ACUSON CV70 US

Troubleshooting Guide System Troubleshooting

07848547 07848539 07848521

© Siemens AG

Siemens Medical Ultrasound lne DivisionSolutions

Print No.: US05-101.840.01.01.02 Replaces: n.a. Part No.: n.a. sd

2004

The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages. All rights, including rights created by patent grant or registration of a utility model or design, are reserved.

English Doc. Gen. Date: 06.04 n.a.

2

Revision / Disclaimer

1Revision / Disclaimer

Copyright Copyright © 2004 by Siemens Corporation. All rights reserved. No part of this publication may be reproduced, transmitted, transcribed, stored in retrieval systems, or translated into any language or computer language, in any form or by any means, electronic, mechanical, magnetic, optical, chemical, manual, or otherwise, without the prior written permission of Siemens Corporation. Siemens Corporation reserves the right to change its products and services at any time. In addition, this manual is subject to change without notice. Siemens Corporation welcomes customer input on corrections and suggestions for improvements to this manual.Although Siemens Corporation has attempted to ensure accuracy throughout this manual, Siemens Corporation assumes no liability for any errors or omissions, nor for any damages resulting from the application or use of this information. Trademarks ACUSON, Sequoia, Cypress, AEGIS, Aspen, 128XP, XP, AcuNav, CV70, SONOLINE, Adara, Antares, G20, G40, G50, G60S, Omnia and KinetDx are trademarks of Siemens Corporation registered in the U.S. Patent and Trademark Office. CWS3000, DS3000, WS3000, and WebPro are trademarks of Siemens Corporation. Windows XP, 2000 and NT are registered trademarks of Microsoft Corporation. Internet Explorer is a trademark of Microsoft Corporation. Netscape and Netscape Navigator are registered trademarks of Netscape Communications Corporation. Java and the JavaScript name are registered trademarks of Sun Microsystems, Inc. Adobe Acrobat is a registered trademark of Adobe Systems, Inc. Other products and brand names are trademarks of their respective owners. Disclaimer The service of equipment described herein is to be performed by qualified personnel who are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or one of its affiliates to provide such services. Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service procedures. License Agreement All computer programs copyright 1990-2004 by Siemens Corporation or its suppliers. Programs are licensed under the following agreement: Siemens or its suppliers retain(s) ownership of and title to any computer program supplied with the equipment and to the trade secrets embodied in such computer programs. Subject to the Buyer’s acceptance and fulfillment of the obligations in this paragraph, Siemens grants the Buyer a personal, non-transferable, perpetual, non-exclusive license to use any computer program supplied with the Equipment that is necessary to operate the Equipment solely on the medium in which such program is delivered for the purpose of operating the equipment in accordance with the instructions set forth in the operator’s manuals supplied with the Equipment and for no other purpose whatsoever. Buyer may not reverse-assemble, reverse-compile or otherwise reverse-engineer such computer programs nor may Buyer make a copy of such program or apply any techniques to derive the trade secrets embodied therein. In the event of a failure by Buyer to comply with the terms

ACUSON CV70

US05-101.840.01.01.02 06.04

Page 2 of 36 sd

Siemens Medical Solutions USA, Inc.

Revision / Disclaimer

3

of this license, the license granted by this paragraph shall terminate. Further, because unauthorized use of such computer programs will leave Siemens without an adequate remedy at law, Buyer agrees that injunctive or other equitable relief will be appropriate to restrain such use, threatened or actual. Buyer further agrees that (i) any of Siemens"s suppliers of software is a direct and intended beneficiary of this end-user sublicense and may enforce it directly against Buyer with respect to software supplied by such supplier, and (ii) No supplier of Siemens shall be liable to buyer for any general, special, direct, indirect, consequential, incidental or other damages arising out of the sublicense of the computer programs supplied with the equipment.

Siemens Medical Solutions USA, Inc.

US05-101.840.01.01.02 06.04

Page 3 of 36 sd

ACUSON CV70

4

Table of Contents

1-

0Table of Contents

1 _______ Troubleshooting_________________________________________________ 5 Service Check List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 System Configuration: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Specific Problem: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Troubleshooting Flow Charts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Using the Flow Chart. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Diagnostic LED Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 System Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 Power Supply Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18 Diagnostic and Error Log Messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19 Troubleshooting the DIMAQ-IP. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24 Review Switch Failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24 Corrupt Images and Clips . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24 Troubleshooting DICOM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 DIMAQ-IP BIOS Corruption - CMOS Error Solution . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 SRAM Memory Clear . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Clearing the SRAM - Software Method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Clearing the SRAM - Hardware Method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Restoring the Software Option Keys and System Presets . . . . . . . . . . . . . . . . . . . .

29 29 30 31 32 33

Service Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36 Changing Video Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36

ACUSON CV70

US05-101.840.01.01.02 06.04

Page 4 of 36 sd

Siemens Medical Solutions USA, Inc.

Troubleshooting

5

2-

Service Check List 1Troubleshooting

0

Use the following checklist to describe identified problems and the conditions under which the problem was observed. Providing this information helps technical support personnel to better understand your specific problem.

System Configuration:

0

• System serial number • • • •

Software version number Product status and upgrade numbers Installed options Enabled features

Specific Problem:

0

• Detailed problem description: For imaging problems include: transducer, exam type, mode settings, etc.

• • • •

Error log and diagnostic messages Specific test failure information Power supply measurements Specific imaging artifacts (capture images to CD or MO disk)

Siemens Medical Solutions USA, Inc.

US05-101.840.01.01.02 06.04

Page 5 of 36 sd

ACUSON CV70

6

Troubleshooting

Troubleshooting Flow Charts

0

The diagnostics flow chart series provide a guide to troubleshooting a malfunctioning CV70. The “Component Designations” section of this manual provides a table called Subsystem and Component Designations that identifies the acronyms used in this chapter to identify devices. Refer to the following chapters in this manual for additional information:

• • • • • •

System Architecture Component Designations Configuring a Remote Network Export Device Configuring DICOM Software Option Keys FRU Removal and Replacement Procedures

WARNING

The CV70 contains hazardous voltages that can range from 90 VAC to 264 VAC. These voltages can cause serious injury or death. ¹ Only authorized service personnel are permitted to remove protective covers and access the energized components of this system.

The system-troubleshooting flow chart is designed as a closed loop. The main flow chart branches to secondary flow charts, each of which returns to the main flow chart for continued troubleshooting. With this method, fault conditions involving more than one error may be resolved. The flow chart is not intended to resolve all possible fault conditions; it is designed to resolve 80% of typical field situations.

Using the Flow Chart

0

Always start at the top and check each item in the Check column. The chart on the next page provides details for each Check item. Example: The system does not boot correctly. Do not go directly to the “Boot-up” flow chart. First confirm power by checking the LED on the Host Module OEM Sub Panel. If this LED is OK, and the monitor LED is on, then proceed to the “Boot-up” flow chart. When the “Boot-up” flow chart tasks are complete, then confirm the remaining items.

ACUSON CV70

US05-101.840.01.01.02 06.04

Page 6 of 36 sd

Siemens Medical Solutions USA, Inc.

Troubleshooting

Fig. 1:

7

Start

Siemens Medical Solutions USA, Inc.

US05-101.840.01.01.02 06.04

Page 7 of 36 sd

ACUSON CV70

8

Troubleshooting

Fig. 2:

Chart A

Fig. 3:

Chart C

ACUSON CV70

US05-101.840.01.01.02 06.04

Page 8 of 36 sd

Siemens Medical Solutions USA, Inc.

Troubleshooting

Fig. 4:

Chart E

Fig. 5:

Chart G

Siemens Medical Solutions USA, Inc.

US05-101.840.01.01.02 06.04

9

Page 9 of 36 sd

ACUSON CV70

10

Troubleshooting

Fig. 6:

ACUSON CV70

Chart G1

US05-101.840.01.01.02 06.04

Page 10 of 36 sd

Siemens Medical Solutions USA, Inc.

Troubleshooting

Fig. 7:

11

Chart G3

Siemens Medical Solutions USA, Inc.

US05-101.840.01.01.02 06.04

Page 11 of 36 sd

ACUSON CV70

12

Troubleshooting

Fig. 8:

ACUSON CV70

Chart I

US05-101.840.01.01.02 06.04

Page 12 of 36 sd

Siemens Medical Solutions USA, Inc.

Troubleshooting

Fig. 9:

13

Chart K

Siemens Medical Solutions USA, Inc.

US05-101.840.01.01.02 06.04

Page 13 of 36 sd

ACUSON CV70

14

Troubleshooting

Fig. 10: Chart M

ACUSON CV70

US05-101.840.01.01.02 06.04

Page 14 of 36 sd

Siemens Medical Solutions USA, Inc.

Troubleshooting

15

Diagnostic LED Indicators

0

Several groups of diagnostic LED indicators are provided; these indicate system or component status.

System Status

0

Diagnostic light emitting diodes (LEDs), located on the side of the host module, provide indications of the boot progress, the system status, and some subsystem failures. The legend shown (Fig. 11 / p. 15) depicts the LED status for the Off (extinguished), On (illuminated), and flashing (On-Off) conditions. In the “Component Designations” section of this manual provides a table called Subsystem and Component Designations that identifies the acronyms used to identify devices.

Fig. 11: LED Legend

The status of the LEDs as they are progressively extinguished during the boot sequence are shown in (Fig. 12 / p. 16). The LED status after the boot sequence completes for systems with and without the DIMAQ-IP module connected, are shown in (Tab. 1 / p. 17).

NOTE

The LED designated with the V character represents the 5V supply. This LED illuminates when 5V power is available. The RTC LED indicates the operation of the real-time clock. The Windows 2000 operating system in the DIMAQ-IP boots during the sequence shown in (Fig. 12 / p. 16).

Siemens Medical Solutions USA, Inc.

US05-101.840.01.01.02 06.04

Page 15 of 36 sd

ACUSON CV70

16

Troubleshooting

Fig. 12: Diagnostic LED Status During the Boot Sequence

After completion of the boot sequence, the LEDs indicate the following conditions:

ACUSON CV70

US05-101.840.01.01.02 06.04

Page 16 of 36 sd

Siemens Medical Solutions USA, Inc.

Troubleshooting Tab. 1

17

System LEDs

FRU

LED

Host Module

Description During

1

On then Off when initial is in progress

2

On when Windows CE and system software is loading

3

On then Off when LCD and startup image displayed

4

On then Off when LCD and startup image displayed

5

On then Off when Control Panel initialized

6

On then Off when graphics initialized

7

Momentarily On then Off when video initialized

8

Momentarily On then Off when CPU initialized

V

+5V (Always On)

Host Module

After 1

Off when completed

2

On when Windows CE and system software initialized

3

On when Control Panel keys are operated

4

On indicates E Module error

5

On when DIMAQ-IP connected

6

Flashes at one-second intervals; indicates Host Module CPU running

7

On indicates cooling fan failure: MO Drive fan, E Module fan, or Power Supply fan

8

On indicates E Module temperature exceeds 80°C (176°F)

V

+5V (Always On)

Monitor

Front bezel, lower

Green indicates power on and video initialized

MO Drive

Front bezel

On (green) during power up and disk access

CD-RW Drive

Front bezel

Green: power on

Siemens Medical Solutions USA, Inc.

Yellow: disk access

US05-101.840.01.01.02 06.04

Page 17 of 36 sd

ACUSON CV70

18

Troubleshooting FRU

LED

Description

Power Supply Module (A07-3)

D53 (left)

D+12V

D52

D+24V

D54

D-10V

D55

D+5V

D56

D+3V

D51 (right)

RC+3V

Power Supply Status

0

Power supply-indicating LEDs are located on the A03 switching regulator of the A07 PCB that is located under the EMI bracket (Fig. 13 / p. 18).

Fig. 13: Location of the A07 PCB

From left to right, these LEDs indicate the presence of the voltages: 12 VDC, 24 VDC, -10 VDC, 5 VDC, 3 VDC, and RC+3 V. These LEDs correspond to the J4 test points located on A07-1.

ACUSON CV70

US05-101.840.01.01.02 06.04

Page 18 of 36 sd

Siemens Medical Solutions USA, Inc.

Troubleshooting

19

Diagnostic and Error Log Messages

0

The following table lists the diagnostic messages and the error messages likely to be encountered when a failure occurs on a specific printed circuit board (PCB). Diagnostic messages, announcing the failed spare part (FRU), are displayed on the monitor, while error messages appear in the error log and usually provide more details. Use the “Expected Failure Location”, “Failed PCB”, and “Failed PCB Name” columns to determine the location and identity of the failed component indicated by each message. No. Diagnostics Message

Error Log Message

Expected Failure Location

Failed Failed PCB PCB Name

1

Fail: Operator: add=0xXXXX val(0x55)=0xXX

Fail: Operator: add=0xXXXX Control Panel A10 val(0x55)=0xXX

OPCT

2

Fail: Operator: add=0xXXXX val(0xaa)=0xXX

Fail: Operator: add=0xXXXX Control Panel A10 val(0xaa)=0xXX

OPCT

3

Fail: Hostbox: HostBox Return Path (Card-Host Test Pat.)

Fail: HostBox: HostBox Return Path (Card-Host Test Pat.)

4

Fail: CardBox: CardBox Return Fail: CardBox: CardBox Path (Card-Host Test Pat.) Return Path (Card-Host Test Pat.)

5

Host Module

A70

DVSC

E Module

-

E Module

Fail: HostBox: HostBox Test Fail: HostBox: HostBox Test Host Module Pat. Gen.(Card-Host Test Pat.) Pat. Gen.(Card-Host Test Pat.)

A70

DVSC

6

Fail: CardBox: CardBox (Card-Host Test Pat.)

Fail: CardBox: CardBox (Card-Host Test Pat.)

E Module

A66

DTPS

7

Fail: Card/Host: CardBox and HostBox (Card-Host Test Pat.)

Fail: Card/Host: CardBox and HostBox (Card-Host Test Pat.)

E Module or Host Module

A66/A 70

DTPS/ DVSC

8

Fail: Cable: Cable between Fail: Cable: Cable between Card and Host (Card-Host Test Card and Host (Card-Host Pat.) Test Pat.)

Cable between E Module and Host Module

-

Cable

9

Fail: HostBox: HostBox (Card-Host Test Pat.)

Fail: HostBox: HostBox (Card-Host Test Pat.)

Host Module

A70

DVSC

10

Fail: HostBox: HostBox Return Path (DIMAQ-IP-Host Test Pat.)

Fail: HostBox: HostBox Return Path (DIMAQ-IP-Host Test Pat.)

Host Module

A70

DVSC

Siemens Medical Solutions USA, Inc.

US05-101.840.01.01.02 06.04

Page 19 of 36 sd

ACUSON CV70

20

Troubleshooting

No. Diagnostics Message

Error Log Message

Expected Failure Location

Failed Failed PCB PCB Name

11

Fail: DIMAQ-IP: DIMAQ-IP Return Path (DIMAQ-IP-Host Test Pat.)

Fail: DIMAQ-IP: DIMAQ-IP Return Path (DIMAQ-IP-Host Test Pat.)

DIMAQ-IP

A75

SPCL

12

Fail: HostBox: HostBox Test Fail: HostBox: HostBox Test Host Module Pat. Gen.(DIMAQ-IP-Host Test Pat. Gen.(DIMAQ-IP-Host Pat.) Test Pat.)

A70

DVSC

13

Fail: DIMAQ-IP: DIMAQ-IP(DIMAQ-IP-Host Test Pat.)

Fail: DIMAQ-IP: DIMAQ-IP(DIMAQ-IP-Host Test Pat.)

DIMAQ-IP

A75

SPCL

14

Fail: DIMAQ-IP/Host: DIMAQ-IP and HostBox (DIMAQ-IP-Host Test Pat.)

Fail: DIMAQ-IP/Host: DIMAQ-IP and HostBox (DIMAQ-IP-Host Test Pat.)

DIMAQ-IP or Host Module

A70/A 75

DVSC/ SPCL

15

Fail: Cable: Cable between DIMAQ-IP and Host (DIMAQ-IP-Host Test Pat.)

Fail: Cable: Cable between DIMAQ-IP and Host (DIMAQ-IP-Host Test Pat.)

Cable between E Module and DIMAQ-IP

-

Cable

16

Fail: HostBox: HostBox (DIMAQ-IP-Host Test Pat.)

Fail: HostBox: Hostbox (DIMAQ-IP-Host Test Pat.)

Host Module

A70

DVSC

17

Fail: HostBox: DIMAQ-IP-Host- Fail: HostBox: Box Test DIMAQ-IP-HostBox Test

Host Module

A70

DVSC

18

Fail: HostBox: CardBox-HostBox Test

Fail: HostBox: CardBox-HostBox Test

Host Module

A70

DVSC

19

Work Memory Error!

-

Recycle -/A70 power. If the error message is displayed again, exchange Host Module.

20

Fail: D+3V

A+3V=Upper Voltage Error!: Power ModV=XXXX etc. ule

21

Fail: A-10V

A-10V=Lower Voltage Error!: V=XXXX etc.

22

Fail: A+5V

A+5V=Lower Voltage Error!: Power ModV=XXXX etc. ule

ACUSON CV70

US05-101.840.01.01.02 06.04

-/DVSC

A03/A 07

SWRG/ PDST

A03/A 07

SWRG/ PDST

Power Module

Page 20 of 36 sd

Siemens Medical Solutions USA, Inc.

Troubleshooting

21

No. Diagnostics Message

Error Log Message

23

Fail: D+5V

D+5V=Lower Voltage Error!: Power ModV=XXXX etc. ule

A03/A 07

SWRG/ PDST

25

Fail: A+12V

A+12V=Lower Voltage Error!: V=XXXX etc.

Power Module

A03/A 07

SWRG/ PDST

26

Fail: A+90V

A+90V=Lower Voltage Error!: V=XXXX etc.

Power Module

A66

DTPS

27

Fail: D+24V

D+24V=Lower Voltage Error!: V=XXXX etc.

Power Module

28

Fail: HV-37V

HV-37V=Lower Voltage Error!: V=XXXX etc.

Power Module

A66

DTPS

29

Fail: A+BV

A+BV=Lower Voltage Error!: Power ModV=XXXX etc. ule

A66

DTPS

30

Fail: A-90V

A-90V=Lower Voltage Error!: V=XXXX etc.

Power Module

A66

DTPS

31

Fail: HV+155V

HV+155V=Lower Voltage Error!: V=XXXX etc.

Power Module

A66

DTPS

32

Fail: No data in EEPROM of DVSC

-

A70

DVSC

33

Fail: HostBox: SDRAM Error

DVSC: SDRAM Read/Write Error

Host Module

A70

DVSC

34

Fail: HostBox: SRAM Error

DVSC: SRAM Read/Write Error

Host Module

A70

DVSC

35

Fail: HostBox: RTC Error

DVSC: RTC function Error

Host Module

A70

DVSC

36

Fail: HostBox: SCSI Error

DVSC: SCSI Controller Register Error

Host Module

A70

DVSC

37

Fail: HostBox: Flash ROM Error

DVSC: Flash ROM Read Error

Host Module

A70

DVSC

38

Fail: CardBox: RXBFA Register RXBF A: Register Error Read/Write Error

E Module

A46

RXBF A

39

Fail: CardBox: RXBFB Register RXBF B: Register Error Read/Write Error

E Module

A46

RXBF B

40

Fail: No data in EEPROM of RXBFB

A46

RXBF B

Siemens Medical Solutions USA, Inc.

Expected Failure Location

-

US05-101.840.01.01.02 06.04

Page 21 of 36 sd

Failed Failed PCB PCB Name

ACUSON CV70

22

Troubleshooting

No. Diagnostics Message

Error Log Message

Expected Failure Location

Failed Failed PCB PCB Name

41

Fail: CardBox: BWPP CMEM(M) Error

BWPP: CMEM(B)63 Read/Write Error

E Module

A51 on A48

CMEM on BWPP

42

Fail: CardBox: BWPP CMEM(M) Error

BWPP: CMEM(B)255 Read/Write Error

E Module

A51 on A48

CMEM on BWPP

43

Fail: CardBox: SUBC RUN Error

SUBC: RACS Run Error

E Module

A50

SUBC

44

Fail: No data in EEPROM of SUBC

-

A50

SUBC

45

Fail: CardBox: CFPR CMEM (Mc) Error

CFPR: CMEM (Mc) Read/Write Error

E Module

A56

CFPR

46

Fail: CardBox: CFPR CMEM(V) Error

CFPR: CMEM(V)63 Read/Write Error

E Module

A56

CFPR

47

Fail: CardBox: CFPR CMEM(V) Error

CFPR: CMEM(V)255 Read/Write Error

E Module

A56

CFPR

48

Fail: CardBox: CFPR CMEM(TP) Error

CFPR: CMEM(TP)63 Read/Write Error

E Module

A56

CFPR

49

Fail: CardBox: CFPR CMEM(TP) Error

CFPR: CMEM(TP)255 Read/Write Error

E Module

A56

CFPR

50

Fail: CardBox: CFPR RAM (NoiseCutFPGA) Error

CFPR: Noise Cut FPGA U38 Error

E Module

A56

CFPR

51

Fail: CardBox: CFPR SDRC(U14, U19) Error

CFPR: SDRC U14,U19 Error

E Module

A56

CFPR

52

Fail: CardBox: CFPR FPGA(U38) Error

CFPR: Noise Cut FPGA RAM Error

E Module

A56

CFPR

53

Fail: CardBox: CFPR ASIC(U20) Error

CFPR: CFP2 ASIC U20 Error

E Module

A56

CFPR

54

Fail: CardBox: CFPR LUTRAM(U4) Error

CFPR: LUT RAM U4 Error

E Module

A56

CFPR

55

Fail: No data in EEPROM of CFPR

-

A56

CFPR

56

Fail: CardBox: DPPR(DSP) Run Error

DPPR: DSP Program Hang up

A56-1

DPPR

ACUSON CV70

US05-101.840.01.01.02 06.04

E Module

Page 22 of 36 sd

Siemens Medical Solutions USA, Inc.

Troubleshooting

23

No. Diagnostics Message

Error Log Message

Expected Failure Location

Failed Failed PCB PCB Name

57

Fail: CardBox: DPPR(DSP) Misloading

DPPR: DSP Program Loading Error

E Module

A56-1

DPPR

58

Fail: CardBox: DPPR(DSP) Ini- DPPR: DSP Program Init tial Error Error

E Module

A56-1

DPPR

59

Fail: CardBox: DPPR CMEM(D) Error

DPPR: CMEM(D) Read/Write Error

E Module

A56-1

DPPR

60

Fail: No data in EEPROM of DPPR

-

A56-1

DPPR

61

-

FAN Stop Error

MO, Software Regulated Fan, or Rear FAN is stopped

MO Fan, Rear Panel Fan, or SWRG Fan

62

-

Temperature Error

E Module LED 8

-

63

-

Warning: Operator is not connected.

Control Panel A10

OPCT

64

-

SRAM Error Code XXXX

Host Module

DVSC

Siemens Medical Solutions USA, Inc.

US05-101.840.01.01.02 06.04

Page 23 of 36 sd

-

A70

ACUSON CV70

24

Troubleshooting

Troubleshooting the DIMAQ-IP

0

Use the following instructions to troubleshoot the DIMAQ-IP.

Review Switch Failure

0

If the DIMAQ-IP screen does not display when the Review button is selected, use the following instructions to troubleshoot the failure. 1. Verify that a patient name and ID number are entered. 2. Verify that HDD is selected (not MOD) for Storage on the Preset Menu. 3. Access DIMAQ-IP through the Service Menu by selecting the DIMAQ-IP Maintenance button; a blank screen indicates a DIMAQ-IP failure. - If the trackball and the Set buttons operate properly, the DIMAQ-IP software and Windows are running, but applications are not active. - Loss of the trackball control indicates the loss of the internal Ethernet communication. - Restart the DIMAQ-IP by rebooting Windows 2000 (Start→ Restart). - Observe the screen during the reboot to detect HDD error messages. - If errors are detected, run the Scan Disk application from the Start→ Programs →Accessories →System Tools → Scan Disk menu item. - If no errors are detected on the HDD, reboot the system. 4. If DIMAQ-IP still does not function properly, reload the DIMAQ-IP software.

Corrupt Images and Clips

0

If corrupt images or clips display after pressing Review, the following failures can be the cause. 1. New images or clips produced by real-time scanning are corrupt, but images loaded from studies on the HDD or from a CD are not corrupt. Corruption of ultrasound images is caused by a failure in the host module or the failure of the image transfer path (serial). Inspect and test the host module and the serial path; replace the host module if necessary. 2. New images from real-time scanning are not corrupt, but images recalled from the HDD or from a CD are corrupt. The corruption of images loaded from the HDD or a CD is caused by the DIMAQ-IP software or by the CD-R/RW drive software. Reload the DIMAQ-IP software to resolve the problem.

ACUSON CV70

US05-101.840.01.01.02 06.04

Page 24 of 36 sd

Siemens Medical Solutions USA, Inc.

Troubleshooting

25

Troubleshooting DICOM

0

Please refer to “Troubleshooting” in the “Configuring DICOM” section for detailed information.

Siemens Medical Solutions USA, Inc.

US05-101.840.01.01.02 06.04

Page 25 of 36 sd

ACUSON CV70

26

Troubleshooting

DIMAQ-IP BIOS Corruption - CMOS Error Solution

0

In the event of DIMAQ-IP CMOS memory loss, the DIMAQ-IP will not boot. The System will display the following Error message: Crescendo application is not available. Check DIMAQ-IP In the DIMAQ-IP Maintenance mode, the BIOS screen will display the following message: Warning! CMOS Save and Restore Error! An error occurred while attempting to erase Flash. Error 0251-System CMOS Checksum Bad-Default Configuration used. Warning 0211-Keyboard Error Press F1 to resume, F2 to set up. The DIMAQ-IP boot is stopped at waiting for input: F1-Resume or F2-Setup. To resolve the problem, perform the following instructions. 1. Power off the system. 2. Locate a PS2 keyboard. 3. Open the upper-rear panel of the system. Refer to the section. 4. Connect the PS2 keyboard to the lower left corner of the DIMAQ-IP (Fig. 14 / p. 26).

Fig. 14: Connect the PS2 Keyboard to the DIMAQ-IP

5. Power on the system, and wait for boot sequence to complete. 6. Press F4 Preset Menu from to display the Preset Menu. 7. Select Service to display the authentication screen. 8. Type the service password and select Ok. 9. From the Service Main Menu, select the DIMAQ-IP Maintenance button to display the DIMAQ-IP boot-up screen. 10. Press the F2 key from the external keyboard.

ACUSON CV70

US05-101.840.01.01.02 06.04

Page 26 of 36 sd

Siemens Medical Solutions USA, Inc.

Troubleshooting

27

11. The BIOS screen is displayed (Main, Advanced, Security, Power, Boot, Exit). Use the left and/or right arrow key(s) to select Boot. 12. Check that the boot sequence has been set as follows: ATAPI CD-ROM + Removable devise + Hard Disk Drive 13. If the boot sequence is not shown as in the previous step, for example, it may be shown as: + Removable devise + Hard Disk Drive ATAPI CD-ROM Then do the following: - Use the up arrow and/or down arrow key(s) to select the ATAPI CD-ROM. - Use the + key to change its position, until it shown as in step 12. 14. Use the left and/or right arrow key(s) to select Exit. 15. Check that the configuration of Exit has been set as follows: CMOS Restore Condition Always) Save CMOS to Flash Erase CMOS from Flash Restore CMOS from Flash 16. If the configuration is not as in the previous step, for example, it may be shown as: CMOS Restore Condition (Never) Save CMOS to Flash Erase CMOS from Flash Restore CMOS from Flash Then, use the + to change the setting until “CMOS Restore Condition” is shown as Always. 17. Use the down arrow key to select Save CMOS to Flash and execute Save CMOS to Flash. 18. Press F10 from the external keyboard to save the setting. 19. Press Set from the external keyboard to confirm saving configuration and exiting from BIOS. 20. DIMAQ-IP will restart automatically, and wait for the boot sequence to complete. 21. Power off the system. 22. Disconnect the PS2 keyboard. 23. Power on the system, and wait for the boot sequence to complete.

Siemens Medical Solutions USA, Inc.

US05-101.840.01.01.02 06.04

Page 27 of 36 sd

ACUSON CV70

28

Troubleshooting 24. Verify the system and DIMAQ-IP function. 25. Replace the upper-rear panel.

ACUSON CV70

US05-101.840.01.01.02 06.04

Page 28 of 36 sd

Siemens Medical Solutions USA, Inc.

Troubleshooting

29

SRAM Memory Clear

0

Overview

0

Background This procedure is valid for use with systems at software versions below 2.0. Battery-backed SRAM memory may become corrupt after an unexpected power loss during system operation. If this happens, the following may occur:

• The system may not power back up to imaging. • The system may display errors or display the message No System. Insert System Disk. Reloading system software may not be possible without first clearing battery-backed SRAM. When the problem occurs the Unknown file error messages are typically displayed before and after software reload attempts. Purpose This procedure describes how to clear the battery-backed SRAM so that the software can be reloaded. Two methods are provided: a software solution (recommended) and a hardware solution.

• Software Solution - This method uses the Memory Repair Disk to clear SRAM memory and is the most effective and recommended method.

• Hardware Solution - This method removes battery power from the SRAM in the Host module to clear the memory and is provided as a alternate to the recommended software method above.

CAUTION

Clearing SRAM memory clears system Software Option Keys, QuickSets, Preset data, and Service data. ¹ Obtain the Option Key MO disks and the Preset MO disk to restore the system to the previous operating condition. If Preset data, Service data, and QuickSets were not previously backed up to the Preset MO disk, manually reconfigured the system to customer preference. Clearing SRAM memory should not effect exam data stored on the DIMAQ-IP hard drive.

CAUTION

Reloading DIMAQ-IP software overwrites all data on the DIMAQ-IP hard drive. ¹ Therefore, before reloading DIMAQ-IP software, customer data on the DIMAQ-IP hard drive that must be maintained should be backed up to CD.

Siemens Medical Solutions USA, Inc.

US05-101.840.01.01.02 06.04

Page 29 of 36 sd

ACUSON CV70

30

Troubleshooting Setup Requirements Kit Contents The following parts are included in the Memory Repair Disk Kit, M/N 78 51 707.

• Memory Repair Disk (M/N 74 84 772) • SRAM Memory Clear Procedure (M/N 78 51 251) Tools and Materials Number 2 Phillips screwdriver Reference Documentation This manual-the SONOLINE G50 and G60 S Service Manual (M/N 74 82 255) Time Required Approximately 1 hour is required to perform this procedure.

Instructions

0

Preparation Before working on the ultrasound system, observe the following guidelines. 1. Power off the system. 2. Disconnect all transducers from the system. 3. Move the system to a level location that provides one meter of work space around the system. 4. Engage each of the four wheel locks to prevent the system from moving during the service activity. Safeguard all hardware and fasteners removed from the system: These items my be needed later in the procedure.

CAUTION

The CV70 contains components that are sensitive to damage by electrostatic discharge (ESD). ¹ Observe ESD protective procedures and use ESD protective equipment including an ESD wrist strap, when the protective covers are removed from the system.

ACUSON CV70

US05-101.840.01.01.02 06.04

Page 30 of 36 sd

Siemens Medical Solutions USA, Inc.

Troubleshooting

31

Clearing the SRAM - Software Method CAUTION

0

Clearing SRAM memory clears system Software Option Keys, QuickSets, Preset data, and Service data. ¹ Obtain the Option Key MO disks and the Preset MO disk to restore the system to the previous operating condition. If Preset data, Service data, and QuickSets were not previously backed up to the Preset MO disk, manually reconfigured the system to customer preference. Clearing SRAM memory should not effect exam data stored on the DIMAQ-IP hard drive.

Perform the following steps to clear the battery-backed SRAM in the CV70. This is the most effective and recommended method. 1. Power off the system. 2. Insert the Memory Repair Disk into the MO drive. 3. Wait 15 seconds from power off, then power on the system. 4. The system boots from the disk, and a message box displays. 5. Eject the Memory Repair Disk from the MO Drive. 6. Select the Clear button in the message box to clear the SRAM. 7. When the progress bar has filled completely to the right, select the Start button in the message box to continue the boot process. NOTE

Print any error messages that occur; then, press Cancel or OK to continue the boot process.

When the boot sequence completes, the normal imaging screen should display. 8. Power off the system, wait 15 seconds, then power on the system again. The system should power on to the normal imaging screen, but all Software Option Keys, Quicksets, Presets data, and Service data will not display as previously set. NOTE

Optionally, choose to reload the system software at this time, but it should not be necessary.

9. Proceed to “Restoring the Software Option Keys and System Presets”.

Siemens Medical Solutions USA, Inc.

US05-101.840.01.01.02 06.04

Page 31 of 36 sd

ACUSON CV70

32

Troubleshooting

Clearing the SRAM - Hardware Method CAUTION

0

Clearing SRAM memory clears system Software Option Keys, Quicksets, preset data, and service data. ¹ Obtain the option key MO disks and the preset MO disk to restore the system to the previous operating condition. If preset data, service data, and Quicksets were not previously backed up to the preset MO disk, manually reconfigured the system to customer preference. Clearing SRAM memory should not effect exam data stored on the DIMAQ-IP hard drive.

If the memory repair disk is not available at the time it is needed, an optional method described below that involves removing battery power from the sram in the Host Module to clear the memory. 1. Power the system off and switch off the MAINS circuit breaker. 2. Remove the upper rear panel as described in . 3. Remove the lower rear panel as described in (US05-101.841.01 / Removing the Rear Lower Panel). 4. Loosen the two screws that secure the access cover to the bottom of the Host Module to expose the SRAM battery (Fig. 15 / p. 32). 5. Disconnect P443 from J5 on the Host Module. observe that the green wire on P443 mates with J5 pin 1.

Fig. 15: Location of the SRAM Battery

6. Short J5 pins 1 and 2 together and to ground. 7. With P443 disconnected, keep the system off for at least 15 minutes. 8. Reconnect P443 to J5.

ACUSON CV70

US05-101.840.01.01.02 06.04

Page 32 of 36 sd

Siemens Medical Solutions USA, Inc.

Troubleshooting

33

9. Switch the MAINS circuit breaker on and power on the system. As the system powers on, the following error message should display: No language. 10. Select Cancel to continue. As the system continues the boot sequence, the following errors may display: - SRAM Error - TV format (twice) If these errors display, again select Cancel (or Ok) to continue the boot sequence. The system should boot to imaging; however, the image and graphics will be offset slightly. 11. Power off the system, wait 15 seconds, then power on the system again. The system should power on to the normal imaging screen, but all Software Option Keys, Quicksets, presets data, and service data, will not display as previously set. NOTE

Optionally, choose to reload the system software at this time, but it should not be necessary to do so.

12. Proceed to “Restoring the Software Option Keys and System Presets”.

Restoring the Software Option Keys and System Presets

0

This section describes how to restore the Software Option Key settings and the system presets after clearing SRAM. 1. Set the correct date and time. With the system powered on to imaging, do the following: -

Press the F4 key. Select General (if it is not already selected). Correct the date and time. Select Set. Select Save.

2. Verify that the date and time are correct on the imaging screen. 3. Verify the system serial number as follows: -

Press the F4 key. Select Service.” Type the service password. Select System and verify that the system serial number is correctly listed. If so, proceed to step 4. - If the serial number is not correct, select Input, select Set, then enter the correct serial number. - Select Ok.

Siemens Medical Solutions USA, Inc.

US05-101.840.01.01.02 06.04

Page 33 of 36 sd

ACUSON CV70

34

Troubleshooting 4. Return to the imaging screen as follows: - Select Home. - Select Exit. The imaging screen should display. 5. Press the F4 key. 6. Select Installation from Key Disk. NOTE

Wait three minutes for the DIMAQ-IP to complete the boot sequence. When complete, “DIMAQ-IP” is listed on the “Installed Option” screen (Fig. 16 / p. 34). If “DIMAQ-IP” is not listed, re-install both the system and DIMAQ-IP software. For additional information on installing software, refer to the corresponding version of the DIMAQ/System Software procedure.

7. Insert each Software Option Key disk and select Install TO restore the system options. 8. Press the Escape/View key to exit the preset menu.

Fig. 16: Installed Option Screen

9. Press Review key to display the DIMAQ-IP study screen. NOTE

If the dimaq-ip study screen is not displayed, you may need to reload DIMAQ-IP software. For additional information on installing software, refer to the corresponding version of the DIMAQ/System Software procedure.

10. Insert the Backup Preset MO disk.

ACUSON CV70

US05-101.840.01.01.02 06.04

Page 34 of 36 sd

Siemens Medical Solutions USA, Inc.

Troubleshooting

35

11. Select the MO radio button from the DIMAQ-IP study screen disk area. 12. Select the Preset tab. 13. From the File Type pull-down menu: -

Select Total System Configuration. Select the file name displayed. Select Recall. Select Ok. Ignore the Restart System message.

14. Press the Review key. 15. Select the MO radio button. 16. Select the Preset tab. 17. FROM the File Type pull-down menu: -

Select DICOM Settings. Select the file name to restore. Select Recall. Select Ok. Ignore the Restart System message.

18. Press the Review key. 19. Select the MO radio button. 20. Select the Preset tab, then proceed as follows: -

Select the Recall Presets button. Select the Quicksets boxes to restore. Select Ok. Select Exit. NOTE

Optionally, restore selected service backups from the SERVICE Backup/restore menu but it should not be necessary to boot the system at this time.

21. Cycle the system power. - Power off the system. - Wait 15 seconds. - Power on the system again. 22. Complete a functional check of the system and peripherals for details (if required), refer to .

Siemens Medical Solutions USA, Inc.

US05-101.840.01.01.02 06.04

Page 35 of 36 sd

ACUSON CV70

36

Troubleshooting

Service Notes

0

Changing Video Format

0

To change the video format from the Imaging application screen: 1. Type the following using all uppercase characters: XAINMO. 2. To change the format to PAL or NTSC: - PAL format, enter tvpal$on. - NTSC format, enter tveia$on. 3. Cycle the system power to apply the video format change. 4. On the System Information page of the sub control panel LCD screen, verify the video format change: - Press LCD Control on the sub control panel (A/Fig. 17 / p. 36). - Press the bottom-right selection button on the sub control panel to display the System Information page (B/Fig. 17 / p. 36).

Fig. 17: Verify Video Format Change

ACUSON CV70

US05-101.840.01.01.02 06.04

Page 36 of 36 sd

Siemens Medical Solutions USA, Inc.