Document No.DRT-MAN-055 Revision No.Rev.05 Released By 2015-11-23
Digital Radiography Technologies
About This Manual The EConsole1 User Manual has prepared for the appropriate use of digital radiography operating console software developed by DRTECH Inc., This manual provides the instructions for the use of EConsole1. The manual contains the following contents.
Operation Environment, Environment, Installation and Method for EConsole1
Functions and Instructions for EConsole1
Please contact us if there are any problems during the use of the product. WebSite
http://www.drtech.co.kr
Customer Suppor Supportt Team
Email Email
[email protected] [email protected] Tel +82 31 730 6805 Fax +82 31 730 6899
To Enhance The Quality Of Human Life!
DRT-MAN-055 (Rev. 06)
Copyright - EConsole1 Manual Information contained in this manual is confidential and is the property of DRTECH Inc., which is provided exclusively for authorized licenses of DRTECH Inc. The purpose of the manual is to assist the users to comprehensively utilize the product of DRTECH Inc. The information contained in this manual may not be altered or distributed without prior written consent of DRTECH DRTECH Inc., Inc., and may not be disclosed to unauthorized personnel. Information contained in the manual may be altered without prior notice. Copyright © 2013 2013 DRTECH DRTECH Corporation. Corporation. All rights reserved.
DRT-MAN-055 (Rev. 06)
Table of Contents 1. INTRODUCTION ................. ................................. ................................. .................................. ................................. ............................ ............ 6 2. SYSTEM REQUIREMENTS ................. ................................. ................................. ................................. ............................... ............... 7 2.1. SUPPORTED DIGITAL X-RAY DETECTOR ......................................... ......................................................... ....................... ...... 7 2.2. PC SPECIFICATION FOR THE SOFTWARE .............................................. ............................................................. .................. 7 3. INSTALLATION .............. ............................... ................................. ................................. .................................. .................................. ................. 8 3.1. INSTALLATION PREREQUISITES .............. ............................... .................................. .................................. ............................ ........... 8 3.2. INSTALLATION ............... ................................ ................................. ................................. .................................. ................................. ................ 14 3.3. LANGUAGE & NUMBER OF DETECTOR SETTING ............................... ............................................... ................... ... 15 3.4. LICENSE PROTECTION TYPE SELECTION ................ ................................. ................................. .............................. .............. 17 3.5. OPTION ............... ................................. .................................. ................................. ................................. .................................. ......................... ....... 18 3.6. HOSPITAL & DEPARTMENT INFORMATION ................. ................................. ................................. ......................... ........ 19 3.7. INSTALLATION PATH CONFIGURATION .............. ............................... .................................. ................................ ............... 20 3.8. START MENU FOLDER CREATION ................ ................................. ................................. ................................. ...................... .....22 3.9. ADDITIONAL TASK SETTING ................ ................................ ................................. .................................. .............................. ............. 23 3.10. INSTALLATION COMPONENT VERIFICATION ................. ................................. ................................. ...................... .....24 3.11. USB H AND S WITCH DRIVER INSTALLATION ............... ............................... ................................. ...................... ..... 25
............................... ................................. .................................. ............................. ............ 26 3.12. INSTALLATION COMPLETE ............... 3.13. M AP FILE & EC ALI1 CONFIGURATION ............... ................................ ................................. .............................. ..............27
3.14. SETTINGS AFTER INSTALLATION ................ ................................. ................................. ................................. ..................... .... 30 ................................. ................................. ................................. ................................. ................ 31 3.15. UNINSTALL ECONSOLE1 ................ ................................. ................................. ................................. ................................. ................32 4. ECONSOLE1 OVERVIEW ................
4.1. PROGRAM START-UP ................ ................................. .................................. ................................. ................................. ...................... ..... 32 4.2. MAIN SCREEN OF THE PROGRAM ............... ............................... ................................. .................................. ...................... .....34
.................................. .................................. ................................. ................................. ................................. .............................. ..............39 5. LIST .................
DRT-MAN-055 (Rev. 06)
5.1. WORKLIST ................. .................................. .................................. ................................. ................................. .................................. ................... 40 5.2. STUDYLIST ................. .................................. .................................. ................................. ................................. .................................. ................... ..54 6. EXAM ................. .................................. ................................. ................................. .................................. .................................. ........................... .......... 74 ................................. .................................. .................................. ................................. ................................. ...................... ..... 77 6.1. TOOLBAR ................
6.2. PROJECTION LIST & IMAGE PROCESSING TAB ............................... ................................................ ..................... .... 84 6.3. TRUVIEW ART ............................... ............................................... ................................. .................................. ................................ ............... 89 7. CONFIGURATION ............... ............................... ................................. .................................. ................................. .......................... .......... 90 7.1. DISPLAY .............. ............................... .................................. .................................. ................................. .................................. ........................ ...... 92 7.2. SYSTEM ............... ................................. .................................. ................................. ................................. .................................. ......................... .......93 7.3. IMAGE ............... ............................... ................................. .................................. .................................. .................................. .......................... ......... 99 7.4. STORAGE ................ ................................. .................................. .................................. ................................. ................................. .................... ... 107 7.5. ACCOUNT ............... ................................ .................................. .................................. ................................. ................................. ..................... .... 112 7.6. BLOCK LIST ................ ................................. .................................. ................................. ................................. .................................. ................... 113 7.7. DATASET .............. ............................... .................................. .................................. ................................. .................................. ........................ ......114 7.8. NETWORK ................. .................................. .................................. ................................. ................................. .................................. ................... 115 7.9. DETECTOR ................................ ................................................. .................................. ................................. ................................. .................... ... 124 7.10. RIS CODE ................ ................................. .................................. ................................. ................................. .................................. ................. 129 7.11. PROJECTION EDITOR ............... ................................ .................................. ................................. ................................. ..................... ....137 7.12. GENERATOR ................ ................................. ................................. ................................. .................................. ............................... .............. 143 APPENDIX A. GLOSSARY ............... ................................ ................................. ................................. ................................. .................. 147 APPENDIX B. TERMINAL EMULATOR .............. ............................... .................................. ............................... .............. 150 APPENDIX C. UPDATE ................ ................................. .................................. ................................. ................................. .................... ... 166 APPENDIX APPEN DIX D. D. PORTABLE PORTABLE FUNCTION FUNCTION............................... ................................................ ............................... .............. 169
DRT-MAN-055 (Rev. 06)
1. Introduction EConsole1
is
digital
radiography
operating
console
software
specialized for the FLAATZ and EVS series digital X-ray detector developed by DRTECH Inc. EConsole1 provides an integrated solution for X-ray projection. It integrates
with the X-ray generator and the digital
detector.
Furthermore, EConsole1 acquires and processes images. In addition, it complies with DICOM standards and is able to transmit and receive data with the PACS system, and print images through the DICOM printer.
Figure 1. X-Ray Projection Solution Utilizing EConsole1 (Wired Detector)
Figure 2. X-Ray Projection Solution Utilizing EConsole1 (Wireless Detector)
DRT-MAN-055 (Rev (Rev.. 06)
6
2. System Requirements 2.1. Supported Digital X-ray Detector EConsole1 supports the following Digital X-ray Detectors developed by DRTECH.
FLAATZ 330N LAN
FLAATZ FLAAT Z 500 LAN
FLAATZ FLAAT Z 560 LAN
FLAATZ 600 Wireless LAN
FLAATZ 750E LAN
FLAATZ FLAAT Z 760 LAN
EVS 4343 LAN
2.2. PC Specification for the Software Table 1. PC Specifications Items
Recommended
CPU
Intel i5-3470
RAM
8 GB
HDD
500 GB
VGA ODD OS
NVIDIA GeForce GT630 1GB DVD Recorder Windows 7 Pro 32 / 64bit
Display Size
23 inch
Display Resoluti Resolution on
1920 x 1080
* Optimal performance is not guaranteed for PCs that do not comply with
the recommended specifications.
DRT-MAN-055 (Rev (Rev.. 06)
7
3. Installation 3.1. Installation Prerequisites 3.1.1. Windows 7 Windows 7 users must set the User Account Control (herein after ‘UAC UAC’’) permission and disable the firewall.
3.1.1.1.
UAC Setting
① Cl Clic ickk tth he ssttart art ico icon n
.
② Click ‘Control Panel’ Panel’. ③ Select ‘User Accounts’ Accounts’. ④ In the User Accounts window, click ‘User Accounts’ Accounts’. ⑤ In the User Accounts task window, click ‘Change User Account Control Setting’ Setting’. ⑥ Adjust the slider to the ‘Never notify,’ notify,’ and then click ‘OK OK’’.
3.1.1.2.
Disabling the Firewall
① Cl Clic ickk tth he S Sttart ico con n
.
② Click ‘Control Panel’ Panel’. ③ Click ‘Security Security’’. ④ Click ‘Windows Firewall’ Firewall’. ⑤ Click ‘ Turns Windows Firewall on or off ’ on the left. ⑥ Click ‘Off (not recommended),‘ recommended),‘ and then click ‘OK OK’’.
3.1.1.3.
Set Power Options
① Click 'Setting' on the right hand menu. ② Click 'Control Panel'. ③ Click 'Power Options' ④ Select
DRT-MAN-055 (Rev (Rev.. 06)
'High
performance'
or
'Energy
Manager
High
8
Performance Mode'
3.1.1.4.
Set Power Options
① Click 'Setting' on the right hand menu. ② Click 'Control Panel'. ③ Click 'Power Options' ④ Select 'Choose when to turn off the display' menu on the left side. ⑤ Set 'Put the computer to sleep' to 'Never' and click Save changes.
3.1.1.5. Disable Windows update setting ① Click 'Setting' on the right hand menu. ② Click 'Control Panel'. ③ Click 'Windows Update' ④ Select 'Change settings' menu on the left side. ⑤ Set 'Important updates' to 'Never check for updates(not recommended'
DRT-MAN-055 (Rev (Rev.. 06)
9
⑥ Uncheck 'Give me recommended updates the same way I receive important updates' ⑦ Uncheck ''Give me updates for other Microsoft products when I update Windows'
DRT-MAN-055 (Rev (Rev.. 06)
10
3.1.2. Windows 8 Windows 8 users must set the User Account Control (herein after ‘UAC UAC’’) permission, disable the firewall and run the program as the system administrator.
3.1.2.1.
UAC Setting
① Click ‘Setting Setting’’ on the right hand menu. ② Click ‘Control Panel’ Panel’. ③ Select ‘User Accounts’ Accounts’. ④ In the User Accounts window, click ‘User Accounts’ Accounts’. ⑤ In the User Accounts task window, click ‘Change User Account Control Setting’ Setting’. ⑥ Adjust the slider to the ‘Never notify,’ notify,’ and then click ‘OK OK’’.
3.1.2.2.
Disabling the Firewall
① Click ‘Setting Setting’’ on the right hand menu. ② Click ‘Control Panel’ Panel’. ③ Click ‘Security Security’’. ④ Click ‘Windows Firewall’ Firewall’. ⑤ Click ‘ Turns Windows Firewall on or off ’ on the left. ⑥ Click ‘Off (not recommended),‘ recommended),‘ and then click ‘OK OK’’.
3.1.2.3.
Run as Administrator
Windows 8 users must run the Installation File and and Program as as the Administrator.
Install the program by right clicking (press and hold, if touch screen) and selecting ‘Run as Administrator’ Administrator ’.
After completing the installation, set the icon, so that it always runs as administrator. ① Right click (press and hold, if touch screen) the shortcut icon. ② Click ‘Properties Properties’’. ③ Click the ‘Advanced Advanced’’ icon in the ‘Shortcut Shortcut’’ tab tab‘‘. ④ Check the ‘Run as administrator’ administrator ’ box, click on ‘OK OK’’.
DRT-MAN-055 (Rev (Rev.. 06)
11
3.1.2.4.
Do not change Compatibility mode
EConsole1.exe's right clicking(press and hold, if touch screen) and select 'Compatibility' tab.
Make sure that
EConsol EConsole1.exe's e1.exe's compatibility mode is not
checked.
3.1.2.5.
Disabling the Windows update
① Click ‘Setting Setting’’ on the right hand menu. ② Click ‘Control Panel’ Panel’. ③ Select ‘Windows Update’ Update’. ④ In the Windows Update window, click ‘Change settings’ settings’. ⑤ In the Change setting window, Change to ‘Never check for updates(not updates(n ot recommended) recommended)’’. ⑥ Uncheck 'Give me recommended updates the same way I receive important updates.
Figure 3. Windows update setting window
3.1.2.6.
Power options setting(Only use Laptop)
Click 'Setting' on the right hand menu.
Click 'Control Panel'.
Click 'Power Options'
DRT-MAN-055 (Rev (Rev.. 06)
12
Figure 4. Windows power options setting window
Select
'High
performance'
or
'Energy
Manager
High
Performance Mode'
Powerr options setting Powe
Click 'Setting' on the right hand menu.
Click 'Control Panel'.
Click 'Power Options'
Select 'Choose when to turn off the display' menu on the left side.
Set 'Put the computer to sleep' to 'Never' and click Save changes.
3.1.2.7. Disable SmartScreen ① Click 'Setting' on the right hand menu. ② Click 'Action Center' ③ Click 'Change Windows SmartScreen setting' ④ Set 'Don't do anything(turn off Windows SmartScreen)
DRT-MAN-055 (Rev (Rev.. 06)
13
⑤ Click 'OK'
3.2. Installation Run the provided setup file.
A minimum of 10GB of available HDD space is required for the proper installation of EConsole1!
0.5.6 is the version of the product, and may change.
Windows 8 users must run the file as ‘administrator ’.
Right-click the file, select ‘Run as administrator’ administrator’ in in the menu.
If the setup file is run, you will be presented with the following screen as in, [Figure 5] 5].. Click ‘Next Next’’ to start the installation installation..
Figure 5. Setup Wizard Window
DRT-MAN-055 (Rev (Rev.. 06)
14
3.3. Language & Number of Detector Setting Select a language to be used in the Econsole1 vet program. Click ‘Next Next’’ after selection.
Figure 6. Language & Number of Detector Setting
① Language Setting Please select the appropriate language.
It may be set after the program installation.
It can be set through Configuration - System - Language.
Please refer to 7.2.1. 7.2.1. Language (p. 94) 94) for setting after installation.
DRT-MAN-055 (Rev (Rev.. 06)
15
② Number of Detector Setting Please select whether to use 1 or 2 detectors.
Be Advised!
For Dual Mode using 2 detectors, there are restrictions to models which are able to use the detectors simultaneously.
Dual Mode (2 detector use) supported detectors
FLAATZ FLAAT Z 500 LAN
FLAATZ FLAAT Z 560 LAN
FLAATZ 600 Wireless LAN
FLAATZ 750E LAN
FLAATZ FLAAT Z 760 LAN
EVS 4343 LAN
EVS 3643 LAN
It may be set after program installation.
It can be set through Configuration - Detector.
Please refer to 7.9 Detector (p. 124) 124) for setting after installation.
DRT-MAN-055 (Rev (Rev.. 06)
16
3.4. License Protection Type Selection
Figure 7. License Protection Type Selection Selects license protection type you purchased.
Hardware License Protection (Dongle Key)
Utilizes a license key embedded within a special USB device. This device must be attached to the PC in order for Econsole1 to operate.
Software License Protection
The license is provided as a serial number to be entered during installat installation. ion.
DRT-MAN-055 (Rev (Rev.. 06)
17
3.5. Option Select the model and input the serial number of the X-ray Detector to be used with the EConsole1. Click ‘Next Next’’ after selection.
It may be set after program installation installation..
It can be configured through Configuration - Detector. Please refer to to 7.9 Detector (p. (p. 124) 124)..
Figure 8. Option: Detector Setting
① X-ray Detector Detector Model Selection Selection
Set the X-ray Detector, which will be used with FConolse1.
Select the appropriat appropriate e model.
② Enter the X-ray Detector Detector Serial Numbe Numberr
Enter the serial number of detector.
DRT-MAN-055 (Rev (Rev.. 06)
18
3.6. Hospital & Department Information Enter the Hospital name and department name.
Figure 9. Hospital & Department Information
DRT-MAN-055 (Rev (Rev.. 06)
19
3.7. Installation Path Configuration 3.7.1. Select EConsole1 Install Folder
You may designate the folder in which to install the program through the ‘Browse Browse’’ icon.
Please verify that there is sufficient space on the HDD where the program will be installed!
The program requires a minimum of 2GB of available space.
After specifying the location location,, click ‘Next Next’’.
Figure 10. EConsole1 Install Folder Selection
DRT-MAN-055 (Rev (Rev.. 06)
20
3.7.2. Select Image Save Folder
You may designate the folder in which to install the program through the ‘Browse Browse’’ icon.
location,, click ‘Next Next’’. After specifying the location If user want to change it is possible only 'drive:\Images' folder.
Figure 11. Image Save Folder Selection
DRT-MAN-055 (Rev (Rev.. 06)
21
3.8. Start Menu Folder Creation Specify a start menu folder name for the program icons, or select not to create a start menu entry. Click ‘Next Next’’.
Figure 12. Start Menu Folder Selection
DRT-MAN-055 (Rev (Rev.. 06)
22
3.9. Additional Task Setting Select whether to create shortcut icons in the desktop and Quick Launch Bar. Click ‘Next’. ‘Next’.
Figure 13. Additional Task Setting
DRT-MAN-055 (Rev (Rev.. 06)
23
3.10. Installation Component Verification Displays the summary of configured component componentss to be installed. Start Sta rt the installation by clicking the ‘Install Install’’ icon, and please wait while installing.
Figure 14. Installation Component Verificat Verification ion
DRT-MAN-055 (Rev (Rev.. 06)
24
3.11. USB Hand Switch Driver Insta Installation llation Install USB Hand Switch Driver for F600, EVS4343 and EVS3643.
Figure 15. USB Hand Switch Driver Installation
DRT-MAN-055 (Rev (Rev.. 06)
25
3.12. Installation Complete Once the installation has completed, it is strongly recommended that you allow the PC to be restarted.
Make sure “Y “Yes, es, restart the
computer now” is selected and click next.
Figure 16. Installation Complete
DRT-MAN-055 (Rev (Rev.. 06)
26
3.13. Map File & ECali1 Configura Configuration tion There are three files that are provided with your detector panel that must be copied to the hard drive in order for the software to operate. These files are unique to your specific panel.
3.13.1.
Map File
DRTECH provide the MAP CD like below.
그림 17. MA MAP P file C CD D
The files can be found on the CD in the folder named MAP. These files consist of the following:
Pixel MAP file: [Detector [Detector serial number].MAP
Gain MAP file: [Detector serial number].GMP
Panel MAP file: [Detector serial number].PMP
Please copy these files to [EConsole1 installation folder]\Map folder. Even you use 2 or 3 detector do not make special folder and copy it to [EConsole1 installation folder]\Map folder. EConsole1 search the Map files using set detector's serial number. DRT-MAN-055 (Rev (Rev.. 06)
27
So, Map file name and the registered serial number must be matched.
그림 18. Serial nu number mber & ma map p files
DRT-MAN-055 (Rev (Rev.. 06)
28
3.13.2.
ECali1 Configuration
You must setup various settings of detector using ECali1. For setting up using ECali1, please refer to the ECali1 manual.
Figure 19. ECali1
DRT-MAN-055 (Rev (Rev.. 06)
29
3.14. Settings after Installation Windows 8 users must set EConsole1 so that it runs as administrator.
Right click (press and hold) the EConsole1 shortcut icon in the desktop.
Click ‘Properties Properties’’.
Select ‘Advanced Advanced’’ in the ‘Shortcut Shortcut’’ tab.
Check the ‘Run as administrator’ administrator ’ box, then click ‘OK OK’’.
DRT-MAN-055 (Rev (Rev.. 06)
30
3.15. Uninstall EConsole1
Click ‘Uninstall EConsole1’ EConsole1’ in Windows start menu – All Programs - EConsole1.
DRT-MAN-055 (Rev (Rev.. 06)
31
4. EConsole1 Overview 4.1. Program Start-up In order to run the program, you can double click the EConsole1 icon in the desktop or select ‘EConsole1 EConsole1’’ by Windows Menu – All Programs – EConsole1.
Figure 20. EConsole1 Shortcut Icon
If the program is properly initiated, a log-in screen as the one you see in [Figure 21] will 21] will appear.
Figure 21. Log-in Screen Enter the ID and Password, then click the Login icon.
DRT-MAN-055 (Rev (Rev.. 06)
32
If you enter an invalid ID or Password, you will be presented with a screen as in [Figure 22] 22] and and will not be able to log in.
Figure 22. Log-in Error
DRT-MAN-055 (Rev (Rev.. 06)
33
4.2. Main Screen of the Program You will be presented with EConsole1 main screen as in [Figure 23] 23] with a successful log-in.
Figure 23. Main Screen of EConsole1
DRT-MAN-055 (Rev (Rev.. 06)
34
4.2.1. Function Tabs It pertains to area ⓐ in [Figure 23] 23],, and there are two tab icons on the upper left corner.
Figure 24. Tab Icons of EConsole1 These tabs are used to carry out various functions of EConsole1. The functions of each tab are as follows. Manages the Worklist and Studylist. Worklist is the list of patients. Studylist is the list of examined studies. Refer to to 5. List (p. (p. 39) 39) for more details Performs X-ray projection of selected a patient in the Worklist, or reviews selected examined study in the Studylist. Refer to to 6. Exam (p. (p. 74) 74) for for more details.
DRT-MAN-055 (Rev (Rev.. 06)
35
4.2.2. Status Information Display It pertains to the lower area ⓑ in [Figure 23] 23],, and displays various status information as in [Figure 25] 25]..
Figure 25. Status Information Bar
① Hardware & Connection Connection Status Display Displays the status of the hardware and the connection status between equipment requiring connections [FLAATZ-600] Displays the acquisition mode in detector. There are AED and Switch box mode. [EVS-4343, EVS-3643] Displays the acquisition mode in detector. There are AED, Switch box, Sync mode. If the user want to change the mode click the icon and the default mode could change using ECali1.(ECali1 - Configuration - Detector X Exposure mode) Mode change sequence. ∙ USB Switch box : USB Switch box AED ∙ Sync T Trigger rigger : Sync Trigger AED ∙ AED : AE AED D only Displays the connection status and the available space of the currently connected HDD. The
displayed
number
is
the
ratio
of
the
remaining HDD volume. Displays the connection status(On/Off) status(On/Off) with the Xray detector. DRT-MAN-055 (Rev (Rev.. 06)
36
[EVS4343, EVS3643] Displays the temperature. Displays the connection status with Storage Server. When you click this icon, the pop-up window will be show. You can change the PACS or do test in it. The UI is almost same as Configuration's PACS menu. Refer to to 7.8.3 PA PACS CS (p. (p. 119) 119) Displays the connection status with the DICOM Printer. Displays the connection status with the Modality Worklist server. [FLAATZ-600, EVS-4343, EVS3643] Displays the battery condition of the wirelessly connected detector. The displayed number indicates the ratio of remaining battery life. [EVS-4343, EVS-3643] EVS-3643] Battery is in wireless charging charging.. [FLAATZ-600, EVS4343, EVS-3643] Displays the signal strength of the wirelessly connected detector.
② Current Time & Log-in Log-in Information Display
Displays the current date, time and user.
The program can be terminated by clicking the Logout icon.
DRT-MAN-055 (Rev (Rev.. 06)
37
③ Configuration Icon
Opens the Configuration screen for various sett settings. ings.
(p. 90) 90) for Refer to 7 to 7.. Configuration (p. for more details.
DRT-MAN-055 (Rev (Rev.. 06)
38
5. List The List tab displays the list of patients and studies as described below.
Worklist Worklist
Displays the list of patients scheduled for examination.
Allows users to add/edit/delete patients on the list.
Allows users to proceed to exam for the selected patient.
When you run the Econsole1 Vet for the first time you will be asked to specify the list length. It will be saved automatically. Please refer to 5.1 to 5.1 Worklist (p.40) (p.40) for more details. details.
Studylist Studylist
Displays the examined studies.
Allows users to edit/delete the examined studies on the list. Note, editing and deletion of images on the list will not affect to the PACS server.
Allows users to transmit examined images to the PACS server.
Please refer to to 5.2 Studylist (p. (p. 54) 54) for for more details.
Please note!
The information on the Worklist and Studylist are stored on the Local Storage (HDD) by default.
Every time Econsole1 is launched, a new Worklist is retrieved from the Worklist server and stored on the Local Storage.
Contents on the list which are edited and deleted in Econsole1 Vet only affects the Local Storage by default. However, edited records may be transmitted to the PACS server using the Send command.
DRT-MAN-055 (Rev (Rev.. 06)
39
5.1. Worklist Displays the list of patients. The list is displayed when the
icon, on the upper
portion of the List tab, is clicked.
5.1.1. Main Screen of the Worklist The main screen of the Worklist is as seen in [Figure 26]. 26].
Figure 26. Worklist Screen
DRT-MAN-055 (Rev (Rev.. 06)
40
① List of Patients
The types of information are as noted in [Table 2] 2]..
The displayed information and its order may be set in
Configuration. The display order may be changed by clicking and dragging the name of the information type(ID, Name and etc.).
If you run the EConsole1 first, please set the column size at your convenience it will save automatically. Table 2. Information Type of the Worklist Type
Description Accession Number.
Acc. No. ID Name
X-ray study registration number Patient ID Patient name
Date/Time
Expected time of study
Study Desc.
Study Description
D.O.B. Sex Order Date
Date of Birth Sex of the patient Time of order
Patient selection in the Worklist.
Patient can be selected by left-clicking the mouse.
The selected patient is displayed in blue .
Patient can be edited and deleted after selection.
Projection can be performed in the Exam tab as a selected patient is double-clicked.
It is equivalent to clicking the ‘Exam Exam’’ icon after selecting a specific patient.
Please refer to to 6. Exam (p. (p. 74) 74) for for more details.
DRT-MAN-055 (Rev (Rev.. 06)
41
② Search Bar & Refresh Icon
Search Bar
Searches the values of Patient ID, Name, and Access Number within the Worklist.
If the
is clicked after entering the value to be
searched, the list will only display relevant works.
If the
icon is clicked without entering a specific value,
an advanced search window will appear.
Advanced Search Window
Figure 27. Advanced Search Window
DRT-MAN-055 (Rev (Rev.. 06)
42
Performs
searches
utilizing
multiple
conditions
simultaneously.
Only displays works satisfying all entered details.
Refresh icon Retrieves recently registered patients from the PACS server.
③ Add/Edit/Delete Icon Table 3. Add/Edit/Delete Icon Icon
Description Adds a new patient. Add Window is opened when the pertinent icon is clicked.
Add
Please refer to to 5.1.2 Add Window Window (p. (p. 45) 45) for for more details. Edits the selected patient information on the Worklist. Only allows editing of patient information added by EConsole1. Edit Window is opened when the patient is selected and the edit
Edit
icon is clicked. Please refer to to 5.1.4 Edit Window (p. (p. 51) 51) for for more details. details. Deletes the selected patient. Only deletes the patients added by EConsole1.
Delete Import the excel file to worklist Please refer to for more details. Excel
Please refer to for more details.
import Save patient list to detector. This function works only FLAATZ-600, EVS-3643 detector. Portable
Please refer to for more details.
DRT-MAN-055 (Rev (Rev.. 06)
43
④ Exam Icon
Switches to the Exam tab to exam of the selected patient. Please refer to to 6. Exam (p. (p. 74) 74) for for more details.
DRT-MAN-055 (Rev (Rev.. 06)
44
5.1.2. Add Window Add Window appears when the
icon is clicked [Figure 28] 28]..
Figure 28. Add Window
DRT-MAN-055 (Rev (Rev.. 06)
45
① Patient Information Study, patient, radiographer information is entered. Details for each category are as described in [Table 4]. 4]. Table 4. Patient Information (Add Window) Category
Description
ID
Patient ID
Accession Number
Registered identification number
Name
Patient Name
Operator
Radiographer Name
Sex
Sex of the patient
Birth Date
Date of birth of the patient
Age
Age of the patient
② Emergency Icon
If this icon is clicked, a default value is set as the ID and Accession Number and switches over to the Exam tab.
③ Projection Information
Displays information of projection. In no information is provided
initially,
the
projection
where
projection
information can be set by clicking ④ Edit icon.
④ Edit Icon
Opens
the
projection
setting
window
information may be entered.
Please refer to to 5.1.3. Edit Projection (p. (p. 49) 49) for for more details.
DRT-MAN-055 (Rev (Rev.. 06)
46
⑤ OK Icon
Saves the patient with the entered information and closes the window.
⑥ Exam Icon
Saves the patient with the entered information and directly switches to the Exam tab to exam.
Please refer to 6 to 6.. Exam (p. (p. 74) 74) for for more details.
⑦ Cancel Icon
Cancels the work addition and closes the window.
* Note If multiple projections are registered for a single patient, each projection receives a different Accession Number [Figure 29] 29]..
Figure 29. Multiple Projections Registration in a Single Event If the OK icon is clicked, each projection registered in the Worklist as 2
DRT-MAN-055 (Rev (Rev.. 06)
47
different studies. [Figure 30] 30]
Figure 30. Two studies for one patient
In order to register projections as a single study, the Accession Numbers for each projection must be equally modified prior to clicking the ‘OK OK’’ icon. [Figure 31] 31]
Figure 31. Enter the Same Accession Number to Register as a Single Study
DRT-MAN-055 (Rev (Rev.. 06)
48
5.1.3. Edit Projection Window It is the window which appears on the Add Window when the Edit icon is clicked.
Figure 32. Edit Projection Window
① Anatomy Information
Select the anatomical region to be examined.
Projection and Position are determined subsequent to the selected anatomical region.
Stand
Receptor Type will be changed by clicking the icon.
Table
DRT-MAN-055 (Rev (Rev.. 06)
Wireless1
Wireless2
Wireless3
49
② Projection & Position List
Specific selection of regions selected in Anatomy.
③ Add Icon
This icon registers the currently set Projection, Position, Receptor Type information.
The currently set Projection, Position and Receptor Type information is also registered when the selected Position is double clicked.
④ Registered Registered Projection Information Information
Displays the list of registered Projection, Position and Receptor Type.
The Receptor Type can be altered by double clicking the Position in Dual Mode.
⑤ Projection Information Information Sequence Edit Icon
Changes the projection order.
Up
Down
⑥ Delete Icon
Deletes the selected item.
⑦ OK Icon
Saves and closes the window.
DRT-MAN-055 (Rev (Rev.. 06)
50
⑧ Cancel Icon
Cancels and closes the window.
5.1.4. Edit Window The Edit Window opens when a study is selected from the Worklistt and the Edit icon is clicked. Worklis (Caution) Edit Window is not able to be modified on the study information, retrieved from the PACS server. server.
Figure 33. Edit Window
DRT-MAN-055 (Rev (Rev.. 06)
51
① Study Information
Edits the information of the radiographer and the patient.
Information are as described in [Table 5] 5].. Table 5. Study Information (Edit Window)
Category ID Name Accession Number Operator
Detail Patient ID Patient Name Registered identification number Radiographer Name
Study Description
Study Description
Sex
Sex of the Patient
Birth Date Age Schedule date & time
Date of Birth of the Patient Age of the Patient Scheduled date and time of imaging
② OK Icon
Saves and closes the window.
③ Cancel Icon
Cancels and closes the window.
5.1.5. Edit Window ① Make a patient list ✓
Use the template found at 'installed path\ Config\Worklist_sample.xlsx' to make a patient list.
DRT-MAN-055 (Rev (Rev.. 06)
52
② Limitation
You can select among the 'M', 'F', 'U' in patient Sex.
Birth Date Date and Study Date's format is 'MM/DD/YYYY'.
It is not support other format. Birth Time and Study Time's format is 'HH:MM:SS'. It is not support other format.
③ Excel import
Click the excel Import and load the file.
DRT-MAN-055 (Rev (Rev.. 06)
53
5.2. Studylist Displays the examined study list. The list is visible if the clicked.
icon at the top of the List tab is
5.2.1. Main Screen of the Studylist The main screen of the Studylist is as seen in [Figure 34]. 34].
Figure 34. Main Screen of the Studylist
DRT-MAN-055 (Rev (Rev.. 06)
54
① List of the Examined Patients
Displays the list of various information of the examined patients.
One row represents one patient information and one ID
may have multiple studies. The column of the list may be changed by dragging the name of the category (ID, Name and etc.).
You can change the list's height using Ctrl key + Mouse wheel or using Ctrl key + arrow key
Types of information are as noted in [Table 6] 6].. Table 6. Information Types of Patient
Category Name
Patient name
ID
Patient ID
Sex
Sex of the patient
D.O.B. Date/Time
Description
Date of Birth Date / time for recent examined study
Patient Selection
Patient can be selected by clicking the left button of the mouse.
Selected patient is displayed in blue. From the ‘① ‘① Examined Examined Patient List’ List’ and ‘② ‘② Examined Examined Study List per Patient’ Patient’.
Selected patient may be edited or deleted.
② Examined Study Study List per Patient Patient
Displays detailed study information of the selected patient on the ‘① ‘① List List of the Examined Patients’ Patients’.
Types of information are as noted in [Table 7] 7]..
DRT-MAN-055 (Rev (Rev.. 06)
55
Table 7. Information Types of Examined Study List Items
Description Accession Number
Acc. No.
Registered identification number
Study ID
RIS code
Study Desc.
Study Description
Status
[Number of images] / [Number of projections]
Date/Time
Date / time for examined study.
③ Thumbnail
Displays the images from the selected patient.
A check mark is added to thumbnail of the selected images for the opening or transmissi transmission. on.
The thumbnail can be selected by double clicking.
Images transmitted to the PACS server have a green outline.
1/1 : Series number / image number
④ Search Bar & Refresh Icon
Search Bar
Searches the values of Patient ID, Name, and Access Number within the Studylist.
If the
is clicked after entering the value to be
searched, the list will only display relevant works.
Refresh Icon Renews the list to the latest state.
DRT-MAN-055 (Rev (Rev.. 06)
56
⑤ Icons Table 8. Studylist Icons Icon
Description Detailed overview of a single selected image from the list of thumbnails.
Open Study
Please refer to to 6. Exam (p. (p. 74) 74) for for more details. Repeat exam for the selected study in the Exam tab.
Repeat Exam
Please refer to to 6. Exam (p. (p. 74) 74) for for more details. Performs merging of two studies or patients. Please refer to to 5.2.2 Merge Window (p. (p. 58) 58) for for more
Merge
details. Stores images on a USB, CD, or DVD, and print through the DICOM Printer. Please refer to to 5.2.3 Export Window (p. (p. 60) 60) for for more
Export
details. Edits the selected patient or study information in the Studylist. Edit Window appears if the Edit Icon is clicked.
Edit
Please refer to to 5.2.4 Edit Window (p. (p. 65) 65) for for more details.
Deletes the selected item.
Delete
Save the patient list to excel. Excel export It makes selected series's report. Please refer to for more details.
Report
⑥ Send Icon
Transmits the selected images to the PACS server.
DRT-MAN-055 (Rev (Rev.. 06)
57
5.2.2. Merge Window
Merges the images and studies.
Allows editing when the image and patient information do not correspond with each other.
Figure 35. Merge Window
① & ② Merge Subjec Subjectt Imaging In Information formation
Study lists appear on each side.
Required items can be easily searched using the search tool.
Identical items selected items in ① are located in ② if the SameList is checked.
DRT-MAN-055 (Rev (Rev.. 06)
58
DRT MAN 055 (Rev (Rev.. 06)
58
③ Merge Icon
Transmits the selected items.
Everything is saved while being transmit transmitted. ted.
④ Close Icon
Closes the Merge window.
DRT MAN 055 (Rev (Rev.. 06)
59
5.2.3. Export Window Saves the images of the selected study from the Studylist to a USB, CD or DVD, and prints it through the DICOM Printer.
5.2.3.1.
Drive
Figure 36. Export USB Screen
Stores the images in the local drive.
Only files in BMP, JPG, PNG and TIF format may be exported.
If the local drive is selected, the current list of stored files are displayed.
Export File Format
Selects Format for storage.
Image is stored to the USB once the OK icon is clicked.
File name is automatically designated and it will be stored It will be stored at 'driver\ID_Name\' folder. .
DRT MAN 055 (Rev (Rev.. 06)
60
5.2.3.2.
Print
Figure 37. Export - Print Screen
Prints through the DICOM Printer.
The DICOM Printer must be connected.
① Print Setting
Printer Selection
Selects a DICOM Printer.
You can register using 'Config' - 'Network' - 'Printer' menu.
Film Orientation
Film Size
Selects a film orientati orientation on between portrait/landscape. Selects the size of the film.
Layout Sets the print layout.
DRT MAN 055 (Rev (Rev.. 06)
61
The ‘Divide Divide’’ icon must be clicked in order to apply any changes made in Layout.
Zoom
Performs zoom in/out of image.
② Preview
Thumbnail Screen
The image to be printed appears in the thumbnail list on the right.
Preview Screen
If an image is selected in the thumbnail, a large image appears in the center preview screen.
Merge Icon
Merges the divided screen which has been separated by ‘Divide Divide’’.
Only adjacent screens of the same sizes can be merged
The merged shape must be a square.
Divide Icon
Rotation Icon
Divides the screen by the value set in ‘Layout Layout’’. Rotates the screen 90 degrees clock-wise.
Flip Icon
Reverses the up and down sides. Print Icon
Run a paper print
③ DICOM Print button
Performs printing.
DRT MAN 055 (Rev (Rev.. 06)
62
5.2.3.3.
CD / DVD
Figure 38. Export - CD / DVD
Stores the images on a CD or DVD.
Only DICOM files can be Exported.
A CD/DVD Recorder device is required.
① CD / DVD Media Configur Configuration ation
Selects the target media to be exported.
Volume Label
Configures the name of the target media.
Total Space
CD, DVD are available.
Displays the total available space of the target media.
Insert Closes the CD / DVD Recorder device.
DRT MAN 055 (Rev (Rev.. 06)
63
Eject
Opens the CD / DVD Recorder device.
② Export File List
Displays the list of files to be Exported.
③ Export Process
Estimated Time
Presents the expected time needed for Export.
Time Left
Displays the time left of the estimated time.
④ Task Icons
Burn
Performs the Export task.
Cancel
Cancels the Export task.
DRT-MAN-055 (Rev (Rev.. 06)
64
5.2.4. Edit Window 5.2.4.1.
Edit Patient Information
Figure 39. Edit Window on Studylist (Patient Information)
Only patient information (within the red outline) can be edited when the Edit icon is clicked in the patient list of the Studylist (upper list).
OK Icon
Saves and closes the window.
Cancel Icon
Cancels and closes the window.
DRT-MAN-055 (Rev (Rev.. 06)
65
5.2.4.2.
Edit Study Information
Figure 40. Edit Window on Studylist (Study Information)
Only study information (within the red outline) can be edited when the Edit icon is clicked in the projection list (lower list) of the Studylist.
OK Icon
Saves and closes the window.
Cancel Icon
Cancels and closes the window.
DRT-MAN-055 (Rev (Rev.. 06)
66
5.2.5. Report 5.2.5.1.
Report
Figure 41. Create Report
Create report using Report button in the Stu StudyList. dyList.
① Patient information
Check or change the Selected series's patient information.
② Attached image
User can attach up to 4 images using the 'Image load' button.
DRT-MAN-055 (Rev (Rev.. 06)
67
③ Write opinion
User can write opinion.
④ Print / Save / Setting
User can print the report.
Save a report to PDF file.
Save a report.
Set the report Format.
DRT-MAN-055 (Rev (Rev.. 06)
68
5.2.5.2.
Image load
Figure 42. Create Report
Layout button
Select 1X1, 1X2, 2X2 layout. Image load button
OK button
Select image Attach the selected image to report
Cancel button
Cancel to attach images
DRT-MAN-055 (Rev (Rev.. 06)
69
5.2.5.3.
Select the image
Figure 43. Select the image
You can select the image when you click "Image load" button. Search button Search the patient using input information. Patient/Series/Image list
OK button
Select the image. Load the selected image.
Cancel button
Cancel to load the image.
DRT-MAN-055 (Rev (Rev.. 06)
70
5.2.5.4.
Input the opinion
Figure 44. Select the image
Input the selected series opinion.
DRT-MAN-055 (Rev (Rev.. 06)
71
5.2.5.5.
Set report form
Figure 45. Set report form
User can set the report form using 'setting' menu.
Address/ Extra Info
Input the Hospital address or further information.
Session
Input the Session.
Default values are "RECOMMENDATION", "FINDING", and "CONCLUSION".
Visible setting
Select the printing report's repor t's contents contents..
All selection is default value.
Save as default button.
OK button
Save current value as default. Use selected form.
Cancel button
Cancel the selected form.
DRT-MAN-055 (Rev (Rev.. 06)
72
5.2.5.6.
Print
Figure 46. Print
User can print the report using print button.
DRT-MAN-055 (Rev (Rev.. 06)
73
6. Exam
The Exam tab processes exam and edit images.
Figure 47. Main Screen of Exam Tab
① Patient Information
Displays ID, patient name, study description.
② Detector Status Display
Displays the status of the detector to be used in exam.
Blue: Ready state
DRT-MAN-055 (Rev (Rev.. 06)
74
Orange: Busy state
③ Image Display Screen
Area - display images.
④ Toolbar
Collection of edit image icons
The width of the column may be changed.
Please refer to to 6.1 Toolbar (p. (p. 77) 77) for for detailed information.
⑤ Projection List & Image Processing Tab
May view the list for images, or set the post-processing parameter which will be applied to the images.
Please refer to to 6.2 Projection List & Image Processing Tab Tab (p. 84) for detailed informat 84) information. ion.
⑥ Generator Condition Setting Setting
Only available with integration with X-ray generator.
Tube Voltage (kVp), Tube Current (mA), Tube Current Amount (mAs) setting.
If the tube current amount icon, mAs, is clicked, the unit may be changed to exposure time (seconds).
Patient Size
Selects the volume of the patient‘ patient‘s body.
Sets value of the tube current, tube voltage and tube current amount, are renewed subsequent to the patient’ patient ’s volume.
AEC
Sets the AEC to be used.
Density
DRT-MAN-055 (Rev (Rev.. 06)
75
Focus
Configures the imaging density. Configures the Focal Size.
Grid
Configures the applied Grid.
⑦ Save / Indicator Indicator / Send Icons
Save Icon
Stores the images on the Local Storage.
Send Icon
Transmits the images to the PACS server.
Transmits all images performed to this icon with a Long-
click to the PACS server. Indicator Icon
Standby
Ready
Exposure
Standby: Default status. May begin exam.
Ready: Ready for the detector to get image.
Exposure: Emits x-ray to the detector.
DRT-MAN-055 (Rev (Rev.. 06)
76
6.1. Toolbar Collection of icons required for image analysis and editing.
6.1.1. Toolbar Overview
① Toolbar Width Adjustment Icon
The width can be changed by clicking this icon.
The width may be set from 1 to 4.
② Tools List
Displays the currently configured list of icons.
The icons can be edited in the Edit Toolbar window.
③ Edit Icon
Opens the Edit Toolbar window for editing the icons.
Please refer to 6.1.2 Edit Toolbar (p. 78) 78) for detailed information information..
DRT-MAN-055 (Rev (Rev.. 06)
77
6.1.2. Edit Toolbar
Figure 48. Edit Toolbar Window
① Applied Icons Box
It is the currently applied icons in the Exam tab.
Icons in this box can be deleted by drag & drop of the icon into ‘② ‘② Tools Tools Collection’ Collection’.
Icons can be added into this box by dragging icons in ‘② ‘② Tools Collection’’ and dropping it into the empty location. Collection
DRT-MAN-055 (Rev (Rev.. 06)
78
② Available Icons
Contains all of icons.
③ Add Shutter Icon
Adds icons with user specified ROI Shutter Sizes.
The following window appears with the click of the icon.
Figure 49. Add Sutter Size Window
④ Shutter Delete Icon
Only deletes user added Shutter.
⑤ OK Icon
Applies the modified tools arrangement and closes the window.
⑥ Cancel Icon
Cancels and closes the window.
DRT-MAN-055 (Rev (Rev.. 06)
79
6.1.3. Toolbar Icons All of toolbar icons, noted in [Table 9] 9].. Table 9. Toolbar Icons Icon
Description Sets the ROI size to 8× 8×10 Sets the ROI size to 10× 10×12 Sets the ROI size to 14× 14×17 Sets the ROI size to 17× 17×17 Sets the ROI size to the user specified size Sets the ROI into a square shape Sets the ROI into a circular shape Sets the ROI shape into a user specified shape Displays the “R" marker on the screen Displays the "L" marker on the screen Displays the "ERECT" marker on the screen Displays the "SUPINE" marker on the screen Displays the "SITTING" marker on the screen Displays the "STRESS" marker on the screen Displays a user entered text on the screen Displays the "PA" marker on the screen
DRT-MAN-055 (Rev (Rev.. 06)
80
Displays the "AP" marker on the screen Displays the "OBL" marker on the screen Displays the "LAT" marker on the screen Displays the "FLEX" marker on the screen Displays the "EXT" marker on the screen Displays the "↑ "↑" marker on the screen Displays the "← "←" marker on the screen Displays the "↓ "↓" marker on the screen Displays the "→ "→" marker on the screen Arrange windows side-by-s side-by-side ide symmetrical symmetrically ly Arrange windows up-and-dow up-and-down n symmetri symmetrically cally Rotates the image 90 degrees clockwise Rotates the image 90 degrees counter-clockwise Zoom-in and out of the image Fits image to screen Partially expands the image Moves the image up/down/left/right Displays the screen in 1× 1×1 Displays the screen in 2× 2×1 Displays the screen in 1× 1×2
DRT-MAN-055 (Rev (Rev.. 06)
81
Displays the screen in 2× 2×2 Measures the length of the image Measures the angle of the image Verifies the DICOM Header information Deletes any annotation or markers indicated on the image using other tools Adjusts the window level and width of the image Inverts the image Stitch images Displays the histogram Captures the image. Prints the image. Resets the edited image into its initial form Deletes the image Ruler calibration Duplicate the image Change mouse sensitivity Run the dark scan manually(EVS-43 manually(EVS-4343, 43, EVS-3643 only) Draw ellipse annotation
Draw rectangle annotatio annotation n
DRT-MAN-055 (Rev (Rev.. 06)
82
Draw polygon annotation Draw free line annotation Draw arrow annotation Select annotation
DRT-MAN-055 (Rev (Rev.. 06)
83
6.2. Projection List & Image Processing Tab The Projection List & Image Processing Tab can be used through the tab icon on the upper-right corner.
Figure 50. Projection List (Left) & Image Processing Tab (Right)
DRT-MAN-055 (Rev (Rev.. 06)
84
6.2.1. Projection List Displays a list of studies categorized by projections.
Figure 51. Projection List Tab
① & ④ Projection an and d Po Position sition
Projection can be selected by clicking ① icon.
④ presents the available positions of the selected projection projection..
Allows verification of thumbnails, position designations, and Receptor type.
POP-UP function
When you click the mouse right button you can change the options.
ROI Marker visible : Set the default ROI and Marker visible when the image comes from detector.
Find auto position ∙ All projection : Find
unexposed position among the all
projections When you save or send a image. ∙ One projection : Find unexposed position in the selected projection when you save or send a image.
DRT-MAN-055 (Rev (Rev.. 06)
85
∙ Do not apply : do not find unexposed postion when you save or send a image.
② Projection Guide Icon
The projection guide for the selected projection.
Figure 52. Projection Guide
The presented projection guide may be closed using the left click of the mouse.
③ Edit Projection Icon
Please refer to 5.1.3 Edit Projection (p. (p. 49) 49) for detailed information.
DRT-MAN-055 (Rev (Rev.. 06)
86
6.2.2. Image Processing Tab * Function for advanced users * It is utilized to acquire a more appropriate image by adjusting the post-process parameter of the completed image.
Figure 53. Image Processing Tab
① Histogram
Displays the histogram of the currently selected image.
Auto WL: Sets the window level automatically.
DRT-MAN-055 (Rev (Rev.. 06)
87
② Image Enhancing Parameters
Adjusts the image enhancing paramet parameters. ers.
The adjustable parameters are as noted in [Table 10] 10].. Table 10. Image Enhancing Parameters Image
Description
Processing Contrast
Adjusts the contrast
Edge
Adjusts the Edge
Mask
Adjusts the mask size
Bright
Adjusts the brightness
Smooth
Adjusts the smoothness
③ Parameter Parameter Function Icon
Wizard
Opens the multi-image viewer window.
8 other types of images may be comprehended based on the currently set parameter.
Apply
Reset
Applies the currently set parameter to the image. Resets the parameter to its initial setting.
Save
Saves the currently set parameter and applies it to ensuing imaging of the pertinent projection-positi projection-position. on.
DRT-MAN-055 (Rev (Rev.. 06)
88
6.3. Truview ART * It is for advanced user. It is works only EVS-3643, EVS3643 detector for sharpness
① Use TRUVIEW ART Default value IS "OFF". If you set "ON", it will be adjusted selected image. ② Magnified screen It magnified selected position in the image viewer. It can select using double click in the image viewer ③ Compare button Compare with not adjusted/adjusted TRUVIEW ART. 2 image come up. Left image is not adjusted TRUVIEW ART, Right image is adjusted TRUVIEW ART. ④ Save button Save a TRUVIEW ART parameter in the Selected Position. It will take more than 3 seconds when you adjust the TRUVIEW ART
DRT-MAN-055 (Rev (Rev.. 06)
89
7. Configuration Configures the user settings of EConsole1. Administrator Administrator privileges are required for configuration. Contents of configuration are saved and applied immediately following changes. Categories of Configuration are as noted in [Table 11] 11].. Table 11. Configuration Category Category
Display
Description The tab (List/Exam) of EConsole1 may be viewed simultaneously by division depending on the number of connected monitors, and performs relevant configurat configurations. ions. Please refer to to 7.1 Display (p. (p. 92) 92) for for detailed informatio information. n. Allows configuration of language, date display, hospital
System
information and etc. Please refer to 7.2 to 7.2 System (p. (p. 93) 93) for for detailed information. Allows configuration of image storage folder, delete policy,
Image
annotation information and etc. Please refer to to 7.3 Image (p. (p. 99) 99) for for detailed information. Performs configuration related to PC Storage (HDD).
Storage
Please refer to to 7.4 Storage (p. (p. 107) 107) for for detailed information. Manages the accounts of EConsole1 users
Account
Please refer to to 7.5 Account (p. (p. 112) 112) for for detailed information. Configures unnecessary RIS code filtering subsequent to
Block List
ultrasound or other devices. Please refer to to 7.6 Block List (p. (p. 113) 113) for for detailed
DRT-MAN-055 (Rev (Rev.. 06)
90
Category
Description information. Configures the DICOM Header Format.
Dataset
Please refer to to 7.7 Dataset (p. (p. 114) 114) for for detailed information. Configures the network information, such as Local, PACS,
Network
Printer and etc. Please refer to to 7.8 Network (p. (p. 115) 115) for for detailed information. Configures the Detector which will be connected with the
Detector
EConsole1. Please refer to to 7.9 Detector (p. (p. 124) 124) for for detailed information. Configures the RIS Code.
RIS Code
Please refer to 7.10 RIS Code Code (p. 129) 129) for detailed information. Allows the configuration of projection and position.
Projection
Please refer to to 7.11 Projection Editor (p. (p. 137) 137) for for detailed information. Configures the Generator which will be connected.
Generator
Please refer to to 7.12 Generator (p. (p. 143) 143) for for detailed information.
DRT-MAN-055 (Rev (Rev.. 06)
91
7.1. Display
Figure 54. Display Configuration Option
Description
1 Monitor
Dual Monitors * Caution: Unstable option. *
DRT-MAN-055 (Rev (Rev.. 06)
92
7.2. System
Performs configurations affecting the overall EConsole1.
Subcategories Language Date Institution Options Worklist
Description Selects system language. Selects date notation. Enters the information of the institution using the EConsole1 Enables and disables automa automatic tic running of EConsole1 at startup of Windows. Edits the type and order of information presented in the Worklist.
DRT-MAN-055 (Rev (Rev.. 06)
93
7.2.1. Language
Figure 55. Language Configuration Screen
Configures the system language.
Items
Description
Korean
Set when you use the Korean window.
Chinese
Set when you use the Chinese window.
Japanese
Set when you use the Japanese window.
Portuguese
Set when you use the Portuguese window.
Spanish
Set when you use the Spanish window.
German
Set when you use the German window.
Turkish
Set when you use the Turkish window.
Swedish
Set when you use the Swedish window.
French
Set when you use the French window.
DICOM character set
Note that DICOM character set is defined separately from system language.
DRT-MAN-055 (Rev (Rev.. 06)
94
Items
DICOM character set
Korean
ISO IR 6\ISO 2022 IR 149
Chinese
ISO IR 6\GB18030
Japanese
ISO 2022 IR 13\ISO 2022 IR 87
Default
ISO IR 100
Turkish
ISO IR 148
Thai Ukrainian, Russian UTF-8
ISO 2022 IR 166 ISO IR 6\ISO IR 144 ISO IR 192
DRT-MAN-055 (Rev (Rev.. 06)
95
7.2.2. Date
Figure 56. Date Configuration Screen
Configures the date notation.
Date
Delimiter
Y: Year , M: Month, D: Day Configures the symbol between Y, M, and D.
Example) If today is May 31, 2012 Date
Delimiter
Final Notation
MM DD YYYY
/
5/31/2012
MM DD YYYY
-
5-31-2012
MM DD YYYY
.
5.31.2012
YYYY MM DD
/
2012/5/31
MM DD YY
-
5-31-12
YY MM DD
.
12.5.31
DRT-MAN-055 (Rev (Rev.. 06)
96
7.2.3. Institution
Figure 57. Institution Configuration Screen
Input the hospital/institution information using EConsole1.
Hospital
Hospital Name
Department
Department name within the hospital.
DRT-MAN-055 (Rev (Rev.. 06)
97
7.2.4. Options
Figure 58. Options Configuration Screen
Auto Start
If
checked,
EConsole1
will
automatically
start
when
Windows start.
Virtual Keyboard
If checked, virtual keyboard will be available for use.
Auto Update
Requires internet connection.
If checked, EConsole1 will update automatically when a new version is released.
DRT-MAN-055 (Rev (Rev.. 06)
98
7.3. Image
Allows configuration relevant to examed images.
Subcategories
Description
Image Folder
Sets the location to save the images.
Delete Option
Sets delete options for image files.
Overlay
Sets overlay on the images.
Import
Imports images to the Studylist.
DRT-MAN-055 (Rev (Rev.. 06)
99
7.3.1. Image Folder
Figure 59. Image Folder Configurati Configuration on Screen
Sets the location to save the images within PC Storage.
Edit Icon
Opens the folder selection window.
Select the folder where the image is to be stored.
DRT-MAN-055 (Rev (Rev.. 06)
100
7.3.2. Delete Option
Figure 60. Delete Option Configuration Screen
① Image Delete Option
Use Password If checked, deletion of image requires a password.
Use Reject Code
If checked, a reject code (reason for deletion) is required every time for an image file is deleted.
② Reject Code
List of reasons for deletion. Statistics Displays the statistical record of each applied Reject Code.
DRT-MAN-055 (Rev (Rev.. 06)
101
The statistics can be exported to the Microsoft Excel file if the Export Icon is clicked within the statistic display window.
Delete
Add
Deletes the selected Reject Code from the list. Adds a new Reject Code.
Edit
Edits the selected Reject Code.
DRT-MAN-055 (Rev (Rev.. 06)
102
7.3.3. Overlay
Allows overlay configurati configuration. on.
Figure 61. Overlay Configuration
Allows editing of types of overlay to be displayed on the Left Top, Right Top, Left Bottom, and Right Bottom.
Editing method is the same while the display location differs.
Figure 62. Overlay Editing
DRT-MAN-055 (Rev (Rev.. 06)
103
① Registered Overlay Type & Font Size
Overlay: Information which will be displaye displayed. d.
Size: Font size of the pertinent per tinent information.
Information is displayed in accordance with the arrangement on the list.
② Edit Icon
Adds a new overlay to the list.
The following window as in [Figure 63] 63] appears when the icon is clicked.
Figure 63. Overlay Option Window
The selected overlay is added after entering the size and clicking ‘OK OK’’.
DRT-MAN-055 (Rev (Rev.. 06)
104
Deletes the selected annotation from the list.
The selected annotation is moved up one row.
The selected annotation is moved down one row.
Change font size of each location.
DRT-MAN-055 (Rev (Rev.. 06)
105
7.3.4. Import
Figure 64. Import Configuratio Configuration n Screen
Imports an external image into the Studylist.
Import Icon
Opens the file selection window.
Select the image to be imported.
DRT-MAN-055 (Rev (Rev.. 06)
106
7.4. Storage
Configurations concerning the deletion and storage of PC where the image will be saved.
Details
Description Configures the deletion and automatic
Delete
deletion function of images stored on Local Storage.
Backup
Compresses and backs-up the images stored on Local Storage.
DRT-MAN-055 (Rev (Rev.. 06)
107
7.4.1. Delete
Configures the deletion and automatic deletion function of images stored on Local Storage.
Figure 65. Delete Configuration Screen
① Storage Delete
The ‘Delete Delete’’ Icon deletes all images stored on the Local Storage.
Deleted images may not be restored, so please perform a backup prior to deleting images!
DRT-MAN-055 (Rev (Rev.. 06)
108
② Auto Delete
Auto Delete
If checked, all images satisfying the specified conditions are automatically deleted from the PC Storage when the user run a Econsole1.
Disk Size
If checked, images on the PC Storage are automatically deleted once the disk space until remain storage is "Disk Size".
Configurable ‘Disk Size’ Size’ is meaning expected remain space on PC storage.
If remain storage is less than ‘Disk Size’ Size’, overflowed images will be automatically deleted. (Whole disk volume except ‘Disk Size’ Size’ is allocat allocated ed for Econsole1.)
Max: Refers to the maximum limit of disk space utilized for storage of images on the PC Storage.
Period
If checked, images on the PC Storage are continuously deleted automatically automatically during the set period.
ex1) Set to "2", "Month" : Delete all images 2 month earlier images based on the Today.
ex2) Set to "1" "Day" : Delete all images 1 day earlier images based on the Today.
The
cycle
may
be
set
by
one
of
the
following;
Day/Week/Month/Year. One of the two methods, Disk Size and Period, may be selected at a time.
DRT-MAN-055 (Rev (Rev.. 06)
109
7.4.2. Backup
Compresses and backs-up the images stored on Local Storage.
Figure 66. Backup Configura Configuration tion Screen
① Database Backup Conf Configuration iguration
Displays the location of the folder used in Data Backup.
Edit
Opens the selection window for the folder to be used in backup.
Compress
Compresses then stores the currently used data based in the selected folder.
Restore
DRT-MAN-055 (Rev (Rev.. 06)
110
File selection window opens.
Restores the selected database file and stores it in the system.
Backup
Backs-up the currently used database in the selected folder.
② Automatic Backup Setting
Auto Backup Check
If checked, automatic backup is performed during the set period below.
Period
Sets the automatic backup period.
Available units consists of Day/Week/Month.
DRT-MAN-055 (Rev (Rev.. 06)
111
7.5. Account
Configures the account of the EConsole1 user.
Figure 67. User Account Configuration Configuration
The currently registered user information is available on the list.
Configures the user ID, Password, and authorization.
Delete Icon
Add Icon
Deletes the selected user account. Adds a new user account.
Edit Icon
Edits the existing user account.
DRT-MAN-055 (Rev (Rev.. 06)
112
7.6. Block List
Filtering the unnecessary RIS code such as ultrasound or other devices.
Figure 68. Block List Configuration Configuration
The currently registered Block List information is available on the list.
Delete Icon
Add Icon
Deletes the selected Block List entry. Adds a new entry to the Block List.
Edit Icon
Edits the existing Block List entry.
DRT-MAN-055 (Rev (Rev.. 06)
113
7.7. Dataset
Configures the DICOM Header Format.
Figure 69. Dataset Configuration
① DICOM Header Header Item Configuration
Deletes the unnecessary entries defined in the existing DICOM Header Format (Full List).
CR and DR are separately applied.
Delete
Deletes the selected entry.
Reset
Initializes the entry list back to Full List.
②
Image format
Configures the Image format information.
DRT-MAN-055 (Rev (Rev.. 06)
114
7.8. Network
Configures network information information such as IP addresses.
Subcategories
Description Requires network connection.
Local
Configures the network information of the PC (or Work Station) currently operating EConsole1. Requires network connection.
MWL
Configures the network information of the server storing the Modality Worklist.
Storage
Requires network connection. Configures the network information of the Storage Server, which stores the image files. Requires printer connection.
Printer
Configures the connection information of the printer.
DRT-MAN-055 (Rev (Rev.. 06)
115
7.8.1. Local
Configures the network information of the PC (or Work Station) currently operating EConsole1.
Figure 70. Local Configuration Screen
AE Title
Enter the name of the local PC (or Work Station) which will be displayed on the Network.
IP
Enter the IPv4 Address of the PC (or Work Station) currently operating EConsole1.
Port
Enter the Port number.
It should be configured in accordance with the server ser ver being linked.
Entered information will be automatically saved and applied.
DRT-MAN-055 (Rev (Rev.. 06)
116
7.8.2. MWL
Modality Worklist Server.
Sets the server retrieving the patient list.
Figure 71. MWL Configuration Screen
① Information of of server being accessed accessed
Manages the information of server being accessed.
Comprised of AE Title(Access Server Name), IP, and Port information.
Test
Verifies the connection with the selected server.
Delete Deletes the selected server from the list.
DRT-MAN-055 (Rev (Rev.. 06)
117
Add
Adds an access server.
Edit
Edits the selected server informatio information. n.
② Option
Time-out
Response standby time with the server.
It is regarded as connection failure if the server is unresponsive unresponsiv e for the set period of time.
③ Query
Configures the automatic refresh setting of MWL.
Auto
Manual
MWL is automatically refreshed during the entered interval. MWL refresh must be performed manually.
Interval
Sets the interval for automati automaticc refresh.
④ Modality
Select the modality type of the imaging equipment.
Add
Example) CR, DR Adds Modality.
Delete
Deletes the selected Modality.
DRT-MAN-055 (Rev (Rev.. 06)
118
7.8.3. PACS
Configures the Storage Server (PACS).
Figure 72. Storage Configuration
① Information of Storage Server to be Accessed
Manages the information of access storage server.
Comprised of AE Title (Access server name), IP, IP, Port
and
Selection informatio information. n.
If selection item is 'O', DICOM image send to it and the user can select more than 1 PACS.
Test
Verifies the connection with the selected server.
Delete Deletes the selected server from the list.
DRT-MAN-055 (Rev (Rev.. 06)
119
Add
Adds an access server.
Edit
Edits the selected server informatio information. n.
② Option
Time-out
Response standby time with the server.
It is regarded as connection failure if the server is unresponsive unresponsiv e for the set period of time.
③ Query
Time-out
Response standby time with the server.
It is regarded as connection failure if the server is unresponsive for the set period of time.
DRT-MAN-055 (Rev (Rev.. 06)
120
7.8.4. Printer
Figure 73. Printer Configuration Screen
① Printer Information & Printer Printer Selection for Connection Connection
Manages the information of the printer to be connected. Selects the printer from the list.
Add
Edit
Edits the selected printer information.
Delete
Adds a printer.
Deletes the selected printer from the list.
Test
Verifies the connection with the selected printer.
DRT-MAN-055 (Rev (Rev.. 06)
121
② Option
Restricts the response time by the Printer.
③ Medium Type
Selects the type of film.
May select one between Clear Film and Blue Film.
④ Border
Sets the color of the border line.
May select one between Black and White.
⑤ Smoothing
Applies Smoothing.
May select one between Medium, Sharp and Smooth.
⑥ Magnification
Sets the Magnification Magnification..
⑦ Density
Adjusts the Gray-Level Scale.
⑧ Printer DPI
Selects the resolution of the printer.
⑨ Overlay information
Select the additional annotation information added on the film.
DRT-MAN-055 (Rev (Rev.. 06)
122
The following pop-up window in [Figure 74] 74] appears appears when Edit is clicked.
Figure 74. Printer Overlay Information Edit
Instructions for use are identical with the overlay configuration of the image. Please refer to to 7.3.3 Overlay (p. (p. 103) 103) for for detailed informatio information. n.
DRT-MAN-055 (Rev (Rev.. 06)
123
7.9. Detector
Performs detector related configuration configurations. s.
Performs Calibration Tool and Terminal Console programs. Subcategory Description Registers and edits the model of panel General
(detector) (detect or) being used. Executes detector software, ECali1 (or Caliview) and Terminal Emulator. Detailed configuration screen concerning Panel A, regist registered ered in
Panel A
‘General General’’. Detailed configuration items for panel are created in accordance with the number of panels registered in ‘General General’’.
DRT-MAN-055 (Rev (Rev.. 06)
124
7.9.1. General
Registers and edits the panel (detector) which will be used.
Executes detector software, ECali1 (or Caliview) and Terminal Emulator.
Figure 75. General Configuration Screen
① Panel(Detector) Panel(Detector) Model Registration Registration and Editing
Registers the panel which will be connected.
Add Icon
Add icon registers the detector selected from Model Select.
Edit Icon
Edit icon replaces the currently registered detector with the detector selected from the Model Select.
Delete
The most recently registered detector is deleted.
DRT-MAN-055 (Rev (Rev.. 06)
125
② ECali1
Run Icon executes the ECali1 (or Caliview) program.
③ Terminal
Run Icon executes the Terminal Emulator.
Please refer to Appendix B. Terminal Emulator (p. 150 150)) for detailed informatio information. n.
DRT-MAN-055 (Rev (Rev.. 06)
126
7.9.2. Detailed Configuration by Panel
Performs detailed configuration for each panel(detector).
A menu such as ‘Panel A’ A’ and ‘Panel B’ B’ are generated in accordance with the number of panels registered in ‘General General’’.
Figure 76. Detector Detailed Configuration Screen
① Receptor
Sets which Receptor type the detector will utilize.
Stand, Table and Wireless mode are available.
Only wireless detectors may utilize the Wireless mode.
② Model
Model types of the detector. Automatically set by the value registered in ‘General General’’.
DRT-MAN-055 (Rev (Rev.. 06)
127
③ Serial Number
Enter the serial number of the Detector.
④ Grid Configuration
Grid Type: Select type of grid.
Line/Inch: Enter the line/inch value of the grid.
Angle: Enter the angle value of the grid.
DRT-MAN-055 (Rev (Rev.. 06)
128
7.10. RIS Code
RIS (Radiology Information System) Code is a collection of
projections. For example, it configures contents such as “A, B, C regions are to be imaged for RIS Codes for certain cancers.” cancers.”
It allows the rearrangement of sequence of projections.
Figure 77. RIS Code Configurati Configuration on
Import Excel Icon
Adds the RIS Code value from an Excel fine.
(Caution) Duplicate code values are overwritten with contents of the Excel file. (Overwrites the existing data).
DRT-MAN-055 (Rev (Rev.. 06)
129
Delete Icon
Deletes the selected RIS Code.
Deletion of information part of the Code, such as the projection, are executed by means of the ‘Edit Edit’’ icon.
Edit Icon
Manages the information, such as detailed projection information and imaging environment, of the RIS Code.
Please refer to 7.10.1. 7.10.1. RIS Code information.
(p. 131) 131) for detailed
DRT-MAN-055 (Rev (Rev.. 06)
130
7.10.1.
RIS Code Edit Window
Figure 78. RIS Code Edit Window
① RIS Code List Management Management
Manages the code and provides description of RIS Codes. Text Box
Are to write the Code and Description when adding or editing the RIS Code.
3 task icons are available. The descriptions are as noted in [Table 12]. 12].
DRT-MAN-055 (Rev (Rev.. 06)
131
Table 12. RIS Code List Management Icon Icon
Description Edits the code and description of the RIS Code selected from the list with the information entered in the text box below the icon. [Editing Sequence]
Edit
1) Selects the code to be edited from the list. 2) Enters in new information in the textbox below the icon. 3) Clicks the Edit icon. Adds the RIS Code with the information entered into the textbox below the icon. [Adding Sequence]
Add
1) Enters in new information in the textbox below the icon. 3) Clicks the Add icon. Deletes the RIS Code selected from the list. [Deleting Sequence]
Delete
1) Selects the code to be deleted from the list. 2) Clicks the Delete icon.
② Editing of projections projections included in the RIS RIS Code
Manages the projections to be included in the selected RIS Code.
Descriptions concerning the task icons are as noted in [Table 13].. 13]
DRT-MAN-055 (Rev (Rev.. 06)
132
Table 13. RIS Code Projection Management Icon Icon
Description Both icons perform identical tasks.
, Add
Adds the projections relevant to the selected Projection/Position in the RIS Code. Deletes the Projection category category from the RIS Code.
Delete Moves the selected Projection category up one row.
Move up
Moves the selected Projection category down one Move down
row.
③ Imaging Environment Environment Configur Configuration ation
Enter the automatically configured environment value when imaging of the pertinent RIS Code is performed.
Patient Size
Selects the patient size.
4 levels of selections are available.
AEC Sets the AEC.
Grid
Selects the Grid Type being used.
May select one of None, Fixed and Moving.
Focal Spot
Sets the focus.
May select from Large/Small.
Density
SID
Sets the density.
DRT-MAN-055 (Rev (Rev.. 06)
133
kVp, mA, mAs(mSec)
Sets the distance between the Detector and X-ray Tube. Sets the Generator condition.
Parameter
Locates the parameter file of the image processing which will be applied after the imaging.
Rotation
Flip H, Flip V
Selects whether to flip the image horizontally or vertically.
Shutter
Sets the rotation of the image.
Selects the ROI Size.
Marker
Sets and manages the marker, which will be automatically inserted in the image.
Descriptions regarding the Marker Icon is as noted in [Table 14].. 14] Table 14. Marker Icon
Icon
Description Adds a new Marker. Please refer to to 7.10.2 Marker Add Window (p.
Add
135) for 135) for detailed informat information. ion.
Deletes the selected marker. Delete
Moves the selected Marker up. Move up
Moves the selected Marker down. Move down
DRT-MAN-055 (Rev (Rev.. 06)
134
7.10.2.
Marker Add Window
Figure 79. Marker Add Window
Location
Select the location the marker will appear on the image
Text
Select the type of Marker
Random texts may be used as a marker, aside from the text which are provided by default
User registered markers may be selected from the list
The Marker list window may be edited using the Add Icon
DRT-MAN-055 (Rev (Rev.. 06)
135
and Delete Icon
If user want to change Maker size, Edit in 'Projection' - 'Marker Size'. Size'.
DRT-MAN-055 (Rev (Rev.. 06)
136
7.11. Projection Editor
Manages the Projection type
Figure 80. Projection Editor Configuration
Marker Size
Change marker size.
Rotation
Set rotation selected receptor's all projection & position.
Rotation's adapted order is each position's 'Flip Vertical', 'Flip Horizontal' is first and 'Rotation' is second.
Delete
Deletes the selected Projection or Position Recent selections which will be deleted by clicking the
DRT-MAN-055 (Rev (Rev.. 06)
137
‘Delete Delete’’ icon are highlighted in yellow
Edit Icon
Edits the Projection and Position after opening the Project Editor window.
Please refer to 7.11.1 Projection Editor Window (p. 139) 139) for detailed informati information. on.
DRT-MAN-055 (Rev (Rev.. 06)
138
7.11.1.
Projection Editor Window
Figure 81. Projection Editor Window
① Anatomy
Must be selected first
Projection
and
position
will
not
appear
unless
an
anatomical region is selected from the Anatomy Anatomy
Select the projection
If the Receptor Type is different, even if the anatomical regions are the same, it will be saved as a different setting
Please be sure to set the anatomical region, as well as the Receptor type.
DRT-MAN-055 (Rev (Rev.. 06)
139
② Projection
Figure 82. Projection Configuration
Manages the Projection list
Please refer to [Table 15] 15] for for instructions of each icon
Table 15. Projection/Position List Management Icon Icon
Description Edits the name of the selected category from the list with the entry in the left lef t textbo textboxx [Edit Sequence]
Edit
1) Select the category to be edited from the list 2) Enter the new information in the textbox on the left of the icon 3) Click the Edit icon Adds the category with the information entered into the textbox on the left of the icon [Add Sequence]
Add
1) Enter the new information in the textbox on the left of the icon
2) Click the Add icon
DRT-MAN-055 (Rev (Rev.. 06)
140
Deletes the selected category from the list [Delete Sequence] 1) Select the category to be deleted from the list Delete
2) Click the Delete icon Moves the selected projection category category up one row in the list
Move up
Moves the selected projection category category down one row in the list
Move down
③ Position
Figure 83. Position Configuratio Configuration n
Manages the Projection list Please refer to [Table 15] 15] for for instructions concerning each icon
DRT-MAN-055 (Rev (Rev.. 06)
141
④ Imaging Environment
Sets the appropriate imaging environment for the selected projection
Detailed information concerning the Imaging Environment Configuration are identical to the Imaging Environment Configuration of the RIS Code Edit Window
Please refer to ‘③ ‘③ Imaging Environment Configuration’ Configuration’ of 7.10.1. RIS Code (p. (p. 131) 131)
DRT-MAN-055 (Rev (Rev.. 06)
142
7.12. Generator
Set the type of X-ray Generator
Set detailed options for each type of X-ray Generator Table 16. Generator Configuration Categories
Subcategories
Details
Sedecal
Sedecal Generator
DRGEM
DRGEM Generator
POSKOM
POSKOM Generator
DRT-MAN-055 (Rev (Rev.. 06)
143
7.12.1.
Sedecal
Setting regarding the Sedecal Generator
Figure 84. Sedecal Configuratio Configuration n Screen
‘Use This Generator’ Generator ’ check box
Check this category in order to use this Generator
Enter the configuration value for the use of Sedecal Generator
DRT-MAN-055 (Rev (Rev.. 06)
144
7.12.2.
DRGEM
For use of the DRGEM Generator
Figure 85. DRGEM Configuration Screen
‘Use This Generator’ Generator ’ check box
Check this category in order to use this Generator
Enter the configuration value for the use of DRGEM Generator
DRT-MAN-055 (Rev (Rev.. 06)
145
7.12.3.
POSKOM
For use of the POSKOM Generator
Figure 86. POSKOM Configurati Configuration on Screen
‘Use This Generator’ Generator ’ check box
Check this category in order to use this Generator
Enter the configuration value for the use of POSKOM Generator
DRT-MAN-055 (Rev (Rev.. 06)
146
Appendix A. Glossary Terms
Description Picture Archiving and Communication System.
PACS
System for the storage, interpreting and access of medical imaging information
DICOM
Digital Imaging and Communication in Medicine. Standard medical digital image and communication DICOM Directory.
DICOMDIR
DICOM directory information file stored in the root folder of the media (CD/DVD) Hospital Information System.
HIS
Hospital information system, registration/administrative system, including OCS, PACS and etc.
OCS RIS
Order Communication System Radiology Information System. Operated as part of HIS or independently Modality Worklist.
MWL
DICOM standards for scheduled examination information sharing Computer Unified Device Architecture. Architecture.
CUDA
High sped parallel processing architecture utilizing a GPU developed by NVIDIA
GPU
Graphic Processing Unit. Graphical User Interface.
GUI
Computer interface for the user, developed by graphic elements
Detector
Flat Panel Digital X-ray Detector Detector..
Compiler
Program which converts source codes in high level language in to specific machine codes
DRT-MAN-055 (Rev (Rev.. 06)
147
Terms Build
API
Description Process where the source code is being converted for computer operation Application Programming Interface. Interface developed for use of specific functions in the application program
Installer Login Coding Convention Code Review Code Coverage Static Code Analysis
Program utilized for the installation, maintenance and removal of a specific application program Process authenticating the user for permission to access the system Style guideline for creating a source code Process where developers inspect and verify the created/edited created/edit ed source code Tested number of code components/Total number of code components * 100 Process of discovering potential defects of the program without operating it by analyzing the source code
Cyclomatic
Digitized level of complexity of the program’ program’s source
Complexity
code Continuous Integration.
CI
Automated integration process of the source code, and the process of improvement and receiving feedback through periodic build, test analysis
Encryption Protection (AntiCrack) SCP
Encrypting the data Protection preventing program hacking through Reverse Engineering Service Contents Provider Text and various shapes marked on the completed
Annotation
image
DRT-MAN-055 (Rev (Rev.. 06)
148
Terms LUT ROI Exception Calibration Map File Pixel Map
Description Lookup Table. Table of previously calculated values Region of Interest Actual region necessary from the completed image Exception subsequent to program error, wrong memory reference or stack overflow Task calibrating the quality of the detector image Generic term for data files used for the calibration of detector image quality Data file used to calibrate bad pixels in detectors Data file used to calibrate the property of the detector
Gain Map Panel Map
subsequent to installation environment Data file used to calibrate the property of the detector panel itself
DRT-MAN-055 (Rev (Rev.. 06)
149
Appendix B. Terminal Emulator Usage of terminal emulator may differ from each detector, please refer to the Detector Service Manual.
Terminal Mode Terminal Emulator Start-up Sequence: 1. Turn the detector power OFF.
Not necessary for F600.
2. Click ‘Run Run’’ button on Configuration - Detector - General tab.
3. Turn the detector power ON.
4. When power is turned on, Terminal Emulator tries to connect to
DRT-MAN-055 (Rev (Rev.. 06)
150
detector automatically.
If not, please click ‘Connect Connect’’ button.
As the Ethernet Terminal Emulator features auto-connect function, it will enter Terminal Mode automatically.
The TERMINAL MODE MENU has 3 options :
USER MODE
Can be accessed by the user.
OPERATOR OPERAT OR MODE
Can be accessed only by the operator.
EXIT TERMINAL MODE
Closes terminal mode.
DRT-MAN-055 (Rev (Rev.. 06)
151
User Mode To activate User Mode, enter “1” in the Terminal Mode menu screen. screen.
Cautions: Number keys on numeric keypad in right-side don’ don ’t work well. Please use number keys in left-side.
DRT-MAN-055 (Rev (Rev.. 06)
152
User Mode has 8 options as below
Check the status of the internal module 1. EXECUTES SELF-TEST in the detector 2. PROGRAM READY DELAY
Check or change the Ready Delay value.
3. PROGRAM EXPOSURE
Check or change the Exposure Window
WINDOW TIME
Time value.
4. SELECT EXPOSURE
Activate/deactivate the Exposure
PROTECTION MODE
Protection Protectio n Mode.
5. SELECT ADAPTIVE ERASE
Activate/deactivate the Adaptive Erase
MODE
Mode.
P. PR PREVIOUS EVIOUS MENU
Return to the previous menu.
?. PRINT MENU
Display all the options for User Mode.
X. EXIT TERMINAL MODE
Exit terminal mode.
DRT-MAN-055 (Rev (Rev.. 06)
153
1. EXECUTES SELF-TEST To run the SELF-TEST , press” press”1 1” in User Mode. If SELF-TEST is successfully completed, all the test results become changed to “OK! OK!””.
If your detector has a problem with the operating temperature, you will get a “Failed” “Failed” answer for the current operating temperature. In this case, you are advised to shut down the power to the detector and allow a cooling time at least more than one hour. If you still get the same error message after one hour of cooling, then report your claim to the DRTECH Customer Support Team
[email protected]). ). (
[email protected]
If your detector has a problem with the "Erase Light" operation, you will get a “Failed” answer. In this case, report your claim using our claim form, CDN(Customer Dissatisfaction Notice / refer to the last page in this Service Manual) to
[email protected] [email protected] immediately.
Your detector should be returned to DRTECH for repair.
DRT-MAN-055 (Rev (Rev.. 06)
154
If your detector has a problem with Read/Write Memory operation, you will get a “Failed” “Failed” answer. In this case, report your claim using CDN to
[email protected] immediately.
Your detector should be returned to DRTECH for repair. If your detector has a problem with the "Data Readout Sequence" operation, you will get a “Failed” answer. In this case, report your claim using CDN to
[email protected] [email protected] immediately. Your detector should be returned to DRTECH for repair.
In all other cases, contact
[email protected] [email protected] for for technical support.
DRT-MAN-055 (Rev (Rev.. 06)
155
2. PROGRAM READY DELAY To change the Ready Delay value, press “2” in User Mode. Mode. The current Ready Delay value and a message for setting a new Ready Delay value are displayed.
You can enter between 0 ~ 3000ms, the default value is 2000ms.
DRT-MAN-055 (Rev (Rev.. 06)
156
To change the Ready Delay value, enter the new value you want and new Ready Delay value will be displayed. To keep existing setting just, press “Enter.” “Enter.”
To return to the previous menu, press “P”.
To see User menu, press” press”?”.
DRT-MAN-055 (Rev (Rev.. 06)
157
3. PROGRAM EXPOSURE WINDOW TIME To change the Exposure Window Time value, press “3 “3” in User Mode. Mode. The current Exposure Window Time value and a message for setting a new Exposure Window Time value are displayed.
You can enter between 0 ~ 3000ms and default value is 500ms.
DRT-MAN-055 (Rev (Rev.. 06)
158
To change the Exposure Window Time value, enter the new value you want and new Exposure Window Time value is displayed. To keep the existing setting, press just “Enter.” “Enter.”
DRT-MAN-055 (Rev (Rev.. 06)
159
4. SELECT EXPOSURE E XPOSURE PRO PROTECTION TECTION MODE To change the Exposure Protection Mode setting, press “4” in User Mode. The current state of Exposure Protection Mode and a message for setting a new value are displayed.
DRT-MAN-055 (Rev (Rev.. 06)
160
To make the Exposure Protection Mode enable, press “1,” or disable, press “2.” “2.” To keep the existing setting, press just “Enter.” “Enter.” The mode is changed successfully, and the current state will appear.
DRT-MAN-055 (Rev (Rev.. 06)
161
5. SELECT ADAPTIVE ERASE MODE To change the Adaptive Erase Mode setting, press “5” in User Mode. Mode. The current state of Adaptive Erase Mode and a message for setting a new value are displayed.
DRT-MAN-055 (Rev (Rev.. 06)
162
To make the Adaptive Erase Mode enable, press “1,” or disable, press “2.” “2.” To keep the existing setting, press just “Enter.” “Enter.” The mode is changed successfully, and the current state will appear.
menu, press “P”. “P”.
DRT-MAN-055 (Rev (Rev.. 06)
163
X. End of Terminal Mode To exit the Terminal Mode, press “X” in the Terminal Mode menu or User Mode. After pressing “X,” “X,” To confirm the exit, press “Y.” “Y.” To cancel the exit, press “N” to return to the previous pre vious menu.
DRT-MAN-055 (Rev (Rev.. 06)
164
When pressing “Y” to confirm the termination of Terminal mode, the Self-Test function of the detector will be executed automatically and the results will be displayed.
Please check the detector operates normally.
Operator Mode Operator mode is used for the detailed setting at the time of producing detector for the first time, and can be accessed by only an authorized operator.
DRT-MAN-055 (Rev (Rev.. 06)
165
Appendix C. Update If current using version has bugs please contact to CS team, CS team will give you a update(patch) installer. If you receive the update installer file please run the provided setup file.
Please back up original Econsole1 folder(ex: C:\Econsole1) in order to prevent a fatal error.
If the setup file is run, you will be present presented ed with the following screen. Click ‘Next Next’’ to start the installation installation..
DRT-MAN-055 (Rev (Rev.. 06)
166
Select license protection type you purchased.
Hardware License Protection (Dongle Key)
Provides special USB, and you can run EConsole1 only when the USB connected.
Software License Protection
Provides the license on the special PC.
DRT-MAN-055 (Rev (Rev.. 06)
167
Do not change installation folder and click 'Next'. It will update the program automatically and installation is finished please run the Econsole1 and check the version.
DRT-MAN-055 (Rev (Rev.. 06)
168
Appendix D. Portable function Some of wireless detector is a function to save a x-ray image. If user shot a x-ray without software the image will be saved in the detector . The saved image control function is called "Portable function".
Figure 87. Portable function summary
Econsole1 Vet's portable has 2 functions.
Download the image saved in detector
Register the exam order to detector
1) Download the image saved in detector. Proceed in the followin following g order ① Select a exam order and go to acquisition.
DRT-MAN-055 (Rev (Rev.. 06)
169
② Click
icon in the ToolBar
If it is not in the toolbar, click the 'Edit' and add it.
③ Save portable image pop-up comes up. Select a image in the Detector.
Click 'OK'.
Detailed function 'Save portable image' pop-up window is described in the below.
Figure 88. Save portable image ① Select a image saved detector. ② Show selected detector's saved image's file name. ③ Selected image's preview image. ④ Function button and UI.
Refresh
Refresh a image list
Delete
DRT-MAN-055 (Rev (Rev.. 06)
170
Delete all
Delete selected image in detector. Delete all images selected detector.
Used(%d) Selected detector's used storage. OK
Get selected image and show it to Econsole1 Vet's image viewer viewer..
Cancel
Close 'Save portable image' window.
2) Register the exam order to detector It is the function that save exam orders to detector and shot a image without PC depending on the given exam list. Save exam order function description is below.
① Click
button.
② 'Save image list' window comes up.
Select a detector. Select List in right side and click '