CORAL LABORATORIES LIMITED Plot No. 27-28, Pharmacity Selaqui Industrial Area Dehradun – 248197 Uttarakhand, India
PROTOCOL FOR CALIBRATION OF JACKETED MANUFACTURING TANK
Equipment Name
Jacketed Manufacturing Tank
Reason for Qualification
Periodic Qualification
Equipment ID No.
LV02
Document No.
QP-PPQ-LV02-02
Supersede No.
Nil
Calibration Frequency
Yearly
Effective Date
PROTOCOL FOR CALIBRATION OF MANUFACTURING TANK Coral Laboratories Ltd. Location: Dehradun
Equipment Name: Jacketed Manufacturing Tank
Equipment ID No.: LV02
Document No.: QP-PPQ-LV02-02
Page No.: 2 of 7
1.0 PROTOCOL APPROVAL: 1.1 This protocol is prepared by validation task force member: Department
Quality Assurance
Sign Name
Rajvar
Designation
Chemist
Date
1.2 This protocol is reviewed by validation core committee: Quality Assurance
Production
Quality Control
Engineering & Maintenance
Name
Puja Aswal
Nilesh Patel
Rajesh Suthar
Nand Kishor
Designation
Executive
Manager
Manager
Incharge Utility
Department
Sign
Date 1.3 This protocol is approved for implementation by Head-QA:
Sign Name
Sunil Kumar Gupta
Designation
Manager
Date 1.4 This protocol is authorized for implementation by Plant -Head:
Sign Name
B.H. Patel
Designation
Plant Head
Date
PROTOCOL FOR CALIBRATION OF MANUFACTURING TANK Coral Laboratories Ltd. Location: Dehradun
Equipment Name: Jacketed Manufacturing Tank
Equipment ID No.: LV02
Document No.: QP-PPQ-LV02-02
Page No.: 3 of 7
2.0 TABLE OF CONTENT: SR. NO.
CONTENT
PAGE NO.
1.0
PROTOCOL APPROVAL
2
2.0
TABLE OF CONTENT
3
3.0
OBJECTIVE
4
4.0
SCOPE
4
5.0
RESPONSIBILITIES
4
6.0
DOCUMENTATION
5
7.0
MATERIAL REQUIREMENT FOR CALIBRATION
5
8.0
PROCEDURE
5
9.0
OBSERVATION
7
10.0
CONCLUSION
7
PROTOCOL FOR CALIBRATION OF MANUFACTURING TANK Coral Laboratories Ltd. Location: Dehradun
Equipment Name: Jacketed Manufacturing Tank
Equipment ID No.: LV02
Document No.: QP-PPQ-LV02-02
Page No.: 4 of 7
3.0 OBJECTIVE: To establish evidence that the manufacturing tank is calibrated for the volume with dip rod marking, equipment ID No-LV02 is acceptably verified in liquid manufacturing area at CORAL LABORATORIES LTD, Dehradun. 4.0 SCOPE: This is applicable at the time of calibration of manufacturing tank ID No-LV02 Liquid manufacturing area at CORAL LABORATORIES LTD, Dehradun. The details are given below: Sr. No.
Equipment Name
Equipment ID
Parameter
1
Jacketed Manufacturing Tank
LV02
Calibration
5.0 RESPONSIBILITIES: 5.1 RESPONSIBILITIES : 5.2.1 Quality Assurance: 5.2.1.1 Preparation and review of Protocol and report. 5.2.1.2 Execute the protocol as per procedure. 5.2.1.3 Approval of Protocol and report. 5.2.2 Quality Control: 5.2.2.1 Providing the equipment for validation. 5.2.2.2 Review the protocol and Report. 5.2.3 Engineering & Maintenance: 5.2.3.1 Providing the utility services. 5.2.3.2 Review the protocol and report. 5.2.4 Production Department:5.2.4.1 Check the calibration protocol 5.2.4.2 Provide the materials used during calibration 5.2.5 Store Department:-Dispensing of material for the calibration of tank
PROTOCOL FOR CALIBRATION OF MANUFACTURING TANK Coral Laboratories Ltd. Location: Dehradun
Equipment Name: Jacketed Manufacturing Tank
Equipment ID No.: LV02
Document No.: QP-PPQ-LV02-02
Page No.: 5 of 7
6.0 DOCUMENTATION All Calibration work will be performed as defined in the protocol. This will include the information, data and analysis of Calibration results. All required documentation (Approved Protocol & Report) would be filled. Any deviation or abnormality observed during the execution of the protocol will be documented. Upon completion of the Calibration process, a final summary report will be prepared, that summarizes the Calibration of the Manufacturing tank. The summary report will be reviewed and approved by Head-QA . 7.0 MATERIAL REQUIREMENT FOR CALIBRATION 7.1 Test Equipment/Instrument Calibration Verification:
1.
Manufacturing tank to be calibrated – I.D. No: LV02
2.
Dip rod for marking of calibrated points
3.
Sodium Chloride IP
4.
Calibrated 50 Ltr SS container.
8.0 PROCEDURE: For the calibration of the tank, maintain the manufacturing area with 25°C. Confirm that there are no traces of chloride in the Purified water used and weigh 49.801 kg of Purified water in the SS container which is equivalent to 50 Lts of water considering the specific gravity of 0.99602. Mark the container with permanent marker at the water level. Weigh the Purified water of 49.801 Kg x 10 and transfer it to the manufacturing tank. Take exact weight of 4.5kg NaCl (Sodium chloride) in calibrated 50 ltrs SS tank & complete dissolve in purified water, make volume up to 50 Ltrs. And transfer it in to manufacturing tank. Now the total volume in the tank is 500 ltr. Stir for 5 minutes. After stirring take out 15 ml of Sodium chloride solution & titrate for content of Sodium chloride content. Based on the content of Sodium chloride calculate the actual volume of purified water content in the tank. If the volume of purified water is within the acceptance criteria as stated in the volume verification chart, permanent mark on Dip rod for 500 ltr. The above same procedure shall followed for calibration point for 1000, 1500, 2000, 2500 and finally 3000 Justification for Addition of Sodium Chloride: The purpose of use of tank is for manufacturing of Oral liquid batches at different batch sizes with appropriate Active Pharmaceutical Ingredients. Hence for to ensure it is essential to determine whether the actual input are getting at different batch size accurately at calibrated points. Hence 0.9%w/v of Sodium Chloride Solution is used as active Pharmaceutical Ingredient in Calibration activity
PROTOCOL FOR CALIBRATION OF MANUFACTURING TANK Coral Laboratories Ltd. Location: Dehradun
Equipment Name: Jacketed Manufacturing Tank
Equipment ID No.: LV02
Document No.: QP-PPQ-LV02-02
Page No.: 6 of 7
Procedure for Estimation of Sodium Chloride: Titrate the water sample with the 0.1N silver nitrate solution using Potassium chromate indicator and note the burette reading for the amount of 0.1N Silver nitrate consumed by NaCl available in the water. Content of Sodium Chloride: Assay =
BR X 0.1N Normality of Ag No3 X 0.005844 X 1000
=
X gm/ltr
For example if we want of verify the volume of purified water for 500 Ltrs :1 X Exact wt. of Sodium Chloride in gms
= Y
Ltr of water in the tank
X gms /Ltr. Y – 500 = + Z ltrs of actual difference of Purified water in tank. (Limit: Difference ± 1%) Determination of % of difference of Purified water in the tank. Z (Actual Difference in Volume) x 100 Calibrated Volume (500)
= W % of Difference)
Repeat the same procedure at different water level listed below in the Observation chart Volume verification Sr.No. 1 2 3 4 5 6
Volume of water 500 1000 1500 2000 2500 3000
Burette Reading
Factor
Assay (g/Ltr)
0.005844
Normality of 0.1N AgNO3 =
9.0 OBSERVATION: Sr. No.
To be
Wt. of water
Volume
NaCl
Assay
Actual
Limit
calibrated
added in Kg
of water
Added
(%w/w)
Water
(± 1.0%)
Result
PROTOCOL FOR CALIBRATION OF MANUFACTURING TANK Coral Laboratories Ltd. Location: Dehradun
Equipment Name: Jacketed Manufacturing Tank
Equipment ID No.: LV02
Document No.: QP-PPQ-LV02-02
Page No.: 7 of 7
at volume
added in
volume
Ltr.
1
(Ltr.) 500
49.801 x 10
Ltr 500
4.5kg
495-505
2
1000
49.801 x 10
1000
4.5kg
990-1010
3
1500
49.801 x 10
1500
4.5kg
1485-1515
4
2000
49.801 x 10
2000
4.5kg
1980-2020
5
2500
49.801 x 10
2500
4.5kg
2475-2525
6
3000
49.801 x 10
3000
4.5kg
2970-3030
10.0 CONCLUSION :
After calibration at different point of manufacturing vessel shall verify and it should be within the 1% of the verification as specified in the Observation table. Than only the tank can be used for the routine commercial batch preparation at calibrated points only.