RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS  ® 

Fabius

Repair Instructions

6

Debug Mode (software 1.00)

The following information is displayed displayed in the debug mode dur during ing normal operation of Fabius: system data, measured values, error log. One debug mode page is available. available. System data cannot be changed.

6.1 6.1

Ac Acce cess ssin ing g tthe he De Debu bug gM Mo ode (so soft ftwa ware re 1. 1.00 00))

The debug mode can be accessed from any operating mode. •

Press the “Menu” key.



Select “keycode” by turning rotary rotar y knob.



Confirm selection by pressing rotary knob.

The selected numeric code is displayed on black background. •

Select keycode “1” by turning rotary rotar y knob one position clockwise.



Confirm selection by pressing rotary knob.



Press rotary knob again.

An additional screen page is displayed. displayed. Fabius remains in the normal mode and can be operated as desired.

6.2 6.2

Ex Exit itin ing g the the De Deb bug Mo Mod de (s (so oft ftwa warre 1. 1.0 00)



Press the “Menu” key.



Select “keycode” by turning rotary rotar y knob.



Confirm selection by pressing rotary knob.

The selected numeric code is displayed on black background. •

Select keycode “0” by turning rotary rotar y knob one position counter-clockwise.



Confirm selection by pressing rotary knob.



Press rotary knob again.

Fabius is now in the normal operating mode.

 3-8

 .   c   n    I  ,    l   a   c    i    d   e    M   r   e   g   e   a     r   m    D    f  .   c    0 _   c    0   u   r    0    t   s    2   n    t    i   r    h    i   a   g   p    i   r   e   r   y _    3   p   c   o   e    C   s

 

RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS

 ® 

Fabius

Repair Instructions

6 .3

Debug M Mo ode TTa able (software 1 1..00)

Table 3: Debug Debug Mode (software (software 1.00)

Plateau pressure in mbar

Pressure sensor value (Paw) in mbar (default) P

 .   c   n    I  ,    l   a   c    i    d   e    M   r   e   g   e     a   r   m    f  .    D   c    0 _   c    0   u    0   r    t   s    2   n    t    i   r    h    i   a   g   p    i   e   r   r _   y    3   p   c   o   e   s    C

Plat XXX (varies with system settings)

XXX (0 to 738)

Flow sensor A/D converter value (default)

Supply voltage (approx. 27 V) in volts volts

FL

U_24

XXX (0 L/min 512)

Oxygen sensor channel 1 in mV

Supply voltage voltage (+12 (+12 V) in volts volts

O 2 _1

U+12

XXX (9 – 20)

XX.X (12.0 ±2.0)

Oxygen sensor channel 2 in mV

Supply voltage (–12 V) iin n volt volts s

O2_2

U–12

XXX (9 – 20)

XX.X (12.0 ± 2.0)

Expiratory Vt during inspiration (mL)

EEPROM error code

Vtix XXX (varies with system settings, ideally 0)

EEPR

Pressure sensor calibration value

Battery charging current in amperes

C al P

LADI

XXX (704 – 768)

XXXX

X.X (approx. 0.0 – 2.0)

Flow sensor calibration value

Motor current in amperes

CalF

MOTI

XXX (482 – 542)

X.X (0.0 – 2.5)

Oxygen sensor sensor calibration calibration value value (21% O2)

Pressure between rolling seals in mbar

CalO

PMEM

XX (39 – 113)

XXX (200 ±50)

Monitoring error code

Ventilation error

MONI

VENT

XXXX

Calculation from controller data (set values): Tidal volume (mL)

frequency (1/min)

V:

F:

XXXX

XX

LOG XX XX XX XX XX XX XX XX XX

3-9

XXXX

System temperature in °C

T

Error log (most recent error entry on o n extreme left)

 

XX.X (22.5 – 35.0)

XX (max. 60°C)

 

RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS  ® 

Fabius

Repair Instructions

7

Debug Mode (software 2.00)

The following information is displayed displayed in the debug mode dur during ing normal operation of Fabius: system data, measured values, error log. Two pages are available in this debug mode. System data cannot be changed.

7.1 7.1

Ac Acce cess ssin ing g tthe he De Debu bug gM Mo ode (so soft ftwa ware re 2. 2.00 00))

The debug mode can be accessed from any operating mode. •

Press the “Menu” key.



Select “keycode” by turning rotary rotar y knob.



Confirm selection by pressing rotary knob.

The selected numeric code is displayed on black background. •

Select keycode “1” (debug mode) or “2” (flow curve) by turning rotary knob one position clockwise.



Confirm selection by pressing rotary knob.



Press rotary knob again.

An additional screen page is displayed. Fabius Fabius remains in the normal nor mal operating mode and can be operated as desired.

7.2 7.2

Ex Exit itin ing g the the De Deb bug Mo Mod de (s (so oft ftwa warre 2. 2.0 00)



Press the “Menu” key.



Select “keycode” by turning rotary rotar y knob.



Confirm selection by pressing rotary knob.

The selected numeric code is displayed on black background. •

Select keycode “0” by turning rotary rotar y knob one position counter-clockwise.



Confirm selection by pressing rotary knob.



Press rotary knob again.

Fabius is now in the normal operating mode.

 3-10

 .   c   n    I  ,    l   a   c    i    d   e    M   r   e   g   e     a   r   m    f  .    D   c    0 _   c    0   u   r    0    t   s    2   n    t    i   r    h    i   a   g   p    i   r   e   r   y _    3   p   c   o   e    C   s

 

RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS

 ® 

Fabius

Repair Instructions

7 .3

Debug M Mo ode TTa able (software 2 2..00)

7.3.1 7.3 .1

Debug Debug Mod Mode e (soft (softwa ware re 2 2.00 .00), ), Page Page 1 (key (keyco code de "1" "1"))

Error log (most recent error entry e ntry on extreme left) LOG XX XX XX XX XX XX XX XX XX XX XX XX XX XX Motor current in amperes A/D pressure sensor value in mbar/ (normal operating range 0 to 1.5 A) calibration value (default) in mbar P XXX (65 (650 0 to 825) / XXX XXX (738 = 0hPa) 0hPa)

Imot

X.XX 

Flow sensor A/D converter value / calibration value (default)

Pressure between the rolling seals in hPa (normal operating range 120 to 250 250 hP hPa) a)

F1 XXX (482 to 542) / XXX (0 L/min L/min = 512) Pvac

Oxygen sensor channel 1 in mV Oxygen sensor channel 2 in mV

System temperature in °C (normal operating range 10 to 70   °C)

(normal operating O2_ XXX / XXX XXX range 5mV to 105 mV)

Tem XX

Oxygen sensor calibration status Oxygen sensor sensor calibration calibration value value (21% O2) (Calibration value in mV

Frequency: calculated from piston movement (1/min) (normal operating range 6 to 60 1/min)

CAL

f_v:

nnnn (XXX) /

Elapsed time meter (full hours displayed)

 .   c   n    I  ,    l   a   c    i    d   e    M   r   e   g   e     a   r   m    f  .    D   c _    0   c    0   u    0   r    t   s    2   n    t    i   r    h    i   a   g   p    i   e   r   r _   y    3   p   c   o   e   s    C

XXX

XX

h XXXX XXXXXX XXXX XXX X

Tidal volume calculated from piston movement (normal operati operating ng range 50 to 1400 mL) V_v: XXXX

Supply voltage voltage (approx. 27 V) in volts volts (normal operating range 23.0 V to 35.0 V) U+24 XX.X 

Motor type, jumper position: Power PCB (motor type: JC = Mavillor, Mavillor, J8 = GEC) XXRXX JC

Supply voltag voltage e (+12 V) in volts volts (normal operating range 10.8 V to 13.2 V) U+12 XX.X (1 (12. 2.0 0 V ±2.0 ±2.0 V)

Monitoring error code

Supply voltage (–12 V) in vo volts lts (normal operating range –10.8 V to –13.2 V)

Ventilator error code

U–12

VEN

XX.X

Battery charging current in amperes (normal operating range –2 to to 3 A) Ibat X.X

MON

XXXX

XXXX

EEPROM error code EEP

XXXX

Control PCB software version X X M X X J6 Rev. A  

3-11

 

RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS  ® 

Fabius

Repair Instructions

7.3.2 7.3 .2

Flow Flow Curv Curve e (sof (softwa tware re 2. 2.00 00), ), Page Page 2 (Ke (Keyc ycode ode “2”)

Flow curve

Measured tidal volume in mL (normal operati operating ng range: 0 to 9999 9999 mL)

(Quit key: toggles between 1:1 and 1:6 resolution)

Vte Vte

XXXX XXXX mL

Expiratory VT during inspiration (in mL, ideally: 0) (normal operating range: 0 to 9999 mL) Vtix

XXXX mL

Expiratory peak flow in L/min (normal operatin operating g range: 0 to 150 L/min) L/min) Flep XXX L/min

Tidal volume corresponding to ventilator’s I/E phase (in mL) (normal operating range: 0 to 9999 mL) Vtev

XXXX mL

Plateau pressure in hPa (normal operati operating ng range 0 to 99 hPa) hPa) Plat

XX hPa

 .   c   n    I  ,    l   a   c    i    d   e    M   r   e   g   e     a   r   m    f  .    D   c    0 _   c    0   u   r    0    t   s    2   n    t    i   r    h    i   a   g   p    i   r   e   r   y _    3   p   c   o   e    C   s

 3-12

 

RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS

Fabius ® 

Repair Instructions

8

"Ventilator" (V (VEN) D Diisplay FFiield

Overview: VEN

0000 ==> no errors –

0001  0008 ==> motor errors 0040 and 0080 ==> pump errors 0100 ==> temperature error 1000 and 2000 ==> communication errors 4000 ==> current error (motor current is too high)

Error

Error Description

Possible Cause

0001

No incrementing.

Motor open or not connected. Incremental encoder not connected.or needs replacement

0002

End position switch activated continuously.

Light barrier not connected.or needs replaced

0004

End En d posi positi tion on swi switc tch h not not det detec ecte ted. d.

Exte Externa rnall li ligh ghtt inte interf rfer erenc ence, e, or lig light ht barrier needs replaced.

0008

Motor direction incorrect (use only upon power-on of Fabius).

Motor conductors reversed. Phase conductors of incremental encoder reversed.

0010

24 V supply not available.

Controller relay (Power PCB) switched off. Fuse F4 on Power Supply PCB open.

Diaphragm pressure too low.

Ventilator cover open.

0020 0040

Vacuum hose not connected. Diaphragm pump needs replaced

 .   c   n    I  ,    l   a   c    i    d   e    M   r   e   g   e     a   r   m    f  .    D   c _    0   c    0   u    0   r    t   s    2   n    t    i   r    h    i   a   g   p    i   e   r   r _   y    3   p   c   o   e   s    C

0080

Diaphragm pres ess sure too high gh..

Diaphr hra agm pressure sen ens sor connector not connected.

0100

Temperature too high.

NTC error on Power PCB. Fan not running

0200

I/E signal from controller (Power PCB) not in expected range.

Ventilation frequency or inspiratory time out of tolerance. Motor and/or motor control error.

0400

Ventilator piston does not no t work in IPPV mode.

Motor open or not connected. Power Pow er PCB needs replaced

0800

Data error (gain, offset, from PEEP valve)

EEPROM or RAM error. New PEEP valve calibration.

Rev. A  

3-13

 

RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS  ® 

Fabius

Repair Instructions

1000

Communication error from Control Contr ol PCB to Powe Powerr PCB.

Line interruption in ribbon cable, bent connector. connector.

2000

Communication error from Power Power PCB to to Control Control PCB.

Line interruption. Power PCB does not work (see error 1000).

4000

Motor Mot or curren currentt temp tempora orarily rily too high. high.

Increm Increment ental al encode encoderr llost ost increments. Contamination, scratches, light barrier needs replaced

8000

9

Motor current current contin continuousl uously y too too hi high. gh.

Power Power amplifier amplifier needs needs replaced. replaced. Motor open.

"Monitor" (MON) Display Field

Overview: MON

0000 ==> no errors 00XX ==> O2 sensor error 0X00 ==> pressure sensor error X000 ==> flow sensor error

Error

Error Description

Possible Cause

0001

 FiO2 se  sensor voltage too low.

O2 sensor needs replaced. O2 sensor calibrated incorrectly. incorrectly.

0002

 FiO2 s  se ensor voltage too high.

O2 sensor needs replaced. O2 sensor calibrated incorrectly. incorrectly. O2 sensor needs replaced

0004

 FiO2 measured values outside permissible range.

0008

FiO2 ti  time-out during calibration.

O2 sensor needs replaced. Sensor not yet stabilized. Calibration of O2 sensor at changing O2 concentration or O2 concentration not corresponding to calibration mode (21%/100%).

0010

FiO2 /relative error between between FiO2  sensor channels > 25%.

O2 sensor needs replaced.

0020

FiO2 calibration constant from

O2 sensor not yet calibrated.

0040

EEPROM incorrect. FiO2 /coding for calibration. calibration.

Data loss in the EEPROM. Current error occurred during calibration.  3-14

 .   c   n    I  ,    l   a   c    i    d   e    M   r   e   g   e     a   r   m    D    f  .   c    0 _   c    0   u   r    0    t   s    2   n    t    i   r    h    i   a   g   p    i   r   e   r   y _    3   p   c   o   e    C   s

 

RETURN TO THIS MANUAL'S TABLE OF CONTENTS RETURN TO CD-ROM TABLE OF CONTENTS

 ® 

Fabius

Repair Instructions

0080

One calibration value lost during operation (pressure sensor, flow sensor, FiO2 sensor calibration values).

RAM error. Electromagnetic interference.

0100

Pressure above maximum value for more than 5 seconds.

Pressure sensor needs replaced. Continuously high pressure.

0200

Pressure below minimum value for more than 5 seconds.

Pressure sensor needs replaced. Continuously low pressure.

0400

Pressure/calibration error.

Pressure sensor needs replaced. Pressure during calibration higher than or lower lower than than 0 mbar. mbar.

0800

Pressure/plausibility error during respiratory phase.

Software error.

1000

Flow above maximum value for more than than 5 seconds seconds (
View more...

Comments