Philips
LIST OF DOCUMENTATION IN THIS BINDER:
Ä
Subsystem Manual bucky DIAGNOST TS
Ä
INALFA non sensing bucky unit
Ä
Manual Collimator Ralco
Note:
indicates document present
LIST OF ALL BINDERS FOR BUCKY DIAGNOST TS --
bucky DIAGNOST TS
--
OPTIMUS generator
--
X-ray tube
--
bucky DIAGNOST VE/VT
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
RADIOGRAPHY
Manual Order No. 4512 984 06663 bucky DIAGNOST TS bdts_01_b002
(b/00.2) ã 2000 Philips Medical Systems ALL RIGHTS RESERVED
0.1
Philips Medical Systems Development and Manufacturing Centre
INTRODUCTION AND TECHNICAL DATA
1
INSTALLATION
2
FAULT FINDING
3
REPLACEMENT
4
PROGRAMMINGS
5
ADJUSTMENTS
6
ACCEPTANCE
7
SERVICE INFORMATION
8
PARTS LIST
P
Philips
SERVICE MANUAL 704 SUBSYSTEM
9848 600 00061
Fix
9848 600 00071
Elevation
9848 600 00081
Tomo
bucky table for routine radiography diagnostics
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
bucky DIAGNOST TS
DMC Hamburg SCHEMATIC DIAGRAMS
Z1-
WIRING DIAGRAMS
Z2-
Printed in Hamburg Federal Republic of Germany 2000 Philips Medical Systems ALL RIGHTS RESERVED
Manual Order No. 4512 984 06663 bucky DIAGNOST TS (b/00.2) reg_984 06663
1
Philips
Bucky DIAGNOST TS
SERVICE MANUAL -- SUBSYSTEM Bucky DIAGNOST TS
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Type No:
Author: V. Neumann
9848 600 00061 9848 600 00071 9848 600 00081
In case there are any questions concerning this manual, please send this LOPAD via fax to 49/(0)40/5078 2481
File:bD_TS_b002
List of pages and drawings (LOPAD)
0.1
(b/00.2)
1 3.1 3.2
(b/00.2) (b/00.2) (b/00.2)
A4
Manual Order No: 4512 984 06663
(Rosa Karton)
_________________________________________ 1-0.1/0.2
(99.0)
1-1 ... 1-15
(99.0)
1Z--1.1 1Z--1.2 1Z--1.3 1Z--1.4 1Z--1.5
(99.0) (99.0) (97.1) (99.0) (99.0)
A3 A3 A3 A3 A3
1Z--5.1 1Z--5.2
(99.0) (99.0)
A3 A3
1Z--6.1 1Z--6.2 1Z--6.3
(99.0) (99.0) (00.0)
A3 A3 A3
1Z--2.1
(99.0)
A3
1Z--7
(00.0)
A3
1Z--2.2.1 1Z--2.2.2
(99.0) (99.0)
A3 A3
1Z--8
(99.0)
A3
1Z--9
(99.1)
A3
1Z--3.1 1Z--3.2
(97.1) (97.1)
A3 A3
1Z--10
(00.0)
A3
Z--50
(00.0)
A3
1Z--4.1 (97.1) A3 1Z--4.2 (97.1) A3 _________________________________________
bucky DIAGNOST TS bdts_lopad_b002
(b/00.2) 2000 Philips Medical Systems ALL RIGHTS RESERVED
4512 982 00081
3.1
Philips
bucky DIAGNOST TS 2-0.1
(99.0)
2-1 ...2- 36
(99.0)
2Z--2.0
(99.0)
2Z--10 (00.0) A3 _________________________________________ 3-0.1
(99.0)
3-1 ... 3-9 (99.0) _________________________________________ P-List 9848 600 02641 _________________________________________ Z0--1
(99.0)
Z1--1 Z1--1.1
(99.0) (99.0)
A3 A3
Z1--2 Z1--2.1
(99.0) (99.0)
A3 A3
Z1--3 Z1--3.1
(99.0) (99.0)
A3 A3
Z1--4 Z1--4.1 Z1--4.2 Z1--4.2
(99.0) (99.0) (00.0) (a/00.0)
A3 A3 A3 A3
Z1--5 Z1--5.1 Z1--5.1.1
(a/00.0) (a/00.0) (00.0)
A3 A3 A3
(99.0)
A3
Z1--6.1
Z1--7.1 (a/00.0) A3 _________________________________________ Z0--2
(99.0)
Z2--1
(00.0)
A4
Z2--2.1 Z2--2.2
(99.0) (99.0)
A3 A3
Z2--3.1 Z2--3.2 Z2--3.3
(99.0) (99.0) (99.0)
A3 A3 A3
Z2--4.1 Z2--4.2
(99.0) (99.0)
A3 A3
(a/00.0)
A3
Z2--5
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
A4
3.2
(b/00.2) 2000 Philips Medical Systems ALL RIGHTS RESERVED
bucky DIAGNOST TS bdts_lopad_b002
bucky DIAGNOST TS
INTRODUCTION AND TECHNICAL DATA
Philips
INTRODUCTION AND TECHNICAL DATA TEXT
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1--0.1 1. 1.1. 1.1.1. 1.1.2. 1.1.3. 1.1.4. 1.2. 1.3. 1.3.1. 1.3.2. 1.4. 1.4.1. 1.4.2. 1.5. 1.5.1. 1.5.2. 1.6. 1.7. 1.8.
Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . System bucky DIAGNOST TS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Standard Table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Tomography table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . X-ray generator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Applications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Tubes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Mechanical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Movements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Electrical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . X-ray generator OPTIMUS 50 / 65 / 80 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Type of power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Calculating the mains resistances . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Earth-leakage circuit breaker . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Emergency-OFF device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Heat emission . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1--1 1--1 1--1 1--1 1--1 1--1 1--2 1--2 1--2 1--2 1--2 1--2 1--3 1--4 1--4 1--5 1--5 1--6 1--6
2. 2.1. 2.2.
Environmental conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . X-ray generator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1--6 1--6 1--6
3. 3.1. 3.2. 3.2.1. 3.2.2. 3.2.3.
Measurements and weights . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Pre-installed system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Transport data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Box dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Roadway . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1--7 1--7 1--8 1--8 1--8 1--9
4. 4.1.
1--10
4.4. 4.4.1. 4.4.2.
Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ”Fixed table” 9848 600 00061 – bucky DIAGNOST TS with floating table top . . . . . . . . . . . . . . Height adjustable 9848 600 00071 bucky DIAGNOST TS with floating table and elevator . . . . . . . . Tomography 9848 600 00081 bucky DIAGNOST TS with floating table and elevator and tomo extension . . . . . . . . . . . . . . . . . . . . . . . . Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . bucky DIAGNOST TS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . bucky DIAGNOST VE / VT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5. 5.1. 5.2. 5.3. 5.4. 5.5. 5.6.
Site planning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Planning data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Determining the site of installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Floor fixing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Loading capacity of floor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Height of the room . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Electrical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1--13 1--13 1--13 1--13 1--13 1--13 1--13
4.2.
Philips
4.3.
bucky DIAGNOST TS INH
1999
(99.0)
Philips Medical Systems ALL RIGHTS RESERVED
1--10 1--11 1--11 1--11 1--11 1--12
1--0.1
Philips
INTRODUCTION AND TECHNICAL DATA
bucky DIAGNOST TS
6.
Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1--14
7. 7.1.
Installation materials required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1--14 1--15
DRAWINGS 1Z--1.1 1Z--1.2 1Z--1.3 1Z--1.4 1Z--1.5 1Z--2.1 1Z--2.2.1 1Z--2.2.2 1Z--3.1 1Z--3.2 1Z--4.1 1Z--4.2 1Z--5.1 1Z--5.2 1Z--6.1 1Z--6.2 1Z--7 1Z--8 1Z--9
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
bucky DIAGNOST TS elevated or tomo version for wall stand left, mechanical dimensions . . . . . . . . . bucky DIAGNOST TS elevated or tomo version for wall stand right, mechanical dimensions . . . . . . . . OPTIMUS L 50, mechanical dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . bucky DIAGNOST TS fixed table version for wall stand left, mechanical dimensions . . . . . . . . . . . . . . . bucky DIAGNOST TS fixed table version for wall stand right, mechanical dimensions . . . . . . . . . . . . . . bucky DIAGNOST TS elevated or tomo version, wall clearance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . bucky DIAGNOST TS fixed table version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . bucky DIAGNOST TS fixed table version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . bucky DIAGNOST TS for wall stand left with bucky DIAGNOST VE typical room layout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . bucky DIAGNOST TS for wall stand right with bucky DIAGNOST VE typical room layout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . bucky DIAGNOST TS for wall stand left with bucky DIAGNOST VE relation data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . bucky DIAGNOST TS for wall stand right with bucky DIAGNOST VE relation data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . bucky DIAGNOST TS elevated or tomo version, connecting surface / floor fixing . . . . . . . . . . . . . . . . . . bucky DIAGNOST TS fixed table version, connecting surface / floor fixing . . . . . . . . . . . . . . . . . . . . . . . . bucky DIAGNOST TS, VE, OPTIMUS L 50 standard version, connection diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . bucky DIAGNOST TS, VE, OPTIMUS L 50 tomo version, connection diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . bucky DIAGNOST TS, VE, OPTIMUS L 50 earthing diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . bucky DIAGNOST TS, VE system installation and cable length . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Wall connection box bucky DIAGNOST TS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1--0.2
1999
(99.0)
Philips Medical Systems ALL RIGHTS RESERVED
bucky DIAGNOST TS INH
1.
Technical data
1.1.
System bucky DIAGNOST TS
INTRODUCTION AND TECHNICAL DATA
1.1.1. Standard Table The bucky DIAGNOST TS is installed with the X-ray generator OPTIMUS L 50. The bucky DIAGNOST TS is characterized by: -- X-ray exposures at the horizontal patient with bucky unit. -- X-ray exposures at the horizontal patient without bucky unit. -- Patient area of table top 2400mm x 600mm (excl. the Alu profile) -- SID 1000mm to 1250mm. 1.1.2. Tomography table Linear tomography at the horizontal patient. -- Angle:
8° ... 40°
-- Time:
0.5 ... 4sec.
-- Layer:
0° ... 250mm
-- SID:
1000 ... 1250mm, 1100mm default
1.1.3. X-ray generator The X-ray generator OPTIMUS L 50 for radiography is based on computer-controlled converter technology. The converter operates in the non-audible frequency range. Applicational options are essentially achieved by releasing software modules. 1.1.4. Applications -- Radiography -- Tomography
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Philips
bucky DIAGNOST TS
bucky DIAGNOST TS bdts_1_990
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
1--1
Philips
INTRODUCTION AND TECHNICAL DATA
1.2.
bucky DIAGNOST TS
Tubes
Versions of the X-ray generator OPTIMUS L 50 50kW 3phase generator for 1 tube. Recommended standard tubes: RO 17 50
0.6 / 1.3 mm
R type stators only for low speed rotor control. ROT 350 Optional tubes:
1.3.
SRO 33 100 / ROT350 (for U.S.A. only) SRO 33 100 / ROT350 SRO 25 50 / ROT350 SRO 09 51 / ROT350 ”SRO” in combination with High Speed Rotor Control
9874 005 16022 9874 005 09022 9874 004 23022 9890 000 64181 9848 600 01101
Mechanical data
1.3.1. Dimensions Mechanical dimensions for wall stand left see
Z--1.1
Mechanical dimensions for wall stand right see
Z--1.2
Mechanical dimensions for OPTIMUS L 50 see
Z--1.3
1.3.2. Movements See
1.4.
Z--1.1 or Z--1.2
Electrical data
1.4.1. Table Operating voltages: 230 V 10% --15% 50/60 Hz Power consumption without tomography: Power consumption with tomography:
appr. 420 W appr. 420 W
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Compatible tube housings:
1--2
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
bucky DIAGNOST TS bdts_1_990
INTRODUCTION AND TECHNICAL DATA
1.4.2. X-ray generator OPTIMUS 50 / 65 / 80 OPTIMUS 50 Power data Mains voltage
Mains frequency Max. mains current -- at 400 V -- at 440 V -- at 460 V -- at 480 V -- at 190 V
OPTIMUS 65
OPTIMUS 80
3 x 400 V ±10% ( 380 V --5%, 415 V +6%) 3 x 400/440/460 V ±10% 3 x 480 V --10% +6% 3 x 190...343 V ±10% with external transformer, max. 50 kW (option) 49 ... 61 Hz 145 A 135 A 125 A 120 A 300 A
190 A 180 A 170 A 160 A --
230 A 215 A 210 A 205 A --
35 A 100 A at £ 240 V
50 A
50 A
Mains resistance -- at 400 V -- at 440/460 V -- at 480 V
£ 300 mW £ 350 mW £ 400 mW
£ 200 mW £ 240 mW £ 300 mW
£ 200 mW £ 240 mW £ 300 mW
--
£ 300 mW
£ 240 mW
£ 180 mW
Fuse protection (slow–blow)
at 480 V, valid for DOD only
Unit supply Max. output power Operating data Tube current Tube voltage mAs product Exposure time Exposure sequency Noise level Heat dissipation Interfacing option for....
230 V / 400 V, max. 5 A 50 kW
65 kW
80 kW
1.0 ... 650 mA
1.0 ... 900 mA
1.0 ... 1100 mA
40 ... 150kV in steps of 1kV or % 0.5 ... 850mAs 1ms ... 6/16s £ 12 exp./s £ 40 dBA 500W door contact, external radiation warning indicator
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Philips
bucky DIAGNOST TS
bucky DIAGNOST TS bdts_1_990
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
1--3
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Philips
INTRODUCTION AND TECHNICAL DATA
1.5.
bucky DIAGNOST TS
Power supply
1.5.1. Type of power supply For 50 kW X-ray generator a three-phase system is needed. The three-phase system must be balanced, i.e. the external conductor voltages against earth must be about equal, see sketch. For reasons of operational safety and compliance with international rules, operation on three-phase mains that are floating or unbalanced to earth (e.g. with one phase earthed instead of the neutral point) is impermissible unless adequate adaptation measures have been taken! Check that the sequence of phases is correct corresponding to the designations L1, L2, L3. L1 3 phase WYE
L2
X-ray installation
-- Working range 400 ... 480V measured between phases: L1 -- L2 / L2 -- L3 / L1 -- L3
L3 (N)
L1 3 phase DELTA, balanced earth or floating X-ray installation
-- Working range 400 ... 480V measured between phases: L1 -- L2 / L2 -- L3 / L1 -- L3
X-ray installation
-- Working range 400 ... 480V measured between phases: L1 -- L2 / L2 -- L3 / L1 -- L3
L2 L3
L1 3 phase DELTA, grounded
Philips
L2 L3
1--4
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
bucky DIAGNOST TS bdts_1_990
Philips
bucky DIAGNOST TS
INTRODUCTION AND TECHNICAL DATA
1.5.2. Calculating the mains resistances If possible, the sum of R1 , R2 and R3 should be smaller than the RXG required. With higher internal mains resistances the generator output is reduced respectively. Note that the cross section of lead 25mm2.
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
+
R0
l3
L1 L2
must not exceed
R1 +
R2
l1
+ R3 l3
L3
< RXG MEX 100 L1 L2 L3 357J94
-- R0 designates the mains resistance on the distributor transformer. -- R1 is dependent upon the length of lead l1 between distributor transformer and main distributor and upon the cross section selected, so: R1 = l1 x RCu
RCu from table
-- R2 is estimated to be 15mΩ EMERGENCY--OFF switch Earth--leakage circuit breaker Fuse
4mΩ 5.5mΩ 5.5mΩ
ü ý þ
15mΩ
-- R3 is dependent upon the length of lead l3 between main distributor and wall junction box and upon the cross section selected, so: R3 = l3 x RCu
RCu from table
The resistances already consider the go and return line so that the calculation can be based on simple cable lengths. Copper cross section [mm2 ]
1.6.
Resistance RCu [mW/m]
16
2.19
25 35
1.4 1.0
50
0.7
70
0.5
95
0.38
120 150
0.30 0.24
Earth-leakage circuit breaker
To be provided between fuse and X-ray installation depending upon local regulations. Siemens earth-leakage circuit breaker FSJ 50 A Rated fault current 30mA Rated current 50A
Philips
Connection terminals for wire cross sections of up to 25mm2.
bucky DIAGNOST TS bdts_1_990
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
1--5
Philips
INTRODUCTION AND TECHNICAL DATA
1.7.
bucky DIAGNOST TS
Emergency-OFF device
To be provided depending upon local regulations. There are 2 possibilities: 1. All the EMERGENCY-OFF buttons are connected in series and looped into the starting circuit (12V DC) of the generator.
1.8.
Heat emission
Table: X-ray generator:
2.
Environmental conditions
2.1.
Table
530W 500W (the generator is air cooled using natural convection)
According to PMS standard UXW 13600 class C1 (indoor temperature controlled) during regular work: -- Ambient temperature:
+ 10 °C ... +40 °C
-- Relative air humidity:
20% ... 80%
-- Air pressure:
700 ... 1100hPA
during storage / transport: -- Ambient temperature:
--25 °C ... +70 °C
-- Relative air humidity:
5% ... 100%
-- Air pressure:
700 ... 1100 hPA
2.2.
X-ray generator
according to PMS standard UXW 13600 class C1 (indoor temperature controlled) during regular work: -- Ambient temperature:
+10 °C ... +40 °C
-- Relative humidity:
0% ... 90%, no condensation
-- Air pressure:
700 ... 1100hPA
during storage / transport: -- Ambient temperature:
--20 °C ... +50 °C Note: Electrolytics require less than 40 °C for storage life.
-- Relative air humidity:
0% ... 95%, non--condensing
-- Air pressure:
700 ... 1100hPA
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
2. The EMERGENCY-OFF circuit acts on an external main contactor which switches off the power before it is fed into the generator.
1--6
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
bucky DIAGNOST TS bdts_1_990
Philips
bucky DIAGNOST TS
INTRODUCTION AND TECHNICAL DATA
3.
Measurements and weights
3.1.
Pre-installed system
Table / column (standard) Weights (kg) Box 1 Contents: manuals, table, table top, rails, SPS, set of cable, covers, tube, control handle, column, set of tomo devices (option)
without shipping box
870
750
with shipping box
270 x 97 x 196
without shipping box
265 x 89 x 190
X-ray generator 50 kW Weights (kg) with shipping box
Dimensions (cm) (L/W/H)
without shipping box
with shipping box
Box 1 Contents: generator cabinet, operating panel
226
178
210 x 82 x 84
Box 2 Contents: H.V. generator 1-tube version
100
73
77 x 67 x 80
without shipping box
195,5 x 55 52 43 x 33,5 x 6
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
with shipping box
Dimensions (cm) (L/W/H)
bucky DIAGNOST TS bdts_1_990
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
1--7
3.2.
bucky DIAGNOST TS
Transport data
3.2.1. Box dimensions Weight Gross
870kg
Loading
816kg
Net
750kg 870kg
Dimension (L x W x H) Box
270cm x 97cm x 196cm
Pallet including loading
265cm x 89cm x 190cm
Sub-pallet Sub-pallet
140cm x 89cm 125cm x 89cm
min 170cm width of roadway for gross pallet
3.2.2. Roadway
min 115cm width of roadway for sub-- pallet
1
1 Roadway for the gross pallet Canting while moving
min 170cm
min 170cm width of the floor
2 Roadway for the sub--pallets Canting while moving
min 115cm width of the floor
2 min 115cm
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Philips
INTRODUCTION AND TECHNICAL DATA
1--8
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
bucky DIAGNOST TS bdts_1_990
Philips
bucky DIAGNOST TS
INTRODUCTION AND TECHNICAL DATA
3.2.3. Unpacking The following steps should be done by four persons. Unpacking
0.5 h 0.5 h
s Dismantle the box, remove the foil.
0.5 h 0.5 h
s 2 Remove from the column the box with documents and distance tool.
1
2 Documents and distance tool
s 3 Erect the column. -- 3a Move the column approximatly 15cm to foot direction of the column. -- 3b Turn the column base about 90°. The column must be ratched in the cam.
3c
3b 90_
-- 3c Lift up the column and place it.
3a
Lay corrugated paper or something similar on the floor in the area where the column is to be moved. Apply extreme care when handling the connection field.
s Remove the wooden carrier where the column was positioned by unscrewing the colored screws.
3
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
s 1 Remove the rails from the pallet.
bucky DIAGNOST TS bdts_1_990
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
1--9
Philips
INTRODUCTION AND TECHNICAL DATA
bucky DIAGNOST TS
s 4 Remove the box including cables from the table.
The pallet is deviable
s 5 Remove the optional tomo devices from the pallet.
6
4 7
5
s Remove the wooden items by removing the colored screws.
7
4 Cables
Note The gross pallet can be splitted in two sub-pallets. s 7 Remove the table from the pallet. Use the shoulder straps. Do not lift the table by means of the table top frame.
8
Table top delivered in the pallet bottom
s 8 After unscrewing the sub–pallets from the pallet you get access to the table top. Remove the table top. s Dispose of the package locally.
4.
Scope of delivery
4.1.
”Fixed table” 9848 600 00061 – bucky DIAGNOST TS with floating table top
-- Table, including column and X-ray tube carriage -- Set of floor fixing material incl. dowels for standard concrete floors -- X-ray generator OPTIMUS L 50 including operating panel -- Complete set of cables incl. plugs at both sides -- Grid -- X-ray tube -- Collimator -- Control handle
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
s 6 Remove the tube assembly from the pallet.
1--10
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
bucky DIAGNOST TS bdts_1_990
Philips
bucky DIAGNOST TS
4.2.
INTRODUCTION AND TECHNICAL DATA
Height adjustable 9848 600 00071 bucky DIAGNOST TS with floating table and elevator
-- Table, including column and X-ray tube carriage -- Set of floor fixing material incl. dowels for standard concrete floors -- X-ray generator OPTIMUS L 50 including operating panel -- Complete set of cables incl. plugs at both sides
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
-- Grid -- X- tube -- Collimator -- Control handle
4.3.
Tomography 9848 600 00081 bucky DIAGNOST TS with floating table and elevator and tomo extension
-- Table including tomography attachment, column and X-ray tube carriage -- Set of floor fixing material incl. dowels for standard concrete floors -- X-ray generator OPTIMUS L 50 including operating panel -- Complete set of cables incl. plugs at both sides -- Grid -- X-ray tube -- Collimator
4.4.
Options
4.4.1. bucky DIAGNOST TS -- wall mounting of operating panel
9890 000 02451
-- stand for operating panel
9890 000 02441
-- Amplimat measuring chamber
9890 000 01612
-- extension main rail
4512 590 30491
-- interface for medio CP / SZ3
4512 590 30351
-- high speed rotor control
9848 600 01101
-- 3phase extension
9848 600 00122
-- SRO 33/100 / ROT 350
9874 005 09022
-- SRO 25/50 / ROT 350
9874 004 23022
-- SRO 09/51 / ROT 350
9890 000 64181
-- SRO 33/100 / ROT 350/HEW (only for U.S.A.)
9874 005 16022
-- bucky unit non sensing left
9848 500 20303
-- bucky unit non sensing right
9848 500 35703
-- manual collimator T303 for TS
9848 600 00172
-- manual collimator for TS (HHS)
9848 600 02092
-- APR
9848 600 00201
-- OPTIMUS 50 (standard) three phase -- OPTIMUS 65
three phase
-- OPTIMUS 80
three phase
-- OPTIMUS 30
single or three phase
bucky DIAGNOST TS bdts_1_990
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
1--11
-- OPTIMUS 40
bucky DIAGNOST TS
single or three phase
-- X-ray tube RO 17/50
9874 001 16002
-- grid 36/08 110F TS
9848 600 00821
-- grid 40/12 110F TS
9848 600 00831
-- H.V. cable, lengths = xx m
9806 402 6xx02
-- set of cables generator--column 14 m
9848 600 01871
-- set of cables generator--table 20 m
9848 600 01901
-- set of cables table--column 20 m
9848 600 01911
-- set of cables table--column 25 m
9848 600 01921
-- TS control grip tomo and laser
9848 600 02081
-- TS control grip standard and laser
9848 600 02071
-- table top Getalit PUR (2.4m)
9890 010 02194
-- table top Getalit (2m)
9890 010 81371
4.4.2. bucky DIAGNOST VE / VT -- grid 36/8 110F
9860 834 90111
-- grid 36/8 140F
9860 834 90141
-- grid 36/8 130F
9860 836 90131
-- grid 36/8 140F
9860 836 90141
-- grid 36/8 180F
9860 836 90181
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Philips
INTRODUCTION AND TECHNICAL DATA
1--12
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
bucky DIAGNOST TS bdts_1_990
Philips
bucky DIAGNOST TS
5.
Site planning
5.1.
Planning data
INTRODUCTION AND TECHNICAL DATA
The preliminary planning work is restricted to the selection of the suitable room. The outfit of the room has to be equipped with the conduits and pipes required and, in cases where the loading capacity of the bearing area plus weight of patient insufficient, the floor preparation for the table. Where DIN specification are not applicable the minimum wall clearances should be determined in compliance with the national regulations concerned.
5.2.
Determining the site of installation
The choice of the suitable room should be based on the layout plan. For detailed information see Z--3.1, Z--3.2, Z--4.1, Z--4.2
5.3.
Floor fixing
The table is fixed by means of 4 bolts and the column rails are fixed by means of 4 + 2 bolts (ext. rail left or right pos.) also. Use e.g. Liebig safety dowels S12 on compound plaster floor concrete class Bn 150 or Upat--PSZ anchor dowels type B or S12 on compound plaster floor concrete class Bn 250 or other national requirements. The mounting points on mounting surface have to be suitable for tensile strength of F ³ 2500N each. In case of ”computer floors” (stilted floors) an appropriate understructure or staybolts have to be used. For detailed information see Z--5
5.4.
Loading capacity of floor
Surface load table and column = 750kg plus 135kg weight of patient. Surface load X--ray generator = 250kg For detailed information see Z--5
5.5.
Height of the room
For detailed information see Z--1.1 or Z--1.2
5.6.
Electrical data
The bucky DIAGNOST TS is supplied incl. all necessary cables incl. plugs with required free length between the main components like X--ray generator and table. For detailed information see Z--6.1 and Z--7, Z--8, Z--50
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
For detailed information about loading area see Z--5.
bucky DIAGNOST TS bdts_1_990
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
1--13
Philips
INTRODUCTION AND TECHNICAL DATA
6.
bucky DIAGNOST TS
Tools
-- Groof key set for column base and transversal tube carriage
4512 590 30362
-- 25mm aluminium for AMPLIMAT adjustment -- chip-ejector for service 2422 487 89772 -- Multimeter -- Mirror (about 30 x 40 cm) -- Standard set of tools -- Fork lift -- ESD antistatic kit -- Hammer drill with 12mm carbide drillbits. -- Spirit level 1m -- Service-PC -- IBM-compatible, 640 kB RAM, 3.5” floppy disk drive, with 1 serial port -- Sub-D Adapter Kit 9/15/25 pole -- OPTIMUS L 50 utility program for generator programming supplied on a floppy disk with the generator. -- PC-Hardkey to use the OPTIMUS L 50 utility program. -- Mains resistance measuring instrument. -- Collimator adjustment tool code No: 4522 980 31521 -- Tomo test tool Code No.: 8112 108 57929 -- PC-XRGSCOPE Cable Code No.: 4512 130 56931 for OPTIMUS generator -- Rail distance tool 4512 590 30501 (included in the delivery)
7.
Installation materials required
To be ordered from the service department via PMS Hamburg: -- Wall junction box for X--ray generator 4512 103 75385 inclusive connection block (25 mm2) for mains supply and connection block (10 mm2) for unit supply. -- Relay for radiation warning indicator 4512 100 45231 1 interface relay with a floating contact (230 V, 1 A) is included in the scope of delivery for the generator. -- Grease – Molykote DX
4512 590 30321 (included with the delivery)
-- Set of three colours
4512 590 30031 (included with the delivery)
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
-- Spring balance 0 ... 200N
1--14
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
bucky DIAGNOST TS bdts_1_990
7.1.
INTRODUCTION AND TECHNICAL DATA
Cables
Cables supplied with the system. Cable length must be ordered via catalogue depending from outfit of the room. H.V. cables
9806 402 6xx02
plugs:
75kVp federal standard female connectors O3/O3
length:
6 ... 20m in steps of 2 m
capacity:
155pF/m
diameter:
16.5mm
The cable length is indicated at the 9th and 10th digits of the code No. Thermal contact cable 2-wire screened for one excess temperature switch
4512 100 66151
10-wire screened for additional supervision like temperature alarm switch, buzzer, selection indicator.
0722 215 19005
Stator cable 3 x 1.31 mm2 , screened
0722 215 02054
AMPLIMAT cable with Sub-D (15pole) 12 m 16 m 20 m 24 m
9890 000 01721 9890 000 01731 9890 000 01741 9890 000 01751
AMPLIMAT cables 9803 507 0xx02 with 3 PLUS plugs at both ends can be connected in the generator via the following adapter for each cable: Adapter for AMPLIMAT cable
4512 108 09041
The generator includes five adapters, 3Plus--Sub-D (15pole) Operating desk cable set
10m 20m 30m
9890 000 02411 9890 000 02421 9890 000 02431
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Philips
bucky DIAGNOST TS
bucky DIAGNOST TS bdts_1_990
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
1--15
Philips Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Philips
Philips Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Philips
Philips Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Philips
Philips Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Philips
Philips Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Philips
Philips Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Philips
Philips Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Philips
Philips Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Philips
Philips Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Philips
Philips Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Philips
Philips Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Philips
Philips Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Philips
Philips Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Philips
Philips Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Philips
Philips Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Philips
Philips Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Philips
S XA C
Philips Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
VE
M9f
6J2
M6f
6J3A
supply
J1
D9m
M12f
6J3B
communication
J3
D15m
D9f
7A2:J4
AEC
J5
D9m
interface
J2
D25f
3TS1 22 3TS1 23 3TS1 24 3TS1 36 3TS1 37 3TS1 51 3TS1 52 3TS1 53 4TS2 4 4TS2 5
22 23 24 36 37 51 52 53 4 5
VP
basic earth lekage circuit breaker L1
switch controller by emergency switch X23 X22
mains fuses X23
X7 X8
PMA
emergency switch
L1 N(L2) 4
Power ON bucky TS
7X11 7
L2 (N) X22
(door contact)
8
7X11
10
da
X13 X14
power supply
230V AC
L1/N
SNX16
M6
supply 4TS2:3
(bk)
4TS2:1
(rd)
4TS2:2
(wt)
XA
tube housing U V W
XA 1 1 2 3
db
stator cable, screened
dc
temperature switch
XF 7020
H.V. cables
XA 1
1 2 3
4TS2:4 4TS2:5 9HT1
O3 plugs
X-ray tube D15m 7A2:J1
SA
X1
D15f
dd
amplimat chamber
PHX1
df
release
SZ1X1
dd
amplimat chamber
VPX4
D15f
df
release
VPX1
D9f
dk
control
Ost./Schr.
bucky examination unit D15m 7A2:J2
SA
X3
bucky wall stand UB1X20 00141z63
Philips A3 00-09-27
E
PMA
bucky DIAGNOST TS C 2000 Philips Medical Systems ALL RIGHTS RESERVED
(00.0)
M12 SZ1X11
M12
bucky DIAGNOST TS, VE OPTIMUS 30 single phase Standard version Connection diagram 1Z-6.3
Philips Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Philips
Philips Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Philips
Philips Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Philips
S
Table
SN X16
4,00 mm2 4,00 mm2 4,00 mm
2
4,00 mm2
4,00 mm2 4,00 mm2
6,00 mm2
T1 T2 61 Z1
VA SA (right) UB
6,00 mm 2
2,50 mm
U
2
4,00 mm 2
UE
A1/A3 00-07-04 Kö
UF
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Philips
XA
2,50 mm
2
4,00 mm
XA
2
LA
bucky DIAGNOST TS Earthing diagram bucky DIAGNOST TS C 1997 Philips Medizin Systeme ALL RIGHTS RESERVED
(00.0)
1Z-10
Philips Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Philips
bucky DIAGNOST TS
INSTALLATION
Philips
INSTALLATION TEXT Safety regulations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2--1
1. 1.1. 1.2.
Transport data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Box dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Roadway . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2--2 2--2 2--2
2. 2.1.
Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Removal of column base from column . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2--3 2--4
3.
Mounting the rails . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2--7
Tomo option only . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2--7
Coarse Adjustment Tomo option only . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.1. Insertion of column to long rail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.2. Installation of tube arm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.3. Final alignment of the column (tomo only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.4. Installation of bucky table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.5. Installation of Tomo pivot tower (tomo only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.6. Installation of bucky-device extension (tomo only) . . . . . . . . . . . . . . . . . . . . . . . . . . 3.7. Installation of position control switches (tomo only) . . . . . . . . . . . . . . . . . . . . . . . . . 3.8. Installation of OPTIMUS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.8.1. Installation of bucky DIAGNOST TS with tomo and Medio CP/CP-H . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.8.2. Programming of APR data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.8.3. Programming tube select at control handle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.9. Installation of tabletop . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.9.1. Installation of tabletop in small rooms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.10. Setting to work . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.11. Adjustment procedure for tomo . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.11.1. Adjustment of the table level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.11.2. Correction for tomo layer height = fine adjustment under X-ray condition . . . . . . . 3.11.3. Adjustment of tomo layer height indication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.11.4. Setting the display language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.11.5. Adjustment of the tomo axis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.11.6. Alignment for center line = center of bucky . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.11.7. Checking drive and position control switches . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.12. Tomo adjustment under exposure conditions (tomo only) . . . . . . . . . . . . . . . . . . . . 3.12.1. Measurement of X-ray zone . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.12.2. Symmetry of tomo axis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.12.3. Tomography resolution test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.12.4. Tomo layer height . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.13. Final work . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.14. Acceptance test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.15. Handing over to customer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.16. Installation for tomo bar parking frame . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2--8 2--9 2--9 2--13 2--14 2--16 2--17 2--17 2--19 2--19 2--20 2--20 2--21 2--21 2--22 2--25 2--25 2--25 2--26 2--27 2--28 2--29 2--29 2--31 2--31 2--32 2--33 2--34 2--34 2--35 2--35 2--36
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2--0.1
bucky DIAGNOST TS INH
1999
(99.0)
Philips Medical Systems ALL RIGHTS RESERVED
2--0.1
27
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Philips
INSTALLATION
Safety regulations Only technically qualified personnel who has received the appropriate instructions and training for radiography equipment in medical applications and in particular for the installation and commissioning of the bucky Diagnost TS are allowed to install, maintain and service this equipment. Therefore, instructions in this chapter are considered as a supplementary guideline for correct installation of the bucky Diagnost TS and are kept very briefly. Furthermore, all personnel involved must be instructed on safety precautions with respect to radiography and the related high voltage devices as well as to all other electrical low-voltage equipment. All relevant standards and regulations must be complied with. The local service organisation is responsible for the safety and training of personnel and prevention of material damage.
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Philips
bucky DIAGNOST TS
bucky DIAGNOST TS bdts_2_990
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
2--1
bucky DIAGNOST TS
1.
Transport data
1.1.
Box dimensions
Weight Gross
870kg
Loading
816kg
Net
750kg 870kg
Dimension (L x W x H) Box
270cm x 97cm x 196cm
Pallet including loading
265cm x 89cm x 190cm
Sub–pallet Sub–pallet
140cm x 89cm 125cm x 89cm
min 170cm width of roadway for gross pallet
1.2.
min 115cm width of roadway for sub-pallet
Roadway
1
min 170cm
1 Roadway for the gross pallet Crossing while moving
min 170cm width of the floor
2 Roadway for the sub--pallets Crossing while moving
min 115cm width of the floor
2 min 115cm
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Philips
INSTALLATION
2--2
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
bucky DIAGNOST TS bdts_2_990
Philips
bucky DIAGNOST TS
2.
INSTALLATION
Unpacking
The following steps should be done by four persons. Unpacking 0.5 h
0.5 h
0.5 h
0.5 h
0.5 h 0.5 h
1. Remove the rails from the pallet. Note For transportation in insufficient rooms only. See 2.1 ”Removal of column base from column”.
Unpacking
The column is supplied with an inserted counterweight. For transport, the counterweight and the transverse arm are secured by a screw. Do not remove these screws before the installation of the tube arm.
0.5 h 0.5 h
1
2 Documents and distance tool
3c
3b 90°
3a
Loosen the transport screws and safety belts on the pallet. 2. Remove the box with documents and distance tool from the column. 3. Set up the column. -- 3a Move the column approximatly 15cm to foot direction of the column. -- 3b Turn the column base about 90°. The column must be ratched in the cam. -- 3c Lift up the column and place it.
Lay corrugated paper or something similar on the floor in the area where the column is to be moved. Apply extreme care when handling the connection field. 3 S Unscrew the colored screws from the wooden carrier on which the column was positioned and remove it.
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
S Dismantle the box, remove the foil.
bucky DIAGNOST TS bdts_2_990
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
2--3
Philips
INSTALLATION
bucky DIAGNOST TS
4. Remove the box including cables from the table.
The pallet is deviable
5. Remove the optional tomo devices from the pallet.
6
4
6. Remove the tube assembly from the pallet.
7
5
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Remove the wooden items by removing the colored screws. 7
4
Note The gross pallet can be splitted in two sub-pallets.
Cables
7. Remove the table from the pallet. Note Use the shoulder straps. Do not lift the table by means of the table top frame.
8
Table top delivered in the pallet bottom
8. After unscrewing the sub-pallets from the pallet you get access to the table top. Remove the table top. S Dispose the package locally.
2.1.
Removal of column base from column
This section is needed if the column is too large to enter the X-ray room. Tools required Groofkey for column base
PMS Code No. 4512 590 30362
Spring scale 200 N Imbus Key 5mm Molykote DX
PMS Code No. 4512 590 30321
Isotrophy alcohol 1. Dismantle base and column 1.1. Measure the breakaway force with a spring scale when the column is in the the lock position. It should measure 110--150 N. Place the hook of the spring scale on the screw head.
Philips
1.2. Repeat point 1.1, only column must be in free position. Should measure £ 60 N.
P1 2--4
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
bucky DIAGNOST TS bdts_2_990
Philips
bucky DIAGNOST TS
INSTALLATION
1.3. Release the pressure on the Beta grid (measure the distance “X” between nut and screwhead. Dimensions needed by the adjustment of reassembling).
P2
1.4. Remove the 2 Imbus (M5) screws and plate (see P 3). 1.5. Loosen the safety screws (pos. 1) on the groove nut (see P6). 1.6. Remove groove nut and steelring.
M5
1.7. Carefully remove column base from column: S Remove the teflon ring. S Clean it and grease it with Molykote DX.
P3
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
X
bucky DIAGNOST TS bdts_2_990
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
2--5
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Philips
INSTALLATION
bucky DIAGNOST TS
1.8. Clean running surface and thread with Isopropyl alcohol (see P4). For transport cover running surface and thread.
3
Note The stop pin pos. 3 can be removed in case of the movement of the column is impaired. (This stop pin will not be needed any longer.)
2
P4
2. Assembling of column base and column 2.1. Grease the running surface of the column and column with Molykote (PMS No. 4512 590 30321). 2.2. Insert the teflon ring. 2.3. Assemble the column base onto the column. (The column must be laid down when assembling the column base. Take extreme care of thread.) 2.4. Assemble the steelring and groove nut. Hand-tighten the groove nut. P5 2.5. Rotate the column base for functional check. 2.6. Tighten groove nut until the inner steelring (spring) is compressed. Tighten safety screws pos. 1 (see P 6).
1
2.7. Adjust the beta grid (see P 2 and step 1.3). 2.8. Repeat step 1.1 and step 1.2. The breakaway force should measure 110--150 Nm and the free running force should measure ≤ 60N. Adjust the force by loosening, resp. tightening the groove nut. P6
2--6
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
bucky DIAGNOST TS bdts_2_990
Philips
bucky DIAGNOST TS
3.
INSTALLATION
Mounting the rails Before starting, refer to your planning room layout! (see PRB.)
S Depending on the installation of the bucky wall stand (right or left), place the extension rail right or lefthand side to the main rail (see Z--1.1, Z--1.2 and Z--5).
position of extension rail for bucky wall stand left
main rail
S Insert the screw (supplied with system) with a washer and tighten with a spanner. position of extension rail for bucky wall stand righ t
pencil
Tomo option only 2
S Loosen the 4 screws of the tomo switch rail (1) and the end switch cams (2) and remove them.
1
S Place the main rail parallel to the wall.
extension rail
S Mark the 5 fixing points on the floor.
min 314 mm
S Remove the rail. S Drill the 8 mm holes approximately 75 mm deep in the middle of the marked area. S Insert the 5 anchor dowels (supplied with system) into the holes.
longitudinal axis table
table position for bucky wall stand left
S Move the rail back to the installation position. Level the rail using a spirit level. Use the spacer if necessary. S Insert the screws (supplied with system) with the washers and tighten with a spanner.
extension rail
min. 314 mm
longitudinal axis table
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
S Assemble main rail.
bucky DIAGNOST TS bdts_2_990
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
table position for bucky wall stand right
2--7
Philips
INSTALLATION
bucky DIAGNOST TS
S Place the short rail parallel to the long rail. S Insert the tool (1 Rail distance tool) (supplied with system) to adjust the distance of the rails. S Measure the distance of the rails using a rule.
position for bucky wall stand left
1
extension rail
S Mark the 2 fixing points on the floor.
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
S Remove the rail.
long
S Drill the 8 mm holes approximately 75 mm deep in the middle of the marked area.
short
short
716 mm
S Insert the 2 anchor dowels (supplied with system) into the holes. S Move the rail back to the installation position. S Level the rail using a spirit level. Use the spacer if necessary.
position for bucky wall stand right
S Insert the screw only with washer (supplied with system) to the opposite side where the column is to be inserted. S Tighten with a spanner.
1
extension rail
long 716 mm
short
short
Coarse Adjustment Tomo option only S Align front and back rails (2, 3) to same level using the spirit level. Use additional spacers if necessary. For a good performance with the tomo option it is essential to align front and back rail very precisely. An alignment test will be made after installation of the column.
2
3
2
S Install the tomo switch rail (1). S Insert both end switch cams (2). S Insert screws with washers and tighten with Allan key.
Philips
1
2--8
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
bucky DIAGNOST TS bdts_2_990
3.1.
INSTALLATION
Insertion of column to long rail
S Remove buffer (1) from long rail. S Insert the column into the rail system (2). Apply extreme care when handling the ball bearings and be careful not to damage the microswitches and optical sensor B1 positioned in area 3. S Install buffer (1) to the rail system. S Check the leveling of the column in both directions using a spirit level.
1
S Use the spacers if necessary. S Check whether the column moves easily and smoothly. S Insert the second screw with washers into the short rail.
2
3
S Tighten with a spanner. S Mount the longitudinal center switch UB S17 (pos.3) standard, or position control switch assembly for tomo.
3.2.
2
Installation of tube arm
S Remove transport screw (2) from the counterweight on the front of the column.
1
Loosen transport screw (1) only and move the complete block downwards.
S Pull down the transverse arm and secure the counterweight in its maximum top position by inserting transport screw (2). 2
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Philips
bucky DIAGNOST TS
bucky DIAGNOST TS bdts_2_990
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
2--9
Philips
INSTALLATION
bucky DIAGNOST TS
S Route the cables from the top of the column to the left and right of the column (1). S Open top cover (2) of the transversal tube arm by removing the 4 screws 1 (Allan key, 2,5 mm).
1 1 1
S Place the pallet with transverse arm, including tube housing and collimator, next to the column.
2 1
1
S Lift up the arm and insert it into column (3).
2
3
S Insert stop screw (1). S Remove the transport screw. The tube arm must be easily movable in Z-direction. S Connect the plug UFX38 (2).
1 2
5
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Do not remove the screws (2) on top of the cover. These screws are fixing the brake rail!
2--10
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
bucky DIAGNOST TS bdts_2_990
INSTALLATION
S Fasten the cable (4) by using a Ty--Rap.
5
S Connect UFX38 (3) 4
S Install the cover of transverse arm (3).
3
S Install the front cover of transverse arm (5) (see previous page).
6
S Install the cable support into the top of the column.
S Install the wall junction box.
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Philips
bucky DIAGNOST TS
bucky DIAGNOST TS bdts_2_990
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
2--11
Philips
INSTALLATION
bucky DIAGNOST TS
S Route the cables on the column as shown in the figure 1. S Replace top cover of the transversal tube arm. S Connect X20 / X22 on stand control UB1 for standard version.
UB1
Figure 1
** All three items, wall box, column cable support und tube arm, should be located as shown in figure 2 (design and function, especially cable lenghts).
*1 = 0.6 meter
*1 Location for ceiling suspension junction box. This box is not included in the system. Please, order this item locally as an option. (PMS Artikel No. 9848 500 26201 ceiling suspension box for CS, this item is not included and must be ordered optional).
**
Note Max. room height with ceiling suspension box = 3.0 m
Figure 2
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
S Connect X20 / X21 / 22 / 23 / 24 on stand control UB1 for tomo version.
2--12
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
bucky DIAGNOST TS bdts_2_990
3.3.
INSTALLATION
Final alignment of the column (tomo only)
S Move the column to the left end position. S Check the alignment of the rails by moving the column smoothly to the other end. The column must remain steadily upright over the whole length. If necessary, adjust the level of the front rail with spacers as described in Section “Mounting the rails”.
(6) (5)
The alignment can be done either with the spirit level or with the following method:
Final adjustment S Apply a vertical line to the wall on the right--hand side, e.g. with a colored adhesive tape, Check the vertical alignment while moving the column along the rail. If corrections should be necessary: S Add adjustment plates (0) under back (1) or front rail (2), or, for vertical correction (5), S adjust the ball bearing (3). Three different holes are available. For easy access move the column to the left or right end position. If the column is not vertical (6), the correction can be made with an eccentric ball bearing (4). (1)
(4)
(3) (0)
(2)
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Philips
bucky DIAGNOST TS
bucky DIAGNOST TS bdts_2_990
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
2--13
Philips
INSTALLATION
3.4.
bucky DIAGNOST TS
Installation of bucky table
Note Four (4) persons are necessary to lift the table. The bucky table is supplied as a complete unit.
1
S Remove the 4 transport screws and push them down.
Do not lift the table by means of the tabletop frame structure.
1
S Insert the shoulder straps into the loops (1). S Lift up the table with the shoulder straps (1). S Remove the pallet and lower the table carefully. S Place the table with rear end towards the short rail. See room layout plan. S Insert screws (1) into the short rail and fit the table to the rail. S Drill two 12mm holes approximately 75mm deep for screws (2). S Insert the 2 anchor dowels (supplied with system) into the holes.
1 2
S Use the spacer if necessary. S Insert screws (2) supplied with system and washers and tighten with a spanner.
table position for bucky wall stand right
S Level the table in both directions using a spirit level. S Use the spacer if necessary. S Insert screws (2) supplied with system and washers and tighten with a spanner.
1 2
S Install the foot-switch assembly.
table position for bucky wall stand left
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
S Remove the wooden blocks from the longitudinal side of the pallet and place them at the rear of the pallet.
2--14
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
bucky DIAGNOST TS bdts_2_990
Philips
bucky DIAGNOST TS
INSTALLATION
S Remove the red screws (2) of the transport securing device of the tabletop carrier on both sides. S Remove the tie-wraps form the Bucky device (1)
2
1
i
Figure 1
S Remove the cassette tray from the bucky.
1
S Move the tabletop to its maximum front position (if it is already mounted).
3
S Loosen 2 screws at the rear (1). S Tilt the tabletop unit gently as shown in figure 2.
4
S Secure the tabletop by fixing the lever (2) via the screw (3).
2
S Remove the tie-wraps (4) from the arm braces on both sides (elevation table only). Figure 2
Always remove cassette tray before opening the tabletop. Tilting the tabletop with inserted tray will cause damages to both table and cassette tray. Tilting the table top beyond the upright position can cause damages to the cover of the table base.
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
2
bucky DIAGNOST TS bdts_2_990
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
2--15
Philips
INSTALLATION
3.5.
bucky DIAGNOST TS
Installation of Tomo pivot tower (tomo only)
S Thread cable W12 (laser cable) through the center holes of table base and pivot tower base. S Insert 4 screws together with washers and tooth lock washer (supplied with system) from the table inside and mount the tomo pivot tower. S Close the tabletop. S Secure tabletop with the 2 screws at the rear. S Reinstall the cassette tray.
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
S Position tomo pivot tower towards the corresponding mounting holes at the rear side of the table.
2--16
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
bucky DIAGNOST TS bdts_2_990
Philips
bucky DIAGNOST TS
3.6.
INSTALLATION
Installation of bucky-device extension (tomo only)
Do not support or level the free hanging end of the bucky-device extension while tightening the screws. It will self-align to the film level in a slightly tilted position when the screws are installed. S Tighten the screws.
3.7.
Installation of position control switches (tomo only)
Pull corner iron with position control switches out of the column base (they are parked in the column base for transportation)
2
3 1
4
S Insert screws with washers (supplied with system) into the column base and fit the position control switches as shown in (1). S Tighten screws with Allan key.
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
S Mount the bucky-device extension and insert the 4 screws with washers and tooth lock washers (supplied with system).
bucky DIAGNOST TS bdts_2_990
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
2--17
INSTALLATION
Philips
Move column with extreme care in order not to damage or shear off the microswitches and the magnetic sensor while passing the cams of the tomo switch rail. S Adjust microswitches for center position and safety end: -- The microswitches (2) (center position) and (3) (safety limit switch) must trigger when passing their respective cams, as shown in the figure below. S Adjust the proper alignment of the magnetic measuring sensor: Slowly move the column along the tomo switch rail: -- the magnetic sensor B1 (1) (angle position) must not touch any of the cams. The distance between magnetic sensor (1) and cam should be 1mm. The functional test will be performed after power up (see Section “Checking drive and position control switches”). The detection of any cam position regarding 40° / 20° / 8° is indicated with the red LED located as shown in the figure.
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
bucky DIAGNOST TS
2--18
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
bucky DIAGNOST TS bdts_2_990
Philips
bucky DIAGNOST TS
3.8.
INSTALLATION
Installation of OPTIMUS The following items are different from the OPTIMUS generator: - Firmware TDL file - PCB low speed rotor control
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
For further information contact the Service Innovation Department in Hamburg. S Place the X-ray generator cabinet in the position shown in the room layout. S Place the keyboard in the position shown in the room layout. For detailed information see manual “Installation of OPTIMUS generator”. S Check cable connectors on all PCBs (Column Controller UB1, Table Controller SZ1). S Route the cables from the wall connection box of the column to the X-ray generator cabinet. S Route the cables from the bucky table to the X-ray generator cabinet. Location at generator site HT cable Stator cable Exposure release Amplimat cable Tomo release Thermo cable Earth Connection Power suply 230V
1 WAX 1 EZX 22 Table EZX 31 BWS 1 WAX 11 EZX 3
3.8.1. Installation of bucky DIAGNOST TS with tomo and Medio CP/CP-H For bucky TS with tomo and Medio CP/CP-H, an Interface board SZ3 is needed. The SZ3 board must be ordered separately (= option): Code No.: 4512 590 30351
Bucky TS & Tomo
Medio CP Exposure rel. X1
X2
SZ3
X1
SZ1
Philips
see Z1-4--2
bucky DIAGNOST TS bdts_2_990
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
2--19
Philips
INSTALLATION
bucky DIAGNOST TS
3.8.2. Programming of APR data The generator is pre-programmed at factory site. For RGDV programming see 2Z--2.0. If special APR assignements are needed, please, proceed as follows:
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
S Optimus / Program / Human interface / Predefined assignments -- without vario focus: AR 46 ENG.TDL AR 46 DEU.TDL AR 46 FRA.TDL AR 46 ESP.TDL
= = = =
English German French Spanish
-- with vario focus for RO 1750: AR 46 V ENG.TDL = English AR 46 V DEU.TDL = German AR 46 V FRA.TDL = French AR 46 V ESP.TDL = Spanish -- and select for Bucky TS RGDV 1 “Bucky GR” RGDV 2 “TOMO BTC” or “TOMO BTS” RGDV 3 “Wall Std GR” RGDV 4 “Free GR”
3.8.3. Programming tube select at control handle S The tube select signal, e.g. for tube one is selected, is coming from generator as shown in diagram Z1--4 with signal RGDV via release decade 1WAX1.
1WA
-- Exposure release for RGDV 2 is also coming over the same 1WAX1 decade. -- BWS is programmed via 1WAX3 for exposure release.
5
-- Free cassette is programmed via 1WAX4. S Modification of tube one select for RGDV 3/4: Usually there is no tube select indication when RGDV 3/4 are selected. If the tube select indication at the control handle is needed, link X1:5 to X3:5 / X4:5 on 1WA.
X1
X2
X3
X4
Philips
In any case the tube 1 select will be present.
2--20
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
bucky DIAGNOST TS bdts_2_990
Philips
bucky DIAGNOST TS
3.9.
INSTALLATION
Installation of tabletop
In rooms with sufficient space on the longitudinal side of the table:
1
S Install the tabletop from the longitudinal side.
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
S Remove end cover (1) and stoppers from one side of the tabletop. S Push the tabletop along the transverse carriage.
Figure 1 Apply extreme care when handling the ball bearings.
S Test the easy and smooth movement of the tabletop, if not correct adjust the eccentric bearing bolts. S Install the stoppers and end cover (1), see figure 1.
3.9.1. Installation of tabletop in small rooms In rooms without sufficient space on the longitudinal side, the tabletop base unit should be rotated to install the tabletop. S Loosen screws (1).
1
S Remove the locking rings from bolts (2). S Remove bolts (2). S Rotate the tabletop base unit until the tabletop can be inserted. S Remove end cover (1) and stoppers from one side of the tabletop. S Insert the tabletop along the transverse carriage. S Rotate the tabletop unit as required.
2
Philips
S Install bolts (2) and secure via locking rings. S Install the stoppers and end cover (1). S Fit screws (1).
bucky DIAGNOST TS bdts_2_990
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
2--21
bucky DIAGNOST TS
3.10. Setting to work The system is supplied fully prefabricated by the factory. Before switching on the X-ray generator: Switch off the auxiliary mains breaker for table. For detailed information see manual installation of the OPTIMUS X-ray generator. S Check the power supply for the table on the PCB SN. Always remove cassette tray before you open the tabletop. Tilting the tabletop with inserted tray will cause damages to both table and cassette tray.
1 3
2
Tilting the tabletop beyond the upright position can cause damages to the cover of the table base. Figure 1 For access to the table control unit: S Remove the cassette tray from the bucky. S Move the tabletop to its maximum front position. S Loosen 2 screws at the rear (1). S Tilt the tabletop unit as shown in figure 1. S Secure the tabletop by fixing the lever (2) via the screw (3). The table power supply must be 230V, 50Hz or 60Hz. Note There is no programming for 50 or 60Hz. The table can only operate with 230V (and not with 110V).
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Philips
INSTALLATION
2--22
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
bucky DIAGNOST TS bdts_2_990
Philips
bucky DIAGNOST TS
INSTALLATION
S Follow the setting-to-work procedure described in the manual OPTIMUS X-ray generator XRG 90.
Programming
1
S Select Registration Devices
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
-- RGDV1 ( = registration device of the place for tables 1) -- RGDV2 ( = registration device free exposure) -- RGDV3 ( = registration device wall stand) -- RGDV4 ( = registration device free exposure)
After finishing the described setting-to-work procedure of the OPTIMUS X-ray generator, check the functions below on the table. After switching on the auxiliary mains breaker for table S Check all functions on the table: -- Green LED on the power pack is on. -- The programmed language. -- Function of foot-switch assembly. -- Function of the rocker switches “table up/down” on both sides. -- Function of end-switch upper position: Motor must stop before mechanical end position is reached. -- Function of end-switch lower position: Motor must stop before mechanical end position is reached. -- Function of safety switch: Apply about 20 kg of manpower against the tabletop during table down movement. S Check the strength of upward table movement. S Check whether the range of movement is 270 mm. S Check the movement of the bucky unit. S Check the functions of the column: -- Function of catches in all directions.
Philips
-- Function of light on the collimator.
bucky DIAGNOST TS bdts_2_990
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
2--23
Philips
INSTALLATION
bucky DIAGNOST TS
S Check the functions of the bucky unit: -- bucky exposure delay S Check delay time of grid release with test exposure time = 4 ... 6 msec. There must not be visible any grid line.
The X-ray tube has been adjusted at factory site. Do not ADAPT the X-ray tube. In case the generator was not powered on during a longer period -- e.g. more than 2 months --, we recommend to recondition the X-ray tube. See OPTIMUS manual, section Installation, 10.1 Tube conditioning. S Check coincidence of X-ray beam and light. -- Make a test exposure. S Check alignment of the stand. S Check alignment of the bucky wall stand.
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Note:
2--24
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
bucky DIAGNOST TS bdts_2_990
Philips
bucky DIAGNOST TS
INSTALLATION
3.11. Adjustment procedure for tomo 3.11.1. Adjustment of the table level S Check whether the table-top is precisely levelled. S Turn on light and laser beam.
-- Raise the table to a position somewhat below maximum level. The distance from the laser beam to front and to rear frame of the table-top must be equal, i.e. the laser beam must be plane-parallel to the table-top. -- If not, turn laser diode slightly with adjusting screw (1). S Raise table-top precisely to laser beam level according to figure 1. Table-top position must be slightly below the maximum level that can be reached. If not, adjust laser diode height (2). If not correct:
Figure 1
S Open and secure the table-top. Set the upper lift limit SNV--S7 switch so that the center of the pivot point meets the 0-level of table-top.
3.11.2. Correction for tomo layer height = fine adjustment under X-ray condition If a tomo test exposer (with a test phantom) deviates from the tomo height, the correction can be done as follows: Solution 1 -- valid for table serial No. =500 If the 0mm layer height = the top of table-top differences to the 0 point of the coupling element. The correction can be done on the axis in both directions. Solution 2 -- valid for table serial No. =499 Correct the tomo layer height = 0 mm via the table height SNVS7 switch = table height for the most upper table position. After the SNVS7 correction follow up with section 2.13.2 Adjustment of tomo layer height.
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
S Adjust laser beam levelling:
bucky DIAGNOST TS bdts_2_990
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
2--25
Philips
INSTALLATION
bucky DIAGNOST TS
3.11.3. Adjustment of tomo layer height indication S Set the service switch (1) on the column control board to service. The following message will appear on the control panel display: Service mode
1 S Press the key on the control panel (1). This will start the calibration menu. Now follow the instructions on the display: Service mode Move to h=0
Figure 2 S Raise the table-top up to laser beam level (corresponding to the minimum objective plane level 0mm).
3
S Press the key on the control panel (2). The position is now stored in the PLC and the display shows: Service mode
4
T
Move to h=200
T
1 2
S Lower the table-top to 200 mm, use scale meter (corresponding to the maximum objective plane level 200 mm). S Press the key on the control panel (2). The position is now stored in the PLC and the display shows: Service mode Diagnosis: 0
Figure 3
You can always quit the calibration menu without saving the values by pressing the key again.
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Diagnosis: 0
2--26
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
bucky DIAGNOST TS bdts_2_990
Philips
bucky DIAGNOST TS
INSTALLATION
S Leave the service switch on the column control board in position service to enter the display language.
3.11.4. Setting the display language
3
4
Service mode Diagnosis: 0 S Press the key on the control panel (3). The following message will appear on the control panel display:
T
T
1
Service mode /V _._ Language: english
2
( V_._ means Softfware--version ) S Press the key (4) to select another laguage. Repeat until the message appears in the desired language. S Press the key on the control panel (3) again to quit the menu.
Figure 4
S Set the service switch on the column control board back to operation.
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
S The service switch should be in mode service.
bucky DIAGNOST TS bdts_2_990
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
2--27
Philips
INSTALLATION
bucky DIAGNOST TS
3.11.5. Adjustment of the tomo axis S Prerequisite: Alignment of standard equipment has been completed. S Select tomo on control handle.
S Move transverse arm to normal position.
4
3
S Insert the coupling bar into central and lower swivels (1, 2). S Fit the coupling bar to the column (3). The locking bolt (4) will be pushed upwards (which will unlock the ß-movement axis of the transverse arm).
2 1
1. Check that the tomo bar activates the tomo sensor switch UB S15 when inserted. 2. Adjustment of ball bearing to tomo bar: Ball bearings must be engaged in direct contact to the tomo coupling bar. So no mechanical clearance is visible.
B
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
S Move column, table-top and bucky tray to the center position (the center position is reached if the 4 LED-indicators on the control handle are lit).
2--28
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
bucky DIAGNOST TS bdts_2_990
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Philips
bucky DIAGNOST TS
INSTALLATION
S Bucky device, light beam localiser and coupling bar must be aligned in a vertical plane. S If necessary, align the system by adjusting central swivel (1), lower swivel (2) and position control bar.
2 1
3 3.11.6. Alignment for center line = center of bucky If the center line 3 is not in center of bucky proceed as follows. S The tomo coupling bar is y centered vertical (4). S Open screws (5) on tomo coupling bar and rotate/twist the tube support arm. S The center line will still be in center of bucky tray (as shown in pos. 3).
3.11.7. Checking drive and position control switches For parameter setting and control please refer to the operating manual. Procedure: S Move the column carefully from its center position towards the left and right end positions and check the following items: -- the magnetic switch does not touch the position cams.
Philips
5
UB S17 S20
-- the micro switches S17/S20 are not damaged or sheared off when passing a cam. -- the magnetic switch is triggered at each angle position cam and the end position cams. This can be checked by observing the red LED on the magnetic switch B1. bucky DIAGNOST TS bdts_2_990
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
2--29
bucky DIAGNOST TS 1
S Test the safety limit switches: Move the column manually to the left and right limit switch positions. The limit switch must be triggered (I/O--LED 35 on tomo controller (PLC) lits) and the system be switched off. S Initiate a tomo test cycle on the control panel: Select mode with 40°/4s. Press the key until the cycle is completed and check the following items: -- the column moves first to the left (1). -- the column moves at constant speed and without jerks (otherwise check: cable routing, rail alignment, dirt on rail surface, spring pressure, driving axle). -- the eccentric disk operates correctly (otherwise check coupling motor (M 3) and position switch UB S16 or both springs are ”fixed too tight”. Therefore, check the distance from screw head to the nut as shown in the figure. The distance as shown must be 15 ± 0.5mm.
Coupling motor
-- the driving wheels grip firmly (otherwise increase spring pressure).
M3
Tomo motor S16
15 ± 0,5mm
M2
Pivot point
-- the column reverses automatically at the end position on each side. -- the bucky device moves smoothly and without jerks (otherwise check alignment of the coupling bar). -- the position control switches (magnetic switch and microswitches) operate correctly, see first step. -- the position of the coupling bar is vertical after completion of a cycle. -- the distance between control switch B1 and tomo position cams should be 1mm. -- If the sequence is aborted, check the function of the position control switches (see first step). -- If the sequence is aborted due to timeout (error code 30, see Chapter ”trouble shooting”), increase the motor supply voltage by switching to the next higher transformer tap (see schematic diagram Z1--2). S Repeat the tomo test cycle with 40°, 20° and 8° sweeps, with high and low speed settings. Again, check whether any anomalities occur during the test.
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Philips
INSTALLATION
2--30
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
bucky DIAGNOST TS bdts_2_990
Philips
bucky DIAGNOST TS
INSTALLATION
3.12. Tomo adjustment under exposure conditions (tomo only) 3.12.1. Measurement of X-ray zone
a
Two exposures are to be taken on the same film.
1
b
First exposure = Bucky -- Small focus
2
-- SID
= tomo SID
-- X-ray area
= 200 mm x 200 mm
-- Exposure values
=
40 kV / 6.3 mAs
Second exposure = Tomo S Set the parameters for exposure 2 (with tomo): -- Small focus -- SID
= tomo SID
-- X-ray area
= 150 mm x 200 mm
-- Tomo height
= 130mm
-- Exposure values
= 40° / 50 kV / 12.5 mA
-- Tomo time
= 2s
S Take the exposure in tomo mode (tomo on) and determine the following values: -- a =
_____mm
-- b =
_____mm
S Check it these values meet the following conditions: -- D a, b =
£ 10 mm
-- S a, b =
£ 80 mm
S If the above conditions are not fulfilled, check whether the free movement of the a--drive and bucky module is impaired. S If under-voltage is indicated on the generator control board, the motor will supply voltage by switching to the next higher transformer tap (see schematic diagram Z1--2).
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
S Set the parameters for exposure 1 (without tomo):
bucky DIAGNOST TS bdts_2_990
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
2--31
Philips
INSTALLATION
bucky DIAGNOST TS
3.12.2. Symmetry of tomo axis S Position the lead plate - 100mm above the X-Ray cassette.
Lead plate
The X-ray cassette ist positioned on the table-top. S Collimate the X-ray field (1) as far as the ”hole” will be exposed during the whole tomo sweep.
100mm
S Take an exposure in center position (bucky mode) = 70 kV / 10 mAs S Take an exposure in tomo mode. 70 kV 2 sec 5 mA 40° .
X-ray cassette
Table top
The result should look like as shown in the figure: -- c » d, i.e. the generator should switch off shortly before the system. -- No undervoltage indication should appear on the generator.
c
d 3
4
S Check whether these values meet the following conditions: -- |c|--|d| < 5mm. S If the result is not symmetrical, proceed as follows: -- In case of overvoltage lower the tomo speed by selecting a higher resistance with jumpers on board SZ2 = fine adjustment, i.e. reduce the supply voltage of the drive or reduce the voltage at transformer T2 side for slow speed = coarse adj. 17.5V / 15.5V / 13.5V ore available. For high speed progr. 31V / 33V or 37V. -- In case of unsymmetry (c < d) increase the tomo speed by selecting a lower resistance with jumpers on board SZ2., i.e raise the supply voltage of the drive
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
¯∅
1
2--32
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
bucky DIAGNOST TS bdts_2_990
Philips
bucky DIAGNOST TS S Programming of symmetriy (please fill out the table) -- The following programming is selected: U On T2 High speed
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
INSTALLATION
Low speed
value for c and d in mm:
R=O ON SZ2
37V
O
33V
O
31V
O
c = .....................mm ................ O
d =.......................mm
17.5V O 15.5V O
c = .....................mm ................ O
d =.......................mm
13.5V O Remarks: Þ
= coarse adj.
= fine adj.
tested with 40° 40°
4 sec and 2 sec
S If necessary, readjust the tomo axis (see Section 3.13.4 Adjusting the tomo axis). S Repeat the above test.
+/-- 5 mm
3.12.3. Tomography resolution test S Take an exposure with the TOMO test phantom with following parameter: TOMO angle
TOMO time
TOMO height
40°
4 sec
50 mm
40°
2 sec
50 mm
40°
4 sec
100 mm
40°
2 sec
100 mm
lp/mm
with
small
Philips
-- Minimum resolution = 2.32 focus.
Actual value Resolution lp/mm
bucky DIAGNOST TS bdts_2_990
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
2--33
bucky DIAGNOST TS
3.12.4. Tomo layer height (Use TOMO test phantom) S Take the following test exposures with inclined bucky tray. Tomo angle
Tomo time
TOMO layer height -default = nominal value
Measured TOMO layer height = actual value
Deviation (+/-- 3mm max.) = nom.--actual value
40°
4 seconds
45 mm
mm
mm
40°
4 seconds
130 mm
mm
mm
40°
4 seconds
200 mm
mm
mm
If the deviation is more than +/-- 3mm, please repeate the table calibration procedure as follows: 1) Raise the table hight to the effective (phantom measured) zero-level. In the display you will see the amont of the deviation. 2) Follow the instruction part 3.13.2 Calibration of the level sensor. Note that your table stays on level zero, do not move the table to zero as requested in the display instruction. 3) Adjust the laser beam.
3.13. Final work S Install all rail covers.
1
S Connect the earth wiring to all covers.
1
S Install all column covers. S Close the X-ray generator cabinet. S Clean all covers with a mild cleaner e.g. soapsuds. S Check the labelling (See drawing Z--3).
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Philips
INSTALLATION
2--34
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
bucky DIAGNOST TS bdts_2_990
Philips
bucky DIAGNOST TS
INSTALLATION
3.14. Acceptance test See national regulations as e.g. Röv / DOD / VA
3.15. Handing over to customer on
( date)
...................
to
( customer)
................................................
by
(system engineer)
................................................
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
The modified system is handed over
bucky DIAGNOST TS bdts_2_990
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
2--35
bucky DIAGNOST TS
3.16. Installation for tomo bar parking frame Tomo bar
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Philips
INSTALLATION
Wall mounting
65 cm ± 2 cm
Philips
10 cm ± 2 cm
Floor 2--36
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
bucky DIAGNOST TS bdts_2_990
Desk :
OPTIMUS OPTIMUS bucky DIAGNOST R CD TS E1999 Philips Medical Systems
ALL RIGHTS RESERVED
(99.0)
] RGDV2 [ ] RGDV3 [ ] ] ] RGDV6 [ ] RGDV7 [ ] ]
For WBX11 : 9 -- 10 (ready) und 1 -- 2 (format size correction contact)
Bucky RGDV :
RGDV1 [ RGDV4 [ RGDV5 [ RGDV8 [
-- Disable time override: -- Tube power factor: -- kV steps: -- mAs steps: -- mA steps: -- time steps: -- Density steps: -- Density correction (6% steps): -- Underexposure display: -- Tube overload protection: Bucky / Scopo 1WB / Decade Bucky 1 (WBX11) :
-- Used for tomo:
Data Set B:
Room: Tube: Release circuit number: Enable handswitch at generator desk: Syncmaster present: Exposure switch type: Bucky format density correction: Cone density correction: Dose measurement input: Dose measurement sensor: Tomo movement (No break after exposure end): Release delay: Mounted radiographical controller: Release circuit adaption unit: Mounted tomo extension: Medium II format kV correction (dose equiv. steps): Medium II format density correction (6% steps): Medium II format mAs correction (6% steps): Small II format kV correction (dose equiv. steps): Small II format density correction (6% steps): Small II format mAs correction (6% steps):
Data Set A :
----------------------
Name :
-
bdts_2z2_0_990
NONE NONE
Bucky RGDV Bucky RGDV
no 100 single 25% 25% 25% 12% 0 yes on WAX12
no
no enable none none none 0 0 0 0 0 0
1 1 1 yes no double step 0 0 none
RGDV4
RGDV6
Tomo time 4 :
Tomo time 3 :
Tomo time 2 :
Tomo time 1 :
RGDV7
Tomo time 8 :
Tomo time 7 :
Tomo time 6 :
Tomo time 5 :
Bucky / Tomo 1WA \ Tomo time [ s ] :
RGDV5
Philips
Tomo RGDV -- switch related RGDV2 Time setting for input at WA X21:1 ......8 Tomo mode switch : X11:3 SL_XG_TO / Bucky RGDV : X11:1 Format + :10 Bucky ready / Tomo RGDV : X11:1 Format + :5 Tomo ready switch related X11:3 -- --> Bucky -- Tomo remote switchover RGDVs
RGDV1
enable
Tomo mode switch Bucky RGDV -- switch related
no 100 single 25% 25% 25% 12% 0 yes on WAX11
no
no enable none 1WA none 0 0 0 0 0 0
1 1 1 yes yes double step 0 0 EZX31
RGDV3
no no 100 100 single single 25% 25% 25% 25% 25% 25% 12% 12% 0 0 yes no on on Bucky / Tomo 1WA : Decade Bucky 1/2
yes
yes enable none 1 WA 1 WA 0 0 0 0 0 0
no enable none 1 WA none 0 0 0 0 0 0 no
1 1 1 yes yes double step 0 0 none
RGDV2
1 1 1 yes yes double step 0 0 EZX21
RGDV1
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
A4 99--12--21 Wa/Re
Philips
RGDV programming OPTIMUS Rel. 3.x
2Z--2.0
RGDV8
1
17a, b, c, d e or
Philips Medical Systems
PHILIPS
PHILIPS
type 9848 600 02641 s/n 99.00.123 TS Family PEI
12a b
LR 37323
US
Philips Medical Systems
230V
50/60Hz
5A
CISPR11-Classification:
Group 1 - Class B
Development and Manufacturing Centre
Röntgenstrasse 24
D-22335 Hamburg/Germany
D-22335 Hamburg/Germany
or
5
type 9848 600 0023. s/n 99.00.123 tube stand
PHILIPS
PHILIPS
INHER. FILTRATION 0,3 Al / 80 IEC 522/1976
Made by Ralco
®
D-22335 Hamburg/Germany
PHILIPS
PHILIPS
Development and Manufacturing Centre
BEAM LIMITING DEVICE type 9848 600 0209. s/n 99.00.123 manual collimator TS (HHS)
C
®
0051
Philips Medical Systems Röntgenstrasse 24 D-22335 Hamburg/Germany
Made by Ralco
INHER. FILTRATION 0,3 Al / 80 IEC 522/1976
CERTIFICATION This Product complies with the Performance Standard under the Radiation Control for Health and Safety Act of 1968, applicable at date of manufacture. Manufactured: September 1992
20
17a
0051
17b 4 12a
5
Philips Medical Systems
PHILIPS
type 9890 010 0219. s/n 99.00.123 tabletop getalit PUR TH
Development and Manufacturing Centre
Röntgenstrasse 24 D-22335 Hamburg/Germany
CERTIFICATION This Product complies with the Performance Standard under the Radiation Control for Health and Safety Act of 1968, applicable at date of manufacture. Manufactured: August 1999
17c
6a b (9, 7, 6 c d)
13a b, 20 10 or
11
PHILIPS
or
or Laser CONTROL GRIP!
1 15, 16 2 3
12b
PHILIPS
Philips Medical Systems
PHILIPS
type 9890 010 8137. s/n 99.00.123 tabletop 2m f.b. table TS
Development and Manufacturing Centre
Röntgenstrasse 24 D-22335 Hamburg/Germany
CERTIFICATION This Product complies with the Performance Standard under the Radiation Control for Health and Safety Act of 1968, applicable at date of manufacture. Manufactured: August 1999
17d
or
4
6a
type 9848 600 0109. s/n 99.00.123 control grip TS
6b type 9848 600 0208. s/n 99.00.123 control grip tomo/laser TS
18
19 or
K
ratio
B
(only for USA)
TUBE HOUSING ASSEMBLY
SA LR37323 C
CERTIFICATION This Product complies with the Performance Standard under the Radiation Control for Health and Safety Act of 1968, applicable at date of manufacture. type 9806 300 7120. s/n 12345
150 kV
X-RAY TUBE HOUSING
ROT 350 10
- X-RAY EQUIPMENT CLASSIFIED BY UNW RITERS LABORATORIES INC ® WITH RESPECT TO ELECTRICAL FIRE SHOCK AND MECHANICAL HAZARDS ONLY 641 B
type 9806 300 7120. s/n 12345
150 kV
X-RAY TUBE HOUSING
ROT 350 10
- X-RAY EQUIPMENT CLASSIFIED BY UNW RITERS LABORATORIES INC ® WITH RESPECT TO ELECTRICAL FIRE SHOCK AND MECHANICAL HAZARDS ONLY 641 B
150 kV
X-RAY TUBE HOUSING
ROT 350 10
type 9806 300 7120. s/n 12345
150 kV
X-RAY TUBE HOUSING
ROT 350 10
type 9806 300 7120. s/n 12345
150 kV
X-RAY TUBE HOUSING
ROT 350 10
D-22335 Hamburg/Germany
150 kV
0.6 IEC 336 1.3 IEC 336
SRO 33 100
PHILIPS
2,5 Al IEC 522/1976
PHILIPS Made in Germany
Philips Medical Systems Development and Manufacturing Centre
Röntgenstrasse 24 0123
D-22335 Hamburg/Germany
150 kV
0.6 IEC 336 1.0 IEC 336
SRO 25 50
PHILIPS
2,5 Al IEC 522/1976
PHILIPS Made in Germany
Philips Medical Systems Development and Manufacturing Centre
Röntgenstrasse 24 0123
TUBE HOUSING ASSEMBLY
CERTIFICATION This Product complies with the Performance Standard under the Radiation Control for Health and Safety Act of 1968, applicable at date of manufacture.
PHILIPS Made in Germany
Röntgenstrasse 24
type 9890 000 6419. s/n 12345 X-Ray TUBE
SA LR37323 C
PHILIPS
Development and Manufacturing Centre 0123
TUBE HOUSING ASSEMBLY
CERTIFICATION This Product complies with the Performance Standard under the Radiation Control for Health and Safety Act of 1968, applicable at date of manufacture.
0.6 IEC 336 1.3 IEC 336
RO 17 50
Philips Medical Systems
type 9806 206 1110. s/n 12345 X-Ray TUBE
SA LR37323 C
D-22335 Hamburg/Germany
150 kV
2,5 Al IEC 522/1976
TUBE HOUSING ASSEMBLY
SA LR37323 C
type 9806 300 7120. s/n 12345
- X-RAY EQUIPMENT CLASSIFIED BY UNW RITERS LABORATORIES INC ® WITH RESPECT TO ELECTRICAL FIRE SHOCK AND MECHANICAL HAZARDS ONLY 641 B
Röntgenstrasse 24
type 9806 206 7000. s/n 12345 X-Ray TUBE
CERTIFICATION This Product complies with the Performance Standard under the Radiation Control for Health and Safety Act of 1968, applicable at date of manufacture.
- X-RAY EQUIPMENT CLASSIFIED BY UNW RITERS LABORATORIES INC ® WITH RESPECT TO ELECTRICAL FIRE SHOCK AND MECHANICAL HAZARDS ONLY 641 B
Development and Manufacturing Centre 0123
TUBE HOUSING ASSEMBLY
CERTIFICATION This Product complies with the Performance Standard under the Radiation Control for Health and Safety Act of 1968, applicable at date of manufacture.
Made in Germany Philips Medical Systems
type 9806 200 6000. s/n 12345 X-Ray TUBE
SA LR37323 C
PHILIPS
PHILIPS
2,5 Al IEC 522/1976
D-22335 Hamburg/Germany
150 kV
0.3 IEC 336 1.0 IEC 336
SRO 0951
PHILIPS
2,5 Al IEC 522/1976
PHILIPS Made in Germany
Philips Medical Systems Development and Manufacturing Centre
Röntgenstrasse 24 0123
type 9806 206 7010. s/n 12345 X-Ray TUBE
D-22335 Hamburg/Germany
150 kV
0.6 IEC 336 1.2 IEC 336
SRO 33 100
6d
or
X-RAY GRID STREUSTRAHLENRASTER GRILLE ANTIOFFUSANTE Lp/cm
17e
type 9848 600 0206. s/n 99.00.123 control grip tomo TS
- X-RAY EQUIPMENT CLASSIFIED BY UNW RITERS LABORATORIES INC ® WITH RESPECT TO ELECTRICAL FIRE SHOCK AND MECHANICAL HAZARDS ONLY 641 B
9
Philips Medical Systems Development and Manufacturing Centre
Röntgenstrasse 24 D-22335 Hamburg/Germany CERTIFICATION This product complies
fo(cm)
with the Performance Standard under the Radiation Control for Health and Safety Act of 1968 applicable at date of manufacture.
MANUFACTURED: AUGUST 1999
type 9848 600 0207. s/n 99.00.123 control grip laser TS
6c
CAUTION LASER RADIATION DO NOT STARE INTO BEAM MAX. OUTPUT < 1 mW EMITTED WAVELENGTH 650 nm FDA CLASS II LASER PRODUCT
LASER RADIATION DO NOT STARE INTO BEAM CLASS 2 LASER PRODUCT IEC 825-1: 1993 MAX. OUTPUT < 1mW EMITTED WAVELENGTH 650 nm
7 according to the release status
10
11
Philips Medical Systems Development and Manufacturing Centre
Röntgenstrasse 24
Remove cassette tray and move tabletop in the front position before tilting the table
5 Kg
bucky DIAGNOST TS C Philips Medical Systems
(00.0)
8
D-22335 Hamburg/Germany
CAUTION LASER RADIATION DO NOT STARE INTO BEAM MAX. OUTPUT < 1 mW EMITTED WAVELENGTH 650 nm FDA CLASS II LASER PRODUCT
CERTIFICATION This product complies
LASER RADIATION DO NOT STARE INTO BEAM CLASS 2 LASER PRODUCT IEC 825-1: 1993 MAX. OUTPUT < 1mW EMITTED WAVELENGTH 650 nm
with the Performance Standard under the Radiation Control for Health and Safety Act of 1968 applicable at date of manufacture.
MANUFACTURED: AUGUST 1999
9
September 1999
BEAM LIMITING DEVICE type 9848 600 0017. s/n 99.00.123 manual collimator TS
C
13b
Development and Manufacturing Centre
Röntgenstrasse 24 D-22335 Hamburg/Germany
Development and Manufacturing Centre
Röntgenstrasse 24
D-22335 Hamburg/Germany
Manufactured :
Philips Medical Systems
PHILIPS
PHILIPS
Development and Manufacturing Centre
Röntgenstrasse 24
September 1999
13a
PHILIPS
Philips Medical Systems
Manufactured :
PHILIPS
type 9848 600 0008. s/n 99.00.123 TOMO Equipment
September 1999
4
Development and Manufacturing Centre
Röntgenstrasse 24 D-22335 Hamburg/Germany
Manufactured :
Philips Medical Systems
PHILIPS
PHILIPS
September 1999
14
Philips Medical Systems
A1/A3 00-02-24 Schr.
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Development and Manufacturing Centre
Röntgenstrasse 24
16
type 9848 500 2030. s/n 99.00.123 bucky device left
Philips
PHILIPS
Philips Medical Systems
Manufactured :
19
SA ® C
PHILIPS
September 1999
18
15
PHILIPS
PHILIPS
type 9848 600 0006. s/n 99.00.123 bucky table fix
Manufactured :
Philips
14
3
type 9848 600 0007. s/n 99.00.123 bucky table elevation
0123
Class I - TYPE B IEC 60601-1
8
2
Development and Manufacturing Centre
Röntgenstrasse 24 D-22335 Hamburg/Germany
bucky DIAGNOST TS Labelling
2Z-10
Product
INTRODUCTION AND TECHNICAL DATA Trouble shooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Safety regulations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Software and PLC faults . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Hardware and tomographic image problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Specific instructions for fault elimination . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . EPROM replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Setting EPROM configuration jumper . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Test exposure with inclined bucky tray . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . PLC input / output allocation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Digital inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Digital outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Indicator LED’s on table control board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Standard version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Signal exchange between generator and table control . . . . . . . . . . . . . . . . . . . . . . Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Additional LED’s for Tomo version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Signals from PLC to table control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Signals from table control to PLC and generator . . . . . . . . . . . . . . . . . . . . . . . . . . . Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3--1 3--1 3--2 3--5 3--6 3--6 3--6 3--6 3--7 3--7 3--8 3--9 3--9 3--9 3--9 3--9 3--9 3--9 3--9
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Philips
1. 1.1. 1.2. 1.3. 1.4. 1.4.1. 1.4.2. 1.4.3. 1.5. 1.5.1. 1.5.2. 1.6. 1.6.1. 1.6.1.1. 1.6.1.2. 1.6.2. 1.6.2.1. 1.6.2.2. 1.6.2.3.
bucky DIAGNOST TS bdts_3_990_inh
1999
(99.0)
Philips Medical Systems ALL RIGHTS RESERVED
3--0.1
27
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Philips
1.
Trouble shooting
1.1.
Safety regulations
FAULTFINDING
Only technically qualified personnel who have received the appropriate instruction and training for radiography equipment in medical applications and for the bucky Diagnost TS in particular are allowed to install, maintain and service this equipment. It is presumed that all personnel involved with trouble shooting are familiar with fault finding and repair of standard radiography equipment. Therefore only topics related to the process controller (PLC) and tomo specific problems are treated in this chapter. Furthermore, all personnel involved must be instructed on safety precautions with respect to radiography and the related high voltage devices as well as to all other electric low voltage equipment. All relevant standards and regulations must be complied with. The local service organisation is responsible for the safety and training of personnel and prevention of material damage.
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Philips
bucky DIAGNOST TS
bucky DIAGNOST TS bdts_3_990
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
3--1
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Philips
FAULTFINDING
1.2.
bucky DIAGNOST TS
Software and PLC faults
s In order to display software error messages, set the service switch on the column control board to service. The following message indicating an error code will appear on the control panel display: Service mode Diagnosis: nn s Follow the relevant suggestions in the table below to eliminate the disturbance. . If the system is o.k., he following message will appear: s Clear error message by pressing Service mode Diagnosis: 0 s Contact the Help--Desk of your local service organization if the fault cannot be corrected.
M1
M2
40° 20° 8°
UB1
VA2
Tomo mode
bucky
X2
E
Tomo ready
1WA X11
X30
bucky ready
X2
X11
SZ1
X3
UE2
Control handle
Format contact
B1
tomo tomo motor coupling SPS tomo controller
1
8° 20° 40°
SZ1
Table controller
X2
Table controller X1
1 2 3 4 5 6 7 8 9 10
X1
1
RQ_XG_EX XG_RD_EX RQ_XG_PR PO_26V_RE RG_DV_1 GND
X1 Code
Fault description
Possible cause
Suggested remedy
--
Display not active
-- RS232 serial link to control panel disturbed
Check cable and plugs
--
LED ”supply” (yellow) not on Power supply missing
1 2 3 4 5 6 7 8 9 10
X11
E 1WA
X1
Notes
1. Check if power supply is on 2. Check supply cable and connection
--
LED ”Run” (yellow) not on
-- Application program not loaded -- EPROM missing
1. Check EPROM 2. Download application program 3. see error code 104
no error message
Philips
0
3--2
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
bucky DIAGNOST TS bdts_3_990
Philips
bucky DIAGNOST TS
FAULTFINDING
Code
Fault description
Possible cause
1
Communication timeout: no -- Cable to table conanswer from PIC--procestrol not correctly sors on table controller (deconnnected centralized I/O) -- Interface module on PLC defective
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
-- RS485 bus nor correctly connected -- Jumpers on RS485 interface board not set correctly -- PIC--processors or interface module defective -- EPROM defective
Suggested remedy
Notes
1. Check cable and plugs 2. Check position of jumpers on RS485 interface board (must be set to CLOSED, see wiring diagram Z2--6) 3. PLC--reset: s Turn PLC off s Disconnect power supply cable of PLC (on the bottom, to the left of the 3 LED) s Switch PLC on s Wait a few seconds s Reconnect the supply cable
power
4. Check the hardware and replace if necessary 2
Incorrect answer from PIC-- see code 1 processors on table controller
see code 1
30
Tomo sequence timeout (se- -- Tomo motor blocked quence time limit exceeded) -- Tomo motor is not or deficiently engaged
1. Check tomo drive (motor, driving wheel, eccentric disk, see Section 4.12.4 ”Checking drive and position control switches”)
-- Speed too low
-- Motor supply voltage 2. Check motor supply too low voltage: -- Angle and limit posivoltager range: ....... If too tion control switch low, change tap of supply disturbed transformer -- Center position 3. 1. Check alignment of switch disturbed position control switches (see Section 4.12.4 ”Checking drive and position control switches”) 31
Column did not stop in the -- After running too center position after complehigh tion of a tomo cycle, i.e. col- -- Center position umn overran the center posiswitch disturbed tion
1. Check function of column brake 2. Check center position switch (incl. Cable and connections) and center position cam.
Philips
3. Check column stop control 100
PLC system RAM test failure -- PLC hardware fault
1. Replace PLC
101
PLC RAM test failure
1. Replace PLC
bucky DIAGNOST TS bdts_3_990
-- PLC hardware fault
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
3--3
bucky DIAGNOST TS
Code
Fault description
Possible cause
Suggested remedy
Notes
104
PLC Program EPROM test -- EPROM defective failure -- EPROM not correctly installed
1. Replace EPROM
PLC serial interface test fail- -- Interface module deure fective
1. Replace interface module in PLC
2. Check EPROM position (see Section EPROM replacement) -- PLC configuration incorrect 3. Check EPROM configuration (see Section Setting EPROM configuration jumper)
120
-- PLC hardware fault
2. Replace PLC
140
PLC system EPROM test -- PLC hardware fault failure
1. Replace PLC
252
Error flag is set
-- Error in PLC applica- 1. Check program release tion program (must match PLC version and generator type) 2. Replace EPROM
253
Index register overflow
-- Error in PLC applica- 1. Replace EPROM tion program
254
Program timeout
-- Error in PLC applica- 1. Replace EPROM tion program 2. Replace PLC -- PLC defective
255
Software watchdog error or -- Error in PLC applica- 1. Replace EPROM power supply error tion program 2. Replace PLC -- PLC defective
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Philips
FAULTFINDING
3--4
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
bucky DIAGNOST TS bdts_3_990
1.3.
FAULTFINDING
Hardware and tomographic image problems
Fault description
Possible cause
Suggested remedy
Tomo sequence aborted
-- Tomo speed too low
1. See error code 30 in Section Software and PLC faults
-- see error code 30 in Section Software and PLC faults Underexposure (indication on generator control desk)
Tomo speed too high
Mains voltage too high
Objective plane level not correct
Notes
1. Check supply voltage and change transformer tap, if necessary 1. Install transductor or voltage stabilizer (contact service organization to order)
-- Calibration maladjust- 1. Check laser alignment (see ment Section Adjusting table level) -- Laser misalignment
2. Calibrate level sensor (see Section Calibrating level sensor) 3. Check level indicator potentiometer in table base: pinion axle must be rectangular to toothed belt 4. Replace level indicator potentiometer
Resolution too low with large objective plane levels
-- Coupling bar has too much play -- Uneven motion of coupling bar -- Uneven motion of bucky device
1. Check coupling bar (movement, fixations, swivels; see Section Adjusting the tomo axis) 2. Check bucky device: -- bearings -- twisted rails -- dirt on rails -- collision with cables
Generally low resolution, even with low objective plane levels
In addition to the possibil- 1. Make test exposure with ities above: inclined bucky tray (see Section Test exposure with -- Column movement inclined bucky tray) and not properly aligned evaluate result with bucky device -- Column jerks -- Column drive not properly adjusted
2. Make test exposure with grating evaluate result
3. Check tomo drive (motor, driving wheel, eccentric disk, -- Transverse arm yaws see Section Checking drive (spring pressure for and position control switches) b-axis lock too low) 4. Increase spring pressure for -- Exposure time too b-axis lock low or too high 5. Check speed adjustment (exposure time coordination with generator)
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Philips
bucky DIAGNOST TS
bucky DIAGNOST TS bdts_3_990
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
3--5
Philips
FAULTFINDING
1.4.
bucky DIAGNOST TS
Specific instructions for fault elimination
1.4.1. EPROM replacement s Switch off PLC power supply. Protection against static electricity: an earth strap must be connected between the wrist and the base of the column before the EPROM is removed. The EPROM could otherwise be destroyed by static electricity. s Replace EPROM, pins 1 and 2 of the socket (see Figure) must be left unused. s Replace cover and switch power on.
1.4.2. Setting EPROM configuration jumper s Switch off PLC power supply. s Wait 1 minute, then remove the cover. s Set jumper J1 to ”E”. s Replace cover and switch power on. 1.4.3. Test exposure with inclined bucky tray s Place bucky tray on table--top as shown in the figure. s Adjust collimator aperture according to the figure. s Make test exposures in tomo mode and interpret the results.
70
Collimator aper max. 1 mm
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
s Wait 1 minute, then remove the cover.
3--6
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
bucky DIAGNOST TS bdts_3_990
Philips
bucky DIAGNOST TS
1.5.
FAULTFINDING
PLC input / output allocation
1.5.1. Digital inputs Signification when LED ON Operation mode
Service mode
¡0
Tomo on
Calibration
¡1
Tomo test
Confirmation of language/level
¡2
Set tomo angle
Language select
¡3
Set tomo time
Error reset
¡4
Raise table top
Raise table top
¡5
Lower table top
Lower table top
¡6
Light beam localiser ON
¡7
FFA ok
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
EA--0
Operation mode / Service mode EA--16 x-axis center lock (transverse arm extension)
¡ 17
y-axis center lock (column movement)
¡ 18
Service switch ON
¡ 19
spare
¡ 20
spare
¡ 21
spare
¡ 22
spare
¡ 23
spare
Philips
¡ 16
bucky DIAGNOST TS bdts_3_990
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
3--7
bucky DIAGNOST TS
1.5.2. Digital outputs Signification when LED ON Tomo mode
Standard mode
¡ 32
Radiation ON command 1 (redundant)
(permanently on)
¡ 33
Radiation ON command 2 (redundant)
(permanently on)
¡ 34
Unlock b--movement axis
¡ 35
Activate brake y-axis (column movement)
¡ 36
Release brake y-axis (column movement)
¡ 37
Disable center position locks
¡ 38
spare
¡ 39
spare
EA--32
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Philips
FAULTFINDING
3--8
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
bucky DIAGNOST TS bdts_3_990
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Philips
bucky DIAGNOST TS
1.6.
FAULTFINDING
Indicator LED’s on table control board
1.6.1. Standard version 1.6.1.1. Signal exchange between generator and table control H12 (yellow)
Signal XG_RD_EX (Exposure) from generator (LED on = signal low)
H15 (yellow)
Signal RG_DV (device active) from generator (LED on = signal low)
H16 (red)
Signal RQ_XG_EX (Radiation on) to generator (LED on = signal low)
1.6.1.2. Power supply H17 (green)
Supply electronics 24VDC
H18 (green)
Supply light beam localiser 24VAC
H20 (green)
Supply motors 24VDC
1.6.2. Additional LED’s for Tomo version 1.6.2.1. Signals from PLC to table control H1 (green)
Speed: low motor speed
H2 (green)
Dir: motor direction reverse
H3 (green)
On1: motor1 on
H4 (green)
On2: motor2 on
H5 (green)
Laser: objective plane indication on
H6 (green)
Bucky_contr: release bucky brake
H7 (green)
Bucky_start: trigger bucky device
H8 (green)
Tomo_ready: tomo start position reached
H9 (green)
Tomo_mode: tomo mode active
H10 (green)
Watchdog
1.6.2.2. Signals from table control to PLC and generator H11 (yellow)
Signal RQ_XG_PR (Preparation) from generator
H12 (yellow)
Signal XG_RD_EX (Exposure) from generator
H13 (yellow)
Coupling bar: coupling bar inserted to PLC
H15 (yellow)
Signal RG_DV (device active) from generator
H16 (red)
Signal RQ_XG_EX (Radiation on) to generator
1.6.2.3. Power supply H17 (green)
Supply electronics 24VDC
H18 (green)
Supply light beam localiser 24VAC
H19 (green)
Supply motors 12VDC
H20 (green)
Supply motors 24VDC
bucky DIAGNOST TS bdts_3_990
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
3--9
Drawings
Block diagram standard version
Z1--1
Block diagram tomo version
Z1--1.1
Power supply standard version
Z1--2
Power supply tomo version
Z1--2.1
Table control standard version
Z1--3
Table control tomo version
Z1--3.1
Bucky control standard version
Z1--4
Bucky control tomo version
Z1--4.1
Exposure Interface SZ3
Z1--4.2
Stand control standard version
Z1--5
Stand control tomo version
Z1--5.1
Bucky DIAGNOST VE with bucky unit non sensing
Z1--6.1
Stand control tomo version
Z1--7.1
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Philips
bucky DIAGNOST TS
bucky DIAGNOST TS bdts_z01_990
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
Z0--1
Philips
EX1100
stator cable, screened
EZX3
temperature switch
XA
tube assembly
H.V. cable
EG
see OPTIMUS manual
H.V. cable
Laser (option) X41
UE
brakes
EZX21
SX..
X1
X1
UE
PH
X32
X1
field
amplimat chamber
LA
UB X3
X1
X2
format contacts bucky 1
1WAX11
X40
X36
X38
VA
See table control
bucky module
centered
UE1
X20
UF
X22
X23
brakes
X37
light
X2
X11
manual collimator
brakes
tomo drive 1WAX1
release 1
X31
X4
brakes and table height
X31
X22
X24
brakes
see Z1-7
X7
SA
UE2
UB1
X4
brakes
stand control
X21
X30
UB1
control handle
stand control
X9
centered
UB
table control
X10
centered
RS 485
SA
X5
XP1
See table control
footswitch
RS 232
SZ1 X5
EA16/32
stand bucky DIAGNOST TS EA0
SC
M X6
table height
see Z1-7
PLC
PGU
tomo control
SNV see Z1-2 mains 230V~
ENX2100
X8
X16
X5
SN
table height
Upgrade kit Center Laser
UF tomo height
power supply
XA
SNV
UE
LA table bucky DIAGNOST TS PLC
ST
UB
VA
UB1 A3 00-02-23 Ost.
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
amplimat chamber 1
X49 X35
LCD display
amplimat chamber 2
EZX31
X4
X1
PH
PH
SA
amplimat chamber
SZ1 release 2
1WAX3
X1
SA
X2
VA bucky module
E
generator
bucky DIAGNOST TS C Philips Medical Systems
VP wall connection box
(99.0)
bucky DIAGNOST VE
SNV
SC
SN
Block diagram Standard version bucky DIAGNOST TS Z1-1
Philips
EX1100
stator cable, screened
EZX3
temperature switch
XA
tube assembly
H.V. cable EG
see OPTIMUS manual
H.V. cable
Laser (option) X41
UE
brakes X35
LCD display amplimat chamber 1
EZX21
SX..
X1
X49
X1
X32
X1
field
PH amplimat chamber
MEDIO CP release adapter (option) release 1 format contacts bucky 1
1WAX1
X1
X2
X3
X2
X40
X36
X38
VA bucky module
X2
centered
UE1
UF
X22
X23
brakes
X37
light
X20
hand collimator
brakes
tomo drive X1
SZ3 1WAX11
LA
UB
X11
X31
X4
brakes and table height
X31
X22
X24
brakes
see Z1-7
X7
SA
UE2
UB1
X4
brakes
stand control
X21
X30
UB1
control handle
stand control
X9
centered
UB
SZ1 table control
X10
X5
centered
RS 485
SA
footswitch
XP1
X5
EA16/32
SC tomo speed control
stand bucky DIAGNOST TS EA0
M X6
table height
see Z1-7
PLC
PGU
tomo control
1/2
SZ2
SNV
see Z1-2 mains 230V~
ENX2100
X16
X8
X5
SN
Upgrade kit Center Laser
table height
UF tomo height
power supply
XA
SNV
UE
LA table bucky DIAGNOST TS PLC
UB VA
UB1 A3 00-02-23 Ost.
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
UE
amplimat chamber 2
EZX31
X4
X1
PH
PH
SA
amplimat chamber
release 2
1WAX3
X1
SA
X2
VA bucky module
E
generator
bucky DIAGNOST TS C Philips Medical Systems
VP wall connection box
(99.0)
bucky DIAGNOST VE
SNV
SC
SN
SZ1
Block diagram tomo version bucky DIAGNOST TS Z1-1.1
Philips
stator cable, screened
EZX3
temperature switch H.V. cable
EG
H.V. cable
XA
tube assembly
Factory adjusted Z1 ENX2100:L1
X16:1
F11
X16:1.1
L1
L1.1
X16:3
5A slow
mains 230V~ ENX2100:N
X16:2
F12
N
T1 230VAC
30VAC
X8:2
28VAC
1.6A slow
mains filter X16:2.1
F14
F5
X16:4
0VAC
0VAC
X22:04
X20:06
X31:04
X32:4
X1:4 24VAC
H18
26VAC N.1
X11:06
6.3A slow
LA manual collimator
X8:3
X11:08
X20:08
X22:07
X31:07
X22:01
X31:01
X32:3
X1:3 0VAC
5A slow
G1 L1
1
X8:4
N
F6
X11:01
X20:01
X11:12
X20:12
24VAC
X49:2
X1:2
5VDC 0VDC
H17 X31:12
X8:5
center laser (optional)
F2 230VAC
X1:1
+24V
+24V
X22:12
5
X49:1
+24V 4A slow
switching power supply 24VDC/5A
X8:6
6.3A slow 24VAC
H20 0VAC
E
SN
generator
A3 99-11-23 Ost.
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Philips
EX1100
power supply
0VAC
X8:8
0VAC
SZ
table control
UB
stand control
UE
control handle
Power supply Standard version bucky DIAGNOST TS bucky DIAGNOST TS C Philips Medical Systems
(99.0)
Z1-2
Philips
stator cable, screened
EZX3
temperature switch H.V. cable
EG
H.V. cable
XA
+24VDC
tube assembly
GND
PLC
tomo control
Factory adjusted F14 ENX2100:L1
X16:1
F11
X16:1.1
L1
L1.1
X16:3
T1
6.3A slow
30VAC
230VAC
mains 230V~ ENX2100:N
X16:2
F12
X8:2
28VAC
5A slow
mains filter X16:2.1
N
F5
X16:4
0VAC
0VAC
X20:06
X22:04
X31:04
X32:4
X1:4 24VAC
LA
H18
26VAC N.1
X11:06
6.3A slow
X8:3
X11:08
X20:08
X22:07
X31:07
X22:01
X31:01
X32:3
hand collimator
X1:3 0VAC
5A slow
G1 L1
1
X8:4
N
X20:01
X1:1
X49:2
X1:2
+24V
+24V
5VDC 0VDC
H17 X22:12
X31:12
X20:12
X8:5
center laser (optional)
F2 37VAC
X49:1
+24V
X11:12 5
T2
X11:01
F6 4A slow
switching power supply 24VDC/5A
X8:6
6.3A slow 31VAC
34VAC
H20
230VAC 31VAC 17.5VAC
F1 X8:7
6.3A slow 15.5VAC
15.5VAC
H19
0VAC 13.5VAC
E
generator
0VAC
SN
A3 99-11-23 Ost.
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Philips
EX1100
power supply
UB
X8:8
stand control
UE
control handle
0VAC
SZ
table control
Power supply tomo version bucky DIAGNOST TS bucky DIAGNOST TS C Philips Medical Systems
(99.0)
Z1-2-1
X10:3
+24V
2
Philips
X10:2
transverse brakes 2
D13
SAH5 SAS5
footswitch
SCS4
SAS21
1
X4:09
X7:7
X4:08
A
2
3
X7:5 X5:03
2 1
X5:02
ENABLE
X4:11
+24V
X4:12
U7
+5V ENABLE
1
X7:3
1b
X7:4
X4:02
D4
SAS2 release bucky brake
U6 &
2 down
DOWN
DOWN
1
U7
K1
X9:2
K2
X4:04
D10
X4:05
D9
X9:3
DK 1.2 X4:03
1
X9:4
DOWN
2
2
D22
ENABLE
X6:05
SAS3 bucky centered
4
U1
2
2 down 1
1
Q11
1
ENABLE
ENABLE
X6:03
1
X9:1
Q10 +24V
U6
SNVS8
Q6
SWGND
+5V
+24V
+24V
ENABLE
JP2B DOWN
1
U6
JP1
D5
2
longitudinal brake
4
1a
SNVS7
SAY3
1
U6
X7:2
SAS22
1
B
& 1b
X4:06
X7:6
SAY2b
+24V
X4:10 1a
B
SAH2
A
bucky centered
SAY1 brake bucky module
X6:04 X6:07
4
safety
1
2 table on 1
X6:06 X6:02
K4
+24V
+24V
Q3 Q7
Q9 F2 24VAC
X8:6
Q9
JP3
6.3A slow
from power supply see Z1-2
X8:2
GL1
0VAC
down
F5
3
brake table motor
SNVY7
table height on
4
off 5
X6:08
3
+
K4
M
-
SNVM1
K2 X6:09
3
5 24VAC
H18
0VAC 0VAC +24VDC
X11:06
X20:06
X11:08
X20:08
X11:01
X20:01
X11:12
X20:12
X11:09
X20:09
X8:3 0VAC
X8:4
F6
+24V
U2 +24V
4A slow
H17
+24V
DC/DC
+5V
X8:5
SZ1
Table up Table down
C Philips Medical Systems
4
up 24VAC
6.3A slow
0VDC
bucky DIAGNOST TS
K1
5 X8:8
2
X6:11 4
24VAC
1
24VAC
H20
0VAC
X6:10
to stand see Z1-5
SNVS9
A3 99-11-26 Ost.
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
1
X4:07
footswitch inactive
Philips
SAY2a
(99.0)
k1/k2 Off k1/k2 ON
k4 ON k4 Off
table control
SNVS7
lowest range = 540mm
highest range = 810mm SNVS8 safety switch ~ 200N without load! SNVS9 SAS21/22 Table up / down
Table control Standard version bucky DIAGNOST TS Z1-3
X11:02
data -
16
Philips
RS485 X11:03
data +
Dk3
k3
2 X10:2
transverse brakes
X4:07 2
D13
X4:06
SAY2b
1
SAH5 footswitch inactive 1
SAS5
footswitch 1
1
Jumper
SCS4
U6
3
X4:09
X7:6
polyswitch RXE185
1
B A
2 X4:08
X7:7
SAY3
1
longitudinal brake
4 3
X7:5 X5:03
2 1
X5:02
+ +24V
K5
1
X4:11
1b
X4:12
4 &
+24V
+5V
1a
ENABLE
1
X7:3
1b
X7:4
D5
X4:02
ENABLE
JP2B DOWN
Q6
SWGND
Q10
+5V
0.5
X9:1
0.5
1
SAS2
+24V
+24V
U6
U6 X6:03
1
1
D4
up
ENABLE
ENABLE
DOWN
DOWN
&
1
Dk1.2 2 down 1
release bucky brake
K1
1
1
X9:2
K2
X4:04
SZ2
2
+24V
SNVS8
1
1
on 4 1
2
off 5
X6:04
X11:07
3
X11:10
4
safety
1
Tomo Motor
2 table on
K4
1
X6:06 X6:02
+24V
31VAC
31VAC
H19
fast slow
F2 17.5VAC
X8:6
4
from power supply see Z1-2
24VAC
down 4 up
GL1
0VAC
down 4
F5
up 24VAC
6.3A slow
0VAC +24VDC
X8:4
K1 3
2
polyswitch RUE700
table motor
SNVY7
A
X5:09 +5VDC
table height on 4 off 5
X6:08
3
+
K4
M
sensor coupling bar
SNV H4 laserdiode
500 E table potentiometer SNV P1
SNVM1
K2 X6:09
3
+24V
+24V
H17
brake bucky module
-
U2 DC/DC
SAY1
SNV S15
X5:10
0VAC
+24V
bucky centered
X5:08
D
24VAC
F6 4A slow
X5:04
SAH2
A
X5:07
5
X11:06
X20:06
X11:08
X20:08
X11:01
X20:01
X11:12
X20:12
X11:09
X20:09
+5V
X8:5
SZ1
Table up Table down
C Philips Medical Systems
1
H18
X8:3
1
B
X5:06
0VAC
0VDC
bucky DIAGNOST TS
brake
X6:11
F3
17.5VAC
X8:8
X6:10
5
5
X8:2
+
Q7 Q9
3
H20
0VAC
+24V
Q9
K5
JP3
6.3A slow
2 X9:4
Q3 6.3A slow
X9:3
X5:05
F1 X8:7
M2
K3
X6:07
SNVS9
D9 D22
ENABLE DOWN
2
X4:05
PIC-Controller
2
D10
to stand see Z1-7 (tomo drive)
SNVS7
down
M
X6:05
bucky centered
4
U1
2
K6
2 down
SAS3
Q11
to stand see Z1-5
SAS22
U6
JP1
1
1
U7
X7:2
1
1 -
ENABLE
fast
SAS21
Dk5
U6
slow
X4:10 1a
A3 99-11-23 Ost.
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
SAY2a
1
F4 I =2A
Philips
X10:3
+24V
to stand see Z1-7 (tomo drive)
U6
(99.0)
k1/k2 Off k1/k2 ON
k4 ON k4 Off
+ Tomo Tomo Start
K3 ON = Tomo Start
K5 Off = Slow k5 ON = fast
table control
SNVS7
lowest range = 540mm
highest range = 810mm SNVS8 safety switch ~ 200N without load! SNVS9 SAS21/22 Table up / down
Table control tomo version bucky DIAGNOST TS Z1-3-1
Philips
SX11:11
FD L CH
X1:01
EZX21:12
SX11:12
FD C CH
X1:02
EZX21:03
SX11:03
FD R CH
X1:03
EZX21:04
SX11:04
REL CH
X1:04
EZX21:05
SX11:05
PO 15/40V
X1:05
EZX21:06
SX11:06
NG 15V
X1:06
EZX21:07
SX11:07
SIGN CH
X1:07
PH
EZX21:08
SX11:08
RF 0V CH
X1:08
amplimat measuring chamber
EZX21:13
SX11:13
0V PW
X1:09
EZX21:01
SX11:01
PO 400V
X1:11
housing
housing
screen
X1:10
VA bucky module
X23 +15V
E/WA
delay new
P1
P2
W1 +26V XG +26V XG
8 +26V XG
X1:05
ACTIVE +26V XG
7
+5V
L5 WAX1:05
1
2
tilt switch
3
IC1A
IC1B
X3:1
+26V XG
X2:1
timer
C7
timer ~20s
RY2
2
D1
RY1 SYNC
T2
+15V
RESET
R8 X24 4
D2 +24V
X25
X3:5
L4
+15V +15V
1
XG
D5
RQ_XG_PR +26V XG
X3:6
X2:6
D3
R11
2 1
X26
X2:5
X1:03
H11
S1
+24V
H15
+5V
DC/DC
WAX1:03
X30
FREE EXPOSURE
6
X28 X21
R4
X2:9
X2:7
RQ_XG_PR
+24V 3
X20
1
X1:04
WAX1:04
H16
RG_DV
5
+24V
R2
+26V XG
4
CASS_PR
generator release 1 "bucky"
VA2 sensing
X22
RY2
T3
R13
R12
X27
Y1 2
& START
C8 +26V XG
+5V
X3:4
X2:4
R13
OC1
D4 +24V
+15V
X29 X31
L7 XG_RD_EX
WAX1:02
JP4 X3:3
X1:02
H12
RQ_XG_EX
GND XG
WAX1:01
X1:01
WAX1:07
X1:07
2
X2:3
XG_RD_EX
D6 4
5 3
JP5 X3:2
RY1
VA2 sensing
RY2 1
CASS_PR
X32 X33
SYNC
X2:2
VA1 bucky control XG
bucky 1 SW SF CF format contact
SL XG T0
WAX11:01 WAX11:02
1)
WAX11:03
tomo mode
A3 99-11-23 Ost.
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Philips
EZX21:11
WAX11:04
SW TO 1
WAX11:05
tomo ready ACTIVE WAX11:06
X11:11
see Z1-5 ACTIVE
to stand
WAX11:09 bucky ready SW BU 1
WAX11:10
1)
SZ1
table control 1) connector delivered with bucky DIAGNOST TS
bucky DIAGNOST TS C Philips Medical Systems
(99.0)
Bucky control standard version bucky DIAGNOST TS with bucky Module non sensing Z1-4
Philips
SX11:11
FD L CH
X1:01
EZX21:12
SX11:12
FD C CH
X1:02
EZX21:03
SX11:03
FD R CH
X1:03
EZX21:04
SX11:04
REL CH
X1:04
EZX21:05
SX11:05
PO 15/40V
X1:05
EZX21:06
SX11:06
NG 15V
X1:06
EZX21:07
SX11:07
SIGN CH
X1:07
PH
EZX21:08
SX11:08
RF 0V CH
X1:08
amplimat measuring chamber
EZX21:13
SX11:13
0V PW
X1:09
EZX21:01
SX11:01
PO 400V
X1:11
housing
housing
Screen
X1:10
VA bucky module
X23 +15V
E/WA
delay new
P1
P2 delay old
WAX1:04
W1
WAX1:05
1
2
+26V XG
tilt switch
3
IC1A
IC1B
2
D1
timer
timer ~20s
RY1
ACTIVE
X2:1
X3:1
+26V XG
FREE EXPOSURE
6
X24 4 +24V
X25
X1:03 RQ_XG_PR +26V XG
X2:6
D3
R11
2 1
X26
timer ~2s
+15V
XG
X3:6
S1
+24V
IC2
+15V
X2:5
X3:5
L4
R8
D2
DC/DC
H11
+15V
RESET
H15
+5V
SYNC
T2
+5V
X1:05 +26V XG
WAX1:03
X30
C7
7
8 +26V XG
X28 X21
R4
X2:9
+26V XG
X2:7
RQ_XG_PR
+24V 3
X20
1
X1:04
H16
RG_DV
5
+24V
R2
+26V XG
4
CASS_PR
generator release 1 "bucky"
VA2 sensing
X22
RY2
D5
R13
R12
X27
Y1
T3 & START
C8 +26V XG
+5V
X2:4
X3:4
R13
OC1
D4 +24V
+15V
X29 X31
L7 XG_RD_EX
WAX1:02
JP4
X1:02
D6 XG_RD_EX
GND XG
WAX1:01
X1:01
WAX1:07
X1:07
5
4 3
JP5 RQ_XG_EX
2
X2:3
X3:3
H12
X3:2
RY1
VA2 sensing
RY2 1
CASS_PR
X32 X33
SYNC
X2:2
VA1 bucky control
bucky 1 SW SF CF format contact
SL XG T0
WAX11:01
X2:01
WAX11:02
X2:02
WAX11:03
X2:03
WAX11:04
X2:04
WAX11:05
X2:05
SW TO 1
controller
tomo mode
A3 99-11-23 Ost.
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Philips
EZX21:11
X11:05
Rad on2
see Z1-7
tomo ready
from stand (PLC) WAX11:06
X2:06
WAX11:08
X2:08
Rad on1
ACTIVE
bucky ready SW BU 1
WAX11:09
X2:09
WAX11:10
X2:10
SZ1
X11:11
see Z1-7 ACTIVE
to stand
table control 1) connector delivered with bucky DIAGNOST TS
bucky DIAGNOST TS C Philips Medical Systems
(99.0)
Bucky control tomo version bucky DIAGNOST TS with bucky Module non sensing Z1-4-1
Philips
MEDIO CP EMX2/5/12/22
4
4
MEDIO CP-H EZAX1/2/3/4 4
0V
X1:4
X2:4 10R
C1
Generator 7
5
7
7
5
5
-(24)V
RDGV
100 m 63V
X1:7
X1:5
X1:4
C2
100n
100n
7924 1µF 40V
K4
0V
K4
100 m X2:7 63V
-(24)V
X1:7
X2:5
RDGV
X1:5
PREP
X1:3
10R 1N5060
3
3
3
PREP
K3
X1:3
1µF 40V
K3
1µF 40V
K2
X2:3
10R
bucky TS table SZ1
1N5060
2
1
2
X1:2
2
EXP 1
1
X1:1
K2 1N5060
10R
K1
X2:2
X1:2
10R
K1
EXP X2:1
X1:1
SZ3 Interface SCP: Programming for EZ X1/2 see EZ106 and EZ102 see B22 and B23 for W40/W41
A3 00-03-01 Schr.
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
SCP EZX1/2
Exposure interface SZ3 bucky DIAGNOST TS /Tomo and Medio CP /CP-H or SCP bucky DIAGNOST TS C Philips Medical Systems
(00.0)
Z1-4.2
Philips
SCP EZX1/2
MEDIO CP EMX2/5/12/22 4
4
0V
X1:4
V1
MP1 0V
R1
C5 100 m 63V
Generator 7
7
7
-(24)V
X1:7
X2:4 C1 100 m 63V
C2 100n 63V
A1
1 GND 3 2 Vin Vout
V2
MP2 -24V AC
5
5
5
RDGV
X1:5
3
3
3
PREP
X1:3
2
2
X1:2
2 EXP
1
1
1
X1:1
V10
X2:5
RDGV
X1:5
X2:3
PREP
X1:3
V6 MP5 RGDV
K4
V5
MP7 EXPout
K2
X2:2
R3
K1
H1 R6
Interface
bucky TS table SZ1
MP6 PREP
K3
V4
K1
SZ3
X1:7
MP4 -24V DC
K2 R8
-(24)V
R4
C8
H3
R10 X2:7
K3 R7
V9
X1:4
R5
C7
H2
0V
H5
C3 100 m 63V
K4 R9
V8
C4 100n 63V
V7
C6
H4
A3 00-03-02 Schr.
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
4
MEDIO CP-H EZAX1/2/3/4
V3 R2
X1:2
EXP MP8 EXPin X2:1
X1:1
SCP: Programming for EZ X1/2 see EZ106 and EZ102 see B22 and B23 for W40/W41
Exposure interface SZ3 bucky DIAGNOST TS /Tomo and Medio CP /CP-H or SCP bucky DIAGNOST TS C Philips Medical Systems
(a/00.0)
Z1-4.2
b b UES21
+5V
U5
+24V
Philips
X35:3
x
z
X1:02
X40:02
X36:2
X38:2
X36:3
X38:3
X36:4
X38:4
+5V +5V
X1:12
X40:12
X1:24
X40:24
X36:5
X38:5
X1:21
+24V
+5V
+5V
U5 +24V
+5V X35:2
U5
DC/DC
U6
+24V
X40:01
X38:7
X36:8
X38:8
1
U6
LED_X
X37:1
Y5
1
U3
x
2
clock
S13
generator
UES23
+5V
U6
U5
+24V
X37:3
1
4
x
X37:4
S23
U4 +5V
UF
tube arm
U7
H4
D1
X1:04
X40:04
LED_Z
UES24
+5V
16-bit micro contoller
driver
U5
X35:6
+5V X1:05
X40:05
X24:5
+24V LED_ b
X31:11
UBS19 1
X31:02
2
X24:3
+24V
LED_ a
column rot. b
X24:6
X22:11
U3
U7
H5
Upgrade kit Center Laser
SID 110 cm
4
X37:2
driver
X35:5
tube support z
X31:9
U7 X1:01
X36:7
2
+5V
+24V
D1
X38:6
S20
+5V
H1
Y6
1 X36:6
column
a
S14
U8
S25
UES20
tube support a
4
U3
X40:21
X35:7
1
+24V +5V
UES25
a
2
S12 U6
LED_Y
driver
a
Y4
1
U7 D1
y
1
JP1
X22:02
X24:4
UBY8
2
y
X24:1
UES22
+5V see Z1-2
X35:4
+24V
U6
U5
X31:10
UBS17
X24:2
X22:10
S22
1
4
column centered X31:09
+5V
D1
X1:03
X40:03
LED_ b
+5V
UB1
LED_ b
driver
H7
X22:09
U7
H3
stand control
U7 active
D1
X1:07
X40:07
LED_ACTIVE
ACTIVE = Tube select
driver
X49:1
X1:1 5V DC
X49:2
0V X1:2
X32:5
X1:5
M4
X32:6
X1:6
M3
Laser *
+5V
time
Q1/Q2
S8 X1:08
X40:14
LIGHT ON
LIGHT ON
UE1
P1
timer X35:1
X11:11
X20:11
X22:05
X31:05
X11:06
X20:06
X22:04
X31:04
X32:4
X1:4
M5
X11:08
X20:08
X22:07
X31:07
X32:3
X1:3
M6
ACTIVE
ACTIVE
24VAC
24VAC
H1 0VAC
A3 00-02-23 Ost.
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
X38:1
U3
+5V
H2
UES26
Philips
X36:1
+24V
+24V
field
0VAC 2
1
X11:01
X20:01
X22:01
X31:01
X11:12
X20:12
X22:12
X31:12
X11:09
X20:09
X22:09
X31:09
K1
+24V
see Z1-2/Z1-3
UE2 SZ1
table control
UB1
stand control
UE
* Optional bucky DIAGNOST TS C Philips Medical Systems
(99.0)
control handle
LA
manual collimator
Stand control standard version bucky DIAGNOST TS Z1-5
b b UES21
+5V
Philips
U5
+24V
S21
X35:3
X36:2
X38:2
X36:3
X38:3
X36:4
X38:4
X36:5
X38:5
X36:6
X38:6
X36:7
X38:7
X36:8
X38:8
a
2
x
z
S12 X41:14
a
U6
LED_Y
X41:01
+24V
VCL +24V
UES25
+5V X35:7
4
a
U6
U8 +24V
+5V
Y6
1 2
z
S14
U5
S25
UES20
DC/DC
U5
+24V
1
SID 110 cm
4
+5V X31:9
S20
X35:2
+24V X41:15
1
U3
X41:09
y
X37:1
Y5
1
U3
U6
LED_X
X37:2
x
2
clock
UES23
+5V
S13
generator
U5
U6
S23
X35:5
+24V
X37:3
1
4
x
X37:4
U4 X41:11
UF
LED_Z
tomo control handle
UES24
+5V
U5
LED_b
X31:11
+5V
U5
JP1
X22:02
X31:02
X24:4
X31:10
LED_ACTIVE
X41:08
LIGHT ON
2
1
X31:09
ACTIVE = Tube select
y
UBS17
X24:2
X22:10
LED_b
X41:13
UBY8
X24:1
+24V
U6 LED_b
1
D1
S22
X35:4
X41:12
b
X24:3
+24V
U3 see Z1-2
UBS19
X24:6
X22:11
LED_a
UES22
Upgrade kit Center Laser
X24:5
+24V
EPLD X41:13
tube arm
S24
X35:6
4
X22:09 LCD
UB1
X41:07
(2x10)
Laser *
UE1
X49:1
X1:1 DC
X49:2
0V X1:2
X32:5
X1:5
signal converter X41:03
display
stand control
X34
X41:06
time
MAX 232
LIGHT ON
P1
M1
M4
Centering laser (option)
X32:6
X35:1
ACTIVE 24VAC
X41:04
X41:05
X41:02
tomo on
timer X1:6
M3
X11:11
X20:11
X22:05
X31:05
X11:06
X20:06
X22:04
X31:04
X32:4
X1:4
M5
X11:08
X20:08
X22:07
X31:07
X32:3
X1:3
M6
M2
ACTIVE 24VAC
H1
X31:12
X11:09
X20:09
X22:09
X31:09
tomo on
UB1
table control
stand control
X30:05
tomo test
X30:04
SZ1
UE control handle EAO:0
bucky DIAGNOST TS
2
K1
+24V
X30:06
EAO:1
X30:07
EAO:2
X30:08
EAO:3
X30:09
EAO:4
X30:10
EAO:5
X30:11
EAO:6
(99.0)
C Philips Medical Systems
PLC
X30:15
EAO:7
X30:02
EA16:0
X30:03
PGU:3
X30:12
X30:01
UE2
PGU:8
LA
hand collimator
Stand control tomo version bucky DAIGNOST TS
5V
X22:12
RTS
X20:12
RX
X11:12
lock X
X31:01
FFA ok
X22:01
light on
X20:01
table down
X11:01
field
0VAC 1
table up
+24V
+24V
tomo time
see Z1-2/Z1-3
tomo angle
0VAC
A3 00-02-23 Ost.
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
1
U3
UES26
Philips
X38:1
X36:1
PGU:9
PGU:1
* Optional
Z1-5.1
b b UES21
+5V
Philips
U5
+24V
S21
X35:3
X38:1
X36A:1
1 V4
R1
X36A:2
X38:2
X36:3
X36A:3
X38:3 1
X36:4
X36:4
X36A:4
X38:4
X36:5
X36:5
X36A:5
X38:5
X36:6
X36:6
X36A:6
X38:6
X36:7
X36:7
X36A:7
X38:7 1
X36:8
X36:8
X36A:8
X38:8
X36:9
X36:9
X36A:9
X36:2
X36:2
X36:3
a
2
x
z
S12 X41:14
a
U6
LED_Y
X41:01
+24V
VCL +24V
UES25
+5V X35:7
a
DC/DC
U5
Y6
1 V5
U6
U8 +24V
+5V
4
R2
2
S14
U5
S25
UES20
+24V
4
z SID 110 cm
+5V
S20
X35:2
+24V X41:15
V2
U3
X41:09
y
X37:1
UE3
U3
U6
LED_X
UF
tube arm
Brake amplifier
X37:2 clock
UES23
+5V
generator
U5
U6
S23
X35:5
+24V
X37:3 X37:4
U4 X41:11
LED_Z
tomo control handle
UES24
+5V
U5
S24
X35:6
LED_b X41:13
X31:11
+5V
U5
JP1
X22:02
X31:02
X24:4
X31:10
LED_ACTIVE
X41:08
LIGHT ON
2
1
X31:09
ACTIVE = Tube select
y
UBS17
X24:2
X22:10
LED_b
X41:13
UBY8
X24:1
+24V
U6 LED_b
1
D1
S22
X35:4
X41:12
b
X24:3
+24V
U3 see Z1-2
UBS19
X24:6
X22:11
LED_a
UES22
Upgrade kit Center Laser
X24:5
+24V
EPLD
4
X22:09 LCD
UB1
X41:07
(2x10)
Laser *
UE1
X49:1
X1:1 DC
X49:2
0V X1:2
X32:5
X1:5
signal converter X41:03
display
stand control
X34
X41:06
time
MAX 232
LIGHT ON
P1
M1
M4
Centering laser (option)
X32:6
X35:1
ACTIVE 24VAC
X41:04
X41:05
X41:02
tomo on
timer X1:6
M3
X11:11
X20:11
X22:05
X31:05
X11:06
X20:06
X22:04
X31:04
X32:4
X1:4
M5
X11:08
X20:08
X22:07
X31:07
X32:3
X1:3
M6
M2
ACTIVE 24VAC
H1
X31:12
X11:09
X20:09
X22:09
X31:09
UB1
table control
stand control
X30:05
tomo test
X30:04
SZ1
UE control handle EAO:0
bucky DIAGNOST TS
2
K1
+24V
X30:06
EAO:1
X30:07
EAO:2
X30:08
EAO:3
X30:09
EAO:4
X30:10
EAO:5
X30:11
EAO:6
(00.0)
C Philips Medical Systems
PLC
X30:15
EAO:7
X30:02
EA16:0
X30:03
PGU:3
X30:12
X30:01
UE2
PGU:8
LA
hand collimator
Stand control Fix tube arm and tomo version bucky DAIGNOST TS
5V
X22:12
RTS
X20:12
RX
X11:12
lock X
X31:01
FFA ok
X22:01
light on
X20:01
table down
X11:01
field
0VAC 1
table up
+24V
+24V
tomo time
see Z1-2/Z1-3
tomo angle
0VAC
tomo on
A3 00-03-29 Ost./Schr.
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
V 1
X36:1
U3
UES26
Philips
X36:1
PGU:9
PGU:1
* Optional
Z1-5.1.1
Philips
X4:11
FD L CH
X1:01
EZX22:12
X4:12
FD C CH
X1:02
EZX22:03
X4:03
FD R CH
X1:03
EZX22:04
X4:04
REL CH
X1:04
EZX22:05
X4:05
PO 15/40V
X1:05
EZX22:06
X4:06
NG 15V
X1:06
EZX22:07
X4:07
SIGN CH
X1:07
PH
VA
EZX22:08
X4:08
RF 0V CH
X1:08
amplimat measuring chamber
EZX22:13
X4:13
0V PW
X1:09
bucky unit "Inalfa"
EZX22:01
X4:01
PO 400V
X1:11
housing
housing
screen
X1:10
X23
VA2 sensing
X22
RY2
+15V
4
5
CASS_PR
E/WA
VP
generator
wall connection box
delay new
+24V
3
tilt switch X20 X30
+24V
R2
R4 1
2
3
IC1A
2
IC1B
D1
timer ~20s
T2
RY1 1
timer
C7
+26V XG
WAX2:07
+15V
R8 X24
X1:1
X2:1
4
D2 +24V
X1:5
X1:6
timer ~2s
+15V
D3
R11
X25
2
S1 1
X26
X2:5
X2:6
+24V
IC2
+15V
DC/DC GND XG
SYNC
RESET
release 2 "BWS" WAX2:04
X28 X21
P2 delay old
P1
W1
1
D5 T3
R13
R12
Y1 X27
2
& START
C8 XG_RD_EX
WAX2:02
X1:4
X2:4
X1:3
X2:3
R13
OC1
D4
+15V
+24V
X29
X31 4
RQ_XG_EX
WAX2:01
X1:2
X2:2
X1:9
X2:9
5
6
VA2 sensing
RY2
D6
CASS_PR
X32
X33
7
RY2 X2:7
2 1
8
X1:7
RY1 SYNC
3
not used
A3 99-11-22 Ost.
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Philips
EZX22:11
FREE EXPOSURE
VA1 bucky control
bucky DIAGNSOT VE with bucky unit non sensing bucky DIAGNOST TS C Philips Medical Systems
(99.0)
Z1-6.1
UB1
X11:01
X20:09
X11:02
X20:02
X21:03
X11:03
X20:03
X21:04
stand control
shielding
W27
XS:11
RS485 data RS485 data +
Philips
X11:04
X21:05
X20:04
EA32:0 Rad on1
X11:05
X20:05
X11:06
X20:06
X11:08
X20:08
X11:01
X20:01
X11:12
X20:12
X21:06
Rad on2
EA32:1 EA32:8
W28
U finder 1
EA32:9
U finder 2
X11:07
GND
X20:07
X11:10
W27
X21:02
X20:10
X21:15
Umot 2X11:11
Rad on2
X30:01
PGU:3
X30:03
PGU:8
X30:02
PGU:9
X30:12
shielding
X20:11
X21:08
W20
SZ1
GND
XPI:INB1 XPI:INB2
X21:09
EA32:2
interrupt1 interrupt2 release clutch
table control
RS232
5V
EA0:0
X30:04
EA0:1
X30:05
EA0:2
X30:06
EA0:3
X30:07
EA0:4
X30:08
EA0:5
X30:09
EA0:6
X30:10
EA0:7
X30:11
EA0:8
X30:15
tomo on tomo test tomo angle tomo time table up table down light on FFA ok
K2 finder 4061.9024 X21:10
EA32:3
X21:01
EA32:4
X21:14
EA32:5
brake Y
EA0:9
X21:11
control handle
EA16:8
X21:12 service switch Nikkai/SS-12 SDP2
X lock
W29
UE EA16:9
S1
RTS
GND
24VDC
XPI:-
X21:07
device active
XPI:+
RX
24VDC
24V
24VDC
Umt 2+
PLC tomo control
X21:13
EA16:1 EA16:3 EA16:0
W28
R1
R2
1R (1W)
1R (1W)
C1
C2
1.5U 100V
1.5U 100V
K1
X22:04
W30
X31:04 U finder 1
X22:07
X31:07
X22:01
X31:01
X22:12
X31:12
U finder 2 24V GND
JP1 Jumper
F1 C3
1.5U 100V 1R
1R
R3
PS 102R (DO-41)
C4
1.5U 100V
(1W)
GL1
X22:06
D1
polyswitch 1.85A
R4 (1W)
~ ~ rectifier + -
W31
X31:06 release brakes
X22:08
X31:08
X22:02
X31:02
X22:03
X31:03
X22:05
X31:05
X22:10
X31:10
X22:11
X31:11
X22:09
X31:09
locks off brake Y spare device active sensor Y sensor beta shielding
UE control handle
X23:01
X23:02
X23:03
X23:04
X23:05
X23:06
X23:08
X23:07
X23:09
X23:10
X23:11
X23:12
X24:1
X24:2
X24:3
X24:1
X24:2
X24:3
X24:4
X24:5
X24:6
X24:4
X24:5
X24:6
R1
UB2
V1
Brake amplifier
V4
A3 00-03-29 Ost./Schr.
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
data +
Rad on1
GND
Philips
PGU:1 data -
XS:12
X24A:1
X24A:2
X24A:3
X24A:4
X24A:5
X24A:6
W24 W26
2
2
M UB
C Philips Medical Systems
(a/00.0)
M
S16: limit switch
M2: tomo motor
W22
W23
4
2
1
1
magnetic sensor 1
1 M3: coupling motor
bucky DIAGNOST TS
W21
W25
S20: safety end-switch
B1: angle sensor
S17: sensor lock Y sensor
Y8: brake Y column axis
S19: sensor beta-lock
Stand control tomo version bucky DIAGNOST TS Z1-7.1
Drawings
Survey of components tomo version
Z2--1
Wiring diagram table standard version
Z2--2.1
Wiring diagram table tomo version
Z2--2.2
Wiring diagram stand standard version
Z2--3.1
Wiring diagram stand tomo version
Z2--3.2
Wiring diagram stand standard version
Z2--3.3
Wiring diagram control handle tomo version
Z2--4.1
Wiring diagram control handle standard version
Z2--4.2
Eathing diagram
Z2--5
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Philips
bucky DIAGNOST TS
bucky DIAGNOST TS bdts_z02_990
(99.0)
1999 Philips Medical Systems ALL RIGHTS RESERVED
Z0--2
Philips
LA
SB SA
SA
VA
PH
SNV SC
SZ3
SZ2
SZ1
SN
UB1 UB UF SB
A4 00-11-10 Ost./Schr.
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
Philips
UE
XA
SA SB SC SN SNV SZ1 SZ2 SZ3 UB UB1 UF PH VA UE LA XA
-
Transverse carriage brakes Pivot tower Footswitch Power supply Motor assy & Motor control switches Table control Tomo speed control Tomo Medio CP interface Column basement Column control & Tomo control Column and transversal tube support arm Amplimat chamber Bucky carriage brake Control handle Collimator X-ray tube
bucky DIAGNOST TS C Philips Medical Systems
(00.0)
UE
Survey of components tomo version bucky DIAGNOST TS Z2-1
SAY2a
SAY2b
bucky module
SAS5/H5 3 B 4
SAY1 SZ1X7:5 SZ1X7:6 SZ1X7:7
WH BN GN
WH BN
SZ1X9:1 SZ1X9:2
2
1a 1 1b
1 2 B A
SZ1X3
SZ1X4:10 SZ1X4:11 SZ1X4:12
gy pk bl
SZ1X4:02 SZ1X4:03 SZ1X4:04 SZ1X4:05
wt bn gn yw
SZ1X9:4
from generator
Identitication of LED´s see chapter 4 "Trouble shooting"
2
1
SZ1X6:08 SZ1X6:09
T2:0V
Z1:L1
Z1:N
G1:L1
T1:0V T1: SN1X8:5
SN1X8:4 +24V 1
0V 5
X16:3 X16:4 X16:
X16:
T2:230V T1:230V
X16:4
X16:1.1
2 6
4 4 SAY1:1
SAY2a:1
SAY2a:2 3 3 SAY1:2
2 2
1 1 SAS3:1
8
9
2 5
3 6
SN1X8:2
X16:3
X16:4
SN1X8:8
SN1X8:3 0VAC
7 1 4
230VAC
9 3 6
0VAC
8 2 5
230VAC
7 1 4
T2
115VAC
9 3 6
SN1X8:6
02
0VAC
10
04
230VAC
13
13
15 12
11
10
12
11
10 6 3
8
5 2
7
4 1
G1 1WAX1
03
Y1
+
screen SAS22:1a SAS22:1 SAS22:1b SAS5:3 SAS5:3 SAH5:B SAS5:4
X7:1 X7:2 X7:3 X7:4 X7:5 X7:6 X7:7 X7:8 X7:9
SNV rd
2.1 1.1
T1
2
motor
blk
X1
9
8
7 yw SZ1X6:10 SZ1X6:11
X5
01 JP6
gn
05
1
X7
L1 N
3
07
14
X6
LINE L1
06
X4 S8
N
08
VAX2
2 5
SZ1X6:04
09
X9
2
10
X3
SAS3:4
X10
1 4
GN
Z1
table control
10A slow
24VAC
4
01
LOAD
G1:
F1
H20
SZ1X6:05
JP5
JP4
SZ1
L1.1
N.1
T2:
1
JP3
YE
02
X16:4
10A slow
8
03
+26V XG
0VAC
F2
6
UB1X20
1
SNT2:0VAC
5
7
X11
4A slow
X16:2.1
SNT2:17.5VAC
H16
F6
4
SNT2:37VAC
1WAX2
5
SNG1:0V
4
05 04
2 3
X16 4
9
SNG1:+24V
1
X2
3
WH
SNT1:0VAC
S7
F5 6.3A slow
8
4 SZ1X6:02
F12
X16:3
1 SNT1:24VAC
2
15
SZ1X6:03
F11
06
24VAC
12
BN
1
07
JP1 +24V
2
08
230VAC
X8
H18
3
09
115VAC
H17
H11 = RQ_XG_PR H12 = XG_RD_EX H15 = ACTIVE
4
X16:3
JP2B
JP2A
10
H15
H11
L
Z1:L1.1
Z1:N.1
H12
7
SZ1X6:07
S9
G1:N
4 PK
N
1
14
SZ1X6:06
11
GY
M1
-
A3 99-11-24 Ost.
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
SAS21
1
SZ1X9:3
2
Philips
1
transverse carriage
S3
4
2
SAS2/H2
SA
1
1
SZ1X4:08
GY PK BU
2
SZ1X4:09
SZ1X7:2 SZ1X7:3 SZ1X7:4
VAX2
SZ1X10:2
1a 1 1b
VA
SZ1X10:3
Philips
SAS22
PH
1 SZ1X4:06
SZ1X4:07
2
SAY3
generator
X1
X6:01 X6:02 X6:03 X6:04 X6:05 X6:06 X6:07 X6:08 X6:09 X6:10 X6:11 X6:12
screen SNVS7:1 SNVS7:2 SNVS8:1 SNVS8:2 SNVS9:1 SNVS9:4 SNVM1:+ SNVM1:SNVY7:1 SNVY7:2
X4:01 X4:02 X4:03 X4:04 X4:05 X4:06 X4:07 X4:08 X4:09 X4:10 X4:11 X4:12 X4:13 X4:14 X4:15
screen SAS2:1 SAS2:2 SAH2:B SAH2:A SAY2a:1 SAY2a:2 SAY3:1 SAY3:2 SAS21:1a SAS21:1 SAS21:1b
X5:01 X5:02 X5:03 X5:04 X5:05 X5:06 X5:07 X5:08 X5:09 X5:10 X5:11 X5:12 X5:13 X5:14 X5:15
screen SCS4:1 SCS4:2 SNVS15:1 SBS 15 SBS 15 SBH 4 SBH 4 SNVP1:1 SNVP1:2
SN
wt bn
SZ1X5:02
SZ1X5:03
4 2
bucky DIAGNOST TS C Philips Medical Systems
(99.0)
S4 1
SC foot switch
power supply
Wiring diagram table Standard version bucky DIAGNOST TS Z2-2.1
SAY2b
SAY2a
bn
SZ1X9:1 SZ1X9:2
SZ1X4:09
SZ1X4:08
SAS21
2
1
SZ1X4:10 gy SZ1X4:11 pk SZ1X4:12 bl
1a 1 1b
SZ1X3
SZ1X4:02 SZ1X4:03 SZ1X4:04 SZ1X4:05
1 2 B A
wt bn gn yw
from generator
Identitication of LED´s see chapter 4 "Trouble shooting"
S9
4
H1
H3
H5
SZ1X6:10 SZ1X6:11
1
rd
SZ1X6:09
+
2
3
4
SAY1:2
SAY1:1
M1
-
screen SAS22:1a SAS22:1 SAS22:1b SAS5:3 SAS5:3 SAH5:B SAS5:4
X7:1 X7:2 X7:3 X7:4 X7:5 X7:6 X7:7 X7:8 X7:9
X6:01 X6:02 X6:03 X6:04 X6:05 X6:06 X6:07 X6:08 X6:09 X6:10 X6:11 X6:12
screen SNVS7:1 SNVS7:2 SNVS8:1 SNVS8:2 SNVS9:1 SNVS9:4 SNVM1:+ SNVM1:SNVY7:1 SNVY7:2
Z1:N
T2:230V
T1:0V
T1:230V
T2:0V
T1: T2:
Z1:L1
X16:3
X16:4
SN1X8:4 +24V 1
G1:L1
SN1X8:5
X16:3
X16:4 X16:
X16:
0V 5
X16:1.1
3
5 9
X4:01 X4:02 X4:03 X4:04 X4:05 X4:06 X4:07 X4:08 X4:09 X4:10 X4:11 X4:12 X4:13 X4:14 X4:15
screen SAS2:1 SAS2:2 SAH2:B SAH2:A SAY2a:1 SAY2a:2 SAY3:1 SAY3:2 SAS21:1a SAS21:1 SAS21:1b
X5:01 X5:02 X5:03 X5:04 X5:05 X5:06 X5:07 X5:08 X5:09 X5:10 X5:11 X5:12 X5:13 X5:14 X5:15
SN1X8:3
X16:4 0VAC
230VAC
0VAC
115VAC
X16:3
SN1X8:8
X16:4
SN1X8:6 230VAC
9 3 6
0VAC
8 2 5
X16:3
04
02
12
11
10 7 1 4
9 3 6
9 3 6
6 3
X16:2.1
1 1
13
13
14 11
10
12
11
10
5 2
8
4 1
G1
T2
motor
blk
L1 N
1WAX1
03
Y1
2
SNV SZ1X6:08
X1
01
JP6
gn
X5
9
8
7
X7
yw
05
1
7
SZ1X6:04
2 5
gn
X6
LINE L1
06
X4 S8
2.1 1.1
07
2 4
N
08
VAX2
3
Z1
09
X9
1 4
SZ1X6:05
SAS3:4
X10
H19
24VAC
yw
LOAD
10
X3
SAS3:1
H20
2
8
SZ1X5:10
P1
01
L1.1
N.1
table control
10A slow
3 1
4
G1:N
F1
7
gn
SZ1X5:09
1WAX2
G1:
SZ1
2 5
bn
JP5
JP4
1
1 4
wt
05
02
F2 10A slow
JP3 SZ1X5:08
X16
+26V XG
4
8
UB1X20
8
7
F12
07
03
15
6
X11
4A slow
7
SNT2:0VAC
5
F11
04
14
SNT2:17.5VAC
4
1
08
H16
F6
1
SNT2:37VAC
X2
2
SNG1:0V
F5 6.3A slow
6
SNG1:+24V
1
2
09
06
24VAC
2 3
3
JP1 +24V
SAY2a:2
SNT1:0VAC
S7
H11 = RQ_XG_PR H12 = XG_RD_EX H15 = ACTIVE
H18
4
10
H15
4
4
JP2B
H13
SAY2a:1
SNT1:24VAC
2
H11
H8
230VAC
1
H6
115VAC
X8
SZ1X6:02
H14
0VAC
H17
wt
H12
H10
SN1X8:2
H4
JP2A
bn
H9
230VAC
H2
SZ1X6:03
H7
L
2
3
SZ1X6:07
2
pk
N Z1:N.1 Z1:L1.1
1
15
SZ1X6:06
12
gy
A3 99-11-25 Ost.
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
1
transverse carriage
S3
4
2
SAS2/H2
SA
1
SZ1X9:4
wt
SZ1X9:3
wt bn gn
2
Philips
VAX2
bucky module
SAY1 SZ1X7:5 SZ1X7:6 SZ1X7:7
1
VA
SAS5/H5 3 B 4
2
SZ1X10:2
PH
SZ1X10:3
gy pk bl
SZ1X4:06
SZ1X7:2 SZ1X7:3 SZ1X7:4
1a 1 1b
1
SZ1X4:07
Philips
SAS22
2
SAY3
generator
X1
T1
screen SCS4:1 SCS4:2 SNVS15:1 SBS 15 SBS 15 SBH 4 SBH 4 SNVP1:1 SNVP1:2
SN
power supply
2 yw gn
SZ1X5:05 SZ1X5:04
4
S15
1 wt SZ1X5:02
bn
bl SZ1X5:03
SZ1X5:07 -
2
H4 wt
4
SZ1X5:06
bucky DIAGNOST TS C Philips Medical Systems
+
SB (99.0)
S4 1
SC foot switch
Wiring diagram table tomo version bucky DIAGNOST TS Z2-2.2
Philips
X38:7 X38:8 1
4
wt bn
2
X38:2
X38:1
S14
Y4:1
2
Y4:2
3
S12:1
4
S12:4
5
Y6:1
1
UE2X37:3 GN UE2X37:4 YE
2
6
Y6:2
7
S14:1
8
S14:4
4
1
2
2
S13
a Y5 X38:3 X38:4 1
UE2X36
4
wt bn
x
X38:5
X38:6 2
UE
tube arm
S12
1
x
UF
2
control handle
see Z2-4 a
Y6
UF tube arm
z
stand control
center position
UBS19:1
UB1
y
9 3 6
2
UBY8 see Z2-6
2
y
4
5
6
1
2
3
X24
UBS17:4 UBY8:2 UBS17:1 UBS19:2 UBY8:1
JP1
4
1
9
UB S19
1
UB1X24:4
8 2 5
X22
UB1X24:3 7 1 4
12
10
9 3 6
11
12
8
11
7
X20
UB1X24:6 WH UB1X24:5 BN
2
8
4
7
UB S17
1
wt bn
2 5
10 UB1X24:2 UB1X24:1
UE2X31
1 4
SZ1X11
A3 99-11-26 Ost.
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
1
Y4
UE2X37:2
1
UE2X37:1
X38
b
Wiring diagram stand Standard version bucky DIAGNOST TS bucky DIAGNOST TS C Philips Medical Systems
(99.0)
Z2-3.1
Philips
X38:7 X38:8 1
4
wt bn
2
X38:2
X38:1
S14
Y4:1
2
Y4:2
3
S12:1
4
S12:4
5
Y6:1
1
UE2X37:3 GN UE2X37:4 YE
2
6
Y6:2
7
S14:1
8
S14:4
4
1
2
2
S13
a Y5 X38:3 X38:4 1
UE2X36
4
wt bn
x
X38:5
X38:6 2
UE
tube arm
S12
1
x
UF
2
control handle
see Z2-4 a
Y6
UF tube arm
z
stand control
center position
JP1
UB1X24:3
UB1X24:4
11
12
4
2
UBY8 see Z2-6
2
y
9
1
1
X24
UBS19:1
UB1
y
15
XS:11 XPI:+ EA32:1 EA32:0 EA32:2 INB2 EA16:3 EA16:1 XPI:EA32:5
3
12
6
9
3
14
6
5
13
11
4
10
8
2
7
2 5
1
7
UB S19
UBS17:4 UBY8:2 UBS17:1 UBS19:2 UBY8:1
X21
1 4
UB1X24:6 WH UB1X24:5 BN
X22
EA16:8
EA32:3
INB1
XS:12
9
10
X23
3 6
15
8
9
7
14
6
2 5
13
11 12
UBS16:1 UBM3:b UBM2:a UBS16:4 UBS20:1 UBM2:b UBB1 UBB1
3 6
X20
2
EA32:4
4
2 5
1 4
UB S17
1
wt bn
UBB1 10
8
1 4
12 9
7
2 5
8
11 8
UBM3:a 10 7
1 4
3
UB1X24:2 UB1X24:1
UBS20:2
UE2X31
UBS16:2
SZ1X11
A3 99-11-26 Ost.
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
1
Y4
UE2X37:2
1
UE2X37:1
X38
b
Wiring diagram stand tomo version bucky DIAGNOST TS bucky DIAGNOST TS C Philips Medical Systems
(99.0)
Z2-3.2
0
UEX30:04
1
UEX30:05
2
UEX30:06
3
UEX30:07
4
UEX30:08
5
UEX30:09
UB1X21:07
6
UEX30:10
UB1X21:08
7
UEX30:11
INB2
INB1
-
+
Philips
XPI
PLC
UB1X21:02
EA16
UB1X21:15
8
PGU D-SUB 9P. Jumpers (bottom side) CLOSED OPEN
8 4
5
7 3
6 2
UEX30:15
1
UB1X21:11
3
RS485
EA32
0
2 1
UB1X21:12
4
EA16
Interface Board
6 7
UB1X21:13
9
UB1X21:4
11
8
1
4
+
bn UB1X23:08
wt
-
a
+ b
M3
-
a
b
M2
GN
2
RD
4
gn
1
UB1X23:3 GY UB1X23:5 PK UB1X23:4 BU 2
rd
UB1X23:9 wt UB1X23:7 bn
UB1X23:06
UB1X23:01 gn
UB1X23:02 yw
UB1X23:10 UB1X23:11 UB1X23:12
UB1X21:3
9
7
EPROM
12
X5
8
5 6
UBX30:01
3 4 5
9
UBX30:03
1
UBX30:02
0
UB1X21:06
UBX30:12
UB1X21:05
2
UB M3 UB S20 Limit position
A3 99-11-26 Ost.
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
9
UB B1
UB S16
UB M2
Wiring diagram stand tomo version bucky DIAGNOST TS bucky DIAGNOST TS C Philips Medical Systems
(99.0)
Z2-3.3
Philips
X31
UB1X22
X30 X36
1
X37 4 3 2 1
1
X41
UFS13:1 UFY5:2
1
UFY5:1
X32 X35 LAX1
3
2 7
1
8
S21:1 S25:1 S20:1 S24:1 S20:4
X40
S23:1
UE2 control handle basic
4 9
S22:1
5 1
6
UE1X1
S8 H6 H1
H5
UE1
H2
H4
control handle LED's
X1
H3
4
S24
2
1 4
S23
2
4
1 4
2
S21
1
2
1
4
UE2X35:2
S21:4 UE2X35:1
UE2X35:3
S22:4 S20:4
UE2X35:4
S23:4 S21:4
UE2X35:5
S24:4 S22:4
S26:4 S23:4
UE2X35:6
UE2X40
2
1
S20
S22
2
UE2X35:7 S26:1
S26:4
S24:4 S25:4 4
S26
S25:1
LAX1
A3 99-11-26 Ost.
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
UFX38
UFS13:4
1
4
2
1
S25
Wiring diagram control handle tomo version bucky DIAGNOST TS bucky DIAGNOST TS C Philips Medical Systems
(99.0)
Z2-4.1
Philips
X34 X31 Display
UB1X22
X30 X36
1
X37 4 3 2 1
1
X41
UFS13:1 UFY5:2
1
UFY5:1
UE1 x 1
X32
Tomo control handle
X35 LAX1
3
2 7
1
8
S21:1 S25:1 S20:1 S24:1 S20:4
X40
S23:1
UE2 control handle basic
4 9
S22:1
5 1
6
UE1X1
S8 H6 H1
H5
UE1
H2
H4
control handle LED's
X1
H3
4
S24
2
1 4
S23
2
4
1 4
2
S21
1
2
1
4
UE2X35:2
S21:4 UE2X35:1
UE2X35:3
S22:4 S20:4
UE2X35:4
S23:4 S21:4
UE2X35:5
S24:4 S22:4
S26:4 S23:4
UE2X35:6
UE2X40
2
1
S20
S22
2
UE2X35:7 S26:1
S26:4
S24:4 S25:4 4
S26
S25:1
LAX1
A3 99-11-26 Ost.
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
UFX38
UFS13:4
1
4
2
1
S25
Wiring diagram control handle Standard version bucky DIAGNOST TS bucky DIAGNOST TS C Philips Medical Systems
(99.0)
Z2-4.2
Philips
XA
2.5 mm 2
UE
2.5 mm 2
LA
2.5 mm2 2.5 mm2
SA
SA
A3 00-03-28 Ost./Schr.
Philips
Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property
UF
4 mm 2 4 mm 2
to generator
8.25 mm 2
X16
U Z1
6 mm 2
SN
6 mm 2
SNV M1
UB 6 mm 2
bucky DIAGNOST TS C Philips Medical Systems
(a/00.0)
T1
G1
T2
Earthing diagram bucky DIAGNOST TS Z2-5