BS EN ISO 4064-3:2014 Incorporating corrigendum November 2014  

BSI Standards Publication

Water meters for cold potable water and hot water Part 3: Test report format

 

BS EN ISO 4064-3:2014

BRITISH STANDARD

National foreword This British Standard is the UK implementation of EN ISO 4064-3:2014. The UK participation in its preparation was entrusted by Technical Committee CPI/30, Measurement Measurement of fluid flow in closed conduits, to Subcommittee CPI/30/7, Volume flow-rate methods. A list of organizations represented on this subcommittee can be obtained on request to its secretary. This publication does not purport to include all the necessary provisions of a contract. Users are responsible for its correct application. © The British Standards Institution 2014. Published by BSI Standards Limited 2014 ISBN 978 0 580 88480 1 ICS 91.140.60

Compliance with a British Standard cannot confer immunity from legal obligations. This British Standard was published under the authority of the Standards Policy and Strategy Committee on 30 June 2014.

Amendments/corrigenda issued since publication Date

Text affected

30 November 2014

Implementation of CEN Correction Notice 20 August 2014: Removal of supersession information from CEN and national forewords

30 November 2014

Implementation of CEN Correction Notice 15 October 2014: Corrected French title in CEN title page

 

EUROPEAN STANDARD

EN ISO 4064-3

NORME EUROPÉENNE EUROPÄISCHE NORM 

June 2014

ICS 91.140.60 

English Version

Water meters for cold potable water and hot water - Part 3: Test report format (ISO 4064-3:2014) Compteurs d'eau potable froide et d'eau chaude - Partie 3: Format du rapport d'essais (ISO 4064-3:2014)

Wasserzähler zum Messen von kaltem Trinkwasser und heißem Wasser - Teil 3: Format des Prüfberichtes (ISO 4064-3:2014)

This European Standard was approved by CEN on 21 September 2013. CEN members are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this European Standard the status of a national standard without any alteration. Up-to-date lists and bibliographical references concerning such national standards may be obtained on application to the CEN-CENELEC Management Centre or to any CEN member. This Standardofexists in member three official (English,and French, German). A version in any Management other language madehas by translation underEuropean the responsibility a CEN into versions its own language notified to the CEN-CENELEC Centre the same status as the official versions. CEN members are the national standards bodies of Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finland, Former Yugoslav Republic of Macedonia, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey and United Kingdom.

 

EUROPEAN COMMITTEE FOR STANDARDIZATION COMITÉ EUROPÉEN DE NORMALISATION EUROPÄISCHES KOMITEE FÜR NORMUNG

CEN-CENELEC Management Centre: Avenue Marnix 17, B-1000 Brussels

© 2014 CEN

All rights of exploitation exploitation in in any form and by any means reserved reserved worldwide for CEN national Members.

Ref. No. EN ISO 4064-3:2014 E

 

BS EN ISO 4064-3:2014

EN ISO 4064-3:2014 (E)

 

Foreword This document (EN ISO 4064-3:2014) has been prepared by Technical Committee ISO/TC 30 “Measurement of fluid flow in closed conduits” in collaboration with Technical Committee CEN/TC 92 “Water meters” the secretariat of which is held by SNV. This European Standard shall be given the status of a national standard, either by publication of an identical text or by endorsement, at the latest by December 2014, and conflicting national standards shall be withdrawn at the latest by June 2017.  Attention is drawn to the th e possibility that some s ome of the elements elem ents of this document may be the subject of patent rights. CEN [and/or CENELEC] shall not be held responsible for identifying any or all such patent rights. This document has been prepared under a mandate given to CEN by the European Commission and the European Free Trade Association.  According to the CEN-CENELEC Internal Regulations, the national standards organizations of the following countries are bound to implement this European Standard: Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finland, Former Yugoslav Republic of Macedonia, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey and the United Kingdom. Endorsement notice The text of ISO 4064-3:2014 has been approved by CEN as EN ISO 4064-3:2014 without any modification.

ii

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)  

Contents

Page

Foreword ........................................................................................................................................................................................................................................ iv Introduction. .................................................................................................................................................................................................................................v 1

Scope ................................................................................................................................................................................................................................. 1

2

Normative Normativ e references ...................................................................................................................................................................................... 1

3

Terms, deinitions, symbols, and abbreviated terms ..................................................................................................... 1

4

Type evaluation report  ..................................................................................................................................................................................  .................................................................................................................................................................................. 1 4.1 General ........................................................................................................................................................................................................... 1 4.2 Information concerning the type ............................................................................................................................................ 2 4.3 General information concerning the test equipment  .........................................................................................  ......................................................................................... 12 4.4 Check list for water meter examination examinationss and performan performance ce tests ............................................................ 13 4.5 Type evaluation tests (for all water meters) .............................................................................................................. 23

4.6

Type evaluation tests (for electronic water meters and mechanical water meters with electronic components)................................................................................................................................................................ 47

5

Initial veriication report .......................................................................................................................................................................... ..........................................................................................................................................................................68 5.1 General ........................................................................................................................................................................................................ 68 5.2 Information concerning the EUT veriied ..................................................................................................................... 68 5.3 Initial veriication test report (ISO 4064-2:2014|OIML R 49-2:2013, Clause 10) ..................... 68

 Annex A (normative)  List of documents concerning the type (ISO 4064‑1:2014|OIML R 49‑1:2013, 7.2.9) .............................................................................................................................................................................................................................. 73  Annex B (normative)  Listing of test equipment used in examinations and tests ...............................................74

© ISO 2014 – All rights reserved

 

iii

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)  

Foreword ISO (the International Organization for Standardization) is a worldwide federation of national standards bodies (ISO member bodies). The work of preparing International Standards is normally carried out through ISO technical committees. Each member body interested in a subject for which a technical committee has been established has the right to be represented on that committee. International

organizations, governmental and non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely with the International Electrotechnical Commission (IEC) on all matters of electrotechnical standardization. The procedures used to develop this document and those intended for its further maintenance are

described in the ISO/IEC Directives, Part 1. In particular the different approval criteria needed for the different types of ISO documents should be noted. This document was drafted in accordance with the editorial rules of the ISO/IEC Directives, Part 2, www.iso.org/directives www.iso.org/directives.. Attention is drawn to the possibility that some of the elements of this document may be the subject of patent rights. ISO shall not be held responsible for identifying any or all such patent rights. Details of any patent rights identiied during the development of the document will be in the Introduction and/or on the ISO list of patent declarations received, www.iso.org/patents www.iso.org/patents.. Any trade name used in this document is information given for the convenience of users and does not constitute an endorsement.

The committees commit tees responsible for this document are Tmethods echnical includi Commng Committe ittee e ISO/TC ISO/TC 30,  and Measurement of luid  low in closed conduits, conduits , Subcommittee SC 7, Volu Volume meTechnical including water meters and meters OIML Technical Subcommittee TC 8/ 8/SC SC 5 Water meters.

This fourth edition of ISO 4064-3 cancels and replaces the third edition (ISO 4064-3:2005), which has been technically revised. Provisions of the third edition are addressed in ISO 4064-2:2014|OIML R 492:2013.

ISO 4064 consists of the following parts, under the general title Water meters for cold potable water and hot water : — Part 1: Metrological and technical requirements — Part 2: Test methods — Part 3: Test report format  4064 -1 — Part 4: Non-metrological requirements not covered in ISO 4064— Part 5: Installation Inst allation requirements requirements

This edition of ISO 4064-3 is identical to the corresponding edition of OIML R 49-3, which has been issued concurrently. OIML R 49-3 was approved for inal publication by the International Committee of Legal Metrology at its 48th 48 th meeting in Ho Chi Minh City Cit y, Vietnam in October 2013. It will be submitted to the International Conference on Legal Metrology in 2016 for formal sanction.

iv 

 

© ISO 2014 – All rights reserved

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)  

Introduction Implementation of this test report format is informative with regard to the implementation of

ISO 40644064 -1|OIML R 49-1 49-1 and ISO 4064-2|O 406 4-2|OIML IML R 49-2 in national regulat r egulations; ions; however, however, its implementation i mplementation is required within the framework of the OIML Certiicate System for Measuring Instruments [ISO 4064-2:2014|OIML R 49-2:2013, 11.1]. Clause 4 shows the required format of a type evaluation report for a complete or combined water meter.

A type evaluation report for a separable calculator (including indicating device) or a measurement transducer (including low or volume sensor) requires a similar format. However, some modiications to the tables may be required because a large number of variations in the design of these separable units is possible. Some examples of tables for presenting the test results for separable units are shown in Clause 5 for 5 for

initial veriications. These tables can also be adapted for type evaluation reports. reports.

© ISO 2014 – All rights reserved

 

v

 

BS EN ISO 4064-3:2014

 

BS EN ISO 4064-3:2014

INTERNATIONAL ST STANDARD ANDARD

ISO 4064-3:2 4064-3:2014(E) 014(E)

Water meters for cold potable water and hot water —

Part 3:  Test report format  1

Scope

This part of ISO 4064|OIML R 49 speciies a test report format to be used in conjunction with ISO 4064-1|OIML R 49-1 and ISO 4064-2|OIML R 49-2 for water meters for cold potable water and hot water.

2

Normative references

The following documents, in whole or in part, are normatively referenced in this document and are indispensable for its application. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies. ISO 4064-1:2014|OIML R requirements 49-1:2013, Water meters for cold potable water and hot water — Part 1: Metrological and technical ISO 4064-2:2014|OIML R 49-2:2013, Water meters for cold potable water and hot water — Part 2: Test methods

3

Terms, deinitions, symbols, and abbreviated terms

For the purposes of this document, the terms and deinitions given in ISO 4064-1|OIML R 49-1 apply. Some symbols and abbreviated terms used in the tables are as follows. +

pass



fail

n/a EU T

not applicable equipment under test  

H

horizont al

MAP

maximum admissible pressure

MAT

maximum admissible temperature

MPE

maximum permissible error

V

vertical

4

Type evaluation report 

4.1 General For each examination and test the checklist shall be completed according to this example: © ISO 2014 – All rights reserved

 

1

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)  

+

 

 



×  n/a 

4.2

Pass  × 

Fail 

n/a 

Not applicable 

Information concerning the type

4.2.1

General

Application number:

_________________________________

Applicant:

_ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __

Authorized 

_________________________________

representative: Address:

_________________________________ _________________________________ _________________________________ _________________________________

Testing

_________________________________

laboratory: Authorized 

_________________________________

representative: Address:

_________________________________ _________________________________ _________________________________ _________________________________

4.2.2

Model submitted

New model:

_________________________________

Variant of approved model(s): Approval number:

_________________________________

Variation of approved model:

_________________________________

See Table 1.

2

 

© ISO 2014 – All rights reserved

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)   Table 1 — Model submitted Yesa

Submit ted for approva l tes t s

Noa

Remarks

Mechanical water meter (complete) Mechanical water meter (combined) Electronic Electr onic water meter (com (complete) plete) Electronic water meter (combined)

Family of water meters Separable calculator (including indicating device)

Separable measurement transducer (including low or volume sensor)

Supplementary electronic device(s) for testing (permanently attached to meter) Supplementary electronic device(s) for data trans mission (permanently attached to meter) Supplementary electronic device(s) for testing (tem porarily attached to meter) Supplementary electronic device(s) for data trans mission (temporarily attached to meter) Ancillary devices a 

Tick as appropriate.

4.2.3

Mechanical water meter (complete or combined)

Manufacturer:

__________________________________________________

Model number:

__________________________________________________

Type details: Q1

_________ m3/h

Q2

_________ m3/h

Q3

_________ m3/h

Q4

_________ m3/h

Q3/Q1

_________

© ISO 2014 – All rights reserved

 

3

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)    

for combination meters  

Qx1

_________ m3/h

 

Qx2

_________ m3/h

Measuring principle:

__________________________

Accuracy class: Temperat ure class:

_____________ _____________

Env ironment al class:

_____________

Elec El ectr tro omagn gne eti ticc envi virronm nme ent:

____________ ______ ________

Maximum admissibl admissiblee 

______ ___ °C °C

temperature:

Maxim Max imum um ad admis missi sibl blee pr pres essu sure re::

______ _____ MPa MPa (_____ ba barr)

Orient at ion limit at ion:

_________________________

EUT testing t esting requirements (ISO ( ISO 4064-2: 4064 -2:201 2014|OIML 4|OIML R 49-2:2013, 49-2:2013, 8.1.8): Category:

_________________________

Case:

_________________________

Installation details:

Connection type (lange, screw thread, concentric manifold):

_________

Minimum straight length of inlet ______ ____ mm mm pipe:

Minimum straight length of out - ______ ____ mm mm let pipe: Flow conditioner (detail (detailss if required):

_________________________

Mounting:

_________________________

Orient ation:

_________________________

Oth the er re relevan antt in inform rma ati tio on:

______ ___ ______ ______ ______ ______ ______ ______ _____

NOTE

If a fami family ly of meters is submitted submitted,, the detai details ls in this subclause are to be given for each size of water

meter.

4

 

© ISO 2014 – All rights reserved

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)   4.2.4

Electronic water meter (complete or combined)

Manufacturer: __________________________________________________ Model number: __________________________________________________ Type details: 3

Q1

 

Q2

_________ m /h _________ m3/h

Q3

_________ m3/h

Q4

_________ m3/h

Q3/Q1

_________

for combination meters  

Qx1

_________ m3/h

 

Q x2

_________ m3/h

Measuring principle: Accuracy class:

__________________________

Temperat ure class:

_____________

Env ironment al class:

_____________

Elect romag net ic env ironment:

_____________

Maximum admissible temperature:

______ ___ °C °C

Ma ximum admissible pressure:

_ _ _ _ _ _ _ _ _ __ __ MPa (_ _ _ _ _ bar)

Orient at ion limit at ion:

_________________________

_____________

EUT testing t esting requirements (ISO ( ISO 4064-2: 4064 -2:201 2014|OIML 4|OIML R 49-2:2013, 49-2:2013, 8.1.8): Category: _________________________ Case:

_________________________

© ISO 2014 – All rights reserved

 

5

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)   Installation details (mechanical):

Connection type (lange, screw thread, th read, 

_________

concentric manifold):

Minimum straight length of inlet pipe:

______ ____ mm mm

Minimum straight length of outlet pipe:

______ ____ mm mm

Flo Fl ow co con ndi diti tio oner (det etai ails ls if re req qui uire red d):

_________________________

Mounting:

_________________________

Orient at ion:

_________________________

Ot her relevant information:

_________________________

Installation details (electrical): Wiring in i nst ruct ions:

_________________________

Mounting arrangement:

_________________________

Orient at ion limit at ions:

_ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __

Power supply: Type (battery, mains AC, mains DC):

_________________________

U max max :

______________________ V

U min min :

______________________ V

Frequenc y:

_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ Hz

NOTE

If a fami family ly of meters is submitted submitted,, the detai details ls in this subclause are to be given for each size of water

meter.

4.2.5

Separable calculator (including indicating device)

Manufacturer:

__________________________________________________

Model number:

__________________________________________________

Type details:

6

Q1

_________ m3/h

Q2

_________ m3/h

Q3

_________ m3/h

Q4

_________ m3/h

Q3/Q1

_________

 

© ISO 2014 – All rights reserved

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)    

for combination meters  

Qx1

_________ m3/h

 

Q x2

_________ m3/h

Measuring principle:

__________________________

Accuracy class: Temperat ure class:

_____________ _____________

Env ironment al class:

_____________

Elect romag net ic env ironment:

_____________

Maximum admissible temperature:

_________°C

Ma ximum admissible pressure:

_ _ _ _ _ _ _ _ _ __ __MPa (_ _ _ _ _ bar)

Orient at ion limit at ion:

_________________________

EUT testing t esting requirements (ISO ( ISO 4064-2: 4064 -2:201 2014|OIML 4|OIML R 49-2:2013, 49-2:2013, 8.1.8): Category: Case:

_________________________ _________________________

Maximum relative error speciied by the manufacturer: manufact urer: Lower low rate zone, Q1 ≤ Q < Q2:

_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _%

Upper low rate zone, Q2 ≤ Q ≤ Q4:

_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _%

Installation details (electrical): Wiring in i nstruct ions:

_________________________

Mounting arrangement:

_________________________

Orient at ion limit at ions:

_ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __

Power supply: Type (battery, mains AC, mains DC):

_________________________

U max max :

________________________ V

U min min :

________________________ V

Frequenc y:

_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ Hz

Approval number(s) of compatible meas-

urement transducer( tra nsducer(s) s) (including (including low  low or volume sensor):

© ISO 2014 – All rights reserved

____________________

 

7

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)   4.2.6

Separable measurement transducer (including low or volume sensor)

Manufacturer: __________________________________________________ Model number: __________________________________________________ Type details: 3

 

Q1 Q2

_________ m /h _________ m3/h

Q3

_________ m3/h

Q4

_________ m3/h

Q3/Q1

_________

for combination meters  

Qx1

_________ m3/h

 

Qx2

_________ m3/h

Measuring principle: Accuracy class:

__________________________

Temperat ure class:

_____________

Environment al class:

_____________

Elect romagnetic env ironment:

_____________

Maximum admissible temperature:

_________°C

Ma ximum admissible pressure:

_ _ _ _ _ _ _ _ _ _ _MPa (_ _ _ _ _ bar)

Orient at ion limit at ion:

_________________________

_____________

EUT testing t esting requirements (ISO ( ISO 4064-2: 4064 -2:201 2014|OIML 4|OIML R 49-2:2013, 49-2:2013, 8.1.8): Category:

_________________________

Case:

_________________________

Maximum relative error speciied by the manufacturer:

8

Lower low rate zone, Q1 ≤ Q < Q2:

_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _%

Upper low rate zone, Q2 ≤ Q ≤ Q4:

_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _%

 

© ISO 2014 – All rights reserved

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)   Installation details (mechanical):

Connection Conne ction type t ype (lange, ( lange, screw thread, concentric concentric mani -

_________

fold):

Minimum straight length of inlet pipe:

__________mm

Minimum straight length of outlet pipe:

__________mm

Flow condit ioner (det ails if required):

_________________________

Mounting:

_________________________

Orient at ion:

_________________________

Other relevant informat ion:

_________________________

Installation details (electrical): Wiring in i nst ruc t ions:

_________________________

Mounting arrangement:

_________________________

Orient at ion limit at ions:

_ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __

Power supply: Type (battery, mains AC, mains DC):

_________________________

U max max :

________________________ V

U min min :

________________________ V

Frequency:

_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ Hz

Approval number(s) of compatible calculator(s) (including in indicat ing de dev ice): _________________________

4.2.7

Supplementary electronic device(s) used for testing (permanently attached to meter) Manufacturer: __________________________________________________

Model number:

__________________________________________________

Power supply: Type (battery, mains AC, mains DC):

_________________________

U max max :

________________________ V

U min min :

________________________ V

Frequenc y:

_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ Hz

© ISO 2014 – All rights reserved

 

9

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)   Installation details (electrical): Wiring in i nst ruct ions:

_________________________

Mounting arrangement:

_________________________

Orient at ion limit at ions:

_ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __

4.2.8 Supplementary electronic device(s) used for for data transmission (permanently attached to meter)

Manufacturer:

__________________________________________________

Model number:

__________________________________________________

Power supply: Type (battery, mains AC, mains DC):

_________________________

U max max :

________________________ V

U min min :

________________________ V

Frequenc y:

_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ Hz

Installation details (electrical): Wiring in i nst ruct ions:

_________________________

Mounting arrangement:

_________________________

Orient at ion limit at ions:

_ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __

4.2.9

Supplementary electronic device(s) used for testing (temporaril (temporarily y attached attached to to meter) meter)

Manufacturer:

__________________________________________________

Model number:

__________________________________________________

Power supply: Type (battery, mains AC, mains DC):

_________________________

U max max :

________________________ V

U min min :

________________________ V

Frequenc y:

_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ Hz

10

 

© ISO 2014 – All rights reserved

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)   Installation details (electrical): Wiring in i nstruct ions:

_________________________

Mounting arrangement:

_________________________

Orient at ion limit at ions:

_ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __ __

4.2.10 Supplementary electronic device(s) used for data transmission (temporarily (temporarily attached to meter)

Manufacturer:

__________________________________________________

Model number:

__________________________________________________

Power supply: Type (battery, mains AC, mains DC):

_________________________

U max max :

________________________ V

U min min :

________________________ V

Frequenc y:

_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ Hz

EUT testing t esting requirements (ISO ( ISO 4064-2: 4064 -2:201 2014|OIML 4|OIML R 49-2:2013, 49-2:2013, 8.1.8): Category:

_________________________

Case:

_________________________

Installation details (electrical): Wiring inst ruc t ions:

_________________________

Mounting arrangement:

_________________________

Orient at ion limit at ions:

_________________________

4.2.11 Ancillary devices

Manufacturer:

__________________________________________________

Model number:

__________________________________________________

© ISO 2014 – All rights reserved

 

11

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)  

Power supply: Type (battery, mains AC, mains DC):

_________________________

U max max :

________________________ V

U min min :

________________________ V

Frequenc y:

_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ Hz

Approval number(s) of compatible calculator(s) (including in indicat ing de dev ice): _________________________

EUT testing t esting requirements (ISO ( ISO 4064-2: 4064 -2:201 2014|OIML 4|OIML R 49-2:2013, 49-2:2013, 8.1.8): Category:

_________________________

Case:

_________________________

Installation details (electrical): Wiring inst ruct ions:

_________________________

Mounting arrangement:

_________________________

Orient ation limit ations:

_________________________

Approval number(s) of compatible water

meters, calculator(s) (including indicating device) and measurement transducer( t ransducer(s) s)

(including low  low or volume sensor): sensor):

_________________________

4.2.12 Documents concerning concerning the type

A list of documents shall be submitted with the type approval application as in Annex A. A. 4.3

General information concerning the test equipment 

Details of all items of measuring equipment and test instruments used for the type examinations, and initial veriications shall be listed in Annex B, including:  

Manufacturer

 

Model number

 

Serial number

 

Date of last calibration

 

Date of next calibrat calibration ion due of e.g. inst instruments ruments for measuring: — linear dimensions — pressure gauges — pressu pressure re transmitters — manometers — temperature transducers

12

 

© ISO 2014 – All rights reserved

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)   — reference meters — volume tank s — weighing machines

— signa signall generators (for pulse, current or voltage) 4.4

Check list for water meter examinations and performance tests

4.4.1

Check list for water meter examinations External examination for all w ater meters

ISO 4064-1:2014|  OIML R 49-1:2013,

Requirement

+   −

Remarks

subclause Function of the indicating device

6.7.1.1

The indicating device shall provide an easily read, reliable and unambiguous visual indication of the indicated volume

6.7 6. 7.1. 1.1 1

The ind The indic icat atin ing g dev devic ice e sha shall ll in incl clud ude e vis visua uall mea means ns fo forr tes testi ting ng and calibratio ca libration. n.

6.7.1.1

The indicating device may include additional elements for testing and calibration by other methods, e.g. for automatic testing and c alibration Unit of measurement and its placement 

6.7 6. 7.1. 1.2 2

The in The indi dica cate ted d vo volu lum me of of wat water er sh shal alll be be exp expre ress ssed ed in cu cubi bicc metres

6.7.1.2

The symbol m3 shall appear on the dial or immediately adja cent to the numbered display Indicating range

6.7.1. 3

For Q3 ≤ 6,3, the minimum indicating range is 0 m 3 to 9 999 m3

6.7.1.3

For 6,3 < Q3 ≤ 63, the minimum indicating range is 0 m3 to 99 999 m3

6.7.1. 3

For 63 < Q3 ≤ 630, the minimum indicating range is 0 m3 to

6.7.1. 3

999 999 m3 For 630 < Q3 ≤ 6 300, the minimum indicating range is 0 m3  to 9 999 999 m3 Colour coding for indicating device

6.7 6. 7.1. 1.4 4

The co The colour bla lack ck sh sho oul uld d be be use used d to to in indi dica cate te th the e cub cubic ic met etre re and its multiples

6.7 6. 7.1. 1.4 4

The co The colo lour ur re red d sh shou ould ld be us used ed to in indi dica cate te su subb-m mul ulti tipl ples es of a cubic metre

6.7.1.4

The colours shall be applied to either the pointers, indexes, numbers, wheels, discs, dials or aperture frames

6.7.1.4

Other means of indicating the cubic metre may be used provided there is no ambiguity in distinguishing between the primary indication and alternative displays, e.g. submultiples for veriication and testing Types of indicating device: Type 1 — Analogue device

© ISO 2014 – All rights reserved

 

13

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)  

External examination for all water meters

ISO 4064-1:2014|  OIML R 49-1:2013,

Requirement

+   −

Remarks

subclause

The indicated volume shall be shown by continuous move -

6.7.2.1

ment of either: a)

one or more pointers moving relative to graduated

scales; or b) one or more circu circular lar scales or drums each passing an index 6.7 6. 7.2 .2..1

The val The value ue ex exp pre ress ssed ed in cu cub bic met etre ress fo for eac each h sca scalle div divis isio ion n n shall be of the form 10 , where n is a positive or a negative

whole number or zero, thereby establishing a system of consecutive decades. 6.7 6. 7.2 .2..1

The Th e sca scalle sha shall ll be gr grad adua uate ted d in in val valu ues ex exp pre ress ssed ed in cu cub bic

metres or accompanied by a multiplying factor (×0,001; ×0,01; ×0,1; ×1; ×10; ×100; ×1 000 etc.) 6.7 6. 7.2 .2..1

Rot otat atio iona nall mo movem emen entt of of the the po poin inte ters rs or cir circu cula larr sca scalles sh shal alll be clockwise

6.7 6. 7.2 .2..1

Line Li near ar movem emen entt of of po poin inte ters rs or sca scalles sh shal alll be be lef leftt to to rig right  ht 

6.7.2.1

Movement of numbered roller indicators shall be upwards Types of indicating device: Type 2 — Digital device

The indicated volume is given by a line of digits appearing in

6.7.2.2

one or more apertures 6.7 6. 7.2 .2.2 .2

The Th e ad adva van nce of on one dig digit it sh shal alll be be co complet eted ed wh whil ile e th the dig digit it

of the next immediately lower decade changes from 9 to 0 6.7 6. 7.2 .2.2 .2

The Th e ap appa pare ren nt hei eigh ghtt of th the e di digi gits ts sha hall ll be at le leas astt 4 mm

6.7.2.2

For non-electronic devices, devices , movement of numbered roller indicators (drums) shall be upwards

6.7.2.2

For non-electronic devices, the lowest value decade may have a continuous movement, the aperture being large enough to permit a digit to be read without ambiguity

6.7.2.2

For electronic devices with non-permanent displays the vol ume shall be able to be displayed at any time for at least 10 s

6.7.2.2

For electronic devices, the meter shall provide visual checking of the entire display which shall have the following sequence:

— for seven segment type displaying all the elements (e.g. an “eights” test); and — for seven segment type blanking all the elements (a “blanks” test).

For graphical displays, an equivalent test is required to dem onstrate that display faults cannot result in any digit being misinterpreted. Each step of the sequence shall last at least 1 s Types of indicating device: Type 3 — Combination of analogue and digital devices

6.7.2.3

The indicated volume is given by a combination of type 1 and type 2 devices and the respective requirements of each shall apply Veriication devices — General requirements

14

 

© ISO 2014 – All rights reserved

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)  

External examination for all w ater meters

ISO 4064-1:2014|  OIML R 49-1:2013,

Requirement

+   −

Remarks

subclause 6.7.3.1

Every indicating device shall provide means for visual, nonambiguous ambiguo us veriication testing tes ting and calibration

6.7.3.1

The visual veriication may have either a continuous or a discontinuous movement 

6.7.3.1

In addition to the visual veriication display, an indicating device may include provisions for rapid testing by the inclusion of complementary elements (e.g. star wheels or discs), providingg signals through externally attached sensors. providin Veriication devices — Visual veriication displays

6.7.3.2.1

The value of the veriication veri ication scale interval, expressed in cubic metres, shall be of the form:1 × 10n, 2 × 10n or 5 × 10 n, where n is a positive or negative whole number, or zero

6.7.3.2.1

The indicated volume is given by a line of digits appearing in one or more apertures

6.7 6. 7.3 .3.2. .2.1 1

For an anal alog ogue ue or di digit gital al in indi dica catin ting g dev devic ices es wit with h con contin tinuo uous us

movement of the irst element, the veriication scale interval may be formed from the division into 2, 5 or 10 equal parts of the interval between two consecutive digits of the irst irs t

element. Numbering shall not be applied to these divisions 6.7 6. 7.3 .3.2. .2.1 1

For di digi gital tal in indi dica catin ting g devi evice cess wit with h di disc scon ontin tinuo uous us mov ovee-

ment of the irst element, the veriication scale interval is the interval between bet ween two consecutive digits or incremental incremental

movements of the irst element  6.7.3.2.2

On indicating devices with continuous movement of the irst element, the apparent scale spacing shall be not less than 1 mm and not more than 5 mm

6.7 6. 7.3 .3.2 .2.2 .2

The Th e sca calle sha hall ll consi sist st of eit ith her: a)

lines of equal thick thickness ness not exceeding one quart quarter er of

the scale spacing and differing only in length; or b) contrasting contrast ing bands of a constant width equal to the scale spacing 6.7 6. 7.3. .3.2.2 2.2

The app The appar aren entt wid width th of th the e poi point nter er at its ti tip p sh shall all no nott exc excee eed d one-quarter of the scale spacing and in no case shall it be

greater than 0,5 mm Resolution of the indicating device

6.7.3.2.3

The sub-divisions of the veriication scale shall be small enough to ensure that the resolution of the indicating device

does not exceed 0,25 % of the actual volume for accuracy class 1 meters, and 0,5 % of the actual volume for accuracy class 2 meters, for a 90 min test at the minimum low rate, Q1.

NOTE 1 When the display of the irst element is continuous, an allowance should be made for a maximum error in each

reading of not more than half of the veriication scale inter val.

NOTE 2 When the display of the irst element is discontinu ous, an allowance should be made for a maximum error in each reading of not more than one digit of the veriication scale

© ISO 2014 – All rights reserved

 

15

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)  

External examination for all water meters

ISO 4064-1:2014|  OIML R 49-1:2013,

Requirement

+   −

Remarks

subclause

NOTE

For combination meters with two indicati indicating ng devices, the above requirements apply to both indicati indicating ng

devices. Marks and inscriptions

6.6.1

A place shall be provided on the meter for afixing the verii cation mark, which shall be visible without dismantling the meter

6.6.2

The water meter shall be clearly and indelibly marked with the information listed in the following, either grouped or distributed on the casing, the indicating device dial, an identiication plate or on the meter cover if is not detachable

6.6. 6.2 2 a)

Unit of meas asu urement: cubic metr tre e

6.6.2 b)

The accuracy class, where it differs from accuracy class 2

6.6.2 c)

The numeric al value of Q3 and the ratio Q3/Q1 (may be

preceded by R). If the meter measures reverse low and Q3  and the ratio Q3/Q1 are different dif ferent in the two directions, both values of Q3 and Q3/Q1 shall be inscribed; the direction of low to which each pair of values refers shall be clear. If the meter has different values of Q3/Q1 in horizontal and verti cal positions, both values of Q3/Q1 shall be inscribed, and the orientation to which each value refers shall be clear 6.6.2 d)

The type approval sign according to national regulations

6.6. 6. 6.2 2 e)

The Th e na nam me or tr tra ademar ark k of th the e man anuf ufa act ctur ure er

6.6.2 f)

The year of manufacture (or the last two digits of the year of manufacture or the month and year of manufacture)

6.6. 6. 6.2 2 g) g)

The ser The seria iall num numbe berr (as (as nea earr as as po poss ssib ible le to th the e in indi dica cati ting ng device)

6.6.2 h)

The direction of low (shown on both sides of the body; or on one side only, provided the direction of low arrow is easily visible under all circumstances)

6.6.2 i)

The maximum admissible pressure (MAP) if it exceeds 1 MPa (10 bar) or 0,6 MPa (6 bar) for nominal diameter dia meter ≥500 mm. (The unit bar may be used where national regulations per mit)

6.6. 6. 6.2 2 j) j)

The le The lett tter er V or H, if th thee met meter er ca can n onl nlyy be ope pera rate ted d in th thee vertical or horizontal position

6.6. 6. 6.2 2 k) k)

The Th e te tempera ratu ture re cla lasss wh where it dif difffers fr fro om T3 T30

6.6.2 l)

The pressure loss class where it differs from Δ p 63

6.6.2 m)

The installation instal lation sensitivity class where it differs from f rom U0/ U0/ D0  Additional markings for water meters with electronic devices

6.6.2 n)

For an external power supply: the voltage and frequency

6.6.2 o)

For a replaceable battery: the latest date by which the bat tery shall be replaced

6.6.2 p)

For a non-replaceable battery: the latest date by which the meter shall be replaced

6.6.2 q)

16

Environmental Environm ental classiication

 

© ISO 2014 – All rights reserved

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)  

External examination for all w ater meters

ISO 4064-1:2014|  OIML R 49-1:2013,

Requirement

+   −

Remarks

subclause 6.6 6. 6. 2 r)

Elec tr tromag ne neti ticc en env ir ironmenta tall cl clas asss Protection devices

6.8. 6. 8.1 1

Wat ater er met eter erss sh shal alll in incl clu ude pro rote tecti ctio on dev evic ices es wh whic ich h ca can n be sealed so as to prevent, both before and after correct instal-

lation of the water meter, dismantling or modiication of the meter, its adjustment device or its correction device, without damaging these devices. In the case of combination meters, this requirement applies to both meters Protection devices — Electronic sealing devices

6.8.2.1

When access to parameters that t hat inluen in luence ce the determinatio determination n of the results of measurements is not protected by mechani cal sealing devices, the protection shall fulil t he following following provisions.

a) Access shall only be allowed to authorized people, e.g. by means of a code (password) or of a special device (hard key, etc.). The code shall be capable of being changed. b)

It shall be possible for at least the last intervention

to be memorized. The record shall include the date and a characteristic element identifying the authorized person making the intervention [see a)]. If it is possible to memorize more than one intervention and if a previous intervention

requires deletion to permit a new record, the oldest record shall be deleted 6.8. 6. 8.2. 2.2 2

For met eter erss wi with th pa parts rts wh whic ich h can can be di disc sco onn nnec ecte ted d one fr fro om another by the user and which are interchangeable, the fol-

lowing provisions provisions shall be fulilled: fu lilled: a) it shall not be possible to access parameter parameterss that participate in the determinatio determination n of results of measurements through disconnected points unless the provisions of

ISO 4064-1:2014|OIML R 49-1:2013, 6.8.2.1 are fulilled; b) interposing any device which may in inluence luence the accuracy shall be prevented by means of electronic and data processing securities or, if this is not possible, by mechanical means 6.8.2.3

For meters with parts which may be disconnected one from the other by the user and which are not interchangeable, the provisions in ISO IS O 4064-1 4064 -1:201 :2014|OIML 4|OIML R 49-1:2013, 49-1:2013, 6.8.2.2 6.8. 2.2 apply. Moreover, these meters shall be provided with devices which do not allow them to operate if the various parts are

not connected according to the approved type. NOTE

Disconnections Disconnect ions which are not allowed to the user

may be prevented, e.g. by means of a device that prevents any measurement after disconnecting and reconnecting Examination and testing of checking facilities General requirements for examining checking facilities

© ISO 2014 – All rights reserved

 

17

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)  

External examination for all water meters

ISO 4064-1:2014|  OIML R 49-1:2013,

Requirement

+   −

Remarks

subclause 5.1. 5. 1.3 3

Wat ater er met eter erss wi with th el elec ectr tro oni nicc dev eviice cess sha hall ll be pro rovi vid ded wi with th

the checking facilities speciied in ISO 4064-1:2014|OIML R 49-1:2013, Annex B, except in the case of non-resettable 5.1. 5. 1.3 3

measurements between t wo constant partners measurements Alll met Al eter erss eq equi uip ppe ped d wi with th ch chec ecki king ng fa faci cili liti ties es sh shal alll pre rev ven entt or

detect reverse low, as laid down in ISO 4064-1:2014|OIML R 49-1:2013, 4.2.7 4.2.7.. 4.4.2

Checklist for water meter performance tests

4.4.2.1

Performance tests for all water meters

ISO 4064-1:201 4064-1:2014| 4|  OIML R 49-1:20 49-1:2013, 13,

Requirement

+   −

Remarks

subclause  Static pressure pre ssure test  te st 

4.2. 4. 2.1 10

The met The mete er sh shal alll be be ca capa pabl ble e of of wit withs hsta tan ndi ding ng th the e fo foll llo owi wing ng te test st pressures without leakage or damage:

— 1,6 times the maximum admissible pressure for 15 min; —

2 times the maxi maximum mum admissible pressure for 1 min Intrinsic errors (of indication)

18

 

© ISO 2014 – All rights reserved

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)  

ISO 4064 4064-1 -1:2014 :2014||  OIML R 49-1:20 49-1:2013, 13,

Requirement

+   −

Remarks

subclause 7.2 .2.3 .3

The Th e er erro rors rs (of in indi dica cati tion on)) of th the e wa wate terr met eter er (i (in n th the e mea easu sure re--

ment of the actual volume), shall be determined at least at the following low rate ranges: a)

Q1 to 1,1Q1;

b)

c)

Q2 to 1,1 Q2; 0,33(Q2 + Q3) to 0,37(Q2 + Q3);

d)

0,67(Q2 + Q3) to 0,74(Q2 + Q3);

e)

0,9Q3 to Q3;

f)

0,95Q4 to Q4;

and for combination meters:; g)

0,85Qx1 to 0,95Q x1;

h)

1,05Q x2 to 1,15Q x2.

The water meter should be tested without its temporary supplementary pleme ntary devices attached (if any). During a test all other in luence factors shall be held at reference conditions.

Other low rates may be tested depending on the shape of the error curve. 1) The relative errors (of indicat ion) observed for each

of the low rates shall not exceed the t he maximum permissible errors (MPEs) given in ISO 4064-1:2014|OIML R 49-1:2013, 4.2.2 or 4.2.3. If the error observed on one or more meters is greater

than the MPE at one low rate only, then if only two results have been taken at that low rate, the test at that t hat low rate shall be repeated. The test shall be declared satisfactory if two out of the three results at that low rate lie within the MPE and the arithmetic mean of the results for the three tests at t hat low rate lies within the MPE. 2)

If all the relative errors (of indicat ion) of the water

meter have the same sign, at least one of the errors shall not exceed one-half one-half of the MPE. In all cases, this t his requirement shall be applied equitably with respect to the water supplier and the consumer (see also ISO 4064-1:2014|OIML R 49-1:2013, 4.3.3 3) and 8) 7.2 .2.4 .4

The Th e met eter er sh shal alllat bethe repe re peat atab able le:: th the e st stan anda dard rd devi de viat atio ion none-third of th thre ree e measurements measureme nts t he same low rate shall not exceed

of the MPEs given in ISO 4064-1:2014|OIML R 49-1:2013, 4.2.2 or 4.2.3. Tests shall be carr ied out at nominal low rates of Q1, Q2, and Q3 Water temperature test 

4.2.8

The requirements relating to the MPEs shall be met for all water temperature variations wit hin the rated operating conditions of temperature the meter Water pressure test 

4.2.8

The requirements relating to the MPEs shall be met for all water pressure variations within t he rated operating conditions of the meter Reverse low test 

© ISO 2014 – All rights reserved

 

19

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)  

ISO 4064-1:201 4064-1:2014| 4|  OIML R 49-1:20 49-1:2013, 13,

Requirement

+   −

Remarks

subclause 4.2.7

A water meter designed to measure reverse low shall either: a) subtract the reverse low volum volumee from the indicated volume; or b)

record the reverse low volume separately.

The MPEs of ISO 4064-1:2014|OIML R 49-1:2013, 4.2.2 or 4.2.3 shall be met for both forward and reverse low 4.2.7

A water meter not designed to measure reverse low shall either:

a)

prevent it; or

b) be capable of withst withstanding anding an accidental reverse low at a low rate up to Q3 without any deterioration or change in it s metrological properties for forward  low Meter characteristics at zero  low rate

4.2.9

The water meter totalization shall not change when the t he low rate is zero Pressure loss test 

6.5

The pressure loss of the water meter, including its ilter where the latter forms an integral part of the water meter, shall not be greater than 0,063 MPa (0,63 bar) between Q1 and Q3 Flow disturbance test 

6.3.4

If the accuracy of water meters is affected by disturbances in the upstream or downstream pipeline, the meter shall be provided with suficient straight straig ht pipe lengths with or without a low str aightener (as (as speciied by the t he manufacturer) so that the indications of the installed water meter do not exceed MPEs according to the accuracy class of the meter. Forward low tests Reverse low tests (where applicable) Overload temperature test 

7.2.5

Water meters with MAT ≥50 °C shall be capable of withstanding withst anding a water temperature of MAT + 10 °C for 1 h Durability Durabili ty test s

7.2.6

The water meter shall undergo a durability test according to to the permanent low rate Q3 and the overload low rate Q4 of the meter,, simulating service conditions meter

7.2.6

Meters with Q3 ≤ 16 m3/h:

7.2.6

a)

100 000 low cycles between 0 and Q3;

b)

100 h at Q4

Meters with Q3 > 16 m3/h: a)

800 h at Q3;

b)

200 h at Q4;

and for combination meters:

c)

20

50 000 low cycles between Q ≥ 2Qx2 and 0

 

© ISO 2014 – All rights reserved

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)  

ISO 4064 4064-1 -1:2014 :2014||  OIML R 49-1:20 49-1:2013, 13,

Requirement

+   −

Remarks

subclause 7.2.6.2

Accuracy class 1 meters The variation in the error cur ve shall not exceed 2 % for low rates in the lower zone (Q1 ≤ Q < Q2) and 1 % for low rates in the upper zone (Q2 ≤ Q ≤ Q4). For the of (of these requirements, arithmetic E   for the value ofpurpose the errors indication) each low rate mean shall apply. For low rates in the lower low rate zone (Q1 ≤ Q < Q2), the error (of indication) curve shall not exceed a maximum error limit

7.2.6.3

of ±4 % for all temperature classes. For low rates in the upper low rate zone (Q2 ≤ Q ≤ Q4), the error (of indication) curve shall not exceed a maximum error limit of ±1,5 % for meters of temperature class T30 and ±2,5 % for all other temperature classes Accuracy class 2 meters The variation in the error cur ve shall not exceed 3 % for low rates in the lower zone (Q1 ≤ Q < Q2) and 1,5 % for low rates in the upper zone (Q2 ≤ Q < Q4). For the purpose of these requirements, the arithmetic mean value of the errors (of indication) E   for each low rate shall apply. Q2), the error For low rates in the lower low rate zone (Q1 ≤ Q  16 m3/h, total hours run = 800 h at Q3 (minimum volume discharged at end of test is [ Q3] × 800, expressed in m3, where [Q3] is the number equal to the value of Q3, expressed in m3/h) and 200 h at Q4 (minimum volume discharged discha rged at end of test is [Q4] × 200, expressed in m3) where [Q4] is the number equal to the value of Q4, expressed in m3/h).

 Ambient conditions at inish i nish A mbient temperat ure °C

A mbient relative

humidity

Ambient atmospheric pressure

%

MPa (bar)

Obse Ob serv rver er:: ___ ______ ______ ______ Da Date te:: ___ ______ ______ ______ ____ Errors (of indication) measured after the continuous low test 

Time

40

 

© ISO 2014 – All rights reserved

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)   Meter Met er serial serial No.:____ No.:____ ___ ___ Actual

Working

Working

Initial

Final

Indicated

Actual

Meter

lowrate

pressure

temp

reading

reading

volume

volume

error

MPEa

Curve

MPE

variation

(of curve

error b

variation error)

Q( ) m3/h

 p w

MPa

T w °C

V i(i) m3

V i(f) m3

V i m3

V a m3

E m %

c

Ē m(B) - Ē m(A)

  %

%

%









(bar)

d

Ē m2 m2 Ē m3 m3 Ē m(B) Comments:



For MPE values refer to ISO 4064-1:2014 4064-1:2014|OIML |OIML R 49-1:20 49-1:2013, 13, 4.2. For accepta acceptance nce criteria criteri a refer to ISO 4064-2:2014|OIML 4064-2:2014|OIML R 49-2:2013, 7.4.5. Ē m(A) is the Mean intrinsic intri nsic error (of indication). indication). See test report 5.3. Ē  5.3. Ē m(B) is the mean error (of indication) measured after this thi s continuous low test (= either Ē m2  or Ē  ). m2 m3 m3 b  c 

For MPE values and accepta nce crit criteria eria refer to ISO 4064-2:2014|OIML 4064-2:2014|OIML R 49-2:2013, 7. 7.11.3.4. 11.3.4.

 Q = Q1, Q2 or Q3 or if the irst or second test is outside the MPE (ISO 4064-2:2014|OIML R 49-2:2013, Perform third test if  Q 7.4.5) d 

4.5.10.3 Discontinuous low test (ISO 4064‑2:2014|OIML 4064‑2:2014|OIML R 49‑2:2013, 49‑2:2013, 7.11.2) (Applicable only to combination meters) Application No. Test method: met hod:

Gravimetric/volumetric

Volume Vol ume measures/weighbridge measur es/weighbridge used — m 3 or kg: Water conductivity (electromagnetic induction meters only) — S/cm: Length of straight pipe before meter (or manifold) — mm: Length of straight pipe after meter (or manifold) — mm: Nominal diameter DN of pipe before and after meter (or manifold) — mm:

Describe low straightener installation if used: Speciied changeover low rate Qx2 Selected test low rate (minimum is twice the changeover low rate Qx2) Readings taken during the test Meter serial Meter serial No.:____ No.:____ ___ ___ NOTE Readings are recorded every 24 h or once for every shorter period, if so divided.

41

 

© ISO 2014 – All rights reserved

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)    Ambient conditions at start  sta rt  A mbient temperature

A mbient relative

humidity

Ambient atmospheric pressure

%

MPa (bar)

°C

Dat e

Time

Observer

Up

Down

Up

stream

stream pressure

stream

MP a

°C

pressure

MPa ( bar)

Actual

Meter

Time

Flow cycle times - s

lowrate reading

temp m3/h

m3

Total Total volume dis- no. of low charged

cycles r i se

on

f a ll

of f

m3

(bar)

Totals at end of test = Theoretical total a = a 

Minimum theoretical volum volumee passed by meters during the test is 0.5 × Q1 × 50000 × 32 / 3600 expressed in m 3. Minimum number of test cycles during the test = 50000.

 Ambient conditions at inish i nish A mbient temperature °C

A mbient relative

humidity

Ambient atmospheric pressure

%

MPa (bar)

Comments: Obse Ob serv rver er:: ______ ______ ______ ___

Date Da te:: ______ ______ ______ ______ _

Errors (of indication) measured after the discontinuous low test 

Time

42

 

© ISO 2014 – All rights reserved

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)   Meter Met er serial serial No.:____ No.:____ ___ ___ Actual

Working

Working

Initial

Final

Indicated

Actual

Meter

lowrate

pressure

temp

reading

reading

volume

volume

error

MPEa

Curve

MPE

variation

(of curve

error

b

variation error)

Q( )

 pw

T w

V i(i)

V i(f)

V i

V a

E m

 

Ē m(B) - Ē m(A)

m3/h

MPa (bar)

°C

m3

m3

m3

m3

%

%

%

%









d

c

Ēm2 Ēm3 Ēm(B) Comments:



For MPE values refer to ISO 4064-1:2014 4064-1:2014|OIML |OIML R 49-1:20 49-1:2013, 13, 4.2. For accepta acceptance nce criteria criteri a refer to ISO 4064-2:2014|OIML 4064-2:2014|OIML R 49-2:2013, 7.4.5. Ē m(A) is the Mean intrinsic intri nsic error (of indication). indication). See test report 5.3. Ē  5.3. Ē m(B) is the mean error (of indication) measured after this thi s discontinuous low test (= either Ē m2 m2 or Ē m3 m3). b  c 

For MPE values and accepta nce crit criteria eria refer to ISO 4064-2:2014|OIML 4064-2:2014|OIML R 49-2:2013, 7. 7.11.3.4. 11.3.4.



Perform third test if Q  = Q1, Q2 or Q3  or if the irst or second test is outside the MPE (ISO 4064-2:2014|OIML

R 49-2:2013, 7.4.5).

4.5.11 Static magnetic ield test (ISO 4064‑2:2014|OIML 4064‑2:2014|OIML R 49‑2:2013, 49‑2:2013, 7.12, 8.16)  

At start

At end

Application No:

______________

Ambient temperature:

°C

Model:

______________

Ambient relative humidity:

%

Date: Observer:

______________ ______________

Ambient atmospheric pressure: Time:

Gravimetric/volumetric

Test method: met hod:

Volume Vol ume measures/weighbridge measur es/weighbridge used — m 3 or kg:: Water conductivity (electromagnetic induction meters only) — S/cm: Length of straight pipe before meter (or manifold) — mm: Length of straight pipe after meter (or manifold) — mm: Nominal diameter DN of pipe before and after meter (or manifold) — mm:

Describe low straightener installation if used: Meter Met er seria seriall No. No.:___ :___ ___ ___ _

MPa

Orientati Orie ntation on (V (V,, H, oth other er):____ ):____ ___ ___ ___ _

43

 

© ISO 2014 – All rights reserved

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)   Flow direction (see Requirement Requirement 1): 1): _ __ __ Location of indicating device (see Requirement Requirement 2):___ _ Application conditions

Nominal

Actual

Initial pressure

Initial inlet water temperature

MPa (bar)

°C

low rate low rate m3/h

  Location 1

Q3

Location 2 (optional)

Q3

Location 3 (optional)

Q3

m3/h

supply

Initial Final reading reading

Indicated volume

Actual volume

Meter MP MPE Ea error

V i(i)

V i(f)

V i

V a

E m

m3

m3

m3

m3

%

%

Comments: Note location of magnet  a 

For a complete water meter this is the maximum permissible error as deined in ISO 4064 4064-1 -1:2014 :2014|OIML |OIML R 49-1:20 49-1:2013, 13, 4.2.2 or 4.2.3 according to the accuracy class of the meter. If the EUT is a separable sub-assembly, the MPE shall be deined by the manufacturer (ISO 4064-2:2014|OIML R 49-2:2013, 9.4).

Requirements

Requirement Require ment 1:

If the low axis is vertical, the low direction (from bottom to top or from top to

bottom) shall be given.

Requirement 2: If the low ax axis is is horizonta horizontall and the meter has an indicat indicating ing device which is integral with the body of the meter, the location of the indicating device (at the side or at the top of the meter) shall be given. 4.5.12 Tests on ancillary anci llary devices of a water meter (ISO (IS O 4064‑2:2014 4064‑2:2014|OIML |OIML R 49‑2:2013, 49‑2:2013, 7.13 7.13))  

At start

At end

Application No:

______________

Ambient temperature:

°C

Model:

______________

Ambient relative humidity:

%

Date:

______________

Ambient atmospheric pressure:

Observer:

______________

Time:

Test method: met hod:

Volume measures/weighbridge used — m3 or kg: Water conductivity (electromagnetic induction meters only) — S/cm: Length of straight pipe before meter (or manifold) — mm: Length of straight pipe after meter (or manifold) — mm: Nominal diameter DN of pipe before and after meter (or manifold) — mm:

Describe low straightener installation if used: u sed: Meter Met er serial serial No.:___ No.:___ ___ ___ _ Orientatio Orientation n (V, (V, H, other):____ other):____ ___ ___ ___ _

MPa

Gravimetric/volumetric

44

 

© ISO 2014 – All rights reserved

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)   Flow direction (see (see Requirement 3): 3): __ __ __ Location of indicating device (see (see Requirement 4):__ 4):____ __ Actual

Initial

Water

Initial

Final

Indicated

Actual

Meter

lowrate

supply

temp.

reading

reading

volume

volume

error

MPEa

pressure Q ( )

 

T w

V i(i)

V i(f)

V i

V a

E m

 

m3/h

MPa(bar)

°C

m3

m3

m3

m3

%

%

 

Standard deviation

MPE/3a

 

%

%

b

Ē m2 m2 Ē m3 m3

s cd

 



For a complete water meter this is the maximum permissible error as deined in ISO 4064-1:2014 4064-1:2014|OIML R 49-1:2013, 4.2.2 or 4.2.3 according to the accuracy class of the meter. If the EUT is a separable sub-assembly the MPE shall be deined by the manufacturer (ISO 4064-2:2014|OIML R 49-2:2013, 9.4). For acceptance criteria refer to ISO 4064-2:2014|OIML R 492:2013, 7.4.5. Perform third test if Q = Q1, Q2 or Q3 or if the irst or second test is outside the MPE (ISO 4064-2:2014 |OIML R 49-2:2013, 49 -2:2013, 7.4.5). b  c 

Calculate standard deviation if Q = Q1, Q2 or Q3 (ISO 4064-2:2014|OIML R 49-2:2013, 7.4.5)



Standard deviation of three measuremen measurements ts of the error (o (off indication) indication) taken at the same nominal nominal lowrate

Meter Met er serial serial No.:___ No.:___ ___ ___ _ Orientati Orientation on (V, (V, H, other):____ other):____ ___ ___ ___ _

45

 

© ISO 2014 – All rights reserved

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)   Flow direction (see (see Requirement 3): 3): __ __ __ Location of indicating device (see (see Requirement 4):___ 4):___ _ Actual

Initial

Water

Initial

Final

Indicated

Actual

Meter

lowrate

supply

temp.

reading

reading

volume

volume

error

MPEa

pressure Q ( )

 

T w

V i(i)

V i(f)

V i

V a

E m

 

m3/h

MPa(bar)

°C

m3

m3

m3

m3

%

%

Standard deviation

MPE/3a

%

%

b

Ē m2 m2 Ē m3 m3  

s cd

 



For a complete water meter this is the maximum permissible error as deined in ISO 4064-1:2014|OIML 4064-1:2014|OIML R 49-1:20 49-1:2013, 13, 4.2.2 or 4.2.3 according to the accuracy class of the meter. If the EUT is a separable sub-assembly the MPE shall be deined by the manufacturer (ISO 4064-2:2014|OIML R 49-2:2013, 9.4). For acceptance criteria refer to ISO 4064-2:2014|OIML R 492:2013, 7.4.5. Perform third test if Q = Q1, Q2 or Q3 or if the irst or second test is outside the MPE (ISO 4064-2:2014|OIML R 49-2:2013, 7.4.5). b  c 

Calculate standard deviation if Q = Q1, Q2 or Q3 (IS  (ISO O 4064-2:201 4064 -2:2014|OIML 4|OIML R 49-2:2013, 7.4.5)



Standard deviation of three measurements of the error (of (of indication) indication) taken at the same nominal lowrate.

Meter Met er serial serial No.:___ No.:___ ___ ___ _ Orientatio Orientation n (V, (V, H, other):____ other):____ ___ ___ ___ _

46

 

© ISO 2014 – All rights reserved

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)

Flow direction (see Requirement 3): ____ Location of indicating device (see Requirement 4):___ Actual

Initial

Water

Initial

Final

Indicated

Actual

Meter

lowrate

supply

temp.

reading

reading

volume

volume

error

MPEa

pressure Q ( )

 

T w

V i(i)

V i(f)

V i

V a

E m

 

m3/h

MPa(bar)

°C

m3

m3

m3

m3

%

%

 

Standard deviation

MPE/3a

 

%

%

b

Ē m2 m2 Ē m3 m3

s cd

 



For a complete water meter this is the maximum permissible error as deined in ISO 4064-1:201 4064-1:2014|OIML 4|OIML R 49-1:20 49-1:2013, 13, 4.2.2 or 4.2.3 according to the accuracy class of the meter. If the EUT is a separable sub-assembly the MPE shall be deined by the manufacturer (ISO 4064-2:2014|OIML R 49-2:2013, 9.4). For acceptance criteria refer to ISO 4064-2:2014|OIML R 492:2013, 7.4.5. Perform third test if Q = Q1, Q2 or Q3 or if the irst or second test is outside the MPE (ISO 4064-2:2014|OIML R 49-2:2013, 7.4.5). b  c 

Calculate standard deviation if Q = Q1, Q2 or Q3 (IS  (ISO O 4064-2:201 4064 -2:2014|OIML 4|OIML R 49-2:2013, 7.4.5)



Standard deviation of three measuremen measurements ts of the error (o (off indication) indication) taken at the same nominal nominal lowrate.

Requirements

Requirement 1:

Tables Ta bles for each  low rate according to ISO 4064-2:201 4064-2:2014|OIML 4|OIML R 49-2:2 49-2:2013, 013, 7.4.4

shall be added.

Requirement 2: Tables Ta bles for each orientat orientation, ion, which shall be as speciied in ISO 4064-2 4064 -2:201 :2014|OIML 4|OIML R 49-2:2013, 49-2:2013, 7.4.2.2 7.4.2.2.7 .7.5 .5 shall sha ll be provided for meters not marked m arked either ‘H’ or ‘V’. ‘V ’. Requirement Requirem ent 3:

If the low axis is vertical, the low direction (from bottom to top or from top to

bottom) shall be given.

Requirement 4: If the low axi axiss is horizontal and the meter has an indicati indicating ng device which is integral with the body of the meter, the location of the indicating device (at the side or at the top of the meter) shall be given. Comments:

4.6 Type evaluation evaluation tests (for (for electronic electronic water water meters meters and mechanical water meters with electronic components) 4.6.1

Dry heat (non‑condensing) (ISO 4064‑2:2014| 4064‑2:2014|OIML OIML R 49‑2:2013, 8.2)

 

At start

At end

Application No:

______________

Ambient temperature:

°C

Model:

______________

Ambient relative humidity:

%

Date:

______________

Ambient atmospheric pressure:

Observer:

______________

Time:

MPa

47

 

© ISO 2014 – All rights reserved

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)

Gravimetric/volumetric

Test method: met hod:

Volume measures/weighbridge used — m3 or kg: Water conductivity (electromagnetic induction meters only) — S/cm: Length of straight pipe before meter (or manifold) — mm: Length of straight pipe after meter (or manifold) — mm: Nominal diameter DN of pipe before and after meter (or manifold) — mm:

Describe low straightener installation if used: u sed: Meter Met er serial serial No.:___ No.:___ ___ ___ _ Orientatio Orientation n (V, (V, H, other):____ other):____ ___ ___ ___ _ Flow direction (see Requirement 1): ____ Location of indicating device (see Requirement 2):___ Application conditions

Actual or simulated

low rate m3/h

 

Meter MP MPE Eb

Working pressurea

Working temperaturea

Initial reading

Final reading

Indicated volume

Actual volume

 pw

T w

V i(i)

V i(j)

V i

V a

E m

MPa (bar)

°C

m3

m3

m3

m3

%

error

%

20 °C 55 °C 20 °C Comments: a 

Temperature and pressure shall be recorded using a data-logging device to ensure conformity conformity with the relevant

IEC standard. b 

For a complete complete water meter this is the maximum permissible error as deined deined in ISO 4064-1: 4064-1:201 2014|OIML 4|OIML R 491:2013, 4.2.2 or 4.2.3 1:2013, 4.2. 3 according to the accuracy class of the meter. If the EUT is i s a separable part of a water meter, the MPE shall be deined by the manufacturer (ISO 4064-2:2014|OIML R 49-2:2013, 9.4).

Requirements

Requirement Require ment 1:

If the low axis is vertical, the low direction (from bottom to top or from top to

bottom) shall be given.

Requirement 2:

If the low ax axis is is horizonta horizontall and the meter has an indicat indicating ing device which is

integral withbe thegiven. body of the meter, the location of the indicating device (at the side or at the top of the meter) shall 4.6.2

Cold (ISO 4064‑2:2014|OIM 4064‑2:2014|OIML L R 49‑2:2013, 8.3)

 

At start

At end

Application No:

______________

Ambient temperature:

°C

Model:

______________

Ambient relative humidity:

%

Date:

______________

Ambient atmospheric pressure:

Observer:

______________

Time:

MPa

48

 

© ISO 2014 – All rights reserved

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)

Gravimetric/volumetric

Test method: met hod:

Volume Vol ume measures/weighbridge measur es/weighbridge used — m 3 or kg: Water conductivity (electromagnetic induction meters only) — S/cm: Length of straight pipe before meter (or manifold) — mm: Length of straight pipe after meter (or manifold) — mm: Nominal diameter DN of pipe before and after meter (or manifold) — mm:

Describe low straightener installation if used: Environmen Enviro nmental tal class:___ class:___ ___ ___ _ Meter Met er seria seriall No. No.:___ :___ ___ ___ _

Orientati Orie ntation on (V (V,, H, oth other er):____ ):____ ___ ___ ___ _

Flow direction (see Requirement 1): ___ Location of indicating device (see Requirement 2):____ Application Actual or conditions simulated

Working pressure a

Working temperature a

Initial reading

Final reading

 pw

T w

V i(i)

V i(j)

V i

V a

E m

 

MPa (bar)

°C

m3

m3

m3

m3

%

%

low rate   m3/h

Indicated Actual Meter MPEb volume volume error

20 °C +5 °C or

−25 °C 20 °C Comments: a 

Temperature Tem perature and pressure shall be recorded using a data-logging device to ensure conformity with the relevant IEC

standard. b 

For a complete water meter this is the maximum permissible error as deined in ISO 4064-1 4064-1:201 :2014|OIML 4|OIML R 49-1:20 49-1:2013, 13, 4.2.2 4.2.2 or 4.2.3 according to the accurac y class of the meter. If the EUT is a separable part of a water meter, the MPE shall be deined by the manufacturer (ISO 4064-2:2014|OIML R 49-2:2013, 9.4).

Requirements

Requirement Requirem ent 1:

If the low axis is vertical, the low direction (from bottom to top or from top to

bottom) shall be given.

Requirement 2: If the low axi axiss is horizontal and the meter has an indicati indicating ng device which is integral with the body of the meter, the location of the indicating device (at the side or at the top of the meter) shall be given. 4.6.3

Damp heat, cyclic (condensing) (ISO 4064‑2:2014 4064‑2:2014|OIML |OIML R 49‑2:2013, 8.4)

 

At start

At end

Application No:

______________

Ambient temperature:

°C

Model:

______________

Ambient relative humidity:

%

Date:

______________

Ambient atmospheric pressure:

Observer:

______________

Time:

MPa

49

 

© ISO 2014 – All rights reserved

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)  

Gravimetric/volumetric

Test method: met hod:

Volume measures/weighbridge used — m3 or kg: Water conductivity (electromagnetic induction meters only) — S/cm: Length of straight pipe before meter (or manifold) — mm: Length of straight pipe after meter (or manifold) — mm: Nominal diameter DN of pipe before and after meter (or manifold) — mm:

Describe low straightener installation if used: u sed: Environmen Enviro nmental tal class:___ class:___ ___ ___ _ Meter Met er seria seriall No. No.:___ :___ ___ ___ _

Orientati Orie ntation on (V (V,, H, oth other er):____ ):____ ___ ___ ___ _

Flow direction (see Requirement 1): ____ Location of indicating device (see Requirement 2):___ Application conditions

Actual or simulated

low rate  

Q( )

m3/h

Working Working prestemperasure turea

Initial reading

Final reading

Indi- Actual cated volvolume ume

Meter error

MPEb Fault SigniiEUT cant functionfault  ing

correctly

 p w

T w

V i (i)

V i(j)

V i

V a

E m

 

E m2) m2)  − E m1) m1)

 

MPa

°C

m3

m3

m3

m3

%

%

%

%





(bar) Reference conditions

1) Before cycling





Precondition meter.

Apply damp heat cycles (duration (duration 24 h), two cycles between 25 °C and 40 °C (environmental class B) or 55 °C (environmental classes O and M). 2) A f ter c ycling

yes

no

Comments: a 

Temperature and pressure shall be recorded using using a data-logging data-logging device to ensure conformity with the relevant relevant IEC IEC

standard. b 

For a complete water meter this is the maximum permissible error as deined in ISO 4064 4064-1 -1:2014 :2014|OIML |OIML R 49-1:201 49-1:2013, 3, 4.2.2 or 4.2.3 according to the accuracy class of the meter. If the EUT is a separable part of a water meter, the MPE shall be deined by the manufacturer (ISO 4064-2:2014|OIML R 49-2:2013, 9.4).

Requirements

Requirement Require ment 1:

If the low axis is vertical, the low direction (from bottom to top or from top to

bottom) shall be given.

Requirement 2: If the low ax axis is is horizonta horizontall and the meter has an indicat indicating ing device which is integral with the body of the meter, the location of the indicating device (at the side or at the top of the meter) shall be given.

50

 

© ISO 2014 – All rights reserved

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)   4.6.4

Power Pow er supply variation (ISO 4064‑2:2014|OIML 4064‑2:2014|OIML R 49‑2:2013, 8.5)

4.6.4.1

General

 

At start

Application No:

______________

Ambient temperature:

Model: Date:

______________ ______________

Ambient relative humidity: Ambient atmospheric pressure:

Observer:

______________

Time:

At end °C % MPa

Gravimetric/volumetric

Test method:

Volume measures/weighbridge used — m3 or kg: Water conductivity (electromagnetic induction meters only) — S/cm: Length of straight pipe before meter (or manifold) — mm: Length of straight pipe after meter (or manifold) — mm: Nominal diameter DN of pipe before and after af ter meter (or manifold) — mm:

Describe low straightener installation if used:

4.6.4.2 Meters powered powered by direct direct AC (single phase) or or AC/DC AC/DC converters, converters, mains power power supply (ISO 4064‑2:2014|OIML R 49‑2:2013, 8.5.2) Meter Met er seria seriall No. No.:___ :___ ___ ___ _

Orientati Orie ntation on (V (V,, H, oth other er):____ ):____ ___ ___ ___ _

Flow direction (see Requirement 1): ___ Location of indicating device (see Requirement 2):____ Application U i conditions (single voltage)  

V

Actual or simulated

low rate

m3/h

Working Wor king pressure  pw

MPa (bar)

Working Initial temperature reading T w V i(i) °C

m3

Final reading V i(j) m3

Indicated Actual Meter MP MPE Ea volume volume error V i

V a

E m

m3

m3

%

%

U nom nom + 10 %  f nom nom + 2 % U nom nom − 15 %  f nom nom − 2 % Comments: a 

For a complete water meter this is the maximum permissible error as deined in ISO 4064-1:2014|OIML 4064-1:2014|OIML R 49-1:2013, 4.2. 4.2.2 2 or 4.2.3 according to the accuracy class of the meter. If the EUT is a separable part of a water meter, the MPE shall be deined by the manufacturer (ISO 4064-2:2014|OIML R49-2:2013, 9.4).

Requirements

Requirement Requirem ent 1:

If the low axis is vertical, the low direction (from bottom to top or from top to

bottom) shall be given.

51

 

© ISO 2014 – All rights reserved

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)  

Requirement 2: If the low ax axis is is horizonta horizontall and the meter has an indicat indicating ing device which is integral with the body of the meter, the location of the indicating device (at the side or at the top of the meter) shall be given. 4.6.4.3 Meters power powered ed by primary batteries batteries or by external external DC voltage voltage (ISO 4064‑2:2014|OIML R 49‑2:2013, 8.5.3) Meter seria Meter seriall No. No.:___ :___ ___ ___ _ Orie Orientati ntation on (V (V,, H, oth other er):____ ):____ ___ ___ ___ _ Flow direction (see Requirement 1): ___ Location of indicating device (see Requirement 2):____ Application con- U i ditions (single voltage)

Actual or simulated

low rate

V

 

Working pressure

m3/h

Working tem- Initial Final perature reading reading

Indicated volume

Actual volume

Meter MP MPE Ea error

 pw

T w

V i(i)

V i(j)

V i

V a

E m

MPa (bar)

°C

m3

m3

m3

m3

%

%

U max max U min min Comments: a 

For a complete water meter this is the maximum permissible error as deined in ISO 4064-1:20 4064-1:2014| 14|OIML OIML R 491:2013, 4.2.2 or 4.2.3 according to the accuracy class of the meter. If the EUT is a separable part of a water meter, the MPE shall be deined by the manufacturer (ISO 4064-2:2014|OIML R 49-2:2013, 9.4).

Requirements

Requirement Require ment 1:

If the low axis is vertical, the low direction (from bottom to top or from top to

bottom) shall be given.

Requirement 2: If the low ax axis is is horizonta horizontall and the meter has an indicat indicating ing device which is integral with the body of the meter, the location of the indicating device (at the side or at the top of the meter) shall be given. 4.6.5

Vibration (random) (ISO 4064‑2:2014 4064‑2:2014|OIML |OIML R 49‑2:2013, 8.6)

 

At start

At end

Application No:

______________

Ambient temperature:

°C

Model:

______________

Ambient relative humidity:

%

Date:

______________

Ambient atmospheric pressure:

Observer:

______________

Time:

Gravimetric/volumetric

Test method: met hod:

Volume measures/weighbridge used — m3 or kg: Water conductivity (electromagnetic induction meters only) — S/cm: Length of straight pipe before meter (or manifold) — mm: Length of straight pipe after meter (or manifold) — mm: Nominal diameter DN of pipe before and after meter (or manifold) — mm:

Describe low straightener installation if used: u sed: Environmen Enviro nmental tal class:___ class:___ ___ ___ _ Meter Met er seria seriall No. No.:___ :___ ___ ___ _

MPa

Orientati Orie ntation on (V (V,, H, oth other er):____ ):____ ___ ___ ___ _

52

 

© ISO 2014 – All rights reserved

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)   Flow direction (see Requirement 1): ___Location of indicating device (see Requirement 2):____ Application conditions

Actual or simulated

Working pressure

low rate Q( )

 p w

Working temperature T w

Initial Final Indi- Actual Meter MPE a read- read- cated vol- error ing ing volume ume V i(i)

V i(j)

V i

V a

E m

 

Fault

Sig-

nii-

EUT functioning

cant fault  correctly E m2) m2)  −

 

E m1) m1)

m3/h

MPa (bar)

°C

m3

m3

m3

m3

%

%

Reference conditions 1) Before vibrations

%

%









Apply random vibrations to the EUT, over the frequency range 10 Hz to 150 Hz, in three mutually perpendicular axes, for a period of at least 2 min per axis. Total RMS level: 7 m.s−2. ASD level at 10 Hz to 20 Hz = 1 m 2 .s−3 and at 20 Hz to 150 Hz = −3 dB/octave. yes

2) After vibrations

no

Comments: a 

For a complete water meter this is the maximum permissible error as deined in ISO 4064-1:201 4064-1:2014|OIML 4|OIML R 49-1:20 49-1:2013, 13, 4.2.2 or 4.2.3 according to the accuracy class of the meter. If the EUT is a separable part of a water meter, the MPE shall be deined by the manufacturer (ISO 4064-2:2014|OIML R 49-2:2013, 9.4).

Requirements

Requirement Requirem ent 1:

If the low axis is vertical, the low direction (from bottom to top or from top to

bottom) shall be given.

Requirement 2: If the low axi axiss is horizontal and the meter has an indicati indicating ng device which is integral with the body of the meter, the location of the indicating device (at the side or at the top of the meter) shall be given. 4.6.6

Mechanical shock (ISO 4064‑2:2014| 4064‑2:2014|OIML OIML R 49‑2:2013, 8.7)

 

At start

At end

Application No:

______________

Ambient temperature:

°C

Model:

______________

Ambient relative humidity:

%

Date:

______________

Ambient atmospheric pressure:

Observer:

______________

Time:

Gravimetric/volumetric

Test method: met hod:

Volume Vol ume measures/weighbridge measur es/weighbridge used — m 3 or kg: Water conductivity (electromagnetic induction meters only) — S/cm: Length of straight pipe before meter (or manifold) — mm: Length of straight pipe after meter (or manifold) — mm: Nominal diameter DN of pipe before and after meter (or manifold) — mm:

Describe low straightener installation if used: Environmen Enviro nmental tal class:___ class:___ ___ ___ _ Meter Met er seria seriall No. No.:___ :___ ___ ___ _

MPa

Orientati Orie ntation on (V (V,, H, oth other er):____ ):____ ___ ___ ___ _

53

 

© ISO 2014 – All rights reserved

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)   Flow direction (see Requirement 1): ___ Location of indicating device (see Requirement 2):____ Application conditions

Actual or Working simulated pressure

Initial reading

Final reading

 p w

T w

V i(i)

V i (j)

V i

V a

E m

 

E m2) m2)  − E m1) m1)

 

MPa

°C

m3

m3

m3

m3

%

%

%

%







low rate  

Q( )

m3/h

Indicated Actual Meter MPEa volume volume error

Working temperature

Fau lt

niicant

Sig-

EUT functioning

fault 

correctly

(bar) Reference conditions 1) Before shock



Place the EUT on a rigid level surface in its normal position of use and tilted towards one bottom edge until the opposite edge of the EUT

is 50 mm above the rigid surface. The angle made by the bottom of the EUT and the test sur face shall not exceed 30°. Allow the EUT to drop freely on to the rigid sur face. Repeat the test for each bottom edge of the EUT. yes

2) After shock 

no

Comments: a 

For a complete water meter this is the maximum permissible error as deined in ISO 4064-1:2014|OIML 4064-1:2014|OIML R 49-1:20 49-1:2013, 13, 4.2.2 or 4.2.3 according to the accuracy class of the meter. If the EUT is a separable part of a water meter, the MPE shall be deined by the manufacturer (ISO 4064-2:2014|OIML R 49-2:2013, 9.4).

Requirements

Requirement Require ment 1:

If the low axis is vertical, the low direction (from bottom to top or from top to

bottom) shall be given.

Requirement 2: If the low ax axis is is horizonta horizontall and the meter has an indicat indicating ing device which is integral with the body of the meter, the location of the indicating device (at the side or at the top of the meter) shall be given. 4.6.7 AC mains voltage dips, short short interruptions and voltage variations (ISO 4064‑2:2014|OIML R 49‑2:2013, 8.8)  

At start

At end

Application No:

______________

Ambient temperature:

°C

Model:

______________

Ambient relative humidity:

%

Date:

______________

Ambient atmospheric pressure:

Observer:

______________

Time:

Gravimetric/volumetric

Test method: met hod:

Volume measures/weighbridge used — m3 or kg: Water conductivity (electromagnetic induction meters only) — S/cm: Length of straight pipe before meter (or manifold) — mm: Length of straight pipe after meter (or manifold) — mm: Nominal diameter DN of pipe before and after meter (or manifold) — mm:

Describe low straightener installation if used: u sed: Meters powered by direct AC (single‑phase) mains power supply Meter Met er seria seriall No. No.:___ :___ ___ ___ _

MPa

Orientati Orie ntation on (V (V,, H, oth other er):____ ):____ ___ ___ ___ _

54

 

© ISO 2014 – All rights reserved

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)   Flow direction (see Requirement 1): ___ Location of indicating device (see Requirement 2):____ Application conditions

Actual or Working Working Initial simulated pressure tempera- readlow rate ture ing

Final reading

Indi- Actual Meter cated volume error volume

MPEa

Signii-

Fault

cant fault b

EUT functioning

correctly Q( )

 

m3/h

 p w

T w

V i (i)

V i(j)

V i

V a

E m

 

E m2) m2)  − E m1) m1)

 

MPa

°C

m3

m3

m3

m3

%

%

%

%





(bar) Reference conditions 1) Before voltage reductions

No voltage reductions.

4064-2:2014| 14|OIML OIML R 49-2:2013, 49-2:2013, 8.8 . 2) During voltage Voltage interruptions and reductions as in ISO 4064-2:20 reduction





yes

no

Comments: a 

For a complete water meter this is the maximum permissible error as deined in ISO 4064-1:201 4064-1:2014|OIML 4|OIML R 49-1:20 49-1:2013, 13, 4.2.2 or 4.2.3 according to the accuracy class of the meter. If the EUT is a separable part of a water meter, the MPE shall be deined by the manufacturer (ISO 4064-2:2014|OIML R 49-2:2013, 9.4). b 

The signiicant signiicant fault is equal to half the MPE MPE in the upper upper low rate zone.

Requirements

Requirement Requirem ent 1:

If the low axis is vertical, the low direction (from bottom to top or from top to

bottom) shall be given.

Requirement 2: If the low axi axiss is horizontal and the meter has an indicati indicating ng device which is integral with the body of the meter, the location of the indicating device (at the side or at the top of the meter) shall be given. 4.6.8

Bursts on signal lines (ISO 4064‑2:2014|O 4064‑2:2014|OIML IML R 49‑2:2013, 8.9)

 

At start

At end

Application No:

______________

Ambient temperature:

°C

Model:

______________

Ambient relative humidity:

%

Date:

______________

Ambient atmospheric pressure:

Observer:

______________

Time:

MPa

Gravimetric/volumetric

Test method: met hod:

Volume Vol ume measures/weighbridge measur es/weighbridge used — m 3 or kg: Water conductivity (electromagnetic induction meters only) — S/cm: Length of straight pipe before meter (or manifold) — mm: Length of straight pipe after meter (or manifold) — mm: Nominal diameter DN of pipe before and after meter (or manifold) — mm:

Describe low straightener installation if used: Meters containing electronics and provided with I/O and communication ports (including its external cables) Meter Met er seria seriall No. No.:___ :___ ___ ___ _

Orientati Orie ntation on (V (V,, H, oth other er):____ ):____ ___ ___ ___ _

55

 

© ISO 2014 – All rights reserved

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)   Flow direction (see Requirement 1): ___ Location of indicating device (see Requirement 2):____ Application conditions

Actual or Working simupressure lated

Working temperature

Initial reading

Final reading

Indicated volume

Actual Meter volume error

MPEa

Fault

Signiicant fault b

low rate

EUT functioning cor-

rectly  

Q( )

 p w

T w

V i(i)

V i(j)

V i

V a

E m

 

E m2) m2)  −

 

m3/h

MPa

°C

m3

m3

m3

m3

%

%

m1) E m1) %

%





(bar) Reference conditions 1) Before burst 





Each spike shall have an amplitude (positive or negative) of 0,5 kV for environmental class E1 inst ruments, or 1 kV for environmental class E2 ins truments (see ISO 4064 -2:2 -2:201 014|OIML 4|OIML R 49-2:2013, 49-2:2013, 8.1.3), phased randomly, with a r ise time of 5 ns a nd a half amplitude duration of 50 ns.

yes no

2) After burst  Comments: a 

For a complete water meter this is the maximum permissible error as deined in ISO 4064-1:2014|OIML 4064-1:2014|OIML R 49-1:20 49-1:2013, 13, 4.2.2 or 4.2.3 according to the accuracy class of the meter. If the EUT is a separable part of a water meter, the MPE shall be deined by the manufacturer (ISO 4064-2:2014|OIML R 49-2:2013, 9.4). b 

The signiicant signiicant fault is equal equal to half the MPE in the upper low rate zone.

Requirements

Requirement Require ment 1:

If the low axis is vertical, the low direction (from bottom to top or from top to

bottom) shall be given.

Requirement 2: If the low ax axis is is horizonta horizontall and the meter has an indicat indicating ing device which is integral with the body of the meter, the location of the indicating device (at the side or at the top of the meter) shall be given. 4.6.9

Bursts (transients) (transients) on AC and DC mains (ISO 4064‑2:2014 4064‑2:2014|OIML |OIML R 49‑2:2013, 8.10)

 

At start

At end

Application No:

______________

Ambient temperature:

°C

Model:

______________

Ambient relative humidity:

%

Date:

______________

Ambient atmospheric pressure:

Observer:

______________

Time:

Gravimetric/volumetric

Test method: met hod:

Volume measures/weighbridge used — m3 or kg: Water conductivity (electromagnetic induction meters only) — S/cm: Length of straight pipe before meter (or manifold) — mm: Length of straight pipe after meter (or manifold) — mm: Nominal diameter DN of pipe before and after meter (or manifold) — mm:

Describe low straightener installation if used: u sed: Meters powered by direct AC (single‑phase) mains power supply Meter Met er seria seriall No. No.:___ :___ ___ ___ _

MPa

Orientati Orie ntation on (V (V,, H, oth other er):____ ):____ ___ ___ ___ _

56

 

© ISO 2014 – All rights reserved

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)   Flow direction (see Requirement 1): ___ Location of indicating device (see Requirement 2):____ Application conditions

Actual or simulated

IndiActual cated volume volume

MPEa

Fault

Signii-

Working temperature

Initial reading

 pw 

w  T w 

V i (i)

V i (j)

V i

V a

E m

 

m2)  − m2) E E  m1) m1

 

MPa

°C

m3

m3

m3

m3

%

%

%

%





low

Final reading

Meter

Working pressure

error

cant fault b

EUT functioning

correctly

rate Q( )

 

m3/h

(bar) Reference conditions 1) Before burst 

With no signi icant noise in mains supply. supply. —



2) After Randomly phased bursts (electromagnetic environment, E1 — 1 000 V peak amplitude electromagnetic environment, E2 — 2 000 V pea k amplitude) applied applied asy nchronously in asymmetr ical mode (common mode). mode). burst 

yes

no

Comments: a 

For a complete water meter this is the maximum permissible error as deined in ISO 4064-1:201 4064-1:2014|OIML 4|OIML R 49-1:20 49-1:2013, 13, 4.2.2 or 4.2.3 according to the accuracy class of the meter. If the EUT is a separable part of a water meter, the MPE shall be deined by the manufacturer (ISO 4064-2:2014|OIML R 49-2:2013, 9.4). b 

The signiicant signiicant fault is equal to half the MPE MPE in the upper upper low rate zone.

Requirements

Requirement Requirem ent 1:

If the low axis is vertical, the low direction (from bottom to top or from top to

bottom) shall be given.

Requirement 2: If the low axi axiss is horizontal and the meter has an indicati indicating ng device which is integral with the body of the meter, the location of the indicating device (at the side or at the top of the meter) shall be given. 4.6.10 Electrostatic discharge discharge (ISO 4064‑2:2014|OIML 4064‑2:2014|OIML R 49‑2:2013, 49‑2:2013, 8.11)  

At start

At end

Application No:

______________

Ambient temperature:

°C

Model:

______________

Ambient relative humidity:

%

Date:

______________

Ambient atmospheric pressure:

Observer:

______________

Time:

Test method: met hod:

Volume Vol ume measures/weighbridge measur es/weighbridge used — m 3 or kg: Water conductivity (electromagnetic induction meters only) — S/cm: Length of straight pipe before meter (or manifold) — mm: Length of straight pipe after meter (or manifold) — mm: Nominal diameter DN of pipe before and after meter (or manifold) — mm:

Describe low straightener installation if used:

MPa

Gravimetric/volumetric

57

 

© ISO 2014 – All rights reserved

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)

Meter Met er seria seriall No. No.:___ :___ ___ ___ _

Orientati Orie ntation on (V (V,, H, oth other er):____ ):____ ___ ___ ___ _

Flow direction (see Requirement 1): ___ Location of indicating device (see Requirement 2):____ Test conditions

Actual or Working simulated preslow rate sure

Working temperature

Initial reading

Final reading

Indi- Actual Meter cated volerror volume ume

MPEa

Fault

Signi-

icant fault b

EUT functioning cor-

rectly Q( )

 

m3/h

 p w

T w

V i (i)

V i(j)

V i

V a

E m

 

E m2) m2)  − E m1) m1)

 

MPa

°C

m3

m3

m3

m3

%

%

%

%





(bar) Reference conditions (no discharges)



2) Discharge Mode d point c



yes no

C

A

yes no

C

A

yes no

C

A

yes no

C

A

yes no

Comments: a 

For a complete water meter this is the maximum permissible error as deined in ISO 4064 4064-1 -1:2014 :2014|OIML |OIML R 49-1:20 49-1:2013, 13, 4.2.2 or 4.2.3 according to the accuracy class of the meter. If the EUT is a separable part of a water meter, the MPE shall be deined by the manufacturer (ISO 4064-2:2014|OIML R 49-2:2013, 9.4). b 

The signiicant signiicant fault is equal equal to half the MPE in the upper low rate zone.



Indicate by drawings if necessary.



C — contact discharge discharge (6 kV); A — air discharge (8 kV).

Requirements

Requirement Require ment 1:

If the low axis is vertical, the low direction (from bottom to top or from top to

bottom) shall be given.

Requirement 2: If the low ax axis is is horizonta horizontall and the meter has an indicat indicating ing device which is integral with the body of the meter, the location of the indicating device (at the side or at the top of the meter) shall be given. 4.6.11 Radiated electromagnetic electromagnetic ield (ISO 4064‑2:2014|OIML 4064‑2:2014|OIML R 49‑2:2013, 49‑2:2013, 8.12)  

At start

At end

Application No:

______________

Ambient temperature:

°C

Model:

______________

Ambient relative humidity:

%

Date:

______________

Ambient atmospheric pressure:

Observer:

______________

Time:

MPa

58

 

© ISO 2014 – All rights reserved

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)  

Gravimetric/volumetric

Test method: met hod:

Volume Vol ume measures/weighbridge measur es/weighbridge used — m 3 or kg: Water conductivity (electromagnetic induction meters only) — S/cm: Length of straight pipe before meter (or manifold) — mm: Length of straight pipe after meter (or manifold) — mm: Nominal diameter DN of pipe before and after meter (or manifold) — mm:

Describe low straightener installation if used: Meter Met er seria seriall No. No.:___ :___ ___ ___ _

Orientati Orie ntation on (V (V,, H, oth other er):____ ):____ ___ ___ ___ _

Flow direction (see Requirement 1): ____ Location of indicating device (see Requirement 2):___ Test conditions

Antenna Actual Work- Work- Initial polariza - or simu- ing ing readtion vertilated pres- tempering low cal/ sure ature horizontal rate

  m3/h

Final reading

Indi- Actual Meter cated volerror volume ume

MPEa

Fault 

Signiicant fault b

EUT functioning cor-

rectly

 pw

T w

V i(i)

V i(f)

V i

V a

E m

 

E m2) m2)  − E m1) m1)

 

MPa

°C

m3

m3

m3

m3

%

%

%

%





(bar) 1) Reference conditions (no disturbance)

V

H

 

26 –4 0 MHz

V

H

40 – 60 MHz

V

6 0 – 80 M H z





 

yes

no

H

 

yes

no

V

H

 

yes

no

80 –100 MHz

V

H

 

yes

no

100 –120 MHz

V

H

 

yes

no

120 –14 4 MHz

V

H

 

yes

no

14 4 –150 MHz

V

H

 

yes

no

150 –160 MHz

V

H

 

yes

no

160 –180 MHz

V

H

 

yes

no

180 –200 MHz

V

H

 

yes

no

200 –250 MHz

V

H

 

yes

no

250 –350 MHz

V

H

 

yes

no

350 – 400 MHz

V

H

 

yes

no

400 –435 MHz

V

H

 

yes

no

435–500 MHz

V

H

 

yes

no

500 – 6 0 0 MH z

V

H

 

yes

no

600 –700 MHz

V

H

 

yes

no

700 –800 MH z

V

H

 

yes

no

8 0 0 – 93 4 MH z

V

H

 

yes

no

934 –1 000 MHz

V

H

 

yes

no

2) Dist urbance



For a complete water meter this is the maximum permissible error as deined in ISO 4064-1:201 4064-1:2014|OIML 4|OIML R 49-1:20 49-1:2013, 13, 4.2.2 or 4.2.3 according to the accuracy class of the meter. If the EUT is a separable part of water meter, the MPE shall be deined by the manufacturer (ISO 4064-2:2014|OIML R 49-2:2013, 9.4). b 

The signiicant signiicant fault is equal to half the MPE MPE in the upper upper low rate zone.

59

 

© ISO 2014 – All rights reserved

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)  

Test conditions

Antenna

Actual Work- Work- Initial or simu- ing ing readtion vertilated pres- tempering cal/ low sure ature horizontal rate

polariza -

  m3/h

Final reading

Indi- Actual Meter cated volerror volume ume

MPEa

Fault 

Signiicant fault b

EUT functioning cor-

rectly

 p w

T w

V i(i)

V i(f)

V i

V a

E m

 

E m2) m2)  − E m1) m1)

 

MPa

°C

m3

m3

m3

m3

%

%

%

%

(bar) 1 000–1 400 MHz

V

H

 

yes

no

1 400–2 000 MHz

V

H

 

yes

no

Comments: a 

For a complete water meter this is the maximum permissible error as deined in ISO 4064 4064-1 -1:2014 :2014|OIML |OIML R 49-1:20 49-1:2013, 13, 4.2.2 or 4.2.3 according to the accuracy class of the meter. If the EUT is a separable part of water meter, the MPE shall be deined by the manufacturer (ISO 4064-2:2014|OIML R 49-2:2013, 9.4). b 

The signiicant signiicant fault is equal equal to half the MPE in the upper low rate zone.

Requirements

Requirement Require ment 1:

If the low axis is vertical, the low direction (from bottom to top or from top to

bottom) shall be given.

Requirement 2:

If the low ax axis is is horizonta horizontall and the meter has an indicat indicating ing device which is

integral withbe thegiven. body of the meter, the location of the indicating device (at the side or at the top of the meter) shall 4.6.12 Conducted electromagnetic electromagnetic ield (ISO 4064‑2:2014|OIML 4064‑2:2014|OIML R 49‑2:2013, 49‑2:2013, 8.13)  

At start

At end

Application No:

______________

Ambient temperature:

°C

Model:

______________

Ambient relative humidity:

%

Date:

______________

Ambient atmospheric pressure:

Observer:

______________

Time:

Gravimetric/volumetric

Test method: met hod:

Volume measures/weighbridge used — m3 or kg: Water conductivity (electromagnetic induction meters only) — S/cm: Length of straight pipe before meter (or manifold) — mm: Length of straight pipe after meter (or manifold) — mm: Nominal diameter DN of pipe before and after meter (or manifold) — mm:

Describe low straightener installation if used: u sed: Meter Met er seria seriall No. No.:___ :___ ___ ___ _

MPa

Orientati Orie ntation on (V (V,, H, oth other er):____ ):____ ___ ___ ___ _

60

 

© ISO 2014 – All rights reserved

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)

Flow direction (see Requirement 1): ___ Location of indicating device (see Requirement 2):____ Test conditions

Actual or Working WorkInitial simulated presing reading low rate sure temperature Q( )

 

m3/h

Final reading

IndiActual Meter cated volerror volume ume

MPEa

Fau lt lt

Si gni gni--

EUT

icant functionfault b

ing cor-

rectly

 p w

T w

V i(i)

V i(f)

V i

V a

E m

 

E m2) m2)  − E m1) m1)

 

MPa

°C

m3

m3

m3

m3

%

%

%

%





(bar) 1) Reference conditions (no disturbance)





0,15– 0, 30 MHz

ye s

no

0, 30 – 0, 57 MHz

ye s

no

0,57–1,1 MHz

ye s

no

1,1–2, 2 MHz

ye s

no

2,2–3,9 MHz

ye s

no

3,9 –7, 5 MHz

ye s

no

7,5 –14 MHz

ye s

no

14 –30 MHz

ye s

no

30 –50 MHz

ye s

no

50 – 8 0 MH z

ye s

no

2) Disturbance

Comments: a 

For a complete water meter this is the maximum permissible error as deined in ISO 4064-1:201 4064-1:2014|OIML 4|OIML R 49-1:20 49-1:2013, 13, 4.2.2 or 4.2.3 according to the accuracy class of the meter. If the EUT is a separable part of water meter, the MPE shall be deined by the manufacturer (ISO 4064-2:2014|OIML R 49-2:2013, 9.4). b 

The signiicant signiicant fault is equal to half the MPE MPE in the upper upper low rate zone.

Requirements

Requirement Requirem ent 1:

If the low axis is vertical, the low direction (from bottom to top or from top to

bottom) shall be given.

Requirement 2: If the low axi axiss is horizontal and the meter has an indicati indicating ng device which is integral with the body of the meter, the location of the indicating device (at the side or at the top of the meter) shall be given. 4.6.13 Surges on signal, data and control lines (ISO 4064‑2:2014|OIML 4064‑2:2014|OIML R 49‑2:2013, 49‑2:2013, 8.14) (ap(applicable only for environmental class E2)  

At start

At end

Application No:

______________

Ambient temperature:

°C

Model:

______________

Ambient relative humidity:

%

Date:

______________

Ambient atmospheric pressure:

Observer:

______________

Time:

MPa

61

 

© ISO 2014 – All rights reserved

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)  

Gravimetric/volumetric

Test method: met hod:

Volume measures/weighbridge used — m3 or kg: Water conductivity (electromagnetic induction meters only) — S/cm: Length of straight pipe before meter (or manifold) — mm: Length of straight pipe after meter (or manifold) — mm: Nominal diameter DN of pipe before and after meter (or manifold) — mm:

Describe low straightener installation if used: u sed: Meter Met er seria seriall No. No.:___ :___ ___ ___ _

Orientati Orie ntation on (V (V,, H, oth other er):____ ):____ ___ ___ ___ _

Flow direction (see Requirement 1): ___ Location of indicating device (see Requirement 2):____ Test conditions

Actual Working Working Initial or simu- pressure reading temp. lated lowrate Q ( )

 

m3/h

Final reading

Indicated volume

Actual Meter volerror ume

MPEa Fault

niicant

Sig-

EUT functioning

fault b

correctly

P w

T w

V i(i)

V i(j)

V i

V a

E m

 

E m2) m2)   -E m1) m1)

 

MPa

°C

m3

m3

m3

m3

%

%

%

%

(bar) 1) Reference conditions (no surges) 2) Surge

Mode c

Posit ive

L

L

yes

no

L

L

yes

no

L

L

yes

no

L

L

yes

no

L

L

yes

no

L

L

yes

no

L

E

yes

no

L

E

yes

no

L

E

yes

no

L

E

yes

no

L

E

yes

no

L

E

yes

no

Negat ive

Posit ive

Negat ive

Comments: a 

For a complete water meter this is the maximum permissible error as deined in ISO 4064-1:2014|OIML 4064-1:2014|OIML R 49-1:20 49-1:2013, 13, 4.2.2 or 4.2.3 according to the accuracy class of the meter. If the EUT is a separable part of the water meter, the MPE shall be deined by the manufacturer (ISO 4064-2:2014|OIML R 49-2:2013, 9.4). b 

The signiicant signiicant fault is equal equal to half the MPE in the upper low rate zone.



L–L — line to line surge; L–E — line to earth surge.

Requirements

Requirement Require bottom)ment shall 1: be given.If the low axis is vertical, the low direction (from bottom to top or from top to

62

 

© ISO 2014 – All rights reserved

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)  

Requirement 2: If the low axi axiss is horizontal and the meter has an indicati indicating ng device which is integral with the body of the meter, the location of the indicating device (at the side or at the top of the meter) shall be given. 4.6.14 Surges on AC and DC mains power power lines (ISO 4064‑2:2014|OIML R 49‑2:2013, 49‑2:2013, 8.15) (ap(applicable only for environmental class E2)  

At start

At end

Application No:

______________

Ambient temperature:

°C

Model:

______________

Ambient relative humidity:

%

Date:

______________

Ambient atmospheric pressure:

Observer:

______________

Time:

Gravimetric/volumetric

Test method: met hod:

Volume Vol ume measures/weighbridge measur es/weighbridge used — m 3 or kg: Water conductivity (electromagnetic induction meters only) — S/cm: Length of straight pipe before meter (or manifold) — mm: Length of straight pipe after meter (or manifold) — mm: Nominal diameter DN of pipe before and after meter (or manifold) — mm:

Describe low straightener installation if used: Meter Met er seria seriall No. No.:___ :___ ___ ___ _

MPa

Orientati Orie ntation on (V (V,, H, oth other er):____ ):____ ___ ___ ___ _

63

 

© ISO 2014 – All rights reserved

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)   Flow direction (see Requirement 1): ___ Location of indicating device (see Requirement 2):____ Actual or Working Working Initial simulated prestemper- reading low rate sure ature

Test conditions

 

Final reading

Indi- Actual Meter cated volerror volume ume

MPEa

Fault

Sig ni-

icant fault b

EUT functioning cor-

rectly

Q( )

 pw

T w

V i(i)

V i(j)

V i

V a

E m

 

E m2) m2)  − E m1) m1)

 

m3/h

MPa

°C

m3

m3

m3

m3

%

%

%

%





(bar) 1) Reference conditions (no surges)





2) DC power

Mode c

Posit ive

L

L

yes

no

L

L

yes

no

L

L

yes

no

L

L

yes

no

L

L

yes

no

L

L

yes

no

L

E

yes

no

L

E

yes

L

E

yes

no no

L

E

yes

no

L

E

yes

no

L

E

yes

no

Negat ive

Posit ive

Negat ive

Comments: a 

For a complete water meter this is the maximum permissible error as deined in ISO 4064 4064-1 -1:2014 :2014|OIML |OIML R 49-1:20 49-1:2013, 13, 4.2.2 or 4.2.3 according to the accuracy class of the meter. If the EUT is a separable part of the water meter, the MPE shall be deined by the manufacturer (ISO 4064-2:2014|OIML R 49-2:2013, 9.4). b 

The signiicant signiicant fault is equal equal to half the MPE in the upper low rate zone.



L–L — line to line surge; L–E — line to earth surge.

Meter Met er seria seriall No. No.:___ :___ ___ ___ _

Orientati Orie ntation on (V (V,, H, oth other er):____ ):____ ___ ___ ___ _

Flow direction (see Requirement 1): ___ Location of indicating device (see Requirement 2):____ Test conditions

Actual or Working simu- Working temlated pressure perature

Initial reading

low

Final reading

Indicated volume

Actual Meter volMPEa error ume

SigniFault 

icant fault b

EUT functioning cor-

rectly

rate  

Q( )

 p w

T w

V i(i)

V i(j)

V i

V a

E m

 

E m2) m2)  − E m1) m1)

 

m3/h

MPa(bar)

°C

m3

m3

m3

m3

%

%

%

%

1) Reference conditions (no surges) a 









For a complete water meter this is the maximum permissible error as deined in ISO 4064-1:2014|OIML 4064-1:2014|OIML R 49-1:20 49-1:2013, 13, 4.2.2 or 4.2.3 according to the accuracy class of the meter. If the EUT is a separable part of the water meter, the MPE shall be deined by the manufacturer (ISO 4064-2:2014|OIML R 49-2:2013, 9.4). b  c 

The signiicant signiicant fault is equal equal to half the MPE in the upper low rate zone. L–L — line to line surge; L–E — line to earth surge.

64

 

© ISO 2014 – All rights reserved

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)  

Test conditions

Actual or Working simu- Working temlated pressure perature

Initial reading

low

Final reading

Indicated volume

Actual Meter MPEa volerror ume

SigniFault 

icant fault b

EUT functioning cor-

rectly

rate  

AC supply

Q( )

 p w

T w

V i(i)

V i(j)

V i

V a

E m

 

E m2) m2)  − E m1) m1)

 

m3/h

MPa(bar)

°C

m3

m3

m3

m3

%

%

%

%

Mode c

voltage

 

0° Posit ive

Negat ive

Posit ive

Negat ive

AC supply

L

L

yes

no

L

L

yes

no

L

L

yes

no

L

L

yes

no

L

L

yes

no

L

L

yes

no

L

E

yes

no

L

E

yes

no

L

E

yes

no

L

E

yes

no

L

E

yes

no

L

E

yes

no

Mode c

 

voltage 90° Posit ive

Negat ive

Posit ive

Negat ive

L

L

yes

no

L

L

yes

no

L

L

yes

no

L

L

yes

no

L

L

yes

no

L

L

yes

no

L L

E E

yes yes

no no

L

E

yes

no

L

E

yes

no

L

E

yes

no

L

E

yes

no

Comments: a 

For a complete water meter this is the maximum permissible error as deined in ISO 4064-1:201 4064-1:2014|OIML 4|OIML R 49-1:20 49-1:2013, 13, 4.2.2 or 4.2.3 according to the accuracy class of the meter. If the EUT is a separable part of the water meter, the MPE shall be deined by the manufacturer (ISO 4064-2:2014|OIML R 49-2:2013, 9.4). b 

The signiicant signiicant fault is equal to half the MPE MPE in the upper upper low rate zone.



L–L — line to line surge; L–E — line to earth surge.

Meter Met er seria seriall No. No.:___ :___ ___ ___ _

Orientati Orie ntation on (V (V,, H, oth other er):____ ):____ ___ ___ ___ _

65

 

© ISO 2014 – All rights reserved

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)   Flow direction (see Requirement 1): ___ Location of indicating device (see Requirement 2):____ Test conditions

Actual or simulated

Working pressure

low rate  

Q( )

m3/h

 pw

MPa(bar)

Working temperature

T w

°C

Initial reading

V i(i)

m3

IndiActual Final cated Meter readvolMPEa volerror ing ume ume V i(j)

m3

1) Reference conditions (no surges)

AC supply voltage

Mode c

Negat ive

Posit ive

Negat ive

AC supply voltage

m3

V a

m3

E m

%

  %

E m2) m2) −

m1) E m1) %



EUT functioning cor-

rectly

  %







 

180° Posit ive

V i

SigniiFault  cant fault b

L

L

yes

no

L

L

yes

no

L

L

yes

no

L

L

yes

no

L

L

yes

no

L

L

yes

no

L

E

yes

no

L

E

yes

no

L

E

yes

no

L

E

yes

no

L

E

yes

no

L

E

yes

no

Mode c

 

270° Posit ive

Negat ive

Posit ive

Negat ive

L

L

yes

no

L

L

yes

no

L

L

yes

no

L

L

yes

no

L

L

yes

no

L L

L E

yes yes

no no

L

E

yes

no

L

E

yes

no

L

E

yes

no

L

E

yes

no

L

E

yes

no

Comments: a 

For a complete water meter this is the maximum permissible error as deined in ISO 4064-1:2014|OIML 4064-1:2014|OIML R 49-1:20 49-1:2013, 13, 4.2.2 or 4.2.3 according to the accuracy class of the meter. If the EUT is a separable part of the water meter, the MPE shall be deined by the manufacturer (ISO 4064-2:2014|OIML R 49-2:2013, 9.4). b 

The signiicant signiicant fault is equal equal to half the MPE in the upper low rate zone.



L–L — line to line surge; L–E — line to earth surge.

Requirements

66

 

© ISO 2014 – All rights reserved

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)  

Requirement Requirem ent 1:

If the low axis is vertical, the low direction (from bottom to top or from top to

bottom) shall be given.

Requirement 2: If the low axi axiss is horizontal and the meter has an indicati indicating ng device which is integral with the body of the meter, the location of the indicating device (at the side or at the top of the meter) shall be given. 4.6.15 Absence of low test (ISO 4064‑2:2014|OIML 4064‑2:2014|OIML R 49‑2:2013, 49‑2:2013, 8.17)  

At start

At end

Application No:

______________

Ambient temperature:

°C

Model:

______________

Ambient relative humidity:

%

Date:

______________

Ambient atmospheric pressure:

Observer:

______________

Time:

MPa

Gravimetric/volumetric

Test method: met hod:

Volume Vol ume measures/weighbridge measur es/weighbridge used — m 3 or kg: Water conductivity (electromagnetic induction meters only) — S/cm: Length of straight pipe before meter (or manifold) — mm: Length of straight pipe after meter (or manifold) — mm: Nominal diameter DN of pipe before and after meter (or manifold) — mm:

Describe low straightener installation if used: Meter Met er seria seriall No. No.:___ :___ ___ ___ _

Orientati Orie ntation on (V (V,, H, oth other er):____ ):____ ___ ___ ___ _

Flow direction (see Requirement 1): ____ Location of indicating device (see Requirement 2):___ Application co conditions

 

Meter illed with water, purg-

Work ing pressure

Working temWorking perature

Initial reading

Final reading after 15 min

Indicated volume

EUT functioning

correctly

 pw

T w

V i(i)

V i(j)

V i

 

MPa (bar)

°C

m3

m3

m3

 

yes

no

yes

no

ing out all air

Waterr fully Wate fu lly discharged from the meter Comments:

The water meter totalization shall not change by more than the value of the veriication scale interval during each test interval. Requirements

Requirement Requirem ent 1:

If the low axis is vertical, the low direction (from bottom to top or from top to

bottom) shall be given.

Requirement 2: If the low axi axiss is horizontal and the meter has an indicati indicating ng device which is integral with the body of the meter, the location of the indicating device (at the side or at the top of the meter) shall be given.

© ISO 2014 – All rights reserved

 

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)  

5

Initial veriication report 

5.1 General

The speciic format layout for reporting initial veriications and subsequent veriications of water meters is left largely to the metrological authorities and the individual organizations carrying out veriication tests. However, the report (records) shall contain the minimum information detailed in ISO 4064-1:2014|OIML R 49-1:2013, 7.3 and ISO 4064-2:2014|OIML R 49-2:2013, 11.2.2. In addition to this, any special requirements and/or restrictions for initial veriication detailed in the type approval certiicate for the EUT shall be applied. A record of equipment and instrumentation used with calibration details (see Annex B) B) shall be kept.

The following basic information should also be included in the veriication report (record) followed by the results of the tests (three examples of how the report may be formatted are given below): 5.2

Information concerning the EUT veriied

Type approval number of the EUT Details of the EUT:

Model number: Accuracy class: Meter designation/s Q3: Ratio Q3/Q1:

Maximum pressure loss Δ pmax  : Flow rate at Δ pmax : Year of manufacture: The manufacturer:

Authorized representative: Address

Testing laboratory: Authorized representative: Address

5.3 5.3.1

Initial veriication test report (ISO 4064‑2:2014|OIML 4064‑2:2014|OIML R 49‑2:2013, 49‑2:2013, Clause Clause 10) Example 1: Approved water meter (complete or combined) (ISO 4064-2:2014|OIML R 49-

2:2013, 10.1)

67

68

 

© ISO 2014 – All rights reserved

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)  

 

At start

At end

Application No:

______________

Ambient temperature:

°C

Model:

______________

Ambient relative humidity:

%

Date:

______________

Ambient atmospheric pressure:

Observer:

______________

Time:

MPa

Error (of indication) tests

EUT testing t esting case (ISO (I SO 4064-2 4064 -2:201 :2014|OIML 4|OIML R 49-2:2013, 49-2:2013, 8.1.8) Category for testing test ing (ISO (IS O 4064-2: 4064 -2:201 2014|OIML 4|OIML R 49-2:2013, 49-2:2013, )

Gravimetric/volumetric

Test method:

Volume measures/we Volume measur es/weighbridge ighbridge used use d — m 3 or kg: Water conductivity (electromagnetic induction meters only) — S/ cm: Length of straight pipe before meter (or manifold) — mm Length of straight pipe after meter (or manifold) — mm Nominal diameter DN of pipe before and after aft er meter (or (or manifold) — mm:

Describe low straightener installation if used: a 

Enter clause clause number number according to one of of the coniguration categories for testing the EUT listed in ISO 4064-2:2014|OIML R 49-2:2013, 8.1.8.2 to 8.1.8.5.

Meter Met er seria seriall No. No.:___ :___ ___ ___ _

Orientati Orie ntation on (V (V,, H, oth other er):____ ):____ ___ ___ ___ _

Flow direction (see Requirement 1): ____ Location of indicating device (see Requirement 2):___

Nominal low ratea

Actual Working Working Initial Final Indicated Actual Meter low pressure temperature reading reading volume volume errorb rate

m3/h

Q( )

 

m3/h

MPa

°C

MPEc  

V i(i)

V i(f)

V i

V a

E c

 

m3

m3

m3

m3

%

%

(bar)

Q1 Q2 Q3

Comments: a 

These low rates shall shall be be applied applied unless alternatives are speciied speciied in the type approval certiicate.



Calculations Calcu lations for the error (of (of indicat ion) are described in ISO 4064 4064-2:201 -2:2014|OIML 4|OIML R 49-2:2013, Annex B.



The maximum permissible error as deined in ISO 4064-1:2014|OIML 4064-1:2014|OIML R 49-1:20 49-1:2013, 13, 4.2. 4.2.2 2 or 4.2. 4.2.3, 3, accordi according ng to the accuracy class of the meter.

Requirements

Requirement Requirem ent 1:

If the low axis is vertical, the low direction (from bottom to top or from top to

bottom) shall be given.

Requirement 2: If the low axi axiss is horizontal and the meter has an indicati indicating ng device which is integral with the body of the meter, the location of the indicating device (at the side or at the top of the meter) shall be given.

69

 

© ISO 2014 – All rights reserved

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)   5.3.2

Example 2: Appro Approved ved calculator (including indicating device) (ISO 4064-2:2014|OIML R 49-

2:2013, 10.2)  

At start

At end

Application No:

______________

Ambient temperature:

°C

Model:

______________

Ambient relative humidity:

%

Date:

______________

Ambient atmospheric pressure:

Observer:

______________

Time:

MPa

Error (of indication) tests

EUT testi t esting ng case (ISO ( ISO 4064-2: 4064 -2:201 2014|OIML 4|OIML R 49-2:2013, 49-2:2013, 8.1.8) Category for testing test ing (ISO 4064-2:201 4064-2:2014|OIML 4|OIML R 49-2:2013, 49-2:2013, ) a 

Enter clause clause number number according to one of of the coniguration categories categories for testing the EUT listed in ISO 4064-2:2014|OIML R 49-2:2013, 8.1.8.2 to 8.1.8.5.

Meter Met er seria seriall No. No.:___ :___ ___ ___ _

Orientati Orie ntation on (V (V,, H, oth other er):____ ):____ ___ ___ ___ _

Flow direction (see Requirement 1): ___ Location of indicating device (see Requirement 2):____

Nominal low ratea

Actual

low

Applied pulse fre-

Initial Final reading reading

rate

quencyb

 

Q( )

 

V i(i)

V i(f)

m3/h

m3/h

Hz

m3

m3

Total pulses

Meter

MPEd

Indicated volume

Actual volume

errorc

V i

V a

E c

 

m3

m3

%

%

injectedb T p

Q1 Q2 Q3

Comments: a 

These low rates rates shall shall be applied unless alternatives are speciied speciied in the type approval approval certi certi icate.



Other types of output signal may be appropriat appropriatee accordi according ng to the design of the water meter.



Calculations Calcul ations for the error (of (of indic indication) ation) are described in ISO 4064-2:2014|OIML 4064-2:2014|OIML R 49-2:2013, Annex B.



The maxi maximum mum error (of (of indicat indication) ion) allowed for the calculator (including indicating indicatin g device) is given in the type approval approval certiicate.

Requirements

Requirement Require ment 1:

If the low axis is vertical, the low direction (from bottom to top or from top to

bottom) shall be given.

Requirement 2: If the low ax axis is is horizonta horizontall and the meter has an indicat indicating ing device which is integral with the body of the meter, the location of the indicating device (at the side or at the top of the meter) shall be given. 5.3.3

Example 3: Approved measurement transducer (including low or volume sensor)

(ISO 4064-2:2014|OIML R 49-2:2013, 10.2)

70

 

© ISO 2014 – All rights reserved

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)  

 

At start

At end

Application No:

______________

Ambient temperature:

°C

Model:

______________

Ambient relative humidity:

%

Date:

______________

Ambient atmospheric pressure:

Observer:

______________

Time:

MPa

Error (of indication) tests

EUT testing t esting case (ISO (I SO 4064-2 4064 -2:201 :2014|OIML 4|OIML R 49-2:2013, 49-2:2013, 8.1.8) Category for testing test ing (ISO (IS O 4064-2: 4064 -2:201 2014|OIML 4|OIML R 49-2:2013, 49-2:2013, )

Gravimetric/volumetric

Test method:

Volume measures/we Volume measur es/weighbridge ighbridge used use d — m 3 or kg: Water conductivity (electromagnetic induction meters only) — S/ cm: Length of straight pipe before meter (or manifold) — mm Length of straight pipe after meter (or manifold) — mm Nominal diameter DN of pipe before and after aft er meter (or (or manifold) — mm:

Describe low straightener installation if used: a 

Enter clause clause number number according to one of of the coniguration categories for testing the EUT listed in ISO 4064-2:2014|OIML R 49-2:2013, 8.1.8.2 to 8.1.8.5.

Meter Met er seria seriall No. No.:___ :___ ___ ___ _

Orientati Orie ntation on (V (V,, H, oth other er):____ ):____ ___ ___ ___ _

Flow direction (see Requirement 1): ___ Location of indicating device (see Requirement 2):____ Nominal

Actual

low ratea

low

Working pressure

rate  

Q( )

 

m3/h

m3/h

MPa

Working temperature °C

Meter MPEd

Initial reading

Final reading

Total output pulsesb

Indicated volume

Actual volume

error c

V i(i)

V i(f)

T p

V i

V a

E c

 

m3

m3

m3

m3

%

%

(bar)

Q1 Q2 Q3

Comments: a 

These low rates shall shall be be applied applied unless alternatives are speciied speciied in the type approval certiicate.



Other types of output signal signal may be appropriate appropriate according to the design design of the water meter meter..



Calculations Calcu lations for the error (of (of indicat ion) are described in ISO 4064 4064-2:201 -2:2014|OIML 4|OIML R 49-2:2013, Annex B.



The maximum error (of indicat ion) allowed for the measurement tran transducer sducer (including low or volume sensor) is given in the type approval certiicate.

Requirements

Requirement Requirem ent 1:

If the low axis is vertical, the low direction (from bottom to top or from top to

bottom) shall be given.

© ISO 2014 – All rights reserved

 

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)  

Requirement 2: If the low ax axis is is horizonta horizontall and the meter has an indicat indicating ing device which is integral with the body of the meter, the location of the indicating device (at the side or at the top of the meter) shall be given.

71

72

 

© ISO 2014 – All rights reserved

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)  

 Annex A (normative)  List of documents concerning the type (ISO 4064‑1:2014|OIML R 49‑1:2013, 7.2.9)

Doc ument reference

Dat e

Brief descript ion

73

 

© ISO 2014 – All rights reserved

 

BS EN ISO 4064-3:2014

ISO 4064-3:2014(E)  

 Annex B (normative)  Listing of test equipment used in examinations and tests

Parameter Instrument measured or Manufacturer or equipment  applied

Comments:

Cali Ca libr brat atio ion n dat date e Model number

Serial number

Used Us ed in te test st No No..

(ISO 4064 4064-2:20 -2:2014| 14| L ast

Nex t   OIML R 49-2:2013, subclause)

74

 

© ISO 2014 – All rights reserved

 

BS EN ISO 4064-3:2014

 

This page deliberately left blank 

 

This page deliberately left blank 

 

NO COPYING WITHOUT BSI PERMISSION EXCEPT AS PERMITTED BY COPYRIGHT LAW

British Standards Institution (BSI) BSI is the national body responsible for preparing British Standards and other standards-related publications, information and services. BSI is incorporated by Royal Charter. British Standards and other standardization products are published by BSI Standards Limited.

About us

Revisions

We bring together business, industry, government, consumers, innovators and others to shape their combined experience and expertise into standards -based solutions.

Our British Standards and other publications are updated by amendment or revision.

The knowledge embodied in our standards has been carefully assembled in a dependable format and refined through our open consultation process. Organizations of all sizes and across all sectors choose standards to help them achieve their goals.

Information on standards We can provide you with the knowledge that your organization needs to succeed. Find out more about British Standards by visiting our website at bsigroup.com/standards or bsigroup.com/standards  or contacting our Customer Services team or Knowledge Centre.

Buying standards You can buy and download PDF versions of BSI publications, including British and adopted European and international standards, through our website at bsigroup.com/shop , where hard copies can also be purchased. If you need international and foreign standards from other Standards Development Organizations, hard copies can be ordered from our Customer Services team.

Subscriptions Our range of subscription services are designed to make using standards easier for you. For further information on our subscription products go to bsigroup.com/subscriptions.. bsigroup.com/subscriptions With British Standards Online (BSOL) you’ll have instant access to over 55,000 British and adopted European and international standards from your desktop. It’s available 24/7 and is refreshed daily so you’ll always be up to date. You can keep in touch with standards developments and receive substantial discounts on the purchase price of standards, both in single copy and subscription format, by becoming a BSI Subscribing Member.

PLUS is an updating service exclusive to BSI Subscribing Members. You will automatically receive the latest hard copy of your standards when they’re revised or replaced. To find out more about becoming a BSI Subscribing Member and the benefits of membership, please visit bsigroup.com/shop bsigroup.com/shop.. With a Multi-User Network Licence (MUNL) you are able to host standards publications on your intranet. Licences can cover as few or as many users as you wish. With updates supplied as soon as they’re available, you can be sure your documentation is current. For further information, email [email protected].

BSI Group Headquarters 389 Chiswick High Road London W4 4AL UK

We continually improve the quality of our products and services to benefit your business. If you find an inaccuracy or ambiguity within a British Standard or other BSI publication please inform the Knowledge Centre.

Copyright All the data, software and documentation set out in all British Standards and other BSI publications are the property of and copyrighted by BSI, or some person or entity that owns copyright in the information used (such as the international standardization standardizatio n bodies) and has formally licensed such information to BSI for commercial publication and use. Except as permitted under the Copyright, Designs and Patents Act 1988 no extract may be reproduced, stored in a retrieval system or transmitted in any form or by any means – electronic, photocopying, recording recording or otherwise – without prior written permission from BSI. Details and advice can be obtained from the Copyright & Licensing Department.

Useful Contacts: Customer Services Tel: +44 845 086 9001 Email (orders): [email protected] Email (enquiries): [email protected] Subscriptions Tel: +44 845 086 9001 Email: [email protected] Knowledge Centre Tel: +44 20 8996 7004 Email: [email protected] Copyright & Licensing Tel: +44 20 8996 7070 Email: [email protected]