DS-451(Ti-1232) Operation Manual Digit)

January 18, 2017 | Author: Sarwar Alam | Category: N/A
Share Embed Donate


Short Description

Download DS-451(Ti-1232) Operation Manual Digit)...

Description

DS451 (TI-1232) Scale key operation detail KEY OPERATIONS : •

S-on Mode : Press [→0 ←] + [→T←] [↑] [←] keys.



Technical specification Mode: Enable S-on mode, Press [→0 ←] + [←] [→T←] [←] keys.



Customer specification Mode: Press [→0 ←] + [←] [←] [←] keys.



Count mode: Enable S-on mode, Press [→0 ←] +[←] [←][→T←] keys



Calibration mode: Enable S-on mode, Press [→0 ←] + [←] [→T←] [→T←] keys.



Set point Programming mode: Press [→0 ←] + [↑] [↑] [↑] keys.

Entering W&M specifications S-on mode to be enabled to enter W&M Specifications (Technical Specification) Sl.No

1

2

OPERATION SEQUENCE Go to S-On mode. To Enter S-on mode Press [→0 ←] + [→T←] [↑] [←]

Press [→0 ←] +[←] [→T←] [←]

3

DISPLAY S – on

T10 XX T10

[←] [↑] 00 T11

4 Press [∗] Key 5

XX T11

[←] [↑] 01 6

[→T←]

S - on

REMARKS Press [Rezero] key and Press [Tare],[Digit Increment] [Digit Select] Keys Depress Press[REZERO] key and press [DIG-SEL] key, [TARE] key & [DIG-SEL] key in sequence. The display toggles between the messages Spec No and the current value Using [DIG-SEL] and [DIG-INC] key Enter the required value Press * key to store spec value entered and advance to next Spec Follow the procedure mentioned in SL.No 3 & 4. and Repeat the same Up to T19 Press [TARE] key, to exit from Technical specification mode

Entering Customer specifications Sl.No

1

OPERATION SEQUENCE Press

[→0 ←] + [←][←][←] 2

DISPLAY

C10 XX

C10 [←] [↑]

4 Press [∗] key 5

02 C11 XX C11

[←] [↑] 6

11 0.000

[→T←]

REMARKS Depress [Rezero] key and Press [Digit Select] Key 3times. The display toggles between the message Spec No and the Current Value. Using [DIG-SEL] and [DIG-INC] key Enter the required value Press * key to store spec value entered and advance to next Spec Follow the procedure mentioned in SL.No 2 & 3. and Repeat the same up to C19 Press [TARE] key, to exit from Technical specification mode

Internal count access This function is provided for checking of internal count of A/D Sl.No

1

2

OPERATION SEQUENCE Go to S-On mode. To Enter S-on mode Press [→0 ←] + [→T←] [↑] [←]

Press [→0 ←] + [←][←][→T←]

DISPLAY S – on

XXXXXX

3 Press [→T←] Key

0.000

REMARKS Press [Rezero] key and Press [Tare], [Digit Increment] [Digit Select] Keys

Depress [REZERO] key and press [DIG-SEL] key twice, [TARE] key once in sequence. The display indicates the Span count. Press [TARE] key, to exit from Count mode and switch for Weighing mode.

Calibration procedure : Calibration using full scale machine capacity / 1/3rd of Machine Capacity Sl.No

1

2

3

OPERATION SEQUENCE Go to S-On mode. To Enter S-on mode Press [→0 ←] + [→T←] [↑] [←]

Press [→0 ←] + [←] [→T←] [→T←]

[←] [↑]

DISPLAY S – on

FS XXXX.XXX

FS 0015.000 CAL 00

4 Press [∗] Key 5 Press [∗] Key 6

XXXXXX CAL SP XXXXXX 600000

Press [∗] Key 7 Press [→T←] Key

15.000

REMARKS Press [Rezero] key and Press [Tare], [Digit Increment] [Digit Select] Keys

Depress [REZERO] key and press [DIG-SEL] key, [TARE] key twice in sequence. The display toggles between the messages FS and the current value / Capacity Using [DIG-SEL] and [DIG-INC] key Enter the required Capcity / value Press * key to accept Capacity entered and to go to next step Press * key to accept the Zero Calibrated value and to go to next step. Place the entered Capacity weight on platter. Press * key to accept span Calibration and display will read span count. Press [TARE] key, to exit from Calibration mode and switch for Weighing mode.

Note : During calibration process, Press [TARE] key to exit from calibration mode.

Digital Tare Programming: Sl.No

1

OPERATION SEQUENCE Go to weighing mode.

DISPLAY

0.000

2 Press [↑] key

Tare XXXX.XXX

3 [←] [↑]

1.000

4

Press [→T←]

-1.000

5

Press [→T←]

REMARKS

0.000

Press [DIG INC] key. The display toggles between the message Tare and current value Using [DIG-SEL] and [DIG-INC] key Enter the required value Press [TARE] key, to Tare weight reading entered

Press [Tare] key to exit from the mode / switch for Weighing more

1.CUSTOMER SPECIFICATION: SPEC C10

C11

C12

MOST SIGNIFICANT DIGIT(MSD) RS232/RS485 DATA BITS & PARITY: 1-> 8 Data Bits & NO Parity 2 -> 7 Data Bits & NO Parity 3 -> 8 Data Bits & ODD Parity 4 -> 7 Data Bits & ODD Parity 5 -> 8 Data Bits & EVEN Parity 6 -> 7 Data Bits & EVEN Parity Note 3 RS232/RS485 Weight Data Transfer: 1 -> All 2 -> Net only 3 -> Gross only

RS232/RS485 Weight Header: 1 -> No

2 -> Numeric C13

C14

C15

C16

3 -> Alpha RS232 One Touch Tare from PC On Stable(T/t): 0 -> No 1 -> Yes RS232 Stream Data Transfer Mode Stable Status Flag: 0 -> No 1 -> Yes Setpoint Output Type: 1 > Active Low 2 -> Active High Negative Setpoint Output: 0 -> No

1 -> Yes C17

No. of CR & LFs after Total Print: 0 -> FORM FEED

1 -> 1 CR LF ~ 9 -> 9 CR LF C18

C19

RS485 Weighing Scale ID: 00 -> Reserved 01 – 99 Not used

LEAST SIGNIFICANT DIGIT(LSD) RS232/RS485 Baud Rate (Stop bit =1): 1 -> 1200 2 -> 2400 3 -> 4800 4 -> 9600 5 -> 19200 Note 3 RS232 Data Transfer Mode: 1 -> Stream 2 -> Manual 3 -> Command (CTRL E) 4 -> Auto Weight Transfer 5 -> Postal(Meghdooth) 6 -> Milkotronics (W1 to W9) 7 -> Manual Serial Print 8 -> Auto Serial Print RS232/RS485 Transfer Stable Weight/Manual Serial Print: 0 -> No 1 -> Yes RS232/RS485 Data : 0 ->Variable 1 -> Fixed ( 8 characters) Skip “STX” Character In RS232 Output: 0 -> No 1 -> Yes Setpoint Output Logic: 1 -> Logic I (Bar Mode) 2 -> Logic II (Dot Mode) VFD Standby Mode (In Battery): 0 -> No 1 -> 5 Minute 2 -> 10 Minute Serial Printout Format: 0 -> 80 col All Wt. 1 -> 40 col All Wt. 2 -> Net Wt only.

Not used

2.TECHNICAL SPECIFICATION: SPEC T10

MOST SIGNIFICANT DIGIT(MSD) Dual Accuracy (Upto 50% of Capacity): 0 - No 1 - Yes

DISPLAY RESOLUTION SHOULD BE LESS THAN OR EQUAL TO 1/15000 & SPEC T11 IS AMONG 02, 05, 10, 20 AND 50 VALUES for dual range.

T11

Accuracy (2 digits):

T12

Capacity:

T13

Power Fail Return: 0 -> No 1 -> Yes

LEAST SIGNIFICANT DIGIT(LSD) Decimal Point Position: 0 -> No Decimal point 1 -> 1 DP 2 -> 2 DP 3 -> 3 DP 4 -> 4 DP

Tare Limit: 1 -> Capacity 2 -> 1/2Capacity – 1 100% TARE LIMIT IS NOT POSSIBLE IF DUAL RANGE IS WORKING.

T14

Power On Start Range: 1 -> 25% of capacity 2 -> Unlimited

T15

RS232 Minimum Weight Transfer Limit: 0 -> No limit 1 -> Net >/ 1d 2 -> Net >/ 5d 3 -> Net >/ 20d S-On Through Key Board: 0 -> Yes 1 -> No

T16

T17

ADC Gain: 1 -> UNITY 2 -> Double Note 3

T18

Not used

Forced Zero Tracking: 0 -> ¼ d 1 -> 1d 2 -> 2 d 2 -> 2d 3 -> 3d 4 -> 4d 5 -> 5d Serial Interface Option: 0 -> No 1 -> RS232 2 -> RS485 Note 3 Second Serial Interface Option: 0 -> No 1 -> RS232 Note 3 Unit Conversion: 0 -> No 1 -> Carat ( 5.0000 ) 2 -> Tola ( 0.085734) 3 -> Litre (Cow Milk) ( 0.9724 ) 4 -> Litre (Soya Oil) (1.0989 ) 5 -> Milkotronics Litre constant Range (0.9000 to 1.0000) Not used

T19

Stable Indicator On: 0 - No 1 - Yes

Display Brightness Level: 3 -> 29% Blue LED 4 -> 71% Red LED 5 -> 77% 6 -> 86% VFD 7 -> 93% 8 -> 100% Note 3

Note 1 : Default is indicated by Highlighting for applicable specs. Note 2 : If a Spec is set to a value other than specified, then default function is used. Note 3 : The Specs will take effect only at power on. So on/off is required whenever changed.