electrical safety book

December 9, 2018 | Author: api-285243890 | Category: Biomedical Engineering, Mains Electricity, Electric Current, Electric Shock, Transformer
Share Embed Donate


Short Description

Download electrical safety book...

Description

I. In Intro troduc ductio tion n to Bio Biomed medica icall Engin Engineeri eering  ng 

Electrical Safety 

• Introduct Introduction ion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 1 • Industry Industry Regulatio Regulation n . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 3 • BMET Responsibilities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page .pag e 4 • History History of Electrica Electricall Safety Safety . . . . . . . . . . . . . . . . . . . . . . . . . . .page 4

Made Easy  II. Fund Fundament amental al Concept Conceptss of Electrical Electrical Safe Safety  ty   . . . .page 6 III.Physiological Effects of Electricity 

Table Of  Contents

• Introduct Introduction ion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 7 • Macroshock Macroshock and and Microsho Microshock ck . . . . . . . . . . . . . . . . . . . . . . . . .page 7 • High Frequ Frequency ency Effects Effects . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 8 • Potential Equipment Safety Safety Hazards . . . . . . . . . . . . . . . . . . .page 8 • Equipment Equipment Safety Safety Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 8 • Electrical Shock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 8 • The Electrical Power Power System System . . . . . . . . . . . . . . . . . . . . . . . . . .page 8 • Contact Contact with with a Ground Ground . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 9 • Skin Skin Resistance Resistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 9 • Leakage Leakage Current Current . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 10 • Electrical safety Power System Devices . . . . . . . . . . . . . . . .page 10 • Ground Fault Fault Circuit Interrupter . . . . . . . . . . . . . . . . . . . . .page 10 • Isolation Transformer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 10 • Equipotenti Equipotential al Grounding Grounding . . . . . . . . . . . . . . . . . . . . . . . . . . .page 11 • Codes Codes and Standar Standards ds . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 11-12

IV.. The AAMI Standa IV Standard rd  . . . . . . . . . . . . . . . . . . . . . . . . . .page 13-32 Bibliography  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 32 Glossary  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 33

 www.bapcoinfo.com  www .bapcoinfo.com

T

he field of medical instrumentation is by no means new. Many instruments were developed as early as the nineteenth century – for example, the electrocardiograph,, first used by  cardiograph Einthoven at the end of that century. Progress was rather slow, however, until after World War II  when a surplus of electronic equipment such as amplifiers and recorders became available. At that time, many technicians and engineers, both within industry  and on their own, started to experiment with and modify  existing equipment for medical use. This process occurred during  the 1950’s and the results were often disappointing. For the experimenters soon learned that physiological parameters are not measured in quite the same way as physical parameters. They also encountered a severe communication problem with the medical profession. During the next decade many  instrument manufacturers entered the field of medical instrumentation, but development costs were high, and the medical profession and hospital staffs were suspicious of new equipment and often uncooperative.. Many developuncooperative ments with excellent potential seemed to have become lost causes. It was during this period that some progressive companies decided that rather than modify  existing hardware, they would design medical instrumentation specifically designed for medical use. Although it is true that many  of the same components were used, the philosophy was changed; equipment analysis and design  were applied directly directly to medical problems.  A large measure of help was provided by the U.S. government, in particular NASA (National  Aeronautics and Space  Administration). The Mercury, Mercury, Gemini and Apollo programs needed accurate physiological monitoring for the astronauts;

consequently, much research and development money went into this area. The aerospace medicine programs were expanded considerably,, both within NASA  considerably facilities and through grants to Universities and Hospital research units. Some of the concepts ad features of patient monitoring  systems presently in use in hospitals throughout the world evolved from the base of astronaut monitoring. The use of adjunct fields, such as biotelemetry biotelemetry,, also found some basis in the NASA  programs.  Also, in the 1960’s, 1960’s, an awareness of  the need for engineers and technicians to work with the medical profession developed. All the major engineering technical societies recognized this need by  forming "Engineering Medicine and Biology" subgroups and new  societies were organized, such as the Biomedical Engineering  Society.. Along with the medical Society research programs at the universities, a need developed for courses and curricula in biomedical engineering and today  almost every major university has some type of biomedical engineering program. However, much of this effort is not concerned with biomedical instrumentation per se. One of the problems of  "biomedical engineering" is defining it. The prefix bio- of  course, denotes something  connected with life. Biophysics and biochemistry are relatively old disciplines in which basic sciences have been applied to living things. One school of thought subdivides bioengineering into different engineering areas – for example, biomechanics, and bioelectronics bioelectronics.. These categories usually indicate the use of that area of engineering  applied to living rather than to physical components. Bioinstrumentation implies measurement of biological variables, and this field of  measurement is often referred to as biometrics, although the latter

Electrical Safety  Made Easy 

I. Introduction to Biomedical Engineering 

-1-

term is also used for mathematical and statistical methods applied to biology.

Electrical Safety  Made Easy 

I.

Naturally, committees have been formed to define these terms; the professional societies have become involved. The latter includes the IEEE Engineering in medicine and biology group, the ASME Biomechanical and Human Factors division, the Instrument society of   America and the American Institute of aeronautics and Astronautics. Many new cross-disciplinary  societies have also been formed. Several years ago an engineering  committee was formed to define bioengineering. This was subcommittee B (Instrumentation) of  the Engineers Joint Council Committee on Engineering on engineering Interactions with biology and Medicine Medicine.. Their recommendation was that bioengineering be bioengineering  be defined as application of the knowledge gained by a cross fertilization of  engineering and the biological sciences so that both will be more fully utilized for the benefit of man. More recently, as new applications have emerged, the field has produced definitions describing  the personnel who work in it. A  tendency has risen to define the biomedical engineer as a person  working in research research or development in the interface area of  medicine and engineering, whereas the practitioner working with the physicians and patients is called a clinical engineer. engineer . One of the societies that has emerged in this interface area is the  Association for the Advancement of  medical Instrumentation (AAMI). This association consists of both engineers and physicians. In late 1974, they developed a definition that is widely accepted:

-2-

"A clin ical engin eer is a professional who brings to health care facilities a level of education, experience and accomplishment  which will enable him to

responsibly, responsibl y, effecti effectively vely,, and safely  manage and interface with medical  devices, instruments and systems  and the use thereof during patient  care,, and who can, becaus care becausee of this  this  level of competence, responsibly and  directly serve the patient and  physician, nurse and and other health care professionals relative to their  use of and other contact with medical instrumentation."  Most engineers go into the profession through the engineering  degree route, but many start out as physicists or physiologists. They  must have at least a B.S. degree and many of them have M.S. or Ph.D. degrees.  Another popular term, also coined in recent years, the biomedical equipment maintenance technician (BMET) is defined as follows: "A biom edical equi equi pm ent  m ain te tenan nan ce te technician chnician (BMET) is 

an individual who is knowledgeable  about the theory of operation, operation, the  underlying physiologic principles  and the practical, safe clinical  application of biomedical  equipment. equipmen t. His capabilities capabilities may  include installation, calibration, inspection, preventive maintenance  and repair of general biomedical  and related technical equipment as  well as operation or supervision of  equipmentt cont equipmen control, rol, safety and  maintenance programs and  systems."  This was also an AAMI definition. Typically, the BMET has two years of training at community college. This person is not to be confused  with a medical technologist. technologist . The latter is usually used in an operative sense, for example in blood chemistry and in the taking  of electrocardiograms. The level of  sophistication of the BMET is usually higher than that of the technologist in terms of  equipment, but possibly lower in terms of the life sciences. In addition, other titles have been used, such as hospital engineer

and medical engineer. engineer . In one hospital the title biophysicist is preferred for their biomedical engineers, for reasons best known to themselves. These definitions are all noteworthy,, but whatever the noteworthy name, this age of the marriage of  engineering to medicine and biology is destined to benefit all concerned. Improved communication among engineers, technicians and doctors, better and more accurate instrumentation to measure vital physiological parameters and the development of interdisciplinary tools to help fight the effects of body malfunctions and diseases are all a part of  this field. The name itself is actually  not all too important; however, however,  what the field can accomplish is important. With this point in mind,  we will be using the term biomedical engineering  for describing the field in general and biomedical instrumentation for the methods of measurement  within the field.

INDUSTRY REGULATION In it’s infancy, biomedical equipment was best serviced and maintained by the original equipment manufacturer as there  were few, few, if any, any, qualified technicians who could do so adequately. A lack of standards linking the technology, an absence of service literature, and the general feeling from hospital personnel that, unless trained by  these OEMs, no one would be able to provide a suitable level of  services. Thus the early BMET was relegated to maintain and repair only the simplest of clinical equipment such as centrifuges, suction pumps and other such equipment. Several factors soon arose to alter this conception. First, agencies such as the Food and Drug Administration (FDA) developed subdivisions under  which fell the area of medical instrumentation. This brought about a set of standards and practices for manufacturing of 

hospital equipment, as well as the maintenance and repair of such. Secondly,, other agencies, under Secondly  which hospital equipment was monitored and controlled, invoked requirements such as the release of  service literature on any new  equipment purchased or manufactured to the hospital. This literature, the law stated, must contain the procedures for calibration and alignment of said equipment. Thirdly, the training of  biomedical equipment technicians began to catch up with the advancing technology so that the level of competence was significantly improved. Finally, through the feeling that they provided the only service available, the OEMs began to charge significantly more for their service contracts each  year. Surprisingly Surprisingly,, this was not a major issue to the hospital community until the middle of  1980 when the government, in an effort to control the rising cost of  medical treatment, and the subsequent rise in medical insurance claims, imposed major restrictions upon hospitals. These restrictions, known as Diagnostic related Groups (DRGs) limited the amount that medical insurance groups  would reimburse both hospitals and physicians for each type of  illness for which a patient might enter the hospital. Without Without going  into great detail about DRGs, suffice it to say that their introduction brought with it a need for hospitals to lower their operating  costs. One area in which these costs could be lowered was in the area of biomedical instrumetation.  Where once a hospital would make a major equipment expenditure, such as buying a new chemistry  analyzer,, every five or so years, they  analyzer began keeping sophisticated equipment such as this for longer periods. So doing, the rising OEM service contract prices suddenly  became an issue and the responsibility for these equipment fell upon the BMET as a lower cost alternative.  As more and more responsibility  responsibility  fell upon the BMET within the

Electrical Safety  Made Easy 

I.

-3-

Electrical Safety  Made Easy 

I.

hospital for larger and more sophisticated instrumentation, new  problems arose for the hospitals. In order for the equipment to be adequately maintained, more BMETs needed to be employed, each having sufficient training in the sophisticated equipment involved. Hospitals were faced with  weighing the additional cost of  salaries, benefits and the costs of  attending additional manufacturers schools and seminars with the cost of the OEM contracts. In very large hospital facilities, this was not such a significant problem as it was with smaller facilities. In many smaller facilities, typically those under 200 patient beds, it was simply not feasible to consider an in-house BMET program and, thus, these facilities were forced to pay for the OEM contracts.

be employed on sensitive, lifesaving medical equipment.

 Again, the government became involved. Federal and state agencies began to require all hospital facilities to comply with stringent safety guidelines for patient related equipment. This included a routine electrical safety  program under which all hospital equipment would be tested, evaluated and forced to comply   with. Typically ypically,, the BMET provides one or more of the following  services to hospital facilities:

In an effort to satisfy applicable federal and state requirements as  well as to insure that clinical equipment are utilized both safely  and to their optimum, some BMET shops provide periodic in-service and education services. This often ranges from simple electrical safety  lectures to the proper use of a newly acquired piece of equipment for which hospital personnel might not yet be familiar.

BMET RESPONSIBILITIES

This brings us to the main purpose of this document. Periodic testing  of patient and non-patient hospital equipment to insure that they meet the safety requirements and guidelines as set fourth by applicable federal and state regulations. This This in no way measures the equipment ability to perform the task for  which is was designed.

Corrective Maintenance

-4-

Non-scheduled repair of hospital equipment other than during  Preventive Maintenance inspections. These repairs range from replacement of minor parts or components to total equipment overhauls as may be required. In an effort to further provide a cost savings to the t he hospital facility, some BMETs attempt to repair equipment to the smallest component level, rather than simply replacing circuit boards or sub-assemblies which may be very  costly. Other BMETs follow a philosophy that component-level repair entails far too many risks of  repeated failure or improper component soldering techniques to

Preventative Maintenance Inspections Periodic testing and evaluation of  patient and non-patient related equipment to insure that they  operate within the guidelines set fourth by federal, state and manufacturer guidelines. This includes calibration and alignment of said equipment along with the repair and/or replacement of  component parts in order to bring  the equipment into compliance. It is the belief of most BMETs that good, quality preventive maintenance inspections further reduce hospital costs by reducing equipment down-time for repairs.

Inservice & Education

Electrical Safety Inspectio Inspections ns

HISTORY OF ELECTRICAL SAFETY  The introduction of electricity into commerce at the close of the 19th century carried with it the need to know how to deal with it safely. For the hospital environment, much of  the electrical safety programs started after World war II. Shortly  thereafter,, federal funds via thereafter v ia the

Hill-Burton Act increased the number of hospital facilities. Grounding for electrical safety was implemented because most electrical accidents in the home and industry occurred because exposed metal was energized. The National Electrical Code stated that "non-current carrying" metal parts of the electrical apparatus could prevent such accidents. Unfortunately, by the end of the 1950’s, electrical power in the hospitals was supplied in a haphazard, eclectic fashion. In the hospital environment of the sixties, electricity came to be used more often on, in and around patients to a degree beyond conception only a few years previously. It was early in 1961 that there appeared the first news that "microshock" (small electric currents applied to a conductor near the heart) was happening in the medical field. In 1969, Carl Walter, M.D., who was at the time a well-known surgeon, stated that, "1,200 patients were being accidentally electrocuted in U.S. hospitals each year." Although many engineers and health care professionals believed that Dr Dr..  Walter’  Wa lter’ss estimates were unrealistically high, the concept of  microshock suddenly became publicized. Then, on June 16th, 1970, Ralph Nader (a household word because of his 1965 book, Unsafe at Any  Speed ) gave a speech in Detroit. In this speech, he stated that, "1200 annual electrocutions in US hospitals was a very least figure" and quoted other experts indicating that the real number might be significantly higher. higher. The event was picked up by a wire service and run under an arresting  headline the next day: Hospital  electrocutions cited Detroit (UPI) Detroit  (UPI) – "Accidental electrocutions claim 5,000 lives in American Hospitals every year but seldom get reported due to the "close nature of 

hospitals" consumer critic Ralph Nader said yesterday. During the 1970’s, several proposals and regulations were introduced to manage this suspected problem in hospitals. In 1971, the National Fire Protection  Association published a recommended standard (76BM) to help hospital engineers understand the principles of electrical safety and coordinate a program of medical equipment electrical testing in their facilities. In the spring of  1972, Underwriters laboratories issued document UL544 Medical  UL544 Medical  and Dental Equipment , which was intended to serve as a guideline for medical equipment manufacturers manufacturers.. In 1970, AAMI (the Association for the Advancement of medical Instrumentation) published a firstdraft standard for electrical leakage current standards that was adopted as an American national standard in 1978.

Electrical Safety  Made Easy 

I.

Probably the most dramatic proposal was the Joint Commission on the Accre Accreditation ditation of Hospitals (now the Joint Commission on the Accre Accreditation ditation of Healthcare Organizations – JCAHO) 1976 recommendation that hospitals maintain equipment control programs to provide for electrical safety training, create a documented preventive maintenance program, and perform semi-annual safety and performance equipment inspections and annual inspections of electrical receptacles. Today, health care institutions in the United States support clinical engineering programs that provide ongoing electrical safety and performance testing as well as preventive maintenance and repair of medical equipment. Typically, these programs use the most recent editions of NFPA 70, NFPA 99, NFPA 101, AAMI Recommended Standards, and Joint Commission accreditation manuals for their reference standards. -5-

S

Electrical Safety 

ome of the most basic concepts of electricity must be understood in order to grasp the potential hazards of  electrical devices as they are used in a clinical setting such as a hospital. Some of these important ideas concerning electrical safety in a medical environment are explained below.

Made Easy  The Nature of Electricity 

II. Fundamental Concepts Of  Electrical Safety 

Electrical current flowing through a conductor is the result of electrons moving from the outer shell of  atoms induced by an electric field that is imposed on a conductor. This field can be caused by any  voltage generating source, such as a local utility company, a battery, or a chemical reaction. In the hospital, this voltage source is provided by the local electrical company and is redirected through a series of transformers to 240, 208, or 120 volts alternating current (AC). The amount of current that flows through an electrical device is determined by the resistance that the device is designed to provide to the applied field. This relationship is called "Ohm’s Law" and is  written as:

I=V  R BAPCO’S SA2115  “Safety Certifier”  is the most complete  safety analyzer  available that provides  everything you need to safety test: • Pa Pati tien entt to no nonnpatient, pati ent, medi medical cal to commerciall devices  commercia • 110 110 V @ 200a 200amp mp to 240V @ 15 amp • Auto Automa mati tica call llyy to AAMI standards 

-6-

 Where "I" is the current in amperes that flows through a device or conductor,, "V" is the conductor t he magnitude in volts of the applied electric field, and "R" is the resistance that the device or conductor has, measured in Ohms. Thus, if an electromedical device such as an aspirator is plugged into a 120 volt receptacle and the unit requires 3.0 amperes, then the resistance that the aspirator provides to the voltage is:

120 volts = 40 Ohms 3 amperes a measure of the amount of power that an electrical device consumes during operation is defined as the product of voltage and current and is recorded in Watts. From the previous example:

Power (watts) = (V ) (I) = (120) (120) (3) = 360 Watts Watts

B

ecause the amount of electricity that will flow through any medium depends on the resistance that it encounters, current can flow  through the human body and cause various effects. If direct current (polarized, nonchanging) or high frequency alternating  current passes through the body body,, heating effects and ultimately  burns will occur. It is this effect that is intentionally created when electrosurgical generators are used to cut tissue and coagulate fluids. If  low frequency alternating current is applied to the body body,, muscular polarization and depolarization take place that can ultimately  create a "circus movement" in the heart muscle, resulting in fibrillation and death. It is this effect that normally accounts for death due to electric shock. Unfortunately, the typical resistance of the human body in combination with the frequency of  commercially generated electricity  (60 hertz) can create a potentially  hazardous situation in the hospital environment.

MACROSHOCK AND MICROSHOCK  The effect of electric shock on the human body can be anything from barely perceptible tinges, to muscle spasms, to death. Each can occur from small or large currents, depending on how the currents are introduced into the body body.. Large currents (milliamperes or larger) that are introduced into the body  from one external point to another (arm to leg, for example) can result in macroshock. If small currents (as low as 10 microamperes) are introduced into the body from an external source such as a catheter or cardiac pacing wires, the resistance to the heart muscle can be very low low,, and electrocution can occur from microshock. The chart below describes the effect of different levels of current that are introduced into the human body by creating a voltage across each arm. For microshock, the same effect can be produced with current levels that are only  1/10,000 as great as those listed.

CURRENT

EFFECT

.001 .00 1 Ampere Ampere (1 (1 Millia Milliampe mpere) re)

"Ting "T inglin ling", g", thre thresho shold ld of perc percept eption ion

.020 Amps (20Ma)

Muscle Spasms, hard to release grip

.050 amps (50 Ma)

Pain, possible fainting, transient interruption of respiration

.100 Amps (100 Ma)

Ventricular fibrillation

>5 Amps

Sustained myocardial contraction, possible burns, temporary respiratory paralysis

Electrical Safety  Made Easy 

III. Physiological Effects Of  Electricity 

-7-

Electrical Safety  Made Easy 

HIGH FREQUENCY  EFFECTS

EQUIPMENT SAFETY TESTING

 As mentioned previously, previously, the primary effect of high frequency  current (500 kilohertz to 2 megahertz) is to heat tissue as it is concentrated in a certain area. The amount of heat that is generated depends on the amount of current applied and the area that the current passes through.

Tests to determine the electrical safety of medical devices include a measurement of the continuity as  well as the leakage current between the chassis and the cord grounding  conductor.. Equipment that has conductor been designed with patient leads or contact points is also measured for current leakage from these points. Figures 1 and 2 show typical circuits that may be used to measure this leakage current.

This relationship is :

Heat He at = I x I  A 

III.

 Where "I" is the applied current and "A" is the area that the current flows through.

POTENTIAL EQUIPMENT SAFETY HAZARDS Because current flow through the body can be hazardous if it is of a certain amplitude and frequency, stray currents must be eliminated from medical equipment. The best method to prevent leakage currents is to ensure that all conductive parts of the equipment are connected to the hospital grounding  system through the power plug.  Also, to minimize patient patient contact  with current leakage on ECG signal leads, electrically isolated amplifiers should be used.

-8-

In theory, the use of line isolated power systems helps to minimize safety hazards by isolating the neutral power line from earth ground. A line isolation monitor is also installed with these systems to identify visually and aurally the presence of leakage current between the isolated power line and the grounded conductor. conductor. Other power system devices that help to identify or disconnect power sources when leakage currents are present include ground fault detectors (GFDs) and ground fault interruptors (GFIs).

ELECTRICAL SHOCK  The three ingredients in the scenario of electrocution are : 1. contac contactt with with the liv live e conductor of a grounded electrical system 2. con contac tactt with with a groun ground d 3. dimin diminished ished skin resi resistanc stance e

THE ELECTRICAL POWER SYSTEM Domestic voltages in the United States are from 120V for lighting  and small appliances, and 240V for electrical ranges and dryers. In the hospital, outlets are 120V while 277V and 208V are often used for fixed lighting and special receptacles. Typically, a voltage of 240V is provided by the service drop to the power meter of the hospital. This line consists of a bare cable plus two wires having black insulation. The bare cable is connected to "earth" ground via a water pipe or grounding rod. It is this connection that makes the electrical service a grounded system. NOTE : The national electric Code defines ground as "a conducting connection,  whether intentional intentional or accidental, between between an electrical circuit or equipment and the earth, or to some conducting body which serves in place of the earth."

There are three reasons to ground electrical systems: 1. Grou Grounding nding of the the system protects against introduction into the structure, via the  wiring, of high voltages (with respect to ground) such as might arise from lightning or insulation failure in a high voltage pole transformer transformer.. 2. Tying the circuit circuit to ground ground obviates the multiple problems that might ensue were the domestic circuits "floating" at some undefined but high voltage relative to ground. 3. To facilitate operation operation of over current protective devices (fuses) which are located in the "live" (ungrounded) side of each circuit.

CONTACT WITH  A GROUND For purpose of illustration, we will use an electrocardiograph (ECG) as our example of a medical device. Back in the 1950s, the standard ECG was fitted with a two-prong  plug. The patient was connected to the chassis ground via the right-leg  electrode, and one wire of the power cord was connected to the chassis through a 200,000-Ohm resistor.. This connected the chassis resistor to ground via the natural conductor in the power system. If the plug was inserted with polarity  reversed, then the patient was connected to the "hot" side of the power system. However, the 200KOhm built-in resistor limited the current to 0.6 milliamperes if the patient or bystander provided a pathway to ground. Added safety to the machine included a fivemilliamp fuse in series with the right-leg electrode. As it turns out this approach was not practical because the fuse kept blowing after the patient was defibrillated. The codes changed this practice and today’s ECG machines do not ground any of the electrodes to earth ground. In fact, the "front end" electronics on most medical

devices have an isolated power supply and grounding system. Plus, the manufacturers use plastic cases minimizing the exposed metal.

SKIN SKI N RESIST RESIS TANC ANCE E  Water affects affects skin resistance, and for a given voltage, resistance determines current. When dry, skin has a resistance of upwards 100,000 ohms. If there is an accidental application of 120 volts between the two hands, only 1.2 milliamperes will flow. In a wet environment or on a hot and humid day, that same current path may come to have a resistance as low as 1,000 ohms, resulting in a current flow of 120 milliamperes milliamperes..

Electrical Safety  Made Easy 

III.

If a patient is exposed to electrical current, it takes about 1 milliampere of 60 Hz AC for a threshold of sensation. The sensation becomes uncomfortable if 5 milliamperes is attained. Strong  muscle spasms appear at currents of 10 to 20 milliamperes ("let-go" current before sustained muscular contraction). A current flow of 100 milliamperes or greater may  induce ventricular fibrillation and death. These values are for currents introduced at the body surface. They are termed macroshocks and require two points of external body  contact. There is a possibility of a direct electrical path to the heart via a needle or catheter in an artery or vain. This directly reduces the resistance and current threshold. Small amounts of current (100 microamperes) can be potentially lethal. Electrical shock  in these circumstances is termed "microshock". The frequency of the current is also important when considering  electrical shock. If the frequency is raised above 1 KHz, these current levels no longer produce such sensations or life-threatening  phenomenon. High frequencies in the megahertz region will not cause shock at all.

-9-

LEAKAGE CURRENT

Electrical Safety  Made Easy 

III.

 All electrically operated devices have some current that flows from the energized electrical portions of  the device to the metal chassis. This current is referred to as leakage current and has two components, capacitive and resistive. Capacitive leakage current results from distributive capacitance between two wires or a wire and a metal chassis case. Components that cause capacitive leakage currents are RF filters, power transformers, power wires, and any  device that has stray capacitance. Resistive leakage current arises from the resistance of the insulation surrounding the power  wires and transformer primary   windings.

ELECTRICAL SAFETY  POWER SYSTEM DEVICES Several techniques are available to protect clinicians and patients from electrical shock. The most common ones are the ground fault circuit interrupter (GFCI), the isolation transformer and equipotential grounding.

GROUND FAULT CIRCUIT INTERRUPTER The GFCI acts like a circuit breaker  when it senses an inequality of as little as 6-ma between the "hot" and neutral wires of the circuit. It is mandated by the national electric code where electrical outlets are so situated that simultaneous contact  with a grounded surface is especially likely. "Wet locations" such as a whirlpool bath or bathroom are examples of GFCI usage.

-10-

In the GFCI, current in the "hot"  wire passes through one transformer coil on the same transformer core. From the design, the net flux is zero zero,, when the

currents are equal. If the current is unbalanced (i.e. current flowing  through ground) a net flux is induced across the third coil. This current will trigger a switch opening the "hot" line. The GFCI affords economical protection against electrical shock. However, the GFCI depends upon an active system, and the integrity  of it’s mechanical operation is crucial.

ISOLATION TRANSFORMER The Isolation transformer offers electrical safety by converting  grounded power into ungrounded power. This is accomplished by  grounding the primary winding of  the transformer and not grounding  g rounding  the secondary winding. Isolation is not perfect for two reasons. First, the isolation transformer has some stray capacitance to ground. Second, every medical device that is attached to the transformer possesses stray capacitance which causes some degree of coupling  between its power power-carrying -carrying wires and the grounded frame. To monitor the system isolation, a line isolation monitor (LIM) is employed. Its function is to analyze the entire isolated circuit and quantify its degree of isolation from ground. The LIM provides visual and audible alarm signals when the predicted ground-seeking current exceeds a specified magnitude. The LIM does not indicate an existing  current flow, rather it predicts the current that would flow if a short circuit were to develop between isolated wire and ground. In the operating room, isolated power systems were first installed as a measure directed against sources of ignition rather than electrical shock. Flammable anesthetics such as diethyl ether and cyclo-propane were used. Today, most anesthesiologists use non-flammable anesthetics and isolated power is not requir required. ed.

EQUIPOTENTIAL GROUNDING  Another technique that reduces electrical shock is equipotential grounding. This is accomplished by  adding another grounding wire from each chassis to a central point that is in parallel with the third wire in the power cord. If the chassis of  all equipment is at the same potential there will be no current leakage to the heart. This technique has its advantages and disadvantages and is not typically used in today’s health care environment.

CODES AND STANDARDS In the 1960s, all aspects of hospital activities involving fire and explosion hazards (including  electrical shock and emergency  electrical power) were were seen to be in need of some guidelines and/or standards. As it turns out, in the standards arena of the 1970s, patient safety was twice corrupted. First, standards were generated in ignorance (solutions were imposed before the problems were defined)

and second, safety was blatantly  exploited for ego-serving, bureaucratic,, and commercial bureaucratic gains. Today, there are two standards that specify electrical safety: NFPA 99 and the AAMI Standard for Safe Current Limits. The 1996 edition of NFPA 99 includes numerous technical changes that relate to leakage current and it now more closely  correlates to the international standards. Originally, NFPA 76B covered the electrical wiring system and electrical appliances. The summary of the standard as it relates to electrical safety is in the Table on the following page.

Electrical Safety  Made Easy 

III.

One further comment about standards. On may 30, 1972, Underwriter’s laboratories (UL) released the first edition of UL-544, Standard for Safety for Medical and  Dental Equipment . UL-544 deals  with details regarding enclosure enclosure safety, mechanical stability, and integrity of insulation. It’s leakage current, isolation requiremen requirements, ts, and test methods are similar to but not identical with those of AAMI and NFPA.

-11-

NFP FPA A - 99 1996 1996

Electrical Safety  Made Easy 

III.

Chassis Source Current, Cord Connected (Portable)

Gro roun und d Open

Gro roun und d Intact

 With Isolated Patient Connection Connection

300µa

300µa

300µa

100µa

 With Nonisolated patient Connection Connection

300µa

300µa

300µa

100µa

Likely to Contact Patient

300µa

300µa

300µa

100µa

No patient Contact

300µa

300µa

300µa

100µa

Chassis Source Current, Permanently Connected

Gro roun und d Open

Gro roun und d Intact

Gro roun und d Gro roun und d Open Intact In

 With Isolated Patient Connection Connection

5000µa

5000µa

5000µa

100µa

 With Nonisolated patient Connection Connection

5000µa

5000µa

5000µa

100µa

Likely to Contact Patient

5000µa

5000µa

5000µa

100µa

No patient Contact

5000µa

5000µa

5000µa

100µa

Lead to Ground Current

Gro roun und d Open

Gro roun und d Intact

Gro roun und d G Gro roun und d Open Intact In

 With Isolated Patient Connection Connection

50µa

10µa

50µa

10µa

 With Nonisolated patient Connection Connection

100µa

100µa

100µa

50µa

Gro roun und d Open

Gro roun und d Intact

 With Isolated Patient Connection Connection

50µa

10µa

n/a

n/a

 With Nonisolated patient Connection Connection

50µa

50µa

n/a

n/a

Sink Current (Isolated Test)

Gro roun und d Open

Gro roun und d Intact

 With Isolated Patient Connection Connection

n/a

20µa

n/a

n/a

50µa

50µa

n/a

n/a

Gro roun und d Open

Gro roun und d Intact

Gro roun und d G Gro roun und d Open Intact

Gro roun und d Intact

Gro roun und d G Gro roun und d Open Intact

Gro roun und d Open 4 Oz

Gro roun und d Intact

Gro roun und d Gro roun und d Open Intact

Gro roun und d Open

Gro roun und d Intact

Gro roun und d G Gro roun und d Open Intact

Lead to Lead Current

 With Nonisolated patient Connection Connection

Ground Impedance

0.5 Ohms

Existing System

0.2 Ohms

New Construction

0.1 Ohms

Receptacle Indicators

Gro roun und d Open

Hospital Grade Emergency System Isolated Ground

Retention Force, Ground Blade

-12-

AAM AA MI/ I/AA AAM MI ESI ESI 1993

GFCI Trip Trip Current

Gro roun und d Gro roun und d Open Intact

Gro roun und d G Gro roun und d Open Intact

Gro roun und d G Gro roun und d Open Intact

Listed Red Orange

6mA 

T

he AAMI Electrical Safety Committee began deliberating the issue of safe risk current limits for electromedical apparatus in 1967 and first published recommended limits and test methods in 1971. The first edition of this American National Standard was approved by  the American National Standards Institute (ANSI) in 1978, and a second edition was approved and published in 1985. Wo Work rk on this third edition involved the dedicated efforts of concerned health care professionals, industrial scientists, and government representatives.. AAMI expresses its representatives gratitude for the service of all persons involved in the development of this standard.

fault conditions and provides risk  current levels for each of these states. Also included included in this standard are test methods for equipment with nonconductive enclosures and for doubleinsulated equipment as well as limits for earth leakage or earth risk currents.

In this third edition of the standard, the risk current limits have been raised to be compatible with, although not identical to, the limits set forth by the International Electrotechnical Electrotec hnical Commission (IEC) in its standard, st andard, Medical electrical equipment—Part 1: General requirements for safety  (IEC 601-1-1988). As in the IEC standard, this edition of Safe current limits for electromedical apparatus (ANSI/AAMI ES1—1993) introduces the concepts of normal operating conditions and single

Suggestions for improving this standard are are invited. invited. Comments and suggested revisions should be sent to Technical Programs, AAMI, 3330 Washington Boulevard, Suite 400, Arlington, VA 22201-4598.

This standard is intended primarily  for the testing of electromedical apparatus intended for use in or near the patient care area. This standard should be considered flexible and and dynamic. As technology advances and new data are brought forward, the standard will be reviewed and, if necessary necessary,, revised.

NOTE — This foreword does not contain provisions of the American National Standard, Safe current limits for electromedical apparatus (ANSI/AAMI ES1—1993), but does provide important information about its development and intended use.

Electrical Safety  Made Easy 

IV. The AAMI Standard Standar d For For Safe Current Limits For Electromedical  Apparatus  Appara tus

-13-

Electrical Safety  Made Easy 

IV.

SAFE CURRENT LIMITS FOR ELECTROMED ELECTROMEDICAL ICAL  APPARATUS  APPARA TUS 1 SC SCOP OPE E 1.1 Inclusions This standard sets risk current limits and referee test methods for electromedical apparatus intended for use in the patient care vicinity  and also sets limits for nonpatient-contact electromedical apparatus. The standard applies to line- and battery-powered battery-power ed apparatus and to apparatus used singly or with properly connected accessory  equipment. When more than one electromedical apparatus is powered by a single power cord, the equipment assembly acts like a single apparatus in terms of risk  current limit requirements, and shall be considered as such for t he purposes of this standard. The safety and performance criteria defined in this standard are intended for use in design qualification by the device manufacturer. NOTE — The referee test methods of Section 5 are intended to provide means by which conformance with the standard can be established. These tests are not intended for use in verifying the performance of  individual devices in routine quality assurance inspections. Also, referee tests allow for the use of  alternative methods for design qualification, provided provided that devices so qualified will also meet the requirements requiremen ts of this standard when tested in accordance with the referee methods.

-14-

1.2 Exclusions This standard does not set limits for the composite risk current  when several devices are performing different functions for the same patient and are independently connected to the utility power system. This standard standard does not apply to therapeutic t herapeutic currents.

In addition, this standard does not apply to apparatus designed primarily for nonmedical applications and used in conjunction with electromedical apparatus, but located outside of  the patient care vicinity. NOTE — As indicated above, all devices cannot be readily covered by this standard. Equipment such as video cassette recorders or computing devices are now being  used in health care care facilities. facilities. Such equipment is designed to other standards. If such equipment equipment is not located in the patient care vicinity, such devices are not considered to pose a risk as they  are not likely to contact the patient.

2 DEF DEFINIT INITION IONS S For the purposes of this standard, the following definitions apply. 2.1 accessory  Device produced or recommended by the manufacturer of an electromedical apparatus, and intended to be electrically connected to that apparatus in order to make the apparatus useful or to improve its efficacy or versatility, and not a modular part of that apparatus. 2.2 auxiliary apparatus Electromedical apparatus used in conjunction with other electromedical apparatus to achieve a common purpose. NOTE — Auxiliary apparatus includes both interconnected apparatus and noninterconnec noninterconnected ted apparatus. 2.3 composite risk current Total risk current derived from the risk currents of all the apparatus associated with the patient that can flow through the patient, medical staff, or bystander. NOTE — This definition is included for reference only. A method of  derivation and limits for composite risk current are not covered in this standard.

2.4 electromedical apparatus Instrument, equipment, system, or device that directly or indirectly  uses electricity for any medical purpose. NOTE — Also included are all parts that are connected to such equipment and are required for the normal use of the equipment, including associated patient wiring  or cables. 2.5 enclosure Exterior surface of the electromedical apparatus, including all accessible parts, knobs, grips, and shafts. 2.6 exposed electrically conductive conduc tive surface External metal or otherwise electrically conductive surface that is connected to the internal circuits, mechanisms, or enclosure of an electromedical apparatus. 2.7 input part Part of the electromedical apparatus, other than a patientapplied part, that is intended to receive input signal voltages or currents from other equipment.  2.8 isolated i solated patient connection connection Connection between the patient and the electromedical apparatus that is isolated from power ground (earth)1), the utility power system, and other supporting circuitry to such a degree that the risk current flowing through the connection does not exceed the limits given in Table 1—Summary of risk current requirements requireme nts in rms microamperes (mA), provided in Section 4.2. 2.9 modular apparatus Electromedical apparatus that includes modules in its construction. 2.10 module Self-contained assembly that performs a function or class of  functions in support of the major function of an electromedical apparatus. NOTE — Modules can generally be

removed or replaced without affecting the operation of other assemblies in the apparatus. 2.11 nonoperational environmental conditions Temperature, humidity, altitude, or acceleration limits specified by the manufacturer for storage or shipment. 2.12 normal condition (NC) (NC) Condition in which all means provided against safety hazards are intact and the device is operating  as desired. 2.13 output part: Part of the electromedical apparatus, other than a patientapplied part, that is intended to deliver output signal voltages or currents to other equipment.

Electrical Safety  Made Easy 

IV.

2.14 patient-applied part Entirety of any part of the equipment that comes intentionally into contact with the patient via a patient connection. 2.15 patient-applied risk current Current flowing from the electromedical apparatus through the patient to power ground (earth) or between patient-applied parts. 2.16 patient care vicinity  Space, within a location intended for the examination or treatment of  patients, extending 6 feet (ft) (1.8 meters [m]) beyond normal location of the bed, chair, table, treadmill or other device that supports the patient during  examination and treatment. treatment. The patient care vicinity extends vertically to 7 ft, 6 inches (in) (2.3 m) above the floor. 2.17 patient connec connection tion Deliberate connection that can carry current between an electromedical apparatus and a patient. This can be a surface contact (e.g., an ECG electrode), an invasive connection (e.g., an implanted wire or catheter), or an incidental longterm connection (e.g., connective tubing).

-15-

Electrical Safety  Made Easy 

IV.

NOTE — As used in this standard, "patient connection" is not intended to include adventitious or casual contacts, such as push buttons, bed surfaces, lamps, and hand-held appliances. 2.18 patient isolation risk current Current flowing flowing from the patient to power ground (earth) through a part applied to the patient due to the unintended introduction of a voltage from an external source on the patient. 2.19 risk current: Nontherapeutic Nontherape utic current that can flow through the patient, medical staff, or bystander as a result of the use of electromedical apparatus. 2.20 single fault condition (SFC) Condition in which a single means of protection against a safety  hazard in equipment is defective, a component failure could increase the risk current, or a single external abnormal condition exists. 2.21 sink current: Current that flows into a device or any part thereof, when an external voltage is applied to it. 2.22 source current Electrical current that flows from any part of an electromedical apparatus to any other part or to power ground (earth), when no external voltages are applied. 2.23 therapeutic current: Current that is intentionally  applied to the patient for treatment of disease or disorder.

3 CLASSIFICATION OF ELECTROMEDICAL  APPARATUS  APPARA TUS AND MEASUREMENT CONDITIONS

-16-

The following Sections define specific classes of electromedical apparatus and measurement conditions and detail how these classifications are applied in this standard.

3.1 Classification of electromedical electromed ical apparatus For purposes of this standard, four categories of electromedical apparatus have been defined. For each category, risk currents are established. These four categories are listed below : a) Electrome Electromedical dical apparatu apparatuss with isolated patient connection: Electromedical apparatus intended to be connected to the patient with the patient circuit isolated from power ground (earth), utility power systems, and other circuitry. circuitry. b) Elec Electromed tromedical ical apparatus apparatus  with nonisolated patient connection: Electromedical apparatus intended to be connected to the patient. c) Electromed Electromedical ical appara apparatus tus likely to contact the patient: Electromedical apparatus that does not have a patientapplied part, but that is intended for use in the patient care vicinity. NOTE — See Section 2 for definition of patient care vicinity. d) Elec Electrome tromedical dical apparatus apparatus with no patient contact: Electromedical apparatus that is intended for use outside the patient care vicinity and that has no patient connections. 3.2 Classification of measurement conditions The following Sections define normal and fault conditions. 3.2.1 Normal operating conditions Under normal operating  conditions, a device is operating as designed with all means provided for protection against safety  hazards intact, connected properly  and securely to an approved power source and, if the device includes patient-applied parts, with such parts applied according to the manufacturer's instructions. The following followin g are considered normal operating conditions:

a) power switch switch on/power on/power switch off; b) power polarity normal/power normal/power polarity reversed (cordconnected apparatus only); c) patient grounded; grounded; patient patient not grounded. 3.2.2 Single fault condition  A single failure of a device's protection mechanism against a safety hazard or the failure of a single device component can introduce a hazard condition or lead to the existence of an external hazardous condition. The following followi ng are considered to be single fault conditions : a) power ground (earth) conductor open; b) short circuit of either barrier barrier of double insulation; c) failure of a single single component that can produce a hazardous current; d) (for equipment that that is not intended to be grounded) the application of line voltage to an input or output part or to accessible conductive hardware of the enclosure; e) (for electromedical apparatus  with isolated patient connections) the application of line voltage on a patientapplied part.

4 REQU REQUIREME IREMENT NTS S 4.1 Labeling and documentation requirements 4.1.1 Isolated patient connections connec tions (labeling) Patient connections that meet the requirements requireme nts of this standard for isolated patient connections shall be identified as being isolated at the connector of the apparatus.

NOTE — Labeling of isolated patient connections shall comply   with symbol number 3 of table DII, page 329, in the t he International Electrotechnical Commission (IEC) standard 601-1, second edition (IEC, 1988). 4.1.2 Information manuals The manufacturer shall supply the user with operating and maintenance instructions specifying how the electromedical apparatus should be operated and maintained to prevent the device's risk current from increasing  beyond the limits set by this standard for its particular category  (refer to Section Section 3.1). In addition, the manufacturer shall disclose the risk current category for which the apparatus is designed and shall identify the specific limits defined by this standard for that category.

Electrical Safety  Made Easy 

IV.

4.2 Risk current requirements requirements (general) Electromedical apparatus shall meet the applicable risk current limits of this standard under normal conditions and under the single fault conditions specified in the test methods of Section 5. Table 1 (see next page) summarizes s ummarizes these requirements requirements.. 4.2.1 Apparatus interconnection interconnection Electromedical apparatus shall meet the risk current limits of this standard when manufacturer manufacturer-designated auxiliary apparatus, modular apparatus, or accessories are attached in the quantity and combinations stipulated by the manufacturer. The manufacturer shall supply the user (and shall label the apparatus) with limitations and with directions for the interconnection of modular apparatus, accessories, and auxiliary apparatus, and with directions for the use of  convenience receptacles. 4.2.2 Cleaning and sterilization Electromedical apparatus shall meet the risk current limits of this standard after exposure to any  disinfection or sterilization process specified by the manufacturer manufacturer..

-17-

Electrical Safety  Made Easy 

4.2.3 Environmental conditions Electromedical apparatus shall meet the risk current limits of this standard after exposure to the nonoperational environmental conditions (e.g., storage, transportation, etc.) and under the  worst-case environmental operating conditions specified by the manufacturer. 4.3 Enclosure risk current 4.3.1 General Enclosure risk current, when measured with the AAMI standard test load, is that current that flows between power ground (earth) and

IV.

a) exposed exposed chassis chassis conductive conductive surfaces or hardware; or b) a 200 cm2 cm2 (centimete (centimeters rs squared) foil in contact with a nonconducting enclosure enclosure..

NOTES : 1) The frequenc frequency-w y-weighte eighted d network compensates for the allowable allowabl e increase in risk current limits with increasing frequency. For measurem measurement ent purposes with a voltmeter as shown, the limit remains constant at 1 mA/mV, independent of frequency.  With the meter connected, the entire circuit is called the "risk current tester." 2) Refe Referr to Section 5.7.2 for component requirements and tolerances. 4.3.2 Risk current limits Limits for enclosure risk current for all categories of electromedical apparatus, whether batterypowered, cord-connected, or permanently connected, and under both normal and single fault conditions, are shown in Table 1 on next page.

10,000 Ω

0.015 µF INPUT

1,000



100 Ω

Figure 1  AAM AA M I standar d test test l oad .

-18-

MILLI VOLTMETER  VOLTMETER

Categor y

Normal Condition

PatientPatient CordApplied Risk Isolation General Other Connected/ Per maCur rent Risk Current CordCordPer maBatter ynent (source (s i n k connected connected)1 nent2) current) current) 1 0 0 mA

100 mA

10 mA

N/A

5 0 0 mA

2,500 mA 5,000 mA  

ISOLATED Single Fault Condition

3 0 0 mA

5,000 mA

5 0 mA

5 0 mA

1,000 mA

5,000 mA 10,000 mA  

Normal Condition

100 mA

100 mA

10 mA

N/A

500 mA

2,500 mA 5,000 mA  

5,000 mA

1 0 0 mA  

N/A 

1,0 00 00 mA

5,0 00 00 mA 10,000 mA  

100 mA

N/A

N/A

500 mA

2,500 mA 5,000 mA  

N/A  

N/A 

1,0 00 00 mA

5,0 00 00 mA 10,000 mA  

100 mA

N/A

N/A

500 mA

2,500 mA 5,000 mA  

5,000 mA

N/A  

N/A 

1,0 00 00 mA

5,0 00 00 mA 10,000 mA  

NON-ISOLATED Single Fault Condition 3 0 0 mA Normal Condition

100 mA

LIKELY TO CONTACT PATIENT Single Fault Condition 300 mA 5,000 mA Normal Condition

100 mA

NO PATIENT CONTACT Single Fault Condition 5 0 0 mA

Electrical Safety  Made Easy 

IV.

Table 1 Sum m ar y of risk risk cur re rent nt re requi qui re rem m ents in rm s m icroamp ere res s (dc to 1 kHz) 

1) Equi Equipment pment that that has no protectively grounded (earthed) accessible parts and no means for protective grounding (earthing) of other medical equipment and which complies with the applicable requirements require ments for enclosure leakage current and patient leakage current; also mobile x-ray equipment and mobile equipment with mineral insulation.

2) Equipment specified specified to be permanently installed with a protective power ground (earth) that is electrically connected and secured at a specific location so that the connection can only be loosened or moved with the aid of a tool.

-19-

4.4 Patient-applied risk current (source current)2)

Electrical Safety  Made Easy 

IV.

4.4.1 General Patient-applied risk current, when measured with the AAMI standard test load, is that t hat current that flows between any patient-applied part and : a) pow power er ground ground (earth) (earth);; b) expos exposed ed chassis chassis conductive conductive surfaces or hardware; c) a 200 cm2 foil foil in contac contactt with a nonconducting enclosure; or d) any other patientpatient-appli applied ed parts. Patient-applied risk current is also that current that flows between all patient connections tied together and (a), (b), (c), and (d) listed above  when measured with the AAMI standard test load.

NOTE — The current shall be measured at the patient end of the cable when connected to the device. 4.6 Earth risk current 4.6.1 General Earth risk current, when measured  with the AAMI standard test load, is that current that flows in the protective earth conductor (ground conductor). NOTE — Not applicable to double insulated devices using a two-wire power cord.

4.4.2 Risk current limits Limits for patient-applied risk  current for all categories of  electromedical apparatus, whether battery-powered, battery-power ed, cord-connected, or permanently connected, under both normal and single fault conditions, are shown in Table 1.

4.6.2 Risk current limits Limits for earth risk current for all categories of electromedical apparatus, whether cordconnected or permanently  connected, and under both normal and single fault conditions, are shown in Table 1.

NOTES :

4.7 Risk current limits versus frequency  The risk current limits specified in Table 1 are for frequencies from dc to 1 kilohertz (kHz). (kHz). Above 1 kHz, kHz, the limit is increased proportionally to a maximum value 100 times the limit at 1 kHz. Above 100 kHz, the limit limit is that which is determined for 100 kHz (see Figure 2). The use of the AAMI AAMI test load automatically compensates for frequency.

1) These limits limits are are not applicable applicable to electromedical apparatus  without direct patient-applied connections. 2) The current current shall shall be measure measured d at the patient end of the cable  when it is attached to the device. The cable is specified by the manufacturer. 4.5 Patient isolation risk current (sink current)

-20-

4.5.2 Risk current limits Limits for patient isolation risk  current for electromedical apparatus with isolated patientapplied part(s), whether cordconnected or permanently  connected, are shown in Table 1.

4.5.1 General Patient isolation risk current, when measured with the AAMI standard test load, is that current that would flow into a patient-applied part if  the patient came into direct contact with a potential of 120 volts (V), 60 hertz (Hz) with respect to power ground (earth).

5 Tests This Section contains referee tests and procedures by which compliance with the requireme requirements nts of Section 4 and Table 1 can be determined.  WARNING — These tests can  WARNING expose personnel to hazardous electric shock and must be carried out with caution.

Electrical Safety 

100 mA 

N  O R  M A  L  I   Z  E  D  C   U R  R  E  N T 

Made Easy  1 mA 

500 Hz

1000 Hz

10 KHZ

100 K HZ

1 MHZ

10 MHZ

FREQUENCY (HZ) Figur e 2 — Norm ali ze zed d cur rent l im it s ve versus rsus frequency 

5.1 Compliance with the labeling requirements Compliance with the labeling and documentation requirements of  Section 4.1 shall be verified by  inspection. 5.2 Compliance with the risk current requirements (general test proced procedures) ures) The risk currents of electromedical apparatus shall be measured by the methods described in this Section. 5.2.1 Test equipment and power system 5.2.1.1 Measuring instruments The risk current tester consists of  the AAMI standard test load and a millivoltmeter as shown in Figure 1. The millivoltmeter shall measure true rms volts; however however,, it may be calibrated to true rms microamperes (mA) by employing a conversion factor of one microampere per millivolt millivolt (mV). The millivoltmeter shall have an input impedance of at least 1 megohm and have a bandwidth of dc to at least 1 megahertz (MHz) (–3 decibels). In the band from dc to 100 kilohertz (kHz), the indicated measurement shall not display an error of greater than 5 percent of  reading, and shall resolve a signal as small as 1 mV.

IV.

Instruments that indicate true rms microamperes and have internal frequency compensation identical to that shown in Figure 1 meet the requirements of this Section if the measurement indicated does not display an error of greater than 5 percent of reading and resolves a signal as small as 1 mA in the band from dc to 100 kHz. 5.2.1.2 Power source 5.2.1.2.1 For line voltage powered equipment, the tests shall be performed on a grounded power system at the rated line voltage plus 10 percent. percent. In the grounded grounded system, the potential between the neutral and grounding conductors at the receptacle selected for the test shall not exceed 3 V. 5.2.1.2.2 The power ground (earth) terminal used in these tests shall be the grounding terminal of the specific receptacle powering the instrument under test. 5.2.1.2.3 Battery-powered Battery-pow ered apparatus shall be tested while powered by the type of  battery recommended by the manufacturer and, if applicable,  while connected to line power. power. -21-

5.2.2 Test conditions

Electrical Safety  Made Easy 

IV.

5.2.2.1 General First, the apparatus shall be disconnected from all other apparatus except auxiliary  apparatus, modular apparatus, or accessories, as defined in the normative definitions (see Section 2). A single- or multi-function multi-function apparatus in a cabinet or in multiple cabinets with a single power cord connection connection is tested as a single apparatus. apparatus. Each individual individual apparatus shall also be tested independently if described by the manufacturer as a stand-alone apparatus. Tests shall be conducted at the rated line voltage plus 10 percent. 5.2.2.2 Nonconducting enclosure The risk current shall be measured from an electrically conductive foil the size of the enclosure, but not to exceed 200 cm2, in immediate contact with the enclosure. The The foil shall be placed at a location— determined by experimentation— such that the current measured to power ground (earth) is a maximum. If exposed exposed chassis hardware is likely to be touched by  personnel, then the hardware shall be treated as an exposed electrically conductive surface. Testing of nonconductive exposed surfaces of patient wiring and cables is not required.

delivers therapeutic energy to the patient (e.g., a pacemaker), the therapeutic energy shall be zero during the test. Otherwise, the instrument shall be in the active or operable mode; i.e., with output switches closed, with electrodes properly connected to dummy  loads, and with final circuit stages properly functioning but without a physiological drive signal. 5.2.2.4 Operation During the test, the apparatus shall run through a normal cycle and activate all accessories and/or auxiliary apparatus.

5.3 Enclosure risk current 5.3.1 Application The enclosure risk current tests shall apply to cord-connected, linepowered apparatus, to batterypowered apparatus with the charger connected, and to permanently connected apparatus. 5.3.2 Cord-connected, normally  grounded apparatus 5.3.2.1 Using the test circuit of Figure 3, the enclosure risk current shall be measured:

5.2.2.3 Controls During risk current tests, all operator-accessible operator-acc essible controls shall be adjusted to yield the largest risk  current found by experiment. If the electromedical apparatus normally  Reversing Switch S1

a) between between enclosure enclosure and and power ground (earth); b) betwe between en electrical electrically ly conductive surfaces and power ground (earth); c) betwe between en a 200 cm2 foil foil in contact with the nonconducting enclosure and power ground (earth).

200 cm 2 fail

Rated Line  Voltage +10%

 Apparatus Under Test

Select Per 5.3.2.1

Ground

M

Ground Switch S2

-22-

Ground Wire

Exposed Conductive Surface

Risk Current Tester

Figure 3 — Enclos Enclosur ur e ri sk cur rent test test circui t (nor m all y grou grou nd ed) 

NOTE — The 200 cm2 foil line leading from the "select" box to the apparatus under test refers to the connective mode with insulated apparatus. The line with an arrow  leading from the "select" box to the apparatus under test refers to connections made with conductive enclosure. 5.3.2.2 Each measurement is performed  when: a) the utility utility electricity supply polarity is normal and when the utility electricity supply polarity is reversed (by reversing S1). These are are normal conditions; b) the apparatus power power switch is on; the apparatus power switch is off. These are normal conditions; c) the ground switch (S2) is open; the ground switch is closed. The first is a single fault condition; the second is a normal condition; d) each barrier of double insulation is short circuited. circuited. These are single fault conditions. NOTE — The test methods for all measurement conditions are not supplied in this standard because they are device- and circuitspecific. 5.3.2.3 The power on/power off  test also applies to apparatus with nonrechargeable nonrecha rgeable batteries.

Rated Line  Voltage +10%

Reversing Switch S1

5.3.3 Cord-connected, Cord-connec ted, normally  ungrounded apparatus 5.3.3.1 Using the test circuit of Figure 4 (see next page), the enclosure risk  current shall be measured: a) between enclosure enclosure and power power ground (earth); b)between electrically conductive surfaces and power ground (earth); c) betwe between en a 200 cm2 foil foil in contact with the nonconducting enclosure and power ground (earth). 5.3.3.2 Each measurement is performed  when:

Electrical Safety  Made Easy 

IV.

a) the utility electricity supply polarity is normal; the utility electricity supply polarity is reversed (by reversing S1). These are normal conditions; b)the apparatus power switch is on; the apparatus power switch is off. These are normal conditions; c) each barrier barrier of double double insulation is short-circuited. These are single fault conditions. NOTE — The test methods for all measurement conditions are not supplied in this standard because they are device- and circuitspecific.

200 cm2 fail  Apparatus Under Test

Select Per 5.3.2.1

Risk Current Tester

Exposed Conductive Surface

Ground

Figur e 4 — Enclos Enclosur ur e ri sk cur rent test test circui t (nor m all y ungrou nd ed) 

-23-

e) all patien patientt connectio connections ns shorted together and power ground (earth); f) all patie patient nt connec connectio tions ns shorted together and any exposed, electrically conductive surface; g) all patien patientt connectio connections ns shorted together and a 200 cm2 foil in contact with the nonconducting enclosure; h) any patient connection and all other patient connections connected together.

5.3.3.3 The power on/power off  test also applies to apparatus with nonrechargeable nonrechargea ble batteries.

Electrical Safety  Made Easy 

5.3.4 Permanently connected apparatus Before line-powered apparatus is permanently installed, the enclosure risk current shall be measured according to the procedures described in 5.3.2. 5.4 Patient-applied risk current (source current)

IV.

5.4.1 Application The patient-applied risk current tests of this Section shall apply to line-powered line-powe red and battery-power battery-powered ed electromedical apparatus that has a patient connection(s).

NOTE — The test methods for all measurement conditions are not supplied in this standard because they are device- and circuitspecific. 5.4.2.2 Each measurement is performed  when:

5.4.2 Apparatus with isolated patient connection 5.4.2.1 Using the test circuit of Figure 5 (see next page), the patient-applied risk current (source current) shall be measured between: a) any patient patient connecti connection on and power ground (earth); b) any patient patient connection connection and and any exposed, electrically conductive surface; c) any patient patient connecti connection on and a 200 cm2 foil in contact with the nonconducting enclosure; d) any patient patient connection connection and and any other patient connection;

a) the utility utility switch switch (S1) (S1) is normal/the utility switch is reversed. These are are normal conditions; b) the apparatus apparatus power power switch is on/the apparatus power switch is off. These are normal conditions; c) the ground ground switch switch (S2) (S2) is open/the ground switch is closed. The first is is a single fault condition; the second is a normal condition;

120K 

Select Per 6.4.2.2.d Reversing Switch S1 Rated Line  Voltage +10%

Ground Switch S2

120V / 60Hz

Patient Connection  Apparatus Under Test

200 cm 2 fail

Select Per 5.3.2.1

Select Per 5.3.2.1 Exposed Conductive Surface

-24-

Figure 5 — Pati ent -ap pl ied ri sk cur rent test test circui t 

Risk Current Tester

d) line voltage voltage is applied applied to an input or output part or to accessible conductive hard ware of the enclosure, if not grounded under normal conditions. These are single fault conditions; e) each barri barrier er of doubl double e insulation is short circuited. These are single fault conditions. NOTE — The 120 K resistance is intended to protect the test operator. 5.4.3 Apparatus with nonisolated patient connection The patient-applied risk current shall be measured by the procedures described in 5.4.2 5.4.2.. 5.5 Patient isolation risk current (sink current) The patient isolation risk current shall be measured in each individual patient connection that is labeled "isolated" when a potential of 120 V rms, 60 Hz, is applied through a series 120 kilohm resistance to the labeled patient connection, as shown in Figure 6. The patient isolation risk current is measured with respect to power ground (earth). For solely batterypowered power ed apparatus, the patient isolation risk current is measured  with respect to an electrically  conductive surface on which the apparatus is positioned, and with an exposed conductive surface or

other external electrical connection on the apparatus apparatus grounded. This test shall be performed with the apparatus both on and off and properly connected to its electrical supply.. The patient cable shall be supply placed 20 cm away from a grounded surface. NOTE — The 120 K resistance is intended to protect the test operator.

Electrical Safety  Made Easy 

5.6 5. 6 Ea Earth rth risk risk curr curren entt 5.6.1 Applicatio Application n The earth risk current test shall apply to cord-connec cord-connected ted apparatus, to battery-powered apparatus with the charger connected, and to permanently connected apparatus.

IV.

5.6.2 Cord-connected, Cord-connec ted, normally  grounded apparatus 5.6.2.1 Using the test circuit of Figure 7 (see next page), the earth risk  current shall be measured in the protective power ground (earth). 5.6.2.2 Measurementt shall be performed Measuremen  when: a) the utility electrical electrical supply is is normal/when the utility electrical supply is reversed (by reversing S1). These are normal conditions;

1:1

Rated Line +10%

 Apparatus Under Test

120V  60Hz

Select Per 5.5 120K 

Ground

Risk Current Tester

Figur e 6 — Pati ent iso isola la ti on r isk cur cur rent test test circui t 

-25-

Electrical Safety  Made Easy 

IV.

b) the apparatus power power switch is on/the apparatus power switch is off. These are normal conditions; c) each supply supply conductor conductor is interrupted, one at a time (opening S2 and S3 in turn). This is a single fault condition; d) each barrier of double insulation is short-circuited. This is a single fault condition. 5.6.3 Permanently connec connected ted apparatus Before the line-power line-powered ed apparatus is permanently installed, the earth risk current shall be measured according to the procedures described in 5.6.2 5.6.2.. 5.7 Risk current limits versus frequenc frequency  y  5.7.1 General  When multiple risk currents of  various frequency and phase relationships are present during a single test, the resultant risk  current is related to the voltage across the AAMI standard test load. The risk current of an apparatus

shall be the largest current measured during any of the required tests and conditions. The apparatus must meet all applicable limits of Table 1. 5.7.2 AAMI standard test load  As shown in Figure 1, the test load shall be constructed using metalfilm resistors with a tolerance of 1 percent or better, and a mica- or plastic-dielectric (extended foil) capacitor with a tolerance of 5 percent or better. The AAMI standard test load has an impedance frequency characteristic (Figure (Figure 8) which is the approximate inverse of the curve of  Figure 2, which shows risk current versus frequency. 5.7.3 Risk current calculation Using the AAMI standard test load of Figure Figure 1 and a voltmeter calibrated to indicate rms millivolts, the weighted risk current is read directly from the meter, because:  V(mV rms)

I(mA rms) =

Line Interruption Switches S2• Reversing  Switch S1 Rated Line  Voltage +10% S3•

Z(k ohms)

 Apparatus Under Test

Protective Earth (Ground Conductor)

Risk  Current Tester

Figure 7 — Eart h ri sk cur rent test test circui t 

-26-

    )     B     d     ( 0    e     d    u    t     i    n    g    a -20     M    e    v     i    t    a     l    e -40     R 10

Figure 8 —  Relative  frequency  char acteri sti cs of  millivoltmeter  readi re adi ng in AAM AAM I  stan dar d t est  load of Figur e 1 

10 1

10 2

10 3

10 4

10 5

 A.1 General The rationale discusses the need for the standard and describes the basic underlying principles, empirical data, assumptions, and sources that support the requirements requireme nts and test methods adopted in the standard.  A.2 Need for the standard st andard This standard seeks to reduce the risk of inadvertent electric shock  from medical medical devices. devices. In particular,, it concerns itself with particular the risk of injury from the small currents that inevitably flow from or to electromedical apparatus. The intent of the AAMI Electrical Safety Committee was to develop a general baseline baseline standard. The extent to which the standard should be applied is to be determined by individual institutions, standards groups, and other authorities.  A.3 Classificat Class ification ion of  electromedical electrome dical apparatus and measurement conditions Changes from the second edition of  the standard (AAMI, 1985) have been made in keeping with changes to the requirements given in Section 4.  A.4 Rationa Rationale le for the t he specific provisions of the standard  A.4.1 Labeling Label ing and an d documentation requirements  A.4.1.1 Isolated pat ient connection connec tion (labeling) Fault conditions can contribute to patient risk due to source and sink  currents. The greatest risk is is with direct cardiac applications; such applications should utilize isolated patient connections. connections. In order order to better manage patients requiring  direct cardiac connection, the user should be able to readily identify  electromedical apparatus with isolated patient connections. Therefore, Therefor e, labels should appear on

the electromedical apparatus itself. The standard does not require nonisolated patient connections to carry any special labeling.

 A.4.1.2 Informa Information tion manuals ma nuals By identifying the risk current classification and risk current limits, the manufacturer is informing the user of the device's intended purpose as that purpose relates to the risk of electric shock.  Any special user actions required to ensure that the risk current limits are maintained throughout the life of the equipment should be described in the operating  instructions or maintenance manuals.

Electrical Safety  Made Easy 

IV.

 A.4.2 Risk R isk current cu rrent requirements requ irements (general) The committee felt that grounding  should not be the primary  approach to limiting the risk of  electric shock because it is possible to have a single fault failure in the grounding system. Redundant means of grounding are possible but are controlled by the user and not the manufacturer of the electromedical apparatus. If the grounding is lost or if other safety  means fail, the risk current available from the enclosure should not represent a substantial hazard to the patient. The risk current limits were changed for certain categories as compared to the previous version of this standard (ANSI/AAMI ES1— 1985). These changes have been made in order to bring this document into closer harmonization with risk current limits specified in the International Electrotechnical Commission standard, Medical electrical equipment—Part 1: General requirements for safety (IEC 601-1, 1988).

-27-

Electrical Safety  Made Easy 

IV.

-28-

In its review of the allowable risk  current levels, the committee considered the followi following: ng: a) Likel Likelihoo ihood d of stimulation stimulation of excitable tissue. The likelihood of stimulation of excitable tissue depends upon: 1) the location location of the sites at at  which current enters and leaves the body; 2) the area area of contac contact; t; 3) the amount amount of curren currentt flowing; 4) the suscep susceptibil tibility ity to mechanical stimulation; 5) the presenc presence e of a fault fault condition; 6) the probabi probability lity of of the current having a given value. Medical devices have been classified into the risk categories described in Section 3 because of the different magnitudes of risk associated with these categories. b) Surve Survey y of published published data. data. During the several years since the publication of the previous version of this standard (ANSI/AAMI ES1—1985), experience has been gained  with respect to the incidence of problems related to risk current and the probability of occurrence of the potential hazard. The available published data on currents causing ventricular fibrillation in humans have also been reexamined. The following have been noted: 1) The combina combination tion of an open power ground (earth)  wire and a person touching both a conductive part of the enclosure and the patient is a low probability event. 2) The combina combination tion of an open power ground (earth)  wire and a person touching both a conductive part of the chassis and the distal end of an invasive cardiac connection is a low probability event. 3) In one study study,, Raftery Raftery (1975) found that the smallest current that produced a

disturbance in rhythm in humans was 80 mA. In a second study, Watson (1973) found that the smallest current that produced ventricular fibrillation in humans was 15 mA. 4) Mec Mechani hanicall cally y induced induced ventricular fibrillation has been observed during cardiac catheterization at zero current. 5) The human human data data obtained obtained by Starmer (1973) and  Watson  Wa tson (1973) follow, follow, reasonably well, a normal distribution for currents to 300 mA. All patients are not equally susceptible to current-induced ventricular fibrillation. According to Figure A.1, there is approximately a 1 percent probability of fibrillation at 30 mA. 6) Cur Current rent perce perception ption is is a function of contact location, contact pressure, skin condition, moisture, and contact contact area. Experiments report a wide range of current perception. Dalziel (1968) reports that only approximately 1 percent of the population can perceive 500 mA passing from the fingers of one hand to the fingers of the other hand. Tan and Johnson (1990) report that 300 mA produces a strong sensation for electrodes placed 10 cm apart on the upper arm. Levin (1991) reports that nearly all the population will perceive 500 mA without any reaction for current passing from the finger on one hand to the underside of the wrist on the other hand. Levin further reports that, on the underside of the  wrist, the stratum corneum (layer of dead material on the skin surface) is not as thick as that on the forefinger and, therefore, the

sensitivity to current perception might be higher higher.. Startle current is that level of perception current that,  when first perceived, might result in a nurse or other clinician's involuntary reaction to the sudden sensation of perception current. This uncontrolled reaction is of great concern. 7) Worldw orldwide, ide, since since the the advent of risk current standards, concern about grounding, and use of better practices in handling catheters and invasive cardiac connections, there have been no reports of incidents involving risk current passing through the patient. NOTE — A minority of the AAMI Electrical Safety Committee were opposed to increasing risk current limits unless scientific studies supported higher limits.  Also, in order to harmonize harmonize with IEC 601-1 and to allow for additional fault conditions as compared to the current standard, the test measurement classifications of "normal condition" (NC) and "single fault condition" (SFC)  were introduced. Measuring risk current under SFC Measuring is important because: — compo components nents can can fail; — singl single e faults exist exist that are not now considered in the standard; — faul faults ts can occur in accessory accessory equipment.  A.4.2.1 Apparatus interconnection i nterconnection The total risk current associated  with a device can be a function of  the modules, accessories, and interconnections interconnectio ns used with the t he device. Voltage differences can occur between different parts of a device, particularly if current flows in the grounding circuit. Thus, a remote accessory powered from the device or from a separate source is, for purposes of the standard, considered part of the device.

 Auxiliary power power outlets may be provided for powering additional devices. The labeling requiremen requirements ts provide some assurance that the user has appropriate guidance about the limitations applying to equipment or accessories connected to an auxiliary power receptacle.  A.4.2.2 Cleaning and sterilizat ster ilization ion The long-term effects of repeated disinfection or sterilization of a device on risk currents must be considered because of possible degradation of insulating materials.  A.4.2.3 Environm Environmental ental conditions condi tions Temperature, humidity, atmospheric pressure, mechanical shock, and similar environmental constraints can have an effect upon the risk currents. To protect the patient, the risk current limits must also be met in the intended environment.

Electrical Safety  Made Easy 

IV.

 A.4.3 Enclosure Encl osure risk ri sk current cur rent The 100 mA NC values were selected for cord-connected and permanently connected apparatus on theoretical grounds. For a typical ground resistance of 0.2 ohms, 100 mA of enclosure risk  current, measured as in Figure 3, requires that 500 mA flow in the power ground (earth) wire. This  would be a major fault condition. The 300 mA SFC values for isolated, nonisolated, or likely-to-co likely-to-contactntactpatient, cord-connected apparatus  were selected on the basis of  reaction current measurements by  Levin and the low probability of  risk current reaching the distal end of an invasive heart connection via another person. person. Levin reported reported that currents of 300 mA will not produce sensations leading to a "startle" reaction (Levin, 1991). The 500 mA limit for SFC for cordconnected, no-patient-contact equipment was selected because there is no concern about this current reaching the patient. The 5,000 mA limit for SFC for permanently connected equipment is basically a power ground (earth)

-29-

Electrical Safety  Made Easy 

IV.

 wire current, because the apparatus is, by definition, permanently grounded. grounded. This is the current allowed from the enclosure of permanently connected equipment if the power ground (earth) wire were opened. opened. The permanently connected power ground (earth) wire is not expected to open.

that produced ventricular fibrillation in humans are considered, the probability is approximately  zero. The 50 mA SFC limit for isolated equipment was selected because the probability of causing ventricular fibrillation is low if the data of  Figure A.1 are extrapolated and is approximately zero if only  observed values of currents producing ventricular fibrillation in humans are considered.

 A.4.4 Patient-appli Pat ient-appli ed risk ris k current curren t (source current) The 10 mA NC limit was selected as the current that may flow directly  into the heart continuously continuously.. Figure  A.1 shows that the probability of  inducing ventricular fibrillation is very small if the data on which the chart is based are extrapolated. extrapolated. If  only observed values of currents

The 100 mA SFC limit was selected for nonisolated equipment, because if this current enters and exits the surface of the body, then only a fraction will reach the heart.

Starmer 60 Hz D = 1.25mm  Watson 60 Hz D = 2.0mm 2 .0mm

99.9 99 95    t    n    e    c    r    e     P    e    v     i    t    a     l    u    m    u     C

+ +

90 50

+ + +

20

+

+ +

+ +

5 1

0.1 0

50

10 0 15 0 2 00 2 50 30 0 35 0 40 0 4 50 5 00 Microamperes

Figure A. A.1 1 — Norm al pr obabili ty plot 

-30-

 A.4.5 Patient Pat ient isolation i solation risk current (sink current) The 50 mA SFC value for isolated equipment was allowed because of  the low probability of line voltage appearing on a patient, and because of the low probability of 50 mA inducing ventricular fibrillation. For line voltage to appear on a patient, a power ground (earth) wire must be open and there must also be a fault in basic insulation. insulation. If the data data in Figure A.1 are extrapolated, then 50 mA has a low probability of  inducing ventricular fibrillation. If  only recorded human data are considered, the probability of  inducing fibrillation at 50 mA is approximately zero.  A.4.6 Ear th risk ri sk (ground risk) current The original standard of 1978 and the revised edition of 1985 did not include earth current as a potential risk current. current. This was not considered to be an issue as most medical devices of the era utilized conductive enclosures requiring  grounding. Thus, earth current current was effectively measured as enclosure (chassis) current under the open ground condition. In the last 25 years, however, the change to nonconductive enclosures negated this equality. The enclosure current is now  measured as the capacitivecoupled current to a 200 cm2 foil in contact with the enclosure. enclosure. This current bears little resemblance to the earth risk current due to the current-limiting current-limitin g characteristics of  the capacitive coupling of the nonconductive enclosure. The earth risk current does not pose a direct risk to the patient or medical personne personnel. l. However, excessive earth risk current, either by design or internal breakdown,  will raise the potential of the device's ground with respect to true power ground (earth) as represented by structural elements, modular wall units, cold water pipes, and other installed piping, or by an adjacent receptacle power

ground (earth). Contact with with such elements and a second device under test will result in current flow.. Thus, the committee felt that flow leaving the earth risk current unmeasured and unlimited constituted a potential hazard that should be avoided. The allowable values for earth risk  current detailed in this standard are not thought to pose a direct hazard as the current is safely  returned to earth. The values selected were chosen to avoid any  significant increase in the current flowing through the protective grounding system of the installation and to be consistent with the limits of IEC 601-1 for power ground (earth) and nonconducting  enclosures. Further, the 2.5 mA  limit in normal mode is within the limit for isolation monitors set by  the National Fire Protection  Association (NFPA, (NFPA, 1993, Section 3.4.3.3), which specifies that isolation monitors should not alarm at 3.75 mA.

Electrical Safety  Made Easy 

IV.

 A.4.7 Risk R isk current cu rrent limits li mits versus frequenc frequency  y  Figure 2 of the standard was derived from strength/freque strength/frequency  ncy  data for perceptible and lethal currents (Geddes and Baker Baker,, 1971). The flat portion between 100 kHz and 1 MHz does not reflect physiological data obtained with purely sinusoidal currents. Stimulation has been observed  with complex waveforms at high frequencies, but little data are available. In the absence of data, it  was deemed prudent not to extrapolate beyond 100 kHz.  A.5 Tests Tests The test procedures documented in Section 5 of the standard s tandard provide referee test methods for verifying  compliance with the requirements of Section 4. These referee referee tests are not necessarily intended for purposes of manufacturing or quality control (although these applications are not precluded), as equivalent measurements may be obtainable by other means.

-31-

 ASSOCIATION FOR THE ADVANCEMENT  ASSOCIATION ADVANCEMENT OF MEDICAL INSTRUMENTATION. Safe current limits for electromedical apparatus.  ANSI/AAMI ES1—1985. Arlington (Vir.): AAMI, 1985. ISBN 0-910275-50-5.

Electrical Safety  Made Easy 

DALZIEL, CF. CF. Reevaluatio Reevaluation n of lethal electric electr ic currents. IEEE Trans Indus Gen Appl, GA-4, 1968, vol. 1, no. 5, p. p. 467-476. GEDDES, LA. and BAKER, LE. Response to the passage of electric currents through the body. J Assn Adv Med Instrum, 1971, vol. 5, p. 13-18. INTERNATIONAL ELECTRO INTERNATIONAL ELECTROTECHNICAL TECHNICAL COMMISSION. Medical electrical equipment—Part 1: General requirements for safety, 2nd ed. IEC 601-1. Geneva: IEC, 1988. LEVIN, M. Perception of chassis leakage current. Biomed Instrumentation and Technolog Technology, y, 1991, vol. 25, no. 2, p.135-140.

IV.

NATIONAL FIRE PROTECTION ASSOCIATION. Standard for health care facilities. NFPA NFPA 99-1993. Quincy (Mass.): NFPA, 1993. RAFTERY, EB., GREEN, HL., and YACOUB, MH. Disturbances of heart rhythm produced by 50 Hz leakage currents in human subjects. Cardiovascular Research, March 1975, vol. 9, no. 2, p. 263-265. STARMER, CF. and WHALEN, RE. Current density and electrically induced ventricular fibrillation. fibrillation. J Assn Adv Med Med Instrum, 1973, vol. vol. 7, no. no. 1, p. 3-6. TAN, KS. and JOHNSON, DL. Threshold of sensation for 60 Hz leakage TAN, current: Results of a survey. Biomed Instrumentation and Technology, 1990, vol. 24, no. 3, p. 207-211.

Bibliography 

 WATSON, AB., WRIGHT  WATSON, WRIGHT,, JS., and LAUGHMAN, J. Electrical thresholds thresholds for ventricular fibrillation in man. Med J Australia, 1973, vol. 1, p.1179-1182.

 You can contact BAPCO: BAPCO 3200 Sencore Drive Sioux Falls, SD 57107 1-800-419-4000

http://www.bapcoinfo.com email: bapco@bapcoinfo [email protected] .com

-32-

Glossary  Of Ter Terms ms  AMPACITY:  AMPA CITY: Current-carrying capacity of electrical conductors expressed in amperes.

 ANESTHETIZING LOCA LOCATION: TION: Any area area of the facility that has been designated for the administration of  any flammable or nonflammable inhalation anesthetic agents in the course of examination or treatment, including the use of such agents for relative analgesia.

CONDUCTIVE: Materials, such as metals, that are commonly considered electrically conductive, and materials that, when tested, have a resistance not exceeding 1,000,000 ohms. Such materials are required  wheree electrostatic  wher electrostatic interconnecti interconnection on is necessary. necessary.

CRITICAL EQUIPMENT:

Equipment that is essential to the safety of the occupants in the facility.

CRITICAL SYSTEM:  A system of of feeders and and branch circuits in nursing homes and custodial care facilities arranged for connection to the alternate power source to restore service to critical receptacles, task illumination, and equipment.

DIRECT ELECTRICAL PATHWAY TO THE HEART:  An external external conductive conductive pathway pathway,, insulated insulated except at its ends, one end of which is in direct contact  with the the heart muscle muscle and the the other outside outside the body body,, that is accessible for inadvertent or intentional contact  with grounded grounded objects objects or energized energized,, ground-refer ground-referenced enced sources. Catheters filled with conductive fluids and electrodes, such as may be used for pacing of the heart, are examples of direct electrical pathways to the heart.

DOUBLE-INSULATED APPLIANCES:  Appliances having  Appliances having an insulatio insulation n system comprising comprising both both basic insulation necessary for the functioning of the appliance and for basic protection against electrical shock and supplementary insulation. The supplementary  insulation is independent insulation provided in addition to the basic insulation to ensure protection against electric shock in case of failure to the basic insulation.

ELECTRODE:  A device intended intended to probe probe an electrically conductive connection through a cable to a patient. There are several types: electrode intended intended to generate generate a  Active Electrode: An electrode surgical effect at its point of application to the patient. electrode intended intended to comDispersive Electrode:  An electrode plete the electrical path between patient and appliance, and at which no surgical effect is intended. It is often called the "indifferent electrode", the "return electrode", the "patient plate", or the "neutral electrode".

EXPOSED CONDUCTIVE SURFACES: Those surfaces that are capable of carrying electric current and that are unprotected, uninsulated, unenclosed, or unguarded, permitting personal contact.

FAULT CURRENT:  A current current in an accidental accidental connection between an energized and a grounded or other conductive element resulting from a failure of  insulation, spacing, or containment of conductors.

FREQUENCE: The number of oscillations, per unit of  time, of a particular current or voltage waveform. The unit of frequency is the Hertz (Hz). (The unit of frequency  used to be "cycles per second", a term no longer may consist of  preferred). Note : The waveform may components having many different different frequencies , in which case it is called a complex or nonsinusoidal waveform.

to a patient, but may be conveyed from exposed metal parts of an appliance to ground or to other accessible parts of an appliance.

LINE ISOLATION MONITOR:  An instrument instrument that continually checks the hazard current from an isolated surface to ground.

GROUND-FAULT CIRCUIT INTERRUPTER:

MACROSHOCK: The effect of large electric current

 A device whose whose function function is to interrupt interrupt the electric electric circuit circuit to the load when a fault current to ground exceeds some predetermined value that is less than that required to operate the overcurrent protective device of the supply  circuit.

(milliamperes or larger) on the body.

GROUNDING SYSTEM: a system of conductors

MICROSHOCK: The effect of small electric currents (as low as 10 microamperes) on the body. To be hazardous, such currents must be applied to a conductor inside or very near the heart.

that provides a low-impedance return path for leakage and fault currents. It coordinates with, but may be locally  more extensive than, the grounding system described in  Article 250 of NFPA NFPA 70, National National Electric Electric Code. Code.

mV:

Millivolt.

mA:

Milliampere.

HAZARD CURRENT: for a given set of connections

facility where patients are examined or treated.  Note : Business offices, offices, corridors, lounges,day rooms, dining rooms, or similar areas are not classified as patient  care areas.

in an isolated power system, the total current that would flow through a low-impedance if it were connected between either isolated conductor and ground. The various hazard currents are: Fault Hazard Current: The hazard current of a given isolated power system with all devices connected except the line isolation monitor. Monitor Hazard Current: The hazard current of the line isolation monitor alone. Total Hazard Current: The hazard current of a given isolated system with all devices, including the line isolation monitor, connected.

IMPEDANCE: Impedance is the ratio of the voltage drop across a circuit element to the current flowing through the same circuit element. The circuit element may consist of any combination of resistance, capacitance, or inductance. The unit of impedance is the Ohm.

INTRINSICALLY SAFE:  As applied applied to equipment equipment and wiring, equipment and wiring that are incapable of  releasing sufficient electrical energy under normal or abnormal conditions to cause ignition of a specific hazardous atmospheric mixture. Abnormal conditions may include accidental damage to any part of the equipment or wiring, insulation or other failure of  electrical components, application of overvoltage, adjustment and maintenance operations, and other similar conditions.

ISOLATED PATIENT LEAD:  A patient patient lead whose whose impedance to ground or to a power line is sufficiently  high that connecting the lead to ground or to either conductor of the power line results in a current flow  below a hazardous limit in the lead.

ISOLATED POWER SYSTEM:  A system comprising an isolating transformer or its equivalent, a line isolation monitor, and ungrounded circuit conductors.

ISOLATION TRANSFORMER: A transformer transformer of  the multiple-winding type, with the primary and secondary windings physically separated, that inductively  couples its ungrounded secondary winding to the grounded feeder system that energizes its primary   winding..  winding

LEAKAGE CURRENT:  Any current, current, including including capacitively coupled current, not intended to be applied

PATIENT CARE AREA:  Any portion portion of a health health care care

PATIENT PA TIENT-CARE-RELATED -CARE-RELATED ELECTRICAL  APPLIANCE:  An electrical electrical appliance appliance that that is intended intended to be used for diagnostic, therapeutic, or monitoring purposes in a patient care area.

PATIENT EQUIPMENT GROUNDING POINT: A jack or terminal terminal that that serves as a collectio collection n point for redundant grounding of electrical appliances serving a patient vicinity or for grounding other items in order to eliminate electromagnetic problems.

PATIENT LEAD:  Any deliberate deliberate electrical electrical connection connection that may carry between an appliance and a patient. This may be a surface contact (such as an ECG electrode), an invasive connection (such as an implanted wire or catheter), or an incidental long-term connection (such as conductive tubing). Adventitious or casual contacts such as a push button, bed surface, lamp, hand-held appliance, and so fourth, are not considered patient leads.

PATIENT VICINITY: In an area in which patients are normally cared for, the patient vicinity is the space with surfaces likely to be touched by the patient or an attendant who can touch the patient. Typically in a patient room, this is a space within the room 6 ft. (1.8m) beyond the perimeter of the bed in its normal location and extending vertically within 7 ft. 6 in. (2.3m) of the floor.

REACTANCE: The component of impedance contributed by inductance or capacitance. The unit of  reactance is the Ohm.

REFERENCE GROUNDING POINT: A terminal terminal bus that is the equipment grounding bus, or an extension of the equipment grounding bus, and is a convenient collection point for installed grounding wires or other bonding wires where used.

 WET LOCAT LOCATIONS: IONS: Those patient care areas that are normally subject to wet conditions, including standing  water on the the floor, floor, or routine dousing dousing or drenchi drenching ng of the  work area. area. Routine Routine housekeepin housekeepingg procedures procedures and incidental spillage of liquids do not define a wet location.

1-800-419-4000 • www.bapcoinfo.com

View more...

Comments

Copyright ©2017 KUPDF Inc.
SUPPORT KUPDF