1093012L_DailyQA3_UG_s
Short Description
RADIO...
Description
Title
Daily QA™ 3 User’s Guide Daily QA Accelerated Model 1093
Your Most Valuable QA & Dosimetry Tools
User’s Guide, Daily QA 3™ Copyright © 2006-2016 by Sun Nuclear Corporation. All rights reserved. The information contained in this guide and the accompanying software program are copyrighted and all rights are reserved by Sun Nuclear Corporation. Copying, duplicating, selling, or otherwise distributing any part of this product without the prior written consent of Sun Nuclear Corporation is prohibited. Sun Nuclear Corporation reserves the right to make periodic modifications to this guide without obligation to notify any person or entity of such revision. This guide is written for: PC software:
version 2.5.4
Database
version 1.4
Firmware:
version 1.7.0 (compatible with all printed circuit board revisions)
Daily QA™ 3, Daily QA Check 2™, PROFILER™ 2, TomoDose™, IMF™ and GMF™ are trademarks of Sun Nuclear Corporation. Other trademarks or trade names are the property of their respective owners. The array calibration procedure for the instrument is protected by U.S. Patent No. 6125335. Document 1093012, Rev L, 3 February 2017
Corporate Headquarters 3275 Suntree Boulevard Melbourne, Florida 32940 USA telephone +1 321-259-6862 http://www.sunnuclear.com
ii
Preface Intended Use The Daily QA 3 is intended to measure and analyze the daily radiation output characteristics of an external treatment machine (linac or Cobalt 60) before the start of daily treatment in a radiation oncology center, and not for calibration of said machines.
CAUTION: Certain federal laws restrict this device to sale by or on the order of a qualified radiation therapy professional, such as a medical dosimetrist, medical physicist, or radiation oncologist.
Application Every morning, the Daily QA 3 is positioned in a standard beam setup for measurement. The parameters measured are: dose output, beam symmetry, beam flatness, beam energy, and light/ radiation field coincidence. These parameter measurements are compared to a reference value using trend graph analysis in the Daily QA 3 software. If a measurement exceeds an action level, an investigation of the cause will take place before treatment resumes for the day.
WARNING: Do not modify a patient’s treatment plan based only on the results from a single QA device; but DO investigate the source of the error in the treatment delivery device (TDD) or treatment planning system (TPS).
Quality and Regulatory Systems Sun Nuclear meets worldwide standards, including ISO 13485, FDA 21 CFR 820 Quality System Regulations, Health Canada's CMDCAS, and the European Medical Device Directive 93/42/EEC. Sun Nuclear is certified and authorized to affix the CE mark on products for marketing and distribution in the Member States of the European Union. Sun Nuclear’s quality systems registrar and European Notified Body for Class II medical products is BSI. Sun Nuclear’s Class I medical products are listed in Europe with the Competent Authority of the Netherlands.
Sun Nuclear Corporation is an ISO 13485 certified company.
Conventions
Preface
• •
Bold typeface indicates a button name or an entry that the user must type. Italicized typeface indicates the title of a manual, a cross-referenced subsection within this manual, or introduces a new term or phrase.
•
Bold italicized typeface indicates a menu option.
•
“Text in double quotes” indicates a message that is displayed to the user.
• •
'Text in single quotes' indicates the name of a window or dialog box. The “Z” suffix in a part number indicates a RoHS2 compliant part.
iii
Health and Safety Instructions WARNING: To avoid the risk of electric shock, this equipment must only be connected to a supply mains with protective earth (ground). CAUTION: Keep the electronics out of the direct beam. Exposing the Daily QA 3 electronics to direct radiation could void the warranty.
iv
•
The device may only be used with the medically approved power supply that is shipped with the device.
•
Inspect the instrument periodically for damage. If errors or damage are suspected, contact Sun Nuclear Corporation.
•
To protect insulation, never pull on a cable to disconnect it. Always grasp the plug or connector. Do not use any power cord or power supply that is damaged or has broken insulation.
•
Do not route cables under or through the treatment room door in a way that would present a tripping hazard.
• •
The instrument is not intended to be used in an oxygen-rich environment. Do not operate the instrument in any strong magnetic fields (i.e., MRI).
Preface
Contents Preface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iii Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iii Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iii Quality and Regulatory Systems . . . . . . . . . . . . . . iii Conventions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iii Health and Safety Instructions . . . . . . . . . . . . . . . iv Introduction . . . . . . . . . . . . . . . . . . . . . . . . . 1 Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1 Parts and Options . . . . . . . . . . . . . . . . . . . . . . . . . . 1 rf-Daily QA 3, P/N 1093000-1Z. . . . . . . . . . . . . . . 1 Daily QA 3, P/N 1093000-0Z . . . . . . . . . . . . . . . . 2 Gantry Mount Accessory . . . . . . . . . . . . . . . . . . . 2 Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Upgrade Installation . . . . . . . . . . . . . . . . . . . . . . . . 3 Backup Before Upgrade . . . . . . . . . . . . . . . . . . . 3 Upgrading Client Software . . . . . . . . . . . . . . . . . 3 Conversion of Database Tables for Daily QA 3 Software Version 2.5 . . . . . . . . . . . . . . 3 New Software Installation . . . . . . . . . . . . . . . . . . . . 3 Prerequisites . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Step 1 – Install Database Software . . . . . . . . . . . 3 Step 2 – Install Client Software . . . . . . . . . . . . . . 4 Hardware Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Charge rf-Daily QA 3 Battery . . . . . . . . . . . . . . . 5 Cable Connections. . . . . . . . . . . . . . . . . . . . . . . . 5 rf-Daily QA 3 Wireless Connection . . . . . . . . 5 Daily QA 3 Wired Connection . . . . . . . . . . . . 6 Software Setup . . . . . . . . . . . . . . . . . . . . . . . 7 Add Users to the Database . . . . . . . . . . . . . . . . . . . 7 Launch Software and Connect to Database . . . . . . 7 Set Up Program Preferences . . . . . . . . . . . . . . . . . 8 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Calibrate Temperature and Pressure – Daily QA 2 and Daily QA 3 . . . . . . . . . . . . . . . . . . . .8 Calibrate Absolute Dose . . . . . . . . . . . . . . . . . . . .8 Absolute Dose Calibration Procedure . . . . . . 9 Calibrate Array . . . . . . . . . . . . . . . . . . . . . . . . . . .9 Create a QA Test Plan, Site, Room, Machine, Template . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 QA Mode View . . . . . . . . . . . . . . . . . . . . . . . . . 10 Set Up the Scheme Tree, Physicist Role . . . . . .10 Set Up Template for Beam Tracking . . . . . . . . . 11 Baseline Options . . . . . . . . . . . . . . . . . . . . . 11 Baseline QA Template. . . . . . . . . . . . . . . . . . . . . . 11
Contents
QA Measurement . . . . . . . . . . . . . . . . . . . . Preparing for Measurements. . . . . . . . . . . . . . . . Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . Printing Reports . . . . . . . . . . . . . . . . . . . . . . . . . Exporting Data . . . . . . . . . . . . . . . . . . . . . . . . . Support and Maintenance . . . . . . . . . . . . . . Maintaining Hardware . . . . . . . . . . . . . . . . . . . . . Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Disposing and Recycling . . . . . . . . . . . . . . . . . Minimizing Radiation Damage. . . . . . . . . . . . . . . Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . Daily QA 3 Troubleshooting . . . . . . . . . . . . . . . Database Connection Troubleshooting . . . . . . Daily QA 3 End Panel . . . . . . . . . . . . . . . . . . . . rf-Daily QA 3 End Panel . . . . . . . . . . . . . . . . . . . PDI Troubleshooting . . . . . . . . . . . . . . . . . . . . . Contacting Sun Nuclear Support . . . . . . . . . . . . . Support Website . . . . . . . . . . . . . . . . . . . . . . . Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Specifications . . . . . . . . . . . . . . . . . . . . . . . Recommended System Requirements . . . . . . . . Daily QA 3 Specifications . . . . . . . . . . . . . . . . . . . rf-Daily QA 3 Specifications . . . . . . . . . . . . . . . . . Software Reference . . . . . . . . . . . . . . . . . . Daily QA 3 Client Software Interface. . . . . . . . . . QA View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Scheme Tree . . . . . . . . . . . . . . . . . . . . . . . Results . . . . . . . . . . . . . . . . . . . . . . . . . . . Trend View . . . . . . . . . . . . . . . . . . . . . . . . . . . . Graph Controls . . . . . . . . . . . . . . . . . . . . . . Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . Appendix A: Regulatory Supplement . . . . . . Sun Nuclear Corporation Symbols . . . . . . . . . . . Operator Responsibility . . . . . . . . . . . . . . . . . . . . Modifications to the Equipment . . . . . . . . . . . . . Environmental Requirements . . . . . . . . . . . . . . . Operating . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Radio Interference . . . . . . . . . . . . . . . . . . . . . . . . Reporting Health or Safety Related Issues or Concerns . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Australian Sponsor . . . . . . . . . . . . . . . . . . . . . . Connection to External Computer Equipment . . EMC Precautions . . . . . . . . . . . . . . . . . . . . . . . . .
13 13 13 14 14 15 15 15 15 15 15 15 16 16 17 17 18 18 19 19 19 20 20 20 22 23 23 24 24 24 25 25 25 27 27 27 28 28 28 28 28 28 28 29 29
v
This page is intentionally left blank.
vi
Contents
1
Introduction
Description Daily QA 3 captures the “beam print” of a radiation field by measuring radiation using ionization chambers and diodes, arranged in a fixed pattern such that a reproducible setup will reveal changes in beam characteristics. Trend analysis is performed with built-in database software. Daily QA 3 software stores measurement results in a database that allows trend analysis of measurements and report generation. The database can be centralized so that any network computer running the Daily QA 3 client software can access the data. Daily QA 3 software can also be used with PROFILER 2, TomoDose, and Daily QA Check 2. For a list of software differences when using other instruments, see the Daily QA 3 Reference Guide or the software Help Index.
Parts and Options rf-Daily QA 3, P/N 1093000-1Z After unpacking, identify the parts. 6 7
1
5 3
4
No. 1
*
Part Number
Qty
2
Description
1093500Z or 1093500
1
rf-Daily QA 3 instrument
2
1093400Z
1
rf Base Station *
3
801073Z
1
Blue Power/data cable, 25 m *
4
0224502Z
2
Power supply, switching, filtered, 110-240 VAC, 50-60 Hz to 18 VDC *
5
801008Z
2
Power cord, detachable, IEC plug to USA style plug *
6
801041Z
1
USB cable, 2 m *
7
0224500Z
1
Power Data Interface (PDI) 3.0 *
–
1093012
1
User’s Guide (not shown)
–
1093061
1
Bulletin, rf-Daily QA 3 Battery Recharging
These items are available in an optional vault pack (part number 1093000-3Z. The vault pack supplies a spare set of cables and hardware so that one can be installed in each room, allowing the user to move only the Daily QA 3 from room-to-room.
Figure 1-1. Parts Furnished with the rf-Daily QA 3
Parts and Options
1
Daily QA 3, P/N 1093000-0Z After unpacking, identify the parts included in package.
6
5 4 3 1
2
No.
Part Number
Qty 1
Description
1
1093300Z or 1093300
Daily QA 3 instrument
2
801073Z
1
Blue Power/data cable, 25 m
3
0224502Z
1
Power supply, switching
4
801008Z
1
Power cord, detachable, IEC plug to USA style plug
5
0224500Z
1
Power Data Interface (PDI) 3.0
6
801041Z
1
USB cable, 2 m
–
1093012
1
User’s Guide (not shown)
Figure 1-2. Parts Furnished with the Daily QA 3
Gantry Mount Accessory See the IMF™ – GMF™ User’s Guide, P/N 1122014, for installation instructions.The user’s guide ships with the gantry mount accessory and is also available on the Sun Nuclear Support page. For instructions to access the Support website, see Support Website on page 19.
2
Linac
P/N
Color
Weight
Dimensions
Elekta
1177000-E
green
7.9 kg (17.5 lb.)
39.4 x 28.7 x 48.8 cm (15.5 x 11.3 x 19.2 in.)
Siemens
1177000-S
red
8.1 kg (17.75 lb.) 39.4 x 31.5 x 48.0 cm (15.5 x 12.4 x 18.9 in.)
Varian
1177000-V
blue
7.4 kg (16.25 lb.) 39.4 x 31.5 x 39.1 cm (15.5 x 12.4 x 15.4 in.)
Section 1. Introduction
2
Setup
Upgrade Installation Backup Before Upgrade Do not install the database software when performing an upgrade to the latest version of Daily QA 3 Software; the database is already installed. Perform a manual backup of the database before beginning the upgrade.
Upgrading Client Software After downloading the client software upgrade, install the upgrade using the instructions in Step 2 – Install Client Software on page 4.
Conversion of Database Tables for Daily QA 3 Software Version 2.5 In the past, Sun Nuclear Support personnel were required to assist in the database conversion, but this is no longer necessary. When the software is first launched following installation of software Version 2.5, there is an automatic database conversion. A dialog box prompts you to confirm the update. 1
Click Yes to continue. The database conversion begins. The update and retrend progress displays on the login screen. Note that the amount of time required for the database update and retrend will depend upon the size of the database.
2
Click No if you do not wish to proceed. For example, if you have not backed up the database, select this option, then perform the backup before continuing.
New Software Installation Prerequisites •
Verify the Daily QA 3 software and Firebird database are not already installed on this computer. If you are currently using a version of the Firebird SQL Database Server for other applications, contact Sun Nuclear Support.
•
Verify the computer meets the system requirements. See Recommended System Requirements on page 20.
•
Administrative rights are required to install the software.
Daily QA 3 software stores data in a database. The database is installed on the computer hosting the client software or a client/server installation. Consult with your network administrator to plan a server installation.
Step 1 – Install Database Software These instructions apply to new installations only! Do not install the database if you are upgrading the Daily QA 3 software. For more details about the database installation, see the Daily QA 3 Reference Guide or the software Help Index. 1
Download the Daily QA 3 Database installation file from the Sun Nuclear Support website and save it to a location that can be accessed from the computer that will host the database. On the ‘Daily QA 3 Product Downloads’ page, click the Software (Current Version) link on the left side of the page to access the current installation file. As of the release of this document, the database version is 1.4.0.
2
Navigate to the location of the downloaded DQA3 Database*.exe file and double-click to begin the installation (where * = the software version).
3
If the ‘User Account Control’ dialog box prompts you to allow the program to make changes to the computer, click Yes.
4
Follow the on-screen instructions to complete the installation.
Upgrade Installation
3
If a ‘Windows Security Alert’ message prompts to keep blocking or unblock the Firebird SQL Server program, select Unblock. For more information about firewalls, see the Daily QA 3 Reference Guide or the software Help Index. 5
When installation completes, read the information about automatic database backups, click OK and then click Finish to exit.
6
Verify that the database software icon, DQA3 DB Admin , appears in the system tray. If the default options were selected when installing the database, the file path for the DBAdmin executable is C:\SNC\QADatabase\DB Admin.
If the firewall does not present the firewall message to unblock the required port, allow the program to communicate through the Windows Firewall as follows (note that the instructions may differ depending upon the version of Windows Operating System): 1
Open the Windows Control Panel.
2
Select Windows Firewall.
3
On the left side of the screen, click Allow a program or feature through Windows Firewall.
4
On the ‘Allowed Programs’ window: •
Windows 7: Click Allow another program. In the ‘Add a Program’ dialog box, select DB Admin, and then click Add.
•
Windows 10: Click Allow another app. In the ‘Add an app’ dialog box, select DB Admin, and then click Add.
Step 2 – Install Client Software Perform the client software installation on the computer(s) which will access the Daily QA 3 Database. When installing the Daily QA 3 client software, the person logged on to the computer must have administrative privileges. The installation fails if the person logged in does not have write access to C:\Windows\System32. 1
Download the Daily QA 3 Database installation file from the Sun Nuclear Support website and save it to a location that can be accessed from the computer on which the client software will be installed. On the ‘Daily QA 3 Product Downloads’ page, click the Software (Current Version) link on the left side of the page to access the current installation file. As of the release of this document, the Daily QA 3 client version is 2.5.4.
2
Navigate to the location of the downloaded DailyQA3Setup_v*.exe file and double-click to begin the installation (where * = the software version).
3
If the ‘User Account Control’ dialog box prompts you to allow the program to make changes to the computer, click Yes.
4
Follow the on-screen instructions to complete the installation.
5
When installation completes, click Finish to exit. After installing the Daily QA 3 client software, ensure that all users on the computer have read and write permissions to the directory where the client software is installed. If the default options were selected when installing the client software, the directory for the DailyQA3 executable is C:\SNC\DailyQA3. A shortcut to the Daily QA 3 executable is placed on the computer’s desktop.
4
Section 2. Setup
Hardware Setup Charge rf-Daily QA 3 Battery If using the rf-Daily QA 3, charge the battery overnight before initial use. The battery is fully charged after about 4.5 hours. Before using the device, press the power switch to lock it in the ON position. For more information about charging the battery, see the Daily QA 3 Reference Guide or the software Help Index. The rf-Daily QA 3 can remain connected to AC power continuously.
Power Supply
Figure 2-1. Charging the rf-Daily QA 3 Battery
Cable Connections Prior to software launch, connect cables. Use the instructions for the type of Daily QA 3 instrument you are using (wireless or wired). rf-Daily QA 3 Wireless Connection If the base station is missing, if the batteries are discharged, or if the rf connection does not appear to be working, plug the 25 m power/data cable from the control room directly into the instrument as shown for the Daily QA 3 wired connection.
Only connect to a power source that matches the ratings on the power supply label
USB Cable
Power Supply (for charging rf-Daily QA battery)
Wireless Base Station (in treatment room)
Power Supply Power/Data Cable (Blue)
Wireless Base Station Notes: • Place base station on a table or shelf out of the direct beam • Point the non-connector end of the base station toward the Daily QA 3 instrument when it is under the beam.
Figure 2-2. Typical Connection for Wireless Daily QA 3 (rf-Daily QA 3)
Hardware Setup
5
Daily QA 3 Wired Connection Only connect to a power source that matches the ratings on the power supply label
USB Cable
Power Supply
Power/Data Cable
Figure 2-3. Typical Connection for Wired Daily QA 3
6
Section 2. Setup
3
Software Setup
Add Users to the Database Before using the Daily QA 3 software, users must be added to the Daily QA 3 database. Adding users populates the User Name drop-down list on the Daily QA 3 ‘Login’ screen. Each user name must be unique but passwords are optional. A user role must be assigned: Physicist or Technician. Physicists have full access to the software; technicians cannot access the setup or calibration menus. At least one physicist role is required to set up a test plan. Perform this procedure on the computer where the Daily QA 3 database software is installed. 1
Double-click the SNC DB Admin icon in the system tray. If the symbol is hidden, click the ‘Show hidden icons’ button in the system tray.
Figure 3-1. Show Hidden Icons 2
In the ‘Daily QA 3 Database Administrator’ window, click the User box in the upper right corner.
Click to open the login data fields
Figure 3-2. Database Administrator Login 3
The user name, ADMIN, is populated by default. Enter the password, photon (all lower case) and press Enter or click Login.
4
Click New and enter the user information.
5
Click Save.
Launch Software and Connect to Database 1
Double-click the Daily QA 3 desktop icon
or go to Start > Programs > SNC Group >
Daily QA 3 > Daily QA 3 to launch the Daily QA 3 software. 2
If this is the first time the software has run on this computer, the ‘Program Preferences’ dialog box opens and prompts for the database location. Enter host name and click OK.
3
Close Program Preferences, exit and restart software. If the software finds the database, the Login screen appears.
4
If a device is connected, a background measurement begins. If the device is connected and not found when the software is launched, continue to the next step and then see the Daily QA 3 Reference Guide or the software Help Index.
5
Select a user name from the list, type the password (if required), and click Login. The software is ready to use when the database connection is established and the instrument is connected.
Add Users to the Database
7
Set Up Program Preferences The Daily QA 3 Program Preferences allow you to customize default options, such as temperature and pressure units, length of the startup background measurement, QA rooms to display on a computer, and trend chart settings. Click Setup > Program Preferences to set default options. For more details, see the Daily QA 3 Reference Guide or the software Help Index.
Calibration The Calibration menu contains five items ordered in a logical sequence that should be followed when using Daily QA 3 software. •
Calibrate Temperature and Pressure – See Calibrate Temperature and Pressure – Daily QA 2 and Daily QA 3 on page 8.
•
Calibrate Absolute Dose – See Calibrate Absolute Dose on page 8.
•
Calibrate Array – See Calibrate Array on page 9.
•
Calibration Management – This function enables import and deletion of qualifying array calibration files and can be performed as needed.See the Daily QA 3 Reference Guide or the software Help Index for details.
•
Export DQA 3 Calibration (for Argus) – This function is used to export an array calibration as a text file for Argus. For information about exporting a Daily QA 3 calibration for Argus, see the Daily QA 3 Reference Guide or the software Help Index.
Calibrate Temperature and Pressure – Daily QA 2 and Daily QA 3 Air density corrections derived from temperature and pressure sensors in the Daily QA device have a direct influence on output measurements, thereby influencing the Dose Calibration Factor of the central axis chamber (CAX). When the Daily QA 3 software is upgraded to version 2.5, the temperature and pressure sensors for each Daily QA 3 device must be calibrated using your temperature and pressure standards as a reference. These reference standards should be the same as those used for dosimetry calibration, thereby enabling a future intercomparison to your dosimetry standard. Any new or replacement Daily QA 3 instruments must also be calibrated for temperature and pressure prior to calibrating the absolute dose response. For more information, see the Daily QA 3 Reference Guide or the software Help Index. 1
From the menu, select Calibration > Calibrate Temperature and Pressure.
2
In the ‘Calibrate Temperature and Pressure’ dialog box, enter actual ambient temperature, to tenths of a degree, obtained from a calibrated thermometer in treatment room.
3
Enter actual pressure, to tenths of a unit, obtained from a calibrated pressure gauge in treatment room and click Apply.
Calibrate Absolute Dose This function converts the CAX ion chamber data from device response to dose. The air density corrections are part of this calibration, therefore the calibration of Temperature and Pressure will influence the CAX dose reference value. The Absolute Dose Calibration results in an Absolute Dose Calibration Factor, or Dose Factor, for short. The Dose Factor, a new database field in version 2.5 labeled as ‘AB_DOSE_CF’, is applied globally to all template measurements for a given device, hence the user will only perform the 'Calibrate Absolute Dose' function for a single energy per machine. For information about calibrating absolute dose, see the Daily QA 3 Reference Guide or the software Help Index.
8
Section 3. Software Setup
Absolute Dose Calibration Procedure 1
Set up the Daily QA 3 using a known absolute dose value for a given MU, field size, depth, and SSD. Note: Water equivalency from the top of the Daily QA 3 instrument to point of measurement is 1.0 cm. The physical distance from the top of the Daily QA 3 instrument to the point of measurement is 0.74 cm.
2
Select Calibration > Calibrate Absolute Dose. In the Available Devices panel the highlighted serial number is the device connected. The instrument image on the right shows the correct orientation for the measurement. Note: The Calibrate Absolute Dose function is only performed for a single energy per device. The resultant Dose Factor is automatically applied to all other energies for that device.
3
Click Start and deliver the MU for the known dose. Follow the instructions on the screen to complete the dose calibration. For additional details, see the Daily QA 3 Reference Guide or the software Help Index.
Calibrate Array This procedure is for Daily QA 3 only. For PROFILER 2 and TomoDose, use the separate instrument software to perform array calibration then import the resulting array calibration files into Daily QA 3 software. For details, see the Daily QA 3 Reference Guide or the software Help Index. If calibrating an array with an Elekta Beam Modulator™ linac limited to 16x21 cm field SAD, see the Daily QA 3 Reference Guide or the software Help Index for additional information. 1
For photons, adjust collimator jaws to 25 x 25; for electrons, install 25 X 25 cm cone.
2
From the menu, select Calibration > Calibrate Array…” The ‘Calibrate Array’ dialog box displays the Daily QA 3 as it should be oriented for the first array calibration measurement and prompts for the energy and beam type. To zoom in and change perspective, click the magnifying glass at the top right of the screen.
3
Enter energy and beam type (MV for photons, MeV for electrons) for the calibration and click Begin. The Daily QA 3 image rotates to the initial position and the ‘Initial Setup’ dialog box displays.
4
Follow the setup instructions and click Next. For additional calibration details, see the Daily QA 3 Reference Guide or the software Help Index.
5
When Step E of the Calibrate Array procedure is complete, click Finish. The calibration factors display. Click Print to print the results or save as a PDF, Accept to save the results to the database, or Close to close the screen without saving the results.
Create a QA Test Plan, Site, Room, Machine, Template Overview Note: Screens in this section show Daily QA 3 only; other device screens have a slightly different appearance. The QA test plan is a Scheme Tree that contains all the test and schedule details. As viewed in QA Mode, the Scheme Tree lists the sites, rooms, machines, and tests; it is linked to a calendar so that recurring tests appear on the days scheduled. Once the device is connected and setup for the test in the Scheme Tree, then as each test is completed, measurement data is stored in the database and the software advances to the next test. Test results are reviewed in the Trend Mode.
Create a QA Test Plan, Site, Room, Machine, Template
9
QA Mode View Scheme tree
Calendar
Switch between: QA Mode – create and execute a test plan Trend Mode – display trends over time in test plan data
Site + Room Machine
Device Setup
Template ID, (i.e., test names)
Machine Setup Data Panel Decision Keys
Results Panel
Figure 3-3. QA View
Set Up the Scheme Tree, Physicist Role The user setting up the Scheme Tree, adding machines, rooms, and templates must be assigned the Physicist role in the SNC Daily QA 3 database. See Add Users to the Database on page 7. Note: Daily QA 3 uses a relational database for recording data and its related site, machine, device, template, and user information. Once in the database, data cannot be deleted. However, obsolete information on site, machine, device, template, and user can be hidden from view when labeled Inactive. 1
Open the ‘Test Setup’ dialog box by clicking Setup > Scheme Tree….
2
Click New, enter all relevant information for the site. The Setup Overview panel is empty until the information is saved.
3
Click Save. The site appears on the Scheme Tree. Rooms can now be set up for the site.
4
In the Scheme Tree, click to select the Site, enter a room name, and then click +Add. Machines can now be added to the room.
5
In the Scheme Tree, select the room to which a machine will be added and then select the manufacturer and model of the machine from the drop-down lists. The manufacturer and model can be entered manually, if needed.
6
Using the drop-down list, select the serial number of the device that will be used for machine QA. If no device has ever been connected, no devices will be available in the list.
7
Select the array calibration file to use with the instrument. For Daily QA 3 instruments manufactured or serviced after June 2007, the factory calibration file may be used to obtain reasonable results.
8
Select energies to be tested. A test template is produced for each selected energy. Note: For accurate determination of change in electron energy, templates for all electron energies must be created prior to data collection. All electron energies do not need to be scheduled in the calendar, but must have templates created.
9
10
Click +Add to save changes in database.
Section 3. Software Setup
Set Up Template for Beam Tracking Test templates are added after a machine is created. Each test requires a specific measurement template defining test setup. 1
From the Daily QA 3 menu, select Setup > Scheme Tree.
2
In the Template Information panel, click New to create a test template,
3
a.
Enter or select relevant information for the measurement, including the calibration file.
b.
Define the schedule for the test and then click Save. See the Daily QA 3 Reference Guide or the software Help Index for template field descriptions.
To edit an existing template, select the template in the Setup Overview panel, and then click Edit in the Template Information panel.
Baseline Options Consider the following when selecting the baseline options for a template. •
Absolute mode evaluates results for axial symmetry, transverse symmetry, field size and field shift against a field with perfect symmetry and size.
•
Relative mode evaluates results against the measured baseline. If you are only interested in deviations from the baseline, use relative mode.
Example: During baseline measurement, if a beam has an axial symmetry of -0.6%, and a daily QA measurement a week later has an axial symmetry of -0.5%, then in absolute mode, it will show an axsym of -0.5%, but in relative mode it will show an axsym of +0.1%.
Baseline QA Template Before using a QA test template, it must be baselined with a QA measurement. All trend parameters are compared with the baseline data set until a new baseline is established. Each time the QA template is baselined, subsequent data is compared to the new baseline data, while all previous data is compared to the previous baseline. 1
From the Daily QA 3 software menu, select Setup > Baseline QA Template.
2
The QA Template Baseline Instructions form appears. Follow the onscreen instructions to complete the baseline procedure. See the Daily QA 3 Reference Guide or software Help Index for more detailed information. The baseline dose corresponds to the value that will represent your desired output baseline. For example, after an output measurement the Daily QA 3 software will subtract the baseline dose from the output measurement to calculate the percent difference. Suppose your output measurement is 102 and you entered 100 for your baseline dose; your percent difference for that output measurement would be 2%. The delivered dose on the other hand allows you to compensate for situations when you know the output of your machine is different from your desired baseline dose. For instance, if you want your baseline dose to be 100 (1.00 cGy/MU), but you know the current output of your machine is 2% hot (1.02 cGy/MU), enter 102 as your delivered dose. These two factors allow you to separate the value used as a baseline from the actual process of taking a baseline measurement.
Baseline QA Template
11
This page is intentionally left blank.
12
Section 3. Software Setup
4
QA Measurement
Preparing for Measurements To maintain a stable temperature, the instrument should remain powered on at all times, connected to power via the PDI 24 hours per day. Store the instrument in the treatment room, but not in a location where a primary beam may occur. If the Daily QA 3 is powered off, power it on and allow it to warm up for 10 minutes before using.
CAUTION: After charging an rf-Daily QA 3, allow electronics to cool down at least 15 minutes before making measurements. For more information about preparing for measurements, see the Daily QA 3 Reference Guide or the software Help Index.
Measurement Screen illustrations, orientation illustrations, and detector layouts in this section show Daily QA 3 only. When connected to other instruments, screens have a slightly different appearance. See the Daily QA 3 Reference Guide or the software Help Index for more information about QA measurements. 1
In the Scheme Tree, select the machine or template (A).
G
D A
B
C F E Figure 4-1. Steps to Make an Exposure 2
Click the drop-down handle (B) above the image of the instrument to display Daily QA 3 setup details. Position instrument as described.
3
For an rf-Daily QA 3, ensure: •
Power switch is ON.
Preparing for Measurements
13
•
Power/data cable is connected to the base station, base station is on a table or shelf out of direct beam, and non-connector end of the base station points toward the Daily QA 3 instrument.
CAUTION: Never place the base station in the direct beam. Direct radiation will damage the base station and void the factory warranty. 4
Set up exposure parameters on the linac to match values under Machine Setup (C).
5
Ensure Background text box on DAILY QA 3 toolbar displays Yes. If not, click Background and perform a background manually. See the Reference Guide or the software Help Index for instructions.
6
Click Start (D) and deliver the prescribed dose for the template. The beam status displays in the Status box in the DAILY QA 3 toolbar.
7
When the beam turns off, review the ‘Results’ panel. Values are color-coded: green = all passed; red = failure levels; yellow = values that fall between warning and failure levels. If desired, enter comments about the test in the Measurement Notes text box.
8
To adjust an output value, apply a correction factor to the measurement by clicking Template Dose Correction (E). The ‘Template Dose Correction’ dialog box opens. Enter the target dose and click Apply. The correction factor is applied automatically. For more information about the Template Dose Correction, see the Daily QA 3 Reference Guide or the software Help Index.
9
Click Record, Re-Do, or Cancel for the measurement (F). When Record is selected, the software automatically advances to the next test template for the machine and initiates data collection. Continue delivering the prescribed dose for each template and record the results until all daily tests are complete.
10 To view trends for all templates, click Trend (G) in the top center of the screen. See the Daily QA 3 Reference Guide or the software Help Index for trending and reporting details.
Printing Reports Reports can be printed for single test or multiple tests. 1
From the Daily QA 3 menu, select File > Reports…
2
At the top of the ‘Reports’ dialog box, select the type of report: All Energies for Day, Multiple Day QA, Output Report, or Array Calibration.
3
•
If you select Array Calibration as the report type, select the calibration from the list of devices and calibrations.
•
For other types of reports, select the desired template, dates, and range of records.
Click Print.
Exporting Data Selected raw data can be exported to a text (.txt) file and then opened using a spreadsheet application such as Microsoft Excel to analyze the data. See the Daily QA 3 Reference Guide or the software Help Index for instructions.
14
Section 4. QA Measurement
5
Support and Maintenance
Maintaining Hardware Repair There are no user-serviceable components in the Daily QA 3 or the Power Data Interface (PDI). If there are problems with the devices, contact Sun Nuclear. See Contacting Sun Nuclear Support on page 19.
CAUTION: Do not open the housing unless you are performing an upgrade that has been approved by Sun Nuclear Corporation. If the Daily QA 3 housing or casing is damaged such that the device integrity is compromised, return the device to Sun Nuclear Corporation for repair. Inspect cables periodically for damage. Return damaged cables to Sun Nuclear Corporation for repair or replacement if any mechanical or electrical degradation is suspected. Additional service information is available on the Sun Nuclear web site: www.sunnuclear.com/support
Cleaning Clean the instruments with a soft dry cloth. Do not use liquid cleaners, solvents, or abrasives.
Storage Store the instrument in an indoor, protected environment where the instrument will not be irradiated by the direct beam.
Disposing and Recycling The Daily QA 3 instrument contains electrical components. In some countries the disposal of electrical components is subject to special requirements. When the components are no longer functional or are otherwise ready to be discarded, recycle or dispose of them according to local waste management or recycling regulations.
Minimizing Radiation Damage Although detectors are designed to be exposed directly to the beam, the electronic processing circuits can be damaged by direct exposure to ionizing radiation. The electronic circuits are located in the raised end of the instrument behind the line that is labeled “KEEP DIRECT BEAM ABOVE THIS LINE”. To minimize damage to electronics: • Never allow any direct radiation beyond the “KEEP DIRECT BEAM ABOVE THIS LINE” mark on top surface of device. • Never expose electronics portion of instrument to beam. • Never leave the device on couch during warm-up. • Never store device where it can accidentally be irradiated, such as a shelf in the treatment room that is irradiated by the beam when the linac is rotated. • Avoid frequent use of thick buildup that may cause scattered radiation that will damage the electronics. Note that adding extra buildup is not normally necessary for daily QA measurements and may increase the scattered radiation to the electronics. Note: The Daily QA 3 firmware cannot be upgraded if the device is connected wirelessly.
Maintaining Hardware
15
Troubleshooting Daily QA 3 Troubleshooting Indication Software does not recognize instrument
Probable Cause Not properly connected to USB port.
Remedy Check USB cable connection to PDI and PC.
Wrong operating system. Must be Windows 10 or Windows 7.
Data appears incorrect
USB drivers not set up correctly.
Reinstall client software. The correct USB drivers are installed with software.
Wrong calibration file.
Use correct calibration file.
Improperly calibrated.
Recalibrate.
No background collected.
Manually collect background and repeat measurement. See the Reference Guide or the software Help Index for instructions.
Conditions changed since background taken. A QA measurement reports 0.00 cm field size
Possible cause: • Setup error. • Electron field size calculation for Daily QA 3 is sensitive to array calibration, so you may need to acquire an array calibration for the specific energy.
Screen display size is incorrect Monitor’s DPI setting is wrong. rf-Daily QA 3 battery is not charging
16
• First, rule out a setup error, such as a slight shift or wrong SSD or linac energy. • If correcting setup doesn’t resolve the issue, acquiring an array calibration for the specific energy may result in a more accurate field size since electron field size calculation for Daily QA 3 is sensitive to array calibration. • If it is impossible to determine field size with a particular cone/linac system, Sun Nuclear recommends examining relationships between the diode detectors that straddle the 20 cm field marks for a commissioned system, and then baseline QA measurements from those relationships. Edit the monitor properties and set it to Normal size (96 DPI). Contact your IT department if you need assistance.
The instrument is stored To reset the charge circuit and allow the device to off AC power for an begin charging again, apply AC power, then lightly extended period of time press and hold the recessed override switch for 3 seconds using a pointed, non-metallic object. See rf-Daily QA 3 End Panel on page 18.
Section 5. Support and Maintenance
Database Connection Troubleshooting Problem
Probable Cause
Client program cannot locate database
User cannot log in to database
Solution
Incorrect address entered in client Enter correct network address in software. client program. Client program set to “localhost” when using a server database.
Enter correct address in client program.
Database not installed on same network with client.
Make sure client computer is on same network as server.
Incorrect password.
Delete user and add another user with same user name and correct password. See Add Users to the Database on page 7.
User name not set up in database. Add user to database. See Add Users to the Database on page 7. User does not have permission to write data to server.
Request that the server’s System Administrator add permissions in Windows to allow users to read and write to database.
No password required for user
No password set up.
Add a password to user. See Add Users to the Database on page 7.
Cannot edit Daily QA 3 templates
User logged in as a technician.
Change user role to a physicist. Only physicists permitted to change templates. See Add Users to the Database on page 7.
Low energy electron measurement Possible cause: • Check for a setup error, such as (6 MeV or lower) reports 0.00 cm • Setup error. a slight shift or the wrong SSD or field size • Electron field size calculation for linac energy Daily QA 3 is sensitive to array • Acquiring an array calibration for calibration, so you may need to the specific energy since acquire an array calibration for electron field size calculation for the specific energy. the Daily QA 3 is sensitive to array calibration. • The cone size will define the electron field size. Therefore, if the corrections above do not address the field size issue, use the electron cone as a guide in determining your electron field size.
Daily QA 3 End Panel
PWR/DATA Connector
LEDs A, B, C, D
Device Action B
C
D
Power On Sequence (bootloader active)
OFF
OFF
OFF
Three (3) toggles 1 sec. apart
Idle Mode
OFF
OFF
Toggle (1/2 sec.)
OFF
Acquiring Measurements
ON
OFF
Toggle (1/2 sec.)
OFF
–
ON
–
OFF
Device Fault
Troubleshooting
LED A
17
The last item, 'Device Faults', shows the pattern if the firmware detects a fault in instrument. If a 'Device Fault' is indicated, contact Sun Nuclear Corporation Support. See Contacting Sun Nuclear Support on page 19. A device fault may be caused by: 1
Pressure reading beyond limits: 115 kPA
2
High voltage reading beyond limits: < -400 V or > -300 V
3
ADC rails, upper or lower limits were hit. If the measurement circuit (including ADC) is intact and LED B comes on after the device is powered up, the cause could be either item 1 or 2. If LED B is lit after starting a measurement, item 3 is the likely cause.
rf-Daily QA 3 End Panel 1 2
3
12 No Description RECEIVE 1 TRANSMIT 2 CHARGE STATUS 3
4
5
6 7 8 9 10
11 Function
Green LED flashes when rf unit is receiving. Green LED flashes when rf unit is transmitting. Ready—Green LED is lit when all battery charging parameters are within limits.
4 5 6 7
CHARGE STATUS
Charge—Yellow LED is lit when batteries are charging.
CHARGE STATUS
Fault—Red LED is lit when a charging fault is detected.
LOW BAT
Red LED turns on when approximately one hour of battery power remains.
Recessed Battery override switch *
Manual override to restart battery charging when batteries have been deeply discharged. See Daily QA 3 Troubleshooting on page 16.
8 9
ON
LED is green when power is applied via battery or PDI cable.
BAT PWR
Push button battery power switch (in = ON; out = OFF). The battery power switch should always be “OFF” when instrument is stored or if instrument is being shipped.
10
+18 VDC
+18 VDC input power jack for charging only. Connecting power to the +18 VDC jack causes the instrument to turn off during charging cycle.
11 12
Status LEDs
Status LEDs for basic Daily QA 3. See Daily QA 3 End Panel on page 17.
PWR/DATA connector
Power/data cable connector. Not used with rf operation but can be used instead of the rf link if the rf link is inoperable. To use as a wired instrument, disconnect the power/data cable from the Base Station and plug it in here.
* To reset the charge circuit and allow the device to begin charging again, apply AC power, then lightly press and hold the recessed override switch for 3 seconds using a pointed, non-metallic object.
PDI Troubleshooting See Technical Bulletin 2-09, Power Data Interface for: •
A list of problems which may occasionally occur when using the PDI.
•
An explanation of the controls and indicators on the PDI.
The bulletin can be downloaded from the Sun Nuclear Support website for Daily QA 3.
18
Section 5. Support and Maintenance
Contacting Sun Nuclear Support You may request support in two ways: •
Contact the Sun Nuclear Support team by telephone: +1 321-259-6862.
•
Submit a support request using our online form. See Support Website below.
Support Website Visit the Sun Nuclear Support web site to request support via an online form and access the latest product updates, such as software/driver downloads, documentation. 1
Open an internet browser and navigate to www.sunnuclear.com/support.
2
Enter your email address and the serial number of any Sun Nuclear product used at your facility, then click Log in to open the Support Dashboard.
Warranty The Sun Nuclear Product Warranty can be found on the Sun Nuclear Support web site, www.sunnuclear.com/support. Log in using your email address and the serial number of any Sun Nuclear instrument used at your facility.
Contacting Sun Nuclear Support
19
6
Specifications
Recommended System Requirements
Category Recommended System Requirements
Characteristic
Specification
Operating System
Windows 10 or Windows 7, any edition, 32- or 64-bit
Computer hardware
• • • •
Pentium III processor Hard disk space: 70 MB 512 MB RAM 4 MB VGA video card capable of at least 1024 x 768 (16 MB video RAM recommended) • USB version: 2 (one available USB port) • OpenGL hardware accelerated video card.
Daily QA 3 Specifications For more information, see the Daily QA 3 Reference Guide or the software Help Index. Category Detectors
Characteristic Type detectors
Specification Primary—Vented ionization chambers, fully guarded Secondary—Diode detectors (for penumbra detection)
Quantity of detectors Ionization chambers (13): • One CAX detector at center of the field • Four primary detectors on X and Y axes, 8 cm from center • Four X-energy detectors, 11.3 cm from center • Four e-Energy detectors, 5.6 cm from center on diagonals Diode detectors (12) (5 mm spacing for penumbra detection): • A row of three detectors on X axis centered at +10 cm • A row of three detectors on X axis centered at –10 cm • A row of three detectors on Y axis centered at +10 cm • A row of three detectors on Y axis centered at –10 cm Ionization chambers
Volume of 9 center, primary, and X-energy chambers—0.3 cm3 Parallel plate separation—4 mm Electrode diameter—9.8 mm diameter, carbon Volume of 4 e-Energy check chambers—0.6 cm3 Parallel plate separation—4 mm CAX collection electrode—1.38 cm diameter, carbon
Performance
Environment
Diodes
Active size—0.8 x 0.8 mm Type—n-type, radiation hardened
Electron energy attenuation
Top right (ETR)—none Bottom right (EBR)—0.216 in. copper (Cu) Bottom left (EBL)—0.216 in. aluminum (Al) Top left (ETL)—0.150 in. iron (Fe)
Radiation measured
Photons—Co-60 to 25 MV Electrons—4 MeV to 25 MeV, ± 1% for 6-20 MeV
Beam limits
Leakage—±0.05 cGy/min Dose range—350 cGy Maximum dose per linac pulse—5.5 cGy/pulse Maximum average dose rate—250 cGy/s
Operating
Ambient Temperature: 18 to 30° C (64.4 to 86° F) Relative Humidity: 20 to 80% relative humidity
Storage
Temperature: –30 to 50° C (–22 to 122° F) Relative humidity: 10 to 90% non-condensing Pressure: 90.6 kPA to 102 kPa
20
Section 6. Specifications
Category Medical Electrical Equipment Classification
Buildup and backscatter
Characteristic
Specification
Basic Safety and Essential Performance
Certified to meet IEC 60601-1.
Electromagnetic Performance Class I
Certified to meet IEC 60601-1-2.
Inherent buildup over Actual: Acrylic, 0.74 cm primary ion chambers Effective: 1.0 ± 0.1 g/cm2 Inherent buildup over Actual: Acrylic, 0.77 cm diode detectors Effective: 1.0 +- 0.1 g/cm2
Temperature and pressure
Physical
Inherent backscatter
Actual: FR-4 0.16 cm; Acrylic, 1.27cm Effective: 1.7 g/cm2
Electron energy chamber buildup
Chamber eTR, air/acrylic buildup, 0.2 g/cm2 Chamber eBR, Cu/acrylic buildup, 4.3 g/cm2 Chamber eBL, Al/acrylic buildup, 0.7g/cm2 Chamber eTL, Fe/acrylic buildup, 3.0 g/cm2 Chamber CAX, acrylic buildup, 1.0 g/cm2
Temperature
Sensor—1 integrated circuit plus 5 thermistors Range—0°C to +50°C Non-linearity—±0.67% over measurement range Accuracy—±2.0°C without user calibration Measurement Resolution—0.05°C Display Resolution—0.1°C
Pressure
Sensor—Temperature compensated on-chip bipolar operational amplifier and thin film resistor network Measurement, Slope Transfer Function between Output voltage and Supply voltage Range—15 to 115 kPa (112 to 862 mmHg) (1 atm = 101.33 kPa) Non-linearity—±1.5% over measurement range Accuracy—±2kPa without user calibration Measurement Resolution—0.0035kPa Display Resolution—0.01kPa
Dimensions
25.6 cm wide, 40.8 cm long, 4.6 cm thick (10.09 x 16.07 x 1.82”)
Weight
5.7 kg (12 lb.)
Connection ports and POWER/DATA—8 pin DIN, provides power to Daily QA 3 and biindicators directional data between PC and Daily QA 3 LEDs—A, B, C, D: Status indicators Power supply
Service life
10 years
Type
Switching, general purpose IEC 320 socket, Class I power inlet for straight cable entry
Supply voltage range 100 to 240 VAC, automatically switching; 47-63 Hz
Cables
DC power output
+18 VDC, 1 amp, 2.5 mm center positive plug
Dimensions
8.7 cm long, 4.7 cm wide, 3.3 cm high
Weight
0.17 kg (6 ounces)
POWER/DATA
25 m power/data, 8 pin DIN, connects Daily QA 3 to PDI
USB
2 m USB, 4 pin, A and B plugs, connects PDI to computer
Power cord
2 m power, IEC straight plug line cord, 3 conductor
Recommended System Requirements
21
rf-Daily QA 3 Specifications Category Power
Physical
Characteristic
Specification
Base station
+18 VDC
Battery charging
+18 VDC
Battery capacity
1.7 or greater amp hours @ 9.6 V
Estimated battery life
5 hours continuous operation
Base station dimensions
4.7 x 2.75 x 1.4 in. (12 x 7 x 3.5 cm)
Base station weight
0.7 lb. (0.32 kg)
Service life Transceiver * Frequency Power
10 years 2.400 to 2.485 GHz 1 mW (0 dBm)
Modulation
O-Q PSK
Baud rate
250 kb/s between transceivers; 19.2 kb/s over wires between PC and/ base station
Antenna
Integral in housing
Duty cycle
Software determined
Interference outside shielded room
Designed to be compatible with existing Wi-Fi
Agency approvals
• • • • •
Antenna
• A24-C1, Surface mount or embedded PCB
FCC Part 15.247—OUR-XBEE (MaxStream) Industry Canada (IC)—4214A-XBEE (MaxStream) Europe—ETSI Japan—R201WW07215214 Australia—C-Tick
* PN 1093500Z contains XBee radio XB24-ACI-001 or XB24-API-001(rf assembly, shipped with rf integrated) PN 1093400Z contains XBee radio XB24-ACI-001 or XB24-API-001 (base station)
22
Section 6. Specifications
7
Software Reference
Daily QA 3 Client Software Interface For more complete software interface details, see the Daily QA 3 Reference Guide or the software Help Index. When launched, the Daily QA 3 client software opens in the main view (Figure 7-1). 6 1 2 3 7
4 8 5 No. 1 2 3
Name
Description
Menu bar
Provides access to commands.
Toolbar
Provides quick access to common program functions and views.
Scheme Tree schedule
A tree-like representation of the schedule for the selected date that shows the sites, rooms, machines, and tests to be performed. See Scheme Tree on page 24.
4 5 6
Data
Displays the real-time values of the primary detectors during an exposure.
7 8
Status bar
Displays temperature and pressure measured by the instrument’s sensors.
Calendar
Displays the current month. The current date is selected by default, but other dates can be selected. The Scheme Tree is linked to the selected date to show only machines and tests for that day.
Test setup details
Shows the details that a technician needs to set up the selected test.
Results
Displays a summary of the calculated results of the completed test. See Results on page 24.
Figure 7-1. Main View of Daily QA Software Window
Daily QA 3 Client Software Interface
23
QA View QA view is used to create a test plan and collect a measurement. Scheme tree
Calendar
Setup
Show/hide the machine setup details
Data Panel
Measurement Buttons
Results Panel
Figure 7-2. QA View Scheme Tree The Scheme Tree is an indented list of sites, rooms, machines, and tests scheduled for the selected date. Select a day on the calendar to display the tests for that day. Test templates are indented under each machine. Items that appear in the Scheme Tree are set up using the ‘Scheme Tree Setup’ forms. To set up or change the tree, log in as a physicist, then select Setup > Scheme Tree from the menu. Tests can be set up on a daily, weekly, monthly, or annual basis. However, tests scheduled for a given day are only visible when that day is selected on the calendar. The Scheme Tree can be altered to only show tests for a single room by configuring Program Preferences (Setup > Program Preferences > Scheme Tree tab). See the Daily QA 3 Reference Guide or the software Help Index for more details. Results The Results panel shows calculated results of a QA measurement. The Results panel appears in both QA view and Trend view, with minor differences. For more details, see the Daily QA 3 Reference Guide or the software Help Index.
Figure 7-3. Measurement Results Panel
24
Section 7. Software Reference
Trend View The Trend view (Figure 7-4) displays a summary of data that was collected for a selected template. It can also display all of the measurements made for that template since the template was created. To access the Trend view from the main QA view, click the Trend button on the toolbar. For more details, see the Daily QA 3 Reference Guide or the software Help Index. Template List
Trend Graph
Trend button
Graph Controls
Selected Point Symbols
Limit Line
Data Panel
Results Panel
Figure 7-4. Trend Graph View Graph Controls Controls above the graph affect the appearance of the data in the graph. If graph controls are not visible, click the arrow at the top, center of the graph; click again to roll them up. For more details, see the Daily QA 3 Reference Guide or the software Help Index. Symbols The symbols displayed in the Results panel match the symbols in the graph. Dose – Central axis (CAX) dose in cGy AxSym – Axial Symmetry percent TrSym – Transverse Symmetry percent QAFlat – Flatness percent e-Energy – Electron energy percent change in flatness X-Energy – Photon energy percent change in flatness Light-field Coincidence – XSize, XShift, YSize, YShift in cm
Figure 7-5. QA Trend Graph Symbols For a detailed explanation of these parameters, see the Calculations section in the Daily QA 3 Reference Guide or the software Help Index.
Daily QA 3 Client Software Interface
25
This page is intentionally left blank.
26
Section 7. Software Reference
Appendix A: Regulatory Supplement Sun Nuclear Corporation Symbols The following symbols are used in this document and in Sun Nuclear Corporation’s product labels.
WARNING: This symbol indicates a risk of electric shock. (EN ISO 7010)
WARNING: This symbol indicates a hazard that could result in major injury or equipment damage. (EN ISO 7010) CAUTION: This symbol indicates a potential hazard that could result in minor injury or equipment damage. (EN ISO 15223-1)
Note: Important or supporting information.
Manufacturer’s Identification (name and address). (EN ISO 15223-1)
Date of Manufacture. (EN ISO 15223-1) Serial Number. (EN ISO 15223-1) Catalog Number. (EN ISO 15223-1) Read the instructions in this manual before using.
Do not throw in trash; dispose of in an environmentally friendly way. (EN 50419) This symbol indicates conformity with EC Directives. Authorized representative in the European Community. (EN ISO 15223-1)
Operator Responsibility This manual is intended for an operator who has a working knowledge in radiation physics and dosimetry and who has experience working with treatment machines and accessories. The operator of the device bears the full responsibility for validating measurement results. The device and its accessories must not be used for any other purpose than described in this manual (see Intended Use on page iii). Violation may result in loss of warranty. •
This equipment must be used in accordance with the instructions in this manual. Read all instructions and safety labels before use.
•
Do not use with any power source that does not match the power ratings listed on the power supply.
Sun Nuclear Corporation Symbols
27
•
When connecting the power supply to AC power, ensure that the plug is not obstructed and can be easily disconnected.
•
Do not permit any short circuit of AC power that may be hazardous to users.
•
Do not permit any liquids to spill onto the instrument or any associated parts or accessories.
WARNING: When using the Isocentric Mounting Fixture (IMF), stay clear of the IMF and the instrument while rotating the gantry to prevent injury. For instructions to report health or safety related concerns, see Reporting Health or Safety Related Issues or Concerns below.
Modifications to the Equipment Any changes or modifications to the device that are not expressly approved by Sun Nuclear Corporation could void your warranty.
Environmental Requirements Operating The device must be operated within normal laboratory conditions maintained for human comfort: Temperature: 18 to 30° C (64 to 86° F); Relative Humidity: 20 to 80%.
Storage The device should be stored in a climate-controlled environment.
Radio Interference The rf-Daily QA 3 complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference and (2) this device must accept any interference received, including interference that may cause undesired operation.
WARNING: To satisfy FCC RF exposure requirements for mobile transmitting devices, a separation distance of 20 cm (8 inches) or more should be maintained between the rf-Daily QA3 and persons during device operation. Operation at closer than this distance is not recommended. The device must not be co-located in conjunction with any other antenna or transmitter.
Reporting Health or Safety Related Issues or Concerns Should the need arise to report any safety or health related issues or concerns regarding the use of Sun Nuclear products, Sun Nuclear should be contacted directly or in the case of our European customers, our Authorized European Representative. Sun Nuclear's Authorized European Representative is: EMERGO EUROPE Prinsessegracht 20 2514 AP The Hague The Netherlands
Australian Sponsor alphaXRT Pty. Ltd. Suite 1.15, Level 1, 90-96 Bourke Road, Alexandria NSW 2015 Australia
28
Appendix A: Regulatory Supplement
Connection to External Computer Equipment To maintain compliance with IEC 60950-1, Safety of Information Technology Equipment, the PDI USB port must be connected to external computer equipment that is compliant with IEC 60950-1.
EMC Precautions The instrument complies with IEC 60601-1-2, Electromagnetic Performance standards, and the operator should ensure that it is used in such an environment. Table 1. Manufacturer’s Declaration - Electromagnetic Emissions Emissions Test
Compliance
Electromagnetic Environment - Guidance
RF Emissions CISPR 11
Group 1
The device uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
Harmonic Emissions IEC 61000-3-2
Class B
Voltage fluctuations flicker emissions IEC 61000-3-3
Class D
The device is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Table 2. Manufacturer’s Declaration - Electromagnetic Immunity for all Equipment Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic Environment - Guidance
Electrostatic discharge ±6 kV contact (ESD) ±8 kV air IEC 61000-4-2
Passed 2 kV, 4 kV and 6 Floors should be wood, concrete or kV contact discharge ceramic tile. If floors are covered with Passed 2 kV, 4 kV and 8 synthetic material, the relative humidity should be at least 30%. kV air discharge
Electrical fast transient/ ±2 kV for burst power supply lines IEC 61000-4-4 ±1 kV for input/ output lines
Passed 2 kV to power lines
Surge IEC 61000-4-5 Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11
Passed 1 kV to signal bundles
±1 kV line(s) to Passed up to 2 kV line(s) (L-GND) ±2 kV line(s) to Passed up to 1 kV earth (L-L)
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment.
Passed < 5%, 40% and Mains power quality should be that of a 70% typical commercial or hospital environment. If the user of the device Passed 95% requires continued operation during interruptions power mains interruptions, it is 40% UT recommended that the 1D Scanner be powered from an uninterruptible power (60% dip in UT) supply or a battery. for 5 cycles
95% dip in UT) for 0.5 cycles
70% UT (30% dip in UT) for 25 cycles 95% dip in UT) for 5 cycles Power frequency (50/60 3 A/m Hz) magnetic field IEC 61000-4-8
Passed up to 3A/m
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.
NOTE: UT is the a.c. mains voltage prior to application of the test level.
Connection to External Computer Equipment
29
Table 3. Manufacturer’s Declaration - Electromagnetic Immunity for Systems that Are Not Life Supporting Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic Environment - Guidance Portable and mobile RF communications equipment should be used no closer to any part of the device including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter. Recommended separation distance
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2.5 GHz
3 Vrms
3 V/m
3.5 d = ------V1
P
Recommended separation distance
3.5 d = ------V1
P =80 MHz to 800 MHz
7 d = -----E1
P =800 MHz to 2.5 GHz
where: • P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer, and • d is the recommended separation distance in meters (m). aField
strengths from fixed RF transmitters, as determined by an electromagnetic site survey, should be less than the compliance level in each frequency range.
CAUTION: bInterference may occur in the vicinity of equipment marked with the following symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the device is used exceeds the applicable RF compliance level above, the device should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the device. b. Over the frequency range of 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
30
Appendix A: Regulatory Supplement
Table 4. Recommended Separation Distances between Portable and Mobile RF Communications Equipment and the Device The device is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the device as recommenced below, according to the maximum output power of the communications equipment.
Rated Maximum Output Power of Transmitter (Watts)
Separation Distance According to Frequency of Transmitter (meters) 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz 3.5 d = ------- P 3.5 7 V1 d = ------- P d = ------ P E1 E1
0.01
0.12
0.12
0.23
0.1
0.37
0.37
0.74 2.33
1
1.17
1.17
10
3.69
3.69
7.38
100
11.67
11.67
23.33
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. NOTE 1. At 80 MHz and 800 MHz, the separation distance for the higher frequency applies. NOTE 2. These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.
EMC Precautions
31
This page is intentionally left blank.
32
Appendix A: Regulatory Supplement
Corporate Headquarters 3275 Suntree Boulevard Melbourne, FL 32940 USA
+1 321 259 6862 sunnuclear.com
View more...
Comments