Philips

LIST OF DOCUMENTATION IN THIS BINDER:

Ä

Subsystem Manual bucky DIAGNOST TS

Ä

INALFA non sensing bucky unit

Ä

Manual Collimator Ralco

Note:

 indicates document present

LIST OF ALL BINDERS FOR BUCKY DIAGNOST TS --

bucky DIAGNOST TS

--

OPTIMUS generator

--

X-ray tube

--

bucky DIAGNOST VE/VT

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

RADIOGRAPHY

Manual Order No. 4512 984 06663 bucky DIAGNOST TS bdts_01_b002

(b/00.2) ã 2000 Philips Medical Systems ALL RIGHTS RESERVED

0.1

Philips Medical Systems Development and Manufacturing Centre

INTRODUCTION AND TECHNICAL DATA

1

INSTALLATION

2

FAULT FINDING

3

REPLACEMENT

4

PROGRAMMINGS

5

ADJUSTMENTS

6

ACCEPTANCE

7

SERVICE INFORMATION

8

PARTS LIST

P

Philips

SERVICE MANUAL 704 SUBSYSTEM

9848 600 00061

Fix

9848 600 00071

Elevation

9848 600 00081

Tomo

bucky table for routine radiography diagnostics

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

bucky DIAGNOST TS

DMC Hamburg SCHEMATIC DIAGRAMS

Z1-

WIRING DIAGRAMS

Z2-

Printed in Hamburg Federal Republic of Germany 2000 Philips Medical Systems ALL RIGHTS RESERVED

Manual Order No. 4512 984 06663 bucky DIAGNOST TS (b/00.2) reg_984 06663

1

Philips

Bucky DIAGNOST TS

SERVICE MANUAL -- SUBSYSTEM Bucky DIAGNOST TS

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Type No:

Author: V. Neumann

9848 600 00061 9848 600 00071 9848 600 00081

In case there are any questions concerning this manual, please send this LOPAD via fax to 49/(0)40/5078 2481

File:bD_TS_b002

List of pages and drawings (LOPAD)

0.1

(b/00.2)

1 3.1 3.2

(b/00.2) (b/00.2) (b/00.2)

A4

Manual Order No: 4512 984 06663

(Rosa Karton)

_________________________________________ 1-0.1/0.2

(99.0)

1-1 ... 1-15

(99.0)

1Z--1.1 1Z--1.2 1Z--1.3 1Z--1.4 1Z--1.5

(99.0) (99.0) (97.1) (99.0) (99.0)

A3 A3 A3 A3 A3

1Z--5.1 1Z--5.2

(99.0) (99.0)

A3 A3

1Z--6.1 1Z--6.2 1Z--6.3

(99.0) (99.0) (00.0)

A3 A3 A3

1Z--2.1

(99.0)

A3

1Z--7

(00.0)

A3

1Z--2.2.1 1Z--2.2.2

(99.0) (99.0)

A3 A3

1Z--8

(99.0)

A3

1Z--9

(99.1)

A3

1Z--3.1 1Z--3.2

(97.1) (97.1)

A3 A3

1Z--10

(00.0)

A3

Z--50

(00.0)

A3

1Z--4.1 (97.1) A3 1Z--4.2 (97.1) A3 _________________________________________

bucky DIAGNOST TS bdts_lopad_b002

(b/00.2) 2000 Philips Medical Systems ALL RIGHTS RESERVED

4512 982 00081

3.1

Philips

bucky DIAGNOST TS 2-0.1

(99.0)

2-1 ...2- 36

(99.0)

2Z--2.0

(99.0)

2Z--10 (00.0) A3 _________________________________________ 3-0.1

(99.0)

3-1 ... 3-9 (99.0) _________________________________________ P-List 9848 600 02641 _________________________________________ Z0--1

(99.0)

Z1--1 Z1--1.1

(99.0) (99.0)

A3 A3

Z1--2 Z1--2.1

(99.0) (99.0)

A3 A3

Z1--3 Z1--3.1

(99.0) (99.0)

A3 A3

Z1--4 Z1--4.1 Z1--4.2 Z1--4.2

(99.0) (99.0) (00.0) (a/00.0)

A3 A3 A3 A3

Z1--5 Z1--5.1 Z1--5.1.1

(a/00.0) (a/00.0) (00.0)

A3 A3 A3

(99.0)

A3

Z1--6.1

Z1--7.1 (a/00.0) A3 _________________________________________ Z0--2

(99.0)

Z2--1

(00.0)

A4

Z2--2.1 Z2--2.2

(99.0) (99.0)

A3 A3

Z2--3.1 Z2--3.2 Z2--3.3

(99.0) (99.0) (99.0)

A3 A3 A3

Z2--4.1 Z2--4.2

(99.0) (99.0)

A3 A3

(a/00.0)

A3

Z2--5

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

A4

3.2

(b/00.2) 2000 Philips Medical Systems ALL RIGHTS RESERVED

bucky DIAGNOST TS bdts_lopad_b002

bucky DIAGNOST TS

INTRODUCTION AND TECHNICAL DATA

Philips

INTRODUCTION AND TECHNICAL DATA TEXT

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1--0.1 1. 1.1. 1.1.1. 1.1.2. 1.1.3. 1.1.4. 1.2. 1.3. 1.3.1. 1.3.2. 1.4. 1.4.1. 1.4.2. 1.5. 1.5.1. 1.5.2. 1.6. 1.7. 1.8.

Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . System bucky DIAGNOST TS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Standard Table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Tomography table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . X-ray generator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Applications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Tubes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Mechanical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Movements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Electrical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . X-ray generator OPTIMUS 50 / 65 / 80 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Type of power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Calculating the mains resistances . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Earth-leakage circuit breaker . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Emergency-OFF device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Heat emission . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1--1 1--1 1--1 1--1 1--1 1--1 1--2 1--2 1--2 1--2 1--2 1--2 1--3 1--4 1--4 1--5 1--5 1--6 1--6

2. 2.1. 2.2.

Environmental conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . X-ray generator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1--6 1--6 1--6

3. 3.1. 3.2. 3.2.1. 3.2.2. 3.2.3.

Measurements and weights . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Pre-installed system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Transport data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Box dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Roadway . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1--7 1--7 1--8 1--8 1--8 1--9

4. 4.1.

1--10

4.4. 4.4.1. 4.4.2.

Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ”Fixed table” 9848 600 00061 – bucky DIAGNOST TS with floating table top . . . . . . . . . . . . . . Height adjustable 9848 600 00071 bucky DIAGNOST TS with floating table and elevator . . . . . . . . Tomography 9848 600 00081 bucky DIAGNOST TS with floating table and elevator and tomo extension . . . . . . . . . . . . . . . . . . . . . . . . Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . bucky DIAGNOST TS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . bucky DIAGNOST VE / VT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5. 5.1. 5.2. 5.3. 5.4. 5.5. 5.6.

Site planning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Planning data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Determining the site of installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Floor fixing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Loading capacity of floor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Height of the room . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Electrical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1--13 1--13 1--13 1--13 1--13 1--13 1--13

4.2.

Philips

4.3.

bucky DIAGNOST TS INH

 1999

(99.0)

Philips Medical Systems ALL RIGHTS RESERVED

1--10 1--11 1--11 1--11 1--11 1--12

1--0.1

Philips

INTRODUCTION AND TECHNICAL DATA

bucky DIAGNOST TS

6.

Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1--14

7. 7.1.

Installation materials required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1--14 1--15

DRAWINGS 1Z--1.1 1Z--1.2 1Z--1.3 1Z--1.4 1Z--1.5 1Z--2.1 1Z--2.2.1 1Z--2.2.2 1Z--3.1 1Z--3.2 1Z--4.1 1Z--4.2 1Z--5.1 1Z--5.2 1Z--6.1 1Z--6.2 1Z--7 1Z--8 1Z--9

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

bucky DIAGNOST TS elevated or tomo version for wall stand left, mechanical dimensions . . . . . . . . . bucky DIAGNOST TS elevated or tomo version for wall stand right, mechanical dimensions . . . . . . . . OPTIMUS L 50, mechanical dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . bucky DIAGNOST TS fixed table version for wall stand left, mechanical dimensions . . . . . . . . . . . . . . . bucky DIAGNOST TS fixed table version for wall stand right, mechanical dimensions . . . . . . . . . . . . . . bucky DIAGNOST TS elevated or tomo version, wall clearance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . bucky DIAGNOST TS fixed table version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . bucky DIAGNOST TS fixed table version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . bucky DIAGNOST TS for wall stand left with bucky DIAGNOST VE typical room layout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . bucky DIAGNOST TS for wall stand right with bucky DIAGNOST VE typical room layout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . bucky DIAGNOST TS for wall stand left with bucky DIAGNOST VE relation data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . bucky DIAGNOST TS for wall stand right with bucky DIAGNOST VE relation data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . bucky DIAGNOST TS elevated or tomo version, connecting surface / floor fixing . . . . . . . . . . . . . . . . . . bucky DIAGNOST TS fixed table version, connecting surface / floor fixing . . . . . . . . . . . . . . . . . . . . . . . . bucky DIAGNOST TS, VE, OPTIMUS L 50 standard version, connection diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . bucky DIAGNOST TS, VE, OPTIMUS L 50 tomo version, connection diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . bucky DIAGNOST TS, VE, OPTIMUS L 50 earthing diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . bucky DIAGNOST TS, VE system installation and cable length . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Wall connection box bucky DIAGNOST TS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1--0.2

 1999

(99.0)

Philips Medical Systems ALL RIGHTS RESERVED

bucky DIAGNOST TS INH

1.

Technical data

1.1.

System bucky DIAGNOST TS

INTRODUCTION AND TECHNICAL DATA

1.1.1. Standard Table The bucky DIAGNOST TS is installed with the X-ray generator OPTIMUS L 50. The bucky DIAGNOST TS is characterized by: -- X-ray exposures at the horizontal patient with bucky unit. -- X-ray exposures at the horizontal patient without bucky unit. -- Patient area of table top 2400mm x 600mm (excl. the Alu profile) -- SID 1000mm to 1250mm. 1.1.2. Tomography table Linear tomography at the horizontal patient. -- Angle:

8° ... 40°

-- Time:

0.5 ... 4sec.

-- Layer:

0° ... 250mm

-- SID:

1000 ... 1250mm, 1100mm default

1.1.3. X-ray generator The X-ray generator OPTIMUS L 50 for radiography is based on computer-controlled converter technology. The converter operates in the non-audible frequency range. Applicational options are essentially achieved by releasing software modules. 1.1.4. Applications -- Radiography -- Tomography

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Philips

bucky DIAGNOST TS

bucky DIAGNOST TS bdts_1_990

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

1--1

Philips

INTRODUCTION AND TECHNICAL DATA

1.2.

bucky DIAGNOST TS

Tubes

Versions of the X-ray generator OPTIMUS L 50 50kW 3phase generator for 1 tube. Recommended standard tubes: RO 17 50

0.6 / 1.3 mm

R type stators only for low speed rotor control. ROT 350 Optional tubes:

1.3.

SRO 33 100 / ROT350 (for U.S.A. only) SRO 33 100 / ROT350 SRO 25 50 / ROT350 SRO 09 51 / ROT350 ”SRO” in combination with High Speed Rotor Control

9874 005 16022 9874 005 09022 9874 004 23022 9890 000 64181 9848 600 01101

Mechanical data

1.3.1. Dimensions Mechanical dimensions for wall stand left see

Z--1.1

Mechanical dimensions for wall stand right see

Z--1.2

Mechanical dimensions for OPTIMUS L 50 see

Z--1.3

1.3.2. Movements See

1.4.

Z--1.1 or Z--1.2

Electrical data

1.4.1. Table Operating voltages: 230 V 10% --15% 50/60 Hz Power consumption without tomography: Power consumption with tomography:

appr. 420 W appr. 420 W

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Compatible tube housings:

1--2

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

bucky DIAGNOST TS bdts_1_990

INTRODUCTION AND TECHNICAL DATA

1.4.2. X-ray generator OPTIMUS 50 / 65 / 80 OPTIMUS 50 Power data Mains voltage

Mains frequency Max. mains current -- at 400 V -- at 440 V -- at 460 V -- at 480 V -- at 190 V

OPTIMUS 65

OPTIMUS 80

3 x 400 V ±10% ( 380 V --5%, 415 V +6%) 3 x 400/440/460 V ±10% 3 x 480 V --10% +6% 3 x 190...343 V ±10% with external transformer, max. 50 kW (option) 49 ... 61 Hz 145 A 135 A 125 A 120 A 300 A

190 A 180 A 170 A 160 A --

230 A 215 A 210 A 205 A --

35 A 100 A at £ 240 V

50 A

50 A

Mains resistance -- at 400 V -- at 440/460 V -- at 480 V

£ 300 mW £ 350 mW £ 400 mW

£ 200 mW £ 240 mW £ 300 mW

£ 200 mW £ 240 mW £ 300 mW

--

£ 300 mW

£ 240 mW

£ 180 mW

Fuse protection (slow–blow)

at 480 V, valid for DOD only

Unit supply Max. output power Operating data Tube current Tube voltage mAs product Exposure time Exposure sequency Noise level Heat dissipation Interfacing option for....

230 V / 400 V, max. 5 A 50 kW

65 kW

80 kW

1.0 ... 650 mA

1.0 ... 900 mA

1.0 ... 1100 mA

40 ... 150kV in steps of 1kV or % 0.5 ... 850mAs 1ms ... 6/16s £ 12 exp./s £ 40 dBA 500W door contact, external radiation warning indicator

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Philips

bucky DIAGNOST TS

bucky DIAGNOST TS bdts_1_990

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

1--3

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Philips

INTRODUCTION AND TECHNICAL DATA

1.5.

bucky DIAGNOST TS

Power supply

1.5.1. Type of power supply For 50 kW X-ray generator a three-phase system is needed. The three-phase system must be balanced, i.e. the external conductor voltages against earth must be about equal, see sketch. For reasons of operational safety and compliance with international rules, operation on three-phase mains that are floating or unbalanced to earth (e.g. with one phase earthed instead of the neutral point) is impermissible unless adequate adaptation measures have been taken! Check that the sequence of phases is correct corresponding to the designations L1, L2, L3. L1 3 phase WYE

L2

X-ray installation

-- Working range 400 ... 480V measured between phases: L1 -- L2 / L2 -- L3 / L1 -- L3

L3 (N)

L1 3 phase DELTA, balanced earth or floating X-ray installation

-- Working range 400 ... 480V measured between phases: L1 -- L2 / L2 -- L3 / L1 -- L3

X-ray installation

-- Working range 400 ... 480V measured between phases: L1 -- L2 / L2 -- L3 / L1 -- L3

L2 L3

L1 3 phase DELTA, grounded

Philips

L2 L3

1--4

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

bucky DIAGNOST TS bdts_1_990

Philips

bucky DIAGNOST TS

INTRODUCTION AND TECHNICAL DATA

1.5.2. Calculating the mains resistances If possible, the sum of R1 , R2 and R3 should be smaller than the RXG required. With higher internal mains resistances the generator output is reduced respectively. Note that the cross section of lead 25mm2.

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

+

R0

l3

L1 L2

must not exceed

R1 +

R2

l1

+ R3 l3

L3

< RXG MEX 100 L1 L2 L3 357J94

-- R0 designates the mains resistance on the distributor transformer. -- R1 is dependent upon the length of lead l1 between distributor transformer and main distributor and upon the cross section selected, so: R1 = l1 x RCu

RCu from table

-- R2 is estimated to be 15mΩ EMERGENCY--OFF switch Earth--leakage circuit breaker Fuse

4mΩ 5.5mΩ 5.5mΩ

ü ý þ

15mΩ

-- R3 is dependent upon the length of lead l3 between main distributor and wall junction box and upon the cross section selected, so: R3 = l3 x RCu

RCu from table

The resistances already consider the go and return line so that the calculation can be based on simple cable lengths. Copper cross section [mm2 ]

1.6.

Resistance RCu [mW/m]

16

2.19

25 35

1.4 1.0

50

0.7

70

0.5

95

0.38

120 150

0.30 0.24

Earth-leakage circuit breaker

To be provided between fuse and X-ray installation depending upon local regulations. Siemens earth-leakage circuit breaker FSJ 50 A Rated fault current 30mA Rated current 50A

Philips

Connection terminals for wire cross sections of up to 25mm2.

bucky DIAGNOST TS bdts_1_990

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

1--5

Philips

INTRODUCTION AND TECHNICAL DATA

1.7.

bucky DIAGNOST TS

Emergency-OFF device

To be provided depending upon local regulations. There are 2 possibilities: 1. All the EMERGENCY-OFF buttons are connected in series and looped into the starting circuit (12V DC) of the generator.

1.8.

Heat emission

Table: X-ray generator:

2.

Environmental conditions

2.1.

Table

530W 500W (the generator is air cooled using natural convection)

According to PMS standard UXW 13600 class C1 (indoor temperature controlled) during regular work: -- Ambient temperature:

+ 10 °C ... +40 °C

-- Relative air humidity:

20% ... 80%

-- Air pressure:

700 ... 1100hPA

during storage / transport: -- Ambient temperature:

--25 °C ... +70 °C

-- Relative air humidity:

5% ... 100%

-- Air pressure:

700 ... 1100 hPA

2.2.

X-ray generator

according to PMS standard UXW 13600 class C1 (indoor temperature controlled) during regular work: -- Ambient temperature:

+10 °C ... +40 °C

-- Relative humidity:

0% ... 90%, no condensation

-- Air pressure:

700 ... 1100hPA

during storage / transport: -- Ambient temperature:

--20 °C ... +50 °C Note: Electrolytics require less than 40 °C for storage life.

-- Relative air humidity:

0% ... 95%, non--condensing

-- Air pressure:

700 ... 1100hPA

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

2. The EMERGENCY-OFF circuit acts on an external main contactor which switches off the power before it is fed into the generator.

1--6

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

bucky DIAGNOST TS bdts_1_990

Philips

bucky DIAGNOST TS

INTRODUCTION AND TECHNICAL DATA

3.

Measurements and weights

3.1.

Pre-installed system

Table / column (standard) Weights (kg) Box 1 Contents: manuals, table, table top, rails, SPS, set of cable, covers, tube, control handle, column, set of tomo devices (option)

without shipping box

870

750

with shipping box

270 x 97 x 196

without shipping box

265 x 89 x 190

X-ray generator 50 kW Weights (kg) with shipping box

Dimensions (cm) (L/W/H)

without shipping box

with shipping box

Box 1 Contents: generator cabinet, operating panel

226

178

210 x 82 x 84

Box 2 Contents: H.V. generator 1-tube version

100

73

77 x 67 x 80

without shipping box

195,5 x 55 52 43 x 33,5 x 6

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

with shipping box

Dimensions (cm) (L/W/H)

bucky DIAGNOST TS bdts_1_990

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

1--7

3.2.

bucky DIAGNOST TS

Transport data

3.2.1. Box dimensions Weight Gross

870kg

Loading

816kg

Net

750kg 870kg

Dimension (L x W x H) Box

270cm x 97cm x 196cm

Pallet including loading

265cm x 89cm x 190cm

Sub-pallet Sub-pallet

140cm x 89cm 125cm x 89cm

min 170cm width of roadway for gross pallet

3.2.2. Roadway

min 115cm width of roadway for sub-- pallet

1

1 Roadway for the gross pallet Canting while moving

min 170cm

min 170cm width of the floor

2 Roadway for the sub--pallets Canting while moving

min 115cm width of the floor

2 min 115cm

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Philips

INTRODUCTION AND TECHNICAL DATA

1--8

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

bucky DIAGNOST TS bdts_1_990

Philips

bucky DIAGNOST TS

INTRODUCTION AND TECHNICAL DATA

3.2.3. Unpacking The following steps should be done by four persons. Unpacking

0.5 h 0.5 h

s Dismantle the box, remove the foil.

0.5 h 0.5 h

s 2 Remove from the column the box with documents and distance tool.

1

2 Documents and distance tool

s 3 Erect the column. -- 3a Move the column approximatly 15cm to foot direction of the column. -- 3b Turn the column base about 90°. The column must be ratched in the cam.

3c

3b 90_

-- 3c Lift up the column and place it.

3a

Lay corrugated paper or something similar on the floor in the area where the column is to be moved. Apply extreme care when handling the connection field.

s Remove the wooden carrier where the column was positioned by unscrewing the colored screws.

3

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

s 1 Remove the rails from the pallet.

bucky DIAGNOST TS bdts_1_990

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

1--9

Philips

INTRODUCTION AND TECHNICAL DATA

bucky DIAGNOST TS

s 4 Remove the box including cables from the table.

The pallet is deviable

s 5 Remove the optional tomo devices from the pallet.

6

4 7

5

s Remove the wooden items by removing the colored screws.

7

4 Cables

Note The gross pallet can be splitted in two sub-pallets. s 7 Remove the table from the pallet. Use the shoulder straps. Do not lift the table by means of the table top frame.

8

Table top delivered in the pallet bottom

s 8 After unscrewing the sub–pallets from the pallet you get access to the table top. Remove the table top. s Dispose of the package locally.

4.

Scope of delivery

4.1.

”Fixed table” 9848 600 00061 – bucky DIAGNOST TS with floating table top

-- Table, including column and X-ray tube carriage -- Set of floor fixing material incl. dowels for standard concrete floors -- X-ray generator OPTIMUS L 50 including operating panel -- Complete set of cables incl. plugs at both sides -- Grid -- X-ray tube -- Collimator -- Control handle

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

s 6 Remove the tube assembly from the pallet.

1--10

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

bucky DIAGNOST TS bdts_1_990

Philips

bucky DIAGNOST TS

4.2.

INTRODUCTION AND TECHNICAL DATA

Height adjustable 9848 600 00071 bucky DIAGNOST TS with floating table and elevator

-- Table, including column and X-ray tube carriage -- Set of floor fixing material incl. dowels for standard concrete floors -- X-ray generator OPTIMUS L 50 including operating panel -- Complete set of cables incl. plugs at both sides

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

-- Grid -- X- tube -- Collimator -- Control handle

4.3.

Tomography 9848 600 00081 bucky DIAGNOST TS with floating table and elevator and tomo extension

-- Table including tomography attachment, column and X-ray tube carriage -- Set of floor fixing material incl. dowels for standard concrete floors -- X-ray generator OPTIMUS L 50 including operating panel -- Complete set of cables incl. plugs at both sides -- Grid -- X-ray tube -- Collimator

4.4.

Options

4.4.1. bucky DIAGNOST TS -- wall mounting of operating panel

9890 000 02451

-- stand for operating panel

9890 000 02441

-- Amplimat measuring chamber

9890 000 01612

-- extension main rail

4512 590 30491

-- interface for medio CP / SZ3

4512 590 30351

-- high speed rotor control

9848 600 01101

-- 3phase extension

9848 600 00122

-- SRO 33/100 / ROT 350

9874 005 09022

-- SRO 25/50 / ROT 350

9874 004 23022

-- SRO 09/51 / ROT 350

9890 000 64181

-- SRO 33/100 / ROT 350/HEW (only for U.S.A.)

9874 005 16022

-- bucky unit non sensing left

9848 500 20303

-- bucky unit non sensing right

9848 500 35703

-- manual collimator T303 for TS

9848 600 00172

-- manual collimator for TS (HHS)

9848 600 02092

-- APR

9848 600 00201

-- OPTIMUS 50 (standard) three phase -- OPTIMUS 65

three phase

-- OPTIMUS 80

three phase

-- OPTIMUS 30

single or three phase

bucky DIAGNOST TS bdts_1_990

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

1--11

-- OPTIMUS 40

bucky DIAGNOST TS

single or three phase

-- X-ray tube RO 17/50

9874 001 16002

-- grid 36/08 110F TS

9848 600 00821

-- grid 40/12 110F TS

9848 600 00831

-- H.V. cable, lengths = xx m

9806 402 6xx02

-- set of cables generator--column 14 m

9848 600 01871

-- set of cables generator--table 20 m

9848 600 01901

-- set of cables table--column 20 m

9848 600 01911

-- set of cables table--column 25 m

9848 600 01921

-- TS control grip tomo and laser

9848 600 02081

-- TS control grip standard and laser

9848 600 02071

-- table top Getalit PUR (2.4m)

9890 010 02194

-- table top Getalit (2m)

9890 010 81371

4.4.2. bucky DIAGNOST VE / VT -- grid 36/8 110F

9860 834 90111

-- grid 36/8 140F

9860 834 90141

-- grid 36/8 130F

9860 836 90131

-- grid 36/8 140F

9860 836 90141

-- grid 36/8 180F

9860 836 90181

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Philips

INTRODUCTION AND TECHNICAL DATA

1--12

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

bucky DIAGNOST TS bdts_1_990

Philips

bucky DIAGNOST TS

5.

Site planning

5.1.

Planning data

INTRODUCTION AND TECHNICAL DATA

The preliminary planning work is restricted to the selection of the suitable room. The outfit of the room has to be equipped with the conduits and pipes required and, in cases where the loading capacity of the bearing area plus weight of patient insufficient, the floor preparation for the table. Where DIN specification are not applicable the minimum wall clearances should be determined in compliance with the national regulations concerned.

5.2.

Determining the site of installation

The choice of the suitable room should be based on the layout plan. For detailed information see Z--3.1, Z--3.2, Z--4.1, Z--4.2

5.3.

Floor fixing

The table is fixed by means of 4 bolts and the column rails are fixed by means of 4 + 2 bolts (ext. rail left or right pos.) also. Use e.g. Liebig safety dowels S12 on compound plaster floor concrete class Bn 150 or Upat--PSZ anchor dowels type B or S12 on compound plaster floor concrete class Bn 250 or other national requirements. The mounting points on mounting surface have to be suitable for tensile strength of F ³ 2500N each. In case of ”computer floors” (stilted floors) an appropriate understructure or staybolts have to be used. For detailed information see Z--5

5.4.

Loading capacity of floor

Surface load table and column = 750kg plus 135kg weight of patient. Surface load X--ray generator = 250kg For detailed information see Z--5

5.5.

Height of the room

For detailed information see Z--1.1 or Z--1.2

5.6.

Electrical data

The bucky DIAGNOST TS is supplied incl. all necessary cables incl. plugs with required free length between the main components like X--ray generator and table. For detailed information see Z--6.1 and Z--7, Z--8, Z--50

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

For detailed information about loading area see Z--5.

bucky DIAGNOST TS bdts_1_990

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

1--13

Philips

INTRODUCTION AND TECHNICAL DATA

6.

bucky DIAGNOST TS

Tools

-- Groof key set for column base and transversal tube carriage

4512 590 30362

-- 25mm aluminium for AMPLIMAT adjustment -- chip-ejector for service 2422 487 89772 -- Multimeter -- Mirror (about 30 x 40 cm) -- Standard set of tools -- Fork lift -- ESD antistatic kit -- Hammer drill with 12mm carbide drillbits. -- Spirit level 1m -- Service-PC -- IBM-compatible, 640 kB RAM, 3.5” floppy disk drive, with 1 serial port -- Sub-D Adapter Kit 9/15/25 pole -- OPTIMUS L 50 utility program for generator programming supplied on a floppy disk with the generator. -- PC-Hardkey to use the OPTIMUS L 50 utility program. -- Mains resistance measuring instrument. -- Collimator adjustment tool code No: 4522 980 31521 -- Tomo test tool Code No.: 8112 108 57929 -- PC-XRGSCOPE Cable Code No.: 4512 130 56931 for OPTIMUS generator -- Rail distance tool 4512 590 30501 (included in the delivery)

7.

Installation materials required

To be ordered from the service department via PMS Hamburg: -- Wall junction box for X--ray generator 4512 103 75385 inclusive connection block (25 mm2) for mains supply and connection block (10 mm2) for unit supply. -- Relay for radiation warning indicator 4512 100 45231 1 interface relay with a floating contact (230 V, 1 A) is included in the scope of delivery for the generator. -- Grease – Molykote DX

4512 590 30321 (included with the delivery)

-- Set of three colours

4512 590 30031 (included with the delivery)

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

-- Spring balance 0 ... 200N

1--14

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

bucky DIAGNOST TS bdts_1_990

7.1.

INTRODUCTION AND TECHNICAL DATA

Cables

Cables supplied with the system. Cable length must be ordered via catalogue depending from outfit of the room. H.V. cables

9806 402 6xx02

plugs:

75kVp federal standard female connectors O3/O3

length:

6 ... 20m in steps of 2 m

capacity:

155pF/m

diameter:

16.5mm

The cable length is indicated at the 9th and 10th digits of the code No. Thermal contact cable 2-wire screened for one excess temperature switch

4512 100 66151

10-wire screened for additional supervision like temperature alarm switch, buzzer, selection indicator.

0722 215 19005

Stator cable 3 x 1.31 mm2 , screened

0722 215 02054

AMPLIMAT cable with Sub-D (15pole) 12 m 16 m 20 m 24 m

9890 000 01721 9890 000 01731 9890 000 01741 9890 000 01751

AMPLIMAT cables 9803 507 0xx02 with 3 PLUS plugs at both ends can be connected in the generator via the following adapter for each cable: Adapter for AMPLIMAT cable

4512 108 09041

The generator includes five adapters, 3Plus--Sub-D (15pole) Operating desk cable set

10m 20m 30m

9890 000 02411 9890 000 02421 9890 000 02431

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Philips

bucky DIAGNOST TS

bucky DIAGNOST TS bdts_1_990

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

1--15

Philips Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Philips

Philips Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Philips

Philips Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Philips

Philips Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Philips

Philips Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Philips

Philips Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Philips

Philips Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Philips

Philips Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Philips

Philips Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Philips

Philips Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Philips

Philips Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Philips

Philips Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Philips

Philips Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Philips

Philips Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Philips

Philips Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Philips

Philips Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Philips

S XA C

Philips Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

VE

M9f

6J2

M6f

6J3A

supply

J1

D9m

M12f

6J3B

communication

J3

D15m

D9f

7A2:J4

AEC

J5

D9m

interface

J2

D25f

3TS1 22 3TS1 23 3TS1 24 3TS1 36 3TS1 37 3TS1 51 3TS1 52 3TS1 53 4TS2 4 4TS2 5

22 23 24 36 37 51 52 53 4 5

VP

basic earth lekage circuit breaker L1

switch controller by emergency switch X23 X22

mains fuses X23

X7 X8

PMA

emergency switch

L1 N(L2) 4

Power ON bucky TS

7X11 7

L2 (N) X22

(door contact)

8

7X11

10

da

X13 X14

power supply

230V AC

L1/N

SNX16

M6

supply 4TS2:3

(bk)

4TS2:1

(rd)

4TS2:2

(wt)

XA

tube housing U V W

XA 1 1 2 3

db

stator cable, screened

dc

temperature switch

XF 7020

H.V. cables

XA 1

1 2 3

4TS2:4 4TS2:5 9HT1

O3 plugs

X-ray tube D15m 7A2:J1

SA

X1

D15f

dd

amplimat chamber

PHX1

df

release

SZ1X1

dd

amplimat chamber

VPX4

D15f

df

release

VPX1

D9f

dk

control

Ost./Schr.

bucky examination unit D15m 7A2:J2

SA

X3

bucky wall stand UB1X20 00141z63

Philips A3 00-09-27

E

PMA

bucky DIAGNOST TS C 2000 Philips Medical Systems ALL RIGHTS RESERVED

(00.0)

M12 SZ1X11

M12

bucky DIAGNOST TS, VE OPTIMUS 30 single phase Standard version Connection diagram 1Z-6.3

Philips Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Philips

Philips Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Philips

Philips Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Philips

S

Table

SN X16

4,00 mm2 4,00 mm2 4,00 mm

2

4,00 mm2

4,00 mm2 4,00 mm2

6,00 mm2

T1 T2 61 Z1

VA SA (right) UB

6,00 mm 2

2,50 mm

U

2

4,00 mm 2

UE

A1/A3 00-07-04 Kö

UF

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Philips

XA

2,50 mm

2

4,00 mm

XA

2

LA

bucky DIAGNOST TS Earthing diagram bucky DIAGNOST TS C 1997 Philips Medizin Systeme ALL RIGHTS RESERVED

(00.0)

1Z-10

Philips Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Philips

bucky DIAGNOST TS

INSTALLATION

Philips

INSTALLATION TEXT Safety regulations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2--1

1. 1.1. 1.2.

Transport data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Box dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Roadway . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2--2 2--2 2--2

2. 2.1.

Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Removal of column base from column . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2--3 2--4

3.

Mounting the rails . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2--7

Tomo option only . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2--7

Coarse Adjustment Tomo option only . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.1. Insertion of column to long rail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.2. Installation of tube arm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.3. Final alignment of the column (tomo only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.4. Installation of bucky table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.5. Installation of Tomo pivot tower (tomo only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.6. Installation of bucky-device extension (tomo only) . . . . . . . . . . . . . . . . . . . . . . . . . . 3.7. Installation of position control switches (tomo only) . . . . . . . . . . . . . . . . . . . . . . . . . 3.8. Installation of OPTIMUS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.8.1. Installation of bucky DIAGNOST TS with tomo and Medio CP/CP-H . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.8.2. Programming of APR data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.8.3. Programming tube select at control handle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.9. Installation of tabletop . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.9.1. Installation of tabletop in small rooms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.10. Setting to work . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.11. Adjustment procedure for tomo . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.11.1. Adjustment of the table level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.11.2. Correction for tomo layer height = fine adjustment under X-ray condition . . . . . . . 3.11.3. Adjustment of tomo layer height indication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.11.4. Setting the display language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.11.5. Adjustment of the tomo axis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.11.6. Alignment for center line = center of bucky . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.11.7. Checking drive and position control switches . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.12. Tomo adjustment under exposure conditions (tomo only) . . . . . . . . . . . . . . . . . . . . 3.12.1. Measurement of X-ray zone . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.12.2. Symmetry of tomo axis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.12.3. Tomography resolution test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.12.4. Tomo layer height . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.13. Final work . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.14. Acceptance test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.15. Handing over to customer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.16. Installation for tomo bar parking frame . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

2--8 2--9 2--9 2--13 2--14 2--16 2--17 2--17 2--19 2--19 2--20 2--20 2--21 2--21 2--22 2--25 2--25 2--25 2--26 2--27 2--28 2--29 2--29 2--31 2--31 2--32 2--33 2--34 2--34 2--35 2--35 2--36

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2--0.1

bucky DIAGNOST TS INH

 1999

(99.0)

Philips Medical Systems ALL RIGHTS RESERVED

2--0.1

27

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Philips

INSTALLATION

Safety regulations Only technically qualified personnel who has received the appropriate instructions and training for radiography equipment in medical applications and in particular for the installation and commissioning of the bucky Diagnost TS are allowed to install, maintain and service this equipment. Therefore, instructions in this chapter are considered as a supplementary guideline for correct installation of the bucky Diagnost TS and are kept very briefly. Furthermore, all personnel involved must be instructed on safety precautions with respect to radiography and the related high voltage devices as well as to all other electrical low-voltage equipment. All relevant standards and regulations must be complied with. The local service organisation is responsible for the safety and training of personnel and prevention of material damage.

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Philips

bucky DIAGNOST TS

bucky DIAGNOST TS bdts_2_990

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

2--1

bucky DIAGNOST TS

1.

Transport data

1.1.

Box dimensions

Weight Gross

870kg

Loading

816kg

Net

750kg 870kg

Dimension (L x W x H) Box

270cm x 97cm x 196cm

Pallet including loading

265cm x 89cm x 190cm

Sub–pallet Sub–pallet

140cm x 89cm 125cm x 89cm

min 170cm width of roadway for gross pallet

1.2.

min 115cm width of roadway for sub-pallet

Roadway

1

min 170cm

1 Roadway for the gross pallet Crossing while moving

min 170cm width of the floor

2 Roadway for the sub--pallets Crossing while moving

min 115cm width of the floor

2 min 115cm

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Philips

INSTALLATION

2--2

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

bucky DIAGNOST TS bdts_2_990

Philips

bucky DIAGNOST TS

2.

INSTALLATION

Unpacking

The following steps should be done by four persons. Unpacking 0.5 h

0.5 h

0.5 h

0.5 h

0.5 h 0.5 h

1. Remove the rails from the pallet. Note For transportation in insufficient rooms only. See 2.1 ”Removal of column base from column”.

Unpacking

The column is supplied with an inserted counterweight. For transport, the counterweight and the transverse arm are secured by a screw. Do not remove these screws before the installation of the tube arm.

0.5 h 0.5 h

1

2 Documents and distance tool

3c

3b 90°

3a

Loosen the transport screws and safety belts on the pallet. 2. Remove the box with documents and distance tool from the column. 3. Set up the column. -- 3a Move the column approximatly 15cm to foot direction of the column. -- 3b Turn the column base about 90°. The column must be ratched in the cam. -- 3c Lift up the column and place it.

Lay corrugated paper or something similar on the floor in the area where the column is to be moved. Apply extreme care when handling the connection field. 3 S Unscrew the colored screws from the wooden carrier on which the column was positioned and remove it.

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

S Dismantle the box, remove the foil.

bucky DIAGNOST TS bdts_2_990

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

2--3

Philips

INSTALLATION

bucky DIAGNOST TS

4. Remove the box including cables from the table.

The pallet is deviable

5. Remove the optional tomo devices from the pallet.

6

4

6. Remove the tube assembly from the pallet.

7

5

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Remove the wooden items by removing the colored screws. 7

4

Note The gross pallet can be splitted in two sub-pallets.

Cables

7. Remove the table from the pallet. Note Use the shoulder straps. Do not lift the table by means of the table top frame.

8

Table top delivered in the pallet bottom

8. After unscrewing the sub-pallets from the pallet you get access to the table top. Remove the table top. S Dispose the package locally.

2.1.

Removal of column base from column

This section is needed if the column is too large to enter the X-ray room. Tools required Groofkey for column base

PMS Code No. 4512 590 30362

Spring scale 200 N Imbus Key 5mm Molykote DX

PMS Code No. 4512 590 30321

Isotrophy alcohol 1. Dismantle base and column 1.1. Measure the breakaway force with a spring scale when the column is in the the lock position. It should measure 110--150 N. Place the hook of the spring scale on the screw head.

Philips

1.2. Repeat point 1.1, only column must be in free position. Should measure £ 60 N.

P1 2--4

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

bucky DIAGNOST TS bdts_2_990

Philips

bucky DIAGNOST TS

INSTALLATION

1.3. Release the pressure on the Beta grid (measure the distance “X” between nut and screwhead. Dimensions needed by the adjustment of reassembling).

P2

1.4. Remove the 2 Imbus (M5) screws and plate (see P 3). 1.5. Loosen the safety screws (pos. 1) on the groove nut (see P6). 1.6. Remove groove nut and steelring.

M5

1.7. Carefully remove column base from column: S Remove the teflon ring. S Clean it and grease it with Molykote DX.

P3

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

X

bucky DIAGNOST TS bdts_2_990

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

2--5

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Philips

INSTALLATION

bucky DIAGNOST TS

1.8. Clean running surface and thread with Isopropyl alcohol (see P4). For transport cover running surface and thread.

3

Note The stop pin pos. 3 can be removed in case of the movement of the column is impaired. (This stop pin will not be needed any longer.)

2

P4

2. Assembling of column base and column 2.1. Grease the running surface of the column and column with Molykote (PMS No. 4512 590 30321). 2.2. Insert the teflon ring. 2.3. Assemble the column base onto the column. (The column must be laid down when assembling the column base. Take extreme care of thread.) 2.4. Assemble the steelring and groove nut. Hand-tighten the groove nut. P5 2.5. Rotate the column base for functional check. 2.6. Tighten groove nut until the inner steelring (spring) is compressed. Tighten safety screws pos. 1 (see P 6).

1

2.7. Adjust the beta grid (see P 2 and step 1.3). 2.8. Repeat step 1.1 and step 1.2. The breakaway force should measure 110--150 Nm and the free running force should measure ≤ 60N. Adjust the force by loosening, resp. tightening the groove nut. P6

2--6

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

bucky DIAGNOST TS bdts_2_990

Philips

bucky DIAGNOST TS

3.

INSTALLATION

Mounting the rails Before starting, refer to your planning room layout! (see PRB.)

S Depending on the installation of the bucky wall stand (right or left), place the extension rail right or lefthand side to the main rail (see Z--1.1, Z--1.2 and Z--5).

position of extension rail for bucky wall stand left

main rail

S Insert the screw (supplied with system) with a washer and tighten with a spanner. position of extension rail for bucky wall stand righ t

pencil

Tomo option only 2

S Loosen the 4 screws of the tomo switch rail (1) and the end switch cams (2) and remove them.

1

S Place the main rail parallel to the wall.

extension rail

S Mark the 5 fixing points on the floor.

min 314 mm

S Remove the rail. S Drill the 8 mm holes approximately 75 mm deep in the middle of the marked area. S Insert the 5 anchor dowels (supplied with system) into the holes.

longitudinal axis table

table position for bucky wall stand left

S Move the rail back to the installation position. Level the rail using a spirit level. Use the spacer if necessary. S Insert the screws (supplied with system) with the washers and tighten with a spanner.

extension rail

min. 314 mm

longitudinal axis table

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

S Assemble main rail.

bucky DIAGNOST TS bdts_2_990

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

table position for bucky wall stand right

2--7

Philips

INSTALLATION

bucky DIAGNOST TS

S Place the short rail parallel to the long rail. S Insert the tool (1 Rail distance tool) (supplied with system) to adjust the distance of the rails. S Measure the distance of the rails using a rule.

position for bucky wall stand left

1

extension rail

S Mark the 2 fixing points on the floor.

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

S Remove the rail.

long

S Drill the 8 mm holes approximately 75 mm deep in the middle of the marked area.

short

short

716 mm

S Insert the 2 anchor dowels (supplied with system) into the holes. S Move the rail back to the installation position. S Level the rail using a spirit level. Use the spacer if necessary.

position for bucky wall stand right

S Insert the screw only with washer (supplied with system) to the opposite side where the column is to be inserted. S Tighten with a spanner.

1

extension rail

long 716 mm

short

short

Coarse Adjustment Tomo option only S Align front and back rails (2, 3) to same level using the spirit level. Use additional spacers if necessary. For a good performance with the tomo option it is essential to align front and back rail very precisely. An alignment test will be made after installation of the column.

2

3

2

S Install the tomo switch rail (1). S Insert both end switch cams (2). S Insert screws with washers and tighten with Allan key.

Philips

1

2--8

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

bucky DIAGNOST TS bdts_2_990

3.1.

INSTALLATION

Insertion of column to long rail

S Remove buffer (1) from long rail. S Insert the column into the rail system (2). Apply extreme care when handling the ball bearings and be careful not to damage the microswitches and optical sensor B1 positioned in area 3. S Install buffer (1) to the rail system. S Check the leveling of the column in both directions using a spirit level.

1

S Use the spacers if necessary. S Check whether the column moves easily and smoothly. S Insert the second screw with washers into the short rail.

2

3

S Tighten with a spanner. S Mount the longitudinal center switch UB S17 (pos.3) standard, or position control switch assembly for tomo.

3.2.

2

Installation of tube arm

S Remove transport screw (2) from the counterweight on the front of the column.

1

Loosen transport screw (1) only and move the complete block downwards.

S Pull down the transverse arm and secure the counterweight in its maximum top position by inserting transport screw (2). 2

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Philips

bucky DIAGNOST TS

bucky DIAGNOST TS bdts_2_990

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

2--9

Philips

INSTALLATION

bucky DIAGNOST TS

S Route the cables from the top of the column to the left and right of the column (1). S Open top cover (2) of the transversal tube arm by removing the 4 screws 1 (Allan key, 2,5 mm).

1 1 1

S Place the pallet with transverse arm, including tube housing and collimator, next to the column.

2 1

1

S Lift up the arm and insert it into column (3).

2

3

S Insert stop screw (1). S Remove the transport screw. The tube arm must be easily movable in Z-direction. S Connect the plug UFX38 (2).

1 2

5

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Do not remove the screws (2) on top of the cover. These screws are fixing the brake rail!

2--10

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

bucky DIAGNOST TS bdts_2_990

INSTALLATION

S Fasten the cable (4) by using a Ty--Rap.

5

S Connect UFX38 (3) 4

S Install the cover of transverse arm (3).

3

S Install the front cover of transverse arm (5) (see previous page).

6

S Install the cable support into the top of the column.

S Install the wall junction box.

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Philips

bucky DIAGNOST TS

bucky DIAGNOST TS bdts_2_990

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

2--11

Philips

INSTALLATION

bucky DIAGNOST TS

S Route the cables on the column as shown in the figure 1. S Replace top cover of the transversal tube arm. S Connect X20 / X22 on stand control UB1 for standard version.

UB1

Figure 1

** All three items, wall box, column cable support und tube arm, should be located as shown in figure 2 (design and function, especially cable lenghts).

*1 = 0.6 meter

*1 Location for ceiling suspension junction box. This box is not included in the system. Please, order this item locally as an option. (PMS Artikel No. 9848 500 26201 ceiling suspension box for CS, this item is not included and must be ordered optional).

**

Note Max. room height with ceiling suspension box = 3.0 m

Figure 2

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

S Connect X20 / X21 / 22 / 23 / 24 on stand control UB1 for tomo version.

2--12

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

bucky DIAGNOST TS bdts_2_990

3.3.

INSTALLATION

Final alignment of the column (tomo only)

S Move the column to the left end position. S Check the alignment of the rails by moving the column smoothly to the other end. The column must remain steadily upright over the whole length. If necessary, adjust the level of the front rail with spacers as described in Section “Mounting the rails”.

(6) (5)

The alignment can be done either with the spirit level or with the following method:

Final adjustment S Apply a vertical line to the wall on the right--hand side, e.g. with a colored adhesive tape, Check the vertical alignment while moving the column along the rail. If corrections should be necessary: S Add adjustment plates (0) under back (1) or front rail (2), or, for vertical correction (5), S adjust the ball bearing (3). Three different holes are available. For easy access move the column to the left or right end position. If the column is not vertical (6), the correction can be made with an eccentric ball bearing (4). (1)

(4)

(3) (0)

(2)

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Philips

bucky DIAGNOST TS

bucky DIAGNOST TS bdts_2_990

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

2--13

Philips

INSTALLATION

3.4.

bucky DIAGNOST TS

Installation of bucky table

Note Four (4) persons are necessary to lift the table. The bucky table is supplied as a complete unit.

1

S Remove the 4 transport screws and push them down.

Do not lift the table by means of the tabletop frame structure.

1

S Insert the shoulder straps into the loops (1). S Lift up the table with the shoulder straps (1). S Remove the pallet and lower the table carefully. S Place the table with rear end towards the short rail. See room layout plan. S Insert screws (1) into the short rail and fit the table to the rail. S Drill two 12mm holes approximately 75mm deep for screws (2). S Insert the 2 anchor dowels (supplied with system) into the holes.

1 2

S Use the spacer if necessary. S Insert screws (2) supplied with system and washers and tighten with a spanner.

table position for bucky wall stand right

S Level the table in both directions using a spirit level. S Use the spacer if necessary. S Insert screws (2) supplied with system and washers and tighten with a spanner.

1 2

S Install the foot-switch assembly.

table position for bucky wall stand left

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

S Remove the wooden blocks from the longitudinal side of the pallet and place them at the rear of the pallet.

2--14

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

bucky DIAGNOST TS bdts_2_990

Philips

bucky DIAGNOST TS

INSTALLATION

S Remove the red screws (2) of the transport securing device of the tabletop carrier on both sides. S Remove the tie-wraps form the Bucky device (1)

2

1

i

Figure 1

S Remove the cassette tray from the bucky.

1

S Move the tabletop to its maximum front position (if it is already mounted).

3

S Loosen 2 screws at the rear (1). S Tilt the tabletop unit gently as shown in figure 2.

4

S Secure the tabletop by fixing the lever (2) via the screw (3).

2

S Remove the tie-wraps (4) from the arm braces on both sides (elevation table only). Figure 2

Always remove cassette tray before opening the tabletop. Tilting the tabletop with inserted tray will cause damages to both table and cassette tray. Tilting the table top beyond the upright position can cause damages to the cover of the table base.

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

2

bucky DIAGNOST TS bdts_2_990

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

2--15

Philips

INSTALLATION

3.5.

bucky DIAGNOST TS

Installation of Tomo pivot tower (tomo only)

S Thread cable W12 (laser cable) through the center holes of table base and pivot tower base. S Insert 4 screws together with washers and tooth lock washer (supplied with system) from the table inside and mount the tomo pivot tower. S Close the tabletop. S Secure tabletop with the 2 screws at the rear. S Reinstall the cassette tray.

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

S Position tomo pivot tower towards the corresponding mounting holes at the rear side of the table.

2--16

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

bucky DIAGNOST TS bdts_2_990

Philips

bucky DIAGNOST TS

3.6.

INSTALLATION

Installation of bucky-device extension (tomo only)

Do not support or level the free hanging end of the bucky-device extension while tightening the screws. It will self-align to the film level in a slightly tilted position when the screws are installed. S Tighten the screws.

3.7.

Installation of position control switches (tomo only)

Pull corner iron with position control switches out of the column base (they are parked in the column base for transportation)

2

3 1

4

S Insert screws with washers (supplied with system) into the column base and fit the position control switches as shown in (1). S Tighten screws with Allan key.

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

S Mount the bucky-device extension and insert the 4 screws with washers and tooth lock washers (supplied with system).

bucky DIAGNOST TS bdts_2_990

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

2--17

INSTALLATION

Philips

Move column with extreme care in order not to damage or shear off the microswitches and the magnetic sensor while passing the cams of the tomo switch rail. S Adjust microswitches for center position and safety end: -- The microswitches (2) (center position) and (3) (safety limit switch) must trigger when passing their respective cams, as shown in the figure below. S Adjust the proper alignment of the magnetic measuring sensor: Slowly move the column along the tomo switch rail: -- the magnetic sensor B1 (1) (angle position) must not touch any of the cams. The distance between magnetic sensor (1) and cam should be 1mm. The functional test will be performed after power up (see Section “Checking drive and position control switches”). The detection of any cam position regarding 40° / 20° / 8° is indicated with the red LED located as shown in the figure.

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

bucky DIAGNOST TS

2--18

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

bucky DIAGNOST TS bdts_2_990

Philips

bucky DIAGNOST TS

3.8.

INSTALLATION

Installation of OPTIMUS The following items are different from the OPTIMUS generator: - Firmware TDL file - PCB low speed rotor control

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

For further information contact the Service Innovation Department in Hamburg. S Place the X-ray generator cabinet in the position shown in the room layout. S Place the keyboard in the position shown in the room layout. For detailed information see manual “Installation of OPTIMUS generator”. S Check cable connectors on all PCBs (Column Controller UB1, Table Controller SZ1). S Route the cables from the wall connection box of the column to the X-ray generator cabinet. S Route the cables from the bucky table to the X-ray generator cabinet. Location at generator site HT cable Stator cable Exposure release Amplimat cable Tomo release Thermo cable Earth Connection Power suply 230V

1 WAX 1 EZX 22 Table EZX 31 BWS 1 WAX 11 EZX 3

3.8.1. Installation of bucky DIAGNOST TS with tomo and Medio CP/CP-H For bucky TS with tomo and Medio CP/CP-H, an Interface board SZ3 is needed. The SZ3 board must be ordered separately (= option): Code No.: 4512 590 30351

Bucky TS & Tomo

Medio CP Exposure rel. X1

X2

SZ3

X1

SZ1

Philips

see Z1-4--2

bucky DIAGNOST TS bdts_2_990

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

2--19

Philips

INSTALLATION

bucky DIAGNOST TS

3.8.2. Programming of APR data The generator is pre-programmed at factory site. For RGDV programming see 2Z--2.0. If special APR assignements are needed, please, proceed as follows:

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

S Optimus / Program / Human interface / Predefined assignments -- without vario focus: AR 46 ENG.TDL AR 46 DEU.TDL AR 46 FRA.TDL AR 46 ESP.TDL

= = = =

English German French Spanish

-- with vario focus for RO 1750: AR 46 V ENG.TDL = English AR 46 V DEU.TDL = German AR 46 V FRA.TDL = French AR 46 V ESP.TDL = Spanish -- and select for Bucky TS RGDV 1 “Bucky GR” RGDV 2 “TOMO BTC” or “TOMO BTS” RGDV 3 “Wall Std GR” RGDV 4 “Free GR”

3.8.3. Programming tube select at control handle S The tube select signal, e.g. for tube one is selected, is coming from generator as shown in diagram Z1--4 with signal RGDV via release decade 1WAX1.

1WA

-- Exposure release for RGDV 2 is also coming over the same 1WAX1 decade. -- BWS is programmed via 1WAX3 for exposure release.

5

-- Free cassette is programmed via 1WAX4. S Modification of tube one select for RGDV 3/4: Usually there is no tube select indication when RGDV 3/4 are selected. If the tube select indication at the control handle is needed, link X1:5 to X3:5 / X4:5 on 1WA.

X1

X2

X3

X4

Philips

In any case the tube 1 select will be present.

2--20

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

bucky DIAGNOST TS bdts_2_990

Philips

bucky DIAGNOST TS

3.9.

INSTALLATION

Installation of tabletop

In rooms with sufficient space on the longitudinal side of the table:

1

S Install the tabletop from the longitudinal side.

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

S Remove end cover (1) and stoppers from one side of the tabletop. S Push the tabletop along the transverse carriage.

Figure 1 Apply extreme care when handling the ball bearings.

S Test the easy and smooth movement of the tabletop, if not correct adjust the eccentric bearing bolts. S Install the stoppers and end cover (1), see figure 1.

3.9.1. Installation of tabletop in small rooms In rooms without sufficient space on the longitudinal side, the tabletop base unit should be rotated to install the tabletop. S Loosen screws (1).

1

S Remove the locking rings from bolts (2). S Remove bolts (2). S Rotate the tabletop base unit until the tabletop can be inserted. S Remove end cover (1) and stoppers from one side of the tabletop. S Insert the tabletop along the transverse carriage. S Rotate the tabletop unit as required.

2

Philips

S Install bolts (2) and secure via locking rings. S Install the stoppers and end cover (1). S Fit screws (1).

bucky DIAGNOST TS bdts_2_990

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

2--21

bucky DIAGNOST TS

3.10. Setting to work The system is supplied fully prefabricated by the factory. Before switching on the X-ray generator: Switch off the auxiliary mains breaker for table. For detailed information see manual installation of the OPTIMUS X-ray generator. S Check the power supply for the table on the PCB SN. Always remove cassette tray before you open the tabletop. Tilting the tabletop with inserted tray will cause damages to both table and cassette tray.

1 3

2

Tilting the tabletop beyond the upright position can cause damages to the cover of the table base. Figure 1 For access to the table control unit: S Remove the cassette tray from the bucky. S Move the tabletop to its maximum front position. S Loosen 2 screws at the rear (1). S Tilt the tabletop unit as shown in figure 1. S Secure the tabletop by fixing the lever (2) via the screw (3). The table power supply must be 230V, 50Hz or 60Hz. Note There is no programming for 50 or 60Hz. The table can only operate with 230V (and not with 110V).

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Philips

INSTALLATION

2--22

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

bucky DIAGNOST TS bdts_2_990

Philips

bucky DIAGNOST TS

INSTALLATION

S Follow the setting-to-work procedure described in the manual OPTIMUS X-ray generator XRG 90.

Programming

1

S Select Registration Devices

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

-- RGDV1 ( = registration device of the place for tables 1) -- RGDV2 ( = registration device free exposure) -- RGDV3 ( = registration device wall stand) -- RGDV4 ( = registration device free exposure)

After finishing the described setting-to-work procedure of the OPTIMUS X-ray generator, check the functions below on the table. After switching on the auxiliary mains breaker for table S Check all functions on the table: -- Green LED on the power pack is on. -- The programmed language. -- Function of foot-switch assembly. -- Function of the rocker switches “table up/down” on both sides. -- Function of end-switch upper position: Motor must stop before mechanical end position is reached. -- Function of end-switch lower position: Motor must stop before mechanical end position is reached. -- Function of safety switch: Apply about 20 kg of manpower against the tabletop during table down movement. S Check the strength of upward table movement. S Check whether the range of movement is 270 mm. S Check the movement of the bucky unit. S Check the functions of the column: -- Function of catches in all directions.

Philips

-- Function of light on the collimator.

bucky DIAGNOST TS bdts_2_990

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

2--23

Philips

INSTALLATION

bucky DIAGNOST TS

S Check the functions of the bucky unit: -- bucky exposure delay S Check delay time of grid release with test exposure time = 4 ... 6 msec. There must not be visible any grid line.

The X-ray tube has been adjusted at factory site. Do not ADAPT the X-ray tube. In case the generator was not powered on during a longer period -- e.g. more than 2 months --, we recommend to recondition the X-ray tube. See OPTIMUS manual, section Installation, 10.1 Tube conditioning. S Check coincidence of X-ray beam and light. -- Make a test exposure. S Check alignment of the stand. S Check alignment of the bucky wall stand.

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Note:

2--24

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

bucky DIAGNOST TS bdts_2_990

Philips

bucky DIAGNOST TS

INSTALLATION

3.11. Adjustment procedure for tomo 3.11.1. Adjustment of the table level S Check whether the table-top is precisely levelled. S Turn on light and laser beam.

-- Raise the table to a position somewhat below maximum level. The distance from the laser beam to front and to rear frame of the table-top must be equal, i.e. the laser beam must be plane-parallel to the table-top. -- If not, turn laser diode slightly with adjusting screw (1). S Raise table-top precisely to laser beam level according to figure 1. Table-top position must be slightly below the maximum level that can be reached. If not, adjust laser diode height (2). If not correct:

Figure 1

S Open and secure the table-top. Set the upper lift limit SNV--S7 switch so that the center of the pivot point meets the 0-level of table-top.

3.11.2. Correction for tomo layer height = fine adjustment under X-ray condition If a tomo test exposer (with a test phantom) deviates from the tomo height, the correction can be done as follows: Solution 1 -- valid for table serial No. =500 If the 0mm layer height = the top of table-top differences to the 0 point of the coupling element. The correction can be done on the axis in both directions. Solution 2 -- valid for table serial No. =499 Correct the tomo layer height = 0 mm via the table height SNVS7 switch = table height for the most upper table position. After the SNVS7 correction follow up with section 2.13.2 Adjustment of tomo layer height.

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

S Adjust laser beam levelling:

bucky DIAGNOST TS bdts_2_990

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

2--25

Philips

INSTALLATION

bucky DIAGNOST TS

3.11.3. Adjustment of tomo layer height indication S Set the service switch (1) on the column control board to service. The following message will appear on the control panel display: Service mode

1 S Press the key on the control panel (1). This will start the calibration menu. Now follow the instructions on the display: Service mode Move to h=0

Figure 2 S Raise the table-top up to laser beam level (corresponding to the minimum objective plane level 0mm).

3

S Press the key on the control panel (2). The position is now stored in the PLC and the display shows: Service mode

4

T

Move to h=200

T

1 2

S Lower the table-top to 200 mm, use scale meter (corresponding to the maximum objective plane level 200 mm). S Press the key on the control panel (2). The position is now stored in the PLC and the display shows: Service mode Diagnosis: 0

Figure 3

You can always quit the calibration menu without saving the values by pressing the key again.

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Diagnosis: 0

2--26

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

bucky DIAGNOST TS bdts_2_990

Philips

bucky DIAGNOST TS

INSTALLATION

S Leave the service switch on the column control board in position service to enter the display language.

3.11.4. Setting the display language

3

4

Service mode Diagnosis: 0 S Press the key on the control panel (3). The following message will appear on the control panel display:

T

T

1

Service mode /V _._ Language: english

2

( V_._ means Softfware--version ) S Press the key (4) to select another laguage. Repeat until the message appears in the desired language. S Press the key on the control panel (3) again to quit the menu.

Figure 4

S Set the service switch on the column control board back to operation.

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

S The service switch should be in mode service.

bucky DIAGNOST TS bdts_2_990

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

2--27

Philips

INSTALLATION

bucky DIAGNOST TS

3.11.5. Adjustment of the tomo axis S Prerequisite: Alignment of standard equipment has been completed. S Select tomo on control handle.

S Move transverse arm to normal position.

4

3

S Insert the coupling bar into central and lower swivels (1, 2). S Fit the coupling bar to the column (3). The locking bolt (4) will be pushed upwards (which will unlock the ß-movement axis of the transverse arm).

2 1

1. Check that the tomo bar activates the tomo sensor switch UB S15 when inserted. 2. Adjustment of ball bearing to tomo bar: Ball bearings must be engaged in direct contact to the tomo coupling bar. So no mechanical clearance is visible.

B

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

S Move column, table-top and bucky tray to the center position (the center position is reached if the 4 LED-indicators on the control handle are lit).

2--28

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

bucky DIAGNOST TS bdts_2_990

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Philips

bucky DIAGNOST TS

INSTALLATION

S Bucky device, light beam localiser and coupling bar must be aligned in a vertical plane. S If necessary, align the system by adjusting central swivel (1), lower swivel (2) and position control bar.

2 1

3 3.11.6. Alignment for center line = center of bucky If the center line 3 is not in center of bucky proceed as follows. S The tomo coupling bar is y centered vertical (4). S Open screws (5) on tomo coupling bar and rotate/twist the tube support arm. S The center line will still be in center of bucky tray (as shown in pos. 3).

3.11.7. Checking drive and position control switches For parameter setting and control please refer to the operating manual. Procedure: S Move the column carefully from its center position towards the left and right end positions and check the following items: -- the magnetic switch does not touch the position cams.

Philips

5

UB S17 S20

-- the micro switches S17/S20 are not damaged or sheared off when passing a cam. -- the magnetic switch is triggered at each angle position cam and the end position cams. This can be checked by observing the red LED on the magnetic switch B1. bucky DIAGNOST TS bdts_2_990

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

2--29

bucky DIAGNOST TS 1

S Test the safety limit switches: Move the column manually to the left and right limit switch positions. The limit switch must be triggered (I/O--LED 35 on tomo controller (PLC) lits) and the system be switched off. S Initiate a tomo test cycle on the control panel: Select mode with 40°/4s. Press the key until the cycle is completed and check the following items: -- the column moves first to the left (1). -- the column moves at constant speed and without jerks (otherwise check: cable routing, rail alignment, dirt on rail surface, spring pressure, driving axle). -- the eccentric disk operates correctly (otherwise check coupling motor (M 3) and position switch UB S16 or both springs are ”fixed too tight”. Therefore, check the distance from screw head to the nut as shown in the figure. The distance as shown must be 15 ± 0.5mm.

Coupling motor

-- the driving wheels grip firmly (otherwise increase spring pressure).

M3

Tomo motor S16

15 ± 0,5mm

M2

Pivot point

-- the column reverses automatically at the end position on each side. -- the bucky device moves smoothly and without jerks (otherwise check alignment of the coupling bar). -- the position control switches (magnetic switch and microswitches) operate correctly, see first step. -- the position of the coupling bar is vertical after completion of a cycle. -- the distance between control switch B1 and tomo position cams should be 1mm. -- If the sequence is aborted, check the function of the position control switches (see first step). -- If the sequence is aborted due to timeout (error code 30, see Chapter ”trouble shooting”), increase the motor supply voltage by switching to the next higher transformer tap (see schematic diagram Z1--2). S Repeat the tomo test cycle with 40°, 20° and 8° sweeps, with high and low speed settings. Again, check whether any anomalities occur during the test.

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Philips

INSTALLATION

2--30

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

bucky DIAGNOST TS bdts_2_990

Philips

bucky DIAGNOST TS

INSTALLATION

3.12. Tomo adjustment under exposure conditions (tomo only) 3.12.1. Measurement of X-ray zone

a

Two exposures are to be taken on the same film.

1

b

First exposure = Bucky -- Small focus

2

-- SID

= tomo SID

-- X-ray area

= 200 mm x 200 mm

-- Exposure values

=

40 kV / 6.3 mAs

Second exposure = Tomo S Set the parameters for exposure 2 (with tomo): -- Small focus -- SID

= tomo SID

-- X-ray area

= 150 mm x 200 mm

-- Tomo height

= 130mm

-- Exposure values

= 40° / 50 kV / 12.5 mA

-- Tomo time

= 2s

S Take the exposure in tomo mode (tomo on) and determine the following values: -- a =

_____mm

-- b =

_____mm

S Check it these values meet the following conditions: -- D a, b =

£ 10 mm

-- S a, b =

£ 80 mm

S If the above conditions are not fulfilled, check whether the free movement of the a--drive and bucky module is impaired. S If under-voltage is indicated on the generator control board, the motor will supply voltage by switching to the next higher transformer tap (see schematic diagram Z1--2).

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

S Set the parameters for exposure 1 (without tomo):

bucky DIAGNOST TS bdts_2_990

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

2--31

Philips

INSTALLATION

bucky DIAGNOST TS

3.12.2. Symmetry of tomo axis S Position the lead plate - 100mm above the X-Ray cassette.

Lead plate

The X-ray cassette ist positioned on the table-top. S Collimate the X-ray field (1) as far as the ”hole” will be exposed during the whole tomo sweep.

100mm

S Take an exposure in center position (bucky mode) = 70 kV / 10 mAs S Take an exposure in tomo mode. 70 kV 2 sec 5 mA 40° .

X-ray cassette

Table top

The result should look like as shown in the figure: -- c » d, i.e. the generator should switch off shortly before the system. -- No undervoltage indication should appear on the generator.

c

d 3

4

S Check whether these values meet the following conditions: -- |c|--|d| < 5mm. S If the result is not symmetrical, proceed as follows: -- In case of overvoltage lower the tomo speed by selecting a higher resistance with jumpers on board SZ2 = fine adjustment, i.e. reduce the supply voltage of the drive or reduce the voltage at transformer T2 side for slow speed = coarse adj. 17.5V / 15.5V / 13.5V ore available. For high speed progr. 31V / 33V or 37V. -- In case of unsymmetry (c < d) increase the tomo speed by selecting a lower resistance with jumpers on board SZ2., i.e raise the supply voltage of the drive

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

¯∅

1

2--32

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

bucky DIAGNOST TS bdts_2_990

Philips

bucky DIAGNOST TS S Programming of symmetriy (please fill out the table) -- The following programming is selected: U On T2 High speed

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

INSTALLATION

Low speed

value for c and d in mm:

R=O ON SZ2

37V

O

33V

O

31V

O

c = .....................mm ................ O

d =.......................mm

17.5V O 15.5V O

c = .....................mm ................ O

d =.......................mm

13.5V O Remarks: Þ

= coarse adj.

= fine adj.

tested with 40° 40°

4 sec and 2 sec

S If necessary, readjust the tomo axis (see Section 3.13.4 Adjusting the tomo axis). S Repeat the above test.

+/-- 5 mm

3.12.3. Tomography resolution test S Take an exposure with the TOMO test phantom with following parameter: TOMO angle

TOMO time

TOMO height

40°

4 sec

50 mm

40°

2 sec

50 mm

40°

4 sec

100 mm

40°

2 sec

100 mm

lp/mm

with

small

Philips

-- Minimum resolution = 2.32 focus.

Actual value Resolution lp/mm

bucky DIAGNOST TS bdts_2_990

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

2--33

bucky DIAGNOST TS

3.12.4. Tomo layer height (Use TOMO test phantom) S Take the following test exposures with inclined bucky tray. Tomo angle

Tomo time

TOMO layer height -default = nominal value

Measured TOMO layer height = actual value

Deviation (+/-- 3mm max.) = nom.--actual value

40°

4 seconds

45 mm

mm

mm

40°

4 seconds

130 mm

mm

mm

40°

4 seconds

200 mm

mm

mm

If the deviation is more than +/-- 3mm, please repeate the table calibration procedure as follows: 1) Raise the table hight to the effective (phantom measured) zero-level. In the display you will see the amont of the deviation. 2) Follow the instruction part 3.13.2 Calibration of the level sensor. Note that your table stays on level zero, do not move the table to zero as requested in the display instruction. 3) Adjust the laser beam.

3.13. Final work S Install all rail covers.

1

S Connect the earth wiring to all covers.

1

S Install all column covers. S Close the X-ray generator cabinet. S Clean all covers with a mild cleaner e.g. soapsuds. S Check the labelling (See drawing Z--3).

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Philips

INSTALLATION

2--34

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

bucky DIAGNOST TS bdts_2_990

Philips

bucky DIAGNOST TS

INSTALLATION

3.14. Acceptance test See national regulations as e.g. Röv / DOD / VA

3.15. Handing over to customer on

( date)

...................

to

( customer)

................................................

by

(system engineer)

................................................

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

The modified system is handed over

bucky DIAGNOST TS bdts_2_990

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

2--35

bucky DIAGNOST TS

3.16. Installation for tomo bar parking frame Tomo bar

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Philips

INSTALLATION

Wall mounting

65 cm ± 2 cm

Philips

10 cm ± 2 cm

Floor 2--36

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

bucky DIAGNOST TS bdts_2_990

Desk :

OPTIMUS OPTIMUS bucky DIAGNOST R CD TS E1999 Philips Medical Systems

ALL RIGHTS RESERVED

(99.0)

] RGDV2 [ ] RGDV3 [ ] ] ] RGDV6 [ ] RGDV7 [ ] ]

For WBX11 : 9 -- 10 (ready) und 1 -- 2 (format size correction contact)

Bucky RGDV :

RGDV1 [ RGDV4 [ RGDV5 [ RGDV8 [

-- Disable time override: -- Tube power factor: -- kV steps: -- mAs steps: -- mA steps: -- time steps: -- Density steps: -- Density correction (6% steps): -- Underexposure display: -- Tube overload protection: Bucky / Scopo 1WB / Decade Bucky 1 (WBX11) :

-- Used for tomo:

Data Set B:

Room: Tube: Release circuit number: Enable handswitch at generator desk: Syncmaster present: Exposure switch type: Bucky format density correction: Cone density correction: Dose measurement input: Dose measurement sensor: Tomo movement (No break after exposure end): Release delay: Mounted radiographical controller: Release circuit adaption unit: Mounted tomo extension: Medium II format kV correction (dose equiv. steps): Medium II format density correction (6% steps): Medium II format mAs correction (6% steps): Small II format kV correction (dose equiv. steps): Small II format density correction (6% steps): Small II format mAs correction (6% steps):

Data Set A :

----------------------

Name :

-

bdts_2z2_0_990

NONE NONE

Bucky RGDV Bucky RGDV

no 100 single 25% 25% 25% 12% 0 yes on WAX12

no

no enable none none none 0 0 0 0 0 0

1 1 1 yes no double step 0 0 none

RGDV4

RGDV6

Tomo time 4 :

Tomo time 3 :

Tomo time 2 :

Tomo time 1 :

RGDV7

Tomo time 8 :

Tomo time 7 :

Tomo time 6 :

Tomo time 5 :

Bucky / Tomo 1WA \ Tomo time [ s ] :

RGDV5

Philips

Tomo RGDV -- switch related RGDV2 Time setting for input at WA X21:1 ......8 Tomo mode switch : X11:3 SL_XG_TO / Bucky RGDV : X11:1 Format + :10 Bucky ready / Tomo RGDV : X11:1 Format + :5 Tomo ready switch related X11:3 -- --> Bucky -- Tomo remote switchover RGDVs

RGDV1

enable

Tomo mode switch Bucky RGDV -- switch related

no 100 single 25% 25% 25% 12% 0 yes on WAX11

no

no enable none 1WA none 0 0 0 0 0 0

1 1 1 yes yes double step 0 0 EZX31

RGDV3

no no 100 100 single single 25% 25% 25% 25% 25% 25% 12% 12% 0 0 yes no on on Bucky / Tomo 1WA : Decade Bucky 1/2

yes

yes enable none 1 WA 1 WA 0 0 0 0 0 0

no enable none 1 WA none 0 0 0 0 0 0 no

1 1 1 yes yes double step 0 0 none

RGDV2

1 1 1 yes yes double step 0 0 EZX21

RGDV1

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

A4 99--12--21 Wa/Re

Philips

RGDV programming OPTIMUS Rel. 3.x

2Z--2.0

RGDV8

1

17a, b, c, d e or

Philips Medical Systems

PHILIPS

PHILIPS

type 9848 600 02641 s/n 99.00.123 TS Family PEI

12a b

LR 37323

US

Philips Medical Systems

230V

50/60Hz

5A

CISPR11-Classification:

Group 1 - Class B

Development and Manufacturing Centre

Röntgenstrasse 24

D-22335 Hamburg/Germany

D-22335 Hamburg/Germany

or

5

type 9848 600 0023. s/n 99.00.123 tube stand

PHILIPS

PHILIPS

INHER. FILTRATION 0,3 Al / 80 IEC 522/1976

Made by Ralco

®

D-22335 Hamburg/Germany

PHILIPS

PHILIPS

Development and Manufacturing Centre

BEAM LIMITING DEVICE type 9848 600 0209. s/n 99.00.123 manual collimator TS (HHS)

C

®

0051

Philips Medical Systems Röntgenstrasse 24 D-22335 Hamburg/Germany

Made by Ralco

INHER. FILTRATION 0,3 Al / 80 IEC 522/1976

CERTIFICATION This Product complies with the Performance Standard under the Radiation Control for Health and Safety Act of 1968, applicable at date of manufacture. Manufactured: September 1992

20

17a

0051

17b 4 12a

5

Philips Medical Systems

PHILIPS

type 9890 010 0219. s/n 99.00.123 tabletop getalit PUR TH

Development and Manufacturing Centre

Röntgenstrasse 24 D-22335 Hamburg/Germany

CERTIFICATION This Product complies with the Performance Standard under the Radiation Control for Health and Safety Act of 1968, applicable at date of manufacture. Manufactured: August 1999

17c

6a b (9, 7, 6 c d)

13a b, 20 10 or

11

PHILIPS

or

or Laser CONTROL GRIP!

1 15, 16 2 3

12b

PHILIPS

Philips Medical Systems

PHILIPS

type 9890 010 8137. s/n 99.00.123 tabletop 2m f.b. table TS

Development and Manufacturing Centre

Röntgenstrasse 24 D-22335 Hamburg/Germany

CERTIFICATION This Product complies with the Performance Standard under the Radiation Control for Health and Safety Act of 1968, applicable at date of manufacture. Manufactured: August 1999

17d

or

4

6a

type 9848 600 0109. s/n 99.00.123 control grip TS

6b type 9848 600 0208. s/n 99.00.123 control grip tomo/laser TS

18

19 or

K

ratio

B

(only for USA)

TUBE HOUSING ASSEMBLY

SA LR37323 C

CERTIFICATION This Product complies with the Performance Standard under the Radiation Control for Health and Safety Act of 1968, applicable at date of manufacture. type 9806 300 7120. s/n 12345

150 kV

X-RAY TUBE HOUSING

ROT 350 10

- X-RAY EQUIPMENT CLASSIFIED BY UNW RITERS LABORATORIES INC ® WITH RESPECT TO ELECTRICAL FIRE SHOCK AND MECHANICAL HAZARDS ONLY 641 B

type 9806 300 7120. s/n 12345

150 kV

X-RAY TUBE HOUSING

ROT 350 10

- X-RAY EQUIPMENT CLASSIFIED BY UNW RITERS LABORATORIES INC ® WITH RESPECT TO ELECTRICAL FIRE SHOCK AND MECHANICAL HAZARDS ONLY 641 B

150 kV

X-RAY TUBE HOUSING

ROT 350 10

type 9806 300 7120. s/n 12345

150 kV

X-RAY TUBE HOUSING

ROT 350 10

type 9806 300 7120. s/n 12345

150 kV

X-RAY TUBE HOUSING

ROT 350 10

D-22335 Hamburg/Germany

150 kV

0.6 IEC 336 1.3 IEC 336

SRO 33 100

PHILIPS

2,5 Al IEC 522/1976

PHILIPS Made in Germany

Philips Medical Systems Development and Manufacturing Centre

Röntgenstrasse 24 0123

D-22335 Hamburg/Germany

150 kV

0.6 IEC 336 1.0 IEC 336

SRO 25 50

PHILIPS

2,5 Al IEC 522/1976

PHILIPS Made in Germany

Philips Medical Systems Development and Manufacturing Centre

Röntgenstrasse 24 0123

TUBE HOUSING ASSEMBLY

CERTIFICATION This Product complies with the Performance Standard under the Radiation Control for Health and Safety Act of 1968, applicable at date of manufacture.

PHILIPS Made in Germany

Röntgenstrasse 24

type 9890 000 6419. s/n 12345 X-Ray TUBE

SA LR37323 C

PHILIPS

Development and Manufacturing Centre 0123

TUBE HOUSING ASSEMBLY

CERTIFICATION This Product complies with the Performance Standard under the Radiation Control for Health and Safety Act of 1968, applicable at date of manufacture.

0.6 IEC 336 1.3 IEC 336

RO 17 50

Philips Medical Systems

type 9806 206 1110. s/n 12345 X-Ray TUBE

SA LR37323 C

D-22335 Hamburg/Germany

150 kV

2,5 Al IEC 522/1976

TUBE HOUSING ASSEMBLY

SA LR37323 C

type 9806 300 7120. s/n 12345

- X-RAY EQUIPMENT CLASSIFIED BY UNW RITERS LABORATORIES INC ® WITH RESPECT TO ELECTRICAL FIRE SHOCK AND MECHANICAL HAZARDS ONLY 641 B

Röntgenstrasse 24

type 9806 206 7000. s/n 12345 X-Ray TUBE

CERTIFICATION This Product complies with the Performance Standard under the Radiation Control for Health and Safety Act of 1968, applicable at date of manufacture.

- X-RAY EQUIPMENT CLASSIFIED BY UNW RITERS LABORATORIES INC ® WITH RESPECT TO ELECTRICAL FIRE SHOCK AND MECHANICAL HAZARDS ONLY 641 B

Development and Manufacturing Centre 0123

TUBE HOUSING ASSEMBLY

CERTIFICATION This Product complies with the Performance Standard under the Radiation Control for Health and Safety Act of 1968, applicable at date of manufacture.

Made in Germany Philips Medical Systems

type 9806 200 6000. s/n 12345 X-Ray TUBE

SA LR37323 C

PHILIPS

PHILIPS

2,5 Al IEC 522/1976

D-22335 Hamburg/Germany

150 kV

0.3 IEC 336 1.0 IEC 336

SRO 0951

PHILIPS

2,5 Al IEC 522/1976

PHILIPS Made in Germany

Philips Medical Systems Development and Manufacturing Centre

Röntgenstrasse 24 0123

type 9806 206 7010. s/n 12345 X-Ray TUBE

D-22335 Hamburg/Germany

150 kV

0.6 IEC 336 1.2 IEC 336

SRO 33 100

6d

or

X-RAY GRID STREUSTRAHLENRASTER GRILLE ANTIOFFUSANTE Lp/cm

17e

type 9848 600 0206. s/n 99.00.123 control grip tomo TS

- X-RAY EQUIPMENT CLASSIFIED BY UNW RITERS LABORATORIES INC ® WITH RESPECT TO ELECTRICAL FIRE SHOCK AND MECHANICAL HAZARDS ONLY 641 B

9

Philips Medical Systems Development and Manufacturing Centre

Röntgenstrasse 24 D-22335 Hamburg/Germany CERTIFICATION This product complies

fo(cm)

with the Performance Standard under the Radiation Control for Health and Safety Act of 1968 applicable at date of manufacture.

MANUFACTURED: AUGUST 1999

type 9848 600 0207. s/n 99.00.123 control grip laser TS

6c

CAUTION LASER RADIATION DO NOT STARE INTO BEAM MAX. OUTPUT < 1 mW EMITTED WAVELENGTH 650 nm FDA CLASS II LASER PRODUCT

LASER RADIATION DO NOT STARE INTO BEAM CLASS 2 LASER PRODUCT IEC 825-1: 1993 MAX. OUTPUT < 1mW EMITTED WAVELENGTH 650 nm

7 according to the release status

10

11

Philips Medical Systems Development and Manufacturing Centre

Röntgenstrasse 24

Remove cassette tray and move tabletop in the front position before tilting the table

5 Kg

bucky DIAGNOST TS C Philips Medical Systems

(00.0)

8

D-22335 Hamburg/Germany

CAUTION LASER RADIATION DO NOT STARE INTO BEAM MAX. OUTPUT < 1 mW EMITTED WAVELENGTH 650 nm FDA CLASS II LASER PRODUCT

CERTIFICATION This product complies

LASER RADIATION DO NOT STARE INTO BEAM CLASS 2 LASER PRODUCT IEC 825-1: 1993 MAX. OUTPUT < 1mW EMITTED WAVELENGTH 650 nm

with the Performance Standard under the Radiation Control for Health and Safety Act of 1968 applicable at date of manufacture.

MANUFACTURED: AUGUST 1999

9

September 1999

BEAM LIMITING DEVICE type 9848 600 0017. s/n 99.00.123 manual collimator TS

C

13b

Development and Manufacturing Centre

Röntgenstrasse 24 D-22335 Hamburg/Germany

Development and Manufacturing Centre

Röntgenstrasse 24

D-22335 Hamburg/Germany

Manufactured :

Philips Medical Systems

PHILIPS

PHILIPS

Development and Manufacturing Centre

Röntgenstrasse 24

September 1999

13a

PHILIPS

Philips Medical Systems

Manufactured :

PHILIPS

type 9848 600 0008. s/n 99.00.123 TOMO Equipment

September 1999

4

Development and Manufacturing Centre

Röntgenstrasse 24 D-22335 Hamburg/Germany

Manufactured :

Philips Medical Systems

PHILIPS

PHILIPS

September 1999

14

Philips Medical Systems

A1/A3 00-02-24 Schr.

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Development and Manufacturing Centre

Röntgenstrasse 24

16

type 9848 500 2030. s/n 99.00.123 bucky device left

Philips

PHILIPS

Philips Medical Systems

Manufactured :

19

SA ® C

PHILIPS

September 1999

18

15

PHILIPS

PHILIPS

type 9848 600 0006. s/n 99.00.123 bucky table fix

Manufactured :

Philips

14

3

type 9848 600 0007. s/n 99.00.123 bucky table elevation

0123

Class I - TYPE B IEC 60601-1

8

2

Development and Manufacturing Centre

Röntgenstrasse 24 D-22335 Hamburg/Germany

bucky DIAGNOST TS Labelling

2Z-10

Product

INTRODUCTION AND TECHNICAL DATA Trouble shooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Safety regulations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Software and PLC faults . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Hardware and tomographic image problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Specific instructions for fault elimination . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . EPROM replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Setting EPROM configuration jumper . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Test exposure with inclined bucky tray . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . PLC input / output allocation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Digital inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Digital outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Indicator LED’s on table control board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Standard version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Signal exchange between generator and table control . . . . . . . . . . . . . . . . . . . . . . Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Additional LED’s for Tomo version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Signals from PLC to table control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Signals from table control to PLC and generator . . . . . . . . . . . . . . . . . . . . . . . . . . . Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3--1 3--1 3--2 3--5 3--6 3--6 3--6 3--6 3--7 3--7 3--8 3--9 3--9 3--9 3--9 3--9 3--9 3--9 3--9

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Philips

1. 1.1. 1.2. 1.3. 1.4. 1.4.1. 1.4.2. 1.4.3. 1.5. 1.5.1. 1.5.2. 1.6. 1.6.1. 1.6.1.1. 1.6.1.2. 1.6.2. 1.6.2.1. 1.6.2.2. 1.6.2.3.

bucky DIAGNOST TS bdts_3_990_inh

 1999

(99.0)

Philips Medical Systems ALL RIGHTS RESERVED

3--0.1

27

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Philips

1.

Trouble shooting

1.1.

Safety regulations

FAULTFINDING

Only technically qualified personnel who have received the appropriate instruction and training for radiography equipment in medical applications and for the bucky Diagnost TS in particular are allowed to install, maintain and service this equipment. It is presumed that all personnel involved with trouble shooting are familiar with fault finding and repair of standard radiography equipment. Therefore only topics related to the process controller (PLC) and tomo specific problems are treated in this chapter. Furthermore, all personnel involved must be instructed on safety precautions with respect to radiography and the related high voltage devices as well as to all other electric low voltage equipment. All relevant standards and regulations must be complied with. The local service organisation is responsible for the safety and training of personnel and prevention of material damage.

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Philips

bucky DIAGNOST TS

bucky DIAGNOST TS bdts_3_990

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

3--1

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Philips

FAULTFINDING

1.2.

bucky DIAGNOST TS

Software and PLC faults

s In order to display software error messages, set the service switch on the column control board to service. The following message indicating an error code will appear on the control panel display: Service mode Diagnosis: nn s Follow the relevant suggestions in the table below to eliminate the disturbance. . If the system is o.k., he following message will appear: s Clear error message by pressing Service mode Diagnosis: 0 s Contact the Help--Desk of your local service organization if the fault cannot be corrected.

M1

M2

40° 20° 8°

UB1

VA2

Tomo mode

bucky

X2

E

Tomo ready

1WA X11

X30

bucky ready

X2

X11

SZ1

X3

UE2

Control handle

Format contact

B1

tomo tomo motor coupling SPS tomo controller

1

8° 20° 40°

SZ1

Table controller

X2

Table controller X1

1 2 3 4 5 6 7 8 9 10

X1

1

RQ_XG_EX XG_RD_EX RQ_XG_PR PO_26V_RE RG_DV_1 GND

X1 Code

Fault description

Possible cause

Suggested remedy

--

Display not active

-- RS232 serial link to control panel disturbed

Check cable and plugs

--

LED ”supply” (yellow) not on Power supply missing

1 2 3 4 5 6 7 8 9 10

X11

E 1WA

X1

Notes

1. Check if power supply is on 2. Check supply cable and connection

--

LED ”Run” (yellow) not on

-- Application program not loaded -- EPROM missing

1. Check EPROM 2. Download application program 3. see error code 104

no error message

Philips

0

3--2

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

bucky DIAGNOST TS bdts_3_990

Philips

bucky DIAGNOST TS

FAULTFINDING

Code

Fault description

Possible cause

1

Communication timeout: no -- Cable to table conanswer from PIC--procestrol not correctly sors on table controller (deconnnected centralized I/O) -- Interface module on PLC defective

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

-- RS485 bus nor correctly connected -- Jumpers on RS485 interface board not set correctly -- PIC--processors or interface module defective -- EPROM defective

Suggested remedy

Notes

1. Check cable and plugs 2. Check position of jumpers on RS485 interface board (must be set to CLOSED, see wiring diagram Z2--6) 3. PLC--reset: s Turn PLC off s Disconnect power supply cable of PLC (on the bottom, to the left of the 3 LED) s Switch PLC on s Wait a few seconds s Reconnect the supply cable

power

4. Check the hardware and replace if necessary 2

Incorrect answer from PIC-- see code 1 processors on table controller

see code 1

30

Tomo sequence timeout (se- -- Tomo motor blocked quence time limit exceeded) -- Tomo motor is not or deficiently engaged

1. Check tomo drive (motor, driving wheel, eccentric disk, see Section 4.12.4 ”Checking drive and position control switches”)

-- Speed too low

-- Motor supply voltage 2. Check motor supply too low voltage: -- Angle and limit posivoltager range: ....... If too tion control switch low, change tap of supply disturbed transformer -- Center position 3. 1. Check alignment of switch disturbed position control switches (see Section 4.12.4 ”Checking drive and position control switches”) 31

Column did not stop in the -- After running too center position after complehigh tion of a tomo cycle, i.e. col- -- Center position umn overran the center posiswitch disturbed tion

1. Check function of column brake 2. Check center position switch (incl. Cable and connections) and center position cam.

Philips

3. Check column stop control 100

PLC system RAM test failure -- PLC hardware fault

1. Replace PLC

101

PLC RAM test failure

1. Replace PLC

bucky DIAGNOST TS bdts_3_990

-- PLC hardware fault

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

3--3

bucky DIAGNOST TS

Code

Fault description

Possible cause

Suggested remedy

Notes

104

PLC Program EPROM test -- EPROM defective failure -- EPROM not correctly installed

1. Replace EPROM

PLC serial interface test fail- -- Interface module deure fective

1. Replace interface module in PLC

2. Check EPROM position (see Section EPROM replacement) -- PLC configuration incorrect 3. Check EPROM configuration (see Section Setting EPROM configuration jumper)

120

-- PLC hardware fault

2. Replace PLC

140

PLC system EPROM test -- PLC hardware fault failure

1. Replace PLC

252

Error flag is set

-- Error in PLC applica- 1. Check program release tion program (must match PLC version and generator type) 2. Replace EPROM

253

Index register overflow

-- Error in PLC applica- 1. Replace EPROM tion program

254

Program timeout

-- Error in PLC applica- 1. Replace EPROM tion program 2. Replace PLC -- PLC defective

255

Software watchdog error or -- Error in PLC applica- 1. Replace EPROM power supply error tion program 2. Replace PLC -- PLC defective

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Philips

FAULTFINDING

3--4

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

bucky DIAGNOST TS bdts_3_990

1.3.

FAULTFINDING

Hardware and tomographic image problems

Fault description

Possible cause

Suggested remedy

Tomo sequence aborted

-- Tomo speed too low

1. See error code 30 in Section Software and PLC faults

-- see error code 30 in Section Software and PLC faults Underexposure (indication on generator control desk)

Tomo speed too high

Mains voltage too high

Objective plane level not correct

Notes

1. Check supply voltage and change transformer tap, if necessary 1. Install transductor or voltage stabilizer (contact service organization to order)

-- Calibration maladjust- 1. Check laser alignment (see ment Section Adjusting table level) -- Laser misalignment

2. Calibrate level sensor (see Section Calibrating level sensor) 3. Check level indicator potentiometer in table base: pinion axle must be rectangular to toothed belt 4. Replace level indicator potentiometer

Resolution too low with large objective plane levels

-- Coupling bar has too much play -- Uneven motion of coupling bar -- Uneven motion of bucky device

1. Check coupling bar (movement, fixations, swivels; see Section Adjusting the tomo axis) 2. Check bucky device: -- bearings -- twisted rails -- dirt on rails -- collision with cables

Generally low resolution, even with low objective plane levels

In addition to the possibil- 1. Make test exposure with ities above: inclined bucky tray (see Section Test exposure with -- Column movement inclined bucky tray) and not properly aligned evaluate result with bucky device -- Column jerks -- Column drive not properly adjusted

2. Make test exposure with grating evaluate result

3. Check tomo drive (motor, driving wheel, eccentric disk, -- Transverse arm yaws see Section Checking drive (spring pressure for and position control switches) b-axis lock too low) 4. Increase spring pressure for -- Exposure time too b-axis lock low or too high 5. Check speed adjustment (exposure time coordination with generator)

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Philips

bucky DIAGNOST TS

bucky DIAGNOST TS bdts_3_990

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

3--5

Philips

FAULTFINDING

1.4.

bucky DIAGNOST TS

Specific instructions for fault elimination

1.4.1. EPROM replacement s Switch off PLC power supply. Protection against static electricity: an earth strap must be connected between the wrist and the base of the column before the EPROM is removed. The EPROM could otherwise be destroyed by static electricity. s Replace EPROM, pins 1 and 2 of the socket (see Figure) must be left unused. s Replace cover and switch power on.

1.4.2. Setting EPROM configuration jumper s Switch off PLC power supply. s Wait 1 minute, then remove the cover. s Set jumper J1 to ”E”. s Replace cover and switch power on. 1.4.3. Test exposure with inclined bucky tray s Place bucky tray on table--top as shown in the figure. s Adjust collimator aperture according to the figure. s Make test exposures in tomo mode and interpret the results.

70

Collimator aper max. 1 mm

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

s Wait 1 minute, then remove the cover.

3--6

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

bucky DIAGNOST TS bdts_3_990

Philips

bucky DIAGNOST TS

1.5.

FAULTFINDING

PLC input / output allocation

1.5.1. Digital inputs Signification when LED ON Operation mode

Service mode

¡0

Tomo on

Calibration

¡1

Tomo test

Confirmation of language/level

¡2

Set tomo angle

Language select

¡3

Set tomo time

Error reset

¡4

Raise table top

Raise table top

¡5

Lower table top

Lower table top

¡6

Light beam localiser ON

¡7

FFA ok

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

EA--0

Operation mode / Service mode EA--16 x-axis center lock (transverse arm extension)

¡ 17

y-axis center lock (column movement)

¡ 18

Service switch ON

¡ 19

spare

¡ 20

spare

¡ 21

spare

¡ 22

spare

¡ 23

spare

Philips

¡ 16

bucky DIAGNOST TS bdts_3_990

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

3--7

bucky DIAGNOST TS

1.5.2. Digital outputs Signification when LED ON Tomo mode

Standard mode

¡ 32

Radiation ON command 1 (redundant)

(permanently on)

¡ 33

Radiation ON command 2 (redundant)

(permanently on)

¡ 34

Unlock b--movement axis

¡ 35

Activate brake y-axis (column movement)

¡ 36

Release brake y-axis (column movement)

¡ 37

Disable center position locks

¡ 38

spare

¡ 39

spare

EA--32

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Philips

FAULTFINDING

3--8

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

bucky DIAGNOST TS bdts_3_990

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Philips

bucky DIAGNOST TS

1.6.

FAULTFINDING

Indicator LED’s on table control board

1.6.1. Standard version 1.6.1.1. Signal exchange between generator and table control H12 (yellow)

Signal XG_RD_EX (Exposure) from generator (LED on = signal low)

H15 (yellow)

Signal RG_DV (device active) from generator (LED on = signal low)

H16 (red)

Signal RQ_XG_EX (Radiation on) to generator (LED on = signal low)

1.6.1.2. Power supply H17 (green)

Supply electronics 24VDC

H18 (green)

Supply light beam localiser 24VAC

H20 (green)

Supply motors 24VDC

1.6.2. Additional LED’s for Tomo version 1.6.2.1. Signals from PLC to table control H1 (green)

Speed: low motor speed

H2 (green)

Dir: motor direction reverse

H3 (green)

On1: motor1 on

H4 (green)

On2: motor2 on

H5 (green)

Laser: objective plane indication on

H6 (green)

Bucky_contr: release bucky brake

H7 (green)

Bucky_start: trigger bucky device

H8 (green)

Tomo_ready: tomo start position reached

H9 (green)

Tomo_mode: tomo mode active

H10 (green)

Watchdog

1.6.2.2. Signals from table control to PLC and generator H11 (yellow)

Signal RQ_XG_PR (Preparation) from generator

H12 (yellow)

Signal XG_RD_EX (Exposure) from generator

H13 (yellow)

Coupling bar: coupling bar inserted to PLC

H15 (yellow)

Signal RG_DV (device active) from generator

H16 (red)

Signal RQ_XG_EX (Radiation on) to generator

1.6.2.3. Power supply H17 (green)

Supply electronics 24VDC

H18 (green)

Supply light beam localiser 24VAC

H19 (green)

Supply motors 12VDC

H20 (green)

Supply motors 24VDC

bucky DIAGNOST TS bdts_3_990

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

3--9

Drawings

Block diagram standard version

Z1--1

Block diagram tomo version

Z1--1.1

Power supply standard version

Z1--2

Power supply tomo version

Z1--2.1

Table control standard version

Z1--3

Table control tomo version

Z1--3.1

Bucky control standard version

Z1--4

Bucky control tomo version

Z1--4.1

Exposure Interface SZ3

Z1--4.2

Stand control standard version

Z1--5

Stand control tomo version

Z1--5.1

Bucky DIAGNOST VE with bucky unit non sensing

Z1--6.1

Stand control tomo version

Z1--7.1

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Philips

bucky DIAGNOST TS

bucky DIAGNOST TS bdts_z01_990

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

Z0--1

Philips

EX1100

stator cable, screened

EZX3

temperature switch

XA

tube assembly

H.V. cable

EG

see OPTIMUS manual

H.V. cable

Laser (option) X41

UE

brakes

EZX21

SX..

X1

X1

UE

PH

X32

X1

field

amplimat chamber

LA

UB X3

X1

X2

format contacts bucky 1

1WAX11

X40

X36

X38

VA

See table control

bucky module

centered

UE1

X20

UF

X22

X23

brakes

X37

light

X2

X11

manual collimator

brakes

tomo drive 1WAX1

release 1

X31

X4

brakes and table height

X31

X22

X24

brakes

see Z1-7

X7

SA

UE2

UB1

X4

brakes

stand control

X21

X30

UB1

control handle

stand control

X9

centered

UB

table control

X10

centered

RS 485

SA

X5

XP1

See table control

footswitch

RS 232

SZ1 X5

EA16/32

stand bucky DIAGNOST TS EA0

SC

M X6

table height

see Z1-7

PLC

PGU

tomo control

SNV see Z1-2 mains 230V~

ENX2100

X8

X16

X5

SN

table height

Upgrade kit Center Laser

UF tomo height

power supply

XA

SNV

UE

LA table bucky DIAGNOST TS PLC

ST

UB

VA

UB1 A3 00-02-23 Ost.

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

amplimat chamber 1

X49 X35

LCD display

amplimat chamber 2

EZX31

X4

X1

PH

PH

SA

amplimat chamber

SZ1 release 2

1WAX3

X1

SA

X2

VA bucky module

E

generator

bucky DIAGNOST TS C Philips Medical Systems

VP wall connection box

(99.0)

bucky DIAGNOST VE

SNV

SC

SN

Block diagram Standard version bucky DIAGNOST TS Z1-1

Philips

EX1100

stator cable, screened

EZX3

temperature switch

XA

tube assembly

H.V. cable EG

see OPTIMUS manual

H.V. cable

Laser (option) X41

UE

brakes X35

LCD display amplimat chamber 1

EZX21

SX..

X1

X49

X1

X32

X1

field

PH amplimat chamber

MEDIO CP release adapter (option) release 1 format contacts bucky 1

1WAX1

X1

X2

X3

X2

X40

X36

X38

VA bucky module

X2

centered

UE1

UF

X22

X23

brakes

X37

light

X20

hand collimator

brakes

tomo drive X1

SZ3 1WAX11

LA

UB

X11

X31

X4

brakes and table height

X31

X22

X24

brakes

see Z1-7

X7

SA

UE2

UB1

X4

brakes

stand control

X21

X30

UB1

control handle

stand control

X9

centered

UB

SZ1 table control

X10

X5

centered

RS 485

SA

footswitch

XP1

X5

EA16/32

SC tomo speed control

stand bucky DIAGNOST TS EA0

M X6

table height

see Z1-7

PLC

PGU

tomo control

1/2

SZ2

SNV

see Z1-2 mains 230V~

ENX2100

X16

X8

X5

SN

Upgrade kit Center Laser

table height

UF tomo height

power supply

XA

SNV

UE

LA table bucky DIAGNOST TS PLC

UB VA

UB1 A3 00-02-23 Ost.

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

UE

amplimat chamber 2

EZX31

X4

X1

PH

PH

SA

amplimat chamber

release 2

1WAX3

X1

SA

X2

VA bucky module

E

generator

bucky DIAGNOST TS C Philips Medical Systems

VP wall connection box

(99.0)

bucky DIAGNOST VE

SNV

SC

SN

SZ1

Block diagram tomo version bucky DIAGNOST TS Z1-1.1

Philips

stator cable, screened

EZX3

temperature switch H.V. cable

EG

H.V. cable

XA

tube assembly

Factory adjusted Z1 ENX2100:L1

X16:1

F11

X16:1.1

L1

L1.1

X16:3

5A slow

mains 230V~ ENX2100:N

X16:2

F12

N

T1 230VAC

30VAC

X8:2

28VAC

1.6A slow

mains filter X16:2.1

F14

F5

X16:4

0VAC

0VAC

X22:04

X20:06

X31:04

X32:4

X1:4 24VAC

H18

26VAC N.1

X11:06

6.3A slow

LA manual collimator

X8:3

X11:08

X20:08

X22:07

X31:07

X22:01

X31:01

X32:3

X1:3 0VAC

5A slow

G1 L1

1

X8:4

N

F6

X11:01

X20:01

X11:12

X20:12

24VAC

X49:2

X1:2

5VDC 0VDC

H17 X31:12

X8:5

center laser (optional)

F2 230VAC

X1:1

+24V

+24V

X22:12

5

X49:1

+24V 4A slow

switching power supply 24VDC/5A

X8:6

6.3A slow 24VAC

H20 0VAC

E

SN

generator

A3 99-11-23 Ost.

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Philips

EX1100

power supply

0VAC

X8:8

0VAC

SZ

table control

UB

stand control

UE

control handle

Power supply Standard version bucky DIAGNOST TS bucky DIAGNOST TS C Philips Medical Systems

(99.0)

Z1-2

Philips

stator cable, screened

EZX3

temperature switch H.V. cable

EG

H.V. cable

XA

+24VDC

tube assembly

GND

PLC

tomo control

Factory adjusted F14 ENX2100:L1

X16:1

F11

X16:1.1

L1

L1.1

X16:3

T1

6.3A slow

30VAC

230VAC

mains 230V~ ENX2100:N

X16:2

F12

X8:2

28VAC

5A slow

mains filter X16:2.1

N

F5

X16:4

0VAC

0VAC

X20:06

X22:04

X31:04

X32:4

X1:4 24VAC

LA

H18

26VAC N.1

X11:06

6.3A slow

X8:3

X11:08

X20:08

X22:07

X31:07

X22:01

X31:01

X32:3

hand collimator

X1:3 0VAC

5A slow

G1 L1

1

X8:4

N

X20:01

X1:1

X49:2

X1:2

+24V

+24V

5VDC 0VDC

H17 X22:12

X31:12

X20:12

X8:5

center laser (optional)

F2 37VAC

X49:1

+24V

X11:12 5

T2

X11:01

F6 4A slow

switching power supply 24VDC/5A

X8:6

6.3A slow 31VAC

34VAC

H20

230VAC 31VAC 17.5VAC

F1 X8:7

6.3A slow 15.5VAC

15.5VAC

H19

0VAC 13.5VAC

E

generator

0VAC

SN

A3 99-11-23 Ost.

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Philips

EX1100

power supply

UB

X8:8

stand control

UE

control handle

0VAC

SZ

table control

Power supply tomo version bucky DIAGNOST TS bucky DIAGNOST TS C Philips Medical Systems

(99.0)

Z1-2-1

X10:3

+24V

2

Philips

X10:2

transverse brakes 2

D13

SAH5 SAS5

footswitch

SCS4

SAS21

1

X4:09

X7:7

X4:08

A

2

3

X7:5 X5:03

2 1

X5:02

ENABLE

X4:11

+24V

X4:12

U7

+5V ENABLE

1

X7:3

1b

X7:4

X4:02

D4

SAS2 release bucky brake

U6 &

2 down

DOWN

DOWN

1

U7

K1

X9:2

K2

X4:04

D10

X4:05

D9

X9:3

DK 1.2 X4:03

1

X9:4

DOWN

2

2

D22

ENABLE

X6:05

SAS3 bucky centered

4

U1

2

2 down 1

1

Q11

1

ENABLE

ENABLE

X6:03

1

X9:1

Q10 +24V

U6

SNVS8

Q6

SWGND

+5V

+24V

+24V

ENABLE

JP2B DOWN

1

U6

JP1

D5

2

longitudinal brake

4

1a

SNVS7

SAY3

1

U6

X7:2

SAS22

1

B

& 1b

X4:06

X7:6

SAY2b

+24V

X4:10 1a

B

SAH2

A

bucky centered

SAY1 brake bucky module

X6:04 X6:07

4

safety

1

2 table on 1

X6:06 X6:02

K4

+24V

+24V

Q3 Q7

Q9 F2 24VAC

X8:6

Q9

JP3

6.3A slow

from power supply see Z1-2

X8:2

GL1

0VAC

down

F5

3

brake table motor

SNVY7

table height on

4

off 5

X6:08

3

+

K4

M

-

SNVM1

K2 X6:09

3

5 24VAC

H18

0VAC 0VAC +24VDC

X11:06

X20:06

X11:08

X20:08

X11:01

X20:01

X11:12

X20:12

X11:09

X20:09

X8:3 0VAC

X8:4

F6

+24V

U2 +24V

4A slow

H17

+24V

DC/DC

+5V

X8:5

SZ1

Table up Table down

C Philips Medical Systems

4

up 24VAC

6.3A slow

0VDC

bucky DIAGNOST TS

K1

5 X8:8

2

X6:11 4

24VAC

1

24VAC

H20

0VAC

X6:10

to stand see Z1-5

SNVS9

A3 99-11-26 Ost.

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

1

X4:07

footswitch inactive

Philips

SAY2a

(99.0)

k1/k2 Off k1/k2 ON

k4 ON k4 Off

table control

SNVS7

lowest range = 540mm

highest range = 810mm SNVS8 safety switch ~ 200N without load! SNVS9 SAS21/22 Table up / down

Table control Standard version bucky DIAGNOST TS Z1-3

X11:02

data -

16

Philips

RS485 X11:03

data +

Dk3

k3

2 X10:2

transverse brakes

X4:07 2

D13

X4:06

SAY2b

1

SAH5 footswitch inactive 1

SAS5

footswitch 1

1

Jumper

SCS4

U6

3

X4:09

X7:6

polyswitch RXE185

1

B A

2 X4:08

X7:7

SAY3

1

longitudinal brake

4 3

X7:5 X5:03

2 1

X5:02

+ +24V

K5

1

X4:11

1b

X4:12

4 &

+24V

+5V

1a

ENABLE

1

X7:3

1b

X7:4

D5

X4:02

ENABLE

JP2B DOWN

Q6

SWGND

Q10

+5V

0.5

X9:1

0.5

1

SAS2

+24V

+24V

U6

U6 X6:03

1

1

D4

up

ENABLE

ENABLE

DOWN

DOWN

&

1

Dk1.2 2 down 1

release bucky brake

K1

1

1

X9:2

K2

X4:04

SZ2

2

+24V

SNVS8

1

1

on 4 1

2

off 5

X6:04

X11:07

3

X11:10

4

safety

1

Tomo Motor

2 table on

K4

1

X6:06 X6:02

+24V

31VAC

31VAC

H19

fast slow

F2 17.5VAC

X8:6

4

from power supply see Z1-2

24VAC

down 4 up

GL1

0VAC

down 4

F5

up 24VAC

6.3A slow

0VAC +24VDC

X8:4

K1 3

2

polyswitch RUE700

table motor

SNVY7

A

X5:09 +5VDC

table height on 4 off 5

X6:08

3

+

K4

M

sensor coupling bar

SNV H4 laserdiode

500 E table potentiometer SNV P1

SNVM1

K2 X6:09

3

+24V

+24V

H17

brake bucky module

-

U2 DC/DC

SAY1

SNV S15

X5:10

0VAC

+24V

bucky centered

X5:08

D

24VAC

F6 4A slow

X5:04

SAH2

A

X5:07

5

X11:06

X20:06

X11:08

X20:08

X11:01

X20:01

X11:12

X20:12

X11:09

X20:09

+5V

X8:5

SZ1

Table up Table down

C Philips Medical Systems

1

H18

X8:3

1

B

X5:06

0VAC

0VDC

bucky DIAGNOST TS

brake

X6:11

F3

17.5VAC

X8:8

X6:10

5

5

X8:2

+

Q7 Q9

3

H20

0VAC

+24V

Q9

K5

JP3

6.3A slow

2 X9:4

Q3 6.3A slow

X9:3

X5:05

F1 X8:7

M2

K3

X6:07

SNVS9

D9 D22

ENABLE DOWN

2

X4:05

PIC-Controller

2

D10

to stand see Z1-7 (tomo drive)

SNVS7

down

M

X6:05

bucky centered

4

U1

2

K6

2 down

SAS3

Q11

to stand see Z1-5

SAS22

U6

JP1

1

1

U7

X7:2

1

1 -

ENABLE

fast

SAS21

Dk5

U6

slow

X4:10 1a

A3 99-11-23 Ost.

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

SAY2a

1

F4 I =2A

Philips

X10:3

+24V

to stand see Z1-7 (tomo drive)

U6

(99.0)

k1/k2 Off k1/k2 ON

k4 ON k4 Off

+ Tomo Tomo Start

K3 ON = Tomo Start

K5 Off = Slow k5 ON = fast

table control

SNVS7

lowest range = 540mm

highest range = 810mm SNVS8 safety switch ~ 200N without load! SNVS9 SAS21/22 Table up / down

Table control tomo version bucky DIAGNOST TS Z1-3-1

Philips

SX11:11

FD L CH

X1:01

EZX21:12

SX11:12

FD C CH

X1:02

EZX21:03

SX11:03

FD R CH

X1:03

EZX21:04

SX11:04

REL CH

X1:04

EZX21:05

SX11:05

PO 15/40V

X1:05

EZX21:06

SX11:06

NG 15V

X1:06

EZX21:07

SX11:07

SIGN CH

X1:07

PH

EZX21:08

SX11:08

RF 0V CH

X1:08

amplimat measuring chamber

EZX21:13

SX11:13

0V PW

X1:09

EZX21:01

SX11:01

PO 400V

X1:11

housing

housing

screen

X1:10

VA bucky module

X23 +15V

E/WA

delay new

P1

P2

W1 +26V XG +26V XG

8 +26V XG

X1:05

ACTIVE +26V XG

7

+5V

L5 WAX1:05

1

2

tilt switch

3

IC1A

IC1B

X3:1

+26V XG

X2:1

timer

C7

timer ~20s

RY2

2

D1

RY1 SYNC

T2

+15V

RESET

R8 X24 4

D2 +24V

X25

X3:5

L4

+15V +15V

1

XG

D5

RQ_XG_PR +26V XG

X3:6

X2:6

D3

R11

2 1

X26

X2:5

X1:03

H11

S1

+24V

H15

+5V

DC/DC

WAX1:03

X30

FREE EXPOSURE

6

X28 X21

R4

X2:9

X2:7

RQ_XG_PR

+24V 3

X20

1

X1:04

WAX1:04

H16

RG_DV

5

+24V

R2

+26V XG

4

CASS_PR

generator release 1 "bucky"

VA2 sensing

X22

RY2

T3

R13

R12

X27

Y1 2

& START

C8 +26V XG

+5V

X3:4

X2:4

R13

OC1

D4 +24V

+15V

X29 X31

L7 XG_RD_EX

WAX1:02

JP4 X3:3

X1:02

H12

RQ_XG_EX

GND XG

WAX1:01

X1:01

WAX1:07

X1:07

2

X2:3

XG_RD_EX

D6 4

5 3

JP5 X3:2

RY1

VA2 sensing

RY2 1

CASS_PR

X32 X33

SYNC

X2:2

VA1 bucky control XG

bucky 1 SW SF CF format contact

SL XG T0

WAX11:01 WAX11:02

1)

WAX11:03

tomo mode

A3 99-11-23 Ost.

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Philips

EZX21:11

WAX11:04

SW TO 1

WAX11:05

tomo ready ACTIVE WAX11:06

X11:11

see Z1-5 ACTIVE

to stand

WAX11:09 bucky ready SW BU 1

WAX11:10

1)

SZ1

table control 1) connector delivered with bucky DIAGNOST TS

bucky DIAGNOST TS C Philips Medical Systems

(99.0)

Bucky control standard version bucky DIAGNOST TS with bucky Module non sensing Z1-4

Philips

SX11:11

FD L CH

X1:01

EZX21:12

SX11:12

FD C CH

X1:02

EZX21:03

SX11:03

FD R CH

X1:03

EZX21:04

SX11:04

REL CH

X1:04

EZX21:05

SX11:05

PO 15/40V

X1:05

EZX21:06

SX11:06

NG 15V

X1:06

EZX21:07

SX11:07

SIGN CH

X1:07

PH

EZX21:08

SX11:08

RF 0V CH

X1:08

amplimat measuring chamber

EZX21:13

SX11:13

0V PW

X1:09

EZX21:01

SX11:01

PO 400V

X1:11

housing

housing

Screen

X1:10

VA bucky module

X23 +15V

E/WA

delay new

P1

P2 delay old

WAX1:04

W1

WAX1:05

1

2

+26V XG

tilt switch

3

IC1A

IC1B

2

D1

timer

timer ~20s

RY1

ACTIVE

X2:1

X3:1

+26V XG

FREE EXPOSURE

6

X24 4 +24V

X25

X1:03 RQ_XG_PR +26V XG

X2:6

D3

R11

2 1

X26

timer ~2s

+15V

XG

X3:6

S1

+24V

IC2

+15V

X2:5

X3:5

L4

R8

D2

DC/DC

H11

+15V

RESET

H15

+5V

SYNC

T2

+5V

X1:05 +26V XG

WAX1:03

X30

C7

7

8 +26V XG

X28 X21

R4

X2:9

+26V XG

X2:7

RQ_XG_PR

+24V 3

X20

1

X1:04

H16

RG_DV

5

+24V

R2

+26V XG

4

CASS_PR

generator release 1 "bucky"

VA2 sensing

X22

RY2

D5

R13

R12

X27

Y1

T3 & START

C8 +26V XG

+5V

X2:4

X3:4

R13

OC1

D4 +24V

+15V

X29 X31

L7 XG_RD_EX

WAX1:02

JP4

X1:02

D6 XG_RD_EX

GND XG

WAX1:01

X1:01

WAX1:07

X1:07

5

4 3

JP5 RQ_XG_EX

2

X2:3

X3:3

H12

X3:2

RY1

VA2 sensing

RY2 1

CASS_PR

X32 X33

SYNC

X2:2

VA1 bucky control

bucky 1 SW SF CF format contact

SL XG T0

WAX11:01

X2:01

WAX11:02

X2:02

WAX11:03

X2:03

WAX11:04

X2:04

WAX11:05

X2:05

SW TO 1

controller

tomo mode

A3 99-11-23 Ost.

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Philips

EZX21:11

X11:05

Rad on2

see Z1-7

tomo ready

from stand (PLC) WAX11:06

X2:06

WAX11:08

X2:08

Rad on1

ACTIVE

bucky ready SW BU 1

WAX11:09

X2:09

WAX11:10

X2:10

SZ1

X11:11

see Z1-7 ACTIVE

to stand

table control 1) connector delivered with bucky DIAGNOST TS

bucky DIAGNOST TS C Philips Medical Systems

(99.0)

Bucky control tomo version bucky DIAGNOST TS with bucky Module non sensing Z1-4-1

Philips

MEDIO CP EMX2/5/12/22

4

4

MEDIO CP-H EZAX1/2/3/4 4

0V

X1:4

X2:4 10R

C1

Generator 7

5

7

7

5

5

-(24)V

RDGV

100 m 63V

X1:7

X1:5

X1:4

C2

100n

100n

7924 1µF 40V

K4

0V

K4

100 m X2:7 63V

-(24)V

X1:7

X2:5

RDGV

X1:5

PREP

X1:3

10R 1N5060

3

3

3

PREP

K3

X1:3

1µF 40V

K3

1µF 40V

K2

X2:3

10R

bucky TS table SZ1

1N5060

2

1

2

X1:2

2

EXP 1

1

X1:1

K2 1N5060

10R

K1

X2:2

X1:2

10R

K1

EXP X2:1

X1:1

SZ3 Interface SCP: Programming for EZ X1/2 see EZ106 and EZ102 see B22 and B23 for W40/W41

A3 00-03-01 Schr.

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

SCP EZX1/2

Exposure interface SZ3 bucky DIAGNOST TS /Tomo and Medio CP /CP-H or SCP bucky DIAGNOST TS C Philips Medical Systems

(00.0)

Z1-4.2

Philips

SCP EZX1/2

MEDIO CP EMX2/5/12/22 4

4

0V

X1:4

V1

MP1 0V

R1

C5 100 m 63V

Generator 7

7

7

-(24)V

X1:7

X2:4 C1 100 m 63V

C2 100n 63V

A1

1 GND 3 2 Vin Vout

V2

MP2 -24V AC

5

5

5

RDGV

X1:5

3

3

3

PREP

X1:3

2

2

X1:2

2 EXP

1

1

1

X1:1

V10

X2:5

RDGV

X1:5

X2:3

PREP

X1:3

V6 MP5 RGDV

K4

V5

MP7 EXPout

K2

X2:2

R3

K1

H1 R6

Interface

bucky TS table SZ1

MP6 PREP

K3

V4

K1

SZ3

X1:7

MP4 -24V DC

K2 R8

-(24)V

R4

C8

H3

R10 X2:7

K3 R7

V9

X1:4

R5

C7

H2

0V

H5

C3 100 m 63V

K4 R9

V8

C4 100n 63V

V7

C6

H4

A3 00-03-02 Schr.

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

4

MEDIO CP-H EZAX1/2/3/4

V3 R2

X1:2

EXP MP8 EXPin X2:1

X1:1

SCP: Programming for EZ X1/2 see EZ106 and EZ102 see B22 and B23 for W40/W41

Exposure interface SZ3 bucky DIAGNOST TS /Tomo and Medio CP /CP-H or SCP bucky DIAGNOST TS C Philips Medical Systems

(a/00.0)

Z1-4.2

b b UES21

+5V

U5

+24V

Philips

X35:3

x

z

X1:02

X40:02

X36:2

X38:2

X36:3

X38:3

X36:4

X38:4

+5V +5V

X1:12

X40:12

X1:24

X40:24

X36:5

X38:5

X1:21

+24V

+5V

+5V

U5 +24V

+5V X35:2

U5

DC/DC

U6

+24V

X40:01

X38:7

X36:8

X38:8

1

U6

LED_X

X37:1

Y5

1

U3

x

2

clock

S13

generator

UES23

+5V

U6

U5

+24V

X37:3

1

4

x

X37:4

S23

U4 +5V

UF

tube arm

U7

H4

D1

X1:04

X40:04

LED_Z

UES24

+5V

16-bit micro contoller

driver

U5

X35:6

+5V X1:05

X40:05

X24:5

+24V LED_ b

X31:11

UBS19 1

X31:02

2

X24:3

+24V

LED_ a

column rot. b

X24:6

X22:11

U3

U7

H5

Upgrade kit Center Laser

SID 110 cm

4

X37:2

driver

X35:5

tube support z

X31:9

U7 X1:01

X36:7

2

+5V

+24V

D1

X38:6

S20

+5V

H1

Y6

1 X36:6

column

a

S14

U8

S25

UES20

tube support a

4

U3

X40:21

X35:7

1

+24V +5V

UES25

a

2

S12 U6

LED_Y

driver

a

Y4

1

U7 D1

y

1

JP1

X22:02

X24:4

UBY8

2

y

X24:1

UES22

+5V see Z1-2

X35:4

+24V

U6

U5

X31:10

UBS17

X24:2

X22:10

S22

1

4

column centered X31:09

+5V

D1

X1:03

X40:03

LED_ b

+5V

UB1

LED_ b

driver

H7

X22:09

U7

H3

stand control

U7 active

D1

X1:07

X40:07

LED_ACTIVE

ACTIVE = Tube select

driver

X49:1

X1:1 5V DC

X49:2

0V X1:2

X32:5

X1:5

M4

X32:6

X1:6

M3

Laser *

+5V

time

Q1/Q2

S8 X1:08

X40:14

LIGHT ON

LIGHT ON

UE1

P1

timer X35:1

X11:11

X20:11

X22:05

X31:05

X11:06

X20:06

X22:04

X31:04

X32:4

X1:4

M5

X11:08

X20:08

X22:07

X31:07

X32:3

X1:3

M6

ACTIVE

ACTIVE

24VAC

24VAC

H1 0VAC

A3 00-02-23 Ost.

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

X38:1

U3

+5V

H2

UES26

Philips

X36:1

+24V

+24V

field

0VAC 2

1

X11:01

X20:01

X22:01

X31:01

X11:12

X20:12

X22:12

X31:12

X11:09

X20:09

X22:09

X31:09

K1

+24V

see Z1-2/Z1-3

UE2 SZ1

table control

UB1

stand control

UE

* Optional bucky DIAGNOST TS C Philips Medical Systems

(99.0)

control handle

LA

manual collimator

Stand control standard version bucky DIAGNOST TS Z1-5

b b UES21

+5V

Philips

U5

+24V

S21

X35:3

X36:2

X38:2

X36:3

X38:3

X36:4

X38:4

X36:5

X38:5

X36:6

X38:6

X36:7

X38:7

X36:8

X38:8

a

2

x

z

S12 X41:14

a

U6

LED_Y

X41:01

+24V

VCL +24V

UES25

+5V X35:7

4

a

U6

U8 +24V

+5V

Y6

1 2

z

S14

U5

S25

UES20

DC/DC

U5

+24V

1

SID 110 cm

4

+5V X31:9

S20

X35:2

+24V X41:15

1

U3

X41:09

y

X37:1

Y5

1

U3

U6

LED_X

X37:2

x

2

clock

UES23

+5V

S13

generator

U5

U6

S23

X35:5

+24V

X37:3

1

4

x

X37:4

U4 X41:11

UF

LED_Z

tomo control handle

UES24

+5V

U5

LED_b

X31:11

+5V

U5

JP1

X22:02

X31:02

X24:4

X31:10

LED_ACTIVE

X41:08

LIGHT ON

2

1

X31:09

ACTIVE = Tube select

y

UBS17

X24:2

X22:10

LED_b

X41:13

UBY8

X24:1

+24V

U6 LED_b

1

D1

S22

X35:4

X41:12

b

X24:3

+24V

U3 see Z1-2

UBS19

X24:6

X22:11

LED_a

UES22

Upgrade kit Center Laser

X24:5

+24V

EPLD X41:13

tube arm

S24

X35:6

4

X22:09 LCD

UB1

X41:07

(2x10)

Laser *

UE1

X49:1

X1:1 DC

X49:2

0V X1:2

X32:5

X1:5

signal converter X41:03

display

stand control

X34

X41:06

time

MAX 232

LIGHT ON

P1

M1

M4

Centering laser (option)

X32:6

X35:1

ACTIVE 24VAC

X41:04

X41:05

X41:02

tomo on

timer X1:6

M3

X11:11

X20:11

X22:05

X31:05

X11:06

X20:06

X22:04

X31:04

X32:4

X1:4

M5

X11:08

X20:08

X22:07

X31:07

X32:3

X1:3

M6

M2

ACTIVE 24VAC

H1

X31:12

X11:09

X20:09

X22:09

X31:09

tomo on

UB1

table control

stand control

X30:05

tomo test

X30:04

SZ1

UE control handle EAO:0

bucky DIAGNOST TS

2

K1

+24V

X30:06

EAO:1

X30:07

EAO:2

X30:08

EAO:3

X30:09

EAO:4

X30:10

EAO:5

X30:11

EAO:6

(99.0)

C Philips Medical Systems

PLC

X30:15

EAO:7

X30:02

EA16:0

X30:03

PGU:3

X30:12

X30:01

UE2

PGU:8

LA

hand collimator

Stand control tomo version bucky DAIGNOST TS

5V

X22:12

RTS

X20:12

RX

X11:12

lock X

X31:01

FFA ok

X22:01

light on

X20:01

table down

X11:01

field

0VAC 1

table up

+24V

+24V

tomo time

see Z1-2/Z1-3

tomo angle

0VAC

A3 00-02-23 Ost.

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

1

U3

UES26

Philips

X38:1

X36:1

PGU:9

PGU:1

* Optional

Z1-5.1

b b UES21

+5V

Philips

U5

+24V

S21

X35:3

X38:1

X36A:1

1 V4

R1

X36A:2

X38:2

X36:3

X36A:3

X38:3 1

X36:4

X36:4

X36A:4

X38:4

X36:5

X36:5

X36A:5

X38:5

X36:6

X36:6

X36A:6

X38:6

X36:7

X36:7

X36A:7

X38:7 1

X36:8

X36:8

X36A:8

X38:8

X36:9

X36:9

X36A:9

X36:2

X36:2

X36:3

a

2

x

z

S12 X41:14

a

U6

LED_Y

X41:01

+24V

VCL +24V

UES25

+5V X35:7

a

DC/DC

U5

Y6

1 V5

U6

U8 +24V

+5V

4

R2

2

S14

U5

S25

UES20

+24V

4

z SID 110 cm

+5V

S20

X35:2

+24V X41:15

V2

U3

X41:09

y

X37:1

UE3

U3

U6

LED_X

UF

tube arm

Brake amplifier

X37:2 clock

UES23

+5V

generator

U5

U6

S23

X35:5

+24V

X37:3 X37:4

U4 X41:11

LED_Z

tomo control handle

UES24

+5V

U5

S24

X35:6

LED_b X41:13

X31:11

+5V

U5

JP1

X22:02

X31:02

X24:4

X31:10

LED_ACTIVE

X41:08

LIGHT ON

2

1

X31:09

ACTIVE = Tube select

y

UBS17

X24:2

X22:10

LED_b

X41:13

UBY8

X24:1

+24V

U6 LED_b

1

D1

S22

X35:4

X41:12

b

X24:3

+24V

U3 see Z1-2

UBS19

X24:6

X22:11

LED_a

UES22

Upgrade kit Center Laser

X24:5

+24V

EPLD

4

X22:09 LCD

UB1

X41:07

(2x10)

Laser *

UE1

X49:1

X1:1 DC

X49:2

0V X1:2

X32:5

X1:5

signal converter X41:03

display

stand control

X34

X41:06

time

MAX 232

LIGHT ON

P1

M1

M4

Centering laser (option)

X32:6

X35:1

ACTIVE 24VAC

X41:04

X41:05

X41:02

tomo on

timer X1:6

M3

X11:11

X20:11

X22:05

X31:05

X11:06

X20:06

X22:04

X31:04

X32:4

X1:4

M5

X11:08

X20:08

X22:07

X31:07

X32:3

X1:3

M6

M2

ACTIVE 24VAC

H1

X31:12

X11:09

X20:09

X22:09

X31:09

UB1

table control

stand control

X30:05

tomo test

X30:04

SZ1

UE control handle EAO:0

bucky DIAGNOST TS

2

K1

+24V

X30:06

EAO:1

X30:07

EAO:2

X30:08

EAO:3

X30:09

EAO:4

X30:10

EAO:5

X30:11

EAO:6

(00.0)

C Philips Medical Systems

PLC

X30:15

EAO:7

X30:02

EA16:0

X30:03

PGU:3

X30:12

X30:01

UE2

PGU:8

LA

hand collimator

Stand control Fix tube arm and tomo version bucky DAIGNOST TS

5V

X22:12

RTS

X20:12

RX

X11:12

lock X

X31:01

FFA ok

X22:01

light on

X20:01

table down

X11:01

field

0VAC 1

table up

+24V

+24V

tomo time

see Z1-2/Z1-3

tomo angle

0VAC

tomo on

A3 00-03-29 Ost./Schr.

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

V 1

X36:1

U3

UES26

Philips

X36:1

PGU:9

PGU:1

* Optional

Z1-5.1.1

Philips

X4:11

FD L CH

X1:01

EZX22:12

X4:12

FD C CH

X1:02

EZX22:03

X4:03

FD R CH

X1:03

EZX22:04

X4:04

REL CH

X1:04

EZX22:05

X4:05

PO 15/40V

X1:05

EZX22:06

X4:06

NG 15V

X1:06

EZX22:07

X4:07

SIGN CH

X1:07

PH

VA

EZX22:08

X4:08

RF 0V CH

X1:08

amplimat measuring chamber

EZX22:13

X4:13

0V PW

X1:09

bucky unit "Inalfa"

EZX22:01

X4:01

PO 400V

X1:11

housing

housing

screen

X1:10

X23

VA2 sensing

X22

RY2

+15V

4

5

CASS_PR

E/WA

VP

generator

wall connection box

delay new

+24V

3

tilt switch X20 X30

+24V

R2

R4 1

2

3

IC1A

2

IC1B

D1

timer ~20s

T2

RY1 1

timer

C7

+26V XG

WAX2:07

+15V

R8 X24

X1:1

X2:1

4

D2 +24V

X1:5

X1:6

timer ~2s

+15V

D3

R11

X25

2

S1 1

X26

X2:5

X2:6

+24V

IC2

+15V

DC/DC GND XG

SYNC

RESET

release 2 "BWS" WAX2:04

X28 X21

P2 delay old

P1

W1

1

D5 T3

R13

R12

Y1 X27

2

& START

C8 XG_RD_EX

WAX2:02

X1:4

X2:4

X1:3

X2:3

R13

OC1

D4

+15V

+24V

X29

X31 4

RQ_XG_EX

WAX2:01

X1:2

X2:2

X1:9

X2:9

5

6

VA2 sensing

RY2

D6

CASS_PR

X32

X33

7

RY2 X2:7

2 1

8

X1:7

RY1 SYNC

3

not used

A3 99-11-22 Ost.

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Philips

EZX22:11

FREE EXPOSURE

VA1 bucky control

bucky DIAGNSOT VE with bucky unit non sensing bucky DIAGNOST TS C Philips Medical Systems

(99.0)

Z1-6.1

UB1

X11:01

X20:09

X11:02

X20:02

X21:03

X11:03

X20:03

X21:04

stand control

shielding

W27

XS:11

RS485 data RS485 data +

Philips

X11:04

X21:05

X20:04

EA32:0 Rad on1

X11:05

X20:05

X11:06

X20:06

X11:08

X20:08

X11:01

X20:01

X11:12

X20:12

X21:06

Rad on2

EA32:1 EA32:8

W28

U finder 1

EA32:9

U finder 2

X11:07

GND

X20:07

X11:10

W27

X21:02

X20:10

X21:15

Umot 2X11:11

Rad on2

X30:01

PGU:3

X30:03

PGU:8

X30:02

PGU:9

X30:12

shielding

X20:11

X21:08

W20

SZ1

GND

XPI:INB1 XPI:INB2

X21:09

EA32:2

interrupt1 interrupt2 release clutch

table control

RS232

5V

EA0:0

X30:04

EA0:1

X30:05

EA0:2

X30:06

EA0:3

X30:07

EA0:4

X30:08

EA0:5

X30:09

EA0:6

X30:10

EA0:7

X30:11

EA0:8

X30:15

tomo on tomo test tomo angle tomo time table up table down light on FFA ok

K2 finder 4061.9024 X21:10

EA32:3

X21:01

EA32:4

X21:14

EA32:5

brake Y

EA0:9

X21:11

control handle

EA16:8

X21:12 service switch Nikkai/SS-12 SDP2

X lock

W29

UE EA16:9

S1

RTS

GND

24VDC

XPI:-

X21:07

device active

XPI:+

RX

24VDC

24V

24VDC

Umt 2+

PLC tomo control

X21:13

EA16:1 EA16:3 EA16:0

W28

R1

R2

1R (1W)

1R (1W)

C1

C2

1.5U 100V

1.5U 100V

K1

X22:04

W30

X31:04 U finder 1

X22:07

X31:07

X22:01

X31:01

X22:12

X31:12

U finder 2 24V GND

JP1 Jumper

F1 C3

1.5U 100V 1R

1R

R3

PS 102R (DO-41)

C4

1.5U 100V

(1W)

GL1

X22:06

D1

polyswitch 1.85A

R4 (1W)

~ ~ rectifier + -

W31

X31:06 release brakes

X22:08

X31:08

X22:02

X31:02

X22:03

X31:03

X22:05

X31:05

X22:10

X31:10

X22:11

X31:11

X22:09

X31:09

locks off brake Y spare device active sensor Y sensor beta shielding

UE control handle

X23:01

X23:02

X23:03

X23:04

X23:05

X23:06

X23:08

X23:07

X23:09

X23:10

X23:11

X23:12

X24:1

X24:2

X24:3

X24:1

X24:2

X24:3

X24:4

X24:5

X24:6

X24:4

X24:5

X24:6

R1

UB2

V1

Brake amplifier

V4

A3 00-03-29 Ost./Schr.

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

data +

Rad on1

GND

Philips

PGU:1 data -

XS:12

X24A:1

X24A:2

X24A:3

X24A:4

X24A:5

X24A:6

W24 W26

2

2

M UB

C Philips Medical Systems

(a/00.0)

M

S16: limit switch

M2: tomo motor

W22

W23

4

2

1

1

magnetic sensor 1

1 M3: coupling motor

bucky DIAGNOST TS

W21

W25

S20: safety end-switch

B1: angle sensor

S17: sensor lock Y sensor

Y8: brake Y column axis

S19: sensor beta-lock

Stand control tomo version bucky DIAGNOST TS Z1-7.1

Drawings

Survey of components tomo version

Z2--1

Wiring diagram table standard version

Z2--2.1

Wiring diagram table tomo version

Z2--2.2

Wiring diagram stand standard version

Z2--3.1

Wiring diagram stand tomo version

Z2--3.2

Wiring diagram stand standard version

Z2--3.3

Wiring diagram control handle tomo version

Z2--4.1

Wiring diagram control handle standard version

Z2--4.2

Eathing diagram

Z2--5

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Philips

bucky DIAGNOST TS

bucky DIAGNOST TS bdts_z02_990

(99.0)

 1999 Philips Medical Systems ALL RIGHTS RESERVED

Z0--2

Philips

LA

SB SA

SA

VA

PH

SNV SC

SZ3

SZ2

SZ1

SN

UB1 UB UF SB

A4 00-11-10 Ost./Schr.

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

Philips

UE

XA

SA SB SC SN SNV SZ1 SZ2 SZ3 UB UB1 UF PH VA UE LA XA

-

Transverse carriage brakes Pivot tower Footswitch Power supply Motor assy & Motor control switches Table control Tomo speed control Tomo Medio CP interface Column basement Column control & Tomo control Column and transversal tube support arm Amplimat chamber Bucky carriage brake Control handle Collimator X-ray tube

bucky DIAGNOST TS C Philips Medical Systems

(00.0)

UE

Survey of components tomo version bucky DIAGNOST TS Z2-1

SAY2a

SAY2b

bucky module

SAS5/H5 3 B 4

SAY1 SZ1X7:5 SZ1X7:6 SZ1X7:7

WH BN GN

WH BN

SZ1X9:1 SZ1X9:2

2

1a 1 1b

1 2 B A

SZ1X3

SZ1X4:10 SZ1X4:11 SZ1X4:12

gy pk bl

SZ1X4:02 SZ1X4:03 SZ1X4:04 SZ1X4:05

wt bn gn yw

SZ1X9:4

from generator

Identitication of LED´s see chapter 4 "Trouble shooting"

2

1

SZ1X6:08 SZ1X6:09

T2:0V

Z1:L1

Z1:N

G1:L1

T1:0V T1: SN1X8:5

SN1X8:4 +24V 1

0V 5

X16:3 X16:4 X16:

X16:

T2:230V T1:230V

X16:4

X16:1.1

2 6

4 4 SAY1:1

SAY2a:1

SAY2a:2 3 3 SAY1:2

2 2

1 1 SAS3:1

8

9

2 5

3 6

SN1X8:2

X16:3

X16:4

SN1X8:8

SN1X8:3 0VAC

7 1 4

230VAC

9 3 6

0VAC

8 2 5

230VAC

7 1 4

T2

115VAC

9 3 6

SN1X8:6

02

0VAC

10

04

230VAC

13

13

15 12

11

10

12

11

10 6 3

8

5 2

7

4 1

G1 1WAX1

03

Y1

+

screen SAS22:1a SAS22:1 SAS22:1b SAS5:3 SAS5:3 SAH5:B SAS5:4

X7:1 X7:2 X7:3 X7:4 X7:5 X7:6 X7:7 X7:8 X7:9

SNV rd

2.1 1.1

T1

2

motor

blk

X1

9

8

7 yw SZ1X6:10 SZ1X6:11

X5

01 JP6

gn

05

1

X7

L1 N

3

07

14

X6

LINE L1

06

X4 S8

N

08

VAX2

2 5

SZ1X6:04

09

X9

2

10

X3

SAS3:4

X10

1 4

GN

Z1

table control

10A slow

24VAC

4

01

LOAD

G1:

F1

H20

SZ1X6:05

JP5

JP4

SZ1

L1.1

N.1

T2:

1

JP3

YE

02

X16:4

10A slow

8

03

+26V XG

0VAC

F2

6

UB1X20

1

SNT2:0VAC

5

7

X11

4A slow

X16:2.1

SNT2:17.5VAC

H16

F6

4

SNT2:37VAC

1WAX2

5

SNG1:0V

4

05 04

2 3

X16 4

9

SNG1:+24V

1

X2

3

WH

SNT1:0VAC

S7

F5 6.3A slow

8

4 SZ1X6:02

F12

X16:3

1 SNT1:24VAC

2

15

SZ1X6:03

F11

06

24VAC

12

BN

1

07

JP1 +24V

2

08

230VAC

X8

H18

3

09

115VAC

H17

H11 = RQ_XG_PR H12 = XG_RD_EX H15 = ACTIVE

4

X16:3

JP2B

JP2A

10

H15

H11

L

Z1:L1.1

Z1:N.1

H12

7

SZ1X6:07

S9

G1:N

4 PK

N

1

14

SZ1X6:06

11

GY

M1

-

A3 99-11-24 Ost.

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

SAS21

1

SZ1X9:3

2

Philips

1

transverse carriage

S3

4

2

SAS2/H2

SA

1

1

SZ1X4:08

GY PK BU

2

SZ1X4:09

SZ1X7:2 SZ1X7:3 SZ1X7:4

VAX2

SZ1X10:2

1a 1 1b

VA

SZ1X10:3

Philips

SAS22

PH

1 SZ1X4:06

SZ1X4:07

2

SAY3

generator

X1

X6:01 X6:02 X6:03 X6:04 X6:05 X6:06 X6:07 X6:08 X6:09 X6:10 X6:11 X6:12

screen SNVS7:1 SNVS7:2 SNVS8:1 SNVS8:2 SNVS9:1 SNVS9:4 SNVM1:+ SNVM1:SNVY7:1 SNVY7:2

X4:01 X4:02 X4:03 X4:04 X4:05 X4:06 X4:07 X4:08 X4:09 X4:10 X4:11 X4:12 X4:13 X4:14 X4:15

screen SAS2:1 SAS2:2 SAH2:B SAH2:A SAY2a:1 SAY2a:2 SAY3:1 SAY3:2 SAS21:1a SAS21:1 SAS21:1b

X5:01 X5:02 X5:03 X5:04 X5:05 X5:06 X5:07 X5:08 X5:09 X5:10 X5:11 X5:12 X5:13 X5:14 X5:15

screen SCS4:1 SCS4:2 SNVS15:1 SBS 15 SBS 15 SBH 4 SBH 4 SNVP1:1 SNVP1:2

SN

wt bn

SZ1X5:02

SZ1X5:03

4 2

bucky DIAGNOST TS C Philips Medical Systems

(99.0)

S4 1

SC foot switch

power supply

Wiring diagram table Standard version bucky DIAGNOST TS Z2-2.1

SAY2b

SAY2a

bn

SZ1X9:1 SZ1X9:2

SZ1X4:09

SZ1X4:08

SAS21

2

1

SZ1X4:10 gy SZ1X4:11 pk SZ1X4:12 bl

1a 1 1b

SZ1X3

SZ1X4:02 SZ1X4:03 SZ1X4:04 SZ1X4:05

1 2 B A

wt bn gn yw

from generator

Identitication of LED´s see chapter 4 "Trouble shooting"

S9

4

H1

H3

H5

SZ1X6:10 SZ1X6:11

1

rd

SZ1X6:09

+

2

3

4

SAY1:2

SAY1:1

M1

-

screen SAS22:1a SAS22:1 SAS22:1b SAS5:3 SAS5:3 SAH5:B SAS5:4

X7:1 X7:2 X7:3 X7:4 X7:5 X7:6 X7:7 X7:8 X7:9

X6:01 X6:02 X6:03 X6:04 X6:05 X6:06 X6:07 X6:08 X6:09 X6:10 X6:11 X6:12

screen SNVS7:1 SNVS7:2 SNVS8:1 SNVS8:2 SNVS9:1 SNVS9:4 SNVM1:+ SNVM1:SNVY7:1 SNVY7:2

Z1:N

T2:230V

T1:0V

T1:230V

T2:0V

T1: T2:

Z1:L1

X16:3

X16:4

SN1X8:4 +24V 1

G1:L1

SN1X8:5

X16:3

X16:4 X16:

X16:

0V 5

X16:1.1

3

5 9

X4:01 X4:02 X4:03 X4:04 X4:05 X4:06 X4:07 X4:08 X4:09 X4:10 X4:11 X4:12 X4:13 X4:14 X4:15

screen SAS2:1 SAS2:2 SAH2:B SAH2:A SAY2a:1 SAY2a:2 SAY3:1 SAY3:2 SAS21:1a SAS21:1 SAS21:1b

X5:01 X5:02 X5:03 X5:04 X5:05 X5:06 X5:07 X5:08 X5:09 X5:10 X5:11 X5:12 X5:13 X5:14 X5:15

SN1X8:3

X16:4 0VAC

230VAC

0VAC

115VAC

X16:3

SN1X8:8

X16:4

SN1X8:6 230VAC

9 3 6

0VAC

8 2 5

X16:3

04

02

12

11

10 7 1 4

9 3 6

9 3 6

6 3

X16:2.1

1 1

13

13

14 11

10

12

11

10

5 2

8

4 1

G1

T2

motor

blk

L1 N

1WAX1

03

Y1

2

SNV SZ1X6:08

X1

01

JP6

gn

X5

9

8

7

X7

yw

05

1

7

SZ1X6:04

2 5

gn

X6

LINE L1

06

X4 S8

2.1 1.1

07

2 4

N

08

VAX2

3

Z1

09

X9

1 4

SZ1X6:05

SAS3:4

X10

H19

24VAC

yw

LOAD

10

X3

SAS3:1

H20

2

8

SZ1X5:10

P1

01

L1.1

N.1

table control

10A slow

3 1

4

G1:N

F1

7

gn

SZ1X5:09

1WAX2

G1:

SZ1

2 5

bn

JP5

JP4

1

1 4

wt

05

02

F2 10A slow

JP3 SZ1X5:08

X16

+26V XG

4

8

UB1X20

8

7

F12

07

03

15

6

X11

4A slow

7

SNT2:0VAC

5

F11

04

14

SNT2:17.5VAC

4

1

08

H16

F6

1

SNT2:37VAC

X2

2

SNG1:0V

F5 6.3A slow

6

SNG1:+24V

1

2

09

06

24VAC

2 3

3

JP1 +24V

SAY2a:2

SNT1:0VAC

S7

H11 = RQ_XG_PR H12 = XG_RD_EX H15 = ACTIVE

H18

4

10

H15

4

4

JP2B

H13

SAY2a:1

SNT1:24VAC

2

H11

H8

230VAC

1

H6

115VAC

X8

SZ1X6:02

H14

0VAC

H17

wt

H12

H10

SN1X8:2

H4

JP2A

bn

H9

230VAC

H2

SZ1X6:03

H7

L

2

3

SZ1X6:07

2

pk

N Z1:N.1 Z1:L1.1

1

15

SZ1X6:06

12

gy

A3 99-11-25 Ost.

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

1

transverse carriage

S3

4

2

SAS2/H2

SA

1

SZ1X9:4

wt

SZ1X9:3

wt bn gn

2

Philips

VAX2

bucky module

SAY1 SZ1X7:5 SZ1X7:6 SZ1X7:7

1

VA

SAS5/H5 3 B 4

2

SZ1X10:2

PH

SZ1X10:3

gy pk bl

SZ1X4:06

SZ1X7:2 SZ1X7:3 SZ1X7:4

1a 1 1b

1

SZ1X4:07

Philips

SAS22

2

SAY3

generator

X1

T1

screen SCS4:1 SCS4:2 SNVS15:1 SBS 15 SBS 15 SBH 4 SBH 4 SNVP1:1 SNVP1:2

SN

power supply

2 yw gn

SZ1X5:05 SZ1X5:04

4

S15

1 wt SZ1X5:02

bn

bl SZ1X5:03

SZ1X5:07 -

2

H4 wt

4

SZ1X5:06

bucky DIAGNOST TS C Philips Medical Systems

+

SB (99.0)

S4 1

SC foot switch

Wiring diagram table tomo version bucky DIAGNOST TS Z2-2.2

Philips

X38:7 X38:8 1

4

wt bn

2

X38:2

X38:1

S14

Y4:1

2

Y4:2

3

S12:1

4

S12:4

5

Y6:1

1

UE2X37:3 GN UE2X37:4 YE

2

6

Y6:2

7

S14:1

8

S14:4

4

1

2

2

S13

a Y5 X38:3 X38:4 1

UE2X36

4

wt bn

x

X38:5

X38:6 2

UE

tube arm

S12

1

x

UF

2

control handle

see Z2-4 a

Y6

UF tube arm

z

stand control

center position

UBS19:1

UB1

y

9 3 6

2

UBY8 see Z2-6

2

y

4

5

6

1

2

3

X24

UBS17:4 UBY8:2 UBS17:1 UBS19:2 UBY8:1

JP1

4

1

9

UB S19

1

UB1X24:4

8 2 5

X22

UB1X24:3 7 1 4

12

10

9 3 6

11

12

8

11

7

X20

UB1X24:6 WH UB1X24:5 BN

2

8

4

7

UB S17

1

wt bn

2 5

10 UB1X24:2 UB1X24:1

UE2X31

1 4

SZ1X11

A3 99-11-26 Ost.

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

1

Y4

UE2X37:2

1

UE2X37:1

X38

b

Wiring diagram stand Standard version bucky DIAGNOST TS bucky DIAGNOST TS C Philips Medical Systems

(99.0)

Z2-3.1

Philips

X38:7 X38:8 1

4

wt bn

2

X38:2

X38:1

S14

Y4:1

2

Y4:2

3

S12:1

4

S12:4

5

Y6:1

1

UE2X37:3 GN UE2X37:4 YE

2

6

Y6:2

7

S14:1

8

S14:4

4

1

2

2

S13

a Y5 X38:3 X38:4 1

UE2X36

4

wt bn

x

X38:5

X38:6 2

UE

tube arm

S12

1

x

UF

2

control handle

see Z2-4 a

Y6

UF tube arm

z

stand control

center position

JP1

UB1X24:3

UB1X24:4

11

12

4

2

UBY8 see Z2-6

2

y

9

1

1

X24

UBS19:1

UB1

y

15

XS:11 XPI:+ EA32:1 EA32:0 EA32:2 INB2 EA16:3 EA16:1 XPI:EA32:5

3

12

6

9

3

14

6

5

13

11

4

10

8

2

7

2 5

1

7

UB S19

UBS17:4 UBY8:2 UBS17:1 UBS19:2 UBY8:1

X21

1 4

UB1X24:6 WH UB1X24:5 BN

X22

EA16:8

EA32:3

INB1

XS:12

9

10

X23

3 6

15

8

9

7

14

6

2 5

13

11 12

UBS16:1 UBM3:b UBM2:a UBS16:4 UBS20:1 UBM2:b UBB1 UBB1

3 6

X20

2

EA32:4

4

2 5

1 4

UB S17

1

wt bn

UBB1 10

8

1 4

12 9

7

2 5

8

11 8

UBM3:a 10 7

1 4

3

UB1X24:2 UB1X24:1

UBS20:2

UE2X31

UBS16:2

SZ1X11

A3 99-11-26 Ost.

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

1

Y4

UE2X37:2

1

UE2X37:1

X38

b

Wiring diagram stand tomo version bucky DIAGNOST TS bucky DIAGNOST TS C Philips Medical Systems

(99.0)

Z2-3.2

0

UEX30:04

1

UEX30:05

2

UEX30:06

3

UEX30:07

4

UEX30:08

5

UEX30:09

UB1X21:07

6

UEX30:10

UB1X21:08

7

UEX30:11

INB2

INB1

-

+

Philips

XPI

PLC

UB1X21:02

EA16

UB1X21:15

8

PGU D-SUB 9P. Jumpers (bottom side) CLOSED OPEN

8 4

5

7 3

6 2

UEX30:15

1

UB1X21:11

3

RS485

EA32

0

2 1

UB1X21:12

4

EA16

Interface Board

6 7

UB1X21:13

9

UB1X21:4

11

8

1

4

+

bn UB1X23:08

wt

-

a

+ b

M3

-

a

b

M2

GN

2

RD

4

gn

1

UB1X23:3 GY UB1X23:5 PK UB1X23:4 BU 2

rd

UB1X23:9 wt UB1X23:7 bn

UB1X23:06

UB1X23:01 gn

UB1X23:02 yw

UB1X23:10 UB1X23:11 UB1X23:12

UB1X21:3

9

7

EPROM

12

X5

8

5 6

UBX30:01

3 4 5

9

UBX30:03

1

UBX30:02

0

UB1X21:06

UBX30:12

UB1X21:05

2

UB M3 UB S20 Limit position

A3 99-11-26 Ost.

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

9

UB B1

UB S16

UB M2

Wiring diagram stand tomo version bucky DIAGNOST TS bucky DIAGNOST TS C Philips Medical Systems

(99.0)

Z2-3.3

Philips

X31

UB1X22

X30 X36

1

X37 4 3 2 1

1

X41

UFS13:1 UFY5:2

1

UFY5:1

X32 X35 LAX1

3

2 7

1

8

S21:1 S25:1 S20:1 S24:1 S20:4

X40

S23:1

UE2 control handle basic

4 9

S22:1

5 1

6

UE1X1

S8 H6 H1

H5

UE1

H2

H4

control handle LED's

X1

H3

4

S24

2

1 4

S23

2

4

1 4

2

S21

1

2

1

4

UE2X35:2

S21:4 UE2X35:1

UE2X35:3

S22:4 S20:4

UE2X35:4

S23:4 S21:4

UE2X35:5

S24:4 S22:4

S26:4 S23:4

UE2X35:6

UE2X40

2

1

S20

S22

2

UE2X35:7 S26:1

S26:4

S24:4 S25:4 4

S26

S25:1

LAX1

A3 99-11-26 Ost.

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

UFX38

UFS13:4

1

4

2

1

S25

Wiring diagram control handle tomo version bucky DIAGNOST TS bucky DIAGNOST TS C Philips Medical Systems

(99.0)

Z2-4.1

Philips

X34 X31 Display

UB1X22

X30 X36

1

X37 4 3 2 1

1

X41

UFS13:1 UFY5:2

1

UFY5:1

UE1 x 1

X32

Tomo control handle

X35 LAX1

3

2 7

1

8

S21:1 S25:1 S20:1 S24:1 S20:4

X40

S23:1

UE2 control handle basic

4 9

S22:1

5 1

6

UE1X1

S8 H6 H1

H5

UE1

H2

H4

control handle LED's

X1

H3

4

S24

2

1 4

S23

2

4

1 4

2

S21

1

2

1

4

UE2X35:2

S21:4 UE2X35:1

UE2X35:3

S22:4 S20:4

UE2X35:4

S23:4 S21:4

UE2X35:5

S24:4 S22:4

S26:4 S23:4

UE2X35:6

UE2X40

2

1

S20

S22

2

UE2X35:7 S26:1

S26:4

S24:4 S25:4 4

S26

S25:1

LAX1

A3 99-11-26 Ost.

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

UFX38

UFS13:4

1

4

2

1

S25

Wiring diagram control handle Standard version bucky DIAGNOST TS bucky DIAGNOST TS C Philips Medical Systems

(99.0)

Z2-4.2

Philips

XA

2.5 mm 2

UE

2.5 mm 2

LA

2.5 mm2 2.5 mm2

SA

SA

A3 00-03-28 Ost./Schr.

Philips

Philips Medical Systems: DECLASSIFIED Customer Service Intellectual Property

UF

4 mm 2 4 mm 2

to generator

8.25 mm 2

X16

U Z1

6 mm 2

SN

6 mm 2

SNV M1

UB 6 mm 2

bucky DIAGNOST TS C Philips Medical Systems

(a/00.0)

T1

G1

T2

Earthing diagram bucky DIAGNOST TS Z2-5

View more...