Hydrocal 1002-1003 Operation Manual English 01.2010
Short Description
Manual do Hidrocal modelo 1002-1003...
Description
HYDROCAL 1002-1003 Gas-in-Oil Analysis System with Transformer Monitoring Functions Manual for installation and operation
HYDROCAL 1002-1003: Manual for Installation and Operation
Edition 01.2010
Copyright MTE Meter Test Equipment AG All rights reserved. The contents of this manual are subject to change without notice. All efforts have been made to ensure the accuracy of this publication but MTE Meter Test Equipment AG can assume no responsibility for any errors or their consequences
Customs tariff number: 9027.1010
HYDROCAL 1002-1003: Manual for installation and operation
Page 1 of 97
Copyright MTE Meter Test Equipment AG All rights reserved. The contents of this manual are subject to change without notice. All efforts have been made to ensure the accuracy of this publication but MTE Meter Test Equipment AG can assume no responsibility for any errors or their consequences
Customs tariff number: 9027.1010
HYDROCAL 1002-1003: Manual for installation and operation
Page 1 of 97
Summary 1. General General............................. ............................................. ................................ ............................... ............................... ................................ .............................. .............................. .................. 4 2. Mounting of HYDROCAL 1002-1003 unit ........................................................................................ 5 2.1 Locations for installation of HYDROCAL 1002-1003 .................................................................. 5 2.2 Security precautions for installation ........................................................................................... 6 2.3 Preparation for installation ......................................................................................................... 8 2.3.1 Power supply for the unit ..................................................................................................... 8 2.3.2 Requested tools:............... tools:............................... .............................. .............................. ................................ ............................... ............................... ...................... ...... 8 2.4 Installation of the HYDROCAL 1002-1003 unit .......................................................................... 9 Note: For additional information for mounting and un-mounting, please refer to the document named: “HC 1002-1003 Special Additional Instructions Mounting & Un-mounting 01-2010.doc”. 01-2010.doc”. .. 11 2.5 Front view of HYDROCAL 1002-1003...................................................................................... 12 2.6 Rear side of HYDROCAL 1002-1003 unit ................................................................................ 13 2.7 Connections on the measuring and controller card .................................................................. 14 2.7.1 HYDROCAL 1002 .............................................................................................................. 14 3. Software on Hydrocal unit ............................................................................................................. 18 3.1. General information ................................................................................................................ 18 3.2. Operation of HYDROCAL 1002-1003 1002-1003 by its keyboard................... ......... .................... ................... ................... .................... ............. ... 19 3.2.1. Contrast Set-up ................................................................................................................ 19 3.3. Main menu.................................. .................................................. ............................... ............................... ................................ .............................. ............................. ............... 20 3.4. 3.5. 3.5.1 3.5.2.
Configuration of user settings ...................................................................................... 21 Display of stored „Gas-in-oil“ measuring measuring values...................... ........... ..................... ................... .................... .................. ....... 22 Graphic display ...................................................................................................... 22 Table presentation................... ......... .................... ................... .................... ...................... ..................... ................... ................... .................. ........ 23
History of alarms ......................................................................................................... 24 3.6. 3.6.1. Acknowledgment of alarms ............................................................................................... 24 3.7. 3.7.1.
Device set – up ................... ........ .................... .................. .................... ..................... ................... ................... .................... ................... ................... ............ .. 25 Menu Transformer data setup ............................................................................... 26
3.7.1.2.
Hot Spot calculation .................. ........ ..................... .................... .................. .................... .................... ................... .................... .................. ........ 27
3.7.1.3.
Ageing Rate / Loss of Life .................... ......... .................... .................. .................... .................... ................... .................... .................. ........ 29
3.7.2.
Configuration of analogue outputs................... ......... ..................... .................... .................. .................... .................... .............. ..... 29
3.7.3.
Configuration of analogue inputs (external sensors) .................... ......... ...................... .................... .............. ..... 30
3.7.4. 3.7.5.
Configuration of alarms alarms and digital outputs ................... .......... ................... .................... ................... ................... ............ .. 33 Unit Setup .................... ......... .................... .................. .................... ..................... ................... ................... .................... ................... .................... ............. 36
3.7.5.1.
Flange Heating Base Temperature .................. ......... .................... ..................... ................... ................... .................. ........ 37
3.7.5.2.
Flange Heating Base Temperature ................... ......... ................... ................... .................... ................... .................... ............. 37
3.7.5.3.
Device options .................... ........... .................. .................... ..................... ................... ................... .................... ................... ................... ............ .. 37
3.7.5.4.
Clock settings .................. ........ .................... ................... ................... ..................... .................... .................. .................... .................... .............. ..... 39 HYDROCAL 1002-1003: Manual for installation and operation
Page 1 of 97
3.7.5.5. 3.7.5.5.1.
Configuration of the communication ................................................................... 40
Alert SMS setup ........................................................................................... 41
3.7.5.5.2. Internal modem setup ................................................................................... 41 3.7.5.5.2.1 Analog Modem ......................................................................................................... 42 3.7.5.5.2.2 GSM Modem ............................................................................................................ 42 3.7.5.5.2.3 IP Modem ................................................................................................................. 43 3.7.5.5.3. RS 232 / RS 485 setup ................................................................................ 44 A. Point to Point ............................................................................................................................. 44 I. RS232 ................................................................................................................................ 44 II. RS485 ................................................................................................................................ 45 III. Modem ............................................................................................................................... 46 B. RS485-Bus ................................................................................................................................ 47 C. Bridge ....................................................................................................................................... 48 I Bridge Setup....................................................................................................................... 49 II RS485-Bus Setup ............................................................................................................... 50 D. MODBUS .................................................................................................................................. 51 Test / Debug Options at the HydroCal Device ............................................................................ 52 I Function, Register and Address Mapping ........................................................................... 53 MODBUS Communication Examples ............................................................................................. 55 I MODBUS/RTU via RS232 (using QuickMod) ..................................................................... 55 II MODBUS/ASCII via RS232 (using Modpoll) ....................................................................... 57 III MODBUS/RTU via RS485 (using QuickMod) ..................................................................... 57 IV MODBUS/RTU on RS485-Bus (using Modpoll) .................................................................. 59 4. Windows Software HydroSoft ................................................................................................. 62 4.1. Purpose of the program .......................................................................................................... 62 4.2. Remarks for use of this operation handbook ........................................................................... 62 4.3. Requirements to the PC.......................................................................................................... 62 4.4. Installation .............................................................................................................................. 63 4.4.1. Microsoft .NET .NET Framework 1.1 Redistributable........................................................ 63 4.4.2. Microsoft Data Access Components ................................................................................. 63 4.4.3. Adaptation of installation ................................................................................................... 64 4.5. Functions of HydroSoft program ............................................................................................. 64 4.6. Start of program ...................................................................................................................... 64 4.7. Preparation / rename of a new customer list ........................................................................... 65 4.8. Opening a new transformer station ......................................................................................... 66 4.9. Opening a file for a new transformer ....................................................................................... 66 4.9.1. Modem settings ................................................................................................................ 68 4.9.2. RS232 interface settings ................................................................................................... 68 4.10. Settings of HYDROCAL 1002-1003 unit ............................................................................... 68 4.10.1. Alert Setup ...................................................................................................................... 69 4.10.2. External sensors (see also chapter 3.7.3) ....................................................................... 72 4.10.3. Alert SMS Setup ............................................................................................................. 73 4.10.4. Input of DGA Level Adjustment ....................................................................................... 75 4.10.5. Analog Output Setup ...................................................................................................... 76 4.10.6. Digital Output Setup........................................................................................................ 77 HYDROCAL 1002-1003: Manual for installation and operation
Page 2 of 97
4.10.7. Device Setup (Clock settings) ......................................................................................... 78 4.11. Data transmission from / to the HYDROCAL 1002-1003 unit unit .................... ........... ................... .................... .................. ........ 79 4.12. View of results ...................................................................................................................... 81 4.12.1. Sensor selection ............................................................................................................. 81 4.12.2. Time range selection ...................................................................................................... 81 4.12.3. Result window ................................................................................................................ 82 4.12.4. Print and export of results ............................................................................................... 82 4.13. View of alarm status ............................................................................................................. 83 4.13.1. Alarm selection ............................................................................................................... 83 4.13.2. Time range selection ...................................................................................................... 83 4.13.3. Alarm window ................................................................................................................. 84 4.13.4. Status of alarms .............................................................................................................. 84 4.13.5. Print and export of results ............................................................................................... 84 4.14. Online Sensors ..................................................................................................................... 85 5. RS232 / RS 485 Wiring ................................................................................................................. 86 6. Vaisala sensors connection on HYDROCAL 1002 ........................................................................ 87 6.1 MMT318 MMT318 ............................ ............................................ ................................ ............................... ............................... ................................ .............................. ...................... ........ 87 6.2 MMT338 MMT338 ............................. ............................................. ................................ ............................... ............................... ................................ .............................. ...................... ........ 88 7. HYDROSOFT Scheduler .............................................................................................................. 89 7.1 General General............................... ............................................. .............................. .............................. .............................. ................................ ................................. ............................ ........... 89 7.2 Scheduler Scheduler .............................. .............................................. ................................ ............................... ............................... ................................ .............................. ...................... ........ 89 7.2.1 Configuring a Transformer for automatic History Data Update ..................... .......... .................... .................. ................ ....... 89 7.2.1.1 Connection Scheduler Dialog ......................................................................................... 89 7.2.2 Scheduler Status Visualization............... Visualization............................... .............................. .............................. ................................ ............................... ................. .. 90 7.2.2.1 Status Icons .................................................................................................................... 90 7.2.2.2 7.2.2.2 Example Example ............................. ........................................... ............................... ................................. .............................. .............................. ............................... ............... 91 7.2.3 Hydrosoft and the Windows Task Scheduler.................... ......... ...................... ..................... ................... ................... ..................... ............. .. 92 7.2.3.1 Command Line Options Options for Windows Task Scheduler .................... .......... ................... ................... .................... ............. ... 93 7.2.4 Internal Operation ................................................................................................................. 93 7.2.5. Data Directory Lock ............................................................................................................. 94 8. Technical data HYDROCAL 1002-1003 ........................................................................................ 95 9. Dimensional Drawings HYDROCAL 1002-1003 ............................................................................ 97
HYDROCAL 1002-1003: Manual for installation and operation
Page 3 of 97
1. General HYDROCAL 1002-1003 is an on-line Gas-in-Oil Monitoring System for continuous analysis of the dissolved gases hydrogen (H 2) and carbon monoxide (CO) of power transformers. HYDROCAL 1003 allows also in addition Analysis of moisture (H 2O) dissolved in transformer oil (absolute humidity in ppm). The possibility to connect additional sensors with an analogue output signal (0..10V or (04..20mA), e.g. for temperature or oil humidity complete the product. The HYDROCAL 1002-1003 system together with the Windows-based software HydroSoft is a powerful tool for detection and diagnosis of incipient faults f aults in power transformers. The present operating instructions are divided in the following sections: •
Instructions for mounting the HYDROCAL 1002-1003 1002-1003 unit unit (chapter 2)
•
Description of of software of the HYDROCAL 1002-1003 unit (chapter (chapter 3)
•
Description of Windows software software HydroSoft HydroSoft (chapter (chapter 4)
•
RS 232 / RS 485 Wiring (chapter (chapter 5) 5)
•
VAISALA sensors sensors connection connection on HYDROCAL 1002 1002 (chapter (chapter 6)
•
HYDROSOFT Scheduler (Chapter 7)
•
Technical data of HYDROCAL HYDROCAL 1002-1003 1002-1003 system (chapter (chapter 8)
•
HYDROCAL 1002-1003 dimensional drawings drawings (chapter (chapter 9)
HYDROCAL 1002-1003: Manual for installation and operation
Page 4 of 97
2. Mounting of HYDROCAL 1002-1003 unit 2.1 Locations for installation of HYDROCAL 1002-1003 To ensure the correct function of the HYDROCAL 1002-1003 the location of mounting is very important. For a typical transformer, three places are possible:
Pos. 1
On return side of cooling circuit: Please install the unit on a straight piece of tube on the cooler output between the cooler pressure , the unit must absolutely be and the main oil tank. In order to avoid negative pressure, mounted on the downstream side (after the bend) of the oil circulation pump. Remark: Choose this place only in case of continuous oil flux through the cooler.
Pos. 2
Flange below on the transformer tank: The flange must be at minimum height of 30 cm in order to allow installation and mounting of HYDROCAL 1002-1003 unit.
Pos. 3
Upper filling valve: It is possible to mount the HYDROCAL 1002-1003 unit in this position, however it may be difficult to operate and read the data of the unit. Remark: Should the unit be mounted on the wall of the cooler, then make sure that the unit is mounted below the oil level. l evel.
Pos. 4
Never install install the HYDROCAL 1002-1003 unit on top of the cooler cooler or on the suction face of the pump.
HYDROCAL 1002-1003: Manual for installation and operation
Page 5 of 97
2.2 Security precautions for installation The following security precautions must be strictly observed during the installation. Otherwise no warranty claims will be accepted. Never connect the sensor head to another HYDROCAL 1002-1003 electronic unit other than the delivered one. The components are delivered as a matched pair and have been calibrated together in our works. You will find the serial numbers on the units, on the packing material and /or on the data sheets. Please ensure you note down these serial numbers for future reference. Never remove the protection cap from the sensor before you are ready to mount the unit on the valve of the transformer . The protection cap protects the membrane of the sensor against dust and other foreign bodies. Additionally it protects the thread of the measuring head and/or adapters. Note: Use the protection cap also after dismounting of the unit. Never touch the membrane with fingers or other objects.
Never close the aeration hole for oxygenation.
Never use or stock chemicals with hydrocarbon contents near the measuring sensor. Never clean the measuring head with solvents, because its gases will strongly impair the measuring accuracy and the sensivity of the sensors.
HYDROCAL 1002-1003: Manual for installation and operation
Page 6 of 97
Never apply a negative pressure to the measuring head. Negative pressure can damage the unit.
Remove the HYDROCAL 1002-1003 unit from the valve every time you degas or refill the transformer. Never install the HYDROCAL unit on a valve that is closed for a long period of time. Remove the unit if this cannot be avoided. Never install the HYDROCAL 1002-1003 on the suction side of the cooling pump! Never install the HYDROCAL 1002-1003 unit on an elbow or curve of a tube. The unit can be damaged by negative pressure produced by turbulences in curves. Never install or remove the HYDROCAL 1002-1003 without opening first the aeration screw. Never use galvanized fittings or valves to mount the HYDROCAL 10021003 unit. Galvanized fittings or valves can react with oil, this will produce incorect measuring values. Never clean the HYDROCAL 1002-1003 unit with high pressure cleaning equipment. High pressure cleaning equipment used to clean transformers can seriously damage the HYDROCAL 1002-1003 unit.
HYDROCAL 1002-1003: Manual for installation and operation
Page 7 of 97
Never use a chaining key tool for mounting. Use of such tools can damage the surface of the measuring head and the threads. Please use an adjustable thin spanner for 1 ½“.
Attention!!! Faulty or inappropriate handling of the sensor head like perforation of the membrane, scratches on the membrane or damages by negative pressure, paint or solvent liquids make the warranty invalid.
2.3 Preparation for installation Before installation, a visual inspection of the membrane must be carried out made. Remove the protection cap and check the membrane. No cuts or tears should exist, the surface must be smooth. Under no circumstances should the membrane be touched with fingers or with any other means. 2.3.1 Power supply for the unit The HYDROCAL 1002-1003 unit requires an auxiliary power supply of 110 / 230V close to the unit. The power needed is max. 350 VA. 2.3.2 Requested tools: •
Long Allen key, size 4 for aeration screw
•
Long Allen key, size 5 to loose the measuring head from the unit (6 x screws).
•
Bucket, oil absorber, cleaning rag
•
Teflon-band
•
Drill / screw driver
•
Digital multimeter
Tin spanner or adjustable tin spanner of 1 ½“ for valve and fittings To seal the thread of the measuring head, only Teflon band must by used. Minimum 4 to 5 layers must be applied. •
HYDROCAL 1002-1003: Manual for installation and operation
Page 8 of 97
2.4 Installation of the HYDROCAL 1002-1003 unit
Attention!!! Mount the measuring head horizontally on the valve. The aeration screw must then be in the 12 o clock position. Please mount the HYDROCAL 1002-1003 unit on a valve with full opening (full bore valve) or a ball valve with an internal thread of 1½” (DIN ISO 228: G 1½). The HYDROCAL unit has a 1½’’ thread (Dimensional drawings, see. chapter 6). A good oil flux should exist on the mounting location. If necessary, use a reduction piece (adapter). In order to have a sufficient oil flux on the membrane, the diameter of the valve should be not less than 25 mm. For optimal working conditions, the valve diameter should not be more than 76 mm. The valve must be on earth potential.
Attention!!! The distance between the oil main stream and the measuring head shall not be more than 6 times the length of valve diameter. (Distance 150 mm at Valve-∅ 25 mm / Distance 228 mm at Valve-∅ 38 mm / Distance 300mm at Valve-∅ 50 mm / Distance 450mm at Valve-∅ 76 mm).
In case there exist strong vibrations on the mounting valve, the valve should be relieved by a support. Then mount the unit according to the following steps: •
Loose the knurled thumbscrews and remove the housing of the HYDROCAL 1002-1003 unit.
•
Disconnect the connection cable of the measuring head from the printed circuit.
•
Loose the measuring head from the unit.
Only remove the protection cap (slowly while turning) when you are ready to mount the measuring head!
HYDROCAL 1002-1003: Manual for installation and operation
Page 9 of 97
•
Clean the internal side of the installation valve.
•
Wind 4 to 5 layers of Teflon band on the thread of the adapter and mount it on the valve. Make sure that it is fixed mechanically.
•
Mount the measuring head on the adapter taking care. Not to use too much force. In case no adapter is required, wind 4 – 5 layers of Teflon band on the thread of the measuring head and mount it on the valve. Make sure that it remains fixed mechanically, but do not use too much force.
•
Turn the aeration screw to the 12 o clock position.
Attention!!! The following steps must be done according to the company procedures. Work carefully in order to prevent air entering into the transformer. Use a bucket to catch oil coming out. •
Unscrew the aeration screw 3 – 4 turns with Allen key No. 4. (Aeration is active).
•
Open the valve slowly until oil comes out on the proving hole.
•
As soon as no air bubbles sort on the proving hole, close the aeration screw. Then fully open the valve. Remove the oil from the measuring head external.
Do not use any solvents for cleaning! •
Make sure that no untaught places exist.
•
Mount the unit on the measuring head, use the 6 pieces of plain washers, lock washers and nuts.
HYDROCAL 1002-1003: Manual for installation and operation
Page 10 of 97
•
Connect the connection cable to the printed circuit. Check that all pins of the plug go straight into the socket; pushing the plug in carefully.
•
Use steel protected cables for power supply, alarm-cables and communication cables and connect them to the appropriate connectors.
•
Put the housing in correct position and secure it with the knurled thumbscrews.
•
Connect the power supply to the HYDROCAL 1002-1003 unit. The display shows the specified conditions confirming that the HYDROCAL 1002-1003 is functioning correctly.
•
The measuring cycle of the HYDROCAL 1002-1003 unit is synchronized on the hour mark, this means the first measuring value is available approx. 2, 22 or 42 min. after the hour mark and can be read on the display.
•
To check the heating function, touch carefully the heat plate after some minutes.
Attention, danger of burning!!! The heating plate should now be warm. Should the resistor heating not function correctly then check its contacts. •
After first start-up, in all cases date and time have to be checked on the HYDROCAL 1002-1003 display and eventually have to be adjusted manually in clock settings (chapter 3.7.5.3) or with HydroSoft (chapter 4.10.7).
Note: For additional information for mounting and un-mounting, please refer to the document named: “HC 1002-1003 Special Additional Instructions Mounting & Un-mounting 012010.doc”.
HYDROCAL 1002-1003: Manual for installation and operation
Page 11 of 97
2.5 Front view of HYDROCAL 1002-1003 On the front side of the HYDROCAL 1002-1003 unit the following operating elements are available.
[1]
LCD-display with 240 x 320 pixels
[2]
Screw mounting to mount the cover
[3]
Alphanumeric keyboard to enter numbers and text
[4]
Arrow-buttons to move in the menus, for data input and contrast set-up from the main display (Up an Down)
[5]
Enter button to confirm and stop functions and data
HYDROCAL 1002-1003: Manual for installation and operation
Page 12 of 97
2.6 Rear side of HYDROCAL 1002-1003 unit
[1] [2]
Measuring sensor with an internal thread of (DIN ISO 228: G1) or as an option (NPT 1½) Connection for power supply 88 VACmin … 276 VACmax Optional: 88 VDCmin … 390 VDCmax max. 350 VA, 50/60 Hz
[3]
System outputs for digital and analogue inputs and outputs Communication output for communication via analogue or GSM/GPRS-modem
[4]
Serial interface RS232 Connector: 5 pin plug Pin 1 Pin 2 Pin 3 Pin 4 Pin 5
TxD CTS RxD RTS GND
HYDROCAL 1002-1003: Manual for installation and operation
Page 13 of 97
2.7 Connections on the measuring and controller card 2.7.1 HYDROCAL 1002
[1]
Analogue inputs Connector: X1:1 X1:2 X1:3 X1:4 X1:5 X1:6 X1:7 X1:8 X1:9 X1:10 X1:11 X1:12 X1:13 X1:14 X1:15 X1:16
(Analogue (Analogue (Analogue (Analogue (Analogue (Analogue (Analogue (Analogue (Analogue (Analogue (Analogue (Analogue (Analogue (Analogue (Analogue (Analogue
input input input input input input input input input input input input input input input input
1 1 2 2 3 3 4 4 5 5 6 6 7 7 8 8
(-)) (+)) (-)) (+)) (-)) (+)) (-)) (+)) (-)) (+)) (-)) (+)) (-)) (+)) (-)) (+))
HYDROCAL 1002-1003: Manual for installation and operation
Page 14 of 97
[2]
Oil temperature sensor Connector: X2:1 X2:2
[3]
Measuring cell Connector: Flat plug 20 pin
[4]
Heating resistors Connector:
(Oil (Oil
temperature sensor (+)) temperature sensor (+))
X4:1 to X4:6 Connection of the heating resistors, depending from the supply voltage: With 230V in series With 110V in series - parallel
[5]
Power supply Connector: X5:1 (L) X5:2 (N) Note: The connection [5] is connected to a filter module from where the connection to the power supply 110VAC/230VAC and PE is made as follow: X5:3 X5:4 X5:5
[6]
(L) (N) (PE)
Digital outputs HYDROCAL 1002-1003: Manual for installation and operation
Page 15 of 97
Connector:
[7]
X6:1 X6:2 X6:3 X6:4 x6:5 X6:6 X6:7 X6:8 X6:9 X6:10 X6:11 X6:12 X6:13 X6:14 X6:15 X6:16 X6:17 X6:18 X6:19 X6:20 X6:21 X6:22 X6:23 X6:24
(Relay (Relay (Relay (Relay (Relay
X7:1 X7:2 X7:3 X7:4 X7:5 X7:6 X7:7 X7:8 X7:9 X7:10 X7:11 X7:
(Analogue
output 1) output 1) output 2) output 2) output 3) (Relay output 3) (Relay output 4) (Relay output 4) (Opto coupler output 1 (-)) (Opto coupler output 1 (+)) (Opto coupler output 2 (-)) (Opto coupler output 2 (+)) (Opto coupler output 3 (-)) (Opto coupler output 3 (+)) (Opto coupler output 4 (-)) (Opto coupler output 4 (+)) (Opto coupler output 5 (-)) (Opto coupler output 5 (+)) (Opto coupler output 6 (-)) (Opto coupler output 6 (+)) (Opto coupler output 7 (-)) (Opto coupler output 7 (+)) (Opto coupler output 8 (-)) (Opto coupler output 8 (+))
Analogue outputs Connector: output 1 (+))
(Analogue output 1 (-)) (Analogue output 2 (+)) (Analogue output 2 (-)) (Analogue output 3 (+)) (Analogue output 3 (-)) (Analogue output 4 (+)) (Analogue output 4 (-)) (Interface RS 485 (-)) (Interface RS 485 (+)) (Analogue Modem (a)) (Analogue Modem (b))
HYDROCAL 1002-1003: Manual for installation and operation
Page 16 of 97
[8]
Connector ST 301: Interface RS232 X8:1 X8:2 X8:3 X8:4 X8:5 X8:6 X8:7 X8:8 X8:9 X8:10
[9]
(RS232 (RS232 (RS232 (RS232 (RS232 (RS232 (RS232 (RS232 (RS232 (RS232
(free)) (free)) (TxD)) (CTS)) (RxD)) (RTS)) (free)) (free)) (GND)) (free))
GSM/GPRS-module / GSM antenna Connector Mini-BNC X9:1 X9:2 X9:3 X9:4 X9:5 X9:6 X9:7 X9:8
(GSM/GPRS-module / (GSM/GPRS-module / (GSM/GPRS-module / (GSM/GPRS-module / (GSM/GPRS-module / (GSM/GPRS-module / (GSM/GPRS-module / (GSM/GPRS-module /
HYDROCAL 1002-1003: Manual for installation and operation
GSN antenna) GSN antenna) GSN antenna) GSN antenna) GSN antenna) GSN antenna) GSN antenna) GSN antenna)
Page 17 of 97
3. Software on Hydrocal unit 3.1. General information The gas-in-oil analyzer HYDROCAL 1002-1003 is based on a microprocessor controlled measuring system. A 32 Bit „Coldfire“ processor with a clock frequency of 50 MHz is used. The software has been developed with the language „C“. The system has a static memory of 4 Mbyte for measuring data. This offers a 2 years time storage capacity for the measuring data. After this period, the oldest data will be overwritten by the latest values. When using additional external sensors, the storage capacity depends from the number of sensors and the time interval selected for storage. The following formula can be used to determine the storage capacity: 4 ⋅10
Memorydept h[days ] ≈
6
[ Byte]
8
[
] ∑ 288[ Byte / day ]⋅ Sensorstor agesn [1 Std ]
5664 Byte / day +
n=1
With one additional sensor with a time interval of 5 minutes the following calculation results: Memorydept h[days ] ≈
4 ⋅10
[
6
]
[ Byte] [
5664 Byte / day + 3456 Byte / day
]
≈
4 ⋅10
6
[
[ Byte]
9120 Byte / day
]
≈ 438[days ]
Also an additional reduction of the storage time takes place by the definition of alarms. Each change of alarm status requires 12 Bytes. Attention: To avoid any loss of data, it is proposed to read out regularly the measuring data by the Windows based software “HydroSoft”.
HYDROCAL 1002-1003: Manual for installation and operation
Page 18 of 97
3.2. Operation of HYDROCAL 1002-1003 by its keyboard The HYDROCAL 1002-1003 unit when in use can be operated by its built-in keyboard. For this its external housing must be removed. Loosen the knurled thumb screws on the front side and remove carefully the housing. The unit can then be operated by the keyboard as shown below: HYDROCAL 1002 HYDROCAL 1003
Following the description of functions of the keyboard: Moves the cursor horizontally Enter button: Selection of active menus or submenus Moves the cursor in the tables and contrast set-up from the main display Alphanumerical input of customer specific data
3.2.1. Contrast Set-up Contrast of HYDROCAL 1002-1003 display can be setup while pushing on up and down arrow keys from the main display.
HYDROCAL 1002-1003: Manual for installation and operation
Page 19 of 97
3.3. Main menu After connection of the HYDROCAL 1002-1003 unit to the power supply, the main menu appears on the display. The actual measuring values of concentration of hydrogen and carbon monoxide in the oil are displayed (In addition moisture (H 2O) if HYDROCAL1003 is used). If other sensors are connected or an un-acknowledged alarms exists, then the display will change every 5 seconds to these measuring values. HYDROCAL 1002 HYDROCAL 1003
In the main menu, the following sub menus can be activated. User menu „Gas-in-oil“-menu Alarm menu / history of alarms Configuration menu Transformer menu: not yet available External sensor menu: not yet available Remark: From the main display under Customer, Station, Transformer and Comment data, you find Serial • Number of the unit (For example: #35414 for the HYDROCAL1002 or #28822 for the HYDROCAL1003) and Firmware version implemented on the unit (For example: V 1.00-1191 for the HYDROCAL 1002 or V 1.00-1031 for the HYDROCAL 1003). To display immediately external sensor screen, you can push button “2”. • To display immediately alert status screen, you can push button “3” •
HYDROCAL 1002-1003: Manual for installation and operation
Page 20 of 97
3.4.
Configuration of user settings
This is the starting page for customer data, data of installations shown on the below page. The system asks for the password before any amendments can be made (Default password is 999):
First of all, you can define the application language while selecting it by the keyboard. You can choice between: English, German, French, Russian, Polish and Turkish. After entering the password by the alphanumeric keyboard and confirmation by the enter key
,
the menu customer data setup is opened (Default password is 999).
By operation of the arrow keys ( firmed by the enter
) the different fields are selected. Selection is con-
button. The customer data can be modified by the alphanumeric keyboard
and the arrow keys. These amendments are stored by selection of selection of the
button or are rejected by
. In the main menu (standard display) up to 26 characters can be entered. Only
the first 10 characters will be displayed.
HYDROCAL 1002-1003: Manual for installation and operation
Page 21 of 97
3.5.
Display of stored „Gas-in-oil“ measuring values
In the menu „Gas-in-oil“ an overview in the form of a bar chart with the composition of measured gases in the oil are shown. The values for H 2, CO and H2O indicate their actual measurement values of the total gas content. This means in this case shown below, that 1663ppm is H 2, 1663ppm is CO and 0.0ppm is H2O. HYDROCAL 1002
HYDROCAL 1003
The following values are available: Hydrogen Carbon monoxide Moisture (Only for HYDROCAL 1003) Temperatures:
- ϑ1 Temperature of oil on the sensor - ϑ2 Temperature in sensor area (ambient temp)
Graphic display 3.5.1 After selection of the measuring quantity, a graph displaying the trend over the time is shown
Following functions are available: Scroll graph to the right Scroll graph to the left Zoom-out with minimum resolution of 3 days / page HYDROCAL 1002-1003: Manual for installation and operation
Page 22 of 97
Zoom-in with maximum resolution of 8 hours / page Go to table presentation Exit
Table presentation 3.5.2. After selection of presentation in form of a table, the measured values are shown with time and current record number. The most recent measured value is on top of the table.
The following functions for tables are available Scroll up one line Scroll up one page Scroll down one page Scroll down one line Scroll one page up / down Go to trend menu Exit
HYDROCAL 1002-1003: Manual for installation and operation
Page 23 of 97
3.6.
History of alarms
In the menu history of alarms, all programmed alarms are presented in a table.
All alarms defined are shown together with their status. In the column „Date / Time” the last change of status is shown. The followings alarm statuses are available: •
OK, Symbol : No active alarm for this measuring value exists. o If a date exists: For this measuring value an acknowledged alarm existed in the past. o If no date exists: For this measuring value no alarm existed in the past.
•
High - Alarm active: o Symbol : The defined “Level 1“ for a High - Alarm has been exceeded at the date/time indicated. o Symbol
: The defined “Level 2“ for a High - High - Alarm has been exceeded in past. The
value has fallen below the limit at the time indicated, however it is still over the limit of “Level 1“. •
High - Alarm inactive, Symbol : The defined level for a High - Alarm has been exceeded in the past. The value has fallen below the limit at the time indicated.
•
High - High - Alarm active, Symbol exceeded at the time indicated.
: The defined level for a High - High - Alarm has been
High - High - Alarm inactive, Symbol : The defined “Level 2“ for a High - High - Alarm has been exceeded in the past. The value has fallen below “Level 1“ at the time indicated. In history of alarms, the following functions are available: •
Acknowledge alarm selected Scroll up one line Scroll down one line Show details of alarm selected: function not yet available Exit
3.6.1. Acknowledgment of alarms For confirmation of an alarm, a password is required (Default password is 999).
HYDROCAL 1002-1003: Manual for installation and operation
Page 24 of 97
3.7.
Device set – up
First of all, you can define the application language while selecting it by the keyboard. You can choice between: English, German, French, Russian, Polish and Turkish. Before any amendment of settings can be made, the system asks for the password (Default password is 999).
After input of password by the alphanumeric keyboard and confirmation with enter
button, the
device setup menu is opened:
The following configurations can be modified: Transformer data, e.g.: allocation of sensors to algorithms for calculation of life expectancy and input of correction factors Configuration of Analogue outputs Configuration of Analogue inputs (extern sensors) Configuration of alarms Configuration of the unit
HYDROCAL 1002-1003: Manual for installation and operation
Page 25 of 97
Menu Transformer data setup 3.7.1. Transformer data can only be programmed by Windows HydroSoft. This data is not necessarily required for operation of the unit. In the menu „Transformer Setup”, input of correction factor, Ho-Spot Calculation as well as Ageing Rate / Loss-of-Life Calculation are possible. Parameters can be set only on the device not in HydroSoft.
Input of correction factor 3.7.1.1. Results of laboratory analysis can be introduced in this menu. These can be used for calculation of correction factors for definition of the gas content.
By operation of the arrow keys (
) the different fields can be chosen. For the input
of results of laboratory analysis, first the date of analysis should be entered in the field „Sample Date“. Then the average value of gas concentration is calculated and presented. After the results of laboratory analysis have been introduced in the fields „Laboratory Value“ for H2 and CO, the button is activated the new correction factor is calculated.
This value can be used. The new values are stored and activated by the button
„Store and
exit“. For the following measured values the new correction factor will then be used. HYDROCAL 1002-1003: Manual for installation and operation
Page 26 of 97
3.7.1.2. Hot Spot calculation HYDROCAL 1002-1003 can allow Hot-Spot Temperature calculation. For this calculation, it is very important to know that system needs external sensor reference in addition to the standard initially purposed with HYDROCAL 1002-1003. These references are: minimum one of Load Current, Top oil Temperature as well as Cooling Stage (this last reference is not necessary required but can improve the calculation).
3.7.1.2.1. External Sensors After selection of the external sensor submenu, following screen will be displayed:
By operation of the arrow keys (
) the different fields can be chosen. First of
all, you must define each field with required information. •
For Load Current, you can select between “Load Current” and “None”
HYDROCAL 1002-1003: Manual for installation and operation
Page 27 of 97
•
For Top Oil Temperature, you can select between “Top oil Temperature”, “Ambient temperature”, “Hot-Spot temperature”, “ ϑ1” (Temperature of oil on the sensor), “ ϑ2” (Temperature in sensor area (ambient temp)) and “None”
For Cooling Stage, you can select between “Cooling Stage”, “Ageing rate”, “Loss of Life” and “None” If necessary, you have to define previously external sensor references as analogue input while push•
ing button (See chapter 3.7.3 for setup). Then, you must define calculation interval. Calculation interval in the field "Calculation interval" is edited by arrow keys and confirmed by the alphanumeric keyboard. The sample interval is made in seconds or by a button in days – hours – min - seconds. For a double point the After confirmation by the
button is used.
button, the value will be converted and displayed in days –hours – min
- seconds. At the end, the new references and calculation interval are stored and activated by the button „Store and exit“.
3.7.1.2.2.
Cooling stage #1
By operation of the arrow keys (
) the different fields can be chosen. First of
all, you must define each field with required information. Followings information must be entered: •
Cooling stage #1 condition o Sensor reading less / equal: in mA
Hot-Spot Calculation parameters o Rated Current: Ir in Amp o Winding Exponent: y o Gradient at rated Current: gr o Hot-Spot-Factor: H o Winding Time Constant: Tw in second Usually, all data can be known through transformer manufacture technical specifications or can be also estimated. •
At the end, conditions and parameters are stored and activated by the button
„Store and exit“.
For Cooling Stage #2 and Cooling Stage #3, please apply exactly the same procedure in order to setup required information.
HYDROCAL 1002-1003: Manual for installation and operation
Page 28 of 97
3.7.1.3. Ageing Rate / Loss of Life After selection of the Ageing Rate / Loss of Life submenu, following screen will be displayed:
By operation of the arrow keys (
) the different fields can be chosen. First of
all, you must define each field with required information. Followings information must be entered: •
Kind of Insulation (Paper): you can select between “Normal” and “Thermally Updated”
Hot-Spot Temperature Sensor: you can select between “None”, “ϑ1” (Temperature of oil on the sensor), “ϑ2” (Temperature in sensor area (ambient temp)), “Top Oil temperature”, “Ambient temperature” and “Hot-Spot temperature” If necessary, you have to define previously Hot-Spot Temperature Sensor as analogue input while •
pushing button
(See chapter 3.7.3 for setup).
At the end, setup is stored and activated by the button
„Store and exit“.
3.7.2. Configuration of analogue outputs The HYDROCAL 1002-1003 has 4 analog outputs for measured gas-in-oil concentration as well as for external sensors values. The output signal is 0-20 mA or 4-20 mA.
First a quantity / variable must be selected. The following functions are available: Scroll up / down marking Enter button: selection of active menus or submenus
HYDROCAL 1002-1003: Manual for installation and operation
Page 29 of 97
Store configuration of outputs and exit Exit
The following measuring quantities are available in the field "Sensor": •
H2 concentration
•
CO concentration
•
H2 Daily trend
•
H2 Weekly trend
•
CO Daily trend
•
CO Weekly trend
•
Temperature 1
•
Temperature 2
•
All quantities / variables defined under external sensors
The following analog outputs signals are available in the field "Mode": •
OFF
•
0…20 mA
•
4…20 mA
•
TEST: Allow simulating an output signal between 0 and 20 mA. The following values can be defined in the field "Sensor”: 0, 1, 3, 4, 5, 10, 15, 19 et 20 mA
3.7.3. Configuration of analogue inputs (external sensors) External sensors within standard outputs (0...10V or 0(4)...20mA) can be connected to the HYDROCAL 1002-1003 analogue inputs. The HYDROCAL 1002-1003 has 8 analogue inputs, the first 4 can be setup in 0(4) ... 20 mA and the last 4 in 0(4)...20 mA or 0...10V (Selectable by jumpers). Note: No power supply is available for external sensors.
First an input number must be selected. The following functions are available: Scroll up marking Scroll down marking Scroll up / down marking HYDROCAL 1002-1003: Manual for installation and operation
Page 30 of 97
Enter button : selection of active menus or submenus Exit As soon as an input is selected and made available for configuration are available:
, the following functions
Selection of different input possibilities for the input mode [Mode], and the unit [Unit] See below. Activation of input over alphanumeric keyboard for : [External Sensor #] • [Sample Interval] • [Min. Value & Max. Value] • [Display Min. & Display Max.]: • • [Smooth] Scroll up Scroll down Scroll up / down Enter button: selection of active menus or submenus Cancel input Store alarm configuration and exit Exit without storage
The following inputs signal can be selected by pressing the
button:
•
The first 4 can be set-up in 0(4)...20 mA or 0..10V (via jumper on electronic board)
•
The last 4 can be set-up in 0(4)...20 mA HYDROCAL 1002-1003: Manual for installation and operation
Page 31 of 97
The following unit s can be selected by pressing •
… (No unit)
•
ppm
•
ppm/Day
•
ppm/Week
•
ºC
•
%
•
mA
•
A
•
KA
•
V
•
KV
button:
The external sensors names in the field "External sensor #" are edited by arrow keys and confirmed by
and the alphanumeric keyboard.
The sample interval in the field "Sample interval" is edited by arrow keys and confirmed by and the alphanumeric keyboard. The sample interval is made in seconds or by a button in days – hours – min - seconds. For minus the After confirmation by the
button and for a double point the
button are used.
button, the value will be converted and displayed in days –hours – min
- seconds. For conversion of input units, the inputs „Minimum Value“ are made under the input signals 0V, 0 mA or 4 mA and „Maximum Value“ are made under the input signals 10V or 20mA. In addition, the display range can be limited by inputting introduction of „Minimum Display“ and „Maximum Display“. Remark: on following display, you could see 4 additional inputs named: •
#9: H.-Spot for Hot-Spot Temperature
•
#10: Ag.Rate for Ageing Rate
•
#11: L.o.life for loss of Life
•
#12: All of them are virtual sensors and any cases have to be considered as real analog inputs. However these values are displayed on the main display as external sensor (See Chapter 3.3: Main Menu “External sensor setup”).
HYDROCAL 1002-1003: Manual for installation and operation
Page 32 of 97
Configuration of alarms and digital outputs 3.7.4. The HYDROCAL 1002-1003 unit has 12 digital outputs for alarms. The first 4 (Out1 ...4) are 12V outputs. The „on“ status is 12V and the „off“ status is 0V. The next 8 outputs (Out 5..12) are optically divided switchers, with status „contact closed“ for “on“ and „contact open“ for “off“. You can have up to 16 possibilities to setup Alarms. Alarms can be triggered either by internal gas or temperature sensors or by external sensors. Configuration of alarms is shown on the following pictures:
First a quantity / variable must be selected. For this choice, the following functions are available: Configuration of SMS transmitting functionality Scroll up marking Scroll down marking Scroll up / down marking Select setting for alarm Select setting for configuration of alarm Exit The following measuring quantities are available: •
H2 concentration
•
CO concentration
•
H2 Daily trend
•
H2 Weekly trend
•
CO Daily trend
•
CO Weekly trend
•
Temperature 1
•
Temperature 2
•
All quantities / variables defined under external sensors
HYDROCAL 1002-1003: Manual for installation and operation
Page 33 of 97
After a measuring variable has been selected and opened with
for configuration, the following
functions are available: Selection of different input possibilities for the alarm mode (mode), Action (action1..2) and choice of the output (clamp1..2) Activation of input over alphanumeric keyboard for the values: Alarm limit (Level1..2), Delay (Delay1..2) Scroll up Scroll down Scroll up / down Cancel alarm Store configuration of alarm and exit Exit without storage Following alarm settings can be chosen by pressing the button
:
•
High: An alarm is triggered when Level 1 is exceeded.
•
High-High: An alarm in 2 steps is triggered. The first alarm is triggered when Level 1 is exceeded; the second alarm is triggered after exceeding Level 2.
•
Low: An alarm is triggered when a value falls below the limit defined in Level 1.
•
Low-Low: An alarm in 2 steps is triggered. Step 1 is alarm is triggered when the measured value falls below the limit defined in Level 1, the second alarm is triggered below Level 2 when the measured value falls below the limit defined in Level 2.
•
Low-high: An alarm in 2 steps is triggered. Step 1 is triggered when the measured value falls below the limit defined in Level 1, the second alarm is triggered after exceeding Level 2.
•
Range: An alarm is triggered when the measured value is outside the limits defined in Levels 1 and 2.
The alarm limits are edited by arrow keys and confirmed by the
button and the alphanumeric
keyboard. The quantity is defined by the unit. This is “ppm” for gas concentration and „°C” for temperatures and for external sensors the unit defined for the respective sensor (e.g. Volt, Ampere ...). The alarm limits must be within the corresponding measuring ranges. This is 0...2000 ppm for gases, 0...+100°C for temperatures and for external sensors the limit is defined for the respective sensor. Delays (Delay1...2) can be programmed to suppress alarms if the limits are exceeded only for a short time. Input of delay time is made in seconds or by a button in days – hours – min - seconds. For miHYDROCAL 1002-1003: Manual for installation and operation
Page 34 of 97
nus the
button and for a double point the
button are used. After confirmation by the
button, the value will be converted and displayed in days –hours – min - seconds. For example: if CO Alarm is setup as follow (delay 1: 10min and delay 2: 20min)
In the event of alarms, CO alarms will be activated as follow: Note: CO measurement interval is 20 min. Moreover, an alarm can be only activated on a record measurement (For CO: every 20 min). In fact, an alarm cannot be activated during 2 measurements interval.
The following actions can be selected by pressing the
button:
•
Log entry: An entry in the history of alarms is made; this has to be confirmed (see history of alarms).
•
Output: An output is switched on (If the measured value exceeded the limits in the field "Level 1 & 2"). The status „on“ described above is activated. If the measured value falls under the limit, the output is switched off again.
•
Out, hold: An output is switched on (If the measured value exceeded the limits in the field "Level 1 & 2"). The status „on“ described above is activated. If the measured value for this output falls under the limit, the output remains switched on. Only a manual deleting of alarm allows that the output is switched off again (see chapter 4.11 or 3.6).
•
Log & SMS: An entry in the history of alarms is made, this has to be confirmed (see history of alarms) and an SMS is sent to the number indicated. HYDROCAL 1002-1003: Manual for installation and operation
Page 35 of 97
•
Output & SMS: An output is switched on (If the measured value exceeded the configured limits in the field "Level 1 & 2"). The status „on“ described above is activated. If the measured value falls under the limit, the output is switched off again. In addition, an SMS is sent to the number indicated.
Out, hold & SMS: An output is switched on (If the measured value exceeded the configured limits in the field "Level 1 & 2"). The status „on“ described above is activated. If the measured value for this output falls under the limit, the output remains switched on. Only a manual deleting of alarm allows that the output is switched off again (see chapter 4.11 or 3.6). In addition, an SMS is sent to the number indicated. If it is possible to use the same output for different alarms and therefore, collection alarms can be defined. •
3.7.5.
Unit Setup
The following setup can be modified: Digital Output Test Flange Heating Base Temperature Device Options setup (Backlight display) Clock settings Communication settings Exit
HYDROCAL 1002-1003: Manual for installation and operation
Page 36 of 97
3.7.5.1. Flange Heating Base Temperature Temperature This is for Test of the 10 Digital Outputs.
3.7.5.2. Flange Heating Base Temperature Temperature The configuration of the Flange Heating Base Temperature is made in the following menu:
Input of Base Temperature is made by arrow key. The followings f ollowings temperature values can be defined: 20°C, 25°C, 30°C, 35°C, 40°C and 45°C.
3.7.5.3. Device options The configuration of the Backlight display as well as User Name and Password (String and Numeric Password) are made in the following menu:
HYDROCAL 1002-1003: Manual for installation and operation
Page 37 of 97
The following functions are available for Display Backlight Timeout: Scroll up / down marking Enter button: selection of active menus or submenus Store the configuration and exit Exit
Input of delay time is made in seconds or by a button in days – hours – min - seconds. For minus the button and for a double point the
buttons are used. After confirmation by the
button,
the value will be converted and displayed in days –hours – min - seconds. Note: 0 means that the screen will be always enlightened. Input of new User Name and Password are made by operation of the arrow keys ( ) the different fields are selected. Selection is confirmed by the enter data can be modified by the alphanumeric keyboard and the arrow keys. Remark:
button. The customer
•
User Name is used for Modem login. login. By Default, “EMH” “EMH” as “User Name” Name” is setup
•
String Password is used used also for Modem Modem Login. By default, “KWA” “KWA” as “String Password” Password” is setup
•
Numeric Password is is used for Device Setup menu, Customer Data Data Setup as well well as for communication from HydroSoft Software to the unit (Default password is 999).
Toggling of the main Display is made by operation of the arrow keys ( ent fields are selected. Selection is confirmed by the enter
) the differ-
button. You have followings possibili-
ties: •
“Standard” means that your Main Main Display will switched every 5 seconds
•
“Slow” means that your Main Display will switch every 15 seconds
•
“Very Slow” means that your Main display will switch every 25 25 seconds
These amendments are stored by selection of
button or are rejected by selection of the
.
Reset
HYDROCAL 1002-1003: Manual for installation and operation
Page 38 of 97
3.7.5.4. Clock settings The HYDROCAL 1002-1003 has a sub menu allowing configuration of time, date as well as clock
The following functions are available: Scroll up / down marking Enter button: selection of active menus or submenus Store the configuration and exit Exit Input of time in the field "System Time" is made in - hours - minutes - seconds. For a double point, "
used the
button.
Input of date in the field "System Date" is made in - year - month - days - . For minus the "
button
is used. The HYDROCAL 1002-1003 has a universal GMT clock in order to modify the t ime zone.
The time zone zone setup in the field "Clock offset" offset" is made by button. The time zone values are between -6 and +6 hours. Variable „Custom" allows manual setup in the field on the right side. Input of time is made in - hours - minutes -. For a double point, used the
button.
Note: "None“ means that GMT clock is not activated. Input of time mode in the field "Daylight Saving Time Mode" is made by
button:
•
MET (Middle (Middle European European Time) adjusts the official local local time forward, forward, one hour from from its official standard time for the duration of the spring and summer months.
•
NONE means that the function is not activated
HYDROCAL 1002-1003: Manual for installation and operation
Page 39 of 97
These amendments are stored by selection of
button or are rejected by selection of the
.
Configuration of the communication 3.7.5.5. The configuration of the communication is made in the following menu:
The following functions are available: Send a test SMS (not yet available) Alert SMS setup Internal modem setup RS 232 / RS 485 setup Status information activation (not yet available) Scroll up / down marking Enter button: selection of active menus or submenus Exit Note: on the main display, you have the currently communication status.
HYDROCAL 1002-1003: Manual for installation and operation
Page 40 of 97
3.7.5.5.1. Alert SMS setup The configuration of the alert SMS transmission is made in the following menu if HYDROCAL 10021003 is equipped within an on-board GSM. This function allows sending SMS in correlation with beforehand definite alarms for gas-in-oil concentrations, temperatures as well as for external sensors values.
Input of operator name is made in the field "Service Center". The system has 5 possibilities to send SMS in the field "Recipient 1-5". Input of telephone number is made by alphanumeric keyboard. A message in the field "Message" can also be pre-write in order to be sent to the recipient in the event of alarms. Note: The SMS mode (in Alarm set-up) must be selected to use this function. Moreover, GSM or Analogue modem must be also installed on HYDROCAL 1002-1003. These amendments are stored by selection of
button or are rejected by selection of the
.
3.7.5.5.2. Internal modem setup The modem configuration is made in the following menu if HYDROCAL 1002-1003 is equipped within an on-board GSM or Analogue modem.
The following functions are available: Scroll up / down marking Enter button: selection of active menus or submenus Store the configuration and exit
HYDROCAL 1002-1003: Manual for installation and operation
Page 41 of 97
Exit Input of Modem type in the field "modem Type" is made by
button:
•
“NONE” means that the function is not activated
•
“Analog (Internal)” means that an analog modem is installed on the unit and you want to use it
•
“GSM (Internal)” means that a GSM modem is installed on the unit and you want to use it
•
“IP” (internal) means that a IP modem is installed on the unit and you want to use it
3.7.5.5.2.1 Analog Modem If you select an Analog Modem, following screen will e displayed:
Input of Number of Ring before answer in the field "No. of RINGs be fore Answer:" is made by button. Input of Timeout in the field "Idle Hang-Up Timeout [s]" is made in - hours - minutes - seconds. For a "
double point, used the
button.
3.7.5.5.2.2 GSM Modem If you select a GSM Modem, following screen will e displayed:
Input of Number of Ring before answer in the field "No. of RINGs be fore Answer:" is made by button. Input of Timeout in the field "Idle Hang-Up Timeout [s]" is made in - hours - minutes - seconds. For a "
double point, used the
button.
Input of Pin code in the field “PIN” is made by meric keyboard and the arrow keys. These amendments are stored by selection of
button. Pin code can be modified by the alphanubutton or are rejected by selection of the
.
HYDROCAL 1002-1003: Manual for installation and operation
Page 42 of 97
3.7.5.5.2.3 IP Modem If you select an IP Modem, following screen will e displayed:
While activating DHCP, it means that IP address of the Hydrocal in DHCP server on the network will be automatically provided, only a port number has to be provided. This one has to be also set up on HydroSoft Software. If DHCP is not enabled need IP address, subnet mask, default gateway and port number in the following menu can be entered.
The input of the idle on the field “Hang up timeout (s)", TCP port number in the "TCP Port Number", the IP address in the "IP Address, subnet mask in the field" Network mask" and the standard gateway in the "Standard Gateway" with the Enter
key and the numeric keypad.
The baud rate for the RS232 interface must be using internal IP-modems to 57600 baud (see following chapter), IP modem IP is connected to the internal RS232 interface.
HYDROCAL 1002-1003: Manual for installation and operation
Page 43 of 97
3.7.5.5.3.
RS 232 / RS 485 setup
The serial interfaces configuration is made in the following menu.
This section illustrates different ways to communicate with HydroCal devices. There are the following main connection / communication scenarios: Host (PC) with one HydroCal device via RS232, RS485 or via Modem • Point-to-Point Host (PC) with many HydroCal devices via RS485 bus • RS485-Bus Host (PC) with many HydroCal devices via RS485 bus using a HydroCal bridge • Bridge SCADA System with online monitoring of one or more HydroCal devices • MODBUS General The following sections describe how to configure the Hydrocal devices and the HydroSoft software. After a short description screen shots of HydroCal and HydroSoft illustrate the setup. Some descriptive comments and hints may follow the screen shots. The setup of the HydroCal has to be carried out manually on the device.
A. Point to Point A point-to-point connection / communication is the simplest setup. The host (PC) is connected directly to one HydroCal device. The connection can use the RS232, the RS485 or the modem. The default configuration for a HydroCal device assumes a point-to-point connection. The same applies to HydroSoft. I. RS232 The RS232 is used for short connection distances (less than 10m). It provides full duplex and handshaking.
HYDROCAL 1002-1003: Manual for installation and operation
Page 44 of 97
HydroCal
HydroSoft
The setup is very simple. Just assure that the baudrate is matching and that the correct COM port is chosen in HydroSoft. II. RS485 The RS485 can be used for long connection distances (up to 200 m). It is implemented as a two-wire interface with half-duplex and a soft-handshake.
HydroCal
HydroSoft
Assure that the baud rate is matching and that the correct COM port is chosen in HydroSoft. Also check the RS485 checkbox to indicate the half-duplex connection. A RS485 adapter might require some special setup of the other checkboxes.
HYDROCAL 1002-1003: Manual for installation and operation
Page 45 of 97
PC RS485 Adapters Connecting a PC to the RS485 of a HydroCal device usually requires an adapter (interface converter). Depending on the properties of the adapter in use, the RS485 related check boxes in HydroSoft must be set accordingly. For communication with a HydroCal device only the two-wire half-duplex mode is of interest. This mode requires a kind of send/receive-control which is either ART (Automatic Receive Transmit control) or a manual control (e.g. via the RTS signal). Further the adapter may echo the send out data. The two checkboxes in the HydroSoft Connection dialog must mirror the properties of the adapter: •
Ignore Echo Must be checked, if the adapter echo the send out data
•
RTS Send Control Must be checked, if the adapter requires a manual send/receive-control via the RTS signal (this assumes a RS232 to RS485 adapter is in use)
Answer Delay If an RS485 adapter with RTS send control is used it may be necessary to set up some answer delay in the HydroCal device. If everything is configured right and still the communication does not work this may be due to problems with the send/receive-control. If the adapter does not release the transmitter quick enough, the answer of the HydroCal may be damaged and thus unrecognized. III. Modem The modem allows a remote connection using the phone network. From a communication point of view the connection via modem is very similar to the point-to-point connection via RS232. The connection supports full duplex, but may be limited in speed.
HydroCal
HydroSoft
The HydroCal modem configuration should have been carried as the device left the production. For HydroSoft select "Connect via:" - "Modem" and fill in the modem connection fields as described in the user manual, i.e. enter user name, password and phone number and select the modem to use for the communication. HYDROCAL 1002-1003: Manual for installation and operation
Page 46 of 97
B. RS485-Bus Using the RS485 interface it is possible to connect more the one HydroCal to a host. Nonetheless it is only possible to communicate with one device at a time. The illustration below shows the schematic topology. Remark: The host (PC) must only be present once (the host connected via the TCP/IP adapter is only present to show an alternate way to connect to the RS485 bus, as it is offered by third party network-to-serial bridges).
HydroCal
HydroSoft
On HydroCal device: •
Configure the device as "Bus Node" HYDROCAL 1002-1003: Manual for installation and operation
Page 47 of 97
•
Set an unique address for each device on the RS485 bus (1..254). Note: Communication with HYDROCAL 1002-1003 can be made right now with up to 32 units
•
Setup "Answer Delay" as required by the RS485 adapter in use (see also chapter: “PC RS485 Adapters” and “Answer delay”)
•
MODBUS must be off
•
Configure RS485 baudrate
In the HydroSoft connection dialog: •
Use "Connect via:" "Com-Port"
•
Select RS485 baudrate chosen for the devices
•
Check RS485
•
Set RS485 related checkboxes are required by the adapter (see also: “Answer delay”)
•
Enter correct bus address
C. Bridge The (RS485-Bus-) bridge is an extension to access a range of HydroCal devices on a RS485 bus. In this setup the host (PC) is not directly connected to the RS485 bus but is using one of the HydroCal devices as communication gateway. The bridge functionality can be used for an RS232 or Modem primary connection as indicated by the illustration below.
The setup of all HydroCal devices must be according to a standard RS485 bus setup, with one exception. The HydroCal that shall serve as the communication gateway must be configured as a (bus-) bridge.
HYDROCAL 1002-1003: Manual for installation and operation
Page 48 of 97
I Bridge Setup HydroCal
HydroSoft
On HydroCal device : •
Configure the device as “Bus Bridge”
•
Set an unique address (1..254). Note: Communication with HYDROCAL 1002-1003 can be made right now with up to 32 units
•
Setup “Answer Delay” as required by the RS485 adapter in use
•
MODBUS must be off
•
Configure RS232 baud rate
In the HydroSoft connection dialog: •
Use “Connect via:” “Com-Port”
•
Select RS232 baud rate chosen for the bridge devices
•
Enter correct bus address
Bridge via Modem The bridge functionality can also be activated, if the primary connection uses the modem. The corresponding setup for the HydroCal and HydroSoft must be carried out as already described in the point-to-point section and is not repeated here (see III).
HYDROCAL 1002-1003: Manual for installation and operation
Page 49 of 97
II RS485-Bus Setup HydroCal
HydroSoft
On HydroCal device : •
Configure the device as “Bus Node”
•
Set an unique address for each device on the RS485 bus (1..254). Note: Communication with HYDROCAL 1002-1003 can be made right now with up to 32 units
•
Setup “Answer Delay” as required by the RS485 adapter in use
•
MODBUS must be off
•
Configure RS485 baud rate
In the HydroSoft connection dialog: •
Use “Connect via:” “Bridge”
•
Select the transformer that is equipped with the bridge HydroCal in the “Device” combo box.
•
Enter correct bus address
HYDROCAL 1002-1003: Manual for installation and operation
Page 50 of 97
D. MODBUS If one or more HydroCal devices shall be integrated into a SCADA System for online monitoring, the MODBUS communication setup might be applicable. MODBUS is currently not supported by HydroSoft, but the setup is shown here for completeness
Since Version V1.00-B0884 of the firmware the MODBUS protocol is available. This section describes the device setup for MODBUS and the MODBUS function code and register mapping. As a reference example, the setup of an ADAM-4572 1-port Modbus® to Ethernet Data Gateway by Advantech (www.advantech.com) is shown below. HydroCal
ADAM-4572
On HydroCal device: •
Activate MODBUS using the RS485 and the desired mode (RTU/ASCII) HYDROCAL 1002-1003: Manual for installation and operation
Page 51 of 97
•
Set an unique MODBUS address for each device (1..247). Note: Communication with HYDROCAL 1002-1003 can be made right now with up to 32 units
•
Set "Baudrate", "Data Bits", "Parity" and "Stop bits"
•
Setup "Answer Delay" if required (i.e. if communication problems occur)
Remark: If the MODBUS protocol on the RS 232 interface is active: * If the data flow control (RTS / CTS) is not available * If the internal modem is not available The following parameters in the configuration menu can be entered •
•
Mode: This is a toggle button to select the transmission mode (RTU / ASCII) and the serial interface to use for MODBUS. Available entries are: - Off (i.e. the proprietary HydroCal protocol is used) - MODBUS RTU via RS485 - MODBUS ASCII via RS485 - MODBUS RTU via RS232 - MODBUS ASCII via RS232
•
Address: This is an edit field to enter the device address. For MODBUS, values in the range from 1 to 247 are allowed. For the proprietary HydroCal protocol the addresses can be configured from 1 to 254. Note: Communication with HYDROCAL 1002-1003 can be made right now with up to 32 units
•
Baud rate, Data Bits, Parity and Stop Bits These toggle buttons allow adjusting the coding format of the serial interface for MODBUS protocol. Remarks: o Change of the transmission mode does not set the coding format parameters to the default values for the selected mode (RTU: 19200,8,E,1; ASCII 19200,7,E,1). The right configuration has to be configured manually. o The proprietary HydroCal protocol allows only configuration of the baud rate. The settings of the Data Bits, Parity and Stop Bits toggle buttons are ignored and the configuration is fixed to 8-N-1.
Answer delay Test / Debug Options at the HydroCal Device In the lower left corner of the configuration screen there is a small MODBUS statistics line with the following information: 0 ; count of all MODBUS messages -> 0 ; count of MODBUS messages for this device •
HYDROCAL 1002-1003: Manual for installation and operation
Page 52 of 97
0x8000, alert #1 -> 0x0001). By writing a '0'-Bit at the corresponding position a probably active alert can be reset. Writing '0000' resets all alerts. Switch MODBUS protocol off MODBUS off
2000
0 .. 1
40001 .. 40002
1000 2000
41001 42001
HydroCal
A value is returned within 2 MODBUS register as a 32-bit unsigned integer in big endian format. Holding Register Test register #1, #2
It is now possible to switch the MODBUS protocol off via MODBUS. This is done by writing 0x00FF (like OFF). Holding Registers 16-Bit test register #1, #2
Alert status bit pattern #16 .. #1 MODBUS protocol status Remark: The reading is always 0x0000. If the MODBUS status is OFF it is not possible to read this status via MODBUS.
Currently the following assignment of sensor numbers applies: 1 gas sensor H2 2, 3 - reserved 4 gas sensor CO 5 temperature sensor (measurement chamber) 6 temperature sensor (oil) 7 humidity/moisture (only HydroCal 1003 devices) 8 - reserved 9 daily trend gas sensor H2 10, 11 - reserved 12 daily trend gas sensor CO 13 weekly trend gas sensor H2 14, 15 - reserved 16 weekly trend gas sensor CO 17 .. 24 external analog sensors (0..10V, 0/4..20mA inputs) 25 .. 28 virtual sensors (hot spot, ageing rate, etc., since firmware V1.00-B0844)
HYDROCAL 1002-1003: Manual for installation and operation
Page 54 of 97
MODBUS Communication Examples This chapter describes working test setups for MODBUS communication. The following hardware and applications were used to test the MODBUS communication with HydroCal. Both test application are free for download. The hardware is not the only hardware that is working with HyroCal. The devices are just an example. Applications (for Microsoft® Windows): •
QuickMod Modbus® scanning tool (Windows based) AzeoTech, Inc. (www.azeotech.com)
Modpoll Modbus® Polling Tool (Command Line Tool) FOCUS Software Engineering (www.focus-sw.com) Hardware: •
•
Roline RS232C / RS485 Converter Roline, IC-485S (12.02.1028), IC-485SI (12.02.1029), (www.rotronic.ch)
•
ADAM-4572 1-port Modbus® to Ethernet Data Gateway Advantech (www.advantech.com)
I MODBUS/RTU via RS232 (using QuickMod) This test of HydroCal MODBUS/RTU via RS232 was executed using the QuickMod application. Connection, setup and results are shown below. •
Physical Connection
•
HydroCal and QuickMod Connection Setup:
•
HydroCal can be set up without any answer delay (due to full duplex on RS232).
•
The timeout for Quickmod can be short. The answer of HydroCal is received within less than 50 ms (see QuickMod Monitor screen shot below).
HYDROCAL 1002-1003: Manual for installation and operation
Page 55 of 97
•
HydroCal and QuickMod Data Screens
•
The QuickMod Tag # field must be filled with the address 1016 to query the HydroCal external sensors 1..8 (address=1016..1023, register=1017..1024).
•
QuickMod Monitor
HYDROCAL 1002-1003: Manual for installation and operation
Page 56 of 97
II MODBUS/ASCII via RS232 (using Modpoll) This test of HydroCal MODBUS/ASCII via RS232 was executed using the Modpoll application. Connection, setup and results are shown below. •
Physical Connection
•
HydroCal Setup and Data Screen Dump
•
HydroCal can be set up without any answer delay (due to full duplex on RS232).
•
Modpoll Result Screen Dump
•
The Modpoll utility expects the start reference to be set to the register 1017 to query the HydroCal external sensors 1..8 (address=1016..1023, register=1017..1024).
III MODBUS/RTU via RS485 (using QuickMod) This test of HydroCal MODBUS/RTU via RS485 was executed using the QuickMod application and the Roline RS2327RS485 adapter (12.02.1028). Connection, setup and results are shown below. •
Physical Connection
•
The RS232-to-RS485 adapter is configured for 2-wire (referred to as MultiDrop Half Duplex in the adapter manual). The terminals pairs T+, R+ and T-, R- are short connected (terminal block connectors 1+4, 2+3). HYDROCAL 1002-1003: Manual for installation and operation
Page 57 of 97
•
The RS232-to-RS485 adapter is configured to use RTS send / receive control (SW2 in middle position to TxRTS/RxRTS, with the underlined text indicating that the signal in not active).
•
HydroCal and QuickMod Connection Setup:
•
HydroCal is set up with an answer delay of 100 ms to give enough time for QuickMod to toggle the RS232-to-RS485 adapter from send to receive.
•
In the QuickMod configuration dialog, setting RTS Control to Toggle will operate the RS232-toRS485 adapter in the desired way.
•
HydroCal and QuickMod Data Screens Dumps
HYDROCAL 1002-1003: Manual for installation and operation
Page 58 of 97
•
QuickMod
Monitor
IV MODBUS/RTU on RS485-Bus (using Modpoll) This test of HydroCal MODBUS/RTU on RS485 bus was executed using the Modpoll application and the ADAM-4572 Ethernet Data Gateway. It demonstrates the ability to connect multiple HydroCal devices to an RS485 bus. Connection, setup and results are shown below. •
Physical Connection / RS485 Bus Topology
HYDROCAL 1002-1003: Manual for installation and operation
Page 59 of 97
•
HydroCal (only device #90) and ADAM-4572 Setup:
•
HydroCal is set up without an answer delay. Each HydroCal on the bus has its own unique address (#90 .. #93, #0 reserved for the bus master).
•
The timeout for ADAM-4572 can be short. The answer of HydroCal is received within less than 100 ms and the ADAM-4572 uses ART that does not require additional delay.
•
Modpoll Result Screen Dump (partial)
HYDROCAL 1002-1003: Manual for installation and operation
Page 60 of 97
HYDROCAL 1002-1003: Manual for installation and operation
Page 61 of 97
4. Windows Software HydroSoft 4.1. Purpose of the program HydroSoft is a program compatible with MS Windows for use with HYDROCAL 1002-1003 to provide a comprehensive transformer monitoring system. HydroSoft not only reads and processes the data of HYDROCAL 1002 but can also used in the configuration of the HYDROCAL 1002 unit. The program is compatible for use with operating systems MS W indows 2000 / XP (32 Bit operating systems). HydroSoft offers the following functions: Reading data from HYDROCAL 1002-1003 unit Data processing Export of data to other standard windows programs Storage on data mediums, read from data medium, archiving of measuring data Configuration of analogue inputs of HYDROCAL 1002-1003 unit Configuration of alarms and respective alarm outputs of HYDROCAL 1002-1003
4.2. Remarks for use of this operation handbook The following operation handbook explains in details the installation and operation of HydroSoft on your PC. It contains the following sections: Program Installation and Start up Program functions General rules for use as transformer monitoring system Export and storage of data
4.3. Requirements to the PC The PC must fulfill the following requirements:
Pentium Processor clock ≥ 300MHz
Free disc space for installation on harddisk ≥ 20 MB, additional memory space for data storage
RAM memory ≥ 256MB Operating system: Windows 2000 / XP CD-ROM drive Interface RS232C Signals used are RxD, TxD, GND Serial interface cable 1:1
USB ↔ RS232C interface - converter; only converters with all RS232-signals can be used. .NET Framework 1.1 redistributable must be supported Microsoft Data Access Components 2.6 (MDAC) or new version must be supported
HYDROCAL 1002-1003: Manual for installation and operation
Page 62 of 97
4.4. Installation All components for installation of HydroSoft software package are available on the CD. All components of the software should be installed by the installation program. Put the CD into the drive and start by Setup.exe. Now the installation program guides you through the different steps of the installation. After installation the program HydroSoft can be started. Should the installation program not be executed and ended, not all components will be installed. The missing components will have to be installed manually (chapter 4.41 and 4.42). Please act as described following. First install the .NET Framework 1.1 4.4.1. Microsoft .NET .NET Framework 1.1 Redistributable Install Microsoft .NET Framework 1.1 Redistributable by: dotnetfx.exe and follow the instructions.
4.4.2. Microsoft Data Access Components Install if necessary Microsoft Data Access Components by execution of: mdac_typ.exe and follow the instructions.
HYDROCAL 1002-1003: Manual for installation and operation
Page 63 of 97
4.4.3. Adaptation of installation If required, you can place the HydroSoft data ("C:\Programme\MTE\HydroSoft\data") into another root, e.g.: "C:\Documents and settings\All Users\HydroSoft\data" Important: The file "root.xml" (contained in the installation data) must exist in the correspondent new root. The data will be stored in the file „Hydrosoft.xml“. In case this file does not exist, the system tries to find a file „Data“ in the installation list. In case this file exists, it will be used as a data list. If this file does not exist, the correspondent file in the menu „File“ under „Set Data Location...“ must be selected.
4.5. Functions of HydroSoft program The program HydroSoft works together with the HYDROCAL 1002-1003 unit. The main functions of the program are: Reading of data from HYDROCAL 1002-1003. The data can be read out through the interface RS232C or from a modem integrated in the HYDROCAL 1002 unit (option). Processing and presentation of measuring data – data can be presented in form of trend graphics or tables (history, see. chapter 4.12). Tables can be exported to MS Excel. Graphs can be printed directly. Configuration of HYDROCAL 1002-1003 units.
4.6. Start of program In case you have introduced a link to the start menu, press Start in the task bar. Open HydroSoft in the submenu “Programs”. It is also possible to start the program by HydroSoft.exe If pre-settings are used; the program is to be found under "C:\Programme\MTE\HydroSoft\ “. First of all, you must to select appropriate application language, by „Extra->Language”
Followings languages can be selected: English, German, French, Spanish, Russian, Chinese and Turkish.
HYDROCAL 1002-1003: Manual for installation and operation
Page 64 of 97
4.7. Preparation / rename of a new customer list Information is presented in a “tree” format with the customer being at the top. Within a customer, transformer stations are defined and within a transformer stations, transformers followed by HYDROCAL 1002-1003’s are defined. See below.
A customer list is opened automatically by the program. It can be renamed, there are two possibilities: •
Either by the menu shown when pressing the right mouse button
•
Or by „File->Customer->Rename“
HYDROCAL 1002-1003: Manual for installation and operation
Page 65 of 97
4.8. Opening a new transformer station After the customer list has been renamed, a file for a new transformer station can be opened as indicated under chapter 4.7 „File->Customer->New Station“ after selection of the customer list. The number of stations is not limited.
After opening a file for a new transformer station, the address of the station can be entered. By double clicking on one of the four „Property” fields, address data can be inputted.
4.9. Opening a file for a new transformer To open a new transformer file, an existing transformer station must be selected. Then, similarly as a transformer station, a new transformer element can be opened and please select which HYDROCAL version you are using (HYDROCAL 1002, 1003, 1005 or 1008).
The number of transformers for a transformer station is not limited. Then additional details/technical data can be introduced.
HYDROCAL 1002-1003: Manual for installation and operation
Page 66 of 97
In addition to the existing fields •
Name
•
Manufacturer
•
Type
•
Year
•
Primary voltage
•
Secondary voltage
•
Prim. current
Sec. current Additional lines can be added by a right mouse click or by the menu „Edit->New”. The settings can be put back by „Revert”. Amendments are stored by „Save” or by selection of another element in the tree diagram. A transformer has the following sub-information: History: Measured data is presented in form of trend charts or tables. (see chapter 4.12) Alarms: Status of alarms are presented in form of trend charts and tables. Alarms can also be acknowledged. (see chapter 4.12) Connection/Communication: Settings can be defined. (see chapter 4.9.1 and 4.9.2) Online Sensors: A direct/online visualisation of the measuring data can be made HYDROCAL 1002-1003 settings plus alarms and external sensors can be configured. (see chapter 4.10) After opening a transformer file and inputting of its data, connection/communication settings should be inputted. •
It must first be defined, whether the connection shall be via modem (option) or by an interface RS232 (Com-Port). Select the „Radio“ – Selection field concerned.
HYDROCAL 1002-1003: Manual for installation and operation
Page 67 of 97
4.9.1. Modem settings During establishment of a modem connection, the „Login“ – Name and the password is requested. The name EMH and the password “KWA” are always used as defaults. In the box „Device“ a modem installed on the PC can be selected for communication. Finally the phone number of the telephone line connected to the HYDROCAL unit must be introduced. Attention: When using a GSM-Modem with HYDROCAL 1002-1003, it must be confirmed that a connection to an analogue modem is possible. The network operator should be able to confirm this. Otherwise a GSM modem (e.g. a cell phone) must be used on the PC side too. When connection is via a GSM Modem it is possible that communication during data reading is interrupted due to a communication faults on the GSM network. Under these circumstances a waiting time of 5 minutes is recommended before re-sending of data. This time is required to restart the modem in the HYDROCAL 1002-1003 and to make a new connection to the GSM network. 4.9.2. RS232 interface settings For local communication over an RS232 interface, the interface number of the PC must be selected. Transmitting speed is 19200 Baud; this must be entered into the field Baud rate.
4.10. Settings of HYDROCAL 1002-1003 unit Settings for alarms and external sensors can be programmed by HydroSoft.
The file HYDROCAL contains the following information, these are read out during a connection: Serial number of HYDROCAL 1002-1003 unit Firmware version of HYDROCAL 1002-1003 unit Contains the list where data are stored on the PC Date of last read data Customer 1st line Customer 2nd line Customer 3rd line Customer 4th line Numerical password to modify settings (Default password is 999). String Password to modify Modem Login (Default String Password is KWA). User Name to modify settings (Default User name is EMH).
HYDROCAL 1002-1003: Manual for installation and operation
Page 68 of 97
4.10.1. Alert Setup In the menu settings to alarms can be programmed by HydroSoft. Alarms can be triggered by internal sensors (H 2, CO, H2O, Oil temperature, temperature of sensor) or by external sensors (see chapter 4.11). The recommended temperatures for fault free operation of HYDROCAL 1002-1003. Temperature for sensors and oil below 30°C and over 60°C are generally not to expected and indicate a problem in the heating circuit. If temperatures are outside this range, then external (air) temperature and oil temperature must be considered. Lower sensor temperatures are possible during severe frosts or when transformers are switched off. Under certain circumstances temperatures higher than 60°C can exist with very hot oil in the transformer without being a problem of the heating circuit. In such a case, the measuring result can have additional error deviations. A total 16 alarms can be defined on the HYDROCAL 1002-1003 unit. If several alarms use the same output line, the status of alarms will be connected by an „OR“- statement, resulting to activate the corresponding output if an alarm is triggered. In the overview of alarms, (Alert Setup Overview) the following displays forms can be selected: •
Either List / View in form of a list
•
Or Details view (listing details of alarm settings)
•
Or Symbol view (presenting alarm settings as symbols)
HYDROCAL 1002-1003: Manual for installation and operation
Page 69 of 97
Selection is made by the menu and with right mouse click
or by „View->View As->...“
In the views „List“ and „Symbol“, the following dialogue can be opened by double clicking on one of the alarms (in view „Details“ it is not possible to open this dialogue):
First the mode of alarms must be selected under „Alert Mode“. Following modes are available: •
High: An alarm is triggered when Level 1 is exceeded.
•
High-High: An alarm in two steps is triggered. The 1 st alarm is triggered when Level 1 is exceeded; the 2nd alarm is triggered after exceeding Level 2.
•
Low: An alarm is triggered when a value falls below the limit defined in Level 1.
•
Low-Low: An alarm in two steps is triggered. Step 1 is alarm is triggered when the measured value falls below the limit defined in Level 1; the second alarm is triggered when the measured value falls below the limit defined in level 2.
HYDROCAL 1002-1003: Manual for installation and operation
Page 70 of 97
•
Low-high: An alarm in two steps is triggered. Step 1 is triggered when the measured value falls below the limit defined in Level 1; the second alarm is triggered after exceeding Level 2.
Range: An alarm is triggered when the measured value is outside the limits defined in Level 1 and Level 2. Once the alarm mode is set then, a sensor must be chosen. All above mentioned internal sensors are available. If external sensors are connected and configured, these can also be used (see chapter 4.11). The following measuring values are available: •
•
H2 Concentration
•
CO Concentration
•
H2 Daily trend
•
H2 Weekly trend
•
CO Daily trend
•
CO Weekly trend
•
Temperature 1
•
Temperature 2
All values defined under external sensors For a clear distinction of an alarm, under „Name“ a name can be inputted. The alarm levels are entered in the fields „Level 1“ and if necessary in „Level 2“. The units defined for concentration of gases „ppm”, for temperatures „°C” and for external sensors the unit defined for the respective sensor (e.g. Volt, Ampere ...) will be used. The alarm limits must be within the respective measuring ranges. This is 0..2000 ppm for gas concentration, 20..100°C for temperatures and for external sensors the maximum and minimum values of the sensor used. Delays1..2 can be introduced to suppress alarms during short-time exceeding of limits. Since the measuring cycle for gas sensors is 20 min, the delay time should not be less than 20 minutes. The input of delay time is made in seconds or directly in the form [days]-[hours]:[min]:[sec]. The value introduced will be converted after confirmation in [days]-[hours]:[min]:[sec]. Following actions can be selected: •
•
Log entry: An entry in the history of alarms is made; this has to be confirmed (see history of alarms).
•
Output: An output is switched on (If the measured value exceeded the configured limits in the field "Level 1 & 2"). The status „on“ described above is activated. If the measured value falls under the limit, the output is switched off again.
•
Out, hold: An output is switched on (If the measured value exceeded the configured limits in the field "Level 1 & 2"). The status „on“ described above is activated. If the measured value for this output falls under the limit, the output remains switched on. Only a manual deleting of alarm allows that the output is switched off again (see chapter 4.11 or 3.6).
•
Log & SMS: An entry in the history of alarms is made, this has to be confirmed (see history of alarms) and an SMS is sent to the number indicated.
•
Output & SMS: An output is switched on (If the measured value exceeded the configured limits in the field "Level 1 & 2"). The status „on“ described above is activated. If the measured value falls under the limit, the output is switched off again. In addition, an SMS is sent to the number indicated.
Out, hold & SMS: An output is switched on (If the measured value exceeded the configured limits in the field "Level 1 & 2"). The status „on“ described above is activated. If the measured value for this output falls under the limit, the output remains switched on. Only a manual deleting of alarm allows that the output is switched off again (see chapter 4.11 or 3.6). In addition, an SMS is sent to the number indicated. Data in the tree are changed after modification of alarm settings. The star indicates that modifications have been made; this must be transferred to the HYDROCAL 1002-1003 unit for activation. This means, the settings indicated are not identical to those of HYDROCAL 1002-1003. With next connection, it is requested whether the settings on HYDROCAL 1002-1003 unit shall be updated (see chapter 4.11). •
HYDROCAL 1002-1003: Manual for installation and operation
Page 71 of 97
4.10.2. External sensors (see also chapter 3.7.3) In the menu , the settings of external sensors are modified. The form of presentation can be chosen as under alarms as
•
„List“ • „Symbol“ • „Details“ The menu „Sensor properties“ is opened by double clicking on a value or a symbol. The settings of internal sensors cannot be modified and are therefore grayed out.
In the box „Input Signal“, the unit of the input signal can be selected. For the first four inputs (Free analog #1 to Free analog #4), the following must be chosen: • •
With option Voltage input: “0...10V anal. In“ Configurable via jumper on the electronic board With option Current input: “0...20mA anal In“ or 4...20mA anal. In“
For the next four inputs (Free analog #5 to Free analog #8) values must either be between “0...20mA anal In“ or “4...20mA anal. In“. For the next four inputs (Free virtual #11 to Free virtual #14), these four inputs are virtual sensors and in any case have to be consider as external sensors or bushing sensors. It is us ed for calculation of Hot Spot Temperature, Aging Rate and Loss of Life
HYDROCAL 1002-1003: Manual for installation and operation
Page 72 of 97
After selection of the input unit, the unit of the sensor has to be chosen from this list: °C: for temperature-sensors e.g. oil temperature, ambient temperature... • ppm: for oil humidity-sensors or other sensors of chemicals for oil humidity sensors or other sensors of chemicals • %: for voltmeters • V: for voltmeters • KV: for ampere meters • MA: for ampere meters • A: for ampere meters • kA: A name can be given to each sensor for ease of use. For conversion of input units, the inputs „Minimum Value“ are made under the input signals 0V, 0 mA or 4 mA and „Maximum Value“ are made under the input signals 10V or 20mA. In addition, the display range can be limited by inputting introduction of „Minimum Display“ and „Maximum Display“. “Smooth calculation can be used, you can activate this function. Data in the tree are changed after modification of sensor settings. The star indicates that modifications have been made; these must be transferred to the HYDROCAL 1002-1003 unit for activation. This means, the settings indicated are not identical to those of HYDROCAL 10021003. With next connection, it is requested whether the settings on HYDROCAL 1002-1003 unit shall be updated (see chapter 4.11). •
4.10.3. Alert SMS Setup In the menu , the configuration of the alert SMS transmission is made if HYDROCAL 1002-1003 is equipped within an on-board GSM / Analogue modem. This function allows sending SMS in correlation with beforehand definite alarms for gas-in-oil concentrations, temperatures as well as for external sensors values
Input of operator name is made in the field "Service Center". The system has 5 possibilities to send SMS in the field "Recipient 1-5". Input of telephone number is made by the alphanumeric keyboard. A message in the field "Message" can be pre-write in order to be sent to the recipient in the event of alarms while using pre-send message. For this, In the field “message” click on right mouse button and click on Insert to visualize followings pre-send messages:
HYDROCAL 1002-1003: Manual for installation and operation
Page 73 of 97
To add pre-send message, you have to click on them. The following possibilities are available: •
{#}- Serial Number: means that in the event of alarms, you will receive a SMS with serial number of the unit
•
{1}- Custom Line 1: means that in the event of alarms, you will receive a SMS with customer 1st line displayed on the main HYDROCAL 1002-1003 Menu
•
{2}- Custom Line 2: means that in the event of alarms, you will receive a SMS with customer 2nd line displayed on the main HYDROCAL 1002-1003 Menu
•
{3}- Custom Line 3: means that in the event of alarms, you will receive a SMS with customer 3rd line displayed on the main HYDROCAL 1002-1003 Menu
•
{4}- Custom Line 4: means that in the event of alarms, you will receive a SMS with customer 4th line displayed on the main HYDROCAL 1002-1003 Menu
•
{T}- Long Time: Means that in the event of alarms, you will receive a SMS with Time of alarm which has been exceeded with f ollowing format: “hh:mm:ss”
•
{t}- Short Time: Means that in the event of alarms, you will receive a SMS with Time of alarm which has been exceeded with f ollowing format: “hh:mm”
•
{D}- Long Date: Means that in the event of alarms, you will receive a SMS with Date of alarm which has been exceeded with f ollowing format: “YYYY:MM:DD”
•
{d}- Short Date: Means that in the event of alarms, you will receive a SMS with Date of alarm which has been exceeded with f ollowing format: “YY:MM:DD”
•
{A}- Alert Name: means that in the event of alarms, you will receive a SMS with the alarm Name
•
{a}- Alert Number: means that in the event of alarms, you will receive a SMS with Alert Number
{l}- Alert Level: means that in the event of alarms, you will receive a SMS with Alert Level (“1” for a High alarm or Level 1 and “2” for a High-High alarm or Level 2) Note: You can combine the variable between them in order to have requested Alarm SMS information (e.g. {#}{1}{T}{D}{A}{l}) A message in the field "Message" can also be free pre-write in order to be sent to the recipient in the event of alarms. •
Note: The SMS mode (in Alarm setup) must be selected to use this function. A GSM / Analogue modem must be also installed in HYDROCAL 1002-1003. Data in the tree
are changed after modification of sensor settings and validation
with . The star indicates that modifications have been made; these must be transferred to the HYDROCAL 1002-1003 unit for activation. This means, the settings indicated are not identical to those of HYDROCAL 1002-1003. With next connection, it is requested whether the settings on HYDROCAL 1002-1003 unit shall be updated (see chapter 4.11).
HYDROCAL 1002-1003: Manual for installation and operation
Page 74 of 97
4.10.4. Input of DGA Level Adjustment Results of laboratory analysis can be introduced in the menu . These can be used for calculation of correction factors for definition of the gas content. After opening of transformer menu page, the correction factors are shown in the fields „Cur. Offset“.
For the input of results of laboratory analysis, first the date of analysis should be entered in the field „Sample Date“. Then the average value of gas concentration is calculated and presented. After the results of laboratory analysis have been introduced in the fields „Laboratory Value“ for H2 and CO, the button
is activated the new correction factor is calculated and displayed in the field „New
Offset“. This value can be used or can be manually adjusted. The new values are stored and activated by
button.
Data in the tree are changed after modification of sensor settings. The star indicates that modifications have been made; these must be transferred to the HYDROCAL 1002-1003 unit for activation. This means, the settings indicated are not identical to those of HYDROCAL 10021003. With next connection, it is requested whether the settings on HYDROCAL 1002-1003 unit shall be updated (see chapter 4.11).
HYDROCAL 1002-1003: Manual for installation and operation
Page 75 of 97
4.10.5. Analog Output Setup In the menu , The HYDROCAL 1002-1003 has 4 analog outputs for measured gas-in-oil concentration, temperatures as well as for external sensors values. The output signal is 020 mA or 4-20 mA.
First a quantity / variable must be selected in the field "Sensor", the following variable are available: •
H2 concentration
•
CO concentration
•
H2 Daily trend
•
H2 Weekly trend
•
CO Daily trend
•
CO Weekly trend
•
Temperature 1
•
Temperature 2
•
All quantities / variables defined under external sensors
The following analog outputs signals are available in the field "Mode": •
OFF
•
0…20 mA
•
4…20 mA
Data in the tree
are changed after modification of sensor settings and validation
with . The star indicates that modifications have been made; these must be transferred to the HYDROCAL 1002-1003 unit for activation. This means, the settings indicated are not identical to those of HYDROCAL 1002-1003. With next connection, it is requested whether the settings on HYDROCAL 1002-1003 unit shall be updated (see chapter 4.11).
HYDROCAL 1002-1003: Manual for installation and operation
Page 76 of 97
4.10.6. Digital Output Setup In the menu
, you can invert digital outputs status from "On" to "Off" while confirm-
ing by a click on the output number and while validating on
.
The following actions configured beforehand in chapter 4.10.1(Alert Setup) will be thus carried out as follows: •
Output: An output is switched "off" (If the measured value exceeded the configured limits in the field "Level 1 & 2"). If the measured value falls under the limit, the output is switched "on" again.
•
Out, hold: An output is switched "off" (If the measured value exceeded the configured limits in the field "Level 1 & 2"). If the measured value for this output falls under the limit, the output remains switched "off". Only a manual deleting of alarm allows that the output is switched "on" again (see chapter 4.11 or 3.6).
•
Output & SMS: An output is switched "off" (If the measured value exceeded the configured limits in the field "Level 1 & 2"). If the measured value falls under the limit, the output is switched "on" again. In addition, an SMS is sent to the number indicated.
•
Out, hold & SMS: An output is switched "off" (If the measured value exceeded the configured limits in the field "Level 1 & 2"). If the measured value for this output falls under the limit, the output remains switched "off". Only a manual deleting of alarm allows that the output is switched "on" again (see chapter 4.11 or 3.6). In addition, an SMS is sent to the number indicated.
Data in the tree
are changed after modification of sensor settings and validation
with . The star indicates that modifications have been made; these must be transferred to the HYDROCAL 1002-1003 unit for activation. This means, the settings indicated are not identical to those of HYDROCAL 1002-1003. With next connection, it is requested whether the settings on HYDROCAL 1002-1003 unit shall be updated (see chapter 4.11).
HYDROCAL 1002-1003: Manual for installation and operation
Page 77 of 97
4.10.7. Device Setup (Clock settings) In the menu
, you can modify the date and the time on each HYDROCAL 1002-1003.
The field "Daylight Saving Time Mode" allows to configure the following time mode : •
MET (Middle European Time) adjusts the official local time forward, one hour from its official standard time for the duration of the spring and summer months.
•
NONE means that the function is not activated
The field "Clock offset" allows modifying time zone. The time zone values are between -6 and +6 hours. Note: "None " means that the function is not activated. Moreover, you can synchronize the date and time on each HYDROCAL 1002-1003 with PC clock while confirming by a click on the field "Synchronize with PC clock" in the following communication box:
HYDROCAL 1002-1003: Manual for installation and operation
Page 78 of 97
Data in the tree
are changed after modification of sensor settings and validation
with . The star indicates that modifications have been made; these must be transferred to the HYDROCAL 1002-1003 unit for activation. This means, the settings indicated are not identical to those of HYDROCAL 1002-1003. With next connection, it is requested whether the settings on HYDROCAL 1002-1003 unit shall be updated (see chapter 4.11).
4.11. Data transmission from / to the HYDROCAL 1002-1003 unit For transmission of data, the respective transformer menu must be selected in the root. In the lower part of the window, there are three selection buttons available for communication.
Pressing button
the connection is initiated and the following selection box is opened:
In the dialogue, the following actions can be selected: •
„Read Configurations“: Read configurations from HYDROCAL 1002-1003 unit. This means that the modifications of alarms and sensor settings made in HydroSoft (indicated by a star) are overwritten by the settings active in the Hydrocal unit. HYDROCAL 1002-1003: Manual for installation and operation
Page 79 of 97
•
„Read History and Alerts“: All measured and stored values, not downloaded previously, are read and stored in the database of HydroSoft.
•
„Update Configuration“: All sensor and alarm modifications made in HydroSoft are transferred to the HYDROCAL unit and activated there. Input of a numerical password is necessary for this action (Default password is 999).
•
„Reset Alert“: This command resets all alarms on the HYDROCAL unit which have not previously been acknowledged. Input of a numerical password is necessary for this action (Default password is 999).
•
„Auto disconnect when done“: Disconnects automatically the line after data transmission. It is recommended that this field is always selected when using modem connections in order to save on transmission costs.
„Synchronize with PC clock“: synchronize automatically the date and time on the unit with internal PC clock. Input of a numerical password is necessary for this action (Default password is 999). After confirmation by the “Start“ – button, the status of communication will be presented in the window. •
By pressing the button
connection is interrupted. This must be avoided during Login. After
interruption, the log-data must be closed by button
. The log-data is used for error tracing in
case of communication problems.
HYDROCAL 1002-1003: Manual for installation and operation
Page 80 of 97
4.12. View of results To view measured results read out, you must select the „History“ field of the respective transformer. Three operation elements are available (see below picture)
4.12.1. Sensor selection This window shows the different sensors available on the HYDROCAL unit. Besides the sensor name, the last measuring value is displayed; this means after having read-out data this is the last measured value. On the left of the sensor name is an activation mark (e.g. : H2 sensor is activated for data presentation). By double clicking on a sensor name, the sensor is activated / deactivated for presentation. In this case, the left vertical scale is used. Open by a right mouse button click on a sensor name the menu for allocation to different Y-axis or deactivation as well as to change the color.
4.12.2. Time range selection The time range for presentation is selected in two steps. •
Selection of the time range the field „Time Span“. Available are: 1 hour, 2 hours, 3 hours, 6 hours, 12 hours, 1 day, 2 days, 3 days, 1 week, 2 weeks, 3 weeks, 1 month, 2 months, 3 months, 6 months, 1 year, 2 years and all data.
•
Selection of date of start by mouse click on a date. Using arrow buttons in the calendar dialogue, the following or the next month is presented.
HYDROCAL 1002-1003: Manual for installation and operation
Page 81 of 97
4.12.3. Result window The result window shows all sensor results activated for presentation. The results are presented in form of time trend or graph tables. When on the result window, click the right mouse button to open the menu and select the form desired. There is also the possibility to activate the print-function (only from trend.....) or export function (only from graph table) and to fade out the sensor and time selection.
4.12.4. Print and export of results There exist two possibilities to process measuring data for the purpose of documentation: •
From the table view, data can be exported to MS Excel © or as a text file. Export to Excel can be activated either from the menu or by „File->Export->Excel…“. Export into a text file is possible only by „File->Export->CSV…“.
• Data can be printed directly from the trend view.
HYDROCAL 1002-1003: Manual for installation and operation
Page 82 of 97
4.13. View of alarm status To view alarms, you must select the list „Alert“ under the corresponding transformer. The operation elements are same as indicated under „View of results. Under „Alarm selection“ the different alarms defined and selected on the HYDROCAL unit can be activated in form of time trend and table views. In the time range selection, the time to be viewed is defined and results are presented in the alarm window.
4.13.1. Alarm selection The alarm selection window presents the different alarms available on the HYDROCAL unit. On the right side of the alarm name, the last alarm status is shown; this is the current status at the time of data retrieval. On the right side of the alarm name is to be found a status and activation mark is displayed (e.g. : : alarm status 0 is activated for presentation, : alarm status 0, is deactivated for presentation, : alarm status 3 is activated). The alarm name can be activated / deactivated for presentation by a double click. The menu is opened with right mouse button.
4.13.2. Time range selection The time range for presentation is selected in two steps. •
Selection of the time range the field „Time Span“. Available are: 1 hour, 2 hours, 3 hours, 6 hours, 12 hours, 1 day, 2 days, 3 days, 1 week, 2 weeks, 3 weeks, 1 month, 2 months, 3 months, 6 months, 1 year, 2 years and all data.
•
Selection of start date by mouse click on a date. Using arrow buttons in the calendar dialogue, the following or the next month is presented.
HYDROCAL 1002-1003: Manual for installation and operation
Page 83 of 97
4.13.3. Alarm window The alarm window shows all alarm status activated for presentation. The results are presented in form of a time trend chart or a table. When on the alarm window, click right mouse button to open the menu and select the form desired. There is also the possibility to activate the print-function (only from time trend chart) or export function (only from table view) and to fade out the sensor and time range selection.
4.13.4. Status of alarms Five different alarm status are presented in both “View as table” and in “View as graph”. The meaning of the different status is described as f ollows: •
Status
•
•
Status
•
Status
3: The limit defined for a High - Alarm is currently exceeded.
•
Status
4: The defined level for a High - High - Alarm is currently exceeded.
0: No alarm active, the limits f or the alarm have not be exceeded in the past or an alarm existing has been acknowledged. Status 1: The limit defined for a High - Alarm was exceeded in the past, but value has fallen below the limit. 2: The defined “Level 2“ for a High - High - alarm has been exceeded in past. The value has fallen below the Level 2, but is still higher than „Level 1“.
4.13.5. Print and export of results There exist two possibilities to process measuring data for the purpose of documentation: •
From the table view, data can be exported to MS Excel © or as a text file. Export to Excel can be activated either from the menu or by „File->Export->Excel…“. Export into a text file is possible only by „File->Export->CSV…“.
HYDROCAL 1002-1003: Manual for installation and operation
Page 84 of 97
4.14. Online Sensors To view a direct/online visualisation, you must select the list „Online Sensors“ under the corresponding transformer. The operation elements are same as indicated under „View of results. Under „visible sensors“ the different measuring values (H2, CO, C2H2, C2H4 and H2O) and external sensors defined in “Sensor setup” and selected on the HYDROCAL unit can be activated in form of graphical view. To activate Online visualisation, you must first of all select the right connection settings (See chapter 4.9) and then click on Start button. You can define followings settings: •
Visible Time Span
•
Graph Width in second (s)
•
Recording Time Span
•
Recording time in hour (h)
HYDROCAL 1002-1003: Manual for installation and operation
Page 85 of 97
5. RS232 / RS 485 Wiring This chapter contains information for the wiring of the connections. RS232 The connection via RS232 is done using the communication cable that is shipped together with the HydroCal device. This cable is designed to be directly plugged into a sub-D 9-pin male connector for an RS232 interface found on standard personal computers (or USB-to-serial adapters). RS485 The RS485 interface is implemented as a two-wire interface. The two signal wires must be clamped onto connector #7, pin 5.1 A(–) and pin 5.2 B(+) inside of the HydroCal. Care must be taken for the right polarity.
Termination, Biasing, Grounding and Protection of an RS485 Bus Termination, biasing, grounding and protection of an RS485 Bus is a complex thing and far beyond the scope of this document. A very good source of information related to these issues can be found at www.bb-elec.com (see the section Tech Notes ).
HYDROCAL 1002-1003: Manual for installation and operation
Page 86 of 97
6. Vaisala sensors connection on HYDROCAL 1002 The connection of Vaisala moisture sensor is depending of the type
6.1 MMT318 The Vaisala moisture sensor type MMT318 has no integrated power supply. In order to energize it on Hydrocal 1002, an integrated GSM-/Net module is used. The wiring of cables for power and signal wires between Hydrocal 1002 and Vaisala moisture sensor is shown as follow on the diagram.
HYDROCAL 1002-1003: Manual for installation and operation
Page 87 of 97
6.2 MMT338 The Vaisala moisture sensor type MMT338 has an integrated power supply. The wiring of signal lines between Hydrocal 1002 and Vaisala moisture sensor is shown as follow on the diagram.
HYDROCAL 1002-1003: Manual for installation and operation
Page 88 of 97
7. HYDROSOFT Scheduler 7.1 General Up to version 1.0.16 the transfer of the history data from the Hydrocal device into the Hydrosoft application could only be started manually. With version 1.0.18 an automatic read-out of the history data is possible. This is managed by the new scheduler integrated into Hydrosoft. In conjunction with the task scheduler provided by the Windows operating systems, the history data can be held up to date automatically.
7.2 Scheduler The scheduler is active in the background. It periodically checks all transformers for pending history data transfers. If an update is awaiting a connection to the Hydrocal is established and the history data read-out is executed. Remark: It is not possible to update the configuration, to reset alerts or to synchronize the real timer clock by the scheduler - these actions still must be carried out manually.
7.2.1 Configuring a Transformer for automatic History Data Update To configure the automatic history data read-out for a transformer, the connection scheduler dialog is used. To launch the dialog, the transformer main page must be activated by selecting the corresponding transformer in the tree view.
The dialog is accessible via the main menu (Transformer -> Scheduler...) or by clicking the rightmost button in the control area of the transformer window. 7.2.1.1 Connection Scheduler Dialog The connection scheduler dialog allow to configure the automatic history data read-out. (For further details on the setting see 0, 7.2.4 Internal Operation.)
HYDROCAL 1002-1003: Manual for installation and operation
Page 89 of 97
The checkbox Scheduled Data Read-Out en- or disables the automatic history data update. The combo box Connect Interval allows the selection of the time span between read-out actions. • After a successful history data update the next connection time is calculated by adding the selected interval time span to the current time. If the Connect Interval is at least one day, a connection time can be set in the text box Connect • Time . The format for the time is hh:mm:00 (seconds must be zero). When a connect time is set, the time-of-day for the next connection will be set to the specified value. If no special time-of day is required, a dash must be entered ('-'). • The Data Read Timeout has the same meaning as in the manual communication and defines a time when the connection has to be regarded as broken - leading to an error abort of the communication. For direct connections 5 seconds should be used - for a modem connection at least 15 seconds are reasonable. The text box Next Connect at shows the date and time of the next connection. • Remark: If a test of the scheduler communication is desired, a double-click on the text box will change the date and time so that the communication will be scheduled in the near future (dialog must be closed with OK to accept the change). The text box Connect Fail Count shows the number of connection failures. • Remark: A double-click on this text box can be used to reset the failure counter to zero (dialog must be closed with OK to accept the change). •
7.2.2 Scheduler Status Visualization The status of the scheduler is visualized in the tree view by changing the icons of the customer, the stations and the transformers. 7.2.2.1 Status Icons The status of transformers, stations and the customer is reflected by special icons. The following description explains the meaning of the different icons. 7.2.2.1.1 Transformer •
The standard icon is used for transformers that are not activated for scheduled history data update. • •
The read-out status is good. (The number of connection failures is less t han 3.) , The read-out status is bad. A connection has been tried several times but was not successful. The icon will be yellow (red) if the number of connection failures is greater then 2 (5).
•
A read-out is pending. The transformer has been marked for a communication and the history data update will be started in the near future. - This status overrules the good/bad-status. •
The transformer is currently busy in scheduler communication. The history data update is in progress. HYDROCAL 1002-1003: Manual for installation and operation
Page 90 of 97
•
The transformer is currently busy in a non-scheduler communication (manual connection or online sensor graph running). - This status overrules the good/bad/pending-status. 7.2.2.1.2 Station The station icon is derived from the status of the transformers that are members of the station and activated for scheduled history data update. •
If a station does not contain any transformer that is activated for scheduled history data update, it is shown with the standard icon. • •
The read-out status of all scheduled transformers of the station is good. , The read-out status of at least one scheduled transformer of the station is bad.
•
A read-out is pending for at least one transformer of the station. - This status overrules both the good/bad-status. •
A transformer in this station is currently busy in a scheduler communication. 7.2.2.1.3 Customer The customer icon is derived from the status of all transformers activated for scheduled history data update. •
The scheduler is active and idle. No communication is pending or active. •
The scheduler is active and waiting for the next pending communication to be started. •
A transformer is currently busy in a scheduler communication. 7.2.2.2 Example The following screen shots illustrate the scheduler operation. The scheduler is active (clock icon) and has scanned the transformers. The pending communications (Station1 and Transformer4 icon with clock) are marked. Station2 does not contain any scheduled transformers.
The scheduler has started the communication for Transformer1 in Station1 (all related icons indicate the active scheduler communication). The communication buttons in the transformer window control HYDROCAL 1002-1003: Manual for installation and operation
Page 91 of 97
area are disabled. The control area is shown with an alternate background colour to indicate the busy status.
The communication with Transformer1 in Station1 has failed more than 5 times. The transformer as well as the station are marked with failure indication icons (red station, red transformer).
7.2.3 Hydrosoft and the Windows Task Scheduler To fully automate the history data update, Hydrosoft can be regularly invoked via the task scheduler available in the Windows operating system. Windows provides a wizard to set up a task. Follow the instructions to create an entry for Hydrosoft. The timing setup for the Windows task depends on the personal needs and the setup of the scheduled history data update in Hydrosoft. If for example a daily update is configured in Hydrosoft the Windows task must also invoke Hydrosoft every day. After the entry has been created with the help of the wizard, the new command line options may be applied to tune the Hydrosoft task. Remark: Prior to setting up a Hydrosoft task it should be checked that the communication setup for the transformers is correct and it should be proved that the communication is working. HYDROCAL 1002-1003: Manual for installation and operation
Page 92 of 97
7.2.3.1 Command Line Options for Windows Task Scheduler Hydrosoft comes with two new command line options for use in conjunction with the scheduled history data update. Example: \Hydrosoft
-IdleClose 0:15:00 -Minimized
7.2.3.1.1 IdleClose The IdleClose command line option can be given to let the application terminate after all history data updates have be carried out. Following the command line option, a time span argument must be supplied, that defines the idle close time. The syntax for the close time is [hh:[mm:]]ss. When the Hydrosoft scheduler scans the transformers for pending history data updates, it determines the shortest time span, when the next communication has to be started. If this time span is greater than the idle close time, the application exits. When a daily update is configured a reasonable default for the idle close time is 15 minutes. This would allow for repeated communication tries (in case of failure), but exit, when the communication fails too often. 7.2.3.1.2 Minimized The Minimized command line option can be given to let the application start minimized. Remark: This option will also suppress the warning message, if the Hydorsoft data directory is in use by a manually started Hydrosoft application. A Windows task scheduler started instance of Hydrosoft can immediately exit, if a manually started Hydrosoft application is already running.
7.2.4 Internal Operation This is a more detailed description of the internal operation of the scheduler (the terms in italics are related to the data configured in the connection scheduler dialog). The scheduler runs as a separate thread in the background of Hydrosoft. Approximately every 15 seconds all transformers are checked if scheduled data read-out is active. Once a transformer is active, a further check is made, whether the current time is greater than the next connect at time. This way a list of pending connections is generated. If the list of pending connections is non empty, it is processed step by step, i.e. the list is traversed and all pending connection requests are executed before a new list is build. For each transformer a try to update the data is done. On success the connect fail count is set to zero and the next connect at time is updated (current time plus connect interval - optionally aligned to the connect time ). In case of an error the connect fail count is incremented and the next connect at time is set based on the current connect fail count . If the count is less than 3, a connection retry with 5 minutes delay is set, for a count less than 5 the retry is scheduled in 10 minutes. If the count exceeds 5, the next retry is scheduled to the current time plus the connect interval (- optionally aligned to the connect time , i.e. the same time as if the connection had succeeded). When the scheduler detects, that a time span of 3 times the connect interval has expired since the last start of the application, the connect fail count is reset to 4. This will lead to a double communication try for a transformer with a bad status when the application has not been used for a longer time period. Because of the internal operation the following notes apply: • •
Changes applied through the connection scheduler dialog take some time until they are reflected in the tree view. Even if the read-out status of a transformer is bad, there will always be a try to update the data with the configured connect interval . HYDROCAL 1002-1003: Manual for installation and operation
Page 93 of 97
•
A manual update of the history data does not update the next connect at time of the scheduler. Even if an update occurs immediately before the next connect at time, the scheduled data readout will still be executed.
A log file 'Scheduler.log' is created in the root if the Hydrosoft data directory. If a problem arises with the scheduler a look into this file may help finding the cause. The log file entries are always written in English.
7.2.5. Data Directory Lock Hydrosoft now protects the data directory by a lock. Once an instance of Hydrosoft is working on a data directory no second instance of Hydrosoft can operate on the same data. This locking is introduced to keep data integrity in case a user is running Hydrosoft and a background instance of Hydrosoft (started via the Windows task scheduler) also tries to work on the data. Hydrosoft will exit with a return code of 2 if the data directory is locked.
HYDROCAL 1002-1003: Manual for installation and operation
Page 94 of 97
8. Technical data HYDROCAL 1002-1003 Measuring quantities Hydrogen H2: Carbon Monoxide CO: Moisture-in-Oil H2O:
Measuring range 0 ppm ... 2.000 ppm 0 ppm ... 2.000 ppm absolute 0 ppm-100ppm
Measuring accuracy ± 15 % of the measuring value ± 25 ppm ± 20 % of the measuring value ± 25 ppm ± 3 % of the measuring value ± 3 ppm (related to oil temperature of +55°C)
Measurement interval: Operation temperature:
Oil pressure: Functions:
Outputs:
20 min. Oil temperature: -20°C ... +90°C Ambient temperature: -50°C ... +55°C (below -10°C display function locked) Temperature coefficient: 1%/K 0-800 kpa, no vacuum allowed Individual gas-in-oil concentration measurement for H 2 and CO Gas trend analysis (chart / bar diagram) Different alarm level configuration RS 232 communication and modem communication (GSM, analogue modem) to central PC Windows� software 3 x analogue outputs: 0/4... 20mA (H2 concentration) 0/4... 20mA (CO concentration) 0/4... 20mA (H2O concentration) with HYDROCAL 1003 1 x analogue outputs: 0/4... 20mA (free configurable) U Max: 6V Impedance: 300 Ohm 12 x digital outputs: 4 x 12 V relay outputs Maximum switching values: 220 V DC, 220 V AC, 2 A, 60 W 8 x opto-coupler outputs Maximum Current: 40 mA UCE 24 V Rated voltage (35 V maximum) UEC 7 V (maximal)
Alarms:
H2 Alarm (H) H2 Daily trend H2 Weekly trend CO Alarm (H) CO Daily trend CO Weekly trend
(1st level and 2n level alarm) 1. Alarm level & 2. Alarm level 1. Alarm level & 2. Alarm level 1. Alarm level & 2. Alarm level 1. Alarm level & 2. Alarm level 1. Alarm level & 2. Alarm level
Inputs:
4 x analogue inputs: 4 x analogue inputs:
0/4 ... 20 mA 0/4 ... 20 mA / 0 ... 10 V (configurable by jumpers)
Gas-in-oil: Moisture-in-oil Monitor:
3 x internal gas sensors (redundant system) 2 x H2, 1 x CO 1 x internal moisture sensor 2 x internal temperature sensors (oil temperature, gas temperature) RS 232 (external connection, no unmounting of protective cover required) RS 485 (bus-operation or point-to-point operation, MODBUS or proprietary protocol) Internal on board modem (GSM, 14.4 kBit / analogue 56 kBit) IP 55 88 VACmin … 276 VACmax Optional: 88 VDCmin … 390 VDCmax
Communication:
Protection class: Power supply:
HYDROCAL 1002-1003: Manual for installation and operation
Page 95 of 97
Dimensions: Weight:
max. 350 VA, 50/60 Hz approx. W 215 x H 215 x D 215 mm approx. 8 kg
HYDROCAL 1002-1003: Manual for installation and operation
Page 96 of 97
View more...
Comments