Document No.DRT-MAN-055 Revision No.Rev.05 Released By 2015-11-23

Digital Radiography Technologies

 

 

About This Manual The EConsole1 User Manual has prepared for the appropriate use of digital radiography operating console software developed by DRTECH Inc., This manual provides the instructions for the use of EConsole1. The manual contains the following contents.  

Operation Environment, Environment, Installation and Method for EConsole1

 

Functions and Instructions for EConsole1

Please contact us if there are any problems during the use of the product. WebSite

http://www.drtech.co.kr

Customer Suppor Supportt Team

Email Email  [email protected] [email protected]   Tel +82 31 730 6805 Fax +82 31 730 6899

To Enhance The Quality Of Human Life! 

DRT-MAN-055 (Rev. 06)

 

  Copyright - EConsole1 Manual Information contained in this manual is confidential and is the property of DRTECH Inc., which is provided exclusively for authorized licenses of DRTECH Inc. The purpose of the manual is to assist the users to comprehensively utilize the product of DRTECH Inc. The information contained in this manual may not be altered or distributed without prior written consent of DRTECH DRTECH Inc.,  Inc., and may not be disclosed to unauthorized personnel. Information contained in the manual may be altered without prior notice. Copyright © 2013 2013 DRTECH  DRTECH Corporation.  Corporation. All rights reserved.

DRT-MAN-055 (Rev. 06)

 

 

Table of Contents 1. INTRODUCTION ................. ................................. ................................. .................................. ................................. ............................ ............ 6  2. SYSTEM REQUIREMENTS ................. ................................. ................................. ................................. ............................... ............... 7  2.1. SUPPORTED DIGITAL X-RAY DETECTOR  ......................................... ......................................................... ....................... ...... 7  2.2. PC SPECIFICATION FOR THE SOFTWARE  .............................................. ............................................................. .................. 7  3. INSTALLATION .............. ............................... ................................. ................................. .................................. .................................. ................. 8  3.1. INSTALLATION PREREQUISITES .............. ............................... .................................. .................................. ............................ ........... 8  3.2. INSTALLATION ............... ................................ ................................. ................................. .................................. ................................. ................ 14  3.3. LANGUAGE & NUMBER OF DETECTOR SETTING ............................... ............................................... ................... ... 15  3.4. LICENSE PROTECTION TYPE SELECTION ................ ................................. ................................. .............................. .............. 17  3.5. OPTION ............... ................................. .................................. ................................. ................................. .................................. ......................... ....... 18  3.6. HOSPITAL & DEPARTMENT INFORMATION ................. ................................. ................................. ......................... ........ 19  3.7. INSTALLATION PATH CONFIGURATION .............. ............................... .................................. ................................ ............... 20  3.8. START MENU FOLDER CREATION ................ ................................. ................................. ................................. ...................... .....22  3.9. ADDITIONAL TASK SETTING ................ ................................ ................................. .................................. .............................. ............. 23  3.10. INSTALLATION COMPONENT VERIFICATION ................. ................................. ................................. ...................... .....24  3.11. USB H AND S WITCH DRIVER INSTALLATION ............... ............................... ................................. ...................... ..... 25   

 

............................... ................................. .................................. ............................. ............ 26 3.12. INSTALLATION COMPLETE ............... 3.13. M AP FILE & EC ALI1 CONFIGURATION ............... ................................ ................................. .............................. ..............27 

3.14. SETTINGS AFTER INSTALLATION ................ ................................. ................................. ................................. ..................... .... 30  ................................. ................................. ................................. ................................. ................ 31  3.15. UNINSTALL ECONSOLE1 ................ ................................. ................................. ................................. ................................. ................32  4. ECONSOLE1 OVERVIEW ................

4.1. PROGRAM START-UP ................ ................................. .................................. ................................. ................................. ...................... ..... 32  4.2. MAIN SCREEN OF THE PROGRAM ............... ............................... ................................. .................................. ...................... .....34   

 

.................................. .................................. ................................. ................................. ................................. .............................. ..............39 5. LIST .................

DRT-MAN-055 (Rev. 06)

 

  5.1. WORKLIST ................. .................................. .................................. ................................. ................................. .................................. ................... 40  5.2. STUDYLIST ................. .................................. .................................. ................................. ................................. .................................. ................... ..54  6. EXAM ................. .................................. ................................. ................................. .................................. .................................. ........................... .......... 74  ................................. .................................. .................................. ................................. ................................. ...................... ..... 77  6.1. TOOLBAR ................

6.2. PROJECTION LIST & IMAGE PROCESSING TAB ............................... ................................................ ..................... .... 84  6.3. TRUVIEW ART ............................... ............................................... ................................. .................................. ................................ ............... 89  7. CONFIGURATION ............... ............................... ................................. .................................. ................................. .......................... .......... 90  7.1. DISPLAY .............. ............................... .................................. .................................. ................................. .................................. ........................ ...... 92  7.2. SYSTEM ............... ................................. .................................. ................................. ................................. .................................. ......................... .......93  7.3. IMAGE ............... ............................... ................................. .................................. .................................. .................................. .......................... ......... 99  7.4. STORAGE ................ ................................. .................................. .................................. ................................. ................................. .................... ... 107  7.5. ACCOUNT ............... ................................ .................................. .................................. ................................. ................................. ..................... .... 112  7.6. BLOCK LIST ................ ................................. .................................. ................................. ................................. .................................. ................... 113  7.7. DATASET .............. ............................... .................................. .................................. ................................. .................................. ........................ ......114  7.8. NETWORK ................. .................................. .................................. ................................. ................................. .................................. ................... 115  7.9. DETECTOR ................................ ................................................. .................................. ................................. ................................. .................... ... 124  7.10. RIS CODE ................ ................................. .................................. ................................. ................................. .................................. ................. 129  7.11. PROJECTION EDITOR ............... ................................ .................................. ................................. ................................. ..................... ....137  7.12. GENERATOR ................ ................................. ................................. ................................. .................................. ............................... .............. 143  APPENDIX A. GLOSSARY ............... ................................ ................................. ................................. ................................. .................. 147  APPENDIX B. TERMINAL EMULATOR .............. ............................... .................................. ............................... .............. 150  APPENDIX C. UPDATE ................ ................................. .................................. ................................. ................................. .................... ... 166   APPENDIX  APPEN DIX D. D. PORTABLE PORTABLE FUNCTION FUNCTION............................... ................................................ ............................... .............. 169 

DRT-MAN-055 (Rev. 06)

 

 

1. Introduction EConsole1

is

digital

radiography

operating

console

software

specialized for the FLAATZ and EVS series digital X-ray detector developed by DRTECH Inc. EConsole1 provides an integrated solution for X-ray projection. It integrates

with the X-ray generator and the digital

detector.

Furthermore, EConsole1 acquires and processes images. In addition, it complies with DICOM standards and is able to transmit and receive data with the PACS system, and print images through the DICOM printer.

Figure 1. X-Ray Projection Solution Utilizing EConsole1 (Wired Detector)

Figure 2. X-Ray Projection Solution Utilizing EConsole1 (Wireless Detector)

DRT-MAN-055 (Rev (Rev.. 06)

6

 

 

2. System Requirements 2.1.  Supported Digital X-ray Detector EConsole1 supports the following Digital X-ray Detectors developed by DRTECH.  

FLAATZ 330N LAN

 

FLAATZ FLAAT Z 500 LAN

 

FLAATZ FLAAT Z 560 LAN

 

FLAATZ 600 Wireless LAN

 

FLAATZ 750E LAN

 

FLAATZ FLAAT Z 760 LAN

 

EVS 4343 LAN

2.2.  PC Specification for the Software Table 1. PC Specifications Items

Recommended

CPU

Intel i5-3470

RAM

8 GB

HDD

500 GB

VGA ODD OS

NVIDIA GeForce GT630 1GB DVD Recorder Windows 7 Pro 32 / 64bit

Display Size

23 inch

Display Resoluti Resolution on

1920 x 1080

* Optimal performance is not guaranteed for PCs that do not comply with

the recommended specifications. 

DRT-MAN-055 (Rev (Rev.. 06)

7

 

 

3. Installation 3.1.  Installation Prerequisites 3.1.1. Windows 7 Windows 7 users must set the User Account Control (herein after ‘UAC UAC’’) permission and disable the firewall.

3.1.1.1. 

UAC Setting

①  Cl Clic ickk tth he ssttart art ico icon n

.

②  Click ‘Control Panel’ Panel’. ③  Select ‘User Accounts’ Accounts’. ④  In the User Accounts window, click ‘User Accounts’ Accounts’. ⑤  In the User Accounts task window, click ‘Change User Account Control Setting’ Setting’. ⑥  Adjust the slider to the ‘Never notify,’ notify,’ and then click ‘OK OK’’.

3.1.1.2. 

Disabling the Firewall

①  Cl Clic ickk tth he S Sttart ico con n

.

②  Click ‘Control Panel’ Panel’. ③  Click ‘Security Security’’. ④  Click ‘Windows Firewall’ Firewall’. ⑤  Click ‘ Turns Windows Firewall on or off ’ on the left. ⑥  Click ‘Off (not recommended),‘ recommended),‘ and then click ‘OK OK’’.

3.1.1.3. 

Set Power Options

①  Click 'Setting' on the right hand menu. ②  Click 'Control Panel'. ③  Click 'Power Options' ④  Select

DRT-MAN-055 (Rev (Rev.. 06)

'High

performance'

or

'Energy

Manager

High

8

 

  Performance Mode'

3.1.1.4. 

Set Power Options

①  Click 'Setting' on the right hand menu. ②  Click 'Control Panel'. ③  Click 'Power Options' ④  Select 'Choose when to turn off the display' menu on the left side. ⑤  Set 'Put the computer to sleep' to 'Never' and click Save changes.

3.1.1.5.  Disable Windows update setting ①  Click 'Setting' on the right hand menu. ②  Click 'Control Panel'. ③  Click 'Windows Update' ④  Select 'Change settings' menu on the left side.   ⑤ Set 'Important updates' to 'Never check for updates(not recommended'

DRT-MAN-055 (Rev (Rev.. 06)

9

 

  ⑥  Uncheck 'Give me recommended updates the same way I receive important updates' ⑦  Uncheck ''Give me updates for other Microsoft products when I update Windows'

DRT-MAN-055 (Rev (Rev.. 06)

10

 

 

3.1.2. Windows 8 Windows 8 users must set the User Account Control (herein after ‘UAC UAC’’) permission, disable the firewall and run the program as the system administrator.

3.1.2.1. 

UAC Setting

①  Click ‘Setting Setting’’ on the right hand menu. ②  Click ‘Control Panel’ Panel’. ③  Select ‘User Accounts’ Accounts’. ④  In the User Accounts window, click ‘User Accounts’ Accounts’. ⑤  In the User Accounts task window, click ‘Change User Account Control Setting’ Setting’. ⑥  Adjust the slider to the ‘Never notify,’ notify,’ and then click ‘OK OK’’.

3.1.2.2. 

Disabling the Firewall

①  Click ‘Setting Setting’’ on the right hand menu. ②  Click ‘Control Panel’ Panel’. ③  Click ‘Security Security’’. ④  Click ‘Windows Firewall’ Firewall’. ⑤  Click ‘ Turns Windows Firewall on or off ’ on the left. ⑥  Click ‘Off (not recommended),‘ recommended),‘ and then click ‘OK OK’’.

3.1.2.3.   

Run as Administrator

Windows 8 users must run the Installation File  and  and Program  as  as the Administrator.

 

Install the program by right clicking (press and hold, if touch screen) and selecting ‘Run as Administrator’ Administrator ’.

 

After completing the installation, set the icon, so that it always runs as administrator. ①  Right click (press and hold, if touch screen) the shortcut icon. ②  Click ‘Properties Properties’’. ③  Click the ‘Advanced Advanced’’ icon in the ‘Shortcut Shortcut’’ tab  tab‘‘. ④  Check the ‘Run as administrator’ administrator ’ box, click on ‘OK OK’’.

DRT-MAN-055 (Rev (Rev.. 06)

11

 

  3.1.2.4.   

Do not change Compatibility mode

EConsole1.exe's right clicking(press and hold, if touch screen) and select 'Compatibility' tab.

 

Make sure that

EConsol EConsole1.exe's e1.exe's compatibility mode is not

checked.

3.1.2.5. 

Disabling the Windows update

①  Click ‘Setting Setting’’ on the right hand menu. ②  Click ‘Control Panel’ Panel’. ③  Select ‘Windows Update’ Update’. ④  In the Windows Update window, click ‘Change settings’ settings’. ⑤  In the Change setting window, Change to ‘Never check for updates(not updates(n ot recommended) recommended)’’. ⑥  Uncheck 'Give me recommended updates the same way I receive important updates.

Figure 3. Windows update setting window

3.1.2.6. 

Power options setting(Only use Laptop)

 

Click 'Setting' on the right hand menu.

 

Click 'Control Panel'.

 

Click 'Power Options'

DRT-MAN-055 (Rev (Rev.. 06)

12

 

 

Figure 4. Windows power options setting window  

Select

'High

performance'

or

'Energy

Manager

High

Performance Mode'

Powerr options setting Powe  

Click 'Setting' on the right hand menu.

 

Click 'Control Panel'.

 

Click 'Power Options'

 

Select 'Choose when to turn off the display' menu on the left side.

 

Set 'Put the computer to sleep' to 'Never' and click Save changes.

3.1.2.7.  Disable SmartScreen ①  Click 'Setting' on the right hand menu. ②  Click 'Action Center' ③  Click 'Change Windows SmartScreen setting' ④  Set 'Don't do anything(turn off Windows SmartScreen)

DRT-MAN-055 (Rev (Rev.. 06)

13

 

  ⑤  Click 'OK'

3.2.  Installation Run the provided setup file.

 

A minimum of 10GB of available HDD space is required for the proper installation of EConsole1!

 

0.5.6 is the version of the product, and may change.

 

Windows 8 users must run the file as ‘administrator ’.  

Right-click the file, select ‘Run as administrator’ administrator’ in  in the menu.

If the setup file is run, you will be presented with the following screen as in, [Figure 5] 5].. Click ‘Next Next’’ to start the installation installation..

Figure 5. Setup Wizard Window

DRT-MAN-055 (Rev (Rev.. 06)

14

 

 

3.3.  Language & Number of Detector Setting Select a language to be used in the Econsole1 vet program. Click ‘Next Next’’ after selection.

Figure 6. Language & Number of Detector Setting

① Language Setting Please select the appropriate language.  

It may be set after the program installation.  

It can be set through Configuration - System - Language.

 

Please refer to 7.2.1. 7.2.1.   Language (p. 94) 94)   for setting after installation.

DRT-MAN-055 (Rev (Rev.. 06)

15

 

  ② Number of Detector Setting Please select whether to use 1 or 2 detectors.  

Be Advised!  

For Dual Mode using 2 detectors, there are restrictions to models which are able to use the detectors simultaneously.

 

 

Dual Mode (2 detector use) supported detectors  

FLAATZ FLAAT Z 500 LAN

 

FLAATZ FLAAT Z 560 LAN

 

FLAATZ 600 Wireless LAN

 

FLAATZ 750E LAN

 

FLAATZ FLAAT Z 760 LAN

 

EVS 4343 LAN

 

EVS 3643 LAN

It may be set after program installation.  

It can be set through Configuration - Detector.

 

Please refer to 7.9 Detector (p. 124) 124)   for setting after installation.

DRT-MAN-055 (Rev (Rev.. 06)

16

 

 

3.4.  License Protection Type Selection

Figure 7. License Protection Type Selection Selects license protection type you purchased.  

Hardware License Protection (Dongle Key)  

Utilizes a license key embedded within a special USB device. This device must be attached to the PC in order for Econsole1 to operate. 

 

Software License Protection  

The license is provided as a serial number to be entered during installat installation. ion. 

DRT-MAN-055 (Rev (Rev.. 06)

17

 

 

3.5.  Option Select the model and input the serial number of the X-ray Detector to be used with the EConsole1. Click ‘Next Next’’ after selection.  

It may be set after program installation installation..  

It can be configured through Configuration - Detector. Please refer to  to 7.9 Detector (p. (p. 124)  124)..

Figure 8. Option: Detector Setting

①  X-ray Detector Detector Model Selection Selection  

Set the X-ray Detector, which will be used with FConolse1.

 

Select the appropriat appropriate e model.

②  Enter the X-ray Detector Detector Serial Numbe Numberr  

Enter the serial number of detector.

DRT-MAN-055 (Rev (Rev.. 06)

18

 

 

3.6.  Hospital & Department Information Enter the Hospital name and department name.

Figure 9. Hospital & Department Information

DRT-MAN-055 (Rev (Rev.. 06)

19

 

 

3.7.  Installation Path Configuration 3.7.1. Select EConsole1 Install Folder  

You may designate the folder in which to install the program through the ‘Browse Browse’’ icon.

 

Please verify that there is sufficient space on the HDD where the program will be installed!  

 

The program requires a minimum of 2GB of available space.

After specifying the location location,, click ‘Next Next’’.

Figure 10. EConsole1 Install Folder Selection

DRT-MAN-055 (Rev (Rev.. 06)

20

 

 

3.7.2. Select Image Save Folder  

You may designate the folder in which to install the program through the ‘Browse Browse’’ icon.



location,, click ‘Next Next’’.   After specifying the location If user want to change it is possible only 'drive:\Images' folder.

 

Figure 11. Image Save Folder Selection

DRT-MAN-055 (Rev (Rev.. 06)

21

 

 

3.8.  Start Menu Folder Creation Specify a start menu folder name for the program icons, or select not to create a start menu entry.   Click ‘Next Next’’.

Figure 12. Start Menu Folder Selection

DRT-MAN-055 (Rev (Rev.. 06)

22

 

 

3.9.  Additional Task Setting Select whether to create shortcut icons in the desktop and Quick Launch Bar. Click ‘Next’. ‘Next’.

Figure 13. Additional Task Setting

DRT-MAN-055 (Rev (Rev.. 06)

23

 

 

3.10.  Installation Component Verification Displays the summary of configured component componentss to be installed. Start Sta rt the installation by clicking the ‘Install Install’’ icon, and please wait while installing.

Figure 14. Installation Component Verificat Verification ion

DRT-MAN-055 (Rev (Rev.. 06)

24

 

 

3.11.  USB Hand Switch Driver Insta Installation llation Install USB Hand Switch Driver for F600, EVS4343 and EVS3643.

Figure 15. USB Hand Switch Driver Installation

DRT-MAN-055 (Rev (Rev.. 06)

25

 

 

3.12.  Installation Complete Once the installation has completed, it is strongly recommended that you allow the PC to be restarted.

Make sure “Y “Yes, es, restart the

computer now” is selected and click next.

Figure 16. Installation Complete

DRT-MAN-055 (Rev (Rev.. 06)

26

 

 

3.13.  Map File & ECali1 Configura Configuration tion There are three files that are provided with your detector panel that must be copied to the hard drive in order for the software to operate. These files are unique to your specific panel.  

3.13.1. 

Map File

DRTECH provide the MAP CD like below.

그림  17. MA MAP P file C CD D

The files can be found on the CD in the folder named MAP.   These files consist of the following:

 

 

Pixel MAP file: [Detector [Detector serial number].MAP

 

Gain MAP file: [Detector serial number].GMP

 

Panel MAP file: [Detector serial number].PMP

Please copy these files to [EConsole1 installation folder]\Map folder. Even you use 2 or 3 detector do not make special folder and copy it to [EConsole1 installation folder]\Map folder. EConsole1 search the Map files using set detector's serial number. DRT-MAN-055 (Rev (Rev.. 06)

27

 

  So, Map file name and the registered serial number must be matched.

그림  18. Serial nu number mber & ma map p files

DRT-MAN-055 (Rev (Rev.. 06)

28

 

 

3.13.2. 

ECali1 Configuration

You must setup various settings of detector using ECali1. For setting up using ECali1, please refer to the ECali1 manual.

Figure 19. ECali1

DRT-MAN-055 (Rev (Rev.. 06)

29

 

 

3.14.  Settings after Installation Windows 8 users must set EConsole1 so that it runs as administrator.  

Right click (press and hold) the EConsole1 shortcut icon in the desktop.

 

Click ‘Properties Properties’’.

 

Select ‘Advanced Advanced’’ in the ‘Shortcut Shortcut’’ tab.

 

Check the ‘Run as administrator’ administrator ’ box, then click ‘OK OK’’.

DRT-MAN-055 (Rev (Rev.. 06)

30

 

 

3.15.  Uninstall EConsole1  

Click ‘Uninstall EConsole1’ EConsole1’  in Windows start menu –  All Programs - EConsole1.

DRT-MAN-055 (Rev (Rev.. 06)

31

 

 

4. EConsole1 Overview 4.1.  Program Start-up In order to run the program, you can double click the EConsole1 icon in the desktop or select ‘EConsole1 EConsole1’’ by Windows Menu – All Programs – EConsole1.

Figure 20. EConsole1 Shortcut Icon

If the program is properly initiated, a log-in screen as the one you see in [Figure 21] will 21] will appear.

Figure 21. Log-in Screen Enter the ID and Password, then click the Login icon.

DRT-MAN-055 (Rev (Rev.. 06)

32

 

  If you enter an invalid ID or Password, you will be presented with a screen as in [Figure 22] 22] and  and will not be able to log in.

Figure 22. Log-in Error

DRT-MAN-055 (Rev (Rev.. 06)

33

 

 

4.2.  Main Screen of the Program You will be presented with EConsole1 main screen as in [Figure 23] 23]   with a successful log-in.

Figure 23. Main Screen of EConsole1

DRT-MAN-055 (Rev (Rev.. 06)

34

 

 

4.2.1. Function Tabs It pertains to area ⓐ  in [Figure 23] 23],, and there are two tab icons on the upper left corner.

Figure 24. Tab Icons of EConsole1 These tabs are used to carry out various functions of EConsole1. The functions of each tab are as follows. Manages the Worklist and Studylist. Worklist is the list of patients. Studylist is the list of examined studies. Refer to  to 5. List (p. (p. 39)  39)  for more details Performs X-ray projection of selected a patient in the Worklist, or reviews selected examined study in the Studylist. Refer to  to 6. Exam (p. (p. 74)  74) for  for more details.

DRT-MAN-055 (Rev (Rev.. 06)

35

 

 

4.2.2. Status Information Display It pertains to the lower area ⓑ  in [Figure 23] 23],, and displays various status information as in [Figure 25] 25]..

Figure 25. Status Information Bar

①  Hardware & Connection Connection Status Display Displays the status of the hardware and the connection status between equipment requiring connections [FLAATZ-600] Displays the acquisition mode in detector. There are AED and Switch box mode. [EVS-4343, EVS-3643] Displays the acquisition mode in detector. There are AED, Switch box, Sync mode. If the user want to change the mode click the icon and the default mode could change using ECali1.(ECali1 - Configuration - Detector X Exposure mode) Mode change sequence. ∙  USB Switch box : USB Switch box AED ∙  Sync T Trigger rigger : Sync Trigger AED ∙  AED : AE AED D only Displays the connection status and the available space of the currently connected HDD. The

displayed

number

is

the

ratio

of

the

remaining HDD volume. Displays the connection status(On/Off) status(On/Off) with the Xray detector. DRT-MAN-055 (Rev (Rev.. 06)

36

 

  [EVS4343, EVS3643] Displays the temperature. Displays the connection status with Storage Server. When you click this icon, the pop-up window will be show. You can change the PACS or do test in it. The UI is almost same as Configuration's PACS menu. Refer to  to 7.8.3 PA PACS CS (p. (p. 119)  119)   Displays the connection status with the DICOM Printer. Displays the connection status with the Modality Worklist server. [FLAATZ-600, EVS-4343, EVS3643] Displays the battery condition of the wirelessly connected detector. The displayed number indicates the ratio of remaining battery life. [EVS-4343, EVS-3643]  EVS-3643]  Battery is in wireless charging charging.. [FLAATZ-600, EVS4343, EVS-3643] Displays the signal strength of the wirelessly connected detector.

②  Current Time & Log-in Log-in Information Display

 

Displays the current date, time and user.

 

The program can be terminated by clicking the Logout icon.  

DRT-MAN-055 (Rev (Rev.. 06)

37

 

  ③ Configuration Icon

 

Opens the Configuration screen for various sett settings. ings.

 

(p. 90)  90) for Refer to 7 to 7.. Configuration (p.  for more details.

DRT-MAN-055 (Rev (Rev.. 06)

38

 

 

5. List The List tab displays the list of patients and studies as described below.  

Worklist   Worklist  

Displays the list of patients scheduled for examination.  

 

Allows users to add/edit/delete patients on the list.  

 

Allows users to proceed to exam for the selected patient.  

 

When you run the Econsole1 Vet for the first time you will be asked to specify the list length. It will be saved automatically.  Please refer to 5.1 to 5.1 Worklist (p.40) (p.40)  for more details.  details. 

 

Studylist   Studylist  

Displays the examined studies.

 

Allows users to edit/delete the examined studies on the list. Note, editing and deletion of images on the list will not affect to the PACS server.

 

Allows users to transmit examined images to the PACS server.

 

 

Please refer to  to 5.2 Studylist (p. (p. 54)  54) for  for more details.

Please note!  

The information on the Worklist and Studylist are stored on the Local Storage (HDD) by default. 

 

Every time Econsole1 is launched, a new Worklist is retrieved from the Worklist server and stored on the Local Storage. 

 

Contents on the list which are edited and deleted in Econsole1 Vet only affects the Local Storage by default. However, edited records may be transmitted to the PACS server using the Send command. 

DRT-MAN-055 (Rev (Rev.. 06)

39

 

 

5.1.  Worklist Displays the list of patients. The list is displayed when the

icon, on the upper

portion of the List tab, is clicked.

5.1.1. Main Screen of the Worklist The main screen of the Worklist is as seen in [Figure 26]. 26].

Figure 26. Worklist Screen

DRT-MAN-055 (Rev (Rev.. 06)

40

 

  ① List of Patients  

The types of information are as noted in [Table 2] 2]..  

The displayed information and its order may be set in

 

Configuration. The display order may be changed by clicking and dragging the name of the information type(ID, Name and etc.).

 

If you run the EConsole1 first, please set the column size at your convenience it will save automatically. Table 2. Information Type of the Worklist Type

Description Accession Number.

Acc. No. ID Name

X-ray study registration number Patient ID Patient name

Date/Time

Expected time of study

Study Desc.

Study Description

D.O.B. Sex Order Date  

 

Date of Birth Sex of the patient Time of order

Patient selection in the Worklist.  

Patient can be selected by left-clicking the mouse.

 

The selected patient is displayed in blue .

 

Patient can be edited and deleted after selection.

Projection can be performed in the Exam tab as a selected patient is double-clicked.  

It is equivalent to clicking the ‘Exam Exam’’ icon after selecting a specific patient.

 

Please refer to  to 6. Exam (p. (p. 74)  74) for  for more details.

DRT-MAN-055 (Rev (Rev.. 06)

41

 

  ② Search Bar & Refresh Icon  

Search Bar

 

Searches the values of Patient ID, Name, and Access Number within the Worklist.

 

If the

is clicked after entering the value to be

searched, the list will only display relevant works.  

If the

icon is clicked without entering a specific value,

an advanced search window will appear.  

Advanced Search Window

Figure 27. Advanced Search Window

DRT-MAN-055 (Rev (Rev.. 06)

42

 

   

Performs

searches

utilizing

multiple

conditions

simultaneously.  

 

Only displays works satisfying all entered details.

Refresh icon   Retrieves recently registered patients from the PACS server.

③ Add/Edit/Delete Icon Table 3. Add/Edit/Delete Icon Icon

Description Adds a new patient. Add Window is opened when the pertinent icon is clicked.

Add

Please refer to  to 5.1.2 Add Window  Window (p. (p. 45)  45) for  for more details. Edits the selected patient information on the Worklist. Only allows editing of patient information added by EConsole1. Edit Window is opened when the patient is selected and the edit

Edit

icon is clicked. Please refer to  to 5.1.4 Edit Window (p. (p. 51)  51) for  for more details.  details.  Deletes the selected patient. Only deletes the patients added by EConsole1.

Delete Import the excel file to worklist Please refer to for more details. Excel

Please refer to for more details.

import Save patient list to detector. This function works only FLAATZ-600, EVS-3643 detector. Portable

Please refer to for more details.

DRT-MAN-055 (Rev (Rev.. 06)

43

 

 

④ Exam Icon    

Switches to the Exam tab to exam of the selected patient. Please refer to  to 6. Exam (p. (p. 74)  74) for  for more details.

DRT-MAN-055 (Rev (Rev.. 06)

44

 

 

5.1.2. Add Window Add Window appears when the

icon is clicked [Figure 28] 28]..

Figure 28. Add Window

DRT-MAN-055 (Rev (Rev.. 06)

45

 

  ① Patient Information Study, patient, radiographer information is entered. Details for each category are as described in [Table 4]. 4]. Table 4. Patient Information (Add Window) Category

Description

ID

Patient ID

Accession Number

Registered identification number

Name

Patient Name

Operator

Radiographer Name

Sex

Sex of the patient

Birth Date

Date of birth of the patient

Age

Age of the patient

② Emergency Icon  

If this icon is clicked, a default value is set as the ID and Accession Number and switches over to the Exam tab.

③ Projection Information  

Displays information of projection.   In no information is provided

initially,

the

projection

where

projection

information can be set by clicking ④ Edit icon.

④ Edit Icon  

Opens

the

projection

setting

window

information may be entered.  

Please refer to  to 5.1.3. Edit Projection (p. (p. 49)  49) for  for more details.

DRT-MAN-055 (Rev (Rev.. 06)

46

 

  ⑤ OK Icon  

Saves the patient with the entered information and closes the window.

⑥ Exam Icon  

Saves the patient with the entered information and directly switches to the Exam tab to exam.

 

Please refer to 6 to 6.. Exam (p. (p. 74)  74) for  for more details.

⑦ Cancel Icon  

Cancels the work addition and closes the window.

* Note If multiple projections are registered for a single patient, each projection receives a different Accession Number [Figure 29] 29]..

Figure 29. Multiple Projections Registration in a Single Event If the OK icon is clicked, each projection registered in the Worklist as 2

DRT-MAN-055 (Rev (Rev.. 06)

47

 

  different studies. [Figure 30] 30]  

Figure 30. Two studies for one patient

In order to register projections as a single study, the Accession Numbers for each projection must be equally modified prior to clicking the ‘OK OK’’  icon. [Figure 31]  31] 

Figure 31. Enter the Same Accession Number to Register as a Single Study

DRT-MAN-055 (Rev (Rev.. 06)

48

 

 

5.1.3. Edit Projection Window It is the window which appears on the Add Window when the Edit icon is clicked.

Figure 32. Edit Projection Window

① Anatomy Information  

Select the anatomical region to be examined.  

Projection and Position are determined subsequent to the selected anatomical region.

 

Stand

Receptor Type will be changed by clicking the icon.

Table

DRT-MAN-055 (Rev (Rev.. 06)

Wireless1

Wireless2

Wireless3

49

 

  ②  Projection & Position List  

Specific selection of regions selected in Anatomy.

③ Add Icon  

This icon registers the currently set Projection, Position, Receptor Type information.

 

The currently set Projection, Position and Receptor Type information is also registered when the selected Position is double clicked.

④  Registered Registered Projection Information Information  

Displays the list of registered Projection, Position and Receptor Type.

 

The Receptor Type can be altered by double clicking the Position in Dual Mode.

⑤  Projection Information Information Sequence Edit Icon  

Changes the projection order.

Up

Down

⑥ Delete Icon  

Deletes the selected item.

⑦ OK Icon  

Saves and closes the window.

DRT-MAN-055 (Rev (Rev.. 06)

50

 

  ⑧ Cancel Icon  

Cancels and closes the window.

5.1.4. Edit Window The Edit Window opens when a study is selected from the Worklistt and the Edit icon is clicked. Worklis (Caution) Edit Window is not able to be modified on the study information, retrieved from the PACS server.  server. 

Figure 33. Edit Window

DRT-MAN-055 (Rev (Rev.. 06)

51

 

  ① Study Information  

Edits the information of the radiographer and the patient.

 

Information are as described in [Table 5] 5].. Table 5. Study Information (Edit Window)

Category ID Name Accession Number Operator

Detail Patient ID Patient Name Registered identification number Radiographer Name

Study Description

Study Description

Sex

Sex of the Patient

Birth Date Age Schedule date & time

Date of Birth of the Patient Age of the Patient Scheduled date and time of imaging

② OK Icon  

Saves and closes the window.

③ Cancel Icon  

Cancels and closes the window.

5.1.5. Edit Window ①  Make a patient list ✓

Use the template found at 'installed path\ Config\Worklist_sample.xlsx' to make a patient list.

DRT-MAN-055 (Rev (Rev.. 06)

52

 

  ②  Limitation  

You can select among the 'M', 'F', 'U' in patient Sex.

 

Birth Date Date and Study Date's format is 'MM/DD/YYYY'.

It is not support other format.   Birth Time and Study Time's format is 'HH:MM:SS'. It is not support other format.

③ Excel import  

Click the excel Import and load the file.

DRT-MAN-055 (Rev (Rev.. 06)

53

 

 

5.2.  Studylist Displays the examined study list. The list is visible if the clicked.

icon at the top of the List tab is

5.2.1. Main Screen of the Studylist The main screen of the Studylist is as seen in [Figure 34]. 34].

Figure 34. Main Screen of the Studylist

DRT-MAN-055 (Rev (Rev.. 06)

54

 

  ① List of the Examined Patients  

Displays the list of various information of the examined patients.  

One row represents one patient information and one ID

 

may have multiple studies. The column of the list may be changed by dragging the name of the category (ID, Name and etc.).

 

You can change the list's height using Ctrl key + Mouse wheel or using Ctrl key + arrow key

 

Types of information are as noted in [Table 6] 6].. Table 6. Information Types of Patient

Category Name

Patient name

ID

Patient ID

Sex

Sex of the patient

D.O.B. Date/Time  

Description

Date of Birth Date / time for recent examined study

Patient Selection  

Patient can be selected by clicking the left button of the mouse.

 

Selected patient is displayed in blue. From the ‘① ‘① Examined  Examined Patient List’ List’ and ‘② ‘② Examined  Examined Study List per Patient’ Patient’.

 

Selected patient may be edited or deleted.

②  Examined Study Study List per Patient Patient  

Displays detailed study information of the selected patient on the ‘① ‘① List  List of the Examined Patients’ Patients’.

 

Types of information are as noted in [Table 7] 7]..

DRT-MAN-055 (Rev (Rev.. 06)

55

 

  Table 7. Information Types of Examined Study List Items

Description Accession Number

Acc. No.

Registered identification number

Study ID

RIS code

Study Desc.

Study Description

Status

[Number of images] / [Number of projections]

Date/Time

Date / time for examined study.

③  Thumbnail

 

Displays the images from the selected patient.

 

A check mark is added to thumbnail of the selected images for the opening or transmissi transmission. on.  

The thumbnail can be selected by double clicking.

 

Images transmitted to the PACS server have a green outline.

 

1/1 : Series number / image number

④ Search Bar & Refresh Icon  

Search Bar  

Searches the values of Patient ID, Name, and Access Number within the Studylist.

 

If the

is clicked after entering the value to be

searched, the list will only display relevant works.  

Refresh Icon   Renews the list to the latest state.

DRT-MAN-055 (Rev (Rev.. 06)

56

 

  ⑤  Icons Table 8. Studylist Icons Icon

Description Detailed overview of a single selected image from the list of thumbnails.

Open Study

Please refer to  to 6. Exam (p. (p. 74)  74) for  for more details. Repeat exam for the selected study in the Exam tab.

Repeat Exam

Please refer to  to 6. Exam (p. (p. 74)  74) for  for more details. Performs merging of two studies or patients. Please refer to  to 5.2.2 Merge Window (p. (p. 58)  58) for  for more

Merge

details. Stores images on a USB, CD, or DVD, and print through the DICOM Printer. Please refer to  to 5.2.3 Export Window (p. (p. 60)  60) for  for more

Export

details. Edits the selected patient or study information in the Studylist. Edit Window appears if the Edit Icon is clicked.

Edit

Please refer to  to 5.2.4 Edit Window (p. (p. 65)  65) for  for more details.

Deletes the selected item.

Delete

Save the patient list to excel. Excel export It makes selected series's report. Please refer to for more details.

Report

⑥ Send Icon  

Transmits the selected images to the PACS server.

DRT-MAN-055 (Rev (Rev.. 06)

57

 

 

5.2.2. Merge Window  

Merges the images and studies.

 

Allows editing when the image and patient information do not correspond with each other.

Figure 35. Merge Window

①  & ②  Merge Subjec Subjectt Imaging In Information formation  

Study lists appear on each side.

 

Required items can be easily searched using the search tool.

 

Identical items selected items in ①  are located in ②  if the SameList is checked.

DRT-MAN-055 (Rev (Rev.. 06)

58

DRT MAN 055 (Rev (Rev.. 06)

58

 

  ③ Merge Icon  

Transmits the selected items.

 

Everything is saved while being transmit transmitted. ted.

④ Close Icon  

Closes the Merge window.

DRT MAN 055 (Rev (Rev.. 06)

59

 

 

5.2.3. Export Window Saves the images of the selected study from the Studylist to a USB, CD or DVD, and prints it through the DICOM Printer.

5.2.3.1. 

Drive

Figure 36. Export USB Screen  

Stores the images in the local drive.

 

Only files in BMP, JPG, PNG and TIF format may be exported.

 

If the local drive is selected, the current list of stored files are displayed.

 

Export File Format  

 

Selects Format for storage.

Image is stored to the USB once the OK icon is clicked.  

File name is automatically designated and it will be stored It will be stored at 'driver\ID_Name\' folder. .

DRT MAN 055 (Rev (Rev.. 06)

60

 

  5.2.3.2. 

Print

Figure 37. Export - Print Screen  

Prints through the DICOM Printer.  

The DICOM Printer must be connected.

① Print Setting  

 

Printer Selection  

Selects a DICOM Printer.

 

You can register using 'Config' - 'Network' - 'Printer' menu.

Film Orientation  

 

Film Size  

 

Selects a film orientati orientation on between portrait/landscape. Selects the size of the film.

Layout   Sets the print layout.

DRT MAN 055 (Rev (Rev.. 06)

61

 

   

The ‘Divide Divide’’  icon must be clicked in order to apply any changes made in Layout.

 

Zoom  

Performs zoom in/out of image.

②  Preview  

Thumbnail Screen  

The image to be printed appears in the thumbnail list on the right.

 

Preview Screen  

If an image is selected in the thumbnail, a large image appears in the center preview screen.

 

 

Merge Icon  

Merges the divided screen which has been separated by ‘Divide Divide’’.

 

Only adjacent screens of the same sizes can be merged

 

The merged shape must be a square.

Divide Icon  

 

Rotation Icon  

 

Divides the screen by the value set in ‘Layout Layout’’. Rotates the screen 90 degrees clock-wise.

Flip Icon  

Reverses the up and down sides.   Print Icon  

Run a paper print

③ DICOM Print button  

Performs printing.

DRT MAN 055 (Rev (Rev.. 06)

62

 

  5.2.3.3. 

CD / DVD

Figure 38. Export - CD / DVD  

Stores the images on a CD or DVD.

 

Only DICOM files can be Exported.

 

A CD/DVD Recorder device is required.

①  CD / DVD Media Configur Configuration ation  

Selects the target media to be exported.  

 

Volume Label  

 

Configures the name of the target media.

Total Space  

 

CD, DVD are available.

Displays the total available space of the target media.

Insert   Closes the CD / DVD Recorder device.

DRT MAN 055 (Rev (Rev.. 06)

63

 

   

Eject  

Opens the CD / DVD Recorder device.

② Export File List  

Displays the list of files to be Exported.

③ Export Process  

Estimated Time  

 

Presents the expected time needed for Export.

Time Left  

Displays the time left of the estimated time.

④ Task Icons  

Burn  

 

Performs the Export task.

Cancel  

Cancels the Export task.

DRT-MAN-055 (Rev (Rev.. 06)

64

 

 

5.2.4. Edit Window 5.2.4.1. 

Edit Patient Information

Figure 39. Edit Window on Studylist (Patient Information)  

Only patient information (within the red outline) can be edited when the Edit icon is clicked in the patient list of the Studylist (upper list).

 

OK Icon  

 

Saves and closes the window.

Cancel Icon  

Cancels and closes the window.

DRT-MAN-055 (Rev (Rev.. 06)

65

 

  5.2.4.2. 

Edit Study Information

Figure 40. Edit Window on Studylist (Study Information)  

Only study information (within the red outline) can be edited when the Edit icon is clicked in the projection list (lower list) of the Studylist.

 

OK Icon  

 

Saves and closes the window.

Cancel Icon  

Cancels and closes the window.

DRT-MAN-055 (Rev (Rev.. 06)

66

 

 

5.2.5.  Report 5.2.5.1. 

Report

Figure 41. Create Report  

Create report using Report button in the Stu StudyList. dyList.

①  Patient information  

Check or change the Selected series's patient information.

②  Attached image  

User can attach up to 4 images using the 'Image load' button.  

DRT-MAN-055 (Rev (Rev.. 06)

67

 

  ③  Write opinion  

User can write opinion.

④  Print / Save / Setting  

User can print the report.

 

Save a report to PDF file.

 

Save a report.

 

Set the report Format.

DRT-MAN-055 (Rev (Rev.. 06)

68

 

  5.2.5.2. 

Image load

Figure 42. Create Report  

Layout button  

Select 1X1, 1X2, 2X2 layout.   Image load button    

OK button  

 

Select image Attach the selected image to report

Cancel button  

Cancel to attach images

DRT-MAN-055 (Rev (Rev.. 06)

69

 

  5.2.5.3. 

Select the image

Figure 43. Select the image    

 

You can select the image when you click "Image load" button. Search button   Search the patient using input information. Patient/Series/Image list  

 

OK button  

 

Select the image. Load the selected image.

Cancel button  

Cancel to load the image.

DRT-MAN-055 (Rev (Rev.. 06)

70

 

  5.2.5.4. 

Input the opinion

Figure 44. Select the image  

Input the selected series opinion.

DRT-MAN-055 (Rev (Rev.. 06)

71

 

  5.2.5.5. 

Set report form

Figure 45. Set report form  

User can set the report form using 'setting' menu.

 

Address/ Extra Info  

 

Input the Hospital address or further information.

Session  

Input the Session.

 

Default values are "RECOMMENDATION", "FINDING", and "CONCLUSION".

 

 

Visible setting  

Select the printing report's repor t's contents contents..

 

All selection is default value.

Save as default button.  

 

OK button  

 

Save current value as default. Use selected form.

Cancel button  

Cancel the selected form.

DRT-MAN-055 (Rev (Rev.. 06)

72

 

  5.2.5.6. 

Print

Figure 46. Print  

User can print the report using print button. 

DRT-MAN-055 (Rev (Rev.. 06)

73

 

 

6. Exam  

The Exam tab processes exam and edit images.

Figure 47. Main Screen of Exam Tab

① Patient Information  

Displays ID, patient name, study description.

② Detector Status Display  

Displays the status of the detector to be used in exam.  

Blue: Ready state

DRT-MAN-055 (Rev (Rev.. 06)

74

 

   

Orange: Busy state

③ Image Display Screen  

Area - display images.

④  Toolbar  

Collection of edit image icons

 

The width of the column may be changed.

 

Please refer to  to 6.1 Toolbar (p. (p. 77)  77) for  for detailed information.

⑤ Projection List & Image Processing Tab  

May view the list for images, or set the post-processing parameter which will be applied to the images.

 

Please refer to  to 6.2 Projection List & Image Processing Tab  Tab (p. 84)  for detailed informat 84) information. ion.

⑥  Generator Condition Setting Setting  

Only available with integration with X-ray generator.

 

Tube Voltage (kVp), Tube Current (mA), Tube Current Amount (mAs) setting.  

If the tube current amount icon, mAs, is clicked, the unit may be changed to exposure time (seconds).

 

Patient Size  

Selects the volume of the patient‘ patient‘s body.

 

Sets value of the tube current, tube voltage and tube current amount, are renewed subsequent to the patient’ patient ’s volume.

 

AEC  

 

Sets the AEC to be used.

Density

DRT-MAN-055 (Rev (Rev.. 06)

75

 

     

Focus  

 

Configures the imaging density. Configures the Focal Size.

Grid  

Configures the applied Grid.

⑦  Save / Indicator Indicator / Send Icons  

Save Icon  

 

Stores the images on the Local Storage.

Send Icon  

Transmits the images to the PACS server.

 

Transmits all images performed to this icon with a Long-

click to the PACS server.   Indicator Icon

Standby

Ready

Exposure

 

Standby: Default status. May begin exam.

 

Ready: Ready for the detector to get image.

 

Exposure: Emits x-ray to the detector.

DRT-MAN-055 (Rev (Rev.. 06)

76

 

 

6.1.  Toolbar Collection of icons required for image analysis and editing.

6.1.1. Toolbar Overview

① Toolbar Width Adjustment Icon  

The width can be changed by clicking this icon.

 

The width may be set from 1 to 4.

② Tools List  

Displays the currently configured list of icons.

 

The icons can be edited in the Edit Toolbar window.

③ Edit Icon  

Opens the Edit Toolbar window for editing the icons.

 

Please refer to 6.1.2 Edit Toolbar (p. 78) 78)   for detailed information information..

DRT-MAN-055 (Rev (Rev.. 06)

77

 

 

6.1.2. Edit Toolbar

Figure 48. Edit Toolbar Window

① Applied Icons Box  

It is the currently applied icons in the Exam tab.

 

Icons in this box can be deleted by drag & drop of the icon into ‘② ‘② Tools  Tools Collection’ Collection’.

 

Icons can be added into this box by dragging icons in ‘② ‘②   Tools Collection’’ and dropping it into the empty location. Collection

DRT-MAN-055 (Rev (Rev.. 06)

78

 

  ② Available Icons  

Contains all of icons.

③ Add Shutter Icon  

Adds icons with user specified ROI Shutter Sizes.

 

The following window appears with the click of the icon.

Figure 49. Add Sutter Size Window

④ Shutter Delete Icon  

Only deletes user added Shutter.

⑤ OK Icon  

Applies the modified tools arrangement and closes the window.

⑥ Cancel Icon  

Cancels and closes the window.

DRT-MAN-055 (Rev (Rev.. 06)

79

 

 

6.1.3. Toolbar Icons All of toolbar icons, noted in [Table 9] 9].. Table 9. Toolbar Icons Icon

Description Sets the ROI size to 8× 8×10 Sets the ROI size to 10× 10×12 Sets the ROI size to 14× 14×17 Sets the ROI size to 17× 17×17 Sets the ROI size to the user specified size Sets the ROI into a square shape Sets the ROI into a circular shape Sets the ROI shape into a user specified shape Displays the “R" marker on the screen Displays the "L" marker on the screen Displays the "ERECT" marker on the screen Displays the "SUPINE" marker on the screen Displays the "SITTING" marker on the screen Displays the "STRESS" marker on the screen Displays a user entered text on the screen Displays the "PA" marker on the screen

DRT-MAN-055 (Rev (Rev.. 06)

80

 

  Displays the "AP" marker on the screen Displays the "OBL" marker on the screen Displays the "LAT" marker on the screen Displays the "FLEX" marker on the screen Displays the "EXT" marker on the screen Displays the "↑ "↑" marker on the screen Displays the "← "←" marker on the screen Displays the "↓ "↓" marker on the screen Displays the "→ "→" marker on the screen Arrange windows side-by-s side-by-side ide symmetrical symmetrically ly Arrange windows up-and-dow up-and-down n symmetri symmetrically cally Rotates the image 90 degrees clockwise Rotates the image 90 degrees counter-clockwise Zoom-in and out of the image Fits image to screen Partially expands the image Moves the image up/down/left/right Displays the screen in 1× 1×1 Displays the screen in 2× 2×1 Displays the screen in 1× 1×2

DRT-MAN-055 (Rev (Rev.. 06)

81

 

  Displays the screen in 2× 2×2 Measures the length of the image Measures the angle of the image Verifies the DICOM Header information Deletes any annotation or markers indicated on the image using other tools Adjusts the window level and width of the image Inverts the image Stitch images Displays the histogram Captures the image. Prints the image. Resets the edited image into its initial form Deletes the image Ruler calibration Duplicate the image Change mouse sensitivity Run the dark scan manually(EVS-43 manually(EVS-4343, 43, EVS-3643 only) Draw ellipse annotation

Draw rectangle annotatio annotation n

DRT-MAN-055 (Rev (Rev.. 06)

82

 

  Draw polygon annotation Draw free line annotation Draw arrow annotation Select annotation

DRT-MAN-055 (Rev (Rev.. 06)

83

 

 

6.2.  Projection List & Image Processing Tab The Projection List & Image Processing Tab can be used through the tab icon on the upper-right corner.

Figure 50. Projection List (Left) & Image Processing Tab (Right)

DRT-MAN-055 (Rev (Rev.. 06)

84

 

 

6.2.1. Projection List Displays a list of studies categorized by projections.

Figure 51. Projection List Tab

①  & ④  Projection an and d Po Position sition  

Projection can be selected by clicking ①  icon.

 

④  presents the available positions of the selected projection projection..  

Allows verification of thumbnails, position designations, and Receptor type.

POP-UP function  

When you click the mouse right button you can change the options.

 

ROI Marker visible : Set the default ROI and Marker visible when the image comes from detector.

 

Find auto position ∙  All projection : Find

unexposed position among the all

projections When you save or send a image. ∙  One projection : Find unexposed position in the selected projection when you save or send a image.

DRT-MAN-055 (Rev (Rev.. 06)

85

 

  ∙  Do not apply : do not find unexposed postion when you save or send a image.

② Projection Guide Icon  

The projection guide for the selected projection.

Figure 52. Projection Guide  

The presented projection guide may be closed using the left click of the mouse.

③ Edit Projection Icon  

Please refer to 5.1.3 Edit Projection (p. (p. 49) 49)   for detailed information.

DRT-MAN-055 (Rev (Rev.. 06)

86

 

 

6.2.2. Image Processing Tab * Function for advanced users * It is utilized to acquire a more appropriate image by adjusting the post-process parameter of the completed image.

Figure 53. Image Processing Tab

①  Histogram  

Displays the histogram of the currently selected image.

 

Auto WL: Sets the window level automatically.

DRT-MAN-055 (Rev (Rev.. 06)

87

 

  ② Image Enhancing Parameters  

Adjusts the image enhancing paramet parameters. ers.

 

The adjustable parameters are as noted in [Table 10] 10].. Table 10. Image Enhancing Parameters Image

Description

Processing Contrast

Adjusts the contrast

Edge

Adjusts the Edge

Mask

Adjusts the mask size

Bright

Adjusts the brightness

Smooth

Adjusts the smoothness

③  Parameter Parameter Function Icon  

Wizard  

Opens the multi-image viewer window.

 

8 other types of images may be comprehended based on the currently set parameter.

 

Apply  

 

Reset  

 

Applies the currently set parameter to the image. Resets the parameter to its initial setting.

Save  

Saves the currently set parameter and applies it to ensuing imaging of the pertinent projection-positi projection-position. on.

DRT-MAN-055 (Rev (Rev.. 06)

88

 

 

6.3.  Truview ART * It is for advanced user. It is works only EVS-3643, EVS3643 detector for sharpness

①  Use TRUVIEW ART Default value IS "OFF". If you set "ON", it will be adjusted selected image. ②  Magnified screen It magnified selected position in the image viewer. It can select using double click in the image viewer ③  Compare button Compare with not adjusted/adjusted TRUVIEW ART. 2 image come up. Left image is not adjusted TRUVIEW ART, Right image is adjusted TRUVIEW ART. ④  Save button Save a TRUVIEW ART parameter in the Selected Position. It will take more than 3 seconds when you adjust the TRUVIEW ART

DRT-MAN-055 (Rev (Rev.. 06)

89

 

 

7. Configuration Configures the user settings of EConsole1. Administrator Administrator privileges are required for configuration. Contents of configuration are saved and applied immediately following changes. Categories of Configuration are as noted in [Table 11] 11].. Table 11. Configuration Category Category

Display

Description The tab (List/Exam) of EConsole1 may be viewed simultaneously by division depending on the number of connected monitors, and performs relevant configurat configurations. ions. Please refer to  to 7.1 Display (p. (p. 92)  92) for  for detailed informatio information. n. Allows configuration of language, date display, hospital

System

information and etc. Please refer to 7.2 to 7.2 System (p. (p. 93)  93) for  for detailed information. Allows configuration of image storage folder, delete policy,

Image

annotation information and etc. Please refer to  to 7.3 Image (p. (p. 99)  99) for  for detailed information. Performs configuration related to PC Storage (HDD).

Storage

Please refer to  to 7.4 Storage (p. (p. 107)  107) for  for detailed information. Manages the accounts of EConsole1 users

Account

Please refer to  to 7.5 Account (p. (p. 112)  112) for  for detailed information. Configures unnecessary RIS code filtering subsequent to

Block List

ultrasound or other devices. Please refer to  to 7.6 Block List (p. (p. 113)  113) for  for detailed

DRT-MAN-055 (Rev (Rev.. 06)

90

 

  Category

Description information. Configures the DICOM Header Format.

Dataset

Please refer to  to 7.7 Dataset (p. (p. 114)  114) for  for detailed information. Configures the network information, such as Local, PACS,

Network

Printer and etc. Please refer to  to 7.8 Network (p. (p. 115)  115) for  for detailed information. Configures the Detector which will be connected with the

Detector

EConsole1. Please refer to  to 7.9 Detector (p. (p. 124)  124) for  for detailed information. Configures the RIS Code.

RIS Code

Please refer to 7.10 RIS Code  Code  (p. 129) 129)   for detailed information. Allows the configuration of projection and position.

Projection

Please refer to  to 7.11 Projection Editor (p. (p. 137)  137) for  for detailed information. Configures the Generator which will be connected.

Generator

Please refer to  to 7.12 Generator (p. (p. 143)  143) for  for detailed information.

DRT-MAN-055 (Rev (Rev.. 06)

91

 

 

7.1.  Display

Figure 54. Display Configuration Option

Description

1 Monitor

Dual Monitors * Caution: Unstable option. *

DRT-MAN-055 (Rev (Rev.. 06)

92

 

 

7.2.  System  

Performs configurations affecting the overall EConsole1.

Subcategories Language Date Institution Options Worklist

Description Selects system language. Selects date notation. Enters the information of the institution using the EConsole1 Enables and disables automa automatic tic running of EConsole1 at startup of Windows. Edits the type and order of information presented in the Worklist.

DRT-MAN-055 (Rev (Rev.. 06)

93

 

 

7.2.1. Language

Figure 55. Language Configuration Screen  

Configures the system language.

Items

Description

Korean

Set when you use the Korean window.

Chinese

Set when you use the Chinese window.

Japanese

Set when you use the Japanese window.

Portuguese

Set when you use the Portuguese window.

Spanish

Set when you use the Spanish window.

German

Set when you use the German window.

Turkish

Set when you use the Turkish window.

Swedish

Set when you use the Swedish window.

French

Set when you use the French window.

 

DICOM character set

 

Note that DICOM character set is defined separately from system language.

DRT-MAN-055 (Rev (Rev.. 06)

94

 

  Items

DICOM character set

Korean

ISO IR 6\ISO 2022 IR 149

Chinese

ISO IR 6\GB18030

Japanese

ISO 2022 IR 13\ISO 2022 IR 87

Default

ISO IR 100

Turkish

ISO IR 148

Thai Ukrainian, Russian UTF-8

ISO 2022 IR 166 ISO IR 6\ISO IR 144 ISO IR 192

DRT-MAN-055 (Rev (Rev.. 06)

95

 

 

7.2.2. Date

Figure 56. Date Configuration Screen  

Configures the date notation.

 

Date  

 

Delimiter  

 

Y: Year , M: Month, D: Day Configures the symbol between Y, M, and D.

Example) If today is May 31, 2012 Date

Delimiter

Final Notation

MM DD YYYY

 /

5/31/2012

MM DD YYYY

-

5-31-2012

MM DD YYYY

.

5.31.2012

YYYY MM DD

 /

2012/5/31

MM DD YY

-

5-31-12

YY MM DD

.

12.5.31

DRT-MAN-055 (Rev (Rev.. 06)

96

 

 

7.2.3. Institution

Figure 57. Institution Configuration Screen  

Input the hospital/institution information using EConsole1.

 

Hospital  

 

Hospital Name

Department  

Department name within the hospital.

DRT-MAN-055 (Rev (Rev.. 06)

97

 

 

7.2.4. Options

Figure 58. Options Configuration Screen  

Auto Start  

If

checked,

EConsole1

will

automatically

start

when

Windows start.  

Virtual Keyboard  

 

If checked, virtual keyboard will be available for use.

Auto Update  

Requires internet connection.

 

If checked, EConsole1 will update automatically when a new version is released.

DRT-MAN-055 (Rev (Rev.. 06)

98

 

 

7.3.  Image 

  Allows configuration relevant to examed images.

Subcategories

Description

Image Folder

Sets the location to save the images.

Delete Option

Sets delete options for image files.

Overlay

Sets overlay on the images.

Import

Imports images to the Studylist.

DRT-MAN-055 (Rev (Rev.. 06)

99

 

 

7.3.1. Image Folder

Figure 59. Image Folder Configurati Configuration on Screen  

Sets the location to save the images within PC Storage.

 

Edit Icon  

Opens the folder selection window.

 

Select the folder where the image is to be stored.

DRT-MAN-055 (Rev (Rev.. 06)

100

 

 

7.3.2. Delete Option

Figure 60. Delete Option Configuration Screen

① Image Delete Option  

Use Password   If checked, deletion of image requires a password.

 

Use Reject Code  

If checked, a reject code (reason for deletion) is required every time for an image file is deleted.

② Reject Code    

List of reasons for deletion. Statistics   Displays the statistical record of each applied Reject Code.

DRT-MAN-055 (Rev (Rev.. 06)

101

 

   

The statistics can be exported to the Microsoft Excel file if the Export Icon is clicked within the statistic display window.

 

Delete  

 

Add  

 

Deletes the selected Reject Code from the list. Adds a new Reject Code.

Edit  

Edits the selected Reject Code.

DRT-MAN-055 (Rev (Rev.. 06)

102

 

 

7.3.3. Overlay  

Allows overlay configurati configuration. on.

Figure 61. Overlay Configuration  

Allows editing of types of overlay to be displayed on the Left Top, Right Top, Left Bottom, and Right Bottom.

 

Editing method is the same while the display location differs.

Figure 62. Overlay Editing

DRT-MAN-055 (Rev (Rev.. 06)

103

 

  ① Registered Overlay Type & Font Size  

Overlay: Information which will be displaye displayed. d.

 

Size: Font size of the pertinent per tinent information.

 

Information is displayed in accordance with the arrangement on the list.

② Edit Icon  

 

Adds a new overlay to the list.

 

The following window as in [Figure 63] 63]   appears when the icon is clicked.

Figure 63. Overlay Option Window  

The selected overlay is added after entering the size and clicking ‘OK OK’’.

DRT-MAN-055 (Rev (Rev.. 06)

104

 

   

 

Deletes the selected annotation from the list.

 

The selected annotation is moved up one row.

 

The selected annotation is moved down one row.

 

Change font size of each location.

 

 

 

DRT-MAN-055 (Rev (Rev.. 06)

105

 

 

7.3.4. Import

Figure 64. Import Configuratio Configuration n Screen  

Imports an external image into the Studylist.

 

Import Icon  

Opens the file selection window.

 

Select the image to be imported.

DRT-MAN-055 (Rev (Rev.. 06)

106

 

 

7.4.  Storage  

Configurations concerning the deletion and storage of PC where the image will be saved.

Details

Description Configures the deletion and automatic

Delete

deletion function of images stored on Local Storage.

Backup

Compresses and backs-up the images stored on Local Storage.

DRT-MAN-055 (Rev (Rev.. 06)

107

 

 

7.4.1. Delete  

Configures the deletion and automatic deletion function of images stored on Local Storage.

Figure 65. Delete Configuration Screen

① Storage Delete  

The ‘Delete Delete’’ Icon deletes all images stored on the Local Storage.

 

Deleted images may not be restored, so please perform a backup prior to deleting images!

DRT-MAN-055 (Rev (Rev.. 06)

108

 

  ② Auto Delete  

Auto Delete  

If checked, all images satisfying the specified conditions are automatically deleted from the PC Storage when the user run a Econsole1. 

 

Disk Size  

If checked, images on the PC Storage are automatically deleted once the disk space until remain storage is "Disk Size". 

 

Configurable ‘Disk Size’ Size’  is meaning expected remain space on PC storage. 

 

If remain storage is less than ‘Disk Size’ Size’, overflowed images will be automatically deleted. (Whole disk volume except ‘Disk Size’ Size’  is allocat allocated ed for Econsole1.) 

 

Max: Refers to the maximum limit of disk space utilized for storage of images on the PC Storage.  

 

Period  

If checked, images on the PC Storage are continuously deleted automatically automatically during the set period. 

 

ex1) Set to "2", "Month" : Delete all images 2 month earlier images based on the Today. 

 

ex2) Set to "1" "Day" : Delete all images 1 day earlier images based on the Today. 

 

The

cycle

may

be

set

by

one

of

the

following;

Day/Week/Month/Year.  One of the two methods, Disk Size and Period, may be selected at a time.

DRT-MAN-055 (Rev (Rev.. 06)

109

 

 

7.4.2. Backup  

Compresses and backs-up the images stored on Local Storage.

Figure 66. Backup Configura Configuration tion Screen

①  Database Backup Conf Configuration iguration  

Displays the location of the folder used in Data Backup.

 

Edit  

Opens the selection window for the folder to be used in backup.

 

Compress  

Compresses then stores the currently used data based in the selected folder.

 

Restore

DRT-MAN-055 (Rev (Rev.. 06)

110

 

   

File selection window opens.

 

Restores the selected database file and stores it in the system.

 

Backup  

Backs-up the currently used database in the selected folder.

② Automatic Backup Setting  

Auto Backup Check  

If checked, automatic backup is performed during the set period below.

 

Period  

Sets the automatic backup period.

 

Available units consists of Day/Week/Month.

DRT-MAN-055 (Rev (Rev.. 06)

111

 

 

7.5.  Account  

Configures the account of the EConsole1 user.

Figure 67. User Account Configuration Configuration  

The currently registered user information is available on the list.

 

Configures the user ID, Password, and authorization.

 

Delete Icon  

 

Add Icon  

 

Deletes the selected user account. Adds a new user account.

Edit Icon  

Edits the existing user account.

DRT-MAN-055 (Rev (Rev.. 06)

112

 

 

7.6.  Block List  

Filtering the unnecessary RIS code such as ultrasound or other devices.

Figure 68. Block List Configuration Configuration  

The currently registered Block List information is available on the list.

 

Delete Icon  

 

Add Icon  

 

Deletes the selected Block List entry. Adds a new entry to the Block List.

Edit Icon  

Edits the existing Block List entry.

DRT-MAN-055 (Rev (Rev.. 06)

113

 

 

7.7.  Dataset  

Configures the DICOM Header Format.

Figure 69. Dataset Configuration

①  DICOM Header Header Item Configuration  

Deletes the unnecessary entries defined in the existing DICOM Header Format (Full List).

 

CR and DR are separately applied.

 

Delete  

 

Deletes the selected entry.

Reset  

Initializes the entry list back to Full List.



 Image format  

Configures the Image format information.

DRT-MAN-055 (Rev (Rev.. 06)

114

 

 

7.8.  Network  

Configures network information information such as IP addresses.

Subcategories

Description Requires network connection.

Local

Configures the network information of the PC (or Work Station) currently operating EConsole1. Requires network connection.

MWL

Configures the network information of the server storing the Modality Worklist.

Storage

Requires network connection. Configures the network information of the Storage Server, which stores the image files. Requires printer connection.

Printer

Configures the connection information of the printer.

DRT-MAN-055 (Rev (Rev.. 06)

115

 

 

7.8.1. Local  

Configures the network information of the PC (or Work Station) currently operating EConsole1.

Figure 70. Local Configuration Screen  

AE Title  

Enter the name of the local PC (or Work Station) which will be displayed on the Network.

 

IP  

Enter the IPv4 Address of the PC (or Work Station) currently operating EConsole1.

 

Port  

Enter the Port number.

 

It should be configured in accordance with the server ser ver being linked.

 

Entered information will be automatically saved and applied.

DRT-MAN-055 (Rev (Rev.. 06)

116

 

 

7.8.2. MWL  

Modality Worklist Server.

 

Sets the server retrieving the patient list.

Figure 71. MWL Configuration Screen

①  Information of of server being accessed accessed  

Manages the information of server being accessed.

 

Comprised of AE Title(Access Server Name), IP, and Port information.

 

Test  

 

Verifies the connection with the selected server.

Delete   Deletes the selected server from the list.

DRT-MAN-055 (Rev (Rev.. 06)

117

 

   

Add  

 

Adds an access server.

Edit  

Edits the selected server informatio information. n.

②  Option  

Time-out  

Response standby time with the server.

 

It is regarded as connection failure if the server is unresponsive unresponsiv e for the set period of time.

③  Query  

Configures the automatic refresh setting of MWL.

 

Auto  

 

Manual  

 

MWL is automatically refreshed during the entered interval. MWL refresh must be performed manually.

Interval  

Sets the interval for automati automaticc refresh.

④  Modality  

Select the modality type of the imaging equipment.  

 

Add  

 

Example) CR, DR Adds Modality.

Delete  

Deletes the selected Modality.

DRT-MAN-055 (Rev (Rev.. 06)

118

 

 

7.8.3. PACS  

Configures the Storage Server (PACS).

Figure 72. Storage Configuration

①  Information of Storage Server to be Accessed  

Manages the information of access storage server.

 

Comprised of AE Title (Access server name), IP, IP, Port

and

Selection informatio information. n.  

If selection item is 'O', DICOM image send to it and the user can select more than 1 PACS.

 

Test  

 

Verifies the connection with the selected server.

Delete   Deletes the selected server from the list.

DRT-MAN-055 (Rev (Rev.. 06)

119

 

   

Add  

 

Adds an access server.

Edit  

Edits the selected server informatio information. n.

②  Option  

Time-out  

Response standby time with the server.

 

It is regarded as connection failure if the server is unresponsive unresponsiv e for the set period of time.

③  Query  

Time-out  

Response standby time with the server.

 

It is regarded as connection failure if the server is unresponsive for the set period of time.

DRT-MAN-055 (Rev (Rev.. 06)

120

 

 

7.8.4. Printer

Figure 73. Printer Configuration Screen

①  Printer Information & Printer Printer Selection for Connection Connection  

Manages the information of the printer to be connected.   Selects the printer from the list.  

Add  

 

Edit  

 

Edits the selected printer information.

Delete  

 

Adds a printer.

Deletes the selected printer from the list.

Test  

Verifies the connection with the selected printer.

DRT-MAN-055 (Rev (Rev.. 06)

121

 

  ②  Option  

Restricts the response time by the Printer.

③ Medium Type  

Selects the type of film.

 

May select one between Clear Film and Blue Film.

④  Border  

Sets the color of the border line.

 

May select one between Black and White.

⑤  Smoothing  

Applies Smoothing.

 

May select one between Medium, Sharp and Smooth.

⑥  Magnification  

Sets the Magnification Magnification..

⑦  Density  

Adjusts the Gray-Level Scale.

⑧ Printer DPI  

Selects the resolution of the printer.

⑨ Overlay information  

Select the additional annotation information added on the film.

DRT-MAN-055 (Rev (Rev.. 06)

122

 

   

The following pop-up window in [Figure 74] 74] appears  appears when Edit is clicked.

Figure 74. Printer Overlay Information Edit  

Instructions for use are identical with the overlay configuration of the image. Please refer to  to 7.3.3 Overlay (p. (p. 103)  103) for  for detailed informatio information. n.

DRT-MAN-055 (Rev (Rev.. 06)

123

 

 

7.9.  Detector  

Performs detector related configuration configurations. s.

 

Performs Calibration Tool and Terminal Console programs. Subcategory Description Registers and edits the model of panel General

(detector) (detect or) being used. Executes detector software, ECali1 (or Caliview) and Terminal Emulator. Detailed configuration screen concerning Panel A, regist registered ered in

Panel A

‘General General’’. Detailed configuration items for panel are created in accordance with the number of panels registered in ‘General General’’.

DRT-MAN-055 (Rev (Rev.. 06)

124

 

 

7.9.1. General  

Registers and edits the panel (detector) which will be used.

 

Executes detector software, ECali1 (or Caliview) and Terminal Emulator.

Figure 75. General Configuration Screen

①  Panel(Detector) Panel(Detector) Model Registration Registration and Editing  

Registers the panel which will be connected.

 

Add Icon  

 

Add icon registers the detector selected from Model Select.

Edit Icon  

Edit icon replaces the currently registered detector with the detector selected from the Model Select.

 

Delete  

The most recently registered detector is deleted.

DRT-MAN-055 (Rev (Rev.. 06)

125

 

  ②  ECali1  

Run Icon executes the ECali1 (or Caliview) program.

③  Terminal  

Run Icon executes the Terminal Emulator.

 

Please refer to Appendix B. Terminal Emulator (p. 150 150))  for detailed informatio information. n.

DRT-MAN-055 (Rev (Rev.. 06)

126

 

 

7.9.2. Detailed Configuration by Panel  

Performs detailed configuration for each panel(detector).

 

A menu such as ‘Panel A’ A’  and ‘Panel B’ B’  are generated in accordance with the number of panels registered in ‘General General’’.

Figure 76. Detector Detailed Configuration Screen

①  Receptor  

Sets which Receptor type the detector will utilize.

 

Stand, Table and Wireless mode are available.

 

Only wireless detectors may utilize the Wireless mode.

②  Model  

Model types of the detector.   Automatically set by the value registered in ‘General General’’.

DRT-MAN-055 (Rev (Rev.. 06)

127

 

  ③ Serial Number  

Enter the serial number of the Detector.

④ Grid Configuration  

Grid Type: Select type of grid.

 

Line/Inch: Enter the line/inch value of the grid.

 

Angle: Enter the angle value of the grid.

DRT-MAN-055 (Rev (Rev.. 06)

128

 

 

7.10.  RIS Code  

RIS (Radiology Information System) Code is a collection of

projections.   For example, it configures contents such as “A, B, C regions are to be imaged for RIS Codes for certain cancers.” cancers.”   

It allows the rearrangement of sequence of projections.

Figure 77. RIS Code Configurati Configuration on

 

Import Excel Icon  

Adds the RIS Code value from an Excel fine.

 

(Caution) Duplicate code values are overwritten with contents of the Excel file. (Overwrites the existing data).

DRT-MAN-055 (Rev (Rev.. 06)

129

 

   

Delete Icon  

Deletes the selected RIS Code.

 

Deletion of information part of the Code, such as the projection, are executed by means of the ‘Edit Edit’’ icon.

 

Edit Icon  

Manages the information, such as detailed projection information and imaging environment, of the RIS Code.

 

Please refer to 7.10.1. 7.10.1.   RIS Code information.

(p. 131) 131)   for detailed

DRT-MAN-055 (Rev (Rev.. 06)

130

 

 

7.10.1. 

RIS Code Edit Window

Figure 78. RIS Code Edit Window

①  RIS Code List Management Management  

Manages the code and provides description of RIS Codes.   Text Box  

Are to write the Code and Description when adding or editing the RIS Code.

 

3 task icons are available. The descriptions are as noted in [Table 12]. 12].

DRT-MAN-055 (Rev (Rev.. 06)

131

 

  Table 12. RIS Code List Management Icon Icon

Description Edits the code and description of the RIS Code selected from the list with the information entered in the text box below the icon. [Editing Sequence]

Edit

1) Selects the code to be edited from the list. 2) Enters in new information in the textbox below the icon. 3) Clicks the Edit icon. Adds the RIS Code with the information entered into the textbox below the icon. [Adding Sequence]

Add

1) Enters in new information in the textbox below the icon. 3) Clicks the Add icon. Deletes the RIS Code selected from the list. [Deleting Sequence]

Delete

1) Selects the code to be deleted from the list. 2) Clicks the Delete icon.

②  Editing of projections projections included in the RIS RIS Code  

Manages the projections to be included in the selected RIS Code.

 

Descriptions concerning the task icons are as noted in [Table 13].. 13]

DRT-MAN-055 (Rev (Rev.. 06)

132

 

  Table 13. RIS Code Projection Management Icon Icon

Description Both icons perform identical tasks.

, Add

Adds the projections relevant to the selected Projection/Position in the RIS Code. Deletes the Projection category category from the RIS Code.

Delete Moves the selected Projection category up one row.

Move up

Moves the selected Projection category down one Move down

row.

③  Imaging Environment Environment Configur Configuration ation  

Enter the automatically configured environment value when imaging of the pertinent RIS Code is performed.

 

Patient Size  

Selects the patient size.

 

4 levels of selections are available.



  AEC   Sets the AEC.

 

 

 

 

Grid  

Selects the Grid Type being used.

 

May select one of None, Fixed and Moving.

Focal Spot  

Sets the focus.

 

May select from Large/Small.

Density  

SID

Sets the density.

DRT-MAN-055 (Rev (Rev.. 06)

133

 

     

kVp, mA, mAs(mSec)  

 

Sets the distance between the Detector and X-ray Tube. Sets the Generator condition.

Parameter  

Locates the parameter file of the image processing which will be applied after the imaging.

 

Rotation  

 

Flip H, Flip V  

 

Selects whether to flip the image horizontally or vertically.

Shutter  

 

Sets the rotation of the image.

Selects the ROI Size.

Marker  

Sets and manages the marker, which will be automatically inserted in the image.

 

Descriptions regarding the Marker Icon is as noted in [Table 14].. 14] Table 14. Marker Icon

Icon

Description Adds a new Marker. Please refer to  to 7.10.2 Marker Add Window (p.

Add

135) for 135)  for detailed informat information. ion.

Deletes the selected marker. Delete

Moves the selected Marker up. Move up

Moves the selected Marker down. Move down

DRT-MAN-055 (Rev (Rev.. 06)

134

 

 

7.10.2. 

Marker Add Window

Figure 79. Marker Add Window  

Location  

 

Select the location the marker will appear on the image

Text  

Select the type of Marker

 

Random texts may be used as a marker, aside from the text which are provided by default

 

User registered markers may be selected from the list

 

The Marker list window may be edited using the Add Icon

DRT-MAN-055 (Rev (Rev.. 06)

135

 

  and Delete Icon  

If user want to change Maker size, Edit in 'Projection' - 'Marker Size'.   Size'.

DRT-MAN-055 (Rev (Rev.. 06)

136

 

 

7.11.  Projection Editor  

Manages the Projection type

Figure 80. Projection Editor Configuration  

Marker Size  

 

Change marker size.

Rotation  

Set rotation selected receptor's all projection & position.

 

Rotation's adapted order is each position's 'Flip Vertical', 'Flip Horizontal' is first and 'Rotation' is second.

 

Delete  

 

Deletes the selected Projection or Position Recent selections which will be deleted by clicking the

DRT-MAN-055 (Rev (Rev.. 06)

137

 

  ‘Delete Delete’’ icon are highlighted in yellow  

Edit Icon  

Edits the Projection and Position after opening the Project Editor window.

 

Please refer to 7.11.1 Projection Editor Window (p. 139) 139)   for detailed informati information. on.

DRT-MAN-055 (Rev (Rev.. 06)

138

 

 

7.11.1. 

Projection Editor Window

Figure 81. Projection Editor Window

①  Anatomy  

Must be selected first  

Projection

and

position

will

not

appear

unless

an

anatomical region is selected from the Anatomy  Anatomy    

Select the projection

 

If the Receptor Type is different, even if the anatomical regions are the same, it will be saved as a different setting  

Please be sure to set the anatomical region, as well as the Receptor type.

DRT-MAN-055 (Rev (Rev.. 06)

139

 

  ②  Projection

Figure 82. Projection Configuration  

Manages the Projection list

 

Please refer to [Table 15] 15] for  for instructions of each icon

Table 15. Projection/Position List Management Icon Icon

Description Edits the name of the selected category from the list with the entry in the left lef t textbo textboxx [Edit Sequence]

Edit

1) Select the category to be edited from the list 2) Enter the new information in the textbox on the left of the icon 3) Click the Edit icon Adds the category with the information entered into the textbox on the left of the icon [Add Sequence]

Add

1) Enter the new information in the textbox on the left of the icon

2) Click the Add icon

DRT-MAN-055 (Rev (Rev.. 06)

140

 

  Deletes the selected category from the list [Delete Sequence] 1) Select the category to be deleted from the list Delete

2) Click the Delete icon Moves the selected projection category category up one row in the list

Move up

Moves the selected projection category category down one row in the list

Move down

③  Position

Figure 83. Position Configuratio Configuration n    

Manages the Projection list Please refer to [Table 15] 15] for  for instructions concerning each icon

DRT-MAN-055 (Rev (Rev.. 06)

141

 

  ④ Imaging Environment  

Sets the appropriate imaging environment for the selected projection

 

Detailed information concerning the Imaging Environment Configuration are identical to the Imaging Environment Configuration of the RIS Code Edit Window  

Please refer to ‘③ ‘③   Imaging Environment Configuration’ Configuration’  of 7.10.1. RIS Code (p. (p. 131)  131)  

DRT-MAN-055 (Rev (Rev.. 06)

142

 

 

7.12.  Generator  

Set the type of X-ray Generator

 

Set detailed options for each type of X-ray Generator Table 16. Generator Configuration Categories

Subcategories

Details

Sedecal

Sedecal Generator

DRGEM

DRGEM Generator

POSKOM

POSKOM Generator

DRT-MAN-055 (Rev (Rev.. 06)

143

 

 

7.12.1.   

Sedecal

Setting regarding the Sedecal Generator

Figure 84. Sedecal Configuratio Configuration n Screen  

‘Use This Generator’ Generator ’ check box  

 

Check this category in order to use this Generator

Enter the configuration value for the use of Sedecal Generator

DRT-MAN-055 (Rev (Rev.. 06)

144

 

 

7.12.2.   

DRGEM

For use of the DRGEM Generator

Figure 85. DRGEM Configuration Screen  

‘Use This Generator’ Generator ’ check box  

 

Check this category in order to use this Generator

Enter the configuration value for the use of DRGEM Generator

DRT-MAN-055 (Rev (Rev.. 06)

145

 

 

7.12.3.   

POSKOM

For use of the POSKOM Generator

Figure 86. POSKOM Configurati Configuration on Screen  

‘Use This Generator’ Generator ’ check box  

 

Check this category in order to use this Generator

Enter the configuration value for the use of POSKOM Generator

DRT-MAN-055 (Rev (Rev.. 06)

146

 

 

Appendix A. Glossary Terms

Description Picture Archiving and Communication System.

PACS

System for the storage, interpreting and access of medical imaging information

DICOM

Digital Imaging and Communication in Medicine. Standard medical digital image and communication DICOM Directory.

DICOMDIR

DICOM directory information file stored in the root folder of the media (CD/DVD) Hospital Information System.

HIS

Hospital information system, registration/administrative system, including OCS, PACS and etc.

OCS RIS

Order Communication System Radiology Information System. Operated as part of HIS or independently Modality Worklist.

MWL

DICOM standards for scheduled examination information sharing Computer Unified Device Architecture. Architecture.

CUDA

High sped parallel processing architecture utilizing a GPU developed by NVIDIA

GPU

Graphic Processing Unit. Graphical User Interface.

GUI

Computer interface for the user, developed by graphic elements

Detector

Flat Panel Digital X-ray Detector Detector..

Compiler

Program which converts source codes in high level language in to specific machine codes

DRT-MAN-055 (Rev (Rev.. 06)

147

 

  Terms Build

API

Description Process where the source code is being converted for computer operation Application Programming Interface. Interface developed for use of specific functions in the application program

Installer Login Coding Convention Code Review Code Coverage Static Code Analysis

Program utilized for the installation, maintenance and removal of a specific application program Process authenticating the user for permission to access the system Style guideline for creating a source code Process where developers inspect and verify the created/edited created/edit ed source code Tested number of code components/Total number of code components * 100 Process of discovering potential defects of the program without operating it by analyzing the source code

Cyclomatic

Digitized level of complexity of the program’ program’s source

Complexity

code Continuous Integration.

CI

Automated integration process of the source code, and the process of improvement and receiving feedback through periodic build, test analysis

Encryption Protection (AntiCrack) SCP

Encrypting the data Protection preventing program hacking through Reverse Engineering Service Contents Provider Text and various shapes marked on the completed

Annotation

image

DRT-MAN-055 (Rev (Rev.. 06)

148

 

  Terms LUT ROI Exception Calibration Map File Pixel Map

Description Lookup Table. Table of previously calculated values Region of Interest Actual region necessary from the completed image Exception subsequent to program error, wrong memory reference or stack overflow Task calibrating the quality of the detector image Generic term for data files used for the calibration of detector image quality Data file used to calibrate bad pixels in detectors Data file used to calibrate the property of the detector

Gain Map Panel Map

subsequent to installation environment Data file used to calibrate the property of the detector panel itself

DRT-MAN-055 (Rev (Rev.. 06)

149

 

 

Appendix B. Terminal Emulator Usage of terminal emulator may differ from each detector, please refer to the Detector Service Manual.

Terminal Mode Terminal Emulator Start-up Sequence: 1.  Turn the detector power OFF.  

Not necessary for F600.

2.  Click ‘Run Run’’ button on Configuration - Detector - General tab.

3.  Turn the detector power ON.

4.  When power is turned on, Terminal Emulator tries to connect to

DRT-MAN-055 (Rev (Rev.. 06)

150

 

  detector automatically.  

If not, please click ‘Connect Connect’’ button.

As the Ethernet Terminal Emulator features auto-connect function, it will enter Terminal Mode automatically.

The TERMINAL MODE MENU has 3 options :

USER MODE

Can be accessed by the user.

OPERATOR OPERAT OR MODE

Can be accessed only by the operator.

EXIT TERMINAL MODE

Closes terminal mode.

DRT-MAN-055 (Rev (Rev.. 06)

151

 

 

User Mode To activate User Mode, enter “1” in the Terminal Mode menu screen.  screen.  

Cautions: Number keys on numeric keypad in right-side don’ don ’t work well. Please use number keys in left-side.

DRT-MAN-055 (Rev (Rev.. 06)

152

 

  User Mode has 8 options as below

Check the status of the internal module 1. EXECUTES SELF-TEST in the detector 2. PROGRAM READY DELAY

Check or change the Ready Delay value.

3. PROGRAM EXPOSURE

Check or change the Exposure Window

WINDOW TIME

Time value.

4. SELECT EXPOSURE

Activate/deactivate the Exposure

PROTECTION MODE

Protection Protectio n Mode.

5. SELECT ADAPTIVE ERASE

Activate/deactivate the Adaptive Erase

MODE

Mode.

P. PR PREVIOUS EVIOUS MENU

Return to the previous menu.

?. PRINT MENU

Display all the options for User Mode.

X. EXIT TERMINAL MODE

Exit terminal mode.

DRT-MAN-055 (Rev (Rev.. 06)

153

 

  1. EXECUTES SELF-TEST To run the SELF-TEST , press” press”1 1” in User Mode. If SELF-TEST   is successfully completed, all the test results become changed to “OK! OK!””.

 

If your detector has a problem with the operating temperature, you will get a “Failed”  “Failed”  answer for the current operating temperature. In this case, you are advised to shut down the power to the detector and allow a cooling time at least more than one hour. If you still get the same error message after one hour of cooling, then report your claim to the DRTECH Customer Support Team [email protected]). ). ([email protected]

 

If your detector has a problem with the "Erase Light" operation, you will get a “Failed” answer. In this case, report your claim using our claim form, CDN(Customer Dissatisfaction Notice / refer to the last page in this Service Manual) to [email protected] [email protected]   immediately.

Your detector should be returned to DRTECH for repair.

DRT-MAN-055 (Rev (Rev.. 06)

154

 

   

If your detector has a problem with Read/Write Memory operation, you will get a “Failed”  “Failed”   answer. In this case, report your claim using CDN to [email protected] immediately.

   

Your detector should be returned to DRTECH for repair. If your detector has a problem with the "Data Readout Sequence" operation, you will get a “Failed” answer. In this case, report your claim using CDN to [email protected] [email protected]   immediately. Your detector should be returned to DRTECH for repair.

 

In all other cases, contact [email protected] [email protected] for  for technical support.

DRT-MAN-055 (Rev (Rev.. 06)

155

 

  2. PROGRAM READY DELAY To change the Ready Delay  value, press “2” in User Mode.  Mode.  The current Ready Delay value and a message for setting a new Ready Delay value are displayed.

 

You can enter between 0 ~ 3000ms, the default value is 2000ms.

DRT-MAN-055 (Rev (Rev.. 06)

156

 

  To change the Ready Delay value, enter the new value you want and new Ready Delay value will be displayed. To keep existing setting just, press “Enter.”  “Enter.”   

To return to the previous menu, press “P”.

 

To see User menu, press” press”?”.

DRT-MAN-055 (Rev (Rev.. 06)

157

 

  3. PROGRAM EXPOSURE WINDOW TIME To change the Exposure Window Time  value, press “3 “3” in User Mode.  Mode.  The current Exposure Window Time value and a message for setting a new Exposure Window Time value are displayed.

 

You can enter between 0 ~ 3000ms and default value is 500ms.

DRT-MAN-055 (Rev (Rev.. 06)

158

 

  To change the Exposure Window Time value, enter the new value you want and new Exposure Window Time value is displayed. To keep the existing setting, press just “Enter.” “Enter.”  

   

DRT-MAN-055 (Rev (Rev.. 06)

159

 

  4. SELECT EXPOSURE E XPOSURE PRO PROTECTION TECTION MODE To change the Exposure Protection Mode   setting, press “4” in User Mode. The current state of Exposure Protection Mode and a message for setting a new value are displayed.

DRT-MAN-055 (Rev (Rev.. 06)

160

 

  To make the Exposure Protection Mode enable, press “1,” or disable, press “2.”  “2.”  To keep the existing setting, press just “Enter.”  “Enter.”   The mode is changed successfully, and the current state will appear.

   

DRT-MAN-055 (Rev (Rev.. 06)

161

 

  5. SELECT ADAPTIVE ERASE MODE To change the Adaptive Erase Mode  setting, press “5” in User Mode.  Mode.   The current state of Adaptive Erase Mode and a message for setting a new value are displayed.

DRT-MAN-055 (Rev (Rev.. 06)

162

 

  To make the Adaptive Erase Mode enable, press “1,” or disable, press “2.”   “2.” To keep the existing setting, press just “Enter.”  “Enter.”   The mode is changed successfully, and the current state will appear.

   

menu, press “P”.  “P”. 

DRT-MAN-055 (Rev (Rev.. 06)

163

 

  X. End of Terminal Mode To exit the Terminal Mode, press “X” in the Terminal Mode menu or User Mode. After pressing “X,”  “X,”  To confirm the exit, press “Y.”  “Y.”   To cancel the exit, press “N” to return to the previous pre vious menu.

DRT-MAN-055 (Rev (Rev.. 06)

164

 

  When pressing “Y” to confirm the termination of Terminal mode, the Self-Test function of the detector will be executed automatically and the results will be displayed.

Please check the detector operates normally.

Operator Mode Operator mode is used for the detailed setting at the time of producing detector for the first time, and can be accessed by only an authorized operator.

DRT-MAN-055 (Rev (Rev.. 06)

165

 

 

Appendix C. Update If current using version has bugs please contact to CS team, CS team will give you a update(patch) installer. If you receive the update installer file please run the provided setup file.

 

Please back up original Econsole1 folder(ex: C:\Econsole1) in order to prevent a fatal error.

If the setup file is run, you will be present presented ed with the following screen. Click ‘Next Next’’ to start the installation installation..

DRT-MAN-055 (Rev (Rev.. 06)

166

 

 

Select license protection type you purchased.  

Hardware License Protection (Dongle Key)  

Provides special USB, and you can run EConsole1 only when the USB connected.

 

Software License Protection  

Provides the license on the special PC.

DRT-MAN-055 (Rev (Rev.. 06)

167

 

  Do not change installation folder and click 'Next'. It will update the program automatically and installation is finished please run the Econsole1 and check the version.

DRT-MAN-055 (Rev (Rev.. 06)

168

 

 

Appendix D. Portable function Some of wireless detector is a function to save a x-ray image. If user shot a x-ray without software the image will be saved in the detector . The saved image control function is called "Portable function".

Figure 87. Portable function summary

Econsole1 Vet's portable has 2 functions.  

Download the image saved in detector

 

Register the exam order to detector

1) Download the image saved in detector. Proceed in the followin following g order ①  Select a exam order and go to acquisition.

DRT-MAN-055 (Rev (Rev.. 06)

169

 

 

②  Click  

icon in the ToolBar

If it is not in the toolbar, click the 'Edit' and add it.

③  Save portable image pop-up comes up.   Select a image in the Detector.  

Click 'OK'.

Detailed function 'Save portable image' pop-up window is described in the below.

Figure 88. Save portable image ①  Select a image saved detector. ②  Show selected detector's saved image's file name. ③  Selected image's preview image. ④  Function button and UI.  

Refresh

   

Refresh a image list

Delete

DRT-MAN-055 (Rev (Rev.. 06)

170

 

     

Delete all  

 

 

Delete selected image in detector. Delete all images selected detector.

Used(%d)   Selected detector's used storage. OK  

Get selected image and show it to Econsole1 Vet's image viewer viewer..

 

Cancel  

Close 'Save portable image' window.

2) Register the exam order to detector It is the function that save exam orders to detector and shot a image without PC depending on the given exam list. Save exam order function description is below.

①  Click

button.

②  'Save image list' window comes up.      

Select a detector. Select List in right side and click '
View more...