SMART PIPETTE - USER GUIDE

CALIBRATION AND ADJUSTMENT All SMART pipettes are been quality tested according to ISO8655/DIN 12650. 1265 0. The quality qualit y control according to ISO I SO 8655/DIN 12650 involves gravimetric testing of each pipette with distilled water (quality 3, DIN ISO3696) at 25°C +- 1° C.   All SMART pipettes are been calibrated in ISO/IEC 17025 accredited laboratory. Each Pipette is calibrated, inspected and validated by qualified technicians according to defined quality system. SMART pipettes are constructed to permit re-adjustment for other pipetting techniques or liquids of different temperature and viscosity.

Devise requirements and test conditions An analytical balance must be used. The balance selection depends upon selected model of the pipette and sensitivity of balance reading. Test liquid: liqui d: Water, Water, distilled or deionized, grade gr ade 3 water conforming confor ming ISO 3696. Calibration should be carried out in a draft-free room at a constant (+- 0.5°C) temperature of water w ater,, pipette and air betwe b etween en 15°C to 30°C. The relative humidity must be above 50%, especially with volumes under 50ul, the air humidly should be as high as possible to reduce the effect of evaporation loss. Special accessories for analytical balance, such as the evaporation trap are recommended for the calibration of volumes under 50ul.

9|

 

SMART PIPETTE - USER GUIDE

Procedure to check calibration The pipette pipet te is checked at maximum volume, at 50% of maximum  volume and an d at minimum or 10% of maximum m aximum volume, vo lume, whiche w hichever ver is higher. • A new tip is first pre-wetted 3-5 times and a series of ten pipetting is done at each volume. • Use of forward pipetting technique is recommended. • Calculate the inaccuracy and imprecision for all three volumes as per EN ISO 8655/ DIN 12650 standards on the basis of the following followin g calculation

Conversion of weight readings to volume  _   _  Mean Me an Va Valu luee V = X · Z  _  Mean Value X = ∑ Xi n Xi n Z

= = =

 

Balance Reading number or reading Conver sion factor  Conversion [example Z= Z=1.0040 1.0040 ul ul/mg /mg at 25°c and 1013 hPa]

Calculation for in-accuracy in-accurac y (Systematic (Systematic Error)  _  A% = V - Vo

· 100

 Vo  _   V  Vo

= =

Mean Value Particul Par ticular ar volume volum e at which whi ch readings rea dings are ar e taken

Calculation for Impression (Random Error) n ∑ ( Vi - V)2 S=

i=1

n-1 CV% CV %=

S = Standard Deviation  V

= Mean Value

n = number of readings

100 · S  _   V

• Compare the results to the limits in the subsequent tables

| 10

 

SMART PIPETTE - USER GUIDE

Variable Volume Pipette CAT No.

S VA-125

S VA-100

S VA-200

S VA-500

S VA-600

S VA-700

S VA-800

S VA-905

S VA-1000

Color code

 Volu me Range  Volume Ran ge [ µl ]

0.1 - 2.5 µl

0.5 -10 µ l

5 -5 0 µ l

10-100 µ l

100-1000 µl

2-20 µl

20 -200 µl

500 -5000 µ l

1000-10000 µl

Increments [ µl ]

0.002

0.02

0.1

0.1

0.1

0 .0 2

0. 2

10

20

Test  Volume  Volu me [µl]

Inaccuracy (±) %

Imprecision (±) %

0 . 25

12

6

1 . 25

3

3

2 .5

2.5

1.6

1

2.5

1. 8

5

2

1

10

1

0. 6

5

2

2

25

0. 8

0. 4

50

0. 6

0. 3

10

3

1

50

1

0 .5

10 0

0.8

0. 2

10 0

3

0. 6

500

1

0. 4

1000

0. 6

0. 2

2

3

2

10

1 .5

1

20

0.9

0. 4

20

2.5

0.7

10 0

0.7

0. 3

200

0. 6

0. 2

500

2.4

0. 6

25 0 0

1. 2

0 . 25

50 0 0

0. 6

0.15

1000

3

0. 6

500 0

0.8

0. 3

10000

0. 6

0. 2

11 |

 

SMART PIPETTE - USER GUIDE

Fix Volume Pipette  Volume  Volu me Range Ran ge [ µl ]

Test  Volume  Volu me [µl]

Inaccuracy (±) %

Imprecision (±) %

S FA-2.5

2 .5 µ l

2.5

2

1. 6

S FA - 0 0 0 5

5 µl

5

1. 3

1. 2

S FA - 0 0 1 0

10 µ l

10

1. 2

0.8

S FA - 0 0 2 0

20 µ l

20

1

0 .5

S FA-0025

25 µ l

25

1

0. 3

S FA-0050

50 µ l

50

0.7

0. 3

SFA-0100

100 µl

10 0

0.6

0. 2

S FA - 0 2 0 0

20 0 µ l

20 0

0.6

0. 2

S FA - 0 2 5 0

25 0 µ l

25 0

0.6

0. 2

S FA-0500

50 0 µ l

500

0.6

0. 2

S FA-1000

1000 µl

1000

0. 6

0. 2

SFA-2000

200 0 µ l

2 000

0. 3

0.15

S FA - 5 0 0 0

500 0 µ l

50 0 0

0. 3

0.15

S FA-10000

10000 µl

10 0 0 0

0. 6

0. 2

CAT No.

Color code

Multi Channel Pipette 8 Channels & 12 Channels

CAT No.

SMA8-010 S M A12-010

SMA8-020 SMA12-020

SMA8-050 SMA12-050

Color code

 Volu me  Volume Range [ µl ]

0.5-10 µl

2-20 µl

5 - 50 µ l

SMA8-100

10 – 100

SMA12-100

µl

| 12

Increments [ µl ]

0. 0 2

0. 0 2

0.1

0.1

Test  Volume

Inaccuracy (±) %

Imprecision (±) %

[µl] 1

8

5

5

4

2

10

2

1

2

7

3

10

3

2

20

2

1. 6

5

3

2

25

1. 5

1

50

1

0.7

10

3

2

50 100

1 0.8

0. 8 0. 3

 

SMART PIPETTE - USER GUIDE

SMA8-200 SMA12-200

SMA8-300 SMA12-300

20-200 µl

30-300 µl

0.2

0.2

20

5

1. 4

100

1

0. 4

2 00

0.7

0 . 25

30

3

1

150

1

0 .5

300

0. 6

0. 3

Calibration Adjustmen Adjustmentt 1. Calibr Calibration ation adjustment is done with the Recalibration (Ellen) key. key. 2.Remove 2. Remove the calibration guard and place the Recalibration (Ellen) key into the opening of the calibration screw scr ew on the side of the handle. 3. Turn the key clockwise clock wise to increase or counterclockw counterclockwise ise to decrease the volume. 4. Af After ter adjustment, check the calibration calibr ation according to the instructions above. 5. Once within permissible error range, place calibration guard in its position to avoid accidental changes in calibration.

For more detail information about Calibration and Adjustment, visit  ww  w w w.accuma w.accumaximum.co ximum.com/cal m/calibr ibration ation Depending upon use, we recommend checking of calibration every six months. However this can be adjusted to individual requirements.

13 |

 

SMART PIPETTE - USER GUIDE

MAINTENANCE When the SMART SMA RT pipette is not in use, us e, please make sure that it is stored in an upright position. Accumax recommends the SMART stand for this purpose.  You should check chec k the pipette pip ette at the t he beginning begi nning of each ea ch day for accuac cumulation of dust and dirt on the outside surfaces. Attention should be paid to the tip cone. No other solvents except Iso-propanol Iso-prop anol should be used to clean the pipette. If the pipette pipet te is used daily, it should be inspected every three months. The servicing procedure starts with the disassembly of the pipette. Please refer to the spare parts lists for better understanding of the components. Shown Show n here is an example of dis-assembly and assembly procedure of 100-1000 100-1000 µl pipette, pipet te, the basic procedure procedur e is to be followed for all other volume pipettes.

1

1. For opening press the tip ejector closer upwards

2

2. Rotate the tip ejector closer in the direction indicated until the snap

3

3. Turn Turn out ou t the tip ejector ejec tor,, slowly

• Pull out the piston and other parts by unscrewing the tip cone. Push out with piston the rest of the piston assembly including the Seal and O-ring. Remember to keep all parts in order for reassembly. • Clean the piston, the piston spring, Seal and the O-ring with a isopropanol and lint free tissue. Allow them to dry. • Check the tip cone for foreign particles par ticles and remove, if any. • Grease the cleaned parts with the approved lubricant provided with each pipette. | 14

 

SMART PIPETTE - USER GUIDE

1 Align the notch on the tip ejector with the slots on tip ejector closer as indicated

2 Ensure the notch on the main handle matches the slots on the tip ejector closer. Push the tip ejector closer upwards as indicated.

3 Insert both parts par ts in the aligned manner as indicated.

4 Once completely closed, twist the tip ejector ejec tor closer in the indicated direction to close

For correct re-assembly, it is crucial that the slots slo ts on the tip ejector closer close r match the notches on the inside of the main handle. Please follow the above procedure.

15 |

 

SMART PIPETTE - USER GUIDE

List of Spare Parts Group 1  Variabl  Var iablee Volume Volum e 100-100µ 100-100µll Fix Volume    

250µl 500µl 1000µl

SP 377-1

SP 302-1

SP 70-1 SP 23-1 SP 317-1 SP 392-1 SP 259-1

SP 44-1

SP 314-1

SP 14-1

| 16

 

SMART PIPETTE - USER GUIDE

G r o up 2

G r ou p 3

 Variab  Var iable le Volume Vol ume 0.1-2.5µl

 Variab  Var iable le Volume Vol ume 0.5-10µl

Fix Volume  

Fix Volume  

0.5µl 2.5µl

S P 41-1

10µl

S P 41-2

SP 301-1 SP 301-1

S P 4 8-2

S P 4 8-2

SP 258-1 SP 48-1

SP 38-1

SP 259-1 SP 48-1

SP 38-1

SP 25-1 SP 25-1

SP 314-1

SP 14-1

SP 314-1

SP 14-1

17 |

 

SMART PIPETTE - USER GUIDE

Gr o up 4

G r o up 5

 Variabl  Var iablee Volume Volum e 2-20µl

 Variabl  Var iablee Volume Volum e 5-50 5-50µl µl

Fix Volume

Fix Volume  

20µl

25µl 50µl

SP 39-3 SP 40-1

SP 301-1 SP 306-1

SP 48-2

SP 45-1

SP 391-1 SP 265-1 SP 260-1 SP 47-1 SP 48-1

SP 38-1 SP 38-1

SP 25-1 SP 24-1

SP 314-1 SP 314-1

SP 14-1

| 18

SP 14-1

 

SMART PIPETTE - USER GUIDE

G r ou p 6

G r o up 7

 Variab  Var iable le Volume Vol ume 10-100µl

 Variabl  Var iablee Volume Volum e 20-200 20-200µl µl

Fix Volume  

Fix Volume  

100µl

SP 39-1

200µl

SP 36 6 -1

SP 303-1 SP 305-1

SP 45-1 SP 45-1

SP 386-1 SP 261-1

SP 388-1 SP 264-1

SP 46-1 SP 45-1

SP 38-1

SP 38-1

SP 24-1

SP 24-1

SP 314-1

SP 314-1

SP 14-1

SP 14-1

19 |

 

SMART PIPETTE - USER GUIDE

G r o up 8

G r ou p 9

 Variabl  Var iablee Volume Volum e 1-5ml

 Variab  Var iable le Volume Vol ume 1-10ml

Fix Volume    

Fix Volume

2ml 5ml

SP 31-2

SP 30 -2

SP 51-1

SP 50 -1

SP 53-1

SP 353-1 SP 263-1

10ml

SP 304-1 SP 57-2 SP 42-1

SP 28-1

SP 26-1

SP 52-1

SP 353-1 SP 262-1

SP 304-2 SP 57-1 SP 43-1

SP 29-1

SP 27-1 SP 312-1

SP 313-1 SP 54-1

| 20

SP 55-1

 

SMART PIPETTE - USER GUIDE

AUTOCLAVING Accumax SMART pipette is completely autoclavable at 2 bar pressure, 121°c 121°c temperature temperatur e for 20 minutes exposure time according to DIN. • Keep digital counter in unlock position. • Autoclave the complete pipette without any disassembling. • After autoclaving, allow pipette to completely cool and dry.

  If the pipette is autoclaved autocl aved frequently, the piston and springs should be greased with the lubricant supplied with each pipette to maintain smooth movement.

TROUBLE SHOOTING GUIDE P r o b l em a r e a

P o s s i b l e C au s e

S o l u ti o n

Pipette is leaking

Wor n O-ring or o r s e al

Repl ace wor n par t s

Foreign particles between tip

Clean tip cone, attach

and tip cone Foreign particles between piston and seal

new tip Clean seal and piston

Wor n O-ring or seal

Repl ace wor n par t s

T ip c o n e is l o o s e

T i g ht e n t i p c o n e

Piston is damaged (Chemically or Mechanically)

Return pipette to Authorised Distributor 

D a ma g e d ti p c on e

Repl ace the tip cone

I mp r op er a s s emb ly

See “Maintenance” section

T ip c o n e is l o o s e

T i g ht e n t i p c o n e

T ip i n c o r r e c t ly a t t a c h e d

At tach fir mly

Calibr ation altered

Recalibr ate according to instructions

Calibration not suitable for particular liquid

Recalibrate with the liquid in question

Pipette wont aspirate

Pipette is inaccurate

Inaccurate dispensing with certain liquids

21 |