028-Procedure for Calibration.pdf

November 21, 2016 | Author: Köksal Patan | Category: N/A
Share Embed Donate


Short Description

Download 028-Procedure for Calibration.pdf...

Description

Pearl GTL Project Shell Project No: HP-3000-QAT Project Document Number: T- 4.250.915

Purchase Order Number:

2MH127

Tag/Item Number(s):

N/A

Unit(s):

ASU

PCWBS:

Z

LINDE ENGINEERING

Contractor Project No : Project Rev.: B

Page 2 of 14

5887

QSGTL PEARL PLANT C2 AIR SEPARATION UNITS CONSTRUCTION

E

B

09.11.08

Ready for Construction

FB

EO

AG

RFC

D

A

09.10.08

Issue for Review

FB

EO

AG

IFR

C

O

02.12.07

Issued for Company Comments

BR

UA

MC /ZG

ICC

B

Issued for Review

MS

MC

ZG

IFR

A

Issued for Internal Review

MS

MC

ZG

IIR

GAMA Rev. No.

Project Rev No.

Date

Prepared Checked Approved By By By ORIGINATOR

Description

Status

Document Title:

PROCEDURE FOR CALIBRATION LINDE ENGINEERING AG Document No.

Contractor

LOC470 Gama Qatar Co. W. L. L.

GAM

PRO

QAC

GTL

028

E

Org. Code

Doc. Type

Disc. Code

Geogr. Area

Seq. No

Rev. No

Project Document Number: T-4.250.915

Vendor Code

Rev No: B

Q A T A R

Doc. Title

:

GAM-PRO-QAC-GTL-028 Rev. E PROCEDURE FOR CALIBRATION

Date

:

09.11.2008

TABLE OF CONTENTS 1.

SCOPE ________________________________________________________________________ 5

2.

PURPOSE _____________________________________________________________________ 5

3.

REFERENCES _________________________________________________________________ 5

4.

DEFINITIONS _________________________________________________________________ 5

5.

RESPONSIBILITIES ____________________________________________________________ 6

6.

PROCEDURE __________________________________________________________________ 6 6.1.

GENERAL REQUIREMENTS_______________________________________________________6

6.2.

IDENTIFICATION OF INSTRUMENTS ______________________________________________6

6.3.

CALIBRATION OF INSTRUMENTS _________________________________________________7

6.3.1. 6.3.2.

7.

Periodic Calibration ____________________________________________________________________ 7 Re-calibration _________________________________________________________________________ 7

6.4.

PRECISION OF CALIBRATION ____________________________________________________7

6.5.

METHOD OF CALIBRATION ______________________________________________________7

6.6.

DOCUMENTATION OF CALIBRATION _____________________________________________8

ATTACHMENTS ________________________________________________________________ 9

T-4.250.915 Rev. B

Page 4 of 14

Q A T A R

Doc. Title

:

GAM-PRO-QAC-GTL-028 Rev. E PROCEDURE FOR CALIBRATION

Date

:

09.11.2008

1. SCOPE This procedure describes the criteria and methods that shall be applied by GAMA in ensuring the accuracy of the measuring and test instruments that shall be used in the erection activities of the QSGTL Pearl Plant C2 Air Separation Units.

2. PURPOSE The purpose of this Quality Control Procedure is to set a guideline of the testing requirements as well as the responsibilities of implementing this procedure.

3. REFERENCES T.4.096.674

: Calibration of Testing and Measuring Tools and Equipment

PQP-08.F01 PQP-08.F02

: List of Measuring and Test Devices Subject to Calibration : Calibration Label

Relevant Manufacturer’s Manual for equipments. Also all relevant ITP’s and QCR’s.

4. DEFINITIONS Company

: QSGTL and/or its PMC

Contractor

: Linde Engineering AG

Sub Contractor

: Gama Qatar

Project

: QSGTL C2 ASU PEARL Construction Works

ITP

: Inspection Test Plan

QCP

: Quality Control Procedure

QA/QC

: Quality Assurance / Quality Control Department

ITL

: Independent Testing Laboratory

MRR

: Material Receiving Report

HSE

: Health and Safety and Environment

T-4.250.915 Rev. B

Page 5 of 14

Q A T A R

Doc. Title

:

GAM-PRO-QAC-GTL-028 Rev. E PROCEDURE FOR CALIBRATION

Date

:

09.11.2008

5. RESPONSIBILITIES The QA/QC Manager is responsible for monitoring and implementing the following activities:  Establishing the calibration schedule for all equipment included in the system.  Keeping a record of all calibrated instruments.  Ensuring that instruments are calibrated at all times during the work.  Storage of calibration instruments, which are defined as secondary "MASTER". Master instruments are generally stored in a locked-up cabinet under the control of the QA/QC Manager. Particularly, over-sized instruments shall be kept at adequate areas and/or in proper containers. The above mentioned activities shall be carried on by the QA/QC Department together with the department that uses the instrument.

6. PROCEDURE 6.1.

GENERAL REQUIREMENTS

Instruments and/or devices having appropriate calibration certificates can be utilised within the expiry period. After the expiry dates, the calibration of these instruments and/or devices should be calibrated. The present procedure is applicable to instruments that require to be calibrated periodically, such as pressure gages, hardness testers, paint thickness gages, optical measurement devices, etc. Calibration intervals for these instruments and/or devices are given in Attachment 1 according to T.4.096.674 Calibration of Testing and Measuring Tools and Equipment.

6.2.

IDENTIFICATION OF INSTRUMENTS

Each instrument that is subject to calibration shall be identified by a serial number. a)

This serial number will be placed on the instrument by such means that shall provide durable identification. All inspection, measuring and test instruments with missing or illegible serial numbers shall be re-calibrated prior to their use.

b)

All inspection, measuring and test instruments that are employed for the inspection or acceptance of the construction work shall be marked with utilising stickers or tags, which shall enable immediate visual confirmation of its calibration status.

T-4.250.915 Rev. B

Page 6 of 14

Q A T A R 6.3.

Doc. Title

:

GAM-PRO-QAC-GTL-028 Rev. E PROCEDURE FOR CALIBRATION

Date

:

09.11.2008

CALIBRATION OF INSTRUMENTS

Instruments subject to calibration shall be calibrated on the following occasions:

6.3.1.

Periodic Calibration

This calibration will be carried out before the defined expiry dates by the QA/QC personnel or by a calibration service or agency under the control of the QA/QC Department.

6.3.2.

Re-calibration

The instruments may be re-calibrated at any time before the expiry date of the calibration if there are doubts as to the accuracy of the equipment.

6.4.

PRECISION OF CALIBRATION

The precision of the calibration shall be defined by the QA/QC Department in accordance with the class of precision stated by the manufacturer of the instrument.

6.5.

METHOD OF CALIBRATION

The calibration of the devices shall be carried out according to the manufacturer instructions. In case of the calibration procedure and/or instruction of the device manufacturer is not available, then one of the following procedures shall apply: a)

Calibration against MASTER devices, Master devices shall possess the manufacturer’s certificates. All primary "MASTER" instruments (2nd level) shall undergo calibration at predetermined periods depending on their characteristics and tendency to lose calibration. Secondary instruments (3rd level) shall be calibrated by comparison with the Masters.

b)

Calibration Control by direct measurement, Where and when applicable, the measuring devices shall be calibrated by direct measurement with appropriate and acceptable measuring methods. Measurements shall be carried out with certified devices. During the controls, if the device has calibration means, then it will be calibrated, supported with an appropriate measurement record.

c)

Calibration by an authorised institute, Calibration of the sophisticated devices, including optical measuring devices shall only be carried out by an authorised institute.

T-4.250.915 Rev. B

Page 7 of 14

Q A T A R 6.6.

Doc. Title

:

GAM-PRO-QAC-GTL-028 Rev. E PROCEDURE FOR CALIBRATION

Date

:

09.11.2008

DOCUMENTATION OF CALIBRATION

A "control list" will be compiled for the measuring and test equipment and shall give the relevant data. The calibration label (Attachment 2) of the calibrated equipment shall indicate the acceptance of the calibration. When this labelling is neither practical nor considered essential for the control purposes, other procedures shall be established to ensure the conformity with this requirement. Any other procedure shall be in compliance with the requirements of related specifications and subject to approval. Prior to use of any inspection, measuring and test instrument for the construction work, the responsible engineer shall assure that the effective calibration period of the device has not yet expired and the instrument has not been damaged to a visible degree.

T-4.250.915 Rev. B

Page 8 of 14

Q A T A R

Doc. Title

:

GAM-PRO-QAC-GTL-028 Rev. E PROCEDURE FOR CALIBRATION

Date

:

09.11.2008

7. ATTACHMENTS Attachment 1 Calibration Intervals Attachment 2 Calibration Labels

T-4.250.915 Rev. B

Page 9 of 14

Q A T A R

T-4.250.915 Rev. B

Doc. Title

:

GAM-PRO-QAC-GTL-028 Rev. E PROCEDURE FOR CALIBRATION

Date

:

09.11.2008

Page 10 of 14

Q A T A R

T-4.250.915 Rev. B

Doc. Title

:

GAM-PRO-QAC-GTL-028 Rev. E PROCEDURE FOR CALIBRATION

Date

:

09.11.2008

Page 11 of 14

Q A T A R

Doc. Title

:

GAM-PRO-QAC-GTL-028 Rev. E PROCEDURE FOR CALIBRATION

Date

:

09.11.2008

Calibration/ Functional Test

Calibration Period (Months)

CAL

6 months

CAL

1 week

CAL

1 month

CAL

12 months and each testing

CAL

12 month

CAL

2 years

CAL

12months

Calibration of Hygro meter

CAL

12 months

Calibration of Elco meter

CAL

12 months

Calibration of Holiday tester

CAL

12 months

Equipment Calibration of survey instruments Calibration of Relief valve Calibration of MT yoke Calibration of optical measuring devices Calibration of ultrasonic thickness measuring device Calibration of surface thermometer Calibration of PMI machine

T-4.250.915 Rev. B

Remarks

Page 12 of 14

Q A T A R

Doc. Title

:

GAM-PRO-QAC-GTL-028 Rev. E PROCEDURE FOR CALIBRATION

Date

:

09.11.2008

ATTACHMENT 2

CALIBRATION LABELS

CALIBRATED

INS

__________

Date :

___/___/___

__________

Sign

__________

S.NO Due

“CALIBRATION"

T-4.250.915 Rev. B

LABEL COLOUR: WHITE

Page 13 of 14

Q A T A R

Doc. Title

:

GAM-PRO-QAC-GTL-028 Rev. E PROCEDURE FOR CALIBRATION

Date

:

09.11.2008

HOLD

"DO NOT USE"

T-4.250.915 Rev. B

LABEL COLOUR: RED

Page 14 of 14

View more...

Comments

Copyright ©2017 KUPDF Inc.
SUPPORT KUPDF