SCHEDULE described
The schedule of periodic maintenance procedures conducted for both the X-ray tube and generator listed below. First Servicing Second Servicing Note:
2J313-002E*B
-------------________
system
here
with
shall
be
the frequency
6 month after installation 1 yaer after installation
All subsequent servicing shall be performed every 12 months When a new X-ray tube is installed, repeat the first thereafter. servicing.
APPENDICES LIST APPENDIX
01
Recording
Form
APPENDIX
02
Relations Between Focal Tube Current Range
APPENDIX
03
Test Method
of Fluoroscopic
APPENDIX
04
Test
of Reproducibility
APPENDIX
05
The Setting Technique Factor for Reproducibility and Linearity Test without
Method
for Tests Spot
Size and
Entrance
Exposure
Rate
and Linearity
A.E.C.
APPENDIX
06
Test
Method
of Automatic
Exposure
Control
Limit
APPENDIX
07
Test Method Time
of Automatic
Exposure
Minimum
Exposure
- 71 -
No.
APPENDIX
01
Recording
Form
2J313-OOZE*B
for Tests
This
data sheets are to be filled out and copied at the time of the Record test results in copied sheets and return installation and maintenance. to: -------------
U.S.A.
TOSHIBA
MEDICAL
SYSTEM
DEVISION
AMERICA INC. 2441 MICHELLE CALIFORNIA 92680 The
---------
others
TOSHIBA'S
Service
Date:
2.
Installed Branch
Maintenance
by:
Office:
Address: 3.
Hospital: Address:
4.
Serial 0
No. of X-ray
Generator
DC-30R:
q DC-30F:
5.
0
DC-50F:
17
DG-30F:
q
DG-5OF:
X-ray
Tube
to be Combined
Manufacturer
6.
Tube
1
Tube
2
Tube
3
X-ray
Table
Model
name
to be Combined
Manufacturer: Model Serial
name: No.:
- 72 -
DRIVE,
TUSTIN,
representative
: Installation,
1.
OF TOSHIBA
Serial
No.
Inspection
No.
X-ray
beam
limiting
device
combined
2J313-002E~kB
with
Manufacturer: Model
name:
Serial: 8
Power
source 0 0
(1) Power
phase
(2) Power
frequency
q 5OHz 0
(3) Power G 0 9.
Single phase Three phase
60Hz
voltage 0 22ov i? 415V
2OOV 38OV
Preliminary
test
0
440v
3
48OV
results Result
Item Section Section Section Section Section Section Section
Section
Tube
3.2 5.1.2 5.1.3 5.1.4 5.1.5 5.1.6 7.1
5.1.5
No.
Power Source Checking voltage and display circuits Checking operations Checking the switching circuit Checking the starter circuit Checking the battery PWB Setting and Adjustment Tables
Check
on starter
Tube model name
circuit
Normal speed boost time
1
- 73 -
starter (set>
OK, OK, OK, OK, OK, OK, OK,
NG NG NG NG NG NG NG
No.
10.
Radiographic
KV, mA
and time
2J313-002E*K
test
TUBE NO. Tube model High
(1) R.kV
potential
cable
R.Time:
and R.mA
I
lx”
71A L
kV
mA
m
length
kV
mA
0.05
kV
set
mA
kV
mA
13 16
L
I
J
- 74 -
No.
TUBE
2J313-OOZE*K
NO.
Tube model High
R.kV
potential
cable
length
R.Time:
and R.mA
m
0.05
set
140
120
kV
-
75 -
mA
kV
mA
No.
2J313-002E*K
TUBE NO.
R.kV
10
and
Tube
model
High
potential
cable
m
R.Time:
R.mA
kV
length
mA
kV
mA
kV
13 16 20 25 32 40 50 64 80 100 125 160 200 250 320 400 500 640
-
76
-
mA
0.05
set
kV
mA
kV
mA
2J313-002E"B
No.
(2)
R.time
R.time
R.kV
0.005sec
60kV
Result
R.mA
1
(set>
320mA
111 +
lOOmA
90kV
1
0.2sec
11.
Fluoroscopic
test
(KXO-30F
or 50F) Tube model Table
model
F. KV and F. mA
/ I
1
I I I j
2
12.
1
F. ABC
mL4
mA
mA
mA
mA
mA
mA
mA
InA I
mA 3
In U.S.A, limit.
’
/ mA
4 Note
m&4 / I
test
should
be performed
within
the entrance
test
. Thickness
of the acrylic
cm
phantom:
V
. Peack value of the video signal waveforms: (including the signal pedestals) . Voltage
at P6 of the PTW unit:
requirements: . Fluoroscopic (during stable operation of the ABC circuit)
V kV,
- 77 -
mA
exposure ,
No.
13.
Automatic exposure control test (in combination with the automatic AEC-lOA/B) Record manual
14.
exposure
2J313-002E*B
. unit, model
control
test results, using data sheets attached to the installation (No. 2C376-519E). for A.E.C. UNIT, model AEC-lOA/B,
ATR adjustment (1) X-ray X-ray
(KXO-30F
or 50F) (FFD
table:
>
tube:
Grid: Intensifying
screen:
Film: (2) Set value R.kV BACK
kV kV
Clamp
DOWN: UP: UP mAs:
(3) Correction
ITAS
value
Correction
Correction mA correction
Division correction
0.5mA 1mA 2mA 3mA 4mA
0
value kV kV kV kV kV kV kV kV kV kV kV
l/2 l/3 l/4 l/6 Ml M2
- 78 -
No.
(4) ATR
data
in FKV-AUTO
--
:
--.-
---
-
i L :; -. . . :-
;_
I!
;_
__
.:
.I
-
F_
f
_. __~
._
.
____L_~-.-
gheet:~‘;~___i_I~”
ATR.d.ata
.-... . . _ _. _-.
-_ .. _.._-.
. -_--
-----
:I
:
::f
_~~~:-
s-___._.-----_ -1. -f -_
--!
._. -. _. :-
.. .-
.--.
I I
_ .
__
_
__. ..__._. . _._ . .
I
I
I
.
.
_.
. _
--‘i
t-
.
. .
_
I
I
I
. , I-
:
__
-.-
&_...- cm22--. -_--_ . . -_. _ ____ . _.________I___-_-_ . _ --.--_ _- ---.__. _ ._____-----Acrylic phantom thickness :L.LL--L- :.__________ ._._.---. __ ] ~ _-
li --__l&___---_-16
-_-
(mA: Manual)
._. . _. - ._ . _._-..__ _ _.
.
._______.&_______
mode
2J313-002E*B
_~____~&-_-__2O__
-L_
T Acrylic phantom 12 cm
Fluoroscopic
conditions
mA
kV
14 16 18 20 22 24
-
79 -
Radiographic
kV
conditions
mA
No.
15.
23313-002E*M
HHS Test The following tests are applied to the X-ray generator. The HSS test is necessary only for ORH regulations.
15.1
Entrance
exposure
(1) Installed (2) Measured a> Manual
rate
system: entrance
limit 1. Under exposure
tube
2. Over
tube
3. C-arm
rate
mode
b) ABC mode Brightness Entrance
control: exposure
limit
F.mA
limit
F.kV
maximum rate:
.
position (R/min) (1 R/min = 2.58 x lo-*C/(kg*min)) mA kV
- 80 -
No.
15.2
Reproducibility
kV
and linearity
60 kV
without
automatic
exposure
TUBE
NO.
TUBE
TYPE
2J313-002E*M
control
60 kV
60 kV
60 kV
60 kV
mA
InA
mA
mA
mA
mA
sec.
S
S
S
S
S
Focus Line
regulation
Exposure No.
I
I
-+%
I
-+%
I
I I
-+%
-+%
1
mR
mR
mR
mR
mR
2
mR
mR
mR
mR
mR
3
mR
mR
mR
mR
mR
4
mR
mR
mR
mR
mR
5
mR
mR
mR
mR
mR
6
mR
mR
mR
mR
mR
7
mR
mR
mR
mR
mR
8
mR
mR
mR
mR
mR
9
mR
mR
mR
mR
mR
10
mR
mR
mR
mR
mR
mR
mR
mR
mR
mR
x S
-+%
I
I
I
I
I
I
I
C
x
I
mR/mAs
mR/mAs
1
mR/mAs
1
mR/mAs
I
1
mR/mAs
1
I
Xl +X2 (1 mR = 2.58 x 10-7C/kg) Note:
Make a sufficient results.
number
of copies
- 81 -
of this
sheet
to record
test
No.
TUBE
NO.
TUBE
TYPE
2J313-002E*M
kV
90 kV
90 kV
90 kV
90 kV
90 kV
mA
mA
mA
mA
mA
mA
sec.
S
S
S
S
S
Focus Line
regulation
Exposure No.
‘j;-
+ -
%
+ -
%
-+
%
+ -
%
-+
%
1
mR
mR
mR
mR
mR
2
mR
mR
mR
mR
mR
3
mR
mR
mR
mR
mR
4
mR
mR
mR
mR
mR
5
mR
mR
mR
mR
mR
6
mR
mR
mR
mR
mR
7
mR
mR
mR
mR
mR
8
mR
mR
mR
mR
mR
9
mR
mR
mR
mR
mR
10
mR
mR
mR
mR
mR
mR
mR
mR
mR
mR
mR/mAs
mR/mAs
mR/mAs
mR/mAs
mR/mAs
S
C ?i I)il - x21 Xl + x2 (1 mR = 2.58 x IO-7C/kg)
- 82 -
No. 2J313-002EkM
TUBE NO. TUBE
TYPE
kV
140 kV
140 kV
140 kV
140 kV
140 kV
mA
mA
mA
mA
mA
mA
sec.
S
S
S
S
S
Focus Line
regulation
Exposure No.
x
+ -
%
+ -
%
-+
%
+ -
%
+ -
%
1
mR
mR
mR
mR
mR
2
mR
mR
mR
mR
mR
3
mR
mR
mR
mR
mR
4
mR
mR
mR
mR
mR
5
mR
mR
mR
mR
mR
6
mR
mR
mR
mR
mR
7
mR
mR
mR
mR
mR
8
mR
mR
mR
mR
mR
9
mR
mR
mR
mR
mR
10
mR
mR
mR
mR
mR
mR
mR
mR
mR
mR
mR/mAs
mR/mAs
mR/mAs
mR/mAs
mR/mAs
S
C 51
Xl +X2
(1 mR = 2.58 x lo-7C/kg)
- 83 -
No.
15.3
Automatic
exposure
Setting technique Measured mAs 15.4
Automatic
exposure
(1) AEC Type: Measured (2) AEC Type:
Field
control
limit.
minumum
mA
50 kV
factors:
exposure
time
phototimer time Solidtimer
Measured
time
1 2 3 1+2 1+3 2+3 /1+2+3/
- 84 -
ms
2J313-002E*B
No.
APPENDIX
02
RELATIONSHIP
BETWEEN
FOCAL
SPOT
SIZE AND TUBE
2J313-002E9cB
CURRENT
RANGE
I
Small X-ray
Large
-r
focus
tube Focal
type
Al92
focus
(EIMK)
spot
The current
size
range
Focal size
spot
0.6 x 0.6mm
20 to 320mA
1.2 x 1.2mm
Tube
current
range
400
to 1OOOmA
Tube voltage range
40 to 150 kV 1
-
85
-
No,
APPENDIX 03
2J313-002E*M
TEST METHOD OF FLUOROSCOPIC ENTRANCE EXPOSURE RATE
A.
Objective Fluoroscopic equipment which is provided with automatic exposure rate of tube potential and current control is not operable at any combination rate in excess of 2.58 x lo-3C/ (kgemin) which results in an exposure of the useful beam enters the {lOR/min} at the point where the center (See the section 1020.32 (d).) patient.
B.
Test method (1) The position as follows. a.
Testing
of
test
when
the
equipment
source
is
and
below
flip
t
I
\
diagnostic
i
table
system
Spot
film
Lead
sheet(300
device x 300 x Zmm)
Aluminum attenuation (200 x 200 x 19mm)
3oomm
assembly
table
Imaging
I
X-ray
block
4Dosimeter
1
P Beam-limiting X-ray
the
Set
2.
Place the attenuation block (200 x 200 x 19mm) at the approximate center of the table and 150mm (6in) above tabletop and conform the center of the X-ray field to observing the image. imaging
to
3.
Set
4.
Place field image.
5.
Place the lead sheet X-ray field between system.
the and
the
system
horizontal
tube
1.
the
tabletop
device
to
probe of the 1Omm (0.4in)
position.
300mm (12in)
dosimeter above the
at the tabletop
above
86 -
tabletop.
center of the by observing
(300 x 300 x 2mm) at the attenuation block
-
the
the and
the it by
X-ray the
center of the the imaging
is
2J313-002E*B
No.
b.
Testing
when
the
source
is
above
the
table
Beam-limiting
device
Aluminum attenuation (200 x 200 x 19mm)
Spot
Lead
sheet
(300
x 300
film
device
Imaging
to
the
horizontal
sys tern
position.
Set
7 _.
If of be
3.
Place the attenuation block (200 x 200 x 19mm) at the center of the table and 150mm (6in) above approximate the center of the X-ray field to tabletop and conform observing the image.
5.
tabletop
x 2mm)
1.
4.
the
the source-to-image the beam-limiting placed provided
block
distance (SID) is variable, set the as close to the tabletop as it device shall not be closer than 30Omm. it
probe of the 30Omm (12in)
dosimeter above the
at
the
center
Place field image.
the and
Place X-ray
(300 x 300 x 2mm) at the lead sheet field on the tabletop.
-
57 -
tabletop
by
the
of
the
the it by
X-ray
observing
center
the
of
the
end can
2J313-002E*M
No.
Testing table)
c.
in
a C-arm
type
fluoroscope
(in
a Gyroscopic
-
4
diagnostic
Imaging Spot
A
system
film
device
Lead sheet (300 x 300 x Zmm) Aluminum attenuation (200 x 200 x 19mm)
300mm 850.5mm (MIN. SID)
Beam-limiting X-ray
t
1.
If necessary, primary beam.
2.
Place the probe of the surface of the imaging
3.
Place the aluminum attenuation block (200 lead sheet (200 x 200 x 2mm) at the center shown in the above figure. open
shutter
the
tabletop
dosimeter system.
(2)
Fully
(3)
Set the Rotary DIP SW, SW5 on the to the position 0. This setting means that the X-ray entrance exposure rate limit.
(4)
Push
(5)
Select the not select
(6)
Set the F.mA control film device.
(7)
Perform
(8)
Monitorig
(9)
When the dosimeter (kgamin) {9R/min}, record it.
(10)
Repeat
the
the
position
blades
of
out
of
the
path
the
the
input
x 200 x 19mm) and the of the X-ray field as
device.
DESK CONTROL PWB in is
of
from
beam-limiting
generator
device
tube
300mm (12in)
the
block
the
excluded
control
from
desk
the
POWER ON switch. spot-film the “ABC”
radiography mode on the for
technique spot film
4mA and F.kV
on the device.
control
for
control
panel,
5OkV on
the
spot
fluoroscopy. increase
a dosimeter,
Steps
(6)
the
reading reaches read the setting
to
(9)
at
the
-
tube
to
125kV.
the limit level 23.22 x lo-*C/ F.kV on the control panel and
F.mA setting
88 -
voltage
of
3,
2,
1 and
0.5mA.
and
No.
(11)
Set the limit kV which is 1 for each F.mA setting. numbers are shown below.
F.mA
i
in Step (9) according to the Table settings and the settings switch
SW No. on the F.CONTROL control box
setting
0.5
recorded The F.mA
2J313-002E*M
DIG.
PWB
in the
SW1
mA 1mA
SW2
2mA
SW3
3mA
SW4
4m.A
SW5
(12)
Select the F.ABC and MA-AUTO mode switches. Set the brightness control to the minimum position.
(13)
Perform
(14)
Monitoring
(15)
When the dosimeter reading reaches the limit level 23.22 x lo-4C/ (kgemin) {9R/min), read the indicated F.kV and F.mA on the control panel and record them.
(16)
Set the limit kV recorded above according to Table 1. Set the limit mA recorded above according to Table 2. The setting switches are shown below.
fluoroscopy. a dosimeter,
(17)
Note:
limit
mA
limit
the brightness
SW No. on the F.CONTROL control box
Limit
kV
increase
DIG. PWB
setting
in the
SW6 SW7
!
Set the Rotary DIP SW, SW5 on the DESK CONTROL to the position 1. When
control.
the DIP
switch,
-
the power
89 -
should
PWB
in the control
be off.
desk
No.
Table
1
Limit
kV setting
2J313-002E*B
code
(1; ON, 0; OFF) F.kV
-
1
2
SW. b
3
No. 5
6
7
8
F.kV
1
2
3
“;’
‘1.
6
7
8
I 50
0
0
1
t
I
I 51 lolo 52
10
1
’ l/1lOll:o
54
0,o
1
55
0
0
010
58
011/o /
1
0
1 1
)
1
(
1
IO
0
I
i
61
0
0
1
1 ; 1
‘l/O
62
0
I 0
1
1 / 1 ( 1 / 1
Note:
When
l~o;O,o
1
1 0
0
0 0
1
74
0
1
75
0
76
0
1
77
0
l/O
1
/
/
/ O/l
93
Ojl
0
11
0
94
0
1
1
95
0
10
01
96
0
1
0
10
IO
L
jl
010 1
1
Oil
1
/
1
1 1
/ 0
0
0
0
0
III
l’lil/l/l 1010
I
011
l/O
0
1 0 /
1
/ 1 1 0
I
1
1
1o7
/
0
/
1
1 1
1
0
/
1
/
!
: 0
l/
1
j
0
1
108
1
j
109
O/l
110
0[1
0
/
0
i0j
011
99 100
1
0
1/00/0/1~0
1
j
l
011
0
1
011/011
(
1
ojo
11
111
0
112
O/l
1
j 0
1 1 1 0
1 0
/
1
/
/
/
1
/
1
1 1
0
10
I
10 1
’
1
11
ljOlO(
-?
l/l
0
0
I
0
1
1
1 1
o/oJo
0
1
80
0
1 / O/
1
81
0
10
1;O
82
011
83
0’1/0/L~0~0/1
84
0
l/O
85
0
1
011/o
86
0
1
0
1
87
0’1
0
l,r)iiil
1
1 l/O
1
I
ilk
Oil
11
115
O/l
1
1
O’Oll/l
116
0
1
1
0
1
101110/l 1
O/
1
o/1/+
1
O/l /
0
/
0
10
0
0111
0
Ijo;1/0 1
0
-
90 -
on
/
11
11
119
0
1
1
01
120
01111
I/
121
Ojl
1
lql;olo~l
0
122
O/l
1
I
IjO
1
123
0
1
1
1
1
121,
0
1
1
I
1
125
oj
1
lllll
all
1
f
I 1 /
0
0
1
11
0
1
1
i
turn
O/l/O
0
/
j 0 / 1/ 1j 0 / I
0
118
0
010/i
125kV,
1
1/1/o
I 1 I 1
0
10
0
011
I
/o
)0~1,1/0/0,0/0/1
198
1)
j
79
than
1 1’1
1 1
0
I
/O/l
1
0
01
0
,l 1
/
0
I
o/1 0
0
o/
73
kV is more
,oo
10
0
j 0
~]
im
!
1
011 !
97
1(1,1 ’
72
1011
0
,
1
j
90
O/l
O/l
1/110!1,1/
0
I.0
(
91
l\l)Oil
0
71
89
, IO I
1
92 111’
60
10
o/101 I
/
I
[ 69
I/
(
10’0
1 011
j
011
l/l
IO,
88
0
,
/
10’0
I59
0 1
ijOll/l/O
1
/
57
I
I
o/o
/
1
I
II
53
0
0
, l!O/Ojl/lj
1
O/l
56
0
I
/
1I 0
switches.
/
0 j 0 1
I
1
0
0
I
1
1
0
10
(
0’ .’
No.
Table
F.mA
SW. No. 4 5
1
2
3
0.2
0
0
0
0
0.3
0
0
0
0.4
0
0
0.5
0
0.6
2
Limit
mA setting
code
SW 7
SW. No. 4 5
6
7
8
F*mA
1
2
3
0
0
1
0
2.6
0
0
0
1
0
0
0
1
1
2.7
0
0
0
0
0
0
1
0
0
2.8
0
0
0
0
0
0
1
0
1
2.9
0
0
0
0
0
0
1
1
0
0.7
0
0
0
0
0
1
1
1
3.1
0.8
0
0
0
0
1
0
0
0
0.9
0
0
0
0
1
_O
0
1.0
0
0
0
0
I
0
1.1
0
0
0
0
1
0
6
7
8
1
0
1
0
1
1
0
1
1
0
1
1
1
0
0
0
0
1
1
1
0
1
0
0
0
1
3.2
0
0
1
0
1
3.3
0
0
1
0
0
0
0
1
1
0
3.4
0
0
0
I
0
1
1
0
1 0
( 0
/ 11
1
0
0
1
0
0
1
3.5
I
1.2
0
LL
0
0
1
1
0
0
1
1
0
1
1
1
1
o(
1.3
0
0
0
0
1.4
0
0
0
0
1.5
0
0
0
0
1
1
1
1
1.6
0
0
0
1
0
0
0
0
1.7
0
0
0
1
0
0
0
1
1.8
0
0
0
1
0
0
1
0
1.9
0
0
0
1
0
0
1
1
2.0
0
0
0
1
0
1
0
0
2.1
0
0
0
1
0
1
0
1
2.2
0
0
0
1
0
1
1
0
2.3
0
0
0
1
0
1
1
i
2.4
0
0
0
1
1
0
0
0
2.5
0
0
0
1
1
0
0
1
1: Note:
ON, 0: OFF When mA is
more
than
I
4.OmA,
0
I
0
1
1 0 1 0 1 1 1 0 1 0
4.0
91 -
I
I
3.8
turn -
0 0 I 1 / 0 / 0 1 1 j
3.6
0
I
2J313-002E*K
on all
0
0
switches.
1
0
1
1 1 1 1
0
0
0
No.
2J313-002E*M
(18) After adjustment of the entrance exposure rate limit, repeat the Steps (6) to (8) for the F.mA setting of 4, 3, 2, 1 and 0.5mA and record the tube voltage and the dosime ter readings at the limit points in manual mode (19) Perform current
fluoroscopy in the FULL-AUTO mode and record and dosimeter readings at the limit point.
the tube voltage,
(20) If the dosimeter reading exceeds the rejection limit, follows and then repeat the Steps (18) to (19). (a) Manual mode If the test results are not within the rejection the limit kV which has been set in Step (11).
as
limit,
decrease
(c) FULL-AUTO mode If the test results are not within the rejection limit, the limit kV and mA which have been set in Step (16).
decrease
(b) kV-AUTO (F.ABC) Same as above.
C.
proceed
Test
instruments
required
(1) Dosimeter This measurement a. Keithley 006A).
mode
requires
model
35020
one of the following
with
96020
b. Victreen model 550 with 550-3, to TOSHIBA APPENDIX 026A.) C.
Victreen APPENDIX
(2) Aluminum
model 555 with 008A.j
attenuation
lmA,
chamber
550-4,
test
(Refer
550-5
instruments:
to TOSHIBA
or 550-6
1OmA or 1DAS chamber
APPENDIX
probe
(Refer
(Refer
to TOSHIBA
block
200 x 200 x 19 (mm> (3) Lead
sheet
300 x 300 x 2 (mm> D.
Rejection
limit
Maximum entrance dose limit: 23.22 x 10_4C/(kg=min) {9R/min). The maximum limit means a reading value on the dosimeter specified C.
-
92 -
in item
No. 2J313-002E*B
APPENDIX A.
04
TEST
METHOD
OF REPRODUCIBILITY
AND LINEARITY
Objective are performed A series of 10 radiation exposure measurements the coefficient of variation and the average exposure ratio. Section 1020.31 (b) and cc>.>
B. B-l
Test
to calculate (See the
method
Testing
without
automatic
exposure
control
(1) Set the X-ray apparatus and test following block diagram.
instrument
according
to the
s Dosimeter
Focus detector distance: 1OOcm Collimator
Primary voltmeter
:
High-voltage generator 1
(2) Turn (3) Set and two and
on the equipment
X-ray control
r~:7.
Power source
&
line switch
and calibrate
test
instruments.
the technique factors (radiographic kV, mA and time selectors) at intervals of one or during the 10 consecutive measurements, factors are adjusted to alternate settings minutes the technique reset the test setting after each measurement.
For a detailed description of this step refer to APPENDIX 05 "Combination of Specified Technique Factors" in this manual. (4) While confirming that the respective percent line voltage regulations are within +l% of the mean value for all the 10 measurements, record the value of a dosimeter. (5) The coefficient of variation "C", 1020.30(b)(8), of the 10 exposure measurements is calculated using the following equation;
where
Xi = the exposure measurement x = mean value of the exposure measurements S = estimated standard deviation
-
93 -
No.
(6)
Confirm
that
the
above
product
(7)
The average ratios of exposure product (mR/mAs), (1020.31(c)), calculated using the following
“C”
is
B-2
= average mR/mAs values tube current settings.
Confirm
above
Testing
with
the
automatic
product
exposure
X- ray apparatus block diagram.
( 1) Set
the following
than
0.045.
in
x lo-TC/kg)
where : xl and x2 consecutive that
greater
to the indicated milliampere-second of the 10 exposure measurements equation;
(1 mR = 2.58
(8)
not
2J313-002E*M
is
not
obtained
greater
at
than
each
0.09
of
two
times.
control
and
instruments
test
according
to
the
Dosimeter
Beam limiting
device __--------_____
___------
-----_ Aluminum attenuation block (200 x 200 x 38mm) SID
(i.e.
Source Image 1ooomm
Distance)
Primary voltmeter Power
H.P. Igenerator
(2)
Set the cassette type) posterior
(3)
Set the tube film density
(4)
Adjust
(5)
Adjust the the inside
(6)
Set
the
X-ray control
(
]
the
with film (Sakura to the detector as
voltage regulator
AEC so beam size
timer
of
for for
that
the
limiting of the the
90kV, “0”
A type) and screen shown in the above
the tube posit ion
film
density
X-ray
control
-
94 -
to
current for lOOmA and and make an exposure. of
device that the attenuation block.
(Toshiba figure.
0.85
to
exposure
0.3sec.
or
1.15 field
longer.
is
EM
the
got.
conforms
to
No.
(7)
Take
10
consecutive
percent
measurements
line-voltage All
measurement. within
-+l%
of
Measurement
/ X-ray 1
I(
conditions The The au
t
I
3
(Tube
1 the
tube
I 1 The
maximum
voltage tube
proper
time
current
to
1 The
proper
; p roper
tube
time
This func
(1)
time t ion.
can
to
be
Exposure
that
greater
than
by
equipment
to
b. C.
1 “0”
Keithley
b.
position
I
/
“0”
pos it ion
time.
that
by
of
1 “0”
position
I
an
ON/OFF
for
equation
!
the
oscilloscope SIGNAL”
with
terminal
a of
storage each
of AEC.) plural detecting fields, test For example, test should field.
detecting fields
get
2.
using
diagram with
“C”
can
1,
the
detector
with
3
fields.
the in
10 exposure step 6 of
measurements, Paragraph B-l,
the
following
dosimeters.
is
not
0.045.
model
35020
one with
of 96020
chamber
(Refer
to
TOSHIBA
APPENDIX
> 550-4,
550 with 550-3, Victreen model to TOSHIBA APPENDIX 026A.j Victreen
model
APPENDIX
008A.
555
with
lmA,
1OmA or
550-5
or
550-6
1DAS chamber
(Refer
probe
(Refer
to
TOSHIBA
>
measurement
ALCO primary 012A.
,
required
Electrical a.
I positions
change
“F:q:POSURE
Radiation measurement This measurement requires
006A.
got,
measured
the product using the
Density
time)
time to 0.3
proper
condition
each
table.
change
current
3 detecting
following
level.
connection the detector
taken
to
be
the
the
system the
not
the
is
the of
the
get
and
be
level.
proper
be
point
done
of
for
got,
The
measurements.
current,
step.) not be
can
should
Confirm calculated
a.
(2)
all
according
proper
the
except
should be
the
voltage
AEC. (See In the case
Test
done
Tube
this can
that
Measuring
C.
be
current
the tube
Note:
to
voltage
/ the
(8)
for
value
voltage,
tube
When
i
mean
should
(See NOTE in proper time
the
The
I
regulation values for
within a time period of 1 hour. shall be determined for each line-voltage regulation shall percent
minimum tube voltage of the system tube current that can get the proper omatic exposure control time of 0.1
1 sec.>. ’ When
12
the
ZJ313-002E$cB
voltmeter
model
PV-401
(Refer
to
TOSHIBA
APPENDIX
>
Tektronix 059B. >
storage
oscilloscope
-
model
95 -
464
(Refer
to
TOSHIBA
APPENDIX
1
No.
D.
Rejection
limit
(1)
Coefficient
(2)
Average
of exposure
variation
c = < 0.045
ratio
1x1 - x2/ Xl
-
+ x2
96 -
2 0.09
2J313402E*B
No.
APPENDIX 05 THE SETTING TECHNIQUE FACTOR FOR REPRODUCIBILITY TEST WITHOUT A. E. C. 1.
2J313-002E+B
AND LINEARITY
General which determines how the combination of technique fundamental idea, d, is to select a peak tube potential factors for the test are selecte middle and high kV) and the product of tube current and exposure (low, time (approximately 5 mAs and 30 mAs).
The
During a series of 10 radiation controls for technique factors reset to the test setting after According
60kV 1OmA 3.0sec
to
the
following
’ I1
-
1
4th exposure
all
variable settings
exposures
are
1
13.0sec
]
:Alternate setting(l))
(Test setting)
1
Resetting 60kV 1OmA 3.0sec
/I
(Alternate setting(2))
-/
10 radiation
procedure,
13.0sec
(Test setting)
Setting 70kV 13mA 2.31sec
exposure measurements, are adjusted to alternate each measurement.
I-
_I
/ (Alternate setting(l))
(Test setting)
_ ____---
----
1
I
Test
Procedure
-
97
-
1OmA / 3.0sec / (Test setting)
-*
and
measured.
No.
2.
Combination
No. Test Settings
of Specified
1 60kV 160mA 0.188sec
Technique
2
2J313-OOZE*K
Factors
4
3
6
7
60kV 20OmA 0.15sec
60kV
60kV
60kV
60kV
60kV
250mA 0.12sec
32OmA 0.094sec
400mA 0.075sec
500mA 0.06sec
640mA 0.047sec
Alternate Settings (1)
50kV 125mA ,0.24sec
50kV 160mA 0.188sec
5OkV 200mA 0.15sec
50kV 250mA 0.12sec
50kV 320mA 0.094sec
50kV 40OmA 0.075sec
50kV 500mA 0.06sec
Alternate Settings (2)
70kV ,200mA 0.15sec
70kV 250mA 0.12sec
70kV 320mA 0.094sec
70kV 400mA 0.075sec
70kV 500mA 0.06sec
70kV 640mA 0.047sec
70kV 640mA 0.047sec
No.
8
T
9
11
10
12
13
14
Test Settings
90kV 1OmA 0.5sec
90kV 13mA 0.385sec
90kV 16mA 0.313sec
90kV 20mA 0.25sec
90kV 25mA 0.2sec
90kV 32mA 0.157sec
90kV 40mA 0.125sec
Alternate Settings (1)
80kV 1OmA 0.5sec
80kV 1OmA 0.5sec
80kV 0.385sec
80kV 16mA 0.313sec
80kV 20mA 0.25sec
80kV 25mA 0.2sec
80kV 32mA 0.157sec
Alternate Settings (2)
1OOkV 13mA 0.385sec
1OOkV 16mA 0.313sec
1OOkV 20mA 0.25sec
1OOkV 25mA 0.2sec
1OOkV 32mA 0.157sec
1OOkV 40mA 0.125sec
1OOkV 50mA O.lsec
No.
15
13mA
18
17
16
19
Test Settings
140kV 50mA O.lsec
140kV 64mA 0.079sec
140kV 80mA 0.063ec
140kV lOOmA 0.05sec
140kV 125mA 0.04sec
Alternate
130kV
130kV
130kV 64mA 0.079sec
130kV 80mA 0.063sec
130kV lOOmA 0.05sec
145kV lOOmA 0.05sec
145kV 125mA 0.04sec
145kV 160mA 0.032sec
(Note)
No. 6 and No. 7 should
be tested
-
98 -
only
for KXO-50F.
No.
APPENDIX
TEST
06
METHOD
OF AUTOMATIC
EXPOSURE
CONTROL
2J313-002E*B
LIMIT
A. Objective To determine that the product of X-ray tube current and exposure time is limited to not more than 6OOmAs per exposure when an automatic exposure control is provided.
B . Test
method
(1) Connect
High resistance voltage MA meter divider (KV-201) (MA-1201DC) /
as follows.
Power SUPPl
X-ray tube --&
\)
High-voltage cables
I
4
KV indicating unit (KV-201)
Exposure time counter (TD-3/-3A) (2) Turn
on the main
(3) Press the control. (4) Turn (5) Set
circuit
technique
on the automatic
switch
which
exposure
can use
control
the POWER
ON pushbutton.
the automatic
exposure
switch.
the exposure condition as follows. : 50kV R.kV : more than GOOmAs R.mA and time
(6) Remove the detector X-ray field. (7) Make back
and press
breaker
for automatic
an exposure and record up operates.
exposure
the mA reading
-
99 -
control
outside
and exposure
the
time when
No.
The operation of over-mAs lamp and the buzzer. (8) Verify
C. Test
that
equipment
the product
: &CO time
(3) High voltage meter
D. Rejection
circuit
of indicated
is indicated
value
is not
over
by the BACK UP
540mAs.
required
(1) MA meter (2) Exposure counter
back-up
2J313-OOZE$cB
Model
MA-120lDC
: ALCO Model
TD-3/-3A
: ALCO Model
KV-201
(Refer (Refer
(Refer
to TOSHIBA to TOSHIBA
to TOSHIBA
APPENDIX APPENDIX
APPENDIX
004B.) 018A.)
003A.)
limit
The product of tube current and exposure time measured on the mA meter the exposure time counter must not exceeded the following limits: Maximum
product
of tube current
and exposure
-
100
-
time
limit
:
540mAs.
and
No.
APPENDIX 07 A.
2J313-002E*B
TEST METHOD OF AUTOMATIC EXPOSURE MINIMUM EXPOSURE TIME
Objective When the X-ray tube potential is equal to or greater than 50kV, the minimum time for field emission equipment rated for pulsed operation shall exposure less than a time interval equivalent to two pulses, and the be equal to or time for all other equipment shall be equal to or less minimum exposure than l/60 second or a time interval required to deliver SmAs, whichever is greater (See 102.31(a)(3)(ii)).
B.
Test (1)
method Set
the
X- ray
apparatus
and
test
instruments
according
X-ray
X-ray
Beam limiting
I
I
device
Acrylic
tube
phantom
to
Fig.
2.
detector
(5cm)
Power SUPPlY I
I
L
I
High resistante voltage divider (KV-201)
H.P. transformer
.
2
AEC selector
diagram
of
test
(2)
Turn
(3)
Set the density
the tube voltage for 90kV, regulator for “0” position.
tube
(4)
Set
acrylic
: 5cm)
(5)
Perform
the
the
Block
phantom and
an exposure
Confirm
that
the
exposure
Measured time on the than l/120 second.
method
switch.
(thickness observe
the
time
is
within
time
exposure
-
101
counter
-
current
actual
In the case of the detector with multiple be taken to each combination of detecting should be done under 7 conditions (23 fields. (6)
-
,
L
L Fig.
on
Automatic exposure control
X-ray control
in
for
lOOmA and
front
of
exposure
detector.
fields, test should For example, test the
rejection
should
film
time.
detecting fields. 1 = 7) for
the
the
the
detector
with
limits.
be equal
to
or
less
3
C.
Test
(2)
reqE:lrea
eqci?men:
Es?osure
tine
-+a couLl,,r
: AK0
?lodel
I’D-I/
-3A
(Eiefer
time
counter
to
TOSF-:I3A
_AI??EIiDIX
0!33.?
D. Rejection
limit
The exposure time following limits. Measured
measured
zirne
on the
5 l/120
exposure
second
-
102
-
must
no:
exceed
the
AP?ENDIX A list
TEST
OS of
test
TOQLS
AND
equipment
required Test
Test
INSTRUMENTS
REQUIRED
is shown for
equipment
below. installation
equipment Remarks
Item
Type I Full
AC Voltmeter Oscilloscope High
Voltage
Analyzer
Tektronix
464
Machlett Dynalizer
II
Digital
V 0 M
Simpson
Primary
Voltmeter
ALCO PV-401
Acrylic phantom Water phantom
or equivalent
or equivalent
thickness
are necessary
300/6OOV
or equivalent
260
or
test equipment The following the above equipment.
range:
cnly
for
18Omm
HHS test
in addition
'Test equipment Remarks Item
Type . Keithley model 35020 with 96020 chamber . Victreen model 550 with 550-3, 550-4, 550-S or 550-6 probe. . Victreen model 555 with ImA, lOmA or 1DAS chamber
Dosemeter
Lead
300 x 300 x 2 (mm>
sh.eet
Al attenuation
200 x 200 x 19 (mm)
block -
Al attenuat:on Acrylic
Note:
These
phantom
are used
for
block
(Note)
thickness
(Note)
testing
200 x 200 x 38 (mm>
of Automatic
- 103 -
Exposure
5Omm
Control.
to
TOSHIBA
TOSHiBA
CORPORATlON
1385,SHIMOISHIGAMI,OTAWARA-SHI,TOCHiGl-KEN
324-8550,JAPAN
View more...
Comments