09 11 2006 Chromatography

May 1, 2018 | Author: api-3696530 | Category: Chromatography, Elution, High Performance Liquid Chromatography, Solvent, Laboratories
Share Embed Donate


Short Description

物理生化學-09-11-2006 Chromatography-黃文鑫...

Description

WenHsin Huang

Principles of Chromatography Principles of Chromatography

•Introduction to Analytical Separations –Solvent Extraction –What is Chromatography? –A Plumber’ s View of Chromatography –Efficiency of Separation –Why Bands Spread

Introduction to Analytical Separations

Wen-Hsin Huang Ph. D. NDMC 2006/9/7

WenHsin Huang, NDMC

1

2006/9/7

WenHsin Huang, NDMC

2

Solvent Extraction

Simple Separations

Solute particles only have one choice of solvent

•Extraction –transfer of solute from one phase to another –phase can be gas, solid, liquid

Solute particles Solute particles now have two choose preferred choices of solvent solvent

•Liquid/liquid extraction –2 immiscible solvents used –typically aqueous solvent and organic solvent

add second immiscible solvent

•you know water and oil don’ t mix

–organic solvents less dense than water –organic solvents more dense than water •chloroform, carbon tetrachloride, dichloromethane 2006/9/7

NDMC

WenHsin Huang, NDMC

shake

Legend: aqueous solvent organic solvent solute particles

•diethyl ether, toluene, hexane

3

2006/9/7

WenHsin Huang, NDMC

4

1

WenHsin Huang

Solvent Extraction

Phase Partitioning If we have a solute (S) that is partitioned between two phases (1 & 2) then we can write an equilibrium expression for this equilibrium.

•Like dissolves like –solute chooses solvent most like itself –polar compounds (and ionic compounds) choose water (which is very polar) –nonpolar compounds choose nonpolar organic solvents

S1  S 2

AS [S ] K 2  2 AS1 [ S1 ] Partition Coefficient (K)

• Convention: Organic phase = phase 2 2006/9/7

WenHsin Huang, NDMC

5

Distribution Coefficient

2006/9/7

WenHsin Huang, NDMC

6

Example of Extraction with n-Hexane The distribution coefficient for compound Z between n-hexane and water is 6.25. Calculate the percent Z remaining in 25.0 mL of water that was originally 0.0600 M in Z after the extraction with the following volumes of n-hexane. A) One 25.0-mL portion 13.8%

The distribution coefficient (D) is used in place of the partition coefficient when dealing with a species that has more than one chemical form.

total _ concentration _ in _ phase _ 2 D total _ concentration _ in _ phase _ 1

B) Two 12.5-mL portions

1.90%

C) Five 5.00-mL portions

4.99 x 10-3%

D) Ten 2.50-mL portions

2.49 x 10-7%

It is more efficient to do several small extractions rather than one large extraction. 2006/9/7

NDMC

WenHsin Huang, NDMC

7

2006/9/7

WenHsin Huang, NDMC

8

2

WenHsin Huang

Solvent Extraction (pH effects)

Solvent Extraction (pH effects)

•Charges of acidic and basic species change with pH

•Distribution of HA between two solvents –two equilibria to consider •Ka equilibrium •D equilibrium

–neutral species generally more soluble in organic solvent –charged species generally more soluble in aqueous solution

•Distribution coefficient (D) describes distribution of species between two phases –Takes into account all forms of a compound (i.e. H2A, HA-, A2-) –Different from partition coefficient (K) 2006/9/7

WenHsin Huang, NDMC

9

organic

HA

aqueous

HA

2006/9/7

Solvent Extraction (pH effects)

D=

[HA]1 + [A-]1

[HA ]2 K [HA ]1 •Substitute and rearrange to get K[H] D  [H ] K a

[B]2 [B]1 + [BH+]1

•Ionic species only found in aqueous layer 2006/9/7

NDMC

WenHsin Huang, NDMC

WenHsin Huang, NDMC

10

[H][ A ] K [HA ] Ka   [ A ]  a  [HA ] [H ] •Also have K expression for [HA]

total conc. in phase 1 (aqueous) D=

H+ + A-

•Distribution of HA between two phases

total conc. in phase 2 (organic)

[HA]2

Ka

Solvent Extraction (pH effects)

•Distribution of acids and bases between two phases D=

H+ + A-

11

2006/9/7

WenHsin Huang, NDMC

Eq’ n 12

3

WenHsin Huang

Solvent Extraction (pH effects) D for a base (B)

K[H] D  [H ] K a

KK a D K a [H]

•Adjust pH to effectively extract an acid or base (and its conjugates) into aqueous layer •Consider base B with a pKa of 9.0

Eq’ ns

–pH < 9.0, B hydrolyzes so majority of B (in the form BH+) in water layer (low D value)

•Distribution of acids and bases between two phases is pH-dependent •To find D, need to know

0

[ B] K  org [ B ]aq [ H ][ B]aq Ka  [ BH ]aq

–pH of solution –Ka of acid or conjugate acid –K of acid or base 2006/9/7

13

History of Chromatography

2006/9/7

pKa

-4

mainly BH+ 2

WenHsin Huang, NDMC

4

6

pH

mainly B 8

10

12

Fig.14

Tswett’ s Experiment

Mikhail Tswett is credited with the invention of chromatography.

The separated species appeared as colored bands on the column.

-2

-6

[ B ]org K Ka D  KB [ B ]aq [ BH ]aq K a [ H ]

WenHsin Huang, NDMC

Developed a technique that separated various plant pigments (e.g., chlorophylls and xanthophylls) by passing solutions through glass columns filled with finely ground CaCO3.

log D

D for an acid (HA)

Solvent Extraction (pH effects)

Ground up plant leaves and added petroleum ether.

Filled column with chalk. M. S. Tswett Russian Scientist (1872-1919)

Method forgotten for many years.

Hence Greek chroma = color and graphein = writing 2006/9/7

NDMC

WenHsin Huang, NDMC

15

2006/9/7

WenHsin Huang, NDMC

16

4

WenHsin Huang

What is Chromatography? Tswett’ s Experiment

•Method to separate components in a mixture based on different distribution coefficients between the two phases •Same principle as solvent extraction (like dissolves like), but one phase is “ stationary” and one phase is “ mobile” –no longer working with organic and aqueous layers in a separatory funnel –working in a column (which is just a tube)

2006/9/7

WenHsin Huang, NDMC

17

2006/9/7

WenHsin Huang, NDMC

18

What is Chromatography? Types of Chromatography

•Two phases: mobile and stationary •Mobile phase is solvent moving through column

•Adsorption Chromatography •Partition Chromatography •Ion-exchange Chromatography •Molecular Exclusion Chromatography •Affinity Chromatography

–liquid (Methanol, water, buffer) –gas (He, H2, N2)

•Stationary phase fixed inside column –viscous liquid coated on inside of column –solid particles packed inside column

•Solutes have different affinities for mobile phase and stationary phase 2006/9/7

NDMC

WenHsin Huang, NDMC

19

2006/9/7

WenHsin Huang, NDMC

20

5

WenHsin Huang

Adsorption Chromatography

Partition Chromatography

Stationary phase Mobile phase

Stationary phase

Mobile phase

Solid

Liquid on solid support

Liquid or Gas

Liquid or Gas

How it separates

How it separates

–solute adsorbs on to stationary phase surface

2006/9/7

–solute dissolves into liquid coating

WenHsin Huang, NDMC

21

2006/9/7

WenHsin Huang, NDMC

Molecular Exclusion Chromatography

Ion-Exchange Chromatography Stationary phase Anions or cations covalently bound to solid stationary phase

22

Mobile phase Liquid

Stationary phase

Mobile phase

Porous gel

Liquid

How it separates –small molecules trapped in pores of stationary phase

How it separates solute ions of opposite charge attracted to stationary phase

2006/9/7

NDMC

WenHsin Huang, NDMC

Also called gel filtration, gel permeation, or size-exclusion chromatography 23

2006/9/7

WenHsin Huang, NDMC

24

6

WenHsin Huang

Classification Liquid chromatography (LC) (Mobile phase a liquid)

Affinity Chromatography Stationary phase

Mobile phase

Immobilized molecules on Liquid liquid or solid stationary phase

How it separates –Molecules with specific shape dock with ligands Gas chromatography (GC) (Mobile phase a gas)

2006/9/7

WenHsin Huang, NDMC

25

2006/9/7

Basic Principles

Specific Method Stationary Phase Liquid-liquid or Liquid adsorbed onto partition solid

Type of equilibrium Partition between immiscible liquids

Liquid-bonded phase

Organic species bonded to solid surface

Partition between liquid and bonded surface

Liquid-solid, or adsorption

Solid

Adsorption

Ion exchange

Ion-exchange resin

Ion exchange

Size exclusion or Interstices of polymeric molecularsolid exclusion

Sieving

Affinity Covalently bonded Chromatography molecule (eg antibody)

Se l e c t i veAf f i ni t y–( e gs pe c i f i c Protein/antibody bind together)

Gas-liquid

Liquid adsorbed onto solid surface

Partition between gas and liquid

Gas-bonded phase

Organic species bonded to solid surface

Partition between gas and bonded liquid

Gas-solid

Solid

Adsorption

WenHsin Huang, NDMC

26

Simple Column

Two phases considered: Fresh eluent Initial band of A and B

1) Mobile Phase: solvent moving through the column. 2) Stationary Phase: stays in place inside of the column.

Column Packing

“ Eluent”

COLUMN

“Eluate”

Porous Disk

Process is called “ elution”

2006/9/7

NDMC

WenHsin Huang, NDMC

27

2006/9/7

WenHsin Huang, NDMC

28

7

WenHsin Huang

Chromatogram

Chromatographic Terms

A chromatogram is a graph showing the detector response (proportional to concentration) as solutes emerge from a chromato-graphic column as a function of time or volume. Detector

tm = “dead time”which is the minimum time for mobile phase to pass through the column tr = “retention time” which is the time required for the solute (analyte) to pass through the column.

• All solutes spend equal time in mobile phase

t’ adjusted retention r= “ time”describes time solutes spend in stationary phase

t rt r t m

2006/9/7

WenHsin Huang, NDMC

29

2006/9/7

•kcapacity factor describes how long solute retained on column

= “relative retention” which is the ratio of adjusted retention times.

t t k'  r m tm

t  r 2 t r 1 >1 always

k’= “capacity factor”

time _ in _ stationary t r k   time _ in _ mobile tm

NDMC

31

=

time spent in stationary phase time spent in mobile phase

•Higher kindicates solute retained longer •If k= 0, solute unretained •If k= 1, solute spent same amount of time in stationary phase as in mobile phase

then separation between two components

WenHsin Huang, NDMC

30

A Plumber’ s View…

Chromatographic Terms

2006/9/7

WenHsin Huang, NDMC

2006/9/7

WenHsin Huang, NDMC

32

8

WenHsin Huang

A Plumber’ s View…

Partition Coefficient Relationship

•kdescribes time spent in two phases •kalso describes number of moles of solute partitioned between two phases

CV k'  s s Cm Vm

time _ in _ stationary moles _ in _ stationary tr k    time _ in _ mobile moles _ in _ mobile tm

[S ] V V t k  s s K s  r [ S ]m Vm Vm t m

Eq’ n

•K Partition coefficient compares concentration of solute in one phase to concentration of solute in the other

2006/9/7

[ S] C V K  2  s  k' K s [S]1 Cm Vm WenHsin Huang, NDMC

t k  K  r 2  2  2 t r k1 K1 1 What this says is that analytes with different partition coefficients will have different retention times.

Eq’ n 33

2006/9/7

Partition coefficient describes conc. of solute in SP and MP

•q fraction of solute in mobile phase molesm Cm Vm q  moless molesm Cm Vm Cs Vs 1 1 q  CV V 1 s s 1 K s Cm Vm Vm

2006/9/7

NDMC

k' (1 q)  1 k '

fraction in MP

fraction in SP

WenHsin Huang, NDMC

34

Review

A Plumber’ s View…

1 q 1 k'

WenHsin Huang, NDMC

35

C K s Cm

Capacity factor describes time solute spends (or moles of solute) in SP and MP

V t t k'  r m K s tm Vm

Relative retention compares retention times (or partition coefficients or capacity factors) of solute 1 and 2

t ' k' K  r 2  2  2 t r1' k '1 K 1

2006/9/7

WenHsin Huang, NDMC

36

9

WenHsin Huang

Chromatographic Peaks

Resolution of peaks

•Solutes elute from a column in a Gaussian peak shape –w –w1/2

Resolution is a measure of how well two peaks are separated.

width of peak at base width of peak at half-height

As tr grows, resolution between two peaks improves.

average retention time   w1/2=2.35

h

1/2h

t ( t t ) resolution  r  r1 r 2 w av ( w 1 w 2 ) / 2

R=0.50

R=0.75

 R=1.00

R=1.50

w=4

t0

tr

time

Peak width (w) is defined as the baseline width (4). 2006/9/7

WenHsin Huang, NDMC

37

2006/9/7

WenHsin Huang, NDMC

38

Plate Height (H)

Resolution Example

Bands broaden as they travel through the column.

Two solutes have retention times and widths of tr1 = 235 s, w1 = 8 sec; tr2 = 250 s, w2 = 10 s. Resolution = 1.7 Plate height (H) relates the amount of broadening to the linear distance traveled. H = plate height x = distance traveled 2 = variance of peak 2006/9/7

NDMC

WenHsin Huang, NDMC

39

2006/9/7

2 H x

WenHsin Huang, NDMC

40

10

WenHsin Huang

Evaluating Separation Efficiency

Evaluating Separation Efficiency

•“ N”in chromatography is analogous to “ n” in liquid-liquid extractions

•Plate theory –Breaks separation up into many discrete stages –Stages represent individual equilibria

–each extraction represents a theoretical plate

[S]2

•N (Theoretical plate) represents each equilibrium between MP and SP

[S]1 •Each time a solute molecule enters the SP from the mobile phase is a theoretical plate 2006/9/7

WenHsin Huang, NDMC

•Higher N corresponds to better separation –indicates solute molecules enter SP a higher number of times 41

2006/9/7

Evaluating Separation Efficiency

•Distribution of retention times for a solute •Peak shape represents differences in behavior of solute molecules during separation

–column –analyte –retention time of analyte –width (at base or at half-height) of peak

t 16t N  r 2  2r  w 5.55t N 2 r w 1/ 2 2006/9/7

NDMC

2

average retention time

2

 

N is dimensionless so measure tr and w (or w1/2) in same units (i.e. time) WenHsin Huang, NDMC

42

Chromatographic Peaks

•N is dependent on

2

WenHsin Huang, NDMC

w1/2=2.35

h

1/2h w=4

43

t0

2006/9/7

tr

WenHsin Huang, NDMC

time 44

11

WenHsin Huang

Evaluating Separation Efficiency

Evaluating Separation Efficiency

•Do not want solute band to spread out as it travels through column •H (Plate height) represents the relationship between the width of a solute band to the distance traveled through column •Lower H corresponds to better separation (higher efficiency) •H also called HETP

•H related to N and length of column (L) Report H in units L H of length (i.e. cm) N •Resolution related to N

R

–Height Equivalent to a Theoretical Plate –One solute equilibrium between SP and MP is one theoretical plate (N) –Equilibrium occurs in one HETP 2006/9/7

WenHsin Huang, NDMC

 k '2  N 1       4   1  k ' av   •Increase L to increase R L L H N N H •Increase to increase R

45

2006/9/7

NDMC

46

The number of theoretical plates (N) increases as the efficiency of the separation increases.

Eq’ n

16t 2 t 2 N  2r  r 2 w 

Derived

or

5.55t 2 t 2 N  2 r  r2 w1 

–to change , must change nature of MP and SP to change relative affinity analytes have for SP WenHsin Huang, NDMC

WenHsin Huang, NDMC

Number of Theoretical Plates (N)

t ' k' K  r 2  2  2 t r1' k'1 K 1

2006/9/7

Eq’ n

–R is dimensionless –k 2 = capacity factor for solute retained longer on column (higher tr)

Evaluating Separation Efficiency R

 k '2  N 1      4   1 k'av   

2

47

2006/9/7

WenHsin Huang, NDMC

48

12

WenHsin Huang

Example

Factors Affecting Resolution The greater the resolution, the better the separation. The more theoretical plates, the better the separation.

A solute with a retention time of 302 s has a width of 11 s on a column that is 15 m long. Find both N and H for this separation.

R

R N  1 k 2 0

N = 1.2 x 10-4 H = 1.2 mm

2006/9/7

WenHsin Huang, NDMC

49

2006/9/7

N 4

 k 2  1          1  k    ave 

R  L R0 R0 WenHsin Huang, NDMC

50

Why Do Bands Broaden?

Principles of Chromatography

Solute invariably spreads apart as it travels (diffusions) through a chromatographic column.

•Introduction to Analytical Separations –Solvent Extraction –What is Chromatography? –A Plumber’ s View of Chromatography –Efficiency of Separation Why Do Bands Spread

The observed variance is a sum of variances from all the broadening mechanisms. 2 obs i2

2006/9/7

NDMC

WenHsin Huang, NDMC

51

2006/9/7

WenHsin Huang, NDMC

52

13

WenHsin Huang

Where Does Broadening Occur?

Where Does Broadening Occur? Inside the column:

Outside the column: Width of injection plug Mixing in detector dead volume

Pump

Multiple paths Longitudinal diffusion Equilibration time

Pump

Injector I

I Column

2006/9/7

Injector

Detector

WenHsin Huang, NDMC

Column

53

2006/9/7

Multiple Paths (A)

Detector

WenHsin Huang, NDMC

54

Multiple Paths (A)

•Solute molecules can choose many different paths through packed columns

Band of 3 solute molecules traveling through packed column

Band spreads

packed column

open tubular column layer of SP coated on inside of column

SP particles packed inside column

time Occurs only in packed columns.

Compare to solute traveling through a straw 2006/9/7

NDMC

Compare to solute traveling through a bean bag WenHsin Huang, NDMC

Smaller particles reduce the plate height.

HA

55

• Open tubular columns do not provide multiple paths for solute during tm (no A term)

2006/9/7

WenHsin Huang, NDMC

56

14

WenHsin Huang

Longitudinal Diffusion (B /ux)

Equilibration Time (Cux)

•S molecules travel through column in a band •[S] varies throughout band

•Finite time required to allow solute molecules to equilibrate between MP and SP •If flow rate too high, MP will “ leave behind” molecules in SP (i.e., Solute in stationary remains

–some solute molecules get ahead of band –some solute molecules lag behind band

“stuck”while solute in mobile phase moves forward.)

•Keep SP thin to decrease equilibration time

Caused by diffusion of solute in the mobile phase. Faster flow rate means less time, thus less diffusion.

H B/ux so, increase ux (flow rate) to reduce problem

start as band travels 2006/9/7

WenHsin Huang, NDMC

57

Direction of travel

MP SP bandwidth

2006/9/7

•Three terms in van Deemter equation correspond to three sources of band spreading or broadening

H A  Cu x ux

•A term Equilibration Time

•B term •C term

• Break equation up into A, B, and C terms

2006/9/7

NDMC

–A

H independent of flow rate

–B/ux

H inversely proportional to flow rate

–Cux

H proportional to flow rate WenHsin Huang, NDMC

58

Band Spreading

Putting the three factors together yields the Van Deemter equation that helps predict how the column flow rate will affect the theoretical plate height. B

Longitudinal Diffusion

bandwidth

H Cux so decrease ux to reduce problem WenHsin Huang, NDMC

Van Deemter Equation

Multiple Paths

MP SP

slow eq.

59

2006/9/7

Multiple Paths term (also called Eddy Diffusion) Longitudinal Diffusion term Equilibration Time term

WenHsin Huang, NDMC

60

15

WenHsin Huang

van Deemter plot

Column Type Affects H

H

B ux

–high resolution (low H) –high sensitivity –low analysis time

A

Flow rate

WenHsin Huang, NDMC

61

2006/9/7

–not useful for large-scale analyses WenHsin Huang, NDMC

62

•Overloaded column (high [S]) –so much S enters SP that the SP starts to look more like S than it looks like original SP –most S molecules retained longer on column Observed peak in chromatogram

Band shape

•In reality, K changes as [S]total changes •If [S] is too high or too low, peak shape deviates from Gaussian shape

NDMC

•Disadvantages

Asymmetric Band Shapes

•Expect Gaussian peak shape from solute S eluting from column, independent of [S]total •Separation based on K C K s Cm

WenHsin Huang, NDMC

–useful for large-scale analyses –A term in van Deemter equation increases H

•Disadvantages

Asymmetric Band Shapes

2006/9/7

•Advantages

•Advantages

Hmin

2006/9/7

SP particles packed inside column

layer of SP coated on inside of column

Cux

uopt

packed column

open tubular column

•Plot dependence of H on individual terms •Sum and find optimum ux to use to minimize band spreading

Majority of S retained on column longer 63

2006/9/7

Faster-moving (less retained) S reaches detector first

WenHsin Huang, NDMC

64

16

WenHsin Huang

Asymmetric Band Shapes

Asymmetric Band Shapes

•Underloading (low [S])

•Load less sample to reduce overloading problem •Protect groups on SP to reduce number of hot sites to reduce tailing problem

–“ hot sites”on SP more available when less S molecules trying to enter SP –some (minority) S get “ stuck”on hot sites –some S retained longer on column Observed peak in chromatogram

Band shape

overloaded

Constant K (slope) results in ideal peak shape

Cs tailed Slower-moving (more retained) S reaches WenHsin Huang, NDMCdetector last

Minority of S retained on column longer 2006/9/7

Cm

Fig. 23-19 65

2006/9/7

WenHsin Huang, NDMC

66

Quantitative Analysis

Chromatographer’ s Triangle Relationship

•In general, detectors can tell us –“ Yes, something is eluting from the column.”

•Use calibration methods to determine how much of a compound is eluting from column Peak Area

Resolution

Speed

Capacity

-b x= m concentration (M)

2006/9/7

NDMC

WenHsin Huang, NDMC

67

2006/9/7

Finding Peak Area •A = ½ wbaseh •A w½ h •cut-and-weigh •computer integration

WenHsin Huang, NDMC

68

17

WenHsin Huang

Quantitative Analysis

Example

•Generally use an internal standard

When 1.06 mmol of 1-pentanol and 1.53 mmol of 1-hexanol were separated by GC, they gave relative peak areas of 922 and 1570 units, respectively. When 0.57 mmol of pentanol was added to an unknown containing hexanol, the relative chromatographic peak areas were 843:816 (pentanol:hexanol). How much hexanol did the unknown contain?

–internal standard is a known amount of a compound different from analyte –compare analytical signal from analyte to analytical signal from internal standard –not the same as method of standard additions –internal standards discussed in Ch. 4

Ax A F s [ X] [ S] 2006/9/7

Ax area of analyte peak As area of standard peak [X] concentration of analyte [S] concentration of standard F response factor WenHsin Huang, NDMC

69

2006/9/7

GUARD

•Analyte

•Constructed of steel or plastic •5 –30 cm length •1 –5 mm i.d. •Easily contaminated and degraded •Guard column

–Soluble in MP (more likely than being volatile)

•Mobile phase –Liquid •Methanol, Water, Acetonitrile, Hexane

–Plays more active role in separation

•Stationary phase

–Contains same SP –Dust, other particles, strongly adsorbed solutes retained on guard column –Expendable –Extends life of analytical column

–Liquid (partition) or solid (adsorption)

•High Performance Liquid Chromatography(HPLC) –High pressure used to force solvent through column (as opposed to gravity)

NDMC

WenHsin Huang, NDMC

70

LC Columns

Liquid Chromatography

2006/9/7

WenHsin Huang, NDMC

71

2006/9/7

WenHsin Huang, NDMC

A N A L Y T I C A L 72

18

WenHsin Huang

LC Columns

LC Columns

•Packing particle diameter (dp) crucial

•Almost exclusively use packed columns •Solutes move much slower through liquids than through gases

–Typical particle diameters are 3 –10 m –Decrease particle diameter to decrease H

Fig. 21-14 S

S

H (m)

•Provide more uniform flow (low A) •Less time needed for solute to get to particle (low C) 10m 60

–Time needed to diffuse to SP in open tubular column is too long

40 20

5m

10

3m 0

2006/9/7

WenHsin Huang, NDMC

73

LC –Adsorption Chromatography

HO

NDMC

Si O

Si O Si

O Si O

Si

Solvent Solute

O

O

Si O

O Si

Fig. 21-10

Si

•Solute is displaced by solvent molecule

OH OH

WenHsin Huang, NDMC

74

SP

OH

OH

Deprotonated silanols (Si –O-) are “ hot sites” that can lead to tailing 2006/9/7

MP Flow

OH

O

8

•Solvent molecules compete with solute molecules for sites on SP

–Pure, spherical, microporous particles

Silica gel with silanol groups on surface

6

Sponge-like Structure

Aggregate of Particles

•Permeable to solvent •Very high surface area •Use when pH < 8

4

LC –Adsorption Chromatography

Microporous Silica Particles

•Solid SP –Silica gel

2

Flow WenHsin rate (mL/min) Huang, NDMC

2006/9/7

OH

OH 75

2006/9/7

WenHsin Huang, NDMC

76

19

WenHsin Huang

LC –Adsorption Chromatography

Isocratic vs. Gradient Elution •Isocratic elution

•Forcible desolvation of solute by solvent is nearly independent of solute identity

–One solvent or one constant solvent mixture used as MP throughout entire separation

–Dependent on solvent identity

•Gradient elution

o) •Eluent strength (

–Adjust solvent mixture through separation –Adjusting eluent strength –May speed up separation process

–Ability of solvent to displace solute

•Elutropic series –Relative eluent strengths of common solvents –Must know about relative polarity of SP/MP

o values of solvents for •Table 21-2 includes  adsorption chromatography on silica o –polarity, 

•Eluent strength increases as MP becomes more like SP 2006/9/7

WenHsin Huang, NDMC

77

2006/9/7

Isocratic vs. Gradient Elution Hexane  °=0.01

Acetonitrile  °=0.52

Start w/ 100% Benzene then add Acetonitrile

2006/9/7

NDMC

WenHsin Huang, NDMC

Fig. 21-18 & Fig. 21-19

isocratic elutions

WenHsin Huang, NDMC

78

Isocratic vs. Gradient Elution •Previous example illustrated balance of good resolution and reasonable retention times •Gradient elution may help –Improve resolution –Shorten retention times (and total analysis time) –Improve peak shapes

gradient elution 79

2006/9/7

WenHsin Huang, NDMC

80

20

WenHsin Huang

LC –Partition Chromatography

LC Phases

•Liquid SP coated on solid support

•Normal-phase Chromatography

–Often, liquid SP covalently attached to surface of silica gel particle

–SP –MP

CH3 Si

O Si

o –More polar solvent has higher 

R

•More attracted to SP and able to displace solute

•Reversed-phase Chromatography

CH3 support

–SP –MP

liquid SP

Common polar R groups

Common nonpolar R groups

(CH2)3NH2

(CH2)17CH3 (or C18)

(CH2)3C N (CH2)2OCH2CH(OH)CH2OH

(CH2)7CH3 (or C8) (CH2)3C6H5

2006/9/7

polar weakly polar or non polar

WenHsin Huang, NDMC

nonpolar or weakly polar polar

o –Less polar solvent has higher 

•More attracted to SP and able to displace solute

81

2006/9/7

LC Phases

WenHsin Huang, NDMC

HPLC Instrumental Design •Already covered columns •Focus now on

•Normal-phase developed first •Reversed-phase is more common

–Injector –Detectors

–Many choices of polar solvents in which solutes are commonly soluble •Water, methanol

Pump

–Nonpolar SP less likely to have hot sites

Injector I

•Less peak tailing

Column

2006/9/7

NDMC

WenHsin Huang, NDMC

82

83

2006/9/7

WenHsin Huang, NDMC

Detector

84

21

WenHsin Huang

Detectors

HPLC 6-way Injection Valve

•Questions to ask when evaluating an LC detector:

Fig. 21-17 waste

waste Sample loop

–Is the detection universal? –Is the detector response linear? –What is the limit of detection (LOD)? –Is the detector useful with gradient elution?

Sample loop

column

•In other words, is the detector insensitive to solvent composition?

2006/9/7

WenHsin Huang, NDMC

85

2006/9/7

UV Detector (w/ flow cell)

WenHsin Huang, NDMC

86

UV Detector

•Common HPLC detector –Many solutes absorb UV light

•Will cover during spectrophotometry

Eluate out

–Any UV spectrophotometers discussed previously are good LC detectors

Flow cell

Light source

•UV cutoff wavelengths –Below given  , MP absorbs UV radiation

Detector

–MP absorbance overwhelms analyte absorbance

Eluate in Fig. 21-20 2006/9/7

NDMC

pathlength WenHsin Huang, NDMC

87

2006/9/7

WenHsin Huang, NDMC

88

22

WenHsin Huang

UV Detector

Fluorescence Detector

•Applicable for UV chromophores Essentially linear

•Similar idea to UV detector •Few molecules fluoresce •Derivatize solutes with fluorescent tag

–Will discuss deviations from linearity in CH 19

LOD ~ 0.1 ng Good detection method with gradient elution

–Derivatize mixture of solutes before separation –Derivatize solutes as they elute from column before detection •Post-column Derivatization

–Use solvents that do not absorb UV radiation

2006/9/7

WenHsin Huang, NDMC

89

2006/9/7

WenHsin Huang, NDMC

90

Refractive Index Detector

Fluorescence Detector

•Compare refractive indices of eluate & reference

Applicable to fluorophores Essentially linear –Will discuss deviations from linearity (Reference)

LOD ~ 0.001 ng Fine with gradient elution

•Cannot use with gradient elution –reference composition cannot exactly match solvent composition during experiment

–Use solvents that do not fluoresce

2006/9/7

NDMC

WenHsin Huang, NDMC

(MP solvent & analyte = eluate)

91

2006/9/7

WenHsin Huang, NDMC

92

23

WenHsin Huang

Comparison of LC Detectors

Refractive Index Detector Type

Essentially universal LOD ~ 100 ng Essentially non-linear

UV-VIS

–Linear range very small

Disaster with gradient elution Finicky –Need constant T 2006/9/7

WenHsin Huang, NDMC

93

Approx LOD Approx Comments Linear range 10 -10 g 10 4 For light absorbing compounds

Fluorescence

10-14 g

10 5

For Fluorescent compounds

MS

10-7-10-9 g

10 5

Universal detector

-8

Electrochemical conductometric

10 g/mL

Electrochemical amperometric

10- 10-10 –11 g 10 5

2006/9/7

10

5

WenHsin Huang, NDMC

Specific detector; for all ions Specfic detector; electroactve compounds 94

Liquid Chromatography (LC) Especially high-performance liquid chromatography (HPLC). Thet e r m“ hi g h-pe r f o r ma nc e ”r e f e r st ot he use of packed columns with very small packing particles (diam. 5-10 m) giving greatly enhance resolution.

Note: several types of LC. In addition to partition (as described so far) - there are also ionexchange and size-exclusion chromatography using liquid mobile phases. We will concentrate only on partition. 2006/9/7

NDMC

WenHsin Huang, NDMC

95

24

View more...

Comments

Copyright ©2017 KUPDF Inc.
SUPPORT KUPDF