Centrifuge 5702 / 5702 R / 5702 RH Service Manual · English
Inhalt / Contents / Sommaire / Indice / Indice
Service Manual Centrifuge 5702 / 5702 R / 5702 RH · English · 5702 920.010-04
Safety precautions 1
PACT information sheets 2
Error lists / Troubleshooting 3
Disassembly / Assembly 4
Alignment / Adjustment 5
Diagrams 6
Spare parts list 7 Technical data 8
(blank) Maintenance 9
Check / Inspection 10
Descriptions 11
(blank) EDP / Printer 12
Service Manual Centrifuge 5702 / 5702 R / 5702 RH · English · 5702 920.010-04
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – Contents
Please read carefully before opening the device! . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
2 PACT information sheets
3 Error list / Troubleshooting Error list . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1 Reset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1 Error list . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1 Warnings and notices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7 Customer service indications 5702 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8 Customer service indications 5702 R . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10 Customer service indications 5702 RH . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-13 Service functions 5702 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-17 Program overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-17 Service functions 5702 R . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-23 Program overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-23 Service functions 5702 R H . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-30 Program overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-30
4 Disassembly / Assembly CENTRIFUGE 5702 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1 Front panel / Control PCB / Standby button . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1 Lid . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1 Converter PCB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1 Motor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2 Microswitch / Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2 Mains input module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3 Rotor chamber . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3 Braking resistor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4 CENTRIFUGE 5702 R / 5702 RH . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5 Front panel PCB / Control PCB / Standby button . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5 Converter PCB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5 Lid . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5 Motor with imbalance switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6 Mains switch / IEC chassis plug . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
MSTM/MMWT/4903
0-1
Contents
1 Safety precautions
eppendorf Service Manual
Contents
Centrifuge 5702 / 5702 R / 5702 RH – Contents Fan cooling unit / Fan motor chamber . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8 Fan cooling unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8 Fan at front plane . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8
5 Alignment / Adjustment Adjusting the microswitch for imbalance cut-off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
6 Drawings Overview Electronics 5702 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5702 901.008 Overview Electronics 5702 R . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5703 901.001 Overview 5702 RH . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5704 900.831
7 Spare parts list Spare parts list . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1 Recommended spare parts – Initial delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-9
8 Technical data Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1
10 Operating Measuring the accessible current . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-1 Measuring circuit for accessible current according to EN 61010-1 and UL 3101-1 (UL 61010A-1) . . . . . . . . . . . . . 10-2 Measuring circuit for accessible current according to NFPA 99 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-3 IQ – Installation Qualification OQ – Operational Qualification PQ – Performance Qualification . . . . . . . . . . . IQ – Installation Qualification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . OQ – Operational Qualification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . PQ – Performance Qualification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
10-4 10-5 10-8 10-9
11 Descriptions General layout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11-1 Functional description of assemblies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Voltage supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Control PCB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Converter assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Rotational speed recording and regulation assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Imbalance recognition assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Temperature control (only with 5702 R / 5702 RH) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Controlling he condenser fan (only with 5702 R / 5702 RH) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Heating function (only 5702 RH). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
0-2
11-1 11-1 11-1 11-2 11-2 11-2 11-2 11-2 11-3
MSTM/MMWT/4903
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 1 Safety precautions Please read carefully before opening the device! The centrifuge must not be operated in a hazardous or flammable environment and must not be used to centrifuge explosive or highly reactive substances.
•
When handling toxic or radioactive liquids or pathogenic bacteria from Risk Group II (see World Health Organization: "Laboratory Biosafety Manual"), we recommend the use of aerosol-tight capsules.
• •
The centrifuge must be thoroughly cleaned when liquids have been spilled in the rotor or rotor chamber.
• •
The rotor must be loaded symmetrically. Opposing tubes should be of the same type and should be filled equally.
• •
Please only use original accessories recommended by Eppendorf.
•
Eppendorf does not honor any warranty or accept any responsibility for damage resulting from the use of parts from other manufacturers or improper repair work.
• • •
All servicing is performed at the technician’s own risk!
•
The service manual is only suitable for repair and maintenance work if you use the latest version and if an appropriate training course has been attended at Eppendorf.
The rotor and rotor lid must always be securely fastened. Do not begin centrifugation before the rotor has been securely fastened!
Before plugging in the centrifuge, compare your power supply with the electrical requirements listed on the identification plate.
Using spare parts and accessories which Eppendorf has not recommended can reduce the precision, accuracy and life of the device.
Disconnect the power supply before opening the device! In order to carry out maintenance or repairs on Eppendorf products, it is necessary to attend an Eppendorf training course on the device.
MSTM/MMWT/4903
1-1
1 Safety precautions
•
eppendorf Service Manual
1 Safety precautions
Centrifuge 5702 / 5702 R / 5702 RH – 1 Safety precautions
1-2
MSTM/MMWT/4903
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 2 PACT information sheets
2 PACT information sheets
MSTM/MMWT/4903
2-1
eppendorf Service Manual
2 PACT information sheets
Centrifuge 5702 / 5702 R / 5702 RH – 2 PACT information sheets
2-2
MSTM/MMWT/4903
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting Error list Note:
R = Response of centrifuge when the error is reported.
Display: Cause and response of centrifuge. The consecutive number (cause of error) provides for more detailed localization of the error. It can be read from service program 8 (5702) or service program 9 (5702 R / 5702 RH).
Reset After power-off the display shows "OFF". The time before the display is cleared depends on the operational status (centrifuge accelerating, braking or stopped). All operational states can only be reset if the "OFF" display is cleared between power-off and power-on ! Only then does the controller perform a reset.
Error list Note:
The error display shows Er 3, 4, etc. Er 3-0, 3-1 only appears when the "speed" dial is held down for six seconds and then turned clockwise/counterclockwise.
Cause
Remedy
Er 2
Imbalance switch, at standstill, defective.
– Check/adjust imbalance switch with service display SE 3.
R: Centrifuge will not start.
– Load rotor symmetrically.
Er 3
Check switch, cable, plug.
– Replace Control PCB.
No rotational speed pulses during running or when starting. Speed measurement system defective. Drive defective. The rotational speed is measured at the motor. Error remains active after power is cut.
– Leave centrifuge switched on for > 5 minutes. Start again. – Check motor speed sensor. – Replace Converter PCB.
R: The centrifuge decelerates unbraked. The centrifuge remains locked for 5 minutes. The lid cannot be opened until the error display is cleared. Tubes collide with lid of centrifuge during start.
– Check tubes. Tubes too long or wrong adapter used? – Check plug connections.
Er 3-0
When the centrifuge is powered up Err. 3 is signaled.
See Er 3.
Er 3-1
No pulses after starting centrifuge. Centrifuge brakes.
See Er 3.
Er 3-2
Pulse failure during running. The centrifuge remains locked for 5 min.
See Er 3.
MSTM/MMWT/4903
3-1
3 Error lists / Troubleshooting
Error
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting Error
Cause
Remedy
Er 5
No acknowledge message from lid latch while centrifuge is running.
– Check lid latch function with Service program 3.
R: The centrifuge decelerates unbraked.
– Check/adjust/replace microswitches. – Check/replace reed magnet/ sensor.
3 Error lists / Troubleshooting
– Check/replace motor lid latch.
Er 5-1
When the centrifuge is powered up the rotor rotates. The lid switch is open.
See Er 5.
Er 5-2
Rotor rotates. The lid switch is open
See Er 5.
Er 5-3
No lid latch acknowledge signal.
See Er 5.
Er 6
Fault in Converter PCB, thermal overload.
– Allow centrifuge to cool and restart.
R: The centrifuge decelerates unbraked.
– Determination of converter error code with customer service indicators. – Check/replace Converter PCB. – Check/replace braking resistor. – Motor sluggish / check / replace.
Er 6-1
Converter PCB not ready.
– See Er 6. – Check connection between control PCB and Converter PCB.
Er 6-2
The Converter PCB has reported an error (error code set).
See Er 6.
Er 6-4
Converter PCB has performed a reset (mains power interrupted).
– Check mains power connection.
Er 6-5
Wrong Converter PCB ID after power-up.
See Er 6.
Er 6-6
Safety circuit for release of Converter PCB defective. Motor is too hot.
See Er 6.
3-2
– Check/replace fuses.
MSTM/MMWT/4903
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting Error
Cause
Remedy
Er 7
Overspeed control.
– Check mains voltage.
R: Centrifuge decelerates unbraked.
– Check smooth running of motor. – Check/replace motor. – Replace Converter PCB. – Replace speed sensor.
Er 7-1
Maximum rotor speed exceeded.
See Er 7.
Er 7-2
5702: Speed deviation ±1000 rpm from nominal. 5702 R: Speed deviation ± 600 rpm from nominal. 5702 RH: Speed deviation ± 600 rpm from nominal.
See Er 7.
Er 8
Ramp error during acceleration.
– Check concentric running of motor / replace.
R: Centrifuge brakes.
– Replace Converter PCB.
Er 8-1
Time error.
See Er 8.
Er 9
EEPROM write error.
– Initialize EEPROM in service program 2.
R: No storage of running data.
– Replace control PCB.
Er 9-1
Save running data.
See Er 9.
Er 9-3
Initialization.
See Er 9.
Er 9-4
Save operating hours.
See Er 9.
Er 10
EEPROM read error.
See Er 9.
R: Set defaults, data of last run are lost.
Er 10-2
Read last nominal values.
See Er 9.
Er 10-5
Read operating hours.
See Er 9.
MSTM/MMWT/4903
3-3
3 Error lists / Troubleshooting
– Motor sluggish, check / replace.
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting Error
Cause
Remedy
Er 11
Mains power interrupted.
– Check mains power connections.
Reset by internal software or hardware.
– Check ground connections.
R: Centrifuge decelerates unbraked.
– Check connection from Converter PCB to control PCB. – If recurring, replace control PCB. – Replace Converter PCB as necessary.
Er 14
EPROM checksum error after power-on. Error in program memory, EPROM defective.
Replace control PCB.
3 Error lists / Troubleshooting
R: No further operation possible.
Er 15 In b
Imbalance shutdown occured during a run.
Er 16
Error in inquiry to Converter PCB.
See Er 2.
R: Centrifuge decelerates unbraked.
See Er 3.
R: Centrifuge shuts down and decelerates unbraked.
Er 16-2
Error after telegram repetition.
See Er 3.
Er 17
Error transmitting select telegram to Converter PCB.
See Er 6.
R: Centrifuge shuts down and decelerates unbraked.
Er 17-1
Converter PCB-telegram queue overflow.
See Er 6.
Er 17-2
Error after three repeats of a select telegram.
See Er 6.
Er 18
Excess temperature in rotor chamber.
– Room temperature too high?
(5702 R / 5702 RH only)
R: Centrifuge shuts down and decelerates unbraked.
– Direct sunlight? – Sufficient space for ventilation? – Condenser fan clogged by dust? – Service program SE 10, check temperature sensors at rotor chamber and condenser. – Check all connections. – Check/replace Converter PCB.
3-4
MSTM/MMWT/4903
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting Error
Cause
Remedy
Er 18-1
After reaching the temperature, the ± 5 °C interval is left.
See Er 18.
Temperature in rotor chamber > +12 °C above nominal temperatur.
See Er 18.
Temperature in rotor chamber > 50 °C.
See Er 18.
Er 19 (5702 only)
Mains interruption.
See Er 25.
Er 19
Excess temperature at condenser.
– Allow cooling unit to cool down, check air inlet
(5702 R / 5702 RH only)
Er 18-2 (5702 R / 5702 RH only)
Er 18-3 (5702 R / 5702 RH only)
R: Cooling unit is switched off, condenser fan keeps running.
– Sufficient space to the wall, to other devices? – Does the condener fan also work when the cooling aggregate is running? Replace, if required. – Check service program SE 10. – Check all connections.
Er 19-1
Temperature measured > 62 °C.
See Er 19.
Temperature sensor at rotor chamber defective.
– Check with service program SE 10.
(5702 R / 5702 RH only)
Er 20 (5702 R / 5702 RH only)
Er 21 (5702 R / 5702 RH only)
Er 22 (5702 R / 5702 RH only)
MSTM/KW/2104
R: Centrifuge shuts down, braked deceleration, cooling unit switches off.
– Replace temperature sensor at rotor chamber.
Temperature sensor at condenser defective. Measured value outside –8 °C.....+83 °C
– Check service program SE 10.
R: Centrifuge shuts down, braked deceleration.
– Replace temperature sensor at condenser.
Mains frequency for condenser fan faulty on Converter PCB.
Replace Converter PCB.
R: Fan is switched on or off. Centrifuge can be operated. Error is displayed in standstill only.
3-5
3 Error lists / Troubleshooting
(5702 R / 5702 RH only)
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting Error
Cause
Remedy
Er 23
Excess temperature motor 5702.
Check / replace motor.
(5702 R / 5702 RH only)
R: Centrifuge decelarates unbraked.
Er 24 (5702 R / 5702 RH only)
Er 24-1
Interference converter – compressor. compressor overheated.
– Check service program SE 10. – Replace Converter PCB.
R: Centrifuge shuts down, braked deceleration.
Compressor fault, cannot be deleted at converter.
See Er 24.
Compressor converter faulty during start.
See Er 24.
Compressor does not start (excess current).
See Er 24.
Mains interruption during a run > 5 seconds.
– Check mains voltage.
Mains voltage generally too low or too high? Variations in mains voltage (with 5702 R / 5702 RH possible during start of cooling unit).
– Check/replace mains switch.
Heating switch relay defective.
Check/replace converter PCB.
(5702 R / 5702 RH only)
Er 24-2
3 Error lists / Troubleshooting
(5702 R / 5702 RH only)
Er 24-3 (5702 R / 5702 RH only)
Er 25 Int
Er 27 (5702 RH only)
3-6
– Check/replace Converter PCB.
MSTM/KW/2104
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting Warnings and notices
☞
Warnings are not saved to the EEPROM!
Error display
Cause of error
Remedy
LID
Lid latch error.
Note: In Service mode Er 30 is displayed.
R: Centrifuge does not start, lid will not open.
– Check/adjust/replace switch/reed sensor/magnet. –
Check/replace lid motor.
–
See Er 5.
Er 30
Lid not locked on starting, Limit switch not reached.
See Er 5.
Er 31
EEPROM error.
– Carry out initialization SE 2. – See Er 5.
Display after power-off. How long the control remains active depends on the operational status of the centrifuge (accelerating, braking, stopped).
Er 40
Internal interface error.
3 Error lists / Troubleshooting
OFF
– Replace control PCB. – Replace Converter PCB.
Er 41
Time incremental rotational transmitter defective.
– Service program SE 4. – Replace control PCB.
Er 42
Speed incremental rotational transmitter defective.
– Service program SE 4. – Replace control PCB.
Er 44
No lid acknowledge message.
– See Er 5.
Er 45
No lid acknowledge message.
– See Er 5.
MSTM/MMWT/4903
3-7
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting Customer service indications 5702 Press and hold down "rpm/rcf" button for > 6 seconds (also possible during running). Switch test displays by tapping button. To return to normal display press and hold down button for > 6 seconds. 1. Frequency converter (FC) status display
1.
1 2 3 4
|
|
1
2
3
|
|
|
|
|
4
5
6
7
8
= Lid closed = Lid latch closed = Converter stand-by message = Collective error message – Converter has shut down with an error
5 6 7 8
= Released = Cycle on = Temperature warning – converter power supply = Converter excess current
3 Error lists / Troubleshooting
2. Converter temperature or converter error code (in case of error)
x
2.
x
x
x xx = Converter temperature (in °C) E xx = Converter error code, if converter shut down with an error Error codes:
1 2 4 7
= Undervoltage = Overvoltage = Excess temperature = Excess current
10 = No release / lid switch open FC = Reset by watchdog FF = Serial interface – timeout
3. Intermediate circuit voltage / speed control
3. Lid closed: Lid opened:
x
x
x
xxx = Intermediate circuit voltage xxx = Actual speed for checking speed (turn rotor manually)
4. Evaluating acknowledges
4.
x
0
y
In standstill:
x = 1 Excess current lid latch motor during opening. y = 1 Excess temperature drive motor. z = 1 Imbalance released.
During run:
oxyz = Actual speed from converter
3-8
MSTM/MMWT/4903
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting Customer service indications 5702 5. Converter version
x
5.
x
x
x
x
xxx = Converter, Version x.xx
6. Software version centrifuge
y
6.
y = Internal program variable xx = Software version centrifuge x.x
3 Error lists / Troubleshooting
7. EEPROM checksum
A
B
C
D
ABCD = Four-digit hex number = Program memory checksum
8. Error number, cause of error as per list
7.
u
x
x
u = Cause of error xx = Error number
MSTM/MMWT/4903
3-9
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting Customer service indications 5702 R Press "rpm/rcf" button > 6 s (as well possible during run). Continue to next test displays with short keystroke. Back to normal display by pressing button > 6 s. 0: Temperatures
y
y
y
min sec
yyy xxx min sec
x
x
x
= corr. nom. value = control value = Actual temperature sensor rotor chamber in °C = Converter – compressor ready = Compressor running
3 Error lists / Troubleshooting
1: Condenser temperature
1.
y
y
min sec
yy xx.x min sec rpm rcf
x
x
x
rpm rcf
= Fan status(-- fan off) = Condenser temperature, cooling unit switches off at 62 °C = Converter – compressor ready = Compressor running = Mains 50Hz = Mains 60Hz
2: Converter – compressor
2.
v
min sec
x v xx yy min sec rpm rcf
3-10
x
y
y
rpm rcf
= Internal program parameter = Internal program parameter = Rotating field frequency compressor = Converter – compressor ready = Compressor running = Limit compr. frequency to max. 65 Hz f. mains volt. approx. 96 V ...110 V = Limit compr. frequency to max. 60 Hz f. mains volt. approx. < 96 V
MSTM/MMWT/4903
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting Customer service indications 5702 R 3: Converter parameter drive
3.
y
z
min sec
Lid opened: During run:
x
x
x
y z xxx xxx
= Drive status (internal) = Machine status (internal) = Actual speed, when turning rotor manually = Actual speed from converter
min sec rpm rcf
= Ready converter – drive = Release for converter – drive = Cycling on = Error message converter – drive
rpm rcf
4: Converter parameter drive
x
x
min
y
y
3 Error lists / Troubleshooting
4.
y
xx = Temperature converter (on PCB Power supply – drive) min = Temperature warning yyy = Intermediate circuit voltage
5: Software version centrifuge
5.
x xx yy
x
y
y
= Software version centrifuge x.x = Temperature (on PCB, Power supply compressor)
6: Lid signals / switches
6.
u
v u v x y z
MSTM/MMWT/4903
x = Internal variable (lid latch off) = Motor-temp. switch = Imbalance switch = Lid-lock switch = Lid switch – Reed magnet
y
z
(0 = Temperature too high, (0 = Imbalance, (0 = open, (0 = open,
1 = Switch OK) 1 = Imbalance OK) 1 = closed) 1 = closed)
3-11
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting Customer service indications 5702 R 7: Last error messages
7.
x
x
°C
y
y
■
xx = Errors occured last yy = Penultimate error message etc. °C, ■ = If a converter error code is recognized together with error 6/16/17, it is displayed alterning with the error no. (converter error is marked with °C or ■)
8: Voltage version / check sum EPROM
3 Error lists / Troubleshooting
8.
y
x
x
x
x
y = Voltage version y=0 = 120 V y=1 = 230 V xxxx = Check sum EPROM 0...FFFDH
9: Current error message (selectable only directly after error)
9.
E
x
x
y
y
rpm rcf
E xx y yy
= Error = Error number = No. for locating the cause = Converter error number, if (with error 6) the converter switches off with error rpm = Converter error number drive rcf = Converter error number compressor 1 – Undervoltage 2 – Overvoltage 4 – Excess temperature converter 5 – Excess temperature motor 7 – Excess current converter / intermediate circuit overload 10 – no release, lid switch open (only drive) 40 – Short-circuit shut off FA – Reset by watch dog FF – Serial interface time out
3-12
MSTM/MMWT/4903
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting Customer service indications 5702 RH Press "rpm/rcf" button > 6 s (as well possible during run). Continue to next test displays with short keystroke. Back to normal display by pressing button > 6 s. H: Settings for heating unit
H.
y
z
min sec
y yz xxx rpm °C min sec
x
°C
x.
x
= Rotor number – with open lid (for input see operating instructions) = Heating intensity with closed lid = Temperature of heating cable = Heating unit on / = Heating unit on – acknowledgement converter = Converter-compressor is ready = Compressor running
3 Error lists / Troubleshooting
0: Temperatures
y
y
y
min sec
yyy xxx min sec
x
x
x
= corr. nom. value = control value = Actual temperature sensor rotor chamber in °C = Converter – compressor ready = Compressor running
1: Condenser temperature
1.
y
min sec
y yy xx.x min sec rpm rcf
MSTM/MMWT/4903
x
x
x
rpm rcf
= Fan status(-- fan off) = Condenser temperature, cooling unit switches off at 62 °C = Converter – compressor ready = Compressor running = Mains 50Hz = Mains 60Hz
3-13
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting Customer service indications 5702 RH 2: Converter – compressor
2.
v
x
min sec
v xx yy min sec rpm rcf
x
y
y
rpm rcf
= Internal program parameter = Internal program parameter = Rotating field frequency compressor = Converter – compressor ready = Compressor running = Limit compr. frequency to max. 65 Hz f. mains volt. approx. 96 V ...110 V = Limit compr. frequency to max. 60 Hz f. mains volt. approx. < 96 V
3 Error lists / Troubleshooting
3: Converter parameter drive
3.
y
z
min sec
Lid opened: During run:
x
x
x
y z xxx xxx
= Drive status (internal) = Machine status (internal) = Actual speed, when turning rotor manually = Actual speed from converter
min sec rpm rcf
= Ready converter – drive = Release for converter – drive = Cycling on = Error message converter – drive
rpm rcf
4: Converter parameter drive
4.
x
x
min
y
y
y
xx = Temperature converter (on PCB Power supply – drive) min = Temperature warning yyy = Intermediate circuit voltage
5: Software version centrifuge / Converter parameters of compressor
5.
u
x xx yy u
3-14
x
y
y
= Software version centrifuge x.x = Temperature (on PCB, Power supply compressor) = Limit of fan step ø ^ = to 45 °C, later display in hexadecimal MSTM/MMWT/4903
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting Customer service indications 5702 RH 6: Lid signals / switches
6.
u
v u v x y z
x
y
= Internal variable (lid latch off) = Motor-temp. switch = Imbalance switch = Lid-lock switch = Lid switch – Reed magnet
z
(0 = Temperature too high, (0 = Imbalance, (0 = open, (0 = open,
1 = Switch OK) 1 = Imbalance OK) 1 = closed) 1 = closed)
7: Last error messages
7.
x
x
°C
y
y
■
3 Error lists / Troubleshooting
xx = Errors occured last yy = Penultimate error message etc. °C, ■ = If a converter error code is recognized together with error 6/16/17, it is displayed alterning with the error no. (converter error is marked with °C or ■)
8: Voltage version / check sum EPROM
8.
y
x
x
x
x
y = Voltage version y=0 = 120 V y=1 = 230 V xxxx = Check sum EPROM 0...FFFDH
MSTM/MMWT/4903
3-15
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting Customer service indications 5702 RH 9: Current error message (selectable only directly after error)
9.
E
x
x
y
y
rpm rcf
= Error = Error number = No. for locating the cause = Converter error number, if (with error 6) the converter switches off with error rpm = Converter error number drive rcf = Converter error number compressor 1 – Undervoltage 2 – Overvoltage 4 – Excess temperature converter 5 – Excess temperature motor 7 – Excess current converter / intermediate circuit overload 10 – no release, lid switch open (only drive) 40 – Short-circuit shut off FA – Reset by watch dog FF – Serial interface time out
3 Error lists / Troubleshooting
E xx y yy
3-16
MSTM/MMWT/4903
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting Service functions 5702 The service functions are accessed by pressing the "rpm/rcf" button and the power-on switch simultaneously (not the "Standby" button). Service function selection SE by rotating the "time" dial. Start a service function by "start/stop" button. Quit service program by switching off mains power.
Program overview 1
Measurement, acceleration and braking times
SE
2
EEPROM initialization
SE
3
Test lid latch/imbalance switch
SE
4
Display and button test
SE
5
Cycle duration test
SE
6
Display error table
SE
7
Display last 4 errors
SE
8
Display operating hours and number of centrifuge runs
SE
9
Test overspeed cut-off
MSTM/MMWT/4903
3 Error lists / Troubleshooting
SE
3-17
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting Service functions 5702 SE 1 Measurement, acceleration and braking times Call SE 1. Press "start/stop" to start the program. A short display test follows, all characters appear for approx. 3 seconds and a long beep sounds (the standby button lights up red). Display:
12
min
4.4
rpm
Start the measurement run by pressing the "start/stop" button. Machine accelerates up to final speed. Display:
HL xx
xx
= Acceleration time 0..4268 rpm (97 %) in s
3 Error lists / Troubleshooting
After 3 seconds, display alternates every second between: 1.
HL
xx
2.
zz
yy
zz yy
Acceleration time
= Remaining time = Rounded actual rotational speed
Stop the run by pressing the "start/stop" button. The centrifuge brakes down to a standstill, the lid opens, 5 beeps sound.
br xx.x
xx
= Braking time in s.
Quit the program with the "short" button.
3-18
MSTM/MMWT/4903
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting Service functions 5702 SE 2 EEPROM initialisation Call SE 2. Start the program by pressing the "start/stop" button. Display:
ın
ıt
Confirm execution with the "rpm/rcf" button. * The default display values are stored. * The error table, operating hours, number of centrifuge runs and the table of the most recent errors are cleared. Display:
5.0
min
4.4
rpm
After initialization the program is automatically terminated. Cancel without initialization by "start/stop" button.
3 Error lists / Troubleshooting
SE 3 Test lid latch and imbalance switch Call SE 3. Start the program by pressing the "start/stop" button.
xy Limit switch display: x=0 = Lid open (R) x=1 = Lid closed (R) y=0 = Unlocked (M) y=1 = Locked (M)
ov v=0 Imbalance switch not actuated v=1 Imbalance switch actuated o= For factory check only (R = Reed magnet; M = Microswitch)
Pressing the "open" button actuates the lid motor (also possible with lid opened). Quit the program with the "short" button.
MSTM/MMWT/4903
3-19
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting Service functions 5702 SE 4 Display and key test Call SE 4. Start the program by pressing the "start/stop" button. A short display test follows, all characters appear for approx. 3 s, a long beep sounds and the standby button lights up red briefly.
ın Press all the buttons in succession, in the sequence: "rpm/rcf" button, "open" and "short". Each time a button is pressed a digit on the display is activated.
ın
I88
When the "Standby" (lights up red) and "start/stop" buttons are pressed the display shows:
3 Error lists / Troubleshooting
88
88
After "start/stop" the program is automatically terminated.
SE5 Cycle duration test Call SE 5. Start the program by pressing the "start/stop" button. Display:
10
min
4.4
rpm
The run-time and the rotational speed can be changed. The cycle duration test is started by pressing the "start/stop" button again. Sequence:
2 programs at 4400 rpm are scheduled, with 15 runs and 36 runs respectively. The switchover is made with the Standby button.
r = 15 runs, 10 min run-time, 3 min break after each run, after 15 runs 5 min break r = 36 runs, 15 min run-time, 1 min break after each run, after 36 runs 5 min break 1. Start run. The other starts are automatic; the lid remains closed. In a break between runs the program can be quit with the "short" button.
3-20
MSTM/MMWT/4903
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting Service functions 5702 SE6 Display error table Call SE 6. Start the program by pressing the "start/stop" button. Display:
xx xx yyy
yyy
= Error number = Number of errors
The display changes once per second. It can be cancelled by pressing the "start/stop" button. The table can be cleared with service program SE 2 (Initialize). When the errors have been run through the program is automatically terminated.
SE7 Display last 4 errors Call SE 7. Start the program by pressing the "start/stop" button.
3 Error lists / Troubleshooting
Display:
x x=1 x=2 yy
yy
= Last error = Last-but-one error, etc. = Error number
When the last 4 errors have been run through the program is automatically terminated.
SE8 Display operating hours and number of centrifuge runs Call SE 8. Start the program by pressing the "start/stop" button. Display:
xx xxxx
min sec
xx
= Operating hours (min + sec)
Display:
yy yyyy
yy
= Number of centrifugation runs : 10
☞ Note: The display switches automatically after 3 seconds. The "Number of centrifugation runs" display also appears for 3 seconds, then the program is automatically terminated.
MSTM/MMWT/4903
3-21
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting Service functions 5702 SE9 Test overspeed cut-off The test is performed without the rotor! Call SE 9. Start the program by pressing the "start/stop" button. Display:
10
min
4.8
rpm
Start test run by pressing "start/stop" button. Centrifuge accelerates up to cut-off speed and must shut down with Error 7. It decelerates unbraked. After standstill, 5 signal tones are emitted.
3 Error lists / Troubleshooting
Quit the program with the "short" button when the rotor has come to a standstill.
3-22
MSTM/MMWT/4903
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting Service functions 5702 R The service functions are accessed by pressing the "rpm/rcf" button and the power-on switch simultaneously (not the "Standby" button). Service function selection SE by rotating the "time" dial. Start a service function by "start/stop" button. Quit service program by switching off mains power.
Program overview 1
Measurement, acceleration and braking times
SE
2
EEPROM initialization
SE
3
Test lid latch/imbalance switch
SE
4
Display and button test
SE
5
Cycle duration test
SE
6
Display error table
SE
7
Display last 4 errors
SE
8
Display operating hours and number of centrifuge runs
SE
9
Test overspeed cut-off
SE 10
Fan and cooling unit / display temperature sensors
SE 11
Filling cooling unit (no service function)
SE 12
Imbalance settings
SE 13
Limitation of the fan speed of the condenser
MSTM/MMWT/4903
3 Error lists / Troubleshooting
SE
3-23
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting Service functions 5702 R 1. Measurement, acceleration and braking times Call SE 1. Press "start/stop" to start the program. Display:
1
2
x
x
4.
4
Start the measurement run by pressing the "start/stop" button. Machine accelerates up to the final speed. The cooling unit is switched on only during braking ( xx=-- oder Kompressorfrequenz).
H xx.x
L
x
x.
x
x.
x
= Acceleration time 0..4268 rpm (97 %) in sec
Press "start/stop" to stop the run. Machine brakes until standstill.
3 Error lists / Troubleshooting
b xx.x
r
x
= Braking time
Press "short" to exit the program.
2. EEPROM initialization Call SE 2. Press "start/stop" to start the program. Display:
I
n
ı
t
Press "rpm/rcf" to carry out initialization. * The error table, operating hours, number of centrifuge runs and the table of the most recent errors are cleared. * Fixed programs 10min / 4°C / 4400 rpm are set. Display:
5.
3-24
0
4
4.
4
MSTM/MMWT/4903
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting Service functions 5702 R 3. Test lid latch Call SE 3. Press "start/stop" to start the program.
L v x y z
u
v
= Motor temp. switch = Imbalance switch = Lid lock switch = Lid switch – Reed magnet
x
y
z
(0 = Temperature too high,1 = Switch OK) (0 = Imbalance, 1 = Imbalance OK) (0 = opened, 1 = closed) (0 = opened, 1 = closed)
Switching off the lid motor u = 1 – Switching off the lid by intermediate step during opening and closing. u = 2 – Lid closed u = 3 – Excess current cut-off, lid open The lid motor is controlled during closing by pressing "rcf" (leave a pause of 1 s between each pulse). Press "short" to exit the program.
Call SE 4. Press "start/stop" to start the program. A short display test follows, all characters appear for approx. 3 s, a long beep sounds and the program buttons light up blue.
I
n
Press all the buttons in succession, in the sequence: "rpm/rcf" button, "open", "fast temp" and "short". Each time a button is pressed a digit on the display is activated.
I
8
8
8
8
When the "Standby" and "start/stop" buttons are pressed the display shows:
8
8
8
8
8
8
After "start/stop" the program is automatically terminated.
MSTM/MMWT/4903
3-25
3 Error lists / Troubleshooting
SE 4 Display and button test
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting Service functions 5702 R SE 5 Cycle duration test Call SE 5. Start the program by pressing the "start/stop" button. Display:
1
5
4
4.
4
The run-time and the rotational speed can be changed. The cycle duration test is started by pressing the "start/stop" button again. Sequence:2 programs at 4400 rpm are scheduled, with 15 min. runs and with 10 min. runs respectively. The switchover is made with the Standby button. r = 36 runs, 10 min run-time, 3 min break after each run, after 36 runs 5 min break r = 36 runs, 15 min run-time, 1 min break after each run, after 36 runs 5 min break
3 Error lists / Troubleshooting
1. Start run. The other starts are automatic; the lid remains closed. In a break between runs the program can be quit with the "short" button.
SE 6 Display error table Call SE 6. Start the program by pressing the "start/stop" button. Display:
E xx yyy
r
x
x
y
y
y
= Error number = Number of errors
The display changes once per second. It can be cancelled by pressing the "start/stop" button. The table can be cleared with service program SE 2 (Initialize). When the errors have been run through the program is automatically terminated.
SE7 Display last 4 errors Call SE 7. Start the program by pressing the "start/stop" button. Display:
x x=1 x=2 yy zz
3-26
E
y
y
z
z
= Last error = Last-but-one error, etc. = Error number = Converter error code (see table "Customer service indications 5702 R", No. 9 "Current error message")
MSTM/MMWT/4903
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting Service functions 5702 R SE8 Display operating hours and number of centrifuge runs Call SE 8. Start the program by pressing the "start/stop" button. Display:
x xxxxx
x
x
x
y
y
x
h
= Operating hours
Display:
y
y
y
y
yyyyyy = Number of centrifuge runs
☞ ☞
The display switches over after 3 s! The number of centrifugations is also displayed for 3 s, after this the program terminates automatically!
3 Error lists / Troubleshooting
SE 9 Test overspeed cut-off The test is performed without rotor! Call SE 9. Start the program by pressing the "start/stop" button. Display:
1
5
4
4.
7
Start test run by pressing "start/stop" button. Centrifuge accelerates up to cut-off speed and must shut down with Error 7. It decelerates unbraked. Quit the program with the "short" button when the motor has come to a standstill.
MSTM/MMWT/4903
3-27
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting Service functions 5702 R SE 10 Fan and cooling unit / display temperature sensors Call SE 10. Start the program by pressing the "start/stop" button. Display unit:
u u=b v=b
x
v
y
z
z
= Converter – compressor problem-free = Converter – drive problem-free
Buttons: fast temp = Compressor, open = Condenser fan, temp = Setting condenser fan status, prog 1 = Fan for Converter PCB,
x=1 (on) z=0 (on)
x=0 (off) x=– (off)
zz = settable from 0 up to 32 (switch on condenser fan with "open" before) y=0 (on) y=1 (off)
"short" button: Selecting Display cooling unit Temperatures
3 Error lists / Troubleshooting
Display temperatures:
x xx.x yy.y
x .
x
y
y.
y
= Condenser temperature with factor "x10" = Chamber temperature
"start/stop" terminates the program
SE 11 Filling cooling unit No service function. "short" key for leaving the program.
SE 12 Setting imbalance Call up SE 12. –
Set rotational speed
–
Insert and load rotor Load rotor on one side with G1 = No error Load rotor on one side with G2 = Error: Imbal
–
Close lid
Table for checking imbalance: Rotor Load A-4-38 4 empty round buckets A-4-38 4 empty rectangular buckets F-35-30-17 A-8-17 Positions 1 – 20 occupied by empty sleeves
G1 7g 7g 8g 10g
G2 15g 15g 17g 21g
Start program with "start/stop" key. Display when emergency switch activated:
I
m
b
a
l
Service program is exited. It is possible to open the lid while the rotor is at a stillstand.
☞ Note: The imbalance setting can be checked with SE 12 without accessing the cooling aggregate. 3-28
MSTM/KW/2104
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting Service functions 5702 R SE 13 Limitation of the fan speed of the condenser Call up SE 13. Start program with "start/stop" key. Display when emergency switch is activated:
2
0
Change setting: "time" dial Save setting and end program: "start/stop" key.
☞ Note: The setting of the condenser fan is possible within a range from "20 – 32". A change makes sense when the room temperature is very high or when the voltage is low The standard lowest setting for the fan speed is "20". "25" must be set for 100 volt devices following the replacement of the PCB Control 5702 R. The setting is maintained after the device is switched off or following initialization.
3 Error lists / Troubleshooting
MSTM/MMWT/4903
3-29
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting Service functions 5702 RH The service functions are accessed by pressing the "rpm/rcf" button and the power-on switch simultaneously (not the "Standby" button). Service function selection SE by rotating the "time" dial. Start a service function by "start/stop" button. Quit service program by switching off mains power.
3 Error lists / Troubleshooting
Program overview SE
1
Measurement, acceleration and braking times
SE
2
EEPROM initialization
SE
3
Test lid latch/imbalance switch
SE
4
Display and button test
SE
5
Cycle duration test
SE
6
Display error table
SE
7
Display last 4 errors
SE
8
Display operating hours and number of centrifuge runs
SE
9
Test overspeed cut-off
SE 10
Fan, cooling and heating unit / display temperature sensors
SE 11
Filling cooling unit (no service function)
SE 12
Imbalance setting
SE 13
Limiting the fan speed of the condenser
3-30
MSTM/MMWT/4903
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting Service functions 5702 RH 1. Measurement, acceleration and braking times Call SE 1. Press "start/stop" to start the program. Display:
1
2
x
x
4.
4
Start the measurement run by pressing the "start/stop" button. Machine accelerates up to the final speed. The cooling unit is switched on only during braking ( xx=-- oder Kompressorfrequenz).
H xx.x
L
x
x.
x
x.
x
= Acceleration time 0..4268 rpm (97 %) in sec
Press "start/stop" to stop the run. Machine brakes until standstill.
b
x
3 Error lists / Troubleshooting
xx.x
r
= Braking time
Press "short" to exit the program.
2. EEPROM initialization Call SE 2. Press "start/stop" to start the program. Display:
I
n
ı
t
Press "rpm/rcf" to carry out initialization. * The error table, operating hours, number of centrifuge runs and the table of the most recent errors are cleared. * Fixed programs 10min / 4°C / 4400 rpm are set. Display:
5.
MSTM/MMWT/4903
0
4
4.
4
3-31
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting Service functions 5702 RH 3. Test lid latch Call SE 3. Press "start/stop" to start the program.
L v x y z
u
v
= Motor temp. switch = Imbalance switch = Lid lock switch = Lid switch – Reed magnet
x
y
z
(0 = Temperature too high,1 = Switch OK) (0 = Imbalance, 1 = Imbalance OK) (0 = opened, 1 = closed) (0 = opened, 1 = closed)
Switching off the lid motor u = 1 – Switching off the lid by intermediate step during opening and closing. u = 2 – Lid closed u = 3 – Excess current cut-off, lid open The lid motor is controlled during closing by pressing "rcf" (leave a pause of 1 s between each pulse).
3 Error lists / Troubleshooting
Press "short" to exit the program.
SE 4 Display and button test Call SE 4. Press "start/stop" to start the program. A short display test follows, all characters appear for approx. 3 s, a long beep sounds and the program buttons light up blue.
I
n
Press all the buttons in succession, in the sequence: "rpm/rcf" button, "open", "fast temp" and "short". Each time a button is pressed a digit on the display is activated.
I
8
8
8
8
When the "Standby" and "start/stop" buttons are pressed the display shows:
8
8
8
8
8
8
After "start/stop" the program is automatically terminated.
3-32
MSTM/MMWT/4903
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting Service functions 5702 RH SE 5 Cycle duration test Call SE 5. Start the program by pressing the "start/stop" button. Display:
1
5
4
4.
4
The run-time and the rotational speed can be changed. The cycle duration test is started by pressing the "start/stop" button again. Sequence:2 programs at 4400 rpm are scheduled, with 15 min. runs and with 10 min. runs respectively. The switchover is made with the Standby button. r = 36 runs, 10 min run-time, 3 min break after each run, after 36 runs 5 min break r = 36 runs, 15 min run-time, 1 min break after each run, after 36 runs 5 min break 1. Start run. The other starts are automatic; the lid remains closed. In a break between runs the program can be quit with the "short" button.
3 Error lists / Troubleshooting
SE 6 Display error table Call SE 6. Start the program by pressing the "start/stop" button. Display:
E xx yyy
r
x
x
y
y
y
= Error number = Number of errors
The display changes once per second. It can be cancelled by pressing the "start/stop" button. The table can be cleared with service program SE 2 (Initialize). When the errors have been run through the program is automatically terminated.
SE7 Display last 4 errors Call SE 7. Start the program by pressing the "start/stop" button. Display:
x x=1 x=2 yy zz
E
y
y
z
z
= Last error = Last-but-one error, etc. = Error number = Converter error code (see table "Customer service indications 5702 R", No. 9 "Current error message")
MSTM/MMWT/4903
3-33
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting Service functions 5702 RH SE 8 Display operating hours and number of centrifuge runs Call SE 8. Start the program by pressing the "start/stop" button. Display:
x xxxxx
x
x
x
y
y
x
h
= Operating hours
Display:
y
y
y
y
yyyyyy = Number of centrifuge runs
3 Error lists / Troubleshooting
☞ ☞
The display switches over after 3 s! The number of centrifugations is also displayed for 3 s, after this the program terminates automatically!
SE 9 Test overspeed cut-off The test is performed without rotor! Call SE 9. Start the program by pressing the "start/stop" button. Display:
1
5
4
4.
7
Start test run by pressing "start/stop" button. Centrifuge accelerates up to cut-off speed and must shut down with Error 7. It decelerates unbraked. Quit the program with the "short" button when the motor has come to a standstill.
3-34
MSTM/MMWT/4903
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting Service functions 5702 RH SE 10 Fan, Cooling aggregate and heating unit/ display temperature sensors Call SE 10. Start the program by pressing the "start/stop" button. Display aggregate:
u u=b v=b
x
v
y
°C
w
z
z
= Converter – compressor problem-free = Converter – drive problem-free
Buttons: fast temp = Compressor, open = Condenser fan, temp = Setting condenser fan status, prog 1 = Fan for Converter PCB prog 2 = Heating unit on/off
x=1 (on) z=0 (on)
x=0 (off) x=– (off)
zz = settable from 0 up to 32 (switch on condenser fan with "open" before) y =0 (an) y=1 (aus) w=1 (on) w=0 (off) °C = Heating unit of PCB converter on
"short" key switchover to display of aggregate Temperatures
3 Error lists / Troubleshooting
Display temperatures:
x xx.x yy.y
x .
x
y
°C
y.
y
y
y
= Condenser temperature = Chamber temperature
2x "short" key switchover to display of heating unit temperature:
yy
= Heating unit temperature
"start/stop" terminates the program
SE 11 Filling cooling unit No service function. "short" key to leave the program.
MSTM/MMWT/4903
3-35
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 3 Error lists / Troubleshooting Service functions 5702 RH SE 12 Setting imbalance Call up SE 12. –
Set rotational speed
–
Insert and load rotor Load rotor on one side with G1 = No error Load rotor on one side with G2 = Error: Imbal
–
Close lid
Table for checking imbalance: Rotor Load A-4-38 4 empty round buckets A-4-38 4 empty rectangular buckets F-35-30-17 A-8-17 Positions 1 – 20 occupied by empty sleeves
G1 7g 7g 8g 10g
G2 15g 15g 17g 21g
Start program with "start/stop" key. Display with activated emergency switch:
I
m
b
a
l
3 Error lists / Troubleshooting
Service program is left. It is possible to open the lid while the rotor is at a standstill.
☞ Note: The imbalance setting can be checked with SE 12 without accessing the cooling aggregate. SE 13 Limiting the fan speed of the condenser Call up SE 13. Start program with the "start/stop" key. Display with activated emergency switch:
2
0
Change setting: "time" dial Save setting and end program: "start/stop" key.
☞ Note: The setting of the condenser fan is possible within a range of "20 – 32". A change makes sense when the room temperature is very high or when the voltage is low. The lowest fan speed of "20" is set as a standard. "25" must be set for 100 volt devices following replacement of the PCB Control 5702 R. The setting is maintained after the device is switched off or following initialization.
3-36
MSTM/KW/2104
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 4 Disassembly / Assembly
Centrifuge 5702 Front panel / Control PCB / Standby button –
Switch off the device.
–
Disconnect mains plug.
–
Lift the lid.
–
Detach the "speed" and "time" dials.
–
Unscrew the 2 Phillips screws at the top on the rear of the front cover.
–
Remove front cover.
–
Disconnect plugs.
–
Disconnect ribbon cable plug.
–
Disconnect ground cable from plug shoe.
–
Unscrew the 4 Phillips screws from the control PCB.
–
Disconnect the 3 plugs.
–
Unscrew the 2 Phillips screws M3x8.
Lid Switch off the device.
–
Disconnect mains plug.
–
Lift the lid.
–
Unscrew the 4 Phillips screws.
–
Remove lid (grey).
–
Relieve pressure on springs.
–
Detach guard ring on one side.
–
Withdraw rod.
–
Pull inside section of lid upwards.
4 Disassembly / Assembly
–
Refit in reverse order. –
Start device.
Converter PCB –
Switch off the device.
–
Disconnect mains plug.
–
Lift the lid.
–
Unscrew the 2 Phillips screws at the top on the rear of the front cover.
–
Remove front cover.
–
Disconnect ribbon cable plug.
–
Disconnect ground cable from plug shoe.
MSTM/MMWT/4903
4-1
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 4 Disassembly / Assembly –
Disconnect plugs.
–
Remove front panel.
–
Unscrew the 5 screws M4x8 and 2 plastic screws on converter PCB.
–
Disconnect plugs on the converter PCB.
Refit in reverse order. –
Start device.
4 Disassembly / Assembly
–
Motor
–
Switch off the device.
–
Disconnect mains plug.
–
Lift the lid.
–
Take out the rotor.
–
Lay the centrifuge on its right side.
–
Unscrew the 4 rubber feet from the ground plate.
–
Unscrew the 1 Phillips screw.
–
Detach the ground plate (remains connected to the emergency cord).
–
Unscrew the 6 Phillips screws from the ground plate.
–
Unscrew the 2 Phillips screws (lid-latch plate) from the ground plate.
–
Unscrew the two Phillips screws (back of front plate).
–
Remove front plate.
–
Take motor cable off converter PCB.
–
Unscrew the 3 nuts M5 by which the motor is fixed to the ground plate and disconnect ground.
–
Take out the motor.
–
Replace the motor. When mounting, pay attention to centric position in rotor chamber. Mark position with felt pen, if neccessary.
Refit in reverse order. –
Start device.
–
Microswitch / Sensor
–
Switch off the device.
–
Disconnect mains plug.
–
Unscrew the 6 Phillips screws from the rear panel.
–
Detach the rear panel.
–
Unscrew from below 2 screws each on the microswitch/sensor.
–
Disconnect the plug connection.
–
Replace the defective assembly.
Refit in reverse order. –
Start device.
4-2
MSTM/KW/2104
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 4 Disassembly / Assembly Mains input module –
Switch off the device.
–
Disconnect mains plug.
–
Unscrew the 6 Phillips screws from the rear panel.
–
Detach the rear panel.
–
Unplug the connections from the IEC mains chassis plug.
–
Replace the IEC mains chassis plug.
Refit in reverse order. –
Start device.
Rotor chamber Switch off the device.
–
Disconnect mains plug.
–
Lift the lid.
–
Take out the rotor.
–
Unscrew the 2 Phillips screws at the top on the rear of the front cover.
–
Lay the centrifuge on its right side.
–
Unscrew the 4 rubber feet from the ground plate.
–
Unscrew the 1 Phillips screw.
–
Detach the ground plate.
–
Unscrew the 6 Phillips screws from the ground plate.
–
Unscrew the 2 Phillips screws (lid-latch plate) from the ground plate.
–
Remove housing.
–
Push out the two foam parts.
–
Replace the rotor chamber. When mounting, pay attention to centric position to motor.
4 Disassembly / Assembly
–
Refit in reverse order. –
Start device.
MSTM/MMWT/4903
4-3
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 4 Disassembly / Assembly Braking resistor –
Switch off the device.
–
Disconnect mains plug.
–
Lift the lid.
–
Unscrew the 2 Phillips screws at the top on the rear of the front cover.
–
Remove front panel.
–
Disconnect ribbon cable plug.
–
Disconnect ground cable from plug shoe.
–
Disconnect plugs.
–
Remove 7 screws from lid-latch plate.
–
Flap lid-latch plate to front.
–
Loosen 2 screws from brake resistor.
Refit in reverse order.
4 Disassembly / Assembly
Start device.
4-4
MSTM/MMWT/4903
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 4 Disassembly / Assembly
Centrifuge 5702 R / 5702 RH Front panel PCB / Control PCB / Standby button –
Switch off the device.
–
Disconnect mains plug.
–
Lift the lid.
–
Detach the "speed" and "time" dials.
–
Unscrew the 2 Phillips screws at the top on the rear of the front cover.
–
Remove 6 Phillips screws M4x8 from Front PCB.
–
Disconnect 4 plugs. (Note orientation of ribbon cable plugs).
–
Remove 2 Phillips screws M3x8 from PCB Standby button.
–
Replace assembly
Refit in reverse order. –
Start device.
Converter PCB Switch off the device.
–
Disconnect mains plug.
–
Lift the lid.
–
Unscrew the 2 Phillips screws at the top on the rear of the front cover.
–
Pull front cover upwards.
–
Disconnect ribbon cable plug.
–
Disconnect ground cable from plug shoe.
–
Disconnect plugs on the converter PCB.
–
Disconnect plugs.
–
Unscrew the 5 screws and 2 plastic screws on converter PCB.
–
Replace assembly
4 Disassembly / Assembly
–
Refit in reverse order. –
Start device.
Lid –
Switch off the device.
–
Disconnect mains plug.
–
Lift the lid.
–
Unscrew the 4 Phillips screws.
–
Remove lid (grey).
MSTM/MMWT/4903
4-5
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 4 Disassembly / Assembly –
Relieve pressure on springs.
–
Detach guard ring on one side.
–
Withdraw rod.
–
Pull lid upwards.
Refit in reverse order. –
Start device.
Motor with imbalance switch –
Switch off the device.
–
Disconnect mains plug.
–
Lift the lid.
–
Take out the rotor.
–
Unscrew the 2 Phillips screws at the top on the rear of the front cover.
–
Pull front cover upwards.
M4x10
M4x10 unscrew
4 Disassembly / Assembly
Device foot remove
Device foot remove
Centrifuge front view place on right hand side only (fan side) !
☞ Device foot remove
M4x10 unscrew
Device foot remove
After tilting the centrifuge, it is required to rest the device for at least 4 hours in normal position to avoid damage to the compressor!
M5x10 unscrew
M5x10 unscrew
M5x10 unscrew Motor ring
4-6
MSTM/MMWT/4903
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 4 Disassembly / Assembly –
Disconnect plugs on Front PCB.
–
Unscrew the 3 nuts by which the motor is fixed to the ground ring and disconnect ground.
–
Take out the motor.
–
Replace assembly
Refit in reverse order. –
Start device.
Mains switch/ IEC chassis plug –
Switch off the device.
–
Disconnect mains plug.
–
Unscrew the 6 Phillips screws from the rear panel.
–
Detach the rear panel.
–
Disconnect plugs.
–
Press latches.
–
Replace assembly.
Refit in reverse order. –
Start device.
4 Disassembly / Assembly
MSTM/MMWT/4903
4-7
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 4 Disassembly / Assembly Fan cooling unit / Fan motor chamber / Remove housing –
Switch off the device.
–
Disconnect mains plug.
–
Lift the lid / remove rotor.
–
Unscrew the 2 Phillips screws M4x8 at the top on the rear of the front cover.
–
Pull front cover upwards.
–
Disconnect plugs.
–
Disconnect ribbon cable plug.
–
Disconnect ground cable from plug shoe.
–
Remove cable binder from ferrite core, replace later with a new cable binder.
–
Unscrew the 6 Phillips screws from the rear panel. M4x10
M4x10 unscrew
Device foot remove
Device foot remove
Centrifuge front view place on right hand side only (fan side) !
4 Disassembly / Assembly
☞ Device foot remove
M4x10 unscrew
Device foot remove
–
Unscrew 6 Phillips screws M5x10 (housing screws) at the sides.
–
Place centrifuge on its ground plate.
–
Unscrew 5 Phillips screws M5x10 (housing screws) at the front.
–
Remove housing by spreading it slightly backwards and upwards.
After tilting the centrifuge, it is required to rest the device for at least 4 hours in normal position to avoid damage to the compressor!
Fan cooling unit –
Unsrew 2 hex screws M4x10.
–
Loosen cable clamp.
–
Unsolder cable in harness.
–
Pull out fan upwards.
Fan at front plane
☞
Loosen 2 Phillips screws at front plane to ease disassembly!
–
Push out 4 plastic pins.
–
Disconnect cable connections.
Refit in reverse order. –
Start device.
4-8
MSTM/MMWT/4903
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 5 Alignment / Adjustment Adjusting the microswitch for imbalance cut-off Place the centrifuge on a horizontal, stable work surface. The imbalance emergency cut-off is effective as soon as e.g. a strong imbalance moves the centrifuge. The microswitch functions as an opener. –
Switch off device.
–
Disconnect mains plug.
–
5702: Disassemble the base plate with the motor or remove the rear wall of the centrifuge.
–
5702 R (up to serial no. 3120): Alignment is possible from the underside of the device, remove both of the screws for the microswitch, see Chapter 4, Disassembly / Assembly.
Aligning microswitch: The microswitch with its platen must always be positioned against the motor. The microswitch should be adjusted in such a way that the microswitch switches in the case of a motor imbalance of 2 – 3 mm. Adjust with the two visible screws on the underside of the device. Carry out an imbalance inspection, see Ch. 10.
–
5702 R (as of serial no. 3121):
–
5702 RH:
Align microswitch: The microswitch can be adjusted with an adjustment screw in the base plate, which is accessible from the left side of the device. Remove the collection tray and unscrew the clamping screw (1). Adjust the microswitch with the adjustment screw (2) in such a way that the microswitch is activated in the case of an imbalance of 2 – 3 mm. The actuator of the microswitch with its platen must always be positioned against the motor. Screw the clamping screw (1) tight. Insert collection tray and carry out an imbalance test, see Chapter 10 "Setting imbalance" or SE 12, Chapter 3, checking without accessing cooling unit.
2
5 Alignment / Adjustment
1
MSTM/KW/2104
5-1
eppendorf Service Manual
5 Alignment / Adjustment
Centrifuge 5702 / 5702 R / 5702 RH – 5 Alignment / Adjustment
5-2
MSTM/MMWT/4903
8
7
6
locking latch position
F
5
LID-motor
LID
4
3
2
1
5702
M2
Micro-switch + rt/red 12V
S2
-
bl/blue
S3 Reed-Contact
ws/white
gr/gray gr/gray
ws/white ws/white
rt/red bl/blue
sw/black sw/black vio vio
ws/white
E
X11 1 2 3
1 2 3
LCD - GLAS X1
C TIME
SHORT
OPEN
SPEED
OFF
LP_"Control 5702" PCB Control 5702
ST 601
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
M1 LP "Umrichtersteuerung" PCB "frequency converter control"
S T
S100
S101
F1/F2 5,0 A 2,5 A F1
A
F2
D
sw (black)
S5 2a
1
1a
Chopper - Resistor
A Main-plug with Fuses
BLACK
Motor
2 3
ST602
Masseband GND
R1
2
gnge ws (white)
BLUE
sw/black sw/black
Version 50/60Hz 50/60Hz 50/60Hz
X10
1
green/yellow
B 100V 120V 230V
BROWN
Drehzahlsensor Speed sensor
1 2
Massebolzen/GND bolt
L1 N PE
1 2 3
gnge
R
5V ENCGND
X10 2 1
CHOP-R ZW+
CONFIDENTIAL
VERTRAULICH
X9 4 3 2 1
Imbalance switch S1 Motortemp.schalter Motor-temp. switch S4
D X2
Unwuchtschalter
Main-switch type 1552.3102
FERRITKERN Ferrite Core
for service purposes only eppendorf Zentrifugen
330 R
Ueberblick Elektronik 5702 Overview Electronics 5702 Blatt 1 von 1 Gepr.: Datum: 14.3.02/Kache
5702_901.008-01
8
7 LID-motor
F
6
4
3
2
1
locking latch position
LID
M2
5
Micro-switch Reed-Contact
R2 temperature condenser
NO
S2
S3
condenser fan X6 A B
E2
BU/bl RD/rt BK/sw BK/sw
M5 BK/sw BK/sw
Kessel chamber
RC SC TC
X1 X4
X2 2
1 2 3 4 5 6 7 8
X3
1 2 3 4
LCD - Display
C
TIME
SPEED
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18
S703 S702 A B
S601
S700
R1
rotortemp. sensor
WH/ws WH/ws
S1 Imbalance
12V VCC(FR+) 12V FR12V VCC(RX+) GND RXGND CL GND TX+ GND(TX-) ML SYN VCC ROT VCC
M1
Motortemp.
LP "Umrichtersteuerung" PCB "frequency converter control"
S400 S401 S402
1 2 3
S T
BN/br BU/bl BK/sw
S102
Motor
Drehzahlmessung 1pol.
1 2 3
L1 N PE
PCB "5702R_Control" for all Versions
Massebolzen
R
5V ENC-
CONFIDENTIAL
S701 VERTRAULICH
E3 BK/sw RD/rt
compressor
S104
1
Zusatz luefter
KondensatorLuefter
BK/sw BK/sw vio vio
RD/rt BU/bl
D
WH/ws WH/ws vio vio BU/bl RD/rt BK/sw BK/sw
BN/br BU/bl BK/sw
E
additional fan
speed sensor
ST703
GN-YE/gn-ge BK/sw
B Version 100V 50/60Hz 120V 50/60Hz 230V 50/60Hz
F1 T-6,0A UL T-5,0A UL T-2,5A IEC
F1
A
F2
D
Cable colors schwarz braun violett rot gelb gruen blau weiss grau
F2 T-6,0A UL T-5,0A UL T-2,5A IEC
BK/sw GN-YE/gn-ge WH/ws
sw br vio rt ge gn bl ws gr
black brown violet red yellow green blue white gray
BK/sw
BK BN vio RD YE GN BU WH GY
Masseband
BN/br RD/rt
WH/ws
A Main-plug with Fuses
GN-YE/gn-ge
S5 main-switch
for service purposes only eppendorf Zentrifugen
Overview 5702R Blatt Gepr.: Datum:
von 17.1.01
5703 901.001-02
8
7
6
5
4
3
2
1
R3 LID-motor
F
locking latch position
LID
M2
heating temperature heating
90degr.C
5704 E1
Micro-switch Reed-Contact
R2 temperature condenser
NO
S2
S3
condenser fan X6 A B BK/sw BK/sw BU/bl RD/rt BK/sw BK/sw
E2
M5 BK/sw BK/sw
RC SC TC
X1 X4
X2 2
1 2 3 4 5 6 7 8
X3
1 2 3 4
LCD - Display
C
TIME
SPEED
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18
S703 S702 A B
S601
S700
WH/ws R1
rotortemp. sensor
WH/ws WH/ws
S1 Imbalance
12V VCC(FR+) 12V FR12V VCC(RX+) GND RXGND CL GND TX+ GND(TX-) ML SYN VCC ROT VCC
M1
Motortemp.
LP "Umrichtersteuerung" PCB "frequency converter control"
S400 S401 S402
1 2 3
S T
BN/br BU/bl BK/sw
S102
Motor
Drehzahlmessung 1pol.
1 2 3
L1 N PE
PCB "5702R_Control" for all Versions
Massebolzen
R
5V ENC-
CONFIDENTIAL
BK/sw Kessel chamber
S701 VERTRAULICH
E3 BK/sw RD/rt
compressor
S104
1
Zusatz luefter
KondensatorLuefter
BK/sw BK/sw vio vio
RD/rt BU/bl
D
WH/ws WH/ws vio vio BU/bl RD/rt BK/sw BK/sw
BN/br BU/bl BK/sw
E
additional fan
speed sensor
ST703
GN-YE/gn-ge BK/sw
B Version 100V 50/60Hz 120V 50/60Hz 230V 50/60Hz
F1 T-6,0A UL T-5,0A UL T-2,5A IEC
F1
A
F2
D
F2 T-6,0A UL T-5,0A UL T-2,5A IEC
Cable colors schwarz braun violett rot gelb gruen blau weiss grau
BK/sw GN-YE/gn-ge WH/ws
sw br vio rt ge gn bl ws gr
BK/sw black brown violet red yellow green blue white gray
BK BN vio RD YE GN BU WH GY
Masseband
BN/br RD/rt
WH/ws
A Main-plug with Fuses
GN-YE/gn-ge
S5 main-switch
for service purposes only eppendorf Zentrifugen
Overview 5702RH Blatt 1 von 1 Gepr.: 17.1.01 Datum:
5704 900.831-00
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 7 Spare parts list
1-3
1-2
7 Spare parts list
1-4
1-1
1-6 1-7 1-8 3-9 3-10
1-5 1-7 1-8
3-1 3-4 3-5 3-5
3-3
3-4
3-6
Fig. 1 Centrifuge 5702
3-14
3-15
Fig. 2
MSTM/KW/2104
7-1
eppendorf Service Manual
7 Spare parts list
Centrifuge 5702 / 5702 R / 5702 RH – 7 Spare parts list
2-2
2-1 1-9 4-1
Fig. 3
4-2 4-3
4-2
Fig. 4
7-2
MSTM/MMWT/4903
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 7 Spare parts list 7 Spare parts list
2-3
2-4 4-5
Fig. 5
5-6 5-1
5-5
4-4
MSTM/MMWT/4903
5-4
Fig. 6
7-3
eppendorf Service Manual
7 Spare parts list
Centrifuge 5702 / 5702 R / 5702 RH – 7 Spare parts list
1-4
1-2
1-3
1-1
3-2
1-6 1-7 1-8
1-5 1-7 1-8 3-9
3-8 3-4
3-1
3-7
3-4
3-3
3-14
3-6
Fig. 7 Centrifuge 5702 R
7-4
MSTM/MMWT/4903
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 7 Spare parts list 7 Spare parts list
7-3 7-2
6-1 6-2
Fig. 8
MSTM/MMWT/4903
3-13
7-5
eppendorf Service Manual
5702 R
5702 RH
x
x
x
x
x
x
x
x
5703 853.058
x
x
Spring, left
5703 853.066
x
x
1-7
Guide bushing (set of 2)
5702 853.070
x
x
x
1-8
Lock washer (set of 2)
5702 853.089
x
x
x
1-9
Emergency lid release, consisting of: Cord, Nylon pointed staple
5702 853.097
x
x
x
New part
5702
7 Spare parts list
Centrifuge 5702 / 5702 R / 5702 RH – 7 Spare parts list
Plastic part of cover, without monitoring glass
5702 853.011
x
1-1
Plastic part of cover, without monitoring glass 5702 R / RH
5703 853.015
1
1-2
Monitoring glass
5425 170.050
7
1-2
Monitoring glass 5702 R / RH
5703 853.023
1
1-3
Metal body
5702 853.038
7
1-3
Metal body 5702 R
5703 853.031
1+7
1-4
Reed magnet
5702 853.046
1
1-5
Spring, right
5702 853.054
x
1
1-6
Spring, left
5702 853.062
x
7
1-5
Spring, right
7
1-6
1+7 1+7 3
Fig.
Pos.
1
1-1
7
1
2
Designation
Lid
x x x
Lid latch
3
2-1
Motor for lid latch
5426 853.100
x
x
x
3
2-2
Reed sensor
5702 853.127
x
x
x
5
2-3
Lid catch (set of 2) with nuts, bushings, washers, springs
5702 853.135
x
x
x
5
2-4
Microswitch
5810 853.044
x
x
x
1
3-1
Housing, with sealing ring for rotor chamber
5702 852.015
x
7
3-1
Housing 5702 R
5703 852.019
x
x
7
3-2
Housing attachment
5703 852.051
x
x
3
Housing
1
3-3
Front panel
5702 852.031
7
3-3
Front panel 5702 R
5703 852.035
x
7
3-3
Front panel 5702 RH
5704 850.036
1+7
3-4
Dial
5702 852.040
x
x x
1
3-5
Push button "short/open"
5702 852.066
x
1+7
3-6
Push button "stand by"
5702 852.074
x
7
3-7
Push button 5702 R / RH, "short", "fast cool", "open", "up", "down"
5703 852.060
7
3-8
Push button 5702 R / RH, prog. 1/2
5703 852.078
1
3-9
Sealing for rotor chamber
5702 852.082
7
3-9
Sealing for rotor chamber 5702 R / RH
5703 852.043
1
3-10
Rotor chamber
5702 852.090
x
3-11
Foam part for rotor chamber
5702 852.104
x
3-12
Sealing profile for rotor chamber
5702 852.112
x
7-6
x
x
x
x
x
x
x
x
x
x
x
MSTM/MMWT/4903
eppendorf Service Manual
x
5703 852.027
3-13 8
3-13
2
3-14
Base skirting
5702 852.139
7
3-14
Base skirting 5702 R / RH
5703 852.124
2
3-15
Device foot (set of 4)
5702 852.139
x
3
4-1
Converter PCB, 230 Volt (without heat sink)
5702 851.051
x
3
4-1
Converter PCB, 120 Volt (without heat sink)
5702 851.060
x
3
4-1
Converter PCB, 230 Volt (without heat sink)
5703 851.055
3
4-1
Converter PCB, 100/120 Volt (without heat sink)
5703 851.063
4
4-2
Control PCB, 5702 (incl. PCB Standby button and prog. software)
5702 851.086
4
4-2
Control PCB 5702 R (incl. PCB Standby button and prog. software)
5703 851.080
4
4-2
Control PCB 5702 RH (incl. PCB Standby button and prog. software)
5704 850.010
4
4-3
EPROM prog. software 5702
5702 852.058
4
4-3
EPROM prog. software 5702 R
5703 851.098
4
4-3
EPROM prog. software 5702 RH
5704 850.028
6
4-4
Imbalance switch
5702 851.108
x
5
4-5
Brake resistor 330 E
5702 851.116
x
4-6
Ribbon cable Converter / PCB Control 5702 R / RH
4-7
Fan (with stopper) for Converter PCB
4
5 6
Designation
5702 RH
5702 852.023
Pos.
5702 R
Rear panel Rear panel 5702 R / RH
Fig.
x
x
x
x
x
x
x
x
x
x
x
Electronics
x x x x x x x
x
5703 851.110
x
x
5703 851.101
x
x
Drive
5-1
Motor 5702, 100–240 Volt, with speed sensor
5-1
Motor 5702 R / RH, 100–240 Volt, with speed sensor
5703 851.012
x
x
5-2
Foam for insulating flange (upon motor)
5703 851.020
x
x
6
5-4
Shock mount (set of 3)
5702 851.035
x
x
x
6
5-5
Copper screen
5425 604.009
x
x
x
6
5-6
O-ring
5702 851.043
x
x
x
6
Mains connection
8
6-1
Mains input module
5702 851.132
x
x
x
8
6-2
Mains switch
0113 202.963
x
x
x
6-3
Fuses for 230 Volt, 2.5 AT (set of 2)
5425 351.003
x
x
x
6-4
Fuses for 100/120 Volt, 5 AT (set of 2)
5425 353.006
x
x
x
MSTM/KW/2104
5702 851.019
x
7-7
7 Spare parts list
New part
5702
Centrifuge 5702 / 5702 R / 5702 RH – 7 Spare parts list
eppendorf Service Manual
5702 R
5702 RH
7-1
Temperature sensor rotor chamber
5811 643.000
x
x
8
7-2
Compressor 100–230 V, 50–60 Hz
5703 852.132
x
x
8
7-3
Condenser fan, 230 Volt
5703 852.159
x
x
8
7-3
Condenser fan, 120 Volt
5703 852.167
x
x
8
7-4
Condenser fan, 100 Volt
5703 852.183
x
x
7-5
Condenser sensor
5703 852.175
x
x
Fig.
Pos.
7
8
7-8
Designation
New part
5702
7 Spare parts list
Centrifuge 5702 / 5702 R / 5702 RH – 7 Spare parts list
Refrigerating system
Accessories
8-1
Swing-bucket rotor A-8-17
5702 700.002
x
x
x
8-2 8-3
Swing-bucket rotor A-4-38 (incl. round buckets)
5702 720.003
x
x
x
Swing-bucket rotor A-4-38 (without buckets)
5702 740.004
x
x
x
8-4
Round buckets 85 ml for A-4-38 (set of 4)
5702 722.006
x
x
x
8-5
Round buckets 85 ml for A-4-38 (set of 1)
5702 743.003
x
x
x
8-6
Rectangular buckets 90 ml for A-4-38 (set of 4)
5702 709.000
x
x
x
8-7
Rectangular buckets 90 ml for A-4-38 (set of 1)
5702 744.000
x
x
x
8-8
Rotor F-35-30-17 (incl. 30 sleeves)
5702 704.008
x
x
x
8-9
Rotor F-35-30-17 (incl. 10 sleeves)
5702 705.004
x
x
x
8-10
Rotor key
5416 301.001
x
x
x
8-11
Rotor key holder with suction cup
5703 852.213
x
x
8-12
Tray for condensation water
5703 852.116
x
x
8-13
Pivot grease (for swing-bucket rotor only)
5810 350.050
x
x
x
8-14
Packing 5702, cpl.
5702 852.201
x
8-15
Packing 5702 R / RH, cpl.
5703 852.205
x
x
MSTM/KW/2104
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 7 Spare parts list 7 Spare parts list
Recommended spare parts – Initial delivery
To:
Eppendorf AG
From:
Fax no.:
+49 40-5 38 01-758
Fax no.: Name: Phone:
Pos.
Designation
5702 R
5702 RH
230 V
100/120 V
Qty.
5702
Order no.:
1+7
1-4
Reed magnet
5702 853.046
x
x
x
x
x
3
1-9
Emergency lid release, consisting of: Cord, Nylon pointed staple
5702 853.097
x
x
x
x
x
3
2-2
Reed sensor
5702 853.127
x
x
x
x
x
5
2-4
Microswitch
5810 853.044
x
x
x
x
x
1
3-9
Sealing for rotor chamber
5702 852.082
x
x
x
7
3-9
Sealing for rotor chamber 5702 R / RH
5703 852.043
x
x
3
4-1
Converter PCB, 230 Volt (without heat sink)
5702 851.051
x
3
4-1
Converter PCB, 100/120 Volt (without heat sink)
5702 851.060
x
3
4-1
Converter PCB, 230 Volt (without heat sink)
5703 851.055
3
4-1
Converter PCB, 100/120 Volt (without heat sink)
5703 851.063
4
4-2
Control PCB 5702 (incl. PCB Standby button and prog. software)
5702 851.086
4
4-2
Control PCB 5702 R (incl. PCB Standby button and prog. software)
5703 851.080
4
4-2
Control PCB 5702 RH (incl. PCB Standby button and prog. software)
5704 850.010
6
4-4
Imbalance switch
5702 851.108
x x
Fig.
5 6
6
MSTM/KW/2104
New part
4-5
Brake resistor 330 E
5702 851.116
4-7
Fan (with stopper) for Converter PCB
5703 851.101
5-1
Motor 5702, 100–240 Volt, with speed sensor
5702 851.019
5-1
Motor 5702 R / RH, 100–240 Volt, with speed sensor
5703 851.012
5-4
Shock mount (set of 3)
5702 851.035
6-3
Fuses, 2.5 AT, 230 Volt (set of 2)
6-4
Fuses, 5 AT, 100/120 Volt (set of 2)
7-1
Temperature sensor rotor chamber
7-4
Condenser sensor
x
x
x x x
x
x
x
x
x x x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x x
x
x
x
x
x
x
x
x
5425 351.003
x
x
x
x
5425 353.006
x
x
x
5811 643.000
x
x
x
x
5703 852.175
x
x
x
x
x
7-9
eppendorf Service Manual
7 Spare parts list
Centrifuge 5702 / 5702 R / 5702 RH – 7 Spare parts list
7-10
MSTM/MMWT/4903
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 8 Technical data Technical data 5702 R
5702 RH
Power supply:
230 V / 50–60 Hz 120 V / 50–60 Hz 100 V / 50–60 Hz
230 V / 50–60 Hz 120 V / 50–60 Hz 100 V / 50–60 Hz
230 V / 50–60 Hz 120 V / 50–60 Hz
Power requirement:
200 W
380 W
380 W
Current consumption 230 V: 120 V: 100 V
0.9 A 2.3 A 2.4 A
1.7 A 3.3 A 3.5 A
1.7 A 3.3 A 3.5 A
Fuse protection
2.5 A T 5.0 A T 5.0 A T
2.5 A T 5.0 A T 5.0 A T
2.5 A T 5.0 A T 5.0 A T
Rotational speed range:
100 – 4400 rpm
100 – 4400 rpm
100 – 4400 rpm
Max. centrifugal force:
3000 rcf
3000 rcf
3000 rcf
Max. kinetic energy:
2280 Nm
2280 Nm
2280 Nm
Permissible density of the material to be centrifuged:
1.2 g/ml
1.2 g/ml
1.2 g/ml
Max. load depending upon rotor:
4x 90 ml / 30x 15 ml
4x 90 ml / 30x 15 ml
4x 90 ml / 30x 15 ml
Ambient temperature:
2–40 °C
10–40 °C
10–40 °C
Max. rel. air humidity:
10 – 75 %
10 – 75 %
10 – 75 %
Setup altitude:
max. 2000 m above sea level
max. 2000 m above sea level
max. 2000 m above sea level
Degree of contamination:
2
2
2
Overvoltage category:
II
II
II
Noise level:
< 58 dB (A)
< 54 dB (A) steady state at 4 °C 23 °C ambient temperature empty rotor
< 54 dB (A) steady state at 4 °C 23 °C ambient temperature empty rotor
Acceleration time 230 V: 100 V / 120 V:
< 25 s < 25 s
< 25 s < 25 s
< 25 s < 25 s
Deceleration time 230 V: 100 V / 120 V:
< 25 s < 25 s
< 25 s < 25 s
< 25 s < 25 s
Weight excluding rotor:
20 kg
36 kg
36 kg
Dimensions (HxDxW)
241x395x320
268x581x381
268x581x381
132 g 129 g
132 g 129 g
230 V: 120 V: 100 V
Refrigerant R 134 a 230 V 100 V / 120 V
8 Technical data
5702
Technical specifications subject to change!
MSTM/KW/2104
8-1
eppendorf Service Manual
8 Technical data
Centrifuge 5702 / 5702 R / 5702 RH – 8 Technical data
8-2
MSTM/MMWT/4903
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 10 Check / Inspection Measuring the accessible current The test is carried out according to IEC 61010-1 or VDE 702-1 (repetition measurement). IEC 61010-1 corresponds to EN 61010-1 in the EC and UL 3101-1 (UL 61010A-1) in the USA. Devices operated in hospitals in the USA are additionally tested according to NFPA 99 using measuring circuit 2. SECUTEST SIII will be used as a reference device. The measurement is carried out on a switched on device. The limits stipulated by the standard must not be exceeded.
Note: The centrifuges 5702 / 5702 R / 5702 RH do not fall under the present status of the norm for medical products IEC 60601-1
10 Check / Inspection
MSTM/MMWT/4903
10-1
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 10 Check / Inspection Measuring circuit for accessible current according to EN 61010-1 and UL 3101-1 (UL 61010A-1) Measuring circuit in the reference device SECUTEST SIII
Test circuit A.1
A
1
C2
C1
R1
R3
Voltmeter
R2
V
2
10 Check / Inspection
B
Example
Plug
Socket
N
N
L
L
PE
PE
A
N L
PE
Test circuit A.1
1
Centrifuge
B
2 V
The accessible current shall be measured with circuit A.1. 1. measuring: see diagram 2. measuring: turn the plug by 180 degrees, meaning exchange L for N. Notice the higher value. Limit value: 3.5 mA in the 1. error condition
10-2
MSTM/MMWT/4903
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 10 Check / Inspection Measuring circuit for accessible current according to NFPA 99 Measuring circuit in the reference device SECUTEST SIII
Test circuit B
A
1
R1
C1
R3
Voltmeter
R5
R2
R4
V
2
B Plug
Socket
N
N
L
L
PE
PE
A
N L
PE
Test circuit B
1
Centrifuge
10 Check / Inspection
Example
B
2 V
The accessible current shall be measured with circuit B. 1. measuring: see diagram 2. measuring: turn the the plug by 180 degree, meaning exchange L for N. Notice the higher value. Limit value: 0.5 mA
MSTM/MMWT/4903
10-3
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 10 Check / Inspection IQ – Installation Qualification OQ – Operational Qualification PQ – Performance Qualification for Eppendorf Centrifuge 5702 / 5702 R / 5702 RH This description is a suggestion from Eppendorf AG for the IQ/OQ/PQ. Only a trained service technician may install and commission the instrument and conduct the IQ/OQ/PQ tests. Only the latest updated version of the Service Manual is applicable. The current operating instructions enclosed should be kept handy and used to operate the instrument.
10 Check / Inspection
• • • •
10-4
MSTM/MMWT/4903
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 10 Check / Inspection IQ – Installation Qualification Installing and commissioning the centrifuge 5702 / 5702 R / 5702 RH 1. Check the condition of the transport packaging The delivery box is labeled with the following content upon receipt: Yes
No
Was the correct model delivered Centrifuge 5702 Centrifuge 5702 R Centrifuge 5702 RH
Yes
No
Which voltage version was delivered 230 Volt, 50 – 60 Hz 120 Volt, 50 – 60 Hz 100 Volt, 50 – 60 Hz
Yes
No
Yes
No
Yes
No
Yes
No
Is the transport packaging undamaged
2. Check delivery The sticker on the shipping package supplies information about the contents Check that the information on the sticker corresponds with the ordered version.
3. Ensure that the location for setting up the device is stable prior to installation Note: The weight of the Centrifuge 5702 with a maximum load is The weight of the Centrifuge 5702 R with a maximum load is The weight of the Centrifuge 5702 RH with a maximum load is
25 kg 41 kg 41 kg
The stability of the location for setting up is adequate
4. Unpacking the devices Procedure Description 4.1 Transport packaging opened. 4.2 Centrifuge removed from the delivery box, Remove 2 styrofoam side parts and protective sleeve. 4.3 Place device on the work surface. Note: Keep the transport packaging for later transport. The original transport packaging is to be used when returning the device to the manufacturer.
MSTM/MMWT/4903
10-5
10 Check / Inspection
The information on the sticker corresponds with the ordered version
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 10 Check / Inspection 5. Check the contents of the delivery for completeness Carry out the following operation steps, follow the operating instructions.
10 Check / Inspection
Quantity 1 1 1 1 1
Description Centrifuge 5702 without rotor Mains cable Operating instructions Allen key Set of fuses
1 1 1 1 1 1
Centrifuge 5702 R without rotor Mains cable Operating instructions Allen key with rotor key holder Tray for condensation water Set of fuses
1 1 1 1 1
Centrifuge 5702 RH without rotor Mains cable Operating instructions Allen key with rotor key holder Set of fuses
Yes
No
Yes
No
6. Setting up the device Note: Following transport, the device must be at room temperature for at least 4 hours before it is operated. Required space for Centrifuge 5702: Required space for Centrifuge 5702 R: Required space for Centrifuge 5702 RH:
320 x 395 x 241 mm (width x depth x height) 381 x 581 x 268 mm (width x depth x height) 381 x 581 x 268 mm (width x depth x height)
At least approx. 30 cm must be kept free around the device. Do not expose device to direct sunshine. Ambient temperature for the Centrifuge 5702: Ambient temperature for the Centrifuge 5702 R: Ambient temperature for the Centrifuge 5702 RH:
2 to 40 °C, max. 10 – 75 % relative air humidity 10 to 40 °C, max. 10 – 75 % relative air humidity 10 to 40 °C, max. 10 – 75 % relative air humidity
Procedure Description 6.1 Visual inspection, centrifuge has not suffered any transport damage. 6.2 Visual inspection, all edges of housing free of burr, free of anything which might cause injury. 6.3 Check stability of lab bench. 6.4 Check horizontal alignment of bench. 6.5 Do the voltage values of 100 – 240 V ± 10 % in the lab agree with the ID plate. 6.6 Do the frequency values of 50 – 60 Hz ± 1 Hz in the lab agree with the IFD plate. 6.7 Do not expose the centrifuge to direct sunlight. 6.8 Inspect positioning height, max. 2000 m above sea level 6.9 Connect mains cable, press ON switch. 6.10 Activate the Open key. Open the centrifuge lid by hand.
10-6
MSTM/MMWT/4903
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 10 Check / Inspection Instrument entries Location of der Centrifuge 5702 / 5702 R / 5702 RH: (building, floor, laboratory etc.): ....................................................... ............................................................................................................................................................................................... Model: ..............................................................................
Serial number: .................................................................
Power supply in volts: ......................................................
Frequency in Hz: .............................................................
Customer’s inventory number: ......................................... Date of installation, day/month/year:................................ Location of operating instructions:........................................................................................................................................
7. Operating the instrument The customer can use this form to become familiar with the Centrifuge 5702 / 5702 R /5702 RH. Carry out the following operating steps, follow the operating instructions. Yes
No
10 Check / Inspection
Procedure Description Function test 5702 / 5702 R / 5702 RH 7.1 Control and operating units 7.2 Mount rotor 7.3 Evenly load rotor with adapters and load 7.4 Centrifuge with time setting 7.5 Change the centrifuge parameters during the run 7.6 Short centrifugation 7.7 Continuous run 7.8 RCF display and calculation 7.9 Switch the brake function on and off 7.10 Only 5702 R / 5702 RH: at set rpm 7.11 Switching the signal tone on and off 7.12 Fixed setting of parameters 7.13 Only 5702 R / 5702 RH: programming the fixed programs 7.14 Only 5702 R / 5702 RH: program selection 7.15 Only 5702 R: cooling, set temperature, Fast Temp, standby cooling unit 7.16 Only 5702 RH: heating and cooling units, temperature profile, set temperature, Fast Temp, standby cooling unit 7.17 Calling up and leaving the customer service display 7.18 Open centrifuge with power cut off 7.19 Instrument fuses
The customer............................................................................................................................................. was initiated Remarks:................................................................................................................................................................................. ................................................................................................................................................................................................ ................................................................................................................................................................................................
Signature of technician: ....................................................
Date: ................................................................................
Signature of customer: .....................................................
Date: ................................................................................
MSTM/MMWT/4903
10-7
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 10 Check / Inspection OQ – Operational Qualification 8. Testing the performance data of the Centrifuge 5702 / 5702 R /5702 RH when putting into operation Function test Device number:
10 Check / Inspection
Procedure 8.1 8.2 8.3 8.4 8.5 8.6 8.7 8.8 8.9 8.10 8.11
8.12
8.13 8.14
Description Accessible current < 3.5 mA Accessible current < 0.5 mA (only valid for USA according to NFPA 99) Protective conductor< 0.2 E Insulation resistance > 2 ME Acceleration time < 25 s Deceleration time < 25 s Overspeed cut-off (test carried out without rotor) = Error 7 Noise level < 58 dB (A) Speed 4.400 1/min ± 44 1/min. Check timer, 5 minutes, tolerance ± 2 s 5702 R / RH only: check temperature – cooling Use rotor F-35-30-17 or rotor A-4-38. Load each bucket with 15 ml of ethylene glycol. Set temperature at 6 °C. Speed can be set at 1,000 rpm to 4,400 rpm. Measuring time after > 60 minutes’ running time. Result: 6 °C ± 2 °C, at room temperature 23 °C ±2 °C 5702 RH only: check temperature – heating Use rotor F-35-30-17 or rotor A-4-38. Create temperature profile. Load each bucket with 15 ml of ethylene glycol. Set temperature at 37 °C. Speed can be set at 2,200 rpm. Measuring time after > 75 minutes running time. Result: 37 °C ± 2 °C, at room temperature 23 °C ±2 °C Check imbalance switch: load rotor with G1 on one side = no error Check imbalance switch: load rotor with G2 on one side = Error: Imbalance Table for checking imbalance: Rotor Load A-4-38 4 empty round buckets A-4-38 4 empty rectangular buckets F-35-30-17 A-8-17 Positions 1 – 20 occupied by empty sleeves
G1 7g 7g 8g 10 g
Yes
No
G2 15 g 15 g 17 g 21 g
Remarks:................................................................................................................................................................................. ................................................................................................................................................................................................ ................................................................................................................................................................................................
Signature of technician: ....................................................
Date: ................................................................................
Signature of customer: .....................................................
Date: ................................................................................
10-8
MSTM/KW/2104
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 10 Check / Inspection PQ – Performance Qualification Periodic check on performance data of the Centrifuge 5702 / 5702 R /5702 RH 9. Performance data Device number:
Procedure 9.1 9.2 9.3 9.4 9.5 9.6 9.7 9.8 9.9 9.10 9.11
9.13 9.14
Table for checking imbalance: Rotor Load A-4-38 4 empty round buckets A-4-38 4 empty rectangular buckets F-35-30-17 A-8-17 Positions 1 – 20 occupied by empty sleeves
MSTM/KW/2104
G1 7g 7g 8g 10 g
Yes
No
10 Check / Inspection
9.12
Description Accessible current < 3.5 mA Accessible current < 0.5 mA (only valid for USA according to NFPA 99) Protective conductor< 0.2 E Insulation resistance > 2 ME Acceleration time < 25 s Deceleration time < 25 s Overspeed cut-off (test carried out without rotor) = Error 7 Noise level < 58 dB (A) Speed 4,400 1/min ± 44 1/min. Check timer, 5 minutes, tolerance ± 2 s 5702 R / RH only: check temperature – cooling Use rotor F-35-30-17 or rotor A-4-38. Load each bucket with 15 ml of ethylene glycol. Set temperature at 6 °C. Speed can be set at 1,000 rpm to 4,400 rpm. Measuring time after > 60 minutes’ running time. Result: 6 °C ± 2 °C, at room temperature 23 °C ±2 °C 5702 RH only: check temperature – heating Use rotor F-35-30-17 or rotor A-4-38. Create temperature profile. Load each bucket with 15 ml of ethylene glycol. Set temperature at 37 °C. Speed can be set at 2,200 rpm. Measuring time after > 75 minutes running time. Result: 37 °C ± 2 °C, at room temperature 23 °C ±2 °C Check imbalance switch: load rotor with G1 on one side = no error Check imbalance switch: load rotor with G2 on one side = Error: Imbalance
G2 15 g 15 g 17 g 21 g
10-9
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 10 Check / Inspection 10. Maintenance of the Centrifuge 5702 / 5702 R /5702 RH Procedure 10.1 10.2 10.3 10.4 10.5 10.6 10.7 10.8 10.9 10.10 10.11 10.12
Description Clean exterior surfaces of the centrifuge with neutral dishsoap solution. Clean rotor chamber with a damp cloth. Wash the rubber seal in the rotor chamber with water and rub in glycerine. Clean motor shaft with a damp cloth and then dry. Clean rotor cone with a damp cloth and then dry. Clean rotors, buckets, sleeves, caps and adapters, remove residues of centrifuged product. Autoclave if necessary (121 ˚C, 20 min) Clean swing-bucket rotor pivots and grooves and grease lightly with pivot grease. Subject rotors to a visual inspection for cracks, damage and corrosion, including inside the rotor bores, take out of commission if necessary. Check date of rotor manufacture, max. service life 7 years. Replace aerosol-tight caps incl. silicone sealing ring as well as rubber plates and rectangular bucket adapters and replacement rings if these are visibly worn. Only 5702 R / 5702 RH: empty condensation water tray and clean. Carefully remove broken glass and splinters.
Yes
No
10 Check / Inspection
Remarks:................................................................................................................................................................................. ................................................................................................................................................................................................ ................................................................................................................................................................................................
Maintenance
Repair
Date of next PQ: ...............................................................
Signature of technician: ....................................................
Date: ................................................................................
Signature of customer: .....................................................
Date: ................................................................................
10-10
MSTM/KW/2104
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 10 Check / Inspection Annex Date
Measure or amendment performed, enclose copy of documents
10 Check / Inspection
MSTM/MMWT/4903
10-11
eppendorf Service Manual
10 Check / Inspection
Centrifuge 5702 / 5702 R / 5702 RH – 10 Check / Inspection
10-12
MSTM/MMWT/4903
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 11 Descriptions General layout The centrifuge 5702 / 5702 R / 5702 RH comprises the following main electronic assemblies:
•
Control PCB with LCD and control elements, central control unit, frequency converter interface, lid latch control, imbalance recognition Temperature control (only with 5702 R / 5702 RH)
•
Converter assembly with integrated rotational speed recording and actuation of a 3-phase asynchronous motor and electronic brake regulation with external braking resistor (only with 5702 R)
•
3-phase asynchronous motor with integrated Hall sensor for rotational speed recording and motor-temperature switch
Functional description of assemblies Voltage supply
11 Descriptions
The voltage is supplied via a mains input module with integrated mains power fuses [230 V version: 2x 2.5 A time-lag (phase and zero)]. [100/120 V version: 2x 5.0 A time-lag (phase and zero)]. A two-pole mains power switch is used for power isolation.
PCB Control The power to the PCB is supplied by the mains power supply unit on the converter board via connector X1. The stabilized 5 V voltage supplies the logic assemblies. The power to the lid motor control is provided by a 15 V voltage supply. The 5 V voltage is monitored by IC4, which triggers a RESET signal for the processor if the voltage falls below the permissible minimum. The functions of this assembly are: –
Input and output of all device running parameters
–
Monitoring of all device running parameters
–
Communication with the converter
Lid latch The lid is locked by catches moved by a geared motor. If the current consumption increases above a permissible level, such as in the event of a blockage, the motor is shut down. The acknowledge signals of a microswitch and a Reed contact are used to evaluate the lock/unlocked status of the lid. The Reed contact is triggered if the lid is properly closed. The microswitch trips if the catches are correctly engaged. The switches are connected via X2 to the Control PCB.
MSTM/MMWT/4903
11-1
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 11 Descriptions Converter communication Data is exchanged between the converter board and Control PCB, based on the RS232 interface. An agreed data protocol is applied to transmit the rotational speed, acceleration and braking ramps, start/stop functions and others.
Display The operating parameters are displayed by way of a custom LCD panel. Current running parameters are stored in an EEPROM. Communication over a bus system is based on a standardized data protocol. The current running parameters are retained when the centrifuge is switched off.
Converter assembly This assembly forms the central core of the centrifuge, because it accommodates the drive of the 3-phase asynchronous motor and the voltage supply to all assemblies. This PCB comprises the following main assemblies: - Mains power input unit with PFC - Switched power supply unit for voltage supply to the control assemblies - Power supply unit for the motor drive - Processor control with interface to the Control PCB
11 Descriptions
Rotational speed recording and regulation assembly An assembly with a Hall-IC, mounted on the underside of the motor, is used to record the rotational speed. For this, a permanent magnet is attached to the motor shaft which generates an alternating magnetic field at the Hall sensor when rotation occurs. The resultant pulse sequence is used to record and regulate the rotational speed in the converter assembly.
Imbalance recognition assembly The imbalance recognition function is implemented mechanically by means of a normally closed momentary contact switch. The switch is mounted on the ground plate of the centrifuge. In case of an imbalance, the momentary contact switch opens and a high signal is connected to the processor (P3.7). At the same time the LH edge of the monoflop (IC8) is evaluated and converted into a pulse (t > 10ms) which is likewise connected to the processor (P1.2).
Temperature control (only with 5702 R / 5702 RH) The temperature is measured in the rotor chamber (chamber temperature) and on the condenser (condenser temperature). The measuring is done by two NTC resistors. Their resistance in series is measured as variable voltage and is converted by the ADC of the microcontroller. The NTC resistors are connected to the Control PCB via X3.
•
Chamber temperature The NTC resistor is mounted in a molded sleeve at the bottom of the rotor chamber.
•
Condenser temperature Another NTC resistor is mounted on a pipe of the condenser. The temperature measured is used to optimize the cooling, to monitor the temperature of the condenser as well as to control the speed of the condenser fan.
Controlling the condenser fan (only with 5702 R / 5702 RH) The condenser fan is driven by a phase-angle control. This allows adjustment of the speed according to the condenser temperature.
11-2
MSTM/MMWT/4903
eppendorf Service Manual
Centrifuge 5702 / 5702 R / 5702 RH – 11 Descriptions Heating function (only 5702 RH) Two heating cables are irremovably mounted around the rotor chamber. The power for the function is supplied by the PCB converter module. With 120 volt devices, both of the heating cables are operated parallel. With 230 volt devices they are operated in series. The resistance of each of the heating cables id approx. 160 Ohm and is measurable with an external measuring device. A temperature sensor has also been integrated for the software control system and is not exchangeable. Another non-exchangeable temperature monitor is mounted on the rotor chamber for safety purposes. The temperature sensor of the rotor chamber and the condenser sensor of the cooling unit continue to be exchangeable for the Centrifuge 5702 RH.
11 Descriptions
MSTM/MMWT/4903
11-3
eppendorf Service Manual
11 Descriptions
Centrifuge 5702 / 5702 R / 5702 RH – 11 Descriptions
11-4
MSTM/MMWT/4903
eppendorf ist eine eingetragene Marke SM Centrifuge 5702 / 5702 R / 5702 RH 5702 920.002-04