Handbook Liquid Mono Propellant Thruster Performance Measurement
Short Description
Download Handbook Liquid Mono Propellant Thruster Performance Measurement...
Description
AFRPL-TR-79-24
_.'". i
JPL
79-32
"" ADA072125
It1111_111111111111111111 f
Handbook of Recommen"deO .......... Practices for the Determination of Liquid Monopropellant Rocket Engine Performance
• _--.__J
c
JET PROPULSION LABORATORY CALIFORNIA INSTITUTE OF TECHNOLOGY PASADENA
s
CALIFORNIA
AUTHORS:
R, R, R.
JUNE
1979
Ao S. K.
91103
BJORKLUND ROGERO BAERWALD
Approved
_or
_i_trib_tion
Pablic
Re£e=_e;
_nlimited
_F
PREPARED AIR
L"
FORCE
DIRECTOR AIR
OF
FORCE
EDWARDS
ROCKET
SCIENCE
SYSTEMS
AFB,
PROPULSION
FOR:
LABORATORY
& TECHNOLOGY
WASHINGTON,
COMMAND
CALIFORNIA
NATIONAL AERONAUTICS SPACE ADMINISTRATION
93523
Ill_ODtR[OBY
NATIONAL TECHNICAL INFORMATION SERVICE U.S.
O[FARIM|III SFIIIIIGFIELD,
O| ¢OIIN£RC[ VA. 22161
D.C.
&
J
NOTICES
I
When U.S. Government drawings, specifications, or other data are used for any purpose other than a definitely related government procurement operation, the Government thereby incurs no responsibility nor any obligation whatsoever, and the fact that the Government may have formulated, furmished, or in any way supplied the said drawings, specifications or other data, is not to be regarded by implication or otherwise, or in any manner licensing the holder or any other person or corporation; or conveying any rights or permission to manufacture, use, or sell any patented invention that may in any way be related thereto.
FOREWORD
The work was prepared by the Control and Energy Conversion Division, Jet Propulsion Laboratory, California Institute of Technology, and was Jointly sponsored by the Air Force Rocket Propulsion Laboratory, Edwards AFB, California, through a MIPR FO-4611-76-X-O053 with NASA and by the National Aeronautics and Space Administration under Contract NAS7-100. This report has been reviewed by the information office/XOJ and is releaseable to the National Technical Information Service (NTIS). At NTIS it will be available to the general public, including foreign nations. This technical report has been reviewed and is approved for publication; it is unclassified and suitable for general public release.
WALTER Chief,
FOR
THE
EDWARD Liquid
A. DETJEN Satellite
Propulsion
Branch
COMMANDER
E. STEIN, Deputy Rocket Division
Chief,
GENERAL
DISCLAIMER
8
This document may have problems that one or more of the following disclaimer statements refer to: t
This document has been reproduced from the best copy furnished sponsoring agency. It is being released in the interest of making available as much information as possible.
by the
This document may contain data which exceeds the sheet parameters. It was furnished in this condition by the sponsoring agency and is the best copy available. •
•
This document
may contain
pictures
have been reproduced
which
The document
is paginated
tone-on-tone
or color graphs,
in black
as submitted
charts and/or
and white.
by the original
source.
Portions of this document are not fully legible due to the historical nature of some of the material. However, it is the best reproduction available from the original submission.
UNCLASSIFIED SECURIT_'-_CL--A'SSIFICATION
OF
THIS
PAGE
(When
Dill
Entered)
,,! AFRPL_'TR-
READINSTRUCTIONS BEFORE COMPLETING
PAGE
,._.' / ,/REPORT DOCUMENTATION -,.. REPOR'i_%MBE-R--: I"
2.
GDVT
ACCESSION
NO.
RECIPIENT'SCATALOG
5.
TYPE
FORM
NUMBER
79- 24 i
....T_-_-L 79-'c_ _.....
\
4. T,ITLE (and Subtitle) .
3,
/-
!;HANDBOOK
OF
RECOMMENDED
_DETERMINATIONOF
PRACTICES
LflQUID
FOR
MONOPROPELLANT
PERFO_NCE,
-
.
1
:
PERFORM , ..............
_ "'.
it
AUTHOR(_
....
'_RT_: ,
f .Bj;rklund,
S._ogero,
R,
Jet
ORGANIZATION
Propulsion
Calif. 4800
/
K./Baerwald
AND ADDRESS
of
OR
- NAS7-100,
Task
10.
Grove
Dr.,
PROGRAM
Force
Rocket
14.
AFB
MONITORING
.r-
12,
,_
•
-i
,
,,
.,
./,-,
!
i.;_. _
? "16.
= I_1S T RI
NAME
BU
,
TI
/
Laboratory/LK
;
\
13."
1979
NUMBER
oF'
J
PAGES
I
265 &
AODRESSOI
.
,
dllfetent
from
Controllln
80lttcs)
15.
SECURITY
,
...... .' •
,
CLASS•
(of
this
report)
Unclassified
, ] /
/
!15a.
DECLASSIFICATION/DOWNGRADING SCHEDULE
,
t
O N_ST'A'T:EI_I_e_N'T
APPROVED
/
REPORT.DAT.E..-
93523
AGENCY
TASK ....
91103
AND ADDRESS Propulsion
CA
No. 240
PROJECT, NUMBERS
•
1 June Edwards
No.
UNIT
/ Alr
Order
ELEMENT, &._WORK
." i
CA
Contract
Amendment
AREA
Pasadena
NUMBER
GRA't_'T'NUM'B'ER(s)
Prime
Technology
I1.--CONTROLLINGOFFICENAME
REPORT
JPL/NASA
Laboratory
Institute Oak
NAME
COVERED
79
RD-65 S: PERFORMING
PERIOD
/,
CON'TRACT
8.
"
R.
&
NG.ORG
:JPL 7.
REPORT
,ROCKET '-6>
li ENGINE
OF
THE
FOR
(of
PUBLIC
thts
Report)
RELEASE:
DISTRIBUTION
UNLIMITED--
\
I7.
DISTRIBUTION
STATEMENT
_'o!
the
abstract
entered
In
Block
20,
It
different
Irom
Report)-'_
•
_,_-
.,
,_"i'_.......
, I.
..:c-._.. -
P~
" "'-,,.... , -_-
: .... lB.
SUPPLEMENTARY
, "
.
..:--:.
..
2,.'*
:":,
..
,.,
.
NOTES .
"-6..
b
'__---_-.'._".'.., .... 19.
KEY
WORDS
(Continua
Performance and
on
reverse
wide
_l necessary
standardization,
steady
state,
Test
and
Identify
by
block'number)
Monopropellant
fixtures,
.... :.
." ,-,
rocket
engines;
:_
."
--Pbl_e
"T:esting "
Instrumentatio_
-.:.
" ,--',:
,'"'"'..-.;
:,.._
,
, t
...-: • L
20.
ABSTRACT
(Continua
on
reverse
side
II
necessary
_d
Identt_
by
block
numbe_
' '
• ,
,,:.
,
" '
.'I
..._ This
handbook
neer
in
direct the
the
is
intended
design,
measurement
determination
of of
defining
relations
are
given.,(OVER)
also
to
serve
installation, those
as and
reducing
which thruster
the
for
operation
quantities
monopropellant
for
a guide
measurements
the
of are
of
experienced.test
systems
to
fundamental
performance. to
be
engi-
used
for
importance
The
performance
the to
procedures,
and
parameters t I I
FORM
DD ,_AN_, 1473
EDITION
OF
1 NOV
6S
IS
OBSOLETE
UNCLASS •
/
SECURITY
CL'ASSIFICATION
IFIED OF
THIS
! PAGE
(When
Data
Entarad)_.J,
. •
,-:_=
F
[
UNCLASSIFIED SECURITY
CLASSIFICA'T'ION
This
handbook
impulse
is
THiS
data
PAGE(When
composed
measurement,
measurement, and
OF
of
glossaries
are
six
propellant
temperature reduction
Detm
and
Eneered_
discrete mass
measurement, performance
included
with
each
-
f._>
. .. -
_ •.
I-'
i-
•
-._-
_"
exhaust section
r
'f.-
and
as
flow gas
determination.
_ -....,.-; .
sections
usage
pertaining measurement, composition References,
necessary.
to
force
and
pressure measuremdnt, appendixes,
and
PREFACE
I
This Handbook of Recommended Practices for the Determination of Liquid Monopropellant Rocket Engine P_rfbrmance has the objective of promoting a uniformity of methodology throughout the monopropellant community as regards to pulse mode and steady state thruster testing and data reduction. The program was conceived within a Joint Army-NavyNASA-Air Force (JANNAF) monopropellant working group, consisting also of representatives from industry and academia, to fill what was felt to be a universal need for such standardization of practices among the
manufacturers
and
users
of monopropellant
rocket
engines.
The JANNAF Rocket Engine Performance Working Group has devoted considerable effort toward the development of a consistent methodology for determining rocket engine performance. However, the publications of the JANNAF Rocket Engine Performance Working Group are directed primarily toward bipropellant engines operating at steady state conditions and are, therefore, not directly applicable to the performance determination of monopropellant engines which may operate primarily in the pulse mode. However, the Handbook of Recommended Practices of Measurement of Liquid Propellant Rocket Engine Parameters, Chemical Propulsion Information Agency (CPIA) Publication No. 179, and Rocket Engine Performance Test Data Acquisition and Interpretation Manual, CPIA Publication No. 245, both published by the JANNAF Rocket Engine Performance Working Group, have served as excellent guides for the development of the current document. This handbook is intended to serve the experienced test engineer in the design, installation, and operation of systems which include the determination, by direct measurement, of rocket engine thrust and impulse, propellant mass flow, pressure, temperature, and exhaust gas composition. The algorithms and defining relations for reducing these measurements to performance parameters are also given. This document is not intended to serve as a primer for the inexperienced engineer, as the depth required for such a publ_cation is clearly beyond the scope of this or any other single document.. Specific design guideline_ are, however, offered for the critical components of each system.
These
This handbook is divided into six sections are presented in the following •
Force
and
•
Propellant Pressure
Impulse Mass
relatively order:
discrete
sections.
Measurement Usage
and
Flow
Measurement
Measurement
•
Temperature
Measurement
•
Exhaust
Composition
•
Data
Gas
Reduction
and
Measurement
Performance
Determination
iv
IPreceding pageblankI l
f
Each and
of the
sections
appendixes
includes
a table
of
contents,
of
data
glossary,
references,
as necessary.
To
achieve
the
uniformity
acquisition
and
interpreta-
tion which is the goal of the handbook, it is recommended that the procedures and practices outlined herein be required for all RFQ's and contracts. This recommendation implic_tly includes also adherence to CPIA Publication No. 180, _RPG Handboo_ for Estimating _he Uncertainty in Measurements Made With Licuid Pro_ellant Rocket Engine SYstems. Deviations from these procedures should be specifically negotiated. This
handbook
has
been
formatted
for
reprint
as
a CPIA
publi-
cation. It is probable that its use by the monopropellant community will result in recommendations for additions or revisions. These comments are welcomed and may be addressed to the authors at JPL, the program sponsors at
AFRPL
and
NASA/OAST, Chemical Johns
Qr_
in the
Propulsion
Hopkins
case
of
Information
University
Applied Physics Laboratory 8621 Georgia Avenue Silver Springs, Maryland
V
20910
the
CPIA
Agency
version,
to:
ACKNOWLEDGMENTS I This handbook was produced under the joint sponsorship of the U.S. Air Force Rocket Propulsion Laboratory (AFRPL) and the National Aeronautics and Space Administration, Office o£ Aeronautics and Space Technology (NASA/CAST). Program managers at AFRPL included Mr. Paul Erickson, Lt. Vince Broderick, and Capt. Dennis Gorman. Mr. Frank Stephenson was the program manager at NASA/CAST. The
development
of
this
handbook
involved
the
efforts
of
almost the entire monopropellant community. A considerable expenditure of time and energy was contributed by those organizations which supplied materials or comments in response to the comprehensive survey questionnaire which was distributed throughout the monopropellant and bipropellant community. The following individuals and organizations deserve much credit for their efforts in this regard (listed alphabetically): Mr. Dan Balzer RCA/Astro Electronics P.O. Box 80 Princeton,
New
York
Division 08540
Mr. Fred Etheridge Rockwell International Space Division 12214 Lakewood Blvd. Downey,
California
Mr.
Freeman
Jim
_-_ 90241
":'
Air Force Rocket Propulsion Edwards Air Force Base Edwards,
6alifornia
Laboratory
935.23
Mr. J. L. Gallagher Vought Corporation System Division P.O. Box 5907 Dallas, Texas 75222 Mr. George Huson COMSAT Laboratories Clarksburg,
Maryland
Mr.
Miller
Richard
2O734
Boeing Aerospace Company P.O. Box 3999 Seattle, Washington 98124 Mr. R. R. Nunamaker NASA/Ames Research Moffett
Field,
'%
Center
California
vl
94035
S
L/ Mr. R. David Steel Martin Marietta Corporation P.O. Box 179 Denver, Colorado 80201 Mr. NormStern Aeroneutronic Ford Corporation OneFord Road Newport Beach, California 92663 Mr. ThomasE. Williams NASA/Goddard SpaceFlight Center Greenbelt, Maryland 20771 By far the most comprehensiveresponses were supplied by the rocket engine manufacturers themselves. This seemsto be due not only to the fact that t_e manufacturers have the most sophisticated and complete rocket engine test facilities, but also to a genuine interest on the part of the manufacturers in the production of this handbook. Excellent responses were received from the following (alphabetically): Mr. T. E. Hudson The Marquardt Company 16555Saticoy Street 91409 Van Nuys, California Mr. Milt Marcus Hamilton Standard Bradley Field Road Windsor Locks, Connecticut
06096
Mr. V. A. Moseley Bell Aerospace Company P.O. Box One Buffalo, NewYork 14240 Mr. Robert Sackheim TRWSystemsGroup OneSpace Park RedondoBeach, California
92078
Mr. Bruce Schmitz Rocket Research Corporation York Center Redmond,Washington 98052 Mr. Walter Schubert Pratt & Whitney Aircraft Florida Researchand DevelopmentCenter P.O. Box 2691 West Palm Beach, Florida 33902 In addition to the written responses, muchadditional information was obtained from visits to select organizations. This provided vii
the authors with the opportunity to see firsthand the test facilities involved and to have explained the rationale behind the various measurement philosophies and data reduction procedures. The following organizations were gracious enoughto give of their time in this manner (alphabetically): Air Force Rocket Propulsion Laboratory Edwards, California Bell Aerospace Buffalo, NewYork GoddardSpace Flight Center Greenbelt, Maryland Hamilton Standard Windsor Locks, Connecticut HughesAircraft Company E1 Segundo, California The Marquardt Company Van Nuys, California Rocket Research Corporation Redmond,Washington Rockwell International Space Division Seal Beach, California TRWSystems, Inc. RedondoBeach, California Vought Corporation SystemsDivision Dallas, Texas The authors would like to take this opportunity to express their appreciation for the cordial way in which they were received and for the hospitality extended to them by the above organizations. Following acquisition of the information which was gathered in the manner described above, a draft Of the six handbooksections was produced and distributed to over 32 membersof the monopropellant community. Several months were allowed for a review and critique of the proposed recommended practices. In addition to those membersand organizations of the monopropellant thruster community, gratitude is expressed to the following individuals for taking the time to review and commentin detail upon various sections of the draft document: Mr. Phil Bliss, Intersociety DireCtor, Instrument Society of America; Mr. Pierre F. Fuselier (retired), viii
Lawrence
Livermore
Mr.
Frederick
Mr.
Phillip
Laboratory;
Williams, Scott,
the
Mr.
Eastern
Jon
Inskeep,
Standards
Jet
Propulsion
Laboratory,
Dept.
Laboratory; of
Foxboro
Company;
and
Mr.
Donald
Bond,
pointed
out
not
all
commentsreceived
the
Jet
Navy;
Propulsion
Laboratory. It the the of
various
and
be
were
acknowledgments the
handbook
suggested of
must
reviewers
the
by task
fiscal
strongly
given by
the and
any
would
complications. and,
in
other
here
do
individual
reviewers have
that
incorporated not or
would
cases,
offered. It is felt, however, have worked to the detriment
imply
of
in
ix
final
an
Some increase
disagreed
endorsement the
changes
in
the
scope
one with
none of the required usefulness or accuracy
and
of
publication
contradicted
from
document
unqualified
unacceptable
authors
that the
the
caused
reviewers the
the
organization.
have
resulted Some
into
delays
another
the
rather
comments
compromises of the document.
CONTENTS
SECTION
I
FORCE
SECTION
II
PROPELLANT
SECTION
III
PRESSURE
SECTION
IV
TEMPERATURE
MEASUREMENT
SECTION
V
EXHAUST
COMPOSITION
SECTIONVI
DATA
AND
IMPULSE
MASS
MEASUREMENT
USAGE
AND
FLOW
I-I
MEASUREMENT
MEASUREMENT
GAS
REDUCTION
AND
2-I
3-I
4-I
MEASUREMENT
PERFORMANCE
DETERMINATION
5-I
6-I
SECTION
FORCE
AND
IMPULSE
e
//-/
I
MEASUREMENT
"
FORCE
AND
SECTION
I
IMPULSE
MEASUREMENT
I
CONTENTS
1.0
INTRODUCTION
I-I
2.0
SCOPE
I-I
2.1
OBJECTIVE
2.2
LIMITATIONS
I-2
3.0
DESIGN
I-2
3.1
MECHANICAL
3.1.1
Thrust
3.1.2
Measurement
Force
3.1.3
Calibration
Equipment
3.2
ELECTRICAL
3.2.1
Signal
3.2.2
Electrical
3.2.3
Recording
3.2.4
Visual
3.2.5
Data
4.0
INSTALLATION
4.1
COMPONENT
4.1.1
Force
4.1.2
Calibrator
4.1.3
Thrust
4.1.4
Dynamic
--
I-I
CONSIDERATIONS COMPONENTS
I-2 I-4
Stand
AND
Transducer
1-12
ELECTRONIC
Conditioning
COMPONENTS
Equipment
Calibration
Equipment
Equipment
AND
CHECKOUT
PROCEDURES
Force r
1-19 1-19
Transducer
Stand
1-18
1-19
CERTIFICATION
Dead
1-15
1-18
Equipment
Processing
1-15
1-18
Equipment
Display
I-6
1-19 Weights
Flexures
1-25 and
Calibrator
Restraints
1-25 I_25
SYSTEM
CERTIFICATION
1-26
ENGINE
INSTALLATION
1-26
5.0
CALIBRATION
5.1
STATIC
5.1.1
Deadweight
5.1.2
Working
Standard
5.1.3
General
Calibration
5.1.4
General
Verification
5.2
DYNAMIC
CALIBRATION
6.0
OPERATING
6.1
PRETEST
6.2
POSTTEST
7.0
DATA
7.1
CALIBRATION
7.2
RUN
8.0
GLOSSARY
1-34
9.0
REFERENCES
1-38
AND
VERIFICATION
PROCEDURES
1-26
CALIBRATION
1-26
Calibrator
Method
Force
1-27
Transducer
Calibration
Method
Procedures
1-29
VERIFICATION
1-29
PROCEDURES
1-30
PROCEDURES
1-30
1-31
PROCEDURES
ACQUISITION AND
1-27 1-28
Procedures AND
--
AND
PROCESSING
VERIFICATION
TECHNIQUES
DATA
1-31 1-31
DATA
1-33
APPENDIX_ I-A
THRUST
I-B
SHORTAND UNCERTAINTY
I-C
SHUNT
MEASUREMENT
SYSTEM
LONG-TERM
CALIBRATION
THRUST
ELEMENTAL
UNCERTAINTIES
MEASUREMENT
IA-I
SYSTEM IB-I
OF FORCE
1-ii
TRANSDUCERS
IC-I
Mechanical Components of a Vertical Axial Thrust Measurement Stand
I-2
Conventional
I-3
Thrust
I-4
Abnormal
I-B-I
Example
Measurement Loading of Control
Uncertainty I-B-2
1-C-1
6-Wire
Strain System
Test
Bridge
Calibration
Element I-3
Transducer
.....
I-9
Diagram
1-16
Installation
1-24
for
Checks
IB-3
Block Diagram of ShortUncertainty Checks Shunt
Block
Effect, Data
Gage
Single
and
Circuit
1-iii
Long-Term IB-4
Configuration
IC-2
i
FORCE
AND
SECTION
I
IMPULSE
MEASUREMENT
1.0
Recommended lation,
checkout,
system
to
Three
be
used
appendixes and
are and
during
of
tests
included:
I-B,
Uncertainty;
practices
calibration, are
Uncertainties;
INTRODUCTION
Short-
I-C,
a
and
This
section
experienced
engineer
Long-Term
measurement
system
liquid
monopropellant
the
detailed
but
rather of
juction
with
good
meets
rocket
each
portion
engine.
current,
performance
2. I
The
thrust
to
of
has the
a
the of
produced
been
to
measurement for
a
by
made
a
specify
system,
the
guidelines,
provide
for
operation
impulse
provided
These
a guide
critical used
in
available
measurement
comcon-
equipment
system
which
specified.
be
of
monopropellant
rocket
at
pressures
tests
are
altitudes
of
These
engine engine
simulated The
recommended
specifically uncertainties.
Thrust
Range
to
44
N
(0.01
to
10
ibf)
measured
30
is
ranging
km
Mode
to
4400 1000
N
69
conducted (100,000
contained
measurements
in
from
of
of
Steady
ft) in
thrust
this and
Operation
Steady
from
the
to
3450
kN/m
in
a vacuum
or
higher section
impulse
the
5 ms
state
catalytic
reaction
chamber 2
(10
to
chamber be
min.
mode, on
50
time
I-I
ms
a psia).
maintained.
are
directed
having
the
in
following
Measurement
+_0.5%
Thrust
+_5.0%
Impulse
+_0.25%
Thrust
±2.0%
Impulse
ibf) Pulse
of 500
where
can
Uncertainty
state
Pulsed mode, min. on time to
derived
hydrazine
normally
practices to
0.044
(I0
System
OBJECTIVE
decomposition
44
as and
commercially
will
criteria
serve
attempt
are
system.
practices,
engine. Elemental
Transducers.
and
portion
state-of-the-art,
engineering
the
No
any
the
to
thrust
guidelines
of
rocket System
Measurement
Force
installation, the
for
design
of
instal-
measurement
Measurement
written
measuring
design,
SCOPE
design,
configuration specific
ponents and
for
been
the
thrust
Thrust
Calibration
has
in
the
a
monopropellant
Thrust
2.0
for of
liquid
I-A,
Shunt
outlined
operation
2.2
LIMITATIONS The
recommended
practices
in this
section
are
limited
to
thrust measurements of a single monopropellant hydrazine rocket engine. There is ho fundamental reason why these practices would not be applicable to other types of monopropellant thrusters or even bipropellant rocket engines if certain environmental factors peculiar to the specified propellant(s) are accommodated. Additional limitations are listed below: (1)
Only those thrust systems that infer a force from a bonded metallic strain gage transducer are considered in this section. Specialized applications requiring higher transducer electrical output, extehded highfrequency response, miniature size and/or weight or some combination of these characteristics may necessitate other transduction techniques. For the vast majority of force measurements related to monopropellant rocket engines, however, the bonded metallic strain gage transduction element is the recommended choice.
(2)
Test Cell simulated altitude pressure should be maintained at a level which assures design exhaust nozzle flow characteristics.
(3)
Temperature
conditioning
should
be
limited
to
a practical
operating range as determined for the specified spacecraft environment designed to use monopropellant hydrazine. (4)
The
minimum
pulsed
thrust
duration
is
limited
by
the
propellant control Valve operating time. For low thrust, below 44 N (10 ibf), this could be as short as 5 ms; however, at higher thrust, above 44 N (10 Ibf), a more practical limit is about 50 ms.
3.0
DESIGN
CONSIDERATIONS
The design of a thrust measuring system requires consideration of certain key mechanical, electrical, and electronic components which must be incorporated. Mechanical components include the thrust stand, a measurement force transducer and the force calibration equipment. Electrical and electronic components include electrical calibration, signal conditioning, recording, visual display and data processing equipment.
3.1
MECHANICAL A typical
in a vertical, in Figure I-I.
_(I)
single These
COMPONENTS arrangement
of the
element, axial components are
Thrust frame is mounted.
into
I-2
major
mechanical
components
thrust measuring stand identified as follows: which
the
rocket
engine
is shown
assembly
DYNAMIC
II
FORCE CALIBRATOR (REMOVA
WORKING
BLEI
STANDARD
FORCE
TRANSDUCER
STATIC
CALIBRATOR
------]
ADJUSTABLE
L I
L..
I
L
MULTIPLE
CALIBRATION
DEAD
FORCE
STATIC
TRANSDUCER
CALIBRATOR
I
(ALTERNATE1
I I
FORCE
WEIGHT
RODS
FORCE
I
GEN£1_TOR
__
SEISMIC MASS
VIBRATION )I, ATORS ,],.,
,;,_,
..."i/
_
GROUND PLANE
LINKAGE FLEXTURES MEASUREMENT FORCE
TRANSDUCER
O,SCO NECT LINKAGE
_
FRAME-_
J
r
_
-_
.THRUST
FRAME
/_uExTUrES
ROCKET
Figure
I-I.
Mechanical Components of a Vertical Axial Thrust Measurement Stand
I-3
Single
Element
-J
(2)
Measurement force and flexures.
transducer
(3)
Seismic
vibration
mass
ground
the
same
isolators
(q)
Thrust frame restraints the seismic mass.
(5)
In-place static force calibrator dead weight type or the working with force generator type,
(6)
Dynamic force during engine
for
the
linkages
attached
to
the
and
calibrator testing.
flexures
attached
to
of either the multiple standard force transducer
which
is
normally
removed
oriented thrust measuring stand will have Additional flexural supports would normally
thrust
As shown
mechanical
plane.
A horizontally basic components.
be required
with
with
frame.
in Figure
I-I,
the measurement
force
transducer
is normally linked mechanically between the thrust frame and the seismic mass with -flexures. An alternate method for the optimum high-response pulsed thrust stand would be to eliminate the flexures and use hard mounting on both ends of a ruggedized force measurement transducer. The rocket engine thrust frame assembly is restrained by struts containing pa_rs of flexures so that movement along only the thrust axis is permitted. The geometric centerline of the axisymmetrical engine is shown to coincide with the thrust measurement axis. Forces-applied by the static calibrator should also coincide with the thrust measurement axis. The interaction of the thrust system components that takes place during engine testing should be faithfully duplicated by in-place calibration techniques.
3.1.I
the
design
Thrust
Stand
The following considerations of a thrust measuring stand:
are
of
primary
interest
in
(1)
The thrust stand should be designed to have a large seismic mass to engine-thrust frame mass ratio (ms/me). Experience has indicated that this ratio should be greater than 20:1 with values around 500:I desirable.
(2)
The
seismic
by the
(3)
mass
test
Transverse
should
chamber
be well
and
restraints
ground
used
isolated
(4)
axial
force
forces
and
be designed than 0.15%
applied.
Transverse restraints should to the main axis of thrust.
I-4
vibration
:
to counteract
moments generated by the engine should to limit their axlal restraint to less of the
from
plane.
be
installed
perpendicular
"7.
(5)
Transverse restraints should be designed at least 20% of the axial force because magnitude or during
I
(6)
may be developed engine failures.
Transverse
restraints
during
should
the
start
be attached
the thrust frame and the seismic mass changes can be made without disturbing stand alignment. (7)
Propellant that they
(8)
Propellant
and purge lines serve as elastic and
purge
perpendicular to and moments, and temperature, and ments. Flexible unless it can be procedures, that are negligible.
(g)
(10)
lines
to withstand forces of this transients
between
so that engine the thrust
should be installed pivots.
so
should
nominally
be
installed
the main thrust axis to balance forces to minimize the effect of fluid flow, pressure variations on thrust measurepropellant lines should be avoided demonstrated, through calibration the effects of system pressurization
The thrust frame structure should be designed to deflection limits rather than stress limits. A very stiff lightweight structure will have a natural frequency high enough to ensure accurate response to pulsed thrust inputs and oscillations that will not influence steady state thrust measurement. A na.tural frequency in excess of 200 Hz is attainable. Environmental changes which may affect the alignment and restraint should be thoroughly evaluated. Safety
stops
should
if a thrust frame fatigue, overload,
be used
to limit
engine
motion
member fails during a test because or other hardware failures.
of
(11)
Safety stops should be set for 4 times the rated deflection.
(12)
The measurement force transducer must be protected from inadvertent overloads during installation and removal of the test engine assembly by suitable mechanical stops.. If this is not practical the force transducer should be removed during these operations.
(13)
Lost
(14)
Clearance
motion
of the system
must
axial
exist
thrust frame deflection.
(15)
Targets
(16)
Locating stand
in the
should Jigs
load
between over
the
be installed and
to facilitate
I-5
indexes proper
a clearance
column
should
fixed
and
range
of
for should
of at
be
alignment.
be eliminated.
movable interest
alignment built
least
parts of
purposes. into
the
(17)
Thermal
load
on the
thrust
stand
should
be
minimized
through the use of radiation shields, conductive insulators, or thermal compensating elements.
Measurement
3.1.2
that
measure
Force
Transducer
These recommended practices are limited to thrust systems force with a bonded metallic strain gage force transducer,
commonly referred to as a load cell. The load-carrying structure of the force transducer is normally machined from a single piece of heattreated metal. The elastic deformation of this structure=is sensed by attached strain gage elements. These strain gages are electrically connected in a Wheatstone bridge circuit which can be balanced at no load. The application of a force to the load structure produces resistance changes in the strain gages, thus unbalancing the bridge circuit. The resultant output signal is directly proportional to the applied force and the excitation voltage. Standard design specifications for strain gage force transducers are given in ISA-$37.8, Reference I-1. The following is a summary of these basic design considerations as they apply to a thrust stand force transducer.
3.1.2.1
Mechanical
(1)
The
Design type
of
force
transducer
is
determined
by
the
direction of the applied load as tension, compression, or universal, combining both tension and compression. The universal type would normally be used in thrust stands. Where preloading is used a compression type could be employed. (2)
The physical dimensions, type of mounting, mounting dimensions, and type of force connection will vary with manufacturer, but all transducers can generally be adapted for thrust stand installation.
(3)
The load-line linkage from the thrust frame to the force transducer should be designed such that only axial loads are transmitted. When using a directly connected load line, flexural linkages are normally used on both ends of the transducer. Preloaded installations contact
(4)
A series thrust
(5)
use a hard mount on the fixed on the active end to eliminate installation measurement
of is not
force
end and point off-axis loads.
transducers
for redundant
recommended.
A parallel installation of force transducers is not recommended for single-component axial thrust measurement. This arrangement reduces the output signal sensitivity and compounds inaccuracies and nonreliabi!ity.
I-6
(6)
The electrical connections standard connectors mounted
can on
be either multipin the transducer outer
case or a flexible cable providing the and shielding for each bridge circuit.
0
required
conductors
(7)
Load range selection should be made such that the transducer is normally operating over the 75 to 90% portion of its full scale range. An overload of 150% of full scale range should produce no degradation in specifications. The ultimate rating for structural failure should be 300% of capacity or greater.
(8)
Derated (downranged) force transducers are used to increase overload capability, increase thrust system stiffness and therefore dynamic measurement response, or extend the static force measurement range to lower values. If derated by a factor of 10:1 or more, the force transducer will probably have to be specially selected and thermally compensated to obtain the desired accuracy.
(9)
Deflection of the force transducer's strain element must be minimized to keep the thrust stand total deflection low and dynamic response high. Total deflection over the full scale load of 0.044 to 4400 N (10 to 1000 ibf) varies from 0.0013 to 0.013 cm (0.0005 to 0.005 in.) depending on the type of strain element used.
(1o)
The natural frequency of the transducer should be several times greater than the thrust stand resonant frequency; a value of 1000 Hz or higher is recommended.
(11)
A safe operating temperature range that is compatible with the test cell environment and that will not cause permanent calibration shift or permanent change in any of its characteristics should be specified for the force transducer. Normal compensated temperature range should be around 0 ° to 66°C (32 ° to 150°F). Maximum operating and storage temperature range should be around -54 ° to +93°C (-65 ° to ÷200°F). Extended temperature ranges are possible with special design provisions. Due to the highly transient nature of the operating temperature that can exist in a test cell, it is recommended that an independent temperature controlled environment be provided for the force transducer. Experience has shown that thermal drift can prove to be the highest single source of error in thrust measurements.
(12)
Barometric test cell transducer
pressure decrease vacuum should not characteristics.
I-7
from cause
ambient to complete any change in
(13)
Material selection for all transducer components should acknowledge possible contact with hydrazine decomposition products. Metal parts which experience temperatures above 93°C (200°F) while exposed to ammonia may be nitrided (Reference I-2). Also, stress corrosion cracking is accelerated under these conditions. An inert gas purge of the transducer case or the surrounding environment will prevent these effects.
(14)
Vibration and acceleration along the load axis resulting from rough engine operations should be considered separately from the static overload requirements. The strain gage mounting and electrical connections are the transducer elements which are most susceptible to these effects. Bonded strain gages with well-designed terminal pads and supported lead wires are recommended.
Electrical
Design
(I)
The
type
of s_rain
_a_e
should
be bonded
metallic
(2)
The number of active strain ga_e bridge arms should be four for a single bridge and eight for a double bridge. A double bridge is recommended for load ranges of 222 N (50 Ibf) and above.
(3)
Recommended excitation should be a regulated voltage of 5 to 28 V dc. Maximum excitation voltage which will not permanently damage the transducer should be 20% or greater than the rated voltage.
(4)
InDu_ and output resistance for dc excitation in ohms at a specific temperature in °C (OF). values range from 120 to 350 ohms.
(5)
Electrical connectiQn_ for conform to pin designation
each bridge circuit shown in the wiring
foil.
is specified Normal
should schematic
diagram (Figure I-2). The output polarities indicated on the wiring diagram apply when an increasing force is applied to the transducer.
(6)
Insulation resistance is specified in megaohms at a specific voltage and temperature between all terminals or leads connected in parallel and the transducer case. A value of 5000 megohms or greater is recommended.
(7)
Shunt calibration resistance should be specified in ohms for a nominal percentage of full scale output at _ specific temperature. The terminal across which the resistor is to be placed shall be specified.
I-8
O
A
(+) EXCITATION
B
(+) OUTPUT
D
(-)
EXCITATION
E
(-)
OUTPUT
SHUNT
F
CALIBRATION
RESISTOR
I
___jJ CONNECTOR
NOTES: 1.
THE OUTPUT POLARITIES INDICATED ON THE WIRING DIAGRAM APPLY WHEN AN INCREASING ABSOLUTE PRESSURE iS APPLIED TO THE PRESSURE PORT (SENSING END) OF AN ABSOLUTE PRESSURE TRANSDUCER. FO R DIFFERENTIAL AND GAGE PRESSURE TRANSDUCERS, THE INDICATED POLARITIES APPLY WHEN THE ABSOLUTE PRESSURE AT MEASURAND PORT IS GREATER THAN THE ABSOLUTE PRESSURE AT THE REFERENCE PRESSURE PORT.
2.
THE BRIDGE ELEMENTS SHALL BE ARRANGED PRODUCING POSITIVE OUTPUT WILL CAUSE IN ARMS 1 AND 3 OF THE BRIDGE.
3.
POSITION OF ANY INTERNAL COMPENSATION INDICATED. SEE APPENDIX III-C.
Figure
I-2.
3.1.2.3 of
performance
measurement
specified, 23
±2°C
pressure
Conventional
they
98
transducers
apply
at
+__.6°F); ±10
(I)
kN/m
the
2
(29
Range
is
End
or
points
citation
are or
SHOULD
BE
Bridge
basic below.
ambient
humidity
usually
compression
(2)
listed
following
+__3 in.
Gage
The
are
relative
NETWORK
Strain
Chgracteristics.
force
(73.4
6-Wire
SO THAT FUNCTIONS INCREASING RESISTANCE
Transducer
performance Unless
parameters otherwise
conditions:
90%
maximum;
temperature barometric
Hg).
expressed tension
in or
expressed
millivolts
I-9
as at
newton
(pounds
force)
both.
millivolts
specified
per volts
volt of
of
ex-
excitation.
(3)
Full scale excitation
(4)
Zero-measured output output at fuI1 rated
(5)
Stability refers to the ability of a transducer to retain its performance throughout its specified operating and storage life.
(6)
L_nearitv is expressed as in direct!on(s) of applied
(7)
Hysteresis is expressed as percent of full scale output upon application of ascending and descending applied forces including rated force.
(8)
Combined hysteresis and linearity are of full scale output upon application and descending forces including rated
(9)
Repeatability is expressed as percent of full over a specified time period with a specified of cycles of load application.
output (FSO) or millivolts
is expressed at specified
as millivolts per volt volts of excitation.
is expressed as percent of full scale excitation and no applied load force.
percent of loading.
full
scale
output
expressed as percent of ascending force. scale output number
(10)
Static error band and repeatability output as referred
(11)
Creep at load is expressed as percent of full output with the transducer subjected to rated for a specified time period.
(12)
Creep recovery is expressed as percent of full scale output measured at no load and over a specified time period immediately following removal of rated force, that force having been applied for the same period of time as specified above.
(13)
Warm-up period the application that subsequent not exceed the
(14)
Static sprin_ (pounds force
is the combined linearity, hysteresis expressed as percent of full scale to a defined type of straight line.
is that time period, starting with of excitation, required to assure shifts in sensitivity and zero will specified percent of full scale output. 90nstant is expressed in newtons per meter per inch) as determined experimentally
or analytically from the dynamic spring mass system representing
(15)
scale force
response the force
of a simplified transducer.
_qNivalent dynamic masses are expressed in kilograms (pounds mass) for both ends of the transducer in the system
described
above.
1-10
(16)
Internal mechanical damping is expressed in newtons per meter/second relative velocity (pounds force per inch/ second relative velocity) between ends at a specified frequency in hertz and a specified dynamic load in newtons '(pounds force).
(17)
Safe overload rating is expressed as an applied force in newtons (pounds force) for a specified time period in minutes which will not cause permanent changes in transducer performance beyond specified static error band.
(18)
Rated axial force
(19)
Thermal sensitivity shift is expressed as percent of sensitivity per specified °C (OF) temperature change over the compensated temperature range in °C (OF).
(2O)
Thermal zero shift is expressed as percent of full scale output per specified °C (OF) temperature change over the compensated temperature range in °C (OF).
(21)
Temperature error b_nd is expressed as output values within a specified percent of full scale output from the straight line establishing the static error band over the compensated temperature range in °C (OF).
(22)
Tem.Perature gradient error is expressed as less than a specified percent of full scale output while at zero load and subjected to a step function temperature change over a specified range in °C (OF) lasting'for a specific time
forc_ of the transducer is the maximum designed force in newtons (pounds force). The rated may be in compression, tension, or both.
period in minutes transducer.
and
applied
to a specific
part
of
the
(23)
Cycling l_fe is expressed as a number of full scale cycles over which the transducer shall operate without change in characteristics beyond its specified tolerance.
(24)
Other environmental CQn_t_ons transducer performance beyond be listed, such as: (a)
Shock,
(b)
High-level
(c)
Humidity
(d)
Corrosive
spray
(e)
Corrosive
gases
which should not change specified limits should
triaxial acoustic
1-11
excitation
(f)
Electromagnetic
(g)
Magnetic
(25)
fields
fields is expressed
as
a specific
(days, months, years) thetransducer in a specified environment without performance characteristics beyond tolerances. (26)
time
period
can be stored changing specific their specifi@d
Concentric angular load effect is expressed as percent of full scale output difference from true output (axially loaded output multiplied by cosine of angle) resulting from a load applied concentric with the primary axis at the point of application and at a specified angle in degrees with respect to the primary axfs.
(27)
Eccentric
an_ular
load
effect
is
expressed
as
percent
of
full scale output difference from true output (axially loaded output multiplied by cosine of angle) resulting from a load applied eccentric with the primary axis and at a specified angle in degrees with respect to the primary
(28)
axis.
Eccentric
load
effect
is expressed
as percent
of
scale output to difference from axially loaded resulting from a load parallel to but displaced specific distance in millimeters (inches) from tricity with the primary axis. (29)
3.1.3
output a concen-
AI_bient pressure effects are expressed as the change in sensitivity and the change in zero measurand output due to subjecting the transducer to a specified ambient pressure change above or below atmospheric.
Calibration
Equipment
Calibration
equipment
for
the
thrust
measuring
system
separated into three types. The in-place static calibrator dynamic calibrator are used on the thrust stand in the test third type involves the laboratory calibration of the force
3.1.3.1
full
_j]m_ce
_atic
Calibrator.
It
is essential
is
and the cell. A transducer.
that
the
thrust
measuring system be provided with in-place static calibration equipment that can produce the same axial deflection on all critical restraints as will be present during the actual engine test. This equipment should be capable of remote operation within the test cell under both pretest and posttest conditions of simulated altitude (vacuum) and temperature and in the presence of all contributing forces including mechanical line and flexure effects, pressurized propellant supply lines, and the
dynamic
influence
of all
supporting
1-12
systems.
static
The following calibrator:
items
should
be
included
in the
design
of
the
in-place
(i)
The calibrator should be either a multipoint dead weight type or a working standard force transducer with a variable force generator.
(2)
The calibration range should be from 0 to 120% of the highest anticipated normal load with values established at three or more intermediate steps.
(3)
The application of the calibration loads should be in both ascending and descending steps that are repeatable in either a preprogrammed or selective manner, so that the transition from one force to the next is accomplished without creating a hysteresis error in the measurement due to overshoot.
(4)
The
|
maximum
inaccuracy
of
the
force
calibrator
not be more than one half the permissible of the measurement force transducer.
should
tolerance
(5)
Traceability (not more than National Bureau of Standards for the force calibrator.
(6)
The calibration force should be applied coincident with the thrust axis of the test engine and in the same direction as the deflection produced by engine operation.
(7)
The calibrator support structure should be independently mounted on the seismic mass, separate from the fixed structure of the measurement force transducer.
(8)
The calibrator force linkage should include a disconnect coupling that will isolate the calibrator from the thrust frame when not in use.
(9)
Provisions of as
(lO)
the the
for adjustment,
calibration measurement
twice removed) to the should be established
alignment
force must force.
and
be given
verification
the
same
emphasis
Protection against damage from shock, vibration, radiation or corrosion caused by the test cell environment should be provided for the calibrator.
3.1.3.2 Dynamic Calibrator. The dynamic response of a thrust stand depends on the stiffness and mass distribution throughout the entire active system. The response characteristic can best be determined by experimentally applying a suitable time-varying force to the complete sxstem and measuring the output versus time response of the measurement force transdu6er. The time-varying of producing
force can _inusoidal
be supplied forces or
by an electromagnetic step function forces.
1-13
forcing coil capable A simpler method is
to use
a single
instantaneous
force
change
of
sufficient
magnitude
to
cause the thrust stand to ring. In either case, data can be developed which provide a measure of thrust stand sensitivity, resonant frequency, and damping factor. The dynamic calibrator should be capable of remote operation with the thrust stand and test cell at simulated engine Operating conditions. The dynamic
force
following
items
should
be included
in the
design
of
the
calibrator:
(1)
The dynamic force should be applied to the thrust frame coincident with the measurement thrust axis and in the same direction operation.
(2)
The
dynamic
as
force
the
deflection
calibrator
produced=by
should
be
the static force calibrator so as not damage to the sensitive equipment.
engine
independent to dause
of
any
(3)
Due to the added weight and complexity, the dynamic calibrator should be disconnected from the active mass of the thrust frame. It may even be removed from the thrust stand completely and only installed when required.
(4)
Magnetic forcing coils should be thoroughly characterized using simulated engine and thrust frame mass before being employed on the thrust stand. The generated force input should be tailored to match the anticipated operating characteristics of test engine thrust pulse whenever possible.
(5)
The application of an instantaneous force change using falling weights, swinging hammer, frangible preloaded linkage, electric solenoid hammer (pinger), etc., should be thoroughly evaluated for causing possible shock overloads to the test engine assembly or the measurement force transducer.
3.1.3.3 Laboratory Calibration. Acceptance and qualification tests should be performed on the strain gage force transducer in a calibration laboratory before the transducer is installed into the thrust stand. The basic equipment consists of a force calibrator, a source of electrical excitation for the strain gage, and a device which measures.the electrical output of the transducer. The error or uncertainties of the laboratory calibration system should be less than one third of the permissible tolerance of the measurement force transducer. Traceability (not more than twice removed) to the National Bureau of Standards should be established for the laboratory equipment. The a laboratory
following
force
is
calibration
a summary system
1-14
of as
the given
design
considerations
in Reference
I-I:
for
(I)
i
3.2
The force calibrator should have a maximum uncertainty of no more than one fifth of that permissible for the measurement force transducer. Typical values for three types of force calibrators in either tension or compression are: (a)
Dead
weight:
(b)
Proving
(c)
Reference standard force error ÷-0.1% full scale.
ring:
maximum
error
maximum
error
±0.01% +_0.1%
transducer:
test
load
full
scale
maximum
(2)
The range of the instrument supplying the calibration force should provide accuracy to 125% of full scale range
(3)
A stable source of electrical excitation of accurately known amplitude is required for the strain gage bridge circuit. DC excitation is normally provided by electronically regulated power supplies utilizing line power. The maximum uncertainty for these excitation power sources should be ±0.02% of reading or better.
(4)
A digital electronic voltmeter with preamplifier is normally used to measure the electrical output signal of the force transducer. The maximum uncertainty for this type of instrument is (a) for 0 to 10 mV range with I _V sensitivity maximum error ±0.02% of reading or +_2 _V, whichever is greater or (b) 0 to 100 mV with 10 wV sensitivity maximum error +_0.01% of reading or ±10 _V, whichever is greater.
ELECTRICAL
AND
ELECTRONIC
or monitoring the necessary of the transducer.
COMPONENTS
The major electrical and electronic components to be considered in the design of a thrust measurement system include (I) signal conditioning equipment, (2) electrical calibration equipment, (3) recording equipment, (4) visual display equipment, and (5) data processing equipment. These components are shown in the thrust measurement system block diagram, Figure I-3. In general all of these components are commercially available, off-the-shelf items. Most of the items are available from more than one manufacturer. The major concern in the selection of these components must be the evaluation of the various manufacturers' general specifications in relation to the specialized force measurement requirement. A subsequent verification that the equipment finally selected conforms to the manufacturers' specifications is essential.
3.2.1
Signal
Conditioning
Equipment
Signal conditioning equipment includes the following functional devices: power supplies, amplifiers, electrical cabling, shielding, signal d£stribution and switching network, and filters. The regulation and stability
1-15
ENGINE ROCKET ASSEMBLY
I
J
CA LIBRAT IO N DYNAMIC SYSTEM
CA LIBRAT ION SYSTEM STATIC
t
J
FRA/vI£ THRUST
J
J
MECHANICAL COMPONENTS FORCE TRANSDUCER MEASUREMENT
1
ELECTRICAL AND ELECTRON!C COMPONENTS
CALIBRATION E I..ECTRICA L EQUIPMENT
t
SIGNAL CONDITIONING EQUIPMENT
i
I RECORDE R OSCILLOGR1APH
/_,NALOG
GRAPH IC RECORDER
1
, FfM
DIGITAL ANALOG-TORECORDER
RECORDER
VISUAL , DISPLAY
J
Figure
I-3.
Thrust
Measurement
i-16
DATA
System
PROCESSING
Block
EQIJIPMENT
Diagram
I
of this equipment should be ±0.05% or better, wherever applicable. The design of these devices varies widely depending on system philosophy and economics; however, certain design principles are universally recommended as follows:
(1)
PoweY Supplies. Constant voltage excitation is the primary type used with hlgh-accuracy thrust measurement systems. Power supplies can be individually rack-mounted units or miniature (several in one card) devices integral with other signal conditioning equipment. Generally, there are provisions for voltage adjustments, less often for zero balance. Ripple should be less than 100 MV peak to peak.
(2)
Amplifiers. The use of high-quality differential amplifiers is now almost universal. With a transducer full scale output of 20 to 40 mV, an amplifier gain of 50 to 500 is sufficient for most conventional analog-todigital conversion and recording systems. It should be verified that peak common mode voltages do not exceed the rejection limits of the amplifier.
(3)
Electrical Cablin_ and Shielding. Electrical noise can be minimized by the use of proper shielding and grounding techniques (Reference I-3). Transmission cables between the transducer and the recording system usually consist of multiple pairs of twisted, shielded, splice-free conductors. A total of 6 conductors should be used for each transducer. The wire gage and corresponding resistance of the excitation and calibration leads should be taken into account when developing calibration techniques and other system design considerations. Each transducer cable should be individually shielded, with continuity of shield to the operational amplifier. The shield grounding connection should be in accordance with the amplifier manufacturer's recommendations. Multichannel cables consisting of inner cable shielding, and overall shielding of the large cable are recommended for long transmission runs. The outer shield of the multiconductor cable should also be terminated at a single point ground. The outer shield and all shields should be insulated from each other.
(4)
e
inner
Other signal conditioning eauipment includes such items as filters, distribution and switching units, and impedance matching devices. The design of these and related devices varies depending on system philosophy, but should in all cases be high-quality equipment providing stability (both with time and temperature), line voltage regulation, and linearity. Thermally induced errors should be minimized in all circuits.
1-17
3.2.2
Electrical
Calibration
Equipment
Some form of electrical simulation of the transducer response to force should be provided. This simulation should track any change in the system sensitivity that is caused by changes in the environmental conditions. The two most commonly used electrical calibration systems that are readily adaptable to automated periodic tests are discussed below. The first of these techniques includes the transducer in the calibration and is thus quasi end to end. The other involves only the:electrical and electronic equipment. The advantages (convenience, technical, or economic) of each system will largely depend upon the user's existing transducers, signal conditioning equipment, cabling, etc.
3.2.3
(i)
Shunt calibration with a constant voltage system requires a 6-wire system to the transducer if an external signal shunt resistor is used: 2 wires are used for excitation, 2 for output, and 2 for shunt simulation; The technique for using a 6-wire shunt resistor calibration method is presented in Appendix III-C.
'(2)
Voltage substitution techniques can be used to calibrate the electronics system (amplifier, recorder, etc.) in addition to, or in lieu of, any transducer electrical or end-to-end calibration. This method requires that the transducer be electrically disconnected (usually by a switching network) and a known voltage substituted. Such a calibration technique will not necessarily provide any information about changes in ambient output nor will it even reveal if thetransducer has been disconnected. It should therefore not be the only type of electrical calibration employed.
Recording
Equipment
• The four commonly used recording systems for recording thrust measurement, data are (I) digital system, (2) graphic recorders, (3) oscillograph and (4) analog magnetic tape recorders. Two or more of these systems may: be combined to provide high accuracy, high-frequency response, and quick readout.
3.2.4
Visual'Display
Equipment
A visual display of real time measured data in engineering units along with other critical operating parameters is required for both pretest and posttest calibrations and for monitoring during the engine test. This alphanumeric data can be presented in either hard copy (printed) or non-hard-copy (no record) form or both. The non-hard-copy is usually displayed on some type of cathode ray tube (CRT) device through a selective preprogrammed format.
1-1S
3.2.5
Data
Processing
The exclusively
data
to
many
other
with
care
the
processing thrust
parameters so
Equipment
that
also.
the
equipment
measurement This
is
not
system, equipment
uncertainty
of
the
generally
but
is
should data
is
dedicated
used
be not
to
chosen
process and
increased
used
during
processing.
4.0 4.1
INSTALLATION
COMPONENT
All should
be
the
critical for
This
all
data
as
transducers such
tests
compliance force one
working-standard
(3)
Calibrator
dead
(4)
Thrust
stand
flexures
(5)
Thrust
stand
restraints
(6)
Dynamic
Force
Transducer
thrust
in makes
may
be
transducer the
Acceptance those
parts, to the
assembly, next.
The I-I
are
summarized
as
transducer is
to
tests
I-I.
for
1-19
(Where
might
the
a vary
force
quantity
of
able
certify
to
working-standard
of
tests of
considerably
transducer
certi-
difficult,
force
function
be
recording
testing The
uncertainty
acceptance
and
the
is
calibrator
should
tests
measurement
are
the
used,
units.)
an
the
which
follows:
be
manufacturer
of
and
to
and
qualification
similar
Individual
adjustment,
acceptance
plus
certified
transducer
qualification
the on
certified
Test.
force
one
of
tolerance
characteristics or
if and
provided be
inspection so
transducer
Reference
testing
should
to
force
justification
permissiSle
4.1.1.1 evaluate
ISA-$37.8,
waived
precision
system before
calibrator
acceptance
previous
measurement
specifications
weights
transducer,
the
used
force
measurement
force given
force
thrust
Components
Calibrator
measurement
the
by
evaluation. items:
(e)
from
half
accomplished
test and following
performing
PROCEDURES
design
Thrust
working-standard by
best
of
with
(i)
The
fied
components
compliance
is
standard laboratory should include the
4.1 .I
CHECKOUT
CERTIFICATION
certified
installation.
AND
not
more
than
transducer.
are
performed
transducer from
from
one
piece unit
Reference
(i)
(2)
Inspect the transducer for mechanical defects, finish and improper identification markings. Install
the
transducer
into
the
force
with axial alignment as specified Connect the excitation source and
poor
calibrator
by the manufacturer. readout instrumenta-
tion to the transducer and turn on for the specified warm-up period. Prior to calibration, it is desirable to exercise the force transducer by applying rated load and returning to zero load for several cycles.
(3)
(4)
(5)
Conduct two or more calibration cycles in succession, including five to eleven points in both ascending and descending directions while monitoring excitation amplitude and recording output signal. From the data obtained in these tests the following transducer characteristics should be determined: (a)
End
points
(b)
Full
scale
(c)
Zero
unbalance
(d)
Linearity
(e)
Hysteresis
(f)
Hysteresis
(g)
Repeatability
(h)
Static
,
and
error
linearity
combined
band
Repeat the calibration cycles over a specffied period of time after warm-up. This data establishes the following characteristics for that period of time: (a)
Zero
drift
(b)
Sensitivity
drift
Apply the rated force to the transducer during a specified short period of time. The measurement of changes in output at constant excitation during the time period should establish: Creep
...
output
at
load
,
(6)
At zero
load,
measure
output
and
sensitivity
over
a period of time, up to one hour, starting with the application of excitation to the transducer. The observed time to stabilize should determine the following
characteristic:
1-20
Warm-up
(7)
B
Apply the specified rated overload for the specified number of times in the specified direction of tension or compression, followed by one complete calibration cycle to establish that the transducer performance characteristics are still within specification. This should verify: Safe
(8)
period
Measure
overload
the
rating
insulation
resistance
between
all
terminals,
or leads connected in parallel, and the case of the transducer with a megohm-meter using a potential of 50 V, unless otherwise specified, at room temperature. This should establish: Insulation
(9)
resistance
A Wheatstone bridge (for dc) or an impedance bridge (for ac) should be used to measure and establish the following: (a)
Input
(b)
Output
impedance impedance
4.1.1.2 Qualification Test. Qualification tests are performed to evaluate those characteristics which are a function of transducer design. They thus would not be expected to vary appreciably from one unit to another for a particular transducer model. The performance of a representative sample of units should represent the performance of an entire lot.
I-I)
are
The qualification tests summarized as follows: (I)
The
force
for the
transducer,
force
while
transducer
still
installed
(Reference
in the
force
calibrator, should be placed in a suitable temperaturecontrolled chamber. After stabilizing the chamber and transducer at a specified temperature, one or more calibration cycles should be performed within the chamber. The procedure should be repeated at an adequate number of selected temperatures within the specified operating range of the transducer, and finally after return to room temperature and stabilization. From these tests the following characteristics should .be determined:
(a)
Thermal
sensitivity
(b)
Thermal
zero
(c)
Temperature
shift
shift error
1-21
band
(2)
With the force transducer at room temperature within the controlled chamber, and with no applied load, it should be subjected to a specified thermal transient above or below room temperature. The observed output over a specified period of time establishes: Temperature
(3)
Install
the
transient
force
error
transducer
into
a suitable
test
fixture with the axial alignment as specified. Attach an electromagnetic forcing coil which is capable of producing an axial load which is 20% of the force transducer rated 10ad or higher. Connect the excitation source, readout instrumentation, and recorder and turn on for the specified warm-up period. Apply a sinusoidal force at a specified load and sweep the frequency from zero This
until the first resonant peak should establish transducer:
(a)
Natural
(b)
Frequency
Apply time
a step period.
resonant
output
is
reached.
frequency
response
function The
(a)
Natural
(b)
Static
(c)
Damping
(d)
Equivalent
force
recorded
resonant spring
at
a specified
output
should
load
and
establish:
frequency
constant
factor dynamic
masses
After completion of the above test, remove the transducer from the test fixture and reinstall it into the force calibrator. Perform one complete calibration cycle to verify the satisfactorily.
(4)
ability
of
the
transducer
to
perform
Install the force transducer into a suitable vacuum chamber with the axial alignment as specified. Connect the excitation source and readout instrumentation to the transducer and turn on for the specified warm-up period. With the transducer at room temperature and no applied load, reduce the barometric pressure within the test chamber in a controlled manner to a specified level below atmospheric and hold. Observe the output over a specified time period to establish: Ambient
pressure
effects
After
completion
of
test,
remove
transducer
the
two or
1-22
/
more from
cycles the
of the vacuum
above
chamber
and
reinstall
into
the
force
calibrator.
Perform
one complete calibration cycle to verify ability of the transducer to perform satisfactorily. O
(5)
Install
the
force
transducer
into
a suitable
test
fixture with the axial alignment as specified. Mount this test assembly onto a vibration and/or shock test machine in the axial loading direction as specified. Connect the excitation source and readout instrumentation to the transducer and turn on for a specified warm-up period. With the transducer at room temperature and zero initial applied load, subject the transducer to a specified series of vibration and/or shock tests
i
at prescribed output during establish: Vibration
frequency and g loading. each of the specification
and/or
shock
effects
After completion of the above test, from the test fixture and reinstall calibrator. Perform one complete verify ability of the transducer
(6)
Observe the tests to
remove the transducer into the force
calibration cycle to to perform satisfactorily.
To determine the effects of concentric angular loading (and side loading), insert wedge blocks above and below the force transducer installed in the force calibrator as shown in Figure I-4a. The angle subtended by the two larger surface areas (b) of each block should be equivalent to the specified angle of interest and should result in the specified side load. Apply rated load and as soon as stabilized, read output and remove the load. A comparison of pretest and posttest zero measurand should establish: Concentric
(7)
loading
effects
I
To determine the effects of eccentric angular loading, remove the upper wedge block as shown in Figure I-4b. Apply rated load and as soon as stabilized, read output and remove the load. A comparison of pretest and posttest zero measurand should establish: Eccentric
(8)
angular
angular
loading
effects
I
To determine the effect of eccentric loading, wedge blocks, install a flat load button into
IThe effects of the various types of loading related conditions can be determined in accordance with the in Figure I-4.
1-23
remove both the transducer,
to axial loading expressions included
c
_u
_< L
\ \
Z U
,..l_
Z
<
-x
>.0
a 8 o <
8
.....
z
\ \
u
c _.-Z
J_
w
c 0
b f_,"l
o_
I o
,.f e-
0
z
113
o
_,_
! -O,--
\
< u
_J
w
,
Z u
i
I
|
I
o_"I
z
,-.1
O
1.0
DAMPING COE'FFICIENT " C
0.6 I=,-
_o _
0.4
IO I
-....4
0.2
,
o 0.1
o -
0.2
0.3
0.4
0.5
0.6
0.7
I
0.8
0.9
1.0
DAMPING RAI"IO, h
Figure
NATURAL
'Damping
pressure
system
containing
transducers)
the
a
where fn constant
is of
at
the
which
called
the
do
free
the the
output
the
natural
not
the
The
natural
of
at at
I/v_,
damping,
no
response
occurs
real
the peak
maximum is
shown
is
often at
Only
natural the
90 ° .
if
in
does
term
reality, affect
applied be
this the
a
(typical
by:
the as
(fn)
the
spring the frequency
is
damping
is at
Figure
3E-6
at
to
damping As
the
occurs zero
all. III-E-7.
a
sometimes
has
period
at
transducer
frequency
maximum
response
response exists
restraint
given
and k is be defined
This
In
frequency.
peak
spring is
nothing which
frequency).
frequency
which
the
but
will
frequency.
occur
frequency
ratio
Overshoot
(III-E-3)
by
input
there
a
of the system It may also
input
(ringing
sinusoidal
a maximum.
response the
a
recorded,
and
frequency
freauencv.
frequency
occur
If is
the
natural
oscillations
at
From
= u 2=
frequency mass m.
lags
undamDe_
with
output is
natural suspended
mass
natural
fn
to
Ratio
FREQUENCY Ina
of
III-E-6.
The
at
and
which
is
the
is
zero
does
damping
is
is
resonant
lowered
frequency. frequency
at
response frequency,
the
maximum
increased, and,
For
the
the
at
a damping
greater
which
maximum
[
I
I
I
I
I
I
1.0 0.8 z 0.6
Z
0.4
°,9, I 0.2
0
I
I
I
I
J
I
0.1
0.2
0.3
0.4
0.5
0.6
IL
1
0.7
O.B
DAMPING RATIO, h
Figure
III-E-7.
If to
execute
decay to the
a
is
it.
transducer
typical
This
is
frequency the
of
oscillation
of
natural
the
frequency)
oscillatory
transient
As and,
nature
of
transient
the at
the
a
oscillation
I
I
I
I
a
is
and
freauency
ratio
at
a
in
I
allowed the
may zero
be
of
(by
The
I
1.0
(5 0.8 Z
8 _
o.6
u.
0 z
RI
5 o.4 _z 0.2
0 0
I
i
I
I
I
I
I
I
i
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
DAMPING RATIO, h
Figure
III-E-8.
Ringing
3E-7
Frequency
vs
Damping
same
definition),
III-E-8.
]
is
freouency
disappeared.
Figure
I
assigned the
the
unity
transient
damping
freauency)
entirely
plotted
then
that
increased,
has
I
Ratio
found
(ringin_
damping
is
be
Only
dampln_
transient
Dampin_
displaced may
and
frequency.
decreases
of
it
natural
frequency.
oscillatory
frequency
the
vs
is
transient,
(ringing
not
Frequency
diaphragm
damped
oscillatory
as the
the
Resonant
Ratio
1.0
FREQUENCY RESPONSE (AMPLITUDE) The amplitude of the system transfer function versus frequency is often referred to as the frequency response. It is freouently normalized
to
show
deviations
illustrates with
the
various
response there
damping
of is
a
a
single
system
(h
mined
<
by
approximation can
be
step
this
input
slightly the
Assume
than ratio
i
the
from a
is
a
good
sinusoidal
Figure order
III-E-9 systems
representation pressure
frequency
the
curves
and
natural
step
of
the
variations
response of
of
Figure
input
and
an
if
underdamped
III-E-9
frequency
pressure
a
40% h
I I I
a
(Figure
can
flush
ringing be
III-E-3).
determined
i
mounted
once
have the
transducer
frequency
_
of
1000
From to
been
analysis
be
Hz
the
whose with
curve
the determethods
response
to
overshoot
of
Figure
III-E-6
in
approximately
0.4.
T
I
2.0
Knowin_
I
the
I
h=C
-
1 1.5
one. second
frequency.
ratio, of
of of
appendix.
indicates
less
damping
made
ratio input
plot to
of
damping
Example: a
This
resonant
application
in
amplitude
sinuscidal
transducer
major
1.0)
the
a
ratios.
frequency,
described
an
to
pressure
An
ringing
from
response
h=0.3
:0
AMPLITUDE RATIO = V( I-/32) 2 + f2hB) 2 h=0.5 h = 0.707
/9= THE RATIO OF THE VARIABLE INPUT PRESSURE FREQUENCY, f, TO THE SYSTEMNATURAL FREQUENCY, f n
I 0.06
_ I I 0.08 0.I
i 0.15
0.3
0.2
0.4
0.5 0.6 0,7
1.0
I .5
2.0
FREQUENCY RATIO B = f/fn
Figure
III-E-9.
Frequency
Response
3E-8
of
a Second
Order
System
3.0
damping
ratio
III-E-8) Figure of
and
is III-E-9
1100
the
it
They
will
be
with
less
than
response
to
10%
can
be
ringing be
can
output
In a
the
found
frequency,
approximately
be
shown
readings high
at
case
10%
overshoot,
of
about
fn
(440
Hz)
an
overdamped
from
the
by
f3dB
50%
during
(3 a
is
the
frequency
dB)
and
T is
test.
While
transducer
under
conversion
is
the plus
95%
by
time
III-E-I,
time
this
with
5%
a
at
high
critically time
Using
natural
0.8 at
(±10%)
(Figure
Hz. fn
0.25 damped
for
the
frequency or
880
fn
(275
the
transducer,
data:
output
amplitude
of
transducer
it T in
3-9).
Hz).
(III-E-4)
the
can
conditions,
Hz.
frequency
I : T
constan_
(Reference
1100
high
response
which
replacing
40% and
frequency
or
value
relationship
test
obtained
response
Figure
this
ideal
at
the
natural
system
or
a coarse
determined
a
be
f3dB where
for
will 0.4
the
that
the 1000/O.91
be
appears
the See
applied
as
to
that
above also
is an
a more
equation, Rise
and
reduced measured ideal realistic with
t95%,
Response
Time
appendix.
REFERENCE
III-E-I.
Burns, Statham
J.,
and
Rosa,
Instrument
G., Notes,
3E-9
CalibratiQn Number
and 6,
Test
December
of 1948.
Accelerometers,
SECTION
TEMPERATURE
e
IV
MEASUREMENT
SECTION TEMPERATURE
IV
MEASUREMENT
CONTENTS D
f
1.0
INTRODUCTION
4-I
2.0
SCOPE
4-I
2.1
OBJECTIVE
4-I
2.2
LIMITATIONS
4-2
3.0
DESIGN
4-2
3.1
TEMPERATURE
3.1.1
Transduction
3.1.2
Performance
3.1.3
Mechanical
Design
4-7
3.1.4
Electrical
Design
4-11
3.2
ELECTRICAL
AND
3.2.1
Signal
3.2.2
Electrical
3.2.3
Recording
3.2.4
Visual
3.2.5
Data
4.0
PERFORMANCE
4.1
TRANSDUCER
4.1 .I
Visual
Inspection
4.1.2
Range,
Output,
4.1.3
Repeatability
o
CONSIDERATIONS TRANSDUCERS
U-2
Element
_-2
Characteristics
ELECTRONIC
Conditionin_
4-5
COMPONENTS
Eauipment
Calibration
4- 16
Equipment
Processing
4-16
Eauipment
4- 16
VERIFICATION TESTING
and
4-12
Equipment
Eauipment
Display
4-12
AND
2-16 CALIBRATION
4-16 4-18
Error
Limits
4-18 4-18
4.1.4
Insulation
4.1.5
Dynamic
4!1.6
Proof
4.1.7
Thermoelectric
4.1.8
Vibration
4.1.9
Other
Surface
4.1.10
Other
Environmental
4.2
SYSTEM
4.3
STANDARDS
4.3.1
Temperature
4.3.2
Readout
5.0
OPERATING
5.1
PRETEST
5.2
POSTTEST
6.0
DATA
6.1
CALIBRATION
6.2
RUN
7.0
GLOSSARY
4-26
8.0
REFERENCES
4-30
4-18
Resistance
4-18
Response Pressure
and
4-19 Potential
4-19
Acceleration
4-19
Temperature
Transducer
Tests
Tests
4-20 4-20 4-20
CALIBRATION
4-21 4-22
Source
4-23
Instrument
4-23
PROCEDURES PROCEDURES
4-23 4-24
PROCEDURES
ACQUISITION AND
AND
PROCESSING
VERIFICATION
TECHNIQUES
4-24 4-24
DATA
DATA
4-25
APPENDIXES IV-A
IV-B
TEMPERATURE MEASUREMENT UNCERTAINTIES
SYSTEM
ELEMENTAL 4A-1
TABLES AND EQUATIONS FOR USE AND RESISTANCE THERMOMETERS
4-ii
WITH
THERMOCOUPLES 4B-I
4-I
Temperature Measuring System
4-2
Basic ThermocoupleCircuit
4-3
Measuring Circuits Transducers
for Resistance Thermometer
IV-B-I
Characteristics
Common
IV-B-2
Callendar-Van
IV-B-3
Thermocouple Materials, Type Letters, and Color Codes for Duplex Insulated Thermocouples and Extension Wires
4_-3
Fixed Points Thermocouples
4B-4
4-14
Tables
IV-B-4
of Dusen
Thermocouple
Materials
Equation
Available for Calibrating and Resistance Thermistors
f
4-iii
4B-I 4B-2
SECTION IV TEMPERATURE MEASUREMENT 1.0 INTRODUCTION Recommended practices are outlined for the design, installation, checkout, calibration, and operation of a temperature measuring system to be used during tests of a liquid monopropellant rocket engine. Twoappendixes are included: IV-A, Temperature MeasurementSystemElemental Uncertainties, and IV-B, Tables and Equations for Use With Thermocouples and Resistance Thermometers. 2.0
SCOPE
This section has been written to serve as a guide for the experienced engineer in the design, installation, and operation of a temperature measurementsystem for measuring the temperatures related to performance evaluation of a liquid monopropellant rocket engine. Design guidelines rather than detailed specifications are provided for the critical componentsof each portion of the system. These guidelines, used in conjunction with current state-of-the-art, commercially available equipment and good engineering practices, will provide an optimum temperature measurement system which meets the performance criteria specified.
2.1
OBJECTIVE
Temperature measurements are made at a number of locations in a monopropellant rocket engine test system, including propellant storage tank, propellant supply lines, propellant flowmeter, thruster catalyst bed, thruster chamber wall and throat, valves, brackets, and flanges. These measurements have different accuracy requirements. One objective of these recommended practices is to provide the propellant temperature data necessary to obtain the density factor which, together with compressibility effects, is required to determine propellant mass flow from simultaneously obtained volumetric flowmeter measurements. In order to do this within the required uncertainty limits for flow, propellant temperature must be determined to within ±0.56°C (±!°F) over the range from 0 ° to 80°C (32 ° to 176°F). Nozzle throat and thrust chamber temperatures, for performance calculations, should be measured to within the range from 705 ° to 871°C (1300 ° to 1600°F).
will
depend
The importance on the user's
attached interest
4_I
when used ±1.2% over
to other temperature measurements and the purpose of the tests.
2.2
LIMITATIoNs
These practices are limited to temperature measurements made with thermocouples and resistance thermometers. I Although not restricted to the Measurement of steady state values, the response of thermocouples and resistance thermometers usually limits temperature measurement to an average of that produced by a pulse train when the engine is operated in the pulsed mode. These direct measurement gas composition.
practices do not of temperatures
3.0
DESIGN
include related
recommendations concerning the to the determination of exhaust
CONSIDERATIONS
A temperature measuring system capable of obtaining accurate data requires that careful consideration be given to selecting and assembling the transducer 2 with its supporting electrical and electronic equipment. The type of transducer and the manner in which it is installed along with the effects of environmental conditions are important from the mechanical viewpoint." Electrical and electronic components include signal conditioning, electrical calibration, recording, visual display, and data processing equipment. They are shown as a block diagram in Figure 4-I.
3.1
TEMPERATURE
TRANSDUCERS
Thermocouples are used for the great majority of temperature measurements on monopropellant rocket engine test systems. This is because they are easy to use, economical, and provide sufficient accuracy for most applications. The resistance thermometer may provide greater accuracy in those cases where measurement Uncertainty is dominated by thermocouple inaccuracy wherever greater accuracy is required. applications may employ such types of transducers pyrometers and temperature sensitive photographic
3.1 .I
Transduction
A
few special measurement as thermistors, scanning film.
Element
3.1.i.1 _j_. Thermoelectric sensing elements are used in thermoCouple circuits. A basic thermocouple circuit, as shown in Figure 4-2, consists of a pair of conductors of dissimilar metal welded or fused £ogether at one end to form a measuring Junction and connected via transmission lines to a reference Junc{ion which is maintained at a well-defined temperature. The circuit is completed with signal conditioning or recording equipment.
IThe resistance bulb (RIB) and
thermometer is also referred as a resistance temperature
2Transducer and sensor can convert physical phenomena
to as a resistance transducer (RTD).
be synonymous words describing devices into measurable electrical signals.
4-2
temperature
which
ROCKET ENGINE
..... ]
[
I
(MECHANICAL
COMPONENTS'
RESISTANCE TH ERMOMETER
THERMOCOUPLE (ELECTRONIC
COMPONENTS)
PRECISION RESISTANCE OR VOLTAGE SUBSTITUTION
REFERENCE JUNCTION
PRECISION VOLTAGE SUBSTITUTION
BRIDGE COMPLETION
CONSTANT - OR
-
NETWORK
CURRENT POWER SUPPLY
If SIGNAL CONDITIONING COMMON EQUIPMENT
1 VISUAL DISPLAY EQUIPMENT
DIGITAL SYSTEM RECORDER
! DATA PROCESSING EQUIPMENT
constructed
Figure
4-1.
By
using
various
to
cover
the
copper/constantan-type S.
temperature
limit
for a in
See
8
AWG
Table
thermocouples;
wire
IV-B-I
to
Table
for
a
chromel/constantan, No.
types
of
+1538°C in
are
primarily
limits
are
thermocouples -253°C with
Appendix
example,
with
decreases for
intended
a
No.
for lower
Junctions.
4-3
from 28
AWG
can
exposed
with
platinum/platinum-t0% The wire 1260°C wire.
enclosed-element for
be
(-424°F)
IV-B.
decreases
(1600°F)
somewhat
System
from
(+2800°F)
IV-B-I
871°C
metals range
thermocouple for
to
Measuring
measurement T
rhodium-type
Temperature
recommended size.
upper The
limit
(2300°F) Limits
for shown
(protected) (bare
wire)
sensing
When used within their optimum ranges the uncertainty values will generally be I% of reading or less with standard materials, construction, etc., down to less than 0.5% of reading with special materials, maximum purities, etc. The Type S thermocouple is the standard by which the international temperature scale from 660 ° to I063°C (1220 ° to 1945°F) is defined by the International Practical Temperature Scale. The
temperature
sensitive
element
in
a thermocouple
may
be made
arbitrarily small thus facilitating precise measurement of temperature a point. The low thermal capacity of the element, resulting from this size, insures quick response. (See Paragraph 3.1.3.1, item 6.)
at small
3.1.1.2 Resistance Thermometers. A resistance thermometer usually consists of a coil or grid of temperature sensitive wire whose resistance, when exposed to the temperature to be measured, varies in a precise and predictable manner. The auxiliary irlstruments which are used to measure these changes in resistance generally consist of a Wheatstone bridge with constant voltage excitation or a constant current with voltage readout proportional to resistance. The resistance thermometer is regarded as the most precise and reliabie device for the range -190 ° to ÷660°C (-310 ° t_ +1220°F). In the vicinity of -18°C (O°F), or at room temperature, measurements with an uncertainty of less than 2.8 x I0-4°C (5 x I0"4°F) can be made; however, the error is normally several thousandths of a degree. At 316°C (600°F)the uncertainty is about I/I0°C or I/5°F.
i
0
LO CONDITIONING AND RECORDING EQUIPMENI
REFERENCE JUNCTION
0
jHERMOCOUPI_E WIRE
Figure
E×TENSIO_ WIRE
4-2.
O
-- COPPER
Basic
TRANSMISSION
Thermocouple
/
4.4
WIRE---e,
Circuit
A winding of pure, strain-free, annealed platinum wire is widely used in transducers for precise temperature measurement. An element of this type is used to define the International Practical Temperature Scale between the limits of -183 ° and +630°C (-298 ° and +1165OF). An element wound from pure nickel wire or from nickel alloy wire such as "Balco" is used +300°C (-148 ° to +572°F); it
3.1.1.3
Thermistors.
A
in the medium temperature range -100 o to is less costly than a platinum wire element.
thermistor
is a semiconductor
whose
resistance
varies strongly with temperature. The semiconductor materials, usually sintered mixtures of sulfides, selenides, or oxides of metals such as nickel, manganese, cobalt, copper, iron, or uranium, are formed into small glass-enclosed beads, disks, or rods. They have high resistivities and high negative temperature coefficients of resistance. Their resistancevs-temperature characteristics are not linear, and it is difficult to maintain narrow resistance tolerances during manufacture. A thermistor must be individually calibrated and is inclined to drift; however, its wide variation in resistance and extremely small size make it desirable for certain applications. The thermistor is generally powered by a few microamps from a constant current source. The resistance is then inferred from voltage measurements. A thermistor is normally used over the temperature range of -75 ° to +250°C (-103 ° to +482°F).
3.1.2
Performance
Characteristics
The transducer performance characteristics (or properties) which are of greatest interest to the user are listed below. A brief commentary and suggested specifications are included where applicable. Unless otherwise specified they apply at the following ambient conditions: temperature 25 ±I0°C (77 ±18°F); relative humidity. 90% maximum; barometric pressure 73 _+7 cm Hg (29 ±2.8 in. Hg).
3.1.2.1 range over frequently capability.
Ra_. Range is usually specified as a nominal temperature which all other performance characteristics apply. It is selected as a range narrower than that given by the transducer's
3.1.2.2 Overload Temoerature. Overload temperature is the maximum or minimum temperature which is beyond the specified range limit, but to which the transducer can be exposed without incurring damage or subsequent performance changes beyond stated tolerances.
3.1.2.3 Ou__u_ID/_. Output can be stated as nominal full-scale output over the transdueer's span between its range limits, e.g., "Full-Scale Output: 36 mV (with reference junction at 0°C) '' or "Full-Scale Output: 250 ohms." Since most temperature transducers have a nonlinear outputvs-temperature relationship, the variation of output over the range
4-5
is best described by a table listing output values at several temperatures including the range limits. The curve established by such a table is often referred to as a theoretical or reference curve. This curve is often based partly or wholly on empirical data. 3.1.2.4 _. Repeatability and bias are usually the only steady state static accuracy characteristics specified for temperature transducers. Linearity is rarely specified since the reference curve of a large majority of temperature transducers is inherently nonlinear. Hysteresis and friction error are virtually nonexistent in temperature transducers. Repeatability, the maximum difference in repeated output readings from one another, can be expressed in percent of full-scale output but is usually stated in output units or output-equivalent .temperature units, e.g., "Repeatability: within 0.01°C ''or "Repeatability: within 1.0 ohm" or "Repeatability: within 0.5 mV." Repeatability has also been expressed in percent of reading (percent of reading-equivalent temperature correlated on the basis of a reference curve), e.g., "Repeatability: within 0.5% of reading." The tolerance specification can be followed by a statement of the period of time over which it is considered applicable. 3.1.2.5 Calibration Interchan_eabilitv. The maximum deviation over the specified range of the calibration curve of any one transducer from a reference curve is established for a group of transducers. A reference curve is often established for a certain transducer part number. Calibration interchangeability then applies to all transducers identified by this part number. It is normally expressed in the same terms as repeatability but with bipolar tolerances, e.g., "Calibration Interchangeability: Within +--2.0 ohms." Calibration interchangeability tolerances can be specified as larger in some portions of the range than in others.
3.1.2.6 Conduction Error. Conduction error iscaused by heat conduction between the sensing element and the mounting of the transducer. It occurs primarily in immersion probes, especially when the stem is short. Conduction error can be specified as the maximum difference in output reading when the transducer is first immersed, to just below the electrical connections, in an agitated temperature bath and when it is immersed up to the mounting fitting while the fitting and leads are artificially cooled or heated so that it attains the specified environmental temperature. The difference between sensing.element and environmental temperatures should be chosen as large as is estimated will occur in the end-use application of the transducer.
3.1.2.7
MQunting
Error.
Mounting
error
tolerances
(also
referred
to as strain effects or strain error) should be specified for surface temperature transducers. This error is introduced mostly by strain in the transducer after it is welded, cemented, or otherwise attached to a surface. It is specified as the maximum difference in output reading before and after mounting the transducer to a sample surface by a specified mounting method.
4-6
3.1.2.8 Time Constant. The time constant is an important dynamic performance parameter of an ideal transducer having a first order characteristic. It is specifically defined as the length of time required for the output of the transducer to rise to 63.2% of its final value as a result of a step change in temperature.
3.1.2.9 Resoonse Time. The time required for the transducer's output to rise to a specified percentage of its final value (say 98 or 99%) under the same test conditions as applicable to time constant determination. The limits of the step change in temperature and type and flow rate of the measured fluid at both limits must also be given, e.g., "from at I m/s."
still
air
3.1.3
Mechanical
at
25 +_5°C to distilled
water
at
80 ±2°C
moving
Design
The mechanical design of a temperature measuring system includes consideration of the characteristics of the type of transduction method selected and installation of the transducer in the system so as to obtain accurate measurement of the test item or medium with minimum effect on its temperature. These factors are discussed in the following paragraphs.
3.1.3.1
Thermoco_ple$
(i)
The performance of a fine gage thermocouple deteriorates more rapidly at elevated temperature than does that of a heavy gage thermocouple. The wire not only physically deteriorates, but the calibration of the wire irreversibly changes.
(2)
Thermocouple materials are normally furnished to a standard accuracy or calibration. There is usually an extra charge for both special accuracy and calibrated wire. The limits of error for standard and special accuracy are given in Table IV-B-I. No standardized limits have been established for calibrated wire.
(3)
Since spurious emfs are produced by temperature gradients in heterogeneous wire, all purchased lots of thermocouple wire should be checked for homogeneity.
(4)
The thermocouple junction may be formed by any of the following techniques which are presented in decreasing order of preference: electric, gas, or discharge welding, brazing, silver or soft soldering, immersion of the leads in a liquid metal pool, or mechanically forcing the wires into contact. For detailed information on thermocouple construction (both probe and surface installation) see References 4-I and 4-2.
4-7
(5)
In
any
near
junction
the
there
Junction
is
ent from the original as a short length of In order ture must (6)
The may
the
time
have
a time
may
(7)
have
(8)
be
If
response
down
in
splices used
The
(1)
to
large
oven
wire,
of
in
or
to
consistent
500
ms.
If
a'thermocouple
less..
known
thermal
the
gradients
thermocouple
matched
errors
thermal
and
due
block
continuously
thermometers units
The
to
material
materials to
must
gradients
temperature
should
monitored.
wire
should
show
metals
an
cobalt
an
which
be
specific
wound
special surface
available.
in
nickel
size
of
strain-
of
way. for
also
element
sensitive
wire
bonded
flexible,
electrically
4-8
resistance 0.003925.
resistivity, and
R,
then
iron,
Although
platinum
a
good
with
increases
particular
nickel is
resistance thermometer
(1200°F).
thermometer
measurements consists between
to
high-accuracy
makes
649°C
resistance
temperature This
In
standard
equal
linear
this
below
form
of
first
used
the
(a)
fashion.
universally
temperatures
to
alpha
increase
is
behave
thermometers,
for
should
conform
with
accelerated
almost
fully
t
Most
A
and
as
application.
element
curve
at
purchased
wire.
thermometer
and
be
performance
specific
transduction
Platinum
in
may
with
the
temperature
(5)
200
ms
be
thermocouples
bare
carefully
avoid
Resis£ance
free
(4)
should
sheath
Thermometers
tailored
(3)
the temperaof uncertainty.
from
50
differ-
4-3).
developed
(2)
to
connections
accurately
R_tance
of gage
vicinity
avoided,
reference
be
3.1.3.2
the
or
be
(Reference
(9)
constant
uncertainty
beregarded a third metal.
thermocouple Closed
small
may of
measurements this region
the
of
materials
avoided.
cannot be
of
from
Connections must
region of
wire] This wire composed
requirements.
constructed
a
consists
to make accurate be uniform over
response
with
always
which
two
insulating
element is
of thin
a
commercially grid pieces
material.
of
temperature of
3.1.3.3 should
.@
Transducer be
temperature general, the
listed
considered
Mouq_ing during
transducers. probes, items (I)
and
Configuration.
the They
surface
pertain
are
design
divided
into
measurements.
to
either
following and
three
Unless
items
installation
otherwise
thermocouples
of
categories
or
-
specified
resistance
thermometers.
General
(a)
Measured
fluids
sensing surface
(b)
material
used
exposure
(exposed,
immersion
identification
the
transducer
the
sensing
mounting
should markings
should
nomenclature,
or
information
number,
manufacturer's
of
external one
essential, at
a
as
to
a
nameplate
of
information temperature,
exposed-element. part
name,
and
identifica-
connections,
and
characteristic
nominal
should
be
be
furnished
throwaway
substrate
a
stated.
range
deemed or
resistance
temperature.
Consideration parts
as
for
includes
electrical
operating
such
stated
for
such
or
nameplate
least
and
transducer,
platinum-wire,
serial tion
element
be
include
e.g.,
on
coated) method
also
Other
at
limitations
stem materials, and should be defined.
enclosed,
or
transducer
resistance,
(e)
in
probe
The
as
, _ ,.-._
0) c,
,,-I :_ 0.,-4
cQ ._
c_ 0 .,.4
.._ O,
O
L._
_._ g., o.
Q;
o cO cQ Q;
L._
,q _u
._
._
c_ -,.-4
:_
.,._
_
._..._ _)
O
Q; e.
E
r,.,
O Q) c, ._m
.g: O.
.g:
L, _0 g_
O
O O
L
r_ _
0 J,.)_..) c_
0,, 0 Q) _
O
0._
c_ 0
CO _)
cO
_0 c_ oE-I
.r-I L
0 0
oJ
¢q
OJ
J -,-t
OJ
O4
c.
°_
.,=4
.,..(
-,-I
_0 0 ,.g:
q_ .0
.O
.O ,--I
0 _0 L
L
::I c. ..g: E-0 c_ .o
O_ 2
O_
C_J E
Z
Z
E
oJ
04
O_
E
._-( Z
Z
=,.
Z
Z
I-.I CO 0 0)
0 Q
"T
_O 0)
cO .,.-4
L
L,
O, n=
O'
.g: O
O L,
,=
_0 g) L, .O
c_ Q; Q) ,.-.4
X
"1
_0 _o Q) c_
.O
_.,
L, Q;
O p.
c_
0 Z
a_
O .O
E
CO JO r-,
O
O
cO
6-28
O O,
r_
Q) _
Q) O C (0 _J _o
cO
¢J
q) _-_ bO ._J c.
_
_ O O. E _
O
_ •
_-_ _
_ _
E
O
.,'_
.._ O r,.
O O
;3 .,-.I CO O9 L, L.
.,-I C .,.-4
_0 (0
:3
c0
.._
_) ._-_
c_ O
E_
"_ '_
0) c.
0 .,'_ ._ _
O •,-_
L _ c0
.,.._ _,.
_. r_ r'_ O
'O
L.
_) O
.._ ._.J
_
_0 I:_ _0 r_ _-,_
O
0) O
O
_.,
(1) (1) L O
0J L
c-_ c0 ," _ L
0
_1 .0
(1)
0
0
_
_
0
c'
_,,_
_._
_)
_-
,_
,...-I ._ i., r_
O _-
_) c'.
O .,-.I aJ cO
O cO
0 .,-4 aJ
c0 O •,-I
,--I _)
O
.C c0 c.
_
03
E
•
_0
q) O
o_-_
"10
E L.
N N 0 c.
L,
b0 c_ .,...i c. O C) v
.,-t _) c--
OJ
b0c0 r, ._.j F,.:2 .,.-_ O O c--,--I []
.O ,--I
cO .,-.I
cq
r_ .,-I
od .,..4
_O
.O
b0 c.
Q) co c_ [-., C0 c.
Od E
E Z
Q. O
Od E
oj E
-,.4 Z
Z
H CO
O c.. O
!
C O
L
,.O
0 f_
o9
S.,
0 cO cO
c_ -,.4
cO
I0 r_
_-, 0 _,._
O
.O
L 0
,O 6.4
O -,-3
.0 0 L.,
m .C
,.0 E
,-_
c0 _-)
c,
E_
0
J-) rO
O G..)
CD
O ..Q
c0
0 0 .C cO
0 ;0
W
0
O. _) L
0
0
0
cO
0._
L..) H
6-29
..c: ,-'4 _ CU t_ c_ c_ _ ,_..., r-,
-el c_ _._
,.4
_
_.-
_
,--_
0
•,_
_'_
_ r_ 0
e-. 0
E
•
C 0
C -,-4 C 0 C_
IlJ e_0_ r.1
r-
0 0 C
.rq
0
L_3
_O
cO ,-.4
0 c_ 0 0
.,_ E 0 ! E --_
L ,...-4
C 0
_ "0
[] co
m 0
0
0
_0
•,_
-_
_ E
-_
_
0 -_
_
_-_
6-30
._
.,.4
,-I
b0_,J
O c-
O
,-_
_J
O
c_
8 O
_ L .,_
,d O .,_
c-, _
°M
o_
_
cO ..t=00 O
b0 c. L _J O (.J
tb0
_0 O O r_ °_ []
r,1 .,.4
b0
c0 L,
CO c,...4 H CO O o .,"4
O
7 _O
O ,-I
" o
b0 -_
_
c_
•_ _._
.,._ _._
o
_0
_
O
O, .,-I O
,O O ,-_
.,._
_
1
A 0
O V O
L aJ
L
CO
6-31
0
,._ o .C
i.z
L
C_
r_ E _; .o
c_
v-0
E ,--I 0 _ _tm _-_ E _0 0
_0 bO
r,
_D .,.4
0
0
c_
0
0
0 ,-4 t_
E-,
t_ F_
v
_9
t_ O r.Q 0
0 IO
(0 cO O. i0
O cO
0 0 O. cO 0
r-, 0 .F-I
,-4
0 -,.4 0 _O 0
e_ .,-.I L O
0
,--4 r-I t_
.o bO
I _JO
:_.0 _ E 0 0 0
bO C 0
C
0
0
•,.4
.o
,_ B
O; B "0
0
I1) r_ B
,-4
•
t_
•
.o
,--I O 0
0 ;..
0 r_ E _J
0
0 >
:>
I:10 0 r_
i0 0
0
0
_
_'_
_
_ 0
0 0 ._-_
E
]
E I.
_,-., _t
0 .o
)
0 r_
_ 0 "_1 .._
._1 rio
oO
tO
0 LD
t..,
OO
e!
_D
t-.
_
OO 0
.,..4 c_ .,-I
0 ._
,--t
0 :_ ._.
C O .,...I
0
>
_, 0
_
0
;:h ,_
0 t) E-,
_o O9
6-32
0
0 CO
cO
I
0 qO c" 0 e" .o ,..-t •_.1 •_.-_
0
_
0
_
_
r...
_
_ 0 ,_
!=1. 0 (I1
._
0
_1 ,-¢ tO -r-4 L
0
CO c.
.4--_ r_
°
Q) O-O E
;_
> (2
O O. 1,_
e-
r_ (2
tlO
o
_) E r-._
O
II
I_
O 0...
c_ 0 °_ b0 c_ -fH L c. 0
c_
[J2
-_ r"
O O r_ S
o
_
°
._
_
_-_
e_o
O)
0
_
r"
E
c-, _ .,-I
_ 0 c-
0 c, ._
a_ O
cO .r-q
r_
L Q) ¢0 r-. E-,
¢0 -r-I
r_ H O0 O_ O r_ O
0
O r" O
4_ E .0..)
r_ L
r_ E
0 E E
0 -,-I
¢) ,-4 .O
O im
c-
c, 0
-,-I
0
0 O) O.
0
e., O 0 _., ..C
t
0 0
>
_., _,_
O r_ E X o0
6-33
_._
_1
_
>
0
_
._
0
o_...I I. a.a C L
C ,,4 c, ,,-,I ,-I
m
O 0
O ,,,-4 O -.4 c.,_
en I
I
I H >
O .,.-I
I. H
+
,Z
gJ .,-I
.,4
c,
c-
C
o_
C_ c_
.r-l.
_" ,_._
O
.,-I 'IO
O
O _
II O
C O &4
C
,O CD
c.c,l I_ -,-,I "El
_0 O
r-
t.
In -,-I ,-4
E r.,u E-4
"7 i= _D ,-4 I--I 0'3 0 I. c_ 0 C O
,,--I 0 .,-I .o L.
. ° v-. I ',,E,
O .,-I
,-I
.-I
_ 0 .,-I
0
ctl C_
_.,-I
c_ O .,.._
O i.
c. 0 .,-I
I. 0 (D
0
C O .,_
0 O
O .,-I
,-.4
,-I
I::
O bO
(tl
C .,4
O I-,
_a
0 0 I. >
-,-4
0 _0 C
.,-I
O ,,,.-t
0
6., I0,-4
O
,-.-I 0 ,,o
c/) -,..I O m.
[-i m,-g-
¢D
6-3_
¢D
,I
oL_.
3
,i._
e'_'
_k
7.0
REFERENCES
6--I.
Gardenshire, April 1964.
6-2.
Conte, S. D., and de Boor, C., Elementary McGraw-Hill Book Company, New York, 1972.
6--3.
Ha_ing, R. W., Numerical McGraw-Hill Book Company,
6--4.
Mechtly, E. A., The International System of Units, _d Conversion Factors, NASA SP-7012, 1969.
6--5.
Green, L., pP. 34-35,
6--6.
Page, _,
6-7.
Roscheke, Nozzle,"
6--8.
Garrett, J. W., Simmons, M., and Gobbell, W. C., Exit Noszle Flow Separation as Influenced by Nozzle Geometry. Fu_l-Oxid_zer Ratio. and Pressure Level, AEDC-TR-67-122, Arnold Engineering Development Center, Tennessee, July 1967.
6--9.
Rothe, D. E., Experimental _tudy of Viscous Low-Density Nozzle Flows, CAL Report AI-2590-A-2, Cornell Aeronautical Laboratory, June 1970.
L.
W.,
"Selecting
Sample
Rates,"
ISA
Journal,
Numerical
pp.
59-64,
Analysis,
J
W
R.
"Flow 1953.
Separation
H., "Flow Vol. 29,
Methods 1973.
for
Scientists
in Rocket
Nozzles,"
and
E. J., and ARS Journal,
Massier, P. F., pp. 1612-1613,
Physical
ARS
Separation in Nozzles," Journal pp. 110-111, January 1962.
Engineers,
Constants
4ournal,
of
"Flow Separation October 1962.
the
Vol.
23,
Aerospace
in a Contour
6-10.
Rae, W. J., Final Report on a Study of Low-Density Nozzle Flows With ADDliC_tion to Microthrust Rockets, CAL Report AI-2590-A-1, Cornell Aeronautical Laboratory, December 1969.
6-11.
Back, L. H., Massier, P. F., and Cuffel, R. F., "Flow and Heat Transfer'Measurements in Subsonic Air Flow Through a Contraction Section," Int. Journal of Heat and Mass'Transfer, Vol. 12, pp. 1-13, 1969.
6-12.
Back, L. H., Cuffel, R. F., and Massier, P. F., "Influence of Contraction Section Shape and Inlet Flow Direction on 'Supersonic Nozzle Flow Performance," Journal of Spacecraf$ _nd Rockets, Vol. No. 6, pp. 420-427, June 1972.
6-13.
JANNAF
Rocket
Dretation 6-14.
Engine
Manual,
Performance
CPIA
Publication
Test
Data
245,
Acquisition
April
and
9,
Inter-
1975.
Arblt, H. A., and Clapp, S. D., Fluorine-Hydrogen Performance, Phase I. Part I: Analysis. Design. and pem_ns_a_on of High Performance In lectors for the Licuid Fluorine - Gaseous _¥drogen Propellant Combination, NASA CR-54978, Rocketdyne Research Department, pp. 135-137, July 1966.
6-35
6-i5. Price,
T_ W., and Evans, D. D.,
Hydrazine Pasadena, 6-i6.
Technology, California,
Glossary, Monopropellant Monopropellant Working Agency, March 1976.
The Status of M6nopropeiiant JPL TR 32-1227, Jet Propulsion LabOratory, February 1968. Hydrazine Engine Technology, JANNAF Group, Chemical Propulsion Information
6-17.
Beltran, M. R._ CPIA Publication
and Kosvic, T. C., Rocket No. 131, January 1967.
6-18.
American Standard Letter Symbols Yi0.14-1959i American Society of
6-36
for Rocket Mechanical
Propulsion
Propulsion, Engineers,
Nomenciathre,
PUblication 1960.
APPENDIX THEORETICAL
PERFORMANCE
VI-A
OF
MONOPROPELLANT
HYDRAZINE
W
This calculations impulse and dissociation VI-A-2. as
a
The
appendix
for
Characteristic
ratio
of
of
of
expansion
product
gases
are
from
results
point
are
the
in
and
of
thrust
the
these
assumin_
assumed
VI-A-3.
Figure
molecular
VI-A-4.
Reference
ammonia
temperature
in and
Figure
by
in
was of
reaction
presented
generation
given
performance specific
functions of percent ammonia in Figures VI-A-I and
conditions
extended
gases
impulse
figures a
VI-A-I.
frozen
during
at
the
coefficient
were
constant
gamma
The
chemical
chamber
conditions
expansion
were
was
calculated
assuming
loss. stagnation
most
pressure here.
pressure
by
in
thermochemical Theoretical
adiabatic
shown
and
product
presented
the
at
from
hydrazine.
chamber
condensation
Specific
results
utilized
last
the
divergence
varying
used VI-A-I
the
of
A all
values
potential
ignored.
and
at
Reference
composition any
of
dissociation
heats
the
from
results
velocity
ammonia
specific
The taken
zero
presents
decomposition
thrust coefficient are shown as and nozzle expansion area ratio
function
weight
and
the
is
quite
stagnatlon
1030
effect
small
monopropellant
various
of
The
kN/m
on
over
the
hydrazlne pressures
2
the
(150
range
of
thrusters; are
psi)
calculated
was
operating however,
presented
used
values
in
for of
conditions tabular
Reference
VI-A-I.
REFERENCE VI-A-I.
Lee,
D.
H.,
Performance
Calculations
HYdrazine
and
Monopropellant
Mixtures,
JPL
TR
California,
32-348,
December
3,
Hydrazine Jet 1962.
t
6A-I
Propulsion
fo_
MonoDroDellant
- Hydrazine Laboratory,
Nitrate Pasadena,
i p
2600
270
I =
l(_30kN./m
2 (1501bf/in.
2)
c
_= 200 260 I00 250O
--
250
"--4
J.
24017
20
--
240
N _-
n_
10
-
--2_ t.)
-- 230
0
2200
--
"I" (-.
-- 220
2100
--
-- 210
2000
--
-- 200
I 20
I 40
I 60
I 80
I00
% rqH 3 DISSOCIATED
Figure VI-A-I.
Theoretical
Specific
6A-2
Impulse
for Monopropellant
Hydrazine
1.95
I
I
P
I
=
1030 kN/m
I
2 (150 Ibf/In.2)
20O P
1.9
oo\ 1.8.5
50
8 u_
z u 1.8--
8 -r .J
o
1.75
I
1.7--
1.65--
1
l
l
[
2O
40
60
80
100
% NH 3 DISSOCIATED
Figure
VI-A-2
Theoretical
Thrust
Coefficient
gr
6A-3
for MonoDropellant
Hydrazine
j
,
,
J
"
j
...........
jsooo
1400 Pc =
5 ¸
!039 kN/m
2 (150 !bf/in.
45OO
2)
D
44O0
1350
9 u
4300 1300 4200
View more...
Comments