Environmental Monitoring R Raghunandanan, Mumbai Director – ISPE India 26th February 2011

Agenda • Introduction

26-02-2011

[email protected]

2

Agenda • Introduction

26-02-2011

[email protected]

2

Environment includes All the factors impacting quality and safety of pharmaceutical products



Facility



Utilities



Machines



Personnel



26-02-2011

[email protected]

3

Why environmental monitoring is important GMP requirement Facilities are designed to protect p rotect products from contamination or deterioration Facility must operate operate to meet the design criteria Only routine monitoring can prove that facilities are functioning as designed. • •





26-02-2011

[email protected]

4

Why environmental monitoring is important • •



Regulatory Requirements, US FDA Microbiological control of drug manufacturing areas: Statutory requirement as per 21 CFR 211 as stated in Sections 211.46, 211.56, and 211.113 211 CFR 211.46, Ventilation, air filtration, air heating and cooling: Adequate control over microorganisms, dust, humidity and temperature. Pre filters and particulate matter air filters for air supplies to production areas





26-02-2011

[email protected]

5

Why environmental monitoring is important • •

• •

Regulatory Requirements EM is a requirement of GMP as per WHO, MHRA, MCC,TGA etc too. Schedule M requirements Sec 8.2.1. The licensee shall prevent mix-up and cross-contamination of drug material and drug product (from environmental dust) by proper airhandling system, pressure differential, segregation, status labeling and cleaning. Proper records and Standard Operating Procedures thereof shall be maintained.

26-02-2011

[email protected]

6

Why environmental monitoring is important Schedule M requirement All environmental parameters listed under para 3.1 to 3.10 shall be verified and established at the time of installation and thereafter monitored at periodic intervals. The recommended frequencies of periodic monitoring shall be as follows :• •

26-02-2011

[email protected]

7

Why environmental monitoring is important Schedule M requirement - recommended frequencies (a) Particulate monitoring in air 6 Monthly. (b) HEPA filter integrity testing (smoke testing) Yearly (c) Air change rates 6 Monthly. (d) Air pressure differentials Daily. (e) Temperature and humidity Daily (f) Microbiological monitoring by settle plates and/or swabs in aseptic areas - daily, and at decreased frequency in other areas •

26-02-2011

[email protected]

8

What impacts from environment •

Product deterioration Microbiological deterioration Thermal deterioration Humidity induced deterioration Product contamination Particulate contamination Cross contamination Increased bio-burden Sterility failures • •







• •



26-02-2011

[email protected]

9

Different environments •





Product exposed areas Aseptic preparation Aseptic filling, blending Preparatory areas Solution preparation Component preparation General areas Component washing

26-02-2011

[email protected]

10

What need to be monitored •

• •

• •



• •



Non-viable air monitoring (particulates) Viable air monitoring (microbiological) Surface monitoring (facility and equipment) Personnel monitoring (garments & gloves) Temperature and humidity Pressure differentials (∆p) Gases and other utilities (including water) Disinfectants Detergents and washing agents

26-02-2011

[email protected]

11

Different classes of environments EU Grade

ISO Class

Class

US FED STD

A

5

100

M 3.5

B

6

1000

M 4.5

C

7

10,000

M 5.5

D

8

100,000

M 6.5

(US FED STD now replaced by ISO 14644-1 though we frequently use the terminology Class 100 etc.) 26-02-2011

[email protected]

12

Available standards and guidance EU Annex 1, 2008 USP 32, General Chapter ISO 14644 Schedule M of D&C Act and Rules of India Guidance for industry, US FDA – Sterile Drug Products produced by Aseptic Processing-Sept 2004 21 CFR Part 211.42 21 CFR Part 211.46 (and others) • • • • •

• •

26-02-2011

[email protected]

13

Available standards and guidance •

It is important to remember that •

All standards are not exactly identical



Manufacturer to follow the market

requirements as a minimum •

You are free to have tighter in-house

standards

26-02-2011

[email protected]

14

Area particulates (non-viable) EU Annex 1 Maximum no of particles per cubic meter At rest

In Operation

Grade

0.5 micron

5.0 micron

0.5 micron

5.0 micron

A

3520

20

3520

20

B

3520

29

352,000

2900

C

352,000

2900

3,520,000

29,000

D

3,520,000

29,000

Not defined

Not defined

26-02-2011

[email protected]

15

‘At rest’ and ‘In Operation’ • ‘At rest’  – Air flow is ON  – Equipment installed but not in use  – No personnel in the area • ‘In Operation’  – Air flow is ON  – Equipment in operation (sometimes simulated)  – Specified no. of personnel in the area • Define these for each clean room 26-02-2011

[email protected]

16

‘Microbiological monitoring standards (EU Annex 1) Limits for microbial contamination Grade

Air sample Cfu/m3

Settle plate Cfu/4 Hrs

Contact plate Cfu/plate

Glove print Cfu/glove

A

View more...